No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight
Technologies, Inc. as governed by
United States and international
copyright laws.
Manual
M9392-90001
Published By
Keysight Technologies
Ground Floor and Second Floor, CP-11
Sector-8, IMT Manesar – 122051
Gurgaon, Haryana, India
Edition
2013
In
Printed
Regulatory Compliance
This product has been designed and
tested in accordance with accepted
industry standards, and has been
supplied in a safe condition. To review
the Declaration of Conformity, go to
http://www.keysight.com/go/conformity.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,” AND
IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE
EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR
CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
Part
USA
Number
FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED
HEREIN. SHOULD KEYSIGHT AND THE
USER HAVE A SEPARATE WRITTEN
AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
KEYSIGHT TECHNOLOGIES DOES NOT
WARRANT THIRD-PARTY SYSTEMLEVEL (COMBINATION OF CHASSIS,
CONTROLLERS, MODULES, ETC.)
PERFORMANCE, SAFETY, OR
REGULATORY COMPLIANCE, UNLESS
SPECIFICALLY STATED.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance with
the terms of such license.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal
Acquisition Regulation (“FAR”) 2.101.
Pursuant to FAR 12.212 and 27.405-3
and Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires commercial
computer software under the same
terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the
Software to U.S. government customers
under its standard commercial license,
which is embodied in its End User
License Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula. The
license set forth in the EULA represents
the exclusive authority by which the
U.S. government may use, modify,
distribute, or disclose the Software. The
EULA and the license set forth therein,
does not require or permit, among other
things, that Keysight: (1) Furnish
technical information related to
commercial computer software or
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the
government rights in excess of these
rights customarily provided to the
public to use, modify, reproduce,
release, perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall apply,
except to the extent that those terms,
rights, or licenses are explicitly required
from all providers of commercial
computer software pursuant to the FAR
and the DFARS and are set forth
specifically in writing elsewhere in the
EULA. Keysight shall be under no
obligation to update, revise or otherwise
modify the Software. With respect to
any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the U.S.
government acquires no greater than
Limited Rights as defined in FAR 27.401
or DFAR 227.7103-5 (c), as applicable in
any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
The following safety precautions should
be observed before using this product
and any associated instrumentation.
This product is intended for use by
qualified personnel who recognize
shock hazards and are familiar with the
Page 4
safety precautions required to avoid
possible injury. Read and follow all
installation, operation, and
maintenance information carefully
before using the product.
If this product is not used as specified,
the protection provided by the
equipment could be impaired. This
product must be used in a normal
condition (in which all means for
protection are intact) only.
The types of product users are:
Responsible body is the individual or
group responsible for the use and maintenanceof equipment, for ensuring that
the equipment is operated within its specifications and operating limits, and for
ensuring operators are adequately trained.
Operators use the productfor its intended
function. They must be trained in electrical
safety procedures and proper use of the
instrument. They must be protected from
electric shock and contact with hazardous
live circuits.
Maintenance personnel perform routine
procedures on the product to keep itoperating properly (for example, setting the line
voltage or replacing consumable materials). Maintenance procedures are
described in the user documentation. The
procedures explicitly stateif theoperator
may perform them. Otherwise, they should
be performed only by service personnel.
Servicepersonnel are trained to work on
live circuits, perform safe installations, and
repair products. Only properly trained servicepersonnel may perform installation
and service procedures.
Operator is responsible to maintain safe
operating conditions. To ensure safe
operating conditions, modules should
not be operated beyond the full
temperature range specified in the
Environmental and physical
specification. Exceeding safe operating
conditions can result in shorter
lifespans, improper module
performance and user safety issues.
When the modules are in use and
operation within the specified full
temperature range is not maintained,
module surface temperatures may
exceed safe handling conditions which
can cause discomfort or burns if
touched. In the event of a module
exceeding the full temperature range,
always allow the module to cool before
touching or removing modules from
chassis.
Keysight products are designed for use
with electrical signals that are rated
Measurement Category I and
Measurement Category II, as described
in the International Electrotechnical
Commission (IEC) Standard IEC 60664.
Most measurement, control, and data
I/O signals are Measurement Category I
and must not be directly connected to
mains voltage or to voltage sources with
high transient over-voltages.
Measurement Category II connections
require protection for high transient
over-voltages often associated with
local AC mains connections. Assume all
measurement, control, and data I/O
connections are for connection to
Category I sources unless otherwise
marked or described in the user
documentation.
Exercise extreme caution when a shock
hazard is present. Lethal voltage may
be present on cable connector jacks or
test fixtures. The American National
Standards Institute (ANSI) states that a
shock hazard exists when voltage levels
greater than 30V RMS, 42.4V peak, or
60VDC are present. A good safety
practice is to expect that hazardous
voltage is present in any unknown
circuit before measuring.
Operators of this product must be
protected from electric shock at all
times. The responsible body must
ensure that operators are prevented
access and/or insulated from every
connection point. In some cases,
connections must be exposed to
potential human contact. Product
operators in these circumstances must
be trained to protect themselves from
the risk of electric shock. If the circuit is
capable of operating at or above 1000V,
no conductive part of the circuit may be
exposed.
Do not connect switching cards directly
to unlimited power circuits. They are
intended to be used with impedancelimited sources. NEVER connect
switching cards directly to AC mains.
When connecting sources to switching
cards, install protective devices to limit
fault current and voltage to the card.
Before operating an instrument, ensure
that the line cord is connected to a
properly-grounded power receptacle.
Inspect the connecting cables, test
leads, and jumpers for possible wear,
cracks, or breaks before each use.
When installing equipment where
access to the main power cord is
restricted, such as rack mounting, a
separate main input power disconnect
device must be provided in close
proximity to the equipment and within
easy reach of the operator.
For maximum safety, do not touch the
product, test cables, or any other
instruments while power is applied to
the circuit under test. ALWAYS remove
power from the entire test system and
discharge any capacitors before:
connecting or disconnecting cables or
jumpers, installing or removing
switching cards, or making internal
changes, such as installing or removing
jumpers.
Do not touch any object that could
provide a current path to the common
side of the circuit under test or power
line (earth) ground. Always make
measurements with dry hands while
standing on a dry, insulated surface
capable of withstanding the voltage
being measured.
The instrument and accessories must
be used in accordance with its
specifications and operating
instructions, or the safety of the
equipment may be impaired.
Do not exceed the maximum signal
levels of the instruments and
accessories, as defined in the
specifications and operating
information, and as shown on the
instrument or test fixture panels, or
switching card.
When fuses are used in a product,
replace with the same type and rating
Page 5
for continued protection against fire
hazard.
Chassis connections must only be used
as shield connections for measuring
circuits, NOT as safety earth ground
connections.
If you are using a test fixture, keep the
lid closed while power is applied to the
device under test. Safe operation
requires the use of a lid interlock.
Instrumentation and accessories shall
not be connected to humans.
Before performing any maintenance,
disconnect the line cord and all test
cables.
To maintain protection from electric
shock and fire, replacement
components in mains circuits –
including the power transformer, test
leads, and input jacks – must be
purchased from Keysight. Standard
fuses with applicable national safety
approvals may be used if the rating and
type are the same. Other components
that are not safety-related may be
purchased from other suppliers as long
as they are equivalent to the original
component (note that selected parts
should be purchased only through
Keysight to maintain accuracy and
functionality of the product). If you are
unsure about the applicability of a
replacement component, call an
Keysight office for information.
No operator serviceable parts inside.
Refer servicing to qualified personnel.
To prevent electrical shock do not
remove covers. For continued
protection against fire hazard, replace
fuse with same type and rating.
PRODUCT MARKINGS:
The CE mark is a registered trademark
of the European Community.
Australian Communication and Media
Authority mark to indicate regulatory
compliance as a registered supplier.
This symbol indicates product
compliance with the Canadian
Interference-Causing Equipment
Standard (ICES-001). It also identifies
the product is an Industrial Scientific
and Medical Group 1 Class A product
(CISPR 11, Clause 4).
South Korean Class A EMC Declaration.
This equipment is Class A suitable for
professional use and is for use in
electromagnetic environments outside
of the home. A 급 기 기 ( 업무 용 방 송 통
신 기 자 재 ) 이 기 기 는 업 무 용 (A 급 ) 전 자
파 적 합 기 기 로 서 판 매 자 또 는 사 용 자 는
이 점 을 주 의하 시 기 바 라 며 , 가 정 외 의
지 역 에 서 사 용 하 는 것 을 목 적 으 로 합 니
다 .
This product complies with the WEEE
Directive marketing requirement. The
affixed product label (above) indicates
that you must not discard this
electrical/electronic product in
domestic household waste. Product
Category: With reference to the
equipment types in the WEEE directive
Annex 1, this product is classified as
“Monitoring and Control
instrumentation” product. Do not
dispose in domestic household waste.
To return unwanted products, contact
your local Keysight office, or for more
information see
This symbol indicates the instrument is
sensitive to electrostatic discharge
(ESD). ESD can damage the highly
sensitive components in your
instrument. ESD damage is most likely
to occur as the module is being
installed or when cables are connected
or disconnected. Protect the circuits
from ESD damage by wearing a
grounding strap that provides a high
resistance path to ground. Alternatively,
ground yourself to discharge any builtup static charge by touching the outer
shell of any grounded instrument
chassis before touching the port
connectors.
