No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N9030-90017
Edition
Edition 1, December 2020
Supersedes: September 2020
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including instrument software
upgrades, application information, and product information, browse to one of the following URLs, according to the name
of your product:
http://www.keysight.com/find/pxa
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search
for software updates for your product, go to the Keysight Technical Support website at:
This chapter contains the specifications for the core signal analyzer. The
specifications and characteristics for the measurement applications and
options are covered in the chapters that follow.
19
PXA Signal Analyzer
Definitions and Requirements
Definitions and Requirements
This book contains signal analyzer specifications and supplemental
information. The distinction among specifications, typical performance, and
nominal values are described as follows.
Definitions
— Specifications describe the performance of parameters covered by the
product warranty (temperature = 0 to 55°C
temperature range" or "Full range", unless otherwise noted).
— 95th percentile values indicate the breadth of the population (2) of
performance tolerances expected to be met in 95% of the cases with a 95
confidence, for any ambient temperature in the range of 20 to 30°C. In
addition to the statistical observations of a sample of instruments, these
values include the effects of the uncertainties of external calibration
references. These values are not warranted. These values are updated
occasionally if a significant change in the statistically observed behavior of
production instruments is observed.
1
also referred to as "Full
— Typical describes additional product performance information that is not
covered by the product warranty. It is performance beyond specification
that 80 of the units exhibit with a 95 confidence level over the
temperature range 20 to 30°C. Typical performance does not include
measurement uncertainty.
— Nominal values indicate expected performance, or describe product
performance that is useful in the application of the product, but is not
covered by the product warranty.
Conditions Required to Meet Specifications
The following conditions must be met for the analyzer to meet its
specifications.
— The analyzer is within its calibration cycle. See the General section of this
chapter.
— Under auto couple control, except that Auto Sweep Time Rules = Accy.
— For signal frequencies < 10 MHz, DC coupling applied.
— Any analyzer that has been stored at a temperature range inside the
allowed storage range but outside the allowed operating range must be
stored at an ambient temperature within the allowed operating range for at
least two hours before being turned on.
1. For earlier instruments (S/N <MY49430337/SG49430006/US49230450), the operating temperature ranges from 5 to 50°C
20 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Definitions and Requirements
— The analyzer has been turned on at least 30 minutes with Auto Align set to
Normal, or if Auto Align is set to Off or Partial, alignments must have been
run recently enough to prevent an Alert message. If the Alert condition is
changed from “Time and Temperature” to one of the disabled duration
choices, the analyzer may fail to meet specifications without informing the
user.
Certification
Keysight Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Keysight Technologies
further certifies that its calibration measurements are traceable to the
International System of Units (SI) via national metrology institutes
(www.keysight.com/find/NMI) that are signatories to the CIPM Mutual
Recognition Arrangement.
Keysight N9030A PXA Specification Guide 21
PXA Signal Analyzer
Frequency and Time
Frequency and Time
DescriptionSpecificationsSupplemental Information
Frequency Range
Maximum Frequency
Option 5033.6 GHz
Option 5088.4 GHz
Option 51313.6 GHz
Option 52626.5 GHz
Option 54343 GHz
Option 54444 GHz
Option 55050 GHz
Preamp Option P033.6 GHz
Preamp Option P088.4 GHz
Preamp Option P1313.6 GHz
Preamp Option P2626.5 GHz
Preamp Option P4343 GHz
Preamp Option P4444 GHz
Preamp Option P5050 GHz
Minimum Frequency
Preamp
AC Coupled
a
DC Coupled
Off10 MHz3 Hz
On10 MHz9 kHz
BandHarmonic
LO Multiple (N
Mixing Mode
b
)Band Overlaps
c
0 (3 Hz to 3.6 GHz)
d
11Options 503, 508, 513, 526, 543,
544, 550
1 (3.5 to 8.4 GHz)11Options 508, 513, 526, 543, 544,
550
2 (8.3 to 13.6 GHz)12Options 513, 526, 543, 544, 550
22 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
3 (13.5 to 17.1 GHz)22Options 526,543, 544, 550
4 (17.0 to 26.5 GHz)24Options 526, 543, 544, 550
5 (26.4 to 34.5 GHz)24Options 543, 544, 550
6 (34.4 to 50 GHz)48Options 543, 544, 550
a. AC Coupled only applicable to Freq Options 503, 508, 513, and 526.
b. N is the LO multiplication factor. For negative mixing modes (as indicated by the “” in the “Harmonic Mixing
Mode” column), the desired 1st LO harmonic is higher than the tuned frequency by the 1st IF.
c. In the band overlap regions, for example, 3.5 to 3.6 GHz, the analyzer may use either band for measurements, in
this example Band 0 or Band 1. The analyzer gives preference to the band with the better overall specifications
(which is the lower numbered band for all frequencies below 26 GHz), but will choose the other band if doing so is
necessary to achieve a sweep having minimum band crossings. For example, with CF = 3.58 GHz, with a span of
40 MHz or less, the analyzer uses Band 0, because the stop frequency is 3.6 GHz or less, allowing a span without
band crossings in the preferred band. If the span is between 40 and 160 MHz, the analyzer uses Band 1, because
the start frequency is above 3.5 GHz, allowing the sweep to be done without a band crossing in Band 1, though
the stop frequency is above 3.6 GHz, preventing a Band 0 sweep without band crossing. With a span greater than
160 MHz, a band crossing will be required: the analyzer sweeps up to 3.6 GHz in Band 0; then executes a band
crossing and continues the sweep in Band 1.
Specifications are given separately for each band in the band overlap regions. One of these specifications is for the
preferred band, and one for the alternate band. Continuing with the example from the previous paragraph
(3.58 GHz), the preferred band is band 0 (indicated as frequencies under 3.6 GHz) and the alternate band is band
1 (3.5 to 8.4 GHz). The specifications for the preferred band are warranted. The specifications for the alternate
band are not warranted in the band overlap region, but performance is nominally the same as those warranted
specifications in the rest of the band. Again, in this example, consider a signal at 3.58 GHz. If the sweep has been
configured so that the signal at 3.58 GHz is measured in Band 1, the analysis behavior is nominally as stated in the
Band 1 specification line (3.5 to 8.4 GHz) but is not warranted. If warranted performance is necessary for this signal, the sweep should be reconfigured so that analysis occurs in Band 0. Another way to express this situation in
this example Band 0/Band 1 crossing is this: The specifications given in the “Specifications” column which are
described as “3.5 to 8.4 GHz” represent nominal performance from 3.5 to 3.6 GHz, and warranted performance
from 3.6 to 8.4 GHz.
d. Band 0 is extendable (set “Extend Low Band” to On) to 3.7 GHz instead of 3.6 GHz in instruments
with frequency option 508, 513 or 526 and serial number is equal with or greater than the following: US54490126, MY54490690, or SG54490127. Band 0 can also be extendable in earlier instruments with firmware of version A.16.13 or later and a field adjustment to perform that extension
Keysight N9030A PXA Specification Guide 23
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
Precision Frequency Reference
Accuracy[(time since last adjustment
aging rate) + temperature
stability + calibration accuracy
a]b
Temperature Stability
8
8
Nominally linear
10
5 10
20 to 30C
Full temperature range
Aging Rate
1.5 10
5 10
Total Aging
7
1 Year
2 Years
Settability
Warm-up and Retrace
d
300 s after turn on
900 s after turn on
Achievable Initial Calibration Accuracy
e
1 10
1.5 10
2 10
4 10
9
8
7
Nominal
1 10
1 10
7
of final frequency
8
of final frequency
Standby power to reference oscillatorNot supplied
Residual FM
(Center Frequency = 1 GHz
0.25 Hz N
(nominal)
10 Hz RBW, 10 Hz VBW)
c
/day (nominal)
f
p-p in 20 ms
a. Calibration accuracy depends on how accurately the frequency standard was adjusted to 10 MHz. If the adjust-
ment procedure is followed, the calibration accuracy is given by the specification “Achievable Initial Calibration
Accuracy.”
b. The specification applies after the analyzer has been powered on for four hours.
c. Narrow temperature range performance is nominally linear with temperature. For example, for
25±3º C, the stability would be only three-fifths as large as the warranted 25±5º C, thus ±0.9 10
8
.
d. Standby mode does not apply power to the oscillator. Therefore warm-up applies every time the power is
turned on. The warm-up reference is one hour after turning the power on. Retracing also occurs every time
warm-up occurs. The effect of retracing is included within the “Achievable Initial Calibration Accuracy” term of
the Accuracy equation.
24 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Frequency and Time
e. The achievable calibration accuracy at the beginning of the calibration cycle includes these effects:
1) Temperature difference between the calibration environment and the use environment
2) Orientation relative to the gravitation field changing between the calibration environment and the use envi-
ronment
3) Retrace effects in both the calibration environment and the use environment due to turning the instrument
power off.
4) Settability
f. N is the LO multiplication factor.
DescriptionSpecificationsSupplemental Information
Frequency Readout Accuracy(marker freq freq ref accy +
a
+ 2 Hz +
b
)
Example for EMC
0.10span + 5 RBW
0.5 horizontal resolution
d
Single detector only
0.0032 (nominal)
a. The warranted performance is only the sum of all errors under autocoupled conditions. Under non-autocoupled
conditions, the frequency readout accuracy will nominally meet the specification equation, except for conditions
in which the RBW term dominates, as explained in examples below. The nominal RBW contribution to frequency
readout accuracy is 2 of RBW for RBWs from 1 Hz to 390 kHz, 4 of RBW from 430 kHz through 3 MHz (the
widest autocoupled RBW), and 30 of RBW for the (manually selected) 4, 5, 6 and 8 MHz RBWs.
First example: a 120 MHz span, with autocoupled RBW. The autocoupled ratio of span to RBW is 106:1, so
the RBW selected is 1.1 MHz. The 5 RBW term contributes only 55 kHz to the total frequency readout accuracy, compared to 120 kHz for the 0.10 span term, for a total of 175 kHz.
Second example: a 20 MHz span, with a 4 MHz RBW. The specification equation does not apply because the
Span: RBW ratio is not autocoupled. If the equation did apply, it would allow 20 kHz of error (0.10) due to the
span and 200 kHz error (5) due to the RBW. For this non-autocoupled RBW, the RBW error is nominally 30,
or 1200 kHz.
b. Horizontal resolution is due to the marker reading out one of the trace points. The points are spaced by
span/(Npts – 1), where Npts is the number of sweep points. For example, with the factory preset value of 1001
sweep points, the horizontal resolution is span/1000. However, there is an exception: When both the detector
mode is “normal” and the span > 0.25 (Npts – 1) RBW, peaks can occur only in even-numbered points, so
the effective horizontal resolution becomes doubled, or span/500 for the factory preset case. When the RBW is
autocoupled and there are 1001 sweep points, that exception occurs only for spans > 750 MHz
c. Specifications apply to traces in most cases, but there are exceptions. Specifications always apply to the peak
detector. Specifications apply when only one detector is in use and all active traces are set to Clear Write. Specifications also apply when only one detector is in use in all active traces and the "Restart" key has been pressed
since any change from the use of multiple detectors to a single detector. In other cases, such as when multiple
simultaneous detectors are in use, additional errors of 0.5, 1.0 or 1.5 sweep points will occur in some detectors,
depending on the combination of detectors in use.
d. In most cases, the frequency readout accuracy of the analyzer can be exceptionally good. As an example, Key-
sight has characterized the accuracy of a span commonly used for Electro-Magnetic Compatibility (EMC) testing
using a source frequency locked to the analyzer. Ideally, this sweep would include EMC bands C and D and thus
sweep from 30 to 1000 MHz. Ideally, the analysis bandwidth would be 120 kHz at 6 dB, and the spacing of the
points would be half of this (60 kHz). With a start frequency of 30 MHz and a stop frequency of 1000.2 MHz and
a total of 16168 points, the spacing of points is ideal. The detector used was the Peak detector. The accuracy of
frequency readout of all the points tested in this span was with 0.0032 of the span. A perfect analyzer with
this many points would have an accuracy of 0.0031 of span. Thus, even with this large number of display
points, the errors in excess of the bucket quantization limitation were negligible.
a. Instrument conditions: RBW = 1 kHz, gate time = auto (100 ms), S/N 50 dB, frequency = 1 GHz
b. If the signal being measured is locked to the same frequency reference as the analyzer, the specified count
accuracy is 0.100 Hz under the test conditions of footnote a. This error is a noisiness of the result. It will
increase with noisy sources, wider RBWs, lower S/N ratios, and source frequencies > 1 GHz.
DescriptionSpecificationsSupplemental Information
Frequency Span
Range
Option 5030 Hz, 10 Hz to 3.6 GHz
Option 5080 Hz, 10 Hz to 8.4 GHz
Option 5130 Hz, 10 Hz to 13.6 GHz
Option 5260 Hz, 10 Hz to 26.5 GHz
Option 5430 Hz, 10 Hz to 43 GHz
Option 5440 Hz, 10 Hz to 44 GHz
Option 5500 Hz, 10 Hz to 50 GHz
Resolution2 Hz
Span Accuracy
Swept
FFT
(0.1 span + horizontal resolution
(0.1 span + horizontal resolution
a
)
a
)
a. Horizontal resolution is due to the marker reading out one of the sweep points. The points are spaced by
span/(Npts 1), where Npts is the number of sweep points. For example, with the factory preset value of 1001
sweep points, the horizontal resolution is span/1000. However, there is an exception: When both the detector
mode is “normal” and the span > 0.25 (Npts 1) RBW, peaks can occur only in even-numbered points, so
the effective horizontal resolution becomes doubled, or span/500 for the factory preset case. When the RBW is
auto coupled and there are 1001 sweep points, that exception occurs only for spans > 750 MHz.
a. Delayed trigger is available with line, video, RF burst and external triggers.
b. Prior to A.19.28 software, zero span trigger delay was limited to -150 ms to 500 ms.
