No part of this manual may be reproduced in
any form or by any means (including
electronic storage and retrieval or
translation into a foreign language) without
prior agreement and written consent from
Keysight Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
M9240-97010
Edition
Third edition, February 2018
Printed in Malaysia
Published by:
Keysight Technologies, Inc.
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Print History
M9240-97000, January 2017
M9240-97007, May 2017
M9240-97010, February 2018
Warranty
The material contained in this document is
provided "as is," and is subject to being
changed, without notice, in future ed itions.
Further, to the maximum extent permitted
by applicable law, Keysight disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, including but not l imited
to the implied warranties of
merchantability and fitness for a particular
purpose. Keysight shall not be liable for
errors or for incidental or consequential
damages in connection with the furnishing,
use, or performance of this document or of
any information contained herein. Should
Keysight and the user have a separate
written agreement with warranty terms
covering the material in this document that
conflict with these terms, the warranty
terms in the separate agreement shall
control.
Technology License
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in
accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer
software," as defined by Federal Acquisition
Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires
commercial computer software under the
same terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the Software
to U.S. government customers under its
stand ard commercial l icense, which is
embodied in its End User License Agreement
(EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify, distribute, or
disclose the Software. The EULA and the
license set forth therein, does not require or
permit, among other things, that Keysight:
(1) Furnish technical information related to
commercial computer software or
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish to,
or otherwise provide, the government rights
in excess of these rights customarily
provided to the public to use, modify,
reproduce, release, perform, display, or
disclose commercial computer software or
commercial computer software
documentation. No additional government
requirements beyond those set forth in the
EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly
required from all providers of commercial
computer software pursuant to the FAR and
the DFARS and are set forth specifically in
writing elsewhere in the EULA. Keysight
shall be under no obligation to update,
revise or otherwise modify the Software.
With respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102,
the U.S. government acquires no greater
than Limited Rights as defined in FAR 27.401
or DFAR 227.7103-5 (c), as applicable in any
technical data.
Safety Notices
This product has been designed and tested
in accordance with accepted industry
standards, and has been supplied in a safe
condition. The documentation contains
information and warnings that must be
followed by the user to ensure safe
operation and to maintain the product in a
safe condition.
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly
performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated
conditions are fully understood and
met.
2Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide5
Page 6
6Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 7
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
WARNING
Startup Guide
1Introduction
Follow the Startup Sequence / 7
Related Documentation / 8
The purpose of this Startup Guide is to detail the processes of receiving and
installing the Keysight M9241/42/43A PXIe oscilloscope modules and the Keysight
M9240A AutoProbe power module.
The chapters in this guide cover the basics of setting up and configuring a PXIe
oscilloscope system, as well as installing the required software.
If you have any questions after reviewing this information, please contact your
local Keysight Technologies representative or contact us through our website at
www.keysight.com/find/pxi-digitizers.
Follow the Startup Sequence
Closely follow the startup process flow in this document. Deviating from the
sequence can cause unpredictable system behavior, damage your system, and may
cause personal injury.
1 Unpack and inspect. See: Chapter 2, “Unpack and Inspect the Module,”
starting on page 11
2 Verify the shipment. See: Chapter 3, “Verify Shipment Contents and Model
Options,” starting on page 15
7
Page 8
1Introduction
3 Install modules. See: Chapter 4, “Install the Modules,” starting on page 17
4 Install drivers and software. See: Chapter 5, “Install the Software,” starting on
page 25
5 Verify operation with the Soft Front Panel (SFP). See: Chapter 6, “Verify
Operation,” starting on page 29
6 Make a measurement. See: Chapter 7, “Make a Measurement,” starting on
page 35
7 (Optional) If you want to coordinate acquisitions between multiple PXIe
8 Installation is complete. Proceed to program your product through the API. See:
Chapter 9, “Applications Programming Interface (API) Overview,” starting on
page 61
Related Documentation
To access documentation related to the Keysight M9241/42/43A PXIe
oscilloscope modules and the Keysight M9240A AutoProbe power module, use
one of the following methods:
•The related documents are available on the product CD:
8Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 9
Introduction1
DocumentDescriptionFile nameFormat
Startup Guide
(this manual)
Soft Front Panel
(SFP) User's
Guide
SCPI
Programmer's
Guide
IVI Programming
Guide
IVI Driver
reference (help
system)
LabVIEW Driver
Reference
Includes procedures to help you to
unpack, inspect, install (hardware
and software), verify operation, and
make a basic measurement.
Shows how to use the InfiniiVision
M9241/42/43A PXIe oscilloscope's
Soft Front Panel (SFP) user
interface.
Shows how to program the
M9241/42/43A PXIe oscilloscopes
using SCPI commands.
Shows you how to use Visual Studio
2010 with the .NET Framework to
write IVI-COM Console Applications
in Visual C#.
Provides detailed documentation of
the IVI-COM and IVI-C driver API
functions, as well as information to
help you get started with using the
IVI drivers in your application
development environment.
Provides detailed documentation of
the LabVIEW G Driver API functions.
M924x_StartupGuide.pdfPDF
M924x_SFP_Users_Guide.pdfPDF
M924x_SFP_Users_Guide.chmCHM (Microsoft
Help Format)
M924x_SCPI_Programmers_Guide.chmCHM (Microsoft
Help Format)
M924x_SCPI_Programmers_Guide.pdfPDF
M924x_IVI_ProgrammingGuide.pdfPDF
AgInfiniiVision.chmCHM (Microsoft
Help Format)
KtInfiniiVision_LabVIEW_Help.chmCHM (Microsoft
Help Format)
•To find the latest versions of the user documentation, go to
www.keysight.com/manuals/M9241A.
See AlsoThe data sheet introduces the product and provides full product specifications.
You can find the data sheet at: www.keysight.com/products/M9241A
The Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power
Module Security Guide is available at www.keysight.com/find/security.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide9
Page 10
1Introduction
10Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 11
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
CAUTION
Startup Guide
2Unpack and Inspect the
Module
Inspect for Damage / 12
Return an Instrument for Service / 12
The modules are shipped in materials which prevent damage from static. Modules
should only be removed from the packaging in an anti-static area ensuring that correct
anti-static precautions are taken. Store all modules in anti-static envelopes when not
in use.
Electrostatic discharge (ESD) can damage or destroy electronic components. Use
a static-safe work station to perform all work on electronic assemblies. The
following figure shows a static-safe work station using two types of ESD
protection: conductive table-mat and wriststrap combination, and conductive
floor-mat and heelstrap combination. Both types, when used together, provide a
significant level of ESD protection. Of the two, only the table-mat and wrist-strap
combination provides adequate ESD protection when used alone. To ensure user
safety, the static-safe accessories must provide at least 1 M
ground.
Ω of isolation from
11
Page 12
2Unpack and Inspect the Module
WARNING
CAUTION
NOTE
DO NOT use these techniques for a static-safe work station when working on
circuitry with a voltage potential greater than 500 volts.
Inspect for Damage
After unpacking an instrument, inspect it for any shipping damage. Report any
damage to the shipping agent immediately, as such damage is not covered by the
warranty.
To avoid damage when handling a module, do not touch exposed connector pins.
Return an Instrument for Service
Should it become necessary to return an instrument for repair or service, follow
the steps below:
It is recommended that you return all modules and cables of the M924xA instrument for repair
and calibration.
12Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 13
Unpack and Inspect the Module2
1 Review the warranty information shipped with your product.
2 Contact Keysight to obtain a Return Material Authorization (RMA) and return
address. For assistance finding Keysight contact information, go to
www.keysight.com/find/assist.
3 Write the following information on a tag and attach it to the malfunctioning
equipment:
•Name and address of owner. A P.O. box is not acceptable as a return
address.
•Module serial number(s). The serial number label is located on the side
panel of the module. The serial number can also be read from the Soft Front
Panel interface - after the hardware is installed.
•Description of failure or service required.
4 Pack the instrument in its original packaging. Include all cables. If the original
packaging material is not available, use anti-static bubble wrap or packing
peanuts and place the instrument in a sealed container and mark the container
-FRAGILE-.
