Keysight M8290A Getting Started Manual

Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide
Notices
CAUTION
WARNING
© Keysight Technologies 2017
Manual Part Number
M8290-90A01
Edition
Edition 1.0, November 2017
Printed in Malaysia
Keysight Technologies Deutschland GmbH Herrenberger Strasse 130, 71034 Böblingen, Germany
Technology Licenses
The hard ware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement
(“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the soft­ware is customarily provided to the public. Accordingly, Keysight provides the Soft­ware to U.S. government customers under its standard commercial license, which is embodied in its End User License Agree­ment (EULA), a copy of which can be found at http://www.keysight.com/find/sweula. The license set forth in the EULA represents
the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Key­sight: (1) Furnish technical information related to commercial computer software or commercial computer software docu­mentation that is not customarily provided to the public; or (2) Relinquish to, or other­wise provide, the government rights in excess of these rights customarily provided to the public to use, mod ify, reproduce, release, perform, display, or d isclose com­mercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as appli­cable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCU­MENT IS PROVIDED "AS IS," AND IS SUB­JECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PAR-
TICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDEN­TAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCU­MENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREE­MENT WITH WARRANTY TERMS COVER­ING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating pro­cedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight M8290A Getting Started Guide

Safety Summary

General Safety
Precautions
Initial Inspection
General This product is a Safety Class 3 instrument. The protective features of this product may
Environment Conditions
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. For safe operation, the general safety precautions for the M9502A and M9505A AXIe chassis, must be followed.
See: http://www.keysight.com/find/M9505A
Keysight Technologies Inc. assumes no liability for the customer's failure to comply with these requirements. Before operation, review the instrument and manual for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
Inspect the shipping container for damage. If there is damage to the container or cushioning, keep them until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically. The Performance Tests give procedures for checking the operation of the instrument. If the contents are incomplete, mechanical damage or defect is apparent, or if an instrument does not pass the operator’s checks, notify the nearest Keysight Technologies Sales/Service Office.
WARNING To avoid hazardous electrical shock, do not perform electrical tests when there are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).
be impaired if it is used in a manner not specified in the operation instructions.
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate within a temperature range of 5 °C – 35 °C (41 °F – 95 °F) at a maximum relative humidity of 80% and at altitudes of up to 2000 meters.
This module can be stored or shipped at temperatures between -40 °C and +70 °C. Protect the module from temperature extremes that may cause condensation within it.
Ground the Instrument
Before Applying Power
Keysight M8290A Getting Started Guide 3
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Verify that all safety precautions are taken including those defined for the mainframe. The power cable inlet of the instrument serves as a device to disconnect from the mains in case of hazard. The instrument must be positioned so that the operator can easily access the power cable inlet. When the instrument is rack-mounted, the rack must be provided with an easily accessible mains switch.
Line Power Requirements The Keysight M829xA modules operate when installed in a Keysight AXIe mainframe.
Do Not Operate in an
Explosive Atmosphere
Do Not Remove the
Instrument Cover
Do not operate the instrument in the presence of flammable gases or fumes.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
4 Keysight M8290A Getting Started Guide

General Safety Considerations

CAUTION
WARNING

Intended Use

This instrument is intended to use in an office or laboratory environment, under the environmental conditions listed in the specifications.

Safety Symbols

The caution sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
The warning sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could resul t in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.

Instrument Markings

Table 1 Instrument Markings
Symbol Description
Indicates warning or caution. If you see this symbol on a product, you must refer to the manuals for specific Warning or Caution information to avoid personal injury or damage to the product.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The electrostatic discharge symbol. This warning symbol is marked on prod ucts which have components that can be damaged by an electrostatic discharge.
Keysight M8290A Getting Started Guide 5
Symbol Description
Indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate d uring normal use. Forty years is the expected useful life of the product.
The C-Tick mark is the certification mark of the Australian Communications Authority.
The RCM Mark is a compliance mark to the ACMA (Australian Spectrum Management Agency). This indicates compliance with all Australian EMC regulatory information.
CSA is the Canadian certification mark to demonstrate compliance with the Safety requirements.
The earthing symbol marks a connection that is connected, through the instrument, to the earth of the line power.
CE compliance marking to the EU Safety and EMC Directives. ISM GRP-1A classification according to the international EMC standard. ICES/NMB-001 compliance marking to the Canad ian EMC standard.
This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001.
6 Keysight M8290A Getting Started Guide