Keysight N9030A PXA Specification Guide 27
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
TriggersAdditional information on some of the triggers and
gate sources
VideoIndependent of Display Scaling and Reference Level
Minimum settable level170 dBmUseful range limited by noise
Maximum usable level
Highest allowed mixer level
a
+ 2 dB (nominal)
Detector and Sweep Type relationships
Sweep Type = Swept
Detector = Normal, Peak, Sample or
Negative Peak
Triggers on the signal before detection, which is
similar to the displayed signal
Detector = AverageTriggers on the signal before detection, but with a
single-pole filter added to give similar smoothing to
that of the average detector
Sweep Type = FFTTriggers on the signal envelope in a bandwidth
wider than the FFT width
RF Burst
Level Range
40 to 10 dBm plus attenuation (nominal)
b
Level Accuracy
Absolute2 dB + Absolute Amplitude Accuracy (nominal)
a. The highest allowed mixer level depends on the IF Gain. It is nominally –10 dBm for Preamp Off and IF Gain =
Low.
b. Noise will limit trigger level range at high frequencies, such as above 15 GHz.
c. Include RF Burst Level Type = Relative.
Keysight N9030A PXA Specification Guide 29
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
Gated Sweep
Gate MethodsGated LO
Gated Video
Gated FFT
Span RangeAny span
Gate Delay Range0 to 100.0 s
Gate Delay Settability4 digits, 100 ns
Gate Delay Jitter33.3 ns p-p (nominal)
Gate Length Range
(Except Method = FFT)
Gated FFT and Gated Video Frequency
and Amplitude Errors
1 s to 5.0 sGate length for the FFT method is fixed at
1.83/RBW, with nominally 2% tolerance.
Nominally no additional error for gated
measurements when the Gate Delay is greater
than the MIN FAST setting
Gated LO Frequency Errors
Gate 10 sNominally no additional error when the Gate
Delay is greater than the MIN FAST setting
1.0 s Gate 10 sNominal error given by 100 ns N (Span/ST) (SpanPosition ST / GateLength); see footnote
Gated LO Amplitude ErrorsNominally no additional error when the Gate
Delay is greater than the MIN FAST setting
Phase Noise EffectsGated LO method overrides the loop
configuration to force single loop in place of dual
loop.
Gate SourcesExternal 1
Pos or neg edge triggered
External 2
Line
RF Burst
Periodic
a
a. ST is sweep time; SpanPosition is the location of the on-screen signal, 0 being the left edge of the screen and 1
being the right edge. N is the harmonic mixing number.
DescriptionSpecificationsSupplemental Information
Number of Frequency Sweep Points
(buckets)
Factory preset1001
Range1 to 100,001Zero and non-zero spans
30 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Frequency and Time
Nominal Measurement Time vs. Span [Plot]
Keysight N9030A PXA Specification Guide 31
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
Resolution Bandwidth (RBW)
Range (3.01 dB bandwidth)
Standard
1 Hz to 8 MHz
Bandwidths above 3 MHz are 4, 5, 6, and
8 MHz.
Bandwidths 1 Hz to 3 MHz are spaced at
10 spacing using the E24 series (24
per decade): 1.0, 1.1, 1.2, 1.3, 1.5, 1.6,
1.8, 2.0, 2.2, 2.4, 2.7, 3.0, 3.3, 3.6, 3.9,
4.3, 4.7, 5.1, 5.6, 6.2, 6.8, 7.5, 8.2, 9.1 in
each decade.
With Option B85 and Option RBE
a
10, 15, 20, 25, 30, 40, 50, 60, and
70 MHz, in Spectrum Analyzer mode and
zero span.
With Option B1X and Option RBE
a
10, 15, 20, 25, 30, 40, 50, 60, 70, 80,
100, and 133 MHz, in Spectrum Analyzer
mode and zero span.
Power bandwidth accuracy
b
RBW RangeCF Range
1 Hz to 100 kHzAll0.5 (0.022 dB)
110 kHz to 1.0 MHz<3.6 GHz1.0 (0.044 dB)
1.1 to 2.0 MHz<3.6 GHz0.07 dB (nominal)
2.2 to 3 MHz<3.6 GHz0 to0.2dB (nominal)
4 to 8 MHz<3.6 GHz0 to0.4dB (nominal)
Noise BW to RBW ratio
Accuracy (3.01 dB bandwidth)
c
d
1.056 2 (nominal)
1 Hz to 1.3 MHz RBW2 (nominal)
1.5 MHz to 3 MHz RBW
CF 3.6 GHz
CF > 3.6 GHz
7 (nominal)
8 (nominal)
4 MHz to 8 MHz RBW
CF 3.6 GHz
CF > 3.6 GHz
15 (nominal)
20 (nominal)
Selectivity (60 dB/3 dB)4.1:1 (nominal)
a. Option RBE enables wider bandwidth filters in zero span in the Signal Analyzer mode. Available detectors are
Peak+ and Average. VBW filtering is disabled. Minimum sweep time is the greater of 200 μS or 200ns/pt. The
filter shape is approximately square. Support for Average detector was first added in SW Version A.23.05.
32 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Frequency and Time
b. The noise marker, band power marker, channel power and ACP all compute their results using the power band-
width of the RBW used for the measurement. Power bandwidth accuracy is the power uncertainty in the results
of these measurements due only to bandwidth-related errors. (The analyzer knows this power bandwidth for
each RBW with greater accuracy than the RBW width itself, and can therefore achieve lower errors.) The warranted specifications shown apply to the Gaussian RBW filters used in swept and zero span analysis. There are
four different kinds of filters used in the spectrum analyzer: Swept Gaussian, Swept Flattop, FFT Gaussian and
FFT Flattop. While the warranted performance only applies to the swept Gaussian filters, because only they are
kept under statistical process control, the other filters nominally have the same performance.
c. The ratio of the noise bandwidth (also known as the power bandwidth) to the RBW has the nominal value and
tolerance shown. The RBW can also be annotated by its noise bandwidth instead of this 3 dB bandwidth. The
accuracy of this annotated value is similar to that shown in the power bandwidth
accuracy specification.
d. Resolution Bandwidth Accuracy can be observed at slower sweep times than auto-coupled conditions. Normal
sweep rates cause the shape of the RBW filter displayed on the analyzer screen to widen by nominally 6. This
widening declines to 0.6 nominal when the Swp Time Rules key is set to Accuracy instead of Normal. The true
bandwidth, which determines the response to impulsive signals and noise-like signals, is not affected by the
sweep rate.
DescriptionSpecificationSupplemental information
Analysis Bandwidth
a
Standard10 MHz
With Option B25
25 MHz
With Option B4040 MHz
With Option B8585 MHz
With Option B1X
160/140 MHz
b
a. Analysis bandwidth is the instantaneous bandwidth available about a center frequency over which the input sig-
nal can be digitized for further analysis or processing in the time, frequency, or modulation domain.
b. Instruments with 160 MHz bandwidth will display "N9030A-B1Y Analysis bandwidth Extension, 160 MHz" after
pressing: System, Show, System on the instrument.
Keysight N9030A PXA Specification Guide 33
PXA Signal Analyzer
Frequency and Time
DescriptionSpecificationsSupplemental Information
Preselector BandwidthRelevant to many options, such as B1X Wide IF Bandwidth, in
Bands 1 and higher. Nominal.
Mean Bandwidth at CF
a
Freq option Freq option > 526
5 GHz58 MHz46 MHz
10 GHz57 MHz52 MHz
15 GHz59 MHz53 MHz
20 GHz64 MHz55 MHz
25 GHz74 MHz56 MHz
35 GHz62 MHz
44 GHz70 MHz
Standard Deviation9%7%
–3 dB Bandwidth–7.5% relative to –4 dB bandwidth, nominal
a. The preselector can have a significant passband ripple. To avoid ambiguous results, the –4 dB bandwidth is
characterized.
DescriptionSpecificationsSupplemental Information
Video Bandwidth (VBW)
RangeSame as Resolution Bandwidth range
plus wide-open VBW (labeled 50 MHz)
Accuracy6 (nominal)
in swept mode and zero span
a. For FFT processing, the selected VBW is used to determine a number of averages for FFT results. That number is
chosen to give roughly equivalent display smoothing to VBW filtering in a swept measurement. For example, if
VBW = 0.1 RBW, four FFTs are averaged to generate one result.
a
34 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
Measurement Range
Preamp OffDisplayed Average Noise Level to +30 dBm
Preamp On
RF (Option 503)Displayed Average Noise Level to +30 dBm
W (Options 508, 513, 526)Displayed Average Noise Level to +24 dBm
mmW (Options 543, 544, 550)Displayed Average Noise Level to +20 dBm
Input Attenuation Range0 to 70 dB, in 2 dB steps
DescriptionSpecificationsSupplemental Information
Maximum Safe Input LevelApplies with or without preamp
(Options P03, P08, P13, P26, P43, P44,
P50)
Average Total Power+30 dBm (1 W)
Peak Pulse Power
10 s pulse width,
1 duty cycle,
input attenuation 30 dB)
DC voltage
DC Coupled0.2 Vdc
AC Coupled100 Vdc
DescriptionSpecificationsSupplemental Information
Display Range
Log ScaleTen divisions displayed;
Linear ScaleTen divisions
+50 dBm (100 W)
0.1 to 1.0 dB/division in 0.1 dB steps, and
1 to 20 dB/division in 1 dB steps
Keysight N9030A PXA Specification Guide 35
PXA Signal Analyzer
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
Marker Readout
Resolution
Log (decibel) units
Trace Averaging Off, on-screen0.01 dB
Trace Averaging On or remote0.001 dB
Linear units resolution1 of signal level (nominal)
36 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
Frequency Response
DescriptionSpecificationsSupplemental Information
Frequency ResponseRefer to the footnote for
(Maximum error relative to reference
condition (50 MHz)
Mechanical attenuator only
Swept operation
c
b
Attenuation 10 dB)
Option 543, 544, or 550 (mmW)
Option 503, 508, 513, or 526 (RF/W)
20 to 30CFull range 95th Percentile (2)
Band Overlaps on
page 22.
Freq Option 526 only:
Modes above 18 GHz
a
3 Hz to 10 MHzx
10 to 20 MHz
d
10 to 20 MHz
d
20 to 50 MHz
20 to 50 MHz
x0.46 dB0.54 dB
x0.35 dB0.44 dB
x0.46 dB0.54 dB
x0.35 dB0.44 dB0.16 dB
x0.35 dB0.44 dB0.19 dB
50 MHz to 3.6 GHzx0.35 dB0.44 dB0.16 dB
50 MHz to 3.6 GHz
3.6 to 3.7 GHz (Band0)x
3.5 to 5.2 GHz
3.5 to 5.2 GHz
5.2 to 8.4 GHz
5.2 to 8.4 GHz
8.3 to 13.6 GHz
8.3 to 13.6 GHz
fg
fg
fg
fg
fg
fg
x0.35 dB0.47 dB0.15 dB
See note
x1.5 dB2.5 dB0.39 dB
x1.7 dB3.5 dB0.70 dB
x1.5 dB2.5 dB0.39 dB
x1.5 dB2.5 dB0.57 dB
x2.0 dB2.7 dB0.45 dB
x2.0 dB2.5 dB0.54 dB
e
13.5 to 17.1 GHz
13.5 to 17.1 GHz
17.0 to 22 GHz
17.0 to 22 GHz
fg
fg
fg
fg
x2.0 dB2.7 dB0.62 dB
x2.0 dB2.7 dB0.64 dB
x2.0 dB2.7 dB0.62 dB
x2.0 dB2.8 dB0.72 dB
Keysight N9030A PXA Specification Guide 37
PXA Signal Analyzer
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
22.0 to 26.5 GHz
22.0 to 26.5 GHz
26.4 to 34.5 GHz
34.4 to 50 GHz
fg
fg
fg
fg
x2.5 dB3.7 dB0.82 dB
x2.5 dB3.5 dB0.71 dB
x2.5 dB3.5 dB0.93 dB
x3.2 dB4.9 dB1.24 dB
a. Signal frequencies above 18 GHz are prone to additional response errors due to modes in the Type-N connector
used. Only analyzers with frequency Option 526 that do not also have input connector Option C35 will have
these modes.With the use of Type-N to APC 3.5 mm adapter part number 1250-1744, there are nominally six
such modes. The effect of these modes with this connector are included within these specifications.
b. See the Electronic Attenuator (Option EA3) chapter for Frequency Response using the electronic attenuator.
c. For Sweep Type = FFT, add the RF flatness errors of this table to the IF Frequency Response errors. An additional
error source, the error in switching between swept and FFT sweep types, is nominally 0.01 dB and is included
within the “Absolute Amplitude Error” specifications.
d. Specifications apply with DC coupling at all frequencies. With AC coupling, specifications apply at frequencies of
50 MHz and higher. Statistical observations at 10 MHz and lower show that most instruments meet the specifica-
tion, but a few percent of instruments can be expected to have errors that, while within the specified limits, are
closer to those limits than the measurement uncertainty guardband, and thus are not warranted. The AC coupling
effect at 20 to 50 MHz is negligible, but not warranted.
e. Band 0 is extendable (set “Extend Low Band” to On) to 3.7 GHz instead of 3.6 GHz in instruments
with frequency option 508, 513 or 526 and serial number is equal with or greater than the follow-
ing: US54490126, MY54490690, or SG54490127. Band 0 can also be extendable in earlier instru-
ments with firmware of version A.16.13 or later and a field adjustment to perform that extension.
Subject to these conditions, statistical observations show that performance nominally fits within
the same range within the 3.6 to 3.7 GHz frequencies as within the next lower specified frequency
range, but is not warranted.
f. Specifications for frequencies > 3.5 GHz apply for sweep rates 100 MHz/ms.
g. Preselector centering applied.
38 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
Nominal Frequency Response Band 0 [Plot]
DescriptionSpecificationsSupplemental Information
IF Frequency Response
a
Freq Option 526 only:Modes above 18 GHz
b
(Demodulation and FFT
response relative to the
center frequency)
Center
Freq (GHz)
(MHz)
Preselector
Max Errord Midwidth Error
(95th
Percentile)
Slope (dB/MHz)
(95th
Percentile)
RMSe
(nominal)
c
Span
3.6100.20 dB0.12 dB0.100.02 dB
3.6, 26.510 On0.23 dB
f
3.6, 26.510
Off
0.25 dB0.12 dB0.100.02 dB
26.5, 5010 On0.12 dB
f
, 5010
Off
0.30 dB±0.12 dB0.100.024 dB
a. The IF frequency response includes effects due to RF circuits such as input filters, that are a function of RF fre-
quency, in addition to the IF passband effects.