This symbol on an instrument means
caution, risk of danger. You should refer
to the operating instructions located in
the user documentation in all cases
where the symbol is marked on the
instrument.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life of
the product.
CLEANING PRECAUTIONS:
To prevent electrical shock, disconnect
the Keysight Technologies instrument
from mains before cleaning. Use a dry
cloth or one slightly dampened with
water to clean the external case parts.
Do not attempt to clean internally. To
clean the connectors, use alcohol in a
well-ventilated area. Allow all residual
alcohol moisture to evaporate, and the
fumes to dissipate prior to energizing
the instrument.
Page 6
Table of Contents
Documentation Map7
M9392A Introduction8
Related Documentation8
Follow the Startup Sequence8
Step 1: Unpack and Inspect the Modules9
ESD9
Inspect for Damage9
Return a Module for Service9
Step 2: Verify M9392A Shipment Contents10
Items included in your M9392A Shipment10
Step 3: Install the Software11
System Requirements11
Hardware Requirements11
Power up the Controller12
Install the Software12
Software Installation Procedure12
Step 4: Install the Modules13
Agilent M9392A Modular Instrument Connections15
Agilent M9392A PXI Vector Signal Analyzer Cabling Diagram15
Agilent M9392A Block Diagram17
M9392A Block Diagram Reference Table for M9302A17
M9392A Block Diagram Reference Table for M9360A18
M9392A Block Diagram Reference Table for M9361A18
M9392A Block Diagram Reference Table for M9351A19
M9392A Block Diagram Reference Table for M9202A19
Step 5: Verify Operation of the Agilent M9392A PXI Vector Signal Generator20
Frequency Range Troubleshooting34
Streaming Data Overview35
Multichannel Overview40
Agilent 89600 Vector Signal Analysis Software45
Page 7
Page 8
Documentation Map
Documentation Map
Agilent M9392A PXI Vector Signal Analyzer Startup Guide7
Page 9
M9392A Introduction
M9392A Introduction
The scope of this Startup Guide is to detail the processes of receiving and installing the modules and cables that comprise
the Agilent M9392A PXI Vector Signal Analyzer. Additionally, installing the required software, and performing a Field
Calibration are documented. Troubleshooting is included to assist you in determining and specifying any failures. If you
have any questions after reviewing this information, please contact your local Agilent Technologies Inc. representative or
contact us through our website at www.agilent.com/find/M9392A.
Related Documentation
This Startup Guide and the documentation listed below are on the CD and at
www.agilent.com/find/M9392A (go to Document Library > Manuals).
l
M9392A Soft Front Panel help system
l
M9392A device driver API help systems (IVI-C/IVI-COM and LabVIEW
l
M9392A specifications (Data Sheet)
l M9392A Streaming DataViewer help system
Follow the Startup Sequence
Closely follow the startup process flow in this document. Deviating from the sequence can
cause unpredictable system behavior, damage your system, and may cause personal injury.
Step 1: Unpack
and Inspect
Step 2: Verify Shipment
Step 3: Install Drivers and Software
Step 4: Install Modules
Step 5: Verify Operation with the Soft Front Panel (SFP)
Step 6: INSTALLATION PROCESS COMPLETE.
Proceed to program your product through the API.
8Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 10
Step 1: Unpack and Inspect the Modules
The modules are shipped in materials which prevent damage from static. The modules should only
be removed from the packaging in an anti-static area ensuring that correct anti-static precautions are
taken. Store all modules in anti-static envelopes when not in use.
ESD
Electrostatic discharge (ESD) can damage or destroy electronic components. Use a static-safe work station to perform all work on electronic assemblies. The figure (left)
shows a static-safe work station using two types of ESD protection: conductive table-mat and wrist-strap combination,
and conductive floor-mat and heel-strap combination. Both
types, when used together, provide a significant level of
ESD protection. Of the two, only the table-mat and wriststrap combination provides adequate ESD protection when
used alone. To ensure user safety, the static-safe accessories must provide at least 1MΩ of isolation from ground.
Step 1: Unpack and Inspect the Modules
DONOT use these techniques for a staticsafe work station when working on circuitry with a voltage potential greater than 500 volts.
Inspect for Damage
After unpacking a module, inspect it for any shipping damage. Report any damage to the shipping agent immediately, as such damage is not covered by the warranty (see warranty information at beginning of this document).
To avoid damage when handling a module, do not touch exposed connector pins.
See www.agilent.com/find/tips for information on preventing damage to your Agilent equipment.
Return a Module for Service
Should it become necessary to return for repair or service, follow the steps below:
1. Review the warranty information shipped with your product.
2. Contact Agilent to obtain a Return Material Authorization (RMA) and return address. For assistance finding
Agilent contact information, go to www.agilent.com/find/assist (worldwide contact information for repair and
service) or refer to the “Support” information on the product web page at www.agilent.com/find/M9392A.
3. Write the following information on a tag and attach it to the malfunctioning equipment:
l Name and address of owner. A P.O. box is not acceptable as a return address.
l Module serial number(s). The serial number label is located on the side panel of the module. The
serial number can also be read from the Soft Front Panel interface – after the hardware is installed.
l Description of failure or service required.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide9
Page 11
Step 2: Verify M9392A Shipment Contents
4. Pack the module in its original ESD bag and packing carton. If the original carton is not available, use bubble
wrap or packing peanuts and place the instrument in a sealed container and mark the container “FRAGILE”.
5. On the shipping label, write ATTENTION REPAIR DEPARTMENT and the RMA number.
If any correspondence is required, refer to the product by serial number and model number.
Step 2: Verify M9392A Shipment Contents
The Agilent M9392A PXI Vector Signal Analyzer is a collection of modules, housed in a PXIe chassis. The minimum Vector Signal Analyzer consists of the software, chassis, an Agilent M9202A PXIe IF Digitizer, an Agilent
M9302A PXI Local Oscillator and either an Agilent M9351A PXI Downconverter or an Agilent M9361A PXI
Downconverter. Your configuration may include both downconverters plus an Agilent M9360A PXI Attenuator/Preselector. Thus, the shipment contents are dependent upon your choice of modules.
Items included in your M9392A Shipment
Ships with
ModuleQty
M9202A19320-6741China RoHS Addendum for Digitizer Products
M9202A1 *M9202-90001Agilent M9202A PXIe IF Digitizer Startup Guide in hard copy
M9302A1 *M9302-90001Agilent M9302A PXI Local Oscillator Startup Guide in hard copy
M9351A1 *M9351-90001Agilent M9351A PXI Downconverter Startup Guide in hard copy
M9360A1 *M9360-90001Agilent M9360A PXI Attenuator/Preselector Startup Guide in hard copy
M9361A1 *M9361-90001Agilent M9361A PXI Downconverter Startup Guide in hard copy
M9360A1 *M9360-20001Cable, semi-rigid, Master LO
M9351A1 *M9351-20001Cable, semi-rigid (SMA-SMA)
M9361A1 *M9361-20001Cable, semi-rigid (SMA-SMA)
M9361A1 *M9361-20002Cable, semi-rigid (SMA-SMA)
M9351A1 *M9351-20002Cable, semi-rigid (SMA-SMA)
M9351A1 *8120-5016Cable, coaxial, SMB-SMB (190 mm)
M9351A1 *8121-2072Cable, coaxial, SMB-SMB (75 mm)
M9202A1 *8120-5531Cable, SMB - SMA coaxial (190 mm)
M9202A1 *8121-2042Cable, A06/A32 80G (SMB-MMCX)
M9202A1 *8121-2041Cable, A06/A32 190G (SMB-MMCX)
M9202A1 *8121-2041Cable, A06/A32 190G (SMB-MMCX)
* Each module shipment includes a hard copy of its Startup Guide and appropriate cables.
Agilent Part
NumberDescription
1M9392-10002
1E2094-60003Agilent IO Libraries Suite CD
1M9392-90001Agilent M9392A PXI Vector Signal Analyzer Startup Guide in hard copy
19320-6698China RoHS Addendum for Vector Signal Analyzer
15023-1450Wrench, socket extension, 5/16 inch, SMA
15002-3361SMB/MMCX Cable Removal Tool
Agilent M9392A PXI Vector Signal Analyzer Software and Product Information CD,
contains: Soft Front Panels, drivers, and all printed documentation in PDF format
In addition to the items described in the preceding table, each module ships with a Certificate of
Calibration and a recalibration letter.
All files contained on the CD are available from at www.agilent.com/find/M9392A.
10Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 12
Step 3: Install the Software
Step 3: Install the Software
System Requirements
TopicWindows®7 and Vista RequirementsWindows®XP Requirements
Operating system
Processor speed
Available memory
Available disk
3
space
Video
BrowserMicrosoft®Internet Explorer 7.0 or greater
1
Note: Not supported when using streaming options (M9202A-V05 or M9202A-V10).
