The information contained in this document is subject to change
without notice.
Agilent Technologies makes no warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
The following safety symbols are used throughout this manual.
Familiarize yourself with the symbols and their meaning before
operating this instrument.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
NOTEAgilent Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Agilent further
certifies that its calibration measurements are traceable to the United
States National Bureau of Standards, to the extent allowed by the
Bureau’s calibration facility, and to the calibration facility of other
International Standards Organization members.
WARNINGBefore the charger of the 85901A is switched on, make sure it
has been properly grounded through the protective conductor
of the ac power cable to a socket outlet provided with the
protective earth contact. Any interruption of the protective
(grounding) conductor inside or outside the instrument, or
disconnection of the protective earth terminal, can result in
personal injury.
WARNINGThere are voltages at many points in the instrument which can,
if contacted, cause personal injury. Be extremely careful. Any
adjustments or service procedures that require operation of the
instrument with the protected covers removed should be
performed only by trained service personnel.
ii
WARNINGALWAYS assume that high voltage is present at the inverter
output and on the power cord adapters. NEVER rely on the
position of the inverter switch or charger switch or on the
front-panel LED indicators to determine if high voltage is
present. Never use a damaged power cord adapter. The inverter
output should be treated as cautiously as you would treat a line
voltage source.
WARNINGWhen the high voltage output of the inverter is exposed, for
example when the 85901A is used in servicing some other
instrument, take extreme safety precautions as you would for
any high voltage exposure. ALWAYS disconnect all inverter
power cords before servicing the load unit. NEVER rely on the
position of the inverter switch or charger switch or on the
front-panel LED indicators to determine if high voltage is
present.
CAUTIONBefore an instrument is connected to the 85901A, verify that the line
voltage selector of the instrument to be connected is set to the same
voltage as the 85901A inverter output voltage. Failure to set the ac
power input to the correct voltage could cause damage to the
instrument.
Regulatory Notices
Herstellerbescheinigung
Hiermit wird bescheinigt, daβ das Gerät/System 85901A in
Übereinstimmung mit den Bestimmungen von Postverfügung 1046/84
funkentstört ist.
Der Deutschen Bundespost wurde das Inverkehrbringen dieses
Gerätes/Systems angezeigt und die Berechtigung Überprüfung der
Serie auf Einhaltung der Bestimmungen eingeräumt.
Zusaatzinformation fur Meβ-und Testgeräte
Werden Meβ- und Testgeräte mit ungeschirmten Kabeln und/oder in
offenen Meβaufbauten verwendet, so ist vom Betreiber sicherzustellen,
daβ die Funk-Entstörbestimmungen unter Betriebsbedingungen an
seiner Grundstücksgrenze eingehalten werden.
iii
Manufacturer’s Declaration
This is to certify that the equipment 85901A is in accordance with the
Radio Interference Requirements of Directive FTZ 1046/84. The
German Bundespost was notified that this equipment was put into
circulation, the right to check the series for compliance with the
requirements was granted.
Additional Information for Test- and Measurement Equipment
If Test- and Measurement Equipment is operated with unscreened
cables and/or used for measurements on open set-ups, the user has to
assure that under operating conditions the Radio Interference Limits
are still met at the border of his premises.
Warranty
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of one year from date
of shipment. During the warranty period, Agilent Technologies will, at
its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies
shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware
designated by Agilent Technologies for use with an instrument will
execute its programming instructions when properly installed on that
instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or
error-free.
iv
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse, operation
outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT
BE LIABLE FOR ANY DIRECT, INDIRECT,SPECIAL,INCIDENTAL,
OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON
This manual contains information required to operate and service the
Agilent Technologies Model 85901A AC Power Source. The 85901A is
shown in Figure 1-1 with all of its externally supplied accessories.
Description
The 85901A is a portable ac power source. It can be operated on the
bench or as a portable power supply for field service. Other uses are for
measurements requiring power line isolation or for measurements
performed during power failure. The instrument has a carrying handle
that can also be used as an adjustable tilt-stand. The handle can be
moved by pushing the sides of the handle in and then turning the
handle to a suitable position.
