Keysight N1914A, N1913 Service Manual

Page 1
Keysight N1913/1914A EPM Series Power Meters
Service Guide
Page 2
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 2009 - 2016 No part of this manual may be repro-
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
N1913-90009
Edition
Edition 14, November 18, 2016
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight N1913/1914A Service Guide
Page 3

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standard and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standard Organization members.

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. No other warranty is expressed or implied. Keysight Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose.

Exclusive Remedies

The remedies provided herein are buyer's sole and exclusive remedies. Keysight Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.
Keysight N1913/1914A Service Guide 3
Page 4

Equipment Operation

WARNING
CAUTION
Warnings and Cautions
This guide uses warnings and cautions to denote hazards.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Personal Safety Considerations
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited. If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means of protection are intact) only. No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers. For continued protection against fire hazard, replace the line fuse(s) only with fuses of the same type and rating (for example, normal blow, time delay, etc.). The use of other fuses or material is prohibited.
4 Keysight N1913/1914A Service Guide
Page 5

Safety Considerations

WARNING
CAUTION
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
Before this instrument is switched on, make sure it has been properly grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact. Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
Any adjustments or service procedures that require operation of the instrument with protective covers removed should be performed only by trained service personnel.
Keysight N1913/1914A Service Guide 5
Page 6

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC) Alternating current (AC)
Off (mains supply) On (mains supply)
Caution, risk of electric shock
Earth (ground) terminal Frame or chassis (ground) terminal
Protective earth (ground) terminal
Both direct and alternating current Out position of a bi-stable push control
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Equipment protected throughout by double insulation or reinforced insulation
Caution, hot surface In position of a bi-stable push control
Operating switch for ‘Stand-by’
[a]
mode
Equipotentiality
[a] Note: The instrument is NOT isolated from the mains when the switch is pressed. To isolate the instrument, the mains coupler
(mains input cord) should be removed from the power supply.
Three-phase alternating current
6 Keysight N1913/1914A Service Guide
Page 7

Environmental Conditions

The N1913/1914A is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Temperature
Humidity
Altitude
Pollution degree 2
Operating condition – 0 °C to 55 °C Storage condition – –40 °C to 70 °C
Operating condition – Up to 80% RH at 40°C (non-condensing) Storage condition – Up to 95% RH at 65°C (non-condensing)
Operating condition – Up to 4600 meters (15,000 feet) Storage condition – Up to 15,240 meters (50,000 feet)
Keysight N1913/1914A Service Guide 7
Page 8

Regulatory Information

The N1913/1914A complies with the following safety and Electromagnetic Compatibility (EMC) compliances:

General

This product complies with the essential requirements of the following applicable European (EC) Directives, and carries the CE marking accordingly to Low Voltage Directive (2006/95/EC) and EMC Directive (2004/108/EC and 89/336/EEC).

Safety compliance

– IEC 61010-1:2010/EN 61010-1:2010 (3rd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-12
– USA: ANSI/UL 61010-1 (3rd Edition)

EMC compliance

The EMC test conforms to the IEC61326-1:2005/EN61326-1:2006 and CISPR11:2003/EN55011:2007 (Group 1, Class A) standards. In order to preserve the EMC performance of the product, any cable which becomes worn or damaged must be replaced with the same type and specification.
– IEC 61326-1:2005/EN 61326-1:2006
– CISPR11:2003/EN 55011:2007, Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004

Low voltage directive

– This product conforms to the requirements of European Council Directive
“2006/95/EC”.
8 Keysight N1913/1914A Service Guide
Page 9

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
The CSA mark is a registered trademark of the Canadian Standards Association.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
Keysight N1913/1914A Service Guide 9
Page 10

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/epm
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
10 Keysight N1913/1914A Service Guide
Page 11

Table of Contents

Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Equipment Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Safety compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
EMC compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Low voltage directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
1 Specifications and Characteristics
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Specification definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Recommended Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Power Meter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Power Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Power Meter Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . .30
Settling Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Measurement Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Keysight N1913/1914A Service Guide 11
Page 12
Measurement speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Rear Panel Inputs and Output Connections . . . . . . . . . . . . . . . . . . . . . 42
Line power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
1 mW Power Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Battery storage conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2Performance Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Complete Equipment List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
1 mW Power Reference Level Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Output Standing Wave Ratio (SWR) Test . . . . . . . . . . . . . . . . . . . . . . . 53
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Zero Set (Average Path) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Absolute Accuracy Test (Average Path) . . . . . . . . . . . . . . . . . . . . . . . . . 57
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Test Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Linearity (Average Path) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
12 Keysight N1913/1914A Service Guide
Page 13
3Adjustments
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
Power Reference Level Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Test method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
4 Theory of Operation
PPMC Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Main Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Calibrator Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Front Panel Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
PSU Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5 Troubleshooting Guide
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Power-Up Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Basic external checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Basic internal checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Keysight N1913/1914A Service Guide 13
Page 14
Possible faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Instrument Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Extended Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Performance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Power Reference Level Adjustment Problems . . . . . . . . . . . . . . . . . . . 76
Possible faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Communication Interface Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Additional Diagnostic Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
6Repair Guide
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Front panel assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Rear panel assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Additional spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Calibrator assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Improved calibrator assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Outer housing components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Sundries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Required Torque Values for Fasteners . . . . . . . . . . . . . . . . . . . . . . . . 105
Disassembly Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Reassembly Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Disassembly vs Part Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Front Panel Disassembly Instructions . . . . . . . . . . . . . . . . . . . . . . . . . 119
Front Panel Reassembly Instructions . . . . . . . . . . . . . . . . . . . . . . . . . 124
Additional Repair Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Replacing the sensor flex assembly . . . . . . . . . . . . . . . . . . . . . . . . 128
Main Board vs. Rear Panel Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 129
Replacing the PPMC Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Replacing the Calibrator Semi-Rigid/Split Ferrite . . . . . . . . . . . . . . . 131
14 Keysight N1913/1914A Service Guide
Page 15
7 Contacting Keysight Technologies
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .134
Contacting Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . .134
Before Calling Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . .135
Check the Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .136
Instrument Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .137
Returning Your Power Meter for Service . . . . . . . . . . . . . . . . . . . . . . .138
Useful Web Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Keysight N1913/1914A Service Guide 15
Page 16
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
16 Keysight N1913/1914A Service Guide
Page 17

List of Figures

Figure 1-1 8480 Series settling time with auto-filter . . . . . . . . . .35
Figure 1-2 E441x Series settling time with auto-filter . . . . . . . . .36
Figure 1-3 E9300 Series settling time with auto-filter . . . . . . . . .37
Figure 1-4 N8480 Series settling time with auto-filter . . . . . . . . .38
Figure 1-5 U2000 Series settling time with auto-filter . . . . . . . . . 40
Figure 2-1 1 mW power reference level test setup connection
Figure 2-2 System calibration connection diagram . . . . . . . . . . .54
Figure 2-3 Output SWR test setup-open connection diagram . . . 54 Figure 2-4 Output SWR test setup-load connection diagram . . .54
Figure 2-5 Absolute accuracy test setup . . . . . . . . . . . . . . . . . . . .58
Figure 6-1 Calibrator assembly and improved calibrator assembly .
Figure 6-2 Separate the clamshells . . . . . . . . . . . . . . . . . . . . . . .106
Figure 6-3 Remove the top clamshell . . . . . . . . . . . . . . . . . . . . .107
Figure 6-4 Remove the PSU safety cover and cable guide . . . . .107
Figure 6-5 Remove the PSU . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
Figure 6-6 Top view with top clamshell removed . . . . . . . . . . . .109
Figure 6-7 Remove the EMI earth wires . . . . . . . . . . . . . . . . . . .110
Figure 6-8 Disconnect the front panel cable . . . . . . . . . . . . . . . .110
Figure 6-9 Disconnect the semi-rigid cable . . . . . . . . . . . . . . . .110
Figure 6-10 Remove the front panel . . . . . . . . . . . . . . . . . . . . . . .111
Figure 6-11 Remove PPMC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111
Figure 6-12 Remove the main board . . . . . . . . . . . . . . . . . . . . . . .112
Figure 6-13 Remove the rear panel . . . . . . . . . . . . . . . . . . . . . . . .112
Figure 6-14 Analog recorder output connections . . . . . . . . . . . . .113
Figure 6-15 Sensor flex connections . . . . . . . . . . . . . . . . . . . . . . .114
Figure 6-16 PSU screw locations . . . . . . . . . . . . . . . . . . . . . . . . . .115
Figure 6-17 PSU cable routing . . . . . . . . . . . . . . . . . . . . . . . . . . .116
Figure 6-18 Main board revision markings . . . . . . . . . . . . . . . . . .129
Figure 6-19 Separate the two halves of the ferrite . . . . . . . . . . . .131
diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
100
Keysight N1913/1914A Service Guide 17
Page 18
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
18 Keysight N1913/1914A Service Guide
Page 19

