No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Keysight Technologies as governed by United
States and international copyright laws.
Manual Part Number
N9911-90002
Edition
August 2014
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Keysight disclaims all warranties, either express or implied, with
regard to this manual and any information contained herein, including
but not limited to the implied warranties of merchantability and fitness for
a particular purpose. Keysight shall
not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Keysight and the user have a
separate written agreement with
warranty terms covering the material
in this document that conflict with
these terms, the warranty terms in
the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Keysight Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
Printing Copies of Documentation from the
Web
To print copies of documentation from the Web, download the PDF file
from the Keysight web site:
•Go tohttp://www.keysight.com.
•Enter the product model number (Ex: N9911X) in the Search bar.
•Click Search.
•Click the Manuals hyperlink.
•Click the hyperlink title for the document you want to print - this
downloads the PDF
•Print the document after the PDF has fully downloaded.
N9911X Economical Waveguide Calibration Components are used to calibrate FieldFox analyzers. With the
calibration data properly loaded in the analyzer and a measurement calibration completed, systematic errors
are minimized. For information on components available to purchase, refer to the online document Keysight N9911X Economical Waveguide Calibration Components Configuration Guide (part number N9911-90003).
1- 2User’s Guide N9911-90002
General Information
Recording the Components Serial Numbers
Recording the Components Serial Numbers
The N9911X Economical Waveguide Components are individually serialized (serial numbers are labeled onto
the body of each component). Record these serial numbers in the appropriate table. Recording the serial
numbers will prevent confusion with similar components.
The N9911X calibration components are manufactured by Flann Microwave. Each component is imprinted
with a Flann logo, part ID number, and serial number. To determine the Keysight option number for a
component, use the cross-references in the following tables.
Table 1-1 Serial Number Record, Waveguide Designators C-Band/WR137/WG14, 5.38 – 8.18 GHz
Description Component
Serial Number
Calibration Components – Metric
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Calibration Components – English (Imperial)
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Keysight
Option Number
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
–110
–110
–111
–112
–113
–115
–115
–116
–117
–118
Flann
Part ID Number
14091–NM70–6332
14091–NM70–6332
14045–6333
14191–6334
14491–02–6335
14091–NM70–6136
14091–NM70–6136
14045–3703
14191–704
14491–02–3483
User’s Guide N9911-90002 1-3
General Information
Recording the Components Serial Numbers
Table 1-2 Serial Number Record, Waveguide Designator X-Band/WR90/WG16, 8.2 – 12.5 GHz
Description Component
Serial Number
Calibration Components – Metric
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Calibration Components – English (Imperial)
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Keysight
Option Number
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
–210
–210
–211
–212
–213
–215
–215
–216
–217
–218
Flann
Part ID Number
16091–NM70–6336
16091–NM70–6336
16045–6337
16191–6338
16491–02–6339
16091–NM70–6141
16091–NM70–6141
16045–5509
16191–2547
16491–02–2401
1- 4User’s Guide N9911-90002
General Information
Recording the Components Serial Numbers
Table 1-3 Serial Number Record, Waveguide Designator Ku-Band/WR62/WG18, 11.9 – 18 GHz
Description Component
Serial Number
Calibration Components – Metric
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Calibration Components – English (Imperial)
Adapter 1, waveguide to type-N (male) coax
Adapter 2, waveguide to type-N (male) coax
Termination
Flush short
¼ wavelength offset shim
Keysight
Option Number
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
–310
–310
–311
–312
–313
–315
–315
–316
–317
–318
Flann
Part ID Number
18091–NM–6340
18091–NM–6340
18045–6341
18191–6342
18491–02–6343
18091–NM–6024
18091–NM–6024
18045–2546
18191–4127
18491–02–4095
User’s Guide N9911-90002 1-5
General Information
Recording the Components Serial Numbers
Table 1-4 Serial Number Record, Waveguide Designator K-Band/WR42/WG20, 17.6 – 26.7 GHz
Description Component
Serial Number
Calibration Components – Metric
Adapter 1, waveguide to 3.5 mm (male) coax
Adapter 2, waveguide to 3.5 mm (male) coax
Termination
Flush short
¼ wavelength offset shim
Calibration Components – English (Imperial)
Adapter 1, waveguide to 3.5 mm (male) coax
Adapter 2, waveguide to 3.5 mm (male) coax
Termination
Flush short
¼ wavelength offset shim
Keysight
Option Number
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
N9911X
–310
–310
–311
–312
–313
–315
–315
–316
–317
–318
Flann
Part ID Number
20091–JM–2371
20091–JM–2371
20045–2376
20191–6349
20491–02–6351
20091–JM–2372
20091–JM–2372
20045–5174
20191–6348
20491–02–6350
1- 6User’s Guide N9911-90002
General Information
Preventive Maintenance
Preventive Maintenance
The best techniques for maintaining the integrity of the components include:
•routine visual inspection
•routine cleaning
•proper connection techniques
All of these are described in Chapter 3. Failure to detect and remove dirt or metallic particles on a mating
plane surface can degrade repeatability and accuracy and can damage any component mated to it. Improper
connections resulting from poor connection techniques, can also damage these components.
