Agilent 8012A Data Sheet

Infiniium 54800 Series Oscilloscope Probes, Accessories, and Options
Selection Guide Data Sheet
Table of Contents
Probe Compatibility Table ................................................................2
General Purpose Probes Overview ................................................3
116XA Family Passive Voltage...................................................4
High-Frequency Voltage Probes Overview ..................................9
1130A/31A/32A/34A InfiniiMax High-Performance
Active Probe System..................................................................10
1168A/69A InfiniiMax II High-Performance
Active Probe System..................................................................13
1155A Active Single-Ended Voltage........................................16
1156A/57A/58A High-Bandwidth, Active Voltage ..............18
1163A Resistive Divider...............................................................4
54006A Resistor Divider Probe ................................................21
Differential Voltage Probes Overview.........................................22
1130A/31A/32A/34A InfiniiMax High-Performance
Active Probe System..................................................................10
1168A/69A InfiniiMax II High-Performance
Active Probe System..................................................................13
1153A Low-Bandwidth Active Differential Voltage..............23
Current Probes Overview...............................................................26
1146A ac/dc Current .................................................................27
1147A High Bandwidth ac/dc Current ...................................29
High-Voltage Passive Probes Overview .....................................31
10076A (100:1)............................................................................32
N2771A (1000:1).........................................................................33
Mixed-Signal Oscilloscope Logic Probe Kit..............................34
E5396A Half-Size Soft Touch Connectorless Logic Probe.......35
Probing Accessories.......................................................................36
Wedge Adapters.........................................................................36
0.5 mm IC Clips, PC Board Mini-Probe Sockets ....................37
EZ-Probe Positioner ...................................................................38
Logic Analyzer/Oscilloscope Time Correlation........................39
VoiceControl .....................................................................................40
Input Devices and Storage Devices.............................................41
E2610A Keyboard, E2609B Rackmount Kit,
1184A Testmobile.......................................................................41
Infiniium Advanced Analysis Options .........................................42
Support, Services, and Assistance..............................................43
To ensure that you have the tools for dependable oscilloscope measurements, Agilent Technologies offers a wide range of oscilloscope probes and accessories. Each is designed for a specific measurement need because the physical and electrical quality of the connection can make the difference between a good measurement and a bad one.
For ordering information when replac­ing your probe or probe accessory:
Refer directly to the page number listed in the table of contents for your probe model.
To assist you in selecting the proper probe for your application:
Use our probe compatibility table below to find the probes that are recommended for use with your Infiniium scope.
Or refer to our probe overview page at the beginning of each section in the table of contents explaining what the different probe types are and the models available for your Infiniium.
Probe Compatibility Table
Agilent Infiniium Oscilloscope Model
Probe Type Probe Model 54830B/31B/32B/33A/ 54835A/45A/ 54810A/15A/ 54852A/53A/54A/55A
30D/31D/32D/33D 45B/46A/46B 20A/25A and DSO80000 Series
General Purpose Passive 1160A, 10:1 Incompatible Incompatible Recommended Incompatible Page 4
1161A, 10:1 Compatible Recommended Incompatible Compatible [1]
1162A, 1:1 Recommended Recommended Recommended Compatible [1]
1164A, 10:1 Incompatible Incompatible Recommended Incompatible
1165A, 10:1 Recommended Compatible Incompatible Incompatible
Low Mass Passive 1171A, 10:1 Recommended Recommended Incompatible Compatible [1] Page 7
