Getting Assistance from Agilent
By internet, phone, or fax, get assistance with all your test and measurement needs.
If you need assistance, refer to “Getting Assistance from Agilent” on page 2.
Description of the Upgrade
This upgrade kit is used to extend the frequency range of the:
• E8362B PNA microwave network analyzer from 20 GHz to 40 GHz
Installation Note E8362-90004 3
Items Included in the Upgrade Kit
Check the contents of your kit against the following list. If any item is missing or damaged,
contact Agilent Technologies. Refer to “Getting Assistance from Agilent” on page 2.
Table 1Option 040 Upgrade Kit (E8362-60109) to Extend the E8362B to 40 GHz
Ref.
Desig.
Parts for all analyzer options (except Option 014)
N/AInstallation note (this document)1E8362-90004
N/ALicense key certificate15964-5136
N/AModel number nameplate (for 40 GHz analyzer, E8363B)1E8363-80002
A21SOMA 5015087-7130
A23Channel R1 detector
A24Channel R2 detector
A25Test port 1 coupler
A26Test port 2 coupler
A27Channel A first converter (mixer)
A30Channel B first converter (mixer)
N/A
Machine screw M3.0 x 18 (for attaching A23 and A24 detectors)
DescriptionQty Part Number
25087-7133
25087-7518
25087-7122
40515-0666
N/A
N/A
N/A
N/A
W1A12 source 20 to A21 SOMA 501E8364-20036
W2A21 SOMA 50 to A22 switch/splitter1E8364-20034
W3A22 switch/splitter to A23 channel R1 detector1E8364-20144
W4A22 switch/splitter to A24 channel R2 detector1E8364-20024
W37A23 channel R1 detector to A16 test set motherboard J2041E8364-60121
W38A24 channel R2 detector to A16 test set motherboard J2051E8364-60122
W39A21 SOMA 50 to A16 test set motherboard J2061E8364-60123
Machine screw M3.0 x 30 (for attaching SOMA 50)
Ribbon cable, A21 SOMA 50 to A16 test set motherboard
Cable, wrapped wire, detectors to A16 test set motherboard
Test set deck front panel
40515-1349
18121-0818
2E7340-60075
1E8364-00014
4Installation Note E8362-90004
Table 1Option 040 Upgrade Kit (E8362-60109) to Extend the E8362B to 40 GHz
Ref.
Desig.
Parts for analyzers WITH NO RF hardware options
W7A23 channel R1 detector to A28 channel R1 mixer1E8364-20025
W8A24 channel R2 detector to A29 channel R2 mixer1E8364-20026
Parts for analyzers WITH Option UNL ONLY
W7A23 channel R1 detector to A28 channel R1 mixer1E8364-20174
W8A24 channel R2 detector to A29 channel R2 mixer1E8364-20175
Parts for analyzers WITH Options UNL and 016 ONLY
W7A23 channel R1 detector to A28 channel R1 mixer1E8364-20149
W8A24 channel R2 detector to A29 channel R2 mixer1E8364-20150
DescriptionQty Part Number
Installation Note E8362-90004 5
Installation Procedure for the Upgrade Kit
The network analyzer must be in proper working condition prior to installing this option. Any
necessary repairs must be made before proceeding with this installation.
WARNING
This installation requires the removal of the analyzer’s protective outer
covers. The analyzer must be powered down and disconnected from the
mains supply before performing this procedure.
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing or connecting
cables or assemblies within the network analyzer.
Static electricity can build up on your body and can easily damage sensitive internal circuit
elements when discharged. Static discharges too small to be felt can cause permanent
damage. To prevent damage to the instrument:
• always wear a grounded wrist strap having a 1 MΩ resistor in series with it when handling
components and assemblies.
• always use a grounded, conductive table mat while working on the instrument.
• always wear a heel strap when working in an area with a conductive floor. If you are
uncertain about the conductivity of your floor, wear a heel strap.
