reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies as governed by
United States and international
copyright laws.
Manual Part Number
U3401-90001
Edition
Edition 14, November 1, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight
products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest
Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal
Acquisition Regulation (“FAR”) 2.101.
Pursuant to FAR 12.212 and 27.405-3
and Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires commercial
computer software under the same
terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the
Software to U.S. government
customers under its standard
commercial license, which is embodied
in its End User License Agreement
(EULA), a copy of which can be found at
http://www.keysight.com/find/sweula.
The license set forth in the EULA
represents the exclusive authority by
which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license set
forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation that
is not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall apply,
except to the extent that those terms,
rights, or licenses are explicitly required
from all providers of commercial
computer software pursuant to the FAR
and the DFARS and are set forth
specifically in writing elsewhere in the
EULA. Keysight shall be under no
obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR
CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING
THE MATERIAL IN THIS DOCUMENT
THAT CONFLICT WITH THESE TERMS,
THE WARRANTY TERMS IN THE
SEPARATE AGREEMENT SHALL
CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight U3401A User’s and Service Guide
Page 3
Safety Symbols
CAT II
300 V
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC)Caution, risk of electric shock
Alternating current (AC)
Both direct and alternating currentOut position of a bi-stable push control
Earth (ground) terminalIn position of a bi-stable push control
Protective earth (ground) terminalFrame or chassis (ground) terminal
IEC Measurement Category II.
Measurement performed on circuits
that are directly connected to the low
voltage installation (up to 300 VAC)
under Category II overvoltage
conditions
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
Keysight U3401A User’s and Service Guide3
Page 4
General Safety Information
WARNING
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
– Do not exceed any of the measurement limits defined in the specifications
to avoid instrument damage and the risk of electric shock.
– Do not defeat power cord safety ground feature. Plug in to a grounded
(earthed) outlet.
– Do not use multimeter in any manner that is not specified by the
manufacturer.
– Double-check the multimeter’s operation by measuring a known voltage.
– For current measurement, turn off circuit power before connecting the
multimeter to the circuit. Always place the multimeter in series with the
circuit.
– When connecting probes, always connect the common test probe first.
When disconnecting probes, always disconnect the live test probe first.
– Do not measure more than the rated voltage (as marked on the
multimeter) between terminals, or between terminal and earth ground.
– Do not use repaired fuses or short-circuited fuse-holders. For continued
protection against fire, replace the line fuses only with fuses of the same
voltage and current rating and recommended type.
– Do not service or perform adjustments alone. Under certain conditions,
hazardous voltages may exist, even with the multimeter switched off. To
avoid dangerous electric shock, service personnel must not attempt
internal service or adjustment unless another person, capable of
rendering resuscitation or first aid, is present.
4Keysight U3401A User’s and Service Guide
Page 5
WARNING
– Do not substitute parts or modify multimeter to avoid the danger of
CAUTION
introducing add itional hazards. Return the product to Keysight
Technologies Sales and Service Office for service and repair to ensure the
safety features are maintained.
– Do not operate damaged equipment as the safety protection features
built into this multimeter may have been impaired, either through
physical damage, excessive moisture, or any other reason. Remove power
and do not use the product until safe operation can be verified by
service-trained personnel. If necessary, return the product to Keysight
Technologies Sales and Service Office for service and repair to ensure the
safety features are maintained.
– Turn off circuit power and discharge all high-voltage capacitors in the
circuit before you perform resistance, continuity, diodes, or capacitance
tests.
– Use the correct terminals, functions, and range for your instruments.
– Do not measure voltage when current measurement is selected.
– Use the multimeter with the cables provided.
– Repair or service that is not covered in this manual should only be
performed by qualified personnels.
Keysight U3401A User’s and Service Guide5
Page 6
Environmental Conditions
NOTE
This multimeter is designed for indoor use only. The table below shows the
general environmental requirements for the multimeter.
Environmental conditionsRequirements
Operating temperatureFull accuracy from 0 °C to 50°C (Operating)
Operating humidity
Storage temperature–20 °C to 60 °C (Non-operating)
AltitudeOperating up to 2,000 metres (6,562 feet)
Pollution degreePollution Degree 2
The Keysight U3401A dual display complies with the following EMC
requirements:
– IEC 61010-1:2001/EN61010-1:2001 (2nd Ed ition)
– Canada: CAN/CSA-C22.2 No. 61010-1-04
– USA: ANSI/UL 61010-1:2004
– IEC 61326-1:2005/EN61326-1:2006
– Canada: ICES/NMB-001: Issue 4, June 2006
Full accuracy up to 80% R.H. (Relative humidity) for
temperature up to 28°C
– Australia/New Zealand: AS/NZS CISPR11:2004
6Keysight U3401A User’s and Service Guide
Page 7
Regulatory Markings
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM
device complies with the
Canadian ICES-001.
Cet appareil ISM est conforme a la
norme NMB-001 du Canada.
The CSA mark is a registered
trademark of the Canadian
Standards Association.
The RCM mark is a registered
trademark of the Spectrum
Management Agency of Australia.
This signifies compliance with the
Australia EMC Framework regulations
under the terms of the Radio
Communication Act of 1992.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical or electronic product in
domestic household waste.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life
of the product.
Keysight U3401A User’s and Service Guide7
Page 8
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/
96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
– www.keysight.com/find/low-cost-dmm
(product-specific information and support, software and documentation
updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight U3401A 4 ½ Digit Dual Display Multimeter
User’s and Service Guide
1Getting Started
Introducing the Keysight U3401A Dual Display Multimeter 18
Connecting power to the instrument 20
Stacking the U3401A 21
Adjusting the Handle 22
Product at a Glance 23
Making measurements 34
Selecting a Range 43
This chapter provides an introduction to the U3401A digital multimeter and a
tutorial showing how to use the front panel in order to make measurements.
17
Page 18
1Getting Started
Introducing the Keysight U3401A Dual Display Multimeter
The key features of the U3401A dual display multimeter are:
– 4 ½-digit dual display measurement
– Ten measurement functions:
–AC voltage
–DC voltage
– AC current
– DC current
–AC + DC voltage
– AC + DC current
– Resistance
– Frequency
– Continuity test
–Diode test
– Six math operations:
–dBm
–Min/Max
–Hold
– Relative (Rel)
– Compare (Comp)
– Percentage (%)
– 50,000 count dual display.
– AC and DC current measurements to 10 A.
– Resistance measurement up to 50 MΩ, 10 mΩ resolution.
– Frequency measurement up to 500 kHz with 0.01 Hz resolution.
– Measurement of dBm with variable reference impedance from 2 Ω to 8000 Ω.
18Keysight U3401A User’s and Service Guide
Page 19
Checking the shipping contents
NOTE
Verify that you have received the following items with your multimeter:
–Power cord
–Standard test lead kit
– Quick start guide
– Product reference CD
–Test report
– Certificate of calibration
Inspect the shipping container for damage. Signs of damage may include a
dented or torn shipping container or cushioning material that shows signs of
unusual stress or compacting.
Carefully remove the contents from the shipping container and verify that any
options ordered are included with the shipment by checking the packing list
included with the shipment.
