Keysight InfiniiVision DSOX1204A, InfiniiVision DSOX1204G User Manual

Keysight InfiniiVision DSOX1204A/G Oscilloscopes
User's Guide
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2005-2018
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
N2137-97000
Edition
First edition, September 2018
Printed in Malaysia Published by:
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Print History
N2137-97000, September 2018
Warranty
The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology License
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or
otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

InfiniiVision DSOX1204A/G Oscilloscopes—At a Glance

Table 1 DSOX1204A/G Model Numbers, Bandwidths
Model: DSOX1204A DSOX1204G
Channels: 4
Bandwidth: 70 MHz
Bandwidth upgrades: 70 MHz to 100 MHz with D1200BW1A upgrade
70 MHz to 200 MHz with D1200BW2A upgrade
100 MHz to 200 MHz with D1200BW3A upgrade
Sampling rate: 2 GSa/s (interleaved), 1 GSa/s (non-interleaved)
Memory: 1 Mpts
Segmented memory: Yes
Waveform generator: No Yes (20 MHz)
Mask/limit test: Yes
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 3
The Keysight InfiniiVision DSOX1204A/G oscilloscopes deliver these features:
7 inch WVGA display.
50,000 waveforms/second update rate.
All knobs are pushable for making quick selections.
Trigger types: edge, pulse width, video, pattern, rise/fall time, and setup and hold.
Serial decode/trigger options for: I
Math waveforms: add, subtract, multiply, divide, FFT (magnitude and phase), and low-pass filter.
Reference waveforms (2) for comparing with other channel or math waveforms.
Many built-in measurements.
G-suffix models have built-in waveform generator with: sine, square, ramp, pulse, DC, noise.
USB/LAN ports makes printing, saving, and sharing data easy.
A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help. Complete instructions for using the quick help system are given in "Access the Built-In Quick Help" on page 29.
For more information about InfiniiVision oscilloscopes, see:
www.keysight.com/find/scope
2
C, UART/RS232, CAN, LIN, and SPI.
4 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

In This Guide

This guide shows how to use the InfiniiVision DSOX1204A/G oscilloscopes.
When unpacking and using the oscilloscope for the first time, see:
When displaying waveforms and acquired data, see:
When setting up triggers or changing how data is acquired, see:
Making measurements and analyzing data:
When using the built-in waveform generator, see:
Chapter 1, “Getting Started,” starting on page 11
"Running, Stopping, and Making Single
Acquisitions (Run Control)" on page 32
"Horizontal Controls" on page 33
"Vertical Controls" on page 36
"Analog Bus Display" on page 39
"FFT Spectral Analysis" on page 40
"Math Waveforms" on page 44
"Reference Waveforms" on page 46
"Display Settings" on page 47
"Triggers" on page 50
"Acquisition Control" on page 54
"Cursors" on page 61
"Measurements" on page 63
"Mask Testing" on page 65
"Digital Voltmeter" on page 72
"Frequency Response Analysis" on page 73
"Waveform Generator" on page 75
When using licensed serial bus decode and triggering features, see:
When saving, recalling, or printing, see:
"Serial Bus Decode/Trigger" on page 76
"Save/Recall (Setups, Screens, Data)" on page 82
"Print (Screens)" on page 85
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 5
When using the oscilloscope's utility
NOTE
functions, see:
For reference information, see: "Specifications and Characteristics" on page 95
Abbreviated instructions for pressing a series of keys and softkeys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:
Press [Key1]> Softkey2 > Softkey3.
The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display.
"Utility Settings" on page 87
"Web Interface" on page 93
"Environmental Conditions" on page 96
"Probes and Accessories" on page 97
"Software and Firmware Updates" on page 98
"Acknowledgements" on page 99
"Product Markings and Regulatory
Information" on page 101
6 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Contents

InfiniiVision DSOX1204A/G Oscilloscopes—At a Glance / 3
In This Guide / 5
1 Getting Started
Inspect the Package Contents / 12
Power-On the Oscilloscope / 13
Connect Probes to the Oscilloscope / 14
Maximum input voltage at analog inputs / 14
Do not float the oscilloscope chassis / 14
Input a Waveform / 15
Recall the Default Oscilloscope Setup / 16
Use Autoscale / 17
Compensate Passive Probes / 18
Learn the Front Panel Controls and Connectors / 20
Front Panel Overlays for Different Languages / 25
Learn the Rear Panel Connectors / 26
Learn the Oscilloscope Display / 27
Access the Built-In Quick Help / 29
2 Quick Reference
Running, Stopping, and Making Single Acquisitions (Run
Control) / 32
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 7
Horizontal Controls / 33
Horizontal Knobs and Keys / 33 Horizontal Softkey Controls / 33 Zoom / 34
Vertical Controls / 36
Vertical Knobs and Keys / 36 Vertical Softkey Controls / 36 Setting Analog Channel Probe Options / 38
Analog Bus Display / 39
FFT Spectral Analysis / 40
FFT Measurement Hints / 40 FFT DC Value / 42 FFT Aliasing / 42 FFT Spectral Leakage / 43
Math Waveforms / 44
Units for Math Waveforms / 45
Reference Waveforms / 46
Display Settings / 47
To load a list of labels from a text file you create / 48
Triggers / 50
Trigger Knobs and Keys / 50 Trigger Types / 50 Trigger Mode, Coupling, Reject, Holdoff / 51 External Trigger Input / 53
Acquisition Control / 54
Selecting the Acquisition Mode / 54 Overview of Sampling / 55
Cursors / 61
Cursor Knobs and Keys / 61
8 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Cursor Softkey Controls / 61
Measurements / 63
Mask Testing / 65
Creating/Editing Mask Files / 65
Digital Voltmeter / 72
Frequency Response Analysis / 73
Waveform Generator / 75
Serial Bus Decode/Trigger / 76
CAN Decode/Trigger / 77 I2C Decode/Trigger / 78 LIN Decode/Trigger / 78 SPI Decode/Trigger / 79 UART/RS232 Decode/Trigger / 80
Save/Recall (Setups, Screens, Data) / 82
Length Control / 83
Print (Screens) / 85
To set up network printer connections / 86
Utility Settings / 87
Setting up the Oscilloscope's LAN Connection / 89 USB Storage Devices / 91 Configuring the [Quick Action] Key / 92
Web Interface / 93
Accessing the Web Interface / 93
Specifications and Characteristics / 95
Environmental Conditions / 96
Declaration of Conformity / 96
Probes and Accessories / 97
Software and Firmware Updates / 98
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 9
Index
Acknowledgements / 99
Product Markings and Regulatory Information / 101
10 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Keysight InfiniiVision DSOX1204A/G Oscilloscopes
User's Guide

1 Getting Started

Inspect the Package Contents / 12 Power-On the Oscilloscope / 13 Connect Probes to the Oscilloscope / 14 Input a Waveform / 15 Recall the Default Oscilloscope Setup / 16 Use Autoscale / 17 Compensate Passive Probes / 18 Learn the Front Panel Controls and Connectors / 20 Learn the Rear Panel Connectors / 26 Learn the Oscilloscope Display / 27 Access the Built-In Quick Help / 29
This chapter describes the steps you take when using the oscilloscope for the first time.
11
1 Getting Started

Inspect the Package Contents

• Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
• Verify that you received the following items and any optional accessories you
may have ordered:
InfiniiVision DSOX1204A/G oscilloscope.
Power cord (country of origin determines specific type).
Oscilloscope probes (one for each analog input channel).
12 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Power-On the Oscilloscope

WARNING
Getting Started 1
Power
Requirements
Ventilation
Requirements
To powe r-on the
oscilloscope
Line voltage, frequency, and power:
• ~Line 100-120 Vac, 50/60/400 Hz
• 100-240 Vac, 50/60 Hz
•50W max
The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is required for proper cooling. Always ensure that the air intake and exhaust areas are free from obstructions.
The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscope.
When using the oscilloscope in a bench-top setting, provide at least 2" clearance at the sides and 4" (100 mm) clearance above and behind the oscilloscope for proper cooling.
1 Connect the power cord to the rear of the oscilloscope, then to a suitable AC
voltage source. Route the power cord so the oscilloscope's feet and legs do not pinch the cord.
2 The oscilloscope automatically adjusts for input line voltages in the range 100
to 240 VAC. The line cord provided is matched to the country of origin.
Always use a grounded power cord. Do not defeat the power cord ground.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 13
3 Press the power switch.
The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds.
1 Getting Started
CAUTION
CAUTION
WARNING

Connect Probes to the Oscilloscope

1 Connect the oscilloscope probe to an oscilloscope channel BNC connector. 2 Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point on the circuit.
Maximum input voltage at analog inputs
150 Vrms, 200 Vpk
Do not float the oscilloscope chassis
Defeating the ground connection and "floating" the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage. The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to measure between two live points, use a differential probe with sufficient dynamic range.
Do not negate the protective action of the ground connection to the oscilloscope. The oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard.
14 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Input a Waveform

WARNING
The Probe Comp signal is used for compensating probes. 1 Connect an oscilloscope probe from channel 1 to the Demo, Probe Comp terminal
on the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the Demo
terminal).
A voltage source should never be connected to the ground terminal of this instrument. If, for any reason, the Protective Conductor Terminal is disconnected or not functioning properly and a voltage source is connected to the equipment's ground terminals, the entire chassis will be at the voltage potential of the voltage source, and the operator or bystanders could receive an electric shock.
Getting Started 1
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 15
1 Getting Started

Recall the Default Oscilloscope Setup

To recall the default oscilloscope setup: 1 Press [Default Setup].
The default setup restores the oscilloscope's default settings. This places the oscilloscope in a known operating condition.
In the Save/Recall menu, there are also options for restoring the complete factory settings or performing a secure erase (see "Save/Recall (Setups, Screens,
Data)" on page 82).
16 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Use Autoscale

Getting Started 1
Use [Auto Scale] to automatically configure the oscilloscope to best display the input signals.
1 Press [Auto Scale].
You should see a waveform on the oscilloscope's display similar to this:
2 If you want to return to the oscilloscope settings that existed before, press Undo
Autoscale.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, press Fast Debug, Channels, or Acq Mode.
These are the same softkeys that appear in the Autoscale Preferences menu. See "Utility Settings" on page 87.
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure "Compensate Passive Probes" on page 18.
If you do not see the waveform, make sure the probe is connected securely to the front panel channel input BNC and to the Demo/Probe Comp terminal.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 17
1 Getting Started
NOTE
Perfectly compensated
Over compensated
Under compensated

Compensate Passive Probes

Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected. A poorly compensated probe can introduce significant measurement errors.
If your probe has a configurable attenuation setting (like the N2140/42A probes do), the 10:1 setting must be used for probe compensation.
1 Input the Probe Comp signal (see "Input a Waveform" on page 15). 2 Press [Default Setup] to recall the default oscilloscope setup (see "Recall the
Default Oscilloscope Setup" on page 16).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
Comp signal (see "Use Autoscale" on page 17).
4 Press the channel key to which the probe is connected ([1], [2], etc.). 5 In the Channel Menu, press Probe. 6 In the Channel Probe Menu, press Probe Check; then, follow the instructions
on-screen.
18 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the trimmer capacitor on the probe for the flattest pulse possible.
On some probes (like the N2140/42A probes), the trimmer capacitor is located on the probe BNC connector. On other probes (like the N2862/63/90 probes), the trimmer capacitor is a yellow adjustment on the probe tip.
7 Connect probes to all other oscilloscope channels.
8 Repeat the procedure for each channel.
Getting Started 1
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 19
1 Getting Started
6. [Auto Scale] key5. [Default Setup] key
10. Tools keys
1. Power switch
2. Softkeys
3. [Intensity] key
4. Entry knob
11. Trigger controls
12. Vertical controls
19. Demo/Probe Comp and Ground
terminals
15. [Help] key20. USB
Host
port
13. [Wave Gen] key
8. Run Control keys
9. Measure controls
7. Horizontal and Acquisition controls
16. Math function controls
18. Waveform generator
output
17. Analog channel
inputs
14. [FFT] key
Back
Back