Figure 1 shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to
“Tools Required for the Installation” on page 7 for part numbers.
Figure 1ESD Protection Setup
6Installation Note E8362-90004
Overview of the Installation Procedure
Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer
Step 2. Remove the Outer Cover
Step 3. Remove the Inner Cover
Step 4. Remove the Front Panel Assembly
Step 5. Disconnect the A12 Source 20 to A22 Switch/Splitter Cable at the A12 Source 20
Step 6. Raise the Receiver Deck
Step 7. Remove the Old Hardware and Cables
Step 8. Replace the A38 and A39 Bias Tees (Option UNL Only)
Step 9. Install the Remaining New Hardware
Step 10. Install the Remaining Cables
Step 11. Lower the Receiver Deck
Step 12. Connect the A12 Source 20 to A21 SOMA 50 Cable at the A12 Source 20
Step 13. Replace the Front Panel Nameplate
Step 14. Reinstall the Front Panel Assembly
Step 15. Reinstall the Inner and Outer Covers
Step 16. Enter the New Model Number
Step 17. Verify that the New Model Number Is Shown Correctly
Step 18. Perform Source and Receiver Calibration Adjustments
Tools Required for the Installation
DescriptionAgilent Part Number
T-8 TORX driver (set to 5 in-lbs)N/A
T-10 TORX driver (set to 9 in-lbs)N/A
T-20 TORX driver (set to 21 in-lbs)N/A
5/16-inch torque wrench (set to 10 in-lbs)N/A
1-inch torque wrench (set to 72 in-lbs)N/A
ESD Equipment and Supplies Required for the Installation
DescriptionAgilent Part Number
ESD grounding wrist strap9300-1367
5-ft grounding cord for wrist strap9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire9300-0797
ESD heel strap (for use with conductive floors)9300-1308
Installation Note E8362-90004 7
Equipment Required for Post-Installation Adjustments
Equipment Type
Power meterE4418B/E4419BE4418A/E4419A
Power sensor, 2.4 mm8487ANone
Adapter, 2.4 mm (f) to 2.4 mm (f)11900B85056-60007
RF cable, 2.4 mm (f) to 2.4 mm (f)85133C85133E
Model or
Part Number
Alternate Model or
Part Number
Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer
Refer to the License Key Certificate supplied with the kit:
• Verify that the analyzer’s model and serial numbers match those on the certificate.
• Verify that the option number on the certificate matches the option described by this
installation note.
If the model, serial, or option number do not match the expected results, you will not be able
to install the option. If this is the case, contact Agilent for assistance. Refer to “Getting
Assistance from Agilent” on page 2.
8Installation Note E8362-90004
Step 2. Remove the Outer Cover
CAUTION
1. Disconnect the power cord (if it has not already been disconnected).
2. Using a T-20 TORX driver, remove the strap handles (item
(item
3. Using a T-20 TORX driver, remove the four rear panel feet (item
screws (item
4. Slide the four bottom feet (item
5. Slide the cover off of the frame.
Figure 2Outer Cover Removal
) on both ends until the handle is free of the analyzer.
②
This procedure is best performed with the analyzer resting on its front
handles in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies.
) by loosening the screws
①
③
).
④
) off the cover.
⑤
) by removing the center
Installation Note E8362-90004 9
Step 3. Remove the Inner Cover
1. Place the analyzer top-side up on a flat surface.
2. Using a T-10 TORX driver, remove the eleven screws (item
3. Lift off the cover.
Figure 3Inner Cover Removal
①
).
10 Installation Note E8362-90004
Step 4. Remove the Front Panel Assembly
1. Using a T-10 TORX driver, remove the eight screws (item ②) from the sides of the frame.
CAUTION
2. Slide the front panel over the test port connectors.
3. Disconnect the front panel interface ribbon cable (item
from the analyzer.
Figure 4Front Panel Assembly Removal
Before removing the front panel from the analyzer, lift and support the front of
the analyzer chassis.
