Keysight E1418A Series, E1418A, E1418A Option 003, E1418A Option 001, E1418A Option002 Service Manual

75000 Series C

Service Manual

Keysight E1418A 8/16 Channel D/A Converter
Notices
© Keysight Technologies, Inc. 1996-2019
No part of this manual may be repro­duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con­sent from Keysight Technologies, Inc. as governed by United States and interna­tional copyright laws.
Manual Part Number
E1418-90010
Edition
Second Edition, October 2019
Published by
Keysight Technologies, Inc. 900 S. Taft Ave. Loveland, CO 80537 USA
Sales and Technical Support
To contact Keysight for sales and techni­cal support, refer to the support links on the following Keysight websites:
www.keysight.com/find/E1418A
(product-specific information and sup­port, software and documentation updates)
www.keysight.com/find/assist (world-
wide contact information for repair and service)
Declaration of Conformity
Declarations of Conformity for this prod­uct and for other Keysight products may be downloaded from the Web. Go to
http://keysight.com/go/conformity and
click on “Declarations of Conformity.” You can then search by product number to find the latest Declaration of Conformity.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDI­TIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WAR­RANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFOR­MATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEP­ARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Keysight Technologies does not warrant third-party system-level (combination of chassis, controllers, modules, etc.) per­formance, safety, or regulatory compli­ance unless specifically stated.
DFARS/Restricted Rights Notices
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ stan­dard commercial license terms, and non­DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Gov­ernment users will receive no greater than Limited Rights as defined in FAR
52.227-14 (June 1987) or DFAR 252.227­7015 (b)(2) (November 1995), as applica­ble in any technical data.
Safety Information
The following general safety precau­tions must be observed during all phases of operation of this instrument. Failure to comply with these precau­tions or with specific warnings or oper­ating instructions in the product manuals violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technolo­gies assumes no liability for the cus­tomer's failure to comply with these requirements.
General
Do not use this product in any manner not specified by the manufacturer. The protec­tive features of this product must not be impaired if it is used in a manner specified in the operation instructions.
Before Applying Power
Verify that all safety precautions are taken. Make all connections to the unit before applying power. Note the external markings described under “Safety Symbols”.
Ground the Instrument
Keysight chassis’ are provided with a grounding-type power plug. The instrument chassis and cover must be connected to an electrical ground to minimize shock hazard. The ground pin must be firmly connected to an electri­cal ground (safety ground) terminal at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in per­sonal injury.
Do Not Operate in an Explosive Atmosphere
Do not operate the module/chassis in the presence of flammable gases or fumes.
Do Not Operate Near Flammable Liquids
Do not operate the module/chassis in the presence of flammable liquids or near containers of such liquids.
Cleaning
Clean the outside of the Keysight mod­ule/chassis with a soft, lint-free, slightly dampened cloth. Do not use detergent or chemical solvents.
Do Not Remove Instrument Cover
Only qualified, service-trained person­nel who are aware of the hazards involved should remove instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Keep away from live circuits
Operating personnel must not remove equipment covers or shields. Proce­dures involving the removal of covers and shields are for use by service­trained personnel only. Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous elec­trical shock, DO NOT perform proce­dures involving cover or shield removal unless you are qualified to do so.
DO NOT operate damaged equipment
Whenever it is possible that the safety protection features built into this prod­uct have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service­trained personnel. If necessary, return the product to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
DO NOT block the primary disconnect
The primary disconnect device is the appliance connector/power cord when a chassis used by itself, but when installed into a rack or system the dis­connect may be impaired and must be considered part of the installation.
Do Not Modify the Instrument
Do not install substitute parts or per­form any unauthorized modification to the product. Return the product to a Keysight Sales and Service Office to ensure that safety features are main­tained.
In Case of Damage
Instruments that appear damaged or defective should be made inoperative and secured against unintended oper­ation until they can be repaired by qualified service personnel
Do NOT block vents and fan exhaust: To ensure adequate cooling and venti­lation, leave a gap of at least 50mm (2") around vent holes on both sides of the chassis.
Do NOT operate with empty slots: To ensure proper cooling and avoid dam­aging equipment, fill each empty slot with an AXIe filler panel module.
Do NOT stack free-standing chassis: Stacked chassis should be rack­mounted.
All modules are grounded through the chassis: During installation, tighten each module's retaining screws to secure the module to the chassis and to make the ground connection.
Operator is responsible to maintain safe operating conditions. To ensure safe operating conditions, modules should not be operated beyond the full temperature range specified in the Environmental and physical specifica­tion. Exceeding safe operating condi­tions can result in shorter lifespan, improper module performance and user safety issues. When the modules are in use and operation within the specified full temperature range is not maintained, module surface tempera­tures may exceed safe handling condi­tions which can cause discomfort or burns if touched. In the event of a module exceeding the full temperature range, always allow the module to cool before touching or removing modules from the chassis.
