Keysight E4990A Installation Manual

Page 1
Keysight E4990A Impedance Analyzer

Installation Guide

Page 2
Notices
© Keysight Technologies
2014-2018
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
E4990-90010
Edition
Edition 3, July 2018
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to
http://www.keysight.com/go/confor mity. You can then search by product
number to find the latest Declaration of Conformity.
perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Printed in Malaysia
Published by:
Keysight Technologies International Japan G.K,
1-3-3 Higashikawasaki-cho Chuo-ku Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/swe ula The license set forth in the EULA
represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release,
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
This NOTE sign denotes important information. It calls attention to a procedure, practice, or condition that is essential for the user to understand.
Page 3

Caution

Keysight E4990A Impedance Analyzer
Installation Guide
Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument’s Function Reference.
Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures are being connected or disconnected. Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground. Alternatively, ground yourself to discharge any static charge built- up by touching the outer shell of any grounded instrument chassis before touching the test port connectors.

Safety Summary

When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable.
Contact your local Keysight Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard to the operator.
Instrument operates abnormally.
Instrument emits abnormal noise, smell, smoke or a spark- like light during
Instrument generates high temperature or electrical shock during
Power cable, plug, or receptacle on instrument is damaged.
Foreign substance or liquid has fallen into the instrument.
operation.
operation.
3
Page 4
-
Caution Manufacturer’s Declaration

Manufacturer’s Declaration

Herstellerbescheinigung
GERA- USCHEMISSION
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer's Declaration
ACOUSTIC NOISE EMISSION
LpA < 70 dB
operator position
normal operation
per ISO 7779

Regulatory Compliance Information

This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly:
The Low Voltage Directive 2006/95/EC
The EMC Directive 2004/108/EEC
To obtain Declaration of Conformity, please contact your local Keysight Technologies sales office, agent or distributor.

Safety Notice Supplement

This equipment complies with EN/IEC61010- 1:2001.
This equipment is of MEASUREMENT CATEGORY I (CAT I). Do not use for
CAT II, III, or IV.
This equipment is a POLLUTION DEGREE 2, INDOOR USE product.
4 Keysight E4990A Impedance Analyzer
Page 5
Caution General Safety Precautions
This equipment is tested in stand- alone condition and in combination with
the accessories supplied by Keysight Technologies against the requirement of the standards described in the Declaration of Conformity. If it is used as a system component, compliance of related regulations and safety requirements are to be confirmed by the builder of the system.
Do not connect the measuring terminals to mains.

General Safety Precautions

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment. Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these precautions.
The E4990A complies with INSTALLATION CATEGORY II as well as POLLUTION DEGREE 2 in IEC61010-1. The E4990A is an INDOOR USE product.
The LEDs in the E4990A are Class 1 in accordance with IEC60825-1, CLASS 1 LED PRODUCT.
Ground the Instrument
To avoid electric shock, the instrument chassis and cabinet must be grounded with the supplied power cable’s grounding prong.
DO NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gases or fumes. Operation of any electrical instrument in such an environment clearly constitutes a safety hazard.
Keep Away from Live Circuits
Operators must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected. Under certain conditions, dangerous voltage levels may remain even after the power cable has been disconnected. To avoid injuries, always disconnect the power and discharge circuits before touching them.
DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
DO NOT Substitute Parts or Modify the Instrument
Keysight E4990A Impedance Analyzer 5
Page 6
-
Caution Safety Symbols
To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
Dangerous voltage levels, capable of causing death, are present in this instrument. Use extreme caution when handling, testing, and adjusting this instrument.

Safety Symbols

General definitions of safety symbols used on the instrument or in manuals are listed below.
Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
A chassis terminal; a connection to the instrument’s chassis, which includes all exposed metal structure.
Standby.

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United
6 Keysight E4990A Impedance Analyzer
Page 7
Caution Safety Symbols
States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility or by the calibration facilities of other International Standards Organization members.

Exclusive Remedies

The remedies provided herein are Buyer’s sole and exclusive remedies. Keysight Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.

