
Quick Reference Guide
HP 8590 EM Series EMC Analyzer
ABCDE
HP Part No. 5963-2929
Printed in USA August 1995

Notice
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and
tness
contained herein or for incidental or consequential damages in connection with
the
for
a
particular
furnishing, performance, or use of this material.
purpose
.
Hewlett-P
ackard shall not be liable for errors
Restricted
Use, duplication, or disclosure by the U.S. Government is subject to restrictions
as set forth in subparagraph (c) (1) (ii) of the Rights of Technical Data and
Computer Software clause at DFARS 252.227-7013 for DOD agencies, and
subparagraphs (c) (1) and (c) (2) of the Commercial Computer Software
Restricted Rights clause at FAR 52.227-19 for other agencies.
Rights Legend.
c
Copyright Hewlett-Packard Company 1995
All Rights Reserved. Reproduction, adaptation, or translation without prior
written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA

Introduction
This
guide
provides
a
quick
reference
for
experienced
analyzer users.
Chapter
measurement
Chapter
Chapter
Appendixes
1
summarizes
and
2
contains
3
contains
A,
B
, C, and D contain helpful charts and tables.
the
how
front-panel
to
perform
features
the
self-calibration
,
and tells how to make a basic
routines.
brief descriptions of the EMC analyzer functions.
the remote programming commands.
Appendix E contains diagrams of the key menus for the HP 8590 EM Series
EMC
analyzers
F
or
additional
Analyzer
User's
.
instrument
Guide
for
information,
your
specic
consult
the
instrument.
HP
8590
EM
Series EMC
Guide Terms and Conventions
The
six
keys
along
the
right
side
of
the
are
An
example
displayed
display
on the screen. The softkeys appear in shaded boxes in this guide.
of
a
softkey
N
N
N
N
N
N
N
N
N
N
N
N
NNNNNNNNNNN
REF
LVL
.
is
The
are called softkeys. Their labels
labeled
keys
that
are
on the front
panel of the analyzer are called front-panel keys. Pressing a front panel key
changes the softkey labels or initiates functions. The front-panel keys appear in
unshaded boxes in this guide. An example of a front-panel key is
The
input
of
the
EMC
Caution
using
a
line
impedance
analyzer can be damaged easily.When
stabilization
network (LISN) device with
4
FREQUENCY
5
.
the EMC analyzer, disconnect the analyzer from the LISN device
before changing either the switch position on, or the voltage to,
the LISN device.
iii

NOTES
iv

Contents
1. Getting Acquainted with the EMC Analyzer
Front-Panel Features . . . . . . . . . . . . . . . . . . . . . . 1-1
EMC Analyzer Screen Annotation . . . . . . . . . . . . . . . . 1-4
Making a Basic Measurement . . . . . . . . . . . . . . . . . . 1-7
Performing Self-Calibration Routines ...... ...... .. 1-11
Warm-Up Time ....................... 1-11
Self-Calibration
Self-Calibration
P
erforming
(Option
Spectrum
2.
EMC Analyzer Functions
Analyzer
Analyzer
Functions ...................... 2-1
the
010)
Routines
Routine
Tracking
.
.
.
Mode
.
.
.
Problems
Generator
.
.
.
.
Screen
.
.
.
.
.
...... ..... 1-11
.
.
.
.
.
.
.
........ 1-12
Self-Calibration
.
...... ...... ... 1-13
Annotation
.
Routine
.
.
.
...... 1-14
3. Programming Commands
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
How to Use This Chapter . . . . . . . . . . . . . . . . . . . . 3-1
Notation Conventions .................... 3-2
Syntax Conventions ...... ...... ...... ... 3-2
Functional Index ....................... 3-6
AMPLITUDE ...... ...... ...... ...... 3-6
AUTO COUPLING ...................... 3-7
AUXILIARY CONTROL
...... ...... ...... .. 3-7
BANDWIDTH .... ...... ...... ...... .. 3-8
CALIBRATION....... ...... ...... ..... 3-8
COMMAND TRIGGER
.................... 3-9
CONFIGURATION .. ...... ...... ...... ..
DISPLAY .......................... 3-10
EMC MEASUREMENTS ..... ..... ...... ... 3-10
EMCOUTPUT.. ...... ..... ...... ..... 3-11
FREQUENCY ........................ 3-11
3-9
Contents-1