The 85901A consists of battery, inverter, charger, and control circuits. It
provides up to 200W maximum continuous power with the internal
battery, at either 115 Vrms or 230 Vrms.
The 85901A weighs 14.2 kg (31.3 lbs) with the battery installed, and 8.0
kg (17.6 lbs) without the battery. The rear panel has terminals for
connecting an external battery or other 12 Vdc source. The dimensions
of the package are 337 mm wide by 125 mm high by 461 mm deep.
Specifications
Instrument specifications are listed in Table 1-1. These specifications
are the performance standards or limits against which the instrument
may be tested.
Performance characteristics are listed in Table 1-2. Characteristics
provide useful information by describing typical, non-warranted
instrument performance.
1-2Chapter 1
General Information
Specifications
Figure 1-185901A w/Complete Set of General Purpose Power-Cord
Adapters
Chapter 11-3
General Information
Safety Considerations
Safety Considerations
The 85901A AC Power Source has been designed to conform to the
safety requirements of an IEC (International Electromechanical
Committee) Safety Class I instrument, and is shipped from the factory
in a safe condition. This Agilent Technologies 85901A A C Power SourceOperation and Service Manual, and all related documentation, must be
reviewed for familiarization with safety markings and instructions
before operation. This unit is to be used in a controlled environment.
WARNINGThis unit contains more than one live circuit. See the schematic
diagrams.
This operation and service manual contains safety information,
cautions, and warnings which must be followed by the user to ensure
safe operation and to maintain the instrument in a safe condition.
Instruments Covered by This Manual
This instrument has a two-part serial number in the form of
00000K00000 (see Figure 1-2). The number is stamped on the serial
plate attached to the rear panel of the instrument. The first five digits
and the letter constitute the serial prefix; the last five digits form the
digits. The prefix is the same for all identical instruments, and it
changes only when a change is made to the instrument. The suffix,
however, is assigned sequentially and is different for each instrument.
The contents of this manual apply directly to instruments having the
same serial number prefix(es) listed under "Serial Numbers” on the
title page. The letter placed between the two sets of digits identifies the
country where the instrument was manufactured.
1-4Chapter 1
Figure 1-2Typical Serial Number Label
An instrument manufactured after the printing of this manual may
have a serial number prefix that is not listed on the title page. This
unlisted serial number prefix indicates that the instrument is different
from those described in the manual.
General Information
Instruments Covered by This Manual
The manual for a newer instrument is accompanied by a Manual
Updating Supplement, which contains the change information that
explains how to adapt the manual to the newer instrument.
In addition to change information, the supplement may contain
information for correcting errors in the manual. Tokeep this manual as
current and accurate as possible, Agilent recommends that you
periodically request Manual Updating Supplement. The supplements
are identified with the manual print date and part number, both of
which appear on the manual title page. For information concerning a
serial number prefix that is not listed on the title page or in the Manual
Updating Supplement, contact your nearest Agilent office.
Chapter 11-5
General Information
Instruments Covered by This Manual
Table 1-185901A Specifications
Specifications describe warranted performance over the temperature
range of 0°C to 55°C after one hour continuous operation, unless
otherwise noted.
INPUTS
Inverter
1
Voltage 10.8 Vdc to 14.5 Vdc
Charger
2
Voltage90 to 250 Vac, auto-selected
Frequency44 to 66 Hz
Power consumption122 VA, maximum
OUTPUTS
Inverter
Voltage135 or 270 V peak 5%
Rectangular waveform with 25%
dead zone (115 or 230 Vrms 5%)
Frequency60 Hz 0.1%, crystal reference
Maximum power200W, continuous
1.
The inverter input must not have any transients higher than 14.5 V.
Otherwise, the overvoltage protection circuitwill automatically turn
the inverter off.
2.
Turning the charger on automatically turns the inverter off.
