Keysight (Agilent) 8594EM Service Note

8594EM-09
SERVICE NOTE
SUPERSEDES: None
Agilent 8591C, 8591E, 8593E, 8594E, 8595E, 8596E, 8591EM, 8593EM, 8594EM, 8595EM, 8596EM, 8594Q, 11757B, and 11758B
859XE Spectrum Analyzer Serial Numbers:
8591C -13 3916A04099 / 3916 A04355 8591E -16 3916A07840 / 3916A08000 8593E -17 3926A04414 / 3926A04567 8594E -16 3911A09049 / 3916A09232 8595E -16 3911A04772 / 3911A04951 8596E -16 3911A01814 / 3911A01861
8591EM-09 3916A01485 / 3916 A01488 8593EM-09 3926A00300 / 3926 A00306 8594EM-09 3911A00388 / 3911 A00391 8595EM-09 3911A00195 / 3911 A00196
Digital Video Broadcast Analyzer Serial Numbers:
8594Q-04 US00 1401 11 / US00140125
Digital Radio Test System Serial Numbers:
11757B-06 US00140105 / US00140 107 11758B-02 US00140103 / US00140 116
Instruments within this serial range can experience a ’smplr unlock’ er­ror message on the analyzer screen.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION AVAILABLE
ACTION CATEGORY:
AGREEABLE TIME
Continued
DATE: January 2001
PERFORMANCE ENHANCEMENT SERVICE/RELIABILITY ENHANCEMENT
LOCATION CATEGORY:
AUTHOR: BD
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
ENTITY: 5330
© 2000 AGILENT TECHNOLOGIES PRINT ED IN U.S.A.
AVAILABLE UNTIL: End of GMS
ADDITIONAL INFORMATION:
1 of 2
Page 2 Service Note 8594EM-09
To Be Performed By: Agilen t Se r v ic e Ce nte r s O n ly
Parts Required:
Remove A9, Th ir d Converter asse mbly and repla ce it ON L Y if the customer com pla int is re­garding the ’smplr unl oc k’ me ssa ge on the a na lyze r di spla y.
Situation:
Manufacturin g disc overed that the plas tic pa ck ag e for A9U1 pr e sc al ar (ve ndor pa rt number MC12090P) that divides down the 600 MHz SAW R os ci lla tor to 300 MHz for use on the A25 Counterlock bo ard wil l not divide correctly and result in a 200 MHz output wit h 600 MHz mix­ing harmonics. The instru me nt di splay will read ’smplr unlock’. The problem ca n be reme died by leaving the system on for 5 minutes to warm it up, and cycle the power. Once the chip is os­cillating cor rec tly, it is stable over te mperature as lo ng a s the pow e r isn’ t c yc le d again while it is cold.
Solutio n / A ction:
Change out A9, Thir d Co nve r ter asse mbly. All field stoc k ha s be en pulle d to verify A9U1 is the correct style of prescalar.