No part of this manual may be reproduced in any
form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written
consent from Agilent Technologies, Inc. as governed by United States and international copyright lays.
Manual Part Number
86100-90094
Technology Licenses
The hardware and/or software described in this
document are furnished under a license and may
be used or copied only in accordance with the
terms of such license.
LZW compression/decompression: Licensed
under U.S. Patent No. 4,558,302 and foreign
counterparts. The purchase or use of LZW graphics capability in a licensed product does not
authorize or permit an end user to use any other
product or perform any other method or activity
involving use of LZW unless the end user is separately licensed in writing by Unisys.
Trademark Acknowledgements
Microsoft is a U.S. registered trademark of
Microsoft Corporation.
Windows and MS Windows are U.S. registered
trademarks of Microsoft Corporation.
MATLAB ® is a U.S. registered trademark of The
Math Works, Inc.
Edition
March 2007
Printed in Malaysia
Agilent Technologies, Inc.
Digital Signal Analysis Division
1400 Fountaingrove Parkway
Santa Rosa, CA 95403, USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed,
without notice, in future editions. Further, to the
maximum extent permitted by applicable law,
Agilent disclaims all warranties, either express or
implied, with regard to this manual and any information contained herein, including but not limited
to the implied warranties of merchantability and
fitness for a particular purpose. Agilent shall not
be liable for errors or for incidental or consequential damages in connection with the furnishing,
use, or performance of this document or of any
information contained herein. Should Agilent and
the user have a separate written agreement with
warranty terms covering the material in this document that conflict with these terms, the warranty
terms in the separate agreement shall control.
Restricted Rights Legend
If software is for use in the performance of a U.S.
Government prime contract or subcontract, Software is delivered and licensed as “Commercial
computer software” as defined in DFAR 252.2277014 (June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19
(June 1987) or any equivalent agency regulation
or contract clause. Use, duplication or disclosure
of Software is subject to Agilent Technologies’
standard commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than Restricted
Rights as defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no
greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227-7015
(b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
Caution denotes a hazard. It calls attention to a
procedure which, if not correctly performed or
adhered to, could result in damage to or destruction of the product. Do not proceed beyond a caution sign until the indicated conditions are fully
understood and met.
WARNING
Warning denotes a hazard. It calls attention to
a procedure which, if not correctly performed or
adhered to, could result in injury or loss of life. Do
not proceed beyond a warning sign until the indicated conditions are fully understood and met.
ii
Contents
1 Installing the 86100C
Installing the 86100C 1–2
Connector Care 1–37
Returning the Instrument for Service 1–42
2 Using the 86100C
A Quick Tour 2–2
Eye/Mask Mode 2–8
Jitter/Amplitude Analysis Mode 2–11
Oscilloscope Mode 2–13
TDR/TDT Mode 2–14
Quick Measure 2–15
Optional Features and Licenses 2–16
Windows XP Professional 2–19
If the Infiniium DCA has a Problem 2–22
3 Using the Built-In Information System
Using the Built-In Information System 3–2
Getting Help While Changing Instrument Settings 3–4
Learning About Measurement Results 3–5
Printing Topics 3–6
4 Working in Comfort
Working in Comfort 4–2
Contents-1
Contents
Contents-2
1
Installing the 86100C1-2
Step 1. Inspect the shipment1-5
Step 2. Position the Infiniium DCA1-7
Step 3. Install the modules1-8
Step 4. Connect the keyboard and mouse1-9
Step 5. Connect a printer (Optional)1-10
Step 6. Connect the power1-11
Step 7. Windows Setup Wizard1-14
Step 8. Change removable hard disk (Option 86100C-090)1-18
Step 9. Install a feature license (Optional)1-19
Step 10. Calibrate the touch screen1-22
Step 11. Avoid costly repairs1-24
Step 12. Calibrate the modules1-25
To configure a LAN connection1-28
To use the removable hard disk option1-30
To connect an external display1-35
To clean the 86100C1-36
Connector Care1-37
Returning the Instrument for Service1-42
Installing the 86100C
Installing the 86100C
Installing the 86100C
Installing the 86100C
This chapter shows you how to properly set up your 86100C. Follow the steps
in the order presented. In addition, you will learn about care and use of electrical and optical connections.