GRAPHICS .......... ...... ...... ... 3-12
INFORMATION ....................... 3-12
INPUT and OUTPUT . . . . . . . . . . . . . . . . . . . . . 3-13
LIMIT LINES . ...... ..... ...... ...... 3-14
MARKER .... ...... ...... ...... .... 3-15
MATH......... ...... ..... ...... .. 3-16
MEASURE/USER ...................... 3-17
MODE............................ 3-17
OPERATOR ENTRY ..................... 3-17
PLOTTER . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
PRESET . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
PRINTER
PROGRAM
RECALL
SIGNAL
SP
AN
SWEEP
.
......................... 3-18
FLOW
or
SA
LIST
.
.
...... ...... ..... ...... ... 3-20
.
.
.
.
.
.
...... ...... ..... .. 3-18
VE
.
..................... 3-19
.
.
...................... 3-19
........................ 3-20
SYNCHRONIZATION ..................... 3-21
TRACE ........................... 3-21
TRACE MATH .... ...... ...... ...... .. 3-22
TRIGGER ........ ...... ...... ...... 3-23
USER-DEFINED . . . . . . . . . . . . . . . . . . . . . . . 3-23
WINDOWS
Programming
.
.
Codes
.
.
.
.
.
.
.
.
.
...... ...... .. 3-24
.
.
.
.
.
.
.
.
.
...... ...... . 3-25
Characters and Secondary Keywords (Reserved Words) Summary . 3-78
A. EMC Analyzer Error Messages
Error
Messages
Nonrecoverable
B
.
AM, FM, and Pulsed RF Reference Charts
.
....................... A-1
System
Errors .... ...... ...... . A-15
C. Cross Reference of Programming Command to Key Function
D. Locating a Softkey
E. Key Menus
Index
Contents-2

Figures
1-1. Front-Panel Feature Overview ................ 1-2
1-2. EMC Analyzer Screen Annotation . . . . . . . . . . . . . . . 1-4
1-3. Placing the Marker on a Signal ................ 1-8
1-4. Displaying a List on the Screen . . . . . . . . . . . . . . . . 1-9
1-5. Displaying the List as a Log Graph .............. 1-10
1-6.
B-1.
B-2.
B-3.
B-4.
Spectrum
P
ercent
Bessel
Loss
IF
BW
Analyzer
Modulation
Null
in
Sensitivity
Setting
Graph
for
Mode
.
.
.
(Pulsed
Pulsed
Screen
.
.
.
.
RF
RF
Annotation
.
.
.
.
.
.
.
.
versus
Computed
.
.
.
.
.
.
.
.
.
.
.
........... B-2
.
.
...... ...... B-4
CW)
.
.
.
.
.
.
..... B-5
B
=
0.1
.
.
.
from
t
o
.
. 1-14
.
Tables
1-1.
EMC
1-2.
Screen
1-3.
Spectrum
1-4.
Screen
B-1.
Determining
Analyzer
Annotation
Annotation
Screen
Analyzer
Amplitude
Annotation . . . . . . . . . . . . . . . 1-5
for
Trace, Trigger, and Sweep Modes .... 1-6
Mode
Screen Annotation ......... 1-15
for Trace, Trigger, and Sweep Modes .... 1-16
Modulation . . . . . . . . . . . . . . B-1
B-2. Carrier and First Sideband Charts for Calibrating Deviation . . . B-3
C-1. Cross Reference of Programming Command to Key Function .. C-1
D-1. Softkey
Locations ...................... D-1
B-6
Contents-3


Getting Acquainted with the EMC
Analyzer
This chapter provides an introduction to the EMC analyzer's front-panel
features, an explanation of screen annotation, the procedure for making a
basic measurement with the EMC analyzer, and the procedure for performing
self-calibration routines. The EMC analyzer also supplies a spectrum analyzer
operating mode with screen annotation described in \Spectrum Analyzer Mode
Screen
Annotation
"
at
the
end
of
this
chapter
.
Front-Panel Features
The following section provides a brief description of front-panel features.
Refer
to
1
Figure
Front-P
provided
A
ctive
1-1
.
anel
Display
to
the
Function
user
is
.
That
Block
the
area on the front-panel where information is
information includes:
area.
Functions
appearing in this block include,
for example, marker, span, center frequency, and reference level.
1
Message Block area. This block is where
(*) appear. If one or more functions are manually set (uncoupled), and
the amplitude or frequency becomes uncalibrated,
(Press
asterisk indicates that the display is not updated for new settings.
Softkey labels. Softkey labels are the annotation on the screen next to
the unlabeled keys. Most of the labeled keys on the EMC analyzer's
front panel (also called front-panel keys) access menus of related
softkeys.
2 Softkeys
MEAS UNCAL
NNNNNNNNNNNNNNNNNNNNNNNNNN
4
AUTO COUPLE
5
, then
AUTO ALL
, to recouple functions.) The
are the unlabeled keys next to the screen.
Getting Acquainted with the EMC Analyzer 1-1
and the asterisk
MEAS UNCAL
appears.