Assistance

Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products.
For any assistance, contact your nearest Keysight Technologies Sales and Service Office. Addresses are provided at the back of this manual.
Keysight E4990A Impedance Analyzer 7
Page 8
-
Caution Manuals for E4990A

Manuals for E4990A

Keysight provides the following three manuals for E4990A. The latest version of all documentations can be downloaded from
http://www.keysight.com/find/E4990A-manual.

Installation Guide

The installation guide (this manual) provides start up setup information when you use the E4990A for the first time and troubleshooting information when the Windows cannot be boot up. See this manual first when you use the E4990A for the first time.

Online Help

The online help provides the information about the quick start, measurement operation, programming, and I/O interface. This is pre-installed in the E4990A. Press [Help] hard key on the front panel to open. Quick Start helps in understanding the E4990A operation quickly.
The latest version of online help is available at:
http://ena.support.keysight.com/e4990a/manuals/webhelp/eng/
The online help has context sensitive help, which is a great feature of the E4990A help. It allows you to get information about the selected softkey by pressing the Help key in the E4990A or by pressing F1 in a keyboard attached to the E4990A or by clicking the help button in a dialog box. With context sensitive help, users can receive information quickly about the area the user is working in the firmware of the E4990A. It provides information relevant to the task that needs to be accomplished and reduces the time to search relevant information required to complete a task.

Service Guide

The service manual provides information about the parts, troubleshooting, performance test, adjustment and service menu.
8 Keysight E4990A Impedance Analyzer
Page 9
Table of Contents
.
Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Regulatory Compliance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Notice Supplement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
General Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Assistance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Manuals for E4990A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Installation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Online Help. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Service Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Contents
1. Introduction
Checking the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Operating Environments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Ventilation Requirement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Protection Against Electrostatic Discharge (ESD). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power
Cable in Case of Emergency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installing Front Handles/Rack Mounting Flanges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Verification of the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Verification and Connection of Power Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Blown Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Starting the E4990A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Turning the Power ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Disconnection from Supply Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Initial Registration of E4990A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2. Troubleshooting
Troubleshooting during Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
System Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Notes on executing the factory recovery function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Keysight E4990A Installation Guide 9
Page 10
Contents
Procedure to execute the factory recovery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
10 Keysight E4990A Installation Guide
Page 11
Keysight E4990A Impedance Analyzer
Installation Guide

1 Introduction

This chapter provides information about how to set up Keysight E4990A

Checking the Shipment

After you receive the analyzer, carry out checks during unpacking according to the following procedure.
When unpacking the analyzer, if the exterior of the analyzer (such as the cover, front/rear panel, LCD screen, power switch, and port connectors) appear to be damaged during transport, do not turn on the power switch. Otherwise, you may get an electric shock.
Step 1. Check that the packing box or shock- absorbing material used to package the
analyzer has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave the packing box and shock-absorbing material as is until other inspection items are checked as follows.
Step 2. Check the packaged items supplied with the analyzer for any damage or
defects.
Step 3. By referring to the furnished contents list, check that all packaged items
supplied with the analyzer have been received as per the specified options.
Step 4. After checking, if one of the following applies, contact your nearest Keysight
Technologies sales and service office.
1. The packing box or shock- absorbing material used to package the analyzer has been damaged or the shock- absorbing material has traces where extreme force has been applied.
2. A packaged item supplied with the analyzer has mechanical damage or defects.
11
Page 12
1-
Introduction Environmental Requirements
3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the analyzer.
If an abnormality is detected in step 1, contact the company that transported the analyzer as well as your nearest Keysight Technologies sales and service office.

Environmental Requirements

Set up theE4990A where the following environmental requirements are met.

Operating Environments

Ensure that the operating environment meets the following requirements.
Temperature 5 ºC to 40 ºC
Temperature range at the error-correction
Humidity 20% to 80% at wet bulb temperature < +29 °C
Altitude 0 to 2,000 m (0 to 6,561 feet)
Vibration 0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and measurement accuracy of the analyzer, but for the operating environment of the analyzer.