If anything is missing, contact your nearest Keysight Sales Office.
Getting Started1
– If the shipping container or packaging material is damaged, it should be kept
until the contents have been checked mechanically and electrically. If there is
mechanical damage, notify the nearest Keysight Technologies office. Keep
the damaged shipping materials (if any) for inspection by the carrier and
Keysight representative. If required, you can find a list of Keysight Sales and
Service Offices on the last page of this guide.
– Ensure you read and understand the preceding safety information before
proceeding.
Original packaging
Containers and materials identical to those used in the factory packaging are
available through Keysight Technologies office. If the multimeter is being returned
to Keysight Technologies for servicing, attach a tag indicating the type of service
required, return address, model number, and serial number. Also mark the
container FRAGILE to assure careful handling. In any correspondence, refer to the
multimeter by model number and serial number.
Keysight U3401A User’s and Service Guide19
Page 20
1Getting Started
NOTE
CAUTION
Connecting power to the instrument
Connect the power cord and press the power switch to turn on the multimeter.
The front panel display illuminates while the multimeter performs its power-on
self-test. (If the multimeter does not power-on, refer “Operating Checklist” on
page 84). During the power-on session, press to hold the full display. Press
any key to resume the power-on self-test.
The multimeter powers up in the DC voltage function with autoranging enabled.
If self-test is successful, the multimeter goes to normal operation.If the self-test
fails, either a full annunciator or a blank display is displayed without entering the
normal operation. If the self-test repeatedly fails, contact your nearest Keysight
Sales and Service Office.
The multimeter will operate at any line voltage between 90 VAC and 264 VAC
when the line voltage selector is set properly with frequency range 50 Hz or 60 Hz.
– Before turning on the multimeter, make sure the line voltage selector is
set to the correct position for the applied line voltage to the power line
cord connector.
– Do not apply a line voltage that exceeds the specified range of the line
cord connector.
20Keysight U3401A User’s and Service Guide
Page 21
Stacking the U3401A
The U3401A is shipped with specially designed anti-slip protective bumpers on
the front and rear panel. The multimeters will not slide off when stacked on top of
each other.
To be able to stack the U3401A multimeters, ensure the attached bumpers are in
correct orientation. See Figure 1-1.
Getting Started1
Figure 1-1Stacking the U3401A
Keysight U3401A User’s and Service Guide21
Page 22
1Getting Started
Carrying position
Handle removal position
Adjusting the Handle
To adjust the handle, grasp the handle by the side and pull outward. Then, rotate
the handle to the desired position. Figure 1-2 below shows the possible handle
positions.
Benchtop
position
Figure 1-2Type of handle position
To attach or detach the handle, rotate the handle upright and pull it out from the
sides of the multimeter. Refer to Figure 1-3.
Figure 1-3Attaching and detaching the handle
22Keysight U3401A User’s and Service Guide
Page 23
Product at a Glance
226.0 mm
105.0 mm
305.0 mm
Product dimensions
Front view
Side view
Getting Started1
Figure 1-4U3401A dimensions
Keysight U3401A User’s and Service Guide23
Page 24
1Getting Started
Measurement
function
keypads
Math
operation
keypads
Primary display
Secondary display
Autoranging,
manual range,
and comparator
operation
Shift
Power
on/off
switch
Input terminals
and current fuse
The front panel at a glance
Figure 1-5Front panel
24Keysight U3401A User’s and Service Guide
Page 25
The display at a glance
Figure 1-6VFD full display with all segments illuminated
The highly visible vacuum fluorescent display (VFD) annunciators are described in
Table 1-1.
Tab le 1-1Display annunciators
AnnunciatorDescription
Primary display
Getting Started1
SReading rate: Slow. Not applicable for the U3401A.
MReading rate: Medium. Not applicable for the U3401A.
FReading rate: Fast. Not applicable for the U3401A.
PEAKPeak measurement. Not applicable for the U3401A.
HOLDData hold
MINMinMax math operation: Minimum value shown on the primary display
MAXMinMax math operation: Maximum value shown on the primary display
RELRelative value
dBmDecibel unit relative to 1 mW
4 2 W4-wire/2-wire resistance. Not applicable for the U3401A.
Diode test
Audible continuity test for resistance
AUTOAutoranging
Keysight U3401A User’s and Service Guide25
Page 26
1Getting Started
Tab le 1-1Display annunciators (continued)
AnnunciatorDescription
DCDirect current
ACAlternating current
DCACAC + DC
Polarity, digits, and decimal points for primary display
°CCelcius temperature unit. Not applicable for the U3401A.
°FFahrenheit temperature unit. Not applicable for the U3401A.
mVVoltage unit: mV, V
μmACurrent unit: μA, mA, A
μmnFCapacitance unit: μF, nF, mF
MkΩResistance unit: Ω, kΩ, MΩ
MkHzFrequency unit: Hz, kHz, MHz
Secondary d isplay
Polarity, digits, and decimal points for secondary display
μmnFCapacitance unit: μF, nF, mF
°CCelcius temperature unit. Not applicable for the U3401A.
°FFahrenheit temperature unit. Not applicable for the U3401A.
%Duty cycle measurement
mVVoltage unit: mV, V
μmACurrent unit: μA, mA, A
SShift mode
MkΩResistance unit: Ω, kΩ, MΩ
2NDSecondary display is enabled. Not applicable for the U3401A.
AUTOAutoranging
DCDirect current
ACAlternating current
26Keysight U3401A User’s and Service Guide
Page 27
Tab le 1-1Display annunciators (continued)
AnnunciatorDescription
DCACAC + DC
COMPCompare operation
dBmDecibel unit relative to 1 mW
CALCalibration mode
EXTExternal. Not applicable for the U3401A.
TRIGTrigger mode
REMOTERemote interface control. For calibration use only.
LOCALLocal mode
STOStore instrument state. Not applicable for the U3401A.
RCLRecall stored instrument state. Not applicable for the U3401A.
Getting Started1
Keysight U3401A User’s and Service Guide27
Page 28
1Getting Started
The keypad at a glance
The operation for each key is shown on Tab l e 1 - 2. Pressing a key changes the
current key operation, illuminates the related symbol on the display and emits a
beep.
Figure 1-7Keypad
Tab le 1-2Keypad functions
KeyDescription
System related operation
Press to power-on or power-off the U3401A multimeter.
Press to select Shift.
Press to return the multimeter to front panel operation when it is in remote state.
Press to step through the Setup menu. See “Using the setup menu” on page 59 for more
+
information.
Press to enable the secondary display.
28Keysight U3401A User’s and Service Guide
Page 29
Tab le 1-2Keypad functions (continued)
KeyDescription
Measurement related operation
Press to disable the secondary display.
+
Press to select the DC voltage measurement.
Press to select the AC voltage measurement.
Press to select the DC current measurement.
Press to select the AC current measurement.
Press to select the AC + DC voltage measurement.
+
Getting Started1
Press to select AC + DC current measurement.
+
Press to select the frequency measurement.
Press to toggle between the diode and continuity measurement.
Press to toggle between resistance or resistance continuity.
– Press to select dBm measurement.
+
– Press to select the reference impedance for dBm measurement.
