Agilent 81635A Users Guide

HP 8163A/4A/6A Lightwave Series Mainframes
HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module User’s Guide
E1299
Notices
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Hewlett-Packard GmbH.
Copyright 1999 by: Hewlett-Packard GmbH Herrenberger Str. 130 71034 Böblingen Germany
Subject Matter
The information in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this printed material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Printing History
New editions are complete revisions of the guide reflecting alterations in the functionality of the instrument. Updates are occasionally made to the guide between editions. The date on the title page changes when an updated guide is published. To find out the current revision of the guide, or to purchase an updated guide, contact your Hewlett-Packard representative.
Control Serial Number: First Edition applies directly to all instruments.
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory.
Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, NIST (formerly the United States National Bureau of Standards, NBS) to the extent allowed by the Institutess calibration facility, and to the calibration facilities of other International Standards Organization members.
ISO 9001 Certification
Produced to ISO 9001 international quality system standard as part of our objective of continually increasing customer satisfaction through improved process control.
81635-90014 E1299 First Edition:
E1299: December 1999
Warranty
This Hewlett-Packard instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, HP will, at its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by HP. Buyer shall prepay shipping charges to HP and HP shall pay shipping charges to return the product to Buyer. However, B uyer shall pay all shipping charges, duties, and taxes for products returned to HP from another country.
HP warrants that its software and firmware designated by HP for use with an instrument will execute its programming instructions when properly installed on that instrument. HP does not warrant that the operation of the instrument, software, or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Hewlett-Packard specifically disclaims the implied warranties of Merchantability and Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive remedies. Hewlett-Packard shall not be liable for any direct, indirect, special, incidental, or consequential damages whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For any assistance contact your nearest Hewlett-Packard Sales and Service Office.
2 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Safety Considerations

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Hewlett­Packard Company assumes no liability for the customer’s failure to comply with these requirements.
Before operation, review the instrument and manual, including the red safety page, for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
WARNING The WARNING sign denotes a hazard. It calls attention to a procedure, practice
or the like, which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.

Safety Symbols

The apparatus will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect the apparatus against damage.

Initial Inspection

Inspect the shipping container for damage. If there is damage to the container or cushioning, keep the m until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically.
The Performance Tests give procedures for checking the operation of the instrument. If the contents are incomplete, mechanical damage or defect is apparent, or if an instrument does not pass the operators checks, notify the nearest Hewlett-Packard Sales/Service Office.
WARNING To avoid hazardous electrical shock, do not perform electrical tests when there
are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).

Line Power Requirements

The HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module operate when installed in the HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System, and HP 8166A Lightwave Multichannel System.
HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 3

Operating Environment

The safety information in the HP 8163/4A Lightwave Measurement System Users Guide summarizes the operating ranges for the HP 81632A Power Sensor Module and HP 81635A Dual Powe r Sensor Module. In order for these modules to meet specifications, the operating environment must be within the limits specified for the HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System, and HP 8166A Lightwave Multichannel System.

Storage and Shipment

This module can be stored or shipped at temperatures between
40°C and +70°C. Protect the module from temperature extremes that may cause condensation within it.
4 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Contents

Safety Considerations 3
Safety Symbols 3 Initial Inspection 3
Line Power Requirements 3
Operatin g Environment 4 Storage and Shipment 4
Contents 5 Accessories 7
Modules and Options 9
Specifications 11
Power Sensor Specifications (Autorange Mode) 13
Performance Tests 15
Equipment Required 17 Test Record 18 Test Failure 18 Instrument Specification 18
Performance Test 18
Accuracy Test 19
Test Setup 19
Linearity Test 21
Test Setup 21 +10 dBm Range 22 0 dBm Range 23
-10 dBm Range 23 Change Setup 23
-20 dBm to -50 dBm Range 24 Calculation 24 Example: Measurement Results 25 Calculations 26
Noise Test 27 Return Loss Test 28 Relative Uncertainty due to Polarization (Optional Test) 29 Relative Uncertainty due to Interference (Optional Test) 31
HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 5
Calculation Sheet for Linearity Measurement 42
Index 43
6 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Accessories

HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 7
The HP 81632A Power Sensor Module and the HP 81635A Dual Power Sensor Module are available in various configurations for the best possible match to the most common applications.
This chapter provides information on the available options and accessories.
8 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Modules and Options

Connector Interfaces
81000AI Diamond HMS-10 81000FI FC/PC/SPC 81000GI D4 81000JI SMA 81000KI SC/APC 81000NI FC/APC 81000PI E-2000 81000SI DIN 47526 81000VI ST 81000WI Biconic
Connector Interfaces
12
Power Sensor Module
HP 8163A Lightwave Multimeter
Power Sensors Economic Mo dules
81632A 10 dBm to -90 dBm 81635A 10 dBm to -110 dBm
Connector Interfaces
Connector Interfaces
Bare
Fiber
Adapters
and Bare Fiber Adapters
81000FI FC/PC/SPC 81000FB Bare Fiber Adapter
81000KI SC/APC 81000KB Bare Fiber Adapt er
81000SI DIN 47526 81000SB Bare Fiber Adapter
81000VI ST 81000SB Bare Fiber Adapter
Figure 1 Recommended Connector Interfaces and Bare Fiber Adapters
HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 9
Product Number Connector InterfaceType
81000AI Diamond HMS-10 81000FI FC/PC/SPC 81000GI D4 81000JI SMA 81000KI SC/APC 81000NI FC/APC 81000PI E-2000 81000SI DIN 47526 81000VI ST 81000WI Biconic
Table 1 Connector Interfaces
Product Number Bare Fiber Adapter Product Number Connector Interface
81000FB Bare Fiber Adapter 81000FI FC/PC/SPC 81000KB Bare Fiber Adapter 81000KI SC/APC 81000SB Bare Fiber Adapter 81000SI DIN 47526 81000SB Bare Fiber Adapter 81000VI ST
Table 2 Connector Interface and Bare Fiber Adapter Combinations
10 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Specifications

HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 11
HP 81632A Power Sensor Module and the HP 81635A Dual Power Sensor Module are produced to the ISO 9001 international quality system standard as part of HPs commitment to continually increasing customer satisfaction through improved quality control.
Specifications describe the modules and heads warranted performance. Supplementary performance characteristics describe the modules and heads non-warranted typical performance.
Because of the modular nature of the instrument, these performance specifications apply to these modules and heads rather than the mainframe unit.
12 HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module
Users Guide, E1299

Power Sensor Specifications (Autorange Mode)

HP 81632A HP 81635A Sensor Element Wavelength Range Power Range Display Resolution Applicable Fiber Type Uncertainty (accuracy) at Refer-
ence Conditions Total Uncertainty Relative Uncertainty
Spectral Ripple (due to interference)
Linearity (power)
1
2
Due to Polarization
5
- at 23°C ±5°C
- at operating temp. range
InGaAs InGaAs (dual)
800 - 1650 nm
¥80 dBm
+10 to
0.0001dB/dBm, 0.01pW to 10 pW (depending on power range)
Standard SM and MM up to 62.5 µm core size, NA ≤ 0.24
< ±3% (1200 nm to 1630 nm)
8
< ±5%
3
4
CW +10 to
(1200 nm to 1630 nm)
± 0.015 dB typical
± 0.015 dB typical
¥60 dBm (1200 nm to 1630 nm)
< ± 0.02 dB ± 20 pW < ± 0.06 dB ± 20 pW
9 9
Return Loss Noise (peak to peak) Averaging Time (minimal) Dimensions (H x W x D) Weight Recalibration Period Operating Temperature Humidity Warm-up time
HP 81632A Power Sensor Module and HP 81635A Dual Power Sensor Module Users Gu ide, E1299 13
7
5, 6
75 mm × 32 mm × 335 mm (2.8× 1.3” × 13.2”)
>40 dB
< 20 pW
100 µs
0.5 kg
2 years
+10°C to +40°C Non-condensing
20 minutes