Ventilation Requirement

To ensure safety requirements, the specifications and measurement accuracy of the analyzer, you must keep the environmental temperature within the specified range by providing appropriate cooling clearance around the analyzer or, for the rack mount type, by forcefully air- cooling inside the rack housing. For more information on environmental temperature to satisfy the specifications and measurement accuracy of the analyzer, see the specification in the E4990A Datasheet.
23 °C ± 5 °C (< 1 °C deviation from the temperature when performing the error-correction)
(non-condensation)
When the environmental temperature around the analyzer is kept within the temperature range of the operating environment specification (See Section
“Operating Environments” on page 12), the analyzer conforms to the
12 Keysight E4990A Impedance Analyzer
Page 13
Introduction Environmental Requirements
requirements of the safety standard. Furthermore, under that temperature requirement, the analyzer still conforms to the requirements of the safety standard even when placing the analyzer with cooling clearance as follows:
Requirements
Back 180 mm
Sides 60mm (both right and left)

Protection Against Electrostatic Discharge (ESD)

Set up a static- safe work- station to protect the electronic components from the damage by the electrostatic discharge (ESD) as shown in Figure 1-1.
Figure 1-1 Example of the static-safe work station

Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency

As described in “Disconnection from Supply Source” on page 20, the power supply is disconnected by removing the power cable’s connector plug from either the AC outlet or the E4990A unit. When installing the E4990A, ensure that there is sufficient free space around the unit to permit quick disconnection of the plug (from AC outlet or E4990A unit) in case of emergency.
Keysight E4990A Impedance Analyzer 13
Page 14
1-
Introduction Environmental Requirements

Installing Front Handles/Rack Mounting Flanges

The E4990A can be installed on a workbench or on a rack.
This section describes how to install the front handles (Option 1CN) used for moving or transporting the instrument, and how to install the analyzer in an equipment rack as part of a measurement system (Option 1CM: without the handles, Option 1CP: with the handles).
Figure 1-2 Installing front handle/rack-mount kits
14 Keysight E4990A Impedance Analyzer
Page 15
Introduction Environmental Requirements
Figure 1-3 E4990A with Option 1CN handle
The handle kit (Option 1CN) is used for transporting and relocating of the E4990A.
The rack- mount kit (Option 1CM) includes two flanges (locking side plates) for mounting the E4990A on a rack (482.6 mm/19 inches) conforming to the EIA Standard.
Option 1CP comes with both, the handle kit and rack- mount kit.
While referring to Figure 1-2 install the handle kit/rack- mount kit by following these steps.
Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame
of the E4990A front panel.
Step 2. Use the provided screws to mount the
front handles (2) - for Option 1CN
rack- mounting flange (4) - for Option 1CM
both - for Option 1CP
on each side of the E4990A front panel frame.
Step 3. (Option 1CN/Option 1CP). Attach the modified trim strip (3) provided to each
front handle in order to cover the front panel locking screws.
Option 1CN contains 2 pieces of standard and 1 special handles ­E4990-20001. Only one of the standard handle is required to be used to mount on the analyzer. Standard handle is used for the left side and special handle is used for the right side. Refer to Figure 1-3.
Keysight E4990A Impedance Analyzer 15
Page 16
1-
Introduction Environmental Requirements
Step 3. (Option 1CM/Option 1CP). Remove the four bottom feet of the E4990A (lift
the bar marked TAB on the inner side of the foot and slide the foot towards the bar).
Step 4. (Option 1CM/Option 1CP). Mount the E4990A on the rack.
If the installed front handle becomes damaged, replace it with a new one immediately. A damaged handle may break while moving or lifting the instrument and cause personal injury or damage to the instrument.
Use both a front handle and a rack-mounting flange in the same time certainly. Do not attempt to install flanges or handles separately with hard ware provided. Serious electrical damage will occur to the instrument.
16 Keysight E4990A Impedance Analyzer
Page 17

Power Supply

Verification of the Power Supply

Introduction Power Supply
Before turning on the E4990A power, check the following.
Confirm that the power supplied to the E4990A meets the following requirements:
Requirements
Voltage
90 to 132 VAC or 198 to 264 VAC
1
Frequency 47 to 63 Hz
Maximum power
300 VA
consumption
1. Switched automatically by the E4990A in conformity to the voltage used.

