Keysight DSO-X 6002A Users Guide

Keysight InfiniiVision 6000 X-Series Oscilloscopes
User's Guide
Notices
CAUTION
© Keysight Technologies, Inc. 2005-2019
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
54609-97048
Edition
Sixth edition, July 2019
Available in electronic format only Published by:
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Print History
54609-97000, April 2014
54609-97012, September 2015
54609-97024, April 2017
54609-97035, November 2017
54609-97047, May 2019
54609-97048, July 2019
Warranty
The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology License
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government
requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
2 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 3

InfiniiVision 6000 X-Series Oscilloscopes—At a Glance

Table 1 6000 X-Series Model Numbers, Bandwidths, Sample Rates
Model Analog
Channels
DSO-X 6002A 2 (none) = 1 GHz
DSO-X 6004A 4
MSO-X 6002A 2 16
MSO-X 6004A 4 16
The Keysight InfiniiVision 6000 X-Series oscilloscopes deliver these features:
1 GHz, 2.5 GHz, 4 GHz, and 6 GHz bandwidth options.
4 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Digital Channels
Bandwidth Options Sample Rate (interleaved,
non-interleaved)
20 GSa/s, 10 GSa/s
BW250 = 2.5 GHz
BW400 = 4 GHz
BW600 = 6 GHz
2- and 4-channel digital storage oscilloscope (DSO) models.
2+16-channel and 4+16-channel mixed-signal oscilloscope (MSO) models.
An MSO lets you debug your mixed-signal designs using analog signals and tightly correlated digital signals simultaneously. The 16 digital channels have a 1 GSa/s sample rate, with a 250 MHz toggle rate.
12.1 inch SVGA capacitive touchscreen display. The touchscreen makes the oscilloscope easier to use:
You can use the pinch and flick multi-touch gestures on waveforms to
change a channel's V/div setting, the time/div setting, or the delay time.
You can "touch" inside alpha-numeric keypad dialogs to enter file, label,
network, and printer names, etc., instead of using softkeys and the Entry knob.
You can drag a finger across the screen to draw rectangular boxes for
zooming in on waveforms or setting up Zone triggers.
You can touch the blue menu icon in the sidebar to display information or
control dialogs. You can drag (undock) these dialogs out of the sidebar, for example, to view cursor values and measurements at the same time.
You can touch other areas of the screen as substitutes for using front panel
keys, softkeys, and knobs.
Voice control commands for hands-free operation.
Interleaved 4 Mpts or non-interleaved 2 Mpts MegaZoom IV memory for the fastest waveform update rates.
All knobs are pushable for making quick selections.
Trigger types: edge, edge then edge, pulse width, pattern, OR, rise/fall time, Nth edge burst, runt, setup & hold, video, and zone.
Reference waveform locations (4) for comparing with other channel or math waveforms.
Many built-in measurements and a measurement statistics display.
Serial decode/trigger options for: CAN/CAN FD/LIN, CXPI, FlexRay, I
2
I
S, Manchester/NRZ, MIL-STD-1553/ARINC 429, SENT, UART/RS232, USB,
and USB PD. There is a Lister for displaying serial decode packets.
Math waveforms: add, subtract, multiply, divide, d/dt, integrate, FFT (Magnitude), FFT (Phase), Ax+B, square, square root, absolute value, common logarithm, natural logarithm, exponential, base 10 exponential, low pass filter, high pass filter, averaged value, smoothing, envelope, magnify, maximum,
2
C/SPI,
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 5
minimum, peak-peak, max hold, min hold, measurement trend, chart logic bus timing, chart logic bus state, and clock recovery.
Jitter measurement analysis and real-time eye analysis option.
Analysis features/options including: Color grade waveform analysis, digital voltmeter (DVM) and counter option, histogram display/statistics, mask testing option, power measurements and analysis option, and precision measurements and math functions.
Built-in, license-enabled 2-channel waveform generator with: arbitrary, sine, square, ramp, pulse, DC, noise, sine cardinal, exponential rise, exponential fall, cardiac, and Gaussian pulse. Modulated waveforms on WaveGen1 except for arbitrary, pulse, DC, and noise waveforms.
USB and LAN ports make printing, saving, and sharing data easy.
VGA port for displaying the screen on a different monitor.
A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help. Complete instructions for using the quick help system are given in "Access the Built-In Quick Help" on page 67.
For more information about InfiniiVision oscilloscopes, see:
www.keysight.com/find/scope
6 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

In This Guide

This guide shows how to use the InfiniiVision 6000 X-Series oscilloscopes.