Learn the Front Panel Controls and Connectors

On the front panel, key refers to any key (button) you can press.
Softkey specifically refers to the six keys next to the display. Menus and softkey labels appear on the display when other front panel keys are pressed. Softkey functions change as you navigate through the oscilloscope's menus.
For the following figure, refer to the numbered descriptions in the table that follows.
1. Power switch Press once to switch power on; press again to switch power off. See "Power-On the
Oscilloscope" on page 13.
2. Softkeys The functions of these keys change based upon the menus shown on the display next to the keys.
The Back key moves back in the softkey menu hierarchy. At the top of the hierarchy, the
Back key turns the menus off, and oscilloscope information is shown instead.
3. [Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to adjust waveform intensity.
You can vary the intensity control to bring out signal detail, much like an analog oscilloscope.
20 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Getting Started 1
4. Entry knob The Entry knob is used to select items from menus and to change values. The function of the Entry knob changes based upon the current menu and softkey selections.
Note that when the Entry knob symbol appears on a softkey, you can use the Entry knob, to select values.
Often, rotating the Entry knob is enough to make a selection. Sometimes, you can push the Entry knob to enable or disable a selection. Also, pushing the Entry knob can also make popup menus disappear.
5. [Default Setup]
key
6. [Auto Scale]
key
7. Horizontal and
Acquisition controls
Press this key to restore the oscilloscope's default settings (details on "Recall the Default
Oscilloscope Setup" on page 16).
When you press the [AutoScale] key, the oscilloscope will quickly determine which channels have activity, and it will turn these channels on and scale them to display the input signals. See "Use
Autoscale" on page 17.
The Horizontal and Acquisition controls consist of:
Horizontal scale knob — Turn the knob in the Horizontal section that is marked to
adjust the time/div setting. The symbols under the knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale.
Push the horizontal scale knob to toggle between fine and coarse adjustment.
Horizontal position knob — Turn the knob marked to pan through the waveform data
horizontally. You can see the captured waveform before the trigger (turn the knob clockwise) or after the trigger (turn the knob counterclockwise). If you pan through the waveform when the oscilloscope is stopped (not in Run mode) then you are looking at the waveform data from the last acquisition taken.
•[Acquire] key — Press this key to open the Acquire menu where you can select the Normal, XY, and
Roll time modes, enable or disable Zoom, and select the trigger time reference point.
Also you can select the Normal, Peak Detect, Averaging, or High Resolution acquisition modes and use segmented memory (see "Selecting the Acquisition Mode" on page 54).
Zoom key — Press the zoom key to split the oscilloscope display into Normal and Zoom
sections without opening the Acquire menu.
For more information see "Horizontal Controls" on page 33.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 21
1 Getting Started
8. Run Control
keys
9. Measure
controls
When the [Run/Stop] key is green, the oscilloscope is running, that is, acquiring data when trigger conditions are met. To stop acquiring data, press [Run/Stop].
When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring data, press [Run/Stop].
To capture and display a single acquisition (whether the oscilloscope is running or stopped), press [Single]. The [Single] key is yellow until the oscilloscope triggers.
For more information, see "Running, Stopping, and Making Single Acquisitions (Run
Control)" on page 32.
The measure controls consist of:
•[Analyze] key — Press this key to access analysis features like:
Trigger level setting.
Measurement threshold setting.
Video trigger automatic set up and display.
Display a bus made up of the analog channel inputs where channel 1 is the least significant bit
and channel 4 is the most significant bit. See also "Analog Bus Display" on page 39.
Enable serial bus decodes. See also "Serial Bus Decode/Trigger" on page 76.
Reference waveforms (see "Reference Waveforms" on page 46).
Mask testing (see "Mask Testing" on page 65).
Digital voltmeter (see "Digital Voltmeter" on page 72).
Frequency response analysis on models with a built-in waveform generator (see "Frequency
Response Analysis" on page 73).
•[Meas] key — Press this key to access a set of predefined measurements. See
"Measurements" on page 63.
•[Cursors] key — Press this key to open a menu that lets you select the cursors mode and source.
Cursors knob — Push this knob select cursors from a popup menu. Then, after the popup menu
closes (either by timeout or by pushing the knob again), rotate the knob to adjust the selected cursor position.
22 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Getting Started 1
T
10. Tools keys The Tools keys consist of:
•[Save/Recall] key — Press this key to save oscilloscope setups, screen images, waveform data, or
mask files or to recall setups, mask files or reference waveforms. See "Save/Recall (Setups,
Screens, Data)" on page 82.
• [Utility] key — Press this key to access the Utility menu, which lets you configure the
oscilloscope's I/O settings, use the file explorer, set preferences, access the service menu, and choose other options. See "Utility Settings" on page 87.
• [Display] key — Press this key to access the menu where you can enable persistence, adjust the
display grid (graticule) intensity, label waveforms, add an annotation, and clear the display (see
"Display Settings" on page 47).
• [Quick Action] key — Press this key to perform the selected quick action: measure all snapshot,
print, save, recall, freeze display. and more. See "Configuring the [Quick Action] Key" on page 92.
•[Save to USB] key — Press this key to perform a quick save to a USB storage device.
11. Trigger controls The Trigger controls determine how the oscilloscope triggers to capture data. These controls consist of:
Level knob — Turn the Level knob to adjust the trigger level for a selected analog channel.
Push the knob to set the level to the waveform's 50% value. If AC coupling is used, pushing the Level knob sets the trigger level to about 0 V.
The position of the trigger level for the analog channel is indicated by the trigger level icon (if the analog channel is on) at the far left side of the display. The value of the analog channel trigger level is displayed in the upper-right corner of the display.
• [Trigger] key — Press this key to select the trigger type (edge, pulse width, video, etc.). See
"Trigger Types" on page 50. You can also set options that affect all trigger types. See "Trigger Mode, Coupling, Reject, Holdoff" on page 51.
•[Force] key — Causes a trigger (on anything) and displays the acquisition.
This key is useful in the Normal trigger mode where acquisitions are made only when the trigger condition is met. In this mode, if no triggers are occurring (that is, the "Trig'd?" indicator is displayed), you can press [Force] to force a trigger and see what the input signals look like.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 23
1 Getting Started
12. Vertical
controls
13. [Wave Gen] key On G-suffix models that have a built-in waveform generator, press this key to access waveform
14. [FFT] key Provides access to FFT spectrum analysis function. See "FFT Spectral Analysis" on page 40.
15. [Help] key Opens the Help menu where you can display overview help topics and select the Language. See also
The Vertical controls consist of:
Analog channel on/off keys — Use these keys to switch a channel on or off, or to access a
channel's menu in the softkeys. There is one channel on/off key for each analog channel.
Vertical scale knob — Use the knob marked to change the vertical sensitivity (gain) for
the selected analog input channel.
Push the vertical scale knob to toggle between fine and coarse adjustment.
The default mode for expanding the signal is about the ground level of the channel; however, you can change this to expand about the center of the display.
Vertical position knob — Use this knob to change the selected analog input channel waveform
vertical position on the display.
The voltage value momentarily displayed in the upper right portion of the display represents the
voltage difference between the vertical center of the display and the ground level ( ) icon. It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground.
For more information, see "Vertical Controls" on page 36.
generator functions. See "Waveform Generator" on page 75.
"Access the Built-In Quick Help" on page 29.
16. Math function
controls
17. Analog channel
inputs
The Math function controls consist of:
•[Math] key — provides access to math (add, subtract, etc.) waveform functions. See "Math
Waveforms" on page 44.
Vertical scale knob — Use the knob marked to change the vertical sensitivity (as with
the analog channel vertical controls).
Vertical position knob — Use the knob to change a math function waveform's vertical position on
the display (as with the analog channel vertical controls).
Attach oscilloscope probes or BNC cables to these BNC connectors.
In the InfiniiVision 1000 X-Series oscilloscopes, the analog channel inputs have 1 M
Also, there is no automatic probe detection, so you must properly set the probe attenuation for accurate measurement results. See "Setting Analog Channel Probe Options" on page 38.
Ω impedance.
24 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Getting Started 1
18. Waveform
generator output
19. Demo/Probe
Comp, Ground terminals
20. USB Host port This port is for connecting USB mass storage devices or printers to the oscilloscope.
On G-suffix models, the built-in waveform generator can output sine, square, ramp, pulse, DC, or noise on the Gen Out BNC. Press the [Wave Gen] key to set up the waveform generator. See
"Waveform Generator" on page 75.
You can also send the trigger output signal or the mask test failure signal to the Gen Out BNC connector. See "Utility Settings" on page 87.
Demo terminal — This terminal outputs the Probe Comp signal which helps you match a probe's
input capacitance to the oscilloscope channel to which it is connected. See "Compensate
Passive Probes" on page 18. The oscilloscope can also output demo or training signals on this
terminal.
Ground terminal — Use the ground terminal for oscilloscope probes connected to the Demo/Probe
Comp terminal. See the warning in "Input a Waveform" on page 15.
Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to save or recall oscilloscope setup files and reference waveforms or to save data and screen images. See
"Save/Recall (Setups, Screens, Data)" on page 82.
To print, connect a USB compliant printer. For more information about printing see "Print
(Screens)" on page 85.
You can also use the USB port to update the oscilloscope's system software when updates are available.
NOTE: You must Eject a USB mass storage device before unplugging it; otherwise, the device will be marked as needing repair when connecting to a computer with the Windows operating system (even though there are no harmful effects to the device).
CAUTION:Do not connect a host computer to the oscilloscope's USB host port. A host computer sees the oscilloscope as a device, so connect the host computer to the oscilloscope's device port (on the rear panel). See "Learn the Rear Panel Connectors" on page 26.

Front Panel Overlays for Different Languages

Front panel overlays, which have translations for the English front panel keys and label text, are available in many languages. The appropriate overlay is included when the localization option is chosen at time of purchase.
To install a front panel overlay:
1 Gently pull on the front panel knobs to remove them. 2 Insert the overlay's side tabs into the slots on the front panel. 3 Reinstall the front panel knobs.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 25
1 Getting Started

Learn the Rear Panel Connectors

For the following figure, refer to the numbered descriptions in the table that follows.
2. Kensington lock hole
3. USB Device port
4. LAN port
5. Ext Trig input
1. Power cord connector
1. Power cord
connector
2. Kensington lock
hole
3. USB Device
port
4. LAN port Lets you print to network printers (see "Print (Screens)" on page 85), access the oscilloscope's
15. Ext Trig input External trigger input BNC connector. See "External Trigger Input" on page 53 for an explanation
Attach the power cord here.
This is where you can attach a Kensington lock for securing the instrument.
This port is for connecting the oscilloscope to a host PC. You can issue remote commands from a host PC to the oscilloscope via the USB device port. See "Utility Settings" on page 87.
built-in web server, and issue remote commands. See "Web Interface" on page 93, "Accessing
the Web Interface" on page 93, and "Utility Settings" on page 87.
of this feature.
26 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Learn the Oscilloscope Display

Analog channel sensitivity
Status line
Analog channels and ground levels
Trigger level
Trigger point, time reference
Delay time
Time/ div
Run/Stop status
Trigger type
Trigger source
Measurements
Trigger level
Softkey labels and information area
Cursors defining measurement
Other waveforms
The oscilloscope display contains acquired waveforms, setup information, measurement results, and the softkey definitions.
Getting Started 1

Figure 1 Interpreting the oscilloscope display

Status line The top line of the display contains vertical, horizontal, and trigger setup information.
Display area The display area contains the waveform acquisitions, channel identifiers, and analog trigger, and
ground level indicators. Each analog channel's information appears in a different color.
Signal detail is displayed using 256 levels of intensity.
For more information about display modes see "Display Settings" on page 47.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 27
1 Getting Started
Back
Back
Softkey labels and information area
Measurements area When measurements or cursors are turned on, this area contains automatic measurement and cursor
When most front panel keys are pressed, short menu names and softkey labels appear in this area. The labels describe the softkey functions. Typically, softkeys let you set up additional parameters for the selected mode or menu.
Pressing the Back key returns through the menu hierarchy until softkey labels are off and the information area is displayed. The information area contains acquisition, analog channel, math function, and reference waveform information.
You can also specify that softkey menus turn off automatically after a specified timeout period ([Utility] > Options > Menu Timeout).
Pressing the Back key when the information area is displayed returns to the most recent menu displayed.
results.
When measurements are turned off, this area displays additional status information describing channel offset and other configuration parameters.
28 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Getting Started 1

Access the Built-In Quick Help

To view Quick Help 1 Press and hold the key, softkey, or knob for which you would like to view help.
Quick Help remains on the screen until another key is pressed or a knob is turned.
To select the user
interface and
Quick Help
language
To select the user interface and Quick Help language:
1 Press [Help], then press the Language softkey. 2 Turn the Entry knob until the desired language is selected.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 29
1 Getting Started
30 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Keysight InfiniiVision DSOX1204A/G Oscilloscopes
User's Guide

2 Quick Reference

Running, Stopping, and Making Single Acquisitions (Run Control) / 32 Horizontal Controls / 33 Vertical Controls / 36 Analog Bus Display / 39 FFT Spectral Analysis / 40 Math Waveforms / 44 Reference Waveforms / 46 Display Settings / 47 Triggers / 50 Acquisition Control / 54 Cursors / 61 Measurements / 63 Mask Testing / 65 Digital Voltmeter / 72 Waveform Generator / 75 Serial Bus Decode/Trigger / 76 Save/Recall (Setups, Screens, Data) / 82 Print (Screens) / 85 Utility Settings / 87 Web Interface / 93 Specifications and Characteristics / 95 Environmental Conditions / 96 Probes and Accessories / 97 Software and Firmware Updates / 98 Acknowledgements / 99 Product Markings and Regulatory Information / 101
31
2 Quick Reference

