This document contains proprietary information that is protected by
copyright. All rights are reserved.
No part of this doc ument may rep roduced in ( including e lectronic s torage
and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies Deutschland
GmbH as governed by United States and international copywright laws.
”Copyright 2001 - 2004 by:
Agilent Technologies Deutschland GmbH
Herrenberger Str. 130
71034 Böblingen
Germany
Subject Matter
The material i n this document is subject to change without notice.
Agil ent Techn ologies makes no warranty of any kind with regard to this
printed materi al, including, but not limite d to, the implied warranties of
merchantability and fitness for a particular purpose.
Agilent Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Printing History
New editions are complete revisions of the guide reflecting alterations in
the functionality of the instrument. Updates are occasionally made to the
guide between editions. The date on the title page changes when an
updated guide is published. To find out the current revision of the guide,
or to purchase an updated guide, contact your Agilent Technologies
representative.
Control Serial Number: First Edition applies directly to all instruments.
Warranty
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of one year from date of
shipment. During the warranty period, Agilent will, at its option, either
repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service
facility designated by Agilent. Buyer shall prepay shipping charges to
Agilent and A gilent shall pay shipping charge s to return the product to
Buyer. However, Buyer shall pay all shipping charges, duties, and taxes
for products returned to Agilent from another country.
Agilent warrants tha t its software and firmware designated by Agilent for
use with an instrument will execute its programming instructions when
properly installed on that instrument. Agilent does not warrant that the
operation of the instrument, software, or firmware will be uninterrupted
or error free.
Certification
Agilent Technologies Inc. certifies that this product met its published
spec ifications at the ti me of ship ment from the fact ory.
Agilent Technologies further certifies that its calibration measurements
are traceable to the United States National Institute of Standards and
Technology, NIST (formerly the United States National Bureau of
Standards, NBS ) to the extent allowed by t he Institutes’s ca libration
facility, and to the calibration facilities of other International Standards
Organization members.
ISO 9001:2000 Certification
Produced to ISO 9001:2000 international quality system standard as part
of our objective of continually increasing customer satisfaction through
improved process contr ol.
Manual Part Number
81570-90A01
Eigth Edition:
81570-90A01: September 01, 2005
Seventh Edition:
81570-90A01: July 20, 2005
Sixth Edition:
81570-90A01: December, 2004
Fifth Edition:
81570-90A01: June 28, 2004
Fourth Edition:
81570-90A01: January 31, 2003
Third Edition:
81560-90A13: October 29, 2002
Seco nd Edition :
81560-90A12: August 04, 2002
First Edition:
81560-90A11: November 2001
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper
or inadequate m aintenance by Buyer, Buyer-suppl ied software or
interfacing, unauthorized modification or mi suse, operation outside of the
environmental specifications for the product, or improper site preparation
or maintenance.
No other warranty is expressed or implied. Agilent Technologies
specifically disclaims the implied warranties of Merchantability and
Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exc lusive remedies.
Agilent Technologies shall not be liable for any direct, indirect, special,
incidental, or conseq uential damages whet her based on contract, tort, or
any other legal theory.
Assistance
Product maintenance agreements and other customer assistance
agreements are available for Agilent Technologies products. For any
assistance contact your nearest Agilent Technologies Sales and Service
Office.
WAR N INGThe WARNING sign denotes a hazard. It calls attention to a procedure, practice or the like,
Safety Summary
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with
specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no
liability for the customer’s failure to comply with these requirements.
Before operation, review the instrument and manual, including the red safety page, for
safety markings and instructions. You must follow these to ensure safe operation and to
maintain the instrument in safe condition.
which, if not correctly performed or adhered to, could result in injury or loss of life. Do not
proceed beyond a WARNING sign until the indicated conditions are fully understood and
met.
CA U TI O NOptical power levels above 100 mW applied to single mode connectors can easily damage
the connector if it is not perfectly clean. Also, scratched or poorly cleaned connectors can
destroy optical connectors mechanically. Always make sure that your optical connectors are
properly cleaned and unscratched before connection. Refer to chapter “Cleaning
Information” on page 77 on appropriate procedures for connector cleaning and inspection.
However, Agilent Technologies Deutschland GmbH assumes no responsibility in case of an
operation that is not compliance with the safety instructions as stated above.
Safety Symbols
The apparatus will be marked with this symbol when it is necessary for the user to refer to
the instruction manual in order to protect the apparatus against damage.
