Keysight (Agilent) 8643A Service Note

8643A-04
SERVICE NOTE
SUPERSEDES: None
8643A Synthesized Signal Generator
Serial Numbers: See Below
Disconnect line power when connecting and disconnecting ribbon cables
Duplicate Service Notes:
8644A-05 Serials: 0000A00000 / 9999Z99999 8644B-05 Serials: 0000A00000 / 9999Z99999 8645A-06 Serials: 0000A00000 / 9999Z99999 8664A-07 Serials: 0000A00000 / 9999Z99999 8665A-08 Serials: 0000A00000 / 9999Z99999 8665B-06 Serials: 0000A00000 / 9999Z99999
To Be Performed By: Agilent-Qualified Personnel
Situation:
DC power is supplied to certain modules for heating and battery charging while line power is connected, even with power off. Connecting or disconnecting ribbon cables or modules plugged into the motherboard with line power connected can result in electrical damage to sensitive components in the instrument. It is recommended that instruments be unplugged for any service requiring the disconnection of ribbon cables or modules plugged into the motherboard.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
MW
ENTITY:
5320
DATE: August 1998
ADDITIONAL INFORMATION:
NONE
© 1998 AGILENT TECHNOLOGIES PRINTED IN U.S.A.