Verification and Connection of Power Cable

The three-wire power cable attached to the E4990A has one wire serving as the ground. Using this power cable allows the E4990A to be grounded, thereby protecting you against any potential electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty cables can cause electrical shock.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 1-4 on page 20) on the rear panel of the E4990A and a three- wire power outlet with the grounding prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely ground the E4990A.
Power cord list, 16000- 99101 shows the power cable options.
Keysight E4990A Impedance Analyzer 17
Page 18
1-
Introduction Blown Fuses

Blown Fuses

If the fuse appears to have blown during operation, this instrument may be subject to failure and must be repaired. For any assistance, contact Keysight Technologies Customer contact center listed at the end of this guide.
The E4990A uses the following fuse types:
UL/CSA Type, Slow- Blo, 10 A- 250 Vac.
DO NOT replace the fuse yourself as doing this may expose yourself to electrical shock.
18 Keysight E4990A Impedance Analyzer
Page 19
Introduction Starting the E4990A

Starting the E4990A

This section describes how to turn ON/OFF the E4990A power and how to cut off the power supply in the case of emergency.

Turning the Power ON and OFF

The standby switch can turn ON/OFF the E4990A. The color on the button shows the status, as indicated below:
Indicator Color Description
Green Normal power on status
Orange Standby status
Red Illegal power on status
Perform the following steps to turn the power ON or OFF.
Turning the Power ON
Step 1. Confirm if the Line Switch on the rear panel is turned ON. The switch should
always be turned ON.
Step 2. Press the standby switch once. It changes to green color.
This operation turns ON the power, and the E4990A starts the self- test.
Step 3. Confirm that the self- test indicates normal operation.
Normal operation is confirmed by the self- test if no error message appears.
Turning the Power OFF
Step 1. To turn OFF the power of the E4990A, first, press this standby switch or send
a shutdown command from the external controller to activate the shutdown process (the process of software and hardware necessary to turn OFF the power supply). This puts the E4990A into the standby state and the button changes to orange color.
Keysight E4990A Impedance Analyzer 19
Page 20
1-
Introduction Starting the E4990A
Step 2. Next, if necessary, turn OFF the power supply to the power cable receptacle
(Line Switch) on the rear panel.
Under normal use, never directly interrupt the power supply to the power cable receptacle on the rear panel when the power supply is turned ON. Always keep the Line Switch ON and never turn it.
If you directly interrupt the power supply to the power cable receptacle when the power supply is turned ON, or turn OFF the Line Switch, the shutdown process will not work. This could damage the software and hardware of the E4990A and lead to device failure.
Turning ON the power supply after a faulty shutdown may cause the system to start up in a condition called "safe mode." If this occurs, first shut down the system to set it to the standby state and then turn ON the power supply again to start up the system in normal mode.
Figure 1-4 Line switch (Always ON) and power cable receptacle

Disconnection from Supply Source

The power supply of the E4990A is cut off by disconnecting the plug of the power cable (on either AC outlet side or E4990A side). When it is necessary to disconnect the power supply in order to avoid shock hazards, etc., pull out the power cable plug from either the AC outlet side or the E4990A side.
20 Keysight E4990A Impedance Analyzer
Page 21
Introduction Starting the E4990A
To allow this operation to be performed smoothly, be sure to follow the guidelines in “Ensuring Adequate Free Space around Analyzer for
Immediate Disconnection of Power Cable in Case of Emergency” on page 13.
When turning the power OFF under normal circumstances, always follow the methods described in “Turning the Power OFF” on page 19.
Keysight E4990A Impedance Analyzer 21
Page 22
1-
Introduction Initial Registration of E4990A

Initial Registration of E4990A

When you start up the E4990A for the first time, you need to perform the initial registration of the Windows operating system of the E4990A.
When the Windows agreement message box appears, click Agree. The rest of the procedure runs automatically.
The Touch Screen does not require initial calibration to work for the very first time. However, user have the option of calibrating the touch screen whenever required. To do so, select System hardkey followed by Misc Setup > Display Setup > Ad just Touch Screen. Refer to Online Help for more details on how to adjust the touch screen.
22 Keysight E4990A Impedance Analyzer
Page 23
Keysight E4990A Impedance Analyzer
Installation Guide

2 Troubleshooting

This chapter describes the troubleshooting process during start up and the procedure of the operating system (OS) recovery when the Windows OS has been damaged.