4
Safety Symbols
A CAUTION denotes a hazard. It calls attention to an operating pro­cedure or practice, that, if not cor­rectly performed or adhered to could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated condi­tions are fully understood and met.
A WARNING denotes a hazard. It calls attention to an operating pro­cedure or practice, that, if not cor­rectly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi­cated conditions are fully under­stood and met.
Products display the following sym­bols:
Warning, risk of electric shock
Refer to manual for addi­tional safety information.
Earth Ground.
Chassis Ground.
Alternating Current (AC).
Direct Current (DC)
5
6
Contents
1 General Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Module Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Recommended Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1 Installation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Initial Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Prepare the Command Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ordering Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Field Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Terminal Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Terminal Module Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Terminal Module Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Shipping Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2 Operating Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Querying Module Identification and Configuration . . . . . . . . . . . . . . . . . . . 15
Configuration Example. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Cleaning Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Main Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Operator’s Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Operator Check Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Keysight E1418A Service Guide 7
3 Verification Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Test Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
About the Verification Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Functional Verification Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Quick Verification Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Performance Verification Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Performance Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
About Program Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
SCPI drivers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
SCPI Command Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Functional Verification Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Functional Verification Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Functional Verification Test Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Performance Verification Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
DC Voltage Output Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
DC Voltage Compliance Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
DC Voltage Short Circuit Output Current. . . . . . . . . . . . . . . . . . . . . . . . . . . 34
DC Current Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Performance Verification Tests Example. . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Performance Test Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Test Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Voltage Output Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
DC Compliance Current Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
DC Short Circuit Current Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Test Accuracy Ratio (TAR). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4 Adjustments
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Keysight Technologies Calibration Services. . . . . . . . . . . . . . . . . . . . . . . . . 57
Calibration Interval. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Closed-Cover Electronic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Calibration Constants and Non-Volatile Memory . . . . . . . . . . . . . . . . . . . . 58
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
If Permanent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
If Temporary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Voltage Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Current Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Optional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8 Keysight E1418A Service Guide
5 Service
Storing the Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Verify the Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Return the Module to Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Example Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Service Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Visual Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Catastrophic Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
General Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Detailed Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Single Channel Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Repair/Maintenance Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Electrostatic Discharge (ESD) Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 79
Surface Mount Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Tools Needed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Disassembly Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
6 Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Assembly Exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
To Exchange an Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7 Manual Changes
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
A Keysight E1418A Specifications
DC Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
DC Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
General Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Cooling Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Keysight E1418A Service Guide 9
10 Keysight E1418A Service Guide