Running, Stopping, and Making Single Acquisitions (Run Control)

To display the results of multiple acquisitions, use persistence. See "Display
Settings" on page 47.
Single vs. Running
and Record Length
Table 2 Run Control Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Run acquisitions [Run/Stop] (the key is green when running)
Stop acquisitions [Run/Stop] (the key is red when stopped)
The maximum data record length is greater for a single acquisition than when the oscilloscope is running (or when the oscilloscope is stopped after running):
Single — Single acquisitions always use the maximum memory available — at
least twice as much memory as acquisitions captured when running — and the oscilloscope stores at least twice as many samples. At slower time/div settings, because there is more memory available for a single acquisition, the acquisition has a higher effective sample rate.
Running — When running (versus taking a single acquisition), the memory is
divided in half. This lets the acquisition system acquire one record while processing the previous acquisition, dramatically improving the number of waveforms per second processed by the oscilloscope. When running, a high waveform update rate provides the best representation of your input signal.
To acquire data with the longest possible record length, press the [Single] key.
For more information on settings that affect record length, see "Length
Control" on page 83.
Single acquisition [Single] (the key is yellow until the oscilloscope triggers)
If the oscilloscope does not trigger, you can press [Force Trigger] to trigger on anything and make a single acquisition.
32 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Horizontal Controls

Trigger point
Time reference
Delay time
Time/ div
Trigger source
Trigger level or threshold
XY or Roll mode
Normal time mode
Zoomed time base
Time reference

Horizontal Knobs and Keys

Horizontal Softkey Controls

The following figure shows the Acquire menu which appears after pressing the [Acquire] key.
Quick Reference 2
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 33
2 Quick Reference
The time reference is indicated at the top of the display grid by a small hollow triangle (∇). Turning the Horizontal scale knob expands or contracts the waveform about the time reference point (∇).
The trigger point, which is always time = 0, is indicated at the top of the display grid by a small solid triangle (▼).
The delay time is the time of the reference point with respect to the trigger. Turning the Horizontal position ( ) knob moves the trigger point () to the left
or right of the time reference () and displays the delay time.
The Acquire menu lets you select the time mode (Normal, XY, or Roll), enable Zoom, set the time base fine control (vernier), and specify the time reference.
Table 3 Horizontal Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Time mode [Acquire] > Time Mode (Normal, XY, or Roll)
XY time mode [Acquire] > Time Mode, XY
Channel 1 is the X-axis input, channel 2 is the Y-axis input. The Z-axis input (Ext Trig) turns the trace on and off (blanking). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off.
Measuring the phase difference between two signals of the same frequency with the Lissajous method is a common use of the XY display mode (see the "XY Display Mode Example" description at
www.keysight.com/find/xy-display-mode).
Roll time mode [Acquire] > Time Mode, Roll

Zoom

Time reference [Acquire] > Time Ref (Left, Center, Right)
See Also "Acquisition Control" on page 54
[Acquire] > Zoom (or press the zoom key)
Zoom
The Zoom window is a magnified portion of the normal time/div window. To turn
on (or off) Zoom, press the zoom key (or press the [Acquire] key and then the Zoom softkey).
34 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
These markers show the beginning and end of the Zoom window
Normal window
Time/div for zoomed window
Time/div for normal window
Delay time momentarily displays when the Horizontal position knob is turned
Zoom window
Signal anomaly expanded in zoom window
Select Zoom
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 35
2 Quick Reference
NOTE
NOTE

Vertical Controls

Vertical Knobs and Keys

Keysight recommends always scaling the signal so that the entire waveform is contained between the top and bottom of the display.
For proper operation of the 1000 X-Series oscilloscope, the channel inputs must not be overdriven more than ±8 divisions. Exceeding this limit may result in signals that appear incorrect and may increase crosstalk between the input channels.
To minimize crosstalk between input channels, make sure the channel is not overdriven. Also, connecting a probe or cable to a channel will reduce crosstalk.

Vertical Softkey Controls

The following figure shows the Channel 1 menu that appears after pressing the [1] channel key.
36 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
Channel, Volts/div
Channel 1 ground level
Trigger source
Trigger level or threshold
Channel 2 ground level
The ground level of the signal for each displayed analog channel is identified by the position of the icon at the far-left side of the display.
Table 4 Vertical Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Channel coupling [1/2/3/4] > Coupling (DC or AC)
Channel bandwidth limit
Vertical scale fine adjustment
Channel Invert [1/2/3/4] > Invert
Note that Channel Coupling is independent of Trigger Coupling. To change trigger coupling see
"Trigger Mode, Coupling, Reject, Holdoff" on page 51.
[1/2/3/4] > BW Limit
[1/2/3/4] > Fine
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 37
2 Quick Reference
CAUTION

Setting Analog Channel Probe Options

In the Channel menu, the Probe softkey opens the Channel Probe menu.
This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe.
For correct measurements, you must match the oscilloscope's probe attenuation factor settings with the attenuation factors of the probes being used.
Table 5 Probe Features
Channel Probe Menu Feature Front Panel Key/Softkey Location (see built-in help for more information)
Channel units [1/2/3/4] > Probe > Units (Volts, Amps)
Probe attenuation
Channel skew
Probe check [1/2/3/4] > Probe > Probe Check
[1/2/3/4] > Probe > Probe, Ratio/Decibels, Entry knob
Changes the vertical scale so that measurement results reflect the actual voltage levels at the probe tip.
[1/2/3/4] > Probe > Skew, Entry knob
Guides you through the process of compensating passive probes (such as the N2140A, N2142A, N2862A/B, N2863A/B, N2889A, N2890A, 10073C, 10074C, or 1165A probes).
38 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Analog Bus Display

You can display a bus made up of the analog channel inputs. Any of the input channels can be assigned to the bus. The bus values display appears at the bottom of the graticule. Channel 1 is the least significant bit and channel 4 is the most significant bit.
Table 6 Analog Bus Display Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Analog bus, display
Analog bus, channel assignment
Analog bus, value number base
Analog bus, threshold source
Analog bus, threshold level for selected source
[Analyze] > Features, Entry knob to select Analog Bus Analysis, push Entry knob to enable or disable
[Analyze] > Channel, Entry knob, push Entry knob to make or clear assignment
[Analyze] > Base, Entry knob (Hex, Binary)
[Analyze] > Source, Entry knob (1, 2, 3, 4), push Entry knob to toggle between source and
threshold
[Analyze] > Threshold, Entry knob, push Entry knob to toggle between source and threshold
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 39
2 Quick Reference

FFT Spectral Analysis

FFT is used to compute the fast Fourier transform using analog input channels. FFT takes the digitized time record of the specified source and transforms it to the frequency domain.
When the FFT function is selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency. The readout for the horizontal axis changes from time to frequency (Hertz) and the vertical readout changes from volts to dB.
Use the FFT function to find crosstalk problems, to find distortion problems in analog waveforms caused by amplifier non-linearity, or for adjusting analog filters.
Table 7 FFT Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
FFT span/center [FFT] > Span
[FFT] > Center
FFT window [FFT] > Settings > Window (Hanning, Flat Top, Rectangular, Blackman Harris, see also "FFT
Spectral Leakage" on page 43)
FFT vertical units [FFT] > Settings > Vertical Units (Decibels, VRMS)
FFT auto setup [FFT] > Settings > Auto Setup
FFT waveform, scale [FFT], turn the Math vertical scale knob
FFT waveform, offset [FFT], turn the Math vertical position knob

FFT Measurement Hints

The number of points acquired for the FFT record can be up to 65,536, and when frequency span is at maximum, all points are displayed. Once the FFT spectrum is displayed, the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater detail. Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution. As frequency span is decreased, the number of points shown is reduced, and the display is magnified.
While the FFT spectrum is displayed, use the [FFT] and [Cursors] keys to switch between measurement functions and frequency domain controls in FFT Menu.
40 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
NOTE
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N). With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the resolution.
Decreasing the effective sampling rate by selecting a greater time/div setting will increase the low frequency resolution of the FFT display and also increase the chance that an alias will be displayed. The resolution of the FFT is the effective sample rate divided by the number of points in the FFT. The actual resolution of the display will not be this fine as the shape of the window will be the actual limiting factor in the FFTs ability to resolve two closely space frequencies. A good way to test the ability of the FFT to resolve two closely spaced frequencies is to examine the sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
• Make sure the probe attenuation is set correctly. The probe attenuation is set from the Channel Menu if the operand is a channel.
• Set the source sensitivity so that the input signal is near full screen, but not clipped.
• Use the Flat Top window.
• Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
• Use the Hanning window.
• Use Cursors to place an X cursor on the frequency of interest.
• Adjust frequency span for better cursor placement.
• Return to the Cursors Menu to fine tune the X cursor.
For more information on the use of FFTs please refer to Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf. Additional
information can be obtained from Chapter 4 of the book Spectrum and Network Measurements by Robert A. Witte.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 41
2 Quick Reference
NOTE

FFT DC Value

FFT Aliasing

The FFT computation produces a DC value that is incorrect. It does not take the offset at center screen into account. The DC value is not corrected in order to accurately represent frequency components near DC.
When using FFTs, it is important to be aware of frequency aliasing. This requires that the operator have some knowledge as to what the frequency domain should contain, and also consider the sampling rate, frequency span, and oscilloscope vertical bandwidth when making FFT measurements. The FFT resolution (the quotient of the sampling rate and the number of FFT points) is displayed directly above the softkeys when the FFT Menu is displayed.
Nyquist Frequency and Aliasing in the Frequency Domain
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. This frequency is half of the sample rate. Frequencies above the Nyquist frequency will be under sampled, which causes aliasing. The Nyquist frequency is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain.
Aliasing happens when there are frequency components in the signal higher than half the sample rate. Because the FFT spectrum is limited by this frequency, any higher components are displayed at a lower (aliased) frequency.
The following figure illustrates aliasing. This is the spectrum of a 990 Hz square wave, which has many harmonics. The horizontal time/div setting for the square wave sets the sample rate and results in a FFT resolution of 1.91 Hz. The displayed FFT spectrum waveform shows the components of the input signal above the Nyquist frequency to be mirrored (aliased) on the display and reflected off the right edge.
42 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Figure 2 Aliasing
Quick Reference 2
Because the frequency span goes from 0 to the Nyquist frequency, the best way to prevent aliasing is to make sure that the frequency span is greater than the frequencies of significant energy present in the input signal.

FFT Spectral Leakage

The FFT operation assumes that the time record repeats. Unless there is an integral number of cycles of the sampled waveform in the record, a discontinuity is created at the end of the record. This is referred to as leakage. In order to minimize spectral leakage, windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT. The FFT Menu provides four windows: Hanning, Flat Top, Rectangular, and Blackman-Harris. For more information on leakage, see Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 43
2 Quick Reference
TIP

Math Waveforms

Math functions can be performed on analog channels and lower math functions. The resulting math waveform is displayed in light purple.
Table 8 Math Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Math operator [Math] > Operator (Add, Subtract, Multiply, Divide, FFT Magnitude, FFT Phase, Low Pass Filter)
Cascaded math functions
Math function waveforms, scale
Math function waveforms, offset
[Math] > Source
[Math], turn the Math vertical scale knob
[Math], turn the Math vertical position knob
Math Operating Hints
Table 9 FFT (Magnitude), FFT (Phase) Operator Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Auto setup [Math] > Auto Setup
Span/center [Math] > More > Span
If the analog channel or math function is clipped (not fully displayed on screen) the resulting displayed math function will also be clipped.
Once the function is displayed, the analog channel(s) may be turned off for better viewing of the math waveform.
The math function waveform can be measured using [Cursors] and/or [Meas].
[Math] > More > Center
Window function [Math] > More > Window (Hanning, Flat Top, Rectangular, Blackman Harris, see also "FFT Spectral
Leakage" on page 43)
44 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
Table 9 FFT (Magnitude), FFT (Phase) Operator Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Vertical units [Math] > More > Vertical Units (For FFT (Magnitude): Decibels or V RMS. For FFT (Phase): Radians or
Degrees.)
FFT (Phase) zero phase reference point
[Math] > More > Zero Phase Ref (Trigger, Entire Display)
Table 10 Low Pass Filter Operator Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Math low-pass filter cutoff frequency
[Math] > Bandwidth

Units for Math Waveforms

Units for each input channel can be set to Volts or Amps using the Units softkey in the channel's Probe Menu. Units for math function waveforms are:
Math function Units
add or subtract V or A
multiply
FFT Magnitude dB (decibels) or V RMS.
FFT Phase degrees or radians
2
, A2, or W (Volt-Amp)
V
A scale unit of U (undefined) will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 45
2 Quick Reference