Figure 1-1. 86100C Front Panel
General Safety
Considerations
NOTEInstall the instrument according to the enclosure protection provided. This
This product has been designed and tested in accordance with the standards
listed in the Manufacturer’s Declaration of Conformity, and has been supplied
in a safe condition. The documentation contains information and warnings
which must be followed by the user to ensure safe operation and to maintain
the product in a safe condition.
instrument does not protect against the ingress of water. This instrument protects against finger access to hazardous parts within the enclosure.
1-2
Installing the 86100C
Installing the 86100C
WARNINGIf this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
WARNINGNo operator serviceable parts inside. Refer servicing to qualified
service personnel. To prevent electrical shock do not remove covers.
WARNINGThis is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
product is likely to make the product dangerous. Intentional
interruption is prohibited.
CAUTIONVENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4° C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
CAUTIONThis product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2 per IEC 1010 and 664 respectively.
Installation Category: Installation Categories (overvoltage categories) are
determined by the transient overvoltage levels that may be expected. CAT I:
Mains isolated. CAT II: Line voltage in appliance and to wall outlet. CAT III:
Line voltage behind wall outlet to next level of distribution.
Instrument MarkingsThe instruction manual symbol. The product is marked with this warn-
ing symbol when it is necessary for the user to refer to the instructions
in the manual.
The laser radiation symbol. This warning symbol is marked on products
which have a laser output.
ON, ❘, OFF, ❍
The ON, ON, STANDBY, OFF, OFF and AC symbols.
The ON, OFF, and STANDBY mark the position of the
instrument power line switch. The AC symbol is used to indicate the required
nature of the line module input power.
The CE mark is a registered trademark of the European Community.
The CSA mark is a registered trademark of the Canadian Standards
Association.
1-3
Installing the 86100C
Installing the 86100C
The C-Tick mark is a registered trademark of the Australian Spectrum
Management Agency.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in the
WEEE Directive Annex I, this product is classed as a “Monitoring and Control
instrumentation” product. Do not dispose in domestic household. To return
unwanted products, contact your local Agilent office, or refer to www.agilent.com for more information.
This text denotes the instrument is an Industrial Scientific and Medical
ISM1-A
Group 1 Class A product.
ICES / NMB-001
This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard.
1-4
Step 1. Inspect the shipment
❒ Inspect the shipping container and instrument for damage.
Keep the shipping container and cushioning material until you have inspected
the contents of the shipment for completeness and have checked the instrument mechanically and electrically.
❒ Locate the shipping list. Verify that you received all the accessories on this list,
and all the options that you ordered. The following list shows some of the items
that may be on the shipping list. The information on your actual shipping list is
more accurate and should supersede the information in this list.
• SMA (f-f) adapter, Agilent part number 1250-1158
• APC 3.5 (f-f) adapter, Agilent part number 5061-5311
Table 1-1 shows the available fiber-optic adapters for optical channels. These
adapters connect onto the front-panel optical channel connectors. FC/PC
adapters come preinstalled on new optical modules. Changing fiber-optic
1-5
Installing the 86100C
Step 1. Inspect the shipment
adapters is easy. Simply hand turn an adapter counter-clockwise to remove it
from a channel connector; to install a different adapter, hand turn the new
adapter clockwise onto the connector.
Table 1-1. Fiber-Optic Adapters for Optical Channels
Front Panel
Fiber-Optic AdaptersDescriptionAgilent Part Number
• 86100C-UK6, Commercial calibration certificate with test data
If anything is missing or defective, refer to “Contacting Agilent Technologies”
on page 1-42. If the shipment was damaged, contact the carrier, then contact
the nearest Agilent Sales Office. Keep the shipping materials for the carrier’s
inspection. The Agilent Sales Office will arrange for repair or replacement at
Agilent’s option without waiting for claim settlement.
1-6
Installing the 86100C
Step 2. Position the Infiniium DCA
Step 2. Position the Infiniium DCA
Position the Infiniium DCA so that it will have sufficient clearance for airflow
around the top, back, and sides. Review the following specifications to ensure
that your operating or storage environment is suitable for the instrument.
NOTEInstall the instrument so that the detachable power cord is readily identifiable
and is easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the
mains supply before other parts of the instrument. The front panel switch is
only a standby switch and is not a LINE switch. Alternatively, an externally
installed switch or circuit breaker (which is readily identifiable and is easily
reached by the operator) may be used as a disconnecting device.