Compliance and Environmental Information

Table 2 Compliance and Environmental Information
Safety Symbol Description
This product complies with WEEE Directive (2002/96/EC) marking requirements. The affixed label ind icates that you must not discard this electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Keysight office, or see
http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more
information.
Keysight M8290A Getting Started Guide 7
8 Keysight M8290A Getting Started Guide

Contents

1 Introduction
2 Software Installation
3 Using the Instrument
Safety Summary 3
General Safety Considerations 5
Compliance and Environmental Information 7
Introduction 11 Getting Started 12 Additional Documents 12
Pre-Requisites 14 Downloading the Photonic Application Suite Package Manager 15 M8290A Software Installation 18 Installing the Keysight OMA Software 25 Post Installation Steps 27
How to use the Instrument 29 Adding the M829xA to the Connection Expert 30 Setting up the M829xA in the Keysight OMA Software 32
4 AXle Chassis
ESM Front Panel Connector 37 Setting the Startup Delay for the Embedded Controller 39
Keysight M8290A Getting Started Guide 9
Contents
5 M829xA Maintenance
Introduction 43 Initial Safety Information 44 Laser Safety Labels 45 ESD Protection 46 Power and Ventilation Requirements 47 Thermal Protection 47 Battery 47 Cleaning Recommendation 47
6 Specifications
M8292A Modular Optical Modulation Analyzer Specifications 50 M8296A Electrical High-Speed Digitizer Specifications 53 General Characteristics 55 Declaration of Conformity 55
10 Keysight M8290A Getting Started Guide
Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide

1 Introduction

Introduction / 11 Getting Started / 12 Additional Documents / 12

Introduction

The M8290A rack-mountable modular coherent test system for the AXIe platform is designed to address the 400G speed class in a significantly narrower form factor and a more attractive price point than today’s oscilloscope-based solutions for this speed class.
For coherent transmitter and receiver testing, the M8292A compact optical modulation analyzer and M8296A high speed digitizer fill the gap between the portable N4392A integrated optical modulation analyzer for 100G and the real-time oscilloscope-based N4391A optical modulation analyzer supporting speed classes of 400G, 600G and 1 Terabit per second.
The compact and modular approach makes the M8290A optical modulation analyzer and high-speed digitizer test solution an ideal system for coherent transmitter signal qualification for EVM and related parameters as well as for component characterization like ICR, PMQ or even fully-assembled CFPx-ACO modules. The modular concept addresses the needs of development teams, new product introduction groups and production test engineers looking for affordable test equipment for 400G. The M8290A modular coherent test solution provides a combination of compactness, affordability and performance that cannot be achieved with current oscilloscope-based solutions in this speed class.
1 Introduction
In the M8290A system configuration, the M8292A optical modulation analyzer module (a 2-slot AXIe module) and a 4-channel single-slot electrical digitizer module are available, both operating at 92 GSa/s sampling rate.

Getting Started

The M829xA is a modular instrument packaged in the AXIe form factor. AXIe is an open standard for high-performance modular instrumentation. Different form factors of AXIe chassis are available: 13-slot, 5-slot and 2-slot AXIe chassis.
The M8290A Coherent Test Solution comprises an M9537A AXIe embedded controller. This controller can control one or multiple M829xA and consumes one module slot of the AXIe chassis. The M8290A Coherent Test Solution comes with all required software pre-installed and configured at the factory. Therefore, the information given in Chapter 2 and Chapter 3 of this guide is only relevant if the user wishes to reinstall a software package or change the system configuration.
Login details for the embedded controller:
• Username: M8290a_admin
• Password: admin!123