2
Note: 4 GB minimum required for streaming capability (M9202A-V05 or M9202A-V10), which requires Windows 7.
3
Note: Because of the installation procedure, less disk space may be required for operation than is required for installation.
Windows 7 (32 bit and 64 bit);
Windows®Vista, SP1 and SP2 (32-bit and 64-bit)
1
1 GHz 32-bit (x86), 1 GHz 64-bit (x64), no support for Itanium64
1 GB minimum
2
1.5 GB available hard disk space, includes:
l 1 GB available for Microsoft .NET Framework 3.5 SP1
l 100 MB for Agilent IO Libraries Suite
Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported)
Window®XP, Service Pack 3
600 MHz or higher required
800 MHz recommended
256 MB minimum (1 GB or greater recommended)
4
Super VGA (800x600) 256 colors or
more
Microsoft®Internet Explorer 6.0 or
greater
The amount of space listed above is required for installation.
4
Note: .NET Framework Runtime Components are installed by default with Windows 7 and Vista. Therefore, you may not
need this amount of available disk space.Hardware Requirements
1
Hardware Requirements
TopicRequirements
ChassiscPCI(J1), PXI-1, or PXIe hybrid chassis slot
Host controllerA PXI or PXI Express embedded controller or remote controller (external PC connected to the chassis
by a PCI-to-PXI interface) is required.
Embedded
controller
Remote controller(for Agilent M9018A chassis use only) Agilent M9021 Cable Interface x8 with one of the following PC
Agilent M9036A or an embedded controller that meets the following requirements:
l PXIe system controller (PXI-1 embedded controllers are not compatible)
l Utilize a 2x8, 4x4, or 1x4 PXIe system slot link configuration.
l Run one of the operating systems listed in System Requirements (above).
interface options:
l Agilent M9045B PCIe ExpressCard Adaptor x1, with cable (for a laptop PC)
l Agilent M9048A PCIe Desktop Adaptor x8, with cable (for desktop PCs)
Or an equivalent remote controller using a PC running one of the operating systems listed in System
Requirements (above).
Agilent M9392A PXI Vector Signal Analyzer Startup Guide11
Page 13
Step 3: Install the Software
Power up the Controller
If you are using a remote controller and you have installed the interface cable, you must power up
the chassis BEFORE you power up the PC. When you power down your chassis, Shut Down the PC
BEFORE you power down the chassis.
If you are using an embedded controller, complete the following steps:
1. Install the embedded controller module into the compatible chassis. The Agilent M9036A PXIe Embedded
Controller and Agilent M9018A PXIe Chassis are recommended. Please refer to the embedded controller
and chassis documentation for further details.
The M9392A software is located on the Agilent M9392A PXI Vector Signal Analyzer Software and Product Information CD (M9392-10002) and the Agilent IO Libraries Suite CD (E2904-60003). The same software is also available for downloading at the Agilent website at www.agilent.com/find/M9392A.
The Agilent 89600 Vector Signal Analyzer Software is available for purchase from www.agilent.com/find/vsa.
This software allows you to measure the RF and modulation quality of digitally modulated signals.
This installation includes the following:
l Agilent IO Libraries Suite (IOLS), which includes the Agilent Connections Expert. This software must be
installed first.
l If you wish to install the Agilent 89600 Vector Signal Analyzer Software, do so AFTER installing the IOLS
and BEFORE installing the drivers located on the Agilent M9392A PXI Vector Signal Analyzer Software and
Product Information CD. See "Agilent 89600 Vector Signal Analysis Software" on page 45.
l Soft Front Panel (SFP), device drivers (IVI-C, IVI-COM, and LabVIEW G) and documentation for all mod-
ules and for the M9392A Vector Signal Analyzer.
Each module has unique device drivers, soft front panel software and documentation. Performing
the software installation offers the opportunity to install software components for one or more, of
the available modules: Agilent M9202A, M9302A, M9351A, M9360A and M9361A. Notice in
the figure that follows Step 5, the software components for the Agilent M9202A PXIe IF Digitizer are called "MD1 Digitizer Installation for M9202".
Version 16.3.16603.3 (or newer) of the Agilent IO Libraries Suite is required for Agilent
M9392A PXI Vector Signal Analyzer operation.
Software Installation Procedure
1. From the Agilent IOLS CD (E2904-60003) browser, launch the installer.
2. Follow the installer prompts to install the IO libraries.
3. If you choose to install the Agilent 89600 Vector Signal Analyzer Software, follow the prompts to complete
the installation.
4. From the Agilent M9392A PXI Vector Signal Analyzer Software and Product Information CD (M9392-
10002), launch the installer.
12Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 14
Step 4: Install the Modules
5. Follow the installer prompts. Choose a "Complete" installation to install all software and documentation, or
a "Custom" installation to select from a listing of modules and other features.
6. After installation is complete, power down the host PC, and then the chassis if you are using a remote controller.
If you are using a remote controller, Shut Down the PC BEFORE you power down the chas-
sis. When you restore power, power up the chassis BEFORE you power up the PC.
Step 4: Install the Modules
PXI hardware does not support "hot-swap" (changing modules while power is applied to the chassis)
capabilities. Before installing or removing a module to/from the chassis, power off the chassis to prevent damage to the module.
The Agilent M9302A PXI Local Oscillator, Agilent M9351A PXI Downconverter, Agilent M9360A
PXI Attenuator/Preselector, and modules are compatible with cPCI(J1), PXI-1, or PXIe hybrid slots.
The Agilent M9202A PXIe IF Digitizer module is compatible with a PXIe or PXIe hybrid slot.
1. Make sure that the line cord is plugged in to establish earth ground and that the chassis power switch is in
the Off (Standby) position.
2. If the chassis has multiple fan speed settings, ensure that the fans are set to automatic. Do not set the fan
speed to low or turn it off.
3. Position the chassis so that there is ample space between the chassis fan intake and exhaust vents. Blockage by walls or obstructions affects the air flow needed for cooling. (Refer to the chassis documentation for
more information about cooling).
4. Before installing a module, be sure to remove the plastic thread protectors from the mounting screws.
Before inserting the module into the chassis, back the mounting screws out to assure that there is no interference between the screws and the mounting rails.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide13
Page 15
Step 4: Install the Modules
5. Holding the module by the injector/ejector handle, slide it into a compatible slot, as shown in the figure
below.
a. Install the module into the PXI slot of the chassis by placing the module card edges into the front
module guides (top and bottom).
b. Assure that the injector/ejector handle is pushed down in the unlatched (downward) position. Slide
the module to the rear of the chassis
c. Slide the module completely into the chassis. When you begin to feel resistance, pull up on the injec-
tor/ejector handle to fully insert and latch the module into the chassis.
6. Secure the front panel to the chassis using the two module front-panel mounting screws. Performance may
suffer if the screws are not tightened properly.
7. Verify that the PXI chassis fans are operable and free of dust and other contaminants that may restrict airflow.
8. Install all filler panels after installing the modules. Missing filler panels may disrupt necessary air circulation in the chassis.
9. If you are using a PC host, such as the Agilent M9021, connect the Cable Interface in the chassis to the
PChost per the instructions that came with the Cable Interface.
10. Power up the PXI chassis.
11. Reboot the PC host.
12. Run the module soft front panel (SFP) software, this initializes the module.
13. Check the status LEDs on the module front panels.
14Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 16
Step 4: Install the Modules
Agilent M9392A Modular Instrument Connections
This section contains a cabling diagram for the Agilent M9392A PXI Vector Signal Analyzer, a cable and module
association table, and a block diagram for the M9392A.
Agilent M9392A PXI Vector Signal Analyzer Cabling Diagram
Recommendation: Install the
semi-rigid ("A") cables first.
Before removing
cables from SMB and MMCX
connectors, See "SMB and
MMCX Cable and Connector
Care" on page 16.
NOTE: For best measurement
performance, connect a highquality 10 MHz reference to the
M9302A 10 MHz REF IN input. If
no other reference is available,
connect the 10 MHz reference
supplied on the back of the PXI
chassis.
Cable and Module Table
Ships
with
Module
M9360AA1M9360-20001M9302A LO OUT – M9360A LO INCable, semi-rigid, Master LO (SMA-SMA)
M9351AA2M9351-20001M9360A RF 1 OUT – M9351A RF INCable, semi-rigid (SMA-SMA)
M9361AA3M9361-20001M9360A RF 2 OUT – M9361A RF INCable, semi-rigid (SMA-SMA)
M9361AA4M9361-20002M9360A LO 2 OUT – M9361A LO INCable, semi-rigid (SMA-SMA)
M9351AA5M9351-20002M9360A LO 1 OUT – M9351A LO INCable, semi-rigid (SMA-SMA)
M9351AB18120-5016M9351A REF IN – M9302A REF 1 OUTCable, coaxial, SMB-SMB (160 mm)
M9351AB28121-2072M9351A IF OUT – M9361A AUX INCable, coaxial, SMB-SMB (75 mm)
M9202AC38120-5531M9361A IF OUT – M9202A INPUT 1Cable, SMB - SMA coaxial (190 mm)
M9202AC48121-2042M9302A REF 2 OUT – M9202A REF INCable, A06/A32 80G (SMB-MMCX)
M9202A*8121-2041n/aCable, A06/A32 190G (SMB-MMCX)
M9202A*8121-2041n/aCable, A06/A32 190G (SMB-MMCX)
Torque specification for all SMA connectors is 8 Lb-In (0.904 Nm).
