Agilent 11974V Data Sheet

Agilent 11974 Series
Preselected Millimeter Mixers
Data Sheet
Whether your millimeter-frequency application is com­ponent or system testing, signal surveillance or EMI measurements, your software development and verifica­tion time will be reduced and your measurement times will be decreased. Simply place a marker on a signal—all responses are real—and perform a preselector peak. The analyzer then displays the amplitude and frequency of the signal. No time is lost creating and running signal identification routines, which used to be required to separate true responses from multiples and images.
Each 11974 series preselected millimeter mixer is char­acterized for conversion loss versus frequency. Calibration accuracy ranges from 1 dB at 26.5 GHz to 2 dB at 75 GHz. Just enter the conversion loss data into your host spectrum analyzer and then make amplitude­calibrated measurements.
Choose Your Host Spectrum Analyzer
Agilent 11974 series preselected millimeter mixers are compatible with Agilent’s 8566B, 8560 series and E4407B of rugged portables
1
, 70000 Series modular spectrum analyzers and the E4440A/46A/48A PSA Series
1
.
1. Requires Option AYZ (external mixing).
For the first time, preselected signal analysis is available to 75 GHz. With preselection, you can quickly and easily characterize all signals of interest. No images or multi­ples are displayed to confuse the measurement. Time consuming signal-identification routines are eliminated.
Figure 1 shows how true signals can be obscured by dis­played multiples and images in an unpreselected sys­tem. Figure 2 shows how only the true signals are dis­played when an Agilent Technologies 11974 series prese­lected millimeter mixer is used in making the measure­ment.
RF Sections Available in Four Bands from 26.5 to 75 GHz
Figure 1.
Figure 2.
2
Added Benefits of Preselected Millimeter Measurements
Reduced Overload from Many Signals
Preselected millimeter mixers reduce susceptibility to overloading by numerous signals. Overloading can cause your measurements to be uncalibrated.
System Amplitude Accuracy
Millimeter preselection improves system amplitude accuracy by reducing the level of local oscillator (LO) emissions back to the device under test. Option 001 further improves system amplitude accuracy by adding an isolator in front of the input filter. With Option 001, the RF input VSWR is <1.6:1 in A, Q, and U bands and <2.3:1 in V Band.
System Configuration
To operate your 11974 series preselected millime­ter mixer, simply connect the tune+span output, local oscillator output,1and the IF input of the spectrum analyzer to the mixer. Select the appro­priate analyzer using the switch on the rear panel of the mixer. Make two easy adjustments, and you're ready to make millimeter measurements.
The 11974 series connect to a standalone power supply with a convenient 6-foot cable. Use the same power supply with any 11974 series prese­lected mixer.
A full set of upgrade kits is available to ensure compatibility with your existing spectrum analyz­ers including Agilent 8560A, 8561A/B, 8562A/B, 8566A/B, and
E4440A/46A/48A PSA Series. Refer
to the compatibility table to determine whether your spectrum analyzer requires an upgrade.
All versions of 8563A spectrum analyzers and 70907B external mixer interface modules are fully compatible with the 11974 series.
Greater Dynamic Range
Preselection improves dynamic range when signals are sufficiently spaced. The preselector allows only one of these signals to reach the mixer at a time. So internally produced distortion products, such as third order intermodulation, are reduced.
Ease of Use
Operators no longer need to learn techniques of signal identification. Anyone familiar with prese­lected microwave spectrum analyzers can quickly and easily make millimeter measurements. Spurs from a device under test are easily found. They are not hidden by images or multiple responses.
1. Operation with 8566B spectrum analyzers requires the use of an 11975A LO amplifier to provide sufficient local oscillator power to the 11974 series LO input.