Keysight N9030A PXA Specification Guide 39
PXA Signal Analyzer
Amplitude Accuracy and Range
b. Signal frequencies above 18 GHz are prone to additional response errors due to modes in the Type-N connector
used. Only analyzers with frequency Option 526 that do not also have input connector Option C35 will have
these modes.With the use of Type-N to APC 3.5 mm adapter part number 1250-1744, there are nominally six
such modes. These modes cause nominally up to 0.35 dB amplitude change, with phase errors of nominally up
to 1.2°.
c. This column applies to the instantaneous analysis bandwidth in use. In the Spectrum Analyzer Mode, this would
be the FFT width.
d. The maximum error at an offset (f) from the center of the FFT width is given by the expression
[Midwidth Error + (f Slope)], but never exceeds Max Error. Here the Midwidth Error is the error at the center
frequency for a given FFT span. Usually, the span is no larger than the FFT width in which case the center of the
FFT width is the center frequency of the analyzer. When using the Spectrum Analyzer mode with an analyzer span
is wider than the FFT width, the span is made up of multiple concatenated FFT results, and thus has multiple centers of FFT widths; in this case the f in the equation is the offset from the nearest center. Performance is nominally
three times better at most center frequencies.
e. The “rms” nominal performance is the standard deviation of the response relative to the center frequency, inte-
grated across the span. This performance measure was observed at a center frequency in each harmonic mixing
band, which is representative of all center frequencies; it is not the worst case frequency.
f. Option MPB is installed and enabled.
DescriptionSpecificationsSupplemental Information
IF Phase LinearityDeviation from mean phase linearity
Freq Option 526 only:Modes above
a
18 GHz
Center Freq
(GHz)
Span
(MHz)
PreselectorPeak-to-peak
(nominal)
RMS (nominal)
0.02, 3.610n/a0.06°0.012°
3.6, 26.510
Off
0.10°0.022°
c
3.610On0.11°0.024°
a. Signal frequencies above 18 GHz are prone to additional response errors due to modes in the Type-N connector
used. Only analyzers with frequency Option 526 that do not also have input connector Option C35 will have
these modes.With the use of Type-N to APC 3.5 mm adapter part number 1250-1744, there are nominally six
such modes. These modes cause nominally up to 0.35 dB amplitude change, with phase errors of nominally up
to 1.2°.
b. The listed performance is the standard deviation of the phase deviation relative to the mean phase deviation
from a linear phase condition, where the rms is computed across the span shown and over the range of center
frequencies shown.
c. Option MPB is installed and enabled.
b
40 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
Absolute Amplitude Accuracy
At 50 MHz
20 to 30C
Full temperature range
At all frequencies
20 to 30C
Full temperature range
95th Percentile Absolute
Amplitude Accuracy
a
0.24 dB
0.13 dB
(95th percentile)
0.28 dB
a
(0.24 dB + frequency response)
(0.28 dB + frequency response)
0.19 dB
b
(Wide range of signal levels,
RBWs, RLs, etc.,
0.01 to 3.6 GHz,
Atten = 10 dB)
Amplitude Reference Accuracy0.05 dB (nominal)
Preamp On
c
(0.36 dB + frequency response)
(P03, P08, P13, P26, P43, P44,
P50)
a. Absolute amplitude accuracy is the total of all amplitude measurement errors, and applies over the following
subset of settings and conditions: 1 Hz RBW 1MHz; Input signal 10 to 50 dBm (details below); Input
attenuation 10 dB; span < 5 MHz (nominal additional error for span 5 MHz is 0.02 dB); all settings auto-coupled except Swp Time Rules = Accuracy; combinations of low signal level and wide RBW use VBW 30 kHz to
reduce noise. When using FFT sweeps, the signal must be at the center frequency.
This absolute amplitude accuracy specification includes the sum of the following individual specifications under
the conditions listed above: Scale Fidelity, Reference Level Accuracy, Display Scale Switching Uncertainty, Resolution Bandwidth Switching Uncertainty, 50 MHz Amplitude Reference Accuracy, and the accuracy with which
the instrument aligns its internal gains to the 50 MHz Amplitude Reference.
The only difference between signals within the range above –50 dBm and those signals below that level is the
scale fidelity. Our specifications and experience show no difference between signals above and below this level.
The only reason our Absolute Amplitude Uncertainty specification does not go below this level is that noise
detracts from our ability to verify the performance at all levels with acceptable test times and yields. So the performance is not warranted at lower levels, but we fully expect it to be the same.
Keysight N9030A PXA Specification Guide 41
PXA Signal Analyzer
Amplitude Accuracy and Range
b. Absolute Amplitude Accuracy for a wide range of signal and measurement settings, covers the 95th percentile
proportion with 95 confidence. Here are the details of what is covered and how the computation is made:
The wide range of conditions of RBW, signal level, VBW, reference level and display scale are discussed in footnote a. There are 44 quasi-random combinations used, tested at a 50 MHz signal frequency. We compute the
95th percentile proportion with 95 confidence for this set observed over a statistically significant number of
instruments. Also, the frequency response relative to the 50 MHz response is characterized by varying the signal
across a large number of quasi-random verification frequencies that are chosen to not correspond with the frequency response adjustment frequencies. We again compute the 95th percentile proportion with 95 confidence for this set observed over a statistically significant number of instruments. We also compute the 95th
percentile accuracy of tracing the calibration of the 50 MHz absolute amplitude accuracy to a national standards organization. We also compute the 95th percentile accuracy of tracing the calibration of the relative frequency response to a national standards organization. We take the root-sum-square of these four independent
Gaussian parameters. To that rss we add the environmental effects of temperature variations across the 20 to
30C range. These computations and measurements are made with the mechanical attenuator only in circuit,
set to the reference state of 10 dB.
A similar process is used for computing the result when using the electronic attenuator under a wide range of
settings: all even settings from 4 through 24 dB inclusive, with the mechanical attenuator set to 10 dB. The 95th
percentile result was 0.21 dB.
c. Same settings as footnote a, except that the signal level at the preamp input is 40 to 80 dBm. Total power at
preamp (dBm) = total power at input (dBm) minus input attenuation (dB). This specification applies for signal
frequencies above 100 kHz.
DescriptionSpecificationsSupplemental Information
Input Attenuation Switching UncertaintyRefer to the footnote for
Band Overlaps on page 22
(Relative to 10 dB (reference setting))
50 MHz (reference frequency), preamp off
Attenuation 12 to 40 dB0.14 dB0.03 dB (typical)
Attenuation 2 to 8 dB, or >40 dB0.18 dB0.05 dB (typical)
Attenuation 0 dB0.05 dB (nominal)
Attenuation > 2 dB, preamp off
3 Hz to 3.6 GHz0.3 dB (nominal)
3.5 to 8.4 GHz0.5 dB (nominal)
8.3 to 13.6 GHz0.7 dB (nominal)
13.5 to 26.5 GHz0.7 dB (nominal)
26.5 to 50.0 GHz1.0 dB (nominal)
42 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
RF Input VSWR
(at tuned frequency, DC coupled)
mmW (Option 543, 544, or 550)
RF/W (Option 503, 508, 513, or 526)
10 dB atten, 50 MHz (ref condition) x1.07:1 (nominal)
10 dB atten, 50 MHz (ref condition)
0 dB atten, 0.01 to 3.6 GHzx
x1.025:1 (nominal)
x<2.2:1 (nominal)
95th Percentile
Band 0 (0.01 to 3.6 GHz, 10 dB atten)x1.139
Band 0 (0.01 to 3.6 GHz, 10 dB atten)
x1.134
Band 1 (3.5 to 8.4 GHz, 10 dB atten)x1.290
Band 1 (3.5 to 8.4 GHz, 10 dB atten)
x1.152
Band 2 (8.3 to 13.6 GHz, 10 dB atten)x1.388
Band 2 (8.3 to 13.6 GHz, 10 dB atten)
x1.178
Band 3 (13.5 to 17.1 GHz, 10 dB atten)x1.403
Band 3 (13.5 to 17.1 GHz, 10 dB atten)
x1.204
Band 4 (17.0 to 26.5 GHz, 10 dB atten)x1.475
Band 4 (17.0 to 26.5 GHz, 10 dB atten)
Band 5 (26.4 to 34.5 GHz, 10 dB atten)
Band 6 (34.4 to 50 GHz, 10 dB atten)
x1.331
x1.321
x1.378
a
Nominal VSWR vs. Freq, 10 dBSee plots following
Atten > 10 dBSimilar to atten = 10 dB
RF Calibrator (e.g. 50 MHz) is OnOpen input
Alignments runningOpen input for some, unless "All but
RF" is selected
Preselector centeringOpen input
a. X-Series analyzers have a reflection coefficient that is excellently modeled with a Rayleigh probability distribu-
tion. Keysight recommends using the methods outlined in Application Note 1449-3 and companion Average
Power Sensor Measurement Uncertainty Calculator to compute mismatch uncertainty. Use this 95th percentile
VSWR information and the Rayleigh model (Case C or E in the application note) with that process.
Keysight N9030A PXA Specification Guide 43
PXA Signal Analyzer
Amplitude Accuracy and Range
Nominal VSWR Low Band [Plot]
44 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Amplitude Accuracy and Range
Nominal VSWR, above 3.5 GHz [Plot]
Keysight N9030A PXA Specification Guide 45
PXA Signal Analyzer
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
Resolution Bandwidth Switching UncertaintyRelative to reference BW of 30 kHz,
1.0 Hz to 1.5 MHz RBW0.03 dB
verified in low band
a
1.6 MHz to 2.7 MHz RBW0.05 dB
3.0 MHz RBW0.10 dB
Manually selected wide RBWs: 4, 5, 6, 8 MHz0.30 dB
a. RBW switching uncertainty is verified at 50 MHz. It is consistent for all measurements made without the prese-
lector, thus in Band 0 and also in higher bands with the Preselector Bypass option. In preselected bands, the
slope of the preselector passband can interact with the RBW shape to make an apparent additional RBW
switching uncertainty of nominally ±0.05 dB/MHz times the RBW.
DescriptionSpecificationsSupplemental Information
Reference Level
Range
Log Units 170 to +30 dBm, in 0.01 dB steps
Linear Units 707 pV to 7.07 V, with 0.01 dB resolution (0.11)
Accuracy
0 dB
a
a. Because reference level affects only the display, not the measurement, it causes no additional error in measure-
ment results from trace data or markers.
DescriptionSpecificationsSupplemental Information
Display Scale Switching Uncertainty
Switching between Linear and Log
Log Scale Switching
0 dB
0 dB
a
a
a. Because Log/Lin and Log Scale Switching affect only the display, not the measurement, they cause no addi-
tional error in measurement results from trace data or markers.
46 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
3
320dB110
SN3dB+20dB–
+log=
Amplitude Accuracy and Range
DescriptionSpecificationsSupplemental Information
Display Scale Fidelity
ab
Absolute Log-Linear Fidelity
(Relative to the reference condition: 25 dBm
input through 10 dB attenuation, thus
35 dBm at the input mixer)
Input mixer level
c
Linearity
Typ ical
18 dBm ML 10 dBm0.10 dB0.04 dB
ML < 18 dBm0.07 dB0.02 dB
Relative Fidelity
d
Applies for mixer levelc range from 10 to
80 dBm, mechanical attenuator only,
preamp off, and dither on.
Sum of the following terms:Nominal
high level term
instability term
slope term
prefilter term
Up to 0.015 dB
0.0019 dBrms
From equation
Up to 0.005 dB
e
f
g
h
a. Supplemental information: The amplitude detection linearity specification applies at all levels below 10 dBm at
the input mixer; however, noise will reduce the accuracy of low level measurements. The amplitude error due to
noise is determined by the signal-to-noise ratio, S/N. If the S/N is large (20 dB or better), the amplitude error
due to noise can be estimated from the equation below, given for the 3-sigma (three standard deviations) level.
The errors due to S/N ratio can be further reduced by averaging results. For large S/N (20 dB or better), the
3-sigma level can be reduced proportional to the square root of the number of averages taken.
b. The scale fidelity is warranted with ADC dither set to Medium. Dither increases the noise level by nominally only
0.28 dB for the most sensitive case (preamp Off, best DANL frequencies). With dither Off, scale fidelity for low
level signals, around 60 dBm or lower, will nominally degrade by 0.2 dB. Dither High will give exceptional linear relative scale fidelity, but increase DANL by 0.63 dB instead of 0.28 dB.
c. Mixer level = Input Level Input Attenuation
d. The relative fidelity is the error in the measured difference between two signal levels. It is so small in many cases
that it cannot be verified without being dominated by measurement uncertainty of the verification. Because of
this verification difficulty, this specification gives nominal performance, based on numbers that are as conservatively determined as those used in warranted specifications. We will consider one example of the use of the error
equation to compute the nominal performance.
Example: the accuracy of the relative level of a sideband around 60 dBm, with a carrier at 5dBm, using attenuation = 10 dB, RBW = 3 kHz, evaluated with swept analysis. The high level term is evaluated with P1 =
15 dBm and P2 = 70 dBm at the mixer. This gives a maximum error within 0.008 dB. The instability term is
0.0019 dB if the measurement is completed within a minute. The slope term evaluates to 0.022 dB. The pre-filter term applies and evaluates to the limit of 0.005 dB. The sum of all these terms is 0.037 dB.
Keysight N9030A PXA Specification Guide 47
PXA Signal Analyzer
Amplitude Accuracy and Range
e. Errors at high mixer levels will nominally be well within the range of 0.015 dB × {exp[(P1 Pref)/(8.69 dB)]
x
exp[(P2 Pref)/(8.69 dB)]} (exp is the natural exponent function, e
). In this expression, P1 and P2 are the pow-
ers of the two signals, in decibel units, whose relative power is being measured. Pref is 10 dBm (10 dBm is
the highest power for which linearity is specified). All these levels are referred to the mixer level.
f. The stability of the analyzer gain can be an error term of importance when no settings have changed. These have
been studied carefully in the PXA. One source of instability is the variation in analyzer response with time when
fully warmed up in a stable lab environment. This has been observed to be well modeled as a random walk process, where the difference in two measurements spaced by time t is given by a sqrt(t), where a is
0.0019 dBrms per root minute. The other source of instability is updated alignments from running full or partial
alignments in the background or invoking an alignment. Invoked alignments (Align Now, All) have a standard
deviation of 0.0018 dB, and performing these will restart the random walk behavior. Partial alignments (Auto
Align set to "Partial") have a standard deviation that is, coincidentally, also 0.0018 dBrms, and only occurs once
every ten minutes. The standard deviation from full background alignment (Auto Align set to "Normal") is 0.015
dBrms; with these alignments on, there is no additional random walk behavior. (Keysight recommends setting
alignments (Auto Align) to Normal in order to make the best measurements over long periods of time or in environments without very high temperature stability. For short term measurements in highly stable environments,
setting alignments to Partial can give the best stability. Setting Alignments to Off is not recommended where
stability matters.)
g. Slope error will nominally be well within the range of 0.0004 × (P1 P2). P1 and P2 are defined in footnote e.
h. A small additional error is possible. In FFT sweeps, this error is possible for spans under 4.01 kHz. For non-FFT
measurements, it is possible for RBWs of 3.9 kHz or less. The error is well within the range of ±0.0021 × (P1 P2) subject to a maximum of ±0.005 dB. (The maximum dominates for all but very small differences.) P1 and P2
are defined in footnote e.