Press to toggle between manual ranging and autoranging.
Press to select a higher range and disable autoranging. See “Selecting a Range” on page 43 for
more information.
Keysight U3401A User’s and Service Guide29
Page 30
1Getting Started
Tab le 1-2Keypad functions (continued)
KeyDescription
Measurement related operation
Press to select a lower range and disable autoranigng. See “Selecting a Range” on page 43 for
more information.
Press to select Compare math operation.
+
Press to select and set the high limit for Compare math operation.
+
Press to select and set the low limit for Compare math operation.
+
Press to enable Hold math operation. See “Hold” on page 53 on for more information.
Press to enable the MinMax math operation.
Press to select the Relative math operation.
Press to select the Percentage operation.
+
30Keysight U3401A User’s and Service Guide
Page 31
The terminals at a glance
CAUTION
NOTE
To avoid damaging this multimeter, do not exceed the rated input limit.
Getting Started1
Figure 1-8Input terminals
Voltages above 300 VAC may be measured only in circuits that are isolated from
mains. However, transient overvoltages are also present on circuits that are
isolated from mains. The Keysight U3401A is designed to safely withstand
occasional transient overvoltages up to 2500 V PEAK. Do not use this
multimeter to measure circuits where transient overvoltages could exceed this
level.
Miliampere (mA), frequency (Hz)mA500 mADC or AC RMS
COM
10 A, frequency (Hz)10 A
Resistance (Ω)
Diode test, continuity test
V-Hz normal mode, or
2x10
6
V-Hz common mode
1x10
10 ADC or AC RMS continuous, and
>10 A DC RMS for 20 seconds maximum
500 VDC or AC RMS
500 VDC or AC RMS
All functionsAny terminal to earth1000 VDC or AC PEAK
32Keysight U3401A User’s and Service Guide
Page 33
The rear panel at a glance
AC line
voltage selector
Line voltage fuse
selection table
AC power
connector
AC line fuseChassis
ground lug
RS232 interface connector
(For calibration use only)
Getting Started1
Figure 1-9Rear panel
Keysight U3401A User’s and Service Guide33
Page 34
1Getting Started
NOTE
Making measurements
The following pages show you how to make measurement connections and how
to select measurement functions from the front panel for each of the
measurement functions.
– After measuring a high voltage measurement of up to 1000 VDC, you are
recommended to wait for approximate two minutes before measuring a
low-level measurement with 1 to 10 μV resolutions
– After measuring a high current measurement using the A input terminal, you
are recommended to wait for approximate ten minutes before measuring a
low-level DC measurements of volts, amperes, or ohms; to achieve accurate
measurement. This is due to the thermal voltages generated during the high
current measurements that may cause errors when measuring the low-level
measurements.
34Keysight U3401A User’s and Service Guide
Page 35
Performing voltage measurements
WARNING
CAUTION
+
–
ACV
voltage
source
Exceeding the crest factor limit may result in an incorrect or a lower reading.
Do not exceed the crest factor limit to avoid instrument damage and the risk
of electric shock.
Ensure that the terminal connections are connected correctly before making
any measurement. To avoid damaging the multimeter, do not exceed the
rated input limit.
AC vol tage
– Five ranges: 500.00 mV, 5.00 V, 50.00 V, 500.00 V, 750.00 V
– Measurement method: AC coupled true RMS—measure the AC component
with up to 400 VDC bias on any range
– Crest factor: Maximum 3:1 at full scale
– Input impedance: 1 MΩ±2% parallel with capacitance <100 pF on all ranges
– Input protection: 750 V RMS on all ranges
Getting Started1
1 Press .
2 Connect the red and black test leads to the respective input terminals as
shown in Figure 1-10.3 Probe the test points and read the display. In autoranging mode, the multimeter
automatically selects the appropriate range and the measurement is displayed.
Figure 1-10ACV terminal connection and display
Keysight U3401A User’s and Service Guide35
Page 36
1Getting Started
+
DCV
voltage
source
–
DC Voltage
– Five ranges: 500.00 mV, 5.00 V, 50.00 V, 500.00 V, 1000.00 V
–Measurement method: Sigma Delta A-to-D converter
– Input impedance: 10 MΩ±2% range (typical)
– Input protection: 1000 V on all ranges
1 Press .
2 Connect the red and black test leads to the respective input terminals as
3 Probe the test points and read the display. In autoranging mode, the
shown in Figure 1-11.
multimeter automatically selects the appropriate range and the measurement
is displayed.
– Range is based on the voltage level of the signal, not frequency
– Measurement method: Reciprocal counting technique. AC coupled input using
AC voltage function
– Signal level: 10% of range to full scale input on all ranges
– Gate time: 0.1 s or 1 period of the input signal, whichever is longer
– Input protection: 1000 V RMS on all ranges
1 Press .
2 Connect the red and black test leads to the input terminal as shown in
Figure 1-14.
3 Probe the test points and read the display. In autoranging mode, the multimeter
automatically selects the appropriate range and the measurement is displayed.
Getting Started1
Keysight U3401A User’s and Service Guide39
Figure 1-14Frequency terminal connection and display
Page 40
1Getting Started
CAUTION
+
–
Test current
Resistance
Performing resistance/continuity measurements
Disconnect circuit power and discharge all high-voltage capacitors before
measuring resistance, or testing circuit continuity, to avoid damaging the
multimeter or the device under test.
– Measurement method: Two-wire ohms, open-circuit voltage limited to < +9 V
– Input protection: 500 V on all ranges
1 Press .
2 Connect the red and black test leads to the input terminal as shown in
3 Probe the test points (by shunting the resistor) and read the display.
Figure 1-15.
In autoranging mode, the multimeter automatically selects the appropriate
range and the measurement is displayed.
Figure 1-15Resistance /continuity terminal connection and display
40Keysight U3401A User’s and Service Guide
Page 41
Performing diode and continuity test
CAUTION
NOTE
The diode test measures the forward voltage of a semiconductor junction of
approximately 0.5 mA. The beeper will emit a single beep tone when the input
voltage is below +0.7 V (approximately 1.4 kΩ) and emits a continuos beep tone
when the input voltage is below 50 mV (approximately 100 Ω).
Disconnect circuit power and discharge all high-voltage capacitors before
testing diodes to avoid damaging the multimeter.
– Measurement method: 0.5 mA ± 0.2% constant current source, open-circuit
voltage limited to < +9 V
– Continuity threshold: 10 Ω fixed
– Input protection: 500 V DC or AC RMS
To test a diode, switch the circuit power off, and remove the diode from the circuit.
Then proceed as follows:
Getting Started1
1 Press to toggle the diode and continuity function. The default function
is diode measurement.
2 Connect the red and black test leads to the input terminal as shown in
Figure 1-16 on page 42.
3 Connect the red test lead to the positive terminal (anode) of the diode and the
black test lead to the negative terminal (cathode). Refer to Figure 1-16 on
page 42.
4 Read the display.
The cathode of a diode is indicated with a band.
5 Reverse the probes and measure the voltage across the diode again as shown
in Figure 1-17. Assess the diode according to the following guidelines:
– A diode is considered good if the multimeter displays OL in reverse bias
mode.