). The front panel is now free
③
Installation Note E8362-90004 11
Step 5. Disconnect the A12 Source 20 to A22 Switch/Splitter Cable at
the A12 Source 20
1. Using a 5/16-inch wrench, disconnect the A12 source 20 to A22 switch/splitter cable, W1, at
the A12 source 20 as shown.
Figure 5A12 Source Cable Disconnection
12 Installation Note E8362-90004
Step 6. Raise the Receiver Deck
1. Place the analyzer bottom-side up on a flat surface.
2. Using a T-10 TORX driver, remove the four screws, (item
3. Pull the latch pin (item
deck.
4. Lift the receiver deck to partially raise it, then release the latch pin (item
receiver deck to its fully raised position and ensure that the latch pin latches in the raised
position.
Figure 6Receiver Deck Raising
②) towards the opposite side of the analyzer to release the receiver
①), that secure the receiver deck.
②). Lift the
Installation Note E8362-90004 13
Step 7. Remove the Old Hardware and Cables
Remove the Old Cables (Refer to Figure 7.)
1. Using a 5/16-inch wrench, remove the following cables in the order listed.
• W1 A12 source 20 to A22 switch splitter
• W3 A22 switch splitter to channel R1 fixed attenuator
• W4 A22 switch splitter to channel R2 fixed attenuator
lifting. Scrape away any excess adhesive from the bracket.
Replace the A27 and A30 Mixers (Refer to Figure 7 and Figure 8.)
5. Disconnect all RF cables (item
6. Remove two RF cables (item
7. Remove four screws (item
later.
8. Remove the mixer bracket, with the mixers attached, from the analyzer. Observe the
CAUTION below. Disconnect the ribbon cable to allow the mixer bracket to be removed.
CAUTION
9. Remove the old A27 and A30 mixers from the mixer bracket by removing three screws
from each, as shown in Figure 8. Retain the screws to install the new mixers.
10. Install the new A27 and A30 mixers using the screws removed from the old mixers. Be
sure to orient the mixers as shown in Figure 8.
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary when removing the assemblies. Do not over-bend them.
①).
②) from the analyzer.
③) by sliding a thin, sharp blade under the clamp and
④) from the mixers.
⑤) from the analyzer and retain for reinstallation later.
⑥) from the mixer bracket. Retain the screws for reinstallation
11. Reconnect the ribbon cable and place the mixer bracket, with all four mixers attached,
into the analyzer. Observe the
12. Reinstall the four screws (item
13. Reconnect all RF cables (item
14 Installation Note E8362-90004
CAUTION above.
⑥) to secure the bracket.
⑤) to the mixers.
Figure 7Mixers Bracket Removal
Figure 8Mixers Replacement
Installation Note E8362-90004 15
Replace the Test Set Deck Front Panel and the A25 and A26 Test Port Couplers
Remove the Existing Test Set Deck Front Panel (Refer to Figure 9.)
1. Disconnect the wrapped wire cable (item
2. Disconnect the remaining RF cable (item
3. Using a T-10 TORX driver, remove eight screws (items
front panel, to release it.
4. Remove the test set deck front panel from the analyzer, with the LED board and both
couplers attached.
5. Using a 1-inch wrench, remove the flange nut (item
flange nuts for installation of the new couplers. Discard the front panel and the couplers.
Figure 9Test Set Deck Front Panel Replacement
①) from the front-panel LED board.
②) from each test port coupler.
❶ through ❽) from the test set deck
③) from each coupler. Retain the
16 Installation Note E8362-90004
Install the Old Front Panel LED Board onto the New Front Panel (Refer to Figure 10.)
1. Remove the LED board from the old front panel by removing the three retaining screws
①).
(item
2. Place the LED board in position on the new front panel and attach it with the three screws
(item
①) removed from the old front panel.
Figure 10 Front Panel LED Board Reinstallation
Install the New Couplers on the New Front Panel (Refer to Figure 9.)
