No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Bluetooth®
Manual Part Number
N9938-90006
Edition
Edition 1, February 2021
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined by
Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires
commercial computer software
under the same terms by which the
software is customarily provided to
the public. Accordingly, Keysight
provides the Software to U.S.
government customers under its
standard commercial license, which
is embodied in its End User License
Agreement (EULA), a copy of which
can be found at
http://www.keysight.com/find/sweu
la. The license set forth in the EULA
represents the exclusive authority by
which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license
set forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation
that is not customarily provided to
the public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall
apply, except to the extent that those
terms, rights, or licenses are
explicitly required from all providers
of commercial computer software
pursuant to the FAR and the DFARS
and are set forth specifically in
writing elsewhere in the EULA.
Keysight shall be under no obligation
to update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated
conditions are fully understood
and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a
WARNING notice until the
indicated conditions are fully
understood and met.
A NOTE calls the user’s attention
to an important point or special
information in the text.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including instrument
software upgrades, application information, and product information, browse to one of the following URLs,
according to the name of your product:
http://www.keysight.com/find/fieldfox
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements.
To search for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/fieldfoxsupport
3
A.12.2x Firmware – Release B Series FieldFox
For customers upgrading FieldFox firmware, the following is a list of changes from the previous release:
— N9950/1/2B Microwave VNA and Spectrum Analyzer and
N9960/1/2B Microwave Spectrum Analyzer FieldFox support
—Improved OTA
— LTE TDD support
—Improved 5G NR
——LTE Coexistence (NSA / DSS) support
— Improved Mapping functionality
4
Contacting Keysight
Assistance with test and measurements needs and information on finding a local Keysight office are available
on the Web at:
If you do not have access to the Internet, please contact your Keysight field engineer.
In any correspondence or telephone conversation, refer to the Keysight product by its model number and full
serial number. With this information, the Keysight representative can determine whether your product is still
within its warranty period.
To contact Keysight for sales and technical support, refer to support links on the following Keysight websites:
http://www.keysight.com/find (product specific information and support, software and documentation
updates)
http://www.keysight.com/find/assist.
http://www.keysight.com/find/assist (worldwide contact information for repair and service).
14. USB Antennas – (Full Capability Requires EMF Option 358, and either SA mode (Option 233 Mixed
Analyzers) or OTA—5G NR / 5G NR EVM Conducted Option 378)
N9913B4Vector Network Analyzer AND Spectrum Analyzer
N9914B6.5Vector Network Analyzer AND Spectrum Analyzer
N9915B9Vector Network Analyzer AND Spectrum Analyzer
N9916B14Vector Network Analyzer AND Spectrum Analyzer
N9917B18Vector Network Analyzer AND Spectrum Analyzer
N9918B26.5Vector Network Analyzer AND Spectrum Analyzer
N9950B32Vector Network Analyzer AND Spectrum Analyzer
N9951B44Vector Network Analyzer AND Spectrum Analyzer
N9952B50Vector Network Analyzer AND Spectrum Analyzer
N9933B4Spectrum Analyzer
N9934B6.5Spectrum Analyzer
N9935B9Spectrum Analyzer
N9936B14Spectrum Analyzer
Description
N9937B18Spectrum Analyzer
N9938B26.5Spectrum Analyzer
N9960B32Spectrum Analyzer
N9961B44Spectrum Analyzer
N9962B50Spectrum Analyzer
Minimum Frequency: 30 kHz for all VNA models; 9 kHz for all SA models (usable to 5 kHz)
FieldFox Options: For a comprehensive list, view the FieldFox Configuration
Guide at: http://literature.cdn.keysight.com/litweb/pdf/5992-3701EN.pdf
25
Accessories
Overview
Accessories
The following accessories are included with every FieldFox. Spare accessories
can be ordered at any time. Refer to Table 1-2. Some of the optional
accessories—not included unless requested—are listed in Table 1-3. For more,
refer to http://www.keysight.com/find/fieldfox.
Table 1-2 Accessories
Accessory Part
Number
N9910X–873
N9910X-876Lithium-Ion Battery (High Capacity)
N9910X–877
N9910X–880Soft-case w/ Backpack & Shoulder Strap
N9910X–891Quick Reference Guide (printed copy)
a. Depending on your region, the FieldFox comes with either a
N9910X-873 AC/DC adapter or a N9910X-877 AC/DC adapter.
Description
AC/DC Adapter (except for US/Canada)
AC/DC Adapter for US / Canada only
Optional Accessories
Table 1-3 Optional Accessories (partial list)
Optional Accessory
Part Number
N9910X-872 External battery charger
N9910X–87617 mm torque wrench
Description
a
a
N9910X-887Replacement Fan Kit
Although not supplied, a USB keyboard and mouse CAN be used with the
FieldFox.
To see a complete list of accessories that are available for the FieldFox, please
visit: http://www.keysight.com/find/fieldfox.
26 Keysight N9938-90006 User’s Guide
Overview
FieldFox Manuals, Software, and Supplemental Help
FieldFox Manuals, Software, and Supplemental Help
The following manuals and software are available for the FieldFox.
To access all of these, visit: www.keysight.com/find/fieldfoxsupport
— User’s Guide –This manual, included with shipment.
— Quick Reference Guide – Printed copy with laminated pages for outdoor
use included with shipment.
— Supplemental Online Help - Concepts and Reference information.
Programming Guide – Online, off line and pdf versions.
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
— FieldFox Data Link Software and Help – Free download. Includes help on
editing calibration kits and cable files. Refer to
http://www.keysight.com/main/software.
— Service Guide – Free download.
— Firmware Updates – Check to see if you have the latest FieldFox firmware.
Conventions that are used in the Manual
— Hardkey indicates a front panel button. The functionality of these buttons
does not change.
The six Softkey menus change dynamically and follow these color
conventions:
— Softkey
— Softkey
— Softkey
— Softkey
— Softkey
Blue indicates an available setting.
Green indicates a change in menu level when selected.
Black indicates the default or selected setting.
Yellow indicates an active entry in process.
Grey indicates a key that is NOT available.
Safety Notes
The following safety notes are used throughout this manual. Familiarize
yourself with each of the notes and its meaning before operating this
instrument. More pertinent safety notes for using this product are located in
Chapter 33, “Safety Considerations”, on page 739.
Denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, would result in damage to or destruction of the
product. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.
Keysight N9938-90006 User’s Guide 27
Overview
FieldFox Manuals, Software, and Supplemental Help
Denotes a hazard. It calls attention to a procedure which, if not correctly
performed or adhered to, could result in injury or loss of life. Do not
proceed beyond a warning note until the indicated conditions are fully
understood and met.
28 Keysight N9938-90006 User’s Guide
Keysight Handheld Analyzers
N99xxB User’s Guide
2 Preparing for Initial Use of Your New FieldFox
Check the Shipment
When you receive your FieldFox, check the shipment according to the
following procedure:
1. Inspect the shipping container for damage. Signs of damage may include
a dented or torn shipping container or cushioning material that indicates
signs of unusual stress or compacting. If not damaged, save the packaging
material in case the FieldFox needs to be returned
2. Carefully remove the contents from the shipping container, and verify that
the standard accessories and your ordered options are included in the
shipment according to the Box Contents List
3. For any question or problems, refer to “Contacting Keysight” on page 5
Meeting Power Requirements for the AC/DC Adapter
Voltage:100 VAC to 240 VAC
Frequency:50 Hz to 60 Hz
Current:1.5 A (100 VAC) to 0.75 A (240 VAC)
The AC/DC adapter supplied with the analyzer is equipped with a three-wire
power cord, in accordance with international safety standards. The power
cable appropriate to the original product shipping location is included with the
FieldFox.
Various AC power cables are available from Keysight that are unique to specific
geographic areas. You can order additional AC power cables that are correct
for use in different areas. For the power cord part number information please
visit: http://www.keysight.com/find/fieldfox
29
Preparing for Initial Use of Your New FieldFox
Install the Lithium-Ion Battery
Install the Lithium-Ion Battery
StepNotes
1. Open the battery doorPush the button on the battery compartment door while sliding the
2. Insert the battery.The terminals end of the battery is inserted into the compartment.
3. Close the battery door.Slide the battery compartment door upwards until it latches.
Battery Usage
When you receive your FieldFox, the lithium-ion battery is not installed, and it
is partially charged to approximately 40% to preserve battery life. A lithium-ion
battery has no memory effect, so it can be used partially charged, as shipped.
A fully charged battery will power your FieldFox for about four hours, so if you
plan to use it for this long, you should fully charge the battery.
The FieldFox will shut down to prevent the battery from discharging to a
level that is damaging. If this occurs, charge the battery either internally or
externally.
Learn more about the lithium-ion battery in Chapter 32, “Working with the
Lithium-Ion Battery.”
Battery charge status is viewable:
door outward.
—In the upper-right corner of the screen.
— On the Battery screen. To access the screen, select System
Diagnostics, and Battery.
30 Keysight N9938-90006 User’s Guide
, Service
Preparing for Initial Use of Your New FieldFox
FieldFox ON/OFF Settings
— On the battery. Open the FieldFox battery compartment door to view the
battery LCD.
To conserve battery power:
—Use Run/Hold to single-trigger a measurement when needed. Hold is
shown on the display.
—Press System
rotary knob, or numeric keypad to adjust the brightness to dim the FieldFox
display as much as possible.
— Briefly press the power button > Standby, to switch to Standby mode when
the FieldFox is not being used.
current settings are preserved.
When powered by the battery only, the FieldFox can stay in Standby mode
for a maximum of four hours and then it powers off automatically. When
the relative battery charge drops by about 20%, the FieldFox will power off
to preserve the remaining charge.
then Display then Brightness. Use the arrows, the
Press power button again to restore power. All
To recharge a battery:
Use ONLY a FieldFox charger to recharge a battery.
— The battery can be fully charged while in the FieldFox in about 4 hours with
the FieldFox either ON or OFF.
— The battery can be fully charged externally using the external battery
charger in about 4 hours.
— When the battery is removed, the FieldFox can still be powered by the
AC/DC adapter.
FieldFox ON/OFF Settings
The FieldFox power button shut down/standby sequence includes a 10
second counter that allows you to either choose a softkey to immediately
initiate the action
counter expire after 10 seconds then perform the action. *Restart can only
be enabled via the Restart softkey.
—To turn power ON, briefly press the power button. Boot-up takes about 1
minute.
—To switch to Standby mode (low battery drain), briefly press the power
button. See the Note above concerning Stand By.
Keysight N9938-90006 User’s Guide 31
(Standby / Shut down / Restart*), or to let the countdown
Preparing for Initial Use of Your New FieldFox
FieldFox High Temperature Protection
—To turn Power OFF (very low battery drain), press the power button briefly
and press the Shutdown softkey. Data and instrument state are NOT
automatically saved when the FieldFox is powered OFF. Learn how to save
data and instrument state in Chapter 31, “File Management”, on page 715.
See the Note above concerning Shut down.
— To restart FieldFox, press the power button and press Restart
above concerning Restart.
—Press Cancel
to exit the power down sequence.
— You can also access the power down softkey menu choices (Standby
down, Restart, and Cancel) by using the Mode hardkey:
—Press Mode
—Then More
—Then Shut down
— You can make a setting to automatically Power ON the FieldFox when a
power source is connected. Learn how on “Power ON” on page 704.
