Keysight P9370B, P9371B, P9372B, P9373B, P9374B User Manual

...
Streamline Series Vector Network Analyzer (B-models)
P937xB 2-port Up to 44 GHz P938xB 4-port Up to 20 GHz P500xB 2-port Up to 53 GHz P502xB 4-port Up to 53 GHz / 6-port Up to 20 GHz
Compact form. Zero compromise.
Find us at www.keysight.com Page 1
Table of Contents
Keysight Streamline Series VNA .......................................................................................................................... 4
P93xxB and P50xxB Configurations..................................................................................................................... 5
Test set options ................................................................................................................................................ 6
Application software ......................................................................................................................................... 6
P93xxB option .................................................................................................................................................. 8
P50xxB software .............................................................................................................................................. 8
Performance comparison between P50xxB and P93xxB ................................................................................... 8
Calibration option ............................................................................................................................................. 9
Training options ............................................................................................................................................... 9
Additional hardware ......................................................................................................................................... 9
Accessories ....................................................................................................................................................10
Test cables .....................................................................................................................................................12
Measurement Application Software .....................................................................................................................13
Automatic fixture removal (S97007B) ..............................................................................................................13
Time domain analysis (S97010B) ....................................................................................................................13
Enhanced time domain analysis with TDR (S97011B) .....................................................................................13
Real time S-parameter and power measurement uncertainty (S97015B) .........................................................14
Basic Pulsed-RF measurements lite (S970240B/S970241B) ...........................................................................14
Basic Pulsed-RF measurements (S970250B/S970251B) .................................................................................14
Noise figure measurements with vector correction (S97029B) .........................................................................15
Scalar mixer/converter measurements (S97082B) ...........................................................................................16
Vector and scalar mixer/converter measurements (S97083B) ..........................................................................16
Embedded LO capability (S97084B) ................................................................................................................16
Gain compression measurements (S97086B) ..................................................................................................17
Intermodulation distortion measurements (S97087B) .......................................................................................17
Source phase control (S97088B) .....................................................................................................................17
Differential and I/Q device measurements (S97089B) ......................................................................................18
Spectrum analysis (S97090xB) .......................................................................................................................18
True-mode stimulus (S97460B) .......................................................................................................................18
Multiple instrument measurements (S97551B).................................................................................................19
Multiport calibration assistant (S97552B) .........................................................................................................20
Multiport calibrated measurements with switch instruments (S97553B) ............................................................20
Banded millimeter-wave network analysis (S97560B) ......................................................................................20
Material measurements ...................................................................................................................................21
N1930xB Physical Layer Test System (PLTS) software ...................................................................................21
Device Measurement eXpert (DMX) (S94601B)...............................................................................................21
Limit assistant (S94602B)................................................................................................................................21
Automated measurement expert (AMX) ...........................................................................................................21
Find us at www.keysight.com Page 2
PathWave Vector Signal Analysis (89600 VSA) software .................................................................................22
Vector Network Analyzer (VNA) simulator (S9405xB) ......................................................................................22
USB Type-C interconnects compliance test software (S94USBCB) .................................................................22
Direct Receiver Access (DRA) Configurations .....................................................................................................23
DRA configuration with high-power ..................................................................................................................23
Keysight Software Licensing Options Provide Flexibility and Support ..................................................................24
Choose your term. Choose your type. Keep control of your budget. .................................................................24
KeysightCare Software Support Subscription provides peace of mind amid evolving technologies. ..................24
Ordering Information ...........................................................................................................................................25
Measurement Accessories ..................................................................................................................................26
Cable and adapter sets ...................................................................................................................................26
Calibration kits ................................................................................................................................................27
For devices with 1.85 mm connectors ..............................................................................................................28
For devices with 2.4 mm connectors................................................................................................................29
For devices with K connectors (2.92 mm) ........................................................................................................30
For devices with 3.5 mm or SMA connectors ...................................................................................................31
For devices with Type-N connectors ................................................................................................................35
For devices with 7 mm connectors ..................................................................................................................39
For devices with waveguide ............................................................................................................................40
Verification kits ................................................................................................................................................42
Accessories ....................................................................................................................................................43
PC Requirements for VNA Control ......................................................................................................................44
Literature Information..........................................................................................................................................44
Web Resources ..................................................................................................................................................44
Find us at www.keysight.com Page 3

Keysight Streamline Series VNA

The freedom of portable network analysis doesn’t have to mean a compromise in performance. The B-models of the Keysight Streamline Series VNAs unlock the full performance of the network analyzer hardware with fast measurements over Thunderbolt 3 connections to your host PC. Gain confidence in your measurements with best-in-class performance offering fast, reliable, and repeatable results. Explore the complete characterization of your devices with a rich portfolio of software applications that transform the compact network analyzer into a complete RF measurement solution.
The P50xxB Series offers the performance required for testing passive components, amplifiers, mixers or frequency converters. The vector network analyzer (VNA) provides best-in-class key specifications such as dynamic range, measurement speed, trace noise and temperature stability. Choose from 2- or 4-port models up to 53 GHz, or 6-port models up to 20 GHz.
The P93xxB Series provides excellent performance in general-purpose network analysis for passive components. With software applications like enhanced time domain analysis with TDR and automatic fixture removal, you can easily characterize passive components with the same performance of a benchtop setup. Choose from 2-port models up to 44 GHz or 4-port models up to 20 GHz.
The Streamline Series utilizes the same measurement science as other Keysight VNAs such as the PNA, ENA and PXI VNA. A common software platform makes it easy to choose the right level of performance to match budget and measurement needs. This commonality guarantees measurement consistency, repeatability, and a common remote-programming interface across multiple instruments in R&D and manufacturing.
This configuration guide describes standard configurations, options, accessories, and compatible peripherals for the P93xxB and P50xxB Keysight Streamline Series vector network analyzer (VNA).
Find us at www.keysight.com Page 4
P93xxB and P50xxB Configurations
Model 1
Description
Test port connectors
P937xB Series
P9370B
9 kHz to 4.5 GHz, 2-port
3.5 mm (f)
P9371B
9 kHz to 6.5 GHz, 2-port
3.5 mm (f)
P9372B
9 kHz to 9 GHz, 2-port
3.5 mm (f)
P9373B
9 kHz to 14 GHz, 2-port
3.5 mm (f)
P9374B
9 kHz to 20 GHz, 2-port
3.5 mm (f)
P9375B
100 kHz to 26.5 GHz, 2-port
3.5 mm (f)
P9377B
100 kHz to 44 GHz, 2-port
2.4 mm (f)
P938xB Series
P9382B
9 kHz to 9 GHz, 4-port
3.5 mm (f)
P9384B
9 kHz to 20 GHz, 4-port
3.5 mm (f)
P500xB Series
P5000B
9 kHz to 4.5 GHz, 2-port
3.5 mm (f)
P5001B
9 kHz to 6.5 GHz, 2-port
3.5 mm (f)
P5002B
9 kHz to 9 GHz, 2-port
3.5 mm (f)
P5003B
9 kHz to 14 GHz, 2-port
3.5 mm (f)
P5004B
9 kHz to 20 GHz, 2-port
3.5 mm (f)
P5005B
100 kHz to 26.5 GHz, 2-port
3.5 mm (f)
P5006B
100 kHz to 32 GHz, 2-port
2.4 mm (f)
P5007B
100 kHz to 44 GHz, 2-port
2.4 mm (f)
P5008B
100 kHz to 53 GHz, 2-port
1.85 mm (f)
P502xB Series 2
P5020B
9 kHz to 4.5 GHz, 4 or 6-port
3.5 mm (f)
P5021B
9 kHz to 6.5 GHz, 4 or 6-port
3.5 mm (f)
P5022B
9 kHz to 9 GHz, 4 or 6-port
3.5 mm (f)
P5023B
9 kHz to 14 GHz, 4 or 6-port
3.5 mm (f)
P5024B
9 kHz to 20 GHz, 4 or 6-port
3.5 mm (f)
P5025B
100 kHz to 26.5 GHz, 4-port
3.5 mm (f)
P5026B
100 kHz to 32 GHz, 4-port
2.4 mm (f)
P5027B
100 kHz to 44 GHz, 4-port
2.4 mm (f)
P5028B
100 kHz to 53 GHz, 4-port
1.85 mm (f)
1. All B -models of Keysight Stream line Series VNAs include: Thunderbolt 3 cable assembly for connection with a host PC (800- mm, P/N: 8121-3588), power supply (external AC- DC adapter. P/N: 0950- 6128 for P937xB/P500xB , or P/ N: 0950-6166 for P938xB/P502xB ), and power cord.
2. P502xB Option 400 (>20 GHz mo dels) or Option 402 (all models) include one flexible control cable (P/N: M9800-61601) and one semi­rigid cable (P/N: M9800- 61602) for the connect ion on the fro nt panel. Both cable assemblies are i ncluded in Y1730A Option 001 (Interconnect cables for multiport configuration of 1-slot M980xA).
Find us at www.keysight.com Page 5

Test set options

Option number
2-port
4-port
6-port
Additional information
Option 200
X
Not available with P937xB, P938xB or P502xB Series.
Option 400
X
Not available with P937xB, P938xB or P500xB Series.
Option 402
X
Adds internal second source. Not available with P937xB, P938xB or P500xB Series.
Option 600
X
Not available with P937xB, P938xB or P500xB Series.
For P937xB,
For P500xB,
Additional information
Automatic fixture removal
S97007B
S97007B
Time domain analysis
S97010B
S97010B
TDR
on the measurement trace.
S970250B.
Basic pulsed-RF measurements lite,
N/A
S970241B
Available with P5005B to P5008B, or
S970251B.
20 GHz 2
P5020B to P5024B.
53 GHz 2
P5025B to P5028B.
correction 3
Scalar mixer/converter measurements3
S97082B
S97082B
Provides SMC measurement class.
Vector and scalar mixer/ converter
N/A
S97083B
Provides SMC + phase
mixer characterization.
Embedded-LO capability 3
N/A
S97084B
Gain-compression measurements 3
N/A
S97086B
Intermodulation distortion
N/A
S97087B
Requires option 402 (internal second
measurements with S97551B.
measurements with S97551B.
Choose one of the frequency models, and one test set option for the P500xB or P502xB Series. Option 200 indicates two test ports, option 40x indicates four test ports and option 600 indicates six test ports. To add options to a product, order the corresponding item number (ex. P5000B-200).

