Agilent 81111A Data Sheet

5
5Agilent 81110A/’11A
Performance Test
1
Agilent 81110A/’11A Performance Test
Introduction
Use the tests in this chapter if you want to check that the Agilent 81110A Pulse Generator Frame with the Agilent 81111A 165 MHz Output Channel(s) is working correctly. Before starting any testing allow all test equipment to warm up for at least 30 minutes.
Conventions Used
When referring to actions that you perform during the tests, the following conventions are used:
FUNCTION This indicates that a labelled button must be pressed
[MODE/TRG] This shows that a soft-key must be pressed. A soft­key is an unlabelled button whose label is shown on the display, and which can vary according to the job that the button is doing
CONTINUOUS PULSES This is an option shown on the dis­play, and is selected by use of the vernier keys. It is shown in upper or lower case to match the case displayed.
Test Results Tables
Tables for entering the results of the tests are included at the end of this chapter. The tests are numbered and reference numbers for each Test Result (TR) are given in a small table at the end of each test. The reference number shows you where the actual results should be entered in the Test Results Tables.
The Test Results tables at the end of the chapter should be photo­copied, and the Test Results entered on the copies. Then, if the tests need to be repeated, the tables can be copied again.
2 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
If Channel 2 has been fitted to your instrument, make an extra copy of the Test Results tables for entry of the results of tests on that channel. In this case, however, it is not necessary to repeat the Period tests, as these are common to both channels.
Agilent 81110A/’11A Performance Test 3
Agilent 81110A/’11A Performance Test
Recommended Test Equipment and Accessories
The following tables list the recommended test equipment you need to perform all the tests in this chapter. You can use alterna­tive instruments if they meet the critical specifications given. The test set-ups and procedures assume you are using the recom­mended equipment.
Test Equipment Model Critical Specifications
Oscilloscope or
Oscilloscope Agilent 54750A +
Counter
or
Counter Agilent 53132A
Digital Voltmeter Agilent 3458A DCV up to 20 V
Pulse Generator Agilent 8110A up to 150 MHz
Delay line Agilent 54008A 22 ns
Digitizing Oscilloscopes Accessories Attenuators
Power Splitter SMA/SMA (m-m)Adapter SMA/BNC Adapter SMA Cable
Agilent 54121T 20 GHz, 10 bit vertical resolution, Histogram
Agilent 54751A
Agilent 5334B #010, 030
#001/010, 030
Accessories Model Critical Specifications
20 GHz, 15 bit vertical resolution, Histogram
Period and Time Interval measurements Oven Osci, 1.3 GHz C-Channel
Frequency measurements > 150 MHz High-Stability Timebase, 3 GHz Channel
33340C#020 33340C#006 11667B 1250-1159 1250-1700 8120-4948
20 dB 6 dB
4 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
Accessories Model Critical Specifications
50 Feedthrough Termination 10100C
See Figure
Adapter 1251-2277 BNC to Banana
Cable Assemblies, BNC 8120-1839
Torque Wrench 8710-1582 5/16 in, 5 lb-in (56 Ncm)
2 W,1% 10 W,0.1%
NOTE: When you connect the test equipment for the first time, and
whenever you change the setup during the course of these tests, use the 8710 - 1582 torque wrench to tighten and loosen SMA connectors. This will ensure that the connectors are at the correct tightness and give the best signal transfer.
Agilent 81110A/’11A Performance Test 5
Agilent 81110A/’11A Performance Test
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following figure provides a schematic and a parts list except for the case. The case must provide shielding and maintain grounding integrity.
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following parts are required:
