Keysight U2722A, U2723A Service Manual

Page 1
Keysight U2722A/U2723A USB Modular Source Measure Units
Service Guide
Page 2
Page 3
Page 4
Notices
CAUTION
WARNING
© Keysight Technologies 2009-2017 No part of this manual may be
reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Trademarks
Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.
Manual Part Number
U2722-90021
Edition
Edition 8, December 1, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/find/
sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
4 Keysight U2722A/U2723A Service Guide
Page 5
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions that must be taken to maintain safe operation of the instrument.
Direct current (DC)
Alternating current (AC) Off (supply)
Both direct and alternating current On (supply)
Three-phase alternating current Caution, risk of electric shock
Earth (ground) terminal
Equipment protected throughout by double insulation or reinforced insulation
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Protective conductor terminal Caution, hot surface
Frame or chassis terminal Out position of a bi-stable push control
Equipotentiality In position of a bi-stable push control
Keysight U2722A/U2723A Service Guide 5
Page 6
General Safety Information
WARNING
CAUTION
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these requirements.
– Do not use this product in any manner not specified by the manufacturer.
The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
– Do not operate the product in an explosive atmosphere or in the presence
of flammable gases or fumes.
– Only qualified, service-trained personnel who are aware of the hazards
involved should remove the instrument casing. Always disconnect the power cable and any external circuits before removing the instrument casing.
– Do not use the equipment if it does not operate properly. Ensure that the
equipment is inspected by qualified service personnel.
– Do not install substitute parts or perform any unauthorized modification
to the product. Return the product to n Keysight Sales and Service Office for service and repair to ensure that the safety features are maintained.
– Make all connections to the unit before applying power.
– Take note of the instrument's external markings described under Safety
Symbols.
– Use the device with the provided cables.
– Repair or service that is not covered in this manual should only be
performed by qualified personnel.
– Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
6 Keysight U2722A/U2723A Service Guide
Page 7
Environmental Conditions
CAUTION
This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental conditions Requirements
Operating temperature 0 °C to 50 °C
Operating humidity 20% to 85% RH non-condensing
Storage temperature –20 °C to 70 °C
Storage humidity 5% to 90% RH non-condensing
The U2722A/U2723A USB Modular Source Measure Units complies with the following safety and EMC requirements.
– IEC 61326-1:2005/EN61326-1:2006
– Canada: ICES-001:2004
– Australia/New Zealand: AS/NZS CISPR11:2004
– IEC 61010-1:2001/EN 61010-1:2001 (2nd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-04
– USA: ANSI/UL 61010-1:2004
Keysight U2722A/U2723A Service Guide 7
Page 8
Regulatory Markings
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical/electronic product in domestic household waste.
8 Keysight U2722A/U2723A Service Guide
Page 9
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product category
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/U2722A
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight U2722A/U2723A Service Guide 9
Page 10
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
10 Keysight U2722A/U2723A Service Guide
Page 11
Table of Contents
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1 Getting Started
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Self-calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Keysight Technologies calibration services . . . . . . . . . . . . . . . . . . . .22
Calibration interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Types of service available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Extended service contracts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Obtaining repair service (worldwide) . . . . . . . . . . . . . . . . . . . . . . . . .23
Repackaging for shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
2 Performance Verification and Adjustment
Performance Verification and Adjustment . . . . . . . . . . . . . . . . . . . . . . .26
Recommended test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Test consideration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Voltage tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Test Record Forms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Source voltage mode programming and readback accuracy . . . . . .32
Source current mode programming and readback accuracy . . . . . . 33
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Keysight U2722A/U2723A Service Guide 11
Page 12
3Characteristics and Specifications
12 Keysight U2722A/U2723A Service Guide
Page 13
List of Figures
Figure 2-1 Voltage test setup (no load condition) . . . . . . . . . . . .28
Figure 2-2 Current test setup (no load condition) . . . . . . . . . . . . .30
Keysight U2722A/U2723A Service Guide 13
Page 14
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight U2722A/U2723A Service Guide
Page 15
List of Tables
Table 2-1 Recommended test equipment . . . . . . . . . . . . . . . . . . 26
Table 2-2 Test settings for source voltage mode programming and
readback accuracy . . . . . . . . . . . . . . . . . . . . . . . . .27
Table 2-3 Test settings for source current mode programming and
readback accuracy . . . . . . . . . . . . . . . . . . . . . . . . .29
Table 2-4 Source voltage mode programming and readback
accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Table 2-5 Source current mode programming and readback
accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Keysight U2722A/U2723A Service Guide 15
Page 16
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
16 Keysight U2722A/U2723A Service Guide
Page 17
Keysight U2722A/U2723A USB Modular Source Measure Units Service Guide
1 Getting Started
Introduction 18
This chapter discusses the self-test and self-calibration procedure for the U2722A/U2723A USB Modular Source Measure Units. It also provides the information for returning your U2722A/U2723A to Keysight for calibration or servicing.