Figure 1-1. Front-Panel Feature Overview
3
4
FREQUENCY
5
,
4
SPAN
5
,
and
4
AMPLITUDE
5
are
the
three
large
that activate the primary EMC analyzer tuning functions and access
menus of related functions.
4
4
5
and
4
5
SETUP
are the dark-gray keys that access menus of functions
TEST
used to setup the instrument, then make fast and accurate EMC
measurements.
5 INSTRUMENT STATE
functions aect the state of the entire EMC
analyzer. Self-calibration routines and special-function menus are
accessed with these keys. The green
4
PRESET
5
key resets the entire EMC
analyzer state and can be used as a \panic" button when you wish to
return to a known state.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
6
4
5
key prints or plots screen data. Use
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Print Config
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COPY DEV PRNT PLT
,and
4
CONFIG
5
,
Plot Config
before pressing
1-2 Getting Acquainted with the EMC Analyzer
dark-gray
4
COPY
5
keys
or
.

7
MEASUREMENT CONTROL
make
manual or automated quasi-peak average measurements, adjust the
functions access menus that allow you to
IF bandwidth, adjust the sweep time, store and manipulate trace data,
demodulate signals, control the markers, control the instrument display,
and make measurements.
8 WINDOWS
4
5
NEXT
functions include:
4ON5
to turn on the windows display mode,
for switching between windows,and
4
ZOOM
5
for controlling the
zone span and center frequency.
9 DATA
keys,
STEP
keys,and
knob
allow you to change the numeric value
of an active function.
10 INPUT 50
Caution
11
PROBE
accessories
12
CAL
OUT
13 VOL-INTEN
is the signal input for the analyzer.
Excessive signal input will damage the EMC analyzer input
attenuator
EMC
PWR
provides
and
analyzer
can
the
the
power
input
mixer
.
The
maximum
tolerate
appears
on
the
for an active probe and other
front
.
provides
a
calibration signal of 300 MHz at 87 dBV.
changes the brightness of the screen display, and the volume
of the speaker.
14 Memory card reader
reads from a read-only (ROM) or random access
(RAM) memory card. The memory card reader writes to a RAM card.
15 100 MHz COMB OUT
HP 8593EM and HP 8596EM EMC analyzers. The output is used to
perform a
NNNNNNNNNNNNNNNNNNNNNNN
CAL YTF
is the output of the comb generator of the
.
power
panel.
that
the
16 RF OUT 50
only
.
17
4
5
turns the instrument on or o and performs an instrument check.
LINE
is the output for the built-in tracking generator.
Getting Acquainted with the EMC Analyzer 1-3
Option 010

EMC
Analyzer
Screen
Annotation
Figure
the
EMC
refers
1-2
analyzer
to
Figure
shows
an
.
1-2
example
T
able
.
of
screen
1-1
lists the screen annotation features numerically and
annotation
as it appears on the screen of
Figure 1-2. EMC Analyzer Screen Annotation
In Figure
1-2, item number 19 refers to the trigger and sweep modes of the EMC
analyzer. The rst letter (\F") indicates the EMC analyzer is in free-run trigger
mode. The second letter (\S") indicates the EMC analyzer is in single-sweep
mode. (\C") indicates the EMC analyzer is in continuous-sweep mode.
Item number 20 refers to the trace modes of the EMC analyzer. The rst letter
(\W") indicates the EMC analyzer is in clear-write mode
. The second letter (\A")
represents trace A. The next two letters (\SB") indicate the store-blank mode
(\S") for trace B (\B"). The trace mode annotation for trace C is displayed under
1-4 Getting Acquainted with the EMC Analyzer

the trace mode annotation of trace A. In Figure 1-2, the trace C trace mode is
\SC," indicating trace C (\C") is in the store-blank mode (\S").
The WINDOWS display mode splits the screen into two separate displays.Only
one of these displays is active at a time. The currently active window will
have a solid line around the graticule rather than a broken line. The complete
annotation is not available for each window because of space limitations.
Table 1-1. EMC Analyzer Screen Annotation
Item Feature Item Feature
1 reference level 16 threshold
2 active function block 17 corrections factors on
3 time/date
and
command
line
18 amplitude
correction
4 screen title factors on
5 autorange 19 trigger
6 active and measurement detector 20 trace mode
7 external preamp 21 display line
8 marker 22 video average
9 softkeys 23 attenuator setting
10 frequency
span
or
stop
24 amplitude
scale
frequency
11 sweep time
12 frequency oset
13 A
VG
bandwidth
14 IF bandwidth
15 center or stop frequency
Getting Acquainted with the EMC Analyzer 1-5

