No part of this manual may be reproduced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement
and written consent from Keysight
Technologies, Inc. as governed by
United States and international
copyright laws.
Manual Part Number
E4981-90000
Edition
Edition 3,September 26, 2019
Printed in Malaysia
Published by:
Keysight Technologies International
Japan G.K,
1-3-3 Higashikawasaki-cho
Chuo-ku
Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE
LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR
ANY INFORMATION CONTAINED
HEREIN. SHOULD KEYSIGHT AND
THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY
TERMS COVERING THE MATERIAL
IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPA-
RATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web. Go to http://www.key-
sight.com/go/conformity. You can
then search by product number to
find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”)
2.101. Pursuant to FAR 12.212 and
27.405-3 and Department of Defense
FAR Supplement (“DFARS”)
227.7202, the U.S. government
acquires commercial computer software under the same terms by which
the software is customarily provided
to the public. Accordingly, Keysight
provides the Software to U.S. government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which
can be found at http://www.key-
sight.com/find/sweula
set forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify, distribute, or disclose the Software. The
EULA and the license set forth
therein, does not require or permit,
among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software documentation that is not
customarily provided to the public; or
(2) Relinquish to, or otherwise provide, the government rights in excess
of these rights customarily provided
to the public to use, modify, reproduce, release, perform, display, or
disclose commercial computer software or commercial computer software documentation. No additional
government requirements beyond
those set forth in the EULA shall
.The license
apply, except to the extent that those
terms, rights, or licenses are explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and
are set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the
like that, if not correctly performed or adhered to, could
result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION
notice until the indicated conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the
like that, if not correctly performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARNING notice until the indicated
conditions are fully understood
and met.
10Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
User Manual
1 Unpacking and Preparation
This chapter describes how to set up and start the Keysight E4981A
Capacitance Meter.
Contents of this Chapter
•“Checking the Shipment” on page 12
After you receive the E4981A, check all the items in the packing
container.
•“Environmental Requirements” on page 13
Describes the system requirements needed to install the E4981A and
how to secure space for heat radiation.
•“How to Remove the Handle” on page 16
Shows how to attach and remove the handle.
•“Preparations Before Use” on page 18
Shows how to check the power supply as well as check and connect
the power cable. This section also describes how to deal with a blown
fuse.
•“Starting the E4981A” on page 20
Describes how to turn on/off the power switch and cut off the power
supply.
1-
Unpacking and Preparation
Checking the Shipment
Checking the Shipment
After you receive the E4981A, carry out checks during unpacking according to
the following procedure.
If the external face of the Capacitance meter (such as the cover, front/rear
panel, LCD screen, power switch, and port connectors) appears to have
been damaged during transport, do not turn on the power switch.
Otherwise, you may get an electrical shock.
Step 1.Check that the packing box or shock-absorbing material used to package the
Capacitance meter has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave
the packing box and shock-absorbing material as is until other inspection
items are checked as follows.
Step 2.Check the packaged items supplied with the Capacitance meter for any
damage or defects.
Step 3.By referring to the packing list, check that all packaged items supplied with the
Capacitance meter have been provided as per the specified options.
Step 4.After checking, if one of the following applies, contact your nearest Keysight
Technologies sales and service office.
•The packing box or shock-absorbing material used to package the
Capacitance meter has been damaged or the shock-absorbing material
displays traces of where extreme pressure has been applied.
•A packaged item supplied with the Capacitance meter has mechanical
damage or defects.
•A packaged item supplied with the Capacitance meter is missing.
•A fault has been detected in the subsequent operation check of the
Capacitance meter.
If an abnormality is detected in Step 1, contact the company that transported
the Capacitance meter as well as your nearest Keysight Technologies sales and
service office. For inspection by the transport company, save the packing box,
shock-absorbing material, and packaged items as you received them.
12Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Unpacking and Preparation
Environmental Requirements
Environmental Requirements
Set up the E4981A where the following environmental requirements are
satisfied.
Operating environment
Ensure that the operating environment meets the following requirements.
Table 1-1 Operating environment requirements
Temperature0 °C to 45 °C
Temperature range at
calibration
Humidity (~40 °C, no
condensation)
Altitude0 to 2,000 m (0 to 6,561 feet)
VibrationMax. 0.5 G, 5 Hz to 500 Hz
The environmental requirements listed above are NOT for the specifications
and measurement accuracy of the E4981A, but for its operating
environment.
Ventilation requirements
To ensure that the safety requirements, the specifications, and the
measurement accuracy of the Capacitance meter are met, you must maintain
the environmental temperature to within the specified range by providing an
appropriate cooling clearance around the Capacitance meter or, for the
rack-mounted type, by forcefully air-cooling inside the rack housing. For more
information on environmental temperatures that satisfy the specifications and
measurement accuracy of the Capacitance meter, see Chapter 10, “Specifications
and Supplemental Information,” on page 167.
23 °C ± 5 °C (|1 °C deviation from the temperature when
performing calibration)
15% to 85% RH
The Capacitance meter conforms to the requirements of the safety
standard.When the environmental temperature around the Capacitance meter
is kept within the temperature range of the operating environment
specifications (“Operating environment” on page 13).
The Capacitance meter still conforms to the requirements of the safety
standard it is installed with the following cooling clearance:
Unpacking and Preparation
Environmental Requirements
Figure 1-1 Ventilation space at installation
Protection against electrostatic discharge (ESD)
Figure 1-2 shows the setup for a static-safe work station to protect the
electronic components from damage by electrostatic discharge (ESD).
Figure 1-2 Example of a static-free work station
14Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Unpacking and Preparation
Environmental Requirements
Ensuring adequate free space around the capacitance meter for immediate
disconnection of the power cable in case of emergency
As described in “Disconnecting from the supply source” on page 21, the plug
attached to the power cable serves as the disconnecting device (device that
cuts off the power supply) for the E4981A. When installing the E4981A, ensure
that there is sufficient free space around the unit to permit quick disconnection
of the plug (from the AC outlet or the E4981A unit) in case of emergency.
Unpacking and Preparation
How to Remove the Handle
How to Remove the Handle
A handle kit is attached to the E4981A Option 600. When using the E4981A
with the rack-mount kit, remove the handle according to the following steps.
Figure 1-3 How to remove the handle
e4981aue0021
Step 1.Lift the handle perpendicular to the unit while pulling it in the direction of 1.
Step 2.While pulling the handle in the direction of 1, lift towards 2.
Install the handle with 3 facing toward the front. Installing the handle the
other way around may damage it.
16Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Unpacking and Preparation
How to Remove the Handle
Caution when using the handle
Follow the instructions below when using the E4981A’s handle, otherwise you
may get your fingers caught in the handle or the E4981A may fall and be
damaged
•When the handle is set up as shown in Figure 1-4, do not put any additional
weight onto the E4981A or lift it suddenly.
•Do not touch the handle while the DUT is attached.
Confirm that the power supplied to the E4981A meets the following
requirements:
Table 1-2 Power supply requirements
Requirements
Voltage90 to 264 Vac
Frequency47 to 63 Hz
Maximum power consumption150 VA
Setting up the fuse
Please use the following fuse type.
UL/CSA type, Slo-Blo, 5×20-mm miniature fuse, 3 A, 250 V (Keysight part
number 2110-1017)
When you need a fuse, contact your nearest Keysight Technologies sales and
service office. To verify and replace the fuse, remove the power cable and pull
out the fuse holder.
Figure 1-5 Fuse holder and power cable receptacle
18Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Unpacking and Preparation
Preparations Before Use
Verifying and connecting the power cable
The three-wire power cable attached to the E4981A has one wire serving as a
ground. Using this power cable allows the E4981A to be grounded, thereby
protecting you against electrical shock from the power outlet.
Step 1.Confirm that the power cable is not damaged.
Never use a power cable showing any sign of damage. Faulty cables can
cause electric shock.
Step 2.Use the supplied cable to connect the power cable receptacle on the real panel
of the E4981A to a three-wire power outlet with the grounding prong firmly
held in the ground slot.
Use the supplied three-wire power cable with a grounding wire to
securely ground the E4981A.
A 3p-2p conversion adapter is not supplied with the Capacitance meter.
When you need a 3p-2p conversion adapter, contact your nearest Keysight
Technologies sales and service office listed in the back of this manual.
Power cord list, 16000-99101 shows the power cable options.
This section describes how to turn on/off the E4981A power and how to cut off
the power supply in an emergency.
Turning the power ON and OFF
Turning the power ON
The ON and OFF status of the power switch is confirmed by the color of the
switch.
Table 1-3 Power switch status indicator
Color of the lightStatus of the power
OrangePower off
Yellow-greenPower on
Light offPower off (The power supply is cut off)
Step 1.Confirm that the power switch light in the lower-left part of the front panel
is lit in orange. If the light is off, the power supply may also be off.
Step 2.Press the power switch. When the light of the power switch turns yellow-green,
the power is turned ON and the E4981A starts the self-test.
The self-test takes approx. 30 seconds.
Step 3.Confirm that the self-test indicates normal operation.
Normal operation is confirmed by the self-test if no error message appears.
Turning the power OFF
Follow the method below to turn the power OFF.
Step 1.Press the power switch in the lower-left part of the front panel.
Do not turn the power OFF while saving into or recalling from the internal
memory of the E4981A or the USB memory. Doing so may clear the
contents from the memory.
20Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Unpacking and Preparation
Starting the E4981A
Disconnecting from the supply source
The plug attached to the power cable (on the power outlet side or device side
of the cable) serves as the disconnecting device (device that cuts off the power
supply) of the E4981A. When the power supply must be cut off to avoid such
danger as electric shock, pull out the power cable’s plug (on the power outlet
side or device side of the cable).