Hazardous laser radiation.
Initial Inspection
Inspect the shipping container for damage. If there is damage to the container or
cushioning, keep them until you have checked the contents of the shipment for
completeness and verified the instrument both mechanically and electrically.
The Performance Tests give procedures for checking the operation of the instrument. If the
contents are incomplete, mechanical damage or defect is apparent, or if an instrument does
not pass the operator’s checks, notify the nearest Agilent Technologies Sales/Service
Office.
WARN INGTo avoid hazardous electrical shock, do not perform electrical tests when there are signs of
shipping damage to any portion of the outer enclosure (covers, panels, etc.).
WARNINGYou MUST return instruments with malfunctioning laser modules to an Agilent Technologies
Sales/Service Center for repair and calibration.
Line Power Requirements
The Agilent 81570A, 71A, 78A Variable Optical Attenuator modules and Agilent 81576A,
77A Variable Optical Attenuator modules with Power Control operate when installed in the
Agilent 8163A and B Lightwave Multimeters, the Agilent 8164A and B Lightwave
Measurement Systems, or the Agilent 8166A and B Lightwave Multichannel Systems.
Within this User’s Guide, these instruments are collectively referred to as ‘mainframes’.
Operating Environment
The safety information in your mainframe’s User’s Guide summarizes the operating ranges
for the Agilent 81570A, 71A, 78A Variable Optical Attenuator modules and Agilent
81576A, 77A Variable Optical Attenuator modules with Power Control. In order for these
modules to meet specifications, the operating environment must be within the limits
specified for your mainframe.
Storage and Shipment
Agilent 81570A, 71A, 78A Variable Optical Attenuator modules and Agilent 81576A, 77A
Variable Optical Attenuator modules with Power Control can be stored or shipped at
temperatures between
-40°C and +70°C.
Protect the module from temperature extremes that may cause condensation within it.
Environmental Information
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/ electronic product in
domestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive Annex I,
this product is classed as a "Monitoring and Control instrumentation" product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see
www.agilent.com/environment/product/ for more information.
For Agilent 8156x modules with a particular firmware revision to operate correctly, your
Agilent 8163A/B, 8164A/B, or 8166A/B mainframe must have a V3.5, or later, firmware
revision installed.
For Agilent 8157x modules with a particular firmware revision to operate correctly, your
Agilent 8163A/B, 8164A/B, or 8166A/B mainframe must have a V4.0, or later, firmware
revision installed.
Firmware files are provided on the OCT Support CD-ROM supplied with the instrument, and
the latest firmware revisions can also be downloaded via
www.agilent.com/comms/octfirmware
The firmware for A and B versions of our mainframes is not binary compatible. Make sure
you select the appropriate firmware for your instrument.
NOTE• Updating mainframe firmware does not require you to update the firmware in every
hosted module. New revisions of mainframe firmware are backwards compatible with
older module firmware.
• You may find it advantageous to update module firmware. To help you determine which
module firmware to update, a readme file is provided with the firmware file detailing its
revision history.
• If the serial number of your 8164A Lightwave Measurement System is DExxxxx339 or
below, problems with writing files to the floppy disk drive may occur in rare cases. If this
is the case, please contact your nearest Agilent Technologies Sales / Service Office.
8163A/B Lightwave Multimeter, 8164A/B Lightwave
Measurement System, or 8166A/B Lightwave Multichannel
System
To check your 8163A/B Lightwave Multimeter, 8164A/B Lightwave Measurement System,
or 8166A/B Lightwave Multichannel System's firmware revision:
1 Press the Config hardkey.
2 Move to the <About Mainframe> menu option and press Enter.
3 The manufacturer, part number, instrument number, and firmware revision are listed. If
the firmware revision number is less than V3.5 for 8156x modules, or is less than V4.0
for 8157x modules, follow the Update Procedure described in the readme.txt in the root
directory of the OCT Support CD-ROM supplied with the instrument to install a later
firmware revision.
8157x Series Modules
To check an 8157x series, module's firmware revision:
1 Press the Config hardkey.
2 Move to the <About Modules> menu option and press Enter. You see a box displaying
the slots that have a module installed.
3 Move to the appropriate slot using the cursor key and press Enter.
4 The part number, module number, and firmware revision of the chosen module are
displayed. If the firmware revision number is less than V4.0 for 8157x modules, follow
the Update Procedure described in the readme.txt in the root directory of the OCT
Support CD-ROM supplied with the instrument to install a later firmware revision.
