Keysight U1273A/U1273AX
Handheld Digital Multimeter
Quick Start Guide
Page 2
Page 3
U1273A/U1273AX Handheld Digital Multimeter
NOTE
Verify that you received the following items in the shipment of
your multimeter:
✔One pair of red and black test leads
✔One pair of 4 mm test probes
✔One K-type thermocouple lead kit
✔Four 1.5 V AAA alkaline batteries (for U1273A) or lithium
batteries (for U1273AX)
✔Printed copies of the Certificate of Calibration (CoC) and
the U1273A/U1273AX Quick Start Guide (this manual)
If any item is missing or damaged, keep the shipping materials
and contact the nearest Keysight Sales Office.
The descriptions and instructions in this guide apply to the
U1273A/U1273AX handheld digital multimeter.
All related documents and software are available for download
at www.keysight.com/find/hhTechLib.
U1273A/U1273AX Quick Start Guide1
Page 4
U1273A/U1273AX Handheld Digital Multimeter
NOTE
Install the Batteries
Install the Batteries
Your multimeter is powered by four 1.5 V AAA batteries
(included with the shipment).
1 Turn the rotary switch to OFF and remove
the test leads from the terminals.
2 Lift the tilt stand and loosen the screws
with a suitable Phillips screwdriver.
3 Remove the battery cover and observe the
polarity markings.
4 Insert the batteries and replace the battery cover and screws.
Turn On the Multimeter
To power ON your multimeter, turn the rotary switch from the
OFF position to any other position.
Your multimeter is capable of remote data logging. To use this
feature, you will need an IR-USB cable (U1173A, purchased
separately) and the Keysight GUI Data Logger Software (downloadable from www.keysight.com/find/hhTechLib).
Auto Dim
By default, the multimeter’s Auto Dim function is enabled. The
multimeter’s backlight will dim automatically after 90 seconds
of inactivity. Press any button to cancel this effect and reset
the Auto Dim timer.
You may change how the multimeter’s backlight behavior
through the Setup menu. Refer to the User’s Guide for further
instructions.
2U1273A/U1273AX Quick Start Guide
Page 5
U1273A/U1273AX Handheld Digital Multimeter
Display screen
Keypad
Rotary switch
Input terminals
Test lead/probe holders
IR communication port
Battery cover
Tilt stand
(Lift tilt stand for access)
The Multimeter at a Glance
The Multimeter at a Glance
U1273A/U1273AX Quick Start Guide3
Page 6
U1273A/U1273AX Handheld Digital Multimeter
NOTE
S
h
i
f
t
V
i
e
w
E
s
c
Z
LOW
Smart
Auto
Understanding the Rotary Switch
Understanding the Rotary Switch
Press to switch between the primary and shifted func-
tions shown on the rotary switch.
LegendDescription
Z
(low input impedance) AC/DC V for eliminating
LOW
ghost voltages
Off
AC V
AC V with Low Pass Filter
AC mV
AC mV with Low Pass Filter
DC V
AC V or AC+DC V
DC mV
AC mV or AC+DC mV
Resistance
Continuity or Smart (offset compensation)
Diode
Auto-diode
Capacitance
Temperature
DC mA (or A)
4U1273A/U1273AX Quick Start Guide
AC mA (or A) or AC+DC mA (or A)
DC A
AC A or AC+DC A
Page 7
U1273A/U1273AX Handheld Digital Multimeter
S
c
a
l
e
N
u
l
l
P
e
a
k
M
a
x
M
i
n
A
u
t
o
T
r
i
g
H
o
l
d
Exit
Dual
Setup
H
z
%
m
s
L
o
g
A
u
t
o
R
a
n
g
e
S
h
i
f
t
V
i
e
w
E
s
c
Understanding the Keypad
Understanding the Keypad
Legend
Key response when pressed for:
Less than 1 secondMore than 1 second
Sets the Null/Relative
mode.
Starts the MaxMin
recording.
Freezes the present
reading in the display.
Switches between available dual-combination
displays.
Sets the Scale mode for
the specified ratio and
unit display.
Starts and stops the
Peak recording.
Automatically freezes
the present reading once
the reading is stable.
Exits the Hold, Null,
MaxMin, Peak, frequency test, and dual
display modes.
Increases the OLED
brightness incrementally when LOW,
MEDIUM, or HIGH set-
Enters and exits the multimeter’s Setup menu.
ting is selected.
U1273A/U1273AX Quick Start Guide5
Switches between frequency, pulse width, and
duty cycle measurements.
Sets a manual range and
disables autoranging.
Switches between the
primary and shifted
functions.
Starts and stops the
Data Logging.
Enables auto-ranging.
Enters and exits the Log
Review menu.
Page 8
U1273A/U1273AX Handheld Digital Multimeter
WARNING
Z
LOW
Smart
Auto
or
Understanding the Input Terminals
Understanding the Input Terminals
Ensure that the terminal connections are correct for that
particular measurement function before starting any
measurement. To avoid damage to the device, do not
exceed the input limit.