5 On the shipping label, write ATTENTION REPAIR DEPARTMENT and the RMA
number.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide13
Page 14
2Unpack and Inspect the Module
14Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 15
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
M9240CMPAComputer Serial Triggering and Analysis (RS232/422/485/UART)
M9240ATOAAutomotive Serial Triggering and Analysis (CAN, CAN-dbc, CAN FD, LIN)
M9240SNSASENT (Single Edge Nibble Transmission) Triggering and Analysis
M9240CXPACXPI Trigger and Decode
M9240FRAAFrequency Response Analyzer
M9240NFCANFC Automated Test for X-Series Oscilloscopes
M9241A-808, M9242A-808,
or M9243A-808
Qty 2 of the N2843-60001 (N2843A passive probe - 500 MHz 10:1)
M9240A AutoProbe Power Module Shipment Contents
Items included in your M9241/42/43A PXIe oscilloscope shipment:
QtyKeysight Part NumberDescription
1M9240AAutoProbe power module, includes one faceplate for oscilloscope module in slot to the
left and one faceplate for oscilloscope module in slot to the right
15185-1605Envelope-Calibration Certificate (230 mm x 153 mm)
15991-3402End User License Agreement
19320-6678China RoHS Addendum for Oscilloscope
19320-6797Manual Safety-booklet Keysight branded
1E2094-60003CD-ROM, IO Libraries Media Suite
1M9240-94xxxSoftware and Product Information CD, contains: Soft Front Panels, drivers, and all
printed documentation in PDF format (also available at
http://www.keysight.com/support/M9241A)
1M9240-97xxxKeysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
Startup Guide in hard copy
16Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 17
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
CAUTION
NOTE
Startup Guide
4Install the Modules
Module Installation Process Overview / 17
Front Panel Features / 19
Cleaning the Instrument / 24
PXI hardware does not support "hot-swap" (changing modules while power is applied
to the chassis) capabilities. Before installing or removing a module to/from the
chassis, power off the chassis to prevent damage to the module.
These modules can be used in a chassis with PXIe or PXI-H chassis slots.
Module Installation Process Overview
The module can be installed in any PXIe or hybrid PXI slot marked with a
peripheral slot compatibility image (solid black circle for PXIe, or solid black circle
with the letter -H- for hybrid).
1 Make sure that the line cord is plugged in to establish earth ground and that
the chassis power switch is Off.
2 If the chassis has multiple fan speed settings, ensure that the fan switch is set
to AUTO and the inhibit switch is set to DEF.
3 Position the chassis to provide ample space between the chassis fan intake and
exhaust vents. Blockage by walls or obstructions affects the air flow needed for
cooling. (Refer to the chassis documentation for more information about
cooling).
4 If using a remote PC controller, install the system interface card into the PC and
connect the chassis to the host PC.
17
Page 18
4Install the Modules
CAUTION
NOTE
5 Before inserting a module into the chassis, back the mounting screws out to
ensure that there is no interference between the screws and the mounting rails.
The M924xA PXIe oscilloscope can use peer-to-peer (module-to-module) triggering
across the PXI Express chassis backplane. Therefore, be sure to install the M924xA
modules into a chassis that supports peer-to-peer PXI Express I/O switch topology.
When using the Keysight M9018A PXIe 18-Slot Chassis, install the M924xA modules
into one of its three slot groups: 1 through 6, 7 through 12, or 13 through 18. For more
information on coordinating multiple M924xA PXIe oscilloscope modules, see
If you are using a non-Keysight chassis, consult your manufacturer's documentation.
When using the Keysight M9018A Chassis you will achieve the highest data throughput for the
Keysight M9241/42/43A PXIe oscilloscopes by using Slots 2, 6, 11 or 15. They are "x8" slots
and the others are "x4". For further details see the Keysight M9018A Chassis
documentation. If you are using a non-Keysight chassis, consult your manufacturer's
documentation.
6 Holding the module by the injector/ejector handle, slide it into an available PXI
(or hybrid) slot, as shown in the following figure.
18Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 19
Install the Modules4
a Install the module into the slot of the chassis by placing the module card
edges into the front module guides (top and bottom).
b Slide the module to the rear of the chassis and ensure that the
injector/ejector handle is pushed down in the unlatched (downward)
position.
c Slide the module completely into the chassis. When you begin to feel
resistance, pull up on the injector/ejector handle to fully inject the module
into the chassis.
7 Secure all modules to the chassis using the module mounting screws. Use a #1
Pozidriv® or a slotted screwdriver and torque to 5 Lb-In (0.57 N-m).
Performance may suffer if the screws are not tightened properly.
8 Verify that the PXI chassis fans are operable and free of dust and other
contaminants that may restrict airflow.
9 Install filler panels and slot blockers after installing the modules. Missing filler
panels or slot blockers may disrupt air circulation in the chassis. The left-most
slot does not accept a slot blocker.
10 If you are using a PCIe Cable Interface, such as the Keysight M9021, connect
the Cable Interface in the chassis to the PC host per the instructions that came
with the Cable Interface.
11 Power up the PXI chassis.
12 If you are using a remote PC controller, reboot the PC host controller.
13 Proceed to Chapter 6, “Verify Operation,” starting on page 29.
Front Panel Features
This section describes the front panel features, that is, connectors and LEDs, for
the M9241/42/43A PXIe oscilloscope and the M9240A AutoProbe power module.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide19
Page 20
4Install the Modules
AutoProbe
Support
Ext
Trig
Gen
Out
Aux
Out
Ref
I/O
1
2
M9243A
PXIe Oscilloscope
1GHz
Right
Module
Left
Module
M9240A
AutoProbe Power
WaveGen/Identify LED
(blue = WaveGen output enabled,
blinking blue = identify PXI module)
Run state LED
(green = runing,
red = stopped,
amber = waiting for
single acquisition
to complete)
Left/Right module
AutoProbe power LED
(green = power supplied,
red = power not supplied)
20Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
LED Operation
LEDs for the M9241/42/43A PXIe oscilloscope module and the M9240A AutoProbe
power module are described in the following table.
LEDColorWhen
Run state LEDGreenAcquisitions are running
WaveGen/Identify LEDBlueThe WaveGen output is enabled
AutoProbe power LED
(Left/Right module)
RedAcquisitions are stopped
AmberWaiting for a single acquisition to complete
OffNo driver is connected and the hardware is shut
Blinking blueThe "Identify PXI Module" feature is on
OffThe WaveGen output is disabled
RedEither of the following conditions is true:
GreenNeither of the red conditions is true
down
• There is a power fault.
• The associated module is shut down (that is,
no driver is connected).
Page 21
AutoProbe Interface
The M9240A AutoProbe power module comes with two faceplates, one for an
oscilloscope module to the left, and one for an oscilloscope module to the right.
The faceplates have connectors that go between the AutoProbe power module
and the oscilloscope's AutoProbe Support connector. Once the faceplate
connections have been established, tighten the faceplate's thumbscrew to secure
it.
Install the Modules4
When the M9240A AutoProbe power module faceplates are connected, the
oscilloscope input channels have AutoProbe interface connectors.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide21
Page 22
4Install the Modules
Ext
Trig
Gen
Out
Aux
Out
Ref
I/O
M9243A
Ext
Trig
Gen
Out
Aux
Out
Ref
I/O
M9243A
RightModule
LeftModule
M9240A
AutoProbe Power
1
2
AutoProbe
Adapter
Left
Module
1
2
AutoProbe
Adapter
Right
Module
AutoProbe interfaces,
right module
AutoProbe interfaces,
left module
CAUTION
CAUTION
Inputs and Outputs
The M9241/42/43A PXIe oscilloscope modules have the following inputs and
outputs.
Channel 1 and 2 inputs
Maximum input voltage at analog inputs
135 Vrms
Ω input: 5 Vrms Input protection is enabled in 50 Ω mode and the 50 Ω load will
50
disconnect if greater than 5 Vrms is detected. However the inputs could still be
damaged, depending on the time constant of the signal. The 50 Ω input protection
only functions when the oscilloscope is powered on.
When measuring voltages over 30 V, use a 10:1 probe.
22Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 23
Install the Modules4
CAUTION
WARNING
CAUTION
Ext Trig input
Maximum voltage at oscilloscope external trigger input
30 Vrms, 60 Vdc
1M ohm input: For steady-state sinusoidal waveforms derate at 20 dB/decade above
100 kHz to a minimum of 5 Vpk
For information on using the Ext Trig input, see the Keysight M9241/42/43A PXIe
Oscilloscopes Soft Front Panel (SFP) User's Guide.
The Probe Comp and chassis ground terminals are used when compensating
passive probes (see "Compensate Passive Probes" on page 39).
If you should inad vertently connect the M9241/42/43A PXIe oscilloscope's chassis
ground terminal to a Hazardous Live voltage and the Protective Earth Ground of the
Agency Certified PXI Chassis has been defeated or is non-existent, then the chassis
of the measuring equipment will also be connected to the Hazardous Live voltage
and you or a bystander could receive an electrical shock from the chassis.
Gen Out output
The Gen Out MMCX connector is for the waveform generator output. For
information on using the waveform generator, see the Keysight M9241/42/43A PXIe Oscilloscopes Soft Front Panel (SFP) User's Guide.
Aux Out output
The Aux Out MMCX connector can be used for trigger output, NFC trigger status,
mask testing status, waveform generator sync pulse, or calibration signal. For
more information, see the Keysight M9241/42/43A PXIe Oscilloscopes Soft Front Panel (SFP) User's Guide.
Ref I/O input/output
Maximum input voltage at Ref I/O connector
Do not apply more than 20 dBm Max (6.32 Vpp Max) at the Ref I/O MMCX connector
or damage to the instrument may occur.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide23
Page 24
4Install the Modules
When used for the 10 MHz reference output signal, the amplitude is 5 Vpp into a
high impedance or 2.5 Vpp into 50 Ohms. It is capable of driving into lower
impedances, but the output will be reduced because of the 50 Ohm source
impedance.
For information on using Ref I/O as an input or output, see the Keysight M9241/42/43A PXIe Oscilloscopes Soft Front Panel (SFP) User's Guide.
Cleaning the Instrument
1 Remove power from the instrument.
2 Clean the external surfaces of the oscilloscope with a soft cloth dampened with
a mixture of mild detergent and water.
3 Make sure that the instrument is completely dry before reconnecting it to a
power source.
24Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 25
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
Startup Guide
5Install the Software
System Requirements / 25
Hardware Requirements / 25
Chassis Requirements / 26
Power Up the Controller / 27
Software Installation Overview / 27
Updating Module Firmware as New Modules Are Installed / 28
System Requirements
ItemRequirements
Operating systemWindows 7 (32 bit and 64 bit) or later
Processor speed1.5 GHz 32-bit (x86), 1.5 GHz 64-bit (x64), 2.4 GHz recommended, no support for Itanium64
Available memory4 GB minimum (8 GB recommended for 64-bit operating systems)
Available disk space1.5 GB available hard disk space, includes:
• 1 GB available for Microsoft .NET Framework 3.5 SP1
• 100 MB for Keysight IO Libraries Suite
VideoSupport for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA
graphics is supported)
BrowserMicrosoft Internet Explorer 7.0 or greater
Hardware Requirements
ItemRequirements
ChassisPXIe or PXI-H chassis slot
ControllersA PXI or PXI Express embedded controller or remote controller (external PC connected to the
chassis by a PCI-to-PXI interface) is required.
25
Page 26
5Install the Software
ItemRequirements
Embedded controllerKeysight M9036A or M9307A or an embedded controller that meets the following
requirements:
• PXIe system controller (PXI-1 embedded controllers are not compatible).
• Utilize a 2x8, or 4x4 PXIe system slot link configuration.
• Run one of the operating systems listed in System Requirements (above).
Remote controller(for Keysight M9018A chassis use only) A PC running one of the operating systems listed in
System Requirements above and a Keysight M9021A Cable Interface x8 with one of the
following PC interface options:
• Keysight M9045B PCIe ExpressCard Adapter x1, with cable (for a laptop PC)
Recommended best practices to ensure proper and safe module operating
conditions:
•Ensure proper chassis air flow is maintained.
•Select a chassis that provides thermal protection if fans become inoperable or
forced air cooling is obstructed
•Use slot blockers (Keysight model Y1212A, 5 per kit) and EMC filler panels in
empty module slots to ensure proper operating temperatures. Keysight chassis
Keysight M9018A chassis and slot blockers optimize module temperature
performance and reliability of test.
•Set chassis fans to high or auto. Do not disable fans.
•Position chassis to allow plenty of space around chassis air intake and fan
exhaust.
•At environment temperatures above 45 °C, set chassis fan speed to high.
Figure 1M9018A Chassis Air Flow
The M9018A has multiple air intakes. They are located at the lower sides, lower
front, and bottom of the chassis.
26Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 27
Power Up the Controller
CAUTION
NOTE
CAUTION
If you are using a remote controller and you have installed the interface cable, you
must power up the chassis BEFORE you power up the PC. When you power down your
chassis, Shut Down the PC BEFORE you power down the chassis.
If you are using an embedded controller, complete the following steps:
1 Install the embedded controller module into the compatible chassis. The
Keysight M9036A PXIe Embedded Controller and Keysight M9018A PXIe
Chassis are recommended. Please refer to the embedded controller and chassis
documentation for further details.
2 Connect peripherals (mouse, keyboard, monitor).
3 Power up the chassis.
Software Installation Overview
Install the Software5
This installation includes the following:
•Keysight IO Libraries Suite (IOLS), which includes the Keysight Connection
Expert. This software is included with your shipment, and is also available at
www.keysight.com/find/iosuite. This software must be installed first.
Version 17.2 (or later) of the Keysight IO Libraries Suite is required.
•Instrument software, which includes the soft front panel (SFP), device drivers
(IVI-C, IVI-COM, and LabVIEW G) and documentation for the Keysight
M9241/42/43A PXIe oscilloscopes and M9240A AutoProbe power module. This
software is included with your shipment and is also available at
www.keysight.com/products/M9241A.
When installing the IO Libraries Suite and instrument software, either use the two CDs
shipped with the product OR the two download files from keysight.com. Installing a
mix of CD and keysight.com download files can result in incompatible software
versions.
Software Installation Procedure:
1 Install the Keysight IO Libraries Suite. Follow the installer prompts to install the
IO libraries.
2 Install the M924x InfiniiVision Oscilloscope software.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide27
Page 28
5Install the Software
NOTE
There are separate install packages for 64-bit and 32-bit Windows operating systems. Be sure
to choose the one that is appropriate for your chassis controller PC.
a Launch the software installer.
b Follow the installer prompts. Choose a "Complete" installation to install all
software and documentation, or a "Custom" installation to select from a
listing of modules and other features.
3 Complete the installation.
Updating Module Firmware as New Modules Are Installed
The software installation process includes FPGA updates to M924x InfiniiVision
oscilloscope modules that are installed in the chassis. Because you can install new
modules after the main software installation, there is a way to perform FPGA
updates separately:
•From the Windows Start menu, choose All Programs > Keysight InfiniiVision
2 In the Keysight InfiniiVision Launcher window, select the oscilloscope module
whose SFP you want to start; then, click the Show Front Panel button.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide31
Page 32
6Verify Operation
The Keysight InfiniiVision M924xA SFP will open:
The Keysight InfiniiVision M924xA SFP has the same user interface as other
standalone Keysight InfiniiVision oscilloscopes.
32Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 33
If not all modules and their slot locations are visible in the Keysight InfiniiVision
Launcher window:
1 Close the Keysight InfiniiVision Launcher.
2 Start the Keysight Connection Expert, by selecting Start > All Programs > Keysight
Connection Expert. If any or all modules and their slot locations are still not visible, in the Instruments tab, click Rescan.
3 Restart the Keysight InfiniiVision Launcher.
Run the Oscilloscope's Hardware Self Test
1 Keysight InfiniiVision M924xA SFP interface, click the top left blue menu icon,
and choose Utilities > Service Menu.
Verify Operation6
2 In the Service Menu, click Hard ware Sel f Test.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide33
Page 34
6Verify Operation
Hardware Self Test performs a series of internal procedures to verify that the
oscilloscope is operating properly. Successfully passing Hardware Self Test does
If the Hardware
Self Test fails
not guarantee 100% of the oscilloscope's functionality. Hardware Self Test is
designed to provide an 80% confidence level that the oscilloscope is operating
properly.