List of Tables

Table 1-1 Zero set specifications . . . . . . . . . . . . . . . . . . . . . . . . .26
Table 1-2 Zero set (internal and external) for U2000 Series . . . .28
Table 1-3 Noise multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Table 1-4 Power sensor specifications . . . . . . . . . . . . . . . . . . . . .30
Table 1-5 U2000 Series power sensors specification . . . . . . . . .33
Table 1-6 8480 Series settling time . . . . . . . . . . . . . . . . . . . . . . .35
Table 1-7 E441x and E9300 Series settling time . . . . . . . . . . . . .36
Table 1-8 N8480 Series settling time . . . . . . . . . . . . . . . . . . . . . .38
Table 1-9 U2000 Series settling time . . . . . . . . . . . . . . . . . . . . . .39
Keysight N1913/1914A Service Guide 19
Page 20
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
20 Keysight N1913/1914A Service Guide
Page 21
Keysight N1913/1914A EPM Series Power Meters Service Guide
1 Specifications and
Characteristics
Introduction 22
Recommended Calibration Interval 23
Power Meter Specifications 24
Power Sensor Specifications 25
Power Meter Supplemental Characteristics 30
Settling Time 35
Measurement Characteristics 41
Rear Panel Inputs and Output Connections 42
1 mW Power Reference 43
Battery storage conditions 44
Physical Characteristics 45
This chapter describes the specifications and characteristics of your EPM Series Power Meters.
21
Page 22
1 Specifications and Characteristics

Introduction

This chapter details the EPM Series Power Meters specifications and supplemental characteristics.

Specification definitions

There are two types of product specifications:
– Warranted specifications
– Characteristic specifications
Warranted specifications
Warranted specifications are covered by the product warranty and apply after a 30 minutes warm- up. These specifications are valid over the power meter’s operating and environmental range unless otherwise stated and after performing a zero and calibration.
Characteristic specifications
Supplemental characteristics which are shown in italics are intended to provide information useful in applying the power meter by giving typical, but non-warranted performance parameters. These characteristics are shown in italics or denoted as “typical”, “nominal”, or “approximate”.
Characteristic information is representative of the product. In many cases, it may also be supplemental to a warranted specification. Characteristic specifications are not verified on all power meters. The types of characteristic specifications can be placed in two groups:
– The first group of characteristic types describes 'attributes' common to all
products of a given model or option.
Examples of characteristics that describe 'attributes' are product weight, and 50 Ω input N-type connector. In these examples, product weight is an approximate value and a 50 Ω input is nominal. These two terms are most widely used when describing a product's 'attributes'.
– The second group of characteristic types describes 'statistically' the aggregate
performance of the population of products.
22 Keysight N1913/1914A Service Guide
Page 23
These characteristics describe the expected behavior of the population of
products. They do not guarantee the performance of any individual product.
No measurement uncertainty value is accounted for in the specification. These
specifications are referred to as typical.
Conditions
The power meter and power sensor meet its specifications when:
– Stored for a minimum of two hours at a stable temperature within the
operating temperature range, and turned on for at least 30 minutes.
– The power meter and power sensor are within their recommended calibration
periods.
– Used in accordance to the information provided in the Keysight N1913/1914A
EPM Series Power Meters User's Guide.

Recommended Calibration Interval

Keysight Technologies recommends a two-year calibration cycle for the N1913/ 1914A EPM Series Power Meters.
Specifications and Characteristics 1
Keysight N1913/1914A Service Guide 23
Page 24
1 Specifications and Characteristics

Power Meter Specifications

Frequency Range
9 kHz to 110 GHz, power sensor dependent
Power Range
–70 dBm to +44 dBm (100 pW to 25 W), power sensor dependent
Power Sensors Compatibility
– Keysight 8480 Series power sensors
– Keysight E9300 E-Series average power sensors
– Keysight E4410 E-Series average power sensors
– Keysight N8480 Series power sensors
– Keysight U2000 Series average USB power sensors
Single Sensor Dynamic Range
– 90 dB maximum (Keysight E-Series power sensors)
– 50 dB maximum (Keysight 8480 Series power sensors)
– 55 dB maximum (Keysight N8480 Series power sensors)
– 80 dB maximum (Keysight U2000 Series USB power sensors)
Display Units
Absolute: Watts (W) or dBm
Relative: Percent (%) or dB
Display Resolution
Selectable resolution of: 1.0, 0.1, 0.01, and 0.001 dB in logarithmic mode; or 1, 2, 3, and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode or three digits in linear mode
24 Keysight N1913/1914A Service Guide
Page 25

Power Sensor Specifications

Definitions

Zero Set
In any power measurement, the power meter must initially be set to zero with no power applied to the power sensor. Zero setting is accomplished within the power meter by digitally correcting for residual offsets.
Zero Drift
This parameter is also called long term stability and is the change in the power meter indication over a long time (usually one hour) for a constant input power at a constant temperature, after a defined warm-up interval.
Measurement Noise
This parameter is also known as short term stability and is specified as the change in the power meter indication over a short time interval (usually one minute) for a constant input power at a constant temperature.
Specifications and Characteristics 1

Accuracy

Instrumentation
Absolute accuracy power sensor linearity specification in your power sensor manual to assess overall system accuracy.)
Relative accuracy sensor linearity specification in your power sensor manual to assess overall system accuracy.)
Zero Set (digital settability of zero): Power sensor dependent (refer to Tab le 1-1 and Tab le 1-2). For Keysight E-Series power sensors, this specification applies when zeroing is performed with the sensor input disconnected from the POWER REF.
[1] Refer to the power sensor linearity specification in your power sensor manual to assess overall
system accuracy.
Keysight N1913/1914A Service Guide 25
[1]
: ±0.02 dB (Logarithmic) or ±0.5% (Linear). (Refer to the
[1]
: ±0.04 dB (Logarithmic) or ±1.0% (Linear). (Refer to the power
Page 26
1 Specifications and Characteristics
Tab le 1-1 Zero set specifications
Power sensor
[b]
8481A
[b]
8481B
[b]
8481D
[b]
8481H
[b]
8482A
[b]
8482B
[b]
8482H
[b]
8483A
[b]
8485A
[b]
8485D
[b]
R8486A
[b]
R8486D
[b]
Q8486A
[b]
Q8486D
[b]
V8486A
[b]
W8486A
[b]
8487A
[b]
8487D
Zero set
±50 nW
±50 mW
±20 pW
±5 mW
±50 nW
±50 mW
±5 mW
±50 nW
±50 nW
±20 pW
±50 nW
±30 pW
±50 nW
±30 pW
±200 nW
±200 nW
±50 nW
±20 pW
[a]
E4412A ±50 pW
E4413A ±50 pW
E9300A ±500 pW
E9301A ±500 pW
E9304A ±500 pW
26 Keysight N1913/1914A Service Guide
Page 27
Specifications and Characteristics 1
Table 1 -1 Zero set specifications (continued)
Power sensor
Zero set
E9300B ±500 nW
E9301B ±500 nW
E9300H ±5 nW
E9301H ±5 nW
N8481A (exclude Option CFT) ±25 nW
N8482A (exclude Option CFT)
N8485A (exclude Option CFT)
N8486A R (exclude Option CFT)
N8486A Q (exclude Option CFT)
N8487A (exclude Option CFT)
N8488A (exclude Option CFT)
N8481B (exclude Option CFT)
N8482B (exclude Option CFT)
N8481H (exclude Option CFT)
N8482H (exclude Option CFT)
N8481A with Option CFT
N8482A with Option CFT
N8485A with Option CFT
N8486A R with Option CFT
N8486A Q with Option CFT
N8487A with Option CFT
N8481B with Option CFT
N8482B with Option CFT
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
[b]
±25 nW
±25 nW
±25 nW
±25 nW
±25 nW
±25 nW
±25 µW
±25 µW
±2.5 µW
±2.5 µW
±63 nW
±63 nW
±63 nW
±63 nW
±63 nW
±63 nW
±63 µW
±63 µW
[a]
Keysight N1913/1914A Service Guide 27
Page 28
1 Specifications and Characteristics
Tab le 1-1 Zero set specifications (continued)
Power sensor
N8481H with Option CFT
N8482H with Option CFT
[a] The zero set specifications are tested with Keysight 11730A power sensor cable, 1.5 m (7.5 ft).
[b] The zero set specifications are tested at 50 MHz.
[b]
[b]
Zero set
±6.3 µW
±6.3 µW
[a]
Tab le 1-2 Zero set (internal and external) for U2000 Series
Power sensor Range Zero set (internal) Zero set (external)
–60 dBm to –35 dBm ±1.5 nW ±600 pW
–38 dBm to –15 dBm ±2 nW ±1.5 nW
U2000/1/2A
U2004A
–20 dBm to –9 dBm ±12 nW ±10 nW
–11 dBm to –5 dBm ±2 µW ±500 nW
–7 dBm to 15 dBm ±4 µW ±1 µW
10 dBm to 20 dBm ±6 µW ±5 µW
–60 dBm to –35 dBm ±2.8 nW ±600 pW
–38 dBm to –15 dBm ±3 nW ±1.5 nW
–20 dBm to –9 dBm ±12 nW ±10 nW
–11 dBm to –5 dBm ±2 µW ±500 nW
–7 dBm to 15 dBm ±4 µW ±1 µW
10 dBm to 20 dBm ±6 µW ±5 µW
–50 dBm to –25 dBm ±15 nW ±8 nW
–28 dBm to –5 dBm ±20 nW ±20 nW
–10 dBm to 1 dBm ±120 nW ±100 nW
U2000/1/2H
–1 dBm to 5 dBm ±20 µW ±20 µW
3 dBm to 25 dBm ±40 µW ±30 µW
20 dBm to 30 dBm ±60 µW ±60 µW
28 Keysight N1913/1914A Service Guide
Page 29
Specifications and Characteristics 1
NOTE
Table 1 -2 Zero set (internal and external) for U2000 Series (continued)
Power sensor Range Zero set (internal) Zero set (external)
–30 dBm to –5 dBm ±1.8 µW ±800 nW
–8 dBm to 15 dBm ±2 µW ±2 µW
U2000/1B
10 dBm to 21 dBm ±12 µW ±10 µW
19 dBm to 25 dBm ±2 mW ±1 mW
23 dBm to 44 dBm ±4 mW ±2 mW
The zero set specifications are only applicable to Keysight U2000 Series USB power sensors with serial prefixes as shown below:
– U2000A serial prefix MY480/SG480 and above
– U2001A serial prefix MY481/SG481 and above
– U2002A serial prefix MY482/SG482 and above
– U2004A serial prefix MY484/SG484 and above
For power sensors with earlier prefixes, refer to the Keysight U2000 Series USB Power Sensors Operating and Service Guide.
Keysight N1913/1914A Service Guide 29
Page 30
1 Specifications and Characteristics