DescriptionSpecificationsSupplemental Information
Available DetectorsNormal, Peak, Sample, Negative Peak,
Average
Average detector works on RMS,
Voltage and Logarithmic scales
48 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
Dynamic Range
Gain Compression
DescriptionSpecificationsSupplemental Information
1 dB Gain Compression Point (Two-tone)
20 to 40 MHz3 dBm4 dBm0 dBm (typical)
40 to 200 MHz+1 dBm1 dBm+3 dBm (typical)
200 MHz to 3.6 GHz+3 dBm+2 dBm+5 dBm (typical)
3.6 to 16 GHz+1 dBm0 dBm+4 dBm (typical)
16 to 26.5 GHz1 dBm3 dBm+2 dBm (typical)
26.5 to 50 GHz0 dBm (nominal)
Clipping (ADC Over-range)
Any signal offset10 dBm
Signal offset > 5 times IF prefilter bandwidth and IF
Gain set to Low
IF Prefilter Bandwidth
abc
Maximum power at mixer
20 to 30CFull range
d
Low frequency exceptions
+12 dBm (nominal)
e
Zero Span orSweep Type = FFT,–3 dB Bandwidth
f
Swept
Keysight N9030A PXA Specification Guide 49
, RBW =
3.9 kHz<4.01 kHz8.9 kHz
4.3 to 27 kHz<28.81 kHz79 kHz
30 to 160 kHz<167.4 kHz303 kHz
180 to 390 kHz<411.9 kHz966 kHz
430 kHz to 8 MHz<7.99 MHz10.9 MHz
a. Large signals, even at frequencies not shown on the screen, can cause the analyzer to incorrectly measure
on-screen signals because of two-tone gain compression. This specification tells how large an interfering signal
must be in order to cause a 1 dB change in an on-screen signal.
b. Specified at 1 kHz RBW with 100 kHz tone spacing. The compression point will nominally equal the specification
for tone spacing greater than 5 times the prefilter bandwidth. At smaller spacings, ADC clipping may occur at a
level lower than the 1 dB compression point.
FFT Width =(nominal)
PXA Signal Analyzer
Dynamic Range
c. Reference level and off-screen performance: The reference level (RL) behavior differs from some earlier analyz-
ers in a way that makes this analyzer more flexible. In other analyzers, the RL controlled how the measurement
was performed as well as how it was displayed. Because the logarithmic amplifier in these analyzers had both
range and resolution limitations, this behavior was necessary for optimum measurement accuracy. The logarithmic amplifier in this signal analyzer, however, is implemented digitally such that the range and resolution greatly
exceed other instrument limitations. Because of this, the analyzer can make measurements largely independent
of the setting of the RL without compromising accuracy. Because the RL becomes a display function, not a measurement function, a marker can read out results that are off-screen, either above or below, without any change
in accuracy. The only exception to the independence of RL and the way in which the measurement is performed
is in the input attenuation setting: When the input attenuation is set to auto, the rules for the determination of
the input attenuation include dependence on the reference level. Because the input attenuation setting controls
the tradeoff between large signal behaviors (third-order intermodulation, compression, and display scale fidelity) and small signal effects (noise), the measurement results can change with RL changes when the input attenuation is set to auto.
d. Mixer power level (dBm) = input power (dBm) input attenuation (dB).
e. The ADC clipping level declines at low frequencies (below 50 MHz) when the LO feedthrough (the signal that
appears at 0 Hz) is within 5 times the prefilter bandwidth (see table) and must be handled by the ADC. For
example, with a 300 kHz RBW and prefilter bandwidth at 966 kHz, the clipping level reduces for signal frequencies below 4.83 MHz. For signal frequencies below 2.5 times the prefilter bandwidth, there will be additional
reduction due to the presence of the image signal (the signal that appears at the negative of the input signal frequency) at the ADC.
f. This table applies without Option FS1, fast sweep. With Option FS1, this table applies for sweep rates that
are manually chosen to be the same as or slower than "traditional" sweep rates, instead of the much faster
sweep rates, such as autocoupled sweep rates, available with FS1. Sweep rate is defined to be span divided by
sweep time. If the sweep rate is 1.1 times RBW-squared, the table applies. Otherwise, compute an "effective
RBW" = Span / (SweepTime × RBW). To determine the IF Prefilter Bandwidth, look up this effective RBW in the
table instead of the actual RBW. For example, for RBW = 3 kHz, Span = 300 kHz, and Sweep time = 42 ms, we
compute that Sweep Rate = 7.1 MHz/s, while RBW-squared is 9 MHz/s. So the Sweep Rate is < 1.1 times
RBW-squared and the table applies; row 1 shows the IF Prefilter Bandwidth is nominally 8.9 kHz. If the sweep
time is 1 ms, then the effective RBW computes to 100 kHz. This would result in an IF Prefilter Bandwidth from
the third row, nominally 303 kHz.
50 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
Displayed Average Noise Level
DescriptionSpecificationsSupplemental
Information
Displayed Average Noise Level
(DANL) (without Noise Floor
Extension)
a
Input terminated
Sample or Average detector
Averaging type = Log
0 dB input attenuation
Refer to the footnote
for
Band Overlaps on
page 22.
IF Gain = High
1 Hz Resolution Bandwidth
mmW with no signal path options
mmW with one or more signal path options
b
c
RF/W (Option 503, 508, 513, or 526)
20 to 30CFull range Typical
3 to 10 Hz
10 to 100 Hz
100 Hz to 1 kHz
1 to 9 kHz
9 to 100 kHz
100 kHz to 1 MHz
xxx–100 dBm (nominal)
xxx–125 dBm (nominal)
xxx–130 dBm (nominal)
xxx–145 dBm (nominal)
xxx146 dBm146 dBm151 dBm
xxx150 dBm150 dBm156 dBm
1 to 10 MHz
10 MHz to 1.2 GHz
1.2 to 2.1 GHz
2.1 to 3 GHz
3.0 to 3.6 GHz
3.6 to 3.7
d
xxx155 dBm152 dBm158 dBm
xxx154 dBm152 dBm155 dBm
xxx153 dBm152 dBm155 dBm
xxx151 dBm150 dBm153 dBm
xxx151 dBm149 dBm153 dBm
x
See note
3.5 to 4.2 GHzx147 dBm146 dBm150 dBm
3.5 to 4.2 GHz
3.5 to 4.2 GHz
4.2 to 8.4 GHz
4.2 to 6.6 GHz
4.2 to 6.6 GHz
6.6 to 8.4 GHz
x143 dBm141 dBm147 dBm
x–145 dBm –143 dBm–148 dBm
x150 dBm148 dBm152 dBm
x144 dBm142 dBm148 dBm
x–146 dBm–144 dBm–149 dBm
x147 dBm145 dBm149 dBm
e
Keysight N9030A PXA Specification Guide 51
PXA Signal Analyzer
Dynamic Range
DescriptionSpecificationsSupplemental
Information
6.6 to 8.4 GHz
8.3 to 13.6 GHz
8.3 to 13.6 GHz
8.3 to 13.6 GHz
13.5 to 16.9 GHz
13.5 to 14 GHz
13.5 to 14 GHz
14 to 17 GHz
14 to 17 GHz
16.9 to 20 GHz
17.0 to 22.5 GHz
17.0 to 22.5 GHz
20.0 to 26.5 GHz
22.5 to 26.5 GHz
22.5 to 26.5 GHz
x–149 dBm–147 dBm–151 dBm
x149 dBm147 dBm151 dBm
x147 dBm145 dBm149 dBm
x–149 dBm–147 dBm–151 dBm
x145 dBm143 dBm147 dBm
x143 dBm141 dBm146 dBm
x145 dBm143 dBm148 dBm
x145 dBm143 dBm148 dBm
x147 dBm145 dBm150 dBm
x143 dBm140 dBm145 dBm
x141 dBm139 dBm146 dBm
x145 dBm143 dBm148 dBm
x137 dBm135 dBm140 dBm
x139 dBm137 dBm143 dBm
x142 dBm140 dBm145 dBm
26.4 to 30 GHz
26.4 to 30 GHz
30 to 34 GHz
30 to 34 GHz
33.9 to 37 GHz
33.9 to 37 GHz
37 to 40 GHz
37 to 40 GHz
40 to 49 GHz
40 to 46 GHz
46 to 49 GHz
49 to 50 GHz
49 to 50 GHz
Additional DANL, IF Gain = Low
x138 dBm136 dBm142 dBm
x141 dBm139 dBm145 dBm
x138 dBm135 dBm142 dBm
x141 dBm138 dBm144 dBm
x134 dBm131 dBm139 dBm
x137 dBm133 dBm142 dBm
x132 dBm129 dBm138 dBm
x136 dBm133 dBm141 dBm
x130 dBm126 dBm135 dBm
x136 dBm132 dBm140 dBm
x133 dBm129 dBm138 dBm
x128 dBm124 dBm133 dBm
x133 dBm129 dBm137 dBm
f
xxx–164.5 dBm (nominal)
52 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
a. DANL for zero span and swept is measured in a 1 kHz RBW and normalized to the narrowest available RBW,
because the noise figure does not depend on RBW and 1 kHz measurements are faster.
b. Specifications marked with an x in this column apply to analyzers with mmW frequency options (Option 543,
544 or 550) and none of the following options that affect the signal path: MPB, LNP, B85, or B1X.
c. Specifications marked with an x in this column apply to analyzers with mmW frequency options (Option 543,
544 or 550) and one or more of the following options that affect the signal path: MPB, LNP, B85, or B1X.
d. DANL below 10 MHz is affected by phase noise around the LO feedthrough signal. Specifications apply with the
best setting of the Phase Noise Optimization control, which is to choose the “Best Close-in Noise" for frequencies below about 150 kHz, and “Best Wide Offset Noise" for frequencies above about 150 kHz.
e. Band 0 is extendable (set “Extend Low Band” to On) to 3.7 GHz instead of 3.6 GHz in instruments
with frequency option 508, 513 or 526 and serial number is equal with or greater than the following: US54490126, MY54490690, or SG54490127. Band 0 can also be extendable in earlier
instruments with firmware of version A.16.13 or later and a field adjustment to perform that
extension. Subject to these conditions, statistical observations show that performance nominally
fits within the same range within the 3.6 to 3.7 GHz frequencies as within the next lower specified frequency range, but is not warranted.
f. Setting the IF Gain to Low is often desirable in order to allow higher power into the mixer without overload, bet-
ter compression and better third-order intermodulation. When the Swept IF Gain is set to Low, either by auto
coupling or manual coupling, there is noise added above that specified in this table for the IF Gain = High case.
That excess noise appears as an additional noise at the input mixer. This level has sub-decibel dependence on
center frequency. To find the total displayed average noise at the mixer for Swept IF Gain = Low, sum the powers
of the DANL for IF Gain = High with this additional DANL. To do that summation, compute DANLtotal = 10 log
(10(DANLhigh/10) + 10(AdditionalDANL / 10)). In FFT sweeps, the same behavior occurs, except that FFT IF Gain
can be set to autorange, where it varies with the input signal level, in addition to forced High and Low settings.
Keysight N9030A PXA Specification Guide 53
PXA Signal Analyzer
Dynamic Range
DescriptionSpecificationsSupplemental
Information
Displayed Average Noise Level with Noise Floor
Extension Improvement
a
mmW with no signal path options
c
Most recent
improvements
b
mmW with one or more signal path optionsd 95th Percentile (2
RF/W (Option 503, 508, 513,or 526)Preamp
Off
Band 0, f > 20 MHz
Band 0, f > 20 MHz
Band 0, f > 20 MHz
g
g
g
x9 dB10 dB
x10 dB9 dB
x10 dB9 dB
Preamp
f
On
Band 1x10 dB9 dB
Band 1
Band 1
x9 dB9 dB
x10 dB9 dB
Band 2x10 dB10 dB
Band 2
Band 2
x9 dB8 dB
x9 dB8 dB
e
Band 3x9 dB10 dB
Band 3
Band 3
x9 dB8 dB
x10 dB8 dB
Band 4x10 dB8 dB
Band 4
Band 4
Band 5
Band 5
Band 6
Band 6
Improvement for CW Signals
h
Improvement, Pulsed-RF Signals
i
x10 dB9 dB
x11 dB9 dB
x11 dB8dB
x12 dB9 dB
x11 dB7 dB
x12 dB8 dB
3.5 dB (nominal)
10.8 dB (nominal)
Improvement, Noise-Like Signals9.1 dB (nominal)
54 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
a. This statement on the improvement in DANL is based on the statistical observations of the error in the effective
noise floor after NFE is applied. That effective noise floor can be a negative or a positive power at any frequency.