Reference Waveforms

Analog channel or math waveforms can be saved to one of two reference waveform locations in the oscilloscope. Then, a reference waveform can be displayed and compared against other waveforms. One reference waveform can be displayed at a time.
Table 11 Reference Waveform Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Reference waveforms, display
Reference waveforms, save
Reference waveforms, skew
Reference waveforms, scale
Reference waveforms, offset
Reference waveforms, clear
Reference waveforms, info
Reference waveforms, info, transparent background
Reference waveforms, save/recall from USB storage device
[Analyze] > Features, Entry knob to select R1 or R2, push Entry knob to enable or disable
[Analyze] > Save/Clear > Source, [Analyze] > Save/Clear > Save to
[Analyze] > Skew, Entry knob
[Analyze] > Scale, Entry knob
[Analyze] > Offset, Entry knob
[Analyze] > Save/Clear > Clear
[Save/Recall] > Default/Erase > Secure Erase
[Analyze] > Save/Clear > Display Info
[Analyze] > Save/Clear > Transparent
[Save/Recall] > Save > Format, Reference Waveform data (*.h5)
[Save/Recall] > Recall > Recall:, Reference Waveform data (*.h5)
46 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Display Settings

You can adjust the intensity of displayed analog input channel waveforms to account for various signal characteristics, such as fast time/div settings and low trigger rates.
You can turn on waveform persistence, where the oscilloscope updates the display with new acquisitions, but does not immediately erase the results of previous acquisitions. All previous acquisitions are displayed with reduced intensity. New acquisitions are shown in their normal color with normal intensity.
Table 12 Display Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Waveform intensity (for analog input channels)
Persistence, single-shot infinite
Persistence, infinite
Persistence, variable [Display] > Persistence > Persistence, Variable Persistence, [Display] > Persistence > Time,
Clear persistence [Display] > Persistence > Clear Persistence
Clear display [Display] > Clear Display
Grid intensity
Grid type [Display] > Grid > Grid (Full, mV, IRE)
Waveform labels [Display] > Labels >
Label library reset [Utility] > Options > Preferences > Default Library
Annotations [Display] > Annotation >
[Intensity] (small round key just below Entry knob)
Increasing the intensity lets you see the maximum amount of noise and infrequently occurring events. Reducing the intensity can expose more detail in complex signals.
[Display] > Persistence > Persistence, Off, [Display] > Capture Waveforms
[Display] > Persistence > Persistence, Persistence
Entry knob
You can also configure the [Quick Action] key to clear the display. See "Configuring the [Quick
Action] Key" on page 92.
[Display] > Grid > Intensity, Entry knob
See also "To load a list of labels from a text file you create" on page 48.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 47
2 Quick Reference
Table 12 Display Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Freeze display You must configure the [Quick Action] key to freeze the display. See "Configuring the [Quick
Action] Key" on page 92.
Many activities, such as adjusting the trigger level, adjusting vertical or horizontal settings, or saving data will un-freeze the display.

To load a list of labels from a text file you create

It may be convenient to create a list of labels using a text editor, then load the label list into the oscilloscope. This lets you type on a keyboard rather than edit the label list using the oscilloscope's controls.
You can create a list of up to 75 labels and load it into the oscilloscope. Labels are added to the beginning of the list. If more than 75 labels are loaded, only the first 75 are stored.
To load labels from a text file into the oscilloscope: 1 Use a text editor to create each label. Each label can be up to ten characters in
length. Separate each label with a line feed.
2 Name the file labellist.txt and save it on a USB mass storage device such as a
thumb drive.
3 Load the list into the oscilloscope using the File Explorer (press [Utility] > File
Explorer).
48 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
NOTE
Label List Management
When you press the Library softkey, you will see a list of the last 75 labels used. The list does not save duplicate labels. Labels can end in any number of trailing digits. As long as the base string is the same as an existing label in the library, the new label will not be put in the library. For example, if label A0 is in the library and you make a new label called A12345, the new label is not added to the library.
When you save a new user-defined label, the new label will replace the oldest label in the list. Oldest is defined as the longest time since the label was last assigned to a channel. Any time you assign any label to a channel, that label will move to the newest in the list. Thus, after you use the label list for a while, your labels will predominate, making it easier to customize the instrument display for your needs.
When you reset the label library list, all of your custom labels will be deleted, and the label list will be returned to its factory configuration.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 49
2 Quick Reference

Triggers

Trigger Knobs and Keys

A trigger setup tells the oscilloscope when to acquire and display data. For example, you can set up to trigger on the rising edge of the analog channel 1 input signal.
You can use any input channel or the Ext Trig input BNC as the source for most trigger types (see "External Trigger Input" on page 53).
Changes to the trigger setup are applied immediately. If the oscilloscope is stopped when you change a trigger setup, the oscilloscope uses the new specification when you press [Run/Stop] or [Single]. If the oscilloscope is running when you change a trigger setup, it uses the new trigger definition when it starts the next acquisition.
You can save trigger setups along with the oscilloscope setup (see "Save/Recall
(Setups, Screens, Data)" on page 82).

Trigger Types

In addition to the edge trigger type, you can set up triggers on pulse widths, patterns, rising and falling edge transition times, setup and hold violations, and video signals.
50 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
Table 13 Trigger Type Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Trigger level Turn the trigger Level knob.
Also: [Analyze] > Features, Trigger Levels.
The edge trigger level for the Line source is not adjustable. This trigger is synchronized with the power line supplied to the oscilloscope.
Trigger type [Trigger] > Trigger Type (Edge, Pulse Width, Pattern, Rise/Fall Time, Setup and Hold, Video, Serial 1)
Edge trigger [Auto Scale] (sets up an Edge trigger)
[Trigger] > Trigger Type, Edge
Pulse width trigger [Trigger] > Trigger Type, Pulse Width
Video trigger [Trigger] > Trigger Type, Video
NOTE: Many video signals are produced from 75 sources, a 75 input.
Pattern trigger [Trigger] > Trigger Type, Pattern
Ω terminator (such as a Keysight 11094B) should be connected to the oscilloscope
Ω sources. To provide correct matching to these
Rise/fall edge transition time trigger
Setup and hold violation trigger
Serial bus trigger [Trigger] > Trigger Type, Serial 1
[Trigger] > Trigger Type, Rise/Fall Time
[Trigger] > Trigger Type, Setup and Hold
See "Serial Bus Decode/Trigger" on page 76.

Trigger Mode, Coupling, Reject, Holdoff

Noisy Signals If the signal you are probing is noisy, you can set up the oscilloscope to reduce the
noise in the trigger path and on the displayed waveform. First, stabilize the displayed waveform by removing the noise from the trigger path. Second, reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display. 2 Remove the noise from the trigger path by turning on high-frequency rejection,
low-frequency rejection, or noise reject.
3 Use "Selecting the Acquisition Mode" on page 54 to reduce noise on the
displayed waveform.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 51
2 Quick Reference
Table 14 Trigger Mode, Coupling, Reject, Holdoff Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Trigger mode [Trigger] > Mode
You can also configure the [Quick Action] key to toggle between the Auto and Normal trigger modes. See "Configuring the [Quick Action] Key" on page 92.
Auto trigger mode [Trigger] > Mode, Auto
If the specified trigger conditions are not found, triggers are forced and acquisitions are made so that signal activity is displayed on the oscilloscope. The Auto trigger mode is appropriate when:
Checking DC signals or signals with unknown levels or activity.
When trigger conditions occur often enough that forced triggers are unnecessary.
Normal trigger mode [Trigger] > Mode, Normal
Triggers and acquisitions only occur when the specified trigger conditions are found. The Normal trigger mode is appropriate when:
You only want to acquire specific events specified by the trigger settings.
Making single-shot acquisitions with the [Single] key.
Often with single-shot acquisitions, you must initiate some action in the device under test, and you do not want the oscilloscope to auto-trigger before that happens. Before initiating the action in the circuit, wait for the trigger condition indicator Trig 'd? to flash (this tells you the pre-trigger buffer is filled).
Force trigger [Force]
When in the Normal trigger mode and no triggers are occurring, you can force a trigger to acquire and display waveforms (which may show why triggers are not occurring).
Trigger coupling [Trigger] > Coupling (DC, AC, LF Reject, TV/Video)
NOTE: Trigger coupling is independent of channel coupling (see "Vertical Controls" on page 36).
Trigger noise reject
Trigger high frequency reject
Trigger holdoff [Trigger] > Holdoff
[Trigger] > Reject, Entry knob to select Noise Reject, push Entry knob to enable or disable
[Trigger] > Reject, Entry knob to select HF Reject, push Entry knob to enable or disable
The correct holdoff setting is typically slightly less than one repetition of the waveform.
52 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

External Trigger Input

CAUTION
The external trigger input can be used as a source in several of the trigger types. The external trigger BNC input is labeled Ext Trig.
Maximum voltage at oscilloscope external trigger input
30 Vrms
The external trigger input impedance is 1M Ohm. This lets you use passive probes for general-purpose measurements. The higher impedance minimizes the loading effect of the oscilloscope on the device under test.
Table 15 External Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
External trigger units [Trigger] > External > Units (Volts, Amps)
Quick Reference 2
External trigger attenuation
External trigger threshold
External trigger range
[Trigger] > External > Probe, Ratio/Decibels, Entry knob
[Trigger] > External > Threshold, Entry knob
[Trigger] > External > Range, Entry knob
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 53
2 Quick Reference

Acquisition Control

This section shows how to use the oscilloscope's acquisition controls.

Selecting the Acquisition Mode

When selecting the oscilloscope acquisition mode, keep in mind that samples are normally decimated (thrown away) at slower time/div settings.
At slower time/div settings, the effective sample rate drops (and the effective sample period increases) because the acquisition time increases and the oscilloscope's digitizer is sampling faster than is required to fill memory.
For example, suppose an oscilloscope's digitizer has a sample period of 1 ns (maximum sample rate of 1 GSa/s) and a 1 M memory depth. At that rate, memory is filled in 1 ms. If the acquisition time is 100 ms (10 ms/div), only 1 of every 100 samples is needed to fill memory.
Table 16 Acquisition Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Acquisition mode [Acquire] > Acq Mode
Normal acquisition mode
Peak detect acquisition mode
Averaging acquisition mode
High resolution acquisition mode
[Acquire] > Acq Mode, Normal
At slower time/div settings, normal decimation occurs, and there is no averaging. Use this mode for most waveforms.
[Acquire] > Acq Mode, Peak Detect
At slower time/div settings when decimation would normally occur, the maximum and minimum samples in the effective sample period are stored. Use this mode for displaying narrow pulses that occur infrequently.
[Acquire] > Acq Mode, Averaging, [Acquire] > # Avgs
At all time/div settings, the specified number of triggers are averaged together. Use this mode for reducing noise and increasing resolution of periodic signals without bandwidth or rise time degradation.
[Acquire] > Acq Mode, High Resolution
At slower time/div settings, all samples in the effective sample period are averaged and the average value is stored. Use this mode for reducing random noise.
54 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Table 17 Segmented Memory Acquisition Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Segmented memory acquisitions
Segmented memory navigation
Segmented memory and persistence
Segmented memory, save to USB storage device