When unpacking and using the oscilloscope for the first time, see:
When displaying waveforms and acquired data, see:
When setting up triggers or changing how data is acquired, see:
Making measurements and analyzing data:
Chapter 1, “Getting Started,” starting on page 31
Chapter 2, “Horizontal Controls,” starting on page 69
Chapter 3, “Vertical Controls,” starting on page 85
Chapter 4, “Math Waveforms,” starting on page 95
Chapter 5, “Reference Waveforms,” starting on page
133
Chapter 6, “Digital Channels,” starting on page 137
Chapter 7, “Serial Decode,” starting on page 155
Chapter 8, “Display Settings,” starting on page 161
Chapter 9, “Labels,” starting on page 171
Chapter 10, “Triggers,” starting on page 177
Chapter 11, “Trigger Mode/Coupling,” starting on
page 215
Chapter 12, “Acquisition Control,” starting on page
223
Chapter 13, “Cursors,” starting on page 241
Chapter 14, “Measurements,” starting on page 251
Chapter 15, “Histogram,” starting on page 285
Chapter 16, “Color Grade Waveform,” starting on page
295
Chapter 17, “Jitter and Real-Time Eye Analysis,”
starting on page 299
Chapter 18, “Mask Testing,” starting on page 311
Chapter 19, “Digital Voltmeter and Counter,” starting
on page 325
Chapter 20, “Frequency Response Analysis,” starting
on page 331
When using the built-in license enabled waveform generator, see:
Chapter 21, “Waveform Generator,” starting on page
337
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 7
When saving, recalling, or printing, see:
Chapter 22, “Save/Email/Recall (Setups, Screens,
Data),” starting on page 357
Chapter 23, “Print (Screens),” starting on page 371
When using the oscilloscope's utility functions or web interface, see:
For reference information, see: Chapter 26, “Reference,” starting on page 411
When using licensed serial bus triggering and decode features, see:
Chapter 24, “Utility Settings,” starting on page 377
Chapter 25, “Web Interface,” starting on page 397
Chapter 27, “CAN/LIN Triggering and Serial Decode,”
starting on page 429
Chapter 28, “CXPI Triggering and Serial Decode,”
starting on page 451
Chapter 29, “FlexRay Triggering and Serial Decode,”
starting on page 461
Chapter 30, “I2C/SPI Triggering and Serial Decode,”
starting on page 471
Chapter 31, “I2S Triggering and Serial Decode,”
starting on page 491
Chapter 32, “Manchester/NRZ Triggering and Serial
Decode,” starting on page 501
Chapter 33, “MIL-STD-1553/ARINC 429 Triggering
and Serial Decode,” starting on page 517
Chapter 34, “SENT Triggering and Serial Decode,”
starting on page 533
Chapter 35, “UART/RS232 Triggering and Serial
Decode,” starting on page 547
Chapter 36, “USB 2.0 Triggering and Serial Decode,”
starting on page 557
Chapter 37, “USB PD Triggering and Serial Decode,”
starting on page 567
8 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
NOTE
Abbreviated instructions for pressing a series of keys and softkeys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:
Press [Key1] > Softkey2 > Softkey3.