CAUTIONIf the Infiniium DCA is mounted in a rack or cabinet, do not use the system’s
power switch to disconnect power from the Infiniium DCA. Instead, use the
Infiniium DCA’s front-panel power switch. Using the system power switch may
corrupt the Windows XP registry requiring you to perform a recovery process
to restore normal operation to the Infiniium DCA.
Table 1-2. Environmental Specifications
Useindoor
Temperature
Operating10 °C to +40 °C (50 °F to + 104 °F)
Non-operating–40 °C to +70 °C (–40 °F to +158 °F)
Altitude
OperatingUp to 4,600 meters (15,000 ft)
Line Power115V 5.7A
230V 3.0A
50/60 Hz
Weight
Mainframe15.5 kg (34 lb)
Modules1.2 kg (2.6 lb)
Dimensions (excluding handle)
Height (characteristic)
Width (characteristic)
Depth (characteristic)
Depth with front connectors
and rear feet
(characteristic)
215 mm (8.47 in)
425 mm (16.75 in)
566 mm (22.2 in)
629 mm (24.8 in)
1-7
Installing the 86100C
Step 3. Install the modules
Step 3. Install the modules
Up to two modules can be inserted into the Infiniium DCA. All documentation
(including specifications) for plug-in modules is located in the 86100C’s Help
system. To access the Help system, complete the installation steps, turn on
the 86100C, and click Contents on the Help menu.
1 Slide the module into an available front-panel slot.
You can remove or install a module while the instrument is operating.
2 Finger tighten the two knurled locking screws on the module’s front panel.
Figure 1-2. Installing a Module
1-8
Installing the 86100C
Step 4. Connect the keyboard and mouse
Step 4. Connect the keyboard and mouse
NOTETo complete the installation, you must connect the keyboard to the 86100C.
Since the 86100C has a touch screen, the use of a mouse is optional.
1 Connect the supplied USB keyboard to either of two USB ports located on the
rear panel. Most users find it more convenient to connect the keyboard to the
rear-panel leaving the front-panel USB port available for a USB flash drive.
Rear-panel PS/2 connectors are also provided in case you want to use a PS/2
keyboard or pointing device. Older AT style (DIN 5) keyboards will require an
adapter that is available at many computer stores.
Figure 1-3. Accessory Connections
2 Connect the optional mouse to any available USB port that is located on the
instrument including the two USB ports located on the keyboard.
1-9
Installing the 86100C
Step 5. Connect a printer (Optional)
Step 5. Connect a printer (Optional)
Perform this step if you plan to connect a printer directly to the serial, parallel,
or USB ports that are located on the 86100C rear panel. If you have a parallel
(Centronics) printer, you will need a parallel printer cable, such as an C2950A
(2 m) or C2951A (3 m) cable. If you have a serial printer, you will need a 9-pin
to 25-pin serial printer cable, such as an 34398A cable, plus the 34399A
adapter kit. Some printers may require other cable configurations, but the
86100C has a 9-pin serial connector.
• Attach the required parallel or serial printer cable between your printer and the
86100C’s rear-panel connector.
Figure 1-4. Location of Parallel and Serial Printer Ports
Support for printers with Plug and Play support will automatically be installed
when the instrument is first turned on. For other printers, use the Add Printer Wizard from the Windows XP desktop. From the Windows Start
menu, click Printers and Faxes, and then click Add a printer.
1-10
Installing the 86100C
Step 6. Connect the power
Step 6. Connect the power
CAUTIONAlways use the three-prong AC power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
product damage.
CAUTIONDo not connect ac power until you have verified the line voltage is correct as
described in Table 1-2 on page 1-7. Damage to the equipment could result.
• Connect the line cord as shown in the figure below.
Figure 1-5. Line Cord Connector on 86100C Rear Panel
The 86100C automatically adjusts for line input voltages of 115 VAC and
230 VAC. There is no voltage selection switch. The line cord provided is
matched by Agilent to the country in which the order originates. Refer to
Table 1-3, “Available Line Cords,” on page 1-13 for a list of other available line
power cords.
1-11
Installing the 86100C
Step 6. Connect the power
CAUTIONThis instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
CAUTIONWindows XP registry: If the Infiniium DCA is mounted in a rack or cabinet, do
not use the system power switch to disconnect power from the instrument.
