THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT
TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL
WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE
USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING
THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is
delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June
1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use,
duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive
no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any technical data.
ii85032F
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Agilent web site:
•Go to www.agilent.com.
•Enter the document’s part number (located on the title page) in the Search box.
•Click Search.
•Click on the hyperlink for the document.
•Click the printer icon located in the tool bar.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent
products.
For any assistance, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-4.
Class Assignments and Standard Definitions Values are Available on the Web. . . . . . . . . .A-2
Contents-vi85032F
1General Information
1-1
General Information
Calibration Kit Overview
Calibration Kit Overview
The Agilent 85032F type-N 50 Ω calibration kit is used to calibrate Agilent network
analyzer systems for measurements of components with type-N connectors up to 9 GHz.
The standards in this calibration kit allow you to perform simple 1- or 2-port calibrations.
This manual describes the 85032F calibration kit and provides replacement part numbers,
specifications, and procedures for using, maintaining and troubleshooting the kit.
NOTE
This manual assumes you practice proper connector care. For information,
refer to Chapter 3,“Use, Maintenance, and Care of the Devices.” Also refer to
the “Connector Care – Quick Reference Card,” located in the back of this
manual.
Kit Contents
The 85032F type-N calibration kit contains the following:
• offset opens and shorts, and broadband load terminations with both male and female
connectors
• a disk that contains the calibration definitions of the devices in the kit for 872x and
875x VNA systems
Broadband Loads
The broadband loads are metrology-grade, 50 Ω terminations that have been optimized for
performance up to 9 GHz. The rugged internal structure provides for highly repeatable
connections. A distributed resistive element on sapphire provides excellent stability and
return loss.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current
state-of-the-art precision machining.
The offset short’s inner conductors have a one-piece construction, common with the
shorting plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong,
low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very
tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts
are designed so that the difference in phase of their reflection coefficients is approximately
180 degrees at all frequencies.
1-285032F
General Information
Calibration Kit Overview
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the
kit installed in the network analyzer prior to performing a calibration. Refer to your
network analyzer user’s guide for instructions on selecting the calibration kit and
performing a calibration.
The calibration definitions can be:
• resident within the analyzer
• loaded from the provided disk
• entered from the front panel
Class assignments and standard definitions may change as more accurate model and
calibration methods are developed. You can download the most recent class assignments
and standard definitions from Agilent’s Calibration Kit Definitions Web page at
www.na.tm.agilent.com/pna/caldefs/stddefs.html.
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal
memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network
analyzer, they must be manually entered. Refer to your network analyzer user’s guide for
instructions.
Options
There are several optional adapters available for the 85032F. See Table 6-1 on page 6-3 for
further information.
Equipment Required but Not Supplied
Connector gage sets, optional adapters, wrenches, and various connector cleaning supplies
are not included with this calibration kit. These or similar items are recommended to
ensure the successful operation of this calibration kit. Refer to Table 6-2 on page 6-4 for
ordering information.
85032F1-3
General Information
Incoming Inspection
Incoming Inspection
Refer to Figure 6-1 on page 6-2 to verify a complete shipment.
Check for damage. The foam-lined storage case provides protection during shipping. If the
case or any device appears damaged, or if the shipment is incomplete, contact Agilent
Technologies. See “Contacting Agilent” on page 5-4. Agilent will arrange for repair or
replacement of incomplete or damaged shipments without waiting for a settlement from
the transportation company. See “Returning a Kit or Device to Agilent” on page 5-3.
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized
(serial numbers are labeled onto the body of each device). For your convenience, the
following table is provided for you to record these serial numbers. Recording the serial
numbers will prevent you from confusing the devices with similar devices.