Additional Documents

Additional documentation can be found at:
http://www.keysight.com/find/M9514A for 14-slot chassis related documentation.
http://www.keysight.com/find/M9505A for 5-slot chassis related documentation.
http://www.keysight.com/find/M9502A for 2-slot chassis related documentation.
http://www.keysight.com/find/M9537A for embedded AXIe controller related documentation.
http://www.keysight.com/find/M8290A for optical modulation analyzer and high-speed digitizer test solution related documentation
http://www.keysight.com/find/M8292A for AXIe based modular optical modulation analyzer related documentation
http://www.keysight.com/find/M8296A for AXIe based electrical high-speed digitizer related documentation
12 Keysight M8290A Getting Started Guide
Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide

2 Software Installation

Pre-Requisites / 14 Downloading the Photonic Application Suite Package Manager / 15 M8290A Software Installation / 18 Installing the Keysight OMA Software / 25 Post Installation Steps / 27
The M8290A coherent test solution uses a layered software architecture, which is illustrated below, that requires multiple software packages to be installed. This chapter explains how to install and configure these packages. However, in general, they are pre-installed at the factory. Therefore, the installation process described here only needs to be completed if the user wishes to reconfigure the system.
2 Software Installation

Pre-Requisites

The following are the pre-requisites for installing the Keysight M8290A software:
• The supported operating systems are:
Windows 10 (32 bit or 64 bit)
Windows 8.1 (32 bit or 64 bit)
Windows 8 (32 bit or 64 bit)
Windows 7 (32 bit or 64 bit)
• Ensure that you have Keysight IO Libraries Suite Version 17.3 or higher installed on your system. The Keysight IO Libraries Suite can be found at http://www.keysight.com/find/iosuite.
14 Keysight M8290A Getting Started Guide
NOTE
Even if a non-Keysight I/O library is already installed on your PC, it is still necessary to install the Keysight I/O library. The Keysight I/O library will install as “secondary” I/O library in this case. This use case is fully supported.

Downloading the Photonic Application Suite Package Manager

The Photonic Application Suite (PAS) Package Manager is required to select and download software packages for proper operation of the OMA. Skip this section if the PAS Package Manager is already present on the system.
1 Visit the PAS Package Manager web page at
www.keysight.com/find/photonic-sw, select the Current Version tab,
and click on the Download button to start the download process.
Software Installation 2
Keysight M8290A Getting Started Guide 15
2 Software Installation
2Click on the Open button to open the zip file.
3 Unzip the file and double-click on the executable file to run the PAS
Package Manager.
16 Keysight M8290A Getting Started Guide
Software Installation 2
4 Start the PAS Package Manager as administrator and select the OMA
tab to see all possible OMA packages.
Keysight M8290A Getting Started Guide 17
2 Software Installation

M8290A Software Installation

Follow the steps below to install the Keysight M8290A software on your system:
1 Start the PAS Package Manager as administrator, if it is not already
open and select the OMA tab to see all possible OMA packages.
2 Select the “M8290A Software”. 3Click on the Download link. 4 Continue following the installation instructions. The installer will first
check for and list some pre-requisites. Click Install to install them. If the embedded controller requests you to reboot, please do so immediately.
18 Keysight M8290A Getting Started Guide
Software Installation 2
5 The Keysight M8290A Setup will prepare the InstallShield Wizard for
the installation process. The following screen will appear.
Keysight M8290A Getting Started Guide 19
2 Software Installation
6Click Next. The following screen will appear:
20 Keysight M8290A Getting Started Guide
Software Installation 2
7Click Next to proceed to the license agreements.
8 Accept the terms of ‘Keysight Software End-User License Agreement
and click Next. The following screen will appear:
9 Select Yes if you want to read the post-installation instructions and
click Next. The following screen will appear:
Keysight M8290A Getting Started Guide 21
2 Software Installation
10 Select a setup type of either Complete or Custom. If you select
Complete, the M8290A documentation will be installed along with the M8290A software and a shortcut to the PAS Package Manager will be added to the Start menu. If you select Custom instead, you can specify which of the optional features to install.
11 Click Next. The following screen will appear:
22 Keysight M8290A Getting Started Guide
Software Installation 2
12 Click Install to begin the software installation. The Setup Wizard
will now install the M8290A software.
Keysight M8290A Getting Started Guide 23
2 Software Installation
13 The following screen will appear once the Keysight M8290A software is
successfully installed on your system.
24 Keysight M8290A Getting Started Guide
14 Click Finish to restart your system. This completes the Keysight
M8290A software installation.