1. R1 = 53.6, 1%, 10 W; Part Number: 0699-0146.
2. R2 = 200, 10%, 0.5 W, Variable trimmer; Part Number: 2100-3350.
3. R3 = 681 ;, 1%, 0.5 W; Part Number: 0757-0816.
4. BNC (M): Part Number: 1250-0045.
5. BNC (F): Part Number: 1250-0083.
6 Agilent 81110A/’11A Performance Test
Getting Started
The Agilent 81110A is controlled by selecting options in a series of pages that are displayed on the instrument's screen. These options vary with the boards that are fitted in the instrument. When the Agilent 81110A is being tested, therefore, different situ­ations can arise, depending on whether you have a standard instru­ment or one that has had additional boards fitted. The following examples illustrate this
Typical Examples of Displayed Screens
Agilent 81110A/’11A Performance Test
Per 1.000µs Normal
OFF
1
Delay Width LeadE TraiE
The OUTPUT Screen in a Standard Agilent 81110A
Agilent 81110A/’11A Performance Test 7
0ps
100.0ns
2.00ns =LeadE
OUTPUT PATTERNLIMITSMODE/TRG
Offset Amplit 50Ω into 50.0
+0.0mV
1.00V
*OFF ON
MODIFY
Agilent 81110A/’11A Performance Test
OFF 1.000µs OFF
1
Delay Width LeadE TraiE
The TIMING Screen in an Agilent 81110A with qty 2 of Agilent 81111A
ON Normal Normal ON Seperate Outputs
High Low
50Ω into 50.0
Per
0ps
100.0ns
2.00ns =LeadE
TIMING PATTERNLEVELSMODE/TRG
+2.50V +0.0mV
2
Delay 0ps Width LeadE TraiE
100.0ns
2.00ns =LeadE
21
High +2.50V Low +0.0mV 50Ω into 50.0
MODIFY
*Period Frequency
MODIFY
Set TTL *High-Low Offs-Ampl Set ECL
TIMING PATTERNLEVELSMODE/TRG
The LEVELS Screen in an Agilent 81110A with qty 2 of Agilent 81111A
8 Agilent 81110A/’11A Performance Test
Instrument Serial Numbers
You will need to write the serial numbers of the instrument at the top of the Test Reports. These can be found as follows:
Press HELP, [SERIAL #]
The Agilent 81110A display lists the instrument's products and serial number.
The display on your instrument should look similar to this:
Agilent 81110A/’11A Performance Test
FRAME : Serial No :
81110A 165 MHz DE38700135
OUTPUTS Ch1-Bd. : Ch2-Bd. :
81111A 81111A
The serial number given for the FRAME applies to the Mainframe, the Power Supply, the Microprocessor Board, and the Timing Board. The number(s) available of the Output Channel(s) applies to the installed numbers of outputs and Model Number.
Agilent 81110A/’11A Performance Test 9
Agilent 81110A/’11A Performance Test
Initial Setup of the Agilent 81110A
In the majority of these tests the initial setting up of the instrument is identical. Therefore, it is described once here, and then referred­to where appropriate. In cases where the initial setup differs, an illustration of the settings is shown.
Set up the Agilent 81110A as follows:
1. Select [MODE/TRG]
• CONTINUOUS PULSES
Single-Pulses at Out 1 (plus Single-Pulses at Out 2, if
second channel is installed
• Pulse-Period:internal Osc
2. If a second output channel is installed, select MORE [CONFIG] screen and set up as follows:
GPIB Address: 10 Perform: Timing Calibration
Group Params by: OUTPUT 1 / 2
PLL-Ref : Internal
TRG-LEV CONFIGMEMCARDLIMITS
CONFIG Screen, Parameters grouped by OUTPUT
10 Agilent 81110A/’11A Performance Test
MODIFY
Tim/Lev *Out 1/2
Agilent 81110A/’11A Performance Test
NOTE: Set-ups are given in all the tests for [OUTPUT 1] and [OUTPUT 2].
If you are testing a single channel instrument set up the [OUTPUT] screen with the settings given for [OUTPUT 1].
Agilent 81110A/’11A Performance Test 11
Agilent 81110A/’11A Performance Test
Test 1: Period (PLL not active)
Test Specifications
Range 6.06 ns to 999.5 s Resolution 3.5 digits, best case 5 ps
Accuracy +3% typical +0.5% after selfcal
Equipment Needed
Counter Cable, 50 , coaxial, BNC
Procedure
1. Connect the Agilent 81110A to the Counter as shown:
81110A UNDER TEST
81110A
165/330 MHz
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
5334B Counter
Connecting the Agilent 81110A to the Counter
2. Set up the Agilent 81110A as described in "Initial Setup of the Agilent 81110A"
12 Agilent 81110A/’11A Performance Test
INPUT A C
Agilent 81110A/’11A Performance Test
On the Agilent 81110A press MORE and set up [OUTPUT 1] and [OUTPUT 2] pages as shown in the following illustrations:
Per Normal ON
Delay DtyCyc LeadE TraiE
Configuring Output 1
Per Normal OFF
Delay2 0ps DtyCyc LeadE TraiE
6.060ns
0ps
50.00%
2.00ns =LeadE
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
6.060ns
50.00%
2.00ns =LeadE
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
Offset Amplit 50Ω into 50.0
Offset Amplit 50Ω into 50.0 Separate Out2
1
+0.0mV
1.00V
+0.0mV
1.00V
MODIFY
6.060
ns
MODIFY
6.060
ns
Configuring Output 2
Agilent 81110A/’11A Performance Test 13
Agilent 81110A/’11A Performance Test
NOTE: When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels. b. For the Period test you can switch OFF the channels that are not being tested.