Power Button LED Status
Solid greenPower is ON
Blinking greenFieldFox in Stand By mode
Blinking amber
Blinking amber and greenStand By mode and battery charging
a
. See the Note
, Shut
Battery charging when the FieldFox power is off
Not litPower is Off and battery is not charging
a. LED showing a repeating ramping-up of the intensity of the LED to indicate the
battery is charging.
FieldFox High Temperature Protection
The following features prevent degradation or damage in the event of high
internal temperatures in the FieldFox.
Do NOT store the FieldFox in the soft-case while powered ON or in
Standby mode.
How to monitor the internal FieldFox temperature:
—Press System, then Service Diagnostics.
—Then Internal Temperatures
.
32 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
FieldFox High Temperature Protection
— The average temperature that is displayed for the Internal Temperature is
the average of the RF1, RF2, SB1, SB2 temperatures. These temperatures
come from internal sensors embedded within the FieldFox.
The average temperature does not include the CPU or the battery
temperatures. The FieldFox does not measure the CPU and battery
temperatures.
Temperature Control Mode
The FieldFox’s internal temperature control preserves measurement accuracy
and maintains the long-term reliability of the FieldFox. See also, “High-Temp
Shutdown Warning (RTSA Mode Only)” on page 35.
Measurement speed specifications do NOT apply in Temperature Control
Mode.
Entering Temperature Control Mode
At an average internal temperature of above approximately 64°C, the FieldFox
enters Temperature Control mode by reducing display intensity. This should
decrease the internal temperature which preserves measurement accuracy and
maintains the long-term reliability of the FieldFox. Refer to Figure 2-1.
In addition, duty cycling is enabled when the internal average temperature of
the FieldFox reaches the duty cycling threshold of above ~64°C. All
applications (except RTSA): Power for internal circuitry is turned off between
sweeps and delaying occurs for the next sweep. In RTSA mode, power is only
turned off, if the FieldFox is in Single Sweep or Hold.
All applications (including RTSA): Display dims to 40% at ~64°C.
For all modes, except RTSA mode, you will see the following messages
displayed for ~3 seconds:
"Automatic Meas Duty Cycling activ ated for thermal
management"
For RTSA mode (where duty cycling is not possible), you will see the following
messages displayed for ~3 seconds:
"Entering Auto Protect mo de for thermal management"
Later if the temperatures are reduced:
"Temperatures reduced, exitin g Automatic Protect Mode "
Keysight N9938-90006 User’s Guide 33
Preparing for Initial Use of Your New FieldFox
FieldFox High Temperature Protection
Entering Auto Protect (Maximum) Threshold Temperature Mode
At above approximately 75°C, the FieldFox enters (Maximum) Auto Protect
temperature control mode by displaying the "Entering Auto Protect mode...".
See also Table 2-1 on page 36.
When you enter the Auto Protect (Maximum) Threshold, the following message
is displayed:
"Entering Maximum Auto Protect mod e for thermal managemen t"
When temperature reaches the maximum protection threshold (~75°C), some
applications take additional measures.
Figure 2-1 Entering Auto Protect mode message (above ~64°C)—(Displays for 3 sec)
Exiting Maximum Auto Protect Threshold Temperature Mode
When the temperature drops to approximately 73.5°C, a message is displayed
indicating that the FieldFox is leaving Temperature Control Mode and normal
operating settings are restored (refer to Figure 2-2). See also, Table 2-1 on
page 36.
When you exit the Auto Protect (Maximum) Threshold control mode, the
following message is displayed:
"Temperatures reduced, exitin g Maximum Auto Protect mod e"
Figure 2-2 Exiting Maximum Auto Protect mode message—(~62.5°C)
34 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
FieldFox High Temperature Protection
High-Temp Shutdown Warning (RTSA Mode Only)
In extreme situations, Temperature Control mode may not stop an increase in
the FieldFox internal temperature. Just prior to shut down, the FieldFox will
display a warning of imminent shut down and the instrument turns off. At
above approximately 77°C, a Maximum Auto Protect message—(Top
"WARNING..." message) displays for 9 sec. Red text at left bottom corner of
display remains on until internal temperature returns to ~75.5°C. At above
~78°C, High-Temperature Shut down will engage and turn OFF the FieldFox.
Refer to Figure 2-3. See also, Table 2-1 on page 36.
FieldFox reaches the duty cycling threshold of ~64°C. Power is turned
off between sweeps and delaying occurs for the next sweep.
In RTSA mode power is only turned off in Single Sweep or Hold. This
enables you to control your instrument’s temperature when you don’t
need to make a measurement by using the "Run/Hold" hardkey to
pause the sweep.
All applications (including RTSA): Display dims to 40% at ~64°C.
(~75°C) some applications take additional measures.
displays for 9 sec). Red text at bottom of display remains on until
internal temperature returns to ~75.5°C. Refer to Figure 2-3 on
page 35.
Avoid Overpowering the FieldFox
The FieldFox can be damaged with too much power or voltage applied.
Exceeding the maximum RF power levels shown below will cause an ADC Over
Range message to appear on the screen.
Maximum Input Voltages and Power:
— RF IN/OUT Connectors: ±50 VDC, +27 dBm RF
— DC Input: -19 to 19 VDC, 60 Watts maximum when charging battery
Learn more about Maximum power and voltages in the FieldFox Data
Sheet.
Very often, coaxial cables and antennas build up a static charge, which, if
allowed to discharge by connecting to the FieldFox, may damage the
instrument input circuitry. To avoid such damage, it is recommended to
dissipate any static charges by temporarily attaching a short to the cable
or antenna prior to attaching to the FieldFox.
36 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
Take the FieldFox Tour
Front Panel
Keysight N9938-90006 User’s Guide 37
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
No.CaptionDescriptionLearn
More:
1PowerON: press momentarily.
STAND BY: with FieldFox power ON, press briefly.
OFF: press and hold until the FieldFox shuts off (about 4 seconds).
2LEDNot lit: FieldFox OFF, not charging
Green: FieldFox ON. Charging status indicated by battery icon on screen
Displays a sub-menu for marker functions“All about
Markers”
on
page 638
Exits and closes the Active Entry dialog box or clears the character input--
Saves the current trace or recalls saved data from memory“Saving
and
Recalling
Files” on
page 716
38 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
No.CaptionDescriptionLearn
More:
11
Limit
12
Run/Hold
13
Cal
14Arrow keysIncreases or decreases a value or setting.--
15Returns to the previous menu selection. --
16Rotary knobHighlights an item for selection, or enables incremental changes to values. --
17SoftkeysAllows selection of settings for configuring and performing measurements, and
Sets limit lines for quick Pass/Fail judgment“All about
Limit
Lines” on
page 653
Toggles between free Run and Hold/Single operation.“Run/Hol
d” on
page 664
Displays a sub-menu for calibration functions“Calibrati
on
Method
Summary
” on
page 142
--
for other FieldFox functions.
18ScreenTransflective screen, viewable under all lighting conditions. If you are using your
FieldFox in direct sunlight, you do not need to shield the display from the
sunlight. In bright lighting conditions, the display is brighter and easier to read
when you allow light to fall directly on the screen. Alternative color modes exist
that maximize viewing in direct sunlight conditions, as well as other conditions
such as nighttime work.
Clean the Transflective screen with gentle and minimal
wiping using Isopropyl alcohol applied to a lint-free cloth.
“Screen
Tour” on
page 43
Keysight N9938-90006 User’s Guide 39
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
Top Panel
CaptionDescriptionsLearn More:
Port 1
RF Output
Port 2
SA RF Input
GPS AntSMA (f) connector for use with built-in GPS. Produces a settable 3.3 VDC or
Ref In
Trig In
For CAT and NA measurements, use to make reflection measurements.
Maximum: ±50 VDC, +27 dBm RF
Also, for SA source output in SA mode.
For SA, use to make all measurements.
For CAT, NA, and VVM mode, use to make Port 2 transmission
measurements.
Maximum: ±50 VDC, +27 dBm RF.
5 VDC bias voltage for the antenna pre-amplifier. Use with a GPS antenna
such as N9910X-825. Other GPS antennas can also be used.
SMA (f) connector for use with Frequency Reference Source and External
Trigger Input signal.
Maximum: 5.5 VDC.
“CAT Mode Settings”
on page 49
“NA Mode Settings” on
page 89
Chapter 9, “SA
(Spectrum Analyzer)
Mode (Option
233–Mixed Analyzers)”,
on page 143
“SA Mode Settings” on
page 146
“GPS” on page 684
“Frequency Reference
Source” on page 698.
“Triggering” on
page 172
40 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
Right Side Panel
ConnectorDescriptionLearn More:
Ethernet cable connector to read trace data using the FieldFox Data Link
Software and connect to the FieldFox remotely.
Download the latest version of the software at:
www.keysight.com/find/fieldfoxsupport
IF Out used in SA mode for external signal processing.“IF Output” on
Frequency Reference Source Output
Trigger Output (ERTA mode).
Secure Digital slot. Use to extend the memory of the FieldFox.“Saving and Recalling
The mini-USB port can be connected to your PC’s standard USB port to
send SCPI commands.
IMPORTANT! It is recommended that only USB
certified cables are used with the FieldFox.
“LAN Settings” on
page 702
page 166
Chapter 24, “ERTA
(Extended Range
Transmission Analysis)
Mode - Option 209”, on
page 535
Files” on page 716
Chapter 30, “Using the
Mini-USB Port to send
SCPI Commands and
Queries”, on page 713
Two standard USB connectors used to connect a power sensor for Power
Meter Mode. Also used to save files to a USB flash drive.
“Saving and Recalling
Files” on page 716
IMPORTANT! It is recommended that only USB
certified cables are used with the FieldFox.
Keysight N9938-90006 User’s Guide 41
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
Left Side Panel
ConnectorDescriptionLearn More:
Audio output jack for use with SA Mode Tune and Listen.“Tune & Listen
(AM/FM)” on page 160
DC Voltage Source for use with external DC Bias.“Variable Voltage
Source (Opt. 309)” on
page 699
DC power connector used to connect to the AC/DC adapter. Maximum: 19
VDC, 4 ADC.
“Internal Charging with
the AC/DC Adapter ” on
page 733
42 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
Screen Tour
CaptionDescriptionLearn More:
1Title – write your own text here“Title” on page 668
“All about Trace Math”
on page 658
“Resolution Bandwidth
(Res BW)” on page 164
10Resolution SettingMode dependent
11Measurement Start Freq or DistanceMode dependent
Keysight N9938-90006 User’s Guide 43
Preparing for Initial Use of Your New FieldFox
Take the FieldFox Tour
CaptionDescriptionLearn More:
12Bandpass / Lowpass setting (Fault Meas)
IF BW in NA Mode
13Output Power Level (CAT and NA)“Output Power” on
14Measurement Stop Freq or DistanceMode dependent
15Actual Sweep TimeMode dependent
16Limit Line Status“All about Limit Lines”
17Time and Date“Date and Time” on
18Marker Readout“All about Markers” on
19Battery Status“Viewing the Battery
20Measurement Type (CAT and NA)--
21acquisitionMode dependent
22Reference PositionMode dependent
“DTF Measurement
Settings” on page 64
page 55
on page 653
page 684
page 638
Charge Status” on
page 731
44 Keysight N9938-90006 User’s Guide
Preparing for Initial Use of Your New FieldFox
How to Enter Numeric Values
How to Enter Numeric Values
Many settings on the FieldFox require the entry of numeric values.
