No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
E4990-90010
Edition
Edition 3, July 2018
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight
products may be downloaded from
the Web. Go to
http://www.keysight.com/go/confor
mity. You can then search by product
number to find the latest Declaration
of Conformity.
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall
apply, except to the extent that those
terms, rights, or licenses are
explicitly required from all providers
of commercial computer software
pursuant to the FAR and the DFARS
and are set forth specifically in
writing elsewhere in the EULA.
Keysight shall be under no obligation
to update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Printed in Malaysia
Published by:
Keysight Technologies International
Japan G.K,
1-3-3 Higashikawasaki-cho
Chuo-ku
Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
U.S. Government Rights
The Software is “commercial
computer software,” as defined by
Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires
commercial computer software
under the same terms by which the
software is customarily provided to
the public. Accordingly, Keysight
provides the Software to U.S.
government customers under its
standard commercial license, which
is embodied in its End User License
Agreement (EULA), a copy of which
can be found at
http://www.keysight.com/find/swe
ula The license set forth in the EULA
represents the exclusive authority by
which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license
set forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation
that is not customarily provided to
the public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
This NOTE sign denotes important
information. It calls attention to a
procedure, practice, or condition that
is essential for the user to
understand.
Page 3
Caution
Keysight E4990A Impedance Analyzer
Installation Guide
Do not exceed the operating input power, voltage, and current level and
signal type appropriate for the instrument being used, refer to your
instrument’s Function Reference.
Electrostatic discharge (ESD) can damage the highly sensitive
microcircuits in your instrument. ESD damage is most likely to occur as the test
fixtures are being connected or disconnected. Protect them from ESD damage
by wearing a grounding strap that provides a high resistance path to ground.
Alternatively, ground yourself to discharge any static charge built- up by
touching the outer shell of any grounded instrument chassis before touching
the test port connectors.
Safety Summary
When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Contact your local Keysight Technologies sales representative or authorized
service company for repair of the instrument. If you continue to operate
without repairing the instrument, there is a potential fire or shock hazard to the
operator.
•Instrument operates abnormally.
•Instrument emits abnormal noise, smell, smoke or a spark- like light during
•Instrument generates high temperature or electrical shock during
•Power cable, plug, or receptacle on instrument is damaged.
•Foreign substance or liquid has fallen into the instrument.
operation.
operation.
3
Page 4
-
Caution
Manufacturer’s Declaration
Manufacturer’s Declaration
Herstellerbescheinigung
GERA- USCHEMISSION
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer's Declaration
ACOUSTIC NOISE EMISSION
LpA < 70 dB
operator position
normal operation
per ISO 7779
Regulatory Compliance Information
This product complies with the essential requirements of the following
applicable European Directives, and carries the CE marking accordingly:
•The Low Voltage Directive 2006/95/EC
•The EMC Directive 2004/108/EEC
To obtain Declaration of Conformity, please contact your local Keysight
Technologies sales office, agent or distributor.
Safety Notice Supplement
•This equipment complies with EN/IEC61010- 1:2001.
•This equipment is of MEASUREMENT CATEGORY I (CAT I). Do not use for
CAT II, III, or IV.
•This equipment is a POLLUTION DEGREE 2, INDOOR USE product.
4 Keysight E4990A Impedance Analyzer
Page 5
Caution
General Safety Precautions
•This equipment is tested in stand- alone condition and in combination with
the accessories supplied by Keysight Technologies against the requirement
of the standards described in the Declaration of Conformity. If it is used as a
system component, compliance of related regulations and safety
requirements are to be confirmed by the builder of the system.
•Do not connect the measuring terminals to mains.
General Safety Precautions
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair
the protection provided by the equipment. Such noncompliance would also
violate safety standards of design, manufacture, and intended use of the
instrument. Keysight Technologies assumes no liability for the customer’s
failure to comply with these precautions.
The E4990A complies with INSTALLATION CATEGORY II as well as
POLLUTION DEGREE 2 in IEC61010-1. The E4990A is an INDOOR USE
product.
The LEDs in the E4990A are Class 1 in accordance with IEC60825-1,
CLASS 1 LED PRODUCT.
•Ground the Instrument
To avoid electric shock, the instrument chassis and cabinet must be
grounded with the supplied power cable’s grounding prong.
•DO NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gases or
fumes. Operation of any electrical instrument in such an environment
clearly constitutes a safety hazard.