Application software

Description 1
Enhanced time domain analysis with
Real-time S-parameter and power measurement uncertainty
Basic pulsed-RF measurements lite, up to 20 GHz
2
up to 53 GHz 2
Basic pulsed-RF measurements, up to
Basic pulsed-RF measurements, up to
Noise figure measurements with vector
P938xB Series
P502xB Series
S97011B S97011B Includes all capabilities of S97010B.
N/A S97015B Displays the measurement
uncertainty dynamically (real-time)
N/A S970240B
Available with P5000B to P5004B, or P5020B to P5024B. Subset of
P5025B to P5028B. Subset of
N/A S970250B Available with P5000B to P5004B, or
N/A S970251B Available with P5005B to P5008B, or
N/A S97029B Standard receivers are used.
measurements 3
measurements
3,4
Source phase control N/A S97088B Requires option 402 (internal second
Find us at www.keysight.com Page 6
measurement class. Provides vector
source option) or multiple instrument
source option) or multiple instrument
Differential and I/Q device
measurements 3
N/A
S97089B
Requires option 402 (internal second
measurements with S97551B.
Spectrum analysis, up to 4.5 GHz
N/A
S970900B
Available with P5000B or P5020B.
Spectrum analysis, up to 6.5 GHz
N/A
S970901B
Available with P5001B or P5021B.
Spectrum analysis, up to 9 GHz
3,5
N/A
S970902B
Available with P5002B or P5022B.
Spectrum analysis, up to 14 GHz
N/A
S970903B
Available with P5003B or P5023B.
Spectrum analysis, up to 20 GHz
N/A
S970904B
Available with P5004B or P5024B.
Spectrum analysis, up to 26.5 GHz
N/A
S970905B
Available with P5005B or P5025B.
Spectrum analysis, up to 32 GHz
N/A
S970906B
Available with P5006B or P5026B.
Spectrum analysis, up to 44 GHz
N/A
S970907B
Available with P5007B or P5027B.
Spectrum analysis, up to 44 GHz
N/A
S970908B
Available with P5008B or P5028B.
S97551B.
Series VNAs
Requires one ECal module.
matrix 9
analysis
extenders
class of Keysight’s VNAs.
Vector Network Analyzer (VNA)
S94051B
S94051B
Runs on your PC and simulates VNA
features of S97xxxB software 10
customizable templates.
masks for production tests.
Automated measurement expert
S94701A
S94701A
A smart software solution for
model and install in a host PC.
or raw cables) using 4-port VNA.
3,5
3,5
3,5
3,5
3,5
3,5
3,5
3,5
source option) or multiple instrument
True-mode stimulus N/A S97460B Requires option 402 or multiple
instruments measurements with
Multiple instruments measurements S97551B S97551B Provides multiport calibrated
measurements using two Streamline
6,7,8
Multiport calibration assistant S97552B S97552B Provides a tool to flexibly define cal
sets of multiport measurements.
Multiport calibrated measurements with switch instruments
S97553B S97553B Provides multiport calibrated
measurements using P9164A/B or P9165A/B USB Solid-State Switch
Banded millimeter-wave network
Vector Network Analyzer (VNA) simulator - Standard
simulator - Advanced
S97560B S97560B Enables the operation with frequency
S94050B S94050B Runs on your PC and simulates
standard S-parameter measurement
functions including advanced
Device measurement expert (DMX) S94601B S94601B Assists the users in setting up
measurements automatically with
Limit assistant S94602B S94602B Allows to acquire the data from the
VNA, and easily generate limit
(AMX)
S94702A KS8400A
S94702A KS8400A
automated S-parameter measurements with the Streamline Series VNA. Order each software
USB Type-C interconnects compliance test software
1. Refer to “Keysight Software Licensing Options Provide Flexibility and Support” for more detail about S97xxxB licenses.
Find us at www.keysight.com Page 7
11,12
N/A S94USBCB
Performs automated compliance tests of USB Type-C interconnects (ex. cable assemblies, connectors,
2. Enables internal pulse generators and pulse modulators on the analyzer’s source(s). Hardware options are not needed for pulsed-RF
Description
P500xB
P502xB
High-performance enabling software 1
S9720xB
S9721xB
VNA activation software 1
S9722xB
S9723xB
Performance
P50xxB
P93xxB
Dynamic range (10 Hz IFBW) 4 GHz
140 dB
115 dB
20 GHz
126 dB
105 dB
4 GHz
-130 dBm
-107 dBm
20 GHz
-122 dBm
-101 dBm
Max output power 4 GHz
+10 dBm
+8 dBm
20 GHz
+4 dBm
+4 dBm
IF bandwidth
1 Hz to 15 MHz
1 Hz to 700 kHz
(2-port cal, 0.8 to 1 GHz, typical)
201 points
1.8 ms (1 MHz IFBW)
15 ms (700 kHz IFBW)
401 points
2.5 ms (1 MHz IFBW)
28 ms (700 kHz IFBW)
measurements.
3. Provides frequency offset mode (FOM) to independe ntly set the frequency of internal source(s) and receivers, and to configure external
sources using External Device Co nfiguration.
4. The most commo n method to use S97087B is to use a 4- port P502xB with t he internal second source option 402, with each built-in
source generating one of th e two tones. Any of the re maining ports can be used as the receive port..
5. Enables spectrum analysis a nd vector signal analysis using the VNA’s receivers. PathWave Vector Signal Analysis (8960 0 VSA)
software is required for vector signal analysis. H ardware options are not n eeded for spectrum analysis or vector signal analysis.
6. When configur ing a multiport VNA using multiple Streamline Series VNAs, Y1701A is recommended for each additional multiport
interconnection.
7. P937xB and P938xB, or P500xB and P502xB can be combined in multiport configurations. P93xxB and P50xxB Series cannot be
combined.
8. A 4- port EC al module (ex. N4431/32/33D) is recommended for multiport measurements (n > 4).
9. Only the 4- port ECa l modules ( ex. N4431/32/33D ) are supported for calibration with S97553B.
10 . Features of S97007B, S97552B and S97553B are not supported w ith the S94051B. 11 . Requires S9701 1B, KS8400B PathWave Test Automation, and KS8104B HTML5 Result Listener Plugin. 12 . Supports L8990M-0LZ 20-port switch matrix to configure a multiport test system which eliminates the need for reconnections of test
cables with DUT.

P93xxB option

VNA activation software option (option 001) enables basic VNA functionalities of P937xB and P938xB
1. Option 001 is mandatory at the time of purchase of P937xB and P938xB Streamline Series VNA.

P50xxB software

S9720xB/S9721xB software improves measurement performance of P500xB and P502xB (ex. dynamic range, noise floor, max output power or cycle time) compared to the performance of P937xB and P938xB.
S9722xB/S9723xB software enables basic VNA functionalities of P500xB and P502xB. The software ensures compatibility with newly added features and keeps your instrument up to date.
1. S972xxB software includes node-locked, perpetual license and one-year support subscription. It is mandatory a t the time of purchase of P5 00xB or P502xB Streamline Series VNA (x = 0 to 8).
Performance comparison between P50xxB and P93xxB
Noise floor (10 Hz IFBW)
Cycle time
1. Refer to data sheet (3121-1235.EN) for m ore details about the difference in performance/specifications between P50xxB and P93xxB.
2. Refer to VNA selection guide (5989- 7603EN) for the comparison of P50xxB and P50xxA, or P93xxB and P937xA.
1,2
Find us at www.keysight.com Page 8

Calibration option

Option number
Description
Additional information
specifications. Conforms to ISO 17025 and ISO 9001.
For P937xB,
P938xB Series
For P500xB,
P502xB Series
Additional information
remotely or on-site.
office.
training needs with the Keysight sales representative.
Model number
Description
Additional information
P9164A
Solid-state switch matrix module, 2x16 full crossbar, 300 kHz to 6.5 GHz
S97553B application software
P9164B
Solid-state switch matrix module, 2x16 full crossbar, 300 kHz to 9 GHz
P9165A
Solid-state switch matrix module, 2x8 full crossbar, 300 kHz to 6.5 GHz
300 kHz to 9 GHz
Option 1A7 ISO 17025 compliant
calibration

Training options

Description
Included - instrument care and operations starter training
Optional - technology and measurement science standard training
N/A PS-S40-01 Provides startup training service that reduces
N/A PS-S40-02 Provides customized technology and measurement
Provides a complete set of measurements which test the unit to manufacturer’s published specifications. Includes calibration label, ISO 17025 calibration certificate and data report, and measurement uncertainties and guardbands on all customer
complexity associated with initial setup and enables users to get to measurements fast.
Recommended topics are product overview, system setup and initial measurements. Can be delivered
science training. Enable users to get deeper insights into the VNA measurement theory and applications.
Recommended topics are VNA theory and applications. Can be delivered on-site or in Keysight
Optional - premium functional and customer specific application training

