Keysight (Agilent) 8657A Specifications Sheet

Spectral performance for general-purpose test
Overview
The Agilent Technologies 8657A and 8657B signal gener­ators are designed to test AM, FM, and pulsed receivers as well as components. With their exceptional analog modulation, good spectral purity, and superb output level performance, these signal generators are ideal for R&D, manufacturing, and support.
Measurement confidence
The 8657A and B signal generators provide the spectral purity and modulation versatility needed for RF testing of communication and navigation equipment.
Low residual FM ensures wide hum and noise test margins
Hum and noise testing requires a signal generator to have low residual FM to ensure no measurement error. The residual FM of the 8657A and B provides the test margins you need for the most demanding hum and noise tests.
Output level range and accuracy for sensitivity testing
With a dynamic range from +13 to –143.5 dBm, the 8657A and B allow sensitivity measurements to be made on even the most sensitive receivers. The 8657A and B couple this with a level accuracy of ±1 dB (typically
0.5 dB) to –127 dBm across the full temperature range of 0 to 55 degrees C, to give you accurate and repeatable measurements every time.
Low RF leakage eliminates measurement interference
Sensitivity measurements can often be masked by RF leakage (radiated emissions). With the 8657A and B, RF leakage is reduced to a level that assures virtually no measurement interference.
DCFM faithfully reproduces digital signals
For radios with digital squelching, the 8657A and B have extremely stable dc-coupled FM (dcFM). The dc coupling of low-rate tones or digital data eliminates droop, and the exceptional stability and center-frequency accuracy when in dcFM mode eliminates the need to retune the signal generator after dcFM is selected.
Pulse modulation with the 8657B
High-performance pulse modulation (available on the 8657B) will give you confidence in your radar or pulsed carrier measurements. This includes rise/fall times of <35 ns, rates from dc to 30 MHz, and better than 70 dB on/off ratios (fc<1030 MHz).
Agilent 8657A/8657B
Signal Generators
Profile
2
Out-of-channel measurements with the 8657A/B
Low single-sideband phase noise for adjacent channel selectivity testing
A receiver’s ability to reject unwanted signals is meas­ured using out-of-channel tests. With its good spectral purity, the 8657A and B make it easy to perform demanding tests such as adjacent channel selectivity measurements.
Adjacent channel selectivity measures a receiver’s ability to pick out a desired signal while rejecting a strong sig­nal one channel away. To measure adjacent channel selectivity the out-of-channel signal generator must have low single-sideband (SSB) phase noise and nonharmonic spurious content at channel spacings, otherwise the in­channel signal is masked. The exceptional phase noise performance of the 8657A and B provides a cost-effective solution for many out-of-channel tests.
General purpose and component tests
Output power to drive high-level inputs
For applications requiring high output power, the 8657A and B can overrange beyond the +13 dBm specified out­put level to >+16 dBm for most frequencies.
DCFM for VCO simulation
State-of-the-art dcFM and wide FM bandwidth make the 8657A and B ideal sources for many VCO simulation applications. For example, the 8657A or B can be used to replace a receiver’s VCO during design.
Phase adjust to characterize phase-sensitive devices
The 8657A and B give you the ability to adjust the phase of the output signal in one-degree increments with respect to a source that is locked to the same reference timebase. This feature makes it easy to characterize phase-detector or phase-interferometer receivers during design or manufacturing.
Manual tests
100 nonvolatile store/recall registers save set-up time
The 8657A and B offer as a standard feature the ability to store 100 complete instrument states. This feature decreases set-up time when performing repetitive tests and reduces operator errors.
Register sequencing provides semi-automation
Stepping through the store/recall registers is easy with the front-panel sequence keys or the rear-panel remote­sequence connector. These features allow the user to sequence through the storage registers in any order.
Automated tests
Reliable output attenuators
With production lines requiring ever-faster throughput, test equipment must be more reliable than ever. Output level cycling requires electromechanical relays to switch in and out of different attenuators to produce varying output levels. The 8657A and B enhance system up-time by using a very reliable attenuator technology. The 8657A is especially dependable with its electronic atten­uator design. Instead of using mechanical relays for switching the attenuators, the 8657A uses solid-state components for setting output levels. The patented design uses PIN-diode switching elements with 3 million hours mean time between failure rate. This exceptional reliability is backed with a 5-year warranty against attenuator failure.