How to enter numeric values
Use any combination of the following keys:
— Numeric 0–9 keys, along with the polarity (+/-
— Up/Down arrow keys to increment or decrement values.
— Rotary knob to scroll through a set of values.
— Back
— Esc
To complete the setting, press Enter
erases previously entered values.
exits data entry without accepting the new value.
Multiplier Abbreviations
Many times after entering numeric values, a set of multiplier or suffix softkeys
are presented. The following explains the meaning of these abbreviations.
Select Frequency multipliers as follows:
— GHz
— MHz
— KHz
— Hz
Select Time multipliers as follows:
Gigahertz (1e9 Hertz)
Megahertz (1e6 Hertz)
Kilohertz (1e3 Hertz)
Hertz
) key.
or a different softkey or hardkey.
Seconds
— s
— ms
— us
— ns
— ps
Keysight N9938-90006 User’s Guide 45
milliseconds (1e–3)
microseconds (1e–6)
nanoseconds (1e–9)
picoseconds (1e-12)
Preparing for Initial Use of Your New FieldFox
Connector Care
Connector Care
— Never store connectors, airlines, or calibration standards loose in a box. This
— Install protective end caps when connectors are not in use.
— Keep connector temperature the same as the test instrument. Holding the
— Do not touch the mating plane surfaces. Natural skin oils and microscopic
— Do not set connectors contact-end down on a hard surface. The plating and
— Wear a grounded wrist strap and work on a grounded, conductive table
is a common cause of connector damage.
connector in your hand or cleaning connector with compressed air can
significantly change the temperature. Wait for connector temperature to
stabilize before using in calibration or measurements.
particles of dirt are difficult to remove from these surfaces.
mating plane surfaces can be damaged if the interface comes in contact
with any hard surface.
mat. This helps protect the analyzer and devices from electrostatic
discharge (ESD).
— Refer to “Connector Care Review” on page 761.
Because of the very small and precise mechanical tolerances of mmWave
connectors, minor defects, damage, and dirt can significantly degrade
repeatability and accuracy. In addition, a dirty or damaged connector can
destroy connectors that are mated to it. For this reason, NEVER use a
damaged connector. See also
http://na.support.keysight.com/pna/connectorcare/Connector_Care.htm
46 Keysight N9938-90006 User’s Guide
Keysight Handheld Analyzers
N99xxB User’s Guide
3CAT (Cable and Antenna Test) Mode
CAT Mode is typically used to test an entire
transmission system, from the transmitter to the
antenna. This process is sometimes referred to as
Line Sweeping.
CAT Mode is similar to NA (Network Analyzer)
Mode. Learn more in the Supplemental Online Help:
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
CAT Mode Settings
Select CAT Mode before making any setting in this chapter.
How to select CAT Mode
—Press Mode.
—Then CAT/TDR.
Measurement Selection
How to select a CAT Mode Measurement
Learn more about the following measurements in the Supplemental Online
Help: http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
Press Measure 1
Then choose one of the following: These softkeys also appear after CAT Mode
is selected.
— Distance To Fault
Transform (IFT) calculations to determine and display the distance to, and
relative size of, a fault or disruption in the transmission line. Units are in
return loss format, expressed as a positive number in dB, unless the
measurement selected is DTF (VSWR). Learn more about DTF
Measurements refer to Chapter 4, “DTF (Distance to Fault)
Measurements.”.
— Return Loss & DTF
measurement. Use this format to display the frequency settings that are
used to make the DTF measurement. The frequency range settings for these
two measurements can be coupled or uncoupled. Learn more on page 56.
Calibrations are applied to both traces
When in Hold mode and Single sweep is performed, only the active
trace is triggered. Use the arrows to activate a trace
— Return Loss
incident signal energy MINUS the amount of energy that is reflected. The
higher the trace is on the screen, the more energy being reflected back to
the FieldFox. Learn how to measure Return Loss on page 58.
.
1-port reflection measurement that uses Inverse Fourier
Displays both a Return Loss measurement and a DTF
1-port reflection measurement that displays the amount of
— VSWR
reflection measurement that displays the ratio of the maximum reflected
voltage over the minimum reflected voltage. The higher the trace is on the
screen, the more energy being reflected back to the FieldFox.
— DTF (VSWR)
Keysight N9938-90006 User’s Guide 49
(Voltage Standing Wave Ratio – also known as SWR) 1-port
Distance to Fault in VSWR format.
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
— Cable Loss (1-Port) 1-port reflection measurement that displays the loss of
a transmission line. Learn more on page 58
— Insertion Loss (2-Port)
displays the loss through a cable or other device in dB. Both ends of the
cable must be connected to the FieldFox. NO phase information is included
in this measurement. Learn more at “2-Port Insertion Loss Measurements”
on page 60. This feature is available only with an option on some FieldFox
models. For detailed information, please view the FieldFox Configuration
Guide at: http://literature.cdn.keysight.com/litweb/pdf/5992-3701EN.pdf
— DTF (Lin)
— TDR (Lin rho)
without reflections shows as 0 (zero). Maximum reflections from an open or
short show as 1.
— TDR (ohm)
— TDR & DTF
Distance to Fault in Linear format.
The Y-axis of the display is impedance (ohms).
TDR and DTF are displayed.
Quick Settings Table
Both CAT and NA Modes allow you to view and change most relevant settings
from a single location. All of these settings are discussed in this chapter and,
unless otherwise noted, ALL of these settings can also be made using the
standard softkey menus.
2-port transmission measurement that accurately
The Y-axis of the display is linear, real, unitless values. A trace
How to view and change Quick Settings
—Press Meas Setup 4.
—Then Settings
—Press Next Page
are NOT available, then all available settings fit on one page.
— To change a setting:
— Use the arrows to highlight a setting.
—Then press Edit
— Some settings require you to press a softkey to change the value.
Otherwise, use the numeric keypad, arrows, or rotary knob to
change the value.
— When finished changing a value, press Done Edit
—Press Dock Window
the trace window. The Dock Window setting persists through a Preset.
Choose from the following
— Full (Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown
.
and Previous Page to view all settings. If these softkeys
. The current setting changes to yellow.
.
to relocate the Settings table to a position relative to
50 Keysight N9938-90006 User’s Guide
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
— Left The Settings table is shown to the left of the trace window
— Bottom The Settings table is shown below the trace window
— When finished changing ALL settings, press Done
Frequency Range
Set the range of frequencies over which you would like to make CAT Mode
measurements
When the frequency range is changed after a calibration is performed, the cal
becomes interpolated. Learn more in “Interpolation *” on page 139.
How to set Frequency Range
—Press Freq/Dist.
— Then choose from the following
— Start
— Center
frequencies (half on either side of center).
— Follow each setting by entering a value using the numeric keypad,
arrows, or the rotary knob.
— After using the keypad, select a multiplier key. Learn about multiplier
abbreviations in Chapter 1, “Overview.”
to save your settings
and Stop frequencies - beginning and end of the sweep.
and Span frequencies – the center frequency and span of
— After using the arrows or the rotary knob, press Enter
Scale Settings
Adjust the Y-axis scale to see the relevant portions of the data trace. The Y-axis
is divided into 10 graticules.
This setting can be changed at any time without affecting calibration accuracy.
How to set Scale
—Press Scale / Amptd.
— Then choose from the following three methods:
1. Autoscale
2. Set Scale, acquisition, and Reference Position:
. The
amount of frequency increment is based on the current span and can
NOT be changed in CAT Mode.
Automatically adjusts the Y-axis to comfortably fit the
Minimum and Maximum amplitude of the trace on the screen.
— Scale
Manually enter a scale per division to view specific
areas of the trace
Keysight N9938-90006 User’s Guide 51
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
— Ref Level Manually set the value of the reference line.
Enter a negative value by pressing Run/Hold (+/-)
before or after typing a value
either
Averaging
— Ref Pos
Manually set the position of the reference line.
Values must be between 0 (TOP line) and 10 (BOTTOM
line)
3. Set Top and Bottom graticule values. The scale per division is
calculated.
— Top
— Bottom
to set the value of the Top graticule.
to set the value of the Bottom graticule.
— Enter a negative value by pressing Run/Hold (+/-)
before or after typing a value.
Scale annotation on the FieldFox screen
Reference Line = red arrow
Ref Level = -40 dB
Ref Position = 1
Scale = 2 dB per division
either
Trace Averaging helps to smooth a trace to reduce the effects of random noise
on a measurement. The FieldFox computes each data point based on the
average of the same data point over several consecutive sweeps.
Average Count determines the number of sweeps to average; the higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as Avg N. This
shows the number of previous sweeps that have been averaged together to
form the current trace. When the counter reaches the specified count, then a
‘running average’ of the last N sweeps is displayed. Average Count = 1 means
there is NO averaging.
This setting can be changed at any time without affecting calibration accuracy.
Averaging is often used to increase the dynamic range of a measurement.
To achieve the highest dynamic range, select NA mode and reduce the IF
Bandwidth setting. Learn more about dynamic range in
Range” on page 110
.
“Increase Dynamic
How to set Trace Averaging
—Press BW 2.
52 Keysight N9938-90006 User’s Guide
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
—Then Average N where N is the current count setting.
— Enter a value using the numeric keypad. Enter 1 for NO averaging.
—Press Enter
— While Trace Averaging is in process, press Sweep 3
Smoothing
Trace smoothing averages a number of adjacent data points to smooth the
peak-to-peak noise values on a displayed trace. The number of adjacent data
points that are averaged is known as the smoothing aperture. Aperture is set
by specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing can be used with any CAT format.
When enabled, Smo appears on the FieldFox screen.
How to set Smoothing
—Press BW 2
—Then Smoothing ON OFF
—Then Sm. Aperture
.
then Restart to restart
the averaging at 1.
.
.
and enter a value between 0 and 25 (percent) using the
numeric keypad.
—Press Enter
.
Single or Continuous Measure
This setting determines whether the FieldFox sweeps continuously or only once
each time the Single
or to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
—Press Sweep 3.
— Then choose one of the following:
— Single
make ONE sweep, then hold for the next Single key press. Hold is
annotated in the upper left corner of the display when NOT sweeping,
and changes to an arrow --> while the sweep occurs
— Continuous
when battery power is not critical.
You can also use Run / Hold +/-
button is pressed. Use Single to conserve battery power
Automatically sets Continuous OFF and causes FieldFox to
Makes continuous sweeps. This is the typical setting
to toggle between Single and Continuous.
Keysight N9938-90006 User’s Guide 53
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the
X-axis to form a trace. Select more data points to increase measurement
resolution. However, more data points require more time to complete an entire
measurement sweep
When the Resolution is changed after a calibration is performed, the cal
becomes interpolated. Learn more in “Interpolation *” on page 139
— Using SCPI, Resolution can be set to any number of points between 3 and
10001. See the Programming Guide at
http://na.support.keysight.com/fieldfox/help/
Sweep Time
The fastest possible sweep time is always used as the default setting. Use the
Min Swp Time setting to slow the sweep time when measuring long lengths of
cable.
