any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
16900-97011
Edition
Ninth edition, December 2007
Printed in Malaysia
Agilent Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Trademark Acknowledgements
Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.
Windows NT is a U.S. registered trademark
of Microsoft Corporation.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
216900 Series Logic Analysis Systems Installation Guide
Installation at a Glance
16900 Series Logic Analysis System Overview
The Agilent Technologies 16900 Series logic analysis systems are modular
systems with slots for logic analyzer and other types of measurement
modules.
16900 Series Logic Analysis Systems Installation Guide3
Tabl e 1Model comparisons
Agilent model
number
16900A6YesUses external display, supports up to four external displays
16901A2YesBuilt-in color touch screen display, 15 inch at 1024 x 768,
16902A6YesBuilt-in color touch screen display, 12.1 inch at 800 x 600,
16902B6YesBuilt-in color touch screen display, 15 inch at 1024 x 768,
16903A3NoBuilt-in color touch screen display, 12.1 inch at 800 x 600,
Number
of slots
Multiframe
Pro
Display and resolutionPCI expansion
at up to 1600 x 1200 (with PCI video card)
supports external monitor (without additional PCI video
card)
supports up to four external monitors at up to 1600 x 1200
(with PCI video card)
supports external monitor (without additional PCI video
card)
supports up to four external monitors at up to 1600 x 1200
(with PCI video card)
All models have a familiar Windows-based user interface which takes the
complexity out of making logic analyzer measurements. You can perform
all operations directly from one window.
• Supplied accessories:
• PS2 mouse.
• PS2 Mini keyboard.
• Accessory pouch and power cord.
• Ten-conductor flying lead cable for target control port (on 16900A,
16902A, 16902B, and 16903A models).
• Optional accessories:
• Multiframe cable (E5861A).
• 1 Gbit low-profile LAN card (Option 014 or E5860A for 16900A,
16902A, and 16903A models).
• Probes.
slots
2 full profile
1 low profile
1 full profile
2 full profile
1 low profile
1 full profile
1 full profile
1 low profile
416900 Series Logic Analysis Systems Installation Guide
In This Guide
Set up hardware and
peripherals (Chapter 1)
Are
measurement
modules installed
?
Yes
First-time set up considerations
and Windows Welcome
questions.
(Chapter 4)
Do you
want to connect
multiple frames
together?
No
You are ready to make
measurements.
No
Yes
Install
measurement
modules
(Chapter 2 & 3)
Go to Setting Up
Multiframe
Configurations
(Chapter 5)
Go to Chapters 6 & 7 for post
set up reference and problem
solving.
16900 Series Logic Analysis Systems Installation Guide5
616900 Series Logic Analysis Systems Installation Guide
Contents
Installation at a Glance 3
In This Guide 5
1Setting Up Hardware
To connect the mouse, keyboard, and monitor 14
To connect to a LAN 17
To connect a printer 18
To make time-correlated measurements with an Agilent oscilloscope 19
2Installing Logic Analyzer Cards and Probes
Software Requirements 23
To connect probes 24
To replace thumb screws on measurement modules 24
To install, remove, or replace measurement modules 26
16960A Logic Analyzer (1-card module) 30
16960A Logic Analyzer (2-card, 3-card, and 4-card modules) 31
Setting Up a Performance Multiframe Configuration 84
Rules for Performance Multiframe Configurations 85
Creating a Performance Multiframe Configuration 85
Remotely Controlling a Performance Multiframe System 88
Other Multiframe Configuration Notes 89
I/O in Multiframe Configurations 89
Licenses in Multiframe Configurations 89
6Using and Updating the Logic Analysis System
To locate information on using the logic analyzer 92
16900 Series Logic Analysis Systems Installation Guide9
Online Help 92
To locate specifications and characteristics 93
Electrical and Operating Environment Characteristics 93
More Specifications and Characteristics 94
To perform backups 95
To install software 95