See: "Return an Instrument for Service" on page 12
If there are communications problems
If you are unable to communicate with the M924xA, verify that the following
installations are correct:
•Keysight IO Libraries Suite
•M924xA SFP programs
•Module and chassis drivers
•System Interface Card, cable and PC PXIe card connections, if you are using an
external host PC
34Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 35
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
CAUTION
Startup Guide
7Make a Measurement
Connect Probes to the Oscilloscope / 35
Input a Signal / 36
Using the Soft Front Panel (SFP) Interface / 36
Recall the Default Oscilloscope Setup / 37
Use Autoscale / 38
Compensate Passive Probes / 39
Turn On Measurements and Statistics / 40
Learn the Oscilloscope Display / 41
This chapter assumes you are using the Keysight InfiniiVision M924xA Soft Front
Panel (SFP). If you have not already opened the SFP, see "Open the M924xA
Oscilloscope Soft Front Panel (SFP)" on page 29.
Connect Probes to the Oscilloscope
1 Connect the oscilloscope probe to an oscilloscope channel BNC connector.
2 Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point
on the circuit.
Maximum input voltage at analog inputs
135 Vrms
50
Ω input: 5 Vrms Input protection is enabled in 50 Ω mode and the 50 Ω load will
disconnect if greater than 5 Vrms is detected. However the inputs could still be
damaged, depending on the time constant of the signal. The 50
only functions when the oscilloscope is powered on.
Ω input protection
35
Page 36
7Make a Measurement
Input a Signal
The first signal to input to the oscilloscope is the Probe Comp signal. This signal is
used for compensating probes.
1 Connect an oscilloscope probe from channel 1 to the Probe Comp terminal on
the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the Probe
Comp terminal).
Using the Soft Front Panel (SFP) Interface
The Keysight InfiniiVision M924xA SFP has the same user interface as other
standalone Keysight InfiniiVision oscilloscopes. While standalone oscilloscopes
have keys, softkeys, and knobs—obviously different than the modular
oscilloscopes—the display interface was also designed to be used with touch
screens and connected USB mice, and it is just as easy to use with a mouse on the
Windows operating system.
The main differences between using the interface on a standalone oscilloscope
and a modular oscilloscope are:
•Instead of pressing front panel keys to open softkey menus, you use the top-left
blue Main Menu button.
•Instead of pressing softkeys, you click the softkey label buttons.
•Instead of turning knobs to adjust scales, offets, delay positions, trigger levels,
and cursor positions, there are clickable up/down, increase/decrease icon
buttons that appear after clicking these values. Also, you can drag waveforms,
trigger levels, and cursors.
Access the Built-In Quick Help
Soft Front Panel (SFP) help for the PXIe oscilloscope is different than in a
traditional Windows application. Instead of a Windows HTML Help file you access
through a Help menu, the PXIe oscilloscope's SFP help is the same quick help
information that is built-in to standalone InfiniiVision oscilloscopes.
To access the built-in quick help, press and hold the left-mouse button over a
softkey label button.
36Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 37
Make a Measurement7
Quick Help remains on the screen until you close the help dialog box or open
another one.
Select the User Interface Language
Another benefit of having the same user interface as standalone InfiniiVision
oscilloscopes is that you have a localized user interface and built-in quick help.
To select the user interface language:
1 Choose Main Menu > Help Menu.
2 In the Help Menu, click the Language softkey label button.
3 In the Language popup menu, select the desired language.
The following languages are available: Czech, English, French, German, Italian,
Japanese, Korean, Polish, Portuguese, Russian, Simplified Chinese, Spanish, Thai,
Traditional Chinese, and Turkish.
Recall the Default Oscilloscope Setup
To recall the default oscilloscope setup:
1 Choose Main Menu > Default Setup.
The default setup restores the oscilloscope's default settings. This places the
oscilloscope in a known operating condition. The major default settings are:
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide37
Page 38
7Make a Measurement
Table 1Default Configuration Settings
In the Default Menu, there are also options for restoring the complete factory
settings.
Use Autoscale
HorizontalNormal mode, 100 µs/div scale, 0 s delay, center time reference.
Vertical (Analog)Channel 1 on, 5 V/div scale, DC coupling, 0 V position, 1 MΩ impedance.
TriggerEdge trigger, Auto trigger mode, 0 V level, channel 1 source, DC coupling, rising
OtherAcquire mode normal, Run/Stop to Run, cursors and measurements off.
LabelsAll custom labels that you have created in the Label Library are preserved (not
erased), but all channel labels will be set to their original names.
Use Autoscale to automatically configure the oscilloscope to best display the input
signals.
1 Choose Main Menu > Autoscale.
You should see a waveform on the oscilloscope's display similar to this:
38Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 39
Make a Measurement7
2 If you want to return to the oscilloscope settings that existed before, click Undo
Autoscale.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, click Fast Debug,
Channels, or Acq Mode.
These are the same softkeys that appear under Main Menu > Utilities > User
Options Menu > Preferences > Autoscale.
If you see the waveform, but the square wave is not shaped correctly as shown
above, perform the procedure "Compensate Passive Probes" on page 39.
If you do not see the waveform, make sure the probe is connected securely to the
front panel channel input BNC and to the left side Probe Comp terminal.
How Autoscale
Works
Autoscale analyzes any waveforms present at each channel and at the external
trigger input.
Autoscale finds, turns on, and scales any channel with a repetitive waveform that
has a frequency of at least 25 Hz, a duty cycle greater than 0.5%, and an
amplitude of at least 10 mV peak-to-peak. Any channels where no signal is found
are turned off.
The trigger source is selected by looking for the first valid waveform starting with
external trigger, then continuing with the lowest number analog channel up to the
highest number analog channel.
During Autoscale, the delay is set to 0.0 seconds, the horizontal time/div (sweep
speed) setting is a function of the input signal (about 2 periods of the triggered
signal on the screen), and the triggering mode is set to Edge.
Compensate Passive Probes
Each oscilloscope passive probe must be compensated to match the input
characteristics of the oscilloscope channel to which it is connected. A poorly
compensated probe can introduce significant measurement errors.
1 Input the Probe Comp signal (see "Input a Signal" on page 36).
2 Choose Main Menu > Default Setup to recall the default oscilloscope setup (see
"Recall the Default Oscilloscope Setup" on page 37).
3 Choose Main Menu > Autoscale to automatically configure the oscilloscope for
the Probe Comp signal (see "Use Autoscale" on page 38).
4 Click the channel button to which the probe is connected (color-coded 1, 2, etc.
at the top of the SFP window).
5 In the Channel Menu, click Probe.
6 In the Channel Probe Menu, click Probe Check; then, follow the instructions
on-screen.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide39
Page 40
7Make a Measurement
Perfectly compensated
Over compensated
Under compensated
NOTE
7 Connect probes to all other oscilloscope channels (that is, channel 2 of a
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On N2894A probes, the trimmer capacitor is located on the probe BNC
connector.
2-channel oscilloscope).
8 Repeat the procedure for each channel.
After a Default Setup, the Probe Comp signal is output to the Probe Comp terminal. However,
the oscilloscope can also output training signals on this terminal. See: Main Menu > Training Signals > Training Signals. On InfiniiVision M924xA oscilloscope modules, the Probe Comp
terminal is also the Demo 1 terminal. There is no Demo 2 terminal (unlike some other
InfiniiVision X-Series oscilloscopes).
Turn On Measurements and Statistics
1 To turn on measurements, choose Main Menu > Measure > Measurements.
2 To turn on measurement statistics, choose Main Menu > Measure > Statistics.
You should see a display similar to this:
40Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 41
Make a Measurement7
Learn the Oscilloscope Display
The oscilloscope display contains acquired waveforms, setup information,
measurement results, and the softkey definitions.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide41
Page 42
7Make a Measurement
Analog channel
sensitivity
Status line
Analog
channels
and ground
levels
Trigger level
Softkey label
buttons
Menu line
Trigger point,
time reference
Delay
time
Time/
div
Run/Stop
status
Trigger
type
Trigger
source
Measurements
Trigger level
Sidebar
information and
controls area
Cursors defining
measurement
Measurement
statistics
Menu
button
Status lineThe top line of the d isplay contains vertical, horizontal, and trigger setup
information.