•Keep Away from Live Circuits
Operators must not remove instrument covers. Component replacement
and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected.
Under certain conditions, dangerous voltage levels may remain even after
the power cable has been disconnected. To avoid injuries, always
disconnect the power and discharge circuits before touching them.
•DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless another person,
capable of rendering first aid and resuscitation, is present.
•DO NOT Substitute Parts or Modify the Instrument
Keysight E4990A Impedance Analyzer 5
Page 6
-
Caution
Safety Symbols
To avoid the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Keysight Technologies Sales and Service Office
for service and repair to ensure that safety features are maintained in
operational condition.
•Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings
must be followed.
Dangerous voltage levels, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and adjusting
this instrument.
Safety Symbols
General definitions of safety symbols used on the instrument or in manuals are
listed below.
Instruction Manual symbol: the product is marked with this symbol when it is
necessary for the user to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
A chassis terminal; a connection to the instrument’s chassis, which includes
all exposed metal structure.
Standby.
Certification
Keysight Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Keysight Technologies
further certifies that its calibration measurements are traceable to the United
6 Keysight E4990A Impedance Analyzer
Page 7
Caution
Safety Symbols
States National Institute of Standards and Technology, to the extent allowed
by the Institution’s calibration facility or by the calibration facilities of other
International Standards Organization members.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive remedies.
Keysight Technologies shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort, or any
other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements
are available for Keysight Technologies products.
For any assistance, contact your nearest Keysight Technologies Sales and
Service Office. Addresses are provided at the back of this manual.
Keysight E4990A Impedance Analyzer 7
Page 8
-
Caution
Manuals for E4990A
Manuals for E4990A
Keysight provides the following three manuals for E4990A. The latest version of
all documentations can be downloaded from
http://www.keysight.com/find/E4990A-manual.
Installation Guide
The installation guide (this manual) provides start up setup information when
you use the E4990A for the first time and troubleshooting information when the
Windows cannot be boot up. See this manual first when you use the E4990A
for the first time.
Online Help
The online help provides the information about the quick start, measurement
operation, programming, and I/O interface. This is pre-installed in the E4990A.
Press [Help] hard key on the front panel to open. Quick Start helps in
understanding the E4990A operation quickly.
The latest version of online help is available at:
The online help has context sensitive help, which is a great feature of the
E4990A help. It allows you to get information about the selected softkey by
pressing the Help key in the E4990A or by pressing F1 in a keyboard attached
to the E4990A or by clicking the help button in a dialog box. With context
sensitive help, users can receive information quickly about the area the user is
working in the firmware of the E4990A. It provides information relevant to the
task that needs to be accomplished and reduces the time to search relevant
information required to complete a task.
Service Guide
The service manual provides information about the parts, troubleshooting,
performance test, adjustment and service menu.
This chapter provides information about how to set up Keysight E4990A
Checking the Shipment
After you receive the analyzer, carry out checks during unpacking according to
the following procedure.
When unpacking the analyzer, if the exterior of the analyzer (such as the
cover, front/rear panel, LCD screen, power switch, and port connectors)
appear to be damaged during transport, do not turn on the power switch.
Otherwise, you may get an electric shock.
Step 1. Check that the packing box or shock- absorbing material used to package the
analyzer has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave
the packing box and shock-absorbing material as is until other inspection
items are checked as follows.
Step 2. Check the packaged items supplied with the analyzer for any damage or
defects.
Step 3. By referring to the furnished contents list, check that all packaged items
supplied with the analyzer have been received as per the specified options.
Step 4. After checking, if one of the following applies, contact your nearest Keysight
Technologies sales and service office.
1. The packing box or shock- absorbing material used to package the analyzer
has been damaged or the shock- absorbing material has traces where
extreme force has been applied.
2. A packaged item supplied with the analyzer has mechanical damage or
defects.
11
Page 12
1-
Introduction
Environmental Requirements
3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the
analyzer.
If an abnormality is detected in step 1, contact the company that
transported the analyzer as well as your nearest Keysight Technologies
sales and service office.
Environmental Requirements
Set up theE4990A where the following environmental requirements are met.
Operating Environments
Ensure that the operating environment meets the following requirements.
Temperature5 ºC to 40 ºC
Temperature range at
the error-correction
Humidity20% to 80% at wet bulb temperature < +29 °C
Altitude0 to 2,000 m (0 to 6,561 feet)
Vibration0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment
of the analyzer.