3. Set the Counter to:
FUNCTION Period A / Freq C INPUT A
SENSE
50 Οn
4. Check the Agilent 81110A period at the following settings:
Period Acceptable Range TR entry
6.060 ns
9.990 ns
10.00 ns
50.00 ns
99.90 ns
100 ns 500 ns 1 µs 500 µs 500 ms
without selfcal!
5.878 ns to 6.242 ns
9.690 ns to 10.290 ns
9.7 ns to 10.3 ns
48.5 ns to 51.5 ns
96.903 ns to 102.897 ns
97 ns to 103 ns 485 ns to 515 ns 970 ns to 1030 ns 485µs to 515 µs 485 ms to 515 ms
1 - 1 1 - 2 1 - 3 1 - 4 1 - 5
1 - 6 1 - 7 1 - 8 1 - 9 1 - 10
14 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
5. To perform the Timing Calibration (shown as selfcal) set up [CONFIG] page as shown in the following illustration:
HP-IB Address: 10
Perform: Timing Calibration
Group Params by: OUTPUT 1 / 2 PLL-Ref : Internal
TRG-LEV CONFIGMEMCARDLIMITS
Selftest *Calibrate
6. Press ENTER and wait till the display shows TIMING CALIBRATION PASSED and gets back to the above shown display .
Test 2: PLL Period
NOTE: This test is only performed if PLL is switched on.
Test Specifications
Range 6.06 ns to 999.5 s Resolution 4 digits, best case 1 ps
Accuracy + 0.01%
MODIFY
Equipment Needed
Counter Agilent 53132A
Agilent 81110A/’11A Performance Test 15
Agilent 81110A/’11A Performance Test
Cable, 50 , coaxial, BNC
NOTE: The Agilent 53132A counter is used in frequency mode to meet
the MIL CAL A uncertainty requirements for TAR (Test Accuracy Ratio) > 4:1.
Procedure
Connect the Agilent 81110A to the counter as follows:
81110A UNDER TEST
165/330 MHz
81110A
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
53132A Counter
3
CHANNEL 1 2
Connecting Agilent 81110A to the Counter
7. Set up the Agilent 81110A as described in "Initial Setup of the Agilent 81110A"
8. Select the [MODE/TRG] screen on the Agilent 81110A and set up as follows:
16 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
CONTINUOUS PULSES
MODIFY
Single-Pulses at Out1 Single-Pulses at Out2
Pulse-Period: internal PLL
int. OSC *int. PLL CLK-IN
TIMING PATTERNLEVELSMODE/TRG
The MODE/TRG Screen Setup
9. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2] pages as shown in the test before!
Agilent 81110A/’11A Performance Test 17
Agilent 81110A/’11A Performance Test
NOTE: When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels. b. For the Period test you can switch OFF the channels that are not being tested.
10. Set the Counter to measure the frequency at the choosen input 1 / 3
11. Check the Agilent 81110A PLL pulse period at the following settings:
Period Frequency Acceptable Range TR Entry
6.061 ns
10.00 ns
50.00 ns 100 ns 500 ns 1 µs 50 µs 5 ms 500 ms 5 s
165.000MHz 100 MHz 20 MHz 10 MHz 2 MHz 1 MHz 20 kHz 200 Hz 2 Hz
0.2 Hz
164.9835 MHz to 165.0165 MHz
99.990 MHz to 100.010 MHz
19.9980 MHz to 20.0020 MHz
9.9990 MHz to 10.0010 MHz
1.9998 MHz to 2.0002 MHz
999.9 kHz to 1.0001 MHz
9.998 kHz to 20.002 kHz
199.980 Hz to 200.020 Hz
1.9998 Hz to 2.0002 Hz
0.19998 Hz to 0.20002 Hz
18 Agilent 81110A/’11A Performance Test
2 - 1 2 - 2 2 - 3 2 - 4 2 - 5 2 - 6 2 - 7 2 - 8 2 - 9 2 - 10
Test 3: Width
Test Specifications
Range 3.03 ns to (period - 3.03 ns) Resolution 3.5 digits, best case 5 ps
Accuracy + 3% + 250 ps typical + 0.5% + 250 ps after selfcal
Equipment Needed
Digitizing Oscilloscope with Accessories Counter Cable, 50 , coaxial, BNC
Procedure
1. Connect Agilent 81110A to the Scope as shown:
Agilent 81110A/’11A Performance Test
81110A UNDER TEST
81110A
165/330 MHz
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
OUT 1
54121T Frontend
INPUT 1 2 3 4 TRIG
6 dB
Attenuator
with SMA/BNC
Adapter
20 dB
Connecting Agilent 81110A to the Scope
2. Set up the Agilent 81110A as described in "Initial Setup of the Agilent 81110A"
Agilent 81110A/’11A Performance Test 19
Agilent 81110A/’11A Performance Test
3. On the Agilent 81110A press MORE and set up [OUTPUT 1] and [OUTPUT 2] pages as shown in the following illustrations:
Per 200 ns Normal ON
Delay
LeadE TraiE
Configuring Output Screen 1
Per 200 ns Normal OFF
Delay Width LeadE TraiE
0ps
100.0nsWidth
2.00ns =LeadE
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
0ps
3.030ns
2.00ns =LeadE
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
Offset Amplit 50Ω into 50.0
Offset Amplit 50Ω into 50.0 Separate Out2
1
+0.0mV
1.00V
2
+0.0mV
1.00V
MODIFY
100.0
ns
MODIFY
3.030
ns
Configuring Output Screen 2
20 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
NOTE: When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels. b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the other channel.