Rotary positionInput terminalsOverload protection
1000 Vrms
1000 Vrms for short
circuit <0.3 A
11 A/1000 V
fast-acting fuse
440 mA/1000 V
fast-acting fuse
6U1273A/U1273AX Quick Start Guide
Page 9
U1273A/U1273AX Handheld Digital Multimeter
DC
1
2
3
4
Performing Measurements and Tests
Performing Measurements and Tests
Voltage measurements
The figure below highlights the primary functions allowing
voltage measurements in your multimeter.
Set up your multimeter as shown in the figure below to perform voltage measurements. When measuring DC voltage from
a mixed signal in DC measurement mode, ensure that the Filter
(
) is enabled.
U1273A/U1273AX Quick Start Guide7
Page 10
U1273A/U1273AX Handheld Digital Multimeter
CAUTIONCAUTION
Setup
H
z
%
m
s
L
o
g
Exit
Dual
Setup
H
z
%
m
s
L
o
g
Setup
S
h
i
f
t
V
i
e
w
E
s
c
Performing Measurements and Tests
1 Press for more than 1 second to enter the multimeter’s
setup menu.
2 Browse to Menu 6>Filter, and press to edit the
value.
3 Press or to enable the Filter.
Refer to the table below for the respective firmware versions.
Version 1.64 or olderVersion 1.95 or newer
onDC
4 Press to save your changes.
5 Press and hold until the multimeter restarts and returns
to normal operation. Turn the multimeter to DC Voltage
mode to verify that the LPF symbol is turned on.
To avoid possible electric shock or personal injury, enable
the Filter
() to verify the presence of hazardous DC
voltages. Displayed DC voltages can be influenced by high
frequency AC components and must be filtered to assure
an accurate reading.
LPF during AC measurements:
Press while performing AC
voltage measurements to pass the
measured signal through a low pass
filter.
• Passing the measured signal through a LPF help blocks
unwanted voltages such as electronic noise.
• Use the LPF function to improve measurement on composite sine waves that are typically generated by inverters and
variable frequency motor drives.
8U1273A/U1273AX Quick Start Guide
Page 11
Z
Z
LOW
measurements:
LOW
U1273A/U1273AX Handheld Digital Multimeter
Performing Measurements and Tests
Rotate the rotary switch’s position to
to enable low impedance mea-
surements.
• Use the Z
(low input impedance) function to remove
LOW
ghost or induced voltages from your measurement.
• Z
can remove ghost voltages from your measurements
LOW
by dissipating the coupling voltage. Use Z
to reduce the
LOW
possibility of false readings in areas where the presence of
ghost voltages are suspected.
U1273A/U1273AX Quick Start Guide9
Page 12
U1273A/U1273AX Handheld Digital Multimeter
Smart
1
2
3
4
S
h
i
f
t
V
i
e
w
E
s
c
Exit
Dual
Performing Measurements and Tests
Resistance measurements
Set up your multimeter as shown in the figure below to perform resistance measurements.
Smart measurements:
• Use the Smart (offset compensation) function to measure
10U1273A/U1273AX Quick Start Guide
resistors affected by DC offset or leakage current.
• Smart removes unexpected DC voltages within the instru-
ment, at the input, or at the circuit being measured, which
will add error to resistance measurements. Press to
switch between the bias voltage (BiAS) display or leakage
current (LEAk) display, calculated based on the bias voltage
and corrected resistance value, on the secondary display.
While performing resistance measurements, press until BiAS is
shown on the secondary display to
enable the Smart function.
Page 13
U1273A/U1273AX Handheld Digital Multimeter
S
h
i
f
t
V
i
e
w
E
s
c
Exit
Dual
Smart
ON
(closed)
OFF
(open)
1
2
4
5
3
S
h
i
f
t
V
i
e
w
E
s
c
NOTE
Performing Measurements and Tests
Continuity tests
Set up your multimeter as shown in the figure below to perform continuity tests. Press to switch to the continuity
test function ( is shown on the display).
The beeper will sound as a continuity indication. Press to
switch between normal open () and normal close ()
contacts.
• Normal open: Circuit is normally open, the beeper will
sound when a short is detected.
• Normal close: Circuit is normally closed, the beeper will
sound when an open is detected.
The continuity function detects intermittent shorts and opens
lasting as short as 1 ms. A brief short or open causes the multimeter to emit a short beep.
U1273A/U1273AX Quick Start Guide11
Page 14
U1273A/U1273AX Handheld Digital Multimeter
1
2
3
4
S
h
i
f
t
V
i
e
w
E
s
c
Performing Measurements and Tests
Diode tests
Set up your multimeter as shown in the figure below to perform diode tests.
Auto
Auto-diode tests:
Press to use the auto diode
function.
• The Auto-diode function tests both the forward bias and
reverse bias directions of your diode simultaneously. The
forward bias voltage is shown on the primary display and
the reverse bias voltage is shown on the secondary display.
•GOOD is shown briefly on the secondary display along with
a brief beep if the diode is found to be in good condition. If
the diode is out of the thresholds, NGOOD is shown instead.