Active Single-Ended 1155A, 750 MHz Recommended Compatible Recommended Compatible Page 10, 16, 18
1156A, 1.5 GHz Recommended Recommended Recommended Compatible
1157A, 2.5 GHz Compatible Recommended Compatible Compatible
1158A, 4 GHz Compatible Recommended Compatible Compatible
Resistive Divider 1163A, 10:1 Recommended Recommended Recommended Incompatible Page 4, 21
54006A, 10:1 or 20:1 Compatible [2] Recommended [2] Compatible [2] Recommended [2]
Active Differential 1153A, 200 MHz Recommended Recommended Recommended Compatible Page 10, 13, 23
1130A, 1.5 GHz [6] Recommended [3] Compatible Incompatible Compatible
1131A, 3.5 GHz [6] Recommended [3] Recommended [4] Incompatible Recommended
1132A, 5 GHz [6] Compatible [3] Compatible [4] Incompatible Recommended
1134A, 7 GHz [6] Compatible [3] Compatible [4] Incompatible Recommended
1168A, 10 GHz [6] Compatible [8] Incompatible Incompatible Recommended [8]
1169A, 12 GHz [6] Compatible [8] Incompatible Incompatible Recommended [8]
Current 1146A, 100 kHz Recommended Recommended Recommended Compatible [1] Page 27, 29
1147A, 50 MHz Recommended Recommended Recommended Incompatible [7]
High Voltage 10076A, 4 kV Recommended Recommended Recommended Compatible [1] Page 32, 33
N2771A, 15 kV Recommended Recommended Recommended Compatible [1]
Mixed-Signal Oscilloscope 54826-68701 Recommended [5] Incompatible Incompatible Incompatible Logic Probe Kit Page 34
Half-Size Soft Touch E5396A Recommended [5] Incompatible Incompatible Incompatible Connectorless Logic Probe Page 35
Table 1.1. Agilent probes compatible with Infiniium oscilloscopes.
2
[1] Requires E2697A 1 Minput adapter. The E2697A includes one 10073C
general-purpose 500 MHz, 10:1 passive probe. [2] Requires E2695A SMA-BNC adapter. [3] Requires 54830 Series system software revision A.03.10 or higher. For A.02.xx or
lower, order N5383A to upgrade system software. [4] Requires 54845/46 system software revision A.04.50 or higher. [5] Recommended for 54830D/31D/32D/33D mixed signal oscilloscopes only.
[6] Each 113XA and 1168A/9A probe amplifier supports both single-ended and
differential measurements.
[7] Order N2774A 50 MHz current probe with N2775A power supply that also requires
the E2697A 1 Minput adapter.
[8] Requires 54830 Series, 54850 Series, DSO80000 Series system software revision
A.03.90 or higher. For A.02.xx or lower, order N5383A to upgrade system software.
General Purpose Probes Overview
Passive voltage probes are used for general purpose probing and are made with passive compo­nents only such as wires, con­nectors, capacitors and resistors (when attenuation is required). There are no active components such as transistors or amplifiers in the probe, and therefore pas­sive voltage probes do not need to be powered.
Model Probe Type Applications and Use Page
116XA Family Passive voltage General purpose, Infiniium replacement probes 4 (excluding the 1163A)
117XA Family Passive voltage Fine-pitched ICs, surface mount devices, and dense circuit boards 7
Passive Voltage Probe Advantages Limitations
116XA are rugged, economical, easy to use, 600 MHz maximum bandwidth have high dynamic range and high input resistance Higher capacitive loading than active probes
117XA feature low mass, low weight, low tip 500 MHz bandwidth capacitance compared to other passive probes Not as rugged as 116XA for browsing
Figure 2.1. Agilent 116XA standard Infiniium replacement probe.
Figure 2.2. Agilent 117XA low mass probe for fine-pitched ICs, SMDs, and dense circuit boards.
General purpose probes are available with attenuation ratios of 1:1, 10:1, and 20:1. The 10:1 passive voltage probe is the most commonly used probe, and is supplied as a standard accessory with all Infiniium oscilloscopes having bandwidths 600 MHz.
3
Built for Reliability
These general purpose replace­ment devices are built and tested for high reliability. Kevlar strengthener has been added to the probe cable for extra pull strength. Durable probe tips are replaceable.
The compact design significantly reduces the problem of probing densely populated integrated cir­cuit components or the character­istically minute conductors on printed circuit boards. These small lightweight probes allow measurements that were previ­ously quite difficult, while reduc­ing the danger of shorting. For tight probing areas, the probe handle can be unscrewed and pulled back along the cable.