Keysight U3401A User’s and Service Guide41
Page 42
1Getting Started
Test current
Forward
bias
+
–
+
–
Reverse
bias
Test current
– A diode is considered shorted if the multimeter displays approximately
0 V in both forward and reverse bias modes, and the multimeter beeps
continuously.
– A diode is considered open if the multimeter displays OL in both forward
and reverse bias modes.
Figure 1-16Forward-biased diode/continuity test terminal connection and
display
Figure 1-17Reverse-biased diode/continuity terminal connection and
42Keysight U3401A User’s and Service Guide
display
Page 43
Selecting a Range
You can allow the multimeter to select the range automatically by using
autoranging, or you can select a fixed range using manual ranging. Autoranging is
convenient because the multimeter automatically selects the appropriate range
for sensing and displaying each measurement. However, manual ranging results
in better performance, since multimeter does not have to determine which range
to use for each measurement.
– The range setting is synchronous for dual display for current or voltage
measurement. In autoranging mode, the range setting for both the primary
and secondary display are corresponding to the higher range of two displays.
Getting Started1
Selects autoranging and disable manual ranging. Press to toggle
in and out of the manual ranging and autoranging.
Selects a higher range and disable autoranging.
Selects a lower range and disable autoranging.
– Manual ranging — If the input signal is greater than can be measured on the
selected range, the multimeter will display an overload indication, OL on the
primary or secondary display front panel.
– The multimeter remembers the selected ranging method (auto or manual) and
the selected manual range for each measurement function.
– Autorange thresholds - The multimeter shifts ranges as follows:
– Down range at < 5% of current range.
– Up range at > full scale of current range.
– Table 1-4 shows the summary of range scale values respectively.
Keysight U3401A 4 ½ Digit Dual Display Multimeter
User’s and Service Guide
2Operations and Features
Operating Math Functions 46
Using the Secondary Display 57
This chapter explains the various functions and features available on the U3401A
digital multimeter.
45
Page 46
2Operations and Features
Operating Math Functions
Table 2-1 presents a summary of the math operations that can be used with each
measurement function.
Tab le 2-1Math operations for different measurement functions
Measurement function
DCV
DCI
Resistance
ACV
ACI
Frequency
Diode/Continuity
dBmRelMinMaxCompHold%
—
——
—
—
——
Allowed math operations
– All math operations can be toggled on and off by selecting the same math
operation.
– Only one math operation can be turned-on at a time. When selecting another
math operation when one is already on, you are required to turn off the first
operation and then turn-on the second math operation.
– All math operations are automatically turned-off when changing the
measurement functions.
– Range changing is allowed for all math operations.
46Keysight U3401A User’s and Service Guide
Page 47
dBm
Operations and Features2
The logarithmic dBm (decibels relative to one milliwatt) scale is often used in RF
signal measurements. The multimeter’s dBm operation takes a measurement and
calculates the power delivered to a reference impedance (typically 50, 75 or 600 Ω).
The formula used for conversion from the voltage reading is:
dBm = 10 x Log
Figure 2-1Typical dBm operation display
The default reference impedance value is 600 Ω.. You can select the following
21 reference impedance values:
The dBm operation can be applied to DCV, ACV and DCV + ACV measurement
functions only. The multimeter displays the dBm modifier on the primary display
and displays the reference impedance selection on the secondary display.
[1000 x (Reading2 / reference impedance)]
10
Procedure 1
1 Press to enable the dBm modifier mode and dBm operation.
2 Use and to select the desired reference impedance.
Keysight U3401A User’s and Service Guide47
Page 48
2Operations and Features
Rel
When making Rel (relative) measurements, each reading is the difference
between a stored relative value and the input signal. For example, this feature can
be used to make more accurate resistance measurements by nulling the test lead
resistance.
After you enable the Rel operation, the multimeter stores the next reading as
a Rel # (reference value) and immediately displays on the primary display:
Primary Display = Reading – Rel #
When the Rel operation is enabled in autoranging mode, enabling the Comp or
Percentage operation will disable the Rel operation.
Procedure
1 Press to exit the relative mode.
Figure 2-2Typical Rel operation display
48Keysight U3401A User’s and Service Guide
Page 49
MinMax
Elapsed time
Elapsed time
Operations and Features2
The MinMax (Minimum/Maximum) operation stores the minimum and maximum
values of reading during a series of measurements.
When enabled, the MinMax operation turns on the MINMAX annunciator and
begins accumulating various statistics of the readings being displayed.
Figure 2-3Typical Max operation display
Figure 2-4Typical Min operation display
Each time a new minimum or maximum value is stored, the multimeter beeps
once (if the beeper is enabled) and briefly turns on the appropriate MAX or MIN
annunciator.
Accumulated statistics are:
– MAX—maximum reading since MinMax was enabled
– MIN—minimum reading since MinMax was enabled
– MINMAX—actual readings
Keysight U3401A User’s and Service Guide49
Page 50
2Operations and Features
The elapsed time is recorded when the MinMax operation is selected. The elapsed
time is shown on the secondary display with HH.MM.SS.
– HH is 0-19 hours
– MM is 0-59 minutes
– SS is 0-59 seconds
Procedure
1 Press to enable MinMax operation.
2 Press continuously until the desired operation appear. The MinMax
operation will display MINMAX > MAX > MIN > MINMAX in sequence when
this key is press continuously.
3 Press to disable the MinMax operation.
50Keysight U3401A User’s and Service Guide
Page 51
Comp
Operations and Features2
The Comp (compare) operation allows you to perform pass/fail testing against
specified upper and lower limits. You can set the upper and lower limits to any
value between 0 and ±100% of the highest range for the present operation.
Figure 2-5Typical Comp operation display
When enabled, the actual readings are shown in primary display and the
comparison results such as HI, LO or PASS is shown in secondary display.
– You should specify the upper limit to always be a more positive number than
the lower limit. The initial factory setting for each limit is 0.
– The secondary display shows PASS when readings are within the specified
limits. The secondary display shows HI when the reading is outside the high
limit and LO when the reading is outside the low limit.
– When the beeper is ON (see “Using the setup menu” on page 59), the beeper
beeps three tones on the transition from PASS to HI or LO. When transitioning
from HI or LO to PASS, it will only beep once.
– Press to enable Comp operation.
Keysight U3401A User’s and Service Guide51
Page 52
2Operations and Features
Procedure
1 Press to enter the upper limit setup mode. The upper limit is
shown on primary display while the HI annunciator is shown on secondary
display.
2 Use , , , and to modify the upper limit.
3 Press to store the specified HI limit value.
4 Press to enter the lower limit setup mode.
The lower limit is shown on primary display while the LO annunciator is shown
on secondary display.
5 Use , , , and to modify the lower limit.
6 Press to store the specified LO limit value.
52Keysight U3401A User’s and Service Guide
Page 53
Hold
Operations and Features2
The reading hold feature allows you to capture and hold a reading on the front
panel display. When enabled, the Hold operation turns on the HOLD annunciator
and hold the reading.
Procedure
1 Press to hold the reading on the display.
Figure 2-6Typical Hold operation display
Refresh Hold
The refresh hold operation allows you to take measurement in dangerous or
difficult measuring field and you could not look at the display. This operation will
update the hold value with a new measurement value automatically, and thus
reminds you with a beep.