1. Place the new couplers onto the new front panel. Orient them the same as the old ones as
shown in Figure 9.
2. Install a flange nut (item
a 1-inch torque wrench, tighten each flange nut to 72 in-lbs.
3. Reconnect the two cables (item
page 16.
Install the New Front Panel on the Analyzer (Refer to Figure 9.)
1. Place the new test set deck front panel, with the new LED board and both new couplers
attached, onto the test set deck as shown in Figure 9.
2. Using a T-10 TORX driver, loosely reinstall the eight screws (items
the front panel to the test set deck. Torque each of the eight screws to 9 in-lbs in the order
indicated (
3. Reconnect the wrapped wire cable (item
Installation Note E8362-90004 17
❶ through ❽). This is important to ensure proper alignment.
③) removed from the old couplers, onto each new coupler. Using
②) to the test port couplers that were disconnected on
❶ through ❽) to secure
①) to the front-panel LED board.
Step 8. Replace the A38 and A39 Bias Tees (Option UNL Only)
Remove the Detector/Bias Tee/Attenuator Brackets (Refer to Figure 11.)
1. Disconnect all remaining RF cables from each attenuator (item
2. Disconnect the bias tee control cables at the A16 test set motherboard connectors,
J22 (P1 BIAS T) and J25 (P2 BIAS T), and release them from the cable clamp (item
3. Disconnect the ribbon cables at the step attenuators (item
4. Remove three screws (item
5. Remove the brackets, with the attenuators and bias tees attached, from the analyzer.
Be careful not to damage the center pins of the semirigid cables. Some flexing of
the cables is necessary when removing the assemblies. Do not over-bend them.
④) from each bracket. Retain the screws for reinstallation.
①) and bias tee (item ②).
③).
①).
18 Installation Note E8362-90004
Replace the Bias Tees (Refer to Figure 12.)
1. Remove the old bias tees from the bracket by removing two screws (item
Retain the screws for installing the new bias tees.
2. Install the new bias tees using the screws (item
to install the ground lug (item
Figure 12 Bias Tees Replacement
②) over the attachment screw as shown.
①) removed from the old bias tees. Be sure
①) from each.
Reinstall the Detector/Bias Tee/Attenuator Brackets (Refer to Figure 11.)
1. Place the brackets, with the attenuators and bias tees attached, into the analyzer as shown
in Figure 11.
2. Reinstall the three screws (item
3. Reconnect the ribbon cables to the attenuators (item
4. Reconnect the bias tee control cables to the A16 test set motherboard. The A38 cable
connects to J22 (P1 BIAS T) and the A39 cable connects to J25 (P2 BIAS T). Be sure to route
them under the cable clamp, (item
5. Reconnect all RF cables to the attenuators (item
④) in each bracket.
①).
③).
①) and bias tees (item ②).
Installation Note E8362-90004 19
Step 9. Install the Remaining New Hardware
Install the A23 and A24 Detectors (Refer to Figure 13.)
1. Place the A23 and A24 detectors on top of the brackets in the analyzer. Be sure to orient
them as shown.
2. Install two screws (item
3. Connect the wrapped wire control cables from the A23 and A24 detectors:
• A23 detector to A16 test set motherboard J5 (R1 DET)
• A24 detector to A16 test set motherboard J6 (R2 DET)
Install the A21 SOMA 50 (Refer to Figure 13.)
1. Place the A21 SOMA 50 in position in the analyzer. Be sure to orient it as shown.
2. Install four screws (item
3. Connect the A21 SOMA 50 ribbon cable to A16 test set motherboard J4 (SOMA50).
Figure 13 A23 and A24 Detectors and A21 SOMA 50 Installation
①) in each detector to secure them to the brackets.
②) to secure the SOMA 50 to the A16 test set motherboard.
20 Installation Note E8362-90004
Step 10. Install the Remaining Cables
Using a 5/16-inch torque wrench set to 10in-lbs, install the following new cables in the order
listed. These new parts are listed in Table 1 on page 4.