Table 1-1Kit and Device Serial Number Record
DeviceSerial Number
Calibration kit
Broadband load –m–
Broadband load –f–
Open –m–
Open –f–
Short –m–
Short –f–
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the gender of cable
connectors and adapters is referred to in terms of the center conductor. For example, a
connector or device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
8510-series, 872x, and 875x ONLY: In contrast, during a measurement calibration, the
network analyzer softkey menus label a 1.85 mm calibration device with reference to the
sex of the analyzer’s test port connector—not the calibration device connector. For
example, the label SHORT(F) refers to the short that is to be connected to the female test
1-485032F
port. This will be a male short from the calibration kit.
Table 1-2 Clarifying the Sex of Connectors: Examples
TerminologyMeaning
General Information
Preventive Maintenance
Short
–f–
Short (f)Male short (male center conductor) to be connected to female port
Female short (female center conductor)
A connector gage is referred to in terms of the connector that it measures. For instance, a
male connector gage has a female connector on the gage so that it can measure male
devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of these are described in Chapter 3,“Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can
degrade repeatability and accuracy, and can damage any connector mated to it.
Improper connections, resulting from pin depth values being out of the observed limits (see
Table 2-3 on page 2-4), or from poor connection techniques, can also damage these devices.
85032F1-5
General Information
Preventive Maintenance
1-685032F
2Specifications
2-1
Specifications
Environmental Requirements
Environmental Requirements
NOTE
Samples of this product have been type-tested in accordance with the Agilent
Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those stresses
include but are not limited to temperature, humidity, shock, vibration,
altitude and power-line conditions. Test methods are aligned with IEC
60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Table 2-1Environmental Requirements
Parameter Limits
Tem per at ur e
Operating
Storage−40 °C to +75 °C
Error-corrected range
Altitude
Operating< 4,500 meters (≈15,000 feet)
Storage< 4,500 meters (≈15,000 feet)
Relative humidityType tested, 0% to 95% at 40 °C, non-condensing
a
b
+20 °C to +26 °C
± 1 °C of measurement calibration temperature
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
Table 2-1.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2-285032F
Specifications
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not
performance specifications. They are, however, important supplemental characteristics
related to electrical performance. Agilent Technologies verifies the mechanical
characteristics of the devices in this kit with special gaging processes and electrical
testing. This ensures that the device connectors do not exhibit any center conductor
protrusion or improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-7 explains how to use gages to determine if the kit devices
have maintained their mechanical integrity. (Refer to Table 2-3 on page 2-4 for typical and observed pin depth limits).
Table 2-2 Mechanical Characteristics
DimensionTypical Value
Inside diameter of outer conductor7.0000 ±0.0076 mm
Outside diameter of center conductor3.0404 ±0.0051 mm
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with
the outer conductor mating plane. See Figure 2-1. Some coaxial connectors (such as
2.4 mm and 3.5 mm) are designed to have these planes nearly flush. Type-N connectors,
however, are designed with a pin depth offset of approximately 5.26 mm (0.207 inch), not
permitting these planes to be flush. The male center conductors are recessed by the offset
value while the female center conductors compensate by protruding the same amount.
This offset necessitates redefining of pin depth with regard to protrusion and recession.
Protrusion refers to a male type-N connector center conductor having a pin depth value
less than 5.26 mm (0.207 inch), or female type-N connector center conductor having a pin
depth value greater that 5.26 (0.207 inch).
Recession refers to a male type-N connector center conductor having a pin depth value
greater than 5.26 mm (0.207 inch), or female type-N connector center conductor having a
pin depth value less than 5.26 (0.207 inch).
The pin depth value of each calibration device in this kit is not specified, but is an
important mechanical parameter. The electrical performance of the device depends, to
some extent, on its pin depth. The electrical specifications for each device in this kit take
into account the effect of pin depth on the device’s performance. Table 2-3 lists the typical
pin depths and measurement uncertainties, and provides observed pin depth limits for the
devices in the kit. If the pin depth of a device does not measure within the observed pin
depth limits, it may be an indication that the device fails to meet electrical specifications.
Refer to Figure 2-1 for a visual representation of proper pin depth in type-N connectors.
85032F2-3
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.