Power Meter Supplemental Characteristics

Zero Drift of Sensors
This parameter is also called long term stability and is the change in the power meter indication over a long time (within one hour) at a constant temperature after a 24-hour warm-up of the power meter.
Power sensor dependent (refer to Tab le 1-4).
Measurement Noise
Power sensor dependent (refer to Tab le 1-3 and Tab le 1-4).
Averaging effects on measurement noise. Averaging over 1 to 1024 readings is available for reducing noise. Table 1-4 provides the measurement noise for a particular power sensor with the number of averages set to 16 for normal mode and 32 for x2 mode. Use the “Noise Multiplier” for the appropriate mode (normal or x2) and number of averages to determine the total measurement noise value.
For example, for a Keysight 8481D power sensor in normal mode with the number of averages set to 4, the measurement noise is equal to:
(<45 pW × 2.75) = <124 pW
Tab le 1-3 Noise multiplier
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Noise multiplier (Normal mode)
Noise multiplier (x2 mode)
5.5 3.89 2.75 1.94 1.0 0.85 0.61 0.49 0.34 0.24 0.17
6.5 4.6 3.25 2.3 1.63 1.0 0.72 0.57 0.41 0.29 0.2
Tab le 1-4 Power sensor specifications
Power sensor
Keysight 8481A
Keysight 8481B
Keysight 8481D
Zero drift
[d]
[d]
[d]
[a]
Measurement
<±10 nW <110 nW
<±10 mW <110 mW
<±4 pW <45 pW
[b]
noise
30 Keysight N1913/1914A Service Guide
Page 31
Specifications and Characteristics 1
Table 1 -4 Power sensor specifications (continued)
Power sensor
Keysight 8481H
Keysight 8482A
Keysight 8482B
Keysight 8482H
Keysight 8483A
Keysight 8485A
Keysight 8485D
Keysight R8486A
Keysight R8486D
Keysight Q8486A
Keysight Q8486D
Keysight V8486A
Keysight W8486A
Keysight 8487A
Keysight 8487D
Zero drift
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[a]
Measurement
<±1 mW <10 mW
<±10 nW <110 nW
<±10 mW <110 mW
<±1 mW <10 mW
<±10 nW <110 nW
<±10 nW <110 nW
<±4 pW <45 pW
<±10 nW <110 nW
<±6 pW <65 pW
<±10 nW <110 nW
<±6 pW <65 pW
<±40 nW <450 nW
<±40 nW <450 nW
<±10 nW <110 nW
<±4 pW <45 pW
Keysight E4412A <±15 pW <70 pW
Keysight E4413A <±15 pW <70 pW
Keysight E9300A
Keysight E9301A
Keysight E9304A
Keysight E9300B
Keysight E9301B
Keysight E9300H
[c]
[c]
[c]
[c]
[c]
[c]
<±150 nW <700 nW
<±150 nW <700 nW
<±150 nW <700 nW
<±150 nW <700 nW
<±150 nW <700 nW
<±1.5 nW <7 nW
[b]
noise
Keysight N1913/1914A Service Guide 31
Page 32
1 Specifications and Characteristics
Tab le 1-4 Power sensor specifications (continued)
Power sensor
Keysight E9301H
Keysight N8481A (exclude Option CFT)
Keysight N8482A (exclude Option CFT)
Keysight N8485A (exclude Option CFT)
[c]
[d]
[d]
[d]
Keysight N8486A R (exclude Option CFT)
Keysight N8486A Q (exclude Option CFT)
Keysight N8487A (exclude Option CFT)
Keysight N8488A (exclude Option CFT)
Keysight N8481B (exclude Option CFT)
Keysight N8482B (exclude Option CFT)
Keysight N8481H (exclude Option CFT)
Keysight N8482H (exclude Option CFT)
Keysight N8481A with Option CFT
Keysight N8482A with Option CFT
Keysight N8485A with Option CFT
Keysight N8486A R with Option CFT
Keysight N8486A Q with Option CFT
Keysight N8487A with Option CFT
Keysight N8481B with Option CFT
Keysight N8482B with Option CFT
Keysight N8481H with Option CFT
Keysight N8482H with Option CFT
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
[d]
Zero drift
[a]
Measurement
[b]
noise
<±1.5 nW <7 nW
<±3 nW <80 nW
<±3 nW <80 nW
<±3 nW <80 nW
[d]
[d]
<±3 nW <80 nW
<±3 nW <80 nW
<±3 nW <80 nW
<±3 nW <80 nW
<±3 µW <80 µW
<±3 µW <80 µW
<±0.3 µW <8 µW
<±0.3 µW <8 µW
<±7 nW <114 nW
<±7 nW <114 nW
<±7 nW <114 nW
<±7 nW <114 nW
<±7 nW <114 nW
<±7 nW <114 nW
<±7 µW <114 µW
<±7 µW <114 µW
<±0.7 µW <11.4 µW
<±0.7 µW <11.4 µW
32 Keysight N1913/1914A Service Guide
Page 33
Specifications and Characteristics 1
[a] Within one hour after zero set, at a constant temperature after a 24-hour warm-up of the power meter.
[b] The number of averages at 16 (for normal mode) and 32 (for x2 mode), at a constant temperature, measured over a
one minute interval and two standard deviations. For Keysight E-Series power sensors, the measurement noise is
measured within the low range. Refer to the relevant power sensor manual for further information.
[c] Specification applies to the low power path, up to 75% relative humidity.
[d] The zero drift and measurement noise specifications are tested at 50 MHz.
Table 1 -5 U2000 Series power sensors specification
Power sensor Range
–60 dBm to –35 dBm 200 pW 1 nW
–38 dBm to –15 dBm 400 pW 1.5 nW
–20 dBm to –9 dBm 1.5 nW 15 nW
U2000/1/2A
–11 dBm to –5 dBm 50 nW 650 nW
–7 dBm to 15 dBm 500 nW 1 µW
10 dBm to 20 dBm 2 µW 10 µW
–60 dBm to –35 dBm 200 pW 1 nW
–38 dBm to –15 dBm 400 pW 1.5 nW
–20 dBm to –9 dBm 1.5 nW 15 nW
U2004A
–11 dBm to –5 dBm 50 nW 650 nW
–7 dBm to 15 dBm 500 nW 1 µW
10 dBm to 20 dBm 2 µW 10 µW
–50 dBm to –25 dBm 2 nW 10 nW
–28 dBm to –5 dBm 4 nW 15 nW
–10 dBm to 1 dBm 15 nW 150 nW
U2000/1/2H
–1 dBm to 5 dBm 500 nW 6.5 µW
3 dBm to 25 dBm 5 µW 10 µW
20 dBm to 30 dBm 20 µW 100 µW
Zero drift
[a]
Measurement
[b]
noise
Keysight N1913/1914A Service Guide 33
Page 34
1 Specifications and Characteristics
NOTE
Tab le 1-5 U2000 Series power sensors specification (continued)
Power sensor Range
–30 dBm to –5 dBm 200 nW 1 µW
–8 dBm to 15 dBm 400 nW 1.5 µW
U2000/1B
[a] Within one hour after zero set, at a constant temperature after a 24-hour warm-up of the power meter.
[b] The number of averages at 16 (for normal mode) and 32 (for x2 mode), at a constant temperature, measured over a
one minute interval and two standard deviations. For Keysight E-Series power sensors, the measurement noise is measured within the low range. Refer to the relevant power sensor manual for further information.
10 dBm to 21 dBm 1.5 µW 15 µW
19 dBm to 25 dBm 50 nW 650 µW
23 dBm to 44 dBm 500 µW 1 mW
Zero drift
[a]
Measurement
[b]
noise
The zero drift and measurement noise specifications are only applicable to Keysight U2000 Series USB power sensors with serial prefixes as shown below:
– U2000A serial prefix MY480/SG480 and above
– U2001A serial prefix MY481/SG481 and above
– U2002A serial prefix MY482/SG482 and above
– U2004A serial prefix MY484/SG484 and above
For power sensors with earlier prefixes, refer to the Keysight U2000 Series Operating and Service Guide.
34 Keysight N1913/1914A Service Guide
Page 35

