Keysight (Agilent) 8665B Service Note

8665B-04
SERVICE NOTE
SUPERSEDES: NONE
8665B Synthesized Signal Generator
Serial Numbers: See Duplicate Service Notes
Return specified transformers to factory for analysis.
Duplicate Service Notes:
8644B-04 Serials: 3546A00973 / 3546A01081 8645A-05 Serials: 3546A00554 / 3546A00569 8664A-05 Serials: 3546A01174 / 3546A01358 8665A-06 Serials: 3546A00850 / 3546A00873 8665B-04 Serials: 3546A00891 / 3546A01000
Parts Required:
P/N DescriptionQty.
9100-4757 Power Transformer 1 2110-0083 Fuse 2.5A 250V 1
DATE: November 1997
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: May 2000
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 2.0 Hours
SERVICE RETURN INVENTORY: SCRAP
SEE TEXT
USED RETURN PARTS: SCRAP
Continued
SEE TEXT
AUTHOR: MW ENTITY: 5320 ADDITIONAL INFORMATION:
© 1997 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 2
Page 2 Service Note 8665B-04
Situation:
This transformer, in the affected serial number range listed above, has demonstrated a potential re­liability problem at the factory, and might blow the main line fuse (P/N 2110-0083) if a properly grounded power cable is not used.
Solution / Action:
It is recommended that any transformer from a power supply that blows its line fuse in the af­fected instruments be replaced, and the old transformer be returned to the factory for analysis. Warranty is extended for the fuse failure and associated transformer replacement until 1 May, 2000 on all instruments specified above.
Please contact MID Source Support at (707)577-6803 for instructions on returning the trans­former.