Keysight E1418A Series, E1418A, E1418A Option 003, E1418A Option 001, E1418A Option002 Service Manual

75000 Series C

Service Manual

Keysight E1418A 8/16 Channel D/A Converter
Notices
© Keysight Technologies, Inc. 1996-2019
No part of this manual may be repro­duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con­sent from Keysight Technologies, Inc. as governed by United States and interna­tional copyright laws.
Manual Part Number
E1418-90010
Edition
Second Edition, October 2019
Published by
Keysight Technologies, Inc. 900 S. Taft Ave. Loveland, CO 80537 USA
Sales and Technical Support
To contact Keysight for sales and techni­cal support, refer to the support links on the following Keysight websites:
www.keysight.com/find/E1418A
(product-specific information and sup­port, software and documentation updates)
www.keysight.com/find/assist (world-
wide contact information for repair and service)
Declaration of Conformity
Declarations of Conformity for this prod­uct and for other Keysight products may be downloaded from the Web. Go to
http://keysight.com/go/conformity and
click on “Declarations of Conformity.” You can then search by product number to find the latest Declaration of Conformity.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDI­TIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WAR­RANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFOR­MATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEP­ARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Keysight Technologies does not warrant third-party system-level (combination of chassis, controllers, modules, etc.) per­formance, safety, or regulatory compli­ance unless specifically stated.
DFARS/Restricted Rights Notices
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ stan­dard commercial license terms, and non­DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Gov­ernment users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227­7015 (b)(2) (November 1995), as applica­ble in any technical data.
Safety Information
The following general safety precau­tions must be observed during all phases of operation of this instrument. Failure to comply with these precau­tions or with specific warnings or oper­ating instructions in the product manuals violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technolo­gies assumes no liability for the cus­tomer's failure to comply with these requirements.
General
Do not use this product in any manner not specified by the manufacturer. The protec­tive features of this product must not be impaired if it is used in a manner specified in the operation instructions.
Before Applying Power
Verify that all safety precautions are taken. Make all connections to the unit before applying power. Note the external markings described under “Safety Symbols”.
Ground the Instrument
Keysight chassis’ are provided with a grounding-type power plug. The instrument chassis and cover must be connected to an electrical ground to minimize shock hazard. The ground pin must be firmly connected to an electri­cal ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in per­sonal injury.
Do Not Operate in an Explosive Atmosphere
Do not operate the module/chassis in the presence of flammable gases or fumes.
Do Not Operate Near Flammable Liquids
Do not operate the module/chassis in the presence of flammable liquids or near containers of such liquids.
Cleaning
Clean the outside of the Keysight mod­ule/chassis with a soft, lint-free, slightly dampened cloth. Do not use detergent or chemical solvents.
Do Not Remove Instrument Cover
Only qualified, service-trained person­nel who are aware of the hazards involved should remove instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Keep away from live circuits
Operating personnel must not remove equipment covers or shields. Proce­dures involving the removal of covers and shields are for use by service­trained personnel only. Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous elec­trical shock, DO NOT perform proce­dures involving cover or shield removal unless you are qualified to do so.
DO NOT operate damaged equipment
Whenever it is possible that the safety protection features built into this prod­uct have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service­trained personnel. If necessary, return the product to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
DO NOT block the primary disconnect
The primary disconnect device is the appliance connector/power cord when a chassis used by itself, but when installed into a rack or system the dis­connect may be impaired and must be considered part of the installation.
Do Not Modify the Instrument
Do not install substitute parts or per­form any unauthorized modification to the product. Return the product to a Keysight Sales and Service Office to ensure that safety features are main­tained.
In Case of Damage
Instruments that appear damaged or defective should be made inoperative and secured against unintended oper­ation until they can be repaired by qualified service personnel
Do NOT block vents and fan exhaust: To ensure adequate cooling and venti­lation, leave a gap of at least 50mm (2") around vent holes on both sides of the chassis.
Do NOT operate with empty slots: To ensure proper cooling and avoid dam­aging equipment, fill each empty slot with an AXIe filler panel module.
Do NOT stack free-standing chassis: Stacked chassis should be rack­mounted.
All modules are grounded through the chassis: During installation, tighten each module's retaining screws to secure the module to the chassis and to make the ground connection.
Operator is responsible to maintain safe operating conditions. To ensure safe operating conditions, modules should not be operated beyond the full temperature range specified in the Environmental and physical specifica­tion. Exceeding safe operating condi­tions can result in shorter lifespan, improper module performance and user safety issues. When the modules are in use and operation within the specified full temperature range is not maintained, module surface tempera­tures may exceed safe handling condi­tions which can cause discomfort or burns if touched. In the event of a module exceeding the full temperature range, always allow the module to cool before touching or removing modules from the chassis.
4
Safety Symbols
A CAUTION denotes a hazard. It calls attention to an operating pro­cedure or practice, that, if not cor­rectly performed or adhered to could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated condi­tions are fully understood and met.
A WARNING denotes a hazard. It calls attention to an operating pro­cedure or practice, that, if not cor­rectly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi­cated conditions are fully under­stood and met.
Products display the following sym­bols:
Warning, risk of electric shock
Refer to manual for addi­tional safety information.
Earth Ground.
Chassis Ground.
Alternating Current (AC).
Direct Current (DC)
5
6
Contents
1 General Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Recommended Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1 Installation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Initial Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Command Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ordering Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Field Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Terminal Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Terminal Module Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Terminal Module Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Shipping Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2 Operating Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Querying Module Identification and Configuration . . . . . . . . . . . . . . . . . . . 15
Configuration Example. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Cleaning Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Main Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Operator’s Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Operator Check Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Keysight E1418A Service Guide 7
3 Verification Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Test Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
About the Verification Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Functional Verification Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Quick Verification Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Performance Verification Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Performance Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
About Program Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
SCPI drivers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
SCPI Command Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Functional Verification Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Functional Verification Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Functional Verification Test Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Performance Verification Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
DC Voltage Output Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
DC Voltage Compliance Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DC Voltage Short Circuit Output Current. . . . . . . . . . . . . . . . . . . . . . . . . . . 34
DC Current Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Performance Verification Tests Example. . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Performance Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Test Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Voltage Output Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
DC Compliance Current Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
DC Short Circuit Current Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Test Accuracy Ratio (TAR). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4 Adjustments
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Keysight Technologies Calibration Services. . . . . . . . . . . . . . . . . . . . . . . . . 57
Calibration Interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Closed-Cover Electronic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Calibration Constants and Non-Volatile Memory . . . . . . . . . . . . . . . . . . . . 58
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
If Permanent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
If Temporary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Voltage Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Current Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Optional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8 Keysight E1418A Service Guide
5 Service
Storing the Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Verify the Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Return the Module to Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Example Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Service Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Visual Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Catastrophic Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
General Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Detailed Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Single Channel Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Repair/Maintenance Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Electrostatic Discharge (ESD) Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 79
Surface Mount Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Tools Needed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
6 Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Assembly Exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
To Exchange an Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7 Manual Changes
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
A Keysight E1418A Specifications
DC Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
DC Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
General Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Cooling Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Keysight E1418A Service Guide 9
10 Keysight E1418A Service Guide

1 General Information

Introduction

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This manual contains information required to test, troubleshoot, and repair the Keysight E1418A 8/16 Channel D/A Converter Module. See the Keysight E1418A User’s Manual for additional module information. Figure 1-1 shows the Keysight E1418A module.

Figure 1-1 Keysight E1418A D/A Converter Module

11
General Information Safety Considerations

Safety Considerations

This product is a Safety Class I instrument that is provided with a protective earth terminal when installed in the mainframe. The instrument, mainframe, and all related documentation should be reviewed for familiarization with safety markings and instructions before operation or service.
Refer to the WARNINGS page (page 6) in this manual for a summary of safety information. Safety information for preventive maintenance, testing, adjusting, and service follows and is also found throughout this manual.

Warnings and Cautions

This section contains WARNINGS which must be followed for your protection and CAUTIONS which must be followed to avoid damage to the equipment when performing instrument maintenance or repair.
SERVICE-TRAINED PERSONNEL ONLY. The information in this manual is for service-trained personnel who are familiar with electronic circuitry and are aware of the hazards involved. To avoid personal injury or damage to the instrument, do not perform procedures in this manual or do any servicing unless you are qualified to do so.
CHECK MAINFRAME POWER SETTINGS. Before applying power, verify that the mainframe setting matches the line voltage and that the correct fuse is installed. An uninterruptible safety earth ground must be provided from the main power source to the supplied power cord set.
GROUNDING REQUIREMENTS. Interruption of the protective (grounding) conductor (inside or outside the mainframe) or disconnecting the protective earth terminal will cause a potential shock hazard that could result in personal injury. (Grounding one conductor of a two-conductor outlet is not sufficient protection.)
IMPAIRED PROTECTION. Whenever it is likely that instrument protection has been impaired, the mainframe must be made inoperative and be secured against any unintended operation.
REMOVE POWER IF POSSIBLE. Some procedures in this manual may be performed with power supplied to the mainframe while protective covers are removed. Energy available at many points may, if contacted, result in personal injury. (If maintenance can be performed without power applied, the power should be removed.)
12 Keysight E1418A Service Guide
Safety Considerations General Information
USING AUTOTRANSFORMERS. If the mainframe is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to neutral (that is, the grounded side of the main’s supply).
CAPACITOR VOLTAGES. Capacitors inside the mainframe may remain charged even when the mainframe has been disconnected from its source of supply.
USE PROPER FUSES. For continued protection against fire hazard, replace the line fuses only with fuses of the same current rating and type (such as normal blow, time delay, etc.). Do not use repaired fuses or short-circuited fuseholders.
SHOCK HAZARD. Only service-trained personnel who are aware of the hazards involved should install, remove, or configure the module. Before you remove any installed module, disconnect AC power from the mainframe and from other modules that may be connected to the module.
CHANNEL WIRING INSULATION. All channels that have a common connection must be insulated so that the user is protected from electrical shock. This means wiring for all channels must be insulated as though each channel carries the voltage of the highest voltage channel.
Keysight E1418A Service Guide 13
General Information

Module Description

MAXIMUM VOLTAGE. The maximum voltage that can be applied to any channel is 42 V Peak/42 Vdc. The maximum voltage from any channel to ground is 42 V Peak/42 Vdc.
STATIC ELECTRICITY. Static electricity is a major cause of component failure. To prevent damage to the electrical components in the module, observe anti-static techniques whenever working on the device.
PARTS REMOVAL. This module uses a surface mount printed circuit assembly. Special soldering equipment is required for parts removal or replacement. Use of standard soldering equipment will cause damage to the printed circuit board and is not covered under warranty.
The Keysight E1418A is an 8 or 16 channel digital-to-analog converter module for use in a VXIbus C-size mainframe. The module is a register based device. The module can be programmed via direct register access or, with the appropriate driver, by high level commands. This manual describes programming the module using SCPI (Standard Commands for Programmable Instruments) and the SCPI driver.
Each Keysight E1418A module is a unique instrument having its own output buffer and error queue. Multiple modules can not be combined into a
single instrument.
Each channel can be configured to either voltage or current output mode. When configured for voltage output, voltages in the range of -16.0 V to
+16.0 V can be set. When configured for current output, current in the range of
-0.02 to + 0.02 Amps can be set. The channel output mode can be programmatically set, or, can be forced to either voltage or current by mechanical jumpers on the terminal block.
Each output channel is individually configurable to be either an isolated output or a non-isolated output. Channel configuration to isolated or non-isolated is made by individual plug-on modules for each channel.