Procedure
1 Press to access to the Setup menu.
2 Use and to select the rHold.
3 Press to enter the second tier menu. Use and to set the
state to ON.
Keysight U3401A User’s and Service Guide53
Page 54
2Operations and Features
NOTE
4 Press to confirm.
5 Press twice quit the Setup menu.
6 Press to select refresh hold mode. The present value will be held and the
HOLD will be lit. The multimeter is now ready to hold a new measurement
value once the variation of measuring value exceeds the setting of the
variation count. The HOLD annunciator will be blinking continuously.
7 When the refresh hold is not in use, repeat step 1 to step 4 to turn off the
refresh hold’s state.
Figure 2-7Typical refresh hold operation display
– For voltage and current measurements, the held value will not be updated
when the reading is below 500 counts.
– For resistance and diode measurements, the held value will not be updated
when the reading is at OL or open state.
54Keysight U3401A User’s and Service Guide
Page 55
Percentage (%)
This operation allows you to transfer the measurement value with a proportional
percentage display. You can set the upper limit (HI) and lower limit (LO) to any
value between 0 and ±100% of the highest range for the present function.
The percentage operation calculates and display the measurement value using
the equation below:
Percentage (%) = [Measurement value – LO/(HI-LO)] x 100%]
When the HI limit is equal to LO limit, the percentage operation will use the
equation as below:
Percentage (%) = [(Measurement value – HI)/HI] x 100%]
When enabled, the actual readings are shown in primary display and the
percentage’s result is shown in secondary display.
– The secondary display shows OL when readings exceed the maximum display
of 999.99%.
– When autoranging is enabled, this operation will be used to lock the current
range.
Keysight U3401A User’s and Service Guide55
Page 56
2Operations and Features
Procedure
1 Press to enter the upper limit setup mode. The upper limit is
shown on primary display while the HI annunciator is shown on secondary
display.
2 Use , , , and to modify the upper limit.
3 Press to store the specified HI limit value.
4 Press to enter the lower limit setup mode.
The lower limit is shown on primary display while the LO annunciator is shown
on secondary display.
5 Use , , , and to modify the lower limit.
6 Press to store the specified LO limit value.
7 Press to enable percentage operation.
8 Press to disable percentage operation.
56Keysight U3401A User’s and Service Guide
Page 57
Using the Secondary Display
Secondary
display
Figure 2-9Secondary display
To enable the secondary d isplay mode:
Repeatedly pressing cycles through the secondary display choices for the
present math operation as shown in Table 2-2 on page 58.
To disable the secondary d isplay mode:
Operations and Features2
Press . The display will remain in primary display mode.
Keysight U3401A User’s and Service Guide57
Page 58
2Operations and Features
Tab le 2-2Description for dual display combination
Secondary display
Primary display
DCVFrequency
ACVFrequency
ACV + DCVFrequency
Defaul t
secondary
display
Press secondary
display once
[b]
ACV
[b]
DCV
[b]
ACV
Press secondary
display twice
dBm—
dBm—
[b]
DCV
dBmACVDVCACV+DCV—
DCIFrequency
ACIFrequency
ACI + DCIFrequency
Frequency
[a]
ACVACI——
ACI
DCI
ACI
[b]
[b]
[b]
——
——
[b]
DCI
COMP (Measuring Value)HI, LO, PASS———
Percentage (Measuring Value)%— ——
Notes:
[a] The Frequency reading is corresponding to the current or voltage input signal respectively. The Frequency measurement is always working
on autoranging mode. The Buttons of AUTO, UP and DOWN are used to select the range of measuring signal. If the current measurement is
set before selecting Hz function, the default will show ACI first, else show ACV.
[b] In autoranging mode, the ranges setting of both the primary and secondary display are corresponding to the higher range of two displays.
Whereas in manual range, the ranges of secondary display are the same as the primary display.
Press
secondary display
three times
dBm
—
58Keysight U3401A User’s and Service Guide
Page 59
Using the setup menu
The Setup menu allows you to customize a number of non-volatile instrument
configurations. The content of the Setup menu are shown in Table 2-3.
.
Tab le 2-3Setup menu and communications parameters
Operations and Features2
First tier
menu
rS232
Ref Ω
bEEPBeeper selection
rHoldRefresh hold
Notes:
[a] For calibration use only.
[b] Variation count is used to recognize new value will be updated once the variation of measuring value is esceeded the setting value.
The parameters in the Setup Menu can be configured by using the following
procedures:
1 Press to access to the Setup Menu.
2 Press and to select the desired configurable items in the first tier
menu.
3 Press to enter second tier menu. The original parameter is indicated in
primary display.
4 Use or to select the desired parameter.
5 Press to confirm the selected parameter or press to quit from the
tier menu without saving the changes.
6 After saving the changes, press to quit from the tier menu.
7 Press again to quit from the setup menu.
8 The configuration settings is saved and main display is displayed.
60Keysight U3401A User’s and Service Guide
Page 61
Trigger mode
This multimeter has two types of trigger mode.
– Internal — triggers measurements continuously.
– External — triggers a measurement only when a direction is given.
The external trigger is used with a delay settling that has been set by the
multimeter automatically. The amount of trigger delay varies depending on
different function. When external trigger is enabled, the multimeter determines
the range for the primary display based on the input at that time.
Autoranging response time is required before a reading is displayed when you
change the input after a trigger is received.
Procedure
1 Press to enter trigger mode and standby condition. The TRIG
annunciator will be lit and display is indicated with “–––––”.
2 Press to get a new value. After a measurement, the result will be
indicated and held on the display.
Operations and Features2
3 Press to get a new value again.
4 Press , or to select autoranging or manual range as
necessary.
5 Press to disable external trigger operation mode.
Figure 2-10Typical trigger mode display
Keysight U3401A User’s and Service Guide61
Page 62
2Operations and Features
NOTE
– All math operations will be disabled when external trigger is enabled.
– Press may also disable external trigger.
– Both the primary and secondary display will only appear when frequency
operation is selected in trigger mode.
62Keysight U3401A User’s and Service Guide
Page 63
Keysight U3401A 4 ½ Digit Dual Display Multimeter
User’s and Service Guide
3Application Tutorial
Applications for Using Dual Display 64
Dual Display Operation Examples 65
This chapter describes the advanced features and the possible applications for
effective operation of the multimeter.
63
Page 64
3Application Tutorial
Applications for Using Dual Display
The dual display feature in the multimeter can be used to enhance test and
measurement capabilities. See Tab le 3-1 for the available combinations and
application when using dual display.
Tab le 3-1Typical combinations and applications when using dual display
No.Primary d isplaySecondary displayApplications
1DCVACV– Test DC to AC or AC to DC converter circuit.
2ACV + DCVDCV
3ACVFrequency– Measuring AC frequency response of amplifier circuit.
4ACIFrequency
5DCIACI– Measuring AC ripple and DC current of power supply.
6ACI + DCIDCI
7dBmReference Ω– Setting dB reference impedance and show dBm.
8dBmDCV– Indicating DC voltage and dBm.
– Measure DC level and AC ripple of power supply.
