8665B-04
SERVICE NOTE
SUPERSEDES: NONE
8665B Synthesized Signal Generator
Serial Numbers: See Duplicate Service Notes
Return specified transformers to factory for analysis.
Duplicate Service Notes:
8644B-04 Serials: 3546A00973 / 3546A01081
8645A-05 Serials: 3546A00554 / 3546A00569
8664A-05 Serials: 3546A01174 / 3546A01358
8665A-06 Serials: 3546A00850 / 3546A00873
8665B-04 Serials: 3546A00891 / 3546A01000
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
P/N DescriptionQty.
9100-4757 Power Transformer 1
2110-0083 Fuse 2.5A 250V 1
DATE: November 1997
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: May 2000
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR 2.0 Hours
SERVICE RETURN
INVENTORY: SCRAP
SEE TEXT
USED RETURN
PARTS: SCRAP
Continued
SEE TEXT
AUTHOR: MW ENTITY: 5320 ADDITIONAL INFORMATION:
© 1997 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
1 of 2

Page 2 Service Note 8665B-04
Situation:
This transformer, in the affected serial number range listed above, has demonstrated a potential reliability problem at the factory, and might blow the main line fuse (P/N 2110-0083) if a properly
grounded power cable is not used.
Solution / Action:
It is recommended that any transformer from a power supply that blows its line fuse in the affected instruments be replaced, and the old transformer be returned to the factory for analysis.
Warranty is extended for the fuse failure and associated transformer replacement until 1 May,
2000 on all instruments specified above.
Please contact MID Source Support at (707)577-6803 for instructions on returning the transformer.