No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval
or translation into a foreign language)
without prior agreement and written consent from Keysight Technologies, Inc. as
governed by United States and international copyright laws.
Manual Part Number
E1418-90010
Edition
Second Edition, October 2019
Published by
Keysight Technologies, Inc.
900 S. Taft Ave.
Loveland, CO 80537 USA
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
www.keysight.com/find/E1418A
(product-specific information and support, software and documentation
updates)
www.keysight.com/find/assist (world-
wide contact information for repair and
service)
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may
be downloaded from the Web. Go to
http://keysight.com/go/conformity and
click on “Declarations of Conformity.” You
can then search by product number to
find the latest Declaration of Conformity.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,” AND
IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM
EXTENT PERMITTED BY APPLICABLE
LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED,
WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED
TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS OR
FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE
FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE
MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Keysight Technologies does not warrant
third-party system-level (combination of
chassis, controllers, modules, etc.) performance, safety, or regulatory compliance unless specifically stated.
DFARS/Restricted Rights
Notices
If software is for use in the performance
of a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause.
Use, duplication or disclosure of Software
is subject to Keysight Technologies’ standard commercial license terms, and nonDOD Departments and Agencies of the
U.S. Government will receive no greater
than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater
than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.2277015 (b)(2) (November 1995), as applicable in any technical data.
Safety Information
The following general safety precautions must be observed during all
phases of operation of this instrument.
Failure to comply with these precautions or with specific warnings or operating instructions in the product
manuals violates safety standards of
design, manufacture, and intended use
of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these
requirements.
General
Do not use this product in any manner not
specified by the manufacturer. The protective features of this product must not be
impaired if it is used in a manner specified in
the operation instructions.
Before Applying Power
Verify that all safety precautions are taken.
Make all connections to the unit before
applying power. Note the external markings
described under “Safety Symbols”.
Ground the Instrument
Keysight chassis’ are provided with a
grounding-type power plug. The
instrument chassis and cover must be
connected to an electrical ground to
minimize shock hazard. The ground pin
must be firmly connected to an electrical ground (safety ground) terminal at
the power outlet. Any interruption of
the protective (grounding) conductor
or disconnection of the protective
earth terminal will cause a potential
shock hazard that could result in personal injury.
Do Not Operate in an Explosive
Atmosphere
Do not operate the module/chassis in
the presence of flammable gases or
fumes.
Do Not Operate Near Flammable
Liquids
Do not operate the module/chassis in
the presence of flammable liquids or
near containers of such liquids.
Cleaning
Clean the outside of the Keysight module/chassis with a soft, lint-free,
slightly dampened cloth. Do not use
detergent or chemical solvents.
Do Not Remove Instrument Cover
Only qualified, service-trained personnel who are aware of the hazards
involved should remove instrument
covers. Always disconnect the power
cable and any external circuits before
removing the instrument cover.
Keep away from live circuits
Operating personnel must not remove
equipment covers or shields. Procedures involving the removal of covers
and shields are for use by servicetrained personnel only. Under certain
conditions, dangerous voltages may
exist even with the equipment
switched off. To avoid dangerous electrical shock, DO NOT perform procedures involving cover or shield removal
unless you are qualified to do so.
DO NOT operate damaged
equipment
Whenever it is possible that the safety
protection features built into this product have been impaired, either through
physical damage, excessive moisture,
or any other reason, REMOVE POWER
and do not use the product until safe
operation can be verified by servicetrained personnel. If necessary, return
the product to a Keysight Technologies
Sales and Service Office for service and
repair to ensure the safety features are
maintained.
DO NOT block the primary
disconnect
The primary disconnect device is the
appliance connector/power cord when
a chassis used by itself, but when
installed into a rack or system the disconnect may be impaired and must be
considered part of the installation.
Do Not Modify the Instrument
Do not install substitute parts or perform any unauthorized modification to
the product. Return the product to a
Keysight Sales and Service Office to
ensure that safety features are maintained.
In Case of Damage
Instruments that appear damaged or
defective should be made inoperative
and secured against unintended operation until they can be repaired by
qualified service personnel
Do NOT block vents and fan exhaust:
To ensure adequate cooling and ventilation, leave a gap of at least 50mm
(2") around vent holes on both sides of
the chassis.
Do NOT operate with empty slots: To
ensure proper cooling and avoid damaging equipment, fill each empty slot
with an AXIe filler panel module.
Do NOT stack free-standing chassis:
Stacked chassis should be rackmounted.
All modules are grounded through the
chassis: During installation, tighten
each module's retaining screws to
secure the module to the chassis and
to make the ground connection.
Operator is responsible to maintain
safe operating conditions. To ensure
safe operating conditions, modules
should not be operated beyond the full
temperature range specified in the
Environmental and physical specification. Exceeding safe operating conditions can result in shorter lifespan,
improper module performance and
user safety issues. When the modules
are in use and operation within the
specified full temperature range is not
maintained, module surface temperatures may exceed safe handling conditions which can cause discomfort or
burns if touched. In the event of a
module exceeding the full temperature
range, always allow the module to cool
before touching or removing modules
from the chassis.
4
Safety Symbols
A CAUTION denotes a hazard. It
calls attention to an operating procedure or practice, that, if not correctly performed or adhered to
could result in damage to the
product or loss of important data.
Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood and met.
A WARNING denotes a hazard. It
calls attention to an operating procedure or practice, that, if not correctly performed or adhered to,
could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood and met.
Products display the following symbols:
Warning, risk of electric
shock
Refer to manual for additional safety information.
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
This manual contains information required to test, troubleshoot, and repair the
Keysight E1418A 8/16 Channel D/A Converter Module. See the Keysight E1418A
User’s Manual for additional module information. Figure 1-1 shows the Keysight
E1418A module.
Figure 1-1 Keysight E1418A D/A Converter Module
11
General InformationSafety Considerations
Safety Considerations
This product is a Safety Class I instrument that is provided with a protective earth
terminal when installed in the mainframe. The instrument, mainframe, and all
related documentation should be reviewed for familiarization with safety
markings and instructions before operation or service.
Refer to the WARNINGS page (page 6) in this manual for a summary of safety
information. Safety information for preventive maintenance, testing, adjusting,
and service follows and is also found throughout this manual.
Warnings and Cautions
This section contains WARNINGS which must be followed for your protection
and CAUTIONS which must be followed to avoid damage to the equipment when
performing instrument maintenance or repair.
SERVICE-TRAINED PERSONNEL ONLY. The information in
this manual is for service-trained personnel who are
familiar with electronic circuitry and are aware of the
hazards involved. To avoid personal injury or damage to
the instrument, do not perform procedures in this manual
or do any servicing unless you are qualified to do so.
CHECK MAINFRAME POWER SETTINGS. Before applying
power, verify that the mainframe setting matches the line
voltage and that the correct fuse is installed. An
uninterruptible safety earth ground must be provided from
the main power source to the supplied power cord set.
GROUNDING REQUIREMENTS. Interruption of the
protective (grounding) conductor (inside or outside the
mainframe) or disconnecting the protective earth terminal
will cause a potential shock hazard that could result in
personal injury. (Grounding one conductor of a
two-conductor outlet is not sufficient protection.)
IMPAIRED PROTECTION. Whenever it is likely that
instrument protection has been impaired, the mainframe
must be made inoperative and be secured against any
unintended operation.
REMOVE POWER IF POSSIBLE. Some procedures in this
manual may be performed with power supplied to the
mainframe while protective covers are removed. Energy
available at many points may, if contacted, result in
personal injury. (If maintenance can be performed without
power applied, the power should be removed.)
12Keysight E1418A Service Guide
Safety ConsiderationsGeneral Information
USING AUTOTRANSFORMERS. If the mainframe is to be
energized via an autotransformer (for voltage reduction)
make sure the common terminal is connected to neutral
(that is, the grounded side of the main’s supply).
CAPACITOR VOLTAGES. Capacitors inside the mainframe
may remain charged even when the mainframe has been
disconnected from its source of supply.
USE PROPER FUSES. For continued protection against fire
hazard, replace the line fuses only with fuses of the same
current rating and type (such as normal blow, time delay,
etc.). Do not use repaired fuses or short-circuited
fuseholders.
SHOCK HAZARD. Only service-trained personnel who are
aware of the hazards involved should install, remove, or
configure the module. Before you remove any installed
module, disconnect AC power from the mainframe and
from other modules that may be connected to the module.