To allow this operation to be performed smoothly, be sure to follow the
guidelines in “Ensuring adequate free space around the capacitance meter
for immediate disconnection of the power cable in case of emergency” on
page 15.
When turning the power OFF under normal circumstances, always follow
the methods described in “Turning the power OFF” on page 20.
22Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
User Manual
2 Overview
This chapter provides the basic procedures for operating the E4981A and
describes names and functions of the front panel, rear panel, and screen
display.
Product Introduction
The Keysight E4981A is a capacitance meter for ceramic capacitor production
tests. The E4981A is used for evaluating ceramic capacitance at frequencies
120 Hz, 1 kHz to 1 MHz and test signal levels (0.1 mV to 1 V).
The E4981A offers C-D measurement with a basic accuracy of ±0.07% (C),
±0.0005 (D) at all frequencies with seven-digit resolution (the dissipation factor
–6
resolution is 1×10
With its built-in comparator, the E4981A can output comparison/decision
results for sorting components into a maximum of ten bins. Furthermore, by
using the handler interface and scanner interface, the E4981A can be easily
combined with a component handler, a scanner, and a system controller to
fully automate component testing, sorting, and quality-control data
processing.
The GPIB/LAN/USB interfaces are standard on the E4981A and enable
automatic testing.
) in every range.
2-
Overview
Front Panel: Names and Functions of Parts
Front Panel: Names and Functions of Parts
This section describes the names and functions of the parts on the E4981A’s
front panel. For more details on the functions displayed on the LCD screen, see
“Screen Area: Names and Functions of Parts” on page 30.
Figure 2-1 Front panel
1. Power switch
Used for choosing between power-on and -off states of the E4981A. When
turned on, the switch lights up in yellow-green and all operating voltages are
applied to the instrument. When turned off, the switch lights up in orange and
no operating voltages are applied to the instrument.
2. LCD
The Liquid Crystal Display (LCD) displays measurement results, test conditions,
etc.
Occasionally there are missing pixels or constantly lit pixels, but this is not
a malfunction and does not affect the performance of your product.
3. Softkeys
Five softkeys are used to select measurement conditions and parameter
functions. Each softkey has a softkey label along its left side.
24Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Overview
Front Panel: Names and Functions of Parts
4. Menu keys
Menu selection keys are used to access the corresponding selection of
instrument controls.
Table 2-1 Menu keys description
[Display Format] keyDisplays measurement results and selections.
[Meas Setup] keySets measurement conditions, a correction function,
[Recall A] keyRecalls the setting information on internal memory 0.
[Recall B] keyRecalls the setting information on internal memory 1.
[Save/Recall] keySaves and recalls the setting information. Saves
[System] keySets the system. Sets the GPIB/LAN/USB interface.
contact check function, and limit values for BIN sorting.
measurement results and screen images.
Performs a self-test.
[Local/Lock] keyLocks and unlocks hard keys and softkeys on the front
5. Cursor keys
Keys used to move the field select cursor from field to field on a displayed
page. When the cursor is moved to a certain field, that field changes to an
inverse video image of the original field. The cursor can only be moved from
field to field.
For the skip key, refer to the “How to use skip keys” on page 33.
6. Entry keys
Keys used to enter numeric data into the E4981A. The entry keys comprise the
digits 0 to 9, a period (.), and a plus/minus (+/–) sign. Entered values are
displayed on the input line (second line from the bottom of the LCD screen),
and pressing the softkey terminates numeric input. The plus/minus key deletes
the last character of the input value.
7. Preset key
A key used to return the Capacitance meter to the initial setup state. There are
three methods for initialization. For details, refer to “Initializing the Instrument”
on page 39.
panel.
8. Trigger key
A key used to manually trigger the E4981A when it is set to the manual trigger
mode.
Overview
Front Panel: Names and Functions of Parts
9. UNKNOWN terminals
These are the UNKNOWN terminals used to connect a four-terminal pair test
fixture or test leads for measuring the device under test (DUT).
When using a four-terminal pair test fixture or test leads with a stopper,
remove the stopper or the bumper of the E4981A.
Do not apply DC voltage or current to the UNKNOWN terminals. Applying
DC voltage or current may lead to device failure. Connect the
measurement sample (DUT) to the test port (or the test fixture, cables, etc.
connected to the test port) after the Capacitance meter has been
completely discharged.
The maximum load the UNKNOWN terminals can withstand is 10 kg (nominal).
The test ports comply with Installation Category I of IEC 61010-1.
10. Front USB port
This port is used to save data in a USB memory.
Connector types: Universal Serial Bus (USB) Jack, Type A (four contact points),
Female.
Compliance Standards: USB 1.1
Do not connect any device other than a USB memory to the USB port. We
do not support connections to the USB port of printers, devices with a
built-in HDD, or USB hubs. To print a screen, refer to “Saving a screenshot
into USB memory” on page 89.
Do not unplug the USB memory while the USB indicator is on.
USB memory type
Use USB memory that is USB mass-storage-class compliant and formatted
with FAT16 or FAT32. For points to notice, refer to “USB memory notes” on
page 78.
Interface: USB 1.1
11. Ground terminal
Connected to the chassis of the E4981A.
26Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Overview
Rear Panel: Names and Functions of Parts
12. Synchronous clock
The internal clock signal is outputted.
Rear Panel: Names and Functions of Parts
This section describes the names and functions of the parts on the E4981A’s
rear panel.
Figure 2-2 Rear panel
1. Handler interface
This interface is for data exchange with an automatic machine (handler) used
on a production line.
Connector type: 36-pin Centronics
2. USB (USBTMC) interface port
Through this port, you can control the E4981A via external controllers.
Connector Types: Universal Serial Bus (USB) jack, type mini-B (five contact
points), Female Compliance Standards: USBTMC-USB488 and USB2.0
The USB port of the E4981A provides (screw holes) to secure the USB cable
with a custom make bracket. The figure describes the dimension required for
the custom bracket. The distance between the surfaces of the rear panel and
USB connector is 2 mm. The length of the screw should be less than 10 mm.
The port to connect the E4981A to a LAN (Local Area Network). Connecting
this instrument to a LAN enables you to control this instrument by using
SICL-LAN or telnet, or from an external PC via a Web server.
This is compliant with LXI standard (LAN eXtensions for Instrumentation):
version 1.2 Class C.
The BNC connector to input the positive/negative TTL pulse to trigger the
E4981A with external trigger signals.(The trigger mode must be set to
EXTernal.)
5. Serial number plate
The seal showing the product’s serial number.
28Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Overview
Rear Panel: Names and Functions of Parts
6. Power cable receptacle (to LINE)
The receptacle (outlet) to which the power cable is connected.
To connect the device to a power source (outlet), use the supplied
three-prong power cable with a ground conductor.
The plug attached to the power cable (on the power outlet side or device
side of the cable) serves as the E4981A’s disconnecting device (device that
cuts off the power supply). When the power supply must be cut off to avoid
such danger as electric shock, pull out the power cable plug (on the power
outlet side or device side of the cable). For the procedure for turning off the
power in normal use, see the description in “1. Power switch” on page 24.
For more on the power supply, see “Verifying the power supply” on page 18.
7. Fan
The cooling fan for controlling the temperature inside the E4981A. This fan
extracts heated air from inside the Capacitance meter.
8. Scanner interface
The interface to connect a scanner to perform up to 256 sets of multi-channel
corrections and measurements.
Connector type: 14-pin Amphenol
9. GPIB interface connector
General Purpose Interface Bus (GPIB). The connection of an external controller
and other devices through this connector allows you to configure an automatic
measurement system.
Overview
Screen Area: Names and Functions of Parts
Screen Area: Names and Functions of Parts
This section describes the names and functions of parts on the E4981A’s LCD
screen.
Figure 2-3 Screen display
1. Display page area
Shows a display page name of the current display page.
2. Comment line area
You can input up to 30 characters in ASCII format by using the front panel or
the DISPlay:LINE command of the SCPI command. The first 22 characters are
displayed in this area.
Comment line is displayed on MEAS DISPLAY, BIN No. DISPLAY, BIN COUNT
DISPLAY and MEAS SETUP display page.
For the input method in the comment line, refer to “Inputting Comment Line”
on page 53.
30Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Overview
Screen Area: Names and Functions of Parts
3. Softkey area
Displays softkey labels corresponding to the field.
A displayed to the right of a softkey indicates that pressing that softkey will
display the softkey label one level lower.
Pressing the Return key when the lower level softkey label is displayed will
display the softkey label one level higher. In this case, a is displayed on the
label to the left of the Return key.
Figure 2-4 Softkey area
4. Measurement data/Conditions area
Displays measurement conditions and measurement results.
5. Input line area
Displays numeric values entered with the entry key.
6. System message area
Displays a system message, warning, and an error message.
7. Status display area
When the front panel key is locked, “LOCK” is displayed in this area. When
sending SCPI commands from an external controller, “RMT” is displayed and
the front panel keys are locked. When E4981A accesses the USB memory,
“USB” is displayed in this area.
The E4981A’s basic operations are described below:
1. Display the desired page using both the MENU keys and softkeys.
2. Move the cursor to the desired field using the cursor keys. When the cursor
is moved to a certain field, the field changes to an inverse video image of
the original. The cursor can be moved from field to field (right and left or up
and down).
3. The softkey labels corresponding to the field indicated by the cursor are
displayed automatically. Press the desired softkey.
Use the entry keys to input numeric data. When one of the entry keys is
pressed, the softkeys will change to the available unit softkeys. Pressing
these unit softkeys terminates numeric input.
The unit changes according to the field selected.
How to use cursor keys
Move the cursor to the desired field using the cursor keys as shown in Figure
2-5.