Additional hardware

P9165B Solid-state switch matrix module, 2x8 full crossbar,
N/A PS-S40-03 Provides an integrated training experience that fits to
the users’ needs on applications or workflow. Custom topics based on the user needs. Discuss the
is required for multiport calibrated measurements.
Find us at www.keysight.com Page 9

Accessories

For P937xB or
For P938xB or
VNA (P938xB or P502xB Series).
VNA.
Y1701A-101
Y1701A-202
Includes latch kit for connecting two instruments 2.
additional VNA.
Hard transit case
Y1710A
Y1710A
of the VNA.
mm cable.
Streamline Series VNA 7.
Description
P500xB Series
P502xB Series
Additional information
Rack mount kit Y1700A-001 Y1700A-002 Order Option 001 for 2-port VNA (P937xB or
P500xB Series). Order Option 002 for 4 or 6-port
Multiple USB instruments configuration kit
PXI adapter module
5,6
Y1701A-002 or 003
Y1701A-302 Y1701A-303
Y1701A-400 or 401
Y1701A-002 or 003
or 304
Y1701A-400 or 401
Includes interconnect cables for configurations using two VNAs
1
. Add one kit for each additional
Includes interconnect cables for multiport configurations in a rack
3
. Add one kit for each
Includes Thunderbolt 3 cable assembly with USB Type-C connectors for connection with a host PC.
Y1731A-001 Y1731A-001 Includes one PXI module to enable coaxial
connections with control signals of Streamline Series VNA. Equipped with twelve SMB connectors. Includes 2 flexible control cables (500 mm and 175 mm) for connection with a control port
4
SMB (f) to MCX (m) cable assembly
BNC (f) to MCX (m) cable assembly
1. Order Y170 1A O ption 00 2 for P93xxB or P50xxB Series up to 20 GHz. Order Y1701A Option 003 f or greater th an 20 GHz models.
2. If 2-port VNA (ex. P500xB) and 4- or 6-port VNA (ex. P502xB) are connected for a multiport VNA, order Y1701A Option 102 latch kit for connecting 1- slot to 2- slot instruments.
3. Order Y170 1A Op tion 302 for 2-port VN A (P937xB or P500xB Ser ies). Order Option 303 for P938xB or P502xB- 400/600 up to 20 GHz. Order option 304 for P502xB- 400 (above 20 GHz) or P502xB- 402 (all models).
4. Y1701 A Option 400 includes on e 500-mm cable (P/N: 8121-3587) and one re tainer (P/N: P5000-29001). Option 401 includes one 800-mm cab le (P/N: 8121- 3588) and one reta iner (P/N: P50 00-29001).
5. Y1731 A does not require power supply. PXI chassis is not needed for the operation.
6. Order Y1731A Option 800 (BN C (m) to SMB (f) cable assembly, 1000 m m) for connection with BNC connectors on external peripherals (ex. pulse generators)
7. Usin g 10 MHz reference ports (10 MHz In/Out) on the rear panel of Streamline Series VNA is not supported.
Y1730A-800 or 801
Y1730A-800 or 801
For connection with 10 MHz external reference ports (Ref In/Ref Out) on the front panel of Streamline Series VNA
7
. Option 800 includes one
300-mm cable and option 801 includes one 1000-
Y1730A-810 Y1730A-810 For connection with 10 MHz external reference
ports (Ref In/Ref Out) on the front panel of
Find us at www.keysight.com Page 10
Figure 1. Rack mount kit for Keysight Streamline Series products (Y1700A option 001)
Figure 2. Transit case for Keysight Streamline Series products (Y1710A)
Figure 3. Multiple instruments configuration kit (Y1701A)
Find us at www.keysight.com Page 11

Test cables

Option number
Description
Additional information
Y1740A-100
Cable Assembly, 3.5 mm (m) to 3.5 mm (m), 26.5 GHz, 36 inches
Y1740A-200
Cable Assembly, 2.92 mm (m) to 2.92 mm (m), 40 GHz, 36 inches
Y1740A-300
Cable Assembly, 2.4 mm (m) to 2.4 mm (m), 50 GHz, 36 inches
Y1740A-310
Cable Assembly, 2.4 mm (m) to 2.92 mm (m), 40 GHz, 36 inches
Y1740A-400
Cable Assembly, 1.85 mm (m) to 1.85 mm (m), 67 GHz, 1 m
Figure 4. Cable assembly with 2.4-mm male connectors (Y1740A-300)
Find us at www.keysight.com Page 12

Measurement Application Software

Solid black series name indicates the feature is available on that series, while gray series name with strikethrough indicates the feature is not available with that series. For example:
P50xxB Series: Available with P500xB or P502xB Series
P93xxB Series: Not available with P937xB or P938xB Series

Automatic fixture removal (S97007B)

P93xxB Series P50xxB Series
Many devices do not have coaxial connectors and are put in fixtures in order to measure them in a coaxial environment. Accurately removing the effects of the fixture is required to get a good measurement of the device under test (DUT). This application adds a powerful application wizard to guide you through characterizing a fixture and removing it from the measurement. Devices can be single-ended or differential. Files can be saved in a variety of formats for later use in ENA, PNA, PXI VNA, Streamline Series VNA, ADS, and PLTS.

Time domain analysis (S97010B)

P93xxB Series P50xxB Series
This application enables the analyzer to view reflection and transmission responses in time or distance. Use time domain to tune filters, gate out the response of fixtures and cables, characterize the impedance of transmission lines and more.

Enhanced time domain analysis with TDR (S97011B)

P93xxB Series P50xxB Series
This application enables the analyzer to perform enhanced time domain analysis for high-speed data applications. All functionalities of the S97010B are included (TDR/TDT mode). In addition, the S97011B enables more detailed measurements and evaluations, such as eye-diagram/mask modes, without adding PLTS software. Jitters and/or emphasis/equalization capabilities enables simulation of real-world signals and environment. S97011B covers up to 53 GHz bandwidth with 8.42 psec rise time. Full calibration is available and automatic deskew ensures easy removal of fixture and probe effects. To get the best accuracy, mechanical calibration kits or ECal with DC option (i.e. N443xD or N469xD with Option 0DC) are recommended.
Find us at www.keysight.com Page 13

Real time S-parameter and power measurement uncertainty (S97015B)

P93xxB Series P50xxB Series
This application provides uncertainties for both S-parameter and power measurements on the P50xxB. The real­time display of the uncertainty associated with power and S-parameter traces increases the confidence in the reproducibility of measurements. This allows users to implement pass/fail tests easily because now the instrument quantifies the gray region that is in between a full pass or a full fail, apply more realistic limit lines which can increase the production yield and reduce the defect percentage on the finished products. This application easily establishes a metric to quantify the quality of the measurement process, so your company’s quality control procedures are simplified. It includes the uncertainty information for most Keysight calibration kits and provides national metrology institute traceability through Keysight’s calibration kits. This application also helps you to include uncertainty information for your product’s specifications and data sheets.

Basic Pulsed-RF measurements lite (S970240B/S970241B)

P93xxB Series P50xxB Series
This application enables internal pulse generators of P500xB or P502xB Series that can be used to control the internal pulse modulators, and it provides an integrated pulse application that uses the wideband-detection method.
The pulse application provides an easy way to set up point-in-pulse measurements with pulse width as narrow as 50 us, and pulse-profile measurements with 1 us minimum timing resolution. Using the built-in pulse modulators, the P50xxB Series VNAs are complete pulsed-RF measurement solutions, eliminating the need for external test sets and pulse generators. S97024xB also controls external pulse generators and modulators and can synchronize to external master pulses. Y1731A I/O Adapter Module is recommended to access pulse signals with SMB connectors if using external master pulses or external pulse modulators.
S970240B is available with P50xxB up to 20 GHz (20 GHz models or less), and S970241B is available for P50xxB greater than 20 GHz. Hardware options are not needed for pulsed-RF measurements with P50xxB.
S97024xB is a subset of S97025xB, so they should not be installed together.

Basic Pulsed-RF measurements (S970250B/S970251B)

P93xxB Series P50xxB Series
This application enables internal pulse generators of P500xB or P502xB Series that can be used to control the internal pulse modulators, and it provides an integrated pulse application that uses the wideband-detection method.
S97025xB extends the capabilities of S97024xB and provides an easy way to set up point-in-pulse measurements with pulse width as narrow as 200 ns, and pulse-profile measurements with 40 ns minimum timing resolution. Using the built-in pulse modulators with pulse width as narrow as 1 us, the P50xxB Series VNAs are complete pulsed-RF measurement solutions, eliminating the need for external test sets and pulse generators. Examples of pulse modulation shapes with 1 us and 100 us pulse width using the internal pulse modulators are included in P50xxB data sheet (3121-1235). S97025xB also controls external pulse generators and modulators and can synchronize to external master pulses. Y1731A I/O Adapter Module is recommended to access pulse signals with SMB connectors if using external master pulses or external pulse modulators.
S970250B is available with P50xxB up to 20 GHz (20 GHz models or less), and S970251B is available for P50xxB greater than 20 GHz. Hardware options are not needed for pulsed-RF measurements with P50xxB.
Find us at www.keysight.com Page 14

Noise figure measurements with vector correction (S97029B)