CHANNEL WIRING INSULATION. All channels that have a
common connection must be insulated so that the user is
protected from electrical shock. This means wiring for all
channels must be insulated as though each channel carries
the voltage of the highest voltage channel.
Keysight E1418A Service Guide13
General Information
Module Description
MAXIMUM VOLTAGE. The maximum voltage that can be applied
to any channel is 42 V Peak/42 Vdc. The maximum voltage from
any channel to ground is 42 V Peak/42 Vdc.
STATIC ELECTRICITY. Static electricity is a major cause of
component failure. To prevent damage to the electrical
components in the module, observe anti-static techniques
whenever working on the device.
PARTS REMOVAL. This module uses a surface mount printed
circuit assembly. Special soldering equipment is required for parts
removal or replacement. Use of standard soldering equipment will
cause damage to the printed circuit board and is not covered
under warranty.
The Keysight E1418A is an 8 or 16 channel digital-to-analog converter module
for use in a VXIbus C-size mainframe. The module is a register based device. The
module can be programmed via direct register access or, with the appropriate
driver, by high level commands. This manual describes programming the module
using SCPI (Standard Commands for Programmable Instruments) and the SCPI
driver.
Each Keysight E1418A module is a unique instrument having its own output
buffer and error queue. Multiple modules can not be combined into a
single instrument.
Each channel can be configured to either voltage or current output mode. When
configured for voltage output, voltages in the range of -16.0 V to
+16.0 V can be set. When configured for current output, current in the range of
-0.02 to + 0.02 Amps can be set. The channel output mode can be
programmatically set, or, can be forced to either voltage or current by
mechanical jumpers on the terminal block.
Each output channel is individually configurable to be either an isolated output
or a non-isolated output. Channel configuration to isolated or non-isolated is
made by individual plug-on modules for each channel.
Module Specifications
Specifications are listed in Appendix A of this manual and in the Keysight E1418A
User’s Manual. These specifications are the performance standards or limits
against which the module may be tested.
14Keysight E1418A Service Guide
Module DescriptionGeneral Information
Module Serial Numbers
Devices covered by this manual are identified by a serial number prefix listed on
the title page. Keysight Technologies uses a two-part serial number in the form
USXXXXYYYYY, where US is the country of origin, XXXX is the serial prefix, and
YYYYY is the serial suffix. The serial number prefix identifies a series of identical
instruments. The serial number suffix is assigned sequentially to each
instrument. The serial number plate is located on the right-hand shield near the
backplane connectors.
Ordering Options
The module may be ordered from Keysight Technologies in a variety of
configurations. As ordered, the module may have the following options:
Description
E1418A8-Channel D/A Converter with Non-Isolated Outputs
E1418A Option
001
E1418A Option
002
E1418A Option
003
16-Channel D/A Converter with Non-Isolated Outputs
8-Channel D/A Converter with Isolated Outputs
16-Channel D/A Converter with Isolated Outputs
Field Kits
The module can also be user configured. The following field expansion and
configuration kits are available. Each field kit contains installation instructions.
DescriptionUse
E1523ASingle Channel Isolated
plug-on module
E1524AExpansion kit, 8-Channel
Non-Isolated Outputs
E1525AExpansion kit, 8-Channel
Isolated Outputs
To change a single channel from
non-isolated to isolated output.
To add 8 additional
non-isolated channels to an
existing 8-channel module.
To add 8 additional isolated
channels to an existing
8-channel module.
Keysight E1418A Service Guide15
General InformationRecommended Test Equipment
Terminal Modules
The standard Keysight E1418A Terminal Module provides screw terminals for
connections. Two other terminal options are available with the Keysight E1418A:
– • Crimp and Insert (Option A3E)
– • Ribbon Cable (Option A3H)
Recommended Test Equipment
Table 1-1 lists the test equipment recommended for testing and servicing the
module. Essential requirements for each piece of test equipment are described in
the Requirements column.
Table 1-1 Recommended Test Equipment
InstrumentRequirementsRecommended
Model
Controller, GPIBGPIB compatibility as defined by IEEE Standard
488-1988 and the identical ANSI Standard
MC1.1: SH1, AH1, T2, TE0, L2, LE0, SR0,
RL0, PP0, DC0, DT0, and C1, 2, 3, 4, 5.
MainframeCompatible with the Keysight E1418AE1400B/T, E1401A/T, or
Command Module,
VXILink, EPC7
Embedded Controller
Digital Multimeter
5 1/2 or 6 1/2 digit
Terminal ModuleCompatible with the Keysight E1418AE1418-60101O, P,T
Resistor10 kΩ , 1%, 0.25 W, metal filmAny available supplier
Resistor600 Ω , 1%, 0.25 W, metal filmP
Compatible with Module
Resource manager in Slot 0
Logical address 0
Typical GPIB address 9
Requires the downloaded driver
“E1418”
Voltage measurements to ± 16.8 Vdc Current
measurements to ± 0.02 A Four-wire resistance
measurements from 50 Ω to 500 Ω
HP 9000 Series 300F,O,P,T
E1401B
E1405B or E1406A/BF,O,P,T
Keysight 3458A or
Keysight 34401A
meeting the requirements
Use
F,O,P,T
O, P, T
P
1
User’s ManualE1418-90001F,O,P,T
1 * F = Functional Verification Tests, O = Operation Verification Tests, P = Performance Verification
Tests, T = Troubleshooting
16Keysight E1418A Service Guide
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
Inspect the shipping container for damage. If the shipping container or
cushioning material is damaged, keep the container until the shipment contents
have been checked and the module has been checked mechanically and
electrically. Chapter 4 of this manual gives procedures to check the electrical
performance.
To avoid possible hazardous electrical shock, do not
perform electrical tests if there are signs of shipping
damage to any portion of the outer enclosure (covers,
panels, etc.).
If the contents are incomplete, if there is mechanical damage or defect, or if the
module does not pass the electrical performance tests, notify your nearest
Keysight Technologies Sales and Service Office. If the shipping container is
damaged or the cushioning material shows signs of stress, notify the carrier as
well as Keysight Technologies. Keep the shipping materials for the carrier’s
inspection.
11
InstallationPreparation for Use
Preparation for Use
Chapter 1 of the Keysight E1418A User’s Manual contains complete instructions
for installing the module in a mainframe and installing a terminal module.
Refer to the mainframe user’s manual, command module user’s manual, and the
Installing Device Drivers installation note to prepare the mainframe.
Prepare the Module
Verify or set the module’s logical address. Refer to the Keysight E1418A User’s
Manual for instructions. In this manual the factory default logical address of 72 is
used.
To prepare the module for the adjustment procedures in Chapter 5, set or verify
the position of the CAL Secure Enable Jumper before installing the module in the
mainframe. (The jumper is factory set to CAL.) Figure 5-2 shows this jumper.
Prepare the Terminal Module
The procedures in this manual show the screw-type terminal module. To prepare
this terminal module, you will need to set the jumpers and make connections. If
you are using other types of terminal modules, consult the Keysight E1418A
User’s Manual for proper connections.
Terminal Module Jumpers
The screw-type terminal module contains jumpers that control how the Keysight
E1418A operates. To prepare the module for the Verification Tests in Chapter 4
and Adjustments in Chapter 5, the terminal module jumpers must be set to allow
the module to be programmed from the backplane. The Keysight E1418A will
operate without any jumpers installed. Figure 2-1 shows the jumper locations
and example settings.
There are two jumpers for each channel; a P/J jumper and a V/I jumper. The two
jumpers work together.
The P/J jumper sets the channel output mode to be either Program Selectable or
Jumper Selectable. When the jumper is in the program selectable position, the
channel output mode can be set via programming commands. The program
selectable position is recommended for the Verification Tests and Adjustment
procedures in this manual. In the jumper selectable position, the channel output
mode is set via the V/I jumper.
The V/I jumper sets the channel output mode to either voltage or current when
the P/J jumper is set to the Jumper Selectable position. When the P/J jumper is
set to Program Selectable, this jumper has no effect.
12Keysight E1418A Service Guide
Preparation for UseInstallation
With no jumpers installed, the module will be in the Program
Selectable output mode.
– If you are using a terminal module dedicated to testing the Keysight
E1418A, you can remove all the jumpers or put the jumpers in the storage
position.
– If you are using the end-user’s terminal module, record the positions of the
jumpers. Set the P/J jumpers to the programmable position. Before
returning the module to use, restore the P/J jumpers to their original
position.
Figure 2-1 P/J and V/I Jumpers
Terminal Module Connections
The Verification Tests require four-wire connections at each channel. The
Adjustment procedures can be performed using a single connections at the CAL
terminals. Additional connection information is given in the procedures in
Chapters 4 and 5 of this manual.
