Agilent Technologies
PNA Series Network Analyzers
E8362A, E8363A, and E8364A
.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT
DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO
THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT
NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR
ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE
MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial license terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users
will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Corrected System Performance.............................................................................................4-5
Table 1. System Dynamic Rangea.................................................................................................................. 4-5
Table 2. Receiver Dynamic Range
E8363/4A Corrected System Performance with 2.4mm Connectors............................................. 4-8
Table 3. 85056A Calibration Kit Standard Configurationand StandardPower Range (E8363/4A)... 4-8
Table 4. 85056A Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees
(E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
Table 5. 85056D Calibration Kit Standard Configurationand StandardPower Range (E8363/4A). 4-10
Table 6. 85056D Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees
(E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
E8363/4A Corrected System Performance with 2.92mm Connectors............................................... 4-12
Table 7. 85056K Calibration Kit Standard Configurationand Standard Power Range (E8363/4A). 4-12
Table 8. 85056K Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees
(E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
E836xA Corrected System Performance with 3.5mm Connectors.............................................. 4-14
Table 9. 85052B Calibration Kit Standard Configurationand StandardPower Range (E836xA).... 4-14
Table 10. 85052B Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 11. 85052C Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-16
Table 12. 85052C Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 13. 85052D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-18
Table 14. 85052D Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
E836xA Corrected System Performance with 7mm Connectors................................................. 4-20
Table 15. 85050B Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-20
Table 16. 85050B Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 17. 85050C Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-22
Table 18. 85050C Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 19. 85050D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-24
Table 20. 85050D Calibration Kit ExtendedConfigurationand Standard Power Range
E836xA Corrected System Performance with Type-N Connectors............................................ 4-26
Table 21. 85054B Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-26
Table 22. 85054B Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 23. 85054D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-28
Table 24. 85054D Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 26. R11644A Calibration Kit Extended Configuration and StandardPower Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees
(E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
Table 28. K11644A Calibration Kit ExtendedConfigurationand Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees
(E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
E836xA Corrected System Performance with WR-62 Connectors............................................. 4-34
Table 29. P11644A Calibration Kit Standard Configurationand Standard Power Range (E836xA)4-34
Table 30. P11644A Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees
(E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
Table 35: Test Port Input............................................................................................................................. 4-42
a
Table 36. Dynamic Accuracy (Specification
Table 37. Test Port Input(GroupDelay)
Test Set Block Diagrams.......................................................................................................4-60
E836xA Standard Configuration and Standard Power Range................................................... 4-60
E836xA - Option UNL Standard Configuration with Extended Power Range and Bias - Tees4-60
Test Set with Option 014 Block Diagrams..........................................................................4-61
E836xA - Option 014 Extended Configuration and Standard Power Range............................ 4-61
E836xA - Option UNL&014 Extended Configuration with Extended Power Range
and Bias - Tees............................................................................................................................... .. 4-62
4-3
This is a complete list of the E8362A, E8363A, and E8364A network analyzer technical specifications.
• To optimize viewing of uncertainty curves, click the Maximize button.
• To view or print the PNA Series Data Sheet (a condensed version of the specifications), visit our web
site at http://www.agilent.com/find/pna
sheet.
• The uncertainty curves contained in this document apply only to the setup conditions listed. Please
download our free Uncertainty Calculator from http://www.agilent.com/find/na_calculator
curves for your PNA setup. View the equations
See Specs for other PNA models
, select your analyzer model, and click on the link for the data
to generate the
used to generate the uncertainty curves.
Definitions
All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes
after the instrument has been turned on.
Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected
statistical performance distribution, measurement uncertainties, and changes in performance due to
environmental conditions.
Characteristic (char.): A performance parameter that the product is expected to meet before it leaves the
factory, but that is not verified in the field and is not covered by the product warranty. A characteristic includes
the same guardbands as a specification.
Typical (typ.): Expected performance of an average unit which does not include guardbands. It is not covered
by the product warranty.
Nominal (nom.): A general, descriptive term that does not imply a level of performance. It is not covered by the
product warranty.
Calibration: The process of measuring known standards to characterize a network analyzer's systematic
(repeatable) errors.
Corrected (residual): Indicates performance after error correction (calibration). It is determined by the quality of
calibration standards and how well "known" they are, plus system repeatability, stability, and noise.
Uncorrected (raw): Indicates instrument performance without error correction. The uncorrected performance
affects the stability of a calibration.
Standard: When referring to the analyzer, this includes no options unless noted otherwise.