P93xxB Series P50xxB Series
This software application enables high-accuracy noise figure and noise-power measurements of amplifiers, frequency converters, and mixers, utilizing Keysight’s unique vector-source-correction technique that uses Keysight N469xD Series ECal module as a source-impedance tuner1 to remove the effects of imperfect system­source match. This approach yields accuracy that surpasses that provided by the Y-factor method and other cold-source implementations, especially for in-fixture, on-wafer, and automated-test environments. S97029B controls N469xD Series ECal modules configured as impedance tuners for use with the P50xxB Series VNA.
A scalar-calibrated method is also available that offers less accuracy but is faster and does not require an impedance tuner. This method requires an external switch to isolate the P50xxB’s source signal during noise figure measurements. A 50-ohm load must be connected to the DUT’s input using the switch.
The instrument’s standard receivers are used for noise figure measurements with the S97029B. An external preamplifier, filter(s) and switches are required for devices with < 30 dB of excess noise (gain plus noise figure in dB). Y1731A I/O Adapter Module is recommended for synchronized measurements using the P50xxB VNA and external switches. A typical block diagram of vector-calibrated noise figure measurements is shown.
For calibration, a standard mechanical cal kit or ECal module is required for the S-parameter portion of the cal (an ECal used as a tuner cannot be shared for calibration). To calibrate a standard receiver for noise figure measurements, a power meter/sensor is required. A 346-serires noise source (Keysight 346C or 346C-K01 recommended) can be used for noise calibration of the instrument’s receiver, when a preamplifier is located before the receiver. All calibration accessories and external hardware must be ordered separately.
Noise figure measurements with the P50xxB are verified between 50 MHz and 45 GHz.
1. 8509x, N443xD or N755xA Series ECal modu les are not su pported.
Find us at www.keysight.com Page 15

Scalar mixer/converter measurements (S97082B)

P93xxB Series P50xxB Series
The S97082B provides frequency offset mode (FOM) to set the frequency of the VNA’s internal source(s) independently from where the receivers are tuned, and to configure external sources using External Device Configuration. This functionality is also included with S97029/083/084/086/087/089/S97090xB.
With a simple setup and calibration, this application delivers the highest accuracy for scalar conversion-loss/gain measurements by combining one-port and power-meter calibrations to remove mismatch errors. S97082B provides an intuitive and easy-to-use user interface for setting up mixer and converter measurements, with single or dual conversion stages. It can control external signal generators for use as LO signals. Supported external sources include the Keysight ESG, PSG, EXG, and MXG Series, as well as other SCPI-controlled signal generators.
S97082B is compatible with S97084B, which enables measurements of converters with internal LOs.

Vector and scalar mixer/converter measurements (S97083B)

P93xxB Series P50xxB Series
This application includes the scalar mixer/converter plus phase (SMC+Phase) measurement class that provides fully calibrated conversion gain/loss, relative phase, and absolute group delay measurements of mixers and converters without the need for reference or calibration mixers. Eliminating the calibration mixer requires a
1
U9391C/F/G comb generator U9391C/F or 800 mA for U9391G. A macro-based series-reference-mixer version of VMC is also included for measuring the phase difference between multiple paths or devices, or for measuring phase shifts within the frequency-converting device. The macro also supports characterization of the calibration mixer that is supplied by the user.
and an external DC power supply capable of sourcing +15 V and 300 mA for
SMC+Phase with the S97083B provides an intuitive and easy-to-use user interface for setting up mixer and converter measurements, with single or dual conversion stages. It can control the analyzer’s built-in source(s) as well as external signal generators for use as LO signals. Supported external sources include the Keysight ESG, PSG, EXG, and MXG Series, as well as other SCPI-controlled signal generators. S97083B is a superset of S97082B, so they should not be installed together. S97083B is compatible with S97084B, which enables measurements of converters with internal LOs.
1. When a comb generator is used as a phase reference for calibrat ion and the s tart frequency of the
measurement is less than 630 MHz, a user-supplied calibration mixer is required.

Embedded LO capability (S97084B)