Table 1-2 shows the dierent screen annotation codes for trace, trigger,and
sweep modes.
Table 1-2.
Screen Annotation for Trace, Trigger, and Sweep Modes
Trace Mode Trigger Mode Sweep Mode
W
=
clear-write (traces A/B/C)
M = maximum hold (traces A/B) L = line S = single sweep
M = minimum hold (trace C) V = video
V = view (traces A/B/C) E = external
S
=
store-blank (traces A/B/C)
F
T
=
=
free run
TV
(Option 102 only)
C
=
continuous
1-6 Getting Acquainted with the EMC Analyzer

Making
Basic
measurements
the
screen,
signals
,
and
a
Basic
then
placing
moving
Measurement
simply
those
involve
a
marker
signals
on
to
tuning
the
instrument
to
place a signal on
one or more signals, measuring those
a list.
Caution
Do not exceed the maximum input power that is printed on the
front panel of the EMC analyzer.
Let's begin using the EMC analyzer by measuring an input signal. Since the
300 MHz calibration signal (CAL OUT) is readily available, we will use it as our
input
signal.
1.
2.
Press
4
LINE
key
.
V
erify
that the start frequency is set to 200 MHz and the stop frequency to
1.0
GHz.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
EMC ANALYZER
5
to
turn
the
instrument
If
not, the instrument has been congured as a spectrum analyzer.
on,
and
then
press
the
green
4
PRESET
and proceed with the next step.
5
3. Connect CAL OUT to the EMC analyzer INPUT 50 connector (on the front
panel) using an appropriate BNC cable and a BNC-to-Type N adapter.
Option 026 only:
Connect the SMA (m) to SMA (m) cable to the EMC
analyzer input with an APC-3.5 connector. Connect the cable to CAL OUT
with the BNC-to-SMA adapter.
4. Press
4
MARKER
5.MARKER
TEST
softkey label appears in inverse video to indicate that the marker is
is now displayed in the active function block, and the
the active function. Place the marker on a signal by pressing the down
keys
.
Refer
to Figure 1-3.
495
Getting Acquainted with the EMC Analyzer 1-7

Figure 1-3. Placing the Marker on a Signal
N
N
N
N
N
N
5.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE
AT MKR
. In a short time, the measured marker values are
displayed in the upper-right corner of the display.
N
N
6.
Press
internal list.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADD TO LIST
Signal Added
. The measure-at-marker results are saved to an
is now displayed on the screen.
7. Press the up
485
key one or more times to place the marker on a second
signal.
8.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
MEASURE AT MKR
, then
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADD TO LIST
to add the second signal to the
internal list. Repeat the process and place a third signal in the internal list.
9.
To display the list on the screen, press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SIG LIST ON OFF
softkey. The measure-at-marker results in the form of a
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
More 1 of 3
, then the
list are displayed on the upper portion of the screen. Refer to Figure 1-4.
1-8 Getting Acquainted with the EMC Analyzer

Figure 1-4. Displaying a List on the Screen
10. The list can be graphed on screen in either log or linear frequency scales by
pressing
11.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
SAVE LOG GRAPH
clear the display's active function area, press
lower-right corner). Refer to Figure 1-5.
N
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Save/Rcl
the
List
softkey.
to view the signals in the list on a log graph. To
4
5
ENTER
Getting Acquainted with the EMC Analyzer 1-9
key (front-panel's

Note
Figure 1-5. Displaying the List as a Log Graph
If
the
EMC
analyzer
is
connected
to
either
a printer or plotter:
a. Press the
4
COPY
5
key to obtain a printed copy of the graph.
Changing the conguration may be required, or
b. Press the
4
OUTPUT
5
key to access the softkeys that will help
you:
Dene a report.
Dene a list.
Add up to two pages of text using a keyboard.
12. T
o return the instrument to a known state,press
1-10 Getting Acquainted with the EMC Analyzer
4
PRESET
5
.

P
erforming
The
self-calibration
circuitry
and
.
The
amplitude
Self-Calibration
routines
addition
specications
of
add
the
correction
.
osets
Routines
,
called
correction
factors, to internal
factors is required to meet frequency
Warm-Up Time
To meet EMC analyzer specications:
1. The EMC analyzer should be stored at a constant temperature, within the
operating temperature range, for at least 2 hours.
2. Turn on the EMC analyzer and allow the EMC analyzer to warm up for
30 minutes.
P
Note
erform
the
attempt
analyzer
the
EMC
analyzer
to
make
self-calibration
EMC analyzer self-calibration routines only
has met the specied warm-up time. Do not
any calibrated measurements until the EMC
routines
have been performed.
after
Self-Calibration
P
erform
1.
the
Connect
following
the
CAL
Routines
steps
to
OUT
connector
self-calibrate
to
the
the
instrument:
INPUT 50 connector, using an
appropriate cable.
2.
Press
Caution
the
following
To avoid losing the data when the EMC analyzer is turned o,
press
EMC
analyzer
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNN
CAL STORE
keys:
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNN
CAL STORE
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
5
CAL
4
CAL
,
FREQ
&
AMPTD
stores the cal data in the area of
.
EMC analyzer memory that is saved when the analyzer is turned
o.
The frequency and amplitude self-calibration routines take approximately
9 minutes to nish, at which time the correction factors will be stored (when
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL STORE
The frequency and amplitude self-calibration functions can be done separately
by using
is pressed) in the EMC analyzer's memory.
NNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FREQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
or
CAL AMPTD
instead of
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FREQ & AMPTD
.
Getting Acquainted with the EMC Analyzer 1-11