Instead, use the Infiniium DCA front-panel power switch. Using the system
power switch may corrupt the Windows XP registry requiring you to perform
the recovery procedure to restore normal operation to the Infiniium DCA.
Refer to “To recover the 86100C Drive-C” on page 2-25.
1-12
Table 1-3. Available Line Cords
Installing the 86100C
Step 6. Connect the power
Plug TypeCable Part No. Plug Description
250V8120-1351
8120-1703
125V8120-1369
Straight *BS1363A
90°
Straight *NZSS198/ASC
90°
8120-0696
250V8120-1689
8120-1692
8120-2857p
125V8120-1378
8120-1521
8120-1992
250V8120-2104
8120-2296
Straight *CEE7-Y11
90°
Straight (Shielded)
Straight *NEMA5-15P
90°
Straight (Medical) UL544
Straight *SEV1011
1959-24507
Type 12 90°
220V8120-2956
8120-2957
250V8120-4211
8120-4600
Straight *DHCK107
90°
Straight SABS164
90°
Length
(in/cm)
90/228
90/228
79/200
87/221
79/200
79/200
79/200
90/228
90/228
96/244
79/200
79/200
79/200
79/200
79/200
79/200
ColorCountry
Gray
Mint Gray
Gray
United Kingdom, Cyprus,
Nigeria, Zimbabwe,
Singapore
Australia, New Zealand
Mint Gray
Mint Gray
Mint Gray
Coco Brown
Jade Gray
Jade Gray
East and West Europe,
Saudi Arabia, So. Africa,
India (unpolarized in many
nations)
United States, Canada,
Mexico, Philippines,
Taiwan
Black
Mint Gray
Switzerland
Mint Gray
Mint Gray
Denmark
Mint Gray
Jade GrayRepublic of South Africa
India
100V8120-4753
8120-4754
* Part number shown for plug is the industry identifier for the plug only. Number shown for
cable is the Agilent Technologies part number for the complete cable including the plug.
Straight MITI
90°
90/230
90/230
Dark GrayJapan
1-13
Installing the 86100C
Step 7. Windows Setup Wizard
Step 7. Windows Setup Wizard
This step requires several minutes to prepare Windows XP using the Windows
XP Setup Wizard. During the following steps, you will be instructed to “click”
with the pointing device. You can use a mouse (if you connected it) or you can
touch the screen with your finger.
CAUTIONBefore switching on this instrument, make sure the supply voltage is in the
specified range.
CAUTIONAvoid touching the screen with a sharp object, as this could result in damage to
the display. Instead, use your finger.
1 Press the ON/OFF switch, , located at the lower left-hand corner of the
front panel.
Figure 1-6. ON/OFF Switch
2 Wait until the Welcome to the Windows XP Setup Wizard appears as shown
in the following figure.
1-14
Figure 1-7. Windows XP Setup Wizard
3 Click Next to display the License Agreement.
Installing the 86100C
Step 7. Windows Setup Wizard
Figure 1-8. Windows XP License Agreement
4 Select I accept this agreement and then click Next.
1-15
Installing the 86100C
Step 7. Windows Setup Wizard
5 Enter your name and the name of your organization or company. Click Next.
Figure 1-9. Personal Information Setup
6 Click Next.
7 Adjust the date, time, and time zone as needed in the displayed screen. Check
automatic clock adjust for daylight saving changes, if it applies to your location.
Figure 1-10. Date, Time, and Time Zone
1-16
Installing the 86100C
Step 7. Windows Setup Wizard
8 Click Next. The setup will take a few moments to complete and to
automatically restart the instrument. In about one or two minutes, you should
see the 86100C screen screen, as shown in the following figure.
Figure 1-11. The 86100C Display
CAUTIONTo ensure proper operation of the 86100C, read “Avoid changing these
Windows system settings” on page 2-21.
NOTEWhen you press the ON/OFF switch to turn the instrument off, there is a small
time delay before you’ll see a change on the instrument. This is normal and
does not indicate a problem.
1-17
Installing the 86100C
Step 8. Change removable hard disk (Option 86100C-090)
Step 8. Change removable hard disk (Option 86100C-090)
If your instrument includes Option 86100C-090, you must repeat the Windows
XP Setup Wizard procedure for the second, identical hard disk. The removable
hard disk option, 86100C-090, is designed for 86100C instruments that are
used in a secure environment. For more information, refer to “To use the
removable hard disk option” on page 1-30.