Ventilation Requirement
To ensure safety requirements, the specifications and measurement accuracy
of the analyzer, you must keep the environmental temperature within the
specified range by providing appropriate cooling clearance around the analyzer
or, for the rack mount type, by forcefully air- cooling inside the rack housing.
For more information on environmental temperature to satisfy the
specifications and measurement accuracy of the analyzer, see the specification
in the E4990A Datasheet.
23 °C ± 5 °C (< 1 °C deviation from the temperature when
performing the error-correction)
(non-condensation)
When the environmental temperature around the analyzer is kept within the
temperature range of the operating environment specification (See Section
“Operating Environments” on page 12), the analyzer conforms to the
12 Keysight E4990A Impedance Analyzer
Page 13
Introduction
Environmental Requirements
requirements of the safety standard. Furthermore, under that temperature
requirement, the analyzer still conforms to the requirements of the safety
standard even when placing the analyzer with cooling clearance as follows:
Requirements
Back ≤ 180 mm
Sides≤ 60mm (both right and left)
Protection Against Electrostatic Discharge (ESD)
Set up a static- safe work- station to protect the electronic components from
the damage by the electrostatic discharge (ESD) as shown in Figure 1-1.
Figure 1-1 Example of the static-safe work station
Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of
Power Cable in Case of Emergency
As described in “Disconnection from Supply Source” on page 20, the power
supply is disconnected by removing the power cable’s connector plug from
either the AC outlet or the E4990A unit. When installing the E4990A, ensure
that there is sufficient free space around the unit to permit quick disconnection
of the plug (from AC outlet or E4990A unit) in case of emergency.
Keysight E4990A Impedance Analyzer 13
Page 14
1-
Introduction
Environmental Requirements
Installing Front Handles/Rack Mounting Flanges
The E4990A can be installed on a workbench or on a rack.
This section describes how to install the front handles (Option 1CN) used for
moving or transporting the instrument, and how to install the analyzer in an
equipment rack as part of a measurement system (Option 1CM: without the
handles, Option 1CP: with the handles).
Figure 1-2 Installing front handle/rack-mount kits
14 Keysight E4990A Impedance Analyzer
Page 15
Introduction
Environmental Requirements
Figure 1-3 E4990A with Option 1CN handle
The handle kit (Option 1CN) is used for transporting and relocating of the
E4990A.
The rack- mount kit (Option 1CM) includes two flanges (locking side plates) for
mounting the E4990A on a rack (482.6 mm/19 inches) conforming to the EIA
Standard.
Option 1CP comes with both, the handle kit and rack- mount kit.
While referring to Figure 1-2 install the handle kit/rack- mount kit by following
these steps.
Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame
of the E4990A front panel.
Step 2. Use the provided screws to mount the
•front handles (2) - for Option 1CN
•rack- mounting flange (4) - for Option 1CM
•both - for Option 1CP
on each side of the E4990A front panel frame.
Step 3. (Option 1CN/Option 1CP). Attach the modified trim strip (3) provided to each
front handle in order to cover the front panel locking screws.
Option 1CN contains 2 pieces of standard and 1 special handles E4990-20001. Only one of the standard handle is required to be used to
mount on the analyzer. Standard handle is used for the left side and
special handle is used for the right side. Refer to Figure 1-3.
Keysight E4990A Impedance Analyzer 15
Page 16
1-
Introduction
Environmental Requirements
Step 3. (Option 1CM/Option 1CP). Remove the four bottom feet of the E4990A (lift
the bar marked TAB on the inner side of the foot and slide the foot towards
the bar).
Step 4. (Option 1CM/Option 1CP). Mount the E4990A on the rack.
If the installed front handle becomes damaged, replace it with a new one
immediately. A damaged handle may break while moving or lifting the
instrument and cause personal injury or damage to the instrument.
Use both a front handle and a rack-mounting flange in the same time
certainly. Do not attempt to install flanges or handles separately with
hard ware provided. Serious electrical damage will occur to the instrument.
16 Keysight E4990A Impedance Analyzer
Page 17
Power Supply
Verification of the Power Supply
Introduction
Power Supply
Before turning on the E4990A power, check the following.