Module Specifications

Specifications are listed in Appendix A of this manual and in the Keysight E1418A User’s Manual. These specifications are the performance standards or limits against which the module may be tested.
14 Keysight E1418A Service Guide
Module Description General Information

Module Serial Numbers

Devices covered by this manual are identified by a serial number prefix listed on the title page. Keysight Technologies uses a two-part serial number in the form USXXXXYYYYY, where US is the country of origin, XXXX is the serial prefix, and YYYYY is the serial suffix. The serial number prefix identifies a series of identical instruments. The serial number suffix is assigned sequentially to each instrument. The serial number plate is located on the right-hand shield near the backplane connectors.
Ordering Options
The module may be ordered from Keysight Technologies in a variety of configurations. As ordered, the module may have the following options:
Description
E1418A 8-Channel D/A Converter with Non-Isolated Outputs
E1418A Option 001
E1418A Option 002
E1418A Option 003
16-Channel D/A Converter with Non-Isolated Outputs
8-Channel D/A Converter with Isolated Outputs
16-Channel D/A Converter with Isolated Outputs
Field Kits
The module can also be user configured. The following field expansion and configuration kits are available. Each field kit contains installation instructions.
Description Use
E1523A Single Channel Isolated
plug-on module
E1524A Expansion kit, 8-Channel
Non-Isolated Outputs
E1525A Expansion kit, 8-Channel
Isolated Outputs
To change a single channel from non-isolated to isolated output.
To add 8 additional non-isolated channels to an
existing 8-channel module.
To add 8 additional isolated channels to an existing
8-channel module.
Keysight E1418A Service Guide 15
General Information Recommended Test Equipment
Terminal Modules
The standard Keysight E1418A Terminal Module provides screw terminals for connections. Two other terminal options are available with the Keysight E1418A:
• Crimp and Insert (Option A3E)
• Ribbon Cable (Option A3H)

Recommended Test Equipment

Table 1-1 lists the test equipment recommended for testing and servicing the module. Essential requirements for each piece of test equipment are described in the Requirements column.
Table 1-1 Recommended Test Equipment
Instrument Requirements Recommended
Model
Controller, GPIB GPIB compatibility as defined by IEEE Standard
488-1988 and the identical ANSI Standard MC1.1: SH1, AH1, T2, TE0, L2, LE0, SR0,
RL0, PP0, DC0, DT0, and C1, 2, 3, 4, 5.
Mainframe Compatible with the Keysight E1418A E1400B/T, E1401A/T, or
Command Module, VXILink, EPC7 Embedded Controller
Digital Multimeter 5 1/2 or 6 1/2 digit
Terminal Module Compatible with the Keysight E1418A E1418-60101 O, P,T
Resistor 10 kΩ , 1%, 0.25 W, metal film Any available supplier
Resistor 600 Ω , 1%, 0.25 W, metal film P
Compatible with Module Resource manager in Slot 0 Logical address 0 Typical GPIB address 9 Requires the downloaded driver “E1418”
Voltage measurements to ± 16.8 Vdc Current measurements to ± 0.02 A Four-wire resistance measurements from 50 Ω to 500 Ω
HP 9000 Series 300 F,O,P,T
E1401B
E1405B or E1406A/B F,O,P,T
Keysight 3458A or Keysight 34401A
meeting the requirements
Use
F,O,P,T
O, P, T
P
1
User’s Manual E1418-90001 F,O,P,T
1 * F = Functional Verification Tests, O = Operation Verification Tests, P = Performance Verification
Tests, T = Troubleshooting
16 Keysight E1418A Service Guide
Keysight E1418A 8/16 Channel D/A Converter Module Service Guide

2 Installation

Introduction

This chapter contains the following topics:

Initial Inspection

Initial Inspection page 11 Preparation for Use page 12 Environment page 14 Shipping Guidelines page 14
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, keep the container until the shipment contents have been checked and the module has been checked mechanically and electrically. Chapter 4 of this manual gives procedures to check the electrical performance.
To avoid possible hazardous electrical shock, do not perform electrical tests if there are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).
If the contents are incomplete, if there is mechanical damage or defect, or if the module does not pass the electrical performance tests, notify your nearest Keysight Technologies Sales and Service Office. If the shipping container is damaged or the cushioning material shows signs of stress, notify the carrier as well as Keysight Technologies. Keep the shipping materials for the carrier’s inspection.
11
Installation Preparation for Use

Preparation for Use

Chapter 1 of the Keysight E1418A User’s Manual contains complete instructions for installing the module in a mainframe and installing a terminal module.
Refer to the mainframe user’s manual, command module user’s manual, and the Installing Device Drivers installation note to prepare the mainframe.

Prepare the Module

Verify or set the module’s logical address. Refer to the Keysight E1418A User’s Manual for instructions. In this manual the factory default logical address of 72 is used.
To prepare the module for the adjustment procedures in Chapter 5, set or verify the position of the CAL Secure Enable Jumper before installing the module in the mainframe. (The jumper is factory set to CAL.) Figure 5-2 shows this jumper.

Prepare the Terminal Module

The procedures in this manual show the screw-type terminal module. To prepare this terminal module, you will need to set the jumpers and make connections. If you are using other types of terminal modules, consult the Keysight E1418A User’s Manual for proper connections.
Terminal Module Jumpers
The screw-type terminal module contains jumpers that control how the Keysight E1418A operates. To prepare the module for the Verification Tests in Chapter 4 and Adjustments in Chapter 5, the terminal module jumpers must be set to allow the module to be programmed from the backplane. The Keysight E1418A will operate without any jumpers installed. Figure 2-1 shows the jumper locations and example settings.
There are two jumpers for each channel; a P/J jumper and a V/I jumper. The two jumpers work together.
The P/J jumper sets the channel output mode to be either Program Selectable or Jumper Selectable. When the jumper is in the program selectable position, the channel output mode can be set via programming commands. The program selectable position is recommended for the Verification Tests and Adjustment procedures in this manual. In the jumper selectable position, the channel output mode is set via the V/I jumper.
The V/I jumper sets the channel output mode to either voltage or current when the P/J jumper is set to the Jumper Selectable position. When the P/J jumper is set to Program Selectable, this jumper has no effect.
12 Keysight E1418A Service Guide
Preparation for Use Installation
With no jumpers installed, the module will be in the Program Selectable output mode.
– If you are using a terminal module dedicated to testing the Keysight
E1418A, you can remove all the jumpers or put the jumpers in the storage
position.
– If you are using the end-user’s terminal module, record the positions of the
jumpers. Set the P/J jumpers to the programmable position. Before
returning the module to use, restore the P/J jumpers to their original
position.
Figure 2-1 P/J and V/I Jumpers
Terminal Module Connections
The Verification Tests require four-wire connections at each channel. The Adjustment procedures can be performed using a single connections at the CAL terminals. Additional connection information is given in the procedures in Chapters 4 and 5 of this manual.
Keysight E1418A Service Guide 13
Installation Environment

Prepare the Command Module

All verification and adjustment procedures in this manual use SCPI (Standard Commands for Programmable Instruments). Prepare the command module by downloading the SCPI driver named “E1418". Downloading instructions are given in the Installing Device Drivers installation note.

Environment

The recommended operating altitude for the module is 2000 m (6000 ft) or less. The recommended operating environment for the module is:
Environment Temperature Humidity
o
Operating 0
Storage and Shipment
C to + 55 oC <65% relative (0 oC to + 40 oC)
non condensing
o
C to + 75 oC <65% relative (0 oC to + 40 oC)
– 40
non condensing

Shipping Guidelines

If you need to return the Keysight E1418A to Keysight Technologies, first remove any terminal modules before packaging for shipment. Attach a tag to the module identifying the owner and indicating the service or repair required. Mark the shipping container “FRAGILE”. If you are exchanging a module, read Assembly Exchange.
In any correspondence, refer to the module by model number and full serial number.
Keysight Technologies recommends using shipping containers and materials identical to those used by the factory. These materials are available through Keysight Technologies Sales and Service Offices.
If you use commercially available shipping materials, place the module in an anti-static bag and wrap the module in heavy paper or plastic. Use a strong shipping container. A double wall carton of 2400 Pa (350 psi) test material is adequate.
Use enough shock-absorbing material (layer at least 75 mm to 110 mm, 3 in to 4 in) around the module to provide firm cushion and to prevent movement in the container. Protect the front panel with cardboard. Seal the container securely and mark the container “FRAGILE.”
14 Keysight E1418A Service Guide

3 Operating Information

Introduction

Operation

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This chapter provides operating information, preventive maintenance instructions, and operators checks.
The Keysight E1418A User’s Guide is your reference for module operation. The User’s Guide contains information about module setup, application examples, and a complete SCPI command reference.