P93xxB Series P50xxB Series
This application tunes the analyzer’s receivers to the output frequency of the converter under test without the need for access to internal LOs or a common reference signal. S97084B is intended to work with S97082B and S97083B measurement applications.
Find us at www.keysight.com Page 16
Gain compression measurements (S97086B)
P93xxB Series P50xxB Series
The gain compression application (GCA) provides input power, output power, gain, and phase at the compression point of an amplifier or frequency converters, over a specified frequency range. GCA’s SMART Sweep is very fast and easy-to-use. GCA also includes a guided calibration that corrects for absolute power levels, frequency response, and mismatch errors.
Intermodulation distortion measurements (S97087B)
P93xxB Series P50xxB Series
The intermodulation distortion (IMD) application makes it very easy to set up and calibrate swept-IMD measurements of both amplifiers and frequency converters. It controls the frequency and power of internal and external sources and tunes the receivers to the main tones as well as the IMD products in a single measurement channel. The user can sweep either the center frequency of the two stimulus signals, the frequency spacing of the two stimulus signals about a fixed center frequency, or the power of one or both stimulus signals or the power of the LO signal. The analyzer can measure intermodulation distortion products of order 2, 3, 5, 7, or 9, and can display the associated intercept points.
The most common method to use S97087B is to use a 4-port P502xB with the internal second source option 402, with each built-in source generating one of the two tones. Any of the remaining ports can be used as the receive port. Alternatively, two cascaded Streamline Series VNAs can be used, with each one generating one of the tones. This scenario requires the S97551B multiple instrument measurements software. An external combiner is required in either case.
To use an external source for the second tone, the Direct Receiver Access (DRA) configuration must be used, which also requires an external dual directional coupler.
Configuration using a 4-port P502xB (left), 6-port configuration using 2-port P500xB and 4-port P502xB (right).
Source phase control (S97088B)
P93xxB Series P50xxB Series
This application allows users to set calibrated, arbitrary phase differences between two internal sources with the P502xB’s option 402 (internal second source option), or multiple instrument measurements using two P50xxB VNAs with S97551B. The phase difference can be fixed or swept between two specified phase values. S97088B also controls the relative power level between the sources using the receiver-leveling feature. S97088B is
Find us at www.keysight.com Page 17
targeted for active-load control, where the analyzer provides a precise, electronically settable impedance to the output port of a device, while gain and output power are measured. This capability can be combined with external load-pull software to create traditional load-pull power contours.
Differential and I/Q device measurements (S97089B)
P93xxB Series P50xxB Series
This application combines source-phase control of multiple internal or external sources with frequency-offset mode, enabling simplified test of I/Q modulators/converters and differential mixers, and harmonic measurements of differential amplifiers. The phase difference between sources can be fixed (for example, at 90 or 180 degrees), or swept between two specified phase values. Providing accurate control of the relative phase between sources eliminates the need for hybrid couplers and baluns to create quadrature or differential signals. After achieving the desired phase alignment, the instrument’s receivers can be tuned to all frequencies needed to characterize the DUT. On an I/Q modulator for example, measurements can be made of both the desired and suppressed conversion bands, along with LO leakage, harmonics and other spurious signals. Phase sweeps can be used to determine a DUT’s phase imbalance versus frequency. Users can specify measurements with individual receivers or multiple receivers combined with a wide range of mathematical operators. Power measurements can employ match correction for increased accuracy. S97089B works on the P502xB with option 402 (internal second source option), or multiple instrument measurements using two P50xxB VNAs with S97551B.
Spectrum analysis (S97090xB)
P93xxB Series P50xxB Series
The spectrum analyzer (SA) application adds high-performance microwave spectrum analysis to P50xxB VNAs. With fast stepped-FFT sweeps resulting from optimized data processing, the SA application provides quick spurious searches over broad frequency ranges. Simultaneous spectrum measurements can be done using test and reference receivers. This multi-channel SA can be used with the internal swept-signal generators for efficient measurements of spurious signals emanating from mixers and frequency converters. The SA application employs source-power and receiver-response calibration as well as fixture de-embedding, providing in-fixture and on-wafer spectrum measurements with the highest level of accuracy. Accurate and fast noise power ratio (NPR) measurement capability is also included in this software application. It provides the ability of setting up and controlling of external signal generator/arbitrary waveform generators and performing the calibration of the flatness of a wideband signal and distortion signals in notches and achieving very fast NPR measurements. Optional external attenuators should be connected to the VNA’s test ports to avoid receiver compression when measuring large signals. The upper frequency range of the SA application is determined by the installed SA license or the stop frequency of the instrument, whichever is smaller. There are several frequency choices intended for standalone instruments: S970900B to 4.5 GHz, S970901B to 6.5 GHz, S970902B to 9 GHz, S970903B to 14 GHz, S970904B to 20 GHz, S970905B to 26.5 GHz, S970906B to 32 GHz, S970907B to 44 GHz, and S970908B to 53 GHz.
True-mode stimulus (S97460B)
P93xxB Series P50xxB Series
The application provides mismatch-corrected true-mode (true differential mode and true common mode) stimulus and enables accurate balanced measurements under real operating conditions. It also provides balanced measurements with forward-only sweeps, reverse-only sweeps, and frequency or power sweeps with arbitrary phase and amplitude offsets.
Find us at www.keysight.com Page 18
S97460B works on the P502xB with option 402 (internal second source option), or multiple instrument measurements using two P50xxB VNAs with S97551B.
Multiple instrument measurements (S97551B)
P93xxB Series P50xxB Series
Keysight Streamline Series VNA offers a capability to extend the number of ports for your measurements by using up to two VNA instruments. The instruments may be identified as one VNA by the firmware on an external host PC. At least one VNA instrument connected to your PC must have one S97551B license to maintain multiport capabilities with multiple instruments.
The frequency of the multiport array is determined by the lowest frequency instrument configured in the array. For example, a 6-port analyzer configuration using a P9372B (9 GHz) and a P9384B (20 GHz) would have a maximum frequency of 9 GHz when performing 6-port measurements. This behavior extends to other measurement application software (S97xxxB). A 4-port analyzer configuration using a P5002B (9 GHz) without SA software and P5004B (20 GHz) with S970904B (SA software up to 20 GHz) would have a maximum frequency of 9 GHz for S-parameter and SA measurements. In a multiport configuration, only one instrument must have a valid license of S97xxxB for these capabilities to function in multiport mode.
Each instrument is connected into the array with Keysight cables. A Y1701A Multiple USB instruments configuration kit must be ordered separately for connection of two 2-port VNA (P937xB or P500xB with Y1701A Option 002) or two 4- or 6-port VNA (P938xB or P502xB with Y1701A Option 002 or 003).
While all models of P937xB and P938xB, or P500xB and P502xB Series can be connected for multiport configurations, P93xxB and P50xxB, P937xA and P93xxB, or P50xxA and P50xxB cannot be combined to configure a multiport VNA due to hardware differences.
Figure 5. 4-port configuration using two P9377B VNAs (left), 6-port configuration using 2-port P9374B and 4-port P9384B (right).
Find us at www.keysight.com Page 19
Multiport calibration assistant (S97552B)
P93xxB Series P50xxB Series
This application software provides a tool to flexibly create and manage cal sets for multiport measurements. For example, two cal sets with independent test ports can be combined as one cal set with a greater number of test ports. The software also offers a way to re-calibrate using a certain set of test ports to refresh the cal set.
At least one ECal module is required for the operation of S97552B.
Multiport calibrated measurements with switch instruments (S97553B)
P93xxB Series P50xxB Series
This application provides a macro which enables full multiport error correction and measurement capabilities using external switch instruments such as the P9164A/B or P9165A/B USB Solid-state Switch Matrix. The software delivers an easy-to-use measurement wizard that simplifies measurement procedures and reduces the setup time of complicated multiport measurements. Measured multiport S-parameters are displayed in the VNA viewer.
Multiport configurations using up to two Streamline Series VNAs and up to two switch matrices are supported.
Only the 4-port ECal modules (i.e. N4431/32/33D) are supported for multiport calibration with the S97553B.
Banded millimeter-wave network analysis (S97560B)
P93xxB Series P50xxB Series
This software enables P93xxB and P50xxB Streamline Series VNA to work with N5252AWxx frequency extenders (up to 170 GHz) or a combination of N5252APXI test set and N5262BWxx frequency extenders (up to 330 GHz) and allows you to configure a banded millimeter-wave network analyzer.
Available bands depend on the combination of the VNA and the frequency extenders. Refer to Banded Millimeter Wave Network Analysis technical overview (literature number 5992-2177EN) for more details.
Find us at www.keysight.com Page 20
Material measurements
P93xxB Series P50xxB Series
The Keysight N1500A materials measurement suite streamlines the process of measuring complex permittivity and permeability with a vector network analyzer. Various type of measurements, such as transmission line and free space, arch reflectivity, resonant cavity, and coaxial probe are available as options. The easy-to-use software guides the user through setup and measurement, instantly converting S-parameter network analyzer data into the data format of your choice and displaying the results within seconds. Results can be charted in a variety of formats: εr’, εr”, tand, μr’, μr”, tandμ, and Cole-Cole.
N1930xB Physical Layer Test System (PLTS) software
P93xxB Series P50xxB Series
The PLTS software platform has become an industry standard for calibration, measurement, and analysis of linear passive interconnects such as cables, connectors, backplanes, and printed circuit boards. Utilizing either a vector network analyzer (VNA) or a time domain reflectometer (TDR), fast and accurate measurements can be obtained without in-depth knowledge of microwave measurement techniques. Refer to
www.keysight.com/find/plts or the technical overview (literature number 5989-6841EN) for more details.
Device Measurement eXpert (DMX) (S94601B)
P93xxB Series P50xxB Series
S94601B assists the users in setting up measurements automatically with customizable templates to use consistent measurement settings in different stages of design and test workflow. It supports various types of DUTs, and once the right DUT is selected, appropriate measurement parameters are listed, and the users can select the parameters of interest.
It provides intelligent algorithms that optimizes measurement setups based on instrument and DUT performance limitations and protects both the DUT and the instrument.
Limit assistant (S94602B)
P93xxB Series P50xxB Series
S94602B allows the users to acquire the data from a VNA or data file in csv, s2p or prn format and easily generate limit masks for complex limit test conditions with an intuitive and convenient graphical interface for production test applications.
Automated measurement expert (AMX)
P93xxB Series P50xxB Series
This software package is a smart software solution for automated S-parameter measurements with the Streamline Series VNA. The S94701A AMX test plan builder provides an interactive GUI which helps you edit your DUT test plans and generate test sequence files for the network analyzers. The S94702A and KS8400A software running on a host PC executes the test sequences according to the optimized test sequence files generated by the S94701A.
Find us at www.keysight.com Page 21
PathWave Vector Signal Analysis (89600 VSA) software
P93xxB Series P50xxB Series
PathWave VSA software is a comprehensive set of tools for demodulation and vector signal analysis. Supporting P500xB or P502xB, the software provides the flexibility and sophisticated measurement tools essential to find and fix signal problems. The multi-channel measurement capability with the P50xxB VNA can configure, execute and display multiple measurements simultaneously to characterize numerous signals. Powerful display tools enable comparisons and correlation of results from different measurements.
S97090xB spectrum analysis software and PathWave Vector Signal Analysis (89600 VSA) software are required to enable vector signal analysis using the P50xxB. The maximum bandwidth of the P50xxB is 29 MHz. To synchronize to external vector signal generators, Y1730A-800/801/810 cables are recommended for connection with 10 MHz external reference ports (Ref In/Ref Out) of the P50xxB.
Refer to www.keysight.com/find/89600 to learn more about 89600 VSA software.
Vector Network Analyzer (VNA) simulator (S9405xB)
VNA simulator runs on your PC and simulates operations of Keysight’s latest VNAs (PNA/ENA/PXI or Streamline Series VNA), eliminating the need for a VNA for your test program development. The simulator has standard and advanced versions. The standard version (S94050B) supports operations of standard S-parameter measurement class. The advanced version (S94051B) gives access to all advanced capabilities with S97xxxB software (ex. spectrum analysis) except S97007B, S97552B or S97553B software for Streamline Series VNAs.
S9405xB is available on subscription licenses only.
USB Type-C interconnects compliance test software (S94USBCB)
This application software provides a fast and automated way to characterize and debug USB Type-C interconnect products such as cable assemblies, connectors, or raw cables on USB4, USB 3.2, USB 3.1, and USB 2.0 technologies. It allows to automatically execute compliance tests with a 4-port VNA based on the latest USB Type-C Compliance Test Specifications (CTS) by USB-IF, and displays the comprehensive test results including test limits, margins, and pass/fails status of each test parameter.
S94USBCB supports multiport configurations using a 4-port VNA and L8990M-0LZ (e.g. 20-port switch matrix), which fully automates the test procedure and dramatically reduces the test time by eliminating manual cable reconnections.
S94USBCB requires, S97011B, KS8400B PathWave Test Automation and KS8401B HTML5 Result Listener Plugin.
Refer to www.keysight.com/find/S94USBCB or S94USBCB USB Type-C Interconnects Compliance Software data sheet (literature number 3122-1253) for more details.
Find us at www.keysight.com Page 22

Upgrading Your System

Upgrade kits are available to add options after initial purchase. To upgrade the Streamline Series VNAs, order the corresponding item number. To add application software, order the appropriate standalone model numbers (S97xxxB).
For P937xB and P938xB Series, a complete list of upgrade kits is available on our Web site:
www.keysight.com/find/p93xxb-upgrades
Find us at www.keysight.com Page 23

Direct Receiver Access (DRA) Configurations

Keysight Streamline Series VNAs support direct receiver access (DRA) configurations to flexibly set up test
1
systems with optimized performance by integrating with external components attenuators, or directional couplers. This function (included in the VNA firmware) combines three physical test ports of the Streamline Series VNA and defines them as a single logical port. You can perform measurements including calibration and post processing using the logical ports.

DRA configuration with high-power

One example using the DRA configuration is S-parameter measurements with high power. When you need an input level higher than the analyzer’s source can provide, pre-amplifiers are necessary to boost the power level prior to the DUT. However, the reference signal is measured before the booster amplifier with a standard configuration of a 2-port VNA, and temperature drift or high reverse isolation of a booster amplifier will prevent accurate reflection measurements of the DUT’s input.
A typical configuration for high-power measurements with the 4-port VNA is shown. A 4-port VNA (ex. P938xB or
2
P502xB with Option 400/402), or a 4-port VNA are needed to setup the DRA configuration on port 1. Reflected and transmitted signals of the DUT are detected with all the logical receivers, and the VNA firmware provides measured high-power S-parameters.
using two 2-port VNAs (ex. P937xB or P500xB with Option 200)
such as booster amplifiers,
External components should be selected based on their high-power limits or frequency range specifications. Be sure that these components can handle the output power level of the booster amplifier.
DRA configurations are supported on all Keysight Streamline Series VNA systems (P93xxA/B and P50xxA/B) that at a minimum have 4-ports, with either 4-ports in one VNA or 4-ports as a result of two VNAs connected together.
1. External componen ts are not included in Streamline Series VNA. Must be purchased separately.
2. The S97551B software is required to configure a multiport VNA with multiple Streamline Series VNAs.
Find us at www.keysight.com Page 24

Keysight Software Licensing Options Provide Flexibility and Support

License term

Descriptions
Perpetual
Perpetual licenses can be used indefinitely.
Subscription
Licenses can be used through the term of the subscription (6, 12, 24, or 36 month)

License type

Descriptions
Node locked1
License can be used on one specified instrument/computer.
Subscription
Description
Perpetual licenses are sold with a 12 (default), 24, 36, or 60-month software support subscription. Support subscriptions may be renewed for a fee after that.
Software subscription licenses include a software support subscription through the term of the license.
Projects ramp up and down, teams grow and shrink, and projects move location. In such a dynamic environment, you need flexible licensing options that allow you to balance your project’s requirements. Whether your software will be a staple for years to come or you have a short-term need for a leading-edge measurement application, Keysight’s licensing puts you in charge.