Microscope with a magnification range about 50X up to 300X83
Ultrasonic bath83
Warm water and liquid soap84
Premoistened cleaning wipes84
Polymer film84
Infrared Sensor Card84
Preserving Connectors85
Cleaning Instrument Housings85
General Cleaning Procedure86
How to clean connectors86
How to clean optical head adapters87
How to clean connector interfaces88
How to clean bare fiber adapters89
How to clean lenses and instruments with an optical glass plate89
How to clean instruments with a fixed connector interface90
How to clean instruments with a physical contact interface90
How to clean instruments with a recessed lens interface92
How to clean optical devices which are sensitive to mechanical stress
and pressure92
How to clean metal filters or attenuator gratings93
Additional Cleaning Information94
How to clean bare fiber ends94
How to clean large area lenses and mirrors94
Figure 5Agilent 81576A High-power Attenuator with Power Control and Straight Connector . 10
Figure 6Agilent 81577A High-power Attenuator with Power Control and Angled Connector . 10
CAUTIONOptical power levels above 100 mW applied to single mode connectors can easily damage
This chapter describes the:
• Agilent 81570A, 81571A VOA modules,
• Agilent 81578A Multimode VOA and the
• Agilent 81576A and 81577A VOA modules with power control
the connector if it is not perfectly clean. Also, scratched or poorly cleaned connectors can
destroy optical connectors mechanically. Always make sure that your optical connectors are
properly cleaned and unscratched before connection. Refer to chapter “Cleaning
Information” on page 77 on appropriate procedures for connector cleaning and inspection.
However, Agilent Technologies Deutschland GmbH assumes no responsibility in case of an
operation that is not compliance with the safety instructions as stated above.
What is an Attenuator?Getting Started with Attenuator Modules
What is an Attenuator?
Agilent 8157x Variable Optical Attenuators attenuate and control the optical power of light
in single and multimode optical fibers. They allow you to set the attenuation factor and/or
power level manually, or via the host instrument’s GPIB interface.
Agilent 81576A and 81577A attenuators include power control functionality that allows
you to set the output power level of the attenuator. When power control mode is enabled,
the module automatically corrects for power changes at the input so that the output power
level set by the user is maintained. See Figure 1 on page 17.
Installation
Agilent 8157xA Variable Optical Attenuators are front-loadable modules for the Agilent
8163A/B Lightwave Multimeter, 8164A/B Lightwave Measurement System, and 8166A/B
Lightwave Multichannel System, collectively referred to as ‘mainframes’.
An Agilent 81570A, 71A or 78A module occupies one slot, while an Agilent 81576A or 77A
module occupies two slots.
For a description of how to install your module, refer to “How to Fit and Remove Modules”
in the Installation and Maintenance chapter of your mainframe’s User’s Guide.
81576A / 81577A
input
Figure 1Agilent 81576A/77A Optical Attenuators with Power Control
What is an Attenuator?Getting Started with Attenuator Modules
Front Panel Controls and Indicators
The front panel contains a push button to switch the shutter open/closed, and a LED that
indicates the state of the shutter. If the shutter is “open”, the LED shows “enable”, and light
can pass through the instrument.
Typical Us e Models
Brief descriptionAgilent's 8157xA Variable Optical Attenuators are instruments that attenuate and control
the optical power level of light in single and multimode optical fibers. As plug-in modules for
Agilent's Lightwave Multichannel platform (8163A/B, 8164A/B, 8166A/B) they allow you
to set the attenuation factor and/or power level manually, or remotely via a common
computer interface. Their high accuracy combined with their flexibility make them ideal as
test and measurement equipment for the modern telecommunication industry.
Modular Design for Multichannel PlatformAgilent's 8157xA variable optical attenuators are a family of plug-in modules for Agilent's
Lightwave Multichannel Platform 8163A/B, 8164A/B and 8166A/B. The attenuator
modules 81570A, 81571A, 81578A occupy one slot, while modules 81576A and 81577A
occupy two slots. The Agilent 8166A/B Lightwave Multichannel System with its 17 slots
can host up to 17 single slot modules (such as the 81570A attenuators) or up to 8 dual slot
modules, (such as the 81577A attenuators).
Variable Optical AttenuatorsAgilent's 81570A, 81571A and 81578A are small and cost effective attenuator modules
with high resolution for single-wavelength applications. Once you have entered the
operating wavelength, the instrument automatically applies the appropriate corrections.