1 General Information

Introduction

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This manual contains information required to test, troubleshoot, and repair the Keysight E1418A 8/16 Channel D/A Converter Module. See the Keysight E1418A User’s Manual for additional module information. Figure 1-1 shows the Keysight E1418A module.

Figure 1-1 Keysight E1418A D/A Converter Module

11
General Information Safety Considerations

Safety Considerations

This product is a Safety Class I instrument that is provided with a protective earth terminal when installed in the mainframe. The instrument, mainframe, and all related documentation should be reviewed for familiarization with safety markings and instructions before operation or service.
Refer to the WARNINGS page (page 6) in this manual for a summary of safety information. Safety information for preventive maintenance, testing, adjusting, and service follows and is also found throughout this manual.

Warnings and Cautions

This section contains WARNINGS which must be followed for your protection and CAUTIONS which must be followed to avoid damage to the equipment when performing instrument maintenance or repair.
SERVICE-TRAINED PERSONNEL ONLY. The information in this manual is for service-trained personnel who are familiar with electronic circuitry and are aware of the hazards involved. To avoid personal injury or damage to the instrument, do not perform procedures in this manual or do any servicing unless you are qualified to do so.
CHECK MAINFRAME POWER SETTINGS. Before applying power, verify that the mainframe setting matches the line voltage and that the correct fuse is installed. An uninterruptible safety earth ground must be provided from the main power source to the supplied power cord set.
GROUNDING REQUIREMENTS. Interruption of the protective (grounding) conductor (inside or outside the mainframe) or disconnecting the protective earth terminal will cause a potential shock hazard that could result in personal injury. (Grounding one conductor of a two-conductor outlet is not sufficient protection.)
IMPAIRED PROTECTION. Whenever it is likely that instrument protection has been impaired, the mainframe must be made inoperative and be secured against any unintended operation.
REMOVE POWER IF POSSIBLE. Some procedures in this manual may be performed with power supplied to the mainframe while protective covers are removed. Energy available at many points may, if contacted, result in personal injury. (If maintenance can be performed without power applied, the power should be removed.)
12 Keysight E1418A Service Guide
Safety Considerations General Information
USING AUTOTRANSFORMERS. If the mainframe is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to neutral (that is, the grounded side of the main’s supply).
CAPACITOR VOLTAGES. Capacitors inside the mainframe may remain charged even when the mainframe has been disconnected from its source of supply.
USE PROPER FUSES. For continued protection against fire hazard, replace the line fuses only with fuses of the same current rating and type (such as normal blow, time delay, etc.). Do not use repaired fuses or short-circuited fuseholders.
SHOCK HAZARD. Only service-trained personnel who are aware of the hazards involved should install, remove, or configure the module. Before you remove any installed module, disconnect AC power from the mainframe and from other modules that may be connected to the module.
CHANNEL WIRING INSULATION. All channels that have a common connection must be insulated so that the user is protected from electrical shock. This means wiring for all channels must be insulated as though each channel carries the voltage of the highest voltage channel.
Keysight E1418A Service Guide 13
General Information

Module Description

MAXIMUM VOLTAGE. The maximum voltage that can be applied to any channel is 42 V Peak/42 Vdc. The maximum voltage from any channel to ground is 42 V Peak/42 Vdc.
STATIC ELECTRICITY. Static electricity is a major cause of component failure. To prevent damage to the electrical components in the module, observe anti-static techniques whenever working on the device.
PARTS REMOVAL. This module uses a surface mount printed circuit assembly. Special soldering equipment is required for parts removal or replacement. Use of standard soldering equipment will cause damage to the printed circuit board and is not covered under warranty.
The Keysight E1418A is an 8 or 16 channel digital-to-analog converter module for use in a VXIbus C-size mainframe. The module is a register based device. The module can be programmed via direct register access or, with the appropriate driver, by high level commands. This manual describes programming the module using SCPI (Standard Commands for Programmable Instruments) and the SCPI driver.
Each Keysight E1418A module is a unique instrument having its own output buffer and error queue. Multiple modules can not be combined into a
single instrument.
Each channel can be configured to either voltage or current output mode. When configured for voltage output, voltages in the range of -16.0 V to
+16.0 V can be set. When configured for current output, current in the range of
-0.02 to + 0.02 Amps can be set. The channel output mode can be programmatically set, or, can be forced to either voltage or current by mechanical jumpers on the terminal block.
Each output channel is individually configurable to be either an isolated output or a non-isolated output. Channel configuration to isolated or non-isolated is made by individual plug-on modules for each channel.