Ref
Agilent Part
Des
NumberConnectionCable Description
* These cables are not required to configure an Agilent M9392A PXI Vector Signal Analyzer. See the Agilent M9392A PXI
Vector Signal Analyzer Cabling Diagram above. These cables are provided for using the Agilent M9202A PXIe IF Digitizer
in other applications.
No cables are shipped with the Agilent M9302A PXI Local Oscillator module.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide15
Page 17
Step 4: Install the Modules
SMB and MMCX Cable and Connector Care
Use the Agilent SMB/MMCX Cable Removal Tool to disconnect SMB and MMCX cables from the module front
panel connectors.
To avoid damage to the cables or connectors, pull the cable straight away from the connector. Do
not use the tool as a pry bar.
16Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 18
Agilent M9392A Block Diagram
Step 4: Install the Modules
M9392A Block Diagram Reference Table for M9302A
M9302A PXI Local Oscillator Operation
InputConnectorOutput
Frequency: 10 MHz
Input level: 0 ± 4 dB
Agilent M9392A PXI Vector Signal Analyzer Startup Guide17
M9351A RF Downconverter operation
Normal IF Path mode
l Frequency: 50 MHz to 2.9 GHz
l Bypassed IF Path mode frequency:
50 MHz to 625 MHz
NOTE: The path selection (downconverter or bypass) depends on the
input frequency and selected BW.
l Downconverter path: Input frequen-
cies ≤ 70 MHz and >605 MHz go
through the downconverter path.
Input frequencies >70 MHz and
≤605 MHz and BW set to ≤ 40 MHz
go through the downconverter
path.
l Bypass path: Input frequencies >70
MHz and ≤605 MHz and BW set to
> 40 MHz go through the bypass
path.
2.75 GHz to 10 GHz
(LO frequency is set to 3.5 GHz above
RF IN for downconverter operation)
Input Power Level: +15 dBm ± 2 dB for
optimal performance.
100 MHz Input Level: 0 dBm ± 3 dBRF IN
RF IN
RF IN
IF OUT
LO IN
Downconverter path:
l Downconverter path is selected when frequency is 50 MHz
l Downconverter path is selected when BW < 40 MHz and
l Normal (Downconverter) path output is-2 dBm (500 mV p-
l The BW is dependent on the RF IN Frequency:
Bypass path only:
l Bypass path is selected when BW > 40 MHz and input
l The bypass path has a maximum BW of 575 MHz and is
l The BW is dependent on the RF IN Frequency:
Step 4: Install the Modules
to ≤ 70 MHz and > 605 MHz to <2.75 GHz
o
IF OUT = 500 MHz @ BW range = 1 kHz to 40 MHz
frequency is >70 MHz to ≤ 605 MHz
o
IF OUT = 500 MHz @ BW range = 1 kHz to 40 MHz
p)
o
The RF IN to IF OUT gain is ~38 dB
o
Lower-side BW = RF IN - 50 MHz
o
Upper-side BW = 625 MHz - RF IN
frequency is >70 MHz to ≤ 605 MHz
o
IF OUT = RF IN
available at an RF IN frequency of 337.5 MHz. The BW
drops linearly to a maximum BW of 40 MHz as RF IN
frequency decreases to a frequency of 70 MHz or
increases to 605 MHz.
o
IF OUT level: Bypass path is -2.5 dB loss
o
Gain: N/A
o
Lower-side BW = RF IN - 50 MHz
o
Upper-side BW = 625 MHz - RF IN
M9392A Block Diagram Reference Table for M9202A
M9202A PXI Digitizer 12 Bit Operation
InputConnector Output
Input level: - 2 dBm
Frequency: 30 MHz to 1 GHz
Input level: +4 dBm ± 3 dB
Frequency: 100 MHz
Agilent M9392A PXI Vector Signal Analyzer Startup Guide19
INPUT 1
REF IN
Page 21
Step 5: Verify Operation of the Agilent M9392A PXI Vector Signal Generator
Step 5: Verify Operation of the Agilent M9392A PXI Vector
Signal Generator
In this step you will verify correct operation of the Agilent M9392A PXI Vector Signal Analyzer. Before running
a Self Test or performing a Field Calibration, assure that all required software is installed, the chassis is powered
on, and all cabling is correct. See "Agilent M9392A Modular Instrument Connections" on page 15 for proper
cabling.
The first step in this process is to conduct a Self Test of the M9392A
1. Open the M9392A soft front panel (SFP).
2. Upon opening the SFP, you are presented with the "Connect to Instrument" dialog. Select all the modules
that are components of the M9392A and press Connect. For example:
If your installed modules are not
listed, restart your PC or embedded controller to
establish these connections.
3. Conduct a Self Test (Utility > Self Test... > Run Self Test). The M9392A Self Test runs each module
through its own Self Test.
a. If the Self Test passes, go to next step.
b. If the M9392A Self Test fails, it indicates which module may need repair. See "M9392A Trou-
bleshooting" on page 23 to clearly determine the source of the failure.
20Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 22
Step 5: Verify Operation of the Agilent M9392A PXI Vector Signal Generator
Conduct M9392A Field Calibration
Required Hardware
Please refer to the following table for recommended hardware. Additionally, some high-quality flexible 3.5 mm
cables will be required, plus adapters to connect the PSG (precision signal generator) and power meter to 3.5 mm
connectors.
HardwareDescription
Agilent E8257D-532 or E8267D-532Precision Signal Generator 31.8 GHz
Agilent N1913A or N1914ASingle / Dual Channel Power Meter
Agilent N8485A, Option 10010 MHz to 26.5 GHz Power Sensor
Agilent 11730APower Sensor Cable
A Field Calibration should be performed upon initial setup and whenever one of more of the modules that compose
the M9392A PXI Vector Signal Analyzer is removed, reconfigured into different slots, or when cables are disconnected or changed. The Field Calibration should be repeated on a regular three month schedule, in order to
achieve the highest level of performance.
Perform an Agilent M9392A PXI Vector Signal Analyzer Field Calibration by using the M9392A SFP and the
included Help system.
1. Open the M9392A soft front panel (SFP).
2. Conduct a Field Calibration (Utility > Field Calibration...).
Use the Help button on the Agilent M9392A Field Calibration screen to access context-sensitive help to guide you
through the procedure.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide21
Page 23
Step 6: Installation is Complete
Step 6: Installation is Complete
Proceed to program your product by means of the application programming interface (API) for the supplied drivers.
API Overview
IVI Drivers
Agilent's IVI drivers simplify the creation and maintenance of instrument control applications in a variety of development environments; they allow programmatic control of instrumentation while providing a greater degree of
instrument interchangeability and code reuse. IVI drivers currently come in two basic types: IVI-COM and IVI-C.
Although the functionality offered by both types of drivers is often very similar, the fundamental differences in
interface technology results in a very different end-user experience. The IVI drivers support compiling application
programs for 32- or 64-bit platforms.
Supported ADEs: Arguably the most important consideration in comparing IVI-COM and IVI-C drivers is the
end user experience in various ADEs. Since IVI-COM drivers are based on Microsoft COM technology, it's not
surprising that IVI-COM drivers offer the richest user experience in Microsoft ADEs. Users working in Visual
C++, Visual C#, Visual Basic.NET, and Visual Basic 6 enjoy a host of features, such as object browsers, IntelliSense, and context-sensitive help.
When you install the product software, the IVI driver files are installed in the standard IVI Foundation directories
(for example, C:\Program Files\IVIFoundation\IVI\Drivers\AgM9392). Example programs are provided to demonstrate most driver functionality (for example, C:\Program Files\IVI Foundation\IVI\Drivers\AgM9392\Examples). The reference material for the driver functions (a Microsoft HTML Help .chm file) is installed with the
IVI driver and is available for Microsoft Visual Studio's IntelliSense context linking. In addition, you can directly
access the chm file (AgM9392.chm) from this Start menu location: Start > All Programs > Agilent IVI Drivers> AgM9392 > Documentation.
LabVIEW Driver
In addition to the IVI drivers, Agilent provides a LabVIEW driver that includes all the functionality of the IVI-C
driver. When you install the product software, the LabVIEW driver is installed to each LabVIEW instr.lib directory for each version of LabVIEW you have on your computer. Example programs are provided to demonstrate
most driver functionality. The reference information for the driver (a Microsoft HTML Help .chm file) is also
installed with the driver and the content is available from LabVIEW's Context Help window. In addition, you can
directly access the chm file (AgM9392_LabVIEW_Help.chm\) from this Start menu location: Start > All Pro-grams > Agilent > M9392 > AgM9392 LabVIEW Help.