If you have a defective module, you can return it to Keysight for repair or replacement.
17
Page 18
1 Getting Started
CAUTION
Introduction
Self-test
– A brief power-on self-test occurs automatically whenever you turn on the
– If the self-test fails, the power indicator on the front panel will start blinking.
– Self-test may be initiated remotely by sending the *TST? SCPI command to
Disconnect all connections from the output terminals before performing the self-test procedure. Ensure that the output terminals are not connected to any load before you turn on the U2722A/U2723A. The U2722A/U2723A must be powered-up for at least 45 minutes before performing the self-test.
instrument. This limited test assures that the instrument is capable of operation.
You can use the SYSTem:ERRor? query command from the remote interface to read the error messages. If repair is required, contact a Keysight Service Center. Please refer to the programmer’s reference guide for more information on the error messages.
the instrument.
Keysight Measurement Manager
To start the self-test procedure using the Keysight Measurement Manager, perform the following steps.
1 Click Tools > Sel f-Test.
18 Keysight U2722A/U2723A Service Guide
Page 19
2 Click Start on the Self-Test form.
Getting Started 1
3 Click OK to continue when the message box appears.
4 Wait for a few minutes for the self-test to complete.
5 The result is displayed on the form once the self-test has completed.
Keysight U2722A/U2723A Service Guide 19
Page 20
1 Getting Started
Self-calibration
To perform self-calibration using the U2722A/U2723A Keysight Measurement Manager, perform the following steps.
1 Click Tools > Sel f-Calibration.
2 Click Start on the Self-Calibration form.
20 Keysight U2722A/U2723A Service Guide
Page 21
Getting Started 1
CAUTION
3 Click OK to continue when the message box appears.
4 Wait for a few minutes for the self-calibration to complete.
5 The result is displayed on the form once the self-calibration has completed.
Disconnect all connections from the output terminals before performing the self-calibration procedure. Ensure that the output terminals are not connected to any load before you turn on the U2722A/U2723A. The U2722A/ U2723A must be powered-up for at least 45 minutes before performing the self-calibration.
Keysight U2722A/U2723A Service Guide 21
Page 22
1 Getting Started
Keysight Technologies calibration services
When your instrument is due for calibration, contact your local Keysight Service Center for a low-cost recalibration. The U2722A/U2723A is supported on automated calibration systems, which allows Keysight to provide this service at a competitive price.
Calibration interval
A one-year interval is adequate for most applications. Accuracy specifications are under warranty only if adjustments are made at regular calibration intervals. Accuracy specifications will not be offered warranty beyond the one-year calibration interval. Keysight does not recommend extending calibration intervals beyond two years for any application. Keysight recommends that a complete readjustment should always be performed at the calibration interval. This will ensure that the U2722A/U2723A will remain within specifications for the next calibration interval. The readjustment provides the best long-term stability and accuracy.
Types of service available
If your instrument fails during the warranty period, Keysight will repair or replace it under the terms of your warranty. After your warranty expires, Keysight offers repair services at competitive prices.
Extended service contracts
Most Keysight products are provided with optional service contracts that extend the coverage period after the standard warranty expires. If you have this service contract and your instrument happens to fail during the coverage period, Keysight will repair or replace it according to the contract.
22 Keysight U2722A/U2723A Service Guide
Page 23
Obtaining repair service (worldwide)
To obtain service for your instrument (in-warranty, under service contract, or post-warranty), contact your nearest Keysight Service Center. They will arrange to have your unit repaired or replaced, and are able to provide warranty or repair cost information where applicable.