11974A
11474Q
11974U
26.5 GHz 33 GHz
50 GHz40 GHz
11974V
60 GHz 75 GHz
3
Frequency
Frequency Range
(LO Harmonic Number) 11974A (n = 8) 26.5 GHz to 40 GHz 11974Q (n = 10) 33 GHz to 50 GHz 11974U (n = 10) 40 GHz to 60 GHz 11974V (n = 14) 50 GHz to 75 GHz
Preselector Bandwidth (3 dB points)
Minimum Typical
11974A 80 MHz 130 MHz 11974Q 100 MHz 150 MHz 11974U 100 MHz 150 MHz 11974V 100 MHz 200 MHz
Center Frequency Accuracy, Residual FM, Span Accuracy
Using specifications of the spectrum analyzer, evaluate characteristics with reference to the harmonic numbers listed below.
Harmonic Number
11974A n = 8 11974Q n = 10 11974U n = 10 11974V n = 14
Amplitude
Displayed Average Noise Level
Spectrum Analyzers: 8566B, 71OOOC, PSA Series (RBW=10 Hz)
Maximum Typical
11974A -111 dBm -118 dBm 11974Q -106 dBm -116 dBm 11974U -109 dBm -117 dBm 11974V -100 dBm -109 dBm
Spectrum Analyzers: 8563A/62A/61B/60A (RBW=300 Hz)
1
Maximum Typical
11974A -96 dBm -103 dBm 11974Q -91 dBm -101 dBm 11974U -94 dBm -104 dBm 11974V -85 dBm -94 dBm
Maximum Conversion Loss
RF to IF port (includes 1-m IF cable, part number 5061-5458) 11974A 44 dB 11974Q 46 dB 11974U 43 dB 11974V 57 dB
Conversion Loss Chart Accuracy
20 °C to 30 °C: RSS of calibration equipment errors (Add 0.7 dB if used with 8563A/62A/61B/60A.)
0 °C to 55 °C: RSS of calibration equipment errors plus worst-case temper­ature effects of 11974 (Add 0.7 dB if used with 8563A/62A/61B/60A.)
0 °C to 55 °C 20 °C to 30 °C
11974A ±2.3 dB ±1.0 dB 11974Q ±2.3 dB +1.0 dB 11974U ±2.6 dB +1.1 dB 11974V ±4.5 dB ±2.0 dB
Image Rejection
(Image positioned 2x fIF above tuned response) Spectrum Analyzers: 8566B, 71000C, PSA Series
O°C to 55°C 20°C to 30°C
11974A ≤ -54 dBc -59 dBc 11974Q ≤ -50 dBc -55 dBc 11974U ≤ -50 dBc -55 dBc 11974V (50 to 67 GHz) ≤ -50 dBc -55 dBc 11974V (67 to 75 GHz) ≤ -40 dBc -45 dBc
Spectrum Analyzers: 8563A/62A/61B/60A
O °C to 55 °C 20 °C to 30 °C
11974A ≤ -50 dBc -55 dBc 11974Q ≤ -45 dBc -50 dBc 11974U ≤ -45 dBc -50 dBc 11974V (50 to 67 GHz) ≤ -45 dBc -50 dBc 11974V (67 to 75 GHz) ≤ -35 dBc -40 dBc
Multiple Response
(Due to in-range applied signals mixing with LO harmonics other than order n) 11974A (n = 8) -63 dBc 11974Q (n = 10) -60 dBc 11974U (n = 10) -60 dBc 11974V (n = 14) (applied signal: 50 to 67 GHz) -60 dBc (applied signal: 67 to 75 GHz) -55 dBc
1 dB Gain Compression (Nominal)
11974A ≥ +6 dBm 11974Q ≥ 0 dBm 11974U ≥ 0 dBm 11974V ≥ +3 dBm
1. The minimum usable resolution bandwidth of the 8563A/61B/60A in external mixing mode is 300 Hz. The 8562A can be used with its minimum resolution band width of 100 Hz in external mixing mode.
Specifications
Unless noted, all specifications describe warranted performance under the following conditions: 5-minute warm-up from ambient conditions; 1-year calibration cycle; environmental requirements met; IF and LO connections between the Agilent 11974 and a compatible spectrum analyzer are made with one length of coax cable (part number 5061-5458). Typical performance is non-warranted. Supplemental characteristics are denoted by "nominal"; these constitute non-warranted functional performance information derived dur­ing the design process and are not tested on a continuing basis.
Loading...
+ 5 hidden pages