NNNNNNNNNNNNNNNNNNNNNNNNNN
If
Note
CAL FREQ
N
N
N
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNN
CAL
the
N
N
N
N
N
N
N
CAL
Interrupting
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ
NNNNNNNNNNNNNNNNNNNNNN
AMPTD
CAL FREQ & AMPTD
in
corrupt
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
data
CAL FREQ & AMPTD
When
the self-calibration routines have successfully nished,
appears
on
the left side of the screen.
and
routine
routine
N
N
N
CAL
the
stored
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL AMPTD
should
self-calibration routines are used,
always
be
performed
.
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNN
AMPTD
NNNNNNNNNNNNNNNNNNNNNNN
CAL
,
FREQ
,or
self-calibration routines may result
in
RAM.
(If this occurs, rerun the
routine.)
CORR
before
the
(corrected)
Self-Calibration Routine Problems
If the correction data has been corrupted or is obviously inaccurate,use
N
N
N
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNNNNN
CAL
FETCH
retrieve correction factor data, press
to
retrieve
the
correction
data that has previously been saved. To
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
5
,
More 1 of 4
4
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNN
CAL FETCH
,
.Ifthe
fetched correction data is corrupt, the following procedure can be used to set
the correction data back to predetermined values:
1.
Press
4
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
DEFAULT CAL DATA
2.
Perform the
N
5,0374Hz5,4
.
N
NNNNNNNNNNNNNNNNNNNNNNNN
CAL FREQ
and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
5
,
More 1 of 4
CAL
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL AMPTD
routines,orthe
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
,
More 2 of 4
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FREQ & AMPTD
,
routine. Be sure CAL OUT is connected to the EMC analyzer input.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Note
Using
DEFAULT CAL DATA
may cause the self-calibration
routine to fail (the frequency span error may interfere with
the EMC analyzer routine that locates the 300 MHz calibration
signal). If this occurs, press
performing the
NNNNNNNNNNNNNNN
N
CAL FREQ
4
NNNNNNNNNN
FREQUENCY
routine, or the
5,0374Hz5
NNNNNNNNNNNNNNN
N
,before
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FREQ & AMPTD
routine.
If the self-calibration routines cannot be performed, see the EMC analyzer's
HP 8590 EM Series EMC Analyzer User's Guide
.
1-12 Getting Acquainted with the EMC Analyzer

Performing the Tracking Generator Self-Calibration Routine
(Option 010)
To meet the tracking generator specications, allow the EMC analyzer to
warm up for 30 minutes after being turned on before attempting to make
any calibrated measurements. Be sure to calibrate the EMC analyzer and the
tracking generator only
after
the EMC analyzer has met operating temperature
conditions.
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Note
Since
level
be calibrated prior to using
CAL
the
of
the EMC analyzer, the EMC analyzer amplitude should
TRK GEN
routine uses the absolute amplitude
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL TRK GEN
.
1. To calibrate the tracking generator, connect the tracking generator output to
the EMC analyzer input connector, using an appropriate cable.
N
N
2.
Press
the
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
following
EMC
analyzer
CAL TRK GEN.TG SIGNAL NOT FOUND
generator
process
3.
Press
memory
4
CAL
will
5
that
output
is
take
approximately
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
,
CAL STORE
is
saved
not
connected
to save this data in the area of EMC analyzer
when
the
N
N
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNN
More
keys:
4
CAL
5
,
1
of
will be displayed if the tracking
to
the
EMC
30
seconds
analyzer
.
is
turned
analyzer
input.
o.
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNN
More
2
of
4
,
The
cal
4
,
Getting Acquainted with the EMC Analyzer 1-13