To obtain warranty, service, or technical support information you can contact Keysight at one of the following telephone numbers.
In the United States: 800 829 4444
In Europe: 31 20 547 2111
In Japan: (81) 426 56 7832
You can also use our Web link for the information on contacting Keysight worldwide:
www.keysight.com/find/assist
Or contact your Keysight representative.
Getting Started 1
Before shipping your instrument, ensure that you acquire shipping instructions, including the components to be shipped, from the Keysight Service Center. Keysight recommends that you retain the original shipping carton for use in such shipments.
Keysight U2722A/U2723A Service Guide 23
Page 24
1 Getting Started
NOTE
Repackaging for shipment
If the unit is to be shipped to Keysight for service or repair, make sure that you do the following.
– Attach a tag to the unit identifying the owner and indicating the required
– Place the unit in its original container with appropriate packaging material for
– Secure the container with strong tape or metal bands.
– If the original shipping container is not available, place your unit in a container
Keysight suggests that you always insure your shipments.
service or repair. Include the model number and full serial number.
shipping.
with at least four inches of compressible packaging material around all sides for the instrument. Use static-free packaging materials to avoid additional damage to your unit.
Cleaning
Clean the outer area of the instrument with a soft, lint-free, and slightly dampened cloth. Do not use detergent. Disassembly is not required for cleaning.
24 Keysight U2722A/U2723A Service Guide
Page 25
Keysight U2722A/U2723A USB Modular Source Measure Units Service Guide
2 Performance Verification
and Adjustment
Performance Verification and Adjustment 26 Performance Verification 27 Test Record Forms 32 Adjustment 34
The performance verification procedures described in this chapter verify that the U2722A/U2723A USB Modular Source Measure Units is operating normally and is within its warranted specifications.
25
Page 26
2 Performance Verification and Adjustment
Performance Verification and Adjustment
This section contains the performance verification and adjustment procedures. If the U2722A/U2723A fails any of the tests or if any abnormal test results are obtained, return the unit to the Keysight Service Center for readjustment. Readjustment can also be carried out using Keysight Measurement Manager revision 1.7 or above.
Recommended test equipment
The recommended test equipment for the performance verification is listed in the table below. If the exact equipment is not available, substitute the performance verification standards of equivalent requirement(s).
Table 2-1 Recommended test equipment
Equipment Requirement(s) Recommended model Purpose Used
Digital multimeter (DMM)
Test consideration
– 8½ digits resolution Keysight 3458A Measures DC voltage
and current
For optimum performance, all procedures should comply with the following recommendations.
– Ensure that the ambient temperature is stable and between 22 °C and 28 °C.
The ideal temperature for performing performance verification should be 25 °C ± 1 °C.
– Make sure that the ambient relative humidity is less than 80%.
– Allow a three-hour warm-up period before performance verification.
– Keep the test connection cables as short as possible.
Performance verification and adjustment
26 Keysight U2722A/U2723A Service Guide
Page 27
Performance Verification and Adjustment 2
NOTE
Performance Verification
Voltage tests
Test each output channel individually. Refer to the appropriate test record form.
Source voltage mode programming and readback accuracy
This test verifies that the voltage programming and readback through the USB are within published specifications.
Test settings
Tab le 2-2 Test settings for source voltage mode programming and readback accuracy
Keysight 3458A settings
Function Range
DC voltage 1 V (for U2722A/U2723A drives 1 V)
10 V (for U2722A/U2723A drives 10 V)
[a] Use the highest resolution available on the DMM.
[a]
Keysight U2722A/U2723A Service Guide 27
Page 28
2 Performance Verification and Adjustment
SMU
+S
+
–S
DMM
+
Test setup
The test setup for no load condition is shown in the following figure.
Figure 2-1 Voltage test setup (no load condition)
Test procedure
1 Turn off the U2722A/U2723A. Connect the DMM to the output to be tested
(refer to Figure 2-1). Ensure that no load is connected. Turn on the U2722A/ U2723A.
2 Configure the selected output as shown in the following.
a Configure the NPLC to 10.
b Configure the voltage range.
c Configure the current range.
d Program the clamp current, I
e Program the drive voltage, V
clamp
drive
.