The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 9
10 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Contents

InfiniiVision 6000 X-Series Oscilloscopes—At a Glance / 4
In This Guide / 7
1 Getting Started
Inspect the Package Contents / 31
Tilt the Oscilloscope for Easy Viewing / 34
Power-On the Oscilloscope / 34
Use only the power cord that came with the oscilloscope / 35
Connect Probes to the Oscilloscope / 35
Maximum input voltage at analog inputs / 36
Do not float the oscilloscope chassis / 36
Input a Waveform / 36
Recall the Default Oscilloscope Setup / 37
Use Autoscale / 37
Compensate Passive Probes / 39
Learn the Front Panel Controls and Connectors / 40
Front Panel Overlays for Different Languages / 48
Learn the Touchscreen Controls / 50
Draw Rectangles for Waveform Zoom or Zone Trigger Set
Up / 51
Pinch, Flick, or Drag to Scale, Position, and Change Offset / 52
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 11
Select Sidebar Information or Controls / 54 Undock Sidebar Dialogs by Dragging / 55 Select Dialog Menus and Close Dialogs / 56 Drag Cursors / 56 Touch Softkeys and Menus On the Screen / 56 Enter Names Using Alpha-Numeric Keypad Dialogs / 57 Change Waveform Offsets By Dragging Ground Reference
Icons / 58 Access Controls and Menus Using the Spark Icon / 59 Turn Channels On/Off and Open Scale/Offset Dialogs / 61 Access the Horizontal Menu and Open the Scale/Delay
Dialog / 61 Access the Trigger Menu, Change the Trigger Mode, and Open the
Trigger Level Dialog / 62 Use a USB Mouse and/or Keyboard for Touchscreen
Controls / 63
Learn the Voice Controls / 63
Learn the Rear Panel Connectors / 64
Learn the Oscilloscope Display / 65
Access the Built-In Quick Help / 67
2 Horizontal Controls
To adjust the horizontal (time/div) scale / 70
To adjust the horizontal delay (position) / 71
Panning and Zooming Single or Stopped Acquisitions / 72
To change the horizontal time mode (Normal, XY, or Roll) / 73
XY Time Mode / 74
To display the zoomed time base / 77
To change the horizontal scale knob's coarse/fine adjustment
setting / 78
12 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
3 Vertical Controls
To position the time reference (left, center, right, custom) / 79
Searching for Events / 79
To set up searches / 80 To copy search setups / 80
Navigating the Time Base / 81
To navigate time / 81 To navigate search events / 82 To navigate segments / 82
To turn waveforms on or off (channel or math) / 86
To adjust the vertical scale / 87
To adjust the vertical position / 87
To specify channel coupling / 88
To specify channel input impedance / 89
To specify bandwidth limiting / 89
To change the vertical scale knob's coarse/fine adjustment
setting / 90
To invert a waveform / 91
Setting Analog Channel Probe Options / 91
To specify the channel units / 92 To specify the probe attenuation / 92 To specify the probe skew / 93 To calibrate a probe / 93
4Math Waveforms
To display math waveforms / 95
To adjust the math waveform scale and offset / 97
Units for Math Waveforms / 97
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 13
Math Operators / 98
Add or Subtract / 98 Multiply or Divide / 99
Math Transforms / 100
Differentiate / 101 Integrate / 102 FFT Magnitude, FFT Phase / 105 Square Root / 115 Ax + B / 115 Square / 116 Absolute Value / 117 Common Logarithm / 117 Natural Logarithm / 118 Exponential / 118 Base 10 Exponential / 119
Math Filters / 119
High Pass and Low Pass Filter / 120 Averaged Value / 121 Smoothing / 121 Envelope / 122
Math Visualizations / 122
Magnify / 122 Maximum/Minimum / 123 Peak-Peak / 124 Max/Min Hold / 124 Measurement Trend / 124 Chart Logic Bus Timing / 126 Chart Logic Bus State / 127 Clock Recovery / 128
The Measurement Record and Waveform Math / 130
14 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
5 Reference Waveforms
To save a waveform to a reference waveform location / 133
To display a reference waveform / 134
To scale and position reference waveforms / 135
To adjust reference waveform skew / 136
To display reference waveform information / 136
To save/recall reference waveform files to/from a USB storage
device / 136
6 Digital Channels
To connect the digital probes to the device under test / 137
Probe cable for digital channels / 138
Acquiring waveforms using the digital channels / 141
To display digital channels using Autoscale / 141
Interpreting the digital waveform display / 142
To change the displayed size of the digital channels / 143
To switch a single channel on or off / 144
To switch all digital channels on or off / 144
To switch groups of channels on or off / 144