Querying Module Identification and Configuration

This example will query the module for the identification string and determine the module configuration.
The module identification is obtained using the IEEE-488 Command *IDN?.
/* Send query to the module*/
*IDN?
/* Enter a string*/
The string returned will be similar to one of the following:
HEWLETT-PACKARD,E1418A_8CH,xxxxxxxxxx,A.01.00
or
HEWLETT-PACKARD,E1418A_16CH,xxxxxxxxxx,A.01.00
The xxxxxxxxxx portion of the response string is the module serial number. The number of channels in the module (8 or 16) is indicated by the E1418A_8CH or E1418A_16CH portion of the returned string.
The module configuration is obtained using the DIAGnostic:CONFiguration? query.
/* Send query to the module*/
DIAG:CONF?
/* Enter six integers and decode the integers*/
The DIAG:CONF? query returns six integers. The six integers returned contain the module configuration and have the following meaning.
15
Operating Information Operation
Integer
Meaning
Returned
1st Expansion Board ID in the form: 0 = present, 7 = none
(expansion board contains channels 9 through 16)
2nd Terminal Module ID in the form:
0 = screw type, 7 = none or other
3rd Isolated/Non-isolated Channel status
A bit set to 0 indicates an isolated channel A bit set to 1 indicates a non-isolated channel or no plug-on module installed (DO NOT operate the module without a plug-on module) Bits 0 – 15 correspond to channels 1 – 16, respectively*
4th Channel Mode
A bit set to 0 indicates a current output channel A bit set to 1 indicates a voltage output channel Bits 0 – 15 correspond to channels 1 – 16, respectively*
5th Channel Output State
A bit set to 0 indicates the channel output relay is closed A bit set to 1 indicates the channel output relay is open Bits 0 – 15 correspond to channels 1 – 16, respectively*
6th Channel Mode Programmable State (P/J Jumper)
A bit set to 0 indicates a channel is not mode programmable A bit set to 1 indicates a channel is mode programmable Bits 0 – 15 correspond to channels 1 – 16, respectively
1
1 For 8-channel configurations, the upper 8 bits of integers 3, 4, 5,
and 6 are set to 1’s.
16 Keysight E1418A Service Guide
Operation Operating Information
For example, if the following six integers are returned, the module has the configuration indicated.
Integer Decimal
Configuration
Value
1st 7 No expansion board installed
(Only channels 1 through 8)
2nd 7 No Terminal Module installed or unknown
terminal module installed
3rd -1 All channels are non-isolated
4th -1 All channels are voltage output channels
5th -1 All channel outputs are disabled
(all output relays are open)
6th -1 All channels are output mode programmable
Configuration Example
The following program segment demonstrates how to read the module identification string and the configuration. The configuration integers are bit manipulated using the C operator for bit shifting
result = result <<1
(a one bit shift to the left). The code shown in this example can be obtained from the file prftest.c on the examples disk provided with this manual.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x);/* VTL error routine*/
void sys_err(ViSession resource);/* Checks for SCPI programming errors */
/** GLOBAL **/ ViStatus err;
ViSession defaultRM, cmd, dac, dmm;
int num_chan;
void main (void)
{
int i,result = {0},config [6]={0},num_chan = {0};
err=viPrintf(dac, “DIAG:CONF?\n”);/* request module configuration */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%,6d”, &config);/* returns six integers*/
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check*/
Keysight E1418A Service Guide 17
Operating Information Operation
/* First integer */
result=config[0];/* Expansion board*/
if (result == 0)
{
}
else
{
printf (“Module is a 16 channel device\n”);
num_chan = 16;
if (result == 7)
{
}
else
{
}
}
printf (“Module is an 8 channel device\n”);
num_chan=8;
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Second integer */
result = config [1];/* Terminal Module */
if (result == 0)
{
}
else
{
printf (“Module has a screw type terminal module installed\n”);
if (result == 7)
{
18 Keysight E1418A Service Guide
Operation Operating Information
}
else
{
}
}
printf (“Module does NOT have a terminal module installed\n”);
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Third integer */
result=config[2];/* Isolated or non-isolated outputs */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d is configured for non-isolated output\n”,i);
printf(“Channel %d is configured for isolated output\n”,i);
result = result << 1;
}
/* Fourth integer */
result=config[3];/* Output Mode Voltage or Current */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
Keysight E1418A Service Guide 19
Operating Information Operation
}
printf (“Channel %d is set to voltage output mode\n”,i);
printf(“Channel %d is set to current output mode\n”,i);
result = result << 1;
}
/* Fifth integer */
result=config[4];/* Channel output ON or OFF */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d output is disabled\n”,i);
printf(“Channel %d output is enabled\n”,i);
result = result << 1;
}
/* Sixth integer */
result=config[5];/* Program or Jumper Selectable channels */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result >= 0x8000)
{
}
else
{
}
printf (“Channel %d output mode is programmable\n”,i);
printf(“Channel %d output mode is fixed and jumper selected\n”,i);
result = result << 1;
}
}
20 Keysight E1418A Service Guide
Preventive Maintenance Operating Information

Preventive Maintenance

Preventive maintenance consists of periodically cleaning the front panel, the terminal module, and performing the Operator’s Check. Clean the module and terminal module yearly, or more often if the module is used in very dusty or humid areas. Disassembly of the module is not recommended for cleaning. However, if the module is disassembled for repair or reconfiguration, the printed circuit board assemblies (PCAs) can be blown off with a properly grounded air gun. Remove the terminal module cover to blow dust off the terminal printed circuit board and case. Table 3-1 shows the recommended cleaning equipment and supplies.
Table 3-1 Recommended Cleaning Equipment and Supplies
Description Recommended Use
Mild Soap Solution Clean face plate panel and terminal covers.
Lint-free Cloth Clean face plate panel and terminal covers.
Airgun with grounded nozzle
To eliminate possible electrical shock, disconnect ac power from the mainframe and disconnect all outputs to the terminal module before removing the module from the mainframe.
The printed circuit assemblies (PCAs) contain static-sensitive devices that can be damaged by careless handling. Use static-control devices (wrist straps, mats, and tools) when handling the module. Chapter 6 contains additional precautions for electrostatic discharge (ESD).
Remove dust from the printed circuit boards and terminal case.
Keysight E1418A Service Guide 21
Operating Information Preventive Maintenance

Cleaning Procedure

Use the following procedure to clean the Keysight E1418A.
Terminal Module
1 Remove the terminal module from the main module.
2 Disconnect all external wiring.
3 Clean the terminal module covers with a mild soap solution and lint-free
cloth.
4 Remove the terminal module cover and blow any dust or debris from the case.
5 Reassemble the terminal module.
6 Reconnect the external wiring.
Main Module
1 Remove the terminal module from the main module.
2 Clean the main module face plate with a mild soap solution and lint-free
cloth.
Disassembling the module exposes the assemblies to ESD damage and is usually unnecessary. Perform steps 3 through 5 only if the module is disassembled for repairs or reconfiguration.
3 Disassemble the module following the instructions in Chapter 6.
4 Using an airgun with a grounded nozzle, remove dust from the PCA
assemblies.
5 Reassemble the module.
22 Keysight E1418A Service Guide
Operator’s Checks Operating Information

Operator’s Checks

The operator’s check for the Keysight E1418A has two parts.
The first part consists of sending the SYStem:ERRor? query to the
module following power-on and entering the response. This check should be performed after every power-on sequence. At power-on, the Keysight E1418A performs a brief self-test that ensures the module can communicate with the backplane. A successful self-test returns an integer and a string:
+0,"No Error"
Any other response indicates a failure in the module. Chapter 6 contains troubleshooting information.
Once the power-on self-test has been checked, the operator may perform a full self-test by sending the *TST? query and entering the response. This query executes a full self-test that checks all circuits of the module except the output relays. A successful self-test returns:
+0
Any other response indicates a failure. Chapter 6 contains additional information about the self-test error codes and troubleshooting information.

Operator Check Example

This example checks the module for errors after power-on and then performs a full self-test.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x);/* VTL error routine */
void sys_err(ViSession resource);/* SCPI error routine */
ViStatus err;/* Global VTL error variable */
void main (void)
{
int selftest;
char selftestresults [80];
err=viPrintf(dac, “*TST?\n”);/* run self test */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d%t”, &selftest,&selftestresults);/* enter integer and string returned */
if(err VI_SUCCESS) err_handler(dac, err);
Keysight E1418A Service Guide 23
Operating Information Operator’s Checks
if (selftest != ’0’)
{
}
else
{
}
}
printf (“Self Test failed\n”);
printf (“The first error encountered was\n”);
printf (“Error number %d %s\n”,selftest,selftestresults);
printf (“This program will terminate\n”);
pause();
abort;
printf (“Self Test PASSED”);
/*** VTL Error handling function ***/
void err_handler (ViSession dac, ViStatus err)
{
char buf[1024]={0}; viStatusDesc(dac,err,buf); printf(“VTL ERROR = %s\n”, buf); return;
} /*End of VTL error handler */
/*** SCPI error checking and reporting function ***/
void sys_err(ViSession resource) /* Test for SCPI Errors */
{
char buf [1024] = {0};
int err_no;
err = viPrintf (resource, “SYST:ERR?\n”); /* Check for an error */
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
err = viScanf (resource,"%d%t",&err_no,&buf);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
while (err_no != 0) /* Report all errors in error que */
{
printf (“\nCommand Error: %d,%s\n”,err_no,buf);
err = viPrintf (resource, “SYST:ERR?\n”);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
err = viScanf (resource,"%d%t",&err_no,&buf);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
}
flushall();/* Clean out the buffers */
24 Keysight E1418A Service Guide
Operator’s Checks Operating Information
err = viFlush(resource,VI_READ_BUF);
if (err < VI_SUCCESS) err_handler (resource,err); /* Check VTL errors */
err = viFlush(resource,VI_WRITE_BUF);
if (err < VI_SUCCESS) err_handler (resource,err); /* Check VTL errors */
} /* End of checking for SCPI errors */
Keysight E1418A Service Guide 25
Operating Information Operator’s Checks
26 Keysight E1418A Service Guide

4 Verification Tests

Introduction

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
Three test procedures are given in this chapter. These test procedures are used to verify that:
the Keysight E1418A is functional (Functional Verification)
the Keysight E1418A meets selected testable specifications (Quick
Verification)
– the Keysight E1418A meets all testable specifications (Performance
Verification)

Test Conditions

Table 1-1 lists the recommended test equipment. When performing the test procedures, observe the following test conditions:
The ambient temperature should be between 18 °C and 28 °C. The
The relative humidity should be < 65%, non-condensing.
Install the module, apply power, and allow the module to warm up for at
Keysight Technologies recommends performance tests be performed at one year intervals. In severe operating environments, or after heavy use, perform the tests more often.
Do not perform any of the following verification tests unless you are a qualified, service-trained technician and have read the WARNINGS and CAUTIONS in Chapter 1 of this manual.
temperature should be stable within ± 1 °C.
least 15 minutes.
27
Verification Tests About the Verification Tests

About the Verification Tests

Three levels of verification tests are described in this chapter:
Functional Verification Tests
Quick Verification Tests
Performance Verification Tests
You should perform the Functional Verification test before either the Quick Verification or full Performance Verification tests.

Functional Verification Test

The Functional Verification test provides a high confidence that the module is operational. A brief self-test is performed when power is first applied to the module. This test assures that module can communicate with the mainframe.
The Functional Verification Test performs a more complete self-test using the SCPI common command *TST?. If this test passes, you have a high confidence level (90%) that the module is operational.
If the module fails the Functional Verification Test, repair is needed.

Quick Verification Tests

The Quick Verification Test combines a Functional Verification Test with an abbreviated set of Performance Tests to give a high confidence level that the module is operational and meets its specifications. The Quick Performance Test is the minimum set of tests recommended after any service activity.
These tests check the module’s performance for normal accuracy and drift mechanisms. These tests do not check for abnormal component failures.
Quick Verification Tests are designated with the letter “Q” in the Performance Verification Tests. To perform the Quick Verification Tests:
1 1. Perform a Functional Verification Test.
2 2. Perform the Performance Verification Tests designated by the letter“Q” in
the table.
If the module fails the Quick Verification Test, adjustment or repair is needed.
28 Keysight E1418A Service Guide
About the Verification Tests Verification Tests

Performance Verification Tests

Performance Verification Tests give a high confidence that the module is operational and meets its specifications. The Performance Verification Tests can be used as acceptance tests when the module is first received.
The Performance Verification Tests should be run at the calibration interval (Keysight Technologies recommends a one year interval). Run the tests to characterize the module against the specifications. Run the Performance Verification Tests after any adjustment to verify the adjustment.
If the module fails any Performance Verification Tests, adjustment or repair is needed.

Performance Test Record

Table 4-1 provides space to enter the results of each Performance Verification test. The table also lists the upper and lower test limits.