M9392A Specifications
The Data Sheet for the M9392A is included on the Agilent M9392A VSA Software and Product Information CD
that came with your module.This document contains specification information. To find the latest update, go to
22Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 24
M9392A Troubleshooting
M9392A Troubleshooting
Communications
If you are unable to communicate with the M9392A, verify that the following installations are correct:
l Agilent IO Libraries Suite
l M9392A SFP
l Module and chassis drivers
l System Interface Card, cable and PC PXI card connections
l If no modules are visible in the SFP "Connect to Instrument" dialog or in Agilent Connection Expert, restart
your PC or embedded controller to establish these connections.
Modular Instrument Troubleshooting Overview
Between the low frequency path (through the M9351A) and the high frequency path (through the M9361A) there
are three common modules and one unique to the path (See " Agilent M9392A Block Diagram" on page 17).
To troubleshoot a modular instrument problem, first:
1. Check cabling.
2. Check the front panel Status LEDs.
3. Check the M9392A SFP (soft front panel) Status Tab and Self Tests.
4. Perform a modular instrument level operational check.
5. If a suspect module still can’t be identified, determine the frequency range at which the problem occurs and
See "Frequency Range Troubleshooting" on page 34 to identify the most likely module(s) and then perform
the module operational checks found in the Startup Guides.
Cables
l Start by verifying ALL cables are properly connected and SMA connectors are torqued to 8 Lb-In (0.904 Nm).
l See "Agilent M9392A Modular Instrument Connections" on page 15 for instrument-level cabling. M9392A
cable connections are also available from the M9392A SFP, by selecting Configure > Cable Configuration.
Modular Instrument Level Troubleshooting
Modular instrument level troubleshooting can be used to identify possible cabling errors and a limited number of
module performance issues without compromising the Field Calibration. The modular instrument level troubleshooting procedure starts by looking at the front panel LEDs, progresses to the modular instrument level soft
front panel (SFP) interface and then offers a limited set of operational checks aimed at identifying a suspect module. Once a suspect module is identified it becomes necessary to remove cables (compromising the Field Calibration) to perform the module operational check to verify the module's performance. At the modular instrument
level you are encouraged to modify frequency points and power levels to cover suspect problem areas.
Configuration/Setups
l Verify all module interconnect cable routings, and SMA connectors are torqued to 8 Lb-In (0.904 Nm).
l No RF signal is routed into the Agilent M9360A PXI Attenuator/Preselector RF IN connector.
l No 10 MHz reference signal is routed into the Agilent M9302A PXI Local Oscillator.
l The M9392A SFP (soft front panel) is active.
l SFP Measurement Frequency is set to 3 GHz (default condition).
Agilent M9392A PXI Vector Signal Analyzer Startup Guide23
Page 25
M9392A Troubleshooting
Status LED States
Once you open the M9392A SFP, select a module, and then choose Connect. The module is initial-
ized. Until it is initialized, the LED behavior is not entirely predictable.The front panel LED behavior is valid only when the soft front panel (SFP) is running or when the Initialize
function/method has been called, using the application programming interface (API).
Module GreenAmberRedOff
Identifies the module asso-
M9202A
Indicates the trigger is
armed and waiting for a
valid trigger to occur.
ciated with the current
channel in the Soft Front
Panel when the “Finder”
button is depressed.
M9302A
M9351A
M9360A
M9361A
All power supplies are
functioning within their tolerances, and all hardware is functioning and a
10 MHz Ref In is connected.
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is functioning.
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is functioning.
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is functioning.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hardware failures.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hardware failures.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hardware failures.
Indicates that all power
supplies are functioning
within their tolerances, but
there are other hardware
failures.
* Once the SFP is used, the LED may be red. When first plugged in, it may or may not be red.
Indicates that the trigger
has occurred, the acquisition is complete, and the
data is waiting to be read.
l Indicates a failure in
the power supplies and
module hardware.
l Indicates the absence
of the 10 MHz REF IN
signal or 10 MHz is
present, but SFP has
not initialized.
l Indicates a failure in
the power supplies and
module hardware.
l * When Meas
Frequency is ≥ 2.75
GHz
l When the M9351A LO
IN is disconnected.
l When the M9360A LO
IN is disconnected.
l When the 100 MHz
REF IN is disconnected.
Indicates a failure in the
power supplies and module
hardware.
Indicates a failure in the
power supplies and module
hardware.
Indicates a failure in the
power supplies. Module hardware health
can’t be determined
until the power supply
failure is resolved.
Indicates a failure in the
power supplies. Module hardware health
can’t be determined
until the power supply
failure is resolved.
Indicates a failure in the
power supplies. Module hardware health
can’t be determined
until the power supply
failure is resolved.
Indicates a failure in the
power supplies. Module hardware health
can’t be determined
until the power supply
failure is resolved.
Indicates a failure in the
power supplies, but
module hardware
seems to be functioning. Module hardware functioning
indication can be erroneous in this mode
since this circuitry can
be compromised by a
power supply failure.
24Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 26
Access LEDs (normal states)
l M9202A - none
l M9302A - blinks on and off when SFP is active
l M9360A - blinks on and off when SFP is active
l M9361A - blinks on and off when SFP is active
l M9351A - blinks on and off when SFP is active
M9392A Soft Front Panel Status Tab (normal states)
M9202AM9302AM9351AM9360AM9361A
l Temperature:
green
l Ref Level: green
l 10 MHz Ref In:
UNLOCKED (yellow) without a 10
MHz Ref In,
LOCKED (green)
with a 10 MHz Ref
In
l Ref: LOCKED
(green)
l Temperature: green
l Voltage (+12, +5,
+3.3, -12): green
l Temperature: green
l Voltage (+12, +5,
+3.3, -12): green
l Ref Level: OK
(green)
l Temperature:
green
l Voltage (+12,
+5, +3.3, -12):
green
l Temperature: green
l Voltage (+12, +5,
l Ref Level: OK (green)
M9392A Troubleshooting
+3.3, -12): green
M9392A Soft Front Panel Self Tests
Start the M9392A SFP self tests by selecting: Utility > Self Test > Run Self Test. Once completed, a dialog box
appears and reports on each of the modules, indicating its serial number and if it passed or failed.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide25
Page 27
M9392A Troubleshooting
Possible Self Test Failures
The chassis must be turned off when installing or removing modules.
1. Power supply failure:
a. If a voltage failure is reported, check the voltages on the other modules, using the SFP Status tab.
Since power supply failures can be related to chassis or module failures, the first step is to check for
a defective module.
b. If the problem is limited to the one module on the SFP Status tab, turn power off and move that mod-
ule to a different slot (cable connections are not required). Turn power back on and check again.
i. Replace the module if the voltage is still bad.
ii. If the voltage is good, after moving the module to a different slot, check the chassis power supply.
c. If more than one module voltage is bad, there is a possibility the chassis is bad or one of the modules
may be pulling down the power supply. Start by turning power off, removing one module at a time
and turning power on.
d. Continue removing one module at a time until only one module remains.
i. If the voltage is good after removing a module, reinstall the other modules.
ii. If the voltage is still good after reinstalling the other modules, the module that was pulling the
power supply down needs to be repaired.
iii. If the voltage is bad after reinstalling the other modules troubleshoot the chassis.
iv. If the voltage is still bad after removing all but the last module, remove the last module and rein-
stall one of the modules that was previously removed.
v. If the voltage is still bad, troubleshoot the chassis.
2. Temperature failure:
a. Temperature failures indicate over temperature.
i. Verify that blank panels have been installed in any open slot on the chassis.
ii. Make sure the chassis air intake and outlets are open and not blocked.
iii. Verify the chassis fans are working and fan speed to is set to Auto.
iv. Verify the ambient temperature is within the instrument operating range (0° C to 50° C).
v. If a single module over-temperature has not been resolved, the module is most likely starting to
fail and should be serviced.
3. M9351A - :LO Sense:
a. LO Sense indicates the M9351A is not able to detect the LO IN signal. If you get an LO Sense indi-
cation, check:
Removing the following cables negates the Field Calibration.
i. A5 cable connections between the M9351A LO IN and the M9360A Attenuator/Preselector LO
1OUT.
ii. A1 cable, M9302A LO OUT to M9360A LO IN
26Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 28
b. 100 MHz Ref Lock - Indicates the M9351A does not detect a 100 MHz REF IN. If you get a 100
MHz Ref Unlock indication, check cable B1, 100 MHz REF 1 OUT to M9351A 100 MHz REF IN.
If this cable is removed it does not affect the Field Calibration.
c. If all cables are properly connected, verify the LO outputs by using the M9302A Operational Check.
4. M9302A- LO Unlock:
a. LO Unlock indicates that the M9302A module has a LO failure and needs to be replaced.
Setting the M9392A SFP Input Power below the actual RF input level creates an OVERLOAD condition that stops sweep and can damage the downconverter. Once the Input Level is reset and the
OVERLOAD condition cleared, Select Sweep: Cont to restart sweep.
1. Select Utility > Reset.
2. Select the Display tab.
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale: 10 dB/div
M9392A Troubleshooting
d. Ref Level: +30 dBm
3. Under the Meas tab, the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 50 MHz
c. Input power: set to +30 dB
d. Time span: 100 µs
e. Trigger - Source Free Run
f. Sweep - Cont.