Module Specifications

Specifications are listed in Appendix A of this manual and in the Keysight E1418A User’s Manual. These specifications are the performance standards or limits against which the module may be tested.
14 Keysight E1418A Service Guide
Module Description General Information

Module Serial Numbers

Devices covered by this manual are identified by a serial number prefix listed on the title page. Keysight Technologies uses a two-part serial number in the form USXXXXYYYYY, where US is the country of origin, XXXX is the serial prefix, and YYYYY is the serial suffix. The serial number prefix identifies a series of identical instruments. The serial number suffix is assigned sequentially to each instrument. The serial number plate is located on the right-hand shield near the backplane connectors.
Ordering Options
The module may be ordered from Keysight Technologies in a variety of configurations. As ordered, the module may have the following options:
Description
E1418A 8-Channel D/A Converter with Non-Isolated Outputs
E1418A Option 001
E1418A Option 002
E1418A Option 003
16-Channel D/A Converter with Non-Isolated Outputs
8-Channel D/A Converter with Isolated Outputs
16-Channel D/A Converter with Isolated Outputs
Field Kits
The module can also be user configured. The following field expansion and configuration kits are available. Each field kit contains installation instructions.
Description Use
E1523A Single Channel Isolated
plug-on module
E1524A Expansion kit, 8-Channel
Non-Isolated Outputs
E1525A Expansion kit, 8-Channel
Isolated Outputs
To change a single channel from non-isolated to isolated output.
To add 8 additional non-isolated channels to an
existing 8-channel module.
To add 8 additional isolated channels to an existing
8-channel module.
Keysight E1418A Service Guide 15
General Information Recommended Test Equipment
Terminal Modules
The standard Keysight E1418A Terminal Module provides screw terminals for connections. Two other terminal options are available with the Keysight E1418A:
• Crimp and Insert (Option A3E)
• Ribbon Cable (Option A3H)

Recommended Test Equipment

Table 1-1 lists the test equipment recommended for testing and servicing the module. Essential requirements for each piece of test equipment are described in the Requirements column.
Table 1-1 Recommended Test Equipment
Instrument Requirements Recommended
Model
Controller, GPIB GPIB compatibility as defined by IEEE Standard
488-1988 and the identical ANSI Standard MC1.1: SH1, AH1, T2, TE0, L2, LE0, SR0,
RL0, PP0, DC0, DT0, and C1, 2, 3, 4, 5.
Mainframe Compatible with the Keysight E1418A E1400B/T, E1401A/T, or
Command Module, VXILink, EPC7 Embedded Controller
Digital Multimeter 5 1/2 or 6 1/2 digit
Terminal Module Compatible with the Keysight E1418A E1418-60101 O, P,T
Resistor 10 kΩ , 1%, 0.25 W, metal film Any available supplier
Resistor 600 Ω , 1%, 0.25 W, metal film P
Compatible with Module Resource manager in Slot 0 Logical address 0 Typical GPIB address 9 Requires the downloaded driver “E1418”
Voltage measurements to ± 16.8 Vdc Current measurements to ± 0.02 A Four-wire resistance measurements from 50 Ω to 500 Ω
HP 9000 Series 300 F,O,P,T
E1401B
E1405B or E1406A/B F,O,P,T
Keysight 3458A or Keysight 34401A
meeting the requirements
Use
F,O,P,T
O, P, T
P
1
User’s Manual E1418-90001 F,O,P,T
1 * F = Functional Verification Tests, O = Operation Verification Tests, P = Performance Verification
Tests, T = Troubleshooting
16 Keysight E1418A Service Guide
Keysight E1418A 8/16 Channel D/A Converter Module Service Guide

2 Installation

Introduction

This chapter contains the following topics:

Initial Inspection

Initial Inspection page 11 Preparation for Use page 12 Environment page 14 Shipping Guidelines page 14
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, keep the container until the shipment contents have been checked and the module has been checked mechanically and electrically. Chapter 4 of this manual gives procedures to check the electrical performance.
To avoid possible hazardous electrical shock, do not perform electrical tests if there are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).
If the contents are incomplete, if there is mechanical damage or defect, or if the module does not pass the electrical performance tests, notify your nearest Keysight Technologies Sales and Service Office. If the shipping container is damaged or the cushioning material shows signs of stress, notify the carrier as well as Keysight Technologies. Keep the shipping materials for the carrier’s inspection.
11
Installation Preparation for Use

Preparation for Use

Chapter 1 of the Keysight E1418A User’s Manual contains complete instructions for installing the module in a mainframe and installing a terminal module.
Refer to the mainframe user’s manual, command module user’s manual, and the Installing Device Drivers installation note to prepare the mainframe.

Prepare the Module

Verify or set the module’s logical address. Refer to the Keysight E1418A User’s Manual for instructions. In this manual the factory default logical address of 72 is used.
To prepare the module for the adjustment procedures in Chapter 5, set or verify the position of the CAL Secure Enable Jumper before installing the module in the mainframe. (The jumper is factory set to CAL.) Figure 5-2 shows this jumper.

Prepare the Terminal Module

The procedures in this manual show the screw-type terminal module. To prepare this terminal module, you will need to set the jumpers and make connections. If you are using other types of terminal modules, consult the Keysight E1418A User’s Manual for proper connections.
Terminal Module Jumpers
The screw-type terminal module contains jumpers that control how the Keysight E1418A operates. To prepare the module for the Verification Tests in Chapter 4 and Adjustments in Chapter 5, the terminal module jumpers must be set to allow the module to be programmed from the backplane. The Keysight E1418A will operate without any jumpers installed. Figure 2-1 shows the jumper locations and example settings.
There are two jumpers for each channel; a P/J jumper and a V/I jumper. The two jumpers work together.
The P/J jumper sets the channel output mode to be either Program Selectable or Jumper Selectable. When the jumper is in the program selectable position, the channel output mode can be set via programming commands. The program selectable position is recommended for the Verification Tests and Adjustment procedures in this manual. In the jumper selectable position, the channel output mode is set via the V/I jumper.
The V/I jumper sets the channel output mode to either voltage or current when the P/J jumper is set to the Jumper Selectable position. When the P/J jumper is set to Program Selectable, this jumper has no effect.
12 Keysight E1418A Service Guide
Preparation for Use Installation
With no jumpers installed, the module will be in the Program Selectable output mode.
– If you are using a terminal module dedicated to testing the Keysight
E1418A, you can remove all the jumpers or put the jumpers in the storage
position.
– If you are using the end-user’s terminal module, record the positions of the
jumpers. Set the P/J jumpers to the programmable position. Before
returning the module to use, restore the P/J jumpers to their original
position.
Figure 2-1 P/J and V/I Jumpers
Terminal Module Connections
The Verification Tests require four-wire connections at each channel. The Adjustment procedures can be performed using a single connections at the CAL terminals. Additional connection information is given in the procedures in Chapters 4 and 5 of this manual.
Keysight E1418A Service Guide 13
Installation Environment

Prepare the Command Module

All verification and adjustment procedures in this manual use SCPI (Standard Commands for Programmable Instruments). Prepare the command module by downloading the SCPI driver named “E1418". Downloading instructions are given in the Installing Device Drivers installation note.