Settling Time

+20 dB
+10 dB
150 ms
350 ms
550 ms
3.5 s
+10 dB
Normal
mode
150 ms
500 ms
1.1 s
6.6 s
x2
mode
Sensor dynamic range
Typical settling times
+10 dB
Max dBm
Min dBm
Specifications and Characteristics 1
For Keysight 8480 Series power sensors
0 to 99% settled readings over the GPIB.
Manual filter, 10 dB decreasing power step (refer to Table 1-6).
Auto filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes (refer to Figure 1-1).
Table 1 -6 8480 Series settling time
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time (s) (Normal mode)
Settling time (s) (x2 mode)
0.15 0.2 0.3 0.5 1.1 1.9 3.4 6.6 13 27 57
0.15 0.18 0.22 0.35 0.55 1.1 1.9 3.5 6.9 14.5 33
Figure 1-1 8480 Series settling time with auto-filter
For Keysight E-Series power sensors
For E441X Series and E9300 Series power sensors in normal and x2 speed modes, manual filter, 10 dB decreasing power step (refer to Table 1-7).
Keysight N1913/1914A Service Guide 35
Auto-filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes (refer to Figure 1-2 for E441X Series sensors and Figure 1-3 for E9300 Series sensors).
Page 36
1 Specifications and Characteristics
+50 dB
+10 dB
70 ms
90 ms
450 ms
3.5 s
+10 dB
Normal
mode
80 ms
130 ms
1.1 s
6.7 s
x2
mode
Sensor dynamic range
Typical settling times
+10 dB
Max dBm
Min dBm
Tab le 1-7 E441x and E9300 Series settling time
[a]
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time (s) (Normal mode)
Settling time (s) (x2 mode)
[a] E-Series power sensors in Fast mode (using free run trigger), within the range –50 dBm to +17 dBm, the settling
time is:
N1913A: 10 ms N1914A: 20 ms
0.08 0.13 0.24 0.45 1.1 1.9 3.5 6.7 14 27 57
0.07 0.09 0.15 0.24 0.45 1.1 1.9 3.5 6.7 14 27
Figure 1-2 E441x Series settling time with auto-filter
36 Keysight N1913/1914A Service Guide
Page 37
Specifications and Characteristics 1
+10 dBm
+2 dBm
40 ms
120 ms
210 ms
400 ms
–4 dBm
x2
mode
70 ms
210 ms
400 ms
1 s
Normal
mode
Sensor dynamic range
Typical settling times
–10 dBm
Max dBm
Min dBm
70 ms
120 ms
1 s
6.5 s
13 s
40 ms
70 ms
–20 dBm
–30 dBm
–40 dBm
–50 dBm
400 ms
3.4 ms
6.8 s
High power path
Low power path
Figure 1-3 E9300 Series settling time with auto-filter
Keysight N1913/1914A Service Guide 37
Page 38
1 Specifications and Characteristics
+20 dB
+10 dB
150 ms
180 ms
400 ms
6.6 s
+10 dB
Normal
mode
150 ms
200 ms
1 s
13.5 s
x2
mode
Sensor dynamic range
Typical settling times
+10 dB
Max dBm
Min dBm
3.6 s +10 dB6.6 s
For Keysight N8480 Series power sensors
Typical Settling time: 0 to 99% settled readings over the GPIB.
Auto filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes (refer to Figure 1-4). Manual filter, 10 dB decreasing power step (refer to
Table 1-8).
Tab le 1-8 N8480 Series settling time
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time (s) (Normal mode)
Settling time (s) (x2 mode)
0.15 0.2 0.3 0.5 1.1 1.9 3.4 6.6 13 27 57
0.15 0.18 0.22 0.35 0.55 1.1 1.9 3.5 6.9 14.5 33
Figure 1-4 N8480 Series settling time with auto-filter
38 Keysight N1913/1914A Service Guide
Page 39
Specifications and Characteristics 1
For Keysight U2000 Series power sensors
In FAST mode (using Free Run trigger), for a 10 dB decreasing power step, the settling time is 25 ms
Table 1 -9 U2000 Series settling time
[1]
.
Number of averages
Settling time (Normal mode)
Settling time (x2 mode)
[a] Manual filter, 10 dB decreasing power step (not across the switching point)
[a]
[a]
1 2 4 8 16 32 64 128 256 512 1024
(s)
0.045 0.09 0.17 0.34 0.66 1.3 2.6 5.2 10.4 20.9 41.9
(s)
0.042 0.05 0.09 0.17 0.34 0.66 1.3 2.6 5.2 10.4 20.9
[1] When a power step crosses the auto-range switch point of the sensor, add 25 ms.
Keysight N1913/1914A Service Guide 39
Page 40
1 Specifications and Characteristics
-50 dBm
+10 dBm
+2 dBm
-40 dBm
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
-4 dBm
-10 dBm
-20 dBm
-30 dBm
Normal
mode
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
x2
mode
Maximum sensor power
within a range
U2000/1/2/4A
U2000/1/2H U2000/1B
+20 dBm +40 dBm
+12 dBm +32 dBm
+6 dBm +26 dBm
0 dBm
-10 dBm
-20 dBm
-30 dBm
-40 dBm
+20 dBm
+10 dBm
0 dBm
-10 dBm
-20 dBm
Minimum sensor power
within a range
Sensor dynamic range
Typical settling times
0 dBm
0 dBm
+12 dBm
+10 dBm
Figure 1-5 U2000 Series settling time with auto-filter
40 Keysight N1913/1914A Service Guide
Page 41

Measurement Characteristics

Measurement speed

Over the GPIB, three measurement speed modes are available as shown, along with the typical maximum measurement speed for each mode:
– Normal: 20 readings/second
– x2: 40 readings/second
–Fast
Maximum measurement speed is obtained using binary output in free run trigger mode.
[1]
: 400 readings/second, for Keysight E-Series power sensors only
Specifications and Characteristics 1
[1] For N1914A, if both channels are used in the fast mode, the measurement speed will be reduced
to 200 readings/second for each channel.
Keysight N1913/1914A Service Guide 41
Page 42
1 Specifications and Characteristics

Rear Panel Inputs and Output Connections

Recorder output(s)
GPIB USB 2.0 10/100Base-T LAN
Trigger input (optional)
Trigger out (optional)
Ground Binding post, accepts 4 mm plug or bare wire connection
USB host (optional) To connect U2000 Series power sensors
VGA out (optional)

Line power

Input voltage range
Input frequency range
Power requirement 70 VA (maximum)
Analog 0 to 1 V, 1 k
Interfaces allow communication with an external controller
Input has TTL compatible logic levels and uses a BNC connector
Output provides TTL compatible logic levels and uses a BNC connector
Standard 15-pin VGA connector, allows connection of external VGA monitor
100 – 240 Vac 100 – 120 Vac Automatic voltage selection Fluctuations not exceeding ±10%
50 – 60 Hz (100 to 240 Vac) 400 Hz (100 to 120 Vac)
Ω output impedance, BNC connectors
42 Keysight N1913/1914A Service Guide
Page 43

1 mW Power Reference

NOTE
The 1 mW Power Reference is provided for calibration of the E-Series, 8480 Series, and N8480 Series power sensors.
Power output
Accuracy
Frequency 50 MHz nominal
SWR
Connector type Type N (f), 50 Ω
Specifications and Characteristics 1
1.00 mW (0.0 dBm) Factory set to ±0.4% traceable to the National Physical
Laboratories (NPL), UK
±1.2% (0 – 55 ºC) ±0.4% (25 ±10 ºC)
1.08 (0 - 55 ºC)
1.05 (typical)
Keysight N1913/1914A Service Guide 43
Page 44
1 Specifications and Characteristics
NOTE

Battery storage conditions

Storage temperature limits: –20°C to 60°C, 80% RH
Refer to “Battery Information (Optional)” on page 90 of Keysight N1913/1914A EPM Series Power Meters User’s Guide for more details on the battery pack.
44 Keysight N1913/1914A Service Guide
Page 45

Physical Characteristics

Dimensions

The following dimensions exclude front and rear panel protrusions:
212.6 mm W x 88.5 mm H x 348.3 mm D
(8.5 in x 3.5 in x 13.7 in)