– Adjusting AC motor control.
– Measuring current dissipation for power supply analysis.
9dBmACV– Indicating AC voltage and dBm.
64Keysight U3401A User’s and Service Guide
Page 65
Dual Display Operation Examples
NOTE
This section describes some practical operations when using dual display feature.
Measure DC Voltage and AC Ripple on a Rectification Circuit
A single measurement for both DC voltage and AC ripple can be displayed
through both displays while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3-1.
+
AC/DC v olta ge
source
–
Application Tutorial3
Figure 3-1Connection terminal when measuring DC voltage and AC ripple
2 Press to select DC voltage measurement in primary display.
3 Press to select AC voltage measurement for secondary display.
4 Press . Use and to select autoranging or manual range for
secondary display.
– Press to disable the secondary display.
– Press to select the suitable range if DCV+ACV ripple is above the scale
of the current range.
Keysight U3401A User’s and Service Guide65
Page 66
3Application Tutorial
WARNING
NOTE
Measure AC and DC Current on a Rectification Circuit
A single measurement for both AC current and DC current can be displayed
through both displays while testing a rectifier circuit.
– Make sure you select the correct input terminal according to the input
range used.
– Do not apply current exceeding specified range to input terminals of mA
or A.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3-2.
+
AC/DC current
source
–
Figure 3-2Connection terminal when using DC current and AC ripple
2 Press to select AC current measurement for primary display.
3 Press to enable the DC current measurement for secondary display.
4 Press , or to select autoranging or manual range for
secondary display.
Press to disable the secondary display.
66Keysight U3401A User’s and Service Guide
Page 67
Measure AC Voltage and Frequency on an AC Circuit
NOTE
A single measurement for both AC voltage and frequency can be displayed
through both displays while testing a rectifier circuit.
1 Connect the red and black test leads to the input terminal and probe the test
points as shown in Figure 3-3.
+
AC vol tage
source
–
Figure 3-3Connection terminal when measuring AC Voltage and
Frequency
Application Tutorial3
2 Press to select AC voltage measurement for primary display.
3 Press to select frequency measurement for secondary display.
4 Press , or to select autoranging or manual range for
secondary display.
Press to disable the secondary display.
Keysight U3401A User’s and Service Guide67
Page 68
3Application Tutorial
WARNING
Measure Resistance
Do not apply voltage exceeding 500 V PEAK between and COM
input terminals.
1 Connect a resistor under test to and LO input terminals as shown in
Figure 3-4.
+
Resistance
–
Figure 3-4Connection terminal when measuring Resistance
2 Press to select resistance measurement.
3 Press . Use and to select autoranging or manual range for
primary display.
68Keysight U3401A User’s and Service Guide
Page 69
Measure True RMS AC + DC
NOTE
The multimeter can measure the true RMS of AC voltage and AC current.
1 Press , and or and simultaneously. The
multimeter will measure the DC and AC signals alternatively, calculate and
display the AC+DC RMS value using the quation below:
AC+DC (RMS) = DC
When AC+DC voltage measurement is selected, the DCV input impedance is
paralleled with the AC coupled 1.1 MΩ AC divider.
2
+ AC
Application Tutorial3
2
Keysight U3401A User’s and Service Guide69
Page 70
3Application Tutorial
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
70Keysight U3401A User’s and Service Guide
Page 71
Keysight U3401A 4 ½ Digit Dual Display Multimeter
WARNING
User’s and Service Guide
4Performance Test
Calibration Overview 72
Recommended Test Equipment 73
Test Considerations 74
Performance Verification Tests Overview 75
This chapter contains performance test procedures and calibration procedures.
The performance tests procedures allow you to verify that the multimeter is
operating within its published specifications.
– Shock hazard. Only service-trained personnel who are aware of the
hazards involved should perform the procedures in this chapter. To avoid
electrical shock and personal injury, make sure to read and follow all test
equipment safety instructions.
– Use only completely electrically insulated test lead sets with connectors
that prevent contact with test voltages.
71
Page 72
4Performance Test
NOTE
Calibration Overview
Make sure you have read “Test Considerations” on page 74 before calibrating the
instrument.
Keysight Technologies calibration services
When your instrument is due for calibration, contact your local Keysight Service
Center. The U3401A is supported on automated calibration systems at Keysight
service centres only.
Calibration interval
A one-year interval is adequate for most applications. Accuracy specifications are
warranted only if adjustment is made at regular calibration intervals. Accuracy
specifications are not warranted beyond the one-year calibration interval.
Keysight does not recommend extending calibration intervals beyond two years
for any application. When an adjustment is required, contact your local Keysight
Service Center.
72Keysight U3401A User’s and Service Guide
Page 73
Performance Test4
Recommended Test Equipment
The test equipments recommended for the performance verification procedures
are listed below. If the exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternate method would be to use the Keysight 3458A 8½ -digit
digital multimeter to measure less accurate yet stable sources. The output value
measured from the source can be entered into the instrument as the target
calibration value.
Zero calibrationShorting plug—Dual banana plug with copper wire short
between the two terminals
DC voltageFluke 5520A< 1/5 instrument 1 year spec
DC currentFluke 5520A< 1/5 instrument 1 year spec
ResistanceFluke 5520A< 1/5 instrument 1 year spec
AC voltageFluke 5520A< 1/5 instrument 1 year spec
AC currentFluke 5520A< 1/5 instrument 1 year spec
—
FrequencyFluke 5520A< 1/5 instrument 1 year spec
DiodeFluke 5520A< 1/5 instrument 1 year spec
Keysight U3401A User’s and Service Guide73
Page 74
4Performance Test
NOTE
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations:
– Ensure that the calibration ambient temperature is stable and between 18°C
and 28°C. Ideally the calibration should be performed at 23°C ±1°C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 30 minutes warm-up period with a Shorting Plug connected to the Hi
and Lo input terminals.
– Use shielded twisted pair PTFE-insulated cables to reduce settling and noise
errors. Keep the input cables as short as possible.
– Connect the input cable shields to earth ground. Connect the calibrator Lo
source to earth ground at the calibrator except where noted in the procedures.
It is important that the Lo to earth ground connection be made at only one
place in the circuit to avoid ground loops.
Please ensure that the calibration standards and test procedures used do not
introduce add itional errors. Ideally, the standards used to verify and adjust the
instrument should be an order of magnitude more accurate than each
instrument range full-scale error specification.
74Keysight U3401A User’s and Service Guide
Page 75
Performance Verification Tests Overview
NOTE
Performance verification tests is an extensive set of tests that are recommended
as an acceptance test when you first receive the instrument.
Use the performance verification test to verify the measurement performance of
the instrument. The performance verification test uses the instrument’s
specifications listed in Chapter 6, “Characteristics and Specifications”.
Performance verification tests
The performance verification test is recommended as acceptance tests when you
first received the instrument. The acceptance test results should be compared
against the one-year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
If the instrument fails performance verification, adjustment or repair is required.
Contact your local Keysight Service Center for details.
Make sure you have read “Test Considerations” on page 74 before doing the
performance verification tests.
Performance Test4
This test checks the full-scale reading accuracy of the instrument.