Various calibration features allow you to set a reference power so that both the attenuation
and the power level, relative to the reference power, can be set and displayed in the user
interface. An integrated shutter can be used for protection purposes or to simulate channel
drops.
Agilent's 81570A and 81571A have excellent wavelength flatness and can handle high
input power levels. These features, combined with low insertion loss, allow you to use these
modules for characterizing EDFAs and Raman amplifiers, as well for other multi-wavelength
applications such as DWDM transmission system testing. They include the reference power
functionality and an integrated shutter.
Agilent's 81578A modular attenuators are available for multimode applications with fiber
interface option for 50µm (#050) and 62.5µm (#062).
Attenuators with Power ControlAgilent's 81576A and 81577A attenuators have the power control functionality combined
with the high-power handling capability and excellent wavelength flatness required for
DWDM applications.
To set the the total power level of a multi-wavelength signal, it is necessary to determine the
convolution of the signal’s spectrum against the sensitivity of the photodiode over
wavelength. An enhanced calibration feature allows you to set the integrated power of the
DWDM signal against a known spectrum.
Getting Started with Attenuator ModulesWhat is an Attenuator?
Calibration ProcessesComprehensive offset functionality in the firmware enhances the calibration of the optical
path in various test set-ups. There is an offset for the attenuation factor, and an
independent offset for the output power level, to calibrate for losses due to the patch cords
and connectors. Additionally, wavelength and offset value pairs can be stored in a table to
compensate for wavelength dependent effects in the optical path of the set-up. This allows
you to set the optical power level at your Device Under Test.
Calibration is even easier and more convenient if the reference powermeter and the
attenuator are hosted by the same mainframe: All power related offsets can be determined
by a firmware function that reads a value from the reference powermeter. The difference
between the power value read by the reference powermeter and the actual value of the
attenuator is automatically stored as the offset.
Optical OutputGetting Started with Attenuator Modules
Optical Output
Angled and Straight Contact Connectors
Agilent 81570A Variable Optical Attenuator modules and 81576A Variable Optical
Attenuator modules with Power Control are designed for straight connector interfaces.
Agilent 81571A Variable Optical Attenuator modules and 81577A Variable Optical
Attenuator modules with Power Control are designed for angled connector interfaces.
The Agilent 81578A (with option 050 or 062) Variable Optical Attenuator modules are
designed for straight (FC/PC) multimode connector.
The inclusion of an angled contact connector is not optionable, and depends on the module
part number.
Angled contact connectors help you to control return loss, since reflected light tends to
reflect into the cladding, reducing the amount of light that reflects back to the source.
CAUTIONIf the contact connector on your instrument is angled, you can only use cables with angled
connectors with the instrument.
Angled Contact
Connector Symbol
Figure 7Angled and Straight Contact Connector Symbols
Figure 7 shows the symbols that tell you whether the contact connector of your attenuator
module is angled or straight. The angled contact connector symbol is colored green.
You should connect straight contact fiber end connectors with neutral sleeves to straight
contact connectors, or connect angled contact fiber end connectors with green sleeves to
angled contact connectors.
NOTEYou cannot connect angled non-contact fiber end connectors with orange sleeves directly
to the instrument.
See “Accessories” on page 25 for further details on connector interfaces and accessories.
Agilent 8157xA Variable Optical Attenuator modules are available in various configurations
for the best possible match to the most common applications.
This chapter describes the options and accessories available for the:
• Agilent 81570A, 81571A VOA modules,
• Agilent 81578A VOA modules for multimode applications; and the
• Agilent 81576A and 81576A VOA modules with power control.
Figure 8 shows all the options that are available for Agilent 8157xA Variable Optical
Attenuator modules, and the instruments that support these modules.
Agilent 8164A or B
Lightwave Measurement System
Optical Attenuators for
Straight Contact Interface
CAUTIONOptical power levels above 100 mW applied to single mode connectors can easily damage
the connector if it is not perfectly clean. Also, scratched or poorly cleaned connectors can
destroy optical connectors mechanically. Always make sure that your optical connectors are
properly cleaned and unscratched before connection. Refer to chapter “Cleaning
Information” on page 77 on appropriate procedures for connector cleaning and inspection.
However, Agilent Technologies Deutschland GmbH assumes no responsibility in case of an
operation that is not compliace with the safety instructions as stated above.