About Program Examples

In this manual test procedures are shown with a portion of an example program that performs the test. These examples are in ANSI C format and complete program files are included on the disk supplied with this manual. Most examples in this chapter are included. The examples are ASCII files with the *.c extension.
The int data type is system dependent. These examples were developed on a system where int is a 16-bit integer. Other systems may define int to be a different width.
In the examples, the Keysight VISA Transition Library is used for I/O operations with the VXIbus. An Keysight command module (Keysight E1405/E1406) is used and controlled via GPIB.
To use the Keysight VISA Transition Library (abbreviated as VTL), include the visa.h header file.
#include visa.h
Keysight VTL function calls and data types typically begin with the lower case letters vi. Output and enter are performed with functions named viPrintf and viScanf. Both these functions require a session (a VTL defined I/O function) to uniquely identify the device being controlled. In the examples, the session has been named dac.
Keysight E1418A Service Guide 29
Verification Tests About the Verification Tests

SCPI drivers

In this manual, all programming examples and procedures assume the use of the SCPI driver and Keysight E1406A Command Module. The Keysight E1418A module is shipped with two 3.5" disks. These disks contain the SCPI driver for the Keysight E1406A Command Module. One disk is in LIF format and one disk is in DOS format. Installing Device Drivers Installation Note is included with the disks. Follow the instructions contained in the installation note to properly install the device driver.
Use the DIAG:DRIV:LIST? query on the command module to verify that the correct device driver is installed. Responses to this query vary depending upon the drivers loaded on your system. A typical response contains a list of all drivers installed and might look like:
E1418,E1418,A.01.00,RAM;SWITCH,SWITCHBOX,A.08.00,RAM; SYSTEM,E1405A,A.08.00,ROM;IBASIC,IBASIC,A.04.02,ROM;VOLT MTR,E1326A,A.05.00,ROM;SWITCH,SWITCHBOX,A.07.00,ROM;COUN TER,E1332A,A.04.02,ROM;COUNTER,E1333A,A.04.02,ROM; DIG_I/O,E1330A,A.04.03,ROM;D/A,E1328A,A.04.02,ROM
The string E1418,E1418,A.01.00, should be located somewhere within the returned string.

SCPI Command Reference

The SCPI (Standard Commands for Programmable Instruments) commands are not documented in this manual except in a general manner. Complete SCPI commands, including syntax and parameters, applicable to the module are documented in the Keysight E1418A User’s Guide.
30 Keysight E1418A Service Guide
Functional Verification Test Verification Tests

Functional Verification Test

The Functional Verification Test quickly verifies that the module is operational. Perform this test any time to verify or check the operation. Perform this test before beginning any Quick Verification or Performance Verification Tests.

Functional Verification Test Procedure

Use the following procedure to verify the module’s operation.
1 Install the module in the mainframe and apply power.
2 Verify the module passed its power-on self-test by executing the
SYStem:ERRor? query and entering the results.
A passed self-test is indicated by the return
+0, “No Error”
any other return value indicates a power-on failure.
If an incorrect module address is used, the module will not respond. Verify the module’s address before troubleshooting.
3 Execute the full self-test by sending the *TST? query and entering the result.
A passed self-test is indicated by the return
+0
Any other return value indicates a failure. Additional information about the
self-test results and self-test error codes is given in Chapter 6

Functional Verification Test Example

An example of the Functional Verification Test is provided in Operator Check Example..
Keysight E1418A Service Guide 31
Verification Tests Performance Verification Tests

Performance Verification Tests

These procedures test the electrical performance of the Keysight E1418A using the specification in Appendix A as the performance standards. These tests are suitable for incoming inspections, troubleshooting, and preventive measures.

DC Voltage Output Verification

This procedure tests the dc voltage output performance of the module.
1 For each channel in the module, connect the 10 kΩ resistor and DMM as
shown in Figure 4-1. Channel 1 connections are shown in the figure. Connect the other channels in a similar manner. Set the DMM to measure dc volts.
Figure 4-1 DC Voltage Output Connections (Channel 1)
2 Set the channel to output each of the following voltages:
+16.00 – 4.00 +12.00 – 8.00 + 8.00 –12.00 + 4.00 –16.00
0.00
32 Keysight E1418A Service Guide
Performance Verification Tests Verification Tests
Use the APPLyn:VOLTage xx.xx command, where n is the channel number
and xx.xx is the voltage to output.
3 For each voltage, measure the output voltage with the DMM, enter it in the
table and compare the voltage output to the limits shown in the table.
4 Open the channel output relay by sending the OUTPutn OFF command, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.

DC Voltage Compliance Current

This test checks the compliance current in dc voltage.
1 For each channel in the module, connect the 600 Ω resistor and the DMM as
shown in Figure 4-2. Channel 1 connections are shown in the figure. Connect the other channels in a similar manner. Set the DMM to measure dc voltage.
Figure 4-2 DC Voltage Output Connections (Channel 1)
2 Set the channel to output +12.00 V. Use the APPLyn:VOLTage 12.00
command, where n is the channel number.
Keysight E1418A Service Guide 33
Verification Tests Performance Verification Tests
3 Measure the output voltage with the DMM, enter it in the table and compare
the measured voltage to the limits shown in the table. The compliance current is inferred from the voltage across the resistor.
4 Open the channel output relay by sending the OUTPutn OFF command, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.

DC Voltage Short Circuit Output Current

This test checks the short circuit output current for each channel.
1 For each channel in the module, connect the DMM as shown in Figure 4-3.
Channel 1 connections are shown in the figure. Connect the other channels in a similar manner. Set the DMM to measure dc current (the DMM should have a current shunt of 100 Ω or less for this test).
Figure 4-3 DC Voltage Output Connections (Channel 1)
2 Set the channel to output +16.00 V. Use the APPLyn:VOLTage 16.00
command, where n is the channel number.
34 Keysight E1418A Service Guide
Performance Verification Tests Verification Tests
3 Measure the output current with the DMM, enter it in the table and compare
the measured current to the limits shown in the table.
4 Open the channel output relay by sending the OUTPutn OFFcommand, where
n is the channel number.
5 Change the connections to the next channel and repeat steps 2, 3, and 4.
6 Repeat steps 2, 3, 4, and 5 for each channel.

DC Current Verification

This test checks the dc current output performance of the module.
1 1. For each channel in the module, connect the DMM as shown in Figure 4-4.
Channel 1 connections are shown in the figure. Connect the other channels in a similar manner. Set the DMM to measure dc current.
Figure 4-4 DC Current Output Connections (Channel 1)
2 2. Set the channel to output each of the following currents:
+0.020 –0.005
+0.015 –0.010
Keysight E1418A Service Guide 35
Verification Tests Performance Verification Tests
+0.010 –0.015
+0.005 –0.020
0.000
Use the APPLyn:CURRent x.xxx command, where n is the channel number and x.xxx is the current to output.
3 3. For each current, measure the output current with the DMM, enter it in the
table and compare the current output to the limits shown in the table.
4 4. Open the channel output relay by sending the OUTPutn OFF command,
where n is the channel number.
5 5. Change the connections to the next channel and repeat steps 2, 3, and 4.
6 6. Repeat steps 2, 3, 4, and 5 for each channel.
36 Keysight E1418A Service Guide
Performance Verification Tests Verification Tests