Display areaThe display area contains the waveform acquisitions, channel identifiers, and
Sidebar
information and
controls area
analog trigger, and ground level indicators. Each analog channel's information
appears in a different color.
Signal detail is displayed using 256 levels of intensity. To adjust waveform
intensity, choose Main Menu > Setup > Waveform Intensity.
To adjust other display modes and settings, choose Main Menu > Setup > Display Menu.
The sidebar information area can contain summary, cursors, measurements, or
digital voltmeter information dialogs or it can contain navigation and other
control dialogs.
To select the type of information or controls you want to see in the sidebar, click
the blue menu icon next to the sidebar dialog box title.
To undock or redock sidebar dialog boxes, drag the dialog box titles out of or
back into the sidebar information area.
42Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 43
Make a Measurement7
Menu lineThis line normally contains menu name or other information associated with the
selected menu.
Softkey labelsThese labels describe softkey functions. Typically, softkeys let you set up
additional parameters for the selected mode or menu.
At the top of the menu hierarchy, clicking the left-side up-arrow button turns off
softkey labels and displays additional status information describing channel
offset and other configuration parameters.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide43
Page 44
7Make a Measurement
44Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 45
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
Startup Guide
8Coordinating Multiple PXIe
Oscilloscope Modules
Step 1: (Optional) Update the Calibration Factors / 46
Step 2: Set Up the Slave Modules / 49
Step 3: Set Up the Master Module / 50
Step 4: (When Necessary) Connect Chassis Bus Segments / 51
Step 5: (Recommended) Phase-Lock Oscilloscope Sampling Clocks / 54
Step 6: (Optional) Turn On Sync Mode / 55
Step 7: Make Coordinated Acquisitions / 60
You can synchronize acquisitions between multiple M924xA oscilloscope modules
by using the PXI Trigger Menu to set up coordinated arming and triggering.
Among all the oscilloscope modules to be synchronized, one module must be the
master and all other modules must be slaves. The PXI Trigger Mode softkey makes
the master or slave selection for a module (or turns off PXI triggering for the
module, making it a standalone module).
The Trigger Menu on the master oscilloscope module is used to define the trigger
condition. The master module outputs its trigger signal to one of the chassis
PXI_TRIG lines (0-7). All slave modules receive the trigger signal over the same
PXI_TRIG line, and their Trigger Menus simply show that they are PXI slaves.
There is a separate "armed" signal from each slave module so that the master can
wait until each slave is ready before triggering. The remaining PXI_TRIG lines can
be used for the arm lines from the slave modules. With arming control, there can
be up to seven slave modules. (With software arming in remote programs, there
can be more than eight modules in a multiple-module system.)
Arming control is not required; however, if it is not used, a slave module may miss
trigger events from the master module (because the master module may send
triggers when the slave module is not armed).
This chapter assumes you are using the Keysight InfiniiVision M924xA Soft Front
Panel (SFP). If you have not already opened the SFP, see "Open the M924xA
Oscilloscope Soft Front Panel (SFP)" on page 29.
45
Page 46
8Coordinating Multiple PXIe Oscilloscope Modules
NOTE
NOTE
Step 1: (Optional) Update the Calibration Factors
This step is optional because there are default calibration factors installed with the
M924x InfiniiVision Oscilloscope software. However, to get the most accurate
calibration factors for your PXIe oscilloscope modules and chassis, you will want to
update the calibration factors.
Calibration factors are updated and saved using the KtScopeCal.exe application
on the chassis controller.
The KtScopeCal.exe application is installed with the M924x InfiniiVision
Oscilloscope software.
To perform trigger routing, the KtScopeCal.exe application requires KtMTrig, which is included
in PXIe chassis driver version 1.6.242.1 and above. You can find PXIe chassis driver software
on the CD that ships with the chassis or at: www.keysight.com/find/pxie-chassis-driver
For the best calibration accuracy, the chassis should be powered on and the oscilloscope
modules enabled with a default setup for at least 30 minutes to stabilize the operating
temperature before running KtScopeCal.exe.
The KtScopeCal.exe application is run in a Command Prompt window. Command
usage information is displayed when the program is run without any arguments or
when the "/?" option is supplied as an argument. For example:
C:\Program Files (x86)\Keysight\InfiniiVision>KtScopeCal.exe /?
This app performs PXI trigger delay calibrations between master and slav
e
modules.Hardware connection is required from the Aux out of master mod
ule
to channel 1 of both master and slave modules using BNC tee adapter with
50ohm cables of equal length.
[a]Chassis bus number.
[b]Master module slot number.
[c]Slave module slot number.
[d]Trigger line number (0-7).
<bus> Optional to configure trigger routing across trigger bus segment
s.
Keysight PXIe chassis will be configured automatically.
[e]Bus segment number where master module is located.
[f]Bus segment number where slave module is located.
Trigger bus segments are assigned as follows:
18-Slot Chassis
Slots 1-6:Trig bus segment number 1
Slots 7-12:Trig bus segment number 2
46Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 47
Coordinating Multiple PXIe Oscilloscope Modules8
Slots 13-18:Trig bus segment number 3
10-Slot Chassis
Slots 1-5:Trig bus segment number 1
Slots 6-10:Trig bus segment number 2
Sync Mode:
KtScopeCal -a
Align the reference clock signals for all modules and save factors.
Perform this action once for new or modified system.
KtScopeCal -p
Align the reference clock signals for all modules.
Perform this action each time InfiniiVision driver is connected.
KtScopeCal -r
Reset the reference clock signals for all modules.
Perform this action only when PXIe 100MHz reference clock is turned off.
This action is not required if -a or -p has been performed.
KtScopeCal -s PXI[a]::[b]::INSTR slot[c] trig[d] <bus[e]-[f]>
Calibrate trigger delay and save factors.
Perform this action after reference clock signals are aligned.
C:\Program Files (x86)\Keysight\InfiniiVision>
The chassis bus number can be found in the Keysight Connection Expert window:
1 Click the Keysight IO Libraries Suite icon in the task bar and choose Connection
Expert from the popup menu.
2 In the Keysight Connection Expert window, PXI/AXIe Chassis > Chassis Content
tab, find the chassis bus number in the Chassis Address string:
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide47
Page 48
8Coordinating Multiple PXIe Oscilloscope Modules
Chassis Bus Number
For each master module, slave module, and trigger line combination that will be
used:
1 Connect a MMCX-to-BNC cable from the master module Aux Out output to a
BNC tee adapter.
2 From the tee adapter, connect equal length cables to the master module
channel 1 input and the slave module channel 1 input.
The calibration information is saved on the chassis controller, so it needs to be run
just once for each combination. However, if you change anything about the
multiple module setup, you should re-run the program.
Because calibration factors are saved per master/slave/trigger line combination, you could
run the KtScopeCal.exe program for every possible combination of master and slave modules
that would ever be used, not just the combination you are planning to use next.
Step 2: Set Up the Slave Modules
Coordinating Multiple PXIe Oscilloscope Modules8
For each slave module in the multiple-oscilloscope system:
1 Choose Main Menu > Trigger > PXI Trigger Menu.
2 In the PXI Trigger Menu, click the PXI Trigger Mode softkey button, and select
Slave.
3 Click the Trigger Line softkey, and select the PXI_TRIG line (0-7) to be used for
the multi-module trigger signal.
The master module and all slave modules must select the same trigger line.
If the modules are in different chassis trigger bus segments (which are isolated
by default), you must use the chassis Soft Front Panel (SFP) application to
connect the selected PXI_TRIG line between trigger bus segments. See "Step 4:
(When Necessary) Connect Chassis Bus Segments" on page 51.
4 Click the Arm Line softkey to select the PXI_TRIG line (0-7) to be used for the
slave module's arm signal.
Each slave module must use a different PXI_TRIG line for its arm signal.