Spectrum
Analyzer
Mode
Screen
Annotation
Figure
1-6
shows
an
example
of
of the EMC analyzer when in the
modes
annotation
are
changed
features
using
the
numerically
4
MODE
screen
annotation
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
as it appears on the screen
SPECTRUM ANALYZER
5
front-panel
and
refers
key
to Figure 1-6.
mode. Instrument
.
T
able 1-3 lists the screen
Figure 1-6. Spectrum Analyzer Mode Screen Annotation
In Figure 1-6, item number 21 refers to the trigger and sweep modes of the
spectrum analyzer
. The rst letter (\F") indicates the spectrum analyzer is in
free-run trigger mode. The second letter (\S") indicates the spectrum analyzer is
in single-sweep mode.
Item number 22 refers to the trace modes of the spectrum analyzer mode.The
rst letter (\W") indicates the spectrum analyzer is in clear-write mode. The
second letter (\A") represents trace A. The next two letters (\SB") indicate the
store-blank mode (\S") for trace B (\B"). The trace mode annotation for trace C
1-14 Getting Acquainted with the EMC Analyzer

is displayed under the trace mode annotation of trace A. In Figure 1-6, the trace
C trace mode is \SC," indicating trace C (\C") is in the store-blank mode (\S").
The WINDOWS display mode splits the screen into two separate displays.Only
one of these displays is active at a time. The currently active window will
have a solid line around the graticule rather than a broken line. The complete
annotation is not available for each window because of space limitations.
Table 1-3. Spectrum Analyzer Mode Screen Annotation
Item Feature Item Feature
1 detector mode 14 video bandwidth
2 reference level 15 resolution bandwidth
3 time
and
date
display
16 center
frequency
or
4 screen title start frequency
5 RF attenuation 17 remote operation
6 preamplier
7 external
keyboard
gain
entry
18 threshold
19 correction
factors
on
8 marker or signal track readout 20 amplitude correction
9 measurement-uncalibrated or factors on
function-in-progress 21 trigger
messages 22 trace
10 service
request
23 video
mode
average
11 frequency span or stop 24 display line
12 frequency 25 amplitude oset
sweep time 26 amplitude scale
13 frequency oset 27 active function block
Getting Acquainted with the EMC Analyzer 1-15

Table 1-4 shows the dierent screen annotation codes for trace, trigger,and
sweep modes.
Table 1-4.
Screen Annotation for Trace, Trigger, and Sweep Modes
Trace Mode Trigger Mode Sweep Mode
W
=
clear-write (traces A/B/C)
M = maximum hold (traces A/B) L = line S = single sweep
M = minimum hold (trace C) V = video
V = view (traces A/B/C) E = external
S
=
store-blank (traces A/B/C)
F
T
=
=
free run
TV
(Option 102 only)
C
=
continuous
1-16 Getting Acquainted with the EMC Analyzer

EMC Analyzer Functions
This section lists the softkey and front-panel functions in alphabetical order.
Next to each key label is a brief description of its function. For more detailed
descriptions, refer to the
softkeys are shown in the menu diagram in Appendix E of this guide.The
functions accessed by
HP 8590 EM Series EMC Analyzer User's Guide
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
N
Service Diag
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
N
and
Service Cal
are not included in
. All
this listing.
2
Analyzer
N
N
N
N
N
N
NNNNNNNN
%
AM
N
N
N
N
N
N
N
NNNNNNN
A/m
NNNNNNNNNNNNN
N
V/m
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
0-2.9 Gz
BAND
0
3
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
2.75-6.5
BAND 1
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
12.4-19.
BAND 3
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
19.1-22
BAND 4
3
3
3
Functions
determines the percent of amplitude modulation of the largest
displayed signal and its two sidebands.
selectsA/m as the transducer conversion units for the antenna
amplitude-correction factors using magnetic eld units.
selectsV/m as the transducer conversion units for the antenna
amplitude-correction factors using electric eld units.
locks onto harmonic band 0. Harmonic band 0 uses low-pass
ltering instead of bandpass preselection. It has a specied
tuning range of 0 to 2.9 GHz.
HP 8593EM, HP 8595EM,and
HP 8596EM only.
locks onto harmonic band 1. Harmonic band 1 is preselected
and has a specied tuning range of 2.75 GHz to 6.5 GHz.
HP 8593EM, HP 8595EM, and HP 8596EM only.
locks onto harmonic band 3. Harmonic band 3 is preselected
and has a specied tuning range of 12.4 to 19.4 GHz.
HP 8593EM only
.
locks onto harmonic band 4. Harmonic band 4 is preselected
and has a specied tuning range of 19.1 to 22 GHz.
HP 8593EM
only.
EMC Analyzer Functions 2-1