Performance Verification Tests Example

This example demonstrates a full performance verification test of the module. The code shown in this example can be obtained from the file prftest.c on the examples disk provided with this manual.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x); /* VTL error routine */
void sys_err(ViSession resource);/* Checks for SCPI programming errors */
void pause();/* Waits for a keystroke to continue program execution */
void dcv_check ();/* Voltage output verification test */
void compliance_check ();/* Compliance current verification test */
void maxcurrent_check ();/* Short circuit output current verification test */
void current_check ();/* Current output verification test */
/* DMM routines*/
void dmm_setup (char function [6], char range [6]); /* Set function and range for DMM */
float dmm_measure ();/* get the dmm measurement */
/** GLOBAL **/ ViStatus err;
ViSession defaultRM, cmd, dac, dmm;
int num_chan;/* 8 or 16 */
float dcv_results [16][9]={0};/* Voltage output verification test results */
float compliance_results [16] = {0}; /* Compliance current verification test results */
float maxcurrent_results [16] = {0};/* Short circuit output current verification test results */
float current_results [16][9] = {0};/* Current output verification test results */
void main (void)
{
dcv_check (); compliance_check (); maxcurrent_check (); current_check ();
}
/*** Routine to perform the dcv performance test ***/
void dcv_check ()
{
float v_out [9] = {16, 12, 8, 4, 0, -4,-8, -12, -16}; /* Voltages to output during test */
int i,j,opc_check;
char function [6] = {“VOLT:DC\0"},range [6] = {”100\0"};
dmm_setup (function,range);
printf (“Voltage Output Check on %d channels”,num_chan);
printf (“\n\n”);
Keysight E1418A Service Guide 37
Verification Tests Performance Verification Tests
printf (“DC VOLTAGE OUTPUT PERFORMANCE TEST\n\n”);
for (i = 0;i != num_chan;i++)
{
/** connection instructions **/
printf (“\tChannel %d test\n”,i+1);
printf (“1.\tMake a four-wire connection from channel %d \n”,i+1);
printf (“\tto a 10,000 Ohm resistor.\n”);
printf (“2.\tMake a two-wire connection from the DMM to\n”);
printf (“\tthe 10,000 Ohm resistor.\n”);
printf (“3.\tThe DMM is set to make DC Volts Measurements on \n”);
printf (“\tthe 100 V range.\n”);
printf (“4.\tFor each channel in the E1418A,\n”);
printf (“\t\tOutput 16 V, 12 V, 8 V, 4 V, 0 V, -4 V, -8 V, -12 V, -16 V.\n”);
printf (“\t\tMeasure each voltage output on the DMM.\n”);
printf (“\n\n”);
pause();
/** voltage output loop **/
for (j=0;j<9;j++)
{
err=viPrintf(dac, “APPLY%d:VOLTAGE %f;*OPC?\n”,i+1,v_out[j]); /* voltage output*/
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check); /* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
dcv_results [i][j] = dmm_measure();
}
printf (“\nChannel %d Voltage Output Test Done\n\n”,i+1); err=viPrintf(dac, “OUTPUT%d OFF;*OPC?\n”,i+1); /* Turn output off*/ if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check */
}
printf (“\n\nEnd of DC Voltage Output Test\n\n”);
err=viPrintf(dac, “*RST\n”);
if(err VI_SUCCESS) err_handler(dac, err);
}
/*** End of dcv_check ***/
/*** Routine to perform the compliance current performance test ***/
void compliance_check ()
{
float v_out = {12};/* Voltage to output */
38 Keysight E1418A Service Guide
Performance Verification Tests Verification Tests
int i,opc_check;
char function [6] = {“VOLT:DC\0"},range [6] = {”100\0"};
dmm_setup (function,range);
printf (“\n\nCompliance Current Verification Test\n\n”);
for (i = 0;i != num_chan;i++)
{
/** connection instructions **/
printf (“\tChannel %d\n”,i+1);
printf (“1.\tMake a four-wire connection from channel %d \n”,i+1);
printf (“\tto a 600 Ohm resistor.\n”);
printf (“2.\tMake a two-wire connection from the DMM to\n”);
printf (“\tthe 600 Ohm resistor.\n”);
printf (“3.\tThe DMM is set to make DC Volts Measurements on \n”);
printf (“\tthe 100 V range.\n”);
printf (“4.\tEach channel is set to output 12 V\n”);
printf (“\t\tMeasure the voltage drop across the resistor output.\n”);
printf (“\n\n”);
pause();
err=viPrintf(dac, “APPLY%d:VOLTAGE %f;*OPC?\n”,i+1,v_out); /* voltage output */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check); /* enter OPC integer */ if(err VI_SUCCESS) err_handler(dac, err); compliance_results [i] = dmm_measure();
err=viPrintf(dac, “OUTPUT%d OFF;*OPC?\n”,i+1); /* Turn output off*/
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check */
}
printf (“\n\nend of Compliance Current Test\n\n”);
err=viPrintf(dac, “*RST\n”);/* reset the dac*/
if(err VI_SUCCESS) err_handler(dac, err);
}
/*** End of compliance_check ***/
/*** Routine to perform the short circuit output current performance test ***/
void maxcurrent_check ()
{
float v_out = {16}; /*voltage to output*/
int i,opc_check;
char function [6] = {“CURR:DC\0"},range [6] = {”0.100\0"};
dmm_setup (function,range);
Keysight E1418A Service Guide 39
Verification Tests Performance Verification Tests
printf (“\n\nShort Circuit Output Current Verification Test\n\n”);
for (i = 0;i != num_chan;i++)
{
/** connection instructions **/
printf (“\tChannel %d\n”,i+1);
printf (“1.\tMake a four-wire connection from channel %d \n”,i+1);
printf (“\tto the DMM current input terminals\n”);
printf (“2.\tThe DMM is set to make DC current Measurements on \n”);
printf (“\tthe 100 mA range.\n”);
printf (“3.\tEach channel is set to output 16 V\n”);
printf (“\t\tMeasure the current flowing through the DMM.\n”);
printf (“\n\n”);
pause();
err=viPrintf(dac, “APPLY%d:VOLTAGE %f;*OPC?\n”,i+1,v_out); /* voltage output */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
maxcurrent_results [i] = dmm_measure();
err=viPrintf(dac, “OUTPUT%d OFF;*OPC?\n”,i+1); /* Turn output off*/
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check */
}
printf (“End of Short Circuit Output Current Test”);
err=viPrintf(dac, “*RST\n”);
if(err VI_SUCCESS) err_handler(dac, err);
}
/*** End of maxcurrent_check ***/
/*** Routine to perform the current performance test ***/
void current_check ()
{
float i_out [9] = {0.02F, 0.015F, 0.01F,0.005F, 0, -0.005F,-0.01F, -0.015F,
-0.02F};
/* Currents to output */
int i,j,opc_check; */
char function [6] = {“CURR:DC\0"},range [6] = {”0.100\0"};
dmm_setup (function,range);
printf (“\n\n”);
printf (“DC CURRENT OUTPUT PERFORMANCE TEST\n\n\n\n”);
40 Keysight E1418A Service Guide
Performance Verification Tests Verification Tests
for (i = 0;i != num_chan;i++)
{
/** connection instructions **/
printf (“Channel %d\n”,i+1);
printf (“1.\tMake a two-wire connection from channel %d \n”,i+1);
printf (“\tto the DMM current input terminals.\n”);
printf (“2.\tThe DMM is set to make DC current measurements on \n”);
printf (“\tthe 100 mA range.\n”);
printf (“3.\tFor each channel in the E1418A,\n”);
printf (“\t\tOutput 20 mA, 15 mA, 10 mA, 5 mA, 0 mA, -5 mA, -10 mA, -15 mA,
-20 mA.\n”);
printf (“\t\tMeasure each current output.\n”);
printf (“\n\n”);
pause();
/** current output loop **/
for (j=0;j<9 ;j++)
{
err=viPrintf(dac, “APPLY%d:CURRENT %f;*OPC?\n”,i+1,i_out[j]); /* voltage output */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
current_results [i][j] = dmm_measure();
}
printf (“Current Output Test on channel %d complete\n\n”,i+1); err=viPrintf(dac, “OUTPUT%d OFF;*OPC?\n”,i+1); /* Turn output off*/ if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &opc_check);/* enter OPC integer */
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check */
}
printf (“Current Output Verification Test completed”); err=viPrintf(dac, “*RST\n”);/* reset the dac*/ if(err VI_SUCCESS) err_handler(dac, err);
}
/*** End of current_check ***/
Keysight E1418A Service Guide 41
Verification Tests Performance Test Record

Performance Test Record

Table 4-1 is a form you can copy and use to record the performance verification test results of the Keysight E1418A. This table shows the output accuracy, measurement uncertainty, and the test accuracy ratio (TAR) values. The test marked with a “Q” in the table are the Quick Verification Tests.

Test Limits

The test limits are defined using the 90-day specifications. Appendix A lists the Keysight E1418A specifications.

Measurement Uncertainty

Measurement uncertainty has been calculated using the Keysight 34401A 90-day specifications. The Keysight 34401A is set to the 61⁄2 digit mode for all calculations. In both the voltage and current measurement mode, the measurement uncertainty is given as:
M.U. = % of reading + % of range

Voltage Output Measurements

The 100 V range is used for the voltage output measurements. The specification is given as:
M.U. = ± (0.0035% of reading + 0.0006% of range)
Keysight E1418A Output
+ 16.00 V 0.00056 V 0.0006 V ± 0.00116 V
+ 12.00 V 0.00042 V 0.0006 V ± 0.00102 V
+ 8.00 V 0.00028 V 0.0006 V ± 0.00088 V
+ 4.00 V 0.00014 V 0.0006 V ± 0.00074 V
0.00 V 0.00000 V 0.0006 V ± 0.00060 V
– 4.00 V 0.00014 V 0.0006 V ± 0.00074 V
– 8.00 V 0.00028 V 0.0006 V ± 0.00088 V
– 12.00 V 0.00042 V 0.0006 V ± 0.00102 V
% of reading % of range Measurement
Uncertainty
– 16.00 V 0.00056 V 0.0006 V ± 0.00116 V
42 Keysight E1418A Service Guide
Performance Test Record Verification Tests

DC Compliance Current Measurements

The 100 V range is used for the compliance current test. (The compliance current is inferred from the voltage developed across the resistor).
M.U. = ± (0.035% of reading + 0.0006% of range)
M.U. = ± (0.00042 + 0.0006) = 0.00102 V

DC Short Circuit Current Measurements

The 100 mA range is used for the short circuit current measurement.
M.U. = ± (0.030% of reading + 0.005% of range)
M.U. = ± (0.000018 + 0.000005) = 0.000023 A
Current Output Measurements
The 100 mA range is used for the current output measurements.
M.U. = ± (0.030% of reading + 0.005% of range)
E1418A Output % of reading % of range Measurement
+ 0.020 A 0.000006 A 0.000005 A ± 0.000011 A
+ 0.015 A 0.0000045 A 0.000005 A ± 0.0000095 A
+ 0.010 A 0.000003 A 0.000005 A ± 0.000008 A
+ 0.005 A 0.0000015 A 0.000005 A ± 0.0000065 A
0.000 A 0.00000 A 0.000005 A ± 0.000005 A
– 0.005 A 0.0000015 A 0.000005 A ± 0.0000065 A
– 0.010 A 0.000003 A 0.000005 A ± 0.000008 A
– 0.015 A 0.0000045 A 0.000005 A ± 0.0000095 A
– 0.020 A 0.000006 A 0.000005 A ± 0.000011 A

Test Accuracy Ratio (TAR)

The Test Accuracy Ratio (TAR) is defined as:
TAR
MaximumAllowe d Input
----------------------------------------------------------------------------=
MeasurementUn certainty
Uncertainty
Where Maximum Allowed, Input, and Measurement Uncertainty are all in Vdc or Amps.
The TAR is listed in table 4-1.
Keysight E1418A Service Guide 43
Verification Tests Performance Test Record
Table 4-1 Performance Test Record
Test Facility:
Name Report No.
Address Date
City/State Customer
Phone Tested by
Model Keysight E1418AAmbient Temperature
Serial No. Relative Humidity
Options
Firmware Rev.
Notes:
Test Equipment Used Model Trace Number Cal Due Date
44 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
1
1
1
1
1
1
1
1
1
2
2
2
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
Min Measured Max M.U. TAR
Voltage Output Check
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
2
Q
Q
Q
Q
Q
2
2
2
2
2
3
3
3
3
3
3
3
3
3
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Keysight E1418A Service Guide 45
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
4
4
4
4
4
4
4
4
4
5
5
5
5
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
Min Measured Max M.U. TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
Q
Q
Q
Q
Q
5
5
5
5
5
6
6
6
6
6
6
6
6
6
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
46 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Q
Q
Q
Q
Q
Q
Channel E1418A
Output
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
Min Measured Max M.U. TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Q
Q
Q
9
9
9
9
9
9
9
9
9
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
Keysight E1418A Service Guide 47
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
10
10
10
10
10
10
10
10
10
11
11
11
11
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
Min Measured Max M.U. TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
Q
Q
Q
Q
Q
11
11
11
11
11
12
12
12
12
12
12
12
12
12
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
48 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
13
13
13
13
13
13
13
13
13
14
14
14
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
Min Measured Max M.U. TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
14
Q
Q
Q
Q
14
14
14
14
14
15
15
15
15
15
15
15
15
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
Q
15
-16.000 V
–16.011
–15.989 V
1.16E-3 V
9.5:1
Keysight E1418A Service Guide 49
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Channel E1418A
Output
16
16
16
16
16
16
16
16
16
1
2
3
4
+16.000 V
+12.000 V
+8.000 V
+4.000 V
0.000 V
-4.000 V
-8.000 V
-12.000 V
-16.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
Min Measured Max M.U. TAR
+15.989 V
+11.991 V
+7.993 V
+3.995 V
– 0.003 V
–4.005
–8.007
–12.009
–16.011
DC Voltage Compliance Current Check
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+16.011 V
+12.009 V
+8.007 V
+4.005 V
+0.003 V
–3.995 V
–7.993 V
–11.991 V
–15.989 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
1.16E-3 V
1.02E-3 V
8.80E-4 V
7.40E-4 V
6.00E-4 V
7.40E-4 V
8.80E-4 V
1.02E-3 V
1.16E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
9.5:1
8.8:1
7.9:1
6.7:1
5:1
6.7:1
7.9:1
8.8:1
9.5:1
8.8:1
8.8:1
8.8:1
8.8:1
5
6
7
8
9
10
11
12
13
14
15
16
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+12.000 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+11.991 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
+12.009 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
1.02E-3 V
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
8.8:1
50 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Channel E1418A
Output
1
2
3
4
5
6
7
8
9
10
11
12
13
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
+16.000 V
Min Measured Max M.U. TAR
DC Voltage Short Circuit Current Check
NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
0.060 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA
14
15
16
Q
Q
Q
1
1
1
1
1
1
1
1
1
+16.000 V
+16.000 V
+16.000 V
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
DC Current Output Check
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
0.060 A
0.060 A
0.060 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
2.3E-5 A
2.3E-5 A
2.3E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide 51
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
2
2
2
2
2
2
2
2
2
3
3
3
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
Min Measured Max M.U. TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
3
Q
Q
Q
Q
3
3
3
3
3
4
4
4
4
4
4
4
4
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
Q
4
–0.02000 A
–0.020023 A
–0.019977 A
1.10E-5 A
2.1:1
52 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
5
5
5
5
5
5
5
5
5
6
6
6
6
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
Min Measured Max M.U. TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
6
6
6
6
6
7
7
7
7
7
7
7
7
7
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide 53
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
8
8
8
8
8
8
8
8
8
9
9
9
9
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
Min Measured Max M.U. TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
9
9
9
9
9
10
10
10
10
10
10
10
10
10
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
54 Keysight E1418A Service Guide
Performance Test Record Verification Tests
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
11
11
11
11
11
11
11
11
11
12
12
12
12
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
Min Measured Max M.U. TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
12
12
12
12
12
13
13
13
13
13
13
13
13
13
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
Keysight E1418A Service Guide 55
Verification Tests Performance Test Record
Quick Check
Q
Q
Q
Q
Channel E1418A
Output
14
14
14
14
14
14
14
14
14
15
15
15
15
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
Min Measured Max M.U. TAR
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
Q
Q
Q
Q
Q
15
15
15
15
15
16
16
16
16
16
16
16
16
16
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
+0.02000 A
+0.01500 A
+0.01000 A
+0.00500 A
0.00000 A
–0.00500 A
–0.01000 A
–0.01500 A
–0.02000 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.019977 A
+0.0149815 A
+0.009986 A
+0.0049905 A
–0.000005 A
–0.0050095 A
–0.010014 A
–0.0150185 A
–0.020023 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
+0.020023 A
+0.0150185 A
+0.010014 A
+0.0050095 A
+0.000005 A
–0.0049905 A
–0.009986 A
–0.0149815 A
–0.019977 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1.10E-5 A
9.50E-6 A
8.00E-6 A
6.50E-6 A
5.00E-6 A
6.50E-6 A
8.00E-6 A
9.50E-6 A
1.10E-5 A
1:1
1.5:1
1.7:1
1.9:1
2.1:1
2.1:1
1.9:1
1.7:1
1.5:1
1:1
1.5:1
1.7:1
1.9:1
2.1:1
56 Keysight E1418A Service Guide

