Keysight UXA Installation Manual

Page 1
Keysight UXA Signal Analyzers
Option YAV Y-Axis Video Out
Installation Note
Page 2
Notices
© Copyright 2016 - 2021 Keysight Technologies, Inc.
Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Manual Part Number
N9040-90008
Edition
Edition 1, January 2021 Supersedes June 2017
Printed in USA/Malaysia
Published by: Keysight Technologies, Inc. 1400 Fountaingrove Parkway Santa Rosa, CA 95403
Page 3

Option YAV Installation Kit

Products Affected: N9040B UXA Signal Analyzer
N9041B UXA Signal Analyzer N9042B UXA Signal Analyzer
Serial Numbers: All
Options
To Be Performed By: ( ) Keysight Service Center
(X) Personnel Qualified by Keysight
(X) Customer
Option YAV Installation Kit
Estimated Installation Time: Estimated Adjustment and Verification Time:
0.5 Hours
0.5 Hours
Installation Note N9040-90008 3
Page 4
Option YAV Installation Kit

Introduction

This kit contains all of the instructions required to install Option YAV on a UXA X-Series signal analyzer. Option YAV enables the ANALOG OUT port and provides various signal selections.
The rear panel ANALOG OUT port is present on all UXA signal analyzers, however Option YAV must be enabled via a license file.
1. To ensure that this newly installed option has been installed properly,
the procedure that follows includes a functional check.
2. The latest revision of the X-Series signal analyzer software may be downloaded from:
http://www.keysight.com/find/Xseries_software
3. This option is licensed for one instrument model/serial number combination. The license key will only install on the designated instrument.
4 Installation Note N9040-90008
Page 5
Option YAV Installation Kit

Contents

Quantity Description Part Number
1 Installation Note This note
1 Option Upgrade Entitlement Certificate 5964-5178

Tools Required for Installation

— Personal computer with internet access and USB port
— USB storage device with > 2 GB free memory

Tools Required for Manual Operation Verification

—Voltmeter
— Test cable, BNC (used to connect voltmeter to the analyzer rear panel ANALOG OUT)
— Banana plug to BNC adapter (used to connect voltmeter to BNC cable)
Installation Note N9040-90008 5
Page 6
Option YAV Installation Kit

Installation Procedure over USB

1. Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit.
2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the
Certificate.
3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File.
4. Locate a USB storage device. Perform a virus scan on this device before use.
5. Save the License File to the root directory of the USB Storage Device.
6. Connect the USB Storage Device to the signal analyzer USB port. Windows will detect the new
hardware and may display the configuration menu shown in Figure 1. This menu may be configured according to your preferences.
Figure 1 USB Storage Device Configuration Menu
7. The signal analyzer will automatically consume the License File. (This may take a few minutes)
When the License File is consumed the Keysight License Manager will display a “Successful License Installation” message as shown in Figure 2.
Figure 2 Successful License Installation
6 Installation Note N9040-90008
Page 7
Option YAV Installation Kit

Alternate Installation Procedure

The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer
C:\Program Files\Agilent\licensing

Verify the Installation

1. Cycle the power on the signal analyzer.
2. Press System, Show, System to display a list of installed options.
3. Verify that the newly installed option, N904xB-YAV, appears on the list.
Installation Note N9040-90008 7
Page 8
Option YAV Installation Kit

Manually Verify Operation

1. Preset the analyzer. Assure mode is Spectrum Analyzer.
2. Display the 50 MHz calibrator signal. Press Input/Output, RF Calibrator, 50 MHz. Press FREQ, Center Frequency, key in 50 MHz. Press Span, Zero Span. You should see a flat line across the
display at 25 dBm.
3. Press AMPTD, Ref Level, and key in 25 dBm. The displayed signal level should be at the top graticule line.
4. Press Input/Output, Output, Analog, Out, then select Screen Video.
5. Connect a voltmeter, set to measure DC volts, to the analyzer rear panel ANALOG OUT
connector.
6. The voltmeter should measure 1V.
7. Increase the Ref Level to 15 dBm. The signal should drop one graticule division on screen and
the voltmeter reading should be 0.9 V.
8. This test is sufficient to determine if the output circuitry is working properly for all Analog Out signal selections.
8 Installation Note N9040-90008
Page 9
Option YAV Installation Kit

Utilities, Adjustments, and Performance Verification Tests

Calibration Software and specified test equipment is required to perform the adjustments, and can be used to automate the performance verification testing. Information on how to obtain this software can be found at
http://www.keysight.com/find/calibrationsoftware

Utilities Required

None

Adjustments Required

None

Performance Testing Required

None
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after the installation of this upgrade or not. If a full calibration is required, arrangements regarding the level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
Installation Note N9040-90008 9
Page 10
This information is subject to change without notice. © Keysight Technologies 2016-2021
Edition 1, January 2021
N9040-90008
www.keysight.com