DC Voltage Verification Test
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in Table 4-2. Provide the
input shown in Table 4-2
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in the Tab le 4-2. (Be certain to allow for
appropriate source settling when using the Fluke 5520A).
Keysight U3401A User’s and Service Guide75
Page 76
4Performance Test
CAUTION
Tab le 4-2DC voltage verification test
FunctionInputRangeError from nominal one year
DC voltage0.00 V500 mV±40 μV
0.0000 V5 V±0.4 mV
0.000 V50 V±4 mV
0.00 V500 V±40 mV
0.0 V1000 V±400 mV
450.00 mV500 mV±130 μV
4.5000 V5 V ±1.3 mV
45.000 V50 V±13 mV
450.00 V500 V±130 mV
900.0 V1000 V±580 mV
Set the calibrator output to 0 V before disconnecting it from the multimeter
input terminals.
DC Current Verification Test
1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in Table 4-3. Provide the
input shown in Table 4-3.
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in Tab le 4 - 3 . (Be certain to allow for
appropriate source settling when using the Fluke 5520A).
76Keysight U3401A User’s and Service Guide
Page 77
Performance Test4
CAUTION
Tab le 4-3DC current verification test
FunctionInputRangeError from nominal one year
DC current0.00 μA500 μA±0.05 μA
0.0000 mA5 mA±0.4 μA
0.000 mA50 mA±4 μA
0.00 mA500 mA±40 μA
450.00 μA500 μA±0.27 μA
4.5000 mA5 mA±2.6 μA
45.000 mA50 mA±26 μA
450.00 mA500 mA±0.26 mA
0.0000 A5 A±0.5 mA
0.000 A10 A±5 mA
4.5000 A5 A±11.7 mA
9.000 A10 A±27 mA
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.
Keysight U3401A User’s and Service Guide77
Page 78
4Performance Test
Resistance Verification Test
Configuration: 2-Wire Ω
1 Select the resistance function.
2 Select each range in the order shown in Table 4-4. Provide the resistance value
indicated. Compare measurement results to the appropriate test limits shown
in Table 4-4. (Be certain to allow for appropriate source settling).
Tab le 4-42-wire Ω verification test
FunctionInputRangeError from nominal one year
2-wire Ω
0.00 Ω500 Ω
0.0000 kΩ5 kΩ
0.000 kΩ50 kΩ±3 Ω
0.00 kΩ500 kΩ±30 Ω
0.0000 MΩ5 MΩ±300 Ω
0.000 MΩ50 MΩ±3 kΩ
450.00 Ω500 Ω
4.5000 kΩ5 kΩ
±50 mΩ
[a]
±0.3 Ω
±500 mΩ
[a]
±4.8 Ω
[a]
[a]
45.000 kΩ50 kΩ±48 Ω
450.00 kΩ500 kΩ±480 Ω
4.5000 MΩ5 MΩ±4.8 kΩ
30.000 MΩ50 MΩ±93 kΩ
Notes:
[a] Specifications are for 2-wire ohms function using the Rel operation enabled to eliminate lead resistance.
78Keysight U3401A User’s and Service Guide
Page 79
Performance Test4
Diode Verification Test
Configuration: Diode
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in Table 4-5. Provide the
input shown in Table 4-5.
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in Table 4-5. (Be certain to allow for
appropriate source settling when using the Fluke 5520A).
Tab le 4-5Diode verification test
FunctionVoltageRangeError from nominal one year
Diode1.0000 V2.3 V±1 mV
2.0000 V2.3 V±1.5 mV
Frequency Verification Test
Configuration: Frequency
1 Select the frequency function.
2 Select each range in the order shown in Table 4-6. Provide the input voltage
and frequency indicated. Compare measurement results to the appropriate
test limits shown in Tab le 4-6. (Be certain to allow for appropriate source
settling).
Tab le 4-6Frequency verification test
FunctionVol tageInput frequencyRangeError from nominal one year
Frequency2 V4.5 kHz5 kHz±0.7 Hz
Keysight U3401A User’s and Service Guide79
Page 80
4Performance Test
CAUTION
AC Voltage Verification Test
Configuration: AC volts
1 Select the AC voltage function.
2 Select each range in the order shown in Tab l e 4-7. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate test
limits shown in Tab le 4 - 7 . (Be certain to allow for appropriate source settling).
Tab le 4-7AC volts verification test
FunctionV RMSInput frequencyRangeError from nominal one year
AC voltage50.00 mV1 kHz500 mV±650 μV
450.00 mV1 kHz500 mV±2.65 mV
0.5000 V1 kHz5 V±3.2 mV
4.5000 V1 kHz5 V±17.2 mV
5.000 V1 kHz50 V±32 mV
45.000 V1 kHz50 V±172 mV
50.00 V1 kHz500 V±400 mV
450.00 V1 kHz500 V±2.4 V
75.0 V1 kHz750 V±1.8 V
675.0 V1 kHz750 V±4.8 V
Set the calibrator output to 0 V before disconnecting it from the multimeter
input terminals.
80Keysight U3401A User’s and Service Guide
Page 81
Performance Test4
AC Current Verification Test
Configuration: AC current
1 Select the AC current function.
2 Select each range in the order shown in Table 4-8. Provide the input current
and frequency indicated. Compare measurement results to the appropriate test
limits shown in Table 4-8. (Be certain to allow for appropriate source settling).
Tab le 4-8AC current verification test
FunctionCurrentInput frequencyRangeError from nominal one year
AC current50.00 μA1 kHz500 μA±0.45 μA
450.00 μA1 kHz500 μA±2.45 μA
0.5000 mA1 kHz5 mA±4.5 μA
4.5000 mA1 kHz5 mA±24.5 μA
5.000 mA1 kHz50 mA±45 μA
45.000 mA1 kHz50 mA±0.245 mA
50.00 mA1 kHz500 mA±0.45 mA
450.00 mA1 kHz500 mA±2.45 mA
0.5000 A1 kHz5 A±5.5 mA
1.000 A1 kHz10 A±35 mA
4.5000 A1 kHz5 A±25.5 mA
9.000 A1 kHz10 A±75 mA
Keysight U3401A User’s and Service Guide81
Page 82
4Performance Test
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
82Keysight U3401A User’s and Service Guide
Page 83
Keysight U3401A 4 ½ Digit Dual Display Multimeter
User’s and Service Guide
5Disassembly and Repair
Operating Checklist 84
Types of Service Available 85
Mechanical Disassembly 89
Replaceable Parts 95
This chapter will help you troubleshoot a faulty multimeter. It describes how to
disassemble the multimeter, how to obtain repair services, and lists the
replaceable parts.
83
Page 84
5Disassembly and Repair
Operating Checklist
Before returning your multimeter to Keysight for service or repair check the
following items:
Is the multimeter inoperative?
Verify the power line voltage setting.
?
Verify the power line fuse is installed.
?
Check the power cord is connected to multimeter and to AC line power.
?
Verify the front panel power switch is depressed.
?
See page 86
Is the mutimeter’s current input inoperative?
Verify the current input fuse.
?