4. Set the Digitizing Oscilloscope Agilent 54121T:
Press AUTOSCALE
Select the Display menu and set the Number of Averages to 32
Select the delta V menu and turn the voltage markers On
Set the preset levels to 50% -50% and press AUTO LEVEL SET
Select the delta t menu and turn the time markers ON
Set START ON EDGE = POS 1 and STOP ON EDGE = NEG1
5. Change the oscilloscope timebase to 1 ns/div
6. Change the Agilent 81110A Ch-1 Width to 3.03 ns
7. Center the pulse in the Scope display
8. Press the PRECISE EDGE FIND key for each new Width set­ting
Agilent 81110A/’11A Performance Test 21
Agilent 81110A/’11A Performance Test
9. Check the Agilent 81110A pulse width at the following set­tings:
Oscilloscope
Timebase
1 ns/div 1 ns/div 2 ns/div 10 ns/div 20 ns/ 100 ns
Period Width Acceptable Range TR Entry
without selfcal!
200 ns 200 ns 200 ns 200 ns 1 µs 1 µs
3.030 ns
6.060 ns
10.00 ns
50.00 ns
100.0 ns
500.0 ns
2.689 ns to 3.371 ns
5.528 ns to 6.492 ns
9.450 ns to 10.55 ns
48.25 ns to 51.75 ns
96.75 ns to 103.25 ns
484.75 ns to 515.25 ns
10. Connect the Agilent 81110A to the Counter as shown:
81110A UNDER TEST
81110A
165/330 MHz
S
Agilent
PULSE-/ PATTERN GENERATOR
OUT 1
5334B Counter
INPUT A
3 - 1 3 - 2 3 - 3 3 - 4 3 - 5 3 - 6
Connecting Agilent 81110A to the Counter
11. Set the Counter to:
22 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
FUNCTION SENSE INPUT A COM A INPUT B
TI A→ B On 50 On 50 , negative slope
12. Check the Agilent 81110A width at the following settings:
Period Width Acceptable Range TR Entry
100 µs 10 ms 999 ms
NOTE: Repeat the entire test for the second channel, if it is installed
50 µs 5 ms 500ms
48.5 µs to 51.5 µs
4.85 ms to 515ms 485 ms to 515 ms
3 - 7 3 - 8 3 - 9
Agilent 81110A/’11A Performance Test 23
Agilent 81110A/’11A Performance Test
Test 4: Delay
Test Specifications
Range Fixed typical Delay of
EXT INPUT to TRIGGER OUT 12 ns TRIGGER OUT to OUTPUT 1/2 14 ns Variable Delay: 0 ns to period - 3.03ns
Resolution Accuracy
Equipment Needed
Digitzing Oscilloscope with Accessories Pulse Generator Counter Cable, 50 , coaxial, BNC
Procedure
Connect Agilent 81110A to the Scope as shown:
3.5 digits, best case 5 ps ±3% ±0.5 ns typical ±0.5% ±0.5 ns after selfcal
24 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
81110A UNDER TEST
165/330 MHz
81110A
S
Agilent
PULSE-/ PATTERN GENERATOR
8110A Pulse Generator
TRIG
OUT 1
OUT
EXT.