4-4
Corrected System Performance
The specifications in this section apply for measurements made with the E836xA analyzer with the following
conditions:
• 10 Hz IF bandwidth
• No averaging applied to data
• Isolation calibration with an averaging factor of 8
Table 1. System Dynamic Range
DescriptionSpecification
a
(dB) at Test Port
Typical (dB) at Direct
b
Receiver Access Input
Dynamic Range (in a 10 Hz BW)
Standard Configuration and Standard Power Range
(E836xA - Standard)
45 MHz to 500 MHz
d
94NA
500 MHz to 2 GHz119NA
2 GHz to 10 GHz122NA
10 GHz to 20 GHz123NA
20 GHz to 30 GHz114NA
30 GHz to 40 GHz110NA
40 GHz to 45 GHz109NA
45 GHz to 50 GHz104NA
Extended Configuration and Standard Power Range
(E836xA - Option 014)
45 MHz to 500 MHz
d
94132
500 MHz to 2 GHz119138
2 GHz to 10 GHz122137
10 GHz to 20 GHz122137
20 GHz to 30 GHz115127
30 GHz to 40 GHz107119
40 GHz to 45 GHz105116
45 GHz to 50 GHz100111
Standard Configuration and Extended Power Range & Bias-Tees
(E836xA - Option UNL)
45 MHz to 500 MHz
d
92NA
500 MHz to 2 GHz117NA
2 GHz to 10 GHz120NA
10 GHz to 20 GHz121NA
20 GHz to 30 GHz112NA
30 GHz to 40 GHz108NA
40 GHz to 45 GHz105NA
45GHzto50GHz99NA
c
4-5
Standard Configuration and Extended Power Range & Bias-Tees
(E836xA - Option UNL&014)
45 MHz to 500 MHz
d
92130
500 MHz to 2 GHz117136
2 GHz to 10 GHz120135
10 GHz to 20 GHz119134
20 GHz to 30 GHz109121
30 GHz to 40 GHz105117
40 GHz to 45 GHz101112
45 GHz to 50 GHz95108
a
The system dynamic range is calculated as the difference between the noise floor and the source maximum output power. The effective
dynamic range must take measurement uncertainties and interfering signals into account.
b
The test port system dynamic range is calculated as the difference between the test port noise floor and the source maximum output
power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
c
The direct receiver access input system dynamic range is calculated as the difference between the receiver access input noise floor and
the source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
This set-up should only be used when the receiver input will never exceed its damage level. When the analyzer is in segment sweep mode,
the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic range is required
(i.e. devices with high insertion loss), and reduced power when receiver damage may occur (i.e. devices with low insertion loss). The
extended range is only available in one-path transmission measurements.
d
May be degraded by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due to spurious receiver residuals. Methods are
available to regain the full dynamic range.
4-6
Table 2. Receiver Dynamic Range
DescriptionSpecification (dB)
a
at Test Port
b
Typical (dB) at Direct Receiver
Access Input
c
Dynamic Range (in a 10 Hz BW)
Standard Configuration and Standard Power Range (E836xA - Standard)
OR
Standard Configuration and Extended Power Range & Bias Tees (E836xA - Option
UNL)
45 MHz to 500 MHz
d
94NA
500 MHz to 2 GHz119NA
2 GHz to 10 GHz122NA
10 GHz to 20 GHz125NA
20 GHz to 30 GHz114NA
30 GHz to 40 GHz111NA
40 GHz to 50 GHz111NA
Extended Configuration and Standard Power Range (E836xA - Option 014)
OR
Extended Configuration and Extended Power Range & Bias Tees (E836xA - Option
UNL&014)
45 MHz to 500 MHz
d
94132
500 MHz to 2 GHz119138
2 GHz to 10 GHz122137
10 GHz to 20 GHz124139
20 GHz to 30 GHz113125
30 GHz to 40 GHz110122
40 GHz to 50 GHz109120
a
The receiver dynamic range is calculated as the difference between the noise floor and the receiver maximum output power. The effective
dynamic range must take measurement uncertainties and interfering signals into account.
b
The test port receiver dynamic range is calculated as the difference between the test port noise floor and the receiver maximum input
level. The effective dynamic range must take measurement uncertainties and interfering signals into account.
c
The direct receiver access input receiver dynamic range is calculated as the difference between the direct receiver access input noise floor
and the receiver maximum input level. The effective dynamic range must take measurement uncertainties and interfering signals into
account. This set-up should only be used when the receiver input will never exceed its damage level. When the analyzer is in segment
sweep mode, the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic
range is required (i.e. devices with high insertion loss), and reduced power when receiver damage may occur (i.e. devices with low insertion
loss). The extended range is only available in one-path transmission measurements.
d
May be degraded by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due to spurious receiver residuals. Methods are
available to regain the full dynamic range.
Note: This E836xA document provides technical specifications for the following calibration kits only: 85056A,
85056D, 85056K, 85052B, 85052C, 85052D, 85050B, 85050C, 85050D, 85054B, 85054D, K11644A, P11644A,
R11644A, and the X11644A.
4-7
E8363/4A Corrected System Performance with 2.4mm Connectors
Table 3. 85056A Calibration Kit
Standard Configuration and Standard Power Range
(E8363/4A)
Applies to the E8363/4A analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-ORStandard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-ORExtended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL&014)
Applies to the, E8363/4A analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Standard Configuration and Standard Power Range
(E8363/4A)
Applies to the, E8363/4A analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-ORStandard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-ORExtended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL & 014)
Applies to the, E8363/4A analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
E8363/4A Corrected System Performance with 2.92mm Connectors
Table 7. 85056K Calibration Kit
Standard Configuration and Standard Power Range
(E8363/4A)
Applies to the, E8363/4A analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-ORStandard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-ORExtended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL&014)
Applies to the, E8363/4A analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
E836xA Corrected System Performance with 3.5mm Connectors
Table 9. 85052B Calibration Kit
Standard Configuration and Standard Power Range
(E836xA)
Applies to the, E836xA analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Extended Configuration and Standard Power Range (E836xA - Option 014)
-ORStandard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL)
-ORExtended Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL&014)
Applies to the, E836xA analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Standard Configuration and Standard Power Range
(E836xA)
Applies to the, E836xA analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Extended Configuration and Standard Power Range (E836xA - Option 014)
-ORStandard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL)
-ORExtended Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL&014)
Applies to the, E836xA analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature