Keysight (Agilent) 85050B Technical Specifications

Technical Specifications
Discontinued Product Information
— For Support Reference Only —
www.agilent.com/find/products
In the US, call Agilent Technologies at 1-800-829-4444
(any weekday between 8am–5pm in any U.S. time zone)
World-wide Agilent sales office contact information is available at:
www.agilent.com/find/contactus
Agilent Technologies
PNA Series Network Analyzers
E8362A, E8363A, and E8364A
.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FUR­THER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FIT­NESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WAR­RANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii
Contacting Agilent
Online assistance: www.agilent.com/find/assist
Americas
Brazil
(tel) (+55) 11 4197 3600 (fax) (+55) 11 4197 3800
Australia
(tel) 1800 629 485 (alt) 1800 143 243 (fax) 1800 142 134
Japan
(tel) 0120 421 345 (alt) (+81) 426 56 7832 (fax) 0120 421 678
Taiwan
(tel) 0800 047 866 (alt) 00801 651 317 (fax) 0800 286 331
Austria
(tel) 0820 87 44 11* (fax) 0820 87 44 22
Canada
(tel) 877 894 4414 (fax) (+1) 905 282-6495
Asia Pacific and Japan
China
(tel) 800 810 0189 (alt) (+86) 10800 650 00 21 (fax) 800 820 2816
Malaysia
(tel) 1800 888 848 (alt) 1800 828 848 (fax) 1800 801 664
Thailand
(tel) 1800 226 008 (alt) (+66) 2 268 1345 (fax) (+66) 2 661 37 14
Belgium
(tel) (+32) (0)2 404 9340 (alt) (+32) (0)2 404 9000 (fax) (+32) (0)2 404 9395
Mexico
(tel) (+52) 55 5081 9469 (alt) 01800 5064 800 (fax) (+52) 55 5081 9467
Hong Kong (tel) 800 930 871 (alt) (+852) 3197 7889 (fax) (+852) 2 506 9233
Singapore
(tel) 1800 375 8100 (alt) (+65) 6 375 8100 (fax) (+65) 683 6 0252
Europe
Denmark
(tel) (+45) 7013 1515 (alt) (+45) 7013 7313 (fax) (+45) 701 3 1555
United States
(tel) 800 829 4444 (alt) (+1) 303 662 3998 (fax) 800 829 4433
India
(tel) 1600 11 2 92 9 (fax) 000800 650 1101
South Korea
(tel) 080 769 0800 (alt) (+82) 2 2004 5004 (fax) (+82) 2 2004 5115
Finland
(tel) (+358) 10 855 2100 (fax) (+358) 10 855 2923
France
(tel) 0825 010 700* (alt) (+33) (0)1 6453 5623 (fax) 0825 010 701*
Italy
(tel) (+39) (0)2 9260 8484 (fax) (+39) (0)2 9544 1175
Spain
(tel) (+34) 91 631 3300 (alt) (+34) 91 631 3000 (fax) (+34) 91 631 3301
Switzerland (Italian)
(tel) 0800 80 5353 opt. 3* (alt) (+39) (0)2 9260 8484 (fax) (+41) (0)22 567 5314
(tel) = primary telephone number; (alt) = alternate telephone nu mber; (fax) = FAX number; * = in country number
Germany
(tel) 01805 24 6333* (alt) 01805 24 6330* (fax) 01805 24 6336*
Luxemburg
(tel) (+32) (0)2 404 9340 (alt) (+32) (0)2 404 9000 (fax) (+32) (0)2 404 9395
Sweden
(tel) 0200 88 22 55* (alt) (+46) (0)8 5064 8686 (fax) 020 120 2266*
United King dom
(tel) (+44) (0)7004 666666 (alt) (+44) (0)7004 123123 (fax) (+44) (0)7004 444555
Ireland
(tel) (+353) (0)1 890 924 204 (alt) (+353) (0)1 890 924 206 (fax)(+353) (0)1 890 924 024
Netherlands
(tel) (+31) (0)20 547 2111 (alt) (+31) (0)20 547 2000 (fax) (+31) (0)20 547 2190
Switzerland (French)
(tel) 0800 80 5353 opt. 2* (alt) (+33) (0)1 6453 5623 (fax) (+41) (0)22 567 5313
Israel
(tel) (+972) 3 9288 500 (fax) (+972) 3 9288 501
Russia
(tel) (+7) 095 797 3963 (alt) (+7) 095 797 3900 (fax) (+7) 095 797 3901
Switzerland (German )
(tel) 0800 80 5353 opt. 1* (alt) (+49) (0)7031 464 63 33 (fax) (+41) (0)1 272 7373
iii
This page intentionally left blank.
iv
TechnicalSpecificationsfortheE836xA

Definitions................................................................................................................................4-4
Corrected System Performance.............................................................................................4-5
Table 1. System Dynamic Rangea.................................................................................................................. 4-5
Table 2. Receiver Dynamic Range
E8363/4A Corrected System Performance with 2.4mm Connectors............................................. 4-8
Table 3. 85056A Calibration Kit Standard Configurationand StandardPower Range (E8363/4A)... 4-8 Table 4. 85056A Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees (E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
(E8363/4A - Option UNL&014)..................................................................................................................... 4-9
Table 5. 85056D Calibration Kit Standard Configurationand StandardPower Range (E8363/4A). 4-10 Table 6. 85056D Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees (E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
(E8363/4A - Option UNL & 014) ................................................................................................................. 4-11
E8363/4A Corrected System Performance with 2.92mm Connectors............................................... 4-12
Table 7. 85056K Calibration Kit Standard Configurationand Standard Power Range (E8363/4A). 4-12 Table 8. 85056K Calibration Kit Extended Configuration and Standard Power Range
(E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees (E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
(E8363/4A - Option UNL&014)................................................................................................................... 4-13
E836xA Corrected System Performance with 3.5mm Connectors.............................................. 4-14
Table 9. 85052B Calibration Kit Standard Configurationand StandardPower Range (E836xA).... 4-14
Table 10. 85052B Calibration Kit Extended Configuration and Standard Power Range (E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-15
Table 11. 85052C Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-16 Table 12. 85052C Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-17
Table 13. 85052D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-18 Table 14. 85052D Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-19
E836xA Corrected System Performance with 7mm Connectors................................................. 4-20
Table 15. 85050B Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-20 Table 16. 85050B Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-21
Table 17. 85050C Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-22 Table 18. 85050C Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-23
Table 19. 85050D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-24 Table 20. 85050D Calibration Kit ExtendedConfigurationand Standard Power Range
a
............................................................................................................... 4-7
4-1
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-25
E836xA Corrected System Performance with Type-N Connectors............................................ 4-26
Table 21. 85054B Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-26 Table 22. 85054B Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-27
Table 23. 85054D Calibration Kit Standard Configurationand StandardPower Range (E836xA).. 4-28 Table 24. 85054D Calibration Kit ExtendedConfigurationand Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-29
E8363/4A Corrected System Performance with WR-28 Connectors.......................................... 4-30
Table 25. R11644A Calibration Kit StandardConfiguration and StandardPower Range
(E8363/4A)..................................................................................................................................................... 4-30
Table 26. R11644A Calibration Kit Extended Configuration and StandardPower Range (E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees (E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
(E8363/4A - Option UNL&014)................................................................................................................... 4-31
E8363/4A Corrected System Performance with WR-42 Connectors.......................................... 4-32
Table 27. K11644A Calibration Kit StandardConfiguration and StandardPower Range
(E8363/4A)..................................................................................................................................................... 4-32
Table 28. K11644A Calibration Kit ExtendedConfigurationand Standard Power Range (E8363/4A - Option 014) -OR- Standard ConfigurationandExtendedPower Range & Bias-Tees (E8363/4A - Option UNL) -OR- Extended Configuration and Extended Power Range & Bias-Tees
(E8363/4A - Option UNL&014)................................................................................................................... 4-33
E836xA Corrected System Performance with WR-62 Connectors............................................. 4-34
Table 29. P11644A Calibration Kit Standard Configurationand Standard Power Range (E836xA)4-34 Table 30. P11644A Calibration Kit Extended Configuration and Standard Power Range
(E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-35
E836xA Corrected System Performance with WR-90 Connectors............................................. 4-36