.
3 Enable the output.
4 Wait for a few seconds for the output to settle.
5 Record the voltage reading on the DMM, V
meas
.
6 Read back the drive voltage over the USB.
7 Record the readback voltage reading over the USB, V
8 Compute (V
meas
– V
drive
) and (V
rdbk
– V
). Compare the values with the test
meas
rdbk
.
limit values.
28 Keysight U2722A/U2723A Service Guide
Page 29
Performance Verification and Adjustment 2
9 Repeat step 2 through step 8 for each iteration to complete the performance
verification for the selected output.
10 Repeat step 1 through step 9 for the rest of the outputs.
Current tests
Source current mode programming and readback accuracy
This test verifies that the current programming and readback through the USB are within published specifications.
Test settings
Tab le 2-3 Test settings for source current mode programming and readback accuracy
Keysight 3458A settings
Function Range
DC current 1 μA (for U2722A/U2723A drives ≤ 1 μA)
10 μA (for U2722A/U2723A drives 10 μA)
100 μA (for U2722A/U2723A drives 100 μA)
1 mA (for U2722A/U2723A drives ≤ 1 mA)
10 mA (for U2722A/U2723A drives 10 mA)
100 mA (for U2722A/U2723A drives 100 mA)
1 A (for U2722A/U2723A drives >100 mA)
[a] Use the highest resolution available on the DMM
[a]
Keysight U2722A/U2723A Service Guide 29
Page 30
2 Performance Verification and Adjustment
SMU
+S
+
–S
DMM
+
Test setup
The test setup for no load condition is shown in the following figure.
Figure 2-2 Current test setup (no load condition)
Test procedure
1 Turn off the U2722A/U2723A. Connect the DMM to the output to be tested
(refer to Figure 2-2). Ensure that there is no load connected. Turn on the U2722A/U2723A.
2 Configure the selected output as shown in the following.
a Configure the NPLC to 10.
b Configure the voltage range.
c Configure the current range.
d Program the clamp voltage, V
e Program the drive current, I
drive
clamp
.
.
3 Enable the output.
4 Wait for a few seconds for the output to settle.
5 Record the current reading on the DMM, I
meas
.
6 Read back the drive current over the USB.
7 Record the readback current reading over the USB, I
8 Compute (I
meas
– I
drive
) and (I
rdbk
– I
). Compare the values with the test
meas
rdbk
.
limit values.
30 Keysight U2722A/U2723A Service Guide
Page 31
Performance Verification and Adjustment 2
9 Repeat step 2 through step 8 for each iteration to complete the performance
verification for the selected output.
10 Repeat step 1 through step 9 for the rest of the outputs.
Keysight U2722A/U2723A Service Guide 31
Page 32
2 Performance Verification and Adjustment
Test Record Forms
Source voltage mode programming and readback accuracy
Table 2-4 Source voltage mode programming and readback accuracy
V
Iteration Test point Vol tage range Clamp current Current range
meas
– V
driveVrdbk
– V
meas
Limit
1
2 10 mA 10 mA
3 1 mA 1 mA
1 V 2 V
4100 μA 100 μA
510 μA 10 μA
61 μA1 μA
7
8 10 mA 10 mA
9 1 mA 1 mA
10 V 20 V
10 100 μA 100 μA
11 10 μA 10 μA
12 1 μA1 μA
120 mA 120 mA
120 mA 120 mA
±2.25 mV
±15 mV
32 Keysight U2722A/U2723A Service Guide
Page 33
Performance Verification and Adjustment 2
Source current mode programming and readback accuracy
Tab le 2-5 Source current mode programming and readback accuracy
Iteration Test point Current range
1
Clamp
voltage
Voltage range
20 V 20 V ±1.275 nA
I
meas
– I
driveIrdbk
– I
meas
0.5 μA1 μA
22 V2 V±1.275 nA
3
20 V 20 V ±12.75 nA
5 μA 10 μA
42 V2 V±12.75 nA
5
20 V 20 V ±112.5 nA
50 μA 100 μA
62 V2 V±112.5 nA
7
20 V 20 V ±1.125 μA
0.5 mA 1 mA
82 V2 V±1.125 μA
9
20 V 20 V ±11.25 μA
5 mA 10 mA
10 2 V 2 V ±11.25 μA
11
20 V 20 V ±0.16 mA
60 mA 120 mA
12 2 V 2 V ±0.16 mA
Limit
Keysight U2722A/U2723A Service Guide 33
Page 34
2 Performance Verification and Adjustment
Adjustment
The following information describes the procedure for voltage and current adjustments for the U2722A/U2723A using Keysight Measurement Manager revision 1.7 or above.