Keysight E1418A Service Guide13
InstallationEnvironment
Prepare the Command Module
All verification and adjustment procedures in this manual use SCPI (Standard
Commands for Programmable Instruments). Prepare the command module by
downloading the SCPI driver named “E1418". Downloading instructions are
given in the Installing Device Drivers installation note.
Environment
The recommended operating altitude for the module is 2000 m (6000 ft) or less.
The recommended operating environment for the module is:
EnvironmentTemperatureHumidity
o
Operating0
Storage and
Shipment
C to + 55 oC<65% relative (0 oC to + 40 oC)
non condensing
o
C to + 75 oC<65% relative (0 oC to + 40 oC)
– 40
non condensing
Shipping Guidelines
If you need to return the Keysight E1418A to Keysight Technologies, first remove
any terminal modules before packaging for shipment. Attach a tag to the
module identifying the owner and indicating the service or repair required. Mark
the shipping container “FRAGILE”. If you are exchanging a module, read Assembly Exchange.
In any correspondence, refer to the module by model number and full serial
number.
Keysight Technologies recommends using shipping containers and materials
identical to those used by the factory. These materials are available through
Keysight Technologies Sales and Service Offices.
If you use commercially available shipping materials, place the module in an
anti-static bag and wrap the module in heavy paper or plastic. Use a strong
shipping container. A double wall carton of 2400 Pa (350 psi) test material is
adequate.
Use enough shock-absorbing material (layer at least 75 mm to 110 mm, 3 in to 4
in) around the module to provide firm cushion and to prevent movement in the
container. Protect the front panel with cardboard. Seal the container securely
and mark the container “FRAGILE.”
14Keysight E1418A Service Guide
3Operating Information
Introduction
Operation
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
This chapter provides operating information, preventive maintenance
instructions, and operators checks.
The Keysight E1418A User’s Guide is your reference for module operation. The
User’s Guide contains information about module setup, application examples,
and a complete SCPI command reference.
Querying Module Identification and Configuration
This example will query the module for the identification string and determine
the module configuration.
The module identification is obtained using the IEEE-488 Command *IDN?.
/* Send query to the module*/
*IDN?
/* Enter a string*/
The string returned will be similar to one of the following:
HEWLETT-PACKARD,E1418A_8CH,xxxxxxxxxx,A.01.00
or
HEWLETT-PACKARD,E1418A_16CH,xxxxxxxxxx,A.01.00
The xxxxxxxxxx portion of the response string is the module serial number. The
number of channels in the module (8 or 16) is indicated by the E1418A_8CH or
E1418A_16CH portion of the returned string.
The module configuration is obtained using the DIAGnostic:CONFiguration?
query.
/* Send query to the module*/
DIAG:CONF?
/* Enter six integers and decode the integers*/
The DIAG:CONF? query returns six integers. The six integers returned contain the
module configuration and have the following meaning.
15
Operating InformationOperation
Integer
Meaning
Returned
1stExpansion Board ID in the form: 0 = present, 7 = none
(expansion board contains channels 9 through 16)
2ndTerminal Module ID in the form:
0 = screw type, 7 = none or other
3rdIsolated/Non-isolated Channel status
A bit set to 0 indicates an isolated channel
A bit set to 1 indicates a non-isolated channel or no plug-on
module installed (DO NOT operate the module without a plug-on
module)
Bits 0 – 15 correspond to channels 1 – 16, respectively*
4thChannel Mode
A bit set to 0 indicates a current output channel
A bit set to 1 indicates a voltage output channel
Bits 0 – 15 correspond to channels 1 – 16, respectively*
5thChannel Output State
A bit set to 0 indicates the channel output relay is closed
A bit set to 1 indicates the channel output relay is open
Bits 0 – 15 correspond to channels 1 – 16, respectively*
6thChannel Mode Programmable State (P/J Jumper)
A bit set to 0 indicates a channel is not mode programmable
A bit set to 1 indicates a channel is mode programmable
Bits 0 – 15 correspond to channels 1 – 16, respectively
1
1 For 8-channel configurations, the upper 8 bits of integers 3, 4, 5,
and 6 are set to 1’s.
16Keysight E1418A Service Guide
OperationOperating Information
For example, if the following six integers are returned, the module has the
configuration indicated.
IntegerDecimal
Configuration
Value
1st7No expansion board installed
(Only channels 1 through 8)
2nd7No Terminal Module installed or unknown
terminal module installed
3rd-1All channels are non-isolated
4th-1All channels are voltage output channels
5th-1All channel outputs are disabled
(all output relays are open)
6th-1All channels are output mode programmable
Configuration Example
The following program segment demonstrates how to read the module
identification string and the configuration. The configuration integers are bit
manipulated using the C operator for bit shifting
result = result <<1
(a one bit shift to the left). The code shown in this example can be obtained from
the file prftest.c on the examples disk provided with this manual.
err=viScanf(dac, “%,6d”, &config);/* returns six integers*/
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check*/
Keysight E1418A Service Guide17
Operating InformationOperation
/* First integer */
result=config[0];/* Expansion board*/
if (result == 0)
{
}
else
{
printf (“Module is a 16 channel device\n”);
num_chan = 16;
if (result == 7)
{
}
else
{
}
}
printf (“Module is an 8 channel device\n”);
num_chan=8;
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Second integer */
result = config [1];/* Terminal Module */
if (result == 0)
{
}
else
{
printf (“Module has a screw type terminal module installed\n”);
if (result == 7)
{
18Keysight E1418A Service Guide
OperationOperating Information
}
else
{
}
}
printf (“Module does NOT have a terminal module installed\n”);
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Third integer */
result=config[2];/* Isolated or non-isolated outputs */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d is configured for non-isolated output\n”,i);
printf(“Channel %d is configured for isolated output\n”,i);
result = result << 1;
}
/* Fourth integer */
result=config[3];/* Output Mode Voltage or Current */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
Keysight E1418A Service Guide19
Operating InformationOperation
}
printf (“Channel %d is set to voltage output mode\n”,i);
printf(“Channel %d is set to current output mode\n”,i);
result = result << 1;
}
/* Fifth integer */
result=config[4];/* Channel output ON or OFF */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d output is disabled\n”,i);
printf(“Channel %d output is enabled\n”,i);
result = result << 1;
}
/* Sixth integer */
result=config[5];/* Program or Jumper Selectable channels */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result >= 0x8000)
{
}
else
{
}
printf (“Channel %d output mode is programmable\n”,i);
printf(“Channel %d output mode is fixed and jumper selected\n”,i);
result = result << 1;
}
•
•
•
}
20Keysight E1418A Service Guide
Preventive MaintenanceOperating Information
Preventive Maintenance
Preventive maintenance consists of periodically cleaning the front panel, the
terminal module, and performing the Operator’s Check. Clean the module and
terminal module yearly, or more often if the module is used in very dusty or
humid areas. Disassembly of the module is not recommended for cleaning.
However, if the module is disassembled for repair or reconfiguration, the printed
circuit board assemblies (PCAs) can be blown off with a properly grounded air
gun. Remove the terminal module cover to blow dust off the terminal printed
circuit board and case. Table 3-1 shows the recommended cleaning equipment
and supplies.
Table 3-1 Recommended Cleaning Equipment and Supplies
DescriptionRecommended Use
Mild Soap SolutionClean face plate panel and terminal covers.
Lint-free ClothClean face plate panel and terminal covers.
Airgun with grounded
nozzle
To eliminate possible electrical shock, disconnect ac
power from the mainframe and disconnect all outputs to
the terminal module before removing the module from the
mainframe.
The printed circuit assemblies (PCAs) contain static-sensitive
devices that can be damaged by careless handling. Use
static-control devices (wrist straps, mats, and tools) when
handling the module. Chapter 6 contains additional precautions
for electrostatic discharge (ESD).
Remove dust from the printed circuit boards and
terminal case.
Keysight E1418A Service Guide21
Operating InformationPreventive Maintenance
Cleaning Procedure
Use the following procedure to clean the Keysight E1418A.
Terminal Module
1 Remove the terminal module from the main module.
2 Disconnect all external wiring.
3 Clean the terminal module covers with a mild soap solution and lint-free
cloth.
4 Remove the terminal module cover and blow any dust or debris from the case.
5 Reassemble the terminal module.
6 Reconnect the external wiring.
Main Module
1 Remove the terminal module from the main module.
2 Clean the main module face plate with a mild soap solution and lint-free
cloth.