When probing about the circuit in debug mode, the probes easily slip inside the included browsers. The browsers feature a crown
General Purpose Probes
Agilent 116XA Family Passive Voltage
• Standard replacement probes
for Infiniium scopes with bandwidths 600 MHz
• Rugged, economical, easy to use
• Compact design, removable probe
handle for tight probing areas
• Agilent 1163A, 500 Ω resistive
divider, 10:1 attenuation
Figure 2.3. Agilent 116XA family general purpose replacement probes.
point that digs into solder and avoids the danger of slipping off the test point and shorting to adjacent leads. A pogo pin allows hand movement on the probes without losing contact with the device under test.
The 116XA family probes are compatible with the AutoProbe interface, which completely con­figures the Infiniium oscilloscope for the probe. A snap-on BNC connector simplifies attaching the probe to the scope. Leads are
available for connecting to a wide variety of test points. See “Ordering Information” for a complete list.
Agilent 1163A Resistive Divider
Agilent 1163A features low capac­itive loading and wide bandwidth, resulting in very accurate timing measurements. Resistive divider probes are useful for probing low­voltage signals such as ECL cir­cuits, 50 transmission lines, and GaAs circuits.
Figure 2.4. No-slip browser crown point.
4
Specifications
System Model Bandwidth Division Scope Compensation Number Type of Probe (scope +probe) Ratio Input R Input C Input R Range Length
1160A High Impedance, 500 MHz 10:1 10 M 9 pF 1 MΩ 6 - 9 pF 1.5 m
Passive
1161A High Impedance, 500 MHz 10:1 10 M 10 pF 1 MΩ 12 - 14 pF 1.5 m
Passive
1162A High Impedance, 25 MHz 1:1 1 M 50 pF + scope 1 M n/a 1.5 m
Passive capacitance
1163A 500 Ω Resistive 1.5 GHz with 10:1 500 Ω 1.5 pF 50 Ω n/a 1.5 m
Divider scope model
54845A/B
1164A High Impedance, 500 MHz 10:1 10 M 10.5 pF 1 M 6 - 9 pF 2.0 m
Passive
1165A High Impedance, 600 MHz 10:1 10 M 10 pF 1 MΩ 12 - 14 pF 1.5 m
Passive typical with
54830B/31B/32B/33A
54830D/31D/32D/33D
Operating Characteristics
Approximate propagation delay 6.7 ns for 1160A/61A/62A/63A/65A probes
8.8 ns for 1164A probe
Maximum input voltage 300 V (dc + peak ac), CAT II
10 V (dc + peak ac), CAT I for 1163A only
Safety Meets IEC1010-2-31
Pulling strength (BNC to barrel) 12 lb static pull
Net weight 2.6 oz
Environmental Characteristics
Temperature (operating) 0° C to +55° C
Humidity (operating) Up to 95% relative humidity at 40° C
Altitude (operating) Up to 4,600 meters (15,000 ft.)
