Keysight (Agilent) 8664A Service Note

8664A-01
SUPERSEDES: None
DATE: 09 October 1992
8664A Synthesized Signal Generator
Serial Numbers: 0000A00000 / 3035A00185
I/O Assembly Replacement requires modification to Reference Assembly
Duplicate Service Notes:
Situation:
When an I/O assembly (08665-69104 or 08665-60204) is replaced in instruments (serial numbers noted above) a component on the Reference assembly (P/N 08665-61006) must be removed to insure proper operation of the reference oscillator.
NOTE: Billings related to replacement of the I/O or Reference assemblies are not covered by this Service Note.
Solution/Action:
Remove L307 on the Reference assembly. Refer to Figure 1 for component location. Re­calibration is not required.
1 of 2
AUTHOR: LHL
ENTITY: 1000
© 1992 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 01 November 1997
ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 1.0 Hour