Figure 2-5 Cursor keys and a field operation example
32Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Overview
Basic Operation
How to use skip keys
The following describes how to use a skip key.
On a display page, three lines are collected together as one. By using a skip
key, you can select the desired field quickly as the selected field moves from
area to area.
34Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
User Manual
3 Getting Started
This chapter describes the basic operations of the Keysight E4981A. “Key
Operation Basics” on page 35 explains how to use the basic key operations.
“Learning Basic Measurement Procedure” on page 38 describes how to take
basic measurements of capacitors (capacitance) with the E4981A.
Key Operation Basics
Three types of operations can take place when you press a front panel key of
the E4981A:
•A single key press completes the operation.
•A menu (for item selection) appears on the screen.
•A value entry screen appears.
The basic operation for each case is described below:
A single key press completes the operation
Single-key operation turns ON/OFF the function or changes the setup.
Specifically, each key press turns the setup ON/OFF. Two examples are given
below:
•Pressing the [Recall A] key recalls configuration states from register 0.
•Pressing the [Local/Lock] key locks and unlocks hard keys and softkeys on
the front panel.
A menu (for item selection) appears on the screen
In this case, pressing a key changes the display of softkey area for selecting
items. Use the cursor keys or menu keys to change the selection. When the
desired item is selected, press the softkey to confirm the selection. An example
is given below.
Step 1.Press the [Meas Setup] key.
Step 2.Use the cursor keys to select the TRIG field.
Step 3.Press the MAN softkey to set the trigger type to manual.
3-
Getting Started
Key Operation Basics
A value entry screen appears
In this case, pressing a key changes the screen display from the measurement
screen to the value entry screen. Use the following keys to enter a value:
Keys used to enter a value. The [+/- <-] key deletes the last character of the
input value.
Softkeys
Use this key to terminate the input.
You can enter values by using only the entry keys and the softkeys.
However, depending on the instrument setup state when you try to enter a
value, some of these keys may be unavailable.
Example of operation used to enter a value
To learn how to enter a value, follow the steps below to change the
measurement range setup the measurement range of 220 pF (when the
measurement frequency is 1 kHz) in the following two ways:
−]
•entry using the numeric keys and softkey, and
•entry using only the softkeys.
Each procedure is described below.
Step 1.Press the [Meas Setup] key.
Step 2.Use the cursor keys to select the RANGE field, as shown in Figure 3-1.
Figure 3-1 Measurement setup screen
36Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Key Operation Basics
Step 3.Enter 220 pF.
Entry using the numeric entry keys and softkeys:
When data is entered with the numeric entry keys, the softkeys change
to units labels (pF, nF, mF, mF, F).
Press the [2] key, [2] key, and [0] key. Then, press the pF softkey.
Entry using only the softkeys:
SoftkeyDescription
INCR+Increments the measurement range in the HOLD mode.
DECR-Decrements the measurement range in the HOLD mode.
Getting Started
Learning Basic Measurement Procedure
Learning Basic Measurement Procedure
This section describes the procedure used to measure a capacitor with the test
fixture. This description is intended to help you learn the basic measurement
procedure of the E4981A.
Connecting test fixture
To avoid failure, do not apply DC voltage or current to the UNKNOWN
terminal. Special care must be taken for capacitors because they may be
charged. Fully discharge DUTs before connecting them to the UNKNOWN
terminal (or the test fixture).
It is difficult to connect the DUT (capacitor) directly to the E4981A. Therefore, a
test fixture is generally used to connect the DUT to the E4981A. Use the test
fixture that is suitable for the shape of the DUT. Figure 3-2 shows an example
of making a connection by using the 16034G, which is suitable for measuring
chip capacitors.
Figure 3-2 Connecting test fixture (16034G)
38Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Learning Basic Measurement Procedure
Setting up basic measurement conditions
This section describes how to set up basic measurement conditions for
capacitor measurement. For an actual measurement (for example,
measurement that provides better accuracy or measurement based on
measurement time), you need to set up measurement conditions that are
appropriate to the task, not necessarily those described here.
Initializing the Instrument
This section describes how to return the settings of the E4981A to their preset
values.
Step 1.Press the [Preset] key.
Step 2.Use the softkeys to select one of the following four default states:
Table 3-1 Four default states of E4981A and how it is initialized
Default stateHow the instrument is initialized
CLEAR SETTINGWhen you initialize the instrument into this state, all
basic parameters configurable through the front panel
and SCPI commands are cleared. (You get the same
result by issuing the :SYST:PRES command).
CLEAR SET&CORR When you initialize the instrument into this state,
calibration data listed in the initial setting list are all
1
cleared.
(You can get the same result by issuing the
*RST command).
FACTORY DEFAULT When you initialize the instrument into this state, it
reverts to factory default settings with all
user-configurable data cleared.
LAN RESETWhen you initialize the instrument into this state, the
LAN setting is returned to the factory default state.
1. It takes a few seconds for the initialization to complete.
For more information on each default state and affected settings, see
Getting Started
Learning Basic Measurement Procedure
Setting up measurement parameters
Set up the primary parameter and secondary parameter you want to measure.
The E4981A allows you to select the parameters from the following
combinations.
Primary parameterSecondary parameter
CpD, Q, G, Rp
CsD, Q, Rs
Each parameter is described below.
Cp: Capacitance value when considering equivalent parallel
resistance
Cs: Capacitance value when considering equivalent series
resistance
D: Dissipation factor
Q: Quality factor (inverse of D)
G: Equivalent parallel conductance
Rp: Equivalent parallel resistance
Rs: Equivalent series resistance
The procedure to set up the measurement parameters is described below.
Step 1.Press the [Display format] key.
Step 2.Press the MEAS DISPLAY softkey.
Step 3.Use the cursor key to select the FUNC field (as shown in Figure 3-3).
Step 4.Press the Cp or Cs softkey to select the primary parameter.
Figure 3-3 Primary parameter selection menu screen
Step 5.Press the softkey to select the secondary parameter (as shown in Figure 3-4).
40Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Learning Basic Measurement Procedure
Figure 3-4 Secondary parameter selection menu screen
Setting up measurement signal frequency
Set up the frequency of the signal applied to the DUT (capacitor) during
measurement.
Step 1.Press the [Display Format] key.
Step 2.Press the MEAS DISPLAY softkey.
Step 3.Use the cursor key to select the FREQ field.
Step 4.Use the softkeys or entry keys to enter the frequency. When data is entered
with the numeric entry keys, the softkeys change to unit labels (Hz, kHz, MHz)
for Option 001 and (Hz, kHz) for Option 002.
Figure 3-5 Measurement signal frequency selection menu screen
Getting Started
Learning Basic Measurement Procedure
Executing measurement to compensate errors
You can compensate for errors in the measurement caused by disturbances
such as stray admittance and residual impedance of the test fixture and cable.
Corrections should be performed before actually connecting the DUT to the
test fixture.
Measuring data for OPEN correction
The OPEN correction is provided to remove stray admittance parallel to the
DUT. The procedure to measure the data for OPEN correction is described
below.
Step 1.Press the [Meas Setup] key.
Step 2.Press the CORRECTION softkey. The CORRECTION display page as shown in
Figure 3-8 appears.
Figure 3-8 Correction screen
Step 3.Use the cursor keys to select the OPEN field.
Step 4.Connect the UNKNOWN terminal and the test fixture with no DUT connected.
Step 5.Press the MEAS OPEN softkey. The data for the OPEN correction is then
measured.
•During the measurement, an “OPEN measurement in progress” message is
shown on the display.
•When the measurement has finished, the “OPEN measurement in progress”
message disappears.
•During the measurement, the ABORT
softkey is shown. Use this key when
you want to abort open correction.
The data for OPEN correction is stored as data for all the measurement
frequencies.
44Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
G2B2+
Learning Basic Measurement Procedure
Step 6.When measurement of the data for OPEN correction is successfully completed,
the OPEN correction turns to ON (as shown in Figure 3-9).
Figure 3-9 Screen upon completion of measuring data for OPEN correction
Description of softkeys
To enable/disable or otherwise control the behavior of open correction, use the
following softkeys:
SoftkeyDescription
ONEnables open correction.
OFFDisables open correction.
MEAS OPENStarts open correction.
If open admittance |Yo| (=) is greater than 20 µS (an unsuitable level for the
data used in OPEN correction), the warning message “Out of limit” appears on the
System message area on the screen.
Even if this warning message appears, the data for OPEN correction is still
used. However, you should recheck the connection between the test fixture
and the UNKNOWN terminal and confirm that the OPEN correction
procedure was done correctly.
If a measurement failure occurs while measuring the data for correction, the
error message “OPEN Measurement incomplete” occurs.
If this error occurs, the data for correction before the measurement remains
without change.
Getting Started
Learning Basic Measurement Procedure
Description of softkeys
To enable/disable or otherwise control the behavior of short correction, use the
following softkeys:
SoftkeyDescription
ONEnables short correction.
OFFDisables short correction.
MEAS
Starts short correction.
SHORT
If SHORT impedance |Zs| (=) is greater than 20 Ω (an unsuitable level
for the data used in SHORT correction), the warning message “Out of limit”
appears on the System message area on the screen
Even if this warning message appears, the data for SHORT correction is still
used. However, you should recheck the connection between the test fixture
and the UNKNOWN terminal and confirm that the SHORT correction
procedure was done correctly.
If a measurement failure occurs while measuring the data for correction, an error
message appears on the screen.
If an error occurs, the data for correction before the measurement remains
without change.
48Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Learning Basic Measurement Procedure
Checking data for SHORT correction
You can check the measured SHORT impedance value (the data for SHORT
correction). The procedure is given below.
Step 1.Press the [Meas Setup] key.