Agilent Logic Analysis System Software 95
Optional Products 95
To change network settings 96
To change Windows Firewall settings 97
To give other applications/ports access through the firewall 97
To restore the logic analysis system firewall defaults 99
To power off the system 102
Using the Windows Shutdown 102
Using a Short Press of the Power Button 102
Using a Long Press of the Power Button 103
Unplugging the Power Cord or Power Loss 103
To remove the hard disk drive (16902B) 104
To install interface cards (16900A, 16902A, 16902B, or 16903A) 105
1 Gbit LAN Card 109
To install interface cards (16901A) 110
To replace the CPU battery 113
7Solving Problems
Running Self-Tests 116
If the logic analysis system powers itself down 117
16900A, 16902A, and 16902B logic analysis systems 117
All 16900 Series logic analysis systems 118
1016900 Series Logic Analysis Systems Installation Guide
If the Power LEDs do not come on 118
If you get a registry error and the system won’t boot 118
If you can’t write/read CDs 118
If there are multiframe connection problems 119
If there are network connection problems 121
Using the System Recovery Software 122
Contacting Agilent Service/Support 122
ASafety Notices
Warnings 123
To clean the instrument 124
Safety Symbols 124
Index
16900 Series Logic Analysis Systems Installation Guide11
1216900 Series Logic Analysis Systems Installation Guide
Agilent 16900 Series Logic Analysis Systems
Installation Guide
1
Setting Up Hardware
To connect the mouse, keyboard, and monitor 14
To connect to a LAN 17
To connect a printer 18
To make time-correlated measurements with an Agilent oscilloscope 19
A
13
1Setting Up Hardware
To connect the mouse, keyboard, and monitor
The 16900A logic analysis system must have the system mouse and
keyboard connected for the system to boot up properly. Once enabled on
the LAN, the system can be operated remotely without a keyboard or
mouse. Use of a monitor is optional for the 16901A, 16902A, 16902B, and
16903A.
(The following rear panel figures are representative; yours may look
different.)
1 Connect the provided PS/2 mouse and keyboard provided to the back of
frame.
16900A and 16902A/B
16901A and 16903A
2 Connect a monitor to the back of the 16900A. A second monitor on the
16901A, 16902A, 16902B, or 16903A is optional.
16900A and 16902A/B
16901A and 16903A
1416900 Series Logic Analysis Systems Installation Guide
Setting Up Hardware1
The touch off button switches the Windows desktop from one display
device to another. A message will appear in one of the monitors telling
you how to use this feature.
3 Connect the power cord supplied with your frame.
NOTE
The 16900A, 16902A, and 16902B have a special notched power cord in order to provide
1300W to the frame.
16901/3A
Frame
Monitor Power Cable
Power
Cables
16900/902
16900 Series Logic Analysis Systems Installation Guide15
1Setting Up Hardware
4 Allow a minimum of 15 cm (6 in.) spacing between instruments for
proper cooling.
15 cm (6 in.)
15 cm (6 in.)
1616900 Series Logic Analysis Systems Installation Guide
To connect to a LAN
Local area networks (LAN) are enabled outside of the logic analyzer
application using the Windows network configuration wizard.
1 Connect the LAN cable(s) to the back of your logic analysis system.
• Agilent recommends using a 1 Gbit LAN card when networking
multiple frames.
The 16902B and 16901A frames have an integrated 1 Gbit LAN
interface.
For information on installing a 1 Gbit LAN card in your system, See
“To install interface cards (16900A, 16902A, 16902B, or 16903A)” on
page 105.
• For information on network configurations when connecting multiple
frames see Chapter 5, “Setting Up Multiframe Configurations on
page 75.
Setting Up Hardware1
16900A and 16902A
10/100BaseT
LAN port
Optional 1Gb
LAN card
NOTE
16900 Series Logic Analysis Systems Installation Guide17
Microsoft recommends the following three steps to ensure the instrument's PC is
protected:
a Use an internet firewall.
b Get computer updates.
c Use up-to-date antivirus software.