Disassembling the module exposes the assemblies to ESD damage
and is usually unnecessary. Perform steps 3 through 5 only if the
module is disassembled for repairs or reconfiguration.
3 Disassemble the module following the instructions in Chapter 6.
4 Using an airgun with a grounded nozzle, remove dust from the PCA
assemblies.
5 Reassemble the module.
22Keysight E1418A Service Guide
Operator’s ChecksOperating Information
Operator’s Checks
The operator’s check for the Keysight E1418A has two parts.
The first part consists of sending the SYStem:ERRor? query to the
module following power-on and entering the response. This check should be
performed after every power-on sequence. At power-on, the Keysight E1418A
performs a brief self-test that ensures the module can communicate with the
backplane. A successful self-test returns an integer and a string:
+0,"No Error"
Any other response indicates a failure in the module. Chapter 6 contains
troubleshooting information.
Once the power-on self-test has been checked, the operator may perform a full
self-test by sending the *TST? query and entering the response. This query
executes a full self-test that checks all circuits of the module except the output
relays. A successful self-test returns:
+0
Any other response indicates a failure. Chapter 6 contains additional information
about the self-test error codes and troubleshooting information.
Operator Check Example
This example checks the module for errors after power-on and then performs a
full self-test.
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
Three test procedures are given in this chapter. These test procedures are used
to verify that:
– the Keysight E1418A is functional (Functional Verification)
– the Keysight E1418A meets selected testable specifications (Quick
Verification)
– the Keysight E1418A meets all testable specifications (Performance
Verification)
Test Conditions
Table 1-1 lists the recommended test equipment. When performing the test
procedures, observe the following test conditions:
– The ambient temperature should be between 18 °C and 28 °C. The
– The relative humidity should be < 65%, non-condensing.
– Install the module, apply power, and allow the module to warm up for at
Keysight Technologies recommends performance tests be performed at one year
intervals. In severe operating environments, or after heavy use, perform the tests
more often.
Do not perform any of the following verification tests unless
you are a qualified, service-trained technician and have
read the WARNINGS and CAUTIONS in Chapter 1 of this
manual.
temperature should be stable within ± 1 °C.
least 15 minutes.
27
Verification TestsAbout the Verification Tests
About the Verification Tests
Three levels of verification tests are described in this chapter:
– Functional Verification Tests
– Quick Verification Tests
– Performance Verification Tests
You should perform the Functional Verification test before either the Quick
Verification or full Performance Verification tests.
Functional Verification Test
The Functional Verification test provides a high confidence that the module is
operational. A brief self-test is performed when power is first applied to the
module. This test assures that module can communicate with the mainframe.
The Functional Verification Test performs a more complete self-test using the
SCPI common command *TST?. If this test passes, you have a high confidence
level (90%) that the module is operational.
If the module fails the Functional Verification Test, repair is needed.
Quick Verification Tests
The Quick Verification Test combines a Functional Verification Test with an
abbreviated set of Performance Tests to give a high confidence level that the
module is operational and meets its specifications. The Quick Performance Test
is the minimum set of tests recommended after any service activity.
These tests check the module’s performance for normal accuracy and drift
mechanisms. These tests do not check for abnormal component failures.
Quick Verification Tests are designated with the letter “Q” in the Performance
Verification Tests. To perform the Quick Verification Tests:
1 1. Perform a Functional Verification Test.
2 2. Perform the Performance Verification Tests designated by the letter“Q” in
the table.
If the module fails the Quick Verification Test, adjustment or repair is needed.
28Keysight E1418A Service Guide
About the Verification TestsVerification Tests
Performance Verification Tests
Performance Verification Tests give a high confidence that the module is
operational and meets its specifications. The Performance Verification Tests can
be used as acceptance tests when the module is first received.
The Performance Verification Tests should be run at the calibration interval
(Keysight Technologies recommends a one year interval). Run the tests to
characterize the module against the specifications. Run the Performance
Verification Tests after any adjustment to verify the adjustment.
If the module fails any Performance Verification Tests, adjustment or repair is
needed.
Performance Test Record
Table 4-1 provides space to enter the results of each Performance Verification
test. The table also lists the upper and lower test limits.
About Program Examples
In this manual test procedures are shown with a portion of an example program
that performs the test. These examples are in ANSI C format and complete
program files are included on the disk supplied with this manual. Most examples
in this chapter are included. The examples are ASCII files with the *.c extension.
The int data type is system dependent. These examples were
developed on a system where int is a 16-bit integer. Other systems
may define int to be a different width.
In the examples, the Keysight VISA Transition Library is used for I/O operations
with the VXIbus. An Keysight command module (Keysight E1405/E1406) is used
and controlled via GPIB.
To use the Keysight VISA Transition Library (abbreviated as VTL), include the
visa.h header file.
#include visa.h
Keysight VTL function calls and data types typically begin with the lower case
letters vi. Output and enter are performed with functions named viPrintf and
viScanf. Both these functions require a session (a VTL defined I/O function) to
uniquely identify the device being controlled. In the examples, the session has
been named dac.
Keysight E1418A Service Guide29
Verification TestsAbout the Verification Tests
SCPI drivers
In this manual, all programming examples and procedures assume the use of the
SCPI driver and Keysight E1406A Command Module. The Keysight E1418A
module is shipped with two 3.5" disks. These disks contain the SCPI driver for the
Keysight E1406A Command Module. One disk is in LIF format and one disk is in
DOS format. Installing Device Drivers Installation Note is included with the disks.
Follow the instructions contained in the installation note to properly install the
device driver.
Use the DIAG:DRIV:LIST? query on the command module to verify that the
correct device driver is installed. Responses to this query vary depending upon
the drivers loaded on your system. A typical response contains a list of all drivers
installed and might look like:
The string E1418,E1418,A.01.00, should be located somewhere within the
returned string.
SCPI Command Reference
The SCPI (Standard Commands for Programmable Instruments) commands are
not documented in this manual except in a general manner. Complete SCPI
commands, including syntax and parameters, applicable to the module are
documented in the Keysight E1418A User’s Guide.
30Keysight E1418A Service Guide
Functional Verification TestVerification Tests
Functional Verification Test
The Functional Verification Test quickly verifies that the module is operational.
Perform this test any time to verify or check the operation. Perform this test
before beginning any Quick Verification or Performance Verification Tests.
Functional Verification Test Procedure
Use the following procedure to verify the module’s operation.
1 Install the module in the mainframe and apply power.
2 Verify the module passed its power-on self-test by executing the
SYStem:ERRor? query and entering the results.
A passed self-test is indicated by the return
+0, “No Error”
any other return value indicates a power-on failure.
If an incorrect module address is used, the module will not
respond. Verify the module’s address before troubleshooting.
3 Execute the full self-test by sending the *TST? query and entering the result.
A passed self-test is indicated by the return
+0
Any other return value indicates a failure. Additional information about the
self-test results and self-test error codes is given in Chapter 6
Functional Verification Test Example
An example of the Functional Verification Test is provided in Operator Check Example..
Keysight E1418A Service Guide31
Verification TestsPerformance Verification Tests
Performance Verification Tests
These procedures test the electrical performance of the Keysight E1418A using
the specification in Appendix A as the performance standards. These tests are
suitable for incoming inspections, troubleshooting, and preventive measures.
DC Voltage Output Verification
This procedure tests the dc voltage output performance of the module.
1 For each channel in the module, connect the 10 kΩ resistor and DMM as
shown in Figure 4-1. Channel 1 connections are shown in the figure. Connect
the other channels in a similar manner. Set the DMM to measure dc volts.
Figure 4-1 DC Voltage Output Connections (Channel 1)
2 Set the channel to output each of the following voltages:
Use the APPLyn:VOLTage xx.xx command, where n is the channel number
and xx.xx is the voltage to output.
3 For each voltage, measure the output voltage with the DMM, enter it in the
table and compare the voltage output to the limits shown in the table.
4 Open the channel output relay by sending the OUTPutn OFF command, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.
DC Voltage Compliance Current
This test checks the compliance current in dc voltage.
1 For each channel in the module, connect the 600 Ω resistor and the DMM as
shown in Figure 4-2. Channel 1 connections are shown in the figure. Connect
the other channels in a similar manner. Set the DMM to measure dc voltage.
Figure 4-2 DC Voltage Output Connections (Channel 1)
2 Set the channel to output +12.00 V. Use the APPLyn:VOLTage 12.00
command, where n is the channel number.