These 95th percentile values are based on the absolute value of that effective remainder noise power.
b. The specifications shown in this table apply to the instruments with N9030A-NF2 (Noise floor extensions, instru-
ment alignment). The N9030A-NF2 ships standard with all PXAs beginning approximately January 2015. The earlier PXAs that shipped standard with NFE (Noise floor extension, factory calibration) can be upgraded with Option N9030AK-NF2 to achieve performance indicated here. Press System> Show> System to ensure the
N9030A-NF2 is licensed. If the N9030A-NF2 is not licensed, then refer to the chapter of Version Dependent Backdated Specifications at the end of this book for the DANL with NFE specifications.
c. Specifications marked with an x in this column apply to analyzers with mmW frequency options (Option543, 544
or 550) and none of the following options that affect the signal path: MPB, LNP, B85, or B1X.
d. Specifications marked with an x in this column apply to analyzers with mmW frequency options (Option543, 544
or 550) and one or more of the following options that affect the signal path: MPB, LNP, B85, or B1X.
e. Unlike other 95th percentiles, these table values do not include delta environment effects. NFE is aligned in the
factory at room temperature. For best performance, in an environment that is different from room temperature,
such as an equipment rack with other instruments, we recommend running the "Characterize Noise Floor"
operation after the first time the analyzer has been installed in the environment, and given an hour to stabilize.
f. DANL of the preamp is specified with a 50 source impedance. Like all amplifiers, the noise varies with the source
impedance. When NFE compensates for the noise with an ideal source impedance, the variation in the remaining
noise level with the actual source impedance is greatly multiplied in a decibel sense.
g. NFE does not apply to the low frequency sensitivity. At frequencies below about 0.5 MHz, the sensitivity is domi-
nated by phase noise surrounding the LO feedthrough. The NFE is not designed to improve that performance. At
frequencies between 0.5 and 20 MHz the NFE effectiveness increases from nearly none to near its maximum.
h. Improvement in the uncertainty of measurement due to amplitude errors and variance of the results is modestly
improved by using NFE. The nominal improvement shown was evaluated for a 2 dB error with 250 traces averaged. For extreme numbers of averages, the result will be as shown in the "Improvement for Noise-like Signals"
and DANL sections of this table.
i. Pulsed-RF signals are usually measured with peak detection. Often, they are also measured with many “max hold”
traces. When the measurement time in each display point is long compared to the reciprocal of the RBW, or the
number of traces max held is large, considerable variance reduction occurs in each measurement point. When the
variance reduction is large, NFE can be quite effective; when it is small, NFE has low effectiveness. For example, in
Band 0 with 100 pulses per trace element, in order to keep the error within 3 dB error 95% of the time, the signal
can be 10.8 dB lower with NFE than without NFE.
Keysight N9030A PXA Specification Guide 55
PXA Signal Analyzer
Dynamic Range
DescriptionSpecificationsSupplemental Information
Displayed Average Noise Level with Noise Floor Extension
mmW with no signal path options
mmW with one or more signal path options
c
d
RF/W (Option 503, 508, 513, or 526)
Band 0, f 20 MHz
g
xxx163 dBm174 dBm
a
95th Percentile (2b
Preamp Off
Preamp On
Band 1 x162 dBm174 dBm
Band 1
Band 1
x159 dBm172 dBm
x160 dBm172 dBm
Band 2 x162 dBm173 dBm
Band 2
Band 2
x159 dBm172 dBm
x161 dBm173 dBm
Band 3 x156 dBm172 dBm
Band 3
x159 dBm173 dBm
c
Band 3
x161 dBm174 dBm
Band 4 x150 dBm166 dBm
Band 4
Band 4
Band 5
Band 5
Band 6
Band 6
x154 dBm169 dBm
x158 dBm171 dBm
x153 dBm167 dBm
x157 dBm168 dBm
x144 dBm158 dBm
x149 dBm161 dBm
a. DANL with NFE is unlike DANL without NFE. It is based on the statistical observations of the error in the effective
noise floor after NFE is applied. That effective noise floor can be a negative or a positive power at any frequency.
These 95th percentile values are based on the absolute value of that effective remainder noise power.
b. Unlike other 95th percentiles, these table values do not include delta environment effects. NFE is aligned in the
factory at room temperature. For best performance, in an environment that is different from room temperature,
such as an equipment rack with other instruments, we recommend running the "Characterize Noise Floor"
operation after the first time the analyzer has been installed in the environment, and given an hour to stabilize.
56 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
c. NFE performance can give results below theoretical levels of noise in a termination resistor at room temperature,
about –174 dBm/Hz. this is intentional and usually desirable. NFE is not designed to report the noise at the input
of the analyzer; it reports how much more noise is at the input of the analyzer than was present in its alignment.
And its alignment includes the noise of a termination at room temperature. So it can often see the added noise
below the theoretical noise. Furthermore, DANL is defined with log averaging in a 1 Hz RBW, which is about 2.3 dB
lower than the noise density (power averaged) in a 1 Hz noise bandwidth.
Keysight N9030A PXA Specification Guide 57
PXA Signal Analyzer
Dynamic Range
Spurious Responses
DescriptionSpecificationsSupplemental Information
Spurious Responses
(see Band Overlaps on page 22)
Residual Responses
b
200 kHz to 8.4 GHz (swept)
Zero span or FFT or other frequencies
13.5 to 17.1 GHzf+645 MHz10 dBm74 dBc80 dBc84 dBc102 dBc
17.0 to 22 GHzf+645 MHz10 dBm70 dBc80 dBc82 dBc100 dBc
e
22 to 26.5 GHzf+645 MHz10 dBm68 dBc80 dBc79 dBc97 dBc
26.5 to 34.5 GHzf+645 MHz30 dBm60 dBc–70 dBc–84 dBc–94 dBc
34.4 to 44 GHzf+645 MHz30 dBm57 dBc–60 dBc–79 dBc–79 dBc
44 to 50 GHzf+645 MHz30 dBm–65 dBc
(nominal)
Other Spurious Responses
Mixer Level
c
Response
–75 dBc
(nominal)
Carrier Frequency 26.5 GHz
First RF Order
g
(f 10 MHz from carrier)
10 dBm
80 dBc + 20 log(N
h
)
Includes IF feedthrough,
LO harmonic mixing
responses
Higher RF Order
i
(f 10 MHz from carrier)
40 dBm
80 dBc + 20
log(Nh)
Includes higher order mixer
responses
58 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
DescriptionSpecificationsSupplemental Information
Carrier Frequency 26.5 GHz
First RF Order
d
30 dBm–90 dBc (nominal)
(f 10 MHz from carrier)
Higher RF Order
f
30 dBm–90 dBc (nominal)
(f 10 MHz from carrier)
LO-Related Spurious Responses
(Offset from carrier 200 Hz to 10 MHz)
Line-Related Spurious Responses
10 dBm
68 dBc
h
log(N
jf
)
+ 20
72 dBc + 20 log(N
(typical)
j
73 dBc
+ 20 log(Nh)
(nominal)
a. The spurious response specifications only apply with the preamp turned off. When the preamp is turned on,
performance is nominally the same as long as the mixer level is interpreted to be: Mixer Level = Input Level
Input Attenuation Preamp Gain
b. Input terminated, 0 dB input attenuation.
c. Mixer Level = Input Level Input Attenuation.
d. The values in this column are for RF/W PXA (Options 503, 508, 513, or 526) or
mmW PXA (Options 543, 544, or 550) with serial number prefix <MY/SG/US5135.
e. The values in this column are for mmW PXA (Options 543, 544, or 550) with serial number
prefix MY/SG/US5135.
f. We also support the following additional spurious responses specifications from 8 to 12 GHz at 20 to 30º C.
Image responses are warranted to be better than –82 dBc, with 95th percentile performance of –88.3 dBc.
LO-related spurious responses are warranted to be better than –83 dBc at 1 to 10 MHz offsets from the carrier,
with phase noise optimization set to Best Wide-Offset.
g. With first RF order spurious products, the indicated frequency will change at the same rate as the input, with
higher order, the indicated frequency will change at a rate faster than the input.
h. N is the LO multiplication factor.
i. RBW=100 Hz. With higher RF order spurious responses, the observed frequency will change at a rate faster
than the input frequency.
j. Nominally 40 dBc under large magnetic (0.38 Gauss rms) or vibrational (0.21 g rms) environmental stimuli.
h
)
Keysight N9030A PXA Specification Guide 59
PXA Signal Analyzer
Dynamic Range
Second Harmonic Distortion
DescriptionSpecificationsSupplemental Information
Second Harmonic Distortion
mmW (Option 543, 544, or 550)
RF/W (Option 503, 508, 513, or 526)
bc
Mixer
Level
Distortion
a
SHI
Distortion
(nominal)
Source Frequency
0 to 100 MHzx
0.1 to 1.8 GHz
d
to 3 GHz
1.75
d
1.75 to 3 GHz
3 to 6.5 GHzx
6.5 to 10 GHzx
10 to 13.25 GHzx
13.2 to 25 GHz
x–15 dBm–57 dBc+42 dBm
x
x–15 dBm–60 dBc+45 dBm
x–15 dBm–77 dBc+62 dBm
x–15 dBm–72 dBc+57 dBm
x15 dBm–77 dBc62 dBm
x15 dBm–70 dBc55 dBm
x15 dBm–62 dBc dBm
x15 dBm65 dBc+50 dBm
a. Mixer level = Input Level Input Attenuation
b. SHI = second harmonic intercept. The SHI is given by the mixer power in dBm minus the second harmonic distor-
tion level relative to the mixer tone in dBc.
c. Performance >3.6 GHz improves greatly with Option LNP enabled. See “Option LNP - Low Noise Path
Specifications” on page 201.
d. These frequencies are half of the band edge frequencies. See Band Overlaps on page 22.
bc
SHI
(nominal)
60 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
Third Order Intermodulation
DescriptionSpecificationsSupplemental Information
Third Order
Intermodulation
(Tone separation > 5 times IF
Prefilter Bandwidth
a
Refer to the footnote for
Band Overlaps on page 22.
Refer to footnote
"Extrapolated Distortion".
Verification conditionsb)
RF/W/mmW (SN prefix MY/SG/US5312)
mmW Option 543, 544, or 550
(SN prefix MY/SG/US5312)
RF/W Option 503, 508, 513, or 526
(SN prefix MY/SG/US5312)
20 to 30C
Intercept
d
Intercept
(typical)
10 to 150 MHzx
150 to 600 MHzx
600 MHz to 1.1 GHzx
1.1 to 3.6 GHzx
xx+13 dBm+16 dBm
xx+18 dBm+21 dBm
xx+20 dBm+22 dBm
xx+21 dBm+23 dBm
3.5 to 8.4 GHzx+17 dBm+23 dBm
c
for the
3.6 to 3.7 GHz (Band 0)x
See note
3.5 to 8.4 GHzx+16 dBm+23 dBm
3.5 to 8.4 GHzx+15 dBm+22 dBm
8.3 to 13.6 GHz x+17 dBm+23 dBm
8.3 to 13.6 GHz
x+16 dBm+23 dBm
8.3 to 13.6 GHz x+15 dBm+23 dBm
13.5 to 17.1 GHz x+15 dBm+20 dBm
13.5 to 17.1 GHz
x+13 dBm+17 dBm
13.5 to 17.1 GHz x+11 dBm+17 dBm
17.0 to 26.5 GHzf
17.0 to 26.5 GHz
17.0 to 26.5 GHz
26.5 to 50 GHz
f
f
x+16 dBm+22 dBm
x+13 dBm+20 dBm
x+11 dBm+17 dBm (nominal)
xx+13 dBm (nominal)
e
Keysight N9030A PXA Specification Guide 61
PXA Signal Analyzer
Dynamic Range
DescriptionSpecificationsSupplemental Information
Full temperature range
10 to 150 MHzxxx+12 dBm
150 to 600 MHzx
600 MHz to 1.1 GHzx
1.1 to 3.6 GHzx
xx+17 dBm
xx+18 dBm
xx+19 dBm
3.5 to 8.4 GHzx+14 dBm
3.5 to 8.4 GHz
xx+13 dBm
8.3 to 13.6 GHz x+14 dBm
8.3 to 13.6 GHz
xx+13 dBm
13.5 to 17.1 GHz x+13 dBm
13.5 to 17.1 GHz
17.0 to 26.5 GHzf
17.0 to 26.5 GHz
17.0 to 26.5 GHz
f
f
xx+10 dBm
x+14 dBm
x+10 dBm
x+9 dBm
a. See the IF Prefilter Bandwidth table in the Gain Compression specifications on page 49. When the tone sepa-
ration condition is met, the effect on TOI of the setting of IF Gain is negligible. TOI is verified with IF Gain set to
its best case condition, which is IF Gain = Low.
b. TOI is verified with two tones spaced by 100 kHz, each at:
16 dBm at the mixer (10 MHz to 26.5 GHz frequency range)
20 dBm at the mixer (26.5 GHz to 50 GHz frequency range)
c. Traditionally, the distortion components from two tones, each at 30 dBm, were given as specifications. When
spectrum analyzers were not as good as they are now, these distortion products were easily measured. As
spectrum analyzers improved, the measurement began to be made at higher levels and extrapolated to the
industry-standard 30 dBm test level. This extrapolation was justified by excellent conformance with the
third-order model, wherein distortion in dBc was given by twice the difference between the test tone level and
the intercept, both given in dBm units. In PXA, we no longer make that extrapolation in this Specifications
Guide.
One reason we don’t extrapolate is that the model does not work as well as it had with higher levels of distortion
in older and less capable analyzers, so that the computation is misleading; distortions at low test levels will be
modestly higher than predicted from the formula. The second reason is that the distortion components are so
small as to be unmeasurable, and thus highly irrelevant, in many cases.
Please note the curvature of the third-order intermodulation line in the 1 GHz graph that follows, which is rep-
resentative of performance below 3.6 GHz.
d. Intercept = TOI = third order intercept. The TOI is given by the mixer tone level (in dBm) minus (distortion/2)
where distortion is the relative level of the distortion tones in dBc.
62 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
e. Band 0 is extendable (set “Extend Low Band” to On) to 3.7 GHz instead of 3.6 GHz in instru-
ments with frequency option 508, 513 or 526 and serial number is equal with or greater than
the following: US54490126, MY54490690, or SG54490127. Band 0 can also be extendable in
earlier instruments with firmware of version A.16.13 or later and a field adjustment to perform
that extension. Subject to these conditions, statistical observations show that performance
nominally fits within the same range within the 3.6 to 3.7 GHz frequencies as within the next
lower specified frequency range, but is not warranted.
f. Intercept performance is nominally 3 dB better in this band in those analyzers which have either Option LNP
or Option MPB installed, or both, when these options are not in use.
