Keysight (Agilent) 4291B Application Note

Correlating Impedance Measurements Among Different Types of Measurement Instruments
Product Note 4291-4
HP4291B RF Impedance/ Material Analyzer
Introduction
As the technology of impedance measuring instruments advances, problems correlating measurements between instruments, instrument configurations, and even instrument platforms become common. The factors affecting the correlation of impedance measurement values and the attention to be paid for achieving higher measurement correlation are studied by making an example of HP 4291B RF impedance/ material analyzer and HP 4191A RF impedance analyzer in this product note.
1
Factors affecting correlation
It is desirable to get the same results when a device is measured by any instrument. However, there is always some difference between measurement instruments. The fundamental difference is a result of the instrument's specified accuracy. The measurement accuracy is specified at the calibration plane of the instrument. This plane is defined by calibration standards, and, in the case of the HP 4291B and the HP 4191A, is defined at the 7mm (APC-7) test port. If the device under test (DUT) is measured at the calibration plane (requires an insertable device), the measurement values are only affected by the specified measurement accuracy, thus there is no correlation problem among different instruments (refer Figure 1). Since most DUTs are non­insertable (do not directly attach to the 7mm connector) at the calibration plain, a test fixture is usually used to contact the device. There are various error factors associated with fixtured measurements, caused by the existence of the test fixture (refer Figure 2). The factors affecting the correlation are discussed as follows.
1. Performing compensation
When a fixtured measurement is made without compensating for the electrical characteristics of the test fixture, the measurement result reflects the characteristics of the DUT as well as the characteristics of the test fixture (fixture parasitics). Several compensation functions are used to remove the error caused by a test fixture. Since different compensation functions are sometimes provided with different instruments, the correlation of measurement values will be affected.
Figure 1. Measuring DUT at Calibration Plane
Figure 2. Factors to Affect Correlation Problem
calibration plane and the DUT. This leads to an impedance measurement error when measuring phase. Residual impedance and phase shift error can be compensated with OPEN/ SHORT compensation and electrical length compensation. The HP 4291B provides both OPEN/SHORT/(LOAD)
Figure 3. Residual Error Model of Test Fixture
compensation and electrical length compensation functions. All error factors mentioned above can
Figure 3 shows the error model caused by a test fixture. There are 2 types of error, one is residual impedance error and the other one is electrical length error. The residual impedance which includes small resistance, inductance, conductance and capacitance makes the measurement value inaccurate. The electrical length error occurs due to a phase shift of the test signal along the
50Ω transmission line between the
be removed effectively. As a result, the real characteristics of a device can be measured, since the device can be equivalently moved to the calibration plane (Figure 4). (Remarks: Strictly speaking, one factor still remains. This factor is regarded as the time fluctuation of compensated values). The HP 4191A provides only the electrical length compensation function (Figure 4(b)). Since the measurement value includes the impedance of the DUT and the test
2
Loading...
+ 2 hidden pages