Choose your term. Choose your type. Keep control of your budget.

Select a node-locked license type.
Select a subscription or perpetual license term, depending on how long you need to use the software.
Each license is sold with a KeysightCare software support subscription which provides technical
support with ensured response time, proactive software updates and enhancements.
Choose a license term and type that best suits your requirements from the table below.
License term
License type
1. Node-locked license is the only license type for S97xxxB.

KeysightCare Software Support Subscription provides peace of mind amid evolving technologies.

Ensure your software is always current with the latest enhancements and measurement standards.
Gain additional insight into your measurement problems with live access to our team of technical
experts.
Stay on schedule with fast turnaround times and priority escalations when you need support.
KeysightCare software support
Find us at www.keysight.com Page 25

Ordering Information

License term
Product
License type
Perpetual
Subscription
License
Support
License and support
Node-locked (fixed)
R-A5C-001-A
+
R-A6C-001-z
R-A4C-001-z
z = Duration
z = Duration
(default)
X 24 months
L 12 months
Y 36 months
X 24 months
Z 60 months
Y 36 months
Step 1. Choose your software product.
Step 2. Choose your license term: perpetual or subscription.
Step 3. Choose your license type: node-locked.
Step 4. Depending on the license term, choose your subscription or support duration.
S97xxxB1
1. For S97xxxB softwar e, the fixed-perp etual with a 1 2-mo nths, support subscription (R-A6C- 001- L) is the only license type that can be ordered as part of the instrument and insta lled. The other license types for S97 xxxB software must be ordered separately and installed from the web after the receipt of the instrument.
L 12 months
F 6 months
Find us at www.keysight.com Page 26

Measurement Accessories

A complete list of RF and microwave test accessories is available on our Web site: www.keysight.com/find/mta
Accessories are available in these connector types: 50 Ω Type-N, 3.5 mm, 7 mm, 2.4 mm, 2.92 mm,
1.85 mm, 1.0 mm, and waveguide. Test port cables and a calibration kit should be added for a complete
measurement system. A verification kit is used to verify corrected system performance.

Cable and adapter sets

Keysight offers cables in the following types:
Single cables in semi-rigid and flexible
Cable sets in semi-rigid and flexible
There are also adapter sets available that protect the test port and convert the port to the desired connector interface. These kits contain:
One male adapter
One female adapter
To attain the best mechanical rigidity for device connection, use a single cable and the appropriate special adapter set. To attain the greatest flexibility for device connection, use a cable set.
Find us at www.keysight.com Page 27
Calibration kits

Coaxial measurements

Mechanical calibration kits include standards, such as opens, shorts and loads, which are measured by the network analyzer for increased measurement accuracy.
Choose a calibration kit for each connector type to be used.
Economy, includes:
Open standards (male and female)
Short standards (male and female)
Fixed-termination standards (male and female)
Standard, includes the devices in the economy kit and adds:
Sliding load standards (male and female) or a series of offset shorts
Precision, includes the devices in the economy kit and adds:
50 Ω airline(s) for TRL calibration
TRL adapters
Electronic calibration (ECal) kits replace mechanical calibration standards with one solid-state calibration module that is controlled by the network analyzer via USB, to present many different impedances to the test ports. A full two-port calibration can be performed quickly with a single connection. This technique reduces operator errors and connector wear and abrasion.
For more information about ECal modules, refer to the technical overview 5963-3743E (N4690, 85090, N4430 and N7550 Series).
Waveguide measurements
For waveguide measurements, Keysight offers mechanical calibration kits that include:
Waveguide-to-coax adapters (X, P, K, R, Q, U, V)
Precision waveguide section
Flush short circuit
Fixed terminations
Straight section
Find us at www.keysight.com Page 28

For devices with 1.85 mm connectors

Mechanical calibration kit

85058B standard: DC to 67 GHz. Includes:
85058-60101 1.85 mm (m) short 5.4 mm
85058-60102 1.85 mm (m) short 6.3 mm
85058-60103 1.85 mm (m) short 7.12 mm
85058-60104 1.85 mm (m) short 7.6 mm
85058-60105 1.85 mm (f) short 5.4 mm
85058-60106 1.85 mm (f) short 6.3 mm
85058-60107 1.85 mm (f) short 7.12 mm
85058-60108 1.85 mm (f) short 7.6 mm
85058-60109 1.85 mm male open
85058-60110 1.85 mm female open
85058-60111 1.85 mm male load
85058-60112 1.85 mm female load
85058-60113 1.85 mm (m) to 1.85 mm (m) adapter
85058-60114 1.85 mm (f) to 1.85 mm (f) adapter
85058-60115 1.85 mm (m) to 1.85 mm (f) adapter
85058E economy: DC to 67 GHz. Includes:
85058-60101 1.85 mm (m) short 5.4 mm
85058-60105 1.85 mm (f) short 5.4 mm
85058-60109 1.85 mm male open
85058-60110 1.85 mm female open
85058-60123 1.85 mm male load
85058-60124 1.85 mm female load
85058-60113 1.85 mm (m) to 1.85 mm (m) adapter
85058-60114 1.85 mm (f) to 1.85 mm (f) adapter
85058-60115 1.85 mm (m) to 1.85 mm (f) adapter

Electronic calibration kits

N4694D Microwave ECal: DC or 10 MHz to 67 GHz, 2-ports. Includes:
Option F0F: Both connectors are 1.85 mm female on module Option M0F: 1 female and 1 male 1.85 mm connector on module Option M0M: Both connectors are 1.85 mm male on module Option 0DC: DC to 67 GHz Option 100: 10 MHz to 67 GHz Option 00A adds:
85058-60113 1.85 mm (m) to 1.85 mm (m) adapter 85058-60114 1.85 mm (f) to 1.85 mm (f) adapter
Find us at www.keysight.com Page 29

For devices with 2.4 mm connectors

Mechanical calibration kit

85056A standard: DC to 50 GHz. Includes:
00901-60003 2.4 mm (m) fixed broadband load
00902-60004 2.4 mm (f) fixed broadband load
00915-60003 2.4 mm (m) sliding load
00915-60004 2.4 mm (f) sliding load
85056-60005 2.4 mm (m) to 2.4 mm (m) adapter
85056-60006 2.4 mm (f) to 2.4 mm (f) adapter
85056-60007 2.4 mm (m) to 2.4 mm (f) adapter
85056-60020 2.4 mm (m) short
85056-60021 2.4 mm (f) short
85056-60022 2.4 mm (m) open
85056-60023 2.4 mm (f) open
85056D economy: DC to 50 GHz. Includes:
00901-60003 2.4 mm (m) fixed broadband load
00902-60004 2.4 mm (f) fixed broadband load
85056-60005 2.4 mm (m) to 2.4 mm (m) adapter
85056-60006 2.4 mm (f) to 2.4 mm (f) adapter
85056-60007 2.4 mm (m) to 2.4 mm (f) adapter
85056-60020 2.4 mm (m) short
85056-60021 2.4 mm (f) short
85056-60022 2.4 mm (m) open
85056-60023 2.4 mm (f) open

Electronic calibration kits

N4693D 1 Microwave ECal: DC or 10 MHz to 50 GHz, 2-ports. Includes:
Option F0F: Both connectors are 2.4 mm female on module Option M0F: 1 female and 1 male 2.4 mm connector on module Option M0M: Both connectors are 2.4 mm male on module Option 0DC: DC to 50 GHz Option 100: 10 MHz to 50 GHz Option 00A adds:
85056-60005 2.4 mm (m) to 2.4 mm (m) adapter 85058-60006 2.4 mm (f) to 2.4 mm (f) adapter
Find us at www.keysight.com Page 30

For devices with K connectors (2.92 mm)

Electronic calibration kits

N4692D Microwave ECal: DC or 10 MHz to 40 GHz, 2-ports. Includes:
Option F0F: Both connectors are 2.92 mm female on module Option M0F: 1 female and 1 male 2.92 mm connector on module Option M0M: Both connectors are 2.92 mm male on module Option 0DC: DC to 40 GHz Option 100: 10 MHz to 40 GHz Option 00A adds:
N4692-60021 2.92 mm (m) to 2.92 mm (m) adapter N4692-60022 2.92 mm (f) to 2.92 mm (f) adapter

Adapters

11904A 2.4 mm (m) to 2.92 mm (m) 11904B 2.4 mm (f) to 2.92 mm (f) 11904C 2.4 mm (m) to 2.92 mm (f) 11904D 2.4 mm (f) to 2.92 mm (m) 11904S 2.4 mm to 2.92 mm:
Adapter set, contains 4 matched adapters
Find us at www.keysight.com Page 31