The actual sweep time is shown on the FieldFox screen. See “Take the FieldFox
Tour” on page 37. To increase the sweep time, enter a value that is higher than
the actual sweep time. The increase will not be exactly the amount that you
enter, as the actual sweep time is the composite of many factors.
Measurement speed specifications do NOT apply in Temperature Control
Mode. Learn more in Chapter 1, “Overview.”
.
.
How to set Sweep Time
—Press Sweep 3.
—Then Min Swp Time
— Enter a value using the numeric keypad.
— Press a multiplier key. Learn more in “Multiplier Abbreviations” on page 45.
54 Keysight N9938-90006 User’s Guide
.
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Output Power
Set the power level out of the FieldFox to High, Low, or manually set power
level to a value between High and Low.
Generally, the high power setting is used when measuring passive, high-loss
devices to place the signal farther from the noise floor. Increasing the power
can also help reduce trace noise. However, for devices that are sensitive to
high power levels such as amplifiers, use the Low power setting.
For best measurement accuracy, use the Manual power setting at -15 dBm.
After calibration, the power level can be decreased for amplifiers, or increased
for higher dynamic range.
Power Level settings in this mode will NOT change Power Level settings in
other modes. To help prevent damage to your DUT, use caution when
changing modes with your DUT connected to the FieldFox test ports.
How to set Output Power
—Press Meas Setup 4.
—Then Power
—Then Output Power
— High Sets output power to the maximum achievable power at all
displayed frequencies. Output power is NOT FLAT across the
displayed FieldFox frequency span. Use the High power setting, when
better dynamic range and lower noise is desired. Please see
“Specifications/Data Sheet” on page 763 for expected power levels.
— Low Sets output power to approximately –50dBm, FLAT across the
displayed FieldFox frequency span.
— Man (Default setting) Sets the output power to –15 dBm, FLAT across
the displayed FieldFox frequency span. If flattened power can NOT be
achieved, a warning message and beep occurs. To achieve a
flattened output power, reduce the power level or stop frequency.
—Then press Power Level
— Then enter a value using the numeric keypad, the arrows, or the
rotary knob.
—Press Enter
.
Keysight N9938-90006 User’s Guide 55
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Interference Rejection
Use this setting when you suspect that other signals in the area are interfering
with a measurement. Interference may look like a spike or lack of stability in the
measurement trace. While monitoring a measurement at a specific frequency,
toggle this setting between ON and OFF. If the measurement result decreases
while ON, then there is an interfering signal in the area. Continue to make
measurements with Interference Rejection ON. However, this will slow the
measurement speed.
Once enabled, up to SIX sweeps may be required before the interfering signal
is neutralized.
This setting can be changed at any time without affecting calibration accuracy.
How to set Interference Rejection
—Press Meas Setup 4.
—Then Interference Rejection [current setting]
— Then choose from the following:
No interference rejection and fastest possible sweep speed
— Off
— Minimum
— Medium
— Maximum
Coupled Frequency
This setting, especially useful for a Return Loss & DTF measurement, allows
both measurements to have different frequency ranges.
How to set Coupled Frequency
With a Return Loss & DTF measurement present:
—Press Meas Setup 4
— Select the DTF measurement (Tr2) using the arrows.
— Then choose from the following:
.
The lowest level of Interference rejection.
The medium level of Interference rejection.
The highest level of Interference rejection.
.
— Coupled Freq ON
frequency range settings.
— Coupled Freq OFF
individual frequency range settings. When set to OFF:
— The Return Loss measurement frequency settings are made in the
usual manner. Learn how at “Frequency Range” on page 51. When a
new Start or Stop frequency is selected, Coupled Frequency is
automatically set to OFF.
56 Keysight N9938-90006 User’s Guide
- Both the Return Loss and DTF traces have the same
(default setting) - Both traces are allowed to have
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
— The DTF measurement is made using the frequencies as determined
by the DTF Frequency Mode setting. Learn more in “DTF
Measurement Settings” on page 64.
Keysight N9938-90006 User’s Guide 57
CAT (Cable and Antenna Test) Mode
Return Loss Measurements
Return Loss Measurements
Return loss can be thought of as the absolute value of the reflected power
compared to the incident power.
When measuring an OPEN or SHORT, all incident power is reflected and
approximately 0 dB return loss is displayed.
When measuring a LOAD, very little power is reflected and values of 40 dB to
60 dB are displayed.
The minus sign is usually ignored when conveying return loss. For example, a
component is said to have 18 dB return loss, rather than –18 dB.
How to Measure Return Loss
— Connect the jumper cable or any adapter used to connect the device under
test (DUT).
— Select Preset
— Select Mode
—Then Return Loss
—Press Freq/Dist
measurement.
—Press Meas Setup 4
calibrating.
— Disconnect the jumper cable or DUT and press Cal 5
calibration prompts.
— Reconnect the jumper cable or DUT.
— The return loss trace is displayed on the FieldFox screen.
then Preset Returns the FieldFox to known settings
then CAT/TDR (Cable and Antenna Test)
(Default measurement).
and enter Start and Stop frequency values of the
then Settings to make appropriate settings before
1-Port Cable Loss Measurements
While all cables have inherent loss, weather and time will deteriorate cables
and cause even more energy to be absorbed by the cable. This makes less
power available to be transmitted.
then follow the
A deteriorated cable is not usually apparent in a Distance to Fault
measurement, where more obvious and dramatic problems are identified. A
Cable Loss measurement is necessary to measure the accumulated losses
throughout the length of the cable.
58 Keysight N9938-90006 User’s Guide
CAT (Cable and Antenna Test) Mode
1-Port Cable Loss Measurements
A 2-port Insertion Loss measurement is usually more accurate than a 1-port
Cable Loss measurement. However, to perform a 2-port Insertion Loss
measurement, both ends of the cable must be connected to the FieldFox.
In high-loss conditions, a Cable Loss measurement becomes ‘noisy’ as the
test signal becomes indistinguishable in the FieldFox noise floor. This can
occur when measuring a very long cable and using relatively high
measurement frequencies. To help with this condition, use High Power
(page 55) and Averaging. (page 52).
How to make a 1-port Cable Loss Measurement
1. Press Preset then Preset.
2. Then More
then Cable Loss (1-Port)
3. Connect the cable to be tested.
4. Press Freq/Dist
and enter Start and Stop frequency values of the
measurement.
5. Press Sweep 3
then Min Swp Time. Increase the Sweep Time until a stable
trace is visible on the screen. The amount of time that is required increases
with longer cable lengths. Learn more in the Supplemental Online Help:
8. Follow the prompts to perform calibration at the end of the jumper cable
or adapter. Learn more about Calibration in “How to Perform a
Calibration” on page 125
9. Connect the cable to be tested.
Low-level standing waves (also known as ‘ripple’) which may be visible in
reflection measurements, can hide the actual loss of the cable. Steps 10
through 13 can minimize the ripple. Perform the measurement with and
without steps 10 through 13 and choose the method with the least amount
of ripple.
10.Connect a LOAD at the end of the cable to be tested. This limits the
reflections to faults that are located in the cable under test.
11.Press Trace 6
then Math and Memory then Data->Mem to store the trace
into Memory.
12.Remove the LOAD and leave the end of the cable to be tested open.
13.Press Data Math
then Data – Mem. The ripple in the measurement is
removed. These minor imperfections in the cable should not be considered
in the Cable Loss measurement.
Keysight N9938-90006 User’s Guide 59
CAT (Cable and Antenna Test) Mode
2-Port Insertion Loss Measurements
14.Use Averaging to remove random noise from high-loss measurements.
Press BW 2
The displayed trace shows the Cable Loss values in one direction through the
cable. A Return Loss measurement would show the loss for both down the
cable and back. Therefore, a Cable Loss measurement is the same as a Return
Loss measurement divided by 2.
The average Cable Loss across the specified frequency range is shown on the
screen below the graticules.
then Average.
2-Port Insertion Loss Measurements
A 2-port Insertion Loss measurement is used to measure the loss through a
DUT (device under test) – or cable – over a specified frequency range. The
FieldFox signal source is transmitted out the RF OUT connector, through the
DUT, and into the RF IN connector. Both ends of the DUT must be connected
to the FieldFox, either directly or indirectly using the cable used in the
normalization cal.
‘Insertion’ loss simply means loss through a device, usually expressed in dB. It
is exactly the same measurement as “S21 Transmission” in NA Mode.
Exception: In CAT mode, if the TDR data are saved in S1P format, the
values represent the real part of the complex transform; the imaginary part
is set to zero. To obtain complex data, either as real/imaginary or
magnitude/phase pairs, use NA mode.
2-port Insertion Loss measurements are generally more accurate than 1-port
Cable Loss measurements.
How to make a 2-port Insertion Loss Measurement
—Press Mode
—Then More
—Press Freq/Dist
measurement.
—Press Sweep 3
—Press Cal 5
Calibration” on page 125.
— Connect the DUT and view the insertion loss measurement results.
then CAT.
then Insertion Loss (2-Port).
and enter Start and Stop frequency values of the
, then select a Resolution setting.
, then perform a calibration. Learn more on “How to Perform a
When measuring very long lengths of cable, it may be necessary to increase
the sweep time. Learn how on page 54. Learn why in the Supplemental Online Help: http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
60 Keysight N9938-90006 User’s Guide
Keysight Handheld Analyzers
N99xxB User’s Guide
4DTF (Distance to Fault) Measurements
CAT Mode Distance to Fault (DTF) measurements are generally used to locate
problems, or faults, in a length of cable or transmission line. In this chapter, the
cable to be tested is referred to as the DUT (Device Under Test).
Settings that are NOT unique to DTF measurements are documented in the
CAT Mode Chapter 3, “CAT (Cable and Antenna Test) Mode.”
In this Chapter
“How to make DTF Measurements” on page 62
“DTF Settings Table” on page 64
“DTF Measurement (Format)” on page 64
“DTF Start and Stop Distance” on page 65
“Frequency Mode” on page 65
61
DTF (Distance to Fault) Measurements
How to make DTF Measurements
“Coupled Frequency” on page 66
“Cable (Correction) Specifications” on page 67
“Window Settings” on page 70
“DTF Units” on page 71
“Calculated DTF values” on page 71
“About Alias Faults” on page 71
Optional settings
“All about Markers” on page 638
“All about Limit Lines” on page 653
“Saving and Recalling Files” on page 716
Trace Math is NOT available in DTF Measurements.
How to make DTF Measurements
Before starting, you may need the following:
— Jumper cable or adapter to connect the beginning of the DUT to the
FieldFox.
— LOAD with correct connector type and gender to terminate the end of the
DUT (if possible).
— The known length and cable type of the DUT. If the cable type is not known,
then the Cable Loss (dB/Meter) and Velocity Factor of the DUT are required.
1. Connect any necessary jumper cable or adapter to the FieldFox RF OUT
port. Do NOT connect the DUT.
2. Press Preset
3. Then Mode
4. Then Meas 1
5. Press Freq/Dist
can optionally set the Start Distance
6. Press Cal 5
“How to Perform a Calibration” on page 125.
then Preset to return the FieldFox to the default settings.
then CAT.
then DTF.