Shock 50 g (400 g tip only)
General Purpose Probes
Agilent 116XA Family Passive Voltage
5
Ordering Information
Probes and Accessories
Part # Description Quantity
1160A 10:1, 10 MΩ, 1.5 m, miniature passive probe 1
1161A 10:1, 10 MΩ, 1.5 m, miniature passive probe 1
1162A 1:1, 1.5 m, miniature passive probe 1
1163A 10:1, 500 Ω, low C, 1.5 m, miniature passive probe 1
1164A 10:1, 10 MΩ, 2 m, miniature passive probe 1
1165A 10:1, 10 MΩ, 1.5 m, miniature passive probe 1
5063-2143 Probe tip to BNC (m) 1
IC clips: See “Probing Accessories”
Horizontal and vertical mini-probe sockets: See “Probing Accessories”
Wedge Probe Adapters: See “Probing Accessories”
Replacement Parts
Part # Description Quantity
5063-2135 General purpose retractable hook tip 2
5063-2140 Alligator ground lead 2
5063-2120 Socketed ground lead 1
5063-2115 Browser 1
5063-2147 Dual lead adapter 1 0
5063-2149 SMD clips 5
01160-68701 Accessory kit (includes four spring grounds, 1
four browser pogo pins, four barrel insulators, one screwdriver)
5063-2136 1160A probe tip, red 5
5063-2137 1161A probe tip, brown 5
5063-2138 1162A probe tip, black 5
5063-2139 1163A probe tip, grey 5
5063-2151 1164A probe tip, orange 5
5063-2137 1165A probe tip, brown 5
One probe
Two probe tips
Two barrel insulators
One browser
One general­purpose
retractable
hook tip
One
dual-lead
adapter
Two SMD clips
Four spring grounds
One alligator ground lead
One socketed ground lead
One screwdriver
One browser pogo pin (spare)
Probe Parts Supplied
Includes user’s guide and three-year warranty.
General Purpose Probes
Agilent 116XA Family Passive Voltage
6
Suited for Today’s ICs
An exceptionally small and light probe tip (<1 gram) and an ultra thin cable make the 117XA family ideal for connecting to and prob­ing fine-pitch ICs, surface mount devices, and dense circuit boards.
The probe fits directly onto stan­dard board headers and IC clips. A range of accessories are avail­able for you to optimize interfac­ing with surface mount devices. Wedge Probe Adapters make con­venient and reliable connections to TQFP/PQFP package leads. See “Ordering Information” for a com­plete list of accessories.
General Purpose Probes
Agilent 117XA Family Low Mass, Passive Voltage
• Easy connection to fine-pitch ICs, SMDs, and dense circuit boards
• Light weight (< 1 gram), low mass probe tip
• 10:1 attenuation, capacitance < 10 pF (1171A)
Figure 2.5. Agilent 117XA family probes for fine-pitch ICs, surface mount devices, and dense circuit boards.
When probing about the circuit in debug mode, the probe easily slips inside the included browsers. The browsers feature a crown point that digs into solder and avoids the danger of slipping off the test point and shorting to adjacent leads. A pogo pin allows hand movement on the probes without losing contact with the device under test.
The 117XA probes are compatible with the AutoProbe interface, which completely configures the Infiniium oscilloscope for the probe. A snap-on BNC connector simplifies attaching the probe to the scope.
Specifications
System Model Bandwidth Division Scope Compensation Number Type of Probe (scope + probe) Ratio Input R Input C Input R Range Length
1171A High Impedance, 500 MHz 10:1 10 M 10 pF 1 MΩ 12 - 14 pF 1.4 m
Passive
7
Specifications
Operating Characteristics
Approximate propagation delay 1171A: 6.5 ns
Maximum input voltage 40 V (dc + peak ac), CAT I
Safety Meets IEC1010-2-31
Pulling strength (BNC to probe tip) 12 lb static pull
Net weight 2.6 oz
Probe tip weight < 1 gram
Environmental Characteristics
Temperature (operating) 0° C to +55° C
Humidity (operating) Up to 95% relative humidity at 40° C
Altitude (operating) Up to 4,600 meters (15,000 ft.)
Shock 50 g (400 g tip only)
Ordering Information
Probes and Accessories
Part # Description Quantity
1171A 10:1, 10 MΩ, 1.4 m, low mass passive probe 1
IC clips: See “Probing Accessories”
Wedge Probe Adapters: See “Probing Accessories”
Replacement Parts
Part # Description Quantity
5063-2122 Browser 1
Probe Parts Supplied
Includes two IC clips, user’s guide, and one-year warranty.