Nominal TOI vs. Input Frequency and Tone Separation [Plot]
Keysight N9030A PXA Specification Guide 63
PXA Signal Analyzer
Dynamic Range
Nominal Dynamic Range [Plot]
64 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
Nominal Dynamic Range vs. Offset Frequency vs. RBW [Plot]
Keysight N9030A PXA Specification Guide 65
PXA Signal Analyzer
Dynamic Range
Phase Noise
DescriptionSpecificationsSupplemental Information
Phase NoiseNoise Sidebands
(Center Frequency = 1 GHz
Optimization
Internal Reference
a. The nominal performance of the phase noise at center frequencies different than the one at which the specifi-
cations apply (1 GHz) depends on the center frequency, band and the offset. For low offset frequencies, offsets
well under 100 Hz, the phase noise increases by 20 × log[(f + 0.3225)/1.3225]. For mid-offset frequencies such
as 50 kHz, phase noise trends as 20 × log[(f + 5.1225)/6.1225], and also varies chaotically an additional nominally ±2 dB versus the center frequency. For wide offset frequencies, offsets above about 500 kHz, phase noise
increases as 20 × log(N). N is the LO Multiple as shown on page 22; f is in GHz units in all these relationships;
all increases are in units of decibels.
b. Noise sidebands for lower offset frequencies, for example, 10 kHz, apply with the phase noise optimization
(PhNoise Opt) set to Best Close-in Noise. Noise sidebands for higher offset frequencies, for example, 1 MHz,
as shown apply with the phase noise optimization set to Best Wide-offset Noise.
c. Specifications are given with the internal frequency reference. The phase noise at offsets below 100 Hz is
impacted or dominated by noise from the reference. Thus, performance with external references will not follow
the curves and specifications. When using an external reference with superior phase noise, we recommend setting the external reference phase-locked-loop bandwidth to wide (60 Hz), to take advantage of that superior
performance. When using an external reference with inferior phase noise performance, we recommend setting
that bandwidth to narrow (15 Hz). In these relationships, inferior and superior phase noise are with respect to
–134 dBc/Hz at 30 Hz offset from a 10 MHz reference. Because most reference sources have phase noise
behavior that falls off at a rate of 30 dB/decade, this is usually equivalent to –120 dBc/Hz at 10 Hz offset. For
more information, see Phase Noise Effects, Ext Ref vs. Loop BW [Plot] on page 70.
d. Nominal phase noise was –75 dBc/Hz for instruments produced before approximately September 1, 2012.
d
66 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
e. Analyzer-contributed phase noise at the low levels of this offset requires advanced verification techniques
because broadband noise would otherwise cause excessive measurement error. Keysight uses a high level low
phase noise CW test signal and sets the input attenuator so that the mixer level will be well above the normal
top-of-screen level (-10 dBm) but still well below the 1 dB compression level. This improves dynamic range
(carrier to broadband noise ratio) at the expense of amplitude uncertainty due to compression of the phase
noise sidebands of the analyzer. (If the mixer level were increased to the "1 dB Gain Compression Point," the
compression of a single sideband is specified to be 1 dB or lower. At lower levels, the compression falls off rapidly. The compression of phase noise sidebands is substantially less than the compression of a single-sideband
test signal, further reducing the uncertainty of this technique.) Keysight also measures the broadband noise of
the analyzer without the CW signal and subtracts its power from the measured phase noise power. The same
techniques of overdrive and noise subtraction can be used in measuring a DUT, of course.
Keysight N9030A PXA Specification Guide 67
PXA Signal Analyzer
Dynamic Range
Nominal Phase Noise of Different LO Optimizations [Plot]
68 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Dynamic Range
Nominal Phase Noise at Different Center Frequencies [Plot]
Keysight N9030A PXA Specification Guide 69
PXA Signal Analyzer
Dynamic Range
Phase Noise Effects, Ext Ref vs. Loop BW [Plot]
The effect of the Ext Ref Loop BW control (Narrow and Wide) is shown in this graphic. When set to Wide, the noise from the
internal circuitry is reduced, but noise in the external reference is subject to being impressed on the LO through the transfer
function shown in the smooth curve labeled "Wide." For an excellent reference, this can give lower overall noise. When the
Narrow selection is made, the internal noise effect is higher, but external reference noise above 20 Hz is rejected.
The noise curves were measured at 1 GHz center frequency with an excellent reference and excellent RF signal, and is thus
a conservative estimate of the residual noise of the PXA circuitry. At that center frequency, the transfer function curves
approach 40 dB gain to phase noise at low offset frequencies for a 10 MHz external reference. (This 40 dB is computed as
20 log
10 ^(fC/f
10
). The measured noise curves will scale with frequency the same way.)
REF
Example: Consider an external reference at 10 MHz with phase noise of 135 dBc/Hz at 20 Hz offset. If the Narrow setting
is chosen, the analyzer noise density will be 86 dBc/Hz at 20 Hz offset, the gain to the reference will be 40 dB, giving
95 dBc/Hz contribution. Add these together on a power scale as 10 x log
(–86/10)
(10
10
+ 10
(–95/10)
) = 85.5 dBc/Hz. If
Wide is chosen, the analyzer noise density will be 97 dBc/Hz at 20 Hz offset, the gain to the reference will be 42 dB, giving
93 dBc/Hz contribution. Add those together on a power scale as 10 x log
(–97/10)
(10
10
+ 10
(–93/10)
) = 91.5 dBc/Hz.
"Wide" will give a 6 dB superior result to the Narrow selection.
70 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Power Suite Measurements
Power Suite Measurements
The specifications for this section apply only to instruments with Frequency
Option 503, 508, 513, or 526. For instruments with higher frequency options,
the performance is nominal only and not subject to any warranted
specifications.
The measurement performance is only slightly different between instruments
with the lower and higher frequency options. Because the hardware
performance of the analyzers is very similar but not identical, you can estimate
the nominal performance of the measurements from the specification in this
chapter.
DescriptionSpecificationsSupplemental Information
Channel Power
Amplitude Accuracy
Case: Radio Std = 3GPP W-CDMA, or IS-95
Absolute Power Accuracy
(20 to 30C, Attenuation = 10 dB)
a. See “Absolute Amplitude Accuracy” on page 41.
b. See “Frequency and Time” on page 22.
c. Expressed in dB.
DescriptionSpecificationsSupplemental Information
Occupied Bandwidth
Frequency Accuracy(Span/1000) (nominal)
0.61 dB
Absolute Amplitude Accuracy
Power Bandwidth Accuracy
0.19 dB (95th percentile)
a
+
bc
Keysight N9030A PXA Specification Guide 71
PXA Signal Analyzer
Power Suite Measurements
DescriptionSpecificationsSupplemental Information
Adjacent Channel Power (ACP)
Case: Radio Std = None
Accuracy of ACP Ratio (dBc)
Accuracy of ACP Absolute Power
(dBm or dBm/Hz)
Accuracy of Carrier Power (dBm), or
Carrier Power PSD (dBm/Hz)
Passband Width
e
Case: Radio Std = 3GPP W-CDMA
Display Scale Fidelity
Absolute Amplitude Accuracy
Power Bandwidth Accuracy
Absolute Amplitude Accuracy
Power Bandwidth Accuracy
MS (UE)5 MHz0.08 dB At ACPR range of 30 to 36 dBc with optimum mixer
h
level
MS (UE)10 MHz0.09 dBAt ACPR range of 40 to 46 dBc with optimum mixer
i
level
BTS5 MHz0.20 dBAt ACPR range of 42 to 48 dBc with optimum mixer
j
level
BTS10 MHz0.18 dBAt ACPR range of 47 to 53 dBc with optimum mixer
i
level
BTS5 MHz0.10 dB
At 48 dBc non-coherent ACPR
k
Option EP0
RadioOffset Freq
MS (UE)5 MHz0.09 dB At ACPR range of 30 to 36 dBc with optimum mixer
l
level
MS (UE)10 MHz0.11 dBAt ACPR range of 40 to 46 dBc with optimum mixer
m
level
BTS5 MHz0.25 dBAt ACPR range of 42 to 48 dBc with optimum mixer
n
level
BTS10 MHz0.25 dBAt ACPR range of 47 to 53 dBc with optimum mixer
i
level
BTS5 MHz0.12 dB
At 48 dBc non-coherent ACPR
o
72 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Power Suite Measurements
DescriptionSpecificationsSupplemental Information
Dynamic RangeRRC weighted, 3.84 MHz noise
bandwidth
Noise
Correction
p
Offset
Freq
Method
ACLR (typical)
q
Optimum MLr (Nominal)
Off5 MHzFiltered IBW82.5 dB8 dBm
Off5 MHzFast81 dB8 dBm
Off10 MHzFiltered IBW87 dB4 dBm
On5 MHzFiltered IBW83.5 dB8 dBm
On5 MHzFiltered IBW
88 dB
s
8 dBm
On10 MHzFiltered IBW89.5 dB4 dBm
RRC Weighting Accuracy
White noise in Adjacent Channel
TOI-induced spectrum
rms CW error
t
0.00 dB nominal
0.001 dB nominal
0.012 dB nominal
a. The effect of scale fidelity on the ratio of two powers is called the relative scale fidelity. The scale fidelity specified
in the Amplitude section is an absolute scale fidelity with –35 dBm at the input mixer as the reference point. The
relative scale fidelity is nominally only 0.01 dB larger than the absolute scale fidelity.
b. See Amplitude Accuracy and Range section.
c. See Frequency and Time section.
d. Expressed in decibels.
e. An ACP measurement measures the power in adjacent channels. The shape of the response versus frequency of
those adjacent channels is occasionally critical. One parameter of the shape is its 3 dB bandwidth. When the
bandwidth (called the Ref BW) of the adjacent channel is set, it is the 3 dB bandwidth that is set. The passband
response is given by the convolution of two functions: a rectangle of width equal to Ref BW and the power
response versus frequency of the RBW filter used. Measurements and specifications of analog radio ACPs are
often based on defined bandwidths of measuring receivers, and these are defined by their 6 dB widths, not their
3 dB widths. To achieve a passband whose 6 dB width is x, set the Ref BW to be x 0.572 RBW.
f. Most versions of adjacent channel power measurements use negative numbers, in units of dBc, to refer to the
power in an adjacent channel relative to the power in a main channel, in accordance with ITU standards. The
standards for W-CDMA analysis include ACLR, a positive number represented in dB units. In order to be consis-
tent with other kinds of ACP measurements, this measurement and its specifications will use negative dBc
results, and refer to them as ACPR, instead of positive dB results referred to as ACLR. The ACLR can be deter-
mined from the ACPR reported by merely reversing the sign.
g. The accuracy of the Adjacent Channel Power Ratio will depend on the mixer drive level and whether the distortion
products from the analyzer are coherent with those in the UUT. These specifications apply even in the worst case
condition of coherent analyzer and UUT distortion products. For ACPR levels other than those in this specifica-
tions table, the optimum mixer drive level for accuracy is approximately 37 dBm (ACPR/3), where the ACPR is
given in (negative) decibels.
Keysight N9030A PXA Specification Guide 73
PXA Signal Analyzer
Power Suite Measurements
h. To meet this specified accuracy when measuring mobile station (MS) or user equipment (UE) within 3 dB of the
required 33 dBc ACPR, the mixer level (ML) must be optimized for accuracy. This optimum mixer level is
22 dBm, so the input attenuation must be set as close as possible to the average input power (22 dBm). For
example, if the average input power is 6 dBm, set the attenuation to 16 dB. This specification applies for the
normal 3.5 dB peak-to-average ratio of a single code. Note that, if the mixer level is set to optimize dynamic
range instead of accuracy, accuracy errors are nominally doubled.
i. ACPR accuracy at 10 MHz offset is warranted when the input attenuator is set to give an average mixer level of
14 dBm.
j. In order to meet this specified accuracy, the mixer level must be optimized for accuracy when measuring node B
Base Transmission Station (BTS) within 3 dB of the required 45 dBc ACPR. This optimum mixer level is 18
dBm, so the input attenuation must be set as close as possible to the average input power (18 dBm). For
example, if the average input power is 6 dBm, set the attenuation to 12 dB. This specification applies for the
normal 10 dB peak-to-average ratio (at 0.01 probability) for Test Model 1. Note that, if the mixer level is set to
optimize dynamic range instead of accuracy, accuracy errors are nominally doubled.
k. Accuracy can be excellent even at low ACPR levels assuming that the user sets the mixer level to optimize the
dynamic range, and assuming that the analyzer and UUT distortions are incoherent. When the errors from the
UUT and the analyzer are incoherent, optimizing dynamic range is equivalent to minimizing the contribution of
analyzer noise and distortion to accuracy, though the higher mixer level increases the display scale fidelity errors.
This incoherent addition case is commonly used in the industry and can be useful for comparison of analysis
equipment, but this incoherent addition model is rarely justified. This derived accuracy specification is based on a
mixer level of 14 dBm.
l. To meet this specified accuracy when measuring mobile station (MS) or user equipment (UE) within 3 dB of the
required 33 dBc ACPR, the mixer level (ML) must be optimized for accuracy. This optimum mixer level is
22 dBm, so the input attenuation must be set as close as possible to the average input power (22 dBm). For
example, if the average input power is 6 dBm, set the attenuation to 16 dB. This specification applies for the
normal 3.5 dB peak-to-average ratio of a single code. Note that, if the mixer level is set to optimize dynamic
range instead of accuracy, accuracy errors are nominally doubled.
m. ACPR accuracy at 10 MHz offset is warranted when the input attenuator is set to give an average mixer level of
14 dBm.
n. In order to meet this specified accuracy, the mixer level must be optimized for accuracy when measuring node B
Base Transmission Station (BTS) within 3 dB of the required 45 dBc ACPR. This optimum mixer level is 18
dBm, so the input attenuation must be set as close as possible to the average input power (18 dBm). For
example, if the average input power is 6 dBm, set the attenuation to 12 dB. This specification applies for the
normal 10 dB peak-to-average ratio (at 0.01 probability) for Test Model 1. Note that, if the mixer level is set to
optimize dynamic range instead of accuracy, accuracy errors are nominally doubled.
o. Accuracy can be excellent even at low ACPR levels assuming that the user sets the mixer level to optimize the
dynamic range, and assuming that the analyzer and UUT distortions are incoherent. When the errors from the
UUT and the analyzer are incoherent, optimizing dynamic range is equivalent to minimizing the contribution of
analyzer noise and distortion to accuracy, though the higher mixer level increases the display scale fidelity errors.