Keysight E1418A Service Guide33
Verification TestsPerformance Verification Tests
3 Measure the output voltage with the DMM, enter it in the table and compare
the measured voltage to the limits shown in the table. The compliance current
is inferred from the voltage across the resistor.
4 Open the channel output relay by sending the OUTPutn OFF command, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.
DC Voltage Short Circuit Output Current
This test checks the short circuit output current for each channel.
1 For each channel in the module, connect the DMM as shown in Figure 4-3.
Channel 1 connections are shown in the figure. Connect the other channels in
a similar manner. Set the DMM to measure dc current (the DMM should have
a current shunt of 100 Ω or less for this test).
Figure 4-3 DC Voltage Output Connections (Channel 1)
2 Set the channel to output +16.00 V. Use the APPLyn:VOLTage 16.00
command, where n is the channel number.
34Keysight E1418A Service Guide
Performance Verification TestsVerification Tests
3 Measure the output current with the DMM, enter it in the table and compare
the measured current to the limits shown in the table.
4 Open the channel output relay by sending the OUTPutn OFFcommand, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.
DC Current Verification
This test checks the dc current output performance of the module.
1 1. For each channel in the module, connect the DMM as shown in Figure 4-4.
Channel 1 connections are shown in the figure. Connect the other channels in
a similar manner. Set the DMM to measure dc current.
Figure 4-4 DC Current Output Connections (Channel 1)
2 2. Set the channel to output each of the following currents:
+0.020–0.005
+0.015–0.010
Keysight E1418A Service Guide35
Verification TestsPerformance Verification Tests
+0.010–0.015
+0.005–0.020
0.000
Use the APPLyn:CURRent x.xxx command, where n is the channel number
and x.xxx is the current to output.
3 3. For each current, measure the output current with the DMM, enter it in the
table and compare the current output to the limits shown in the table.
4 4. Open the channel output relay by sending the OUTPutn OFF command,
where n is the channel number.
5 5. Change the connections to the next channel and repeat steps 2, 3, and 4.
6 6. Repeat steps 2, 3, 4, and 5 for each channel.
36Keysight E1418A Service Guide
Performance Verification TestsVerification Tests
Performance Verification Tests Example
This example demonstrates a full performance verification test of the module.
The code shown in this example can be obtained from the file prftest.c on the
examples disk provided with this manual.
err=viPrintf(dac, “APPLY%d:CURRENT %f;*OPC?\n”,i+1,i_out[j]); /* voltage
output */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
current_results [i][j] = dmm_measure();
}
printf (“Current Output Test on channel %d complete\n\n”,i+1);
err=viPrintf(dac, “OUTPUT%d OFF;*OPC?\n”,i+1); /* Turn output off*/ if(err
VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check */
}
printf (“Current Output Verification Test completed”); err=viPrintf(dac,
“*RST\n”);/* reset the dac*/ if(err VI_SUCCESS) err_handler(dac, err);
}
/*** End of current_check ***/
Keysight E1418A Service Guide41
Verification TestsPerformance Test Record
Performance Test Record
Table 4-1 is a form you can copy and use to record the performance verification
test results of the Keysight E1418A. This table shows the output accuracy,
measurement uncertainty, and the test accuracy ratio (TAR) values. The test
marked with a “Q” in the table are the Quick Verification Tests.
Test Limits
The test limits are defined using the 90-day specifications. Appendix A lists the
Keysight E1418A specifications.
Measurement Uncertainty
Measurement uncertainty has been calculated using the Keysight 34401A
90-day specifications. The Keysight 34401A is set to the 61⁄2 digit mode for all
calculations. In both the voltage and current measurement mode, the
measurement uncertainty is given as:
M.U. = % of reading + % of range
Voltage Output Measurements
The 100 V range is used for the voltage output measurements. The specification
is given as:
M.U. = ± (0.0035% of reading + 0.0006% of range)
Keysight
E1418A Output
+ 16.00 V0.00056 V0.0006 V± 0.00116 V
+ 12.00 V0.00042 V0.0006 V± 0.00102 V
+ 8.00 V0.00028 V0.0006 V± 0.00088 V
+ 4.00 V0.00014 V0.0006 V± 0.00074 V
0.00 V0.00000 V0.0006 V± 0.00060 V
– 4.00 V0.00014 V0.0006 V± 0.00074 V
– 8.00 V0.00028 V0.0006 V± 0.00088 V
– 12.00 V0.00042 V0.0006 V± 0.00102 V
% of reading% of rangeMeasurement
Uncertainty
– 16.00 V0.00056 V0.0006 V± 0.00116 V
42Keysight E1418A Service Guide
Performance Test RecordVerification Tests
DC Compliance Current Measurements
The 100 V range is used for the compliance current test. (The compliance current
is inferred from the voltage developed across the resistor).
M.U. = ± (0.035% of reading + 0.0006% of range)
M.U. = ± (0.00042 + 0.0006) = 0.00102 V
DC Short Circuit Current Measurements
The 100 mA range is used for the short circuit current measurement.
M.U. = ± (0.030% of reading + 0.005% of range)
M.U. = ± (0.000018 + 0.000005) = 0.000023 A
Current Output Measurements
The 100 mA range is used for the current output measurements.
Where Maximum Allowed, Input, and Measurement Uncertainty are all in Vdc or
Amps.
The TAR is listed in table 4-1.
Keysight E1418A Service Guide43
Verification TestsPerformance Test Record
Table 4-1 Performance Test Record
Test Facility:
NameReport No.
AddressDate
City/StateCustomer
PhoneTested by
ModelKeysight E1418AAmbient Temperature
Serial No.Relative Humidity
Options
Firmware Rev.
Notes:
Test Equipment UsedModelTrace NumberCal Due Date
44Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
1
1
1
1
1
1
1
1
1
2
2
2
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
MinMeasuredMaxM.U.TAR
Voltage Output Check
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
2
Q
Q
Q
Q
Q
2
2
2
2
2
3
3
3
3
3
3
3
3
3
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Keysight E1418A Service Guide45
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
4
4
4
4
4
4
4
4
4
5
5
5
5
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
MinMeasuredMaxM.U.TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
Q
Q
Q
Q
Q
5
5
5
5
5
6
6
6
6
6
6
6
6
6
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
46Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
Q
Q
Q
Q
Q
Q
ChannelE1418A
Output
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
MinMeasuredMaxM.U.TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Q
Q
Q
9
9
9
9
9
9
9
9
9
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Keysight E1418A Service Guide47
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
10
10
10
10
10
10
10
10
10
11
11
11
11
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
MinMeasuredMaxM.U.TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
Q
Q
Q
Q
Q
11
11
11
11
11
12
12
12
12
12
12
12
12
12
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
48Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
13
13
13
13
13
13
13
13
13
14
14
14
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
MinMeasuredMaxM.U.TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
14
Q
Q
Q
Q
14
14
14
14
14
15
15
15
15
15
15
15
15
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
Q
15
-16.000 V
–16.011
–15.989 V
1.16E-3 V
9.5:1
Keysight E1418A Service Guide49
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
ChannelE1418A
Output
16
16
16
16
16
16
16
16
16
1
2
3
4
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
MinMeasuredMaxM.U.TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
DC Voltage Compliance Current Check
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
8.8:1
8.8:1
8.8:1
8.8:1
5
6
7
8
9
10
11
12
13
14
15
16
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
50Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
ChannelE1418A
Output
1
2
3
4
5
6
7
8
9
10
11
12
13
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
MinMeasuredMaxM.U.TAR
DC Voltage Short Circuit Current Check
NA NA NA NA NA
NA NA NA NA NA
NA NA NA NA NA
NA
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA
NA
14
15
16
Q
Q
Q
1
1
1
1
1
1
1
1
1
+16.000 V
+16.000 V
+16.000 V
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
DC Current Output Check
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
0.060 A
0.060 A
0.060 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide51
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
2
2
2
2
2
2
2
2
2
3
3
3
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
MinMeasuredMaxM.U.TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
3
Q
Q
Q
Q
3
3
3
3
3
4
4
4
4
4
4
4
4
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
Q
4
–0.02000 A
–0.020023 A
–0.019977 A
1.10E-5 A
2.1:1
52Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
5
5
5
5
5
5
5
5
5
6
6
6
6
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
MinMeasuredMaxM.U.TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
6
6
6
6
6
7
7
7
7
7
7
7
7
7
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide53
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
8
8
8
8
8
8
8
8
8
9
9
9
9
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
MinMeasuredMaxM.U.TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
9
9
9
9
9
10
10
10
10
10
10
10
10
10
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
54Keysight E1418A Service Guide
Performance Test RecordVerification Tests
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
11
11
11
11
11
11
11
11
11
12
12
12
12
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
MinMeasuredMaxM.U.TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
12
12
12
12
12
13
13
13
13
13
13
13
13
13
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide55
Verification TestsPerformance Test Record
Quick
Check
Q
Q
Q
Q
ChannelE1418A
Output
14
14
14
14
14
14
14
14
14
15
15
15
15
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
MinMeasuredMaxM.U.TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
15
15
15
15
15
16
16
16
16
16
16
16
16
16
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
56Keysight E1418A Service Guide
5Adjustments
Introduction
Keysight Technologies Calibration Services
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
This chapter contains procedures for adjusting the calibration constants in the
Keysight E1418A. Run the Performance Tests (described on Chapter 4) before
any adjustments to characterize the module. Run the Performance Tests after
any adjustments to verify the adjustments made. The module should be adjusted
following any reconfiguration or repair.