Step 2.Press the CORRECTION softkey.
Step 3.Use the cursor keys to select the SHORT R-X field (as shown in Figure 3-12).
Figure 3-12 SHORT correction data
Step 4.Use the following softkeys:
SoftkeyDescription
R-XDisplays the measured R-X value on the screen.
Ls-RsDisplays the measured Ls-Rs value on the screen.
For information on measuring data for Load correction, refer to “Obtaining
(measuring) data for LOAD correction” on page 102.
For checking data for Load correction, refer to “Checking
displaying/setting up data for LOAD correction” on page 111.
Getting Started
Learning Basic Measurement Procedure
Connecting the DUT (capacitor)
Mount a capacitor on the test fixture. The measurement result for the
parameter selected in “Setting up measurement parameters” on page 40 is
displayed. Figure 3-13 shows an example when the primary parameter is Cp
and the secondary parameter is D.
Figure 3-13 Measurement result display screen (when primary parameter is Cp and
secondary parameter is D)
50Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
User Manual
4 Setting up Measurement Conditions and Display
This chapter describes how to initialize the instrument, set up the
measurement conditions and display. It also describes how to save/recall the
instrument setup state, including the measurement conditions.
Selecting Measurement Parameters
You can select one of the measurement parameter combinations shown in
Table 4-1.
Table 4-1 Measurement parameters
Primary parameterSecondary parameter
CpD, Q, G, Rp
CsD, Q, Rs
Each parameter is described below.
Cp: Capacitance value measured using the parallel equivalent circuit model
Cs: Capacitance value measured using the series equivalent circuit model
D:Dissipation factor
Q:Quality factor (inverse of D)
G:Equivalent parallel conductance measured using the parallel equivalent circuit model
Rp: Equivalent parallel resistance measured using the parallel equivalent circuit model
Rs: Equivalent series resistance measured using the series equivalent circuit model
4-
e4981aue0049
Cp
CsRs
Rp
D =
2Ǹf Cp Rp
1
=
2Ǹf Cp R p
D =
2Ǹf Cs Rs
Q =
G =
Q =
1
D
Rp
1
1
D
2Ǹf Cs Rs
1
=
Parallel equivalent
circuit model
Series equivalent
circuit model
f: M e a sure m e n t signa l f re q u e ncy ( 120 Hz / 1 k H z / 1 M Hz )
Setting up Measurement Conditions and Display
Selecting Measurement Parameters
Figure 4-1 Relationship between equivalent circuit model and measurement parameters
Setup procedure
For information on the selection criteria for the equivalent circuit model, refer
to “Selection criteria of parallel/series equivalent circuit models” on
page 247.
For details on setup procedure, refer to “Setting up measurement parameters”
on page 40.
52Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Inputting Comment Line
Inputting Comment Line
You can enter a comment in the comment line by using the softkeys to enter
letters and the entry keys to enter numbers from 0 through 9, +, -, and period
(.). Your entered comment is saved in the internal memory or external USB
memory along with the control settings of the E4981A. When you load the
control settings, your saved comment is loaded as well.
The comment can be up to 30 characters in length. However, only the first 22
characters are displayed in the area.
Until you enter a comment into the comment line, the default text “USER
COMMENT” appears in the comment line.
You can also use the :DISPlay:LINE command to enter ASCII characters into
the comment line.
To enter a comment into the comment line:
Step 1.Press the [Meas Setup].
Step 2.Press MEAS SETUP softkey.
Step 3.Using the cursor keys, select the USER COMMENT field.
If there is already a comment, select that comment (field).
Step 4.To input a letter, use the following softkeys to cycle through letters in
alphabetical order and then select your desired letter:
SoftkeyDescription
NEXTDisplays the next letter to the letter currently displayed in the ADD
CHAR softkey.
PREVDisplays the previous letter to the letter currently displayed in the ADD
CHAR softkey.
To input a number, use the entry keys.
Step 5.Press the ADD CHAR softkey.Your selected single character appears in the
input line area.
Step 6.Repeat
Step 7.Press the ENTER softkey to enter your text in the USER COMMENT field.
Setting up Measurement Conditions and Display
Setting Up Measurement Signals (Frequency And Level)
Setting Up Measurement Signals (Frequency And Level)
Setting up frequency
You can set the frequency of the measurement signal applied to the DUT as
either
120 Hz, 1 kHz, or 1 MHz.
In Option 002, 1 MHz frequency is not available.
Setup procedure
For details on setup procedure, refer to “Setting up measurement signal
frequency” on page 41.
Setting up level
You can set up the level of the measurement signal applied to the DUT within a
range of 0.1 V to 1.0 V in steps of 10 mV.
Setup procedure
For details on setup procedure, refer to “Setting up measurement signal level”
on page 42.
Setting up signal level compensation (SLC) function
Functional description
If the impedance of DUT is small, then the output resistance of the signal
source, the resistance of the measurement cable can cause the voltage applied
to the DUT to decrease below the specified voltage of the signal source. The
signal level compensation feature adjusts the voltage across the DUT to the
same level as the signal voltage level setting. Using this feature, you can
maintain a constant level (voltage) of measurement signals applied to the DUT.
SLC is available on:
•120 Hz: 220 µF, 470 µF, 1 mF range
•1 kHz: 22 µF, 47 µF, 100 µF range
For all other range/frequency SLC is not executed even if LVL COMP is ON.
54Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Setting Up Measurement Signals (Frequency And Level)
To set up the signal level compensation feature:
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the LVL COMP field.
Step 3.Use the following softkeys:
SoftkeyDescription
ONTurns ON the signal level compensation feature.
OFFTurns OFF the signal level compensation feature.
Level error of the signal level compensation function
Even if you use the signal level compensation function, signal level drop occurs
at larger capacitance with larger D (refer to “Signal Level Compensation (SLC)
function” on page 193).
In E4981A, the AC level monitor is always ON.
The status in which capacity further increases and the output current of the
signal source approaches the limit (approximately 1 Arms) is judged as
overload. In this case, OVLD appears on the display.
Executing signal level check
When level error becomes over 10%. If the applied signal level for DUT is
dropped more than 10% from the setting level the following error occurs:
•When both SLC and contact check are ON, generates contact check error.
•When SLC is ON and contact check is OFF, generates OVLD (overload)
error.
For more information on contact check, refer to “Setting up Contact Check” on
Setting up Measurement Conditions and Display
Selecting Measurement Range
Selecting Measurement Range
Setting measurement range to be automatically selected (auto ranging)
Two modes can be used to select the measurement range as shown in Table
4-2. To enable automatic selection, select the auto range mode.
Table 4-2 Measurement range mode
ModeFunction overviewAd vantageDisadvantage
The instrument
automatically
Auto range
(auto selection)
selects a proper
measurement
range depending
on the value of the
DUT and performs
measurement.
You do not need to
select the
measurement
range.
The measurement
time is longer due
to the ranging time.
Hold
(fixed) range
(manual selection)
Measurement is
performed with a
fixed measurement
range regardless of
the value of the
DUT.
No ranging time is
required.
You need to select a
proper range
depending on the
value of the DUT.
56Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Selecting Measurement Range
Selecting your desired measurement range (hold range)
The selectable measurement ranges are shown in Table 4-3. As this table
shows, selectable measurement ranges differ depending on the measurement
signal frequency. Therefore, if a newly selected measurement frequency
conflicts with the current measurement range setting, the setting is
automatically changed to cover the allowable range.
Table 4-3 Selectable measurement ranges
For 120 Hz measurement
signal frequency:
For 1 kHz measurement
signal frequency:
For 1 MHz measurement
signal frequency:
1 pF2.2 pF4.7 pF
10 pF22 pF47 pF
100 pF220 pF470 pF100 pF220 pF470 pF
1 nF2.2 nF4.7 nF1 nF
10 nF22 nF47 nF10 nF22 nF47 nF
100 nF220 nF470 nF100 nF220 nF470 nF
1 µF2.2 µF4.7 µF1 µF2.2 µF4.7 µF
10 µF22 µF47 µF10 µF22 µF47 µF
100 µF220 µF470 µF100 µF
1 mF
The procedure to set up the measurement range mode is described below.
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor key to select the RANGE field.
Step 3.Use the numeric keys or following softkeys:
SoftkeyDescription
AUTOSets the measurement range mode to AUTO.
HOLDSets the measurement range mode to HOLD.
INCR+Increments the measurement range in the HOLD mode.
DECR-Decrements the measurement range in the HOLD mode.
Setting up Measurement Conditions and Display
Selecting Measurement Time
Selecting Measurement Time
Measurement time of 1, 2, 4, 6 and 8 are available in E4981A.
For details on the actual measurement time, refer to “Measurement time” on
page 188.
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the MEAS TIME field.
Step 3.Use the numeric keys or following softkeys:
SoftkeyDescription
INCR+Increments the measurement time.
DECR-Decrements the measurement time.
Selecting Cable Length
Set up the length of the measurement cable to compensate errors due to the
extension of the measurement cable. The E4981A lets you select the length
from 0 m, 1 m, or 2 m, depending on the measurement cable you use.
0 m When you do not use the test lead (in other words, when you
1 m When you use the
2 m When you use the
Setup procedure
For details on setup procedure, refer to “Setting up cable length” on page 43.
Cable length information is saved in internal FLASH memory when power is
OFF with backup function enabled.
connect the test fixture directly to the UNKNOWN terminal).
Keysight 16048A/B Keysight test lead.
Keysight 16048D test lead.
58Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Setting Up Averaging Count
Setting Up Averaging Count
Functional description
The averaging feature of the E4981A allows you to obtain moving average
values of successive measurement results.
You can specify the averaging factor within the range of 1 to 256 in steps of 1.
