Keysight (Agilent) 8662A Service Note

8662A-19
SERVICE NOTE
SUPERSEDES: NONE
8662A Synthesized Signal Generator
Serial Numbers: 3909A4697 / 3909A4716
Procedure to resolve 2-4 MHz spurs out from 640-1280 MHz carrier
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
P/N Description Quantity
Situation:
A possible misload collector bias DC-Feed within the 08662-67009 package for the Doubler Module (08662-60303). When this occurs, the spurs from 2-4 MHz with a level between
-35dB to -84dB will genarate.
Solution / Action:
Order the replacement Doubler Module (08662-60309). Replace the part and recalibrate the 8662A Synthesized Signal Generator. All 08662-67009' s with the misload DC-feed have been purged from inventory.
DATE: July 1999
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: July 2001
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 2.0 Hours
SERVICE RETURN INVENTORY: SCRAP
SEE TEXT
USED RETURN PARTS: SCRAP
SEE TEXT
AUTHOR: SN ENTITY: 5320 ADDITIONAL INFORMATION:
© 1999 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 1
Loading...