12U1273A/U1273AX Quick Start Guide
Page 15
U1273A/U1273AX Handheld Digital Multimeter
1
2
3
4
NOTE
Performing Measurements and Tests
Capacitance measurements
Set up your multimeter as shown in the figure below to perform capacitance measurements.
is shown on the bottom left of the display when the capacitor is charging, and is shown when the capacitor is discharging.
U1273A/U1273AX Quick Start Guide13
Page 16
U1273A/U1273AX Handheld Digital Multimeter
S
h
i
f
t
V
i
e
w
E
s
c
WARNING
1
2
4
5
3
S
h
i
f
t
V
i
e
w
E
s
c
K-type thermocouple
Heat source
NOTE
Performing Measurements and Tests
Temperature measurements
Set up your multimeter as shown in the figure below to perform temperature measurements. Press to switch to the
temperature measurement function.
Do not connect the thermocouple to electrically live
circuits. Doing so will potentially cause fire or electric
shock.
The multimeter uses a type-K (default setting) temperature
probe for measuring temperature.
14U1273A/U1273AX Quick Start Guide
Page 17
U1273A/U1273AX Handheld Digital Multimeter
S
h
i
f
t
V
i
e
w
E
s
c
WARNING
AC
LOAD
1
2
3
4
NOTE
Performing Measurements and Tests
Current measurements
Set up your multimeter as shown in the figure below to perform current measurements. Press to switch between
AC, DC, AC+DC, or % scale current measurements. When measuring DC current from a mixed signal in DC measurement
mode, ensure that the Filter (
Always use the proper function, range, and terminals for
current measurements. Set the positive input terminal to
the terminal for currents below 440 mA, and the
terminal for currents above 440 mA.
) is enabled.
For better accuracy when measuring low currents (up to A),
turn the rotary switch to the position.
U1273A/U1273AX Quick Start Guide15
Page 18
U1273A/U1273AX Handheld Digital Multimeter
Smart
Smart
To check Fuse 1To check Fuse 2
Check the Fuse
Check the Fuse
Follow the instructions below for a quick check on the fuses
(Fuse 1 and Fuse 2) of your multimeter.
Fuse
Part
number
12110-1400440 mA/1000 V102 OL
22110-140211 A/1000 V0.05 OL
Displayed readings
Fuse rating
Fuse healthy Replace fuse
16U1273A/U1273AX Quick Start Guide
Page 19
U1273A/U1273AX Handheld Digital Multimeter
NOTE
Check the Fuse
• To c he ck Fu se 1: Ensure that the probe tip is touching the
top half metal contact inside the µA mA terminal.
• To c he ck Fu se 2: Ensure that the probe tip is touching the
left half metal contact inside the A terminal.
The multimeter will sound an input warning alert if the probe
tip is in contact with any other sides of the µA mA or A terminal other than the sides specified in the instructions above.
U1273A/U1273AX Quick Start Guide17
Page 20
U1273A/U1273AX Handheld Digital Multimeter
Check the Fuse
18U1273A/U1273AX Quick Start Guide
Page 21
Contacting Keysight
CAUTIONCAUTION
WARNING
CAT III
1000 V
CAT IV
600 V
To obtain service, warranty, or technical assistance, contact us at the following phone numbers:
• United States Call Center: 800-829-4444
• Canada Call Center: 877-894-4414
• China Call Center: 800-810-0189
• Europe Call Center: 31-20-547-2111
• Japan Call Center: (81) 426-56-7832
For other countries, contact your country’s Keysight support organization. A
list of contact information for other countries is available on the Keysight
Web site: www.keysight.com/find/assist
Safety NoticesSafety and EMC Information
A CAUTION notice denotes
a hazard. It calls attention
to an operating procedure,
practice, or the like that, if
not correctly performed or
adhered to, could result in
damage to the product or
loss of important data. Do
not proceed beyond a CAU-TION notice until the indicated conditions are fully
understood and met.
A WARNING notice
denotes a hazard. It calls
attention to an operating
procedure, practice, or the
like that, if not correctly
performed or adhered to,
could result in personal
injury or death. Do not proceed beyond a WARNING
notice until the indicated
conditions are fully understood and met.
This meter is safety-certified in compliance
with EN/IEC 61010-1:2001, ANSI/UL
61010-1:2004, and CAN/CSA-C22.2 No.
61010-1-04 for CAT III 1000 V/CAT IV 600 V
pollution degree 2 environment. EMC is
designed in compliance with IEC
61326-1:2005/EN 61326-1:2006. Use with
standard or compatible test probes.
Safety Symbols
Earth (ground) terminal
Equipment protected
throughout by double insulation or reinforced insulation
Caution, risk of electric
shock
Caution, risk of danger (refer
to the instrument manual for
specific Warning or Caution
information)
Category III 1000 V
overvoltage protection
For further safety information details, refer to the
Keysight U1273A/U1273AX Handheld Digital Multimeter User’s Guide.
Category IV 600 V
overvoltage protection
Page 22
Page 23
This information is subject to change without notice.