1 Establish the connection between the U2722A/U2723A and the PC. Launch
the Keysight Measurement Manager as shown below.
34 Keysight U2722A/U2723A Service Guide
Page 35
Performance Verification and Adjustment 2
2 Select Tools > Source Measure > Ad justment to start the adjustment.
3 A message box will appear to inform that the adjustment operation will take
approximately two hours. During this period, you will not be able to operate the U2722A/U2723A. Click Yes to continue with the adjustment operation.
Keysight U2722A/U2723A Service Guide 35
Page 36
2 Performance Verification and Adjustment
4 Establish the connection between the 3458A and the PC to proceed with the
adjustment operation. Click OK to continue.
If the 3458A is not detected, an error message will appear as shown in the following figure. Reconnect the 3458A to the PC and click Yes to proceed. Select No to exit the adjustment operation.
5 The 3458A and U2722A/U2723A will be triggered to perform self-test and
self-calibration. During this process, make sure that there is no connection on the instruments terminals. If self-test or self-calibration fails, cycle the U2722A/U2723A power and repeat step 1 through step 5. In the unlikely event that self-test or or self-calibration repeatedly fails, contact your nearest Keysight Sales Office.
6 After the self-test and self-calibration have completed, a message box shown
in the following will appear to indicate you to connect the respective U2722A/ U2723A output to the voltage terminal of the 3458A. After the connection is established, click OK to continue. The Keysight Measurement Manager will now perform voltage adjustment on channel 1.
36 Keysight U2722A/U2723A Service Guide
Page 37
Performance Verification and Adjustment 2
7 When the voltage adjustment has completed, connect the respective U2722A/
U2723A output to the current terminal of the 3458A. After the connection is established, click OK to continue with current adjustment for channel 1.
8 After the voltage and current adjustments for channel 1 have completed, the
3458A will proceed with voltage and current adjustments for channel 2 followed by voltage and current adjustments for channel 3. Follow the instructions on the message box to complete the adjustment operation.
9 When the adjustment operation has completed, a message box as shown in
the following will appear. Click OK to end the adjustment operation.
10 A message box will appear as shown below prompting you to power cycle the
U2722A/U2723A for the adjustment to take effect. To power cycle the U2722A/U2723A, power off the U2722A/U2723A and restart it. Ensure that the U2722A/U2723A is powered-off properly before restarting.
Keysight U2722A/U2723A Service Guide 37
Page 38
2 Performance Verification and Adjustment
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
38 Keysight U2722A/U2723A Service Guide
Page 39
Keysight U2722A/U2723A USB Modular Source Measure Units Service Guide
3 Characteristics and
Specifications
For the characteristics and specifications of the U2722A/U2723A USB Modular Source Measure Units, refer to the datasheet at
http://literature.cdn.keysight.com/litweb/pdf/5990-7416EN.pdf.