1-6Chapter 1
Instruments Covered by This Manual
Table 1-2 85901A Characteristics (1 of 2)
ENVIRONMENTAL
General Information
Temperature
1
Operating0°C to +55°C
Storage:with battery–20°C to +40°C
TypeSealed lead acid
Voltage12Vdc
Capacity17 ampere-hours
MECHANICAL
Dimensions: width 337mm (13.3 in)
height125 mm (4.9 in)
depth461 mm 18.1 in)
Weight: with battery14.2 kg (31.3 lb)
without battery8.0 kg (17.7 lb)
Shipping weight16.3 kg (36.0 lb)
BATTERY SHELF LIFE
Storage TemperatureRecharge Interval
<20°C18 months
20°C to 30°C12 months
0°C to 40°C6 months
CERTIFICATION
The 85901A is certified for use with the following Agilent test and
measurement equipment:
HP/Agilent 8562A/B, Agilent 8590A, and Agilent 8592A, and their
options.
1.
Shelf life of battery decreases with either extended operation or storage at
higher than 40C
2.
Battery cyclic life (charge-discharge) is nominally rated at more
than 200 cycles.
3.
Battery is a Matsushita Electric LCR12V17P sealed lead acid. Battery is available from Agilent Technologies as HP/Agilent p/n
1420-0373.
Chapter 11-7
General Information
Instruments Covered by This Manual
Figure 1-385901A Characteristics (2 of 2)
Table 1-3Recommended Test Equipment
InstrumentCritical SpecificationsRecommended
Model
Power SupplyOutput voltage: 9 to 20 VdcHP/Agilent 6032A√√
Digital
Multimeter
Electronic LoadInput voltage: 20 Vdc
4.5 digits display
Input current: 10A minimum
1-8Chapter 1
HP/Agilent 3487A
Fujitsu
EUL-300D
Where Used
Perf.
Tests
√√√
√√
Adj.Ser-
vice
2Installation
2-1
Installation
Introduction
Introduction
This chapter provides the information needed to install the 85901A
AC PowerSource. Topics covered include:initial inspection, line voltage
and fuse selection, operating environment, battery charging, battery
replacement, storage, and shipment.
Initial Inspection
The 85901A AC Power Source, as shipped from the factory, meets all
the specifications listed in Table 1-1.
Inspect the shipping container for damage. If the shipping container or
cushioning material is damaged, notify the carrier as well as the
Agilent office and be sure to keep the shipping materials for the
carrier's inspection until the contents of the shipment have been
checked for completeness and the instrument has been checked
mechanically and electrically.
Complete shipment contents are shown in Figure 2-1. A test procedure
for checking general electrical operation may be found under "Basic
Operation Check" in Chapter 3. Performance tests to check the 85901A
against its specifications are given in Chapter 4. If the instrument does
not pass the performance tests, notify the nearest Agilent office (or, to
obtain the address of the Agilent office nearest you, contact an office
listed on the inside rear cover of this manual). The office will arrange
for repair or replacement without waiting for a claim settlement.
CAUTIONBefore using the 85901A for the first time, the internal battery should
be charged. Follow the procedure for charging the battery in Chapter 3.
If the shipping materials are in good condition, retain them for possible
future use. You may wish to ship the instrument to another location or
return it to Agilent for service.
2-2Chapter 2
Figure 2-185901A Shipping Container and Contents
Installation
Initial Inspection
Chapter 22-3
Installation
Preparation for Use
Preparation for Use
Power Requirements
The inverter of the 85901A requires a power source of 10.8 Vdc to 14.5
Vdc. Power consumption is approximately 330W when the 85901A
provides its maximum continuous output capacity of 200W.
WARNINGIf the instrument is to be energized via an external
autotransformer for voltage reduction, make sure the common
terminal is connected to the neutral pole of the charger input.
CAUTIONBefore turning the 85901A charger switch on, verify that the line
voltage is within the charger input's voltage range.
The charger circuit of the 85901A requires a power source of 90 to 250
Vac, 47 to 66 Hz, single-phase. You do not need to select the correct
input voltage; the charger circuit automatically does so. Maximum
power consumption in the charging mode is approximately 122 VA.
Inverter Line Voltage and Fuse Selection
CAUTIONBefore turning the 85901A inverter switch on, verify that the connected
instrument's line voltage selection is set to the same voltage as the
85901A inverter output voltage.
The 85901A voltage selection switch selects the output voltage of the
inverter (115V or 230V).
The appropriate fuses are factory-installed for the instrument's
destination. Figure 2-2 illustrates replacement fuses, and Table 2-1
lists their ratings and part numbers.