If the modules are in different chassis trigger bus segments (which are isolated
by default), you must use the chassis Soft Front Panel (SFP) application to
connect the selected PXI_TRIG line between trigger bus segments. See "Step 4:
(When Necessary) Connect Chassis Bus Segments" on page 51.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide49
Page 50
8Coordinating Multiple PXIe Oscilloscope Modules
5 Click the Master Slot # softkey; then, enter the slot number of the master
M924xA oscilloscope module.
It is important to select the correct master slot number because it identifies the
proper set of calibration factors to use when aligning waveforms.
Calibration factors are created and saved using the KtScopeCal.exe application
on the chassis controller. See "Step 1: (Optional) Update the Calibration
Factors" on page 46.
As a slave module in a multiple M924xA oscilloscope module system, this module
gets it trigger signal from the master module. As a result, the Trigger Type softkey in
the Trigger Menu simply shows that it is a PXI Slave.
Step 3: Set Up the Master Module
For the master module in the multiple-oscilloscope system:
1 Choose Main Menu > Trigger > PXI Trigger Menu.
2 In the PXI Trigger Menu, click the PXI Trigger Mode softkey button, and select
Master.
3 Click the Trigger Line softkey, and select the PXI_TRIG line (0-7) to be used for
the multi-module trigger signal.
The master module and all slave modules must select the same trigger line.
If the modules are in different chassis trigger bus segments (which are isolated
by default), you must use the chassis Soft Front Panel (SFP) application to
connect the selected PXI_TRIG line between trigger bus segments. See "Step 4:
(When Necessary) Connect Chassis Bus Segments" on page 51.
4 Click the Arm Lines softkey to select the PXI_TRIG lines (0-7) used for the arm
signals of the slave modules in the multiple-module system. Select the line and
then click it again to enable or disable it.
Each slave module must use a different PXI_TRIG line for its arm signal.
The master module will wait until all slave modules are armed before triggering.
If the modules are in different chassis trigger bus segments (which are isolated
by default), you must use the chassis Soft Front Panel (SFP) application to
connect the selected PXI_TRIG lines between trigger bus segments. See "Step
4: (When Necessary) Connect Chassis Bus Segments" on page 51.
50Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 51
Coordinating Multiple PXIe Oscilloscope Modules8
The Trigger Menu on the master oscilloscope module is used to define the trigger
condition. You can select from all the usual trigger types (Edge, Edge Then Edge,
Pulse Width, etc.). You are limited to master module input sources.
Step 4: (When Necessary) Connect Chassis Bus Segments
If the PXI oscilloscope modules are in different chassis trigger bus segments
(which are isolated by default), you must use the chassis Soft Front Panel (SFP)
application to connect the selected PXI_TRIG lines between trigger bus segments.
For example, if:
•The master module is in segment 1 and it is using PXI TRIG 0 as the trigger line.
•Two slave modules are in segment 3 and they will use PXI TRIG 1 and
PXI TRIG 2 as arm lines.
You would fllow these steps:
1 Click the Keysight IO Libraries Suite icon in the task bar and choose Connection
Expert from the popup menu.
2 In the Keysight Connection Expert window, PXI/AXIe Chassis > Chassis Triggers
tab, hover over the lines in the bus segments you would like to reserve, and
then click Click to reserve.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide51
Page 52
8Coordinating Multiple PXIe Oscilloscope Modules
When lines in bus segments are reserved, it will look something like:
52Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 53
Coordinating Multiple PXIe Oscilloscope Modules8
3 Next, click the route directions to be used:
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide53
Page 54
8Coordinating Multiple PXIe Oscilloscope Modules
CAUTION
4 Finally, click OK.
The Keysight M9018A PXIe Chassis does not remember the trigger bus configuration
upon cycling power. After a power cycle, you must reconfigure the bus triggers by
using the M9018A Soft Front Panel (SFP) or by creating a persistent configuration by
using the AgM9018 API.
It is also recommended that each module in the system phase-lock its timebase to
the PXIe 100 MHz reference.
For each module in the multiple-oscilloscope system:
1 Choose Main Menu > Utilities > User Options Menu.
2 In the User Options Menu, click the Auxiliary I/O softkey button.
54Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 55
3 In the Auxiliary I/O Menu, click the 10MHz Ref Signal softkey button, and select
PXIe 100MHz.
Step 6: (Optional) Turn On Sync Mode
Optionally, to lessen the jitter between channels in different modules so that it is
more like the jitter between channels within a single module, you can turn on Sync
mode. Sync mode uses the trigger system in the master module for all modules.
When Sync mode is off, this is called Asynchronous mode. Asynchronous mode
uses the trigger systems in all modules where the trigger output from the master
module is fed to the trigger input of the slave modules.
To turn on and set up Sync mode:
1 For each M924xA oscilloscope module in the multi-module system (that is, the
master module and the slave modules), click the Sync softkey button to enable
Sync mode.
Coordinating Multiple PXIe Oscilloscope Modules8
When you turn on Sync mode for the first time (after InfiniiVision software
start-up), there is a red unlocked image shown in the top right of the screen.
For example:
•On the master module, you see:
•On the slave modules, you see:
This red unlocked image indicates that the clock phase for the Sync mode
trigger is not aligned.
2 Open a Command Prompt window on the chassis controller.
3 To align the reference clock signals, you must run the KtScopeCal.exe
application using the "-a" or "-p" option:
•If setting up Sync mode triggers on a new or modified system, use the
KtScopeCal -a command. For example:
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide55
Page 56
8Coordinating Multiple PXIe Oscilloscope Modules
C:\Program Files (x86)\Keysight\InfiniiVision>KtScopeCal.exe -a
Searching for connected modules...
Measuring PLL Band for module at slot 6...
Band = 3
PLL Band measurement completed.
Measuring PLL Band for module at slot 5...
Band = 2
PLL Band measurement completed.
Calibration signal reset completed.
Performing clock phase alignment for module at slot 5...
Median = 2883
Iteration = 1
Mean = 1
Iteration = 2
Mean = 2439
Iteration = 3
Mean = 3081
Temperature = 52.25
Clock phase alignment completed.
Performing clock phase alignment for module at slot 6...
Median = 2927
Mean = 3117
Temperature = 52.00
Clock phase alignment completed.
C:\Program Files (x86)\Keysight\InfiniiVision>
•If the system is not new or modified but you have reconnected the
Searching for connected modules...
Calibration signal reset completed.
Performing clock phase reset for module at slot 5...
Temperature change = 0.25
XML mean = 3081
Target = 3080
Iteration = 1
Mean = 2916
Mean change = -165
Iteration = 2
Mean = 2439
Mean change = -642
Iteration = 3
Mean = 4031
Mean change = 950
Iteration = 4
Mean = 4031
Mean change = 950
Iteration = 5
Mean = 3717
Mean change = 636
Iteration = 6
Mean = 2595
Mean change = -486
KtScopeCal -p command. For example:
56Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 57
Coordinating Multiple PXIe Oscilloscope Modules8
Iteration = 7
Mean = 4031
Mean change = 950
Iteration = 8
Mean = 4031
Mean change = 950
Iteration = 9
Mean = 3551
Mean change = 470
Iteration = 10
Mean = 2271
Mean change = -810
Iteration = 11
Mean = 3717
Mean change = 636
Iteration = 12
Mean = 3717
Mean change = 636
Iteration = 13
Mean = 2270
Mean change = -811
Iteration = 14
Mean = 3873
Mean change = 792
Iteration = 15
Mean = 3078
Mean change = -3
Clock phase reset completed.
Performing clock phase reset for module at slot 6...
Temperature change = 0.25
XML mean = 3117
Target = 3116
Iteration = 1
Mean = 3110
Mean change = -7
Clock phase reset completed.
C:\Program Files (x86)\Keysight\InfiniiVision>
When running the KtScopeCal.exe application with the Sync mode options
"-a", "-p", or "-r", the oscilloscope modules automatically phase-lock their
timebases to the chassis' PXIe 100 MHz reference. Using the PXIe 100 MHz
reference signal is a requirement for Sync mode triggering (see "Step 5:
(Recommended) Phase-Lock Oscilloscope Sampling Clocks" on page 54).
If you turn off the PXIe 100MHz ref signal on any module, clock phase
alignment between modules is lost, and the red unlocked image will re-appear.