5 Adjustments

Introduction

Keysight Technologies Calibration Services

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This chapter contains procedures for adjusting the calibration constants in the Keysight E1418A. Run the Performance Tests (described on Chapter 4) before any adjustments to characterize the module. Run the Performance Tests after any adjustments to verify the adjustments made. The module should be adjusted following any reconfiguration or repair.
Contact the local Keysight Technologies Service Center for low cost calibration services. The Keysight E1418A is supported on automated calibration systems which allow calibration at competitive prices. Calibrations to MIL-STD-45662 are also available.
Calibration Interval
The Keysight E1418A 8/16 Channel D/A Converter Module should be calibrated on a regular interval determined by the accuracy requirements of you application. Keysight Technologies recommends calibration every 90 days for most applications. Demanding applications may require a daily calibration cycle. Keysight Technologies does not recommend extending the calibration interval beyond one year in any application.
Whatever calibration interval you use, Keysight Technologies recommends that a complete adjustment be performed. This increases confidence that the module will remain in specification for the next calibration interval and provides the best measure of the module’s long-term stability. Performance Verification Test data can be used to extend or determine future calibration intervals.

Closed-Cover Electronic Calibration

The Keysight E1418A features closed-cover electronic calibration. There are no internal mechanical adjustments. Once properly installed, the module can be tested and adjusted using connections to a terminal module.
Do not perform any of the following adjustment procedures unless you are a qualified, service trained technician and have read the warnings and cautions.
57
Adjustments Introduction

Calibration Constants and Non-Volatile Memory

The accuracy of a channel output depends upon a number of calibration constants. Calibration constants are stored in two locations within the module; in non-volatile memory and in RAM. The constants in RAM are used by the module to adjust all outputs during use.
When the module is used with the SCPI driver, the RAM constants are loaded from the non-volatile memory at power-up. You can adjust the RAM calibration constants without disturbing the non-volatile memory constants (creating a temporary adjustment).
Non-volatile memory has a finite number of writes. Writing the calibration constants to non-volatile memory, therefore, reduces the life of this memory. If you are calibrating the module at 90 day or 1 year intervals, write the new constants into non-volatile memory. The new constants will be used following power-up. If you are adjusting the module more frequently (i.e., daily), write the new calibration constants into RAM, but do not write to non-volatile memory.
The adjustment procedures in this chapter demonstrate both methods of writing calibration constants.
58 Keysight E1418A Service Guide
Making Connections Adjustments

Making Connections

All adjustments can be performed using the CAL output terminals. The CAL output terminals provide one set of connections for both voltage and current adjustment. Procedures in this chapter assume connections are made at the CAL output terminals on the screw-type terminal module.
For demanding applications or for adjustment using the intended load, you can perform the adjustment at each channel output (to include the output relay contacts, wiring, and the load in the path). To make the adjustments at each channel output, use the DIAG:CAL:OUTP CHAN command to set the adjustment point at the channels instead of the CAL output terminals. Adjustment at each channel output will require individual four-wire connections.

Figure 5-1 CAL Output Terminals

Keysight E1418A Service Guide 59
Adjustments Adjustment Procedure

Adjustment Procedure

Adjustment is performed on each channel, one at a time. Voltage output and current output each use unique calibration constants.

Preparation

Before performing any adjustments, determine if the adjustments are to be temporary or permanent.
If Permanent
1 Turn off the mainframe.
2 Remove the terminal module.
3 Remove the module from the mainframe.
4 Set the Cal Store Enable Jumper to the CAL position as shown below.
Figure 5-2 Cal Store Enable Jumper
5 Re-install the module in the mainframe.
6 Make the CAL output terminal connections on the terminal module.
7 Install the terminal module.
8 Apply power and allow a 15 minute warm-up.
60 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
If Temporary
1 Make the CAL output terminal connections on the terminal module.
2 Install the terminal module.
3 Apply power and allow a 15 minute warm-up.

Voltage Adjustment

Voltage output adjustment uses two of the CALibration subsystem queries.
CALibrationn:CONFigure:VOLTage?
CALibrationn:VALue:VOLTage? <value>
The general procedure for channel voltage adjustment is:
1 Make the connections shown in figure 5-3 or 5-4.
2 Send the CALn:CONF:VOLT? query to the module and read the response until
a 1 is returned (typically the first query response).
3 Read the voltage output on the DMM.
4 Send the DMM value read with the CALn:VAL:VOLT? value query.
5 Enter the query response.
6 Repeat steps 3, 4, and 5 until the CALn:VAL:VOLT? value query returns a 0.
7 Repeat steps 1 through 6 for each channel.
The first query, CALn:CONF:VOLT?, sets the channel to the calibration mode and returns a 1 when the channel is ready for calibration. When a 1 is returned, the channel output voltage is set to the first adjustment point.
Read the output voltage on the DMM and return the value obtained using the CALn:VAL:VOLT? value query. This query returns an integer indicating the calibration state of the channel. Any non-zero return from this query indicates additional values are needed.
Each channel will require multiple iterations of the CALn:VAL:VOLT? value query. A minimum of 9 queries, to a maximum of 50 queries, will be required at each channel. During the process, the output voltage will range from +16 V to -16 V and the last values output will be at or near 0 volts.
Keysight E1418A Service Guide 61
Adjustments Adjustment Procedure
Figure 5-3 Voltage Adjustment Connections (CAL)
62 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
Figure 5-4 Optional Voltage Adjustment Connections

Current Adjustment

Current output uses an internal resistance value to calibrate all channels. This resistance value is approximately 244 Ω.. The current is output adjusted by measuring the voltage developed across this resistor.
To set the internal resistance value use the following CALibration commands:
CALibration:CONFigure:RESistance?
CALibration:VALue:RESistance <value>
You may also, optionally, use an external resistance value (or load) on each channel for current output and calibration. If you wish to specify an external resistor value, you must send the DIAG:CAL:OUTP CHAN command and perform the adjustment at each channel output. Any load value may be used but values in the range of 50 Ω to 500 Ω are recommended.
Note that during the adjustment process self-heating of the external load may affect the adjustment.
Keysight E1418A Service Guide 63
Adjustments Adjustment Procedure
The resistance value set, either internal or external, is not stored. When using the internal resistor value, the general procedure is:
1 Make the connections shown in Figure 5-5.
2 Send the CALibration:CONFigure:RESistance? query to the module and read
the response until a 1 is returned (typically the first query response).
3 Read the resistance on the DMM.
4 Send the DMM value read to the module using the
CALibration:VALue:RESistance <value> command.
The resistor value is only set once for current calibration of all channels.
Optional
If you are using an external resistor value, for each channel:
1 Send the CALibration:CONFigure:RESistance? query to the module and read
the response until a 1 is returned (typically the first query response).
2 Measure the channel’s external resistor value on the DMM.
3 Send the DMM value read to the module using the
CALibration:VALue:RESistance <value> command.
64 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
Figure 5-5 Internal Resistor Adjustment Connections
Current output adjustment uses two of the CALibration subsystem queries. Current adjustment requires voltage measurements.
CALibrationn:CONFigure:CURRent?
CALibrationn:VALue:CURRent? <value>
The current adjustment procedure measures and inputs a voltage value developed across a known resistor value (set at the start of the procedure).
Once the calibration resistance value is set as described earlier, the general procedure for each channel current output adjustment is:
1 Make the connections shown in figure 5-6 or 5-7.
2 Send the CALn:CONF:CURR? query to the module and read the response until
a 1 is returned (typically the first query response).
3 Read the voltage on the DMM.
Keysight E1418A Service Guide 65
Adjustments Adjustment Procedure
4 Send the DMM value read with the CALn:VAL:CURR? <value> query.
5 Enter the query response.
6 Repeat steps 3, 4, and 5 until the CALn:VAL:CURR? <value> query returns a 0.
7 Repeat steps 1 through 6 for each channel.
The first query, CALn:CONF:CURR?, sets the channel to the calibration mode and returns a 1 when the channel is ready for calibration. When a 1 is returned, the channel output voltage is set to the first adjustment point.
Read the voltage on the DMM and return the value obtained using the CALn:VAL:CURR? <value> query. This query returns an integer indicating the calibration state of the channel. A value other than 0 or 2 from this query indicates additional values are needed.
Each channel will require multiple iterations of the CALn:VAL:CURR? <value> query. A minimum of 9 queries, to a maximum of 50 queries, will be required at each channel. During the process, the output voltage will range from -5 to +5 V and the last values output will be at or near 0 volts.
Figure 5-6 Current Adjustment Connections (CAL)
66 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
Figure 5-7 Optional Current Adjustment Connections

Storing the Adjustments

Once all adjustments have been made, you can store the new calibration constants in non-volatile memory by sending the following command:
/* Store the new constants */
CALibration:STORe
This command will generate an error if the Cal Store Enable Jumper is not set to the CAL position.
The Keysight E1418A will be busy storing the new calibration constants following the CAL:STOR command. You should not perform any bus resets or module resets until the store is complete. You can use the *OPC? query to determine when the command has finished.