Two pogo pins
One alligator ground lead
One socketed ground lead
Two probing pins
One ground extender
One probe
One browser
One screwdriver
One walking stick ground
General Purpose Probes
Agilent 117XA Family Low Mass, Passive Voltage
8
High-Frequency Voltage Probes Overview
Active voltage probes contain an active component, usually a field­effect transistor (FET), and there­fore need to be powered. A FET input has the advantage of pro­viding a very low input capaci­tance, typically from less than 1 pF to a few pF. This low capaci­tance results in a high input impedance on frequencies up to 7 GHz. With such low loading, active probes can be used on high-impedance circuits that would be seriously loaded by passive probes.
New InfiniiMax probe family.
The innovative InfiniiMax probe system provides either differential­or single-ended probing solutions for the most demanding connec­tion requirements, without sacri­ficing performance.
Figure 3.1. Agilent 1130A/31A/32A/34A InfiniiMax high-performance active probe.
Figure 3.2. Agilent 1156A/57A/58A active probes for signals up to 4 GHz.
Model Probe Type Applications and Use Page
1130A/31A/32A/34A Active Diff/SE Measure both differential and single-ended signals up to 7 GHz 10
1155A Active Single-Ended Measure fast transitions on low-voltage signals, 2 channels 16
1156A/57A/58A Active Single-Ended Measure fast transitions on low-voltage signals up to 4 GHz 18
1163A Resistive divider Measure fast transitions on a wide range of signal voltages 4
54006A Resistive divider Low-cost alternative for high-frequency probing 21
High-Frequency Probe Advantages Limitations
Timing and voltage measurements more accurate at Active probes are more expensive than general purpose passive probes high bandwidths
Resistive divider probes cost less than active probes Relatively heavy resistive loading with resistive divider probes
Active probes are least intrusive to circuit under test, Active probes have lower dynamic range, lower maximum voltage and high input resistance are less rugged compared to passive probes
Resistive divider probes are passive probes. They feature low capacitive loading and accurate timing measurements with high­bandwidth signals at a much lower cost than active probes.
9
High-Frequency Active Differential/Single-Ended Probe System
Agilent 1130A/31A/32A/34A InfiniiMax High-Performance Active Probe System
• The Agilent InfiniiMax 1130A series probe system supports
a wide variety of real-world applications with an exten­sive line up of probe heads and accessories. The accessories
can meet the most demanding mechanical access requirements. Small probe heads can be placed between densely packed PC boards. Solder-in sockets are available for signals that need frequent measurement. A dif­ferential SMA probe head is available to connect to fixtures that have SMA connections. A smart ergonomic design allows users to set the spacing between the probe pins (variable span). When not concerned with min­imum probe size, designers can use a browsing sleeve to make holding the probe more com­fortable. Both probe tips of the differential probe can “flex” to support various probing angles and target system characteristics (z-axis compliance). Innovative damped-wire accessories com­pensate for the inductance and capacitance associated with the leads, and prevent distor­tion of the measured signal.
• InfiniiMax 7 GHz, 5 GHz, 3.5 GHz, and 1.5 GHz probing system
• Each InfiniiMax probe amplifier supports both differential and single-ended measurements for a more cost-effective solution
• Unrivaled InfiniiMax probing accessories support browsing, solder-in, and socket use models at the maximum performance available
The Agilent InfiniiMax 1134A, 1132A, 1131A and 1130A probe systems provide 7 GHz, 5 GHz,
3.5 GHz and 1.5 GHz of band-
width respectively, and offer the following benefits:
• The new probes have a flat
frequency response over the entire bandwidth specification,
eliminating the distortion and loading that affect probes with in-band resonance. The probing system enables engineers to utilize their oscilloscope’s entire bandwidth without being limited to measuring only 50 trans­mission lines or using passive resistive divider probes that produce voltage measurement error and circuit loading. Designers can achieve system measurement bandwidths of
4.5 to 6 GHz even when manu­ally “browsing” with the probe. Solder-in probe heads and solder-in sockets provide even higher bandwidths.
Figure 3.3. InfiniiMax offers you the highest performance available for measuring differential and single-ended signals.