•When clock phase alignment is lost, use the
KtScopeCal -r command to
reset/realign the clock phase. For example:
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide57
Searching for connected modules...
Clock phase reset for module at slot 5...
Temperature change = 0.00
XML mean = 3081
Target = 3081
Iteration = 1
Mean = 3400
Mean change = 319
Iteration = 2
Mean = 2273
Mean change = -808
Iteration = 3
Mean = 3080
Mean change = -1
Clock phase reset completed.
Clock phase reset for module at slot 6...
Temperature change = 0.00
XML mean = 3117
Target = 3117
Iteration = 1
Mean = 4070
Mean change = 953
Iteration = 2
Mean = 3273
Mean change = 156
Iteration = 3
Mean = 2305
Mean change = -812
Iteration = 4
Mean = 1986
Mean change = -1131
Iteration = 5
Mean = 3109
Mean change = -8
Clock phase reset completed.
C:\Program Files (x86)\Keysight\InfiniiVision>
The red unlocked image will disappear.
4 Finally, to perform Sync mode trigger delay calibration on each trigger line and
save the calibration factors, run the
are the same as when using the
Trigger line PXI TRIG 0 bus segment configured: Isolated
Arm line PXI TRIG 1 bus segment configured: Isolated
Connect Aux Out from master module at slot 5 to channel 1 of master m
odule
and channel 1 of slave module at slot 6 using BNC tee adapter with
50ohm cables of equal length.
58Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 59
Coordinating Multiple PXIe Oscilloscope Modules8
Press enter to start calibration...
Mean change = -2926
Mean change = -2926
Mean change = 1192
Mean change = 1167
Mean change = 1134
Mean change = 1108
Mean change = 1073
Mean change = 1051
Mean change = 1017
Mean change = 992
Mean change = 957
Mean change = 933
Mean change = 899
Mean change = 872
Mean change = 837
Mean change = 810
Mean change = 774
Mean change = 747
Mean change = 711
Mean change = 685
Mean change = 652
Mean change = 631
Mean change = 600
Mean change = 576
Mean change = 542
Mean change = 517
Mean change = 481
Mean change = 457
Mean change = 422
Mean change = 402
Mean change = 367
Mean change = 341
Mean change = 305
Mean change = 281
Mean change = 247
Mean change = 224
Mean change = 185
Mean change = 163
Mean change = 125
Mean change = 99
Iteration = 40
OutSel = 0
Tdelay = 39
Sync mode calibration passed. Master slot 5, Slave slot 6, Trigger li
ne 0.
C:\Program Files (x86)\Keysight\InfiniiVision>
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide59
Page 60
8Coordinating Multiple PXIe Oscilloscope Modules
CAUTION
CAUTION
Step 7: Make Coordinated Acquisitions
After calibration factors are updated, trigger and arm lines are properly configured,
the master slot is identified in all slaves, and the PXIe 100 MHz reference signal is
selected, you are ready to use the multiple-oscilloscope system to make
coordinated acquisitions.
If arming control is used, follow this sequence to keep acquisitions correlated:
1 Stop all M924xA oscilloscope modules.
2 Start acquisitions (single or run) on the slave module (it will wait for triggers
from the master).
3 Start acquisitions (single or run) on the master module.
4 When running, stop acquisitions on the master module first. Then, stop
acquisitions on the slave modules.
•If running, and you want to change settings on the master module (such as
time/div, volts/div, or the trigger setting), stop the master module before
changing its settings. The slave modules do not need to be stopped.
Single Acquisitions
for Deep Data
Messages When
Stopping
Acquisitions
•If running, and you want to change settings on slave modules (such as
time/div or volts/div), stop the master module first. Then, stop the slave
modules and make their setting changes.
When you are using remote programs to control multi-module systems and data
synchronization is important, not only must you stop acquisitions in the proper order,
but you must also query a module's run status to make sure that acquisitions have
completed and are truly stopped.
For the deepest acquisitions possible, use single acquisitions. When you stop
running acquisitions in a multi-module configuration, the data length can be
smaller (depending on the setup) than when stopping a running acquisition in a
single module.
When stopping acquisitions, particularly acquisitions that take a long time, you
may see this message:
Please wait...
To abort, press Stop.
This message means a "soft stop" is in progress. When stopping multi-module
acquisitions, "soft stop" makes sure that data across the multiple modules are
from the same trigger.
If you abort a "soft stop", be aware that the module's data may not be from the same
trigger as other modules' data.
60Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 61
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
Startup Guide
9Applications Programming
Interface (API) Overview
IVI Drivers / 61
LabVIEW Driver / 62
SCPI Interface (M9241/42/43A PXIe Oscilloscope Only) / 62
Using the SFP and the Remote API at the Same Time / 63
When you have completed installation, you can use the oscilloscope Soft Front
Panel (SFP) or program the instrument using the applications programming
interface (API) for the supplied drivers.
IVI Drivers
Supported ADEs
(application
development
environments)
Keysight's IVI drivers simplify the creation and maintenance of instrument control
applications in a variety of development environments; they allow programmatic
control of instrumentation while providing a greater degree of instrument
interchangeability and code reuse. IVI drivers currently come in two basic types:
IVI-COM and IVI-C. Although the functionality offered by both types of drivers is
often very similar, the fundamental differences in interface technology results in a
very different end-user experience. The IVI drivers support compiling application
programs for 32- or 64-bit platforms.
Arguably the most important consideration in comparing IVI-COM and IVI-C
drivers is the end user experience in various ADEs. Because IVI-COM drivers are
based on Microsoft COM technology, it is not surprising that IVI-COM drivers offer
the richest user experience in Microsoft ADEs. Users working in Visual C++, Visual
C#, Visual Basic.NET, and Visual Basic 6 enjoy a host of features, such as object
browsers, IntelliSense, and context-sensitive help.
When you install the product software, the IVI driver files are installed in the
standard IVI Foundation directories (for example, C:\Program Files\IVI Foundation\
IVI\Drivers\).
Example programs are provided to demonstrate most driver functionality (for
example, C:\Program Files\IVI Foundation\IVI\Drivers\AgInfiniiVision\Examples).
The reference material for the driver functions (a Microsoft HTML Help .chm file) is
installed with the IVI driver and is available for Microsoft Visual Studio's
IntelliSense context linking. In addition, you can directly access the .chm file
(AgInfiniiVision.chm) from this Start menu location: Start > All Programs > Keysight Instrument Drivers > IVI-COM-C AgInfiniiVision Oscilloscope.
LabVIEW Driver
In addition to the IVI drivers, Keysight provides a LabVIEW driver that includes all
the functionality of the IVI-C driver.
When you install the product software, the LabVIEW driver is installed to each
LabVIEW instr.lib directory for each version of LabVIEW you have on your
computer (for example, C:\Program Files (x86) \National Instruments\<LabVIEW
version>\instr.lib\<Keysight product model>).
If you install LabVIEW drivers before you install LabVIEW itself, the drivers will be
installed in the Keysight directory instead of the National Instruments directory
(for example, C:\Program Files (x86)\Keysight\<Keysight product model>\
LabVIEW Driver\<LabVIEW version>\...).
Example programs are provided to demonstrate most driver functionality.
The reference information for the driver (a Microsoft HTML Help file named
KtInfiniiVision_LabVIEW_Help.chm) is also installed with the driver, and the
content is available from LabVIEW's context Help menu or in the LabVIEW help
directory.
In general, you can use the Soft Front Panel (SFP) and the remote applications
programming interface (API) together, at the same time. This can be useful, for
example, because it lets you change oscilloscope settings while debugging remote
programs.
However, the SFP can become "locked out" while the oscilloscope waits for a
waveform to become fully acquired. This can happen, for example, after these
SCPI commands/queries: :WAVeform:DATA?, :WAVeform:PREamble?,
:LISTer:DATA?, :SEARch:COUNt?, :DIGitize, or :SINGle. In these situations, you can
clear the remote operation in progress by sending a Device Clear to the remote
user interface.
Sending a Device Clear to the Remote Interface
Every remote programming library gives you a way to send an IEEE 488.1-style
Device Clear. For example, the IVI-COM library provides the System2.ClearIO
method. Another easy way to send a Device Clear is through the IO Libraries
Suite's Interactive IO application.