Verify the Adjustments

Run the Performance Verification Tests to verify the adjustments made.
Keysight E1418A Service Guide 67
Adjustments Adjustment Procedure

Return the Module to Use

If you have changed the position of the Cal Store Enable Jumper and you wish to prevent writing calibration constants to non-volatile memory, set the jumper to the SECURE position.
If you have changed the position of the terminal jumpers return the jumpers to their original positions.

Example Program

The following program, written in ANSI C illustrates a full calibration of all channels for both voltage and current. The internal calibration resistor is used. The disk provided with this manual contains a calibration example file. This example uses function calls to the Keysight VISA Transition Library for I/O operations.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x); /* VTL error routine */
void sys_err(ViSession resource); /* Checks for SCPI programming errors */ void pause(); /* Waits for a keystroke to continue program execution */ int config_check (); /* Checks and reports the module configuration */
/* returns the number of channels in module*/ float get_voltage (); /* Obtains the measured voltage*/ float get_resistance ();/* Obtains the measured resistance */
void main (void)
{
char cal_point [5],int_ext [4];
int num_chan,result,compare,i,condition [4];
float meas_volt,meas_res;
/*** VOLTAGE OUTPUT ADJUSTMENT ***/
for (i=1;i<num_chan+1;i++)
68 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
{
result = 0;
while (result < 1)
{
err=viPrintf(dac, “CAL%d:CONF:VOLT?\n”,i); /* Configure for calibration */
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 1 when ready to calibrate */
if(err < VI_SUCCESS) err_handler(dac, err);
}
result = 1;
while (result !=0)
{
meas_volt=get_voltage ();/* Get the voltage measurement */
err=viPrintf(dac, “CAL%d:VAL:VOLT? %f\n”,i,meas_volt); /* Send the measurement*/
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 0 when calibration complete */
if(err < VI_SUCCESS) err_handler(dac, err);
if (result == 5)/* Calibration for this channel aborted*/
{
printf (“Calibration for channel %d aborted\n”,i);
result = 0;
}
}
}
/***CURRENT OUTPUT ADJUSTMENT***/
/*** set the calibration resistor value***/
result = 0;
while (result != 1)
{
err=viPrintf(dac, “CAL:CONF:RES?\n”);/* Configure for resistance calibration */
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 1 when ready to calibrate*/
if(err < VI_SUCCESS) err_handler(dac, err);
}
meas_res = get_resistance();
Keysight E1418A Service Guide 69
Adjustments Adjustment Procedure
err=viPrintf(dac, “CAL:VAL:RES %f\n”,meas_res); /* Send the resistance value */
if(err < VI_SUCCESS) err_handler(dac, err);
/*** channel current output adjustment ***/
for (i=1;i<num_chan+1;i++)
{
result = 0;
while (result != 1)
{
err=viPrintf(dac, “CAL%d:CONF:CURR? \n”,i); /* Configure for calibration */
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/*, returns a 1 when ready to calibrate */
if(err < VI_SUCCESS) err_handler(dac, err);
}
result = 1;
while (result !=0 && result !=2)
{
meas_volt=get_voltage ();/* Get the voltage measurement */
err=viPrintf(dac, “CAL%d:VAL:CURR? %f\n”,i,meas_volt);
if(err < VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d”, &result);/* Enter integer returned*/
/* returns a 0 when calibration complete */
if(err < VI_SUCCESS) err_handler(dac, err);
if (result == 5)/* Calibration aborted */
{
printf (“Calibration for channel %d aborted\n\n”,i);
result = 0;
}
}
}
/*** STORE THE NEW CALIBRATION CONSTANTS***/
err=viPrintf(dac, “CAL:STOR;*OPC?\n”);/* Store the new cal constants */
if(err < VI_SUCCESS) err_handler(dac, err);
err=(viScanf(dac,"%d",&result);
if(err < VI_SUCCESS) err_handler(dac, err);
sys_err(dac);
70 Keysight E1418A Service Guide
Adjustment Procedure Adjustments
}
Keysight E1418A Service Guide 71
Adjustments Adjustment Procedure
72 Keysight E1418A Service Guide

6 Service

Introduction

Equipment Required

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This chapter contains troubleshooting, repair and maintenance guidelines.
Do not perform any of the service procedures shown unless you are a qualified, service-trained technician, and have read the warnings cautions .
Test equipment required is listed in Table 1-1. Any equipment that satisfies the requirements listed in the table may be used.
The Keysight E1418A uses Torx head screws. To avoid damage to the screw head slots, use only a Torx driver for disassembly and assembly. You will need a T10 Torx driver (Keysight Part Number 8710-1284).

Service Aids

Service aids, manual updates, and service literature may be available for the Keysight E1418A. For information, contact your nearest Keysight Technologies Sales and Service Office.
73
Service Troubleshooting

Troubleshooting

This section will help isolate a failing assembly.

Visual Checks

Before installing or operating the module, visually inspect the rear panel and front panel connectors for bent pins. Inspect circuit boards for evidence of arcing or excessive heat.
If a module does not respond to commands, verify the Logical Address setting. The Keysight E1418A User’s Guide describes this switch setting.

Catastrophic Failures

If a module does not respond, verify the power fuses F1, F2, F3 and F4.

Self-Test

The internal self-test can be used to isolate most failures of the module. The major assemblies of the Keysight E1418A are available on an exchange basis.
Use the following SCPI Common Command query to initiate a module self-test.
/* Start the Self-test*/
*TST?
/* Enter the 16-bit integer returned*/
The query will return a +0 if no errors are encountered. Any return other than +0 indicates a failure. The command returns the first test to fail, additional failures are possible.
The *TST? query initiates a full self-test. You can obtain the complete test results by using the TEST:TST:RESults? query. This query returns an array of 100 integers representing the results of each self-test.
Use the following SCPI commands to initiate a self-test and obtain complete results.
/* Start the Self-test*/
*TST?
/* Enter the 16-bit integer returned*/
/* If the returned value is not +0, send*/ TEST:TST:RES?
/*Enter an array of 100 integers*/
74 Keysight E1418A Service Guide
Troubleshooting Service
The integers returned from the self test query have the form:
Bit 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Use Test Number A24 Test Information

General Self-Test

The test number returned by the *TST? query can be used to isolate a failing assembly. The possible test numbers and probable failing assembly are shown below.
Test Number Failing Assembly Indicated Notes
0 None All Self-tests passed
1 Main Board
2 through 9 Main Board or Plug-On Modules Test 2 = channel 1, Test 3 = channel 2, etc.
10 through 17 Expansion Board or Plug-on Modules Test 10 = channel 9, Test 11 = channel 10, etc.
18 through 25 Main Board
26 through 33 Main Board
34 through 41 Expansion Board
42 through 43 Main Board
44 through 51 Main Board
52 through 59 Expansion Board
60 through 67 Main Board
68 through 75 Expansion Board
Since the *TST? query returns only the first failure test number, additional failures are possible but not reported by this query. The complete test results can be obtained using the TEST:TST:RES? query after the *TST? query.
Keysight E1418A Service Guide 75
Service Troubleshooting

Detailed Self-Test

Following the *TST? query, the TEST:TST:RESults? query returns complete self-test details. The following table lists each self-test and the areas checked.
Test Number A24 Value Test Information
Value If Failed
1 0 1 Relay Control Register
0 2 A/D multiplexer
2 through 17 0 Channel Number First DAC check of each channel
18 0 for A16 read
1 for A16 read in A24 window 2 for A16 read in A24
19 0 for A16 read/write
1 for A16 read/write via A24 window
20 0 Register Value VXI Control Register
21 0 Register Value Cal Control Register
22 Lower 8 bits of failed register value Channel Mode Register
23 0 for A16 writes
A16 reads 1 for A16 writes A24 reads 2 for A24 writes A16 reads 3 for A24 writes A24 reads
Register Address ID Register, Device Type Register, VXI Offset
Register Value VXI Control Register
Register Value Card Control Register
Areas Checked
Register, Card Configuration Register, Isolation Status
Register, Channel Program Jumper Register
VXI Status Register
24 0 Checksum Error Non-volatile memory checksum
25 Register Value Trigger Control Register
76 Keysight E1418A Service Guide
Troubleshooting Service
Test Number A24 Value Test Information
Value If Failed
26 through 41
(Test 26 = Ch 1 Test 41 = Ch 16)
42 0 Register Value Interrupt Control Register
43 0 1 through 5 Interrupt Status Register, SW Trigger Register,
0 1 Channel SW Trigger Register
0 2 Channel WriteThru Register
0 3 Channel Output
0 4 Channel Triggered Register
0 5 Channel Output
0 6 Channel SW and Trigger Registers
0 7 through 16 Channel SW and Triggered Registers
0 17 Channel Triggered Register
0 18 Channel Output
0 19 Channel SW and Triggered Registers
0 20 Channel Output
Areas Checked
and trig ignored bit
44 through 59
(Test 44 = Ch 1 Test 59 = Ch 16)
0 1 through 5 Voltage mode, Set the Gain/Offset Register to
nominal and 5 incrementing values, read the values in the Calibration Control Register
0 6 Check linearity of values
0 7 Gain Register set to minimum, Channel
Writethru register set to 1 V, read the Calibration Control Register
0 8 Gain Register set to maximum,
read the Calibration Control Register, calculate change from previous test
0 9 Gain Register set to nominal, Offset Register
set to minimum, Channel Writethru register set to -1 V, read the Calibration Control Register
0 10 Offset Register set to maximum,
read the Calibration Control Register, calculate change from previous test
Keysight E1418A Service Guide 77
Service Troubleshooting
Test Number A24 Value Test Information
Value If Failed
60 through 75
(Test 60 = Ch 1 Test 75 = Ch 16)
76 through 100 0 0 Not Used
0 1 through 5 Current mode, Set the Gain/Offset Register to
0 6 Check linearity of values
0 7 Gain Register set to minimum, Channel Write
0 8 Gain Register set to maximum,
0 9 Gain Register set to nominal, Offset Register
0 10 Offset Register set to maximum,
Areas Checked
nominal and 5 incrementing values, read the values in the Calibration Control Register
thru register set to 1 V, read the Calibration Control Register
read the Calibration Control Register, calculate change from previous test
set to minimum, Channel Write thru register set to -1 V, read the Calibration Control Register
read the Calibration Control Register, calculate change from previous test

Single Channel Failures

A single channel failure, not caught by the self-test, can be caused by one of two components: the channel plug-on module or the channel output relay.
To help isolate a plug-on module, disassemble the module and swap two plug-on modules. DO NOT operate the module without a plug-on module installed for every channel.
If a channel still fails after swapping the plug-on modules, suspect the channel output relay.
78 Keysight E1418A Service Guide
Repair/Maintenance Guidelines Service

Repair/Maintenance Guidelines

Electrostatic Discharge (ESD) Precautions

Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 volts. Observe the following guidelines when servicing the Keysight E1418A.
Disassemble and reassemble only in a static-free work area.
Remove all plastic, Styrofoam, vinyl, paper, and other static-generating
materials from the immediate work area.
Use a conductive work area to dissipate static charge.
Use a conductive wrist strap to dissipate static charge.
Minimize handling.
Do not stack assemblies.
Use only anti-static solder suckers, soldering irons, and tools.
Keep replacement parts in original static-free packaging.