Installing the Keysight OMA Software

1 Start the PAS Package Manager as administrator, if it is not already
open and select the OMA tab to see all possible OMA packages.
Software Installation 2
2 Select the “Keysight OMA Software”. 3Click on the Download link. 4 Continue following the installation instructions and always select the
default settings. The PAS Package Manager will install additional drivers and software required to install the OMA software. During the OMA installation, select M8290A Hardware as setup type. You may postpone all system restarts until the installation has been completed.
5 When the installation has been completed, verify that the PAS
Package Manager lists all installed components with an “Installed Version”. If any of them is not listed, please launch the installation of the missing package again.
Keysight M8290A Getting Started Guide 25
2 Software Installation
26 Keysight M8290A Getting Started Guide

Post Installation Steps

NOTE
NOTE
Software Installation 2
Follow the post installation steps as shown below: 1 Shut down the embedded controller. 2 Reboot the AXIe chassis. 3 The embedded controller should automatically recognize all available
M829xA modules. You may check this in the device manager; e.g. via
Start > Control Panel > Device Manager, or right-click Computer > Manage > Device Manager.
The instruments should be visible in the device tree as Keysight Technologies Modular Devices > M8296A or M8292A respectively.
4 Check if all installed M829xA modules are visible in the Connection
Expert. The connection expert can be opened by clicking its icon in the
system tray.
If something went wrong and any of the instruments does not appear in the PXI section, it may be necessary to reboot the embedded controller once more.
To successfully connect the M829xA to the PC, the post installation steps mentioned above must be strictly followed.
Reboot your PC immediately, if requested to do so.
Keysight M8290A Getting Started Guide 27
Keysight M8290A Optical Modulation Analyzer and
NOTE
NOTE
High-Speed Digitizer Test Solution
Getting Started Guide

3 Using the Instrument

How to use the Instrument / 29 Adding the M829xA to the Connection Expert / 30 Setting up the M829xA in the Keysight OMA Software / 32

How to use the Instrument

In order to use the instrument: 1 Turn on the AXIe chassis. 2 Start a separate instance of the M8290A Software
(Start > All Programs > Keysight M8290 > Keysight M8290 Software) for each M829xA module installed in the system. The user interface will display the VISA resource details required to set up the instrument in the Connection Expert.
3 Add the M829xA to the Connection Expert as described in section
Adding the M829xA to the Connection Expert on page 30. This step is
required to access the instrument from the Keysight OMA software.
4 Start the Keysight OMA software and set up the instrument connection
as described in the section Setting up the M829xA in the Keysight OMA
Software on page 32.
To show in the Keysight OMA Software, the module must have been added to the Connection Expert as a LAN instrument. Socket connections are not supported.
Always close the M8290A Software and shut down the embedded controller prior to switching off the AXIe chassis.
3 Using the Instrument