x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
120
kHz
EMI BW
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
150
0
30 MHz
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
200 Hz
EMI BW
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
200
0
1GHz
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
30 MHz
0
300 MHz
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
3
POINTS
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
3rd
MEAS
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
6dB
POINTS
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
6.0-12.8
3
kHz
3
MHz
3
dB
3
ORD
3
3
BAND 2
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
9 kHz
EMI
BW
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
9 kHz
0
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
99%
PWR BW
3
150 kHz
3
NNNNNNNNNNNNNNNNNNNNNNNN
A
!
B
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
A0B
!
A
ON OFF
NNNNNNNNNNNNNNNNNNN
A
!
C
allows a 6 dB IF bandwidth of 120 kHz.
presets
3
for EMC measurements.
the
instrument
to
the
predened
CISPR
band
B
settings
allows a 6 dB IF bandwidth of 200 Hz.
presets
measurements
the
receiver to the predened settings for EMC
.
presets the receiver to the predened settings for EMC
measurements
3
.
initiates an automatic 3 dB bandwidth measurement on the
highest peak displayed on screen.
initiates an automatic frequency and amplitude measurement
on
a
and
the
third-order
third-order
product.
product
At
least
must
one
be
displayed
of
the
fundamental
on
the
tones
screen.
initiates an automatic 6 dB bandwidth measurement on the
highest peak displayed on the screen.
locks onto harmonic band 2. Harmonic band 2 is preselected
3
and has a specied tuning range of 6.0 to 12.8 GHz.
HP 8593EM, and HP 8596EM only.
allows a 6 dB IF bandwidth of 9 kHz.
3
presets
for
initiates
the
EMC
measurements
instrument
to the predened CISPR band A settings
.
an automatic measurement of the 99% power
bandwidth of the largest signal displayed on the screen.
exchanges the contents of the trace A register with the trace B
register and puts trace A in view mode
.
when ON is underlined, subtracts the data in trace B from the
3
measured data in trace A.
copies trace A into trace C.
2-2 EMC Analyzer Functions

N
N
N
N
N
NNNNNNNNNNNNNNN
ABCDEF
N
N
N
N
N
N
N
NNNNNNNNNN
Abort
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ADD TO
LIST
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ALC
INT
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ALC
EXT
MTR
3
3
INT XTAL
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ALL
DLP
!
CARD
3
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNN
Amp
Cor
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AMP COR
ON OFF
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AMPL SCL
3
LOG LIN
accesses
A
through F.
allows
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO-MEASURE
adds a signal to the internal signal list. The signal added is
either for the last
at
the
analyzer
the softkey menu for selecting screen title characters
you
to
exit from an
OUTPUT
REPORT
or
an
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE AT MKR
result or the measurement
marker position after a stepped measurement.
mode
only.
activates internal (INT) leveling or external (EXT) leveling.
Only
available
HP
8593EM
Option
010
activates
internal
3
HP 8591EM with Option 010 only.
,
only
the
(INT)
in
EMC
HP
8594EM
.
automatic
leveling
Analyzer
,
HP
8595EM
leveling
or
external
measurement
mode
, or HP 8596EM with
control
(XT
(ALC)
AL
or MTR) leveling.
functions
saves all the downloadable programs and key denitions
that are in analyzer memory onto the memory card. If the
downloadable program was stored using a prex, the le name
for the downloadable program consists of d(prex) (register
number). If no prex was specied, the data is stored with the
le name d (register number).
accesses the menus for entering and editing the current
amplitude-correction factors. This function is supplied to
maintain compatibility with the HP 8590EL Series spectrum
analyzers
and
other
.
The
EM-series
amplitude
analyzers
correction
also include antenna, cable,
factors
.
turns the current amplitude-correction factors on and o.
species whether the limit line is derived from a logarithmic or
linear amplitude axis. Underline LIN to set the amplitude axis
3
to linear or LOG to set the amplitude axis to logarithmic.
EMC
.
for
4
AMPLITUDE
5
activates the reference level function and accesses the
amplitude menu.
EMC Analyzer Functions 2-3

x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
Amptd
Units
3
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ANALOG+
ON
OFF
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ANALYZER
ADDRESS
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ANNOTATN
ON OFF
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ANTENNA
!
CARD
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
Antenna
Factors
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ANTENNA
ON
OFF
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
Antenna
Units
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
APND
CAT ITEM
accesses a menu to select the desired amplitude units.
Amplitude units can be selected by pressing,
NNNNNNNNNNNNNN
dBuV
NNNNNNNNNNNNNNNNN
,
Volts
,or
NNNNNNNNNNNNNNNNN
Watts
.
N
NNNNNNNNNN
dBm
N
NNNNNNNNNNNNN
dBmV
,
,
turns on the analog+ display mode. This is a digital
3
3
3
3
implementation of an analog display, combining the advantages
of both types of displays.
allows you to set the HP-IB address of the instrument. The
address is set to 18 by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BAUD RATE
turns
the
4
DISPLAY
and
o
saves
the
card.
#
and
for option 043.
screen
5
key
.
the
printing
current
T
o
save
PREFIX=
annotation
When
antenna amplitude-correction factors to a
antenna
are displayed on the screen. Use the data keys
to enter the desired register number then press
Requires Option 101 or 301.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT CONFIG.Changes to
on
and
o
when
accessed
of
the
data
using
the
annotation.
N
N
N
N
N
N
press
ANTENNA
,
4
OUTPUT
N
N
N
N
N
N
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
accessed
5
key
!
CARD
4
ENTER
using
,
turns
.
REGISTER
5
. The
the
on
antenna data is then saved to a card.
accesses a menu to turn on, recall, edit, or save antenna
3
amplitude-correction factors. Antenna factors are conversion
factors relating eld strength to measured voltage.
turns on or o antenna amplitude-correction factors.
3
accesses a menu of the transducer conversion units for the
3
antenna-correction factors. This species the units of the
physical attribute to which an antenna actually responds and
which will be indicated on the display for reference level,
display line level, threshold level, and marker readouts.
starts the DLP editor function and allows the highlighted item
3
from the catalog of analyzer memory to be appended to the
end of the item that is currently in the analyzer's DLP editor
memory.
2-4 EMC Analyzer Functions