When to Calibrate the Analyzer
An analyzer calibration remains valid as long as the changes in the systematic error are insignificant. This
means that changes to the uncorrected leakages (directivity and isolation), mismatches (source match and
load match), and frequency response of the system are small (<10%) relative to accuracy specifications.
Change in the environment (especially temperature) between calibration and measurement is the major
cause in calibration accuracy degradation. The major effect is a change in the physical length of external and
internal cables. Other important causes are dirty and damaged test port connectors and calibration
standards. If the connectors become dirty or damaged, measurement repeatability and accuracy is affected.
Fortunately, it is relatively easy to evaluate the general validity of the calibration. To test repeatability,
remeasure one of the calibration standards. If you can not obtain repeatable measurements from your
calibration standards, maintenance needs to be performed on the test port connectors, cables and
calibration standards. Also, maintain at least one sample of the device under test or some known device as
your reference device.
How to Calibrate the Analyzer
Your analyzer’s calibration user interface prompts you through each step in a 1-port or a 2-port calibration
procedure. For detailed information on using your analyzer, refer to the appropriate user guide. See “Where
to Look for More Information” on page 4-3.
User’s Guide N9911-90002 1-7
General Information
Preventive Maintenance
1- 8User’s Guide N9911-90002
2 Specifications
User’s Guide N9911-900022- 1
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Temperature
Operating
Storage-20 °C to +70 °C
1. The temperature range over which the calibration components maintain performance to their specifications.
1
+4 °C to +40 °C
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a
critical factor in performance. During a measurement calibration, the temperature of the calibration
components must be stable and within the range shown in
Tab le 2- 1.
IMPORTANTAvoid unnecessary handling of the components during calibration because your fingers act
as a heat source and may increase the temperature of the component.
2- 2User’s Guide N9911-90002
Specifications
Typical VSWR Values
Typical VSWR Values
Tab le 2- 2 and Tab le 2 - 3 list the typical VSWR values for the N9911X terminations and waveguide-to-coax
adapters.
Tab le 2- 2 Ty p i c a l VS WR Va l u e s f or Ter mi n a t i o ns
Keysight
Option Number
for Termination
N9911X–111
N9911X–116 English
N9911X–211
N9911X–216 English
N9911X–311
N9911X–316 English
N9911X–411
N9911X–416 English
Waveguide Band DesignatorThread
Ty p e
Metric
C-Band, WR137, WG14
(Imperial)
Metric
X-Band, WR90, WG16
(Imperial)
Metric
Ku-Band, WR62, WG18
(Imperial)
Metric
K-Band, WR42, WG20
(Imperial)
Frequency VSWR,
Maximum
1.10
5.38 to 8.18 GHz
8.2 to 12.5 GHz1.02
11.9 to 18 GHz1.02
17.6 to 26.7GHz1.03
1.02
Table 2-3 Typical VSWR Values for Waveguide-to-Coax Adapters
Flush shortShorting face flatness25 um25 um25 um25 um
1/4 wavelength offset
shim
Waveguide aperture
tolerance
tolerance
Waveguide aperture
tolerance
46 um MAX25 um MAX23 um MAX20 um MAX
46 um MAX25 um MAX23 um MAX20 um MAX
46 um MAX,
14.711 mm
+/– 45 um
25 um MAX,
9.63 mm
+/– 35 um
23 um MAX,
6.668 mm
+/– 30 um
20 um MAX,
4.501 mm
+/– 25 um
NOTERefer to the Appendix for graphics showing the component dimensions.
User’s Guide N9911-90002 2-5
Specifications
Dimension Specifications
2- 6User’s Guide N9911-90002
3 Use, Maintenance, and Care of the Components
User’s Guide N9911-900023- 1
Use, Maintenance, and Care of the Components
Electrostatic Discharge
Electrostatic Discharge
Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning
connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration
components and devices under test (DUT), can also carry an electrostatic charge. To prevent damage to the
test set, components, and devices:
•always wear a grounded wrist strap having a 1 MW resistor in series with it when handling components
and devices or when making connections to the test set.
•always use a grounded antistatic mat in front of your test equipment.
•always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the
conductivity of your floor, wear a heel strap.
•always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or
test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
•always ground the center conductor of a test cable before making a connection to the analyzer test port
or other static-sensitive device. This can be done as follows:
1. Connect a short to one end of the cable to short the center conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
•Remove the short from the cable.
Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap.
Figure 3-1 ESD Protection Setup
3- 2User’s Guide N9911-90002
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.