Contact the local Keysight Technologies Service Center for low cost calibration
services. The Keysight E1418A is supported on automated calibration systems
which allow calibration at competitive prices. Calibrations to MIL-STD-45662 are
also available.
Calibration Interval
The Keysight E1418A 8/16 Channel D/A Converter Module should be calibrated
on a regular interval determined by the accuracy requirements of you
application. Keysight Technologies recommends calibration every 90 days for
most applications. Demanding applications may require a daily calibration cycle.
Keysight Technologies does not recommend extending the calibration interval
beyond one year in any application.
Whatever calibration interval you use, Keysight Technologies recommends that a
complete adjustment be performed. This increases confidence that the module
will remain in specification for the next calibration interval and provides the best
measure of the module’s long-term stability. Performance Verification Test data
can be used to extend or determine future calibration intervals.
Closed-Cover Electronic Calibration
The Keysight E1418A features closed-cover electronic calibration. There are no
internal mechanical adjustments. Once properly installed, the module can be
tested and adjusted using connections to a terminal module.
Do not perform any of the following adjustment procedures unless
you are a qualified, service trained technician and have read the
warnings and cautions.
57
AdjustmentsIntroduction
Calibration Constants and Non-Volatile Memory
The accuracy of a channel output depends upon a number of calibration
constants. Calibration constants are stored in two locations within the module; in
non-volatile memory and in RAM. The constants in RAM are used by the module
to adjust all outputs during use.
When the module is used with the SCPI driver, the RAM constants are loaded
from the non-volatile memory at power-up. You can adjust the RAM calibration
constants without disturbing the non-volatile memory constants (creating a
temporary adjustment).
Non-volatile memory has a finite number of writes. Writing the calibration
constants to non-volatile memory, therefore, reduces the life of this memory. If
you are calibrating the module at 90 day or 1 year intervals, write the new
constants into non-volatile memory. The new constants will be used following
power-up. If you are adjusting the module more frequently (i.e., daily), write the
new calibration constants into RAM, but do not write to non-volatile memory.
The adjustment procedures in this chapter demonstrate both methods of writing
calibration constants.
58Keysight E1418A Service Guide
Making ConnectionsAdjustments
Making Connections
All adjustments can be performed using the CAL output terminals. The CAL
output terminals provide one set of connections for both voltage and current
adjustment. Procedures in this chapter assume connections are made at the CAL
output terminals on the screw-type terminal module.
For demanding applications or for adjustment using the intended
load, you can perform the adjustment at each channel output (to
include the output relay contacts, wiring, and the load in the
path). To make the adjustments at each channel output, use the
DIAG:CAL:OUTP CHAN command to set the adjustment point at
the channels instead of the CAL output terminals. Adjustment at
each channel output will require individual four-wire connections.
Figure 5-1 CAL Output Terminals
Keysight E1418A Service Guide59
AdjustmentsAdjustment Procedure
Adjustment Procedure
Adjustment is performed on each channel, one at a time. Voltage output and
current output each use unique calibration constants.
Preparation
Before performing any adjustments, determine if the adjustments are to be
temporary or permanent.
If Permanent
1 Turn off the mainframe.
2 Remove the terminal module.
3 Remove the module from the mainframe.
4 Set the Cal Store Enable Jumper to the CAL position as shown below.
Figure 5-2 Cal Store Enable Jumper
5 Re-install the module in the mainframe.
6 Make the CAL output terminal connections on the terminal module.
7 Install the terminal module.
8 Apply power and allow a 15 minute warm-up.
60Keysight E1418A Service Guide
Adjustment ProcedureAdjustments
If Temporary
1 Make the CAL output terminal connections on the terminal module.
2 Install the terminal module.
3 Apply power and allow a 15 minute warm-up.
Voltage Adjustment
Voltage output adjustment uses two of the CALibration subsystem queries.
CALibrationn:CONFigure:VOLTage?
CALibrationn:VALue:VOLTage? <value>
The general procedure for channel voltage adjustment is:
1 Make the connections shown in figure 5-3 or 5-4.
2 Send the CALn:CONF:VOLT? query to the module and read the response until
a 1 is returned (typically the first query response).
3 Read the voltage output on the DMM.
4 Send the DMM value read with the CALn:VAL:VOLT? value query.
5 Enter the query response.
6 Repeat steps 3, 4, and 5 until the CALn:VAL:VOLT? value query returns a 0.
7 Repeat steps 1 through 6 for each channel.
The first query, CALn:CONF:VOLT?, sets the channel to the calibration mode and
returns a 1 when the channel is ready for calibration. When a 1 is returned, the
channel output voltage is set to the first adjustment point.
Read the output voltage on the DMM and return the value obtained using the
CALn:VAL:VOLT? value query. This query returns an integer indicating the
calibration state of the channel. Any non-zero return from this query indicates
additional values are needed.
Each channel will require multiple iterations of the CALn:VAL:VOLT? value query.
A minimum of 9 queries, to a maximum of 50 queries, will be required at each
channel. During the process, the output voltage will range from +16 V to -16 V
and the last values output will be at or near 0 volts.
Keysight E1418A Service Guide61
AdjustmentsAdjustment Procedure
Figure 5-3 Voltage Adjustment Connections (CAL)
62Keysight E1418A Service Guide
Adjustment ProcedureAdjustments
Figure 5-4 Optional Voltage Adjustment Connections
Current Adjustment
Current output uses an internal resistance value to calibrate all channels. This
resistance value is approximately 244 Ω.. The current is output adjusted by
measuring the voltage developed across this resistor.
To set the internal resistance value use the following CALibration commands:
CALibration:CONFigure:RESistance?
CALibration:VALue:RESistance <value>
You may also, optionally, use an external resistance value (or
load) on each channel for current output and calibration. If you
wish to specify an external resistor value, you must send the
DIAG:CAL:OUTP CHAN command and perform the adjustment at
each channel output. Any load value may be used but values in
the range of 50 Ω to 500 Ω are recommended.
Note that during the adjustment process self-heating of the
external load may affect the adjustment.
Keysight E1418A Service Guide63
AdjustmentsAdjustment Procedure
The resistance value set, either internal or external, is not stored. When using the
internal resistor value, the general procedure is:
1 Make the connections shown in Figure 5-5.
2 Send the CALibration:CONFigure:RESistance? query to the module and read
the response until a 1 is returned (typically the first query response).
3 Read the resistance on the DMM.
4 Send the DMM value read to the module using the
CALibration:VALue:RESistance <value> command.
The resistor value is only set once for current calibration of all channels.
Optional
If you are using an external resistor value, for each channel:
1 Send the CALibration:CONFigure:RESistance? query to the module and read
the response until a 1 is returned (typically the first query response).
2 Measure the channel’s external resistor value on the DMM.
Current output adjustment uses two of the CALibration subsystem queries.
Current adjustment requires voltage measurements.
CALibrationn:CONFigure:CURRent?
CALibrationn:VALue:CURRent? <value>
The current adjustment procedure measures and inputs a voltage
value developed across a known resistor value (set at the start of
the procedure).
Once the calibration resistance value is set as described earlier, the general
procedure for each channel current output adjustment is:
1 Make the connections shown in figure 5-6 or 5-7.
2 Send the CALn:CONF:CURR? query to the module and read the response until
a 1 is returned (typically the first query response).