To change the logic threshold for digital channels / 145
To reposition a digital channel / 145
To display digital channels as a bus / 146
Digital channel signal fidelity: Probe impedance and
grounding / 149 Input Impedance / 150 Probe Grounding / 152
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 15
7Serial Decode
8 Display Settings
Best Probing Practices / 154
Serial Decode Options / 155
Lister / 157
Searching Lister Data / 159
To adjust waveform intensity / 161
To set or clear persistence / 163
To clear the display / 165
To select the grid type / 165
To adjust the grid intensity / 165
To add an annotation / 166
To display waveforms as vectors or dots / 168
To disable/enable antialiasing / 169
To freeze the display / 170
9Labels
To turn the label display on or off / 171
To assign a predefined label to a channel / 172
To define a new label / 173
To load a list of labels from a text file you create / 174
To reset the label library to the factory default / 175
10 Triggers
Adjusting the Trigger Level / 178
Forcing a Trigger / 179
16 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Edge Trigger / 180
Edge then Edge Trigger / 182
Pulse Width Trigger / 184
Pattern Trigger / 187
Hex Bus Pattern Trigger / 189
OR Trigger / 190
Rise/Fall Time Trigger / 191
Nth Edge Burst Trigger / 193
Runt Trigger / 194
Setup and Hold Trigger / 197
Video Trigger / 198
To set up Generic video triggers / 203 To trigger on a specific line of video / 204 To trigger on all sync pulses / 205 To trigger on a specific field of the video signal / 206 To trigger on all fields of the video signal / 207 To trigger on odd or even fields / 208
Serial Trigger / 211
Zone Qualified Trigger / 212
11 Trigger Mode/Coupling
To select the Auto or Normal trigger mode / 216
To select the trigger coupling / 217
To enable or disable trigger noise rejection / 219
To enable or disable trigger HF Reject / 219
To set the trigger holdoff / 220
External Trigger Input / 221
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 17
12 Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run
Overview of Sampling / 224
Selecting the Acquisition Mode / 229
Data Acquisition (Sampling) Mode / 236
Maximum voltage at oscilloscope external trigger input / 221
Control) / 223
Sampling Theory / 225 Aliasing / 225 Oscilloscope Bandwidth and Sample Rate / 225 Oscilloscope Rise Time / 227 Oscilloscope Bandwidth Required / 228 Memory Depth and Sample Rate / 229
Normal Acquisition Mode / 230 Peak Detect Acquisition Mode / 230 Averaging Acquisition Mode / 233 High Resolution Acquisition Mode / 235
Realtime Sampling and Oscilloscope Bandwidth / 237
Acquiring to Segmented Memory / 238
Navigating Segments / 239 Measurements, Statistics, and Infinite Persistence with
Segmented Memory / 239 Segmented Memory Re-Arm Time / 240 Saving Data from Segmented Memory / 240
13 Cursors
To make cursor measurements / 242
Cursor Examples / 245
18 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
14 Measurements
To make automatic measurements / 252
To edit measurements / 254
Measurements Summary / 254
Snapshot All / 258
Voltage Measurements / 259
Peak-Peak / 259 Maximum / 260 Minimum / 260 Y at X / 260 Amplitude / 260 Top / 260 Base / 261 Overshoot / 261 Preshoot / 263 Average / 263 DC RMS / 264 AC RMS / 264 Ratio / 266
Time Measurements / 266
Period / 267 Frequency / 267 Counter / 268 + Width / 269 – Width / 269 Burst Width / 269 Duty Cycle / 269 Bit Rate / 270 Rise Time / 270 Fall Time / 270 Time at Edge / 270
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 19
Delay / 271 Phase / 272 X at Min Y / 274 X at Max Y / 274
Count Measurements / 274
Positive Pulse Count / 274 Negative Pulse Count / 275 Rising Edge Count / 275 Falling Edges Count / 275
Mixed Measurements / 276
Area / 276 Slew Rate / 276
FFT Analysis Measurements / 277
Channel Power / 277 Occupied Bandwidth / 277 Adjacent Channel Power Ratio (ACPR) / 277 Total Harmonic Distortion (THD) / 278
Measurement Thresholds / 279
Measurement Window / 281
Measurement Statistics / 281
Precision Measurements and Math / 284
15 Histogram
Waveform Histogram Set Up / 286
Defining the Histogram Limits Window / 288
Measurement Histogram Set Up / 290
Histogram Data Graph / 291
Histogram Data Statistics / 292
20 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
16 Color Grade Waveform
Enabling a Color Grade Waveform / 296
Color Grade Themes / 298
17 Jitter and Real-Time Eye Analysis
Setting Up Jitter Analysis / 300
Jitter Measurements / 303
Data TIE / 303 Clock TIE / 304 N-Period / 304 Period-Period / 305 +Width to +Width / 306