Table 31. X11644A Calibration Kit Standard Configuration and Standard Power Range
(E836xA)........................................................................................................................................................ 4-36
Table 32. X11644A Calibration Kit Extended Configuration and StandardPower Range (E836xA - Option 014) -OR- StandardConfiguration and ExtendedPower Range & Bias-Tees (E836xA - Option UNL) -OR- Extended Configuration and ExtendedPowerRange& Bias-Tees
(E836xA - Option UNL&014)...................................................................................................................... 4-37
Table 33. Uncorrected System Performance .............................................................................................. 4-38
Table 34. Test Port Output
a
......................................................................................................................... 4-40
Table 35: Test Port Input............................................................................................................................. 4-42
a
Table 36. Dynamic Accuracy (Specification Table 37. Test Port Input(GroupDelay)
) ............................................................................................ 4-45
a
.................................................................................................. 4-46
General Information.............................................................................................................4-46
Table 38. Miscellaneous Information..........................................................................................................4-46
Table 39. Front Panel Information.............................................................................................................. 4-47
Table 40. Rear Panel Information............................................................................................................... 4-47
Table 41. Analyzer Environment and Dimensions..................................................................................... 4-49
Measurement Throughput Summary..................................................................................4-50
4-2
Table 42. Typical Cycle Time Table43.CycleTimevsIFBandwidth Table44.CycleTimevsNumberofPoints Table 45. Data Transfer Time (ms)
a,b
(ms) for Measurement Completion.......................................................4-50
a
...................................................................................................... 4-51
a
a
............................................................................................... 4-51
............................................................................................................ 4-52
Specifications: Front-Panel Jumpers...................................................................................4-53
Table 46: Measurement Receiver Inputs (Rcvr A In, Rcvr B In)............................................................. 4-53
Table 47: Reference Receiver Inputs (Rcvr R1, Rcvr R2).............................................................. 4-55
Table 48: Reference Outputs (Reference 1 Source Out, Reference 2 Source Out)..................................4-56
Table 49: Source Outputs (Port 1 Source Out, Port 2 Source Out).......................................................... 4-57
Table 50: Coupler Inputs (Port 1 Cplr Thru, Port 2 Cplr Thru) ............................................................. 4-58
Table 51: Coupler Outputs (Port 1 Cplr Arm, Port 2 Cplr Arm)............................................................4-59
Test Set Block Diagrams.......................................................................................................4-60
E836xA Standard Configuration and Standard Power Range................................................... 4-60
E836xA - Option UNL Standard Configuration with Extended Power Range and Bias - Tees4-60
Test Set with Option 014 Block Diagrams..........................................................................4-61
E836xA - Option 014 Extended Configuration and Standard Power Range............................ 4-61
E836xA - Option UNL&014 Extended Configuration with Extended Power Range
and Bias - Tees............................................................................................................................... .. 4-62
4-3
This is a complete list of the E8362A, E8363A, and E8364A network analyzer technical specifications.
To optimize viewing of uncertainty curves, click the Maximize button.
To view or print the PNA Series Data Sheet (a condensed version of the specifications), visit our web
site at http://www.agilent.com/find/pna sheet.
The uncertainty curves contained in this document apply only to the setup conditions listed. Please download our free Uncertainty Calculator from http://www.agilent.com/find/na_calculator curves for your PNA setup. View the equations
See Specs for other PNA models
, select your analyzer model, and click on the link for the data
to generate the
used to generate the uncertainty curves.
Definitions
All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument has been turned on.
Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Characteristic (char.): A performance parameter that the product is expected to meet before it leaves the factory, but that is not verified in the field and is not covered by the product warranty. A characteristic includes the same guardbands as a specification.
Typical (typ.): Expected performance of an average unit which does not include guardbands. It is not covered by the product warranty.