Weight

Specifications and Characteristics 1
N1913/1914A weight (net)
N1913/1914A weight (shipping)
3.60 kg (approximately)
8.20 kg (approximately)
Keysight N1913/1914A Service Guide 45
Page 46
1 Specifications and Characteristics
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
46 Keysight N1913/1914A Service Guide
Page 47
Keysight N1913/1914A EPM Series Power Meters Service Guide

2 Performance Tests

Introduction 48
Complete Equipment List 49
1 mW Power Reference Level Test 51
Output Standing Wave Ratio (SWR) Test 53
Zero Set (Average Path) 56
Absolute Accuracy Test (Average Path) 57
Linearity (Average Path) 61
This chapter contains procedures that allows you to test the power meter’s electrical performance to its specifications.
47
Page 48
2 Performance Tests
NOTE

Introduction

The performance tests described in this chapter test the power meter’s electrical performance against the specifications detailed in Chapter 1. They are used for incoming inspection, during calibration cycle (also called periodic maintenance), or after repairs have been made.
– This document does not provide a complete breakdown for these tests; it only
gives a brief overview of each, in line with Keysight’s recommendation that the Keysight N7800 Series calibration software should be used at all times.
– Performance testing is limited to the measurement and verification of
warranted specifications.
– Some tests cannot be performed manually, and so the N7800 Series
calibration software is essential.
– Measurement uncertainty will not be addressed in this document (this is
handled by the N7800 Series software).
The following performance tests are described in this chapter:
“1 mW Power Reference Level Test” on page 51
“Output Standing Wave Ratio (SWR) Test” on page 53
“Zero Set (Average Path)” on page 56
“Absolute Accuracy Test (Average Path)” on page 57
“Linearity (Average Path)” on page 61
48 Keysight N1913/1914A Service Guide
Page 49

Complete Equipment List

Performance Tests 2
Instrument Critical specifications Recommended Keysight
model number
Meters
Power Meter Dual channel
Absolute accuracy: 0.5%
Power reference accuracy: 0.9% – (a best
capability measurement is required for the
power reference output — the power level
must be accurately measured, and the
uncertainty of this measurement must also be
known)
Power Sensor Frequency: 50 MHz
Amplitude range: –70 dBm to –20 dBm
SWR: 1.15 at 50 MHz
Power Sensor Frequency: 50 MHz
Amplitude range: –30 dBm to +20 dBm
SWR: 1.1 at 50 MHz
Power Sensor Frequency: 50 MHz
Amplitude range: –70 dBm to +20 dBm
SWR: 1.06 maximum
Attenuator
20 dB Fixed Attenuator N-type (m,f) 8491B (Option 020)
E4419B E4419A
8481D
N8482A 8482A
E4412A
Alternative Keysight model number
Keysight N1913/1914A Service Guide 49
Page 50
2 Performance Tests
Instrument Critical specifications Recommended Keysight
model number
Miscellaneous
Power Splitter Frequency: DC to 18 GHz insertion
Loss: 7 dB Equivalent output SWR: 1.20 DC to 8 GHz
Coaxial Termination DC to 18 GHz 909A
BNC Cable Cable Assy-Coaxial RG/223 cable 50-Ohm
straight BNC male to straight BNC 24-in LG
Calibration Test Cable required for N1913A and N1914A
Sensor Cable 85032B
N-Type Calibration Kit 85032B
Assorted accessories (cables and adapters) required
11667A (Option 001)
8120-1839
11730A
Alternative Keysight model number
50 Keysight N1913/1914A Service Guide
Page 51

1 mW Power Reference Level Test

Ch A
N1913/1914A
POWER METER
E4419B/E4417A POWER METER
NOTE

Description

The 1 mW power reference is used for the calibration of 8480 Series, N8480 Series and E-Series power sensors, and is traceable to national standards. This test uses an N8482A power sensor to transfer the power measured on an accurately calibrated E4419B or E4417A power meter to the DUT reference.

Equipment

– Required test equipment:
– 1 unit of E4419B or E4417A dual channel power meter
– 1 unit of N8482A power sensor
– 11730A power sensor cable
– Either of these E4419B or E4417A power meters can be used. This specific
power meter model must be used.
Performance Tests 2
Figure 2-1 1 mW power reference level test setup connection diagram
For rear panel options, the connections will differ from the illustration shown here. Refer to the connector identification markings on the rear panel for further details.
Keysight N1913/1914A Service Guide 51
Page 52
2 Performance Tests
NOTE

Test method

1 Enter the recorded measurement uncertainty of the E4419B or E4417A 1 mW
power reference.
2 Using the E4419B or E4417A power meter and the N8482A sensor, measure
the 1 mW power reference of the E4419B or E4417A.
3 Using the E4419B or E4417A power meter and the N8482A sensor, measure
the 1 mW power reference of the DUT.
4 Using all of these values, the N7800 Series software will calculate the power
reference level of the DUT.
– The 1 mW reference of the E4419B or E4417A power meter must be precisely
calibrated at a standards accredited lab, and the uncertainty of this measurement known.
– Anyone who has a basic understanding of metrology should be able to
perform this test manually; it is simply the transfer of known power level with a known calibration uncertainty to the DUT.
– An adjustment is available for this test if it fails (see Chapter 3,
"Adjustments").
52 Keysight N1913/1914A Service Guide
Page 53

Output Standing Wave Ratio (SWR) Test

Description

Connector mismatch is the largest single contributor to measurement uncertainty, so this specification must be warranted to provide assurance of instrument accuracy. The 1 mW power reference level test must be carried out prior to this test, as the VSWR specification is only valid at 1 mW. This test measures VSWR by equating relative powers (measured by the test system power meter and its sensors) when the power reference is exercised under different load conditions.

Equipment

– Required test equipment:
– 1 unit of 8753ES/ET network analyzer
– 1 unit of 85032B Type N calibration kit
– 1 unit of E4419B or E4417A dual channel power meter
– 2 units of 8481D power sensor
Performance Tests 2
– 2 units of 11667A #001 power splitter
– 1 unit of 20 dB pad, male to female (e. g. 8491A)
– 1 unit of 30 dB pad (e. g. 11708A reference attenuator)
– 2 units of 11730A power sensor cable
– An alternative network analyzer can be used, as long as it can measure S11 in
the 45 MHz to 55 MHz range
– These specific models of power sensors and power splitters must be used
– Any type of pad can be used (as long as there are no additional mating
connections, or differing pad values)
– 1 unit of 11667A, 1 unit of 8481D, and the 30 dB pad combine to create the
‘Calibration System’
– 1 unit of 11667A, 1 unit of 8481D, and the 20 dB pad combine to create the
‘Measurement System’
Keysight N1913/1914A Service Guide 53
Page 54
2 Performance Tests
POWER METER
POWER SENSOR
Load Port
20 dB
TYPE N MALE PORT
ATTE NUATOR
30 dB
ATTENUATOR
POWER SPLITTERS
Ch A
CH B
Ch A
POWER METER
EPM SERIES
POWER METER
POWER SENSOR
N-TYPE MALE OPEN
Load Port
20 dB
ATTE NUATO R
Ch A
POWER METER
POWER SENSOR
N-TYPE MALE LOAD
Load Port
20 dB
ATTE NUATO R
EPM SERIES
POWER METER
Figure 2-2 System calibration connection diagram
Figure 2-3 Output SWR test setup-open connection d iagram
54 Keysight N1913/1914A Service Guide
Figure 2-4 Output SWR test setup-load connection d iagram
Page 55
NOTE
For rear panel options, the connections will differ from the illustration shown
NOTE
here. Refer to the connector identification markings on the rear panel for further details.

Test method

1 Obtain the S11 parameter of the calibration system.
2 Connect the measurement system to the calibration system and obtain its S21
(load) and S21 (open) parameters.
3 Using only the measurement system, terminated with the OPEN connector
from the 85032B calibration kit, measure the 1 mW power reference level of
the DUT.
4 Remove the OPEN connector from the measurement system, terminate it with
the 50 R load from the 85032B calibration kit, and repeat the 1 mW power
reference level measurement.
5 Using all of these values, the N7800A Series software will calculate the VSWR
of the power reference output.
Performance Tests 2
– This test cannot be performed manually, due to the complexity of the
equipment calibration procedure, and the complexity of the measurement
algorithm.
– No adjustment is available for this test if it fails (see Chapter 5,
"Troubleshooting Guide").
Keysight N1913/1914A Service Guide 55
Page 56
2 Performance Tests
NOTE

Zero Set (Average Path)

Description

Zero set is defined as the amount of residual offset error that is present following a zero operation. This offset error is caused by contamination from several sources, including circuit noise. This test measures the effectiveness of zero set by performing 15 back-to-back zero operations of the average path (with no sensor attached), after which the standard deviation of the results is calculated and returned as the measured value.