-Width to -Width / 306
Real-Time Eye Analysis / 307
18 Mask Testing
To create a mask from a "golden" waveform (Automask) / 311
Mask Test Setup Options / 313
Mask Statistics / 316
To manually modify a mask file / 317
Building a Mask File / 321
How is mask testing done? / 324
19 Digital Voltmeter and Counter
Digital Voltmeter / 326
Counter / 328
20 Frequency Response Analysis
To make connections / 331
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 21
To set up and run the analysis / 332
To view and save the analysis results / 334
21 Waveform Generator
To select generated waveform types and settings / 337
To edit arbitrary waveforms / 341
To output the waveform generator sync pulse / 348
To specify the expected output load / 349
To use waveform generator logic presets / 349
To add noise to the waveform generator output / 350
To add modulation to the waveform generator output / 350
Creating New Arbitrary Waveforms / 342 Editing Existing Arbitrary Waveforms / 343 Capturing Other Waveforms to the Arbitrary Waveform / 348
To set up Amplitude Modulation (AM) / 351 To set up Frequency Modulation (FM) / 353 To set up Frequency-Shift Keying Modulation (FSK) / 354
To restore waveform generator defaults / 355
To set up dual channel tracking / 355
22 Save/Email/Recall (Setups, Screens, Data)
Saving Setups, Screen Images, or Data / 357
To save setup files / 359 To save BMP or PNG image files / 360 To save CSV, ASCII XY, or BIN data files / 360 Length Control / 362 To save Lister data files / 363 To save reference waveform files to a USB storage device / 364 To save masks / 364 To save arbitrary waveforms / 365
22 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
23 Print (Screens)
To navigate storage locations / 365 To enter file names / 365
Emailing Setups, Screen Images, or Data / 366
Recalling Setups, Masks, or Data / 367
To recall setup files / 368 To recall mask files / 368 To recall reference waveform files from a USB storage
device / 368 To recall arbitrary waveforms / 369
Recalling Default Setups / 369
Performing a Secure Erase / 370
To print the oscilloscope's display / 371
To set up network printer connections / 373
To specify the print options / 374
To specify the palette option / 375
24 Utility Settings
I/O Interface Settings / 377
Setting up the Oscilloscope's LAN Connection / 378
To establish a LAN connection / 379 Stand-alone (Point-to-Point) Connection to a PC / 380
File Explorer / 380
Setting Oscilloscope Preferences / 383
To choose "expand about" center or ground / 383 To disable/enable transparent backgrounds / 384 To set voice recognition and speaker options / 384 To set up the screen saver / 385
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 23
To set Autoscale preferences / 386 Jitter-Free Trigger / 387
Setting the Oscilloscope's Clock / 387
Setting the Rear Panel TRIG OUT Source / 388
Setting the Reference Signal Mode / 389
To supply a sample clock to the oscilloscope / 389
Maximum input voltage at 10 MHz REF connector / 389 To synchronize the timebase of two or more instruments / 390
Enabling Remote Command Logging / 391
Performing Service Tasks / 392
To perform user calibration / 392 To perform hardware self test / 393 To perform front panel self test / 393 To display oscilloscope information / 394 To display the user calibration status / 394 To clean the oscilloscope / 394 To check warranty and extended services status / 394 To contact Keysight / 395 To return the instrument / 395
Configuring the [Quick Action] Key / 395
25 Web Interface
Accessing the Web Interface / 398
Browser Web Control / 399
Browser-Based Remote Front Panel / 400 Remote Programming via the Web Interface / 400 Remote Programming with Keysight IO Libraries / 402
Save/Recall / 403
Saving Files via the Web Interface / 403
24 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
26 Reference
Recalling Files via the Web Interface / 404
Get Image / 405
Identification Function / 406
Instrument Utilities / 407
Setting a Password / 408
Specifications and Characteristics / 411
Measurement Category / 411
Oscilloscope Measurement Category / 411 Measurement Category Definitions / 412 Maximum Input Voltages / 413
Maximum input voltage at analog inputs / 413
Maximum input voltage at digital channels / 413
Environmental Conditions / 413
Probes and Accessories / 414
Loading Licenses and Displaying License Information / 415
Licensed Options Available / 415 Other Options Available / 418 Upgrading to an MSO / 418
Software and Firmware Updates / 418
Binary Data (.bin) Format / 418