This incoherent addition case is commonly used in the industry and can be useful for comparison of analysis
equipment, but this incoherent addition model is rarely justified. This derived accuracy specification is based on a
mixer level of 14 dBm.
p. The dynamic range shown with Noise Correction = Off applies with Noise Floor Extension On. (Noise Correction is
the process within the measurement of making a calibration of the noise floor at the exact analyzer settings used
for the measurement. Noise Floor Extension is the factory calibration of the noise floor.)
q. Keysight measures 100 of the signal analyzers for dynamic range in the factory production process. This mea-
surement requires a near-ideal signal, which is impractical for field and customer use. Because field verification is
impractical, Keysight only gives a typical result. More than 80 of prototype instruments met this “typical” spec-
ification; the factory test line limit is set commensurate with an on-going 80 yield to this typical.
The ACPR dynamic range is verified only at 2 GHz, where Keysight has the near-perfect signal available. The
dynamic range is specified for the optimum mixer drive level, which is different in different instruments and dif-
ferent conditions. The test signal is a 1 DPCH signal.
The ACPR dynamic range is the observed range. This typical specification includes no measurement uncertainty.
74 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Power Suite Measurements
r. ML is Mixer Level, which is defined to be the input signal level minus attenuation.
s. All three production units hand-measured had performance better than 88 dB with a test signal even better than
q
the "near-ideal" one used for statistical process control in production mentioned in the footnote
above. Therefore, this value can be considered "Nominal" not "Typical" by the definitions used within this document. These
observations were done near 2 GHz because that is a common W-CDMA operating region in which the analyzer
third-order dynamic range is near its best.
t. 3GPP requires the use of a root-raised-cosine filter in evaluating the ACLR of a device. The accuracy of the pass-
band shape of the filter is not specified in standards, nor is any method of evaluating that accuracy. This footnote
discusses the performance of the filter in this instrument. The effect of the RRC filter and the effect of the RBW
used in the measurement interact. The analyzer compensates the shape of the RRC filter to accommodate the
RBW filter. The effectiveness of this compensation is summarized in three ways:
White noise in Adj Ch: The compensated RRC filter nominally has no errors if the adjacent channel has a spectrum that is flat across its width.
TOIinduced spectrum: If the spectrum is due to thirdorder intermodulation, it has a distinctive shape. The
computed errors of the compensated filter are 0.001 dB for the 100 kHz RBW used for UE testing with the IBW
method. It is 0.000 dB for the 27 kHz RBW filter used for BTS testing with the Filtered IBW method. The worst
error for RBWs between 27 and 390 kHz is 0.05 dB for a 330 kHz RBW filter.
rms CW error: This error is a measure of the error in measuring a CWlike spurious component. It is evaluated
by computing the root of the mean of the square of the power error across all frequencies within the adjacent
channel. The computed rms error of the compensated filter is 0.012 dB for the 100 kHz RBW used for UE testing
with the IBW method. It is 0.000 dB for the 27 kHz RBW filter used for BTS testing. The worst error for RBWs
between 27 kHz and 470 kHz is 0.057 dB for a 430 kHz RBW filter.
a. Coher = no means that the specified accuracy only applies when the distortions of the device under test are not
coherent with the third-order distortions of the analyzer. Incoherence is often the case with advanced
multi-carrier amplifiers built with compensations and predistortions that mostly eliminate coherent third-order
effects in the amplifier.
b. Optimum mixer level (MLOpt). The mixer level is given by the average power of the sum of the four carriers
minus the input attenuation.
Keysight N9030A PXA Specification Guide 75
PXA Signal Analyzer
Power Suite Measurements
DescriptionSpecificationsSupplemental Information
Power Statistics CCDF
Histogram Resolution
a
0.01 dB
a. The Complementary Cumulative Distribution Function (CCDF) is a reformatting of a histogram of the power
envelope. The width of the amplitude bins used by the histogram is the histogram resolution. The resolution of
the CCDF will be the same as the width of those bins.
DescriptionSpecificationsSupplemental Information
Burst Power
MethodsPower above threshold
Power within burst width
ResultsOutput power, average
Output power, single burst
Maximum power
Minimum power within burst
Burst width
DescriptionSpecificationsSupplemental Information
TOI (Third Order
Intermodulation)
Measures TOI of a signal with two
dominant tones
ResultsRelative IM tone powers (dBc)
Absolute tone powers (dBm)
Intercept (dBm)
DescriptionSpecificationsSupplemental Information
Harmonic Distortion
Maximum harmonic number10th
ResultsFundamental Power (dBm)
Relative harmonics power (dBc)
Total harmonic distortion (%, dBc)
76 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Power Suite Measurements
Description SpecificationsSupplemental Information
Spurious EmissionsTable-driven spurious signals;
search across regions
Case: Radio Std = 3GPP W-CDMA
Dynamic Range
a
, relative (RBW=1 MHz)
88.8 dB 92.1 dB (typical)
(1 to 3.6GHz)
Dynamic Range
b
, relative (RBW=1 MHz)
86.9 dB 89.9 dB (typical)
(1 to 3.6GHz) (Option EP0)
Sensitivity
c
, absolute (RBW=1 MHz)
88.5 dBm91.5 dBm (typical)
(1 to 3.6 GHz)
Sensitivity
d
, absolute (RBW=1 MHz)
865 dBm89.5 dBm (typical)
(1 to 3.6 GHz) (Option EP0)
AccuracyAttenuation = 10 dB
20 Hz to 3.6 GHz0.19 dB (95th percentile)
3.5 to 8.4 GHz1.08 dB (95th percentile)
3.5 to 8.4 GHz (Option EP0)1.15 dB (95th percentile)
8.3 to 13.6 GHz1.48 dB (95th percentile)
8.3 to 13.6 GHz (Option EP0)1.52 dB (95th percentile)
a. The dynamic range is specified at 12.5 MHz offset from center frequency with mixer level of 1 dB compression
point, which will degrade accuracy 1 dB.
b. The dynamic range is specified at 12.5 MHz offset from center frequency with mixer level of 1 dB compression
point, which will degrade accuracy 1 dB.
c. The sensitivity is specified at far offset from carrier, where phase noise does not contribute. You can derive the
dynamic range at far offset from 1 dB compression mixer level and sensitivity.
d. The sensitivity is specified at far offset from carrier, where phase noise does not contribute. You can derive the
dynamic range at far offset from 1 dB compression mixer level and sensitivity.
a. The dynamic range specification is the ratio of the channel power to the power in the offset specified. The
dynamic range depends on the measurement settings, such as peak power or integrated power. Dynamic range
specifications are based on default measurement settings, with detector set to average, and depend on the mixer
level. Default measurement settings include 30 kHz RBW.
b. This dynamic range specification applies for the optimum mixer level, which is about 18 dBm. Mixer level is
defined to be the average input power minus the input attenuation.
c. The sensitivity is specified with 0 dB input attenuation. It represents the noise limitations of the analyzer. It is
tested without an input signal. The sensitivity at this offset is specified in the default 30 kHz RBW, at a center frequency of 2 GHz.
d. The relative accuracy is a measure of the ratio of the power at the offset to the main channel power. It applies for
spectrum emission levels in the offsets that are well above the dynamic range limitation.
e. The absolute accuracy of SEM measurement is the same as the absolute accuracy of the spectrum analyzer. See
“Absolute Amplitude Accuracy” on page 41 for more information. The numbers shown are for 0 to 3.6
GHz, with attenuation set to 10 dB.
78 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Options
Options
The following options and applications affect instrument specifications.
Option 503:Frequency range, 3 Hz to 3.6 GHz
Option 508:Frequency range, 3 Hz to 8.4 GHz
Option 513:Frequency range, 3 Hz to 13.6 GHz
Option 526:Frequency range, 3 Hz to 26.5 GHz
Option 543:Frequency range, 3 Hz to 43 GHz
Option 544:Frequency range, 3 Hz to 44 GHz
Option 550:Frequency range, 3 Hz to 50 GHz
Option ALV:Auxiliary Log Video output
Option B1X:
Option B25:Analysis bandwidth, 25 MHz
Option B40:Analysis bandwidth, 40 MHz
Option B85:Analysis bandwidth, 85 MHz
Option BBA:BBIQ inputs, analog
Option C35:APC 3.5 mm connector (for Freq Option 526 only)
N9083A:Multi-Standard Radio (MSR) measurement application
80 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Options
a. Instruments with 160 MHz bandwidth will display "N9030A-B1Y Analysis bandwidth Extension, 160 MHz" after
pressing: System, Show, System on the instrument.
Keysight N9030A PXA Specification Guide 81
PXA Signal Analyzer
General
General
DescriptionSpecificationsSupplemental Information
Calibration Cycle1 year
DescriptionSpecificationsSupplemental Information
Temperature Range
Operating
a
Altitude 2,300m0 to 55C
Altitude = 4,500 m0 to 47C
Derating
Storage
Altitude
b
c
d
40 to +70C
4,500 m (approx 15,000 feet)
Humidity
Relative humidityType tested at 95%, +40°C
(non-condensing)
a. For earlier instruments (S/N <MY49430337/SG49430006/US49230450), the operating temperature ranges
from 5 to 50C.
b. The maximum operating temperature derates linearly from altitude of 4,500 m to 2,300 m.
c. For earlier instruments (S/N <MY49430337/SG49430006/US49230450), and installed with hard disk drives,
the storage temperature ranges from –40 to +65C.
d. For earlier instrument (S/N <MY49430337/SG49430006/US49230450), the altitude was specified as 3,000 m
(approximately 10,000 feet).
DescriptionSpecificationsSupplemental Information
Environmental and Military
Specifications
Samples of this product have been type tested in
accordance with the Keysight Environmental Test
Manual and verified to be robust against the
environmental stresses of Storage, Transportation
and End-use; those stresses include but are not
limited to temperature, humidity, shock, vibration,
altitude and power line conditions. Test Methods are
aligned with IEC 60068-2 and levels are similar to
MIL-PRF-28800F Class 3.
82 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
General
DescriptionSpecifications
EMCComplies with the essential requirements of the European EMC Directive as well as
current editions of the following standards (dates and editions are cited in the
Declaration of Conformity):
— IEC/EN 61326-1 or IEC/EN 61326-2-1
— CISPR 11 Group 1, Cass A
— AS/NZS CISPR 11
—ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
Values given are per ISO 7779 standard in the "Operator Sitting"
position
considered suitable for use in quiet office environments.
considered suitable for use in noisy office environments. (The fan
speed, and thus the noise level, increases with increasing ambient
temperature.)
Keysight N9030A PXA Specification Guide 83
PXA Signal Analyzer
General
DescriptionSpecificationSupplemental Information
Power Requirements
Low Range
Voltage100 to 120 V
Frequency50, 60 or 400 Hz
High Range
Voltage220 to 240 V
Frequency 50 or 60 Hz
Power Consumption, On630 WMaximum
Power Consumption, Standby45 WStandby power is not supplied to the
frequency reference oscillator but to
the CPU.
Typical instrument configurationPower (nominal)
Base PXA instrument330 W
Adding Option B85/B1X to base instrument+45 W
Adding Option BBA to base instrument+46 W
DescriptionSupplemental Information
Measurement Speed
a
Local measurement and display update rate
Remote measurement and LAN transfer rate
bc
bc
Nominal
10 ms (100/s)
10 ms (100/s)
Marker Peak Search2.5 ms
Center Frequency Tune and Transfer (Band 0)43 ms
Center Frequency Tune and Transfer (Bands 1-4)69 ms
Measurement/Mode Switching40 ms
W-CDMA ACLR measurement timeSee page 72
Measurement Time vs. SpanSee page 31
a. Sweep Points = 101.
b. Factory preset, fixed center frequency, RBW = 1 MHz, 10 MHz < span 600 MHz, stop frequency 3.6 GHz,
Auto Align Off.
c. Phase Noise Optimization set to Fast Tuning, Display Off, 32 bit integer format, markers Off, single sweep, mea-
sured with IBM compatible PC with 2.99 GHz Pentium® 4 with 2 GB RAM running Windows® XP, Keysight I/O
Libraries Suite Version 14.1, one meter GPIB cable, National Instruments PCI-GPIB Card and NI-488.2 DLL.
84 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
General
DescriptionSpecificationsSupplemental Information
Display
a
Resolution1024 768XGA
Size213 mm (8.4 in) diagonal (nominal)
a. The LCD display is manufactured using high precision technology. However, there may be up to six bright points
(white, blue, red or green in color) that constantly appear on the LCD screen. These points are normal in the
manufacturing process and do not affect the measurement integrity of the product in any way.
DescriptionSpecificationsSupplemental Information
Data Storage
Internal Total
Removeable solid state drive (80 GB
a
Internal User9 GB available on separate partition for user
data
a. For earlier instruments (S/N <MY49430337/SG49430006/US49230450), a hard disk (>80 GB) drive is installed
as a standard feature unless ordered with a solid state drive.
DescriptionSpecificationsSupplemental Information
WeightWeight without options
Net22 kg (48 lbs) (nominal)
Shipping 34 kg (75 lbs) (nominal)
Cabinet DimensionsCabinet dimensions exclude front and rear
Height177 mm (7.0 in)
protrusions.
Width426 mm (16.8 in)
Length556 mm (21.9 in)
Keysight N9030A PXA Specification Guide 85
PXA Signal Analyzer
Inputs/Outputs
Inputs/Outputs
Front Panel
DescriptionSpecificationsSupplemental Information
RF Input
Connector
StandardType-N femaleFrequency Option 503, 508, 513, and 526
2.4 mm maleFrequency Option 543, 544, and 550
Option C353.5 mm maleFrequency Option 526 only
Impedance50 (nominal)
DescriptionSpecificationsSupplemental Information
Probe Power
Voltage/Current+15 Vdc, 7 at 0 to 150 mA (nominal)
12.6 Vdc, 10 at 0 to 150 mA (nominal)
GND
DescriptionSpecificationsSupplemental Information
USB 2.0 Ports
Host (2 ports)
ConnectorUSB Type “A” (female)
Output Current0.5 A (nominal)
DescriptionSpecificationsSupplemental Information
Headphone Jack
Connectorminiature stereo audio jack3.5 mm (also known as "1/8 inch")
Output Power90 mW per channel into 16 (nominal)
See Rear Panel for other ports
86 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Inputs/Outputs
Rear Panel
DescriptionSpecificationsSupplemental Information
10 MHz Out
ConnectorBNC female
Impedance50 (nominal)
Output Amplitude0 dBm (nominal)
Output ConfigurationAC coupled, sinusoidal
Frequency10 MHz
(1 + frequency reference accuracy)
DescriptionSpecificationsSupplemental Information
Ext Ref In
ConnectorBNC femaleNote: Analyzer noise sidebands and spurious
response performance may be affected by the
quality of the external reference used. See
footnote
within the Dynamic Range section on page 66.