• W1A12 source 20 to A21 SOMA 50
Connect this cable to the A21 SOMA 50 connector as shown with the other end of the
cable routed through the opening to the top side of the analyzer. The other end of this
cable will be connected later.
• W2A21 SOMA 50 to A22 switch/splitter
• W3A22 switch/splitter to A23 channel R1 detector
• W4A22 switch/splitter to A24 channel R2 detector
• W37 A23 channel R1 detector to A16 test set motherboard J204 (R1)
• W38 A24 channel R2 detector to A16 test set motherboard J205 (R2)
• W39 A21 SOMA 50 to A16 test set motherboard J206 (SOMA 50)
• W7A23 channel R1 detector to A28 channel R1 mixer
The shape of this cable varies depending upon the option installed. Be sure to install
the correct cable for your analyzer as specified in Table 1 on page 4.
• W8A24 channel R2 detector to A29 channel R2 mixer
The shape of this cable varies depending upon the option installed. Be sure to install
the correct cable for your analyzer as specified in Table 1 on page 4.
Figure 14Cables Installation
Installation Note E8362-90004 21
Step 11. Lower the Receiver Deck
1. Pull the latch pin (item ②) towards the opposite side of the analyzer to release the receiver
deck.
2. Lift the receiver deck to partially lower it, then release the latch pin (item
receiver deck to its fully lowered position and ensure that the latch pin latches in the
lowered position.
3. Using a T-10 TORX driver, reinstall the four screws, (item
Figure 15Receiver Deck Lowering
①), to secure the receiver deck.
②). Lower the
22 Installation Note E8362-90004
Step 12. Connect the A12 Source 20 to A21 SOMA 50 Cable at the A12
Source 20
1. Place the analyzer top-side up on a flat surface.
2. Using a 5/16-inch wrench, connect the A12 source 20 to A21 SOMA 50 cable, W1, at the
A12 source 20 as shown.
Figure 16 A12 Source Cable Connection
Installation Note E8362-90004 23
Step 13. Replace the Front Panel Nameplate
Remove the Old Nameplate
1. Using a sharp object such as a knife blade, lift one end of the old nameplate (item
careful not to scratch the painted surface.
2. Using your fingers, pull the nameplate completely off of the front panel and discard it.
3. If any adhesive remains on the front panel surface, it can be removed by rubbing it with
your finger. Do not use a sharp object to remove this adhesive. If necessary, the surface can
be cleaned with isopropol alcohol.
Install the New Nameplate
IMPORTANT
1. Remove the protective backing from the new nameplate to expose the adhesive. Avoid
touching the adhesive with your fingers as this may affect its ability to adhere to the front
panel surface.
2. Loosely align the new nameplate over the recess in the front panel and apply lightly.
3. Once the nameplate is properly aligned, press it firmly onto the surface to secure it.
Figure 17Model Number Nameplate Replacement
If the Option 050 upgrade kit will be installed to extend the frequency to
50 GHz, do not install the new nameplate provided in this upgrade kit.
Instead, use the one provided in the Option 050 upgrade kit.
①). Be
24 Installation Note E8362-90004
Step 14. Reinstall the Front Panel Assembly
CAUTION
1. Reconnect the ribbon cable (item
2. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
cable (item
3. Using a T-10 TORX driver, install the eight screws (item
Figure 18Front Panel Assembly Reinstallation
Before installing the front panel assembly onto the analyzer, lift and support the
front of the analyzer chassis.
) to the A3 front panel interface board.
①
) is located below the fan to prevent it from being damaged by the fan blades.
①
) in the sides of the frame.
②
Installation Note E8362-90004 25
Step 15. Reinstall the Inner and Outer Covers
Reinstall the Inner Cover
1. Place the inner cover on the analyzer as shown. There are two alignment pins on the front
frame that align with holes in the front of the cover to ensure proper alignment.