Confirm that the power supplied to the E4990A meets the following
requirements:
Requirements
Voltage
90 to 132 VAC or 198 to 264 VAC
1
Frequency47 to 63 Hz
Maximum power
300 VA
consumption
1. Switched automatically by the E4990A in conformity to the voltage used.
Verification and Connection of Power Cable
The three-wire power cable attached to the E4990A has one wire serving as the
ground. Using this power cable allows the E4990A to be grounded, thereby
protecting you against any potential electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 1-4 on page 20) on the rear panel of the E4990A and a three- wire
power outlet with the grounding prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely ground the
E4990A.
Power cord list, 16000- 99101 shows the power cable options.
Keysight E4990A Impedance Analyzer 17
Page 18
1-
Introduction
Blown Fuses
Blown Fuses
If the fuse appears to have blown during operation, this instrument may be
subject to failure and must be repaired. For any assistance, contact Keysight
Technologies Customer contact center listed at the end of this guide.
The E4990A uses the following fuse types:
UL/CSA Type, Slow- Blo, 10 A- 250 Vac.
DO NOT replace the fuse yourself as doing this may expose yourself to
electrical shock.
18 Keysight E4990A Impedance Analyzer
Page 19
Introduction
Starting the E4990A
Starting the E4990A
This section describes how to turn ON/OFF the E4990A power and how to cut
off the power supply in the case of emergency.
Turning the Power ON and OFF
The standby switch can turn ON/OFF the E4990A. The color on the button
shows the status, as indicated below:
Indicator ColorDescription
GreenNormal power on status
OrangeStandby status
RedIllegal power on status
Perform the following steps to turn the power ON or OFF.
Turning the Power ON
Step 1. Confirm if the Line Switch on the rear panel is turned ON. The switch should
always be turned ON.
Step 2. Press the standby switch once. It changes to green color.
This operation turns ON the power, and the E4990A starts the self- test.
Step 3. Confirm that the self- test indicates normal operation.
Normal operation is confirmed by the self- test if no error message appears.
Turning the Power OFF
Step 1. To turn OFF the power of the E4990A, first, press this standby switch or send
a shutdown command from the external controller to activate the shutdown
process (the process of software and hardware necessary to turn OFF the
power supply). This puts the E4990A into the standby state and the button
changes to orange color.
Keysight E4990A Impedance Analyzer 19
Page 20
1-
Introduction
Starting the E4990A
Step 2. Next, if necessary, turn OFF the power supply to the power cable receptacle
(Line Switch) on the rear panel.
Under normal use, never directly interrupt the power supply to the power cable receptacle on the rear panel when the power supply is turned ON.
Always keep the Line Switch ON and never turn it.
If you directly interrupt the power supply to the power cable receptacle
when the power supply is turned ON, or turn OFF the Line Switch, the
shutdown process will not work. This could damage the software and
hardware of the E4990A and lead to device failure.
Turning ON the power supply after a faulty shutdown may cause the
system to start up in a condition called "safe mode." If this occurs, first
shut down the system to set it to the standby state and then turn ON the
power supply again to start up the system in normal mode.
Figure 1-4 Line switch (Always ON) and power cable receptacle
Disconnection from Supply Source
The power supply of the E4990A is cut off by disconnecting the plug of the
power cable (on either AC outlet side or E4990A side). When it is necessary to
disconnect the power supply in order to avoid shock hazards, etc., pull out the
power cable plug from either the AC outlet side or the E4990A side.
20 Keysight E4990A Impedance Analyzer
Page 21
Introduction
Starting the E4990A
To allow this operation to be performed smoothly, be sure to follow the
guidelines in “Ensuring Adequate Free Space around Analyzer for
Immediate Disconnection of Power Cable in Case of Emergency” on
page 13.
When turning the power OFF under normal circumstances, always follow
the methods described in “Turning the Power OFF” on page 19.
Keysight E4990A Impedance Analyzer 21
Page 22
1-
Introduction
Initial Registration of E4990A
Initial Registration of E4990A
When you start up the E4990A for the first time, you need to perform the initial
registration of the Windows operating system of the E4990A.
When the Windows agreement message box appears, click Agree. The rest of
the procedure runs automatically.
The Touch Screen does not require initial calibration to work for the very
first time. However, user have the option of calibrating the touch screen
whenever required. To do so, select System hardkey followed by Misc
Setup > Display Setup > Ad just Touch Screen. Refer to Online Help for
more details on how to adjust the touch screen.