2 Refer to “First- Time Set Up” on page 69 to answer the Windows
Welcome when you power up the logic analyzer.
10/100BaseT
LAN port
Optional 1Gb
LAN card
16903A
1Setting Up Hardware
To connect a printer
Local and network printers are added outside of the logic analyzer
application using the Windows add printer wizard.
1 Connect the printer cable to the back of your logic analysis system. If
2 Click Start>Settings>Printers>Add Printer.
3 Follow the Windows add printer wizard instructions.
using a network printer ensure the LAN cable is connected.
1816900 Series Logic Analysis Systems Installation Guide
Setting Up Hardware1
To make time-correlated measurements with an Agilent oscilloscope
You can make time-correlated measurements between an Agilent 16900
Series logic analyzer and Agilent 5000, 6000, 548xx, 8000, or 80000 Series
oscilloscopes.
Measurements between the logic analyzer and Agilent oscilloscope are
automatically de- skewed. The instruments communicate with one another
through a LAN connection and Trigger In/Trigger Out.
The instruments connect to your device under test through separate
probes, just as when they are used independently. The oscilloscope time
markers and two logic analysis system global markers are time- correlated
and track each other.
To set up and use the time-correlated connection, choose Setup>Add External Scope... in the Agilent Logic Analyzer application.
16900 Series Logic Analysis Systems Installation Guide19
1Setting Up Hardware
2016900 Series Logic Analysis Systems Installation Guide
Agilent 16900 Series Logic Analysis Systems
Installation Guide
2216900 Series Logic Analysis Systems Installation Guide
Software Requirements
The following table gives you the Agilent Logic Analyzer software version
required for logic analyzer measurement modules in the 16900A, 16902A,
16902B, or 16903A frames. For information on updating your logic
analysis system software go to “To install software” on page 95.
Tabl e 2Software Version Required, Frames
Model NumberSoftware Version
16903A02.00 or higher
16902B03.70 or higher
16902A02.00 or higher
16901A03.60 or higher
16900A02.00 or higher
Installing Logic Analyzer Cards and Probes2
Tabl e 3Software Version Required, Logic Analyzer Modules
Model NumberSoftware Version
16960A03.70 or higher
16951B03.65 or higher
16950B03.55 or higher
16950A02.00 or higher
16911A, 16910A02.00 or higher
16760A03.20 or higher
16756/55/54/53A02.00 or higher
16752/51/50B/A, 16742/41/40A02.00 or higher
16900 Series Logic Analysis Systems Installation Guide23
2Installing Logic Analyzer Cards and Probes
To connect probes
Probing is the key to effective and efficient use of logic analyzers. Agilent
Technologies offers a wide variety of probing accessories that support
general- purpose and application- specific measurement needs. We provide
reliable, electrically and mechanically unobtrusive probes that make it easy
to connect your Agilent logic analyzer to your system under test. For more
information on Agilent probes:
• See the quick reference card that came with your logic analysis system
for an overview of probing solutions.
• Go to www.agilent.com/find/logic_analyzer_probes for the most
up- to- date information on probe compatibility with a given logic
analyzer and purchasing information.
• Go to www.agilent.com and search for Probing Solutions for Logic
Analyzers (publication 5968-4632E) for a catalog of detailed probing
information.
To connect and set up a probe, follow the instructions in the User’s Guide
that came with the probe.
To replace thumb screws on measurement modules
If you ordered measurement modules with your logic analysis system, they
will have the proper thumb screws and you can skip this procedure.
If you are using a measurement module from a 16700 Series logic analysis
system, you may need to replace the thumb screws. The rigid screws on
these modules are not compatible with the new ejector levers.
1 Remove the retaining ring from the back of the panel then remove the
thumb screw.