Overview of Sampling

[Acquire] > Segmented > Segmented, # of Segs, [Run] or [Single]
After each segment fills, the oscilloscope re-arms and is ready to trigger in about 19 µs. Remember though, for example, if the horizontal time per division control is set to 5 µs/div, and the Time Reference is set to Center, it will take at least 50 µs to fill all ten divisions and re-arm. (That is 25 µs to capture pre-trigger data and 25 µs to capture post-trigger data.)
[Acquire] > Segmented > Current Seg
[Display] > Persistence, Infinite
[Acquire] > Segmented > Analyze Segments
[Save/Recall] > Save > Format (CSV, ASCII XY, or BIN) > Settings > Save Seg (Current, All)
Persistence or Variable Persistence
To understand the oscilloscope's sampling and acquisition modes, it is helpful to understand sampling theory, aliasing, oscilloscope bandwidth and sample rate, oscilloscope rise time, oscilloscope bandwidth required, and how memory depth affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth (band-limited) signal with maximum frequency f
must be greater than twice the maximum frequency f signal be uniquely reconstructed without aliasing.
, the equally spaced sampling frequency fS
MAX
, in order to have the
MAX
f
= fS/2 = Nyquist frequency (fN) = folding frequency
MAX
Aliasing
Aliasing occurs when signals are under-sampled (fS < 2f distortion caused by low frequencies falsely reconstructed from an insufficient
number of sample points.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 55
). Aliasing is the signal
MAX
2 Quick Reference
Figure 3 Aliasing
Oscilloscope Bandwidth and Sample Rate
56 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
An oscilloscope's bandwidth is typically described as the lowest frequency at which input signal sine waves are attenuated by 3 dB (-30% amplitude error).
At the oscilloscope bandwidth, sampling theory says the required sample rate is f
. However, the theory assumes there are no frequency components above
= 2f
BW
(fBW in this case) and it requires a system with an ideal brick-wall frequency
f
MAX
response.
S
Quick Reference 2
[
H
[
C
"(Y7
6iiZcjVi^dc
;gZfjZcXn
%Y7
Figure 4 Theoretical Brick-Wall Frequency Response
However, digital signals have frequency components above the fundamental frequency (square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics), and typically, for 500 MHz bandwidths and below, oscilloscopes have a Gaussian frequency response.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 57
2 Quick Reference
Limiting oscilloscope bandwidth (fBW) to 1/4 the sample rate (fS/4) reduces frequency components above the Nyquist frequency (f
N).
[
H
[
C
[H$)
"(Y7
6iiZcjVi^dc
6a^VhZY[gZfjZcXn
XdbedcZcih
;gZfjZcXn
%Y7
Figure 5 Sample Rate and Oscilloscope Bandwidth
So, in practice, an oscilloscope's sample rate should be four or more times its bandwidth: f
= 4fBW. This way, there is less aliasing, and aliased frequency
S
components have a greater amount of attenuation.
See Also Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to Make the
Most Accurate Digital Measurements, Keysight Application Note 1587
(http://literature.cdn.keysight.com/litweb/pdf/5989-5732EN.pdf)
Oscilloscope Rise Time
Closely related to an oscilloscope's bandwidth specification is its rise time specification. Oscilloscopes with a Gaussian-type frequency response have an approximate rise time of 0.35/f
An oscilloscope's rise time is not the fastest edge speed that the oscilloscope can accurately measure. It is the fastest edge speed the oscilloscope can possibly produce.
58 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
based on a 10% to 90% criterion.
BW
Quick Reference 2
Oscilloscope Bandwidth Required
The oscilloscope bandwidth required to accurately measure a signal is primarily determined by the signal's rise time, not the signal's frequency. You can use these steps to calculate the oscilloscope bandwidth required:
1 Determine the fastest edge speeds.
You can usually obtain rise time information from published specifications for devices used in your designs.
2 Compute the maximum "practical" frequency component.
From Dr. Howard W. Johnson's book, High-Speed Digital Design – A Handbook of Black Magic, all fast edges have an infinite spectrum of frequency components. However, there is an inflection (or "knee") in the frequency spectrum of fast edges where frequency components higher than f
insignificant in determining the shape of the signal.
= 0.5 / signal rise time (based on 10% - 90% thresholds)
f
knee
= 0.4 / signal rise time (based on 20% - 80% thresholds)
f
knee
3 Use a multiplication factor for the required accuracy to determine the
oscilloscope bandwidth required.
knee
are
Required accuracy Oscilloscope bandwidth required
20% fBW = 1.0 x f
10% fBW = 1.3 x f
3% fBW = 1.9 x f
knee
knee
knee
See Also Choosing an Oscilloscope with the Right Bandwidth for your Application, Keysight
Application Note 1588 (http://literature.cdn.keysight.com/litweb/pdf/5989-5733EN.pdf)
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed, and there is a maximum sample rate associated with oscilloscope's analog-to-digital converter; however, the actual sample rate is determined by the time of the acquisition (which is set according to the oscilloscope's horizontal time/div scale).
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 59
2 Quick Reference
sample rate = number of samples / time of acquisition
For example, when storing 50 µs of data in 50,000 points of memory, the actual sample rate is 1 GSa/s.
Likewise, when storing 50 ms of data in 50,000 points of memory, the actual sample rate is 1 MSa/s.
The actual sample rate is displayed in the right-side information area.
The oscilloscope achieves the actual sample rate by throwing away (decimating) unneeded samples.
60 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Cursors

X Cursors X cursors are vertical dashed lines that adjust horizontally and can be used to
Y Cursors Y cursors are horizontal dashed lines that adjust vertically and can be used to
Quick Reference 2
Cursors are horizontal and vertical markers that indicate X-axis values and Y-axis values on a selected waveform source. You can use cursors to make custom voltage, time, phase, or ratio measurements on oscilloscope signals.
Cursor information is displayed at the bottom of the screen.
measure time (s), frequency (1/s), phase (°), and ratio (%).
When used with the FFT math function as a source, the X cursors indicate frequency.
In XY horizontal mode, the X cursors display channel 1 values (Volts or Amps).
measure Volts or Amps, dependent on the channel Probe Units setting, or they can measure ratios (%). When math functions are used as a source, the measurement units correspond to that math function.
The Y cursors adjust vertically and typically indicate values relative to the waveform's ground point, except math FFT where the values are relative to 0 dB.
In XY horizontal mode, the Y cursors display channel 2 values (Volts or Amps).

Cursor Knobs and Keys

Cursor Softkey Controls

Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 61
2 Quick Reference
Table 18 Cursor Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Cursors mode [Cursors] > Mode
Manual cursors mode [Cursors] > Mode, Manual (and use Cursors knob to select and adjust)
Track Waveform cursors mode
Measure cursors mode
Binary cursors mode [Cursors] > Mode, Binary
Hex cursors mode [Cursors] > Mode, Hex
Cursors X units [Cursors] > Units > X Units (Seconds, Hz, Phase, Ratio)
Cursors Y units [Cursors] > Units > Y Units (Base, Ratio)
[Cursors] > Mode, Track Waveform
[Meas] (cursors show locations used for most recently added measurement)
62 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Measurements

NOTE
NOTE
The [Meas] key lets you make automatic measurements on waveforms. Some measurements can be made only on analog input channels.
If a portion of the waveform required for a measurement is not displayed or does not display enough resolution to make the measurement (approximately 4% of full scale), the result will be displayed as greater than a value, less than a value, not enough edges, not enough amplitude (low signal), or waveform is clipped.
The results of the most recent measurements are displayed in the Measurements information area at the bottom of the screen.
Cursors are turned on to show the portion of the waveform being measured for the most recently selected measurement (right-most in the bottom measurement area).
Post Acquisition Processing
Quick Reference 2
In addition to changing display parameters after the acquisition, you can perform all of the measurements and math functions after the acquisition. Measurements and math functions will be recalculated as you pan and zoom and turn channels on and off. As you zoom in and out on a signal using the horizontal scale knob and vertical volts/division knob, you affect the resolution of the display. Because measurements and math functions are performed on displayed data, you affect the resolution of functions and measurements.
The units of math waveforms are described in "Units for Math Waveforms" on page 45.
All measurements are available for analog channel waveforms. All measurements except Counter are available for math waveforms other than FFT. A limited set of measurements is available for math FFT waveforms. Use the cursors to make other measurements on FFT.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 63
2 Quick Reference
Table 19 Measurement Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Measurement source [Meas] > Source
Displayed channels or math functions are available for measurements. The External source is available only for the Counter measurement when the trigger type is Edge.
Measurement type [Meas] > Type:
Snapshot all measurements
Voltage measurements
Time measurements [Meas] > Type: (Period, Frequency, Counter, + Width, – Width, Bit Rate, + Duty Cycle, - Duty Cycle,
Count measurements [Meas] > Type: (Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count),
Measurement thresholds
Measurement window [Meas] > Settings > Meas Window (Auto Select, Main, Zoom)
Clear measurements [Meas] > Clear Meas >
[Meas] > Type: Snapshot All, Add Measurement
You can also configure the [Quick Action] key to display the Snapshot All popup. See "Configuring
the [Quick Action] Key" on page 92.
[Meas] > Type: (Peak-Peak, Maximum, Minimum, Amplitude, Top, Base, Overshoot, Preshoot, Average, DC RMS, AC RMS), Add Measurement
Rise Time, Fall Time, Delay, Phase, X at Min Y, X at Max Y), Add Measurement
The Counter measurement is available when the edge or pulse width trigger mode is selected and the measurement source is the same as the trigger source.
Add Measurement
[Meas] > Settings > Thresholds >
Also: [Analyze] > Features, Measurement Thresholds.
64 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Mask Testing

One way to verify a waveform's compliance to a particular set of parameters is to use mask testing. A mask defines a region of the oscilloscope's display in which the waveform must remain in order to comply with chosen parameters. Compliance to the mask is verified point-by-point across the display. Mask test operates on displayed analog channels; it does not operate on channels that are not displayed.
Table 20 Mask Testing Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Mask testing enable/disable
Mask statistics [Analyze] > Statistics >
Mask test run until [Analyze] > Setup > Run Until (Forever, Minimum # of Tests, Minimum Time, Minimum Sigma)
Mask test error action [Analyze] > Setup > On Error (Stop, Save, Print, Measure)
Mask test source lock [Analyze] > Setup > Source Lock
Auto mask creation [Analyze] > Automask >
Clear mask [Analyze] > Clear Mask
Mask files, save/recall from USB storage device
[Analyze] > Features, Entry knob to select Mask Test, push Entry knob to enable or disable
[Save/Recall] > Save > Format, Mask (*.msk)
[Save/Recall] > Recall > Recall:, Mask (*.msk)

Creating/Editing Mask Files

A mask file contains the following sections:
• Mask File Identifier.
•Mask Title.
• Mask Violation Regions.
• Oscilloscope Setup Information.
Mask File Identifier The Mask File Identifier is MASK_FILE_548XX.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 65
2 Quick Reference
Mask Title The Mask Title is a string of ASCII characters. Example: autoMask CH1 OCT 03
Mask Violation
Regions
09:40:26 2008
When a mask file contains the keyword "autoMask" in the title, the edge of the mask is passing by definition. Otherwise, the edge of the mask is defined as a failure.
Region 1
Region 2
Up to 8 regions can be defined for a mask. They can be numbered 1-8. They can appear in any order in the .msk file. The numbering of the regions must go from top to bottom, left to right.
An Automask file contains two special regions: the region "glued" to the top of the display, and the region that is "glued" to the bottom. The top region is indicated by y-values of "MAX" for the first and last points. The bottom region is indicated by y-values of "MIN" for the first and last points.
The top region must be the lowest numbered region in the file. The bottom region must be the highest numbered region in the file.
Region number 1 is the top mask region. The vertices in Region 1 describe points along a line; that line is the bottom edge of the top portion of the mask.
Similarly, the vertices in Region 2 describe the line that forms the top of the bottom part of the mask.
66 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
The vertices in a mask file are normalized. There are four parameters that define how values are normalized:
•X1
ΔX
•Y1
•Y2
These four parameters are defined in the Oscilloscope Setup portion of the mask file.
The Y-values (normally voltage) are normalized in the file using the following equation:
Y
= (Y - Y1)/ΔY
norm
where ΔY = Y2 - Y1
To convert the normalized Y-values in the mask file to voltage:
Oscilloscope
Setup Information
Y = (Y
norm
* ΔY) + Y1
where ΔY = Y2 - Y1
The X-values (normally time) are normalized in the file using the following equation:
X
= (X - X1)/ΔX
norm
To convert the normalized X-values to time:
X = (X
norm
* ΔX) + X1
The keywords "setup" and "end_setup" (appearing alone on a line) define the beginning and end of the oscilloscope setup region of the mask file. The oscilloscope setup information contains remote programming language commands that the oscilloscope executes when the mask file is loaded.
Any legal remote programming command can be entered in this section.
The mask scaling controls how the normalized vectors are interpreted. This in turn controls how the mask is drawn on the display. The remote programming commands that control mask scaling are:
:MTES:SCAL:BIND 0 :MTES:SCAL:X1 -400.000E-06 :MTES:SCAL:XDEL +800.000E-06
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 67
2 Quick Reference
:MTES:SCAL:Y1 +359.000E-03 :MTES:SCAL:Y2 +2.35900E+00
Building a Mask File
The following display shows a mask that uses all eight regions.
This mask is created by recalling the following mask file:
MASK_FILE_548XX
"All Regions"
/* Region Number */ 1 /* Number of vertices */ 4
-12.50, MAX
-10.00, 1.750
10.00, 1.750
12.50, MAX
/* Region Number */ 2 /* Number of vertices */ 5
-10.00, 1.000
-12.50, 0.500
-15.00, 0.500
-15.00, 1.500
-12.50, 1.500
/* Region Number */ 3 /* Number of vertices */ 6
-05.00, 1.000
-02.50, 0.500
68 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
02.50, 0.500
05.00, 1.000
02.50, 1.500
-02.50, 1.500
/* Region Number */ 4 /* Number of vertices */ 5
10.00, 1.000
12.50, 0.500
15.00, 0.500
15.00, 1.500
12.50, 1.500
/* Region Number */ 5 /* Number of vertices */ 5
-10.00, -1.000
-12.50, -0.500
-15.00, -0.500
-15.00, -1.500
-12.50, -1.500
/* Region Number */ 6 /* Number of vertices */ 6
-05.00, -1.000
-02.50, -0.500
02.50, -0.500
05.00, -1.000
02.50, -1.500
-02.50, -1.500
/* Region Number */ 7 /* Number of vertices */ 5
10.00, -1.000
12.50, -0.500
15.00, -0.500
15.00, -1.500
12.50, -1.500
/* Region Number */ 8 /* Number of vertices */ 4
-12.50, MIN
-10.00, -1.750
10.00, -1.750
12.50, MIN
setup :CHANnel1:RANGe +8.00E+00 :CHANnel1:OFFSet +2.0E+00 :CHANnel1:DISPlay 1 :TIMebase:MODE MAIN :TIMebase:REFerence CENTer :TIMebase:RANGe +50.00E-09 :TIMebase:POSition +10.0E-09 :MTESt:SOURce CHANnel1 :MTESt:ENABle 1 :MTESt:LOCK 1 :MTESt:SCALe:X1 +10.0E-09 :MTESt:SCALe:XDELta +1.0000E-09 :MTESt:SCALe:Y1 +2.0E+00 :MTESt:SCALe:Y2 +4.00000E+00 end_setup
Quick Reference 2
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 69
2 Quick Reference
NOTE
In a mask file, all region definitions need to be separated by a blank line.
Mask regions are defined by a number of (x,y) coordinate vertices (as on an ordinary x,y graph). A "y" value of "MAX" specifies the top of the graticule, and a "y" value of "MIN" specifies the bottom of the graticule.
If there are more than 1000 vertices in a mask region, only the first 1000 vertices will be processed.
The mask x,y graph is related to the oscilloscope graticule using the :MTESt:SCALe setup commands.
The oscilloscope's graticule has a time reference location (at the left, center, or right of the screen) and a trigger (t=0) position/delay value relative to the reference. The graticule also has a vertical ground 0 V reference (offset relative to the center of the screen) location.
The X1 and Y1 setup commands relate the mask region's x,y graph origin to the oscilloscope graticule's t=0 and V=0 reference locations, and the XDELta and Y2 setup commands specify the size of the graph's x and y units.
• The X1 setup command specifies the time location of the x,y graph's x origin.
• The Y1 setup command specifies the vertical location of the x,y graph's y origin.
• The XDELta setup command specifies the amount of time associated with each x unit.
• The Y2 setup command is the vertical location of the x,y graph's y=1 value (so in effect, Y2 – Y1 is the YDELta value).
For example:
• With a graticule whose trigger position is 10 ns (before a center screen reference) and whose ground reference (offset) is 2 V below the center of the screen, to place the mask region's x,y graph's origin at center screen, you would set X1 = 10 ns and Y1 = 2 V.
• If the XDELta parameter is set to 5 ns and Y2 is set to 4 V, a mask region whose vertices are (-1, 1), (1, 1), (1, -1), and (-1, -1) goes from 5 ns to 15 ns and from 0V to 4V.
• If you move the mask region's x,y graph origin to the t=0 and V=0 location by setting X1 = 0 and Y1 = 0, the same vertices define a region that goes from
-5 ns to 5 ns and from -2 V to 2 V.
70 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
NOTE
Although a mask can have up to 8 regions, in any given vertical column, it is only possible to define 4 regions. When there are 4 regions in a vertical column, one region must be tied to the top (using the MAX y value) and one must be tied to the bottom (using the MIN y value).
How is mask testing done? InfiniiVision oscilloscopes start mask testing by
creating a database that is 200 x 640 for the waveform viewing area. Each location in the array is designated as either a violation or a pass area. Each time a data point from a waveform occurs in a violation area a failure is logged. If Test All was selected, every active analog channel is tested against the mask database for each acquisition. Over 2 billion failures can be logged per-channel. The number of acquisitions tested is also logged and displayed as "# of Tests".
The mask file allows greater resolution than the 200 X 640 database. Some quantization of data occurs to reduce the mask file data for display on-screen.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 71
2 Quick Reference