3 Read the voltage on the DMM.
Keysight E1418A Service Guide65
AdjustmentsAdjustment Procedure
4 Send the DMM value read with the CALn:VAL:CURR? <value> query.
5 Enter the query response.
6 Repeat steps 3, 4, and 5 until the CALn:VAL:CURR? <value> query returns a 0.
7 Repeat steps 1 through 6 for each channel.
The first query, CALn:CONF:CURR?, sets the channel to the calibration mode and
returns a 1 when the channel is ready for calibration. When a 1 is returned, the
channel output voltage is set to the first adjustment point.
Read the voltage on the DMM and return the value obtained using the
CALn:VAL:CURR? <value> query. This query returns an integer indicating the
calibration state of the channel. A value other than 0 or 2 from this query
indicates additional values are needed.
Each channel will require multiple iterations of the CALn:VAL:CURR? <value>
query. A minimum of 9 queries, to a maximum of 50 queries, will be required at
each channel. During the process, the output voltage will range from -5 to +5 V
and the last values output will be at or near 0 volts.
Figure 5-6 Current Adjustment Connections (CAL)
66Keysight E1418A Service Guide
Adjustment ProcedureAdjustments
Figure 5-7 Optional Current Adjustment Connections
Storing the Adjustments
Once all adjustments have been made, you can store the new calibration
constants in non-volatile memory by sending the following command:
/* Store the new constants */
CALibration:STORe
This command will generate an error if the Cal Store Enable Jumper is not set to
the CAL position.
The Keysight E1418A will be busy storing the new calibration
constants following the CAL:STOR command. You should not
perform any bus resets or module resets until the store is
complete. You can use the *OPC? query to determine when the
command has finished.
Verify the Adjustments
Run the Performance Verification Tests to verify the adjustments made.
Keysight E1418A Service Guide67
AdjustmentsAdjustment Procedure
Return the Module to Use
If you have changed the position of the Cal Store Enable Jumper and you wish to
prevent writing calibration constants to non-volatile memory, set the jumper to
the SECURE position.
If you have changed the position of the terminal jumpers return the jumpers to
their original positions.
Example Program
The following program, written in ANSI C illustrates a full calibration of all
channels for both voltage and current. The internal calibration resistor is used.
The disk provided with this manual contains a calibration example file. This
example uses function calls to the Keysight VISA Transition Library for I/O
operations.
void sys_err(ViSession resource); /* Checks for SCPI programming
errors */ void pause(); /* Waits for a keystroke to
continue program execution */ int config_check ();
/* Checks and reports the module configuration */
/* returns the number of channels in module*/ float get_voltage (); /*
Obtains the measured voltage*/ float get_resistance ();/* Obtains the
measured resistance */
•
•
•
void main (void)
{
char cal_point [5],int_ext [4];
int num_chan,result,compare,i,condition [4];
float meas_volt,meas_res;
•
•
•
/*** VOLTAGE OUTPUT ADJUSTMENT ***/
for (i=1;i<num_chan+1;i++)
68Keysight E1418A Service Guide
Adjustment ProcedureAdjustments
{
result = 0;
while (result < 1)
{
err=viPrintf(dac, “CAL%d:CONF:VOLT?\n”,i); /* Configure for calibration */
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 1 when ready to calibrate */
if(err < VI_SUCCESS) err_handler(dac, err);
}
result = 1;
while (result !=0)
{
meas_volt=get_voltage ();/* Get the voltage measurement */
err=viPrintf(dac, “CAL%d:VAL:VOLT? %f\n”,i,meas_volt); /* Send the
measurement*/
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 0 when calibration complete */
if(err < VI_SUCCESS) err_handler(dac, err);
if (result == 5)/* Calibration for this channel aborted*/
{
printf (“Calibration for channel %d aborted\n”,i);
result = 0;
}
}
}
/***CURRENT OUTPUT ADJUSTMENT***/
/*** set the calibration resistor value***/
result = 0;
while (result != 1)
{
err=viPrintf(dac, “CAL:CONF:RES?\n”);/* Configure for resistance calibration
*/
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 1 when ready to calibrate*/
if(err < VI_SUCCESS) err_handler(dac, err);
}
meas_res = get_resistance();
Keysight E1418A Service Guide69
AdjustmentsAdjustment Procedure
err=viPrintf(dac, “CAL:VAL:RES %f\n”,meas_res); /* Send the resistance value
*/
if(err < VI_SUCCESS) err_handler(dac, err);
/*** channel current output adjustment ***/
for (i=1;i<num_chan+1;i++)
{
result = 0;
while (result != 1)
{
err=viPrintf(dac, “CAL%d:CONF:CURR? \n”,i); /* Configure for calibration */
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/*, returns a 1 when ready to calibrate */
if(err < VI_SUCCESS) err_handler(dac, err);
}
result = 1;
while (result !=0 && result !=2)
{
meas_volt=get_voltage ();/* Get the voltage measurement */
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 0 when calibration complete */
if(err < VI_SUCCESS) err_handler(dac, err);
if (result == 5)/* Calibration aborted */
{
printf (“Calibration for channel %d aborted\n\n”,i);
result = 0;
}
}
}
/*** STORE THE NEW CALIBRATION CONSTANTS***/
err=viPrintf(dac, “CAL:STOR;*OPC?\n”);/* Store the new cal constants */
if(err < VI_SUCCESS) err_handler(dac, err);
err=(viScanf(dac,"%d",&result);
if(err < VI_SUCCESS) err_handler(dac, err);
sys_err(dac);
•
70Keysight E1418A Service Guide
Adjustment ProcedureAdjustments
•
•
}
Keysight E1418A Service Guide71
AdjustmentsAdjustment Procedure
72Keysight E1418A Service Guide
6Service
Introduction
Equipment Required
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
This chapter contains troubleshooting, repair and maintenance guidelines.
Do not perform any of the service procedures shown unless you are
a qualified, service-trained technician, and have read the warnings
cautions .
Test equipment required is listed in Table 1-1. Any equipment that satisfies the
requirements listed in the table may be used.
The Keysight E1418A uses Torx head screws. To avoid damage to the screw head
slots, use only a Torx driver for disassembly and assembly. You will need a T10
Torx driver (Keysight Part Number 8710-1284).
Service Aids
Service aids, manual updates, and service literature may be available for the
Keysight E1418A. For information, contact your nearest Keysight Technologies
Sales and Service Office.
73
ServiceTroubleshooting
Troubleshooting
This section will help isolate a failing assembly.
Visual Checks
Before installing or operating the module, visually inspect the rear panel and
front panel connectors for bent pins. Inspect circuit boards for evidence of arcing
or excessive heat.
If a module does not respond to commands, verify the Logical Address setting.
The Keysight E1418A User’s Guide describes this switch setting.
Catastrophic Failures
If a module does not respond, verify the power fuses F1, F2, F3 and F4.
Self-Test
The internal self-test can be used to isolate most failures of the module. The
major assemblies of the Keysight E1418A are available on an exchange basis.
Use the following SCPI Common Command query to initiate a module self-test.
/* Start the Self-test*/
*TST?
/* Enter the 16-bit integer returned*/
The query will return a +0 if no errors are encountered. Any return other than +0
indicates a failure. The command returns the first test to fail, additional failures
are possible.
The *TST? query initiates a full self-test. You can obtain the complete test results
by using the TEST:TST:RESults? query. This query returns an array of 100
integers representing the results of each self-test.
Use the following SCPI commands to initiate a self-test and obtain complete
results.
/* Start the Self-test*/
*TST?
/* Enter the 16-bit integer returned*/
/* If the returned value is not +0, send*/ TEST:TST:RES?
/*Enter an array of 100 integers*/
74Keysight E1418A Service Guide
TroubleshootingService
The integers returned from the self test query have the form:
Bit1514131211109876543210
UseTest NumberA24Test Information
General Self-Test
The test number returned by the *TST? query can be used to isolate a failing
assembly. The possible test numbers and probable failing assembly are shown
below.
Test NumberFailing Assembly IndicatedNotes
0NoneAll Self-tests passed
1Main Board
2 through 9Main Board or Plug-On ModulesTest 2 = channel 1, Test 3 = channel 2, etc.
10 through 17Expansion Board or Plug-on ModulesTest 10 = channel 9, Test 11 = channel 10, etc.
18 through 25Main Board
26 through 33Main Board
34 through 41Expansion Board
42 through 43Main Board
44 through 51Main Board
52 through 59Expansion Board
60 through 67Main Board
68 through 75Expansion Board
Since the *TST? query returns only the first failure test number,
additional failures are possible but not reported by this query. The
complete test results can be obtained using the TEST:TST:RES?
query after the *TST? query.