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the AVG field
Step 3.Use the numeric keys or following softkeys:
SoftkeyDescription
ONTurns ON the averaging feature.
OFFTurns OFF the averaging feature.
INCR+Increments the averaging factor in steps of 1.
DECR-Decrements the averaging factor in steps of 1.
Setting Up Trigger Delay Time
You can set up a wait time (trigger delay time) between when a trigger is
detected and when the measurement is started. You can set up the trigger
delay time within the range of 0 s to 1 s in steps of 100
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the TRIG DLY field.
Step 3.Use the numeric keys or following softkeys:
SoftkeyDescription
INCR++Increments the trigger delay time in steps of 1 msec.
μs.
INCR+Increments the trigger delay time in steps of 0.1 msec.
DECR-Decrements the trigger delay time in steps of 0.1 msec.
DECR--Decrements the trigger delay time in steps of 1 msec.
Setting up Measurement Conditions and Display
Outputting the Measurement Signal Only During the Measurement to Protect the
Contact Pin (Synchronous Source Function)
When source delay is available, trigger delay executes after source delay
time.
Outputting the Measurement Signal Only During the Measurement to
Protect the Contact Pin (Synchronous Source Function)
You can use this function to prevent large current from flowing through the
contact pin at the moment of contact with DUT to avoid damaging the contact
pin. This function, after a trigger, outputs the measurement signal so that it is
applied only during measurement (synchronous source function).
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the SYNC SRC field.
Step 3.Use the following softkeys:
SoftkeyDescription
ONTurns ON the synchronous source function.
OFFTurns OFF the synchronous source function.
Step 4.Use the cursor keys to select the SRC DLY field.
Step 5.Use the following softkeys:
SoftkeyDescription
INCR++Increments the source delay in steps of 1 ms.
INCR+Increments the source delay in steps of 0.1 ms.
DECR-Decrements the source delay in steps of 0.1 ms.
DECR--Decrements the source delay in steps of 1 ms.
The specified source delay time (to delay the start of signal output) is valid
only when the synchronous source function is ON.
When the source delay time and trigger delay time are both specified, the
source delay time precedes the trigger execution.
Figure 4-2 shows the measurement signal output timing.
60Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Measurement Status
Measurement Signal
Trigger Signal
Trigger Signal
Measurement Signal
Measurement Status
Output
Halted
Halted
(Index)
(Index)
Source Delay Time
Trigger Delay Time
Trigger Delay Time
Halted
Analog Meas.
Output continues
Halted
Analog Meas.
Digital
Computing
Digital
Computing
Synchronous Source Function = ON
Synchronous Source Function = OFF
Setting Up Frequency Shift
Figure 4-2 Measurement signal output timing
e4981aue0043
Setting Up Frequency Shift
When two or more E4981A units are integrated into a single system, you can
shift the 1 MHz measurement frequency by +1%, –1%, 0%, –2%, or +2% to
avoid interference between the measurement signals.
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the FREQ SHFT field.
Step 3.Use the following softkeys:
SoftkeyDescription
+1%Shifts the 1 MHz measurement frequency by +1%.
–1%Shifts the 1 MHz measurement frequency by –1%.
0%Shifts the 1 MHz measurement frequency by 0%.
–2%Shifts the 1 MHz measurement frequency by –2%.
+2%Shifts the 1 MHz measurement frequency by +2%.
For Option 002, the frequency shift function is not available.
Setting up Measurement Conditions and Display
Setting Up Display
Setting Up Display
Turning ON/OFF display
You can turn ON/OFF the display with the [Display Format] key.
If you turn OFF the display, DISPLAY NORMAL is always displayed in the
display area for the above items, and you cannot read the measurement result.
However, the measurement time is shortened because no additional time is
required for updating the display.
Setup procedure
Step 1.Press [Display Format].
Step 2.Press the DISPLAY BLANK softkey.
Step 3.If you turn OFF the display, DISPLAY NORMAL is always displayed in the
display area. To return to normal display press the DISPLAY NORMAL softkey.
Using fixed point display for measurement result
The following two display modes can be used to display the measurement
result of the measurement parameter.
Description
Floating point display
(default setup)
Fixed point displayThe value displayed at each digit is fixed to the value defined by
The fixed point display is always used to display the deviation percentage
(refer to “Displaying measurement result in deviation from reference value
(deviation measurement mode)” on page 63) as well as to display D and Q.
The value displayed at each digit is not fixed and changes
depending on the measured value.
the user.
62Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Setting Up Display
Setup procedure
Selecting floating/fixed point display
Step 1.Press [Display Format].
Step 2.Use the cursor keys to select the field in which a measurement result of
primary parameter or secondary parameter is displayed.
Step 3.Use the following softkeys:
SoftkeyDescription
D.P. AUTOAutomatically sets the decimal point
D.P. FIXFixes the decimal point position
D.P. POS INCR+Increments the position of decimal point
D.P. POS DECL-Decrements the position of decimal point
Displaying measurement result in deviation from reference value (deviation
measurement mode)
Functional description
The deviation measurement feature allows you to display deviation values
instead of actual measurements. A deviation is expressed as the difference
between the actual measurement and the stored reference value. The deviation
measurement feature is useful when you observe how a particular value of a
device/component changes under varying conditions of temperature,
frequency, bias, and other influences.
You can apply the deviation measurement feature to the primary or secondary
parameter or both. The deviation measurement feature supports the following
two modes:
ΔABS (absolute value) deviation measurement
•
Identifies and displays the difference between the actual measurement of
the DUT and the stored reference value. This value is calculated based on
the following formula:
Setting up Measurement Conditions and Display
Setting Up Display
•Δ% (percentage) deviation measurement
Identifies the difference between the actual measurement of the DUT and
the stored reference value and displays it as a percentage of the reference
value. This percentage deviation value is calculated based on the following
formula:
Δ%= (X-Y)/Y×100 (%)
Xactual measurement of the DUT
Ystored reference value
To set up the deviation measurement feature:
Step 1.Press [Meas Setup].
Step 2.Use the cursor keys to select the REF A field.
Step 3.Enter the reference value using the softkeys or entry keys. If you use the entry
keys to enter the value, the softkey labels change to unit labels (p, n, u, m, x1,
k).
SoftkeyDescription
MEASUREIf you want to use a particular device/component as the reference,
connect the DUT and press this key. Then the instrument measures the
DUT once and automatically fills the REF A and REF B fields with the
measured values, which now serve as the reference values.
Step 4.Use the cursor keys to select the DEV A field.
Step 5.Use the softkey to select the deviation mode for the primary parameter:
SoftkeyDescription
ABSDisplays the deviation as the difference from the reference value.
%Displays the deviation as a percentage of the reference value.
OFFTurns OFF deviation measurement.
Step 6.Using the cursor keys, select the DEV B field.
Step 7.Select the deviation mode for the primary parameter using the softkeys
described in
64Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Step 5.
Setting up Measurement Conditions and Display
Setting Up Display
Difference between deviation measurement mode and offset correction
The offset correction function is similar to the deviation measurement mode
with respect to the behavior when a predefined value is subtracted from the
measured value. On the other hand, the two functions differ as explained
below.
PurposeEffect on comparator
Not affected.
The absolute value of the
measurement result is
always used for sorting
judgment.
Deviation
measurement
mode
Used when calculating the
deviation of the measured value
from any given value (for
example, nominal value).
Used to bring the measured
Offset correction
value close to any given value
(for example, to compensate
variations of the measured value
among instruments for the same
Values compensated by this
function are used for sorting
Affected.
judgment.
DUT).
If the deviation measurement mode and the offset correction are both ON, the
offset correction is executed first and then, using the result as the
measurement result, the value calculated is displayed. (Refer to Figure 5-1 on
Setting up Measurement Conditions and Display
Setting up Contact Check
Setting up Contact Check
The E4981A contact check function can be used to check for contact failure
between the E4981A connector and the DUT.
When contact failure is detected, N.C. will appear on the display and a fixed
value (9.9E37) is always read out via GPIB/LAN/USB as the measurement
value, and the /LOWC_OR_NC signal of the handler interface goes Low.
The contact check function is available only at 120 Hz and 1 kHz
measurement frequency. This function does not works when range is set as
AUTO.
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Press the CONT CHECK softkey.
Step 3.Use the cursor keys to select the CONT CHK1 field.
Step 4.Use the following softkeys:
SoftkeyDescription
ONTurns ON the contact check function to detect loose connection
OFFTurns OFF the contact check function
Step 5.Use the cursor keys to select the CC1 TH1 field. TH1 is a parameter
proportional to the contact resistance of Hp or Hc.
Step 6.Use the recommended setting value 0.1 (default value) for CC1 TH1.
Table 4-4 Recommended setting value for TH1 and TH2
Range
TH1TH2
120 Hz1 kHz
10 nF100 pF - 1 nF
22 nF - 100 nF2.2 nF - 10 nF0.08
0.08
220 nF - 1 µF22 nF - 100 nF0.08
1
2.2 µF - 10 µF220 nF - 1 µF0.1
22 µF - 100 µF2.2 µF - 10 µF
0.1
0.15
1
220 µF - 1 mF22 µF - 100 µFN/A
1. Default value.
66Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Setting Up Condition To Make A Beep
Step 7.Use the cursor keys to select the CC1 TH2 field. TH2 is a parameter
proportional to the contact resistance of Lp or Lc.
Step 8.Use the softkeys or entry keys to enter the TH2 value according to Table 4-4.
When data is entered with the entry keys, the softkeys change to units labels
(m, x1).
SoftkeyDescription
INCR++Increments the threshold value as 0, 10 m, 20 m, 50 m, 100 m, 200 m,
500 m, 1.
INCR+Increments the threshold value in steps of 10.