22 Keysight E4990A Impedance Analyzer
Page 23
Keysight E4990A Impedance Analyzer
Installation Guide
2Troubleshooting
This chapter describes the troubleshooting process during start up and the
procedure of the operating system (OS) recovery when the Windows OS has
been damaged.
Troubleshooting during Startup
When you encounter problems during start up, see Table 2-1. System recovery
saves most of problems.
Table 2-1 Troubleshooting during startup
SymptomSolution
Turning ON ( | ) the standby switch does not
start up the system.
The system starts up, but it automatically
shuts down immediately.
The system starts up, but it enters the
service mode (The status bar at the lower
right of the screen displays SVC in red).
The measurement screen appears after
start up, but the power-on test fails with
Error Message 241 prompting on a red
background in the instrument’s
message/warning area at the lower left of
the screen.
Confirm that the power cable is properly
plugged in.
Confirm that the line switch at the rear
panel is turned ON.
Execute the system recovery.
23
Page 24
2-
Troubleshooting
System Recovery
System Recovery
By executing system recovery, you can return the E4990A’s system (the
Windows operating system and the firmware) to the factory state (at the time
1
of purchase
Strictly follow the steps described below. If you do any operation other
than the following steps, the system may not be recovered.
Notes on executing the factory recovery function
Executing the factory recovery function causes the following conditions:
•In addition to the Windows operating system and the firmware, the
following settings of the E4990A are returned to the factory state.
—Network setting
).
— GPIB setting
—Printer setting
•The driver for the supported printer which is installed after purchase is
deleted.
•You need to execute initial registration again.
Files that you have created using the save function (files in the D drive) are not
affected. However, Keysight recommends backing them up before executing
system recovery for precautionary purposes. For more information on backup,
refer to “Backing Up the Data” as described in E4990A Online Help.
1. The default setting of the hard disk is recovered during the system recovery. For the hard
disk that had failed earlier and had been replaced (after purchase), its default setting
may not be the same as the factory state.
24 Keysight E4990A Impedance Analyzer
Page 25
Troubleshooting
System Recovery
Procedure to execute the factory recovery
This section describes how to return the contents of the C drive to the factory
state.
You need a keyboard for this operation.
Step 1. Shut down the E4990A.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E4990A.
Step 4. Press the standby switch of the E4990A to turn it ON.
Step 5. When the screen as shown in the figure below appears, select Keysight
Recovery System and press [Enter].
Figure 2-1 Keysight Recovery System selection screen
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 6. In the Keysight Software End User License Agreement dialog box, select the
Agree button.
Keysight E4990A Impedance Analyzer 25
Page 26
2-
Troubleshooting
System Recovery
Step 7. Type 2 to select Recover Factory Backup Image option and press [Enter].
Figure 2-2 System recovery selection screen
Step 8. A confirmation message appears. Click OK to continue the process.
Figure 2-3 System recovery confirmation screen
Step 9. The progress of the system configuration is displayed on the screen. The
recovery takes a few minutes depending on the amount of data.
Figure 2-4 System configuration progress screen
26 Keysight E4990A Impedance Analyzer
Page 27
Troubleshooting
System Recovery
Step 10. Once the recovery process is completed, completion message prompts. Click
OK. The E4990A restarts automatically.
Figure 2-5 Recovery completion dialog box
Step 11. After the restart, execute initial registration. For information on the execution
procedure, refer to “Initial Registration of E4990A” on page 22.
Never turn OFF the power during the system recovery because doing so
may cause serious damage to the E4990A.
Keysight E4990A Impedance Analyzer 27
Page 28
2-
Troubleshooting
System Recovery
28 Keysight E4990A Impedance Analyzer
Page 29
Index
A
assistance
7
D
Documentation warranty
2
E
Environmental requirements
Operating Environments
Ventilation Requirement
12
12
error-correction
temperature range
exclusive remedies
12
7
F
front handles
how to install front hanles
Index
14
I
inspection
checking the shipment
11
O
option
Option 1CM
Option 1CN
Option 1CP
15
15
15
P
power
disconnection from supply source
power cable options
17
20
verification and connection of power cable
verification of the power supply
17
S
safety
symbols
6
shutdown
turning the power OFF
starting the E4990A
19
19
system
installing rack mounting flanges
14
System Recovery
how to execute factory recovery
25
17
Keysight E4990A Installation Guide
29
Page 30
Index
30
Keysight E4990A Installation Guide
Page 31
This information is subject to change without notice.