Surface Mount Repair

The Keysight E1418A uses surface mount components. Surface mount components should only be removed using soldering irons or desoldering stations expressly designed for surface-mount components. Use of conventional soldering equipment will almost always result in permanent damage to the printed circuit board and will void your Keysight Technologies warranty.

Disassembly

Tools Needed
T10 Torx Driver.
Static-Safe Workstation.
Static Wrist Strap.
Keysight E1418A Service Guide 79
Service Repair/Maintenance Guidelines
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 volts. Disassemble and re-configure only in a static free work area. Minimize handling of assemblies and components. Keep all assemblies and replacement parts in the original static free packaging.
These procedures should be performed by qualified service personnel at approved static-safe workstations.

Disassembly Procedure

Refer to Figures 6-1 and 6-2 during these procedures.
1 Remove the 8 T-10 Torx screws in the top cover and remove the top cover.
2 Remove the T-10 Torx screw holding the sheet metal spacer or expansion
board.
a If the Keysight E1418A is an 8 channel device, lift the sheet metal spacer
off the main board.
b If the Keysight E1418A is a 16 channel device, carefully separate the
expansion board and main board printed circuit assemblies. The two boards are joined by three electrical connectors.
3 The main board and expansion board each have 8 plug-on assemblies. To
remove the plug-on assemblies, remove the T-10 Torx screw securing the plug-on assembly. Lift the plug-on assembly off the electrical connector. The plug-on assembly locations are shown in Figure 6-3.
Reassembly procedure
1 Replace all plug-on assemblies. DO NOT operate the module without a
plug-on assembly installed. The plug-on assemblies have a locator hole to ensure correct orientation, see Figure 6-4.
2 Replace the expansion board or sheet-metal spacer.
3 Replace the top cover.
80 Keysight E1418A Service Guide
Repair/Maintenance Guidelines Service
Figure 6-1 8-Channel Disassembly
Keysight E1418A Service Guide 81
Service Repair/Maintenance Guidelines
Figure 6-2 16-Channel Disassembly
82 Keysight E1418A Service Guide
Repair/Maintenance Guidelines Service
Figure 6-3 Plug-on Channel Locations
Figure 6-4 Installing Plug-on Modules
Keysight E1418A Service Guide 83
Service Repair/Maintenance Guidelines
84 Keysight E1418A Service Guide
Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
7 Replaceable Parts

Introduction

This chapter contains the following topics:
Assembly Exchange
Ordering Information

Replaceable Parts

Assembly Exchange

The following assemblies are available for exchange.
Part number Assembly
E1418-69201 Main board and all sheet metal. Does NOT include plug-on
assemblies or expansion board.
E1418-69502 Expansion board. Does NOT include plug-on assemblies.
E1418-66503 Isolated plug-on module (replacement, not an exchange).
When exchanging either the main board or the expansion board, you must disassemble the module and remove all plug-on assemblies before sending the exchange part. Keep all parts removed in a safe place.
If you are exchanging a main board from a module configured as a 16-channel device, you must remove the expansion board before sending the exchange part.
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 volts. Disassemble and re-configure only in a static free work area. Minimize handling of assemblies and components. Keep all assemblies and replacement parts in the original static free packaging.
85
Replaceable Parts Ordering Information
These procedures should be performed by qualified service personnel at approved static safe workstations.

To Exchange an Assembly

1 Contact the nearest Keysight Technologies Sales and Service Office for
instructions on packaging and shipping.
2 Disassemble the module, using the procedures given in Chapter 6 of this
manual, until just the exchange part is obtained. Remove the plug-on modules and expansion board if any.
3 Keep all removed parts in a static-safe place. Do not send plug-on modules
with the exchange assembly. Plug-on modules should be individually replaced, if necessary.
4 Package the assembly in static-safe packaging material.
5 Include a description of the problem encountered as well as your phone
number and return address.

Ordering Information

To order a part listed as an exchange assembly or a part from Table 7-1, 7-2 or 7-3, specify the Keysight part number and the quantity desired. Send the order to your nearest Keysight Technologies Sales and Service Office.
Table 7-1 Keysight E1418A Exchange/Replaceable Assemblies
Description Reference Designator Exchange
Main board and sheet metal. Does NOT include plug-on assembles or expansion board.
Expansion Board. Does NOT include plug-on assemblies.
Isolated Plug-On Assembly PCB1 through PCB16 None E1418-66503
Non-Isolated plug-On Assembly PCB1 through PCB16 None E1418-66504
A1 E1418-69201 E1418-66501
A2 E1418-69502 E1418-66502
Part Number
New Part Number
86 Keysight E1418A Service Guide
Ordering Information Replaceable Parts
Table 7-2 Keysight E1418A Replaceable Parts
Reference Designator
F1001 - 1004 2110-0863 4 Fuse 5A 125V SMT 04703 R459005
HDL1 E1400-45102 1 Hndl-Bottom Metal Injection 28480 E1400-45102
HDL2 E1400-45101 1 Hndl-Top Metal Injection 28480 E1400-45101
MP11 8160-0686 1 Clip-EMC, VXI Module 28480 8160-0686
PCA1 E1418-66501 1 PCA-8CH D/A Converter 28480 E1418-66501
PCA2 E1418-66502 1 PCA-8CH Piggyback 28480 E1418-66502
PCB1-16 E1418-66503 16 PCA-Isolated CH 28480 E1418-66503
PNL1 E1418-00201 1 Panel Front 28480 E1418-00201
SCR1-8 0515-0372 8 SCR-PANM3x.5TXSC 00000
SCR1-2 E1400-00610 2 Shoulder Screw Assembly 28480 E1400-00610
SCR3-10 0515-1135 8 SCRFHM3.0x25TX 00000
SCR14-15 0515-2733 2 SCRPHM2.5 17MM-LG 00000
SCR16-23 0515-0372 9 SCRPANM3 .5TXSC 00000
Keysight Part Number
E1418-66504 PCA-NON Isolated CH 28480 E1418-66504
QTY Description Mfr
Code
Mfr Part Number
SCR26 0515-1410 1 SCRPHM3.0x20TXSC 00000
SHD2 E1418-00601 1 SHIELD-TOP 28480 E1418-00601
SHD3 E1418-00604 1 Internal Spacer 28480 E1418-00604
SHD4 E1418-00602 1 SHIELD-BOTTOM 28480 E1418-00602
XJ3001 1258-0141 1 JMPR-REM .025P 28480 1258-0141
Keysight E1418A Service Guide 87
Replaceable Parts Ordering Information
Table 7-3 Terminal Module Replaceable Parts
Reference Designator
Terminal Module - Common Parts
A1 E1400-84405 1 Case Assembly-Term Lexus 24840 E1400-84405
MP1-2 1390-1027 2 Receptical Quick Fastener 24840 1390-1027
MP3 1460-2552 1 Torsion Spring Left Hand 24840 1460-2552
MP4 1460-2553 1 Torsion Spring Right Hand 24840 1460-2553
MP5 E1400-45103 1 Top Lever 24840 E1400-45103
MP6 E1400-45104 Bottom Lever 24840 E1400-45104
Standard Screw Terminal Module
JM1-32 1259-0141 32 JMPR-REM .025P 24840 1259-0141
MP11-14 1400-0507 4 Cable Tie Nat Nyl 00000
PCA1 E1418-66510 1 PCA Terminal Board 24840 E1418-66510
Crimp and Insert Terminal Module Option A3E
Keysight Part Number
E1418-60101 1 Terminal Block Assembly 24840 E1418-60101
E1418-60102 1 Terminal Block Asembly 24840 E1418-60102
QTY Description Mfr Code Mfr Part
Number
MP7-10 E1400-21204 4 Crimp & Insert Connector Sup 24840 E1400-21204
P1-2 1252-6532 2 Female Conn Housing Din 24840 1252-6532
SCR1-4 0515-0905 4 SCRPHM2 .5x06PZ 00000
Ribbon Cable Terminal Module Option A3H
E1418-60103 1 Term Block Ribbon Cable 24840 E1418-60103
MP7-10 E1400-21204 4 Crimp & Insert Connector Sup 24840 E1400-21204
MP11-14 1400-0507 4 Cable Tie Nat Nyl 00000
P1-2 1252-6894 2 DIN-IDC 64 CONT CONN 24840 1252-6894
SCR1-4 0515-0905 4 SCRPHM2 .5x06PZ 00000
88 Keysight E1418A Service Guide
Ordering Information Replaceable Parts
Table 7-4 Keysight E1418A Reference Designators
Reference Designators
A Assembly
F Fuse
HDL Handle
J Electrical Connector
JM Jumper
MP Miscellaneous Mechanical Part
P Electrical Connector
PCA Printed Circuit Assembly
PCB Printed Circuit Board
PNL Panel
SCR Screw
SHD Shield
XJ Removable Jumper
Table 7-5. Keysight E1418A Code List of Manufacturers
Mfr. Code Manufacturer’s Name Manufacturer’s Address Zip Code
00000 Any suitable supplier may be used
04703 Little Fuse Inc. Arcola, IL, U.S.A. 61910
28480 Keysight Technologies - Corporate Palo Alto, CA, U.S.A. 94304
Keysight E1418A Service Guide 89
Replaceable Parts Ordering Information

Figure 7-1 Keysight E1418A Replaceable Parts

90 Keysight E1418A Service Guide
Ordering Information Replaceable Parts

Figure 7-2 Keysight E1418A Replaceable Parts

Keysight E1418A Service Guide 91
Replaceable Parts Ordering Information

Figure 7-3 Keysight E1418A Replaceable Parts

92 Keysight E1418A Service Guide
Ordering Information Replaceable Parts

Figure 7-4 Keysight E1418A Replaceable Parts

Keysight E1418A Service Guide 93
Replaceable Parts Ordering Information
94 Keysight E1418A Service Guide
Loading...