, then Stop Distance and enter the length of the DUT. You
and follow the Cal prompts. Learn all about Calibration in
7. Disconnect any components or antenna that should NOT be measured and
connect a LOAD at the end of the DUT.
8. Press Meas Setup 4
62 Keysight N9938-90006 User’s Guide
then DTF Cable Specifications.
DTF (Distance to Fault) Measurements
How to make DTF Measurements
9. Either press Edit/Save/Recall Cables then Recall Cable, or directly enter
the Velocity Factor/Cable Loss of the DUT using the Velocity Facto
Cable Loss
keys.
r and
10. Connect the start end of the DUT to the FieldFox.
11. Press Meas Setup 4
then Settings then Next Page. If the Alias-free Range
setting is False, then you may see Alias faults on the screen. Learn more
on page 71.
Keysight N9938-90006 User’s Guide 63
DTF (Distance to Fault) Measurements
DTF Measurement Settings
DTF Measurement Settings
DTF Settings Table
You can set and view all of the DTF settings, including some calculated values,
on the DTF Settings table. Learn about the calculated values in “Calculated
DTF values” on page 71.
How to make settings on the DTF settings table
—Press Meas Setup 4.
—Then Settings
—Press Next Page
— To change a setting:
— Use the arrows or rotary knob to highlight a setting.
— Numeric settings can be changed by pressing numbers using the
numeric keypad. Then press Enter or select a suffix if available.
— Other settings require you to press Edit
change the value.
— When finished changing a value, press Done Edit
—Press Dock Window
the trace window. The setting persists through a Preset. Choose from the
following:
— Full (Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
— Left The Settings table is shown to the left of the trace window.
— Bottom The Settings table is shown below the trace window.
.
and Previous Page to view all settings.
, then press a softkey to
.
to relocate the Settings table to a position relative to
— When finished changing ALL settings, press Done
to save your settings.
DTF Measurement (Format)
You can select from 3 different DTF Formats.
—Press Measure 1
— Then choose from:
— Distance to Fault (dB)
loss format, expressed as a positive number in dB.
— DTF (VSWR)
about SWR at the FieldFox Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.ht
m
64 Keysight N9938-90006 User’s Guide
Faults are displayed on the Y-axis in return
Faults are displayed on the Y-axis in SWR. Learn more
DTF (Distance to Fault) Measurements
DTF Measurement Settings
— More then DTF (Lin) Faults are displayed on the Y-axis in linear
(unitless) format.
DTF Start and Stop Distance
In DTF measurements, you set the physical length of cable or other device to
be tested. The FieldFox calculates the frequency range of the measurement
from this distance. The longer the cable to be tested, the lower the frequencies
that are used. You can also set the frequencies manually using the Frequency
Mode [Bandpass] setting.
How to set Start and Stop Distance
— With a DTF measurement present, press Freq/Dist.
— Then choose from the following:
— Start Distance
— Stop Distance
Frequency Mode
All DTF measurements are made with frequency settings and, using Inverse
Fourier Transform (IFT), the time and distance to faults are calculated.
The start and stop frequencies for the measurement are always annotated
on the screen below the start and stop distances.
How to set Frequency Mode
With a DTF measurement present,
—Press Meas Setup 4
Enter a value using the numeric keypad, the
arrows, or the rotary knob, then Enter
is set to 0 Meters. This means that the measurement will display
faults starting at the point at which calibration standards are
connected.
Enter a value between the start distance and 5 km (or
16,404 ft.) using the numeric keypad, the arrows, or the rotary
knob, then Enter
.
. By default, the Start Distance
—Then Frequency Mode
— Choose one of the following:
— Lowpass
automatically based on the Start and Stop Distances. Use Lowpass
mode when the DUT is a cable ONLY.
Keysight N9938-90006 User’s Guide 65
The frequency range of a DTF measurement is set
DTF (Distance to Fault) Measurements
DTF Measurement Settings
— Bandpass (Default setting) The frequency range of a DTF
measurement is set manually. Use Bandpass mode when the DUT
contains a diplexer or other filtering device which does not pass
some frequencies.
Typically, you will set the frequency range of the measurement to the passband
of the filter. However, you may also want to test the ability of the filter to reject
unwanted frequencies. In this case, set the frequency range to include those
frequencies which the filter may not be adequately rejecting.
How to manually set Frequencies in Bandpass Mode
—Press Freq/Dist
—Then Min Start Freq
measurement.
—Then Max Stop Freq
measurement.
—OR
—Press More
—Then Max Freq Span and type the frequency range to use for the DTF
measurement.
—Then Center Frequency
for the DTF measurement.
When the DTF frequencies are set manually, they may not be the optimum
frequencies for measuring the distance to fault. The distance may no longer be
alias-free. Learn more about alias-free range on page 71.
These settings specify the minimum and maximum frequencies to be used for
the DTF measurement. These exact frequencies may not be used, but a
narrower frequency range may be used that will still pass through the
bandpass filter.
Coupled Frequency
and type the start frequency to use for the DTF
and type the stop frequency to use for the DTF
and type the center frequency of the range to use
When a Return Loss & DTF measurement is present, this setting allows you to
have different frequency ranges for each measurement. Learn more in
“Coupled Frequency” on page 56.
66 Keysight N9938-90006 User’s Guide
DTF (Distance to Fault) Measurements
DTF Measurement Settings
Cable (Correction) Specifications
By default, the FieldFox does NOT correct DTF measurements to account for
the inherent loss of a cable. However, to make more accurate DTF
measurements, the Cable Loss and Velocity Factor values should be
considered.
The following describes Coax Media type. However, you can create or edit
Waveguide Media Standards. Learn how in “Waveguide Calibrations” on
page 137
About Velocity Factor and Cable Loss
— Velocity Factor is a property of the physical material of a cable. A VF of 1.0
corresponds to the speed of light in a vacuum, or the fastest VF possible. A
polyethylene dielectric cable has VF = 0.66 and a cable with PTFE dielectric
has VF = 0.7.
— Cable Loss is specified in dB/meter. In addition to the length of the cable,
loss is also directly proportional to the frequency of the signal that passes
through the cable.
.
The following is an example showing how DTF cable correction works:
The DUT is a 100 meter transmission cable. The Cable Loss value is
0.1 dB/meter. This means that a signal traveling ONE WAY through the cable
will lose 10 dB of power (100 m * 1dB/m). Because the FieldFox performs this
measurement with 1 port, the test signal travels down the cable and then back,
for a total loss of 20 dB.
After a calibration has been performed, for the purpose of illustrating this
point, connect an OPEN to the end of the cable – a maximum-sized fault - for
100% reflection of the 300 MHz test signal.
Without compensation for the loss of the cable, a –20 dB response would be
visible at 100 meters, which is the OPEN at the end of the DUT. This is from
10 dB of loss through the cable in each direction.
With compensation for the loss using the manufacturer’s specification, the
FieldFox compensates the trace as though the signal traveling through 100
meters was increased by +20 dB. Therefore the response will show 0 dB for
100% reflection.
How to enter Cable Loss and Velocity Factor
Cable Loss and Velocity factor can be entered using one of the following
methods:
— Manually enter cable loss and velocity factor for the measurement.
— Select or create a cable file which contains the cable loss and velocity
factor.
With a DTF measurement present:
Keysight N9938-90006 User’s Guide 67
DTF (Distance to Fault) Measurements
DTF Measurement Settings
—Press Meas Setup 4.
—Then DTF Cable Specifications
— Select Cable Corr
— Auto Use Cable Loss and Velocity Factor values from a Cable file. See
“How to Edit, Save, and Recall a Cable File” below. This will overwrite
a manually-entered value.
— Man
(Default setting) Manually enter a value for Cable Loss and
Velocity Factor.
—Then:
— Velocity Factor
0.01 and 1. Then press Enter
— Cable Loss
value in dB/m, then press Enter
Using the numeric keypad, enter a value between
.
Using the numeric keypad, enter a positive Cable Loss
.
How to Edit a Cable File
The FieldFox includes many predefined cable files with the manufacturer’s
specifications. You can edit these files or create new cable files using the
following procedure or using the FieldFox Data Link Software.
The Cable correction data survives a Mode Preset and Preset.
With a DTF measurement present:
—Press Meas Setup 4
—Then DTF Cable Specifications
—Then Edit/Save/Recall Cables
—Press New Cable
then Yes to clear all data from the existing DTF Cable
table and reset header information to default settings.
—Then Edit Cable
to open the Cable Editor.
— Then use the arrows to select a field,
— When editing Cable Description information, press Edit
then modify
the selected field using the FieldFox labeler.
— When editing Frequency/Loss pairs, enter numbers using the
numeric keypad, then select a frequency suffix. Then Enter
. Learn
more about “How the Freq/Loss pairs are applied” below.
— Optionally choose from the following:
— Previous / Next Page
Quickly scrolls through pages of Freq/Loss
data.
— Add Data
68 Keysight N9938-90006 User’s Guide
Add a blank Freq/Loss pair to the table,
DTF (Distance to Fault) Measurements
DTF Measurement Settings
— Delete/Clear then:
— Delete Row
— Clear All
Remove the selected Freq/Loss pair from the table.
then Yes. Remove all Freq/Loss pairs from the table and
resets header information
—Press Done
to close the Cable Editor.
How to Save or Recall a Cable
—Press Save Cable to saves your changes to the specified Storage Device.
Enter a filename using the FieldFox labeler (learn more on “How to use the
FieldFox labeler ” on page 718). Learn more about Cable files below.
—Press Recall Cable
— Storage Device
to load a Cable file from the specified Storage Device
Changes the device used to save or recall Cable files. This
is a different setting from the Save/Recall Storage Device setting. Choose
from Internal (default setting), USB (must be connected) or SD card.
About Cable files
— Cable files are saved to, and recalled from, the Cables folder. If the folder
does not already exist on a USB or SD card, it is created automatically
before storing the file.
— Cable files are stored as *.xml files. Existing cable files that are preloaded
into the FieldFox firmware can be overwritten. Your edited file will NOT be
overwritten when firmware is updated.
—To DELETE a Cable file, use the Manage Files feature. Learn more in
“Manage Files” on page 724.
How the Freq/Loss pairs are applied
When the cable file contains one Freq/Loss pair, that correction value is
applied to the entire displayed frequency span.
When the cable file contains two or more Freq/Loss pairs, the Loss value that is
used is interpolated from the Freq/Loss pairs and the DTF center frequency.
For example, using a cable file with the following Freq/Loss pairs:
1 GHz: 0.1 dB/m
2 GHz: 0.2 dB/m
Keysight N9938-90006 User’s Guide 69
DTF (Distance to Fault) Measurements
DTF Measurement Settings
The center frequency for the measurement is determined from the
calculated (Stop – Start) frequency values (seen on the second page of
DTF Settings):
Calculated Start = 2.0 MHz
Calculated Stop = 3.598 GHz
Center Freq = 1.80 GHz
The Loss value for the measurement is interpolated from the Freq/Loss pairs at
the Center Freq:
1 GHz = 0.1 dB/m
1.8 GHz = 0.18 dB/m
2 GHz = 0.2 dB/m
The correction for loss at 5 meters in one direction: 0.18 dB/m * 5m = 0.9 dB.
All DTF measurements correct for loss for travel down the DUT and back, so
double the correction: 0.9 dB * 2 = 1.8 dB.