4. Connect an RF signal to the M9360A RF IN. Set the source to the following frequency and power:
a. Frequency: 50 MHz
b. Power Level: -20 dBm
5. With the M9392A set to display frequency:
a. Monitor the signal level while stepping the source and M9392A up in frequency in 100 MHz steps to
2.74 GHz.
b. The power level should stay within ± 3 dB.
c. If power levels vary more than expected, check that cables are properly connected and torqued.
d. Recheck areas of concern.
e. If power levels still vary greater than expected, perform the M9351A, M9360A, and M9302A mod-
ule level operational checks.
6. Select Utility > Reset.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide27
Page 29
M9392A Troubleshooting
7. Select the Display tab:
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale: 10 dB/div
d. Ref Level: +30 dBm
8. Under the Meas tab, the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 2 GHz
c. Input power: set to +30 dBm.
d. Bandwidth: set to 40 MHz
e. Time span: 100 µs
f. Trigger - Source: Free Run
g. Sweep - Cont.
9. Connect an RF signal to the M9360A RF IN. Set the source to the following frequency and power:
a. Frequency: 2 GHz
b. Power level: -20 dBm
10. On the Meas tab:
a. Step the Input Power down in 10 dB increments to 0 dBm.
b. The path is alright if:
i. The noise floor drops approximately 10 dB each time the Input Power level is decreased.
ii. The attenuator clicks as it switches. The attenuator switches when stepping between the levels
displayed in the table in Step 11.
11. To check all levels, set the following conditions, and set the input power level first to avoid an OVERLOAD condition.
Set Display Ref-
RF Input Power
-20 dBm+30 dBm+30 dBm+20 dBm
-20 dBm+30 dBm+20 dBm+10 dBm
-20 dBm+0 dBm+10 dBm+0 dBm
-20 dBm+0 dBm+0 dBm-10 dBm
-20 dBm-20 dBm-10 dBm-20 dBm
-50 dBm-30 dBm-20 dBm-30 dBm
-50 dBm-30 dBm-30 dBm-40 dBm
Level
Change Meas Input level
FromTo
12. Reset Meas Input Power to +30 dBm.
13. On the source:
a. Step the power level down to +30 dBm.
b. The power level on the M9392A should drop by 10 dB with each decrease in power level.
28Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 30
M9392A Troubleshooting
Measure the IF OUT levels <2.75 GHz
1. Source input frequency: 2 GHz
2. Power level: -10 dBm
3. M9392A SFP settings:
a. Frequency: 2 GHz
b. Input power: -10 dBm
4. At the M9351A IF OUT: ~260 mV p-p @ 500 MHz (~ -1.5 dBm to -4.0 dBm)
5. At the M9361A IF OUT: ~250 mV p-p @ 500 MHz (~ -1.5 dBm to -4.5 dBm)
Setting the M9392A SFP Input Power below the actual RF input level creates an OVERLOAD condition that stops sweep and can damage the downconverter. Once the Input Level is reset and the
OVERLOAD condition cleared, Select Sweep: Cont to restart the sweep.
1. Select Utility > Reset
2. Select the Display tab.
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale 10 dB/div
d. Ref Level: +30 dBm
3. Under the Meas Tab the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 2.25 GHz
c. Input power: set to +30 dBm
d. Time span: 100 µs
e. Trigger – Source Free Run
f. Sweep – Cont.
4. Connect an RF signal to the M9360A RF IN. Set the bandwidth to 250 MHz and the source to the following
frequency and power levels:
a. Frequency: 2.25 GHz
b. Power level: -20 dBm
Agilent M9392A PXI Vector Signal Analyzer Startup Guide29
Page 31
M9392A Troubleshooting
5. With the M9392A set to display frequency:
a. Monitor the signal level while stepping the source and M9392A up in frequency in 100 MHz steps to
2.70 GHz.
b. The power level should stay within ± 3 dB.
c. If power levels vary more than expected, check that cables are properly connected and torqued.
d. Recheck areas of concern.
e. If power levels still vary greater than expected, perform the M9351A, M9360A, and M9302A mod-
ule level operational checks.
6. Select Utility > Reset.
7. Select the Display tab:
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale: 10 dB/div
d. Ref Level: +30 dBm
8. Under the Meas tab, the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 2.25 GHz
c. Input power: set to +30 dBm.
d. Bandwidth: set to 250 MHz
e. Time span: 100 µs
f. Trigger - Source: Free Run
g. Sweep - Cont.
9. Connect an RF signal to the M9360A RF IN. Set the source to the following frequency and power:
a. Frequency: 2.25 GHz
b. Power level: -20 dBm
10. On the Meas tab:
a. Step the Input Power down in 10 dB increments to 0 dBm.
b. The path is alright if:
i. The noise floor drops approximately 10 dB each time the Input Power level is decreased.
ii. The attenuator clicks as it switches. The attenuator switches when stepping between the levels
displayed in the table in Step 11.
30Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 32
M9392A Troubleshooting
11. To check all levels, set the following conditions, and set the input power level first to avoid an OVERLOAD condition.
Set Display Ref-
RF Input Power
-20 dBm+30 dBm+30 dBm+20 dBm
-20 dBm+30 dBm+20 dBm+10 dBm
-20 dBm+0 dBm+10 dBm+0 dBm
-20 dBm+0 dBm+0 dBm-10 dBm
-20 dBm-20 dBm-10 dBm-20 dBm
-50 dBm-30 dBm-20 dBm-30 dBm
-50 dBm-30 dBm-30 dBm-40 dBm
erence Level
Change Meas Input level
FromTo
12. Reset Meas Input Power to +30 dBm.
13. On the source:
a. Step the power level down to +30 dBm.
b. The power level on the M9392A should drop by 10 dB with each decrease in power level.
Measure the IF OUT levels from 2.25 GHz to 2.75 GHz
1. Source input frequency: 2.25 GHz
2. Power level: -10 dBm
3. M9392A SFP settings:
a. Frequency: 2.25 GHz
b. Bandwidth: 250 MHz
c. Input power:-10 dBm
4. At the M9361A IF OUT: ~250 mVp-p @ 500 MHz (~-2.5 dBm to -3.0 dBm)
Setting the M9392A SFP Input Power below the actual RF input level creates an OVERLOAD condition that stops sweep and can damage the downconverter. Once the Input Level is reset and the
OVERLOAD condition cleared, Select Sweep: Cont to restart the sweep.
1. Select Utility > Reset
2. Select the Display tab.
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale 10 dB/div
d. Ref Level: +30 dBm
Agilent M9392A PXI Vector Signal Analyzer Startup Guide31
Page 33
M9392A Troubleshooting
3. Under the Meas Tab the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 3 GHz
c. Input power: set to +30 dBm
d. Time span: 100 µs
e. Trigger – Source Free Run
f. Sweep – Cont.
4. Connect an RF signal to the M9360A RF IN. Set the source to the following frequency and power levels:
a. Frequency: 3 GHz
b. Power level: -20 dBm
5. With the M9392A set to display frequency:
a. Use the M9392A SFP to monitor the signal level while stepping the source and M9392A up in
frequency in 500 MHz steps to 26.5 GHz.
b. The power level should stay within ± 3 dB.
c. If power levels vary more than expected check that all cables are properly connected and torqued.
d. Recheck areas of concern.
e. If power levels still vary more than expected, perform the M9361A, M9360A, and M9302A module
level operational checks.
6. Select the Display tab.
a. Set Display: Frequency
b. Frequency Display: Auto Scale (unchecked)
c. Scale 10 dB/div
d. Ref Level: +30 dBm
7. Under the Meas Tab the Settings should be as follows:
a. Use Preselector: (unchecked)
b. Frequency: 3 GHz
c. Input power: set to +30 dBm
d. Time span: 100 µs.
e. Trigger Source: Free Run
f. Sweep – Cont.
8. Connect an RF signal to the M9360A RF IN. Set the source to the following frequency and power:
a. Frequency: 3 GHz
b. Power level: -20 dBm
32Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 34
M9392A Troubleshooting
9. On the Meas Tab:
a. Step the Input Power down in 10 dB increments to 0 dBm.
b. The path is performing properly if:
i. The noise floor drop approximately 10 dB each time the Input Power level is decreased.
ii. The attenuator clicks as it switches. The attenuator switches when stepping between the levels
listed in the table in Step 10.
10. To check all power levels set the following conditions and set input power level first to avoid an OVERLOAD condition:
Set Display Ref-
RF Input Power
-20 dBm+30 dBm+30 dBm+20 dBm
-20 dBm+30 dBm+20 dBm+10 dBm
-20 dBm+0 dBm+10 dBm+0 dBm
-20 dBm+0 dBm+0 dBm-10 dBm
-20 dBm-20 dBm-10 dBm-20 dBm
-50 dBm-30 dBm-20 dBm-30 dBm
-50 dBm-30 dBm-30 dBm-40 dBm
erence Level
Change Meas Input level
FromTo
11. On the Meas Tab, select Use Preselector. The power level should be close to the same level when in the
thru path (Use Preselector unchecked) and the preselector path.
