Network Analyzer Accessories
Overview
Accessories for the HP 8753D, 8720D, and the
8510C series network analyzers include a variety
of calibration kits, verification kits, cables, and
adapters from dc to 110 GHz.
Calibration Kits
Error-correction procedures require that the
systematic errors in the measurement system be
characterized by measuring known devices (standards) on the system over the frequency range of
interest. HP offers two types of calibration kits:
mechanical and electronic.
Electronic Calibration (ECal) Kits
ECal kits consist of connector specific modules,
adapters for test ports and a torque wrench for
proper connection.
75
Mechanical Calibration Kits
All network analyzer mechanical calibration kits
contain precision standard devices to characterize
the systematic errors of the HP 8753 series,
8720 series or 8510C network analyzer system.
Each mechanical calibration kit also contains
adapters for test ports and a torque wrench for
proper connection.
Verification Kits
Measuring known devices, other than the calibration standards, is a straightforward way of verifying
that the network analyzer system is operating
properly. HP offers verification kits that include
precision airlines, mismatch airlines, and precision
fixed attenuators. Traceable measurement data is
shipped with each kit on disk. Verification kits may
be recertified by Hewlett-Packard. This recertification
includes a new measurement of all standards and
new data with uncertainties.
11

76
Network Analyzer Accessories
Electronic Calibration Kits
1
Frequency
HP Model Connector Range (GHz) Description
85060A 7 mm 1 to 18 One two-port ECal
module and collet
remover
Option 001 7 mm 0.045 to 2 Adds lowband
ECal module
85062A 3.5 mm (m-f) 1 to 26.5 One two-por t
ECal module and
m-m and f-f adapters
Option 001 3.5 mm (m-f) 0.045 to 2 Adds lowband
ECal module
Option 00F 3.5 mm (f-f) Changes
connectors to f-f
Option 00M 3.5 mm (m-m) Changes
connectors to m-m
85064A Type-N (m-f) 1 to 18 One two-port
ECal module
Option 001 Type-N (m-f) 0.045 to 2 Adds lowband
ECal module
Option 00F Type-N (f-f) Changes
connectors to f-f
Option 00M Type-N (m-m) Changes
connectors to m-m
1
Must have an HP 85060C control unit to perform electronic calibration.
Mechanical Calibration Kits
2,3,4,5
Frequency
HP Model Connector Range (GHz) Type
85036B Type-N (75 Ω) DC to 6 Economy
85039A Type-F (75 Ω) DC to 3 Economy
85031B 7 mm DC to 6 Economy
85050B 7 mm 0.045 to 18 Standard
85050C 7 mm 0.045 to 18 Precision
85050D 7 mm 0.045 to 18 Economy
85033D 3.5 mm DC to 6 Economy
85052B 3.5 mm 0.045 to 26.5 Standard
85052C 3.5 mm 0.045 to 26.5 Precision
85052D 3.5 mm 0.045 to 26.5 Economy
85032B Type-N DC to 6 Economy
85054B Type-N 0.045 to 18 Standard
85054D Type-N 0.045 to 18 Economy
85056A 2.4 mm 0.045 to 50 Standard
85056D 2.4 mm 0.045 to 50 Economy
6
85056K
2.92 mm 0.045 to 50 Economy
85058D 1.85 mm 0.045 to 65 Standard
X11644A WR-90 8.2 to 12.4 Precision
P11644A WR-62 12.4 to 18 Precision
K11644A WR-42 18 to 26.5 Precision
R11644A WR-28 26.5 to 40 Precision
Q11644A WR-22 33 to 50 Precision
U11644A WR-19 40 to 60 Precision
V11644A WR-15 50 to 75 Precision
W11644A WR-10 75 to 110 Precision
2
Unless stated otherwise, all calibration kits are 50 ohms.
3
Economy calibration kits contain fixed loads.
4
Standard calibration kits contain sliding loads.
5
Precision calibration kits contain TRL calibration standards.
6
Option 001 adds a sliding load.
Verification Kits
Verification Kit 85029B 85051B 85053B 85055A 85057B R11645A Q11645A U11645A V11645A W11645A
Connector Type 7 mm 7 mm 3.5 mm Type-N 2.4 mm WR-28 WR-22 WR-19 WR-15 WR-10
Frequency Range DC to 6 0.045 0.045 0.045 0.045 26.5 to 40 33 to 50 40 to 60 50
(GHz) to 18 to 26.5 to 18 to 50
to
75 75 to 110