CAUTIONBefore using the 85901A, check that the proper fuse for your line
voltage and selected inverter-output voltage is installed.
2-4Chapter 2
Figure 2-2Line Voltage and Fuse Selection
Installation
Inverter Line Voltage and Fuse Selection
Table 2-1Replacement Fuse Ratings and Part Numbers
The internal battery for the 85901A is not installed when the
instrument is shipped from the factory. When unpacking the 85901A,
you will find a package containing two battery connectors in the main
container. You will also find two screws and two washers in the battery
container. These connectors and screws are used to install the internal
battery. You will need two screwdrivers, both slotted and Phillips.
WARNINGThis instrument is independent of the power mains. HIGH
VOLTAGE is present whenever the internal battery has been
installed and the inverter switch is on.
Connecting the battery may cause electrical arcing at the
terminal. Do not connect the battery in an explosive
atmosphere or near inflammable materials.
CAUTIONThe voltage ratings of the battery should be the same as the input
voltage range of the inverter. Verify that you are installing a battery of
the correct voltage rating.
Connect the internal battery's positive and negative terminals to the
85901A front-panel positive (red wire) and negative (black wire)
internal battery terminals, respectively.
1. Install the two separate battery connectors on the battery.
2. Open the battery door in the right-hand side of the front panel.
3. Place the battery in the 85901A battery cage with its terminals
toward the top. Correct battery position and connection polarity are
shown in Figure 2-3
4. Insert the two internal battery cable terminals into the positive and
negative battery connectors, respectively. Fasten them using the two
screws.
5. Refasten the battery cage door.
Confirm the correct placement of the wide four-terminal shorting strap
at the rear-panel terminal block. Refer to Figure 2-4 for correct
terminal block configurations with either internal or external batteries.
2-6Chapter 2
Installation
Connecting External Battery
Connecting External Battery
The 85901A AC Power Source can use an external battery by
connecting the external battery to the rear-panel terminal block. Car
and truck 12V batteries can be used for the external battery. Figure 2-4
shows the correct connections from the 85901A terminal block to the
external battery.
WARNINGThis instrument is independent of the power mains. HIGH
VOLTAGE is present whenever a 12 Vdc source has been
connected and the inverter switch is on.
Connecting the battery may cause electrical arcing at the
terminal. Do not connect the battery in an explosive
atmosphere or near inflammable materials.
NOTEWhen an external DC voltage source is used, voltage transients must
not exceed 14.5V or the Over Voltage Protection circuitry will shut the
inverter off.
Likewise, if an external charger is connected to an external battery,
voltage transients above 14.5V may turn the 85901A off. This may
happen, for example, when a car battery is used while the car's
alternator is operating.
Chapter 22-7
Installation
Connecting External Battery
Figure 2-3Battery Position and Connector Polarity
2-8Chapter 2
Figure 2-4Battery Terminal Block
Installation
Connecting External Battery
CAUTIONThe voltage ratings of the battery should be the same as the input
voltage range of the inverter. Verify that you are installing a battery of
the correct voltage rating
Connect the external battery's positive and negative terminals to the
85901A rear-panel positive and negative external-battery terminals,
respectively.
Do not remove the wide two-terminal shorting bar on rear-panel
positive and negative external- battery terminals.
Chapter 22-9
Installation
Power Cable for Charger Input
Power Cable for Charger Input
The 85901A is equipped with a three-connector power cable.When
connected to an appropriate ac power receptacle, this cable grounds the
instrument cabinet.
WARNINGThe mains plug may only be inserted in a socket outlet
provided with a protective earth contact. The protective action
must not be negated by the use of an extension cord (power
cable) without a protective conductor (grounding). Grounding
one conductor of a two-conductor outlet is not sufficient
protection. Various power cables are available to connect the
instrument to the types of ac power outlets unique to specific
geographic areas. The cable appropriate for the area to which
the analyzer is originally shipped is included with the unit.
NOTEWhile the inverter output is used, the 85901A will meet its EMI
specification with the charger input power cable removed.