Nominal (nom.): A general, descriptive term that does not imply a level of performance. It is not covered by the product warranty.
Calibration: The process of measuring known standards to characterize a network analyzer's systematic (repeatable) errors.
Corrected (residual): Indicates performance after error correction (calibration). It is determined by the quality of calibration standards and how well "known" they are, plus system repeatability, stability, and noise.
Uncorrected (raw): Indicates instrument performance without error correction. The uncorrected performance affects the stability of a calibration.
Standard: When referring to the analyzer, this includes no options unless noted otherwise.
4-4
Corrected System Performance
The specifications in this section apply for measurements made with the E836xA analyzer with the following conditions:
10 Hz IF bandwidth
No averaging applied to data
Isolation calibration with an averaging factor of 8
Table 1. System Dynamic Range
Description Specification
a
(dB) at Test Port
Typical (dB) at Direct
b
Receiver Access Input
Dynamic Range (in a 10 Hz BW) Standard Configuration and Standard Power Range (E836xA - Standard)
45 MHz to 500 MHz
d
94 NA 500 MHz to 2 GHz 119 NA 2 GHz to 10 GHz 122 NA 10 GHz to 20 GHz 123 NA 20 GHz to 30 GHz 114 NA 30 GHz to 40 GHz 110 NA 40 GHz to 45 GHz 109 NA 45 GHz to 50 GHz 104 NA
Extended Configuration and Standard Power Range (E836xA - Option 014)
45 MHz to 500 MHz
d
94 132 500 MHz to 2 GHz 119 138 2 GHz to 10 GHz 122 137 10 GHz to 20 GHz 122 137 20 GHz to 30 GHz 115 127 30 GHz to 40 GHz 107 119 40 GHz to 45 GHz 105 116 45 GHz to 50 GHz 100 111
Standard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL)
45 MHz to 500 MHz
d
92 NA 500 MHz to 2 GHz 117 NA 2 GHz to 10 GHz 120 NA 10 GHz to 20 GHz 121 NA 20 GHz to 30 GHz 112 NA 30 GHz to 40 GHz 108 NA 40 GHz to 45 GHz 105 NA 45GHzto50GHz 99 NA
c
4-5
Standard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL&014)
45 MHz to 500 MHz
d
92 130 500 MHz to 2 GHz 117 136 2 GHz to 10 GHz 120 135 10 GHz to 20 GHz 119 134 20 GHz to 30 GHz 109 121 30 GHz to 40 GHz 105 117 40 GHz to 45 GHz 101 112 45 GHz to 50 GHz 95 108
a
The system dynamic range is calculated as the difference between the noise floor and the source maximum output power. The effective
dynamic range must take measurement uncertainties and interfering signals into account.
b
The test port system dynamic range is calculated as the difference between the test port noise floor and the source maximum output
power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
c
The direct receiver access input system dynamic range is calculated as the difference between the receiver access input noise floor and the source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account. This set-up should only be used when the receiver input will never exceed its damage level. When the analyzer is in segment sweep mode, the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic range is required (i.e. devices with high insertion loss), and reduced power when receiver damage may occur (i.e. devices with low insertion loss). The extended range is only available in one-path transmission measurements.
d
May be degraded by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due to spurious receiver residuals. Methods are available to regain the full dynamic range.
4-6
Table 2. Receiver Dynamic Range
Description Specification (dB)
a
at Test Port
b
Typical (dB) at Direct Receiver Access Input
c
Dynamic Range (in a 10 Hz BW) Standard Configuration and Standard Power Range (E836xA - Standard)
OR Standard Configuration and Extended Power Range & Bias Tees (E836xA - Option
UNL)
45 MHz to 500 MHz
d
94 NA 500 MHz to 2 GHz 119 NA 2 GHz to 10 GHz 122 NA 10 GHz to 20 GHz 125 NA 20 GHz to 30 GHz 114 NA 30 GHz to 40 GHz 111 NA 40 GHz to 50 GHz 111 NA
Extended Configuration and Standard Power Range (E836xA - Option 014) OR
Extended Configuration and Extended Power Range & Bias Tees (E836xA - Option UNL&014)
45 MHz to 500 MHz
d
94 132 500 MHz to 2 GHz 119 138 2 GHz to 10 GHz 122 137 10 GHz to 20 GHz 124 139 20 GHz to 30 GHz 113 125 30 GHz to 40 GHz 110 122 40 GHz to 50 GHz 109 120
a
The receiver dynamic range is calculated as the difference between the noise floor and the receiver maximum output power. The effective
dynamic range must take measurement uncertainties and interfering signals into account.