Adding the M829xA to the Connection Expert

A standard M8290A system is pre-configured at the factory. This section shows how to modify the factory configuration or to configure a stand-alone module.
1Open the Connection Expert by clicking its icon ( ) in the system
tray and selecting the corresponding item from the context menu.
2Click on the Add ( ) button in the “My Instruments” section and
select LAN instrument from the drop-down list. 3 Go to the ‘Enter Address’ tab. 4 Enter the hostname or the IP address of the embedded controller
running the M8290A software. Then set the Protocol to HiSlip and
enter the remote name as displayed on the M8290A Software user
interface.
30 Keysight M8290A Getting Started Guide
Using the Instrument 3
5Hit the OK button. You should now be able to view the instrument
details within the Connection Expert as indicated below.
6 Verify that the instrument details shown in the Connection Expert are
correct and that the instrument is listed with a green check mark in the
My Instruments” section. If so, the instrument is ready to be
connected to the Keysight OMA Software as described in the following
section.
Keysight M8290A Getting Started Guide 31
3 Using the Instrument

Setting up the M829xA in the Keysight OMA Software

A standard M8290A system is pre-configured at the factory. This section shows how to modify the factory configuration or to configure a stand-alone module.
1 In the Keysight 89600 VSA Software interface, select Utilities >
Hardware > Discovered Instruments to view a list of all available
instruments. A screen similar to the following is displayed.
However, if your module is not listed, please add it to the Connection
Expert as described in section Adding the M829xA to the Connection
Expert on page 30 and hit the refresh button ( ).
2Switch to the Configurations tab and click on the plus ( ) button.
32 Keysight M8290A Getting Started Guide
Using the Instrument 3
3From the Possible Logical Instruments list, select Keysight M8290
Optical Modulation Analyzer then click on the button to add it to
the Configuration list.
Keysight M8290A Getting Started Guide 33
3 Using the Instrument
4 In the ADC selection, click on the down arrow and select the desired
instrument.
34 Keysight M8290A Getting Started Guide
Using the Instrument 3
5In the Synthesizer selection, click on the down arrow and select the LO
configuration (internal or external LO). With the M8296A only an
external manually controlled LO is supported. This setting can also be
used if the test setup does not actually include an LO.
Keysight M8290A Getting Started Guide 35
3 Using the Instrument
6 In the name field, enter the desired name for this configuration then
click on the OK button. 7 Repeat this procedure to add more configurations.
36 Keysight M8290A Getting Started Guide
Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide

4 AXle Chassis

ESM Front Panel Connector / 37 Setting the Startup Delay for the Embedded Controller / 39
This chapter describes the usage of the AXIe chassis in combination with the M829xA.
The detailed documentation for the AXIe chassis can be found at:
http://www.keysight.com/find/M9514A for 14-slot chassis
http://www.keysight.com/find/M9505A for 5-slot chassis
http://www.keysight.com/find/M9502A for 2-slot chassis

ESM Front Panel Connector

The ESM Front Panel Connector is shown in the figure below:
4 AXle Chassis
1 PCle Connects a host PC to the chassis via PCle.
2 USB Connects a host PC to the chassis via
3 Multiframe Input Synchronizes timing signals with multiple
4 Multiframe Output
5 Trigger In External Trigger connections.
6 Trigger Out
7 Clock In External clock connections.
8 Clock Out
9 LAN Connects the host PC to the chassis, via
10 Status Light Indicates the chassis status.
USB 2.0. The USB 2.0 port is only available for AXIe chassis with option -U20. Using USB to connect to the M8290A is not supported.
daisy-chained chassis.
The Trigger In of the AXIe ESM cannot be used to trigger the M829xA. The Trigger Out of the AXIe ESM cannot be controlled by the M829xA.
10/100/1000 Ethernet. In particular, the LAN connector is used for ESM configuration, but NOT to communicate to the M829xA.
38 Keysight M8290A Getting Started Guide

