8662A-19
SERVICE NOTE
SUPERSEDES: NONE
8662A Synthesized Signal Generator
Serial Numbers: 3909A4697 / 3909A4716
Procedure to resolve 2-4 MHz spurs out from 640-1280 MHz carrier
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
P/N Description Quantity
08662-60309 Restr. Doubler Module 1
Situation:
A possible misload collector bias DC-Feed within the 08662-67009 package for the Doubler
Module (08662-60303). When this occurs, the spurs from 2-4 MHz with a level between
-35dB to -84dB will genarate.
Solution / Action:
Order the replacement Doubler Module (08662-60309). Replace the part and recalibrate the
8662A Synthesized Signal Generator. All 08662-67009' s with the misload DC-feed have
been purged from inventory.
DATE: July 1999