1 On a networked PC that has the IO Libraries Suite installed, click the Keysight
IO Control icon in the taskbar and choose Utilities > Interactive IO from the popup
menu.
2 In the Keysight Interactive IO application, choose Connect > Connect....
3 In the Connect dialog box, enter the oscilloscope's HiSLIP address into the
Resource Name field and click OK.
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide63
The oscilloscope's HiSLIP address can be cut-and-pasted from your remote
program or you can piece it together from the controller PC's IP address (or
hostname) and the oscilloscope's PXI address. For example if the controller
PC's IP address is "10.112.93.141" (or hostame is
"lab-pxi-3.cos.is.keysight.com") and the oscilloscope's PXI address is
"PXI37::0::0::INSTR", the oscilloscope's HiSLIP address is
"TCPIP0::10.112.93.141::hislip37-0.0::INSTR" (or
"TCPIP0::lab-pxi-3.cos.is.keysight.com::hislip37-0.0::INSTR").
If you are running the Interactive IO application on the controller PC, you can
use the "127.0.0.1" localhost IP address, for example,
"TCPIP0::127.0.0.1::hislip37-0.0::INSTR".
4 When connected, click Device Clear.
64Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
The Keysight M9241/42/43A PXIe oscilloscopes and M9240A AutoProbe power
module products have been safety-evaluated to the M9018A PXIe 18-slot chassis
environmental characteristics.
EnvironmentIndoor use only.
Ambient
temperature
Humidity95% rH, non-condensing to temperatures up to 40 °C decreasing linearly to
AltitudeOperating: 3,000 m (9,842 ft)
Pollution DegreeThe InfiniiVision M9241/42/43A PXIe oscilloscopes may be operated in
Pollution Degree
Definitions
Operating: 0 °C to +55 °C
Non-operating: –40°C to +70°C
50% rH at 55 °C
Non-operating: 4,500 m (14,764 ft)
environments of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.
67
Page 68
10Reference
Probes and Accessories
For a list of the probes and accessories that are compatible with the
M9241/42/43A PXIe oscilloscopes, see the data sheet at:
www.keysight.com/products/M9241A
AutoProbe
Interface
See AlsoFor more information on probes and accessories, see www.keysight.com for:
The AutoProbe interface is enabled with the M9240A AutoProbe power module.
Most Keysight single-ended active, differential, and current probes are compatible
with the AutoProbe interface. Active probes that do not have their own external
power supply require substantial power from the AutoProbe interface.
If too much current is drawn from the AutoProbe interface, an error message will
be displayed, indicating that you must momentarily disconnect all probes to reset
the AutoProbe interface, then connect only the supported quantity of active
probes.
•Probes and Accessories Selection Guide (5989-6162EN)
•InfiniiVision Oscilloscope Probes and Accessories Selection Guide Data Sheet
(5968-8153EN)
•For compatibility information, manuals, application notes, data sheets,
selection guides, SPICE models, and more for oscilloscope probes, see the
Probe Resource Center at: www.keysight.com/find/PRC
Specifications and Characteristics
Please see the InfiniiVision oscilloscope data sheets for complete, up-to-date
specifications and characteristics. To download the data sheet for the
M9241/42/43A PXIe oscilloscopes and M9240A AutoProbe power module, please
visit: www.keysight.com/products/M9241A
Measurement Category
•"Oscilloscope Measurement Category" on page 68
•"Measurement Category Definitions" on page 69
Oscilloscope Measurement Category
The InfiniiVision oscilloscopes are not intended to be used for measurements in
Measurement Category II, III, or IV.
68Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 69
WARNING
Use this instrument only for measurements within its specified measurement
category (not rated for CAT II, III, IV). No transient overvoltages allowed.
Measurement Category Definitions
The "Not rated for CAT II, III, IV" measurement category is for measurements
performed on circuits not directly connected to MAINS. Examples are
measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS derived circuits. In the latter case, transient stresses are variable;
for that reason, the transient withstand capability of the equipment is made known
to the user.
Measurement category II is for measurements performed on circuits directly
connected to the low voltage installation. Examples are measurements on
household appliances, portable tools and similar equipment.
Measurement category III is for measurements performed in the building
installation. Examples are measurements on distribution boards, circuit-breakers,
wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the
fixed installation, and equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to the fixed installation.
Reference10
Measurement category IV is for measurements performed at the source of the
low-voltage installation. Examples are electricity meters and measurements on
primary overcurrent protection devices and ripple control units.
Product Markings
These symbols are used on the M9241/42/43A PXIe oscilloscopes.
SymbolDescription
Caution, refer to accompanying documentation
Caution, hot surface
Avoid static discharge
This symbol indicates separate collection for electrical and electronic
equipment mandated under EU law as of August 13, 2005. All electric
and electronic equipment are required to be separated from normal waste
for disposal (Reference WEEE Directive 2002/96/EC).
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide69
Page 70
10Reference
.4*13&.
,TU"
SymbolDescription
South Korean Certification (KC) mark; includes the marking's identifier
code which follows this format:
MSIP-REM-YYY-ZZZZZZZZZZZZZZ.
The RCM mark is a registered trademark of the Australian
Communications and Media Authority.
Indicates the time period during which no hazardous or toxic substance
elements are expected to leak or deteriorate during normal use. Forty
years is the expected useful life of the product.
The CE mark is a registered trademark of the European Community.
ICES / NMB-001 Cet appareil ISM est conforme a la norme NMB du
Canada. This is a marking to indicate product compliance with the
Industry Canadian Interference-Causing Equipment Standard
(ICES-001).
This is also a symbol of an Industrial Scientific and Medical Group 1 Class
A product (CISPR 11, Clause 4).
The CSA mark is a registered trademark of the CSA International.
70Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
Page 71
Index
Numerics
10 MHz Ref I/O, 19
A
accessories, 68
altitude, 67
API overview, 61
ARINC 429 serial triggering and
analysis, 16
asynchronous mode, 55
AultoProbe interface, 21
AutoProbe power LED, 20
AutoProbe support interface, 19
Autoscale, undo, 39
Aux Out, 19
Aux Out output, 23
B
built-in help, 36
C
calibration factors for multiple oscilloscope
modules, 46
CAN serial triggering and analysis, 16
Channel 1 input, 22
Channel 2 input, 22
channel input BNC connectors, 19
characteristics, 68
chassis bus segments, connecting, 51
chassis ground terminal, 19, 23
cleaning, 24
communications problems, 34
compensate passive probes, 39
controller, power up, 27
copyright, 2
CXPI trigger and decode, 16
Czech user interface, 37
D
damage, inspect for, 12
data sheet, 68
default configuration, 37
default setup, 37
Demo 1 terminal, 40
Device Clear to remote interface, 63
display, area, 42
display, interpreting, 41
display, softkey labels, 43
display, status line, 42
driver not connected, 20
driver not connected, LED off, 20
E
English user interface and Quick Help, 37
environmental characteristics, 67
Ext Trig input, 19, 23
F
firmware update, 28
French user interface and Quick Help, 37
frequency response analyzer, 16
front panel features, 19
G
Gen Out, 19
Gen Out output, 23
German user interface and Quick Help, 37
graphical user interface language, 37
H
hard ware sel f test, 33
help, built-in, 36
humidity, 67
I
I2C serial triggering and analysis, 15
Identify LED, 20
Identify PXI Module blinking blue LED, 20
indoor use only, 67
information area, 42
inputs, 22
Interactive IO, 63
introduction, 7
Italian user interface and Quick Help, 37
IVI drivers, 61
J
Japanese user interface and Quick Help, 37
K
Keysight Connection Expert, 29
Keysight InfiniiVision Launcher, 31
Keysight Interactive IO application, 63
Keysight IO Control icon, 63
Korean user interface and Quick Help, 37
KtScopeCal.exe application, 46
L
LabVIEW driver, 62
language, user interface, 37
LED operation, 20
LEDs, AutoProbe power module, 19
LEDs, oscilloscope module, 19
LIN serial triggering and analysis, 16
M
markings, product, 69
mask limit testing, 15
master module, setting up, 50
measurement category, definitions, 69
measurement, making a, 35
menu line, 43
MIL-STD 1553 serial triggering and