DECR-Decrements the threshold value in steps of 10.
DECR--Decrements the threshold value 1, 500 m, 200 m, 100 m, 50 m, 20 m,
10 m, 0.
Setting Up Condition To Make A Beep
The condition to make a beep sound differs depending on the beep mode, as
shown in Table 4-5.
Table 4-5 Condition to make a beep
ModeCondition to make a beep
OffNever makes a beep.
Fail
Pass
Setup procedure
Step 1.Press [Meas Setup].
Step 2.Press the LIMIT TABLE softkey.
When wrong key operation is
performed.
When an error, alarm, or other
message
is outputted.
When the sorting judgment result of
the comparator is OUT_OF_BIN,
AUX_BIN, No contact or OVLD.
When the sorting judgment result of
the comparator is from BIN1 to
BIN9.
Step 3.Use the cursor keys to select the BEEP field.
Setting up Measurement Conditions and Display
Configuring the System Date
SoftkeyDescription
DAY DECR-Decrements the day in steps of 1.
When you enter the year through the entry keys, use a four-digit value.
Step 7.If you have selected the time field, use the following keys to edit the hour,
minute, and second digits.
SoftkeyDescription
HOURAllows you to change the hour digit.
MINUTEAllows you to change the minute digit.
SECONDAllows you to change the second digit.
Step 8.Enter the hour/minute/second using the softkeys or entry keys. If you use the
entry keys to enter the value, the softkey labels change to unit labels (x1).
SoftkeyDescription
HOUR INCR+Increments the hour in steps of 1.
HOUR DECR-Decrements the hour in steps of 1.
MINUTE INCR+Increments the minute in steps of 1.
MINUTE DECR-Decrements the minute in steps of 1.
SECOND INCR+Increments the second in steps of 1.
SECOND DECR-Decrements the second in steps of 1.
When you enter the hour with the entry keys, use a value within the range
of 0 (midnight) through 23 (11:00 pm).
72Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Configuring the GPIB Address
Configuring the GPIB Address
Functional description
Before you can control the E4981A by issuing SCPI commands from an external
controller connected via its GPIB connector, you have to configure the GPIB
address of your E4981A. For information on the concept and implementation of
automatic configuration by use of GPIB, refer to Programming Manual.
To configure the GPIB address:
Step 1.Press [System].
Step 2.Press the SYSTEM CONFIG softkey.
Step 3.Use the cursor keys to select the GPIB ADDR field.
Step 4.Enter a value with the entry keys in the range of 0 through 30.
Configuring the LAN IP Address
Functional description
To enable the E4981A to communicate over a local area network (LAN), you
have to configure its IP address and connect a LAN cable. The IP address can
be either automatically obtained or manually configured.
When the IP address is set to AUTO, AUTO-IP address is set. When the IP
address is set to MANUAL, the address manually configured is set.
For information on the concept and implementation of automatic configuration
by use of a LAN, refer to Programming Manual.
SoftkeyDescription
AUTOAutomatically obtains the IP address.
MANUALManually configures the IP address.
Table 4-6 IP address configuration methods
MethodDescription
AUTO-IPYou can automatically obtain an available IP address.
MANUALYou can manually configure the IP address as well as subnet mask and
gateway settings.
1. Consult your network administrator for relevant settings of your network.
Setting up Measurement Conditions and Display
Configuring the LAN IP Address
Once you have automatically obtained an IP address, you can check the
following monitor areas in the SYSTEM CONFIG page to see the address,
subnet mask, and gateway:
•CURRENT IP ADDR
•CURRENT SUBNET MASK
•CURRENT GATEWAY
To automatically obtain the IP address:
Step 1.Press [System].
Step 2.Press the SYSTEM CONFIG softkey.
Step 3.When you want to obtain the IP address automatically, select the IP CONFIG
field using the cursor keys and then press the AUTO softkey.
To manually configure the IP address:
Step 1.Press [System].
Step 2.Press the SYSTEM CONFIG softkey.
Step 3.Use the cursor keys to select the IP CONFIG field and then press the MANUAL
softkey.
Step 4.Use the cursor keys to select the MANUAL IP ADDR field.
Step 5.Enter the IP address using the entry keys.
Example: 192.168.1.101
Step 6.Press the ENTER softkey.
Step 7.Use the cursor keys to select the MANUAL SUBNET MASK field.
Step 8.Enter the subnet mask using the entry keys.
Step 9.Press the ENTER softkey.
Step 10.Use the cursor keys to select the MANUAL GATEWAY field.
Step 11.Enter the gateway using the entry keys.
Step 12.Press the ENTER softkey.
Step 13.Press the RESTART NETWORK softkey.
74Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Configuring the LAN IP Address
Checking the LAN connection status
You can check the CURRENT LAN STATUS monitor area to see the LAN
connection status of the E4981A. The status is expressed as one of the
following:
Table 4-7 LAN connection status
StatusDescription
NORMALLAN connection is OK.
FAILEDDisconnected from the LAN or LAN connection has failed.
IDENTIFYLAN connection is being initialized.
To reconnect to the network:
Step 1.Press [System].
Step 2.Press the SYSTEM CONFIG softkey.
Step 3.Press the RESTART NETWORK softkey on the IP CONFIG field, MANUAL IP
ADDR field, MANUAL SUBNET MASK field, or MANUAL GATEWAY field.
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Saving/Recalling Instrument Setup State (Save/Recall Function)
You can save/recall up to 20 instrument setup states into/from the built-in
FLASH memory (0 through 9) and external USB memory (10 through 19).
The ninth instrument setting corresponds to Auto Recall. Auto recall is not
executed when power is ON, by pressing the [Preset] key.
Overview of save/recall functionality
Configurations and measurement results can be saved into, and recalled from,
the E4981A’s internal memory or external USB memory through the save/recall
functionality.
Save methods and their uses
Table 4-8 shows available save methods and their uses:
Table 4-8 Save methods and their uses
Save method
Type
File format
(extension)
Configuration
Save (internal
-----------Yes
memory)
Configuration
Save (USB
(.sta)Yes
memory)
Data Save (USB
memory)
Screen Save
(USB memory)
CSV format (.csv)No
GIF format (.gif)No
For information on configuration states that can be saved, see Table C-1,
“Initial settings, settings that can be saved/recalled, settings that can be
backed up,” on page C-228.
RecallableUse
To save E4981A’s configuration
states into the internal memory.
To save E4981A’s configuration
states into USB memory.
To save measurement results into
USB memory.
To save E4981A’s screenshot into
USB memory.
76Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
\
\
\
\
Saving/Recalling Instrument Setup State (Save/Recall Function)
Folder/File structure on USB memory
When you save information into USB memory, a predetermined structure
scheme is used to organize folders and files on the memory, as shown in Figure
4-3.
Folder
Max. number
of files
Description
data999Contains measurement results as .csv files.
image999Contains screenshots as .gif files.
state10Contains instrument configuration states.
1
system
1. This folder can be operated from the SYSTEM page.
1
This single file, whose name is always “system,”
contains the system information.
These folders are automatically created in the memory.
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
USB memory notes
Attention should be given to the following points when using a USB memory
device with the E4981A
•Use a USB memory whose interface is USB 1.1.
•Use a USB memory that is USB mass storage class compliant and
formatted with FAT16 or FAT32.
•Use the USB memory solely for the E4981A. Otherwise, other data
previously saved in the USB memory could be erased.
•If you cannot save into or recall from the USB memory, use another USB
memory device.
•Keysight Technologies shall not be responsible for nor assume any liability
for data loss in your USB memory device after using it with the E4981A.
Saving/Recalling instrument configuration states
Overview of instrument configurations
Pressing the [Save/Recall] opens the CATALOG page. You can save/recall
instrument configuration states in either of two ways:
•Save into the internal memory
•Save into USB memory
You can save up to 10 registers into the internal memory and into a single USB
memory device. Also, register numbers have extensions as shown in Table 4-9:
Table 4-9 Extensions for register numbers
Classification
Internal memory
Register number
(No. field)
0Recalled when the [Recall A] key is pressed.
1Recalled when the [Recall B] key is pressed.
2 through 8-
9Auto recall
Extension
USB memory10 through 19-
In this page, you can configure each of the following controls with the cursor
placed in the corresponding field (denoted in parentheses).
•Medium mode (MEDIA field)
•Register number (No. field)
78Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Figure 4-4 CATALOG page
Medium mode
Functional description
You have to specify the medium type of the destination or source before saving
or recalling instrument configurations.
To select the medium mode:
Step 1.Press [Save/Recall].
Step 2.Use the cursor keys to select the MEDIA field.
Step 3.Select the medium mode by pressing the appropriate softkey:
SoftkeyDescription
INTUses the internal memory as the destination or source. Once you have
selected this mode, the register numbers (in the No. field) change to 0
through 9.
EXTUses USB memory as the destination or source. Once you have selected
this mode, the register numbers (in the No. field) change to 10 through
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Choosing a register number
Functional description
You have to choose one of the register numbers (in the No. field) before saving
or recalling instrument configurations.
Available choices include 0 through 9 when the medium mode is INT (internal
memory), or 10 through 19 when the medium mode is EXT (USB memory).
To choose a register number and save/recall the configurations:
Step 1.Press [Save/Recall].
Step 2.Use the cursor keys to select the No. field of your desired register number.
Step 3.Select your desired action by pressing the appropriate softkey:
SoftkeyDescription
RECALLRecalls the configurations previously saved in the register that
corresponds to your specified register number.
SAVESaves the configurations into the register that corresponds to your
specified register number.
DELETEDeletes the configurations previously saved in the register that
corresponds to your specified register number.
80Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Memory status information
Functional description
Each register number is associated with one of the following memory status
values:
Status valueDescription
–1
0The register contains no configurations.
1The register does contain configurations.
2The register contains any configuration that was saved under a
1. This value is not applicable when the medium mode is INT (internal memory).
2. An “Incompatible state file” warning message is displayed. Status information may
fail to be correctly recalled.
The ninth instrument setting corresponds to Auto Recall. Auto recall is not
executed when power is ON, by pressing the [Preset] key.
Comment information
Functional description
You can view comments entered into the comment line (USER COMMENT field)
on the Measurement Conditions screen. For information on how to enter a
comment, see “Inputting Comment Line” on page 53.
The register contains non-E4981A configuration information.
1
different firmware version or from another E4981A with a different
option(s) equipped.
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Functional description
When you save configuration states into USB memory, they are saved as
predefined files in predefined locations. You cannot manually define the file
names or locations.
Each state file is assigned a file name in the format of “register number.sta”.
To save configuration states into USB memory:
Step 1.Press [Save/Recall].
Step 2.Use the cursor keys to select the MEDIA field.
Step 3.Press the EXT softkey.
Step 4.Use the cursor keys to select the No. field (10 through 19) for your desired
register.
Step 5.Press the SAVE softkey to save configuration states into the USB memory.
If you have selected the No. field for a register that already contains
configuration states, the new configuration states overwrite the existing
ones.
To recall configuration states from USB memory:
Step 1.Press [Save/Recall].
Step 2.Use the cursor keys to select the MEDIA field.
Step 3.Press the EXT softkey.
Step 4.Use the cursor keys to select the No. field (10 through 19) for your desired
register.
Step 5.Press the RECALL softkey to recall configurations from the USB memory.
State files are automatically assigned file names 10.sta through 19.sta, and
you cannot change the file names.
If you attempt to recall configuration states under any of the following
conditions, an “Incompatible state file” warning message is displayed.
•The configuration states were saved under a different firmware version.
•The configuration states were saved from another E4981A with a different
option(s) equipped.
84Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Using the auto recall feature
Functional description
You can have the E4981A, at start-up, automatically recall the configuration
states previously saved in register number 9 on Internal memory.
To use the auto recall feature:
Step 1.Press [Save/Recall].
Step 2.Use the cursor keys to select the No. 9 [A] field.
Step 3.Press the RECALL softkey, to configure the instrument using the instrument
configuration information previously saved in register 9 on the Internal
memory.
Transferring auto recall setting to another E4981A
Using external USB memory
Step 1.Perform auto recall.
Step 2.Connect external USB memory.
Step 3.Press [Save/Recall].
Step 4.Use the cursor keys to select MEDIA to EXT.
Step 5.Use the cursor keys to select memory location from 10 to 19.
Step 6.Press the SAVE softkey to save the data in the USB memory.
Step 7.Connect the USB memory to another E4981A, recall the saved state.
Step 8.After recalling the state, save it to internal memory No. 9[A].
Using *LRN? command
Step 1.Perform auto recall.
Step 2.Retrieve the instrument configuration using *LRN? command via
GPIB/LAN/USB.
Step 3.Send the instrument configurations (retrieved in Step 2) to another E4981A.
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Saving measurement results into USB memory
You can save measurement results obtained by the E4981A into USB memory
as .CSV files. You can later load your saved files into an application program
running on a PC.
Data buffer 3 is used to perform save data operation through front panel
and web server. Also, it is not possible to load data through web server.
Before saving measurement results into USB memory, you need to complete
some tasks preparatory to initiating measurement so that the data buffer
memory contains the necessary data. For more information, see “To save
measurement results into USB memory:” on page 88.
You cannot load measurement results from USB memory into the E4981A.
Measurement result format
Measurement results are output in the format shown in Table 4-10.
The Data A and Data B fields of a measurement result file differ depending on
whether the “effective digits” setting is on or off (controlled by the
:FORMat:ASCii:LONG command).
Table 4-10 Measurement result format
Measurement
Screen
ComparatorResult Format
one pointOFF<Data A>,<Data B>,<Status>
one pointON<Data A>,<Data B>,<Status>,<BIN No.>
Data A
Outputs the measurement data for the primary parameter.
The Data A field uses the following two different fixed-length ASCII formats:
When the “effective digits”
setting is off:
(:FORMat:ASCii:LONG OFF)
SN.NNNNNESNN
(S:+/-, N:0 to 5, E: exponent character)
When the “effective digits”
setting is on:
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
(:FORMat:ASCii:LONG ON)
86Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Data B
Outputs the measurement data for the secondary parameter.
The Data B field uses the following two different fixed-length ASCII formats:
When the “effective digits”
setting is off:
SN.NNNNNESNN
(S:+/-, N:0 to 5, E: exponent character)
(:FORMat:ASCii:LONG OFF)
When the “effective digits”
setting is on:
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
(:FORMat:ASCii:LONG ON)
Status
Represents the measurement result status by using one of the following status
values:
0Measurement successfully completed.
+1Overload.
+2Low C or No contact.
The Status field uses the following two-character fixed-length ASCII format:
SN (S:+/-, N:0 to 2)
If the value is 1, the measurement data is 9.9E37; if the value is 0 or 2, the
actual measurement data is output.
The data buffer memory contains a measurement result with no data (as
represented by the status value of –1), but it is excluded when the data is
saved into USB memory.
Bin no. (IN/OUT)
Represents the bin sorting results as well as IN/OUT evaluation results, as
shown below:
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
The <BIN No.> data output format is either 2- or 3-character fixed-length
ASCII format:
SN or SNN (S:+/-, N:0 to 9)
Example 4-1 Example of saved measurement result data
+1.059517689E-24,+1.954963777E+00,+0,+0
+9.706803904E-25,+2.095857894E-01,+0,+0
+2.172725184E-24,+2.072965495E-01,+0,+0
+3.660460872E-25,+7.172688291E+00,+0,+0
+1.135428381E-24,+6.490636201E-01,+0,+0
+1.384790632E-24,+2.193020669E+00,+0,+0
+3.829879310E-26,+2.788435221E+01,+0,+0
To save measurement results into USB memory:
Step 1.Plug a USB memory stick into the front USB port.
Step 2.Press [Save/Recall].
Step 3.Press the SAVE DATA softkey.
Step 4.Press the START LOG softkey and then press the following softkeys to enter the
measurement results into the data buffer memory.
SoftkeyDescription
START LOGStarts logging the measurement results into the data buffer memory.
SAVE & STOPCopies the data from the data buffer memory into the USB memory.
Then stops saving the measurement results into the data buffer memory
and clears the data buffer memory.
The above mentioned steps stores the data in buffer 3, enabling to save the
data to USB memory.
Step 5.Start measurement. The data buffer memory is filled with up to 1000 sets of
measurement results.
Step 6.Press the SAVE & STOP softkey to save results into the USB memory.
Step 7.When the data has been saved into the USB memory, a “Storing data
completed. : E498xXXX.csv” message appears in the system message area.
88Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
㪼㪋㪐㪏㪇㪸㫌㪼㪈㪈㪌㪇
㪛㪸㫋㪸㩷㪙㫌㪽㪽㪼㫉
㪤㪼㫄㫆㫉㫐
㪤㪼㪸㫊㫌㫉㪼㫄㪼㫅㫋
㪬㪪㪙㩷㪤㪼㫄㫆㫉㫐
㪲㪪㪫㪘㪩㪫㩷㪣㪦㪞㪴
㪲㪪㪘㪭㪜㩷㩽㩷㪪㪫㪦㪧㪴
㪣㫆㪾㪾㫀㫅㪾㩷㫊㫋㪸㫉㫋
㪛㪸㫋㪸㩷㫆㫌㫋㫇㫌㫋
㪚㫃㪼㪸㫉
Saving/Recalling Instrument Setup State (Save/Recall Function)
Figure 4-7 Saving measurement results into USB memory
Measurement result files are automatically assigned file names
E498x001.csv through E498x999.csv, and you cannot change the file
names.
Saving a screenshot into USB memory
You can save a screenshot of the E4981A’s display into USB memory as a .GIF
file. You can later load your saved file into an application program running on a
PC.
To save a screenshot into USB memory
Step 1.Display the screen you want to save.
Step 2.Plug a USB memory device into the front USB port.
Step 3.Press [Save/Recall].
Step 4.Press the SAVE DISPLAY softkey.
Step 5.When the data has been saved into the USB memory device, a “Storing image
completed. : E498xXXX.gif” message appears in the system message area.
Screenshot files are automatically assigned file names E498x001.gif
through E498x999.gif, and you cannot change the file names.
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
90Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Preparation for Accurate Measurement (Executing Correction)
Overview of Correction Function
5 Preparation for Accurate Measurement (Executing
Correction)
This chapter gives an overview and the operational procedures of the different
types of correction function available in the E4981A.
Overview of Correction Function
The E4981A provides several types of corrections such as OPEN, SHORT,
LOAD, OFFSET, and Cable correction. This section explains various types of
corrections available in the E4981A.
OPEN/SHORT/LOAD/OFFSET correction
The E4981A provides four types of correction functions: OPEN correction,
SHORT correction, LOAD correction, and OFFSET correction. The following
table gives a brief description of each correction function:
Type of correctionDescription
Corrects errors caused by parallel stray admittance of the test
fixture. This correction is based on the results of admittance
OPEN correction
SHORT correction
LOAD correction
measurement in the OPEN state which is performed in advance.
OPEN correction can be performed at any frequency (120 Hz/
1 kHz/1 MHz) and the correction data remains valid for all
measurement frequencies.
Corrects errors caused by series residual impedance of the test
fixture. This correction is based on the results of impedance
measurement in the SHORT state which is performed in advance.