Window Settings
Window settings provide the ability to choose between optimizing DTF
measurements for resolving closely-spaced faults or for the ability to measure
low-level faults.
How to select Window settings
—Press Meas Setup 4.
—Then Settings
— Then press arrows to move to the Window row.
—Then press Edit
—Then press Window
— Maximum
is lowered to provide the ability to measure low-level responses.
(Default setting)
— Medium
settings.
— Minimum
between two closely-spaced responses.
.
.
and choose from the following:
(Maximum – Optimized for dynamic range) the noise floor
– Compromise between Minimum and Maximum window
– Best Response Resolution, providing the ability to resolve
—Then press Done Edi
—Again press Done
70 Keysight N9938-90006 User’s Guide
t.
.
DTF (Distance to Fault) Measurements
DTF Measurement Settings
DTF Units
The DTF Units setting is available ONLY on the DTF Settings table.
By default, X-axis units for DTF measurement settings are displayed in Meters.
How to change DTF units
— With a DTF measurement present, press Freq/Dist.
—Then More
— The current selection is underlined m (meters) Feet
then Distance Unit m feet.
Calculated DTF values
Press Next Page on the DTF Settings Table to view the following calculated
Values noted on the FieldFox screen with c - <setting>
Start Frequency – Start frequency that is used to calculate DTF.
Stop Frequency – Stop frequency that is used to calculate DTF.
Range Resolution. Indicates the accuracy of the distance to fault
measurement. For example, with range resolution of 500 mm, if the distance to
fault is 10 meters, this value could be inaccurate by +/- 500 mm or between
9.5 to 10.5 meters. This value is calculated from frequency span / resolution
(points).
Response Resolution, not displayed, indicates the distance that could be
between two faults and still show as separate faults. Learn more in “Window
Settings” on page 70.
Maximum Distance. The distance that could be viewed with the current
settings. Defined by: Vf*c*Points/(2*Bandwidth) where:
.
— Vf = velocity factor
— c = speed of light
— Points = resolution
— Bandwidth = frequency range
Alias-free Range (On/Off)
—On = No Alias images
— Off = Alias images may appear in the response.
About Alias Faults
An alias fault is not a true device response. An alias fault appears because of
the method used to convert frequency to time.
On the DTF Settings page (above) the c - Alias-free Range = Off setting
indicates alias images MAY appear on the screen.
Keysight N9938-90006 User’s Guide 71
DTF (Distance to Fault) Measurements
DTF Measurement Settings
Shorter stop distances (less than 10 meters) and a higher resolution (1001
points) will be more likely to result in Alias-free Range = Off.
When the Alias-free Range = Off, the following procedure will help to
determine if a response is true or an alias response:
1. Put a marker on the response in question and note the distance to the
fault.
2. Change the start or stop distance.
A true fault response will not move in distance. That is, if a true fault is present
at 10.3 meters, changing the stop distance from 15 m to 20 m will not move
the fault; the fault will remain at 10.3 meters. However, an alias response will
appear to move.
An un-terminated cable (with NO perfect load at the end) will show faults that
appear to be beyond the end of the cable. These are NOT alias faults. These
faults appear as the signal reflects off the open at the end of the cable and
travels back down the cable toward the connection at the FieldFox.
Re-reflections are measured at the FieldFox as mirror images of the original
faults. The largest fault is the open end of the cable. To avoid confusion, set
the Stop distance shortly after that fault.
72 Keysight N9938-90006 User’s Guide
Keysight Handheld Analyzers
N99xxB User’s Guide
5TDR (Time Domain Reflectometry) - Option 215
CAT Mode Time Domain Reflectometry (TDR) measurements are generally
used to locate problems, and identify the type of problem, in a length of cable
or transmission line.
Settings that are NOT unique to TDR measurements are documented in
Chapter 3, “CAT (Cable and Antenna Test) Mode”, on page 47.
In this chapter, the cable to be tested is referred to as the DUT (Device Under
Test) .
In this Chapter
“TDR Measurement (Format)” on page 76
“TDR Settings Table” on page 76
“TDR Start and Stop Distance” on page 77
“TDR Sweep” on page 77
73
TDR (Time Domain Reflectometry) - Option 215
“Frequency Mode” on page 78
“Cable (Correction) Specifications” on page 79
“Window Settings” on page 82
“TDR Units” on page 83
“Calculated TDR values” on page 83
“About Alias Faults” on page 84
“How to Prevent Undersampling” on page 85
Optional Settings
“All about Markers” on page 638
“All about Limit Lines” on page 653
“Saving and Recalling Files” on page 716
Trace Math is NOT available in DTF Measurements.
How to make TDR Measurements
Before starting, you may need the following:
— Jumper cable or adapter to connect the beginning of the DUT to the
FieldFox.
— LOAD with correct connector type and gender to terminate the end of the
DUT (if possible).
— The known length and cable type of the DUT. If the cable type is not known,
then the Cable Loss (dB/Meter) and Velocity Factor of the DUT are required.
1. Connect any necessary jumper cable or adapter to the FieldFox RF
OUT port. Do NOT connect the DUT.
2. Press Preset
3. Then Mode
4. Then Meas 1
5. Press Freq/Dist
then Preset to return the FieldFox to the default settings.
then CAT/TDR.
then More then DTF (Lin) or TDR (Lin rho) or (ohm).
, then Stop Distance and enter the length of the DUT. You
can optionally set the Start Distance
6. Press Cal 5
and follow the Cal prompts. Learn all about Calibration in
“How to Perform a Calibration” on page 125.
7. Disconnect any components or antenna that should NOT be measured and
connect a LOAD at the end of the DUT.
8. Press Meas Setup 4
74 Keysight N9938-90006 User’s Guide
then DTF Cable Specifications.
TDR (Time Domain Reflectometry) - Option 215
9. Either press Edit/Save/Recall Cables then Recall Cable, or directly enter
the Velocity Factor/Cable Loss of the DUT using the Velocity Facto
Cable Loss
keys.
r and
10. Connect the start end of the DUT to the FieldFox.
11. Press Meas Setup 4
then Settings then Next Page. If the Alias-free Range
setting is Off, then you may see Alias faults on the screen. Learn more in
Select CAT Mode, then TDR before making any setting in this chapter.
How to select CAT Mode
—Press Mode
—Then CAT/TDR
.
.
TDR Measurement (Format)
You can select from 3 different TDR Formats.
—Press Measure 1
—Then More
— Then choose from:
— TDR (Lin rho)
A trace without reflections shows as 0 (zero). Maximum reflections
from an open or short show as 1.
— TDR (ohm)
— DTF (Lin)
X-Axis displays distance to fault in meters.
— TDR & DTF TDR and DTF are displayed.
.
The Y-axis of the display is linear, real, unitless values.
The Y-axis of the display is impedance (ohms).m
The Y-axis of the display is linear, real, unitless values.
TDR Settings Table
You can set and view all of the TDR settings, including some calculated values,
on the TDR Settings table. Learn more in “Calculated TDR values” on page 83.
How to make settings on the TDR settings table
—Press Meas Setup 4.
—Then Settings
—Press Next Page
— To change a setting:
— Use the arrows or rotary knob to highlight a setting.
— Numeric settings can be changed by pressing numbers using the
numeric keypad. Then press Enter or select a suffix if available.
—Press Dock Window to relocate the Settings table to a position relative to
the trace window. The setting persists through a Preset. Choose from the
following:
— Full (Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
— Left The Settings table is shown to the left of the trace window.
— Bottom The Settings table is shown below the trace window.
— When finished changing ALL settings, press Done
The following TDR settings are available ONLY on the settings table:
Mkr Format – The format of marker readout. Choose from:
— Z Magnitude – Impedance (magnitude only)
— Default – The current TDR measurement format.
Auto: Max Freq – When TDR Sweep is set to Auto, this is the maximum stop
frequency of the measurement. The minimum value of this setting is 100 MHz.
TDR Start and Stop Distance
In TDR measurements, you set the physical length of the DUT. The FieldFox
calculates the frequency range of the measurement from this distance. The
longer the DUT, the lower the frequencies that are used.
You can NOT set the frequencies manually in TDR. However, you CAN limit the
maximum frequency using the TDR Sweep setting. Learn more in “TDR Sweep”
on page 77.
How to set Start and Stop Distance
— With a TDR measurement present, press Freq/Dist.
to save your settings.
— Then choose from the following:
— Start Distance
arrows, or the rotary knob, then Enter
is set to 0 Meters. This means that the measurement will display
faults starting at the point at which calibration standards are
connected.
— Stop Distance
16,404 ft.) using the numeric keypad, the arrows, or the rotary
knob, then Enter
Enter a value using the numeric keypad, the
. By default, the Start Distance
Enter a value between the start distance and 5 km (or
.
TDR Sweep
This setting allows you to choose the method by which the Stop Frequency and
Resolution are set.
By default, the FieldFox does NOT correct TDR measurements to account for
the inherent loss of a cable. However, to make more accurate TDR
measurements, the Cable Loss and Velocity Factor values should be
considered.
The following describes Coax Media type. Waveguide Media can NOT be
measured with TDR as waveguide is inherently a high pass device.
About Velocity Factor and Cable Loss
— Velocity Factor is a property of the physical material of a cable. A VF of 1.0
corresponds to the speed of light in a vacuum, or the fastest VF possible. A
polyethylene dielectric cable has VF = 0.66 and a cable with PTFE dielectric
has VF = 0.7.
— Cable Loss is specified in dB/meter. In addition to the length of the cable,
loss is also directly proportional to the frequency of the signal that passes
through the cable.
The following is an example showing how DTF cable correction works:
The DUT is a 100 meter transmission cable. The Cable Loss value is 0.1
dB/meter. This means that a signal traveling ONE WAY through the cable will
lose 10 dB of power (100 m * 1dB/m). Because the FieldFox performs this
measurement with 1 port, the test signal travels down the cable and then back,
for a total loss of 20 dB.
After a calibration has been performed, for the purpose of illustrating this
point, connect an OPEN to the end of the cable – a maximum-sized fault - for
100% reflection of the 300 MHz test signal.
Without compensation for the loss of the cable, a –20 dB response would be
visible at 100 meters, which is the OPEN at the end of the DUT. This is from 10
dB of loss through the cable in each direction.
With compensation for the loss using the manufacturer’s specification, the
FieldFox compensates the trace as though the signal traveling through 100
meters was increased by +20 dB. Therefore the response will show 0 dB for
100% reflection.
How to enter Cable Loss and Velocity Factor
Cable Loss and Velocity factor can be entered using one of the following
methods:
— Manually enter cable loss and velocity factor for the measurement.
— Select or create a cable file which contains the cable loss and velocity
— Auto Use Cable Loss and Velocity Factor values from a Cable file. See
“How to Edit, Save, and Recall a Cable File” below. This will overwrite
a manually-entered value.
— Man
(Default setting) Manually enter a value for Cable Loss and
Velocity Factor.
—Then:
— Velocity Factor
0.01 and 1. Then press Enter
— Cable Loss
value in dB/m, then press Enter
Using the numeric keypad, enter a value between
.
Using the numeric keypad, enter a positive Cable Loss
.
How to Edit a Cable File
The FieldFox includes many predefined cable files with the manufacturer’s
specifications. You can edit these files or create new cable files using the
following procedure or using the FieldFox Data Link Software.