12. Repeat Step 6 at 10 GHz. 20 GHz and 26.5 GHz.
13. Reset Meas Input Power to +30 dBm.
14. On the Source:
a. Step the power level down in 10 dB steps.
b. The power level on the M9392A should drop by 10 dB with each decrease in power level.
15. If a problem occurs:
a. If no output, for frequencies >2.75 GHz , check the M9351A path.
b. If the M9351A is good, go to the module level operational check and measure the M9360A RF 2 and
LO 2 OUT.
c. If the M9360A fails, check the M9302A.
d. If the M9361A passes the operational check, check the signals at M9361A end of A3.
e. If the M9360A is good, using the module level operational check, repeat the M9361A operational
check.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide33
Page 35
M9392A Troubleshooting
Measure the IF OUT levels ≥2.75 GHz
1. Source input frequency: 20 GHz
2. Power level: -10 dBm
3. M9392A SFP settings:
a. Frequency: 20 GHz
b. Input power:-10 dBm
4. At the M9361A IF OUT: ~250 mVp-p @ 500 MHz (~-1.5 dBm to -4.0 dBm)
Settable Bandwidths and IF frequencies
Maximum
Settable
Frequency
≤ 625 MHz500 MHz500 MHz or equal to the signal frequency if <500 MHz
>625 MHz to 2.74 GHz (M9351A path) 40 MHz500 MHz
>2.75 GHz to 26.5 GHz
Preselector - off
>2.75 GHz to 26.5 GHz
Preselector - on
2.25 GHz to 2.75 GHz (M9361A path)250 MHz500 MHz
BandwidthIF Frequency
250 MHz
30 MHz468 MHz
500 MHz when bandwidth is >50 MHz
468 MHz when bandwidth is <50 MHz
Frequency Range Troubleshooting
When problems occur that are related to a specific frequency range, proceed by:
1. Checking cables for proper connection and torque.
2. Performing module level operational checks and using the following table.
FrequencyMost Likely Module Next Most Likely module
50 MHz to 450 MHzM9351AM9360A
50 MHz to 2.74 GHzM9351AM9360A
>2.75 GHzM9361AM9360A
50 MHz to 26.5 GHzM9302AM9360A
34Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 36
Streaming Data Overview
Streaming Data Overview
The Agilent M9392A PXI Vector Signal Analyzer with streaming data allows you to capture continuous (gapless)
IQ data to files. Streaming data is acquired from the Agilent M9202A PXIe IF Digitizer to a circular RAM buffer
on the system controller (which handles latency in disk file writes), and is simultaneously written from the RAM
buffer to the controller file system. When used with an Agilent-qualified RAID storage system, up to 100 MHz
bandwidth data can be continuously captured for unlimited durations (limited only by available disk space). When
used with slower disk drives, streaming data can be captured to available system RAM (circular capture buffer)
and then subsequently written to the slow disk drive.
l For product-specific information and support, visit www.agilent.com/find/pxi-streaming.
l To read an application note about M9392A streaming, see http://cp.literature.agilent.com/litweb/pdf/5990-
8872EN.pdf.
Basic Steps to Using the M9392AStreaming Capability
1. Select the storage capability. See "Streaming System Configuration" on page 38.
2. Capture the streaming data. This can be accomplished with the M9392A Soft Front Panel (see the SFP help
for details), or with one of the M9392A drivers (see the IVI driver help or LabVIEW driver help for details).
l You can download the M9392A SFP help from www.agilent.com/find/M9392A. If you've run the full
M9392A installer, you can access the help file from this program group in the Windows Start menu:
Agilent > M9392 > M9392 Help.
l You can download the driver help files from www.agilent.com/find/M9392A. If you've run the full
M9392A installer, you can access the respective driver help file from these program groups in the Windows Start menu:
LabVIEW Driver Help: Agilent > M9392 > M9392 LabVIEW Help
3. Use the Agilent DataViewer application to review the captured data and "snip" out content of interest to be
played back in the Agilent 89600 Series VSA software, or converted to play on one of Agilent's arbitrary
waveform generators (AWG) or vector signal generators (VSG). For details on using the DataViewer application, refer to the DataViewer help system. You can download the DataViewer help from www.ag-
ilent.com/find/M9392A; or if you've run the full M9392A installer, you can access the help file from this
program group in the Windows Start menu: Agilent > M9392 > DataViewer Help.
4. Analyze the captured data. Instructions are provided in the DataViewer help and the M9392A SFP help.
Agilent 89600 Series VSA Software is recommended to process the recorded data files.
System Requirements
In addition to the standard system requirements See "Step 3: Install the Software" on page 11), your system must
have the software and firmware revisions listed below. If you need to upgrade, see Upgrading your system below.
l
Agilent IO Libraries Suite (IOLS) Version 16.3.16603.3 or higher. This software must be installed first.
l
If you use Agilent 89600 Series VSA Software, you must have version B (89600B) before updating the
M9392A software.
l
M9392A software revision 1.1 or higher
l
M9202A firmware revision 1.25 or higher
If you purchased your M9392A VSA instrument after September 1, 2011, it will have one of the following
M9202A IF Digitizer module options, depending on your order:
l
M9202AU-V05: This option is capable of 50 MHz streaming (standard).
l
M9202AU-V10: This option is capable of 100 MHz streaming (option).
Agilent M9392A PXI Vector Signal Analyzer Startup Guide35
Page 37
Streaming Data Overview
To use streaming capture capabilities, properly configure the Agilent M9021A PCIe Cable Interface.
See the M9021A Installation Guide at http://cp.literature.agilent.com/litweb/pdf/M9021-90001.pdf.
Upgrading your System
Use the following procedure if you need to upgrade your M9392A instrument for streaming. (NOTE: You can perform the upgrade at your site; there is no need to return any equipment to Agilent.)
1.
Contact an Agilent representative (see www.agilent.com/find/assist) to purchase one of these upgrade kits:
l
M9202AU-V05: 50 MHz streaming update kit
l
M9202AU-V10: 100 MHz streaming update kit
2.
When Agilent receives payment for the Streaming Upgrade Kit, Agilent sends you the following:
l
License file (*.epr file), provided on a USB flash drive (This file is also sent via email.)
l
Agilent M9392A Product Software and Information CD (M9392-10002)
l
M9202A – Option Upgrade Procedure (M9202-90011)
3.
Make sure you have M9392A software version 1.1 or later installed. Also, make sure the M9202A module
is installed in the chassis. If you have a version older than 1.1, install the latest software from the supplied
CD, or from the Drivers & Software tab on the Technical Support page at www.agilent.com/find/M9392A.
If the installer indicates it cannot proceed without removing the currently installed M9392A
software, uninstall the M9392A software using Windows "Add or Remove Programs" utility.
4. Update the firmware and license files on the M9202A IF Digitizer:
a. Launch the AcqEepromProg.exe file (EEPROM Programmer application). The file location depends
on your operating system:
l 32-bit OS: C:\Program Files\Agilent\MD1\bin\AcqEepromProg.exe
l 64-bit OS: C:\Program Files (x86)\Agilent\MD1\bin\AcqEepromProg.exe
b. In the EEPROM Programmer's File choice dialog, open the *.vme file (example: m9202_ddc_
02000381.vme) from one of these locations:
l 32-bit OS: C:/Program Files/IVI Foundation/IVI/Drivers/AgMD1/Firmware
Once you Open the .vme file, it takes approximately 30 to 60 minutes to com-
plete the firmware update. DO NOT STOP THE PROGRAM during this process; doing so could damage your M9202A module.
36Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 38
Streaming Data Overview
c. Close the EEPROM Programmer application, and all other applications.
d. Cycle power to the chassis.
If you are using a remote controller, Shut Down the PC BEFORE you power down the
chassis. When you restore power, power up the chassis BEFORE you power up the
PC.
e. Run the EEPROM Programmer application again, and load the license (*.epr) file sent to you by Agi-
lent Support (see Step 2 above). The license file update should only take several seconds.
5. To verify the streaming upgrade is complete and operating correctly, use the M9392A Soft Front Panel
(SFP) to conduct a streaming data capture test:
a. Open the M9392 SFP by selecting: Start > All Programs > Agilent > M9392 > M9392 SFP.
b. In the SFP, select the Streaming Enabled checkbox () to enable streaming.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide37
Page 39
Streaming Data Overview
c.
Create a test.bin file to store the streaming data. To do this, click the(File Save As) button.
In the Save As dialog, type in the file name “test.bin” and click Save.
d. Under Acquisition Mode, select Duration Limited.
e. Set the Acquisition Duration to 4 seconds.
f. If you ordered the M9202AU-V10 upgrade, set the bandwidth between 50 MHz and 100 MHz.
Otherwise, leave the bandwidth at the default of 40 MHz.
g. Click the Start button to initiate the data capture. When the streaming session is complete, the
status bar shows "Streaming Complete" and the Start button is re-enabled (Stop and Abort buttons are disabled).
You have completed the Streaming upgrade for your M9392A Vector Signal Analyzer and M9202A IF Digitizer.
Streaming System Configuration
The Agilent M9392A streaming solution uses a controller disk drive, and (optionally) an external RAID solution.