2 Insert the module sleeve into the front of the module panel and screw
the nut onto it.
2416900 Series Logic Analysis Systems Installation Guide
Installing Logic Analyzer Cards and Probes2
3 Screw the module screw into the sleeve.
Hex nut
Thumb
screw
Module
Retaining
ring
The 16900A/16902A logic analysis system comes with a set of 6
replacement thumb screws (16900-68713), and the 16903A logic analysis
system comes with a set of 2 replacement thumb screws (16903-68713).
sleeve
Module
screw
16900 Series Logic Analysis Systems Installation Guide25
2Installing Logic Analyzer Cards and Probes
To install, remove, or replace measurement modules
If you ordered your measurement modules with the frame, they will
already be installed and you can skip this procedure.
If you are using a measurement module from a 16700 Series logic analysis
system you will need to replace the thumb screws. Begin with the
instructions on page 24.
The following procedure guides you through installing, removing, or
replacing measurement modules in the logic analysis system frame.
Information on connecting specific multi- card measurement modules is in
the next sections of this chapter. Measurement modules with different
model numbers cannot be connected together in multi- card
(Master/Expander) modules unless stated otherwise.
CAUTION
CAUTION
Electrostatic discharge can damage electronic components. Use grounded wrist straps
and mats when performing any service to measurement modules.
When removing cards that are connected together as multiple-card modules, be sure
to eject the multiple cards simultaneously, otherwise, the flex cables and/or
connectors could be damaged.
1 Power down the system and disconnect the power cable before
installing, removing or replacing measurement modules.
Power Off
Disconnect
OR
Disconnect
2616900 Series Logic Analysis Systems Installation Guide
Installing Logic Analyzer Cards and Probes2
2 Remove filler panels as shown in the following figure.
a Loosen the thumb screws on the filler panels.
b Rotate the ejector levers from the locked position to the open
position.
c Remove filler panels.
Thumb screw
Locked position
3 If you are using a measurement module from a 16700 Series logic
analysis system go to “To replace thumb screws on measurement
modules” on page 24. If not, skip this step.
4 Ensure all ejectors are fully open and out of the way then slide the
module(s) into the logic analysis system frame.
The following figure shows a 6-card frame. The insertion/removal order
would be the same for a 3- card frame. A single- module configuration
can be installed in any available slot.
Open position
16900 Series Logic Analysis Systems Installation Guide27
2Installing Logic Analyzer Cards and Probes
5 Seat the modules then rotate the ejector tabs to the locked position.
Repeat for all modules or filler panels.
1
6
2
5
3
4
4
3
5
2
6
1
Insertion Order
Removal Order
Locking Order
Ejecting Order
6
5
4
3
2
1
6 Tighten the thumb screws until finger tight in the same order used in
step 5 above.
7 For correct air circulation, install filler panels in all unused card slots.
Correct air circulation keeps the instrument from overheating. Keep any
extra filler panels for future use.
2816900 Series Logic Analysis Systems Installation Guide
1
2
3
4
5
6
Installing Logic Analyzer Cards and Probes2
8 Refer to “First- Time Set Up” on page 69 to answer the Windows
Welcome questions when you power up the logic analysis system for the
first time.
16900 Series Logic Analysis Systems Installation Guide29
2Installing Logic Analyzer Cards and Probes
16960A Logic Analyzer (1-card module)
A single- card module can be installed in any available slot.
For a single card, the two 16960-61601 flex cables (kit 16960- 60001 in the
accessory pouch) are not used.
NOTE
Turn off the mainframe power before removing, replacing, or installing modules. For
instructions on installing modules into the logic analysis system frame, refer to page 26.
The order of pods on the card’s rear panel is:
Pod 4Pod 3Pod 2Pod 1
In a multi- card set, pods are indexed from the bottom up. Because clock
signals can be input to any card in the set (on pod 1 or 3), cards do not
have master and expander designations.