TRIG
OUT 1
OUT
IN
54121T Frontend
INPUT 1 2 3 4 TRIG
Attenuator
3 x 20 dB
with SMA/BNC
Adaptor
Connecting Agilent 81110A to the Scope
13. Set up the Agilent 81110A as described in "Initial Setup of the Agilent 81110A"
14. Set the Pulse Generator to:
Period Width Amplitude Offset Output
1 µs 100 ns 1 V +1.0 V Enable
15. Select the [MODE/TRG] screen on the Agilent 81110A and set up as follows:
Agilent 81110A/’11A Performance Test 25
Agilent 81110A/’11A Performance Test
PULSES
TRIGGERED
Single-Pulses at Out1 Single-Pulses at Out2
Trg'd by: EXT-IN
The TRG MODE Screen Setup
16. On the Agilent 81110A press MORE and set up [TRIG-LEV] page as shown:
MODIFY
Continous *Triggered Gated Ext-Width
TIMING PATTERNLEVELSMODE/TRG
+1.0V 50
EXT-IN: Threshold
CLK-IN: Threshold +1.0V 50
TRIGGER-OUT: TTL STROBE-OUT : TTL
TRG-LEV CONFIGMEMCARDLIMITS
The TRG-LEV Screen Setup
26 Agilent 81110A/’11A Performance Test
Ω Ω
Set TTL Set ECL *Voltage
MODIFY
Agilent 81110A/’11A Performance Test
17. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2] pages as shown in the following illustrations:
Per -------- Normal ON
Delay Width LeadE TraiE
Configuring Output Screen 1
Per -------- Normal OFF
Delay Width LeadE TraiE
0ps
100ns
2.00ns =LeadE
0ps
100ns
2.00ns =LeadE
Offset Amplit 50Ω into 50.0
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
Offset Amplit 50Ω into 50.0 Separate Out2
1
+0.0mV
1.00V
2
+0.0mV
1.00V
MODIFY
0
ps
MODIFY
0
ps
OUTPUT 1 PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 2
Agilent 81110A/’11A Performance Test 27
Agilent 81110A/’11A Performance Test
NOTE: When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels. b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the other channel.
Set the Digitizing Oscilloscope Agilent 54121T:
Press AUTOSCALE
Set timebase to TIME/DIV = 10 ns/div
Center the positive-going edges of the two signals
Select the Display menu and set the screen function to single; set the number of averages to 32
Select the Delta V menu and turn the voltage markers ON and assign marker 1 to channel 3 and marker 2 to channel 4
Set Preset levels to 50% - 50% and press AUTO LEVEL SET
Select the Delta t menu and turn the time markers ON
Set START ON EDGE= POS1 and STOP ON EDGE= POS 1
Press the PRECISE EDGE FIND key
18. Check the Agilent 81110A delay at the following settings:
28 Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
NOTE: Record the value of the fixed delay and subtract it from the other
readings.
Oscilloscope Timebase Delay Acceptable Range TR Entry
10 ns/div
10 ns/div 20 ns/div 20 ns/div 50 ns/div 200 ns/div
0 ps
without selfcal!
5.000 ns
10.00 ns
50.00 ns
100.0 ns
500.0 ns
fixed Delay of TRIG OUT to OUT 1/2: 14 ns typ.
4.35 ns to 5.65 ns
9.200 ns to 10.80 ns
48.00 ns to 52.00 ns
96.50 ns to 103.50 ns
484.50 ns to 515.50 ns
4 - 1
4 - 2 4 - 3 4 - 4 4 - 5 4 - 6
19. Connect the Agilent 81110A to the Counter as follows:
81110A UNDER TEST
165/330 MHz
81110A
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
OUT 1
5334B Counter
INPUT A INPUT B
Connecting Agilent 81110A to the Counter
20. Set Agilent 81110A to Continuous-Pulses on the MODE/TRG screen
21. Set the Counter to:
Agilent 81110A/’11A Performance Test 29
Agilent 81110A/’11A Performance Test
FUNCTION TI SENSE INPUT A INPUT B
A → B On 50 50
22. Check the Agilent 81110A delay at the following settings:
NOTE: Subtract the fixed delay from the other readings
Period Delay Acceptable Range TR Entry
100 µs 10 ms 999 ms
NOTE: Repeat the entire test for the second channel, if it is installed.
50 µs 5 ms 500ms
48.5 µs to 51.5 µs
4.85 ms to 515ms 485 ms to 515 ms
4 - 7 4 - 8 4 - 9
30 Agilent 81110A/’11A Performance Test