Equipment

No test equipment required

Test method

1 Execute the internal zero set measurement procedure for channel A.
2 Read back the result of the measurement from the DUT.
3 If the DUT model number is N1914A, then repeat this procedure for channel B.
4 The test will take a few minutes to complete.
5 The measurement result should be less than 0.0000175. The smaller the
measurement result, the smaller the amount of residual offset error.
This test can be performed manually via the commands:
SERV:BIST:CW[1|2]:ZSET
SERV:BIST:CW[1|2]:ZSET:NUM?
(Refer to the N1913/1914A EPM Series Power Meters Programming Guide for further details on the use of these commands)
56 Keysight N1913/1914A Service Guide
Page 57

Absolute Accuracy Test (Average Path)

Description

The absolute accuracy test checks the ability of the power meter to accurately measure the power sensor voltage and display the appropriate power level.

Equipment

– Required test equipment:
– 1 unit of 3458A digital multimeter
– 1 unit of 33250A function generator
– 1 unit of 11683A (Option H01) range calibrator
– 1 unit of 11730A power sensor cable
– 2 units of 10503A BNC cable
– 1 unit of BNC T-joint connector (BNC female, male, female)
– 1 unit of BNC (female) to dual banana connector
Performance Tests 2
Keysight N1913/1914A Service Guide 57
Page 58
2 Performance Tests
11683A range
calibrator front panel
11683A range
calibrator rear panel
BNC T-joint connector
BNC cable
BNC cable
Banana
connector
Banana
connector
33250A function
generator
3458A digital
multimeter
N1913/1914A EPM
Series power meters
NOTE
58 Keysight N1913/1914A Service Guide
Figure 2-5 Absolute accuracy test setup
The procedure details the key presses required on the Keysight N1913A EPM Series power meters. For the Keysight N1914A EPM Series power meters, the equivalent key presses should be performed on both channels.

Test Method

1 Connect the equipment as shown in Figure 2-5.
2 Unplug the power cord from the range calibrator. Eliminate ground loops to
enable the 11683A (Option H01) range calibrator to operate properly.
Page 59
Performance Tests 2
NOTE
NOTE
Disconnect the power cord to stop the operation of the range calibrator as the
range calibrator will continue to operate with no power applied.
When switching the range calibrator to STANDBY, allow enough time for the range calibrator to settle to its zero value before attempting to zero the Keysight N1913/1914A EPM Series power meters. This settling would appear on the Keysight N1913/1914A EPM Series power meters display as downward drift. When the drift has reached minimum, (typically less than 60 seconds), the range calibrator is settled.
3 Turn on the voltmeter and allow it to warm up for 4 hours.
4 Turn on the DC source and allow it to warm up. Set the output voltage to 0 V.
5 Turn on the DUT and allow it to warm up for 30 minutes.
6 Ensure the range calibrator is not plugged in or powered on.
7 Connect the range calibrator to the DUT using the power meter interconnect
cable.
8 Connect the DC source to the range calibrator external voltage input and the
voltmeter using cables and a BNC T-joint connector at the range calibrator.
9 Configure the DUT as shown below.
Parameter Value
Filter/Averaging On
Filter Mode/Measurement Average Manual
Filter Length/Average Number 16
Resolution 4 digits
All other settings use the default setup settings. The *RST command sets the default setup.
10 Configure the voltmeter using the default setup settings.
Keysight N1913/1914A Service Guide 59
Page 60
2 Performance Tests
11 Perform a power meter zero on the DUT. The voltage of the range calibrator is
assumed to be 0 V.
12 Set the DC voltage to 89.6056 mV as measured by the voltmeter.
13 Perform a power meter calibration on the DUT.
14 Measure and record the absolute accuracy of the N1913/1914A in a table as
shown below.
Effective
power
-12 dBm 0.00565 V 0.1 V 256 ±0.5%
-5 dBm 0.02834 V 0.1 V 64 ±0.5%
5 dBm 0.28400 V 1 V 64 ±0.5%
8 dBm 0.56700 V 1 V 16 ±0.5%
10 dBm 0.90100 V 1 V 16 ±0.5%
12 dBm 1.43500 V 10 V 16 ±0.5%
14 dBm 2.29000 V 10 V 16 ±0.5%
16 dBm 3.66700 V 10 V 16 ±0.5%
17 dBm 4.65200 V 10 V 16 ±0.5%
18 dBm 5.91500 V 10 V 16 ±0.5%
19 dBm 7.53000 V 10 V 16 ±0.5%
20 dBm 9.62300 V 10 V 16 ±0.5%
DC Voltage
Voltmeter
range
Power meter filter /
voltmeter NRDNS
CH A % error CH B % error Specification
60 Keysight N1913/1914A Service Guide
Page 61

Linearity (Average Path)

NOTE

Description

Linearity over the full input voltage range of the measurement path is warranted to provide assurance of instrument accuracy. This test measures linearity by using a calibration DAC and a calibration ADC (built into the DUT) to stimulate and compare performance of the average path against the measurement ADC, returning the worst case percentage error.

Equipment

No test equipment required

Test method

1 Execute the internal linearity measurement procedure for channel A.
2 Read back the result of the measurement from the DUT.
3 If the DUT model number is N1914A, then repeat this procedure for channel B.
4 The test will take a few minutes to complete.
Performance Tests 2
5 The measurement result should be less than 0.5 and greater than –0.5. The
optimum measurement result for this test is 0.
This test can be performed manually via the commands:
SERV:BIST:CW[1|2]:LIN
SERV:BIST:CW[1|2]:LIN:PERR?
(Refer to the N1913/1914A EPM Series Power Meters Programming Guide for further details on the use of these commands)
Keysight N1913/1914A Service Guide 61
Page 62
2 Performance Tests
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
62 Keysight N1913/1914A Service Guide
Page 63
Keysight N1913/1914A EPM Series Power Meters Service Guide

3 Adjustments

Introduction 64
Power Reference Level Adjustment 65
This chapter contains checks and adjustments that ensure proper performance of the power meter.
63
Page 64
3Adjustments

Introduction

The adjustment attempts to correct the power reference level if the performance test has failed. Power reference level is controlled by the coarse and fine settings of a digital potentiometer. Adjustment of the coarse and fine settings can only be carried out via remote commands. Adjustment can be carried out without having to remove the outer covers from the DUT.
64 Keysight N1913/1914A Service Guide
Page 65

Power Reference Level Adjustment

NOTE

Equipment

As per the test equipment list for the power reference level performance test.

Test method

1 Set: Coarse = 834, Fine = 550
2 Measure power ref. level as per the performance test:
a If the result is > 1 mW, then increment COARSE by 1
b If the result is < 1 mW, then decrement COARSE by 1
3 Repeat step 2 until the result crosses the 1 mW boundary (in either direction)
4 Measure power ref. level as per the performance test:
a If the result is > 1 mW, then decrement FINE by 1
a If the result is < 1 mW, then increment FINE by 1
Adjustments 3
5 Repeat step 4 until the result crosses the 1 mW boundary (in either direction)
6 The adjustment is completed
– This adjustment can be performed manually via the commands:
SERV:CAL:ADJ:COUR <value>
SERV:CAL:ADJ:COUR?
SERV:CAL:ADJ:FINE <value>
SERV:CAL:ADJ:FINE?
(Refer to the Keysight N1913/1914A EPM Series Power Meters Programming
Guide for further details on the use of these commands.)
– COARSE and FINE values are valid in the range of 0 to 1023
– If adjustment is not possible, then a fault may be present in the DUT (see
Chapter 5, "Troubleshooting Guide").
Keysight N1913/1914A Service Guide 65
Page 66
3Adjustments
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
66 Keysight N1913/1914A Service Guide
Page 67
Keysight N1913/1914A EPM Series Power Meters Service Guide

4 Theory of Operation

PPMC Assembly 68
Main Board Assembly 69
Calibrator Assembly 70
Front Panel Assembly 71
PSU Assembly 72
This chapter describes how each of the power meter’s individual assemblies operate.
67
Page 68
4 Theory of Operation

PPMC Assembly

Purpose

– Provides the main processor and memory for the power meter
– Provides external interfaces for LAN and USB
– Stores the power meter firmware in flash EEPROM
– Stores the power meter serial number and option data

Inputs

– Power supplies [from PSU, via main board]
– Control and data lines [from main board and front panel]
– LAN/USB communications [from external equipment]
– GPIB communications [from external equipment, via main board]

Outputs

– Control address and datalines [to main board and front panel]
68 Keysight N1913/1914A Service Guide
Page 69

Main Board Assembly

Purpose

– Provides the average measurement path(s)
– Provides external trigger input/output and recorder output(s)
– Provides the driver and the LVDS serialiser for the LCD display
– Provides signal routing between the PPMC and front panel

Inputs

– Power supplies [from PSU]
– Sensed power level(s) [from sensor flex(s)]
– Trigger input [from external equipment]
– Control, address, and data lines [from PPMC]

Outputs

Theory of Operation 4
– Processed average path measurement [to PPMC]
– Trigger output and recorder output(s) [to external equipment]
– LVDS LCD display control lines [to front panel]
– Control and data lines [to PPMC]
Keysight N1913/1914A Service Guide 69
Page 70
4 Theory of Operation

Calibrator Assembly

Purpose

– Provides a 1 mW (0 dBm) power reference level at 50 MHz

Inputs

– Power supplies [from PSU, via main board]
– Control, address, and data lines [from PPMC]