SHORT correction can be performed at any frequency (120 Hz/
1 kHz/1 MHz) and the correction data remains valid for all
measurement frequencies.
Corrects complex errors related to amplitude/phase errors, the
scanner, and other factors caused by the cable and test fixture.
This correction is based on the results of impedance
measurement of the standard having a known value which is
performed in advance.
LOAD correction is frequency dependent and the correction data
remains valid only for the measurement frequency at which the
LOAD correction is applied. If the frequency is changed, the LOAD
correction needs to be performed again.
Preparation for Accurate Measurement (Executing Correction)
Overview of Correction Function
Type of correctionDescription
Corrects errors between the actual measured value and the ideal
value that the user desires. This is done by subtracting any value
OFFSET correction
that the user enters from the measured result (for example,
difference between a known standard value and its measured
value, or instrument-by-instrument differences in the measured
values of the same DUT).
In Option 002, there is no correction for 1 MHz frequency.
As shown in the data processing flow of Figure 5-1, the OPEN/SHORT/LOAD
correction is first performed for the measured impedance result (complex
number), and the result is converted to the primary/secondary parameters.
Then, for the primary/secondary parameter values (real number), the OFFSET
correction is performed.
Figure 5-1 Data processing flow
The measurement value displayed for correction or LOAD REF uses the
internal data for calculation. The value is set depending upon the state of
MULTI CORRECTION, LOAD REF, or CHANNEL.
92Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Preparation for Accurate Measurement (Executing Correction)
Overview of Correction Function
Cable correction
Cable correction avoids measurement error due to the differences of each
cable when you use a cable length of 1 m or 2 m in 1 MHz.
In Option 002, 1 MHz frequency is not available.
Measurement error due to an individual cable is proportional to the square of
the test frequency. Accordingly, the difference of each test cables may not be
ignored at high frequencies, such as 1 MHz. When the E4981A is shipped from
the factory, it contains typical correction data for both 1 m and 2 m cables.
Cable correction can compensate the error for your own cable.
The following equipment are required for cable correction:
Open Termination 42090A Open termination (do not use BNC adapter as an
Load Standard 6383A 100 pF (capacitor) or 42037A 1 kΩ (resistor)
alternative)
Cable correction is not required in the following conditions:
•For 120 Hz and 1 kHz measurement frequency.
•When Open, Short, and Load correction is executed at 1 MHz
measurement frequency.
Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions
Turning ON/OFF Correction Functions
You can turn ON/OFF each correction function separately.
Turning on OPEN correction
Use one of the following two ways to set the OPEN correction to ON.
Measuring data for OPEN correction
Execute measurement of the data for the OPEN correction. When the
measurement finishes successfully, the OPEN correction is automatically set to
ON. For information on the measurement procedure for the data used in the
OPEN correction, refer to “Measuring data for OPEN correction” on page 44.
Setting up ON/OFF directly
Use the following procedure to set the OPEN correction to ON/OFF.
Step 1.Press [Meas Setup].
Step 2.Press the CORRECTION softkey.
Step 3.Use the cursor keys to select the OPEN field.
Step 4.Press the ON softkey to turn ON the OPEN correction function.
If you change the setup of the cable length or frequency shift (1 MHz) with
the OPEN correction ON, the warning message “Need corr meas” appears
and the OPEN correction is automatically set to OFF. Along with OPEN
correction, SHORT and LOAD correction is set to OFF as well.
If you set the OPEN correction to ON by using the above procedure when
the setup of the cable length or frequency shift differs from that during
measurement of the data for OPEN correction, the warning message “Need
open meas” appears. Even if this warning message appears, the OPEN
correction is set to ON. However, you must again measure the data for
OPEN correction to ensure accurate measurement.
When the measurement result becomes overload, “Measurement failed”
error is displayed and OPEN correction data is not updated.
94Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions
Turning on SHORT correction
Use one of the following two ways to set the SHORT correction to ON.
Measuring data for SHORT correction
Execute measurement of the data for the SHORT correction. When the
measurement finishes successfully, the SHORT correction is automatically set
to ON. For information about the measurement procedure for the data used in
the SHORT correction, refer to “Measuring data for SHORT correction” on
page 47.
Setting up ON/OFF directly
Use the following procedure to set the SHORT correction to ON/OFF.
Step 1.Press [Meas Setup].
Step 2.Press the CORRECTION softkey.
Step 3.Use the cursor keys to select the SHORT field.
Step 4.Press the ON softkey to turn ON the SHORT correction function.
If you change the setup of the cable length or frequency shift (1 MHz) with
the SHORT correction ON, the warning message “Need corr meas” appears
and the SHORT correction is automatically set to OFF. Along with SHORT
correction, OPEN and LOAD correction is set to OFF as well.
If you set the SHORT correction to ON using the above procedure when the
setup of the cable length or frequency shift d iffers from that during
measurement of the data for SHORT correction, the warning message
“Need short meas” appears. Even if this warning message appears, the
SHORT correction is set to ON. However, you must again measure the data
for SHORT correction to ensure accurate measurement.
When the measurement result becomes overload, “Measurement failed”
error is displayed and SHORT correction data is not updated.
Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions
Turning on LOAD correction
Use one of the following two ways to set the LOAD correction to ON.
Measuring LOAD correction data
Execute measurement of the data for the LOAD correction. When the
measurement finishes successfully, the LOAD correction is automatically set to
ON. For information about the measurement procedure for the data used in the
LOAD correction, refer to “Obtaining (measuring) data for LOAD correction” on
page 102.
Setting up ON/OFF directly
Use the following procedure to set the LOAD correction to ON/OFF.
Step 1.Press [Meas Setup].
Step 2.Press the CORRECTION softkey.
Step 3.Use the cursor keys to select the LOAD field.
Step 4.Press the ON softkey to turn ON the LOAD correction function.
If you change the setup of the cable length or frequency shift (1 MHz) with
the LOAD correction ON, the warning message “Need corr meas” appears
and the LOAD correction is automatically set to OFF. Along with LOAD
correction, OPEN and SHORT correction is set to OFF as well.
If you set up the LOAD correction to ON using the above procedure when
the setup of the cable length or frequency shift differs from that during
measurement of the data for LOAD correction, the warning message “Need
load meas” appears. Even if this warning message appears, the LOAD
correction is set to ON. However, you must again measure the data for the
LOAD correction to ensure accurate measurement.
When the measurement result becomes overload, “Measurement failed”
error is displayed and LOAD correction data is not updated.
96Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Preparation for Accurate Measurement (Executing Correction)
Turning ON/OFF Correction Functions
Turning on OFFSET correction
Use the following procedure to set the OFFSET correction to ON/OFF.
Step 1.Press [Meas Setup].
Step 2.Press the CORRECTION softkey.
Step 3.Use the cursor keys to select the OFFSET field.
Step 4.Press the ON softkey to turn ON the OFFSET correction function.
If you change the setup of the measurement parameter with the OFFSET
correction ON, the OFFSET correction is automatically set to OFF.
You cannot separately set the ON and OFF states as the primary parameter
and secondary parameter. However, if you set the correction value to 0, the
state is effectively the same as OFF even if the OFFSET correction is ON.
Therefore, you can set up separate ON/OFF states by setting the correction
value of either parameter to 0.
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
Obtaining Correction Data
Obtaining (measuring) data for OPEN correction
Data structure
The 120 Hz, 1 kHz, and 1 MHz data used in the OPEN correction is divided into
the data for normal operation and the data of each channel (256 channels) for
multi-correction, as shown in Figure 5-2.
Figure 5-2 Structure of data for OPEN correction
Measured data
For Open correction, 120 Hz, 1 kHz, and 1 MHz are measured and each data
frequency is saved. Therefore, even if the measurement frequency is changed
98Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
there is no need to reset the correction. The result is set up as the data for
normal operation when the multi-correction function is OFF and as the data for
multi-correction when the function is ON (for the channel that has been
selected at execution).
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
The table below summarizes how data is set up, depending on the setup at
execution.
Multicorrection
OFF
ON
Setup of
measurement
signal frequency
120 Hz
1 kHz
1 MHz
120 Hz
1 kHz
1 MHz
Data
120 Hz correction data and 120 Hz cable length for
normal operation
1 kHz correction data and 1 kHz cable length for
normal operation
1 MHz correction data, 1 MHz cable length, and 1
MHz frequency shift for normal operation
120 Hz correction data and 120 Hz cable length for
the channel selected at the time of multi-correction
measurement
1 kHz correction data and 1 kHz cable length for the
channel selected at the time of multi-correction
measurement
1 MHz correction data, 1 MHz cable length and
1 MHz frequency shift for the channel selected at
the time of multi-correction measurement
Measurement conditions during data measurement
The data for the OPEN correction is measured under the following conditions:
•Measurement range mode: auto range mode
•Measurement time mode: 8
For other settings, such as averaging and trigger delay, the measurement
conditions set up at execution are used for measurement.
Measurement procedure
For more information on the measurement procedure for the data used in the
OPEN correction, refer to “Measuring data for OPEN correction” on page 44.
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
Obtaining (measuring) data for SHORT correction
Data structure
The 120 Hz, 1 kHz, and 1 MHz data used in the SHORT correction is divided
into the data for normal operation and the data of each channel (256 channels)
for multi-correction, as shown in Figure 5-3.
Figure 5-3 Structure of data for SHORT correction
Measured data
For SHORT correction, 120 Hz, 1 kHz, and 1 MHz are measured and each data
frequency is saved. Therefore, even if the measurement frequency is changed
there is no need to reset the correction. The result is set up as the data for
normal operation when the multi-correction function is OFF and as the data for
multi-correction when the function is ON (for the channel that has been
100Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
selected at execution).
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.