8662A-21
SERVICE NOTE
SUPERSEDES: NONE
8662A Synthesized Signal Generator
Serial Numbers: 3909A04711 / 3909A04746
Procedure to resolve A7A3C10 polarity misprinted.
Duplicate Service Notes:
8663A-15 Serial Numbers: 3909A02711 / 3909A02789
To Be Performed By: A gilent-Qua lif ie d Pe r sonnel or Custome r
Parts Required:
P/N Description Quantity
08662-69016 Repair kit C-10 1
The 08662-6901 6 co ntains:
- one "+" stick e r
- one "Rev B 4038" sticker
- one 270uF 25V capacitor (0180-2749)
DATE: September 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: April 2003
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR .0.5 Hours
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
Continued
RETURN
SCRAP
SEE TEXT
AUTHOR: SN ENTITY: 5331 ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES
PRINT ED IN U.S.A.
1 of 3

Page 2 Service Note 8662A-21
Situation:
There is a misprinted A7A3C 10 polarity sign on A7A3, pa rt number 08662-60604 whic h c a uses
C10 to be inserted in the reverse-bias position. The A7A3C10, 270 uf electrolatic capacitor, will
fail when the 8662A/63A operates over time.
Solution / Ac t ion :
All defective boa rds ha ve be e n pur ge d f ro m sto ck and reworked. Th e fix ed boa rd s w ill ha ve a
new revision, Re v B 403 8. The bo ard pa rt number remains the sa me, p art nu mbe r 08662-60604.
Customers, who own the 8662A/63A with the above se ria l num be r will be notif ied f or repair . The
repair can be done by Agilent-Qualified pe rsonnel or the custom er.
Contact Son Nguyen a t telne t 1-577-5977 or 707-577- 5977 or
E-mail at son_c_nguyen@agilent.com
to get the repair kit C-10, 08662-69016 one per 8662A or 8663A.
Repair procedure:
1. Turn off power of 8662/6 3A.
2. Pull out AC power co rd from 8662/ 63A
3. Take off topsid e of St an d-of fs and cov er with T-15 Torx drive r.
4. Locate black powe r suppl y co ve r that is loca te d at lowe r righ t side of co rn er when you face
to rear side o f th e in str u m ent.
5. Unscrew 6 screws wit h #2 Po zi or Phill ip dr ive r , and tak e th e to p co ve r off.
6. Pull out A7A3 (08662-60604) by pull ing out orange colored tap s.
7. Locate C10 (270uF 25V) that is located on the lower left side of A7A3.
* Inspect A7A3C10
- If C10 did not leak, conti nue ste p 8
- If C10 leaked and damaged the board(s), return unit to Agilent Service Center for board(s)
replacement
Continued