Outputs

– 1 mW (0 dBm) power reference [to external equipment]
– Control and data lines [to PPMC, via main board]
70 Keysight N1913/1914A Service Guide
Page 71

Front Panel Assembly

NOTE

Purpose

– Provides a keypad as the manual user interface
– Provides an LCD display to assist with manual setups and measurements
– Provides mounting for the sensor and power reference connectors (Option
101)

Inputs

– Power supplies [from PSU, via main board]
– Front panel control interface [from the PPMC LVDS LCD control lines, via main
board]
– Front USB port available for Option 008 (this USB port can only be used for
U2000 Series power sensors)
Front USB port is not to be used with USB flash storage devices.
Theory of Operation 4

Outputs

– Keypress data [to PPMC, via main board]
– Information on the LCD display
– Control and data lines [to PPMC, via main board]
Keysight N1913/1914A Service Guide 71
Page 72
4 Theory of Operation
PSU
Charger + Battery
Switching regulation, distribution and filtering
Battery Option
+
1
2
V
1
2
V
+
5
V
5
V
+
3
.
3
V

PSU Assembly

Purpose

– Provides various DC power supplies

Inputs

– 100 Vac ~240 Vac, 50 Hz ~60 Hz, 150 VA Max [from an external source]
– Control lines [from front panel, via main board]

Outputs

72 Keysight N1913/1914A Service Guide
Page 73
Keysight N1913/1914A EPM Series Power Meters Service Guide

5 Troubleshooting Guide

Introduction 74
Power-Up Problems 74
Instrument Self-Test 75
Extended Self-Test 76
Performance Test 76
Power Reference Level Adjustment Problems 76
Communication Interface Failures 77
Additional Diagnostic Tests 77
This chapter contains troubleshooting procedures for the EPM Series Power Meters.
73
Page 74
5 Troubleshooting Guide

Introduction

This chapter contains general troubleshooting guide to detect failures for the EPM Series Power Meters.

Power-Up Problems

Basic external checks

– Check that the mains power source is live
– Check that the mains fuse is operational
– Check the mains cable for any obvious damage
– Check that the line module fuse in the instrument is operational

Basic internal checks

– Check/reseat the cable between the line module and the PSU
– Check/reseat the cable between the PSU and the main board
– Green LED DS1: If this is off, then the PSU may be faulty
– Green LED DS4: This should come on when the power button is pressed
– Green LEDs DS2/DS3: These will flash on and off during normal operation

Possible faults

–PSU
– Main board
– Front panel (defective keymat, key flex circuit, or display)
– Loose front panel cable (connection to main board)
74 Keysight N1913/1914A Service Guide
Page 75
Troubleshooting Guide 5

Instrument Self-Test

Instrument Purpose Debug tips Possible faults
Test point voltages Checks that all of the supply voltages
are present
Calibrator Verifies that the calibrator is working
(Note: This test does not check that the calibrator meets its specifications)
Fan Verifies that the fan is working Check/reseat the cable between the
RTC battery Checks that the lithium manganese
battery on the main board is working
CW path/ChA CW path Verifies that the average path of
channel A is working (Note: This does not prove that the
average path meets its specifications)
ChB CW path Verifies that the average path of
channel B is working (Note: This does not prove that the
average path meets its specifications)
Replace the PSU to see if this clears the faults
Check/reseat that cable between the calibrator assembly and the main board
Attempt to adjust the 1 mW power reference level
fan assembly and the main board Check visually if the fan is functioning
Replace the battery to see if this clears the fault
Not applicable Main board
Not applicable Main board
PSU (low probability) Main board (high
probability)
Calibrator assembly (high probability)
Main board (low probability)
Fan assembly (high probability)
Main board (low probability)
Lithium manganese battery (high probability)
Main board (low probability)
Keysight N1913/1914A Service Guide 75
Page 76
5 Troubleshooting Guide

Extended Self-Test

Instrument Purpose Debug tips Possible faul ts
Front panel (defective keymat or key flex circuit)
Front panel (defective display, display interface board, or inverter board)
Keypad
Bitmap display
Verifies the operation of every key (apart from the power button)
Verifies that all pixels in the display can be illuminated in various colors
Not applicable
Not applicable

Performance Test

Type of failures Debug tips Possible faults
1 mW power reference level failures Attempt to adjust the 1 mW power
reference level
VSWR failures Not applicable Calibrator assembly
Calibrator assembly (high probability) Main board (low probability)
Zero set (average Path) failures Not applicable Main board
Linearity (average path) failures Not applicable Main board

Power Reference Level Adjustment Problems

Possible faults

– Calibrator assembly (high probability)
– Main board (low probability)
76 Keysight N1913/1914A Service Guide
Page 77
Troubleshooting Guide 5

Communication Interface Failures

Type of communication Debug tips Possible faul ts
GPIB communication Check/reseat the ribbon cable
connecting the PPMC to the main board
LAN/USB communication Check visually to see whether or not the
connector is obstructed/damaged
Ribbon cable (low probability) Main board (high probability)
PPMC assembly

Additional Diagnostic Tests

Type of functionality Reason Recommended test method Possible faults
USB/LAN functionality The N7800 Series software only tests
functionality over GPIB
Sensor functionality The N7800 Series software does not
prove both paths of the sensor flex assembly
Check the DUT responds when *RST is sent to it via the USB/LAN interfaces
Connect an E4412A sensor to the DUT and ensure it can be zeroed/ calibrated
PPMC assembly
Sensor flex assembly
Keysight N1913/1914A Service Guide 77
Page 78
5 Troubleshooting Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
78 Keysight N1913/1914A Service Guide
Page 79
Keysight N1913/1914A EPM Series Power Meters Service Guide

6 Repair Guide

Introduction 80
Replaceable Parts 81
Tools Required 104
Required Torque Values for Fasteners 105
Disassembly Instructions 106
Reassembly Instructions 113
Disassembly vs Part Replacement 117
Front Panel Disassembly Instructions 119
Front Panel Reassembly Instructions 124
Additional Repair Notes 128
Main Board vs. Rear Panel Assembly 129
Replacing the PPMC Assembly 130
Replacing the Calibrator Semi-Rigid/Split Ferrite 131
This chapter details the power meter's replaceable parts. It also explains how to assemble and disassemble the power meter.
79
Page 80
6 Repair Guide

Introduction

This chapter contains details of some of the higher level components and assemblies which can be ordered from Keysight Technologies. It also details how to assemble and disassemble the power meter for repair. The contents included are:
1 Replaceable Parts
2 Tools Required
3 Disassembly Instructions
4 Reassembly Instructions
5 Disassembly vs Part Replacement
6 Front Panel Disassembly Instructions
7 Front Panel Reassembly Instructions
8 Additional Repair Notes
9 Replacing the PPMC Assembly
10 Replacing the Calibrator Semi-Rigid/Split Ferrite
You can order replaceable parts from Keysight by contacting your local Keysight Technologies Sales and Service Office.
You can return your power meter for servicing at a qualified service center. Refer to Chapter 7, "Contacting Keysight Technologies".
80 Keysight N1913/1914A Service Guide
Page 81