1 Turn off the 86100C.
2 Locate the hard disk lock, which is located at the upper-left corner of the
86100C's rear panel.
3 Insert one of the supplied keys into the hard disk lock, and turn the key counter
clockwise to unlock the hard disk.
NOTEIdentical locks and keys are provided for all removable hard disks. The lock
mechanically secures the disk and is not intended for security purposes.
4 Press the button that is located beneath the lock, and pull out the hard disk.
5 Install the alternate hard disk. Make sure that the hard disk is oriented so that
the connector on the hard disk lines up with the connector in the mainframe.
6 Turn the key clockwise to lock the hard disk.
7 Repeat “Step 7. Windows Setup Wizard” on page 1-14 with the new hard disk.
8 Continue with “Step 9. Install a feature license (Optional)” on page 1-19.
1-18
Installing the 86100C
Step 9. Install a feature license (Optional)
Step 9. Install a feature license (Optional)
If you ordered an optional feature, such as Option 200 Enhanced Jitter Analysis, with your 86100C, it was probably installed at the factory. If you ordered it,
but it wasn’t installed, you received a Software Entitlement Certificate that
you must use to install the license on your 86100C.
To confirm if your license was installed
1 On the 86100C Utilities menu, click Licenses.
2 Check the installation status of the license in the Licenses dialog box.
License installed
Host ID and
Serial number
Figure 1-12. Licenses Installed Status
3 If the license that you purchased is installed, no further action is required.
Continue with “Step 10. Calibrate the touch screen” on page 1-22. If your
license was not installed, continue with “To install a license” on page 1-20.
1-19
Installing the 86100C
Step 9. Install a feature license (Optional)
To install a license
1 On the 86100C Utilities menu, click Licenses. Write down the HostID and
Serial # identifiers listed on the displayed dialog box. You will need this
information in the next step.
2 Go to the following web site to obtain your license file. You will need to enter
the HostID and Serial# identifiers for your 86100C. The license file will be sent
to you via email.
www.agilent.com/find/softwarelicense
3 After you receive your license file, copy the file to a memory location that is
available to the 86100C. This could be an external USB device (for example,
flash drive or CD drive) or a network drive.
4 On the 86100C Utilities menu, click Licenses to view the Licenses dialog box.
5 In the dialog box, click to highlight the feature that you want to activate. If the
feature is not listed, click the <new> listing.
6 Click Install License for Feature.
7 In the Install New License dialog box, click Browse for License File, find
the file, and then click Install This License File.
Figure 1-13. Install New License Dialog Box
If the license file is not available, some licenses can be manually entered by
clicking Enter 60 Digit Key Code. (This button may be unavailable for some
licenses.) Manually entering the key code fully fulfills the installation requirements. Open the license file on your PC using a text editor such as Notepad.
Look for the string SIGN= which marks the beginning of the 60-digit key code,
1-20
Installing the 86100C
Step 9. Install a feature license (Optional)
as shown in the following picture. In the Install New License dialog box,
click Enter 60 Digit Key Code and enter your key code. The letters in the key
code are not case sensitive, so you can enter letters in either upper or lower
case.
Figure 1-14. Example of the contents of a license file
.
Figure 1-15. Enter License Key Dialog Box
1-21
Installing the 86100C
Step 10. Calibrate the touch screen
Step 10. Calibrate the touch screen
The touch screen configuration utility aligns and orientates the touch screen
with the display so that when you touch an element on the display screen, the
instrument can detect the task that you want to perform. The display will
always function as a touch screen, even when a mouse and keyboard are connected.
1 On the Utilities menu, choose Touch Screen Config.
2 On the displayed dialog box’s Main tab, click the Calibrate button.
Figure 1-16. Touch Screen Calibration
3 Follow the on-screen instructions, while touching the displayed touch targets.
The calibration will time out within 20 seconds, if you do not touch the screen.
4 Each time that you touch the screen, an audible beep is heard. To turn the
audible click sound off, select the Touch Settings tab. Then, repeatedly click
the button shown in Figure 1-17 on page 23 (in the Touch Sound area) until the
button’s label reads “No beep”.
5 Close the dialog box.
1-22
Figure 1-17. Turning the Touch Beep Off
Installing the 86100C
Step 10. Calibrate the touch screen
1-23