Impedance50 (nominal)
Input Amplitude Range
sine wave
square wave
Input Frequency1 to 50 MHz (nominal)
6
Lock range
DescriptionSpecificationsSupplemental Information
SyncReserved for future use
ConnectorBNC female
2 10
reference input frequency
of ideal external
5 to +10 dBm (nominal)
0.2 to 1.5 V peak-to-peak (nominal)
(selectable to 1 Hz resolution)
c
in the Phase Noise specifications
Keysight N9030A PXA Specification Guide 87
PXA Signal Analyzer
Inputs/Outputs
DescriptionSpecificationsSupplemental Information
Trigger Inputs
(Trigger 1 In, Trigger 2 In)
ConnectorBNC female
Impedance10 k (nominal)
Trigger Level Range5 to +5 V1.5 V (TTL) factory preset
DescriptionSpecificationsSupplemental Information
Trigger Outputs
(Trigger 1 Out, Trigger 2 Out)
ConnectorBNC female
Impedance50 (nominal)
Level0 to 5 V (CMOS)
DescriptionSpecificationsSupplemental Information
Monitor Output
Connector
Format
VGA compatible,
15-pin mini D-SUB
Either trigger source may be selected
XGA (60 Hz vertical sync rates, non-interlaced)
Analog RGB
Resolution
DescriptionSpecificationsSupplemental Information
Analog Out
ConnectorBNC female
Impedance50 (nominal)
DescriptionSpecificationsSupplemental Information
Noise Source Drive +28 V (Pulsed)
ConnectorBNC female
Output voltage on28.0 ± 0.1 V60 mA maximum current
Output voltage off< 1.0 V
1024 768
88 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Inputs/Outputs
DescriptionSpecsSupplemental Information
SNS Series Noise SourceFor use with Keysight Technologies SNS Series noise sources
DescriptionSpecificationsSupplemental Information
Digital BusThis port is intended for use with the Keysight N5105 and N5106 products
only. It is not available for general purpose use.
ConnectorMDR-80
DescriptionSpecificationsSupplemental Information
USB 2.0 Ports
Host (3 ports)
a
See Front Panel for additional ports
ConnectorUSB Type “A” (female)
Output Current
0.5 A (nominal)
Device (1 port)
ConnectorUSB Type “B” (female)
a. Earlier instruments with SN prefix MY/SG/US55330000 shipped with 4 USB ports.
This product is designed for use in Installation Category II and Pollution Degree
2 per IEC 61010 3rd ed, and 664 respectively.
This product has been designed and tested in accordance with accepted
industry standards, and has been supplied in a safe condition. The instruction
documentation contains information and warnings which must be followed by
the user to ensure safe operation and to maintain the product in a safe
condition.
This product is intended for indoor use.
The CE mark is a registered trademark of the European Community (if accompanied
by a year, it is the year when the design was proven). This product complies with all
relevant directives.
ccr.keysight@keysight.com
ICES/NMB-001“This ISM device complies with Canadian ICES-001.”
ISM 1-A (GRP.1 CLASS A)This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
The CSA mark is a registered trademark of the CSA International.
The Keysight email address is required by EU directives applicable to our product.
“Cet appareil ISM est conforme a la norme NMB du Canada.”
(CISPR 11, Clause 4)
The RCM mark is a registered trademark of the Australian Communications and
Media Authority.
This symbol indicates separate collection for electrical and electronic equipment
mandated under EU law as of August 13, 2005. All electric and electronic equipment
are required to be separated from normal waste for disposal (Reference WEEE
Directive 2002/96/EC).
China RoHS regulations include requirements related to packaging, and require
compliance to China standard GB18455-2001.
This symbol indicates compliance with the China RoHS regulations for
paper/fiberboard packaging.
South Korean Certification (KC) mark; includes the marking’s identifier code which
follows this format:
MSIP-REM-YYY-ZZZZZZZZZZZZZZ.
90 Keysight N9030A PXA Specification Guide
PXA Signal Analyzer
Regulatory Information
EMC: Complies with the essential requirements of the European EMC Directive
as well as current editions of the following standards (dates and editions are
cited in the Declaration of Conformity):
— IEC/EN 61326-1
— CISPR 11, Group 1, Class A
— AS/NZS CISPR 11
— ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
This is a sensitive measurement apparatus by design and may have some performance
loss (up to 25 dBm above the Spurious Responses, Residual specification of -100 dBm)
when exposed to ambient continuous electromagnetic phenomenon in the range of
80 MHz -2.7 GHz when tested per IEC 61000-4-3.
South Korean Class A EMC declaration:
This equipment has been conformity assessed for use in business
environments. In a residential environment this equipment may cause radio
interference.
This EMC statement applies to the equipment only for use in business
environment.
SAFETY: Complies with the essential requirements of the European Low
Voltage Directive as well as current editions of the following standards (dates
and editions are cited in the Declaration of Conformity):
To find a current Declaration of Conformity for a specific Keysight product, go
to: http://www.keysight.com/go/conformity
92 Keysight N9030A PXA Specification Guide
Keysight X-Series Signal Analyzer
N9030A
Specification Guide
2I/Q Analyzer
This chapter contains specifications for the I/Q Analyzer measurement
application (Basic Mode).
93
I/Q Analyzer
Specifications Affected by I/Q Analyzer
Specifications Affected by I/Q Analyzer
Specification NameInformation
Number of Frequency Display Trace Points
(buckets)
Resolution BandwidthSee “Frequency” on page 95 in this chapter.
Video BandwidthNot available.
Clipping-to-Noise Dynamic RangeSee “Clipping-to-Noise Dynamic Range” on page 96 in this
Resolution Bandwidth Switching UncertaintyNot specified because it is negligible.
Available DetectorsDoes not apply.
Spurious ResponsesThe “Spurious Responses” on page 58 of core specifications still
IF Amplitude FlatnessSee “IF Frequency Response” on page 39 of the core
IF Phase LinearitySee “IF Phase Linearity” on page 40 of the core specifications for
Does not apply.
chapter.
apply. Additional bandwidth-option-dependent spurious responses are
given in the Analysis Bandwidth chapter for any optional bandwidths in
use.
specifications for the 10 MHz bandwidth. Specifications for wider
bandwidths are given in the Analysis Bandwidth chapter for any
optional bandwidths in use.
the 10 MHz bandwidth. Specifications for wider bandwidths are given
in the Analysis Bandwidth chapter for any optional bandwidths in use.
Data AcquisitionSee “Data Acquisition” on page 97 in this chapter for the 10 MHz
bandwidth. Specifications for wider bandwidths are given in the
Analysis Bandwidth chapter for any optional bandwidths in use.
94 Keysight N9030A PXA Specification Guide
I/Q Analyzer
Frequency
Frequency
DescriptionSpecificationsSupplemental Information
Frequency Span
Standard instrument10 Hz to10 MHz
Option B2510 Hz to 25 MHz
Option B4010 Hz to 40 MHz
Option B8510 Hz to 85 MHz
Option B1X
Resolution Bandwidth
(Spectrum Measurement)
Range
Overall
Span = 1 MHz
Span = 10 kHz
Span = 100 Hz
Window ShapesFlat Top, Uniform, Hanning, Hamming,
Analysis Bandwidth (Span)
(Waveform Measurement)
Standard instrument10 Hz to 10 MHz
Option B2510 Hz to 25 MHz
Option B4010 Hz to 40 MHz
Option B8510 Hz to 85 MHz
Option B1X
10 Hz to 160 MHz
100 mHz to 3 MHz
50 Hz to 1 MHz
1 Hz to 10 kHz
100 mHz to 100 Hz
Gaussian, Blackman, Blackman-Harris, Kaiser
Bessel (K-B 70 dB, K-B 90 dB & K-B 110 dB)
10 Hz to 160 MHz
a
a
a. Instruments with 160 MHz bandwidth will display "N9030A-B1Y Analysis bandwidth Extension, 160 MHz" after
pressing: System, Show, System on the instrument.
Keysight N9030A PXA Specification Guide 95
I/Q Analyzer
Clipping-to-Noise Dynamic Range
Clipping-to-Noise Dynamic Range
DescriptionSpecificationsSupplemental Information
Clipping-to-Noise Dynamic Range
a
Excluding residuals and spurious
responses
Clipping Level at MixerCenter frequency 20 MHz
IF Gain = Low10 dBm8 dBm (nominal)
IF Gain = High20 dBm17.5 dBm (nominal)
Noise Density at Mixer
at center frequency
c
(DANL
+ IFGainEffectd) + 2.25
b
dB
e
Example
f
a. This specification is defined to be the ratio of the clipping level (also known as “ADC Over Range”) to the noise
density. In decibel units, it can be defined as clipping_level [dBm] noise_density [dBm/Hz]; the result has units
of dBFS/Hz (fs is “full scale”).
b. The noise density depends on the input frequency. It is lowest for a broad range of input frequencies near the
center frequency, and these specifications apply there. The noise density can increase toward the edges of the
span. The effect is nominally well under 1 dB.
c. The primary determining element in the noise density is the “Displayed Average Noise Level” on
page 51.
d. DANL is specified with the IF Gain set to High, which is the best case for DANL but not for Clipping-to-noise
dynamic range. The core specifications “Displayed Average Noise Level” on page 51, gives a line entry
on the excess noise added by using IF Gain = Low, and a footnote explaining how to combine the IF Gain noise
with the DANL.
e. DANL is specified for log averaging, not power averaging, and thus is 2.51 dB lower than the true noise density.
It is also specified in the narrowest RBW, 1 Hz, which has a noise bandwidth slightly wider than 1 Hz. These two
effects together add up to 2.25 B.
f. As an example computation, consider this: For the case where DANL = 151 dBm in 1 Hz, IF Gain is set to low,
and the “Additional DANL” is 160 dBm, the total noise density computes to 148.2 dBm/Hz and the Clipping-to-noise ratio for a 10 dBm clipping level is 138.2 dBFS/Hz.
96 Keysight N9030A PXA Specification Guide
I/Q Analyzer
Data Acquisition
Data Acquisition
DescriptionSpecificationsSupplemental Information
Time Record Length
Analysis Tool
IQ Analyzer4,000,000 IQ sample pairsWaveform measurement
Advanced ToolsData Packing89600 VSA software or Fast
b
32-bit 64-bit
Length (IQ sample pairs)536 MSa (2
29
Sa)268 MSa (228 Sa)2 GB total memory
Maximum IQ Capture TimeData Packing
(89600 VSA and Fast Capture)32-bit 64-bit Calculated by: Length of IQ
10 MHz IFBW42.94 s21.47 s
Sample Rate (IQ Pairs)1.25 × IFBW
ADC Resolution
16 bits
Capture
sample pairs/Sample Rate (IQ
c
Pairs)
a
a. This can also be accessed with the remote programming command of "read:wav0?".
b. This can only be accessed with the remote programming command of "init:fcap" in the IQ Analyzer (Basic) wave-
form measurement.
c. For example, using 32-bit data packing at 10 MHz IF bandwidth (IFBW) the Maximum Capture Time is calculated
29
using the formula: "Max Capture Time = (2
)/(10 MHz × 1.25)".
Keysight N9030A PXA Specification Guide 97
I/Q Analyzer
Data Acquisition
98 Keysight N9030A PXA Specification Guide
Keysight X-Series Signal Analyzer
N9030A
Specification Guide
3VXA Vector Signal Analysis Application
This chapter contains specifications for the N9064A1 VXA vector signal
modulation analysis measurement application.
Additional
Definitions and
Requirements
Specs & Nominals These specifications summarize the performance for the X-Series Signal
Because digital communications signals are noise-like, all measurements will
have variations. The specifications apply only with adequate averaging to
remove those variations.
The specifications apply in the frequency range documented in In-Band
Frequency Range.
The specifications for this chapter apply only to instruments with Frequency
Option 503, 508, 513 or 526. For Instruments with higher frequency options,
the performance is nominal only and not subject to any warranted
specifications.
The measurement performance is only slightly different between instruments
with the lower and higher frequency options. Because the hardware
performance of the analyzers is very similar but not identical, you can estimate
the nominal performance of the measurements from the specifications in this
chapter.
Analyzer and apply to the VXA measurement application inside the analyzer.
Values shown in the column labeled "Specs & Nominals" are a mix of warranted
specifications, guaranteed-by-design parameters, and conservative but not
warranted observations of performance of sample instruments.
1. In software versions prior to A.06.00, the VXA measurement application product
number was 89601X. Software versions A.06.00 and beyond have renamed 89601X
to N9064A.
99
VXA Vector Signal Analysis Application
Vector Signal Analysis Performance (N9064A-1FP/1TP)
Vector Signal Analysis Performance (N9064A-1FP/1TP)
Frequency
DescriptionSpecs & NominalsSupplemental Information
RangeSee “Frequency Range” on
page 22
Center Frequency
Tuning Resolution1 mHz
Frequency Span,
Maximum
FFT Spectrum25 MHz (Option B25)
Frequency Points per
Span
FFT Window Type
Window
Flat Top0.410.01 dB>95 dBc
Gaussian Top0.250.68 dB>125 dBc
Hanning0.111.5 dB>31 dBc
10 MHz (standard)
40 MHz (Option B40)
85 MHz (Option B85)
160/140 MHz (Option B1X)
Calibrated points: 51 to 409,601
Displayed points: 51 to 524,288
Passband
Selectivity
Flatness
a
Rejection
The window choices allow the user to
optimize as needed for best amplitude
accuracy, best dynamic range, or best
response to transient signal
characteristics.
Uniform0.00144.0 dB>13 dBc
a. Instruments with 160 MHz bandwidth will display "N9030A-B1Y Analysis bandwidth Extension, 160 MHz"
after pressing: System, Show, System on the instrument.
100 Keysight N9030A PXA Specification Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.