Troubleshooting during Startup

When you encounter problems during start up, see Table 2-1. System recovery saves most of problems.
Table 2-1 Troubleshooting during startup
Symptom Solution
Turning ON ( | ) the standby switch does not start up the system.
The system starts up, but it automatically shuts down immediately.
The system starts up, but it enters the service mode (The status bar at the lower right of the screen displays SVC in red).
The measurement screen appears after start up, but the power-on test fails with Error Message 241 prompting on a red background in the instrument’s message/warning area at the lower left of the screen.
Confirm that the power cable is properly plugged in.
Confirm that the line switch at the rear panel is turned ON.
Execute the system recovery.
23
Page 24
2-
Troubleshooting System Recovery

System Recovery

By executing system recovery, you can return the E4990A’s system (the Windows operating system and the firmware) to the factory state (at the time
1
of purchase
Strictly follow the steps described below. If you do any operation other than the following steps, the system may not be recovered.

Notes on executing the factory recovery function

Executing the factory recovery function causes the following conditions:
In addition to the Windows operating system and the firmware, the following settings of the E4990A are returned to the factory state.
—Network setting
).
— GPIB setting
—Printer setting
The driver for the supported printer which is installed after purchase is deleted.
You need to execute initial registration again.
Files that you have created using the save function (files in the D drive) are not affected. However, Keysight recommends backing them up before executing system recovery for precautionary purposes. For more information on backup, refer to “Backing Up the Data” as described in E4990A Online Help.
1. The default setting of the hard disk is recovered during the system recovery. For the hard disk that had failed earlier and had been replaced (after purchase), its default setting may not be the same as the factory state.
24 Keysight E4990A Impedance Analyzer
Page 25
Troubleshooting System Recovery

Procedure to execute the factory recovery

This section describes how to return the contents of the C drive to the factory state.
You need a keyboard for this operation.
Step 1. Shut down the E4990A.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E4990A.
Step 4. Press the standby switch of the E4990A to turn it ON.
Step 5. When the screen as shown in the figure below appears, select Keysight
Recovery System and press [Enter].
Figure 2-1 Keysight Recovery System selection screen
After several seconds, the next screen appears automatically even if you do not press any key.
Step 6. In the Keysight Software End User License Agreement dialog box, select the
Agree button.
Keysight E4990A Impedance Analyzer 25
Page 26
2-
Troubleshooting System Recovery
Step 7. Type 2 to select Recover Factory Backup Image option and press [Enter].
Figure 2-2 System recovery selection screen
Step 8. A confirmation message appears. Click OK to continue the process.
Figure 2-3 System recovery confirmation screen
Step 9. The progress of the system configuration is displayed on the screen. The
recovery takes a few minutes depending on the amount of data.
Figure 2-4 System configuration progress screen
26 Keysight E4990A Impedance Analyzer
Page 27
Troubleshooting System Recovery
Step 10. Once the recovery process is completed, completion message prompts. Click
OK. The E4990A restarts automatically.
Figure 2-5 Recovery completion dialog box
Step 11. After the restart, execute initial registration. For information on the execution
procedure, refer to “Initial Registration of E4990A” on page 22.
Never turn OFF the power during the system recovery because doing so may cause serious damage to the E4990A.
Keysight E4990A Impedance Analyzer 27
Page 28
2-
Troubleshooting System Recovery
28 Keysight E4990A Impedance Analyzer
Page 29
Index
A
assistance
7
D
Documentation warranty
2
E
Environmental requirements
Operating Environments Ventilation Requirement
12
12
error-correction
temperature range
exclusive remedies
12
7
F
front handles
how to install front hanles
Index
14
I
inspection
checking the shipment
11
O
option
Option 1CM Option 1CN Option 1CP
15
15
15
P
power
disconnection from supply source power cable options
17
20
verification and connection of power cable verification of the power supply
17
S
safety
symbols
6
shutdown
turning the power OFF
starting the E4990A
19
19
system
installing rack mounting flanges
14
System Recovery
how to execute factory recovery
25
17
Keysight E4990A Installation Guide
29
Page 30
Index
30
Keysight E4990A Installation Guide
Page 31
This information is subject to change without notice.
© Keysight Technologies 2014-2018
Edition 3, July 2018
*E4990-90010*
E4990-90010
www.keysight.com