Keysight E1418A Service Guide75
ServiceTroubleshooting
Detailed Self-Test
Following the *TST? query, the TEST:TST:RESults? query returns complete
self-test details. The following table lists each self-test and the areas checked.
Test NumberA24 ValueTest Information
Value If Failed
101Relay Control Register
02A/D multiplexer
2 through 170Channel NumberFirst DAC check of each channel
180 for A16 read
1 for A16 read in
A24 window
2 for A16 read in
A24
190 for A16 read/write
1 for A16 read/write
via A24 window
200Register ValueVXI Control Register
210Register ValueCal Control Register
22Lower 8 bits of failed register valueChannel Mode Register
230 for A16 writes
A16 reads
1 for A16 writes
A24 reads
2 for A24 writes
A16 reads
3 for A24 writes
A24 reads
Register AddressID Register, Device Type Register, VXI Offset
Register ValueVXI Control Register
Register ValueCard Control Register
Areas Checked
Register,
Card Configuration Register, Isolation Status
Register,
Channel Program Jumper Register
VXI Status Register
240Checksum ErrorNon-volatile memory checksum
25Register ValueTrigger Control Register
76Keysight E1418A Service Guide
TroubleshootingService
Test NumberA24 ValueTest Information
Value If Failed
26 through 41
(Test 26 = Ch 1
Test 41 = Ch 16)
420Register ValueInterrupt Control Register
4301 through 5Interrupt Status Register, SW Trigger Register,
01Channel SW Trigger Register
02Channel WriteThru Register
03Channel Output
04Channel Triggered Register
05Channel Output
06Channel SW and Trigger Registers
07 through 16Channel SW and Triggered Registers
017Channel Triggered Register
018Channel Output
019Channel SW and Triggered Registers
020Channel Output
Areas Checked
and trig ignored bit
44 through 59
(Test 44 = Ch 1
Test 59 = Ch 16)
01 through 5Voltage mode, Set the Gain/Offset Register to
nominal and 5 incrementing values, read the
values in the Calibration Control Register
06Check linearity of values
07Gain Register set to minimum, Channel
Writethru register set to 1 V, read the
Calibration Control Register
08Gain Register set to maximum,
read the Calibration Control Register, calculate
change from previous test
09Gain Register set to nominal, Offset Register
set to minimum, Channel Writethru register set
to -1 V, read the Calibration Control Register
010Offset Register set to maximum,
read the Calibration Control Register, calculate
change from previous test
Keysight E1418A Service Guide77
ServiceTroubleshooting
Test NumberA24 ValueTest Information
Value If Failed
60 through 75
(Test 60 = Ch 1
Test 75 = Ch 16)
76 through 10000Not Used
01 through 5Current mode, Set the Gain/Offset Register to
06Check linearity of values
07Gain Register set to minimum, Channel Write
08Gain Register set to maximum,
09Gain Register set to nominal, Offset Register
010Offset Register set to maximum,
Areas Checked
nominal and 5 incrementing values, read the
values in the Calibration Control Register
thru register set to 1 V, read the Calibration
Control Register
read the Calibration Control Register,
calculate change from previous test
set to minimum, Channel Write thru register
set to -1 V, read the Calibration Control
Register
read the Calibration Control Register,
calculate change from previous test
Single Channel Failures
A single channel failure, not caught by the self-test, can be caused by one of two
components: the channel plug-on module or the channel output relay.
To help isolate a plug-on module, disassemble the module and swap two
plug-on modules. DO NOT operate the module without a plug-on module
installed for every channel.
If a channel still fails after swapping the plug-on modules, suspect the channel
output relay.
78Keysight E1418A Service Guide
Repair/Maintenance GuidelinesService
Repair/Maintenance Guidelines
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 volts. Observe the following guidelines when servicing the
Keysight E1418A.
– Disassemble and reassemble only in a static-free work area.
– Remove all plastic, Styrofoam, vinyl, paper, and other static-generating
materials from the immediate work area.
– Use a conductive work area to dissipate static charge.
– Use a conductive wrist strap to dissipate static charge.
– Minimize handling.
– Do not stack assemblies.
– Use only anti-static solder suckers, soldering irons, and tools.
– Keep replacement parts in original static-free packaging.
Surface Mount Repair
The Keysight E1418A uses surface mount components. Surface mount
components should only be removed using soldering irons or desoldering
stations expressly designed for surface-mount components. Use of conventional
soldering equipment will almost always result in permanent damage to the
printed circuit board and will void your Keysight Technologies warranty.
Disassembly
Tools Needed
– T10 Torx Driver.
– Static-Safe Workstation.
– Static Wrist Strap.
Keysight E1418A Service Guide79
ServiceRepair/Maintenance Guidelines
Almost all electrical components can be damaged by electrostatic
discharge (ESD) during handling. Component damage can occur
at electrostatic discharge voltages as low as 50 volts.
Disassemble and re-configure only in a static free work area.
Minimize handling of assemblies and components. Keep all
assemblies and replacement parts in the original static free
packaging.
These procedures should be performed by qualified service
personnel at approved static-safe workstations.
Disassembly Procedure
Refer to Figures 6-1 and 6-2 during these procedures.
1 Remove the 8 T-10 Torx screws in the top cover and remove the top cover.
2 Remove the T-10 Torx screw holding the sheet metal spacer or expansion
board.
a If the Keysight E1418A is an 8 channel device, lift the sheet metal spacer
off the main board.
b If the Keysight E1418A is a 16 channel device, carefully separate the
expansion board and main board printed circuit assemblies. The two
boards are joined by three electrical connectors.
3 The main board and expansion board each have 8 plug-on assemblies. To
remove the plug-on assemblies, remove the T-10 Torx screw securing the
plug-on assembly. Lift the plug-on assembly off the electrical connector. The
plug-on assembly locations are shown in Figure 6-3.
Reassembly procedure
1 Replace all plug-on assemblies. DO NOT operate the module without a
plug-on assembly installed. The plug-on assemblies have a locator hole to
ensure correct orientation, see Figure 6-4.
2 Replace the expansion board or sheet-metal spacer.
3 Replace the top cover.
80Keysight E1418A Service Guide
Repair/Maintenance GuidelinesService
Figure 6-1 8-Channel Disassembly
Keysight E1418A Service Guide81
ServiceRepair/Maintenance Guidelines
Figure 6-2 16-Channel Disassembly
82Keysight E1418A Service Guide
Repair/Maintenance GuidelinesService
Figure 6-3 Plug-on Channel Locations
Figure 6-4 Installing Plug-on Modules
Keysight E1418A Service Guide83
ServiceRepair/Maintenance Guidelines
84Keysight E1418A Service Guide
Keysight E1418A 8/16 Channel D/A Converter Module
Service Guide
7Replaceable Parts
Introduction
This chapter contains the following topics:
– Assembly Exchange
– Ordering Information
– Replaceable Parts
Assembly Exchange
The following assemblies are available for exchange.
Part numberAssembly
E1418-69201Main board and all sheet metal. Does NOT include plug-on
assemblies or expansion board.
E1418-69502Expansion board. Does NOT include plug-on assemblies.
E1418-66503Isolated plug-on module (replacement, not an exchange).
When exchanging either the main board or the expansion board,
you must disassemble the module and remove all plug-on
assemblies before sending the exchange part. Keep all parts
removed in a safe place.
If you are exchanging a main board from a module configured as a
16-channel device, you must remove the expansion board before
sending the exchange part.
Almost all electrical components can be damaged by electrostatic
discharge (ESD) during handling. Component damage can occur
at electrostatic discharge voltages as low as 50 volts.
Disassemble and re-configure only in a static free work area.
Minimize handling of assemblies and components. Keep all
assemblies and replacement parts in the original static free
packaging.
85
Replaceable PartsOrdering Information
These procedures should be performed by qualified service personnel at
approved static safe workstations.
To Exchange an Assembly
1 Contact the nearest Keysight Technologies Sales and Service Office for
instructions on packaging and shipping.
2 Disassemble the module, using the procedures given in Chapter 6 of this
manual, until just the exchange part is obtained. Remove the plug-on
modules and expansion board if any.
3 Keep all removed parts in a static-safe place. Do not send plug-on modules
with the exchange assembly. Plug-on modules should be individually
replaced, if necessary.
4 Package the assembly in static-safe packaging material.
5 Include a description of the problem encountered as well as your phone
number and return address.
Ordering Information
To order a part listed as an exchange assembly or a part from Table 7-1, 7-2 or
7-3, specify the Keysight part number and the quantity desired. Send the order
to your nearest Keysight Technologies Sales and Service Office.