Caution
Note
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
ATTEN
AUTO MAN
x3x3x3x3x3x3x3x3x3x3x3x3x3x
AUTO
ALL
3
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AUTO CAL
ON
OFF
4
AUTO COUPLE
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AUTO-
MEASURE
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AUTOEXEC
3
ON OFF
To prevent damage to the input mixer, the power level at the
input mixer must not exceed +30 dBm. To prevent signal
compression, power at the input to the input mixer must be
kept below010 dBm.
T
o
protect the mixer from possible damage, 0 dB RF attenuation
(no input power reduction to the mixer) can be selected only
from the number/units keypad.
sets the input attenuation in 10 dB increments. The instrument
input attenuator, which is normally coupled (linked) to the
3
reference
signal
level control, reduces the power level of the input
at the input mixer. The attenuator is recoupled when
AUTO is underlined. Toset0dB, 0 then dB must be entered on
the data entry keys.
couples the following functions:
intermediate frequency bandwidth
average video bandwidth
attenuation
sweep time
center-frequency
average-video-bandwidth
bandwidth
turns
the
automatic
ratio
step
.
calibration
to
intermediate-frequency-
feature
on
or
o.
3
5
accesses the softkey menu of functions that can be coupled.
(Coupled functions are functions that are linked: if one
function is changed, the other function is changed.)
nds all signals on the display that are above the margin and
that meet signal criteria, then
using specied detectors
makes an EMC measurement
.
EMC analyzer mode only.
enables and disables the automatic loading and execution of
the le named
3
instrument, upon a power-up sequence, will search the card for
an
"eAUTOEXEC"
"eAUTOEXEC"
from a card. When enabled, the
le and if found will load and execute it.
EMC Analyzer Functions 2-5

x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AUTORANG
ON OFF
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
Aux
Conn
Control
turns
the
current
When
on,
3
the
attenuation
state of the auto-range function on or o.
auto-range function automatically adjusts RF
or
reference
level (IF step-gain) in response to
either an IF or RF overload condition detected during the
previous sweep.
RF overload detection is only available when
used with an HP 85420E Option 1EM, RF lter section.
accesses the softkey menu used to control the auxiliary outputs
3
and input.
1EM RF lter section.
Not available when used with an HP 85420E Option
4
AUX/USER
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AV
TIME
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AV/IF
BW
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AVG BW
AUTO
x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x3x
AVG
ON OFF
DWELL
RATIO
MAN
3
5
3
3
3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
B0DL
!
B
NNNNNNNNNNNNNNNNNNN
B
!
C
accesses the softkey menu used for control of the auxiliary
interface connector and user menu. The softkey also accesses
the comb generator (
tracking generator functions (
sets
the
measurement
is measured. This key is used in conjunction with
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE AT MKR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
STEPPED
HP 8593EM, and HP 8596EM only
time
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
,
RE-measure
Option 010
when
the
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
,
AUTO-MEASURE
).
average
.
)and
detector
, and
determines the automatic setting of averaging (video)
bandwidth by multiplying the parameter by the intermediate
frequency (IF) bandwidth.
Ratio
values
other
than
the
values
in
the
1,
3,
10
sequence are
rounded to the nearest permissible value.
species the averaging bandwidth, which is a post-detection,
low-pass lter.
Frequency
are
rounded
A
VG
values
other than the values in the 1, 3, 10 sequence
to the nearest permissible value.Auto couples the
BW to the IF BW.
toggles the average detector on and o. When on, the
instrument is placed in linear amplitude detection, the
detector is turned on, and the system settings are optimized to
accurately measure the average amplitudes.
All except option
703.
subtracts the display line from trace B and places the result in
trace B
. The B
0DL!
B function is a math operation.
copies trace B into trace C.
2-6 EMC Analyzer Functions