For devices with 3.5 mm or SMA connectors

Mechanical calibration kits

85033E economy: DC to 9 GHz. Includes:
85033-60016 3.5 mm (m) load
85033-60017 3.5 mm (f) load
85033-60018 3.5 mm (m) open
85033-60019 3.5 mm (f) open
85033-60020 3.5 mm (m) short
85033-60021 3.5 mm (f) short
8710-1761 torque wrench
Option 85033E-100 adds: 85027-60005 3.5 mm (f) to 3.5 mm (f) adapter
Option 85033E-200 adds: 85027-60007 3.5 mm (m) to 3.5 mm (m) adapter
Option 85033E-300 adds: 85027-60006 3.5 mm (m) to 3.5 mm (f) adapter
Option 85033E-400 adds:
1250-1744 3.5 mm (f) to 50 Ω Type-N (m) adapter
1250-1743 3.5 mm (m) to 50 Ω Type-N (m) adapter
1250-1745 3.5 mm (f) to 50 Ω Type-N (f) adapter
1250-1750 3.5 mm (m) to 50 Ω Type-N (f) adapter
Option 85033E-500 adds:
1250-1746 3.5 mm (m) to 7 mm adapter (two included)
1250-1747 3.5 mm (f) to 7 mm adapter (two included)
85052B standard: DC to 26.5 GHz. Includes:
00902-60003 3.5 mm (m) fixed load
00902-60004 3.5 mm (f) fixed load
00911-60019 3.5 mm (m) sliding load
00911-60020 3.5 mm (f) sliding load
85052-60006 3.5 mm (m) short
85052-60007 3.5 mm (f) short
85052-60008 3.5 mm (m) open
85052-60009 3.5 mm (f) open
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
85052-60013 3.5 mm (f) to 3.5 mm (m) adapter
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
Find us at www.keysight.com Page 32
85052C precision TRL: DC to 26.5 GHz. Includes:
Connector type
Female
Male
Connector type
Female
Male
3.5 mm
101
102
Type-N 50 Ω
203
204
7-16
205
206
00902-60003 3.5 mm (m) fixed load
00902-60004 3.5 mm (f) fixed load
85052-60006 3.5 mm (m) short
85052-60007 3.5 mm (f) short
85052-60008 3.5 mm (m) open
85052-60009 3.5 mm (f) open
85052-60032 3.5 mm (f) to 3.5 mm (f) adapter
85052-60033 3.5 mm (m) to 3.5 mm (m) adapter
85052-60034 3.5 mm (f) to 3.5 mm (m) adapter
85052-60035 3.5 mm short TRL line
85052-60036 3.5 mm long TRL line
85052D economy: DC to 26.5 GHz. Includes:
00902-60003 3.5 mm (m) fixed load
00902-60004 3.5 mm (f) fixed load
85052-60006 3.5 mm (m) short
85052-60007 3.5 mm (f) short
85052-60008 3.5 mm (m) open
85052-60009 3.5 mm (f) open
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
85052-60013 3.5 mm (f) to 3.5 mm (m) adapter
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter

Electronic calibration (ECal) modules

85093C RF ECal: 300 kHz to 9 GHz, 2-ports. Standard module includes: Option 00F: Both 3.5 mm connectors are female Option 00M: Both 3.5 mm connectors are male Option M0F: One female and one male connector, both 3.5 mm Option 00A adds:
o
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter
o
85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
Option 150: Replaces standard storage container with a wooden box

85093C mixed-connector options

Port A Option Port B Option
Find us at www.keysight.com Page 33
85093D RF Electronic Calibration Module (ECal), DC/300 kHz to 9 GHz, 3.5 mm, 2 Ports.
Type-N
203
204
7-16
205
206
4.3-10
207
208
Connector Type
Port A Option
Port B Option
Port C Option
Port D Option
Four 3.5 mm (f)
010
Four Type-N 50 Ω (f)
020
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Type-N 50 Ω (f)
103
203
303
403
Type-N 50 Ω (m)
104
204
304
404
7-16 (f)
105
205
305
405
7-16 (m)
106
206
306
406
4.3-10 (f)
107
207
307
407
4.3-10 (m)
108
208
308
408
Options available
Standard connector option of 3.5 mm
Option 003: 300 kHz to 9 GHz
Option 0DC: DC to 9 GHz
Option 00A: Add Type-N 50-Ohm adapters
Option 00F: Both 3.5 mm connectors are female on module
Option 00M: Both 3.5 mm connectors are male on module
Option MOF: One female and one male connector, both 3.5 mm
Option 1A7: ANSI Z540-1-1994 Calibration
Option A6J: Calibration + Uncertainties + Guardbanding (Not Accredited)
Option UK6: Commercial calibration certificate with test data
85093D mixed-connector options
Port A Option Port B Option
Connector type Female
3.5 mm 101 102
Male
Connector type Female
Male

N4431D Microwave ECal: DC to 13.5 GHz, 4 ports

Frequency option (mandatory)
Option 0DC: DC to 13.5 GHz
Connector options
Option 010: 3.5 mm female connector on four ports on module
Option 020: Type-N female connector on four ports on module
Find us at www.keysight.com Page 34

N4433D ECal: DC/300 kHz to 26.5 GHz, 4-ports

Connector Type
Port A Option
Port B Option
Port C Option
Port D Option
Four 3.5 mm (f)
010
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Frequency options
Option 0DC: DC to 26.5 GHz
Option 003: 300 kHz to 26.5 GHz
Connector Options:
Option 010: 3.5 mm female connector on four ports on module
N4691D Microwave ECal, DC or 300 kHz to 26.5 GHz, 2-ports. Includes:
Option F0F: Both 3.5 mm connectors are female Option M0F: One female and one male connector, both 3.5 mm Option M0M: Both 3.5 mm connectors are male Option 0DC: DC to 26.5 GHz Option 003: 300 kHz to 26.5 GHz Option 00A adds:
85052-60012 3.5 mm (f) to 3.5 mm (f) adapter 85052-60014 3.5 mm (m) to 3.5 mm (m) adapter
N755XA Series Economy ECal, 2-ports:
N7550A DC to 4 GHz, 2-ports
N7551A DC to 6.5 GHz, 2-ports
N7552A DC to 9 GHz, 2-ports
N7553A DC to 14 GHz, 2-ports
N7554A DC to 18 GHz, 2-ports
N7555A DC to 26.5 GHz, 2-ports
Find us at www.keysight.com Page 35
N755xA Series includes:
Option 3FF: Both 3.5 mm connectors are female Option 3MF: One female and one male connector, both 3.5 mm Option 3MM: Both 3.5 mm connectors are male Option 150: Plastic storage box N7550X-151: 3.5 mm or 2.92 mm torque wrench

For devices with Type-N connectors

Mechanical calibration kits

85032F standard, DC to 9 GHz. Includes:
85032-60017 Type-N (m) fixed load
85032-60018 Type-N (f) fixed load
85032-60013 Type-N (m) open
85032-60014 Type-N (f) open
85032-60016 Type-N (m) short
85032-60015 Type-N (f) short
Option 85032F-100 adds:
85032-60021 Type-N (f) to Type-N (f) adapter
Option 85032F-200 adds:
85032-60019 Type-N (m) to Type-N (m) adapter
Option 85032F-300 adds:
85032-60020 Type-N (m) to Type-N (f) adapter
Option 85032F-500 adds:
85054-60001 Type-N (f) to 7 mm adapter (two included)
85054-60009 Type-N (m) to 7 mm adapter (two included)
85054B standard: DC to 18 GHz. Includes:
00909-60011 Type-N (m) fixed lowband load
00909-60012 Type-N (f) fixed lowband load
85054-60025 Type-N (m) short
85054-60026 Type-N (f) short
85054-60027 Type-N (m) open
85054-60028 Type-N (f) open
85054-60031 Type-N (f) to 7mm adapter
85054-60032 Type-N (m) to 7mm adapter
85054-60037 Type-N (f) to Type-N (f) adapter
85054-60038 Type-N (m) to Type-N (m) adapter
85054-80010 Type-N (f) sliding load
85054-80009 Type-N (m) sliding load
85054-60050 Type-N (f) connector gage
85054-60052 Type-N (f) gage master
85054-60051 Type-N (m) connector gage
85054-60053 Type-N (m) gage master
Find us at www.keysight.com Page 36
85054D economy: DC to 18 GHz. Includes:
3.5 mm
201
202
7-16
205
206
4.3-10
207
208
85054-60025 Type-N (m) short
85054-60026 Type-N (f) short
85054-60027 Type-N (m) open
85054-60028 Type-N (f) open
85054-60031 Type-N (f) to 7mm adapter
85054-60032 Type-N (m) to 7mm adapter
85054-60037 Type-N (f) to Type-N (f) adapter
85054-60038 Type-N (m) to Type-N (m) adapter
85054-60046 Type-N (m) fixed load
85054-60047 Type-N (f) fixed load

Electronic calibration (ECal) modules

85092C RF ECal: 300 kHz to 9 GHz, 2 ports. Includes:
Option 00F: Both Type-N connectors are female
Option 00M: Both Type-N connectors are male
Option M0F: One female and one male connector, both Type-N
Option 00A adds:
85054-60037 Type-N (f) to Type-N (f) adapter 85054-60038 Type-N (m) to Type-N (m) adapter
85092D RF Electronic Calibration Module (ECal), DC/300 kHz to 9 GHz, Type-N, 2 Ports.
Options available
Standard connector option of Type-N
Option 003: 300 kHz to 9 GHz
Option 0DC: DC to 9 GHz
Option 00A: Add Type-N 50-Ohm adapters
Option 00F: Both Type-N 50-Ohm connectors are female on module
Option 00M: Both Type-N 50-Ohm connectors are male on module
Option MOF: One female and one male connector, both type-N 50 ohm
Option 1A7: ANSI Z540-1-1994 Calibration
Option A6J: Calibration + Uncertainties + Guardbanding (Not Accredited)
Option UK6: Commercial calibration certificate with test data
85092D mixed-connector options
Port A Option Port B Option
Connector type Female Male Connector type Female Male
Type-N 103 104
Find us at www.keysight.com Page 37
Connector Type
Port A Option
Port B Option
Port C Option
Port D Option
Four 3.5 mm (f)
010
Four Type-N 50 Ω (f)
020
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Type-N 50 Ω (f)
103
203
303
403
Type-N 50 Ω (m)
104
204
304
404
7-16 (f)
105
205
305
405
7-16 (m)
106
206
306
406
4.3-10 (f)
107
207
307
407
4.3-10 (m)
108
208
308
408
Connector Type
Port A Option
Port B Option
Port C Option
Port D Option
Four Type-N 50 Ω (f)
020
3.5 mm (f)
101
201
301
401
3.5 mm (m)
102
202
302
402
Type-N 50 Ω (f)
103
203
303
403
Type-N 50 Ω (m)
104
204
304
404