Controller and RAID selection depends on the data capacity, and data read/write rate requirements. The following
are Agilent-recommended storage packages:
Agilent's predefined external RAID packages have been tested to guarantee sustained data rates
with the M9392A. These predefined mass storage packages ensure performance when used with the
recommended controllers.
For details on connecting and configuring the above JMR RAID solutions to work with the M9392A,
refer to the white paper (located on the web at http://cp.literature.agilent.com/litweb/pdf/5990-
9483EN.pdf). If you installed the M9392A software, you can access the white paper from
38Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 40
System diagram:
Streaming Data Overview
Agilent M9392A PXI Vector Signal Analyzer Startup Guide39
Page 41
Multichannel Overview
Multichannel Overview
The M9392A provides these multichannel capabilities:
l Multichannel Wideband Signal Analysis: Provides wide-band multichannel analysis capability by com-
bining two M9392A vector signal analyzers in one chassis.See "Agilent 89600 Vector Signal Analysis Soft-
ware" on page 45. The M9392A modules can be configured to use a common LO, or independent LOs. See
http://cp.literature.agilent.com/litweb/pdf/5991-0135EN.pdf.
l Multichannel Streaming: Combines the M9392A’s multichannel capability with gapless recording to
address a wide variety of test needs in wireless communications and aerospace industries. Agilent’s
predefined packages have been tested to guarantee sustained data rates. These predefined mass storage
packages ensure performance when used with the recommended controller. See
l Two M9392A Vector Signal Analyzers (full configuration) installed into one chassis. You have the option
of using two independent M9302A LO modules, or sharing one M9302A LO module. For details on these
configurations, see http://cp.literature.agilent.com/litweb/pdf/5990-8254EN.pdf.
o
(Optional) Agilent's predefined packages for streaming data storage have been tested to guarantee
sustained data rates. For details see http://cp.literature.agilent.com/litweb/pdf/5990-9483EN.pdf.
l Agilent M9392A software version 1.2 or greater,and Agilent M9202A firmware version 512 or greater. If
you need to upgrade your system, install the latest software from the supplied CD or from the Drivers and
Software tab on the Technical Support page at www.agilent.com/find/M9392A. If you intend to use mul-
tichannel streaming but have not performed a streaming upgrade, See "Upgrading your System" on page 36.
l Agilent IOLibrary Suite (IOLS) Version 16.3.16603.3 – see http://www.agilent.com/find/iosuite
l Agilent 89600 VSA software version 15.0 – see http://www.agilent.com/find/VSA
Basic Steps to Using the M9392AMultichannel Capability
M9392A Soft Front Panel (SFP)
The M9392A SFP is used primarily to configure and control a complete Agilent M9392A PXI Vector Signal
Analyzer, or a single Agilent M9202A PXIe IF Digitizer in digitizer-only mode. It can only be used to control one
channel of M9392A or M9202A digitizer at any one time. It is possible to launch multiple running SFPs to control
multiple channels concurrently. However, the multichannel trigger capability cannot be controlled by way of the
SFP, so the SFP is not very useful when synchronized multichannel operation is required.
Configuration Assumptions
l All software is installed – see "Step 3: Install the Software" on page 11.
l All hardware is installed and cabled – see "Step 4: Install the Modules" on page 13.
If are using a version of Agilent 89600 VSA software previous to release 15.0, do not use the pro-
cedure below. Instead, use the Connection procedure found in the M9392A SFP Help system. Open
the help system (Programs > Agilent > M9392 > M9392 Help) and navigate to the "Setting Up Mul-
tichannel Analysis" topic.
40Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 42
Multichannel Overview
Configuration with Agilent 89600 VSA Software (Recommended for VSA version 15.0 and later)
To create a new M9392A instrument using the modules method, use the Agilent 89600 VSA Software and follow
these steps:
1. Open the "New Hardware Configuration" window in the Agilent 89600 VSA software by selecting: Util-ities > Hardware > Configurations from the menu bar. When you hover over the last element in this
sequence you will see a listing of the relevant configurations. Click on the "Add New Hardware" icon.
2. In the "New Hardware Configuration" window:
a.
Select the Agilent M9392A Analyzer (Modules) entry and click on theicon to copy the
entry to the "Configuration" window.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide41
Page 43
Multichannel Overview
b. Configure the new M9392A (Modules) instrument by selecting the modules identified by their
VISA address in the "2. Select the instrument(s) to use for each logical instrument in the configuration" section.
c. Name the new configuration in the "3. Name the configuration" section. The default name
"Analyzer1"may be changed. Use a meaningful name for later identification.
d. Save the new configuration by selecting the "OK" button.
42Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 44
Multichannel Overview
For purposes of illustration, the chassis has been populated with two M9392A logical instru-
ments (one in slots 2 through 9 and one in slots 11 through 18. Notice in the figure above, that,
by default, the Agilent 89600 VSA software has configured a logical instrument by using the
modules in the lowest slot numbers. With this arrangement you can establish two independent-LOlogical instruments.
You have now configured one "logical instrument". Follow the rest of this procedure to create a second logical instrument and establish the Master/Slave relationship between the two configurations.
3. From the same "New Hardware Configuration window: select (highlight) "Agilent M9392A Analyzer (Modules)" under "Possible Logical Instruments" and click on theicon to copy the entry to the "Configuration"
window to create a new logical instrument.
This is now the second entry in the Configuration window and is highlighted. Notice the red Hint in the following figure. Since the VSAsoftware has selected the modules in the lowest numbered (left-most) slots, it
appears that the M9202A (Adc) is being used in both configurations.
The relative position of the two logical instruments in the Configuration window determines
the Master/Slave relationship The instrument at the top is the master and the one at the bottom
is the slave. Use the up/down icons shown below to change this relationship
4. Click on the Agilent M9202A IFDigitizer dropdown menu(see bottom arrow below). The dropdown displays the slot of both M9202A modules in the chassis. Select the one in Slot 11 at this time. The red "Hint"
will persist until you select a module from each of the choices (two downconverters, one synthesizer and
one Input) and assign the unused module to complete the second configuration.
5. Name the configuration. Double-click on the "3. Name the configuration dialog and give the two-channel
configuration a descriptive name, such as the one shown in the following figure:
Agilent M9392A PXI Vector Signal Analyzer Startup Guide43
Page 45
Multichannel Overview
You have now created a two-channel independent LO configuration.
In step 2 you could have created a shared LOconfiguration by leaving the Synthesizer (LO)
module in the "slave" instrument the same as the LOin the "master" instrument. Remember
to use a descriptive name for all configurations.
6. To create a digitizer-only logical instrument, you can repeat Step 1 and when presented with the "2. Select
the instrument(s) to use for each logical instrument in the configuration" section:
a. Select the "Agilent M9202A IF Digitizer" dropdown and make your choice.
b. For each of the of the other dropdown choices select None.
c. When you have configured this digitizer-only logical instrument, Name it appropriately.
7. To create a two-channel digitizer-only configuration, repeat Step 4 and when you have both digitizer-only
configurations in the Configuration window, you can establish the Master/Slave relationship.
Work Processes
l For more detail about implementing Multichannel Wideband Analysis see
l A comprehensive collection of resources for multichannel PXI analysis is provided at
http://www.agilent.com/find/pxi-vsa-dualchannel.
44Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Page 46
Agilent 89600 Vector Signal Analysis Software
Agilent 89600 Vector Signal Analysis Software
To use the Agilent 89600 Series VSA Software measurement capabilities, you must first create a connection
between the software and the hardware configuration that is your M9392A Vector Signal Analyzer. If you are
using M9392A instruments in a multi-channel configuration, see Multichannel Overview (page 40).
If are using Version 14.2 of Agilent 89600 VSA software, do not use the procedure below. Instead,
use the Connection procedure found in the M9392A SFP Help system. Open the help system (Pro-grams > Agilent > M9392 > M9392 Help) and navigate to the "Using Measurement Capabilities of
the Agilent 89600 Series VSA Software" topic.
To create a new M9392A instrument using the modules method, use the Agilent 89600 VSA Software and follow
these steps:
1. Open the "New Hardware Configuration" window in the Agilent 89600 VSA software by selecting: Util-ities > Hardware > Configurations from the menu bar. When you hover over the last element in this
sequence you will see a listing of the relevant configurations. Click on the "Add New Hardware" icon.
2. In the "New Hardware Configuration" window:
Agilent M9392A PXI Vector Signal Analyzer Startup Guide45
Page 47
Agilent 89600 Vector Signal Analysis Software
a.
Select the Agilent M9392A Analyzer (Modules) entry and click on theicon to copy the
entry to the "Configuration" window.
b. Configure the new M9392A (Modules) instrument by selecting the modules identified by their
VISA address in the "2. Select the instrument(s) to use for each logical instrument in the configuration" section.
c. Name the new configuration in the "3. Name the configuration" section. The default name
"Analyzer1"may be changed. Use a meaningful name for later identification.
d. Save the new configuration by selecting the "OK" button.
46Agilent M9392A PXI Vector Signal Analyzer Startup Guide