Setting the Startup Delay for the Embedded Controller

For the PCIe connection between the embedded controller and the M829xA module to work properly, the AXIe modules must boot in a certain order. To assure that this order is maintained at all times, a startup delay needs to be configured on the AXIe chassis’ system module. All M8290A systems are pre-configured accordingly at the factory. However, if a different AXIe chassis is used with any M829xA module, it may be necessary to set the startup delay manually. To do this, please complete the process described below.
1 On the embedded controller, go to the AXIe chassis' web interface (e.g.
by typing “169.254.1.0” into the address bar of a web browser):
AXle Chassis 4
Keysight M8290A Getting Started Guide 39
4 AXle Chassis
2Click on the “Modules” button ( ) on the left side:
40 Keysight M8290A Getting Started Guide
AXle Chassis 4
3 Browse to the bottom of the “Chassis Module Configuration” box:
4If the “Active Configuration” shows a wait time below 15 seconds,
select “Auto” in the “New Configuration” list and enter 15 seconds as
the wait time. Then press “Apply New PCIe Settings”.
Keysight M8290A Getting Started Guide 41
4 AXle Chassis
5 Shutdown the embedded controller and power cycle the AXIe chassis.
This setting is now shown as the “Active Configuration”:
42 Keysight M8290A Getting Started Guide
Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide

5 M829xA Maintenance

Introduction / 43 Initial Safety Information / 44 ESD Protection / 46 Power and Ventilation Requirements / 47 Thermal Protection / 47 Battery / 47 Cleaning Recommendation / 47

Introduction

This chapter explains how to install and maintain the M829xA. It covers the following topics:
• Laser Safety
• ESD Protection
• Power & Ventilation Requirements
• Thermal Protection
• Battery
• Operating Environment
• Cleaning Recommendation
5 M829xA Maintenance

Initial Safety Information

The laser sources classified by this guide are classified as Class 1M according to IEC 60825-1 (2014).
All laser sources comply with 21 CFR 1040.10 except for deviations pursuant to Laser Notice No. 50, dated 2007-June-24.
Table 3 Initial Safety Information
M8292A Options
Max. CW output power
Beam waist diameter < 10 µm -
Numerical aperture 0.1 -
Laser class according to IEC 60825-1 (2014) Class 1M Class 1
Max. permissible CW output power
* Max. CW output power is defined as the highest possible optical CW power that the laser
source can produce at its output.
† Max. permissible CW output power is defined as the highest optical power that is
permitted within the appropriate IEC laser class.
*
C92+800
< 100 mW - (internal Laser only)
163 mW 10 mW
Options C92
44 Keysight M8290A Getting Started Guide

Laser Safety Labels

Laser Class 1 Label

Laser Class 1M Label

M829xA Maintenance 5
A sheet of laser safety labels is included. In order to meet the requirements of IEC 60825-1 we recommend that you stick the laser safety labels, in your language, onto a suitable location on the outside of the instrument where they are clearly visible to anyone using the instrument.
Keysight M8290A Getting Started Guide 45
5 M829xA Maintenance
WARNING
CAUTION
CAUTION
Please pay attention to the following laser safety warnings:
Under no circumstances look into the end of an optical cable attached
to the optical output when the device is operational. The laser radiation can seriously damage your eyesight.
Do not enable the laser when there is no fiber attached to the optical
output connector.
The laser is enabled by the software. The laser is on when the green
LED above the LO Output connector is lit.
The use of the instruments, such as microscopes or spectacles, with
this product will increase the hazard to your eyes.
The laser module has built-in safety circuitry which will disable the
optical output in the case of a fault condition.
Refer servicing only to qualified and authorized personnel.

ESD Protection

All the connectors are very sensitive to electrostatic discharge (ESD). When you connect a device or cable that is not fully discharged to these connectors, you risk damage to the instrument and expensive instrument repairs.
Electrostatic discharge (ESD) can damage the circuits of the M829xA. Avoid applying static discharges to the front-panel connectors. Before connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel connectors without first touching the frame of the instrument. Be sure the instrument and all connected devices (DUT, etc.) are properly earth-grounded (to a common ground) to prevent buildup of static charge and electrical over-stress.
46 Keysight M8290A Getting Started Guide

Power and Ventilation Requirements

WARNING
For power and ventilation requirements, refer to:
http://www.keysight.com/find/M9514A for 14-slot chassis related
documentation.
http://www.keysight.com/find/M9505A for 5-slot chassis related
documentation.
http://www.keysight.com/find/M9502A for 2-slot chassis related
documentation.