The Cable correction data survives a Mode Preset and Preset.
With a TDR measurement present:
—Press Meas Setup 4
—Then DTF Cable Specifications
—Then Edit/Save/Recall Cables
—Press New Cable
then Yes to clear all data from the existing DTF Cable
table and reset header information to default settings.
—Then Edit Cable
to open the Cable Editor.
— Then use the arrows to select a field,
— When editing Cable Description information, press Edit
then modify
the selected field using the FieldFox labeler.
— When editing Frequency/Loss pairs, enter numbers using the
numeric keypad, then select a frequency suffix. Then Enter
. Learn
more about “How the Freq/Loss pairs are applied” below.
— Optionally choose from the following:
— Previous / Next Page
Quickly scrolls through pages of Freq/Loss
data.
— Add Data
— Delete/Clear
— Delete Row
80 Keysight N9938-90006 User’s Guide
Add a blank Freq/Loss pair to the table,
then:
Remove the selected Freq/Loss pair from the table.
— Clear All then Yes. Remove all Freq/Loss pairs from the table and
resets header information
— Press Done to close the Cable Editor.
How to Save or Recall a Cable
—Press Save Cable to save your changes to the specified Storage Device.
Enter a filename using the FieldFox labeler (see “How to use the FieldFox
labeler ” on page 718. Learn more about Cable files below
—Press Recall Cable
— Storage Device
to load a Cable file from the specified Storage Device
Changes the device used to save or recall Cable files. This
is a different setting from the Save/Recall Storage Device setting. Choose
from Internal (default setting), USB (must be connected) or SD card.
About Cable files
— Cable files are saved to, and recalled from, the Cables folder. If the folder
does not already exist on a USB or SD card, it is created automatically
before storing the file.
— Cable files are stored as *.xml files. Existing cable files that are preloaded
into the FieldFox firmware can be overwritten. Your edited file will NOT be
overwritten when firmware is updated.
—To DELETE a Cable file, use the Manage Files feature. Learn more in
“Manage Files” on page 724.
How the Freq/Loss pairs are applied
When the cable file contains one Freq/Loss pair, that correction value is
applied to the entire displayed frequency span.
When the cable file contains two or more Freq/Loss pairs, the Loss value that is
used is interpolated from the Freq/Loss pairs and the TDR center frequency.
For example, using a cable file with the following Freq/Loss pairs:
1 GHz: 0.1 dB/m
2 GHz: 0.2 dB/m
The center frequency for the measurement is determined from the calculated
(Stop – Start) frequency values (seen on the second page of TDR Settings):
The Loss value for the measurement is interpolated from the Freq/Loss pairs at
the Center Freq:
1 GHz = 0.1 dB/m
1.8 GHz = 0.18 dB/m
2 GHz = 0.2 dB/m
The correction for loss at 5 meters in one direction: 0.18 dB/m * 5m = 0.9 dB.
All TDR measurements correct for loss for travel down the DUT and back, so
double the correction: 0.9 dB * 2 = 1.8 dB.
Window Settings
Window settings provide the ability to choose between optimizing TDR
measurements for resolving closely-spaced faults or for the ability to measure
low-level faults.
How to select Window settings
—Press Meas Setup 4.
—Then Settings
.
— Then press arrows to move to the Window row.
—Then press Edit
—Then press Window
— Maximum
.
and choose from the following:
(Maximum – Optimized for dynamic range) the noise floor
is lowered to provide the ability to measure low-level responses.
(Default setting)
— Medium
— Minimum
– Compromise between Min and Max window settings.
– Best Response Resolution, providing the ability to resolve
Network Analyzer Windows Frequency Span and TDR Window
Selection for Step Rise Time (Approximated)
The following table shows the approximated relationship between the
frequency span and the window selection for impulse width and step rise time.
Refer to Figure 5-1 on page 83.
WindowLow-pass step
(10% to 90%)
Minimum0.45/ f span0.60 / f span1.20 / f span
Medium0.99 / f span0.98 / f span1.95 / f span
Maximum1.48 / f span1.39 / f span2.77 / f span
Figure 5-1 Network Analyzer TDR Rise Time
Low-pass impulseBandpass impulse
TDR Units
By default, X-axis units for TDR measurement settings are displayed in Meters.
How to change TDR units
— With a TDR measurement present, press Freq/Dist
—Then More
then Distance Unit m feet.
— The current selection is underlined m (meters) Feet
.
.
The TDR Units setting is also available on the TDR Settings table.
Calculated TDR values
Press Next Page on the TDR Settings Table to view the following calculated Values noted on the FieldFox screen with c - <setting>
Start Frequency – Start frequency that is used to calculate TDR.
Stop Frequency – Stop frequency that is used to calculate TDR.
Range Resolution. Indicates the accuracy of the distance to fault
measurement. For example, with range resolution of 500 mm, if the distance to
fault is 10 meters, this value could be inaccurate by +/- 500 mm or between
9.5 to 10.5 meters. This value is calculated from frequency span / resolution
(points).
Response Resolution, not displayed, indicates the distance that could be
between two faults and still show as separate faults. Learn more in Window
Settings in “Window Settings” on page 82.
Maximum Distance. The distance that could be viewed with the current
settings. Defined by: Vf*c*Points/(2*Bandwidth) where:
— Vf = velocity factor
— c = speed of light
— Points = resolution
— Bandwidth = frequency range
Alias-free Range (On/Off)
—On = No Alias images
— Off = Alias images may appear in the response.
About Alias Faults
An alias fault is not a true device response. An alias fault appears because of
the method used to convert frequency to time.
On the TDR Settings page (above) the c - Alias-free Range = Off setting
indicates alias images MAY appear on the screen.
Shorter stop distances (less than 10 meters) and a higher resolution (1001
points) will be more likely to result in Alias-free Range = Off.
When the Alias-free Range = Off, the following procedure will help to
determine if a response is true or an alias response:
1. Put a marker on the response in question and note the distance to the
fault.
2. Change the start or stop distance.
A true fault response will not move in distance. That is, if a true fault is present
at 10.3 meters, changing the stop distance from 15 m to 20 m will not move
the fault; the fault will remain at 10.3 meters. However, an alias response will
appear to move.
An un-terminated cable (with NO perfect load at the end) will show faults that
appear to be beyond the end of the cable. These are NOT alias faults. These
faults appear as the signal reflects off the open at the end of the cable and
travels back down the cable toward the connection at the FieldFox.
Re-reflections are measured at the FieldFox as mirror images of the original
faults. The largest fault is the open end of the cable. To avoid confusion, set
the Stop distance shortly after that fault.
How to Prevent Undersampling
Undersampling is when the TDR measurement is inaccurate because the
frequency step size (resolution) is too large. The predominant symptom of this
is when the measured impedance is incorrect. For example, when measuring a
75Ω cable, a marker that is on a part of the trace without reflections shows
56Ω.
To prevent undersampling, set TDR Sweep to Cable. If undersampling still
occurs, set TDR Sweep to Auto. Then set Auto Max Frequency to a lower
value. Manually increase the Resolution so that the Start Frequency setting is
less than about 500 kHz.
Select NA Mode before making any setting in this chapter.
How to select NA Mode
—Press Mode_
—Then NA
.
About S-parameters
S-parameters (scattering parameters) are used to describe the way a device
modifies a signal. The FieldFox can measure four S-parameters. The syntax for
each parameter is described by the following:
S (out | in)
out = FieldFox receiver port
in = FieldFox source port
.
The FieldFox automatically switches the internal source and receivers to make
both forward and reverse measurements. Therefore, the FieldFox can measure
all four S-parameters with a single connection
When the source comes from port 1, the measurement is said to be in the
forward direction
When the source comes from port 2, the measurement is said to be in the
reverse direction
S11 and S22 reflection measurements are used to measure the amount of
reflections off the corresponding DUT port. Low reflections means there is a
good impedance match between the source and DUT.
Keysight N9938-90006 User’s Guide 89
NA (Network Analyzer) Mode
NA Mode Settings
S21 and S12 transmission measurements are used to measure the loss or gain
through a DUT over a specified frequency range. Both ends of the DUT must be
connected to the FieldFox. The FieldFox signal source is transmitted out one of
the test port connectors, through the DUT, and into the other test port
connector
How to measure S-parameters
1. Press Preset then Mode Preset (Factory)
2. Press Measure 1 then choose from the following:
— S11
— S21
— S12
— S22
OR select a multi-trace configuration. Learn more on “Multi-Trace
Configurations” on page 93
3. Press Freq/Dist
Frequency Range for the measurement.
4. Press BW 2
Narrower bandwidths require more time to sweep, but lowers trace noise.
5. Press Sweep 3
measurement. More data points require more time to sweep.
6. Press Cal 5
a Calibration” on page 125.
7. All other settings can be made AFTER calibration without compromising
measurement accuracy.
Reflection measurement at port 1.
Forward 2-port transmission measurement.
Reverse 2-port transmission measurement. May require an
option.
Reflection measurement at port 2. May require an option.
then either Start and Stop or Center and Span to enter a
then IFBW to select the IF Bandwidth for the measurement.
then Resolution to select the number of data points for the
to calibrate the measurement. Learn more in “How to Perform
Mixed-Mode S-Parameters
In NA Mode ONLY, mixed-mode S-parameters (also known as Balanced
measurements) are available with Option 212.
Because the FieldFox has only two test ports, only reflection measurements are
available. Connect the balanced input or output of your DUT to the FieldFox
ports 1 and 2
For highest accuracy, a Full 2-port calibration is required.
90 Keysight N9938-90006 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
All FieldFox settings and features are supported (except Parameter Conversion)
with mixed-mode S-parameters.
Learn more about Balanced Measurements with the FieldFox in the
FieldFox Supplemental Online Help:
To make balanced reflection measurements at the DUT output, connect the
DUT output to the FieldFox ports.
.
Common reflect/common incident for logical port 1.
Differential reflect/differential incident for logical port 1.
Differential reflect/common incident for logical port 1.
Common reflect/differential incident for logical port 1.
Parameter Conversion
In NA Mode ONLY, converts the active S-parameter trace to an equivalent
impedance (Z), admittance (Y), or reciprocal 1/S-parameter
How to select parameter conversions
—Press Measure 1
—Then select an S-parameter
—Then More
—Then Conversion
— Then choose from the following:
(default) No conversion is performed.
—Off
—Z Conv (--) or Y Conv (--)
admittance (Y). Then choose from the following for either:
——Auto
——Refl
Keysight N9938-90006 User’s Guide 91
The displayed S-parameter is converted to the
appropriate Z or Y parameter: Refl for S11 and S22; Trans for
S21 and S12. When the S-parameter is changed, the
appropriate conversion changes automatically.
The displayed S-parameter is converted to Z or Y
reflection, regardless of whether the S-parameter is reflection
(S11 or S22) or transmission (S21 or S12).
Perform conversion for impedance (Z) or
NA (Network Analyzer) Mode
NA Mode Settings
——Trans The displayed S-parameter is converted to Z or Y
transmission, regardless of whether the S-parameter is
reflection (S11 or S22) or transmission (S21 or S12).
—1/1/S
The displayed S-parameter is converted to 1/S-parameter.