b
The test port receiver dynamic range is calculated as the difference between the test port noise floor and the receiver maximum input
level. The effective dynamic range must take measurement uncertainties and interfering signals into account.
c
The direct receiver access input receiver dynamic range is calculated as the difference between the direct receiver access input noise floor and the receiver maximum input level. The effective dynamic range must take measurement uncertainties and interfering signals into account. This set-up should only be used when the receiver input will never exceed its damage level. When the analyzer is in segment sweep mode, the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic range is required (i.e. devices with high insertion loss), and reduced power when receiver damage may occur (i.e. devices with low insertion loss). The extended range is only available in one-path transmission measurements.
d
May be degraded by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due to spurious receiver residuals. Methods are available to regain the full dynamic range.
Note: This E836xA document provides technical specifications for the following calibration kits only: 85056A, 85056D, 85056K, 85052B, 85052C, 85052D, 85050B, 85050C, 85050D, 85054B, 85054D, K11644A, P11644A, R11644A, and the X11644A.
4-7
E8363/4A Corrected System Performance with 2.4mm Connectors

Table 3. 85056A Calibration Kit
Standard Configuration and Standard Power Range (E8363/4A)
Applies to the E8363/4A analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity 42 42 42 38 36 Source Match 41 38 38 33 31 Load Match 42 42 42 37 35 Reflection Tracking
Transmission Tracking
±0.001 +0.02/°C ±0.014 +0.02/°C
±0.008 +0.02/°C ±0.033 +0.02/°C
±0.008 +0.02/°C ±0.039 +0.02/°C
±0.020 +0.03/°C ±0.105 +0.03/°C
±0.027 +0.04/°C ±0.200 +0.04/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-8
Table 4. 85056A Calibration Kit
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-OR­Standard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-OR­Extended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL&014)
Applies to the, E8363/4A analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity 42 42 42 38 36 Source Match 41 38 38 33 31 Load Match 42 42 42 37 35 Reflection Tracking ±0.001
+0.02/°C
Transmission Tracking
±0.019 +0.02/°C
±0.008
+0.02/°C ±0.039 +0.02/°C
±0.008
+0.02/°C ±0.053 +0.02/°C
±0.020
+0.03/°C ±0.114 +0.03/°C
±0.027
+0.04/°C ±0.215 +0.04/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-9
Table 5. 85056D Calibration Kit
Standard Configuration and Standard Power Range (E8363/4A)
Applies to the, E8363/4A analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity 42 34 34 26 Source Match 40 30 30 23 Load Match 42 34 34 25 Reflection Tracking
Transmission Tracking
±0.002 +0.02/°C ±0.016 +0.02/°C
±0.029 +0.02/°C ±0.081 +0.02/°C
±0.029 +0.03/°C ±0.095 +0.03/°C
±0.075 +0.04/°C ±0.544 +0.04/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-10
Table 6. 85056D Calibration Kit
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-OR­Standard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-OR­Extended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL & 014)
Applies to the, E8363/4A analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity 42 34 26 26 Source Match 40 30 24 23 Load Match 42 33 25 25 Reflection Tracking
Transmission Tracking
±0.002 +0.02/°C ±0.022 +0.02/°C
±0.029 +0.02/°C ±0.130 +0.02/°C
±0.079 +0.03/°C ±0.384 +0.03/°C
0.075 +0.04/°C
0.589 +0.04/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-11
