Network Analyzers
275Test Sets
External S-Parameter Test Sets 3
The S-parameter test sets provide the capability to measure reflection
and transmission characteristics (including S-parameters) of two-port
devices in either direction with a single connection. The test sets are
controlled from the analyzer and include programmable step attenuators. These test sets are used with the HP 8753A/B/C or the HP
8753D/E/ES Option 011 only.
HP 85046A/B S-Parameter Test Sets
The HP 85046A/B test sets provide the capability to simultaneously
measure the transmission and reflection characteristics of 50-ohm and
75-ohm devices, respectively.
Specifications Summary
HP 85046A HP 85046B
Impedance 50Ω 75Ω
Frequency Range 300 kHz to 3 GHz 300 kHz to 2 GHz
Directivity 35 dB to 1.3 GHz 35 dB to 1.3 GHz
Typical Tracking Transmission Magnitude, Phase
0.3 MHz to 2.0 MHz ±1.5 dB, ±20˚ ±1.5 dB, ±20˚
2.0 MHz to Fmax ±1.5 dB, ±10° ±1.5 dB, ±10˚
Reflection Magnitude, Phase
0.3 MHz to 2.0 MHz ±1.5 dB, ±25˚ ±1.5 dB, ±25˚
2.0 MHz to Fmax ±1.5 dB, ±10˚ ±1.5 dB, ±10˚
Effective Source Match3(test ports)
0.3 MHz to 2.0 MHz 14 dB 14 dB
2.0 MHz to 1.3 GHz 20 dB 17 dB
2.0 MHz to Fmax 16 dB 16 dB
RF Connectors
Test Ports Precision 7 mm 75Ω type-N (female)
All Others 50Ω Type-N (female) 50Ω Type-N (female)
1
Degrees, specified as deviation from linear phase.
2
Fmax is the upper frequency limit of the associated test set.
3
Can be improved through accuracy enhancement.
Includes: Four 190-mm (7.5 in) cables with Type-N (male) connectors
for connection to the HP 8753. One HP 8753 test set interconnect cable.
Physical Characteristics
Size: 426 mm W x 90 mm H x 508 mm D (16.75 in x 3.5 in x 20 in)
Weight: Net, 6.8 kg (15 lb); shipping, 9.1 kg (20 lb)
HP 85047A S-Parameter Test Set
The HP 85047A test set includes a frequency doubler that can be switched
in to measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in a single sweep. The HP 8753B/C controls the
frequency doubler. (The HP 8753D/E/ES Option 006 and 011 with
built-in 6 GHz source does not use the frequency doubler,but is still
compatible with the HP 85047A.) Option 006 (6 GHz receiver) is
required to activate the HP 85047A.
Specifications Summary
Impedance: 50Ω
Frequency Ranges
300 kHz to 3 GHz and 3 MHz to 6 GHz (HP 8753B/C);
300 kHz to 6 GHz (HP 8753D/E/ES Option 006 and 011)
Directivity
300 kHz to 1.3 GHz: 35 dB;
1.3 GHz to 3 GHz: 30 dB
3 GHz to 6 GHz: 25 dB
Typical Tracking
Transmission Magnitude, Phase
300 kHz to 3 GHz: ±1.5 dB, ±10˚;
3 GHz to 6 GHz: +0.5, – 2.5 dB, ±20˚
Reflection Magnitude, Phase
00 kHz to 3 GHz: ±1.5 dB, ±10°;
3 GHz to 6 GHz: ±1.5 dB, ±20°
30 dB to 3.0 GHz 30 dB to 2.0 GHz
1, 2, 3
1, 2, 3
Effective Source Match
300 kHz to 1.3 GHz: 20 dB;
1.3 GHz to 3 GHz: 16 dB
3 GHz to 6 GHz: 14 dB
RF Connectors
Test Ports: Precision 7 mm
All Others: 50Ω Type-N (female)
Includes: Four 190 mm (7.5 in) cables with Type-N (male) connectors for
connection to the HP 8753, one HP 8753 test set interconnect cable.
Physical Characteristics
Size: 426 mm W x 90 mm H x 508 mm D (16.75 in x 3.5 in x 20 in)
Weight: Net, 10 kg (22 lb); shipping, 15 kg (33 lb)
External Test Set Switching
Option 009 replaces the standard solid-state RF test port switch with a
mechanical RF switch. HP 8753 system specifications for standard and
Option 009 test sets are identical. Nominal insertion loss of the solidstate switch is less than 2 dB (at 3 GHz) or 3 dB (at 6 GHz), relative to a
mechanical switch.
Special Test Sets
Special test sets are available to configure the HP 8753ES for specific
applications. Some examples are listed below. Contact HP for details
about these products or for information about additional special options
for HP 8753 network analyzers.
Option H16 Low Noise Floor
Adds the ability to reverse the port 2 coupler to increase the forward
dynamic range by about 13 dB.
Option H36 Duplexer Test Set
Adds a third test port to enable single-connection duplexer measurements. Does not provide Tx-to-Rx and Rx-to-Tx measurements.
Option H39 Three-Port Test Set
Adds a third test port and switching to provide all transmission and
reflection measurements for three-port devices.
Option H68 Extended frequency range above 6 GHz
Allows characterization of components up to 6.8 GHz. Higher frequencies up to 8 GHz can be quoted upon request (may be ordered for HP
8753ET or HP 8753ES).
Option H85 High power test set
Adds access to signal paths to allow the addition of an external amplifier,
high power attenuators or isolators for handling up to 20 watts of power
at the test ports. Standard solid-state transfer switch is replaced by
mechanical switch and internal attenuators are added.
Accessories
HP 11930A/B Power Limiters
The HP 11930A/B limiters protect the input circuits of network analyzers,
spectrum analyzers and sources from transients and short-duration
overloads.
Specifications Summary
HP 11930A HP 11930B
Frequency Range DC to 6 GHz 5 MHz to 6.5 GHz
Input/output Connector APC-7 Type-N
Insertion Loss 1.0 dB (dc to 3 GHz) 1.0 dB (16 MHz to 3 GHz)
Return Loss 22 dB (30 kHz to 3 GHz) 21 dB (16 MHz to 3 GHz)|
Impedance 50-ohm nominal 50-ohm nominal
Fax-on-demand (24 hr) 1-800-800-5281 Internet URL www.hp.com/go/tmc00 Product and
order Information 8am-8pm EST 1-800-452-4844.
1.5 dB (3 to 6 GHz) 1.5 dB (3 to 6.5 GHz)
20 dB (3 to 6 GHz) 17 dB (3 to 6.5 GHz)
5
Internet URL www.hp.com/go/tmc00 Product & Order Info See page 607