39
Page 40
3 Characteristics and Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
40 Keysight U2722A/U2723A Service Guide
Page 41
Index
A
accuracy
Source Current Mode
Programming and Measurement
Source Voltage Mode
Programming and Measurement
specifications
adjustment procedure
procedure
ambient
ideal temperature relative humidity temperature, 26 warm-up period
, 29, 33
, 27, 32
, 22
. See
, 26
, 26
, 26
C
Calibration
interval performance. See performance procedure self. See Keysight
services. See Keysight standards
channel Characteristics, 39 cleaning, 24
compliance connection cable connector, output. See output consideration, test. See test
, 22,
. See procedure
Measurement Manager
, 26
, 27, 28, 30
. See test
contact contract, service, 22, 23
current
, 18, 22, 23
clamp, 28 DC, 29
, 30, 30
drive programmable programming, 29 read back
, 28, 30
range source. See source
, 29, 30
test
, 26
, 29, 30
D
DC
current voltage, 27
degree, pollution
Digital Multimeter
disassembly, 24 drive
current voltage
. See current
. See pollution
degree
, 26, 27, 28,
29, 30, 30
. See current
. See voltage
E
ELV power source, isolated. See
power
equipment
, test. See test
H
humidity, relative
ambient. See ambient
I
ideal temperature. See ambient interface, remote. See remote interval, calibration. See
Calibration
isolated
iteration
, ELV power source. See
power
, 29, 31, 32, 32
K
Keysight
, 26, 26
3458A calibration services service
Keysight Measurement Manager
self-calibration, 20, 21 self-test
, 22, 23
, 18, 19
, 22
L
limit, test. See test load, 18, 21, 28, 30
M
material, packaging, 24 MEAS:CURR?, 30
Keysight U2722A/U2723A Service Guide 41
Page 42
measurement
accuracy manager. See Keysight
Measurement Manager
source current mode, 29, 33 source voltage mode, 27, 32
message box multimeter, digital. See Digital
Multimeter
. See accuracy
, 19, 21
N
no load condition, 28, 30
O
operating
environment
operating OUTP ON output
channel, 27 guard resistance. See guard terminals
, 28, 29, 30
test
. See environment,
, 28, 28
, 18, 21
P
packaging materials. See material,
packaging
performance
calibration characteristics
Characteristics
test
, 26
power procedure
adjustment, 34 calibration test
, current, 30 , voltage, 28
test
, 25, 26
. See
, 27, 28, 30
product
Keysight characteristics. See
Characteristics
programming
source current mode source voltage mode voltage/current
voltage
, 22
, 29, 33
, 27, 32
. See current or
R
range, current/voltage. See
current or voltage
ratings read back. See current or voltage recommended, test equipment, 26 record form, test. See test Regulatory markings, 8 relative humidity. See humidity,
remote
repackaging for shipment, 24 repair service resolution, 26, 27, 29 result, test, 26
, output. See output
relative
interface sense operating range. See
operating range sense
, 18
, remote
, 22, 23
S
safety. See compliance SCPI commands
self-test
Self-calibration
Measurement Manager
, 18
. See Keysight
Self-test
sense service
settings, test. See test setup, test. See test shipment, 24
source
specifications
standard
storage support, technical, 23 SYSTEM:ERROR?, 18
, 18
Keysight Measurement
Manager Measurement Manager
. See Keysight
, remote. See remote
calibration, 22 contract, 22, 23
, 23
repair
, 22
types
current mode, 29, 33
, isolated ELV. See power
power voltage mode
accuracy. See accuracy performance product. See product
calibration commands, programming
language commands
warranty, 22
, 27, 32
. See performance
, 26
. See SCPI
. See compliance
T
technical support. See support,
technical
temperature
ambient/ideal operating
terminals
output
. See output
. See ambient
. See operating
42 Keysight U2722A/U2723A Service Guide
Page 43
test
connection cable consideration, 26 current equipment limit, 28, 30 procedure
procedure
procedure, voltage. See
record form result self settings, current, 29 settings setup setup, voltage, 28 voltage
tools
self-calibration, 20 self-test
*TST? types, service. See service
. See current
, calibration. See
procedure
, current. See
procedure
procedure
. See result, test
. See Self-test
, voltage, 27
, current, 30
. See voltage
, 18
, 18
, 26
, 26
, 27, 32, 33
voltage
clamp, 30 DC, 27
, 28, 28
drive
, overshoot. See output
output programmable, 26 programming
, 28, 30
range read back, 27, 28
. See source
source
, 27, 28, 28
test
W
warm-up, 26 warranty, 22
in, 23 post, 23
worldwide
, repair service, 23
, 27, 29
U
USB, 27, 28, 30
Hi-Speed 2.0. See Hi-Speed
USB 2
.0
V
vibration. See shock and vibration
Keysight U2722A/U2723A Service Guide 43
Page 44
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
44 Keysight U2722A/U2723A Service Guide
Page 45
This information is subject to change without notice. Always refer to the English version at the Keysight website for the latest revision.
© Keysight Technologies 2009-2017 Edition 8, December 1, 2017
Printed in Malaysia
*U2722-90021*
U2722-90021
www.keysight.com