Digital Voltmeter

The Digital Voltmeter (DVM) analysis feature provides 3-digit voltage and 5-digit frequency measurements using any analog channel. DVM measurements are asynchronous from the oscilloscope's acquisition system and are always acquiring.
Table 21 Digital Voltmeter Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Digital voltmeter enable/disable
Digital voltmeter mode
Digital voltmeter display, transparent background
Auto range enable/disable
[Analyze] > Features, Entry knob to select Digital Voltmeter, push Entry knob to enable or disable
[Analyze] > Mode (DC, DC RMS, AC RMS, Frequency)
The Frequency mode requires the Edge or Pulse Width trigger type, and the DVM source and the trigger source must be the same analog channel.
[Analyze] > Transparent
[Analyze] > Auto Range
Auto range can be used when the DVM input channel is not used in oscilloscope triggering.
72 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Frequency Response Analysis

On G-suffix oscilloscope models (that have a a built-in waveform generator), the Frequency Response Analysis (FRA) feature controls the built-in waveform generator to sweep a sine wave across a range of frequencies while measuring the input to and output from a device under test (DUT). At each frequency, gain (A) and phase are measured and plotted on a frequency response Bode chart.
When the frequency response analysis completes, you can move a marker across the chart to see the measured gain and phase values at each frequency point. You can also adjust the chart's scale and offset settings for the gain and phase plots.
Table 22 Frequency Response Analysis Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Frequency Response Analysis (FRA) enable/disable
Channels probing DUT input V and output V
Frequency sweep min and max values
Waveform generator amplitude and expected output load
Points per decade [Analyze] > Setup > Pts Per Decade (10, 20, 30, 40, 50)
Run the analysis [Analyze] > Run Analysis
Adjust scale and offset of Bode plots
Gain, Phase, or both plots in chart
Autoscale gain and phase plots
[Analyze] > Features, Entry knob to select Frequency Response Analysis, push Entry knob to enable or disable
[Analyze] > Setup > Sources > Input V
[Analyze] > Setup > Sources > Output V
[Analyze] > Setup > Min/Max Freq, Entry knob
[Analyze] > Setup > Amplitude, Entry knob
[Analyze] > Setup > Output Load (50
Frequency Response Analysis automatically changes the acquisition mode during the sweep to provide a good balance between a clean signal and high throughput at each frequency. For most frequencies, averaging is enabled with a varying number of averages.
[Analyze] > Chart > Gain/Phase Scale/Offset, Entry knob
[Analyze] > Chart > Trace (Gain, Phase, both)
[Analyze] > Chart > Autoscale
Ω, High-Z)
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 73
2 Quick Reference
Table 22 Frequency Response Analysis Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
View measured gain and phase values
Frequency response analysis display, transparent background
Save data to USB storage device
[Analyze] > Move Marker, Entry knob
[Analyze] > Transparent
[Save/Recall] > Save > Format, Frequency Response Analysis data (*.csv)
74 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Waveform Generator

On G-suffix oscilloscope models, a waveform generator is built into the oscilloscope. The waveform generator gives you an easy way to provide input signals when testing circuitry with the oscilloscope.
Waveform generator settings can be saved and recalled with oscilloscope setups. See "Save/Recall (Setups, Screens, Data)" on page 82.
Table 23 Waveform Generator Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Waveform generator waveform type
Waveform generator expected output load
Waveform generator logic presets
Invert shape of waveform
Add noise to output [Wave Gen] > Settings > Add Noise
Modulation, enable/disable
AM modulated output [Wave Gen] > Settings > Modulation > Type, Amplitude Modulation (AM)
FM modulated output [Wave Gen] > Settings > Modulation > Type, Frequency Modulation (FM)
FSK modulated output [Wave Gen] > Settings > Modulation > Type, Frequency-Shift Keying Modulation (FSK)
Restore waveform generator defaults
[Wave Gen] > Waveform (Sine, Square, Ramp, Pulse, DC, Noise)
[Wave Gen] > Settings > Output Load (50
[Wave Gen] > Settings > Logic Presets > (TTL, CMOS 5.0V, CMOS 3.3V, CMOS 2.5V, ECL)
[Wave Gen] > Settings > Invert Output
[Wave Gen] > Settings > Modulation > Modulation
[Wave Gen] > Settings > Default Wave Gen
Ω, High-Z)
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 75
2 Quick Reference

Serial Bus Decode/Trigger

Depending on the oscilloscope model, the following hardware-accelerated serial decode and trigger options are available:
Serial decode and trigger type: With license:
CAN (Controller Area Network) AUTO
I2C (Inter-IC) EMBD
LIN (Local Interconnect Network) AUTO
SPI (Serial Peripheral Interface) EMBD
UART (Universal Asynchronous Receiver/Transmitter) protocols including RS232 (Recommended Standard 232)
To determine whether these licenses are installed on your oscilloscope, press [Help] > About Oscilloscope.
Triggering on
Serial Data
To order serial decode licenses, go to www.keysight.com and search for the product number (for example, DSOX1AUTO) or contact your local Keysight Technologies representative (see www.keysight.com/find/contactus).
When triggering on a slow serial signal (for example, I2C, SPI, CAN, LIN, etc.) it may be necessary to switch from the Auto trigger mode to the Normal trigger mode to prevent the oscilloscope from Auto-triggering and stabilize the display. You can select the trigger mode by pressing the [Trigger] key, then the Mode softkey.
Also, the threshold voltage level must be set appropriately for each source channel. The threshold level for each serial signal can be set in the Signals menu (press [Analyze] > Features, select Serial Bus, then press the Signals softkey).
EMBD
76 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Table 24 Serial Bus Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
Serial bus select, enable/disable
Serial bus mode
Serial bus trigger [Trigger] > Trigger Type, Serial 1
[Analyze] > Features, Entry knob to select Serial Bus, push Features softkey or Entry knob to enable or disable
[Analyze] > Mode, Entry knob (CAN, I2C, LIN, SPI, UART/RS232)
The built-in help for the Mode softkey describes the decode waveforms.

CAN Decode/Trigger

The CAN serial decode and triggering option can be enabled with the AUTO license.
When interpreting the decode waveforms, see the Mode softkey built-in help.
Table 25 CAN Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
CAN signal setup After selecting the Serial Bus and the CAN serial bus mode, press [Analyze] > Signals > to open the
CAN Signals menu. In this menu, you can select the oscilloscope source channel probing the signal and the appropriate threshold voltage to use when decoding/triggering the signal, as well as other signal options.
CAN baud rate
[Analyze] > Signals > Baud, Entry knob
CAN sample point
CAN signal type/polarity
CAN counters [Analyze] > Reset CAN Counters
[Analyze] > Signals > Sample Point, Entry knob
[Analyze] > Signals > Signal, Entry knob (Rx, Tx, CAN_H, CAN_L, Differential (L-H), Differential
(H-L))
Counters run even when the oscilloscope is stopped (not acquiring data).
When an overflow condition occurs, the counter displays OVERFLOW.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 77
2 Quick Reference
Table 25 CAN Decode/Trigger Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
CAN trigger [Trigger] > Trigger Type, Serial 1 (CAN)
[Trigger] > Trigger on: (SOF - Start of Frame, Remote Frame ID (RTR), Data Frame ID (~RTR), Remote
or Data Frame ID, Data Frame ID and Data, Error Frame, All Errors, Acknowledge Error, Overload Frame)
For triggers where you can specify frame ID or data values, press [Trigger] > Bits > to open the CAN Bits menu where you can enter the values.

I2C Decode/Trigger

On all 1000 X-Series oscilloscopes, the I2C serial decode and triggering option can be enabled with the EMBD license.
When interpreting the decode waveforms, see the Mode softkey built-in help.
Table 26 I2C Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
I2C signal setup After selecting the Serial Bus and the I2C serial bus mode, press [Analyze] > Signals > to open the
I2C Signals menu. In this menu, you can select the oscilloscope source channels probing the serial clock and serial data signals. You can also specify the appropriate threshold voltages to use when decoding and triggering on the signals.
I2C address size [Analyze] > Addr Size (7 Bit, 8 Bit)
I2C trigger [Trigger] > Trigger Type, Serial 1 (I2C)
[Trigger] > Trigger on: (Start Condition, Stop Condition, Missing Acknowledge, Address with no Ack, Restart, EEPROM Data Read, Frame (Start: Addr7: Read: Ack: Data), Frame (Start: Addr7: Write: Ack: Data), Frame (Start: Addr7: Read: Ack: Data: Ack: Data2), Frame (Start: Addr7: Write: Ack: Data: Ack: Data2), 10-bit Write)
For triggers where you can specify address or data values, there are additional softkeys you can use to enter the values.

LIN Decode/Trigger

The LIN serial decode and triggering option can be enabled with the AUTO license.
When interpreting the decode waveforms, see the Mode softkey built-in help.
78 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
Table 27 LIN Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
LIN signal setup After selecting the Serial Bus and the LIN serial bus mode, press [Analyze] > Signals > to open the
LIN Signals menu. In this menu, you can select the oscilloscope source channel probing the signal and the appropriate threshold voltage to use when decoding/triggering the signal, as well as other signal options.
LIN baud rate
LIN sample point
LIN standard
LIN sync break
LIN show parity [Analyze] > Show Parity
LIN trigger [Trigger] > Trigger Type, Serial 1 (LIN)
[Analyze] > Signals > Baud > Baud, Entry knob
[Analyze] > Signals > Sample Point, Entry knob
[Analyze] > Signals > Standard, Entry knob (LIN 1.3, LIN 2.X)
[Analyze] > Signals > Sync Break, Entry knob (>= 11, >= 12, >= 13)
[Trigger] > Trigger on: (Sync - Sync Break, ID - Frame ID, ID & Data - Frame ID and Data, Parity Error,
Checksum Error)
For triggers where you can specify frame ID or data values, there are additional softkeys you can use to enter the values.