84Keysight U3401A User’s and Service Guide
Page 85
Types of Service Available
If your instrument fails during the warranty period, Keysight Technologies will
repair or replace it under the terms of your warranty. After your warranty expires,
Keysight offers repair services at competitive prices.
Extended service contracts
Many Keysight products are available with optional service contracts that extend
the covered period after the standard warranty expires. If you have such a service
contract and your instrument fails during the covered period, Keysight
Technologies will repair or replace it in accordance with the contract.
Obtaining repair service (worldwide)
To obtain service for your instrument (in-warranty, under service contract, or
post-warranty), contact your nearest Keysight Technologies Service Center. They
will arrange to have your instrument repaired or replaced, and can provide
warranty or repair-cost information where applicable.
To obtain warranty, service, or technical support information you can contact
Keysight Technologies at one of the following telephone numbers:
In the United States: (800) 829–4444
Disassembly and Repair5
In Europe: 31 20 547 2111
In Japan: 0120–421–345
Or use our Web link for information on contacting Keysight worldwide:
www.keysight.com/find/assist
Or contact your Keysight Technologies representative.
Before shipping your instrument, ask the Keysight Technologies Service Center to
provide shipping instructions, including what components to ship. Keysight
recommends that you retain the original shipping carton for use in such
shipments.
Keysight U3401A User’s and Service Guide85
Page 86
5Disassembly and Repair
Repackaging for shipment
If the instrument is to be shipped to Keysight for service or repair, be sure to:
– Attach a tag to the unit identifying the owner and indicating the required
service or repair. Include the model number and full serial number.
– Place the unit in its original container with appropriate packaging material for
shipping.
– Secure the container with strong tape or metal bands.
– If the original shipping container is not available, place your instrument in a
container which will ensure at least 4 inches of compressible packaging
material around all sides for the instrument. Use static-free packaging
materials to avoid additional damage to your instrument.
Keysight suggests that you always insure shipments.
Cleaning
Clean the outside of the multimeter with a soft, lint-free, slightly dampened cloth.
Do not use detergent. Disassembly is not required or recommended for cleaning.
86Keysight U3401A User’s and Service Guide
Page 87
Disassembly and Repair5
2 Remove the line voltage selector from
fuse holder assembly.
1 Disconnect the power cord. Depress tab 1 and
2 and pull the fuse holder from rear panel.
1
2
4 Replace fuse holder assembly in
rear panel.
3 Rotate the line voltage selector and
reinstall it back, so that the correct voltage
appears in the fuse holder window.
To replace the power line fuse
The power line fuse is located within the multimeter’s fuse-holder assembly on
the rear panel. The multimeter is shipped from the factory with a power-line fuse
installed (according to country of destination). See Table 5-1. If you determine
that the fuse is faulty, replace it with one of the same size and rating.
Tab le 5-1Type of supplied fuse (according to country of destination)
Type of fuse (time-lag, low breaking fuse)Input line vol tagePart number
0.25A, 250V, 5x20 mm100 V to 120 VA02-62-25592-3U
0.125A, 250V, 5x20 mm220 V to 240 VA02-62-25648-1U
Keysight U3401A User’s and Service Guide87
Page 88
5Disassembly and Repair
To replace a current input fuse
Both the mA and the A current input terminals are fuse protected. The fuse for the
mA input terminal is located on the front panel (see page 32). The fuse is a 0.63 A,
500 V fuse (refer to Table 5-2). If you determine that the fuse is faulty, replace it
with one of the same size and rating.
The fuse for the A current input terminal is located inside the multimeter (see
page 93) and requires partial disasembly of the multimeter. The fuse is a 25 A,
440 V fast-acting fuse (refer to Table 5-2). If you determine that the fuse is faulty,
replace it with one of the same size and rating.
Tab le 5-2Type of current input fuse
Type of fusePart number
0.63 A, 500 V fuse2110-1432
25 A, 440 V fast-acting fuse2110-1431
Electrostatic Discharge (ESD) precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when servicing the
instrument or any electronic device.
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastics, foam, vinyl, paper, and other static-generating materials
from the immediate work area.
– Use only anti-static solder suckers.
88Keysight U3401A User’s and Service Guide
Page 89
Mechanical Disassembly
WARNING
For procedures in this manual, the following tools are required for disassembly:
– T15 Torx driver
– T20 Torx driver (most disassembly)
– #2 Pozi-drive screw driver
Shock hazard. Only service–trained personnel who are aware of the hazards
involved should remove the instrument covers. To avoid electrical shock and
personal injury, make sure to disconnect the power cord from the instrument
before removing the covers. Some circuits are active and have power applied
even when the power switch is turned off.
General Disassembly
1 Remove power and all cables from the instrument.
2 Remove the carrying handle by rotating the handle upright and pulling out
from the sides of the instrument.
Disassembly and Repair5
Keysight U3401A User’s and Service Guide89
Page 90
5Disassembly and Repair
3 Remove the instrument bumpers. Pull from a corner and stretch the bumpers
off the instrument.
4 Remove the rear bezel. Loosen the two captive screws in the rear bezel and
remove the rear bezel.
90Keysight U3401A User’s and Service Guide
Page 91
Disassembly and Repair5
5 Remove the cover. Remove the screw in the bottom of the cover and slide the
cover off the instrument.
Front Panel Removal
1 Remove on/off switch push rod. Gently move the power switch push rod
toward the front of the instrument to disengage it from the switch. Be careful
not to twist or bend the push rod.
Keysight U3401A User’s and Service Guide91
Page 92
5Disassembly and Repair
2 Remove the screw holding the front panel.
3 Disconnect the two ribbon cable connectors from the front panel.
92Keysight U3401A User’s and Service Guide
Page 93
Disassembly and Repair5
25 A current
fuse
4 Disconnect the individual front panel wires shown below.
5 There is now enough play to allow the side of the front panel to be pried from
the chassis and removed as an assembly.
Keysight U3401A User’s and Service Guide93
Page 94
5Disassembly and Repair
Front Panel Disassembly
1 Remove the keyboard and display assembly. Remove the two screws holding
the circuit board. Lift the keyboard and display assembly from the plastic
housing.
a The rubber keypad can now be pulled from the plastic housing.
94Keysight U3401A User’s and Service Guide
Page 95
Replaceable Parts
This section contains information for ordering replacement parts for your
instrument. You can find the instrument support part list at Keysight’s
Test & Measurement Parts Catalog at http://www.keysight.com/find/parts.
This parts list includes a brief description of each part with applicable Keysight
part number.
To order replaceable parts
You can order replaceable parts from Keysight using the Keysight part number.
Note that not all parts listed are available as field-replaceable parts.
To order replaceable parts from Keysight, do the following:
1 Contact your nearest Keysight Sales Office or Service Center.
2 Identify the parts by the Keysight part number shown in the support parts list.
3 Provide the instrument model number and serial number.
Disassembly and Repair5
Keysight U3401A User’s and Service Guide95
Page 96
5Disassembly and Repair
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
96Keysight U3401A User’s and Service Guide
Page 97
Keysight U3401A 4 ½ Digit Dual Display Multimeter
User’s and Service Guide
6Characteristics and
Specifications
For the characteristics and specifications of the U3401A 4 ½ Digit Dual Display
Multimeter, refer to the datasheet at