Ordering information
Please contact your Agilent Technologies Sales and Service Office for more information or visit our website at www.agilent.com/find/tmdir
Option Description
001 High-stability timebase 002 Rear-panel connections 003 Pulse modulation ( 8657B only) 1BN Mil std 45662A calibration certification 1BP Mil std 45622A calibration with test data 907 Front handle kit 908 Rack flange kit 909 Rack flange kit w/ front handle 910 Adds operation/calibration manual and two service manuals 915 Adds service manual W30 3-year return repair service W32 3-year return calibration service W34 3-year standards compliant calibration service W50 5-year return repair service W52 5-year return calibration service W54 5-year standards compliant calibration service
3
Technical specifications
Specifications describe the instruments warranted per-
formance and apply after a 30-minute warm-up. All specifications are valid over the signal generator’s entire operating/environment range unless otherwise noted.
Supplemental characteristics (indicated by italic type) are intended to provide information useful in estimating instrument capability in your application by describing typical, but not warranted, performance.
Note: The upper frequency range of the 8657A is 1.04 GHz. Specifications above 1.04 GHz apply only to the 8657B.
Frequency
Range
8657A 100 kHz to 1.04 GHz 8657B 100 kHz to 2.6 GHz Underrange To 10 kHz with uncalibrated output and
modulation.
Resolution 8657A 10 Hz 8657B 1 Hz Accuracy and stability Same as timebase Switching speed
1
<35 ms.2(30 ms typical at 25 °C)
Phase offset Output signal phase is adjustable in
1-degree nominal increments.
Internal reference oscillator
Std. (typ.) High-stability Option 001
Aging rate ±2 ppm/yr 8657A, 1.5 x 10-8parts/day after 10 days
1.0 x 10-9parts/day after 180 days
8657B, 1.0 x 10-9parts/day after 45 days Temperature (0 to 55 °C) ±10 ppm 7 x 10
-9
Line voltage 2 x 10-9(+5%, –10%) Frequency 50 MHz 10 MHz Timebase reference Available at a level of >0.15 V
rms
into
output (rear panel) 50 (output of 10, 5, or 1 MHz is selec-
table via internal jumper). If external ref-
erence is used, output will be the same
frequency.
External reference input Accepts any 10, 5, or 1 MHz ±0.002%) (rear panel) Frequency standard at a level
>0.15 Vrms into 50
Output
Range (dBm)
8657A +13 dBm to –143.5 dBm into 50 ,
+10 dBm to –143.5 dBm for frequencies from 100 kHz to 1 MHz
8657B +13 dBm to –143.5 dBm into 50 ,
+10 dBm to –143.5 dBm with pulse
modulation installed at fc<1.03 GHz Resolution 0.1 dB Absolute level accuracy
3
8657A <±1.5 dB (>+7 dBm)
<±1.0 dB (+7 to –127 dBm)
<±1.5 dB (<–127 dBm)
8657B <±1.5 dB (>+3.5 dBm)
<±1.0 dB (+3.5 to –127 dBm)
<±1.5 dB (<–127 dBm)
Level flatness
100 kHz to 2.06 GHz ±0.5 dB, output level setting of 0 dBm Reverse power protection
to maximum output frequency) 50 watts (from a 50 source)
Maximum DC voltage 8657A, 50 V
8657B, 25 V
SWR
8657A (fc(400 kHz) <1.5 for levels <–3.5 dBm
<2.0 for levels +13 dBm 8657B <1.5 for levels –6.5 dBm
<2.0 for levels +13 dBm
Output impedance 50 nominal
1. To be within 100 Hz of carrier frequency.
2. Add 5 ms when switching to fc>1.03 GHz for the 8657B.
3. Absolute level accuracy includes allowances for detector linearity, temperature,
flatness, attenuator accuracy and measurement uncertainty.