Thermal Protection

Overheating Detection

The instrument monitors its internal temperature. If the temperature exceeds approximately 80°C the power supply is switched off. The instrument will not turn on automatically if the temperature is decreasing again.

Fan Failure

M829xA Maintenance 5
If a fan is broken or prevented from operating by a blockage the temperature will increase. When the temperature exceeds approximately 80°C the overheating detection switches off the instrument for safety reasons. For reliability, it is recommended to send instruments with broken or defective fans immediately to Keysight Service for repair.

Battery

The M829xA does not have a battery.

Cleaning Recommendation

To prevent electrical shock, disconnect the instrument from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean external case parts. Do not attempt to clean internally.
Keysight M8290A Getting Started Guide 47
Keysight M8290A Optical Modulation Analyzer and High-Speed Digitizer Test Solution
Getting Started Guide

6 Specifications

M8292A Modular Optical Modulation Analyzer Specifications / 50 M8296A Electrical High-Speed Digitizer Specifications / 53 General Characteristics / 55 Declaration of Conformity / 55
6 Specifications

M8292A Modular Optical Modulation Analyzer Specifications

Table 4 M8292A Modular Optical Modulation Analyzer specifications
M8292A Modul ar optical modulation analyzer Typical values
Maximum detectable symbol rate 74 Gbaud
Sample rate range 83 to 92 GSa/s
Maximum record length per channel 512 kSa
ADC resolution 8 bit
Operating frequency range
Analog bandwid th, uncorrected 37 GHz (3 dB)
4
1
1 MHz to 40 GHz
Optical wavelength operating range 1527.60 to 1570.01 nm
(196.25 to 190.95 THz)
Average input power monitor accuracy ±0.4 dB
Optical phase angle of I-Q mixer after correction 90° ±0.5°
Relative skew after correction < ±1 ps
Image suppression
5
> 30 dB
Error vector magnitude noise floor < 2.4% EVM rms at 2.5 GHz freq. offset
< 3.5% EVM rms at 10 GHz freq. offset
2
Sensitivity
–20 dBm –14 dBm (with Option 800)
Internal local oscillator (built-in)
Wavelength settling time < 30 s
Wavelength uncertainty ±4.5 pm (±560 MHz), guaranteed
±2.5 pm (±310 MHz), typical
Sidemode Suppression Ratio (SMSR) 50 dB
Relative Intensity Noise (RIN)
3
–145 dB/Hz (10 MHz to 40 GHz)
5
6
50 Keysight M8290A Getting Started Guide
Specifications 6
External local oscillator input and output (Option 800)
LO input wavelength range 1527.60 to 1570.01 nm
LO input power range –3 dBm to +17 dBm
LO output power > +9 dBm
1 128 samples are unavailable, resulting in 511872 samples per channel effectively
available.
2 Valid at EVM = 32.5% for 32 GBaud DP-QPSK corresponding to raw BER = 1E-3,
boost mode off.
3 At maximum laser power.
4 Adjusted baseband frequency range available for signal analysis.
5 Valid at the following reference conditions
Sampling rate 92 GSa/s
Optical continuous wave signal at optical input port
Signal power > 0 dBm
Optical frequency is offset by 2.5 GHz from local oscillator frequency
Vector analyzer I-Q spectrum span set to 12.5 GHz
QPSK demodulation
10 Gbaud symbol rate
PolStokesAlign set to “Single Polarization”
KFPhaseTrack with carrier phase variance set to 1E-4
Result length set to 500 symbols
Raised cosine fil ter selected as reference filter
25° C ±5 K environmental temperature
6 Valid at reference conditions as stated above, except for:
Optical frequency is offset by 10 GHz from local oscillator frequency
Vector analyzer I-Q spectrum span set to 50 GHz
40 Gbaud symbol rate
Table 5 M8292A Modular Optical Modulation Analyzer specifications (Continued)
Trigger input Typical values
Input range –4 V to +4 V
Threshold –4 V to +4 V
Range 10 mV
Resolution 100 mV
Keysight M8290A Getting Started Guide 51
6 Specifications
Sensitivity 1 MHz to 40 GHz
Polarity Selectable: positive, negative, either edge
Timing uncertainty 8 ns
Reference clock input
Input frequency range 10 MHz to 17 GHz
Amplitude range 500 mV
pp
to 2 V
Impedance 50 Ohm (nominal)
Connector type SMA (female)
Reference clock output
Frequency with respect to sample rate fSa / 512
Amplitude 0.9 Vpp (nominal)
Impedance 50 Ohm (nominal)
Connector SMA (female)
Table 6 M8292A maximum ratings
Maximum ratings
Maximum signal input power +14.5 dBm
Signal input damage level +15 dBm
External LO input power
Maximum +17 dBm
Damage level +18 dBm
pp
Reference clock input damage level 3 V
pp
52 Keysight M8290A Getting Started Guide