Environment

The recommended operating altitude for the module is 2000 m (6000 ft) or less. The recommended operating environment for the module is:
Environment Temperature Humidity
o
Operating 0
Storage and Shipment
C to + 55 oC <65% relative (0 oC to + 40 oC)
non condensing
o
C to + 75 oC <65% relative (0 oC to + 40 oC)
– 40
non condensing

Shipping Guidelines

If you need to return the Keysight E1418A to Keysight Technologies, first remove any terminal modules before packaging for shipment. Attach a tag to the module identifying the owner and indicating the service or repair required. Mark the shipping container “FRAGILE”. If you are exchanging a module, read Assembly Exchange.
In any correspondence, refer to the module by model number and full serial number.
Keysight Technologies recommends using shipping containers and materials identical to those used by the factory. These materials are available through Keysight Technologies Sales and Service Offices.
If you use commercially available shipping materials, place the module in an anti-static bag and wrap the module in heavy paper or plastic. Use a strong shipping container. A double wall carton of 2400 Pa (350 psi) test material is adequate.
Use enough shock-absorbing material (layer at least 75 mm to 110 mm, 3 in to 4 in) around the module to provide firm cushion and to prevent movement in the container. Protect the front panel with cardboard. Seal the container securely and mark the container “FRAGILE.”
14 Keysight E1418A Service Guide

3 Operating Information

Introduction

Operation

Keysight E1418A 8/16 Channel D/A Converter Module Service Guide
This chapter provides operating information, preventive maintenance instructions, and operators checks.
The Keysight E1418A User’s Guide is your reference for module operation. The User’s Guide contains information about module setup, application examples, and a complete SCPI command reference.