SPI Decode/Trigger

The SPI serial decode and triggering option can be enabled with the EMBD license.
When interpreting the decode waveforms, see the Mode softkey built-in help.
Table 28 SPI Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
SPI signal setup After selecting the Serial Bus and the SPI serial bus mode, press [Analyze] > Signals > to open the
SPI Signals menu. In this menu, there are separate softkeys and submenus for specifying the Clock, MOSI/MISO, and CS (chip select) signal oscilloscope source channels, threshold voltages, and edge for the Clock source.
The Display Info softkey lets you display or hide signal setup and timing diagram information.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 79
2 Quick Reference
Table 28 SPI Decode/Trigger Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
SPI framing
SPI word size [Analyze] > Word Size
SPI bit order [Analyze] > Bit Order (MSB, LSB)
SPI trigger [Trigger] > Trigger Type, Serial 1 (SPI)
[Analyze] > Signals > CS > Frame by, Entry knob (~CS - Not Chip Select, CS - Chip Select, Clock Timeout)
[Trigger] > Trigger Setup > Trigger Type (Master-Out, Slave-In (MOSI) Data)
The additional softkeys let you specify the number of data bits and the value of each bit.

UART/RS232 Decode/Trigger

On all 1000 X-Series oscilloscopes, the UART/RS232 serial decode and triggering option can be enabled with the EMBD license.
When interpreting the decode waveforms, see the Mode softkey built-in help.
Table 29 UART/RS232 Decode/Trigger Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
UART/RS232 signal setup
After selecting the Serial Bus and the UART/RS232 serial bus mode, press [Analyze] > Signals > to open the UART Signals menu. In this menu, you can select the oscilloscope source channels probing the Rx and Tx signals and the appropriate threshold voltage for each.
UART/RS232 bus configuration
UART/RS232 number of bits
UART/RS232 parity
UART/RS232 baud rate
UART/RS232 bus polarity
After selecting the Serial Bus and the UART/RS232 serial bus mode, press [Analyze] > Bus Config > to open the UART Bus Config menu. In this menu, you can select the oscilloscope source channels probing the Rx and Tx signals and the appropriate threshold voltage for each.
[Analyze] > Bus Config > # Bits, Entry knob (5, 6, 7, 8, 9)
[Analyze] > Bus Config > Parity, Entry knob (Even, Odd, None)
[Analyze] > Bus Config > Baud Rate > Baud, Entry knob
[Analyze] > Bus Config > Polarity, Entry knob (Idle low, Idle high)
80 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Table 29 UART/RS232 Decode/Trigger Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick Reference 2
UART/RS232 bit order
UART/RS232 display base
UART/RS232 framing
UART/RS232 counters [Analyze] > Settings > Reset UART Counters
UART/RS232 trigger [Trigger] > Trigger Type, Serial 1 (UART/RS232)
[Analyze] > Bus Config > Bit Order, Entry knob (LSB, MSB)
[Analyze] > Settings > Base, Entry knob (Hex, Binary, ASCII)
[Analyze] > Settings > Framing, Entry knob (Off, 8-bit hex value)
Counters run even when the oscilloscope is stopped (not acquiring data).
When an overflow condition occurs, the counter displays OVERFLOW.
[Trigger] > Trigger Setup > Trigger (Rx Start Bit, Rx Stop Bit, Rx Data, Rx 1:Data, Rx 0:Data, Rx X:Data, Tx Start Bit, Tx Stop Bit, Tx Data, Tx 1:Data, Tx 0:Data, Tx X:Data, Rx or Tx Parity Error)
For triggers where you can specify data values, there are additional softkeys for specifying the data comparison operator, the data value, the data value base (Hex or ASCII), and the Nth frame burst count.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 81
2 Quick Reference

Save/Recall (Setups, Screens, Data)

Oscilloscope setups, reference waveforms, and mask files can be saved to internal oscilloscope memory or to a USB storage device and recalled later. You can also recall default or factory default setups.
Oscilloscope screen images can be saved to a USB storage device in BMP or PNG formats.
Acquired waveform data can be saved to a USB storage device in comma-separated value (CSV), ASCII XY, and binary (BIN) formats.
There is also a command to securely erase all the oscilloscope's non-volatile internal memory.
Table 30 Save/Recall Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Save setup files, screen images, waveform data, mask files
Save setup files [Save/Recall] > Save > Format, Setup (*.scp)
Save screen images [Save/Recall] > Save > Format,
[Save/Recall] > Save >
You can also configure the [Quick Action] key to save setups, screen images, or data. See
"Configuring the [Quick Action] Key" on page 92.
Press to Save
• 8-bit Bitmap image (*.bmp)
• 24-bit Bitmap image (*.bmp)
• 24-bit image (*.png)
Settings >
•Setup Info
• Invert Grat
•Palette (Color, Grayscale)
82 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Table 30 Save/Recall Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Save waveform data [Save/Recall] > Save > Format,
•CSV data (*.csv)
• ASCII XY data (*.csv)
• Multi Channel Waveform data (*.h5)
• Binary data (*.bin)
Settings > Length (to select number of data points to be saved, see "Length Control" on page 83)
When saving waveform data, the save times depend on the chosen format: BIN=fastest, ASCII XY=medium, CSV=slowest.
Quick Reference 2
Quick save to USB storage device
Recall setups, mask files, or reference waveforms
Recall setup files [Save/Recall] > Recall > Recall:, Setup (*.scp)
Factory default setup [Save/Recall] > Default/Erase > Factory Default
Secure erase [Save/Recall] > Default/Erase > Secure Erase
[Save to USB] (once "save to USB" settings have been configured under [Save/Recall] > Save)
[Save/Recall] > Recall > Recall:

Length Control

The Length control is available when saving data to CSV, ASCII XY, or BIN format files. It sets the number of data points that will be output to the file. Only displayed data points are saved.
The maximum number of data points depends on these things:
• Whether acquisitions are running. When stopped, data comes from the raw acquisition record. When running, data comes from the smaller measurement record.
• Whether the oscilloscope was stopped using [Stop] or [Single]. Running acquisitions split memory to provide fast waveform update rates. Single acquisitions use full memory.
• Whether only one channel of a pair is turned on. (Channels 1 and 2 are one pair, channels 3 and 4 are the other.) Acquisition memory is divided among the channels in a pair.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 83
2 Quick Reference
• Whether reference waveforms are on. Displayed reference waveforms consume acquisition memory.
• Whether segmented memory is on. Acquisition memory is divided by the number of segments.
• The horizontal time/div (sweep speed) setting. At faster settings, fewer data points appear on the display.
• When saving to a CSV format file, the maximum number of data points is 50,000.
When necessary, the Length control performs a "1 of n" decimation of the data . For example: if the Length is set to 1000, and you are displaying a record that is 5000 data points in length, four of each five data points will be decimated, creating an output file 1000 data points in length.
84 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Print (Screens)

You can print the complete display, including the status line and softkey labels, to a USB printer or a network printer.
To set up a printer:
1 Connect a printer. You can:
2 To open the Print Config menu:
Quick Reference 2
Connect a USB printer to the rectangular USB host port on the front panel.
For the most up-to-date information on printers that are compatible with the InfiniiVision oscilloscopes, please visit
www.keysight.com/find/InfiniiVision-printers.
Set up a network printer connection. See "To set up network printer
connections" on page 86.
Press [Save/Recall] > Print.
Select the Quick Print quick action ([Utility] > Quick Action > Action, Quick Print);
then, press Settings.
Some softkeys in the Print Config menu are ghosted (not available) until a printer is connected.
Table 31 Print Config Menu Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Printer selection Print to, (printer)
Print options Options,
• Setup Information
• Invert Graticule Colors
• Form Feed
• Landscape
Color or grayscale printing
Print the current screen
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 85
Palette, (Color, Grayscale)
Press to Print
2 Quick Reference

To set up network printer connections

To print the current screen once the printer is configured (and Quick Print is selected as the quick action), simply press the [Quick Action] key.
You can set up network printer connections. A network printer is a printer attached to the network or a print server on the network.
1 Press [Save/Recall] > Print. 2 In the Print Config menu, press the Print to softkey; then, turn the Entry knob to
select the network printer you want to configure (either #0 or #1).
3 Press the Network Setup softkey. 4 In the Network Printer Setup Menu, press the Modify softkey; then, turn the
Entry knob to select the network parameter you want to enter.
The setting that must be entered is:
Printer Address — this is the printer or print server's address in one of the
following formats:
IP address of a network-enabled printer (for example: 192.168.1.100 or
192.168.1.100:650). Optionally, a non-standard port number can be specified following a colon.
IP address of a print server followed by the path to the printer (for
example: 192.168.1.100/printers/printer-name or
192.168.1.100:650/printers/printer-name).
5 Use the Spell, Enter, and Delete Character softkeys to enter the network printer
settings:
Spell — press this softkey and turn the entry knob to select the character at
the current position.
Enter — press this softkey to enter characters and move the cursor to the next
character position.
Delete Character — press the Enter softkey until the desired character is
highlighted; then, press this softkey to delete the character.
6 Press the Apply softkey to make the printer connection.
A message appears to tell you whether the connection was successful.
86 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Utility Settings

This section explains oscilloscope utility functions.
Table 32 Utility Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
I/O setup [Utility] > I/O >
The oscilloscope can be accessed and/or controlled remotely using these interfaces:
USB device port on the rear panel (square shaped USB port).
Connect the USB device port to a computer only when the oscilloscope is turned off or when the oscilloscope is fully booted up and running. If you make this connection while the oscilloscope is booting up, you will likely get a "USB device not recognized" error.
LAN interface on the rear panel.
The oscilloscope can be controlled using both I/O interfaces at the same time.
For more information, see the Programmer's Guide.
File explorer [Utility] > File Explorer >
From the internal file system, under "/User Files", you can load oscilloscope setup files (from 10 locations) or mask files (from four locations).
From a connected USB storage device, you can load setup files, mask files, license files, firmware update (*.ksx) files, label files, etc. Also, you can delete files on a connected USB storage device. See also "USB Storage Devices" on page 91.
The rectangular USB port on the front panel is a USB Series A receptacle to which you can connect USB mass storage devices and printers.
NOTE: To properly unmount a USB storage device before removal, press the Eject softkey. If you remove the device without pressing the Eject softkey first, the device will be marked as needing repair when connecting to a computer with the Windows operating system (even though there are no harmful effects to the device).
Quick Reference 2
V/div expansion options
Transparent backgrounds
Screen saver [Utility] > Options > Preferences > Screen Saver >
Autoscale preferences [Utility] > Options > Preferences > Autoscale >
[Utility] > Options > Preferences > Expand,
• Ground
• Center
[Utility] > Options > Preferences > Transparent
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 87
2 Quick Reference
Table 32 Utility Features (continued)
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Undo Autoscale [Utility] > Options > Preferences > Autoscale > Undo Autoscale
Fast debug Autoscale [Utility] > Options > Preferences > Autoscale > Fast Debug
Channels to be autoscaled
Acquisition mode during autoscale
Oscilloscope clock [Utility] > Options > Clock >
Softkey menu timeout [Utility] > Options > Menu Timeout
Gen Out signal [Utility] > Options > Auxiliary > Gen Out,
User calibration protect
Remote command logging
Service task support [Utility] > Service >
[Utility] > Options > Preferences > Autoscale > Channels (All Channels, Only Displayed Channels)
[Utility] > Options > Preferences > Autoscale > Acq Mode (Use normal acquisition mode, Preserve
acquisition mode)
• Triggers
•Mask
•WaveGen
[Utility] > Options > Auxiliary > Cal Protect
[Utility] > Options > Remote Log >
See also the Keysight InfiniiVision DSOX1204A/G Oscilloscopes Service Guide. The service guide also tells you how to:
Clean the oscilloscope
Check warranty and extended services status
Contact Keysight (www.keysight.com/find/contactus)
Return the instrument
User calibration [Utility] > Service > Start User Calibration
User calibration status [Utility] > Service > User Calibration Status
Hardware self test [Utility] > Service > Diagnostics > Hardware Self Test
Front panel self test [Utility] > Service > Diagnostics > Front Panel Self Test
Crash log export [Utility] > Service > Diagnostics > Export Logs >
Crash log files are used by Keysight Technical Support when debugging oscilloscope problems.
88 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Table 32 Utility Features (continued)
NOTE
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Host ID export [Utility] > Service > Export Host ID >
The oscilloscope's Host ID is needed when redeeming licenses for oscilloscope upgrades or other licensed features.
About oscilloscope [Help] > About Oscilloscope
Displays: model number, serial number, bandwidth, software version, and installed licenses
Quick Reference 2
User interface language