2. Using a T-10 TORX driver, reinstall the eleven screws (item
Figure 19 Inner Cover Reinstallation
①
).
26 Installation Note E8362-90004
Reinstall the Outer Cover
CAUTION
1. Slide the cover onto the frame.
2. Using a T-20 TORX driver, install the four rear panel feet (item
screws (item
3. Slide the four bottom feet (item
4. Using a T-20 TORX driver, install the strap handles (item
(item
②) on both ends.
Figure 20 Outer Cover Reinstallation
This procedure is best performed with the analyzer resting on its front
handles in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies.
).
④
) onto the cover.
⑤
) by tightening the screws
①
) by installing the center
③
Installation Note E8362-90004 27
Step 16. Enter the New Model Number
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A keyboard is required for this procedure.
• A mouse is recommended for this procedure but is not required.
Mouse Procedure
1. On the analyzer’s System menu, point to Service, then click Option Enable.
2. Click Repair.
3. Check Model shown is incorrect!, then click Begin Repair.
4. In the Model Change box, select the new model number, E8363B.
5. Using the keyboard, enter the keyword in the box provided. The keyword is printed on the
License Key Certificate. Enter this keyword exactly as it is printed on the certificate.
6. Click Enter Model.
7. Verify that the new model number displayed is correct, then click YES. If it is not correct,
click NO and repeat the repair.
8. When the installation is complete, click Exit.
Front Panel Keys Procedure
1. In the COMMAND block, press Menu/Dialog.
2. In the NAVIGATION block, press the right tab and arrow keys to move over to the System
menu and down to the Service selection. Press the right tab to display the extended menu
and the arrows to select Option Enable. Press Click.
3. Press Menu/Dialog, then press R on the keyboard to select Repair.
4. Tab to Model shown is incorrect!, then press Click.
5. Tab to the Model Change box, then use the arrows to select the new model number,
E8363B.
6. Tab to the Keyword box and, using the keyboard, enter the keyword. The keyword is
printed on the License Key Certificate. Enter this keyword exactly as it is printed on the
certificate.
7. Tab to Enter Model, then press Click.
8. Verify that the new model number displayed is correct, then tab to Yes and press Click. If
it is not correct, tab to NO, press Click, and repeat the repair.
9. When the installation is complete, Tab to Exit, then press Click.
28 Installation Note E8362-90004
Step 17. Verify that the New Model Number Is Shown Correctly
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse is recommended for this procedure but is not required.
Mouse Procedure
1. On the analyzer’s Help menu, click About Network Analyzer.
2. Verify that the new model number, E8363B, is shown correctly in the display. Click OK.
3. On the analyzer front panel, press Preset. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.
Front Panel Keys Procedure
1. In the COMMAND block, press Menu/Dialog.
2. In the NAVIGATION block, press the right tab and arrow keys to move over to the Help
menu, and down to the About Network Analyzer selection. Press Click.
3. Verify that the new model number, E8363B, is shown correctly, then press Click.
4. On the analyzer front panel, press Preset. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.
Installation Note E8362-90004 29
Step 18. Perform Source and Receiver Calibration Adjustments
The following adjustments must be made due to the change in the full frequency range of the
analyzer.
• source calibration
• receiver calibration
These adjustments are described in the PNA service guide and in the PNA on-line HELP. A
list of equipment required to perform these adjustments can be found at “Tools Required for
the Installation” on page 7.
Performance Tests and System Verification
The analyzer should now operate and phase lock over its entire frequency range.
If you experience difficulty with the basic functioning of the analyzer, contact Agilent. Refer to
“Getting Assistance from Agilent” on page 2.
Although the analyzer functions, its performance relative to its specifications has not been
verified for the additional frequency range enabled by this upgrade.
It is recommended that a full instrument calibration be performed using the N2721A
performance test software.
If the testing of the analyzer’s full range of specifications is not required, a system verification
can be performed.
Refer to the analyzer’s service guide for information on performance tests and system
verification.
30 Installation Note E8362-90004