• The groundbreaking design of Agilent InfiniiMax 1130A probe system also enables users to make either single-ended or
differential measurements from a single probe amplifier,
depending on their choice of probe head and accessory. This can result in significant savings in cost and time. The common mode rejection of the differen­tial probe head reduces a mea­surement’s noise floor. Overall, the Agilent 1130 series probing system delivers unmatched performance, accuracy and connectivity.
InfiniiMax: The World’s Best High-Speed Oscilloscope Probing System
EDN Magazine has awarded Agilent’s InfiniiMax active probe system the 2002 Innovation of the Year Award.
10
High-Frequency Active Differential/Single-Ended Probe System
Agilent 1130A/31A/32A/34A InfiniiMax High-Performance Active Probe System
Specifications
Operating Characteristics
Probe bandwidth* 1134A: > 7 GHz
1132A: > 5 GHz 1131A: > 3.5 GHz 1130A: > 1.5 GHz
Rise and fall time (10% to 90%) 1134A: 60 psec
1132A: 86 psec 1131A: 100 psec 1130A: 233 psec
System bandwidth (–3 dB) 1134A with 54855A: 6 GHz
1132A with 54854A: 4 GHz 1131A with 54853A: 2.5 GHz 1131A with 54852A: 2 GHz 1130A with 54832B/D, 33A/D: 1 GHz
Input capacitance** Cm = 0.1 pF Cm is between tips.
Cg = 0.34 pF Cg is to ground for each tip. Cdiff = 0.27pF Differential mode capacitance = Cm + Cg/2 Cse = 0.44 pF Singe-ended mode capacitance = Cm + Cg
Input resistance* Differential mode resistance = 50 k±1%
Single-ended mode resistance = 25 k±1%
Input dynamic range ±2.5 V
Input common mode range ±6.75 V dc to 100 Hz; ±1.25 V >100 Hz
Maximum signal slew rate 18 V/ns when probing a single-ended signal
30 V/ns when probing a differential signal
DC attenuation 10:1 ± 3% before calibration on oscilloscope
10:1 ± 1% after calibration on oscilloscope
Zero offset error referred < 30 mV before calibration on oscilloscope to input < 5 mV after calibration on oscilloscope
Offset range* ± 12.0 V when probing single-ended
Offset accuracy < 3 % setting before calibration on oscilloscope
< 1 % setting after calibration on oscilloscope
Noise referred to input 3.0 mVrms
Propagation delay ~6 nsec (This delay can be deskewed relative to
other signals.)
Maximum input voltage* 30 Vpeak, CAT I
ESD tolerance > 8 kV from 100 pF, 300 HBM
* Denotes warranted specifications, all others are typical. ** Measured using the probe amplifier and solder-in differential probe head with full bandwidth resistors.
Figure 3.4. Swept frequency response.
Figure 3.5. Common mode rejection vs. frequency.
Example of characterized performance plots: differential solder-in probe head
11
6
3
0
dB
-3
-6
-9
-12
8
10
0
-10
-20
dB
-30
-40
-50
-60
8
10
9
10
Frequency (Hz)
9
10
Frequency (Hz)
10
10
10
10
High-Frequency Active Differential/Single-Ended Probe System
Agilent 1130A/31A/32A/34A InfiniiMax High-Performance Active Probe System
Ordering Information
Probe Amplifier Model
Part # Description Quantity
1134A 7 GHz InfiniiMax Probe Amplifier (order one or more probe heads or connectivity kits per amplifier). 1
1132A 5 GHz InfiniiMax Probe Amplifier (order one or more probe heads or connectivity kits per amplifier). 1
1131A 3.5 GHz InfiniiMax Probe Amplifier (order one or more probe heads or connectivity kits per amplifier). 1
1130A* 1.5 GHz InfiniiMax Probe Amplifier (order one or more probe heads or connectivity kits per amplifier). 1
* Note: Requires 54830 Series system software revision A.03.10 or higher. For A.02.xx or lower, order N5383A to upgrade system software.