Setting up the Oscilloscope's LAN Connection

[Help] > Language
You can place the oscilloscope on the network and set up its LAN connection. Once that is done, you can use the oscilloscope's web interface or remotely control the oscilloscope via the LAN interface.
The oscilloscope supports methods for automated LAN configuration or manual LAN configuration (see "To establish a LAN connection" on page 89). It is also possible to set up a point-to-point LAN connection between a PC and the oscilloscope (see "Stand-alone (Point-to-Point) Connection to a PC" on page 91).
Once the oscilloscope is set up on the network, you can use the oscilloscope's web page to view or change its network configuration and access additional settings (like the network password). See "Web Interface" on page 93.
Any time you modify the oscilloscope's hostname, it breaks the connection between the oscilloscope and the LAN. You need to re-establish communication to the oscilloscope using the new hostname.
To establish a LAN connection
Automatic
Configuration
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 89
1 Press [Utility] > I/O. 2 Press the LAN Settings softkey. 3 Press the Config softkey; then, turn the Entry knob to select Automatic, and press
the softkey again to enable it.
2 Quick Reference
If your network supports DHCP or AutoIP, enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings
4 If your network provides Dynamic DNS, you can enable the Dynamic DNS option
to let the oscilloscope register its hostname and use the DNS server for name resolution.
5 You can enable the Multicast DNS option to let the oscilloscope use Multicast
DNS for name resolution on small networks without a conventional DNS server.
6 Connect the oscilloscope to the local area network (LAN) by inserting the LAN
cable into the "LAN" port on the rear panel of the oscilloscope.
In a few moments the oscilloscope will connect to the network automatically.
If the oscilloscope does not automatically connect to the network, press [Utility] > I/O > LAN Reset. In a few moments the oscilloscope will connect to the network.
Manual
Configuration
1 Get the oscilloscope's network parameters (host name, IP address, subnet
mask, gateway IP, DNS IP, etc.) from your network administrator.
2 Press [Utility] > I/O. 3 Press the LAN Settings softkey. 4 Press the Config softkey; then, turn the Entry knob to select Automatic, and press
the softkey again to disable it.
If Automatic is not enabled, the oscilloscope's LAN configuration must be set up manually using the Addresses and Host name softkeys
5 Configure the oscilloscope's LAN interface:
a Press the Addresses softkey. b Use the Modify softkey (and the other softkeys and the Entry knob) to enter
the IP Address, Subnet Mask, Gateway IP, and DNS IP values. When you are done, go back up in the menu hierarchy.
c Press the Host name softkey. Use the softkeys and the Entry knob to enter the
Host name. When you are done, go back up in the menu hierarchy.
d Press the Apply softkey.
6 Connect the oscilloscope to the local area network (LAN) by inserting the LAN
cable into the "LAN" port on the rear panel of the oscilloscope.
90 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Quick Reference 2
Stand-alone (Point-to-Point) Connection to a PC
The following procedure describes how to establish a point-to-point (stand alone) connection to the oscilloscope. This is useful if you want to control the oscilloscope using a laptop computer or a stand-alone computer.
1 Press [Utility] > I/O. 2 Press the LAN Settings softkey. 3 Press the Config softkey; then, turn the Entry knob to select Automatic, and press
the softkey again to enable it.
If your network supports DHCP or AutoIP, enabling Automatic lets the oscilloscope use those services to get its LAN configuration settings
4 Connect your PC to the oscilloscope using a cross-over LAN cable such as
Keysight part number 5061-0701, available on the web at
www.keysight.com/find/parts.
5 Cycle power on the oscilloscope. Wait until the LAN connection is configured:
Press [Utility] > I/O and wait until the LAN status shows "configured".
This may take a few minutes.
Now, the instrument is connected, and the instrument's web interface or remote control via LAN may be used.

USB Storage Devices

Use your PC to create directories on a USB storage device.
Most USB mass storage devices are compatible with the oscilloscope. However, certain devices may be incompatible, and may not be able to be read or written to. USB storage devices must be formatted with the FAT/FAT16, FAT32, NTFS, EXT2, EXT3, or EXT4 file system format. The exFAT format is not supported. Any given storage device may not support all of these formats.
When the USB mass storage device is connected to the oscilloscope's USB host port, a small four-color circle icon may be displayed briefly as the USB storage device is read.
You must "eject" a USB mass storage device before unplugging it; otherwise, the device will be marked as needing repair when connecting to a computer with the Windows operating system (even though there are no harmful effects to the device).
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 91
2 Quick Reference
Do not connect USB storage devices that identify themselves as hardware type "CD" because these devices are not compatible with the InfiniiVision X-Series oscilloscopes.

Configuring the [Quick Action] Key

The [Quick Action] key lets you perform common, repetitive actions by pressing a single key.
Table 33 Quick Action Features
Feature Front Panel Key/Softkey Location (see built-in help for more information)
Quick action setup [Utility] > Quick Action > Action,
•Off
• Quick Measure All
• Quick Mask Statistics Reset — see "Mask Testing" on page 65.
•Quick Print — see "Print (Screens)" on page 85.
•Quick Save — see "Save/Recall (Setups, Screens, Data)" on page 82.
• Quick Recall
• Quick Freeze Display
• Quick Trigger Mode — see "Trigger Mode, Coupling, Reject, Holdoff" on page 51.
• Quick Clear Display
Quick action perform [Quick Action]
92 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Web Interface

Quick Reference 2
When Keysight InfiniiVision 1000 X-Series oscilloscopes are set up on the LAN, you can access the oscilloscope's built-in web server using a Java-enabled web browser. The oscilloscope's web interface lets you:
• View information about the oscilloscope like its model number, serial number, host name, IP address, and VISA (address) connect string.
• Control the oscilloscope using the Remote Front Panel.
• Send SCPI (Standard Commands for Programmable Instrumentation) remote programming commands via the SCPI Commands applet window.
• Save setups, screen images, waveform data, and mask files.
• Recall setup files, reference waveform data files, or mask files.
• Get screen images and save or print them from the browser.
• Activate the Identification function to identify a particular instrument by causing a message to be displayed or a front panel light to blink.
• View installed options, view firmware versions and install firmware upgrade files, and view calibration status (via the Instrument Utilities page).
• View and modify the oscilloscope's network configuration.
The web interface for InfiniiVision X-Series oscilloscopes also provides help for each of its pages.
Microsoft Internet Explorer is the recommended Web browser for communication and control of the oscilloscope. Other Web browsers may work but are not guaranteed to work with the oscilloscope. The Web browser must be Java-enabled.
Before you can use the web interface, you must place the oscilloscope on the network and set up its LAN connection.

Accessing the Web Interface

To access the oscilloscope's web interface: 1 Connect the oscilloscope to your LAN (see "To establish a LAN connection" on
page 89) or establish a point-to-point connection (see "Stand-alone
(Point-to-Point) Connection to a PC" on page 91).
It is possible to use a point-to-point connection, but using a normal LAN connection is the preferred method.
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 93
2 Quick Reference
2 Type the oscilloscope's hostname or IP address in the web browser.
The oscilloscope's web interface Welcome Page is displayed.
94 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Specifications and Characteristics

For up-to-date specifications and characteristics on the 1000 X-Series oscilloscopes, see the data sheet at: www.keysight.com/find/1200X-Series
Quick Reference 2
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 95
2 Quick Reference

Environmental Conditions

Environment Indoor use only.
Ambient temperature
Humidity Operating: Up to 95% RH, non-condensing to temperatures up to +40 °C
Altitude Operating: to 3,000 m
Overvoltage Category
Pollution Degree The InfiniiVision 1000 X-Series oscilloscopes may be operated in environments
Pollution Degree Definitions
Operating: 0 °C to +50 °C
Non-operating: -40 °C to +70 °C
decreasing linearly to 50% RH at +50 °C
Non-operating: Up to 90% RH up to +65 °C (non condensing)
Non-operating to 15,300 m
This product is intended to be powered by MAINS that comply to Overvoltage Category II, which is typical of cord-and-plug connected equipment.
of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. Example: A clean room or climate controlled office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. Example: Sheltered outdoor environment.

Declaration of Conformity

For Declarations of Conformity for Keysight products, go to:
www.keysight.com/go/conformity
96 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Probes and Accessories

For a list of the probes and accessories that are compatible with the 1000 X-Series oscilloscopes, see the data sheet at: www.keysight.com/find/1200X-Series
Because the 1000 X-Series oscilloscopes do not have a ring around the BNC connector for identifying probes, you must set the probe attenuation factor manually. See "Setting Analog Channel Probe Options" on page 38.
See Also For more information on probes and accessories, see www.keysight.com for:
Probes and Accessories Selection Guide (5989-6162EN)
InfiniiVision Oscilloscope Probes and Accessories Selection Guide Data Sheet
(5968-8153EN)
• For compatibility information, manuals, application notes, data sheets, selection guides, SPICE models, and more for oscilloscope probes, see the Probe Resource Center at: www.keysight.com/find/PRC
Quick Reference 2
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 97
2 Quick Reference

Software and Firmware Updates

From time to time Keysight Technologies releases software and firmware updates for its products. To search for firmware updates for your oscilloscope, direct your web browser to www.keysight.com/find/1200X-Series-sw.
To view the currently installed software and firmware press [Help] > About Oscilloscope.
Once you have downloaded a firmware update file, you can place it on a USB storage device and load the file using File Explorer (see "Utility Settings" on page 87).
98 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide

Acknowledgements

Table 34 Third-Party Software
Quick Reference 2
Software Description and Copyright License
asio Copyright (c) 2003-2018 Christopher M. Kohlhoff Boost Software License
Boost Libraries Copyright © 2008 Beman Dawes, Rene Rivera Boost Software License
Breakpad Copyright (c) 2006, Google Inc.
Copyright 2001-2004 Unicode, Inc.
Haru Free PDF Library Copyright (C) 1999-2006 Takeshi Kanno
Copyright (C) 2007-2009 Antony Dovgal
HDF5 HDF5 was developed by The HDF Group and by the National Center
for Supercomputing Applications at the University of Illinois at Urbana-Champaign.
Copyright 2006-2016 by The HDF Group.
Copyright 1998-2006 by the Board of Trustees of the University of Illinois.
JSON for Modern C++ Copyright (c) 2013-2018 Niels Lohmann MIT License
libpng Copyright (c) 1998-2002,2004,2006-2016 Glenn Randers-Pehrson
(Version 0.96 Copyright (c) 1996, 1997 Andreas Dilger)
(Version 0.88 Copyright (c) 1995, 1996 Guy Eric Schalnat, Group 42, Inc.)
1
(BSL-1.0)
(BSL-1.0)
BSD 3-clause "New" or "Revised" License
zlib License
BSD-style open source
libpng License
noVNC Copyright (C) 2011 Joel Martin <github@martintribe.org>
Source code can be obtained from the third party or by contacting Keysight. Keysight will charge for the cost of physically performing the source distribution.
SWFObject Copyright (c) 2007-2009 The SWFObject team MIT License
TCLAP Copyright (c) 2003 Michael E. Smoot MIT License
Mozilla Public License Version 2.0
Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide 99
2 Quick Reference
Table 34 Third-Party Software (continued)
Software Description and Copyright License
U-Boot (C) Copyright 2000 - 2013 Wolfgang Denk, DENX Software
Engineering, wd@denx.de.
Source code can be obtained from the third party or by contacting Keysight. Keysight will charge for the cost of physically performing the source distribution.
udev Copyright (C) 2003 Greg Kroah-Hartman
Copyright (C) 2003-2010 Kay Sievers
vnc Copyright (C) 2002-2005 RealVNC Ltd. All Rights Reserved.
(In separate Remote Front Panel user install)
websockify Copyright 2010 Joel Martin (github.com/kanaka)
Source code can be obtained from the third party or by contacting Keysight. Keysight will charge for the cost of physically performing the source distribution.
zlib Copyright (C) 1995-2013 Jean-loup Gailly and Mark Adler zlib License
1
These licenses are located on the Keysight InfiniiVision oscilloscopes manuals website at
www.keysight.com/find/1200X-Series-manual.
GNU General Public License (GPL or GPLv2)
GNU Lesser General Public License v2.1 or later
GNU General Public License v3.0 or later
GNU Lesser General Public License v3.0
Additionally, there is third-party software in the oscilloscope's embedded Linux operating system. For this list, see:
www.keysight.com/find/embedded-linux-os-third-party-software
1
100 Keysight InfiniiVision DSOX1204A/G Oscilloscopes User's Guide
Loading...