Querying Module Identification and Configuration

This example will query the module for the identification string and determine the module configuration.
The module identification is obtained using the IEEE-488 Command *IDN?.
/* Send query to the module*/
*IDN?
/* Enter a string*/
The string returned will be similar to one of the following:
HEWLETT-PACKARD,E1418A_8CH,xxxxxxxxxx,A.01.00
or
HEWLETT-PACKARD,E1418A_16CH,xxxxxxxxxx,A.01.00
The xxxxxxxxxx portion of the response string is the module serial number. The number of channels in the module (8 or 16) is indicated by the E1418A_8CH or E1418A_16CH portion of the returned string.
The module configuration is obtained using the DIAGnostic:CONFiguration? query.
/* Send query to the module*/
DIAG:CONF?
/* Enter six integers and decode the integers*/
The DIAG:CONF? query returns six integers. The six integers returned contain the module configuration and have the following meaning.
15
Operating Information Operation
Integer
Meaning
Returned
1st Expansion Board ID in the form: 0 = present, 7 = none
(expansion board contains channels 9 through 16)
2nd Terminal Module ID in the form:
0 = screw type, 7 = none or other
3rd Isolated/Non-isolated Channel status
A bit set to 0 indicates an isolated channel A bit set to 1 indicates a non-isolated channel or no plug-on module installed (DO NOT operate the module without a plug-on module) Bits 0 – 15 correspond to channels 1 – 16, respectively*
4th Channel Mode
A bit set to 0 indicates a current output channel A bit set to 1 indicates a voltage output channel Bits 0 – 15 correspond to channels 1 – 16, respectively*
5th Channel Output State
A bit set to 0 indicates the channel output relay is closed A bit set to 1 indicates the channel output relay is open Bits 0 – 15 correspond to channels 1 – 16, respectively*
6th Channel Mode Programmable State (P/J Jumper)
A bit set to 0 indicates a channel is not mode programmable A bit set to 1 indicates a channel is mode programmable Bits 0 – 15 correspond to channels 1 – 16, respectively
1
1 For 8-channel configurations, the upper 8 bits of integers 3, 4, 5,
and 6 are set to 1’s.
16 Keysight E1418A Service Guide
Operation Operating Information
For example, if the following six integers are returned, the module has the configuration indicated.
Integer Decimal
Configuration
Value
1st 7 No expansion board installed
(Only channels 1 through 8)
2nd 7 No Terminal Module installed or unknown
terminal module installed
3rd -1 All channels are non-isolated
4th -1 All channels are voltage output channels
5th -1 All channel outputs are disabled
(all output relays are open)
6th -1 All channels are output mode programmable
Configuration Example
The following program segment demonstrates how to read the module identification string and the configuration. The configuration integers are bit manipulated using the C operator for bit shifting
result = result <<1
(a one bit shift to the left). The code shown in this example can be obtained from the file prftest.c on the examples disk provided with this manual.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x);/* VTL error routine*/
void sys_err(ViSession resource);/* Checks for SCPI programming errors */
/** GLOBAL **/ ViStatus err;
ViSession defaultRM, cmd, dac, dmm;
int num_chan;
void main (void)
{
int i,result = {0},config [6]={0},num_chan = {0};
err=viPrintf(dac, “DIAG:CONF?\n”);/* request module configuration */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%,6d”, &config);/* returns six integers*/
if(err VI_SUCCESS) err_handler(dac, err);
sys_err (dac);/* SCPI error check*/
Keysight E1418A Service Guide 17
Operating Information Operation
/* First integer */
result=config[0];/* Expansion board*/
if (result == 0)
{
}
else
{
printf (“Module is a 16 channel device\n”);
num_chan = 16;
if (result == 7)
{
}
else
{
}
}
printf (“Module is an 8 channel device\n”);
num_chan=8;
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Second integer */
result = config [1];/* Terminal Module */
if (result == 0)
{
}
else
{
printf (“Module has a screw type terminal module installed\n”);
if (result == 7)
{
18 Keysight E1418A Service Guide
Operation Operating Information
}
else
{
}
}
printf (“Module does NOT have a terminal module installed\n”);
printf (“Error in DIAG:CONF ? command\n”);
printf (“First value returned was %d \n”,config[0]);
printf (“Program will terminate\n”);
pause();
abort;
/* Third integer */
result=config[2];/* Isolated or non-isolated outputs */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d is configured for non-isolated output\n”,i);
printf(“Channel %d is configured for isolated output\n”,i);
result = result << 1;
}
/* Fourth integer */
result=config[3];/* Output Mode Voltage or Current */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
Keysight E1418A Service Guide 19
Operating Information Operation
}
printf (“Channel %d is set to voltage output mode\n”,i);
printf(“Channel %d is set to current output mode\n”,i);
result = result << 1;
}
/* Fifth integer */
result=config[4];/* Channel output ON or OFF */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result > = 0x8000)
{
}
else
{
}
printf (“Channel %d output is disabled\n”,i);
printf(“Channel %d output is enabled\n”,i);
result = result << 1;
}
/* Sixth integer */
result=config[5];/* Program or Jumper Selectable channels */
if (num_chan == 8) result = result << 8; /* strip upper 8 bits */
for (i=num_chan;i>0;i—)
{
if (result >= 0x8000)
{
}
else
{
}
printf (“Channel %d output mode is programmable\n”,i);
printf(“Channel %d output mode is fixed and jumper selected\n”,i);
result = result << 1;
}
}
20 Keysight E1418A Service Guide
Preventive Maintenance Operating Information

Preventive Maintenance

Preventive maintenance consists of periodically cleaning the front panel, the terminal module, and performing the Operator’s Check. Clean the module and terminal module yearly, or more often if the module is used in very dusty or humid areas. Disassembly of the module is not recommended for cleaning. However, if the module is disassembled for repair or reconfiguration, the printed circuit board assemblies (PCAs) can be blown off with a properly grounded air gun. Remove the terminal module cover to blow dust off the terminal printed circuit board and case. Table 3-1 shows the recommended cleaning equipment and supplies.
Table 3-1 Recommended Cleaning Equipment and Supplies
Description Recommended Use
Mild Soap Solution Clean face plate panel and terminal covers.
Lint-free Cloth Clean face plate panel and terminal covers.
Airgun with grounded nozzle
To eliminate possible electrical shock, disconnect ac power from the mainframe and disconnect all outputs to the terminal module before removing the module from the mainframe.
The printed circuit assemblies (PCAs) contain static-sensitive devices that can be damaged by careless handling. Use static-control devices (wrist straps, mats, and tools) when handling the module. Chapter 6 contains additional precautions for electrostatic discharge (ESD).
Remove dust from the printed circuit boards and terminal case.
Keysight E1418A Service Guide 21
Operating Information Preventive Maintenance