Connectivity Kits Model
Part # Description Quantity
E2669A InfiniiMax connectivity kit for differential/single-ended measurements. Includes one 1
differential browser, four solder-in differential probe heads and two socketed differential probe heads. Includes all necessary accessories.
E2668A InfiniiMax connectivity kit for single-ended measurements. Includes one single-ended browser, 1
one solder-in probe heads and one socketed probe heads. Includes all necessary accessories.
Individual Probe Heads
Part # Description Quantity
E2675A InfiniiMax differential browser probe head and accessories. Includes 20 replaceable tips and 1
ergonomic handle. Order E2658A for replacement accessories.
E2676A InfiniiMax single-ended browser probe head and accessories. Includes 2 ground collar 1
assemblies, 10 replaceable tips, a ground lead socket and ergonomic handle. Order E2663A for replacement accessories.
E2677A InfiniiMax differential solder-in probe head and accessories. Includes 20 full bandwidth and 1
10 medium bandwidth damping resistors. Order E2670A for replacement accessories.
E2678A InfiniiMax single-ended/differential socketed probe head and accessories. Includes 48 full 1
bandwidth damping resistors, 6 damped wire accessories, 4 square pin sockets and socket heatshrink. Order E2671A for replacement accessories.
E2679A InfiniiMax single-ended solder-in probe head and accessories. Includes 16 full bandwidth and 1
8 medium bandwidth damping resistors and 24 zero ohm ground resistors. Order E2672A for replacement accessories.
E2695A Differential SMA probe head. Includes semi-rigid coax to change span between SMA connectors. 1
Adapters
Part # Description Quantity
N1022A Adapts 113X/115X active probes to 86100 Infiniium DCA. 1
E2696A General purpose probing solution for use with instruments such as spectrum analyzers and 1
network analyzers. Includes N1022A probe adapter, N1022-60014 connection adapter, and 01143-61602 probe power extension cable. Items that must be purchased separately for a complete probing solution are an InfiniiMax probe amplifier and desired probe heads, and 1143A probe power supply.
12
13
Two new high-bandwidth InfiniiMax II Series probe amplifiers have been added to the InfiniiMax product line. InfiniiMax I probe amplifiers and probe heads can also be used with DSO 80000 Series scopes for lower performance applications.
High-Frequency Active Differential/Single-Ended Probe System
Agilent 1168A/69A InfiniiMax II High-Performance Active Probe System
12 GHz Hi-BW differential SMA probe
head provides maximum bandwidth for SMA fixtured differential pairs.
12 GHz Hi-BW solder-in differential probe head provides maximum bandwidth and
minimizes capacitive loading to 210 fF. Variable spacing from 0.2 to 3.3 mm (8 to 130 mills).
12 GHz Hi-BW differential browser provides maximum bandwidth for hand-held or probe holder use. Variable spacing from 0.2 to
3.3 mm (8 to 130 mills).
InfiniiMax II: The World’s Best High-Speed Probing System Just Got Better
InfiniiMax offers you the highest performance
available for measuring differential and single-ended signals, with flexible connectivity solutions for today’s high-density ICs and circuit boards.
InfiniiMax probes have fully characterized performance for all of their various probe heads.
This includes:
• Swept frequency response plot
• Common mode rejection vs. frequency plot
• Impedance vs. frequency plot
• Time-domain probe loading plot
• Time-domain probe tracking plot
One-year standard warranty on active probes and a variety of Agilent support options to choose from.
Controlled impedance transmission lines in every probe head deliver full performance versus the performance limitations produced by traditional wire accessories.
Probe interface software allows you to save the calibration information for up to 10 different probe heads per channel and will automatically retrieve calibration data for a probe amplifier as it is attached to the scope.
High-input impedance active probes minimize loading, support differential measurements and DC offset, and can compensate for cable loss.
Probe calibration software delivers the most accurate probe measurements, linear phase response and allows various probe combinations to be deskewed to the same reference time.
InfiniiMax II probe heads
Loading...
+ 30 hidden pages