Keysight (Agilent) 8662A Service Note

8662A-20
SERVICE NOTE
SUPERSEDES: NONE
8662A Synthesized Signal Generator
Serial Numbers: 3842A04698 / 3909A04738
Procedure to resolve 230V misprinted fuse rating on fuse plate.
Duplicate Service Notes: 8663A-14 S/N 3842A02691 / 3909A02 751
To Be Performed By: A gilent-Qua lif ie d Pe r sonnel or Custome r
Parts Required:
N/A Sticker with "2.5A" rating 1
Situation:
There is a misprint ed fus e rating value on the fu se p late in rea r o f the 8662A/8663A for 230V. The printed value should be 2.5A instead of 1. 5A. If the cu stom e r wou ld used 1. 5A fuse, the worst case woul d be that the fu se wo uld bl ow without any further dama de to the unit.
Solutio n / A ction:
The correct stic ke r la be l with "2.5A" wil l be p ut on the fuse plate to cove r the "1. 5 A" label. Customer can make the label by typing "2.5A" on a sticker label and place over the "1.5A" on fuse plate of the 8662 A/8663A or contact Son Nguye n at 707- 577-5977 for a pre-made Sticker labe l.
DATE: March 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: April 2003
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 0.25 Hours
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
RETURN SCRAP SEE TEXT
AUTHOR: SN ENTITY: 5331 ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES PRINT ED IN U.S.A.