E8363/4A Corrected System Performance with 2.92mm Connectors

Table 7. 85056K Calibration Kit
Standard Configuration and Standard Power Range (E8363/4A)
Applies to the, E8363/4A analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to 40 GHz
Directivity 42 42 42 40 Source Match 40 40 40 35 Load Match 42 42 42 38 Reflection Tracking
Transmission Tracking
±0.018 +0.02/°C ±0.016 +0.02/°C
±0.018 +0.02/°C ±0.028 +0.02/°C
±0.018 +0.03/°C ±0.033 +0.03/°C
±0.067 +0.04/°C ±0.089 +0.04/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-12
Table 8. 85056K Calibration Kit
Extended Configuration and Standard Power Range (E8363/4A - Option 014)
-OR­Standard Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL)
-OR­Extended Configuration and Extended Power Range & Bias-Tees (E8363/4A - Option UNL&014)
Applies to the, E8363/4A analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to 40 GHz
Directivity 42 42 42 40 Source Match 40 40 40 35 Load Match 42 42 41 38 Reflection Tracking
Transmission Tracking
±0.018 +0.02/°C ±0.021 +0.02/°C
±0.018 +0.02/°C ±0.033 +0.02/°C
±0.018 +0.03/°C ±0.046 +0.03/°C
±0.067 +0.04/°C ±0.098 +0.04/°C
Transmission Uncertainty (Specifications
Reflection Uncertainty (Specifications)
4-13
E836xA Corrected System Performance with 3.5mm Connectors

Table 9. 85052B Calibration Kit
Standard Configuration and Standard Power Range (E836xA)
Applies to the, E836xA analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to
26.5 GHz
Directivity 48 44 44 44 Source Match 40 31 31 31 Load Match 48 44 44 44 Reflection Tracking
Transmission Tracking
±0.003 +0.02/°C ±0.013 +0.02/°C
±0.006 +0.02/°C ±0.057 +0.02/°C
±0.006 +0.02/°C ±0.065 +0.02/°C
±0.006 +0.03/°C ±0.104 +0.03/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-14
Table 10. 85052B Calibration Kit
Extended Configuration and Standard Power Range (E836xA - Option 014)
-OR­Standard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL)
-OR­Extended Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL&014)
Applies to the, E836xA analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to
26.5 GHz
Directivity 48 44 44 44 Source Match 40 31 31 31 Load Match 48 44 44 44 Reflection Tracking
Transmission Tracking
±0.003 +0.02/°C ±0.017 +0.02/°C
±0.006 +0.02/°C ±0.065 +0.02/°C
±0.006 +0.02/°C ±0.091 +0.02/°C
±0.006 +0.03/°C ±0.109 +0.03/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-15
Table 11. 85052C Calibration Kit
Standard Configuration and Standard Power Range (E836xA)
Applies to the, E836xA analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to
26.5 GHz
Directivity 48 50 50 50 Source Match 40 50 50 50 Load Match 48 50 50 50 Reflection Tracking
Transmission Tracking
±0.003 +0.02/°C ±0.013 +0.02/°C
±0.000 +0.02/°C ±0.010 +0.02/°C
±0.000 +0.02/°C ±0.012 +0.02/°C
±0.000 +0.03/°C ±0.018 +0.03/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-16
Table 12. 85052C Calibration Kit
Extended Configuration and Standard Power Range (E836xA - Option 014)
-OR­Standard Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL)
-OR­Extended Configuration and Extended Power Range & Bias-Tees (E836xA - Option UNL&014)
Applies to the, E836xA analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)Description
0.045 to 2GHz
2to 10 GHz
10 to 20 GHz
20 to
26.5 GHz
Directivity 48 50 50 50 Source Match 40 50 50 50 Load Match 48 50 50 50 Reflection Tracking
Transmission Tracking
±0.003 +0.02/°C ±0.017 +0.02/°C
±0.000 +0.02/°C ±0.012 +0.02/°C
±0.000 +0.02/°C ±0.016 +0.02/°C
±0.000 +0.03/°C ±0.021 +0.03/°C
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)
4-17