N4431D ECal: DC to 13.5 GHz, 4-ports

Frequency option (mandatory)
Option 0DC: DC to 13.5 GHz
Connector options
Option 010: 3.5 mm female connector on four ports on module
Option 020: Type-N female connector on four ports on module

N4432D ECal: DC or 300 kHz to 18 GHz, 4-ports

Frequency options
Option 0DC: DC to 18 GHz
Option 003: 300 kHz to 18 GHz
Connector options
Option 020: Type-N female connector on four ports on module
Find us at www.keysight.com Page 38
N4690D Microwave ECal, DC or 300 kHz to 18 GHz, 2-ports. Includes:
Option F0F: Both Type-N connectors are female Option M0F: One female and one male connector, both Type-N connectors Option M0M: Both Type-N connectors are male Option 0DC: DC to 18 GHz Option 003: 300 kHz to 18 GHz Option 00A adds:
85054-60037 Type-N (f) to Type-N (f) adapter 85054-60038 Type-N (m) to Type-N (m) adapter
N755XA Series Economy ECal, 2-ports:
N7550A DC to 4 GHz, 2-ports
N7551A DC to 6.5 GHz, 2-ports
N7552A DC to 9 GHz, 2-ports
N7553A DC to 14 GHz, 2-ports
N7554A DC to 18 GHz, 2-ports
N755xA Series includes:
Option NFF: Both Type-N connectors are female
Option NMF: One female and one male connector, both Type-N
Option NMM: Both Type-N connectors are male
Option 150: Plastic storage box
N7550X-152: Type-N torque wrench

Adapter sets

11853A 50 Ω Type-N accessory kit. Includes:
1250-1472 Type-N (f) to Type-N (f) adapter (two included)
1250-1475 Type-N (m) to Type-N (m) adapter (two included)
11511A Type-N (f) short
11512A Type-N (m) short
11878A Type-N to 3.5 mm adapter kit. Includes:
1250-1744 3.5 mm (f) to 50 Ω Type-N (m) adapter
1250-1743 3.5 mm (m) to 50 Ω Type-N (m) adapter
1250-1745 3.5 mm (f) to 50 Ω Type-N (f) adapter
1250-1750 3.5 mm (m) to 50 Ω Type-N (f) adapter
11524A 7 mm to Type-N (f) adapter
11525A 7 mm to Type-N (m) adapter
Find us at www.keysight.com Page 39

For devices with 7 mm connectors

Mechanical calibration kits

85050B standard: DC to 18 GHz. Includes:
00909-60008 7 mm coax termination
85050-60006 7 mm fixed broadband load
85050-80007 7 mm short
85050-80010 7 mm open
85050-80011 7 mm sliding load
85050C precision TRL: DC to 18 GHz. Includes:
00909-60008 7 mm coax termination
85050-60003 7 mm to 7 mm airline
85050-60005 7 mm to 7 mm TRL adapter
85050-60006 7 mm fixed broadband load
85050-80008 7 mm short
85050-80009 7 mm short with collect
85050-80010 7 mm open
85050D economy: DC to 18 GHz. Includes:
85050-60006 7 mm fixed broadband load
85050-80007 7 mm short
85050-80010 7 mm open

Electronic calibration (ECal) modules

N4696D Microwave ECal: DC or 300 kHz to 18 GHz, 2-ports, 7 mm connectors. Includes:
Option 0DC: DC to 18 GHz
Option 003: 300 kHz to 18 GHz
85091D RF Electronic Calibration Module (ECal), DC/300 kHz to 9 GHz, 7 mm, 2 Ports.
Options available
Standard connector option of 7 mm
Option 003: 300 kHz to 9 GHz
Option 0DC: DC to 9 GHz
Option 1A7: ANSI Z540-1-1994 Calibration
Option A6J: Calibration + Uncertainties + Guardbanding (Not Accredited)
Option UK6: Commercial calibration certificate with test data
Find us at www.keysight.com Page 40

For devices with waveguide

Mechanical calibration kits

X-Band
X11644A standard, WR-90: 8.2 to 12.4 GHz. Includes:
00896-60008 X-band standard section
00910-60003 X-band termination
11644-20018 X-band short
11644-20021 X-band shim
X281C adapter (included in calibration kit): WR-90 to 7 mm
P Band
P11644A standard, WR-62: 12.4 to 18 GHz. Includes:
00896-60007 P-band standard section
00910-60002 P-band termination
11644-20017 P-band short
11644-20020 P-band shim
P281C adapter (included in calibration kit): WR-62 to 7 mm
K Band
K11644A standard, WR-42: 18 to 26.5 GHz. Includes:
00896-60006 K-band standard section
00910-60001 K-band termination
11644-20016 K-band short
11644-20019 K-band shim
K281C adapter (included in calibration kit): WR-42 to 3.5 mm (f) Option 012 WR-42 to 3.5 mm (m)
R Band
R11644A standard, WR-28: 26.5 to 40 GHz. Includes:
00914-20028 R-band termination
11644-20005 R-band short
11644-20003 R-band shim
11644-60001 R-band 10 cm straight waveguide
11644-60016 R-band 5 cm straight waveguide
Find us at www.keysight.com Page 41
Q Band
Q11644A standard, WR-22: 33 to 50 GHz. Includes:
11644-60005 Q-band termination
11644-20004 Q-band short
11644-20001 Q-band shim
11644-60002 Q-band 10 cm straight waveguide
11644-60017 Q-band 5 cm straight waveguide
U Band
U11644A standard, WR-19: 40 to 60 GHz. Includes:
11644-60006 U-band termination
11644-20004 U-band short
11644-20002 U-band shim
11644-60003 U-band 10 cm straight waveguide
11644-60018 U-band 5 cm straight waveguide
Find us at www.keysight.com Page 42

Verification kits

All Keysight verification kits include:
Precision Z0 airline or match thru
Mismatched airline or mismatch thru
Fixed attenuators (except 85059V)
Traceable measured data and uncertainties

85051B 45 MHz to 18 GHz 7 mm kit

Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming calibrated system performance, traceable to national standards. Test procedure is provided in the service manual.

85053B 300 kHz to 26.5 GHz 3.5 mm kit

Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming calibrated system performance, traceable to national standards. Test procedure is provided in the service manual.

85055A 300 kHz to 18 GHz Type-N kit

Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming calibrated system performance, traceable to national standards. Test procedure is provided in the service manual.

85057B 45 MHz to 50 GHz 2.4 mm kit

Includes attenuators, airline and mismatch airline with data on a USB drive for use in confirming calibrated system performance, traceable to national standards. Test procedure is provided in the service manual.

85058V 45 MHz to 67 GHz 1.85 mm kit

Includes attenuators, match thru and mismatch thru with data on a USB drive for use in confirming calibrated system performance, traceable to national standards. Test procedure is provided in the service manual.
Find us at www.keysight.com Page 43

Accessories

P93xxB Series
P50xxB Series
CalPod calibration refresh modules (8553xB/4xB)
CalPod calibration refresh modules allow in-situ calibration refreshes to be performed at the push of a button without removing the DUT or re-connecting calibration standards. CalPods are designed for any measurement situation where it is desirable to ensure that a valid calibration is present before recording measurement data. For example, they are useful for removing environmental effects such as cable movement and thermal-based changes of cables, connectors, and adapters, as well as connector and switch-matrix repeatability errors. The modules are especially useful in temperature, or thermal-vacuum chamber testing. CalPod modules are equipped with 2.92 mm connectors.
85530B 20 GHz ambient-temperature CalPod
85531B 20 GHz temperature-compensated CalPod
85532B 20 GHz thermal-vacuum compatible CalPod
85540B 40 GHz ambient-temperature CalPod
85541B 40 GHz temperature-compensated CalPod
85542B 40 GHz thermal-vacuum compatible CalPod
85523B CalPod controller (provides control for up to 4 CalPods)
85556A CalPod drive-cable splitter (allows control for up to 12 CalPods)
85554A CalPod drive cable extension (10-meter cascadable extension cable)
Find us at www.keysight.com Page 44
Learn more at: www.keysight.com
This inf ormation i s subject to c hange wit hout noti ce. © Keysight Techn ologies, 2 021 - 202 2, Published i n USA, S eptember 19, 2022, 3 121-1254.EN
Operating systems
Windows 10 (64-bit)
Available memory
4 GB minimum, 16 GB recommended
Available disk space
4 GB minimum
Display resolution
1024 x 768 minimum
Connection with VNA
Thunderbolt 3 and Thunderbolt 4
Recommended CPU
Intel Core i7 10th Generation or later recommended
later

PC Requirements for VNA Control

Requirement 1
Keysight IO Libraries
1. Visit www.keysight.com/find/VNATBTConnectivity for troubleshooting the co nnection of a host PC with Streamline Series VNA.
Keysight IO Libraries Suite 2022 Update 1 (18.2.28014.7) or

Literature Information

Keysight P93xxB and P50xxB Series Vector Network Analyzer – Data Sheet, 3121-1235.EN
Keysight P93xxB Series Vector Network Analyzer - Technical Overview, 3121-1262.EN
Keysight Vector Network Analyzer – Selection Guide, 5980-7603EN
Electric Calibration (ECal) Modules – Technical Overview, 5963-7343E

Web Resources

www.keysight.com/find/usb-vna
www.keysight.com/find/na
www.keysight.com/find/vnasoftware
www.keysight.com/find/ecal
For more information on Keysight Technologies’ products, applications, or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus
Find us at www.keysight.com Page 45