Power Cable for Inverter Output
The 85901A standard connector is two CEE22-V connectors in place of
the USA standard NEMA-15P connector. The interconnection cable,
Part Number 8120-5220, CEE22-V male to NEMA-15P female, is
provided. This cable is used to connect external instruments to the
inverter output.
Other general-purpose power-cord adapters are available.The adapters
listed below allow country-specific line cords to be plugged into the
85901A AC Power Source.
EuropeHP/Agilent Part Number 8120-5210
U.S.A.HP/Agilent Part Number 8120-5211
United KingdomHP/Agilent Part Number 8120-5212
2-10Chapter 2
Installation
Operating Environment
Operating Environment
Temperature
The 85901A may be operated over temperatures from O°C to +55°C.
Temperatures above +400C may reduce the life of the battery.
Humidity
The 85901A may be operated in environments with relative humidities
up to 95% and to +40°C. However, the 85901A should be protected from
temperature extremes, which cause condensation within the
instrument.
Installation
The 85901A can be operated on the bench or as a portable power supply
for field use. The instrument has a carrying handle that can also be
used as an adjustable tilt-stand. The handle can be moved by pushing
its sides in and then turning the handle to the desired position.
The battery must be installed before the 85901A can be operated. See
the battery-installation instructions, above. Before turning on the
85901A inverter switch, verify that the same line voltage is selected for
both the 85901A inverter output and the equipment connected to the
inverter output.
Chapter 22-11
Installation
Storage and Shipment
Storage and Shipment
Environmental
The 85901A with battery may be stored and shipped in environments
within the following limits:
Temperature–20°C to +40°C
Humidity<95%
Altitude<50,000 ft
The 85901A should be protected from temperature extremes, which
cause condensation inside the equipment.
Packaging
Original packaging is shown in Figure 2-1. Containers and materials
identical to those used in factory packaging are available through
Agilent offices. If the instrument is being returned to Agilent for
servicing, attach a tag indicating the type of service required, return
address, model number, and full serial number. Mark the container
FRAGILE to assure careful handling. In any correspondence, refer to
the instrument by model number and full serial number.
Other packaging may be used if these general instructions for
repackaging with commercially-available materials are followed:
a. Wrap instrument in heavy paper or plastic. If it is being shipped to
an Agilent office or service center, attach tag indicating the type of
service required, return address, model number, and full serial
number.
b. Use a strong shipping container. A double-walled carton made of
350-pound test material is adequate.
c. Use enough shock-absorbing material (3- to 4-inch layer) around all
sides of instrument to provide firm cushioning and prevent
movement inside the container.
d. Seal shipping container securely.
e. Mark shipping container FRAGILE to ensure careful handling.
2-12Chapter 2
3Operation
3-1
Operation
Introduction
Introduction
This chapter provides operating instructions for the 85901A AC Power
Source. It also provides a brief description of the front- and rear-panel
features.
Upon receipt of the 85901A AC Power Source, perform the procedure
described under "Basic Operation Check" in this chapter. This check
should also be performed after the instrument has been repaired. LED
indicators, which represent the status of the battery, inverter, and
charger circuits, are located on the left-hand side of the front panel. See
Figure 3-1.
Front-Panel Features
LED indicators, which represent the status of the battery, inverter, and
charger circuits, are located on the left-hand side of the front panel. See
Figure 3-1.
The battery function area has two LEDs: the yellow LED indicates
whether the charger is on or off, and the red LED indicates the state of
the battery charger circuit. When the charger is operating, the red LED
will be on if the battery-charging current is greater than 0.45 A 0.15 A,
and it will be off if the battery-charging current is less than
0.45 A 0.15 A.
The inverter function area includes a yellow LED and the inverter
on/off switch. The LED should be on when the inverter switch is on and
the inverter is operating, and it should be off when the inverter is not
operating. The inverter will not operate while the battery charger is
operating.
During inverter operation, the red LED in the battery function area
also is used to indicate the stateofthe battery output voltage. When the
inverter is operating, the red LED will be off when battery voltage is
greater than 10.8 V typical, and it will blink (accompanied by a buzzer)
when battery voltage is less than 10.8 V typical.
The battery door occupies the remaining part of the front panel. This
door provides access for battery service.
3-2Chapter3
Loading...
+ 77 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.