M8296A Electrical High-Speed Digitizer Specifications

Table 7 M8296A Electrical High-Speed Digitizer Specifications
M8296A Electrical high-speed digitizer Typical values
Maximum detectable symbol rate 74 GBaud
Sample rate range 83 to 92 GSa/s
Maximum record length per channel 512 kSa
ADC resolution 8 bit
Operating frequency range
Analog bandwid th, uncorrected 37 GHz (3 dB), guaranteed
Skew between different input channels < ±250 ps
Skew between normal and complement < ±1 ps
2
Specifications 6
1
50 kHz to 42 GHz
3
Input amplitude ranges 150 mV
Input impedance 50 Ohm (nominal)
Number of input channels 4
Trigger input
Input range –4 V to +4 V
Threshold
Range –4 V to +4 V
Resolution 10 mV
Sensitivity 100 mV
Polarity Selectable: positive, negative, either edge
Timing uncertainty 8 ns
Keysight M8290A Getting Started Guide 53
300 mV 500 mV 800 mV
pp,diff
pp,diff
pp,diff
pp,diff
6 Specifications
Reference clock input
Input frequency range 10 MHz to 17 GHz
Amplitude range 500 mVpp to 2 V
Impedance 50 Ohm (nominal)
Connector type SMA (female)
Reference clock output
Frequency with respect to sample rate fSa / 512
Amplitude 0.9 Vpp (nominal)
Impedance 50 Ohm (nominal)
Connector SMA (female)
1 128 samples are unavailable, resulting in 511872 samples per channel effectively
available.
2 Adjusted baseband frequency range available for signal analysis.
3 Determined from a 9th order polynomial fit to the measured amplitude response.
pp
Table 8 M8296A maximum ratings
Maximum ratings
Damage level (single ended) 0.9 Vpp / 3 VDC
Damage level 3 V
pp
54 Keysight M8290A Getting Started Guide

General Characteristics

Table 9 General characteristics
Specifications 6
M8292A Modular 92 GSa/s optical modul ation analyzer
Dimensions (W x H x D)
Weight 3.6 kg 3.3 kg
Storage temperature range
Operating temperature range
Humidity 15% to 80% relative humidity, non-condensing
Operating altitude 0 to 2000 m
Power consumption
Safety designed to IEC61010-1, UL61010, CSA22.2 61010.1 tested
EMC tested to IEC61326-1
Form factor 2-slot AXIe 1-slot AXIe
Warm-up time 30 minutes
Recommended re-calibration interval
322.25 mm x 60 mm x 281.5 mm 322.25 mm x 30 mm x 281.5 mm
–40° C to +70° C
+5° C to +35° C
100 W at 92 GSa/s
2 years
M8296A Modular 92 GSa/s 4-channel electrical high-speed digitizer

Declaration of Conformity

Click the following link to view or download the latest version of DoC:
http://www.keysight.com/go/conformity
Keysight M8290A Getting Started Guide 55
This information is subject to change without notice. © Keysight Technologies 2017 Edition 1.0, November 2017
www.keysight.com
Loading...