Cleaning Procedure

Use the following procedure to clean the Keysight E1418A.
Terminal Module
1 Remove the terminal module from the main module.
2 Disconnect all external wiring.
3 Clean the terminal module covers with a mild soap solution and lint-free
cloth.
4 Remove the terminal module cover and blow any dust or debris from the case.
5 Reassemble the terminal module.
6 Reconnect the external wiring.
Main Module
1 Remove the terminal module from the main module.
2 Clean the main module face plate with a mild soap solution and lint-free
cloth.
Disassembling the module exposes the assemblies to ESD damage and is usually unnecessary. Perform steps 3 through 5 only if the module is disassembled for repairs or reconfiguration.
3 Disassemble the module following the instructions in Chapter 6.
4 Using an airgun with a grounded nozzle, remove dust from the PCA
assemblies.
5 Reassemble the module.
22 Keysight E1418A Service Guide
Operator’s Checks Operating Information

Operator’s Checks

The operator’s check for the Keysight E1418A has two parts.
The first part consists of sending the SYStem:ERRor? query to the
module following power-on and entering the response. This check should be performed after every power-on sequence. At power-on, the Keysight E1418A performs a brief self-test that ensures the module can communicate with the backplane. A successful self-test returns an integer and a string:
+0,"No Error"
Any other response indicates a failure in the module. Chapter 6 contains troubleshooting information.
Once the power-on self-test has been checked, the operator may perform a full self-test by sending the *TST? query and entering the response. This query executes a full self-test that checks all circuits of the module except the output relays. A successful self-test returns:
+0
Any other response indicates a failure. Chapter 6 contains additional information about the self-test error codes and troubleshooting information.

Operator Check Example

This example checks the module for errors after power-on and then performs a full self-test.
/** FUNCTION PROTOTYPES **/
void main (void);
void err_handler(ViSession vi, ViStatus x);/* VTL error routine */
void sys_err(ViSession resource);/* SCPI error routine */
ViStatus err;/* Global VTL error variable */
void main (void)
{
int selftest;
char selftestresults [80];
err=viPrintf(dac, “*TST?\n”);/* run self test */
if(err VI_SUCCESS) err_handler(dac, err);
err=viScanf(dac, “%d%t”, &selftest,&selftestresults);/* enter integer and string returned */
if(err VI_SUCCESS) err_handler(dac, err);
Keysight E1418A Service Guide 23
Operating Information Operator’s Checks
if (selftest != ’0’)
{
}
else
{
}
}
printf (“Self Test failed\n”);
printf (“The first error encountered was\n”);
printf (“Error number %d %s\n”,selftest,selftestresults);
printf (“This program will terminate\n”);
pause();
abort;
printf (“Self Test PASSED”);
/*** VTL Error handling function ***/
void err_handler (ViSession dac, ViStatus err)
{
char buf[1024]={0}; viStatusDesc(dac,err,buf); printf(“VTL ERROR = %s\n”, buf); return;
} /*End of VTL error handler */
/*** SCPI error checking and reporting function ***/
void sys_err(ViSession resource) /* Test for SCPI Errors */
{
char buf [1024] = {0};
int err_no;
err = viPrintf (resource, “SYST:ERR?\n”); /* Check for an error */
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
err = viScanf (resource,"%d%t",&err_no,&buf);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
while (err_no != 0) /* Report all errors in error que */
{
printf (“\nCommand Error: %d,%s\n”,err_no,buf);
err = viPrintf (resource, “SYST:ERR?\n”);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
err = viScanf (resource,"%d%t",&err_no,&buf);
if (err < VI_SUCCESS) err_handler(resource,err); /* Check VTL errors */
}
flushall();/* Clean out the buffers */
24 Keysight E1418A Service Guide
Loading...
+ 76 hidden pages