8665A-07
SERVICE NOTE
SUPERSEDES: None
8665A Synthesized Signal Generator
A5 LF Driver
Serial Numbers: See Below
Deceptive LF Driver failure codes
Duplicate Service Notes:
8664A-06 Serial Numbers: 0000A00000 / 3546A99999
8665A-07 Serial Numbers: 0000A00000 / 3546A99999
8665B-05 Serial Numbers: 0000A00000 / 3546A99999
To Be Performed By: Agilent-Qualified Personnel
Situation:
When activating special 320 (service diagnostics), certain result codes sometimes appear
falsely in the instruments with serial numbers listed above. These codes may not appear using Special Functions 170 through 172.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION AVAILABLE
ACTION
CATEGORY:
AGREEABLE TIME
Continued
DATE: December 1997
PERFORMANCE ENHANCEMENT
SERVICE/RELIABILITY ENHANCEMENT
LOCATION
CATEGORY:
AUTHOR:
MW
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
ENTITY:
5320
© 1997 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
AVAILABLE UNTIL:
December 1999
ADDITIONAL INFORMATION:
1 of 2

Page 2 Service Note 8665A-07
-10,138,xxx
-10,144,xxx
-10,150,xxx
-10,156,xxx
The problem may be with the diagnostic firmware’s set voltage limits, and not the instrument’s
performance.
Solution / Action:
A. The first recommended solution to this problem is a firmware upgrade (to version 4.2.2
or higher). If the firmware upgrade does not correct the problem, then replace the LF
Driver assembly (P/N 08665-61205). Do not order the LF Driver Upgrade Kit (P/N
08665-60328) if the firmware is already at version 4.2.2 or higher.
B. As a result of this firmware change, and the subsequent elimination of an old LF Driver
assembly (08665-61005), it is necessary to upgrade the firmware to version 4.2.2 or
higher when replacing the LF Driver. A universal LF Driver Upgrade Kit (P/N 08665-
60328) is available, which includes the new LF Driver assembly and firmware
upgrade kits.