Keysight (Agilent) 8664A Service Note

8664A-03
SUPERSEDES: None
Continued
DATE: 23 April 1993
8664A Synthesized Signal Generator
Serial Numbers: 3203A00254/3203A00275 (Refer to specific serial numbers in text)
Potential RPP Assembly Failures
Duplicate Service Notes: 8665A-04
Situation:
Instruments with specific serial numbers noted below may have defective RPP (Reverse Power Protection) assemblies. The RPP assembly protects the RF output circuitry from damage due to externally applied power. Failure symptoms include no RF output or a low output level (relative to front-panel settings).
The RPP assembly consists of several components, one of which is reed relay encased in a glass tube with teflon plugs at both ends. An elastomer envelope is used to secure coil windings around the glass tube and to provide consistent electrical characteristics.
The RPP fails when the elastomer expands beyond its nominal size and contacts the external connections of the reed relay. This causes a low impedance RF path to ground.
1 of 3
AUTHOR: LHL
ENTITY: 1000
© 1993 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
RESPONSIBILE UNTIL: 01 June 1995
ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 2.5 hrs
Page 2 Service Note 8664A-03
NOTE:BILLINGS WILL ONLY BE ACCEPTED FOR INSTRUMENTS WITH SERI­AL NUMBERS LISTED BELOW. INSTRUMENTS WITH SERIAL NUMBERS NOT LISTED UTILIZE THE REDESIGNED RPP ASSEMBLY AND DO NOT REQUIRE MODIFICATION.
Affected Serial Numbers: Standard Instruments P/N 08664-61003, RF/RPP Assy
3203A00254 3203A00257 3203A00258 3203A00259 3203A00260 3203A00261 3203A00262 3203A00263 3203A00264 3203A00265 3203A00266 3203A00268 3203A00271 3203A00272 3203A00273
Instruments with Option 008 (Pulse Modulation) P/N 08664-61005, RF/RPP Assy
3203A00256 3203A00267 3203A00274 3203A00275
Solution/Action:
1. Remove the original RPP assembly and replace it with the part number listed above. Refer to Figure 1 for assembly and component location.
2. Run the internal diagnostics after replacing the RPP assembly.
3. Perform RF Level Accuracy Performance Test to insure that the instrument meets specifications.