Replaceable Parts

Front panel assembly

Main assembly
The standard N1913/1914A power meter has the reference calibrator at the front panel. An option is available to move the reference calibrator to the rear panel.
Keysight part number Description Visual
Repair Guide 6
N1914-60200 N1914-60201
N1914-60203 Front panel assembly (rear calibrator option)
Front panel assembly (front calibrator option)
Front panel assembly - Front calibrator Front panel assembly - Front calibrator and
USB
Note: – The front panel assembly must be
customized to suit the hardware configuration of the unit being repaired
– Refurbished front panel assembly are not
available
Front panel assembly - Rear calibrator
Keysight N1913/1914A Service Guide 81
Page 82
6 Repair Guide
Customization details
The standard EPM Series Power Meters has an input sensor connector(s) and a reference calibrator connector on the front panel. Option 003 is available to move the input sensor connector(s) and reference calibrator connector to the rear panel. Below are the customization details on front panel assembly.
Connector option Details Part number
N1913A (Front connectors option)
1 unit of sensor flex assembly 1 unit of calibrator plug 1 unit of PLUG-HOLE TR-HD FOR
.688-D-HOLE NYL 1 unit of blank front panel dress label 1 unit of N1913A nameplate
N1913-67300 N1912-21003 6960-0024
N1912-00027 N1913-34300
N1913A (Rear connectors option)
N1914A (Front connectors option)
N1914A (Rear connectors option)
2 units of PLUG-HOLE TR-HD FOR .688-D-HOLE NYL
1 unit of front panel plug (small) 1 unit of blank front panel dress label 1 unit of N1913A nameplate
2 units of sensor flex assembly 1 unit of calibrator plug 1 unit of blank front panel dress label 1 unit of N1914A nameplate
2 units of PLUG-HOLE TR-HD FOR .688-D-HOLE NYL
1 unit of front panel plug (small) 1 unit of blank front panel dress label 1 unit of N1914A nameplate
6960-0024
N1912-21005 N1914-34302 N1913-34300
N1913-67300 N1912-21003 N1912-00027 N1914-34300
6960-0024
N1912-21005 N1912-00025 N1914-34300
82 Keysight N1913/1914A Service Guide
Page 83
Customization parts
E4418-00044
E4418-00045
N1912- 21005
Keysight part number Description Visual
N1913-67300 (Front panel - Long)
N1913-67301 (Rear panel - short)
N1912-21003 Calibrator plug
Sensor flex assembly
Note: – The same assembly is used for all four
sensor positions
– The sensor flex assembly is supplied
straight, and so it must be folded to match the assembly being replaced (see
“Additional Repair Notes” on page 128)
Power sensor cable assembly - Front Power sensor cable assembly - Rear
Repair Guide 6
N1912-21005 Plug - Front panel small
Keysight N1913/1914A Service Guide 83
Page 84
6 Repair Guide
N1914-34302
N1912-00027
Keysight part number Description Visual
N1914-34302 N1912-00027
Label dual channel - Rear cal Label - Dual
N1913-34300 N1914-34300
N1913A name plate N1914A name plate
84 Keysight N1913/1914A Service Guide
Page 85
Replaceable parts
Keysight part number Description Visual
N1913-40200 Front panel
Note:
This front panel sub-frame is used on all variants of the front panel assembly
N1913-36600 Display support
Note:
This display support molding is used on all variants of the front panel assembly
Repair Guide 6
N1913-38300 Keypad
Note:
This keypad is used on all variants of the front panel assembly
Keysight N1913/1914A Service Guide 85
Page 86
6 Repair Guide
Keysight part number Description Visual
N1913-66503 Key flex circuit
Note:
This key flex circuit is used on all variants of the front panel assembly
N1912-20005 Window EMI shield
Note:
This EMI Shielded Window is used on all variants of the Front Panel Assembly
N1913-00600 EMI screen
Note:
This EMI screen is used on all variants of the front panel assembly
86 Keysight N1913/1914A Service Guide
Page 87
Keysight part number Description Visual
2090-0825 Sharp display
Note: This display is used on all variants of the front
panel assembly
Front view
Rear view
Repair Guide 6
N1912-60002 Display interface board
Note: This display interface board is used on all
variants of the front panel assembly
Keysight N1913/1914A Service Guide 87
Page 88
6 Repair Guide
Keysight part number Description Visual
0950-4111 Inverters 5-Watt 1-output
Note: This inverter interface board is used on all
variants of the front panel assembly
N1912-61002 Cable assembly backlight
Note: This backlight cable is used on all variants of
the front panel assembly
N1912-00038 EMC split washer
88 Keysight N1913/1914A Service Guide
Page 89
Keysight part number Description Visual
N1913-68303 USB assembly - Front
Note: This USB assembly is only available for
Option 105, 106, and 167
Repair Guide 6
N1913-68301 (Single) N1913-68302 (Dual)
8121-0936 Flat-Ribbon-Assy- 28-AWG 30-COND
USB assembly - Single USB assembly - Dual
03-IN-LG
Note: This flat ribbon cable is used on all variants of
the front panel assembly
Keysight N1913/1914A Service Guide 89
Page 90
6 Repair Guide
Main board assembly
Keysight part number Description Visual
N1913-66502 N1913A mother board [New]
N1914-66502 N1914A mother board [New]
Note: – Refurbished main boards are not available – The part number for the Lithium
Manganese battery (upper- right of both photographs) is 1420-0394
90 Keysight N1913/1914A Service Guide
Page 91
PPMC (Processor PCI Mezzanine) assembly
Keysight part number Description Visual
– N1913-66501 (Tested single channel
processor peripheral component interconnect mezzanine card PCA)
– N1914-66501 (Tested dual channel
processor peripheral component interconnect mezzanine card PCA)
Note: – The same assembly is used for both
N1913A and N1914A models
– N1913-66501 (Tested single channel
processor peripheral component interconnect mezzanine card PCA) comes pre-programmed with N1913A firmware
– N1914-66501 (Tested dual channel
processor peripheral component interconnect mezzanine card PCA) comes pre-programmed with N1914A firmware
– Refurbished PPMC Assemblies are not
available
– The PPMC assembly must be programmed
once it has been installed (see“Additional
Repair Notes” on page 128)
– Ribbon Cable 8121-1076 is supplied
separately
Repair Guide 6
Keysight N1913/1914A Service Guide 91
Page 92
6 Repair Guide
Keysight part number Description Visual
PSU [New]
N1913-00100 (Non-battery option)
N1913-00105 (Battery option)
0950-5015
N1913-34102
N1913-37900 Battery pack
Deck assembly
Deck assembly - Battery option
Power supply AC-DC Adapter 120-WATT 1-OUTPUT
Power supply cover
N1913-61301 N1913-61604
Ribbon cable assembly - PSU Cable assembly - PSU
92 Keysight N1913/1914A Service Guide
Page 93

Rear panel assembly

Main assembly
Keysight part number Description Visual
N1913-00200 Rear panel assembly [New]
Revised rear panel assembly [New]
Note: – The rear panel assembly must be
customized to suit the hardware configuration of the unit being repaired
– Refurbished rear panel assemblies are not
available
– The same assemblies are used for both
N1913A and N1914A models
Repair Guide 6
Keysight N1913/1914A Service Guide 93
Page 94
6 Repair Guide
Customization details
The standard EPM Series Power Meters has an input sensor connector(s) and a reference calibrator connector on the front panel. Option 003 is available to move the input sensor connector(s) and reference calibrator connector to the rear panel. Below are the customization details on the rear panel assembly.
Connector option Details Part number
N1913A (Front connectors option)
N1913A (Rear connectors option)
N1914A (Front connectors option)
N1914A (Rear connectors option)
1 unit of rear panel plug (BNC) 2 units of rear panel plug (Sensor) 1 unit of rear panel plug (Calibrator)
1 unit of sensor flex assembly 1 unit of N-type connector 1 unit of lock washer 1 unit of Hex nut 1 unit of washer 1 unit of rear panel plug (BNC) 1 unit of rear panel plug (Sensor)
2 units of recorder output cable 2 units of rear panel plug (Sensor) 1 unit of rear panel plug (Calibrator)
2 units of sensor flex assembly 1 unit of N-type connector 1 unit of lock washer 1 unit of Hex nut
6960-0081 6960-0024 6960-0178
N1913-67300 E4418-20009 E4418-00016 2950-0132 3050-0916 6960-0081 6960-0024
E4418-61015 6960-0024 6960-0178
N1912-61806 E4418-20009 E4418-00016 2950-0132
94 Keysight N1913/1914A Service Guide
Page 95
Customization parts
050-0916
2950-0132
E4418-00016
E4418-20009
Keysight part number Description Visual
E4418-20009 E4418-00016 2950-0132 3050-0916
N- Type connector Lock washer Hex nut Washer
Repair Guide 6
N1913-67300 N1913-67301
E4418-61015 Recorder output cable
Sensor flex assembly
Keysight N1913/1914A Service Guide 95
Page 96
6 Repair Guide
Keysight part number Description Visual
N1913-67912 BNC option
6960-0081 6960-0024 6960-0178 N1913-36200 N1913-36201 N1913-36202
BNC plug (rear panel) Sensor plug (front and rear panels) Calibrator plug (rear panel) USB plug (front panel) USB plug (rear panel) VGA plug (rear panel)
96 Keysight N1913/1914A Service Guide
Page 97

Additional spare parts

Keysight part number Description Visual
N1913-62700 Line module assembly
N1913-67910 VGA option
Repair Guide 6
Keysight N1913/1914A Service Guide 97
Page 98
6 Repair Guide
Keysight part number Description Visual
N1912-00016 Spring contact - Perpendicular
N1913-61601 N1913-61600 N1913-60002
Wire hardness assembly - 10 to10 Pins Ribbon cable assembly - 6 to 4 Pins PCA - Battery charger
98 Keysight N1913/1914A Service Guide
Page 99

Calibrator assembly

Keysight part number Description Visual
N1913-62000 Calibrator assembly - Front
(Front connectors option) For more information on the improved
calibrator assembly, refer to “Improved
calibrator assembly” on page 100.
N1911-61002 Calibrator assembly (Rear connectors option)
Note: Semi-rigid cable N1912-61004 is not
included with assembly N1911-61002; if this is required, it is available as a separate item
Repair Guide 6
Keysight N1913/1914A Service Guide 99
Page 100
6 Repair Guide
NOTE

Improved calibrator assembly

Calibrator assembly
Improved calibrator assembly
The N1913/1914A calibrator assembly is improved with a new type-N calibrator connector for a better fit with the front panel assembly as shown in Figure 6-1. The calibrator plug (N1912-21003) will not be needed with the improved calibrator assembly.
The N1913/1914A performance will not be affected with the improved calibrator assembly.
100 Keysight N1913/1914A Service Guide
Figure 6-1 Calibrator assembly and improved calibrator assembly
Loading...