Receiver Measurements
In NA Mode ONLY, you can make unratioed, uncorrected receiver
measurements. These measurements are typically used as diagnostic tools for
service personnel.
Receiver measurements are NOT corrected for absolute power. They are
only useful for making relative measurements. The Y-Axis and markers are
labeled in dB - NOT dBm. Specifically, R1 and R2, do NOT indicate the
actual power levels out of the source ports.
In NA Mode you can display multiple traces on the FieldFox screen.
The graphic above shows a x3H configuration. Tr2 is the ACTIVE trace as
indicated by the highlighted Tr2.
Trace Setting Notes
— The Frequency Range, IF BW, Resolution, Average, and Output Power
settings are common for all displayed traces.
— All other trace settings, such as measurement, format, and limit lines, are
applied individually to the ACTIVE trace in the same manner as when a
single trace is present.
IMPORTANT! For CAT and NA modes, limit lines do not apply
where F1 = F2.
— By default, a marker is created on ALL traces. However, they can be created
individually by disabling Coupled Markers. Learn more in “Coupled Markers
(NA Mode)” on page 641.
Keysight N9938-90006 User’s Guide 93
NA (Network Analyzer) Mode
NA Mode Settings
How to select a multi-trace configuration
—Press Trace 6
—Then Num of Traces
— Then choose from the following:
The default measurements depend on the options that are installed
x1
x2
x2H
x3H
x3
x4
x4H
1 trace standard configuration
2 traces overlayed on a single graticule
2 traces on separate graticules
3 traces on separate graticules
3 traces overlayed on a single graticule
4 traces overlayed on a single graticule
4 traces on separate graticules
How to activate a trace in order to change a setting
—Use the arrows OR
—Press Trace 6
then select Trace 1, Trace 2, Trace 3, or Trace 4, Only traces
that are shown can be activated.
How to maximize the viewing of the active trace
—Press System 7
—Then Full Screen.
— Press any key to return to the standard display.
Quick Settings
Both CAT and NA Modes allow you to view and change most relevant settings
from a single location. All of these settings are discussed in this chapter and,
unless otherwise noted, ALL of these settings can also be made using the
standard softkey menus.
94 Keysight N9938-90006 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
How to view and change Quick Settings
—Press Meas Setup 4
—Then Settings
—Press Next Page
.
and Previous Page to view all settings. If these softkeys
are NOT available, then all available settings fit on one page.
— To change a setting:
—Use the arrows to highlight a setting.
—Then press Edit
. The current setting changes to yellow.
—Some settings require you to press a softkey to change the value.
Otherwise, use the numeric keypad, arrows, or rotary knob to
change the value.
—When finished changing a value, press Done Edit
—Press Dock Window
to relocate the Settings table to a position relative to
.
the trace window. The Dock Window setting persists through a Preset.
Choose from the following:
—Full (Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
—Left The Settings table is shown to the left of the trace window.
—Bottom The Settings table is shown below the trace window.
When finished changing ALL settings, press Done
Calibration Settings
NA Mode has a page specifically for making settings that pertain to Port
Extensions, including Velocity Factor and Media Type.
Learn more in “Port Extensions” on page 106.
Learn more about Media Type in “Waveguide Calibrations” on page 137.
How to view and change Calibration Settings
—Press Meas Setup 4
—Then Calibration Settings
— Make these settings in the same manner as Quick Settings in the previous
section.
to save your settings.
.
Keysight N9938-90006 User’s Guide 95
Format
NA (Network Analyzer) Mode
NA Mode Settings
Select the display format in which to present measurement results. This setting
can be changed at any time without affecting calibration accuracy.
Learn more about Display Formats in the FieldFox Supplemental Online
Help:
The marker on screen readout can be changed to formats other than the
display format. Learn more in “Marker Format” on page 643.
How to select a NA Format
—Press Measure 1
—Then Format [current setting]
.
.
— Then choose from the following:
—Log Magnitude
—Linear
Displays positive values only. Y-axis: Unitless (U) for ratioed
Displays magnitude in dB
measurements; Watts (W) for unratioed measurements.
—VSWR
—Phase
Used mainly for S11 and S22. Displays unitless reflection data.
Displays phase in degrees. The trace ‘wraps’ every 360 degrees,
from +180 to –180, for easy scaling.
—Smith
Used mainly for S11 and S22, which are plotted on a Smith
Chart. Displays series resistance and reactance.
— More
—Polar
then…
Used mainly for S11 and S22. Displays magnitude and phase
of the reflection coefficient.
— Group Delay
Used mainly for S21 and S12. Displays signal
transmission (propagation) time through a device in seconds. The
Group Delay aperture is the current Smoothing aperture. Default is
1.5% of the X-axis.
——Learn how to set Smoothing aperture on “Smoothing” on
page 101.
——Learn more about Group Delay measurements at the FieldFox
Displays only the real (resistive) portion of the measured complex
—Real
data. The Y-axis is Unitless. Often used for Time Domain
measurements.
96 Keysight N9938-90006 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
—Imaginary Displays only the reactive portion of the measured complex
data. The Y-axis is Unitless.
—Z Magnitude
—Unwrap Phase
Phase is unwrapped by comparing the phase from one data point to the
next. If the phase difference between two data points is greater than 180
degrees, or if the phase of the first data point is greater than 180 degrees
from DC, than the phase measurement is probably NOT accurate. To
ensure that the phase measurement is accurate, increase the resolution
setting. When making a narrow band measurement, reduce the start
frequency for the unwrapped phase measurement to ensure the first data
point is less than 180 degrees from DC.
Frequency Range
Set the range of frequencies over which you would like to make measurements.
When the frequency range is changed after a calibration is performed, the cal
becomes interpolated. Learn more in “Interpolation *” on page 139.
How to set Frequency Range
This can be done in two ways:
Displays the effective impedance in ohms.
Same as Phase, but without 180 degree wrapping.
—Press Freq/Dist
— Then choose from the following:
— Start
— Center
— Follow each by entering a value using the numeric keypad, the arrows,
or the rotary knob.
— After using the arrows or the rotary knob, press Enter
setting of the arrows is based on the current span and can NOT be changed
in NA Mode.
— After using the keypad, select a multiplier key. Learn more in “Multiplier
Abbreviations” on page 45.
Scale Settings
Adjust the Y-axis scale to see the relevant portions of the data trace. The Y-axis
is divided into 10 graticules.
and Stop frequencies – Specify the beginning and end of the
sweep.
and Span frequencies - Specify the center frequency and
span of frequencies (half on either side of center).
. The increment
This setting can be changed at any time without affecting calibration accuracy.
Keysight N9938-90006 User’s Guide 97
NA (Network Analyzer) Mode
NA Mode Settings
How to set Scale
—Press Scale / Amptd
.
— Then choose from the following method
—Autoscale
Automatically adjusts the Y-axis to comfortably fit the
Minimum and Maximum amplitude of the trace on the screen.
—Autoscale All
Autoscales all of the traces on the screen, useful only for
multi-trace configurations.
—3.Set Scale, acquisition, and Reference Position
——Scale
Manually enter a scale per division to view specific areas
of the trace.
——Ref Level
Manually set the value of the reference line. Enter a
negative value by pressing Run/Hold (+/-)
after typing a value.
——Ref Pos
Manually set the position of the reference line. Values
must be between 0 (TOP line) and 10 (BOTTOM line)
Scale annotation on the FieldFox screen
—Reference Line = red arrow
—Ref Level = –40 dB
—Ref Position = 1
—Scale = 2 dB per division
either before or
Magnitude Offset
Magnitude Offset allows you to offset the magnitude (not phase) data by a
fixed and/ or sloped value in dB. If the display format is Linear Magnitude or
Real (unitless), the conversion from dB is performed and the correct amount of
offset is implemented.
The Magnitude offset setting affects only the active trace.
How to set Magnitude Offset
—Press Scale / Amptd
—Then More
—Then Magnitude Offset
value.
98 Keysight N9938-90006 User’s Guide
- Offsets the entire data trace by the specified
NA (Network Analyzer) Mode
NA Mode Settings
—Then enter a positive or negative offset value in dB using the numeric
keypad, the arrows, or the rotary knob. Press a multiplier key.
Learn about multiplier abbreviations.
—And/or press Slope
with frequency. The offset slope begins at 0 Hz. and continues to the stop
frequency. The offset value at the start frequency is then calculated from 0
Hz.
—Then enter a positive or negative offset slope value in dB/GHz using
Electrical Delay
Electrical delay is a mathematical function that simulates a variable length of
lossless transmission line. Use the electrical delay feature to compensate for
the linear phase shift through a device and view only the deviation from linear
phase of the device.
You can set the electrical delay independently for each measurement trace. To
apply an electrical delay to all measurement traces, use Port Extensions. Learn
how in “Port Extensions” on page 106.
Learn how to set Phase formats in “Phase Offset” on page 99.
How to set Electrical Delay
- Offsets the data trace by a value that changes linearly
the numeric keypad, the arrows, or the rotary knob. Press a
multiplier key. Learn about multiplier abbreviations in “Multiplier
Abbreviations” on page 45.
—Press Scale / Amptd
—Then More
—Then Electrical Delay
— Enter a time value using the numeric keypad, the arrows, or the rotary
knob.
— Press a multiplier key. Learn about multiplier abbreviations in “Multiplier
Abbreviations” on page 45.
Electrical Delay can also be set using the Mkr->Delay feature. Learn how on
“Marker Functions” on page 651.
Phase Offset
Phase offset mathematically adjusts the phase measurement by a specified
amount, up to 360°. Use this feature in the following ways:
Improve the display of a phase measurement. This is similar to the way you
would change the acquisition in an amplitude measurement. Change the
phase response to center or the response on the screen.
Keysight N9938-90006 User’s Guide 99
NA (Network Analyzer) Mode
NA Mode Settings
Emulate a projected phase shift in your measurement. For example, if you
know that you need to add a cable and that the length of that cable will add a
certain phase shift to your measurement, you can use phase offset to add that
amount and simulate the complete device measurement.
You can set the phase offset independently for each measurement trace.
How to set Phase Offset
—Press Scale / Amptd
—Then More
—Then Phase Offset
— Enter a value in degrees using the numeric keypad, the arrows, or the
Averaging
Averaging helps to reduce the effects of random noise on a measurement. You
specify the number of measurements to be averaged. The more measurements
averaged, the greater the amount of noise reduction. An average counter is
shown in the left edge of the screen as Avg <n> where <n> is the number of
measurements that are averaged.
rotary knob. Press Enter
Averaging can be set before or after calibration. When set before calibration,
each calibration standard is measured <n> times and averaged. More time is
needed to perform the calibration, but there will be less noise in the resulting
error terms which means that subsequent measurements will also have less
noise. In addition, noise is further reduced by continuing to average after
calibration.
How to set Averaging
—Press BW 2
—Then Average <n>
.
where <n> is the number of measurements to average.
— Enter a value using the numeric keypad. Enter 1 for NO averaging.
—Press Enter
—Then Average Mode
.
Choose from the following:
—Sweep - Each data point is based on the average of the same data
point being measured over <n> consecutive sweeps. The average
counter shows the number of previous sweeps that have been
averaged together to form the current trace. When the counter
reaches the specified count, then a ‘running average’ of the last <n>
sweeps is displayed.
100 Keysight N9938-90006 User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.