Binary Data in MATLAB / 419 Binary Header Format / 420 Example Program for Reading Binary Data / 422 Examples of Binary Files / 422
CSV and ASCII XY files / 425
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 25
CSV and ASCII XY file structure / 426 Minimum and Maximum Values in CSV Files / 426 Frequency Data in CSV Files / 427
Acknowledgements / 427
Product Markings and Regulatory Information / 427
27 CAN/LIN Triggering and Serial Decode
Setup for CAN/CAN FD Signals / 429
Loading and Displaying CAN Symbolic Data / 432
CAN/CAN FD Triggering / 433
CAN/CAN FD Serial Decode / 436
Interpreting CAN/CAN FD Decode / 437 CAN Totalizer / 438 Interpreting CAN Lister Data / 440 Searching for CAN Data in the Lister / 441
Setup for LIN Signals / 442
Loading and Displaying LIN Symbolic Data / 443
LIN Triggering / 444
LIN Serial Decode / 446
Interpreting LIN Decode / 448 Interpreting LIN Lister Data / 449 Searching for LIN Data in the Lister / 450
28 CXPI Triggering and Serial Decode
Setup for CXPI Signals / 451
CXPI Triggering / 452
CXPI Serial Decode / 455
Interpreting CXPI Decode / 457 Interpreting CXPI Lister Data / 459
26 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
29 FlexRay Triggering and Serial Decode
Setup for FlexRay Signals / 461
FlexRay Triggering / 462
Triggering on FlexRay Frames / 463 Triggering on FlexRay Errors / 464 Triggering on FlexRay Events / 464
FlexRay Serial Decode / 465
Interpreting FlexRay Decode / 466 FlexRay Totalizer / 467 Interpreting FlexRay Lister Data / 468 Searching for FlexRay Data in the Lister / 469
30 I2C/SPI Triggering and Serial Decode
Setup for I2C Signals / 471
I2C Triggering / 472
I2C Serial Decode / 476
Interpreting I2C Decode / 477 Interpreting I2C Lister Data / 479 Searching for I2C Data in the Lister / 479
Setup for SPI Signals / 480
SPI Triggering / 484
SPI Serial Decode / 486
Interpreting SPI Decode / 487 Interpreting SPI Lister Data / 488 Searching for SPI Data in the Lister / 489
31 I2S Triggering and Serial Decode
Setup for I2S Signals / 491
I2S Triggering / 494
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 27
I2S Serial Decode / 497
Interpreting I2S Decode / 498 Interpreting I2S Lister Data / 499 Searching for I2S Data in the Lister / 500
32 Manchester/NRZ Triggering and Serial Decode
Setup for Manchester Signals / 501
Manchester Triggering / 504
Manchester Serial Decode / 506
Interpreting Manchester Decode / 506 Interpreting Manchester Lister Data / 508
Setup for NRZ Signals / 509
NRZ Triggering / 512
NRZ Serial Decode / 513
Interpreting NRZ Decode / 514 Interpreting NRZ Lister Data / 515
33 MIL-STD-1553/ARINC 429 Triggering and Serial Decode
Setup for MIL-STD-1553 Signals / 517
MIL-STD-1553 Triggering / 519
MIL-STD-1553 Serial Decode / 520
Interpreting MIL-STD-1553 Decode / 521 Interpreting MIL-STD-1553 Lister Data / 522 Searching for MIL-STD-1553 Data in the Lister / 523
Setup for ARINC 429 Signals / 524
ARINC 429 Triggering / 525
ARINC 429 Serial Decode / 527
Interpreting ARINC 429 Decode / 529 ARINC 429 Totalizer / 530
28 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Interpreting ARINC 429 Lister Data / 531 Searching for ARINC 429 Data in the Lister / 532
34 SENT Triggering and Serial Decode
Setup for SENT Signals / 533
SENT Triggering / 538
SENT Serial Decode / 540
Interpreting SENT Decode / 541 Interpreting SENT Lister Data / 543 Searching for SENT Data in the Lister / 545
35 UART/RS232 Triggering and Serial Decode
Setup for UART/RS232 Signals / 547
UART/RS232 Triggering / 549
UART/RS232 Serial Decode / 551
Interpreting UART/RS232 Decode / 552 UART/RS232 Totalizer / 553 Interpreting UART/RS232 Lister Data / 554 Searching for UART/RS232 Data in the Lister / 554
36 USB 2.0 Triggering and Serial Decode
Setup for USB 2.0 Signals / 557
USB 2.0 Triggering / 559
USB 2.0 Serial Decode / 561
Interpreting USB 2.0 Decode / 562 Interpreting USB 2.0 Lister Data / 564 Searching for USB 2.0 Data in the Lister / 565
37 USB PD Triggering and Serial Decode
Setup for USB PD Signals / 567
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 29
Index
USB PD Triggering / 568
USB PD Serial Decode / 570
Interpreting USB PD Decode / 571 Interpreting USB PD Lister Data / 572
30 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Keysight InfiniiVision 6000 X-Series Oscilloscopes
User's Guide

1 Getting Started

Inspect the Package Contents / 31 Tilt the Oscilloscope for Easy Viewing / 34 Power-On the Oscilloscope / 34 Connect Probes to the Oscilloscope / 35 Input a Waveform / 36 Recall the Default Oscilloscope Setup / 37 Use Autoscale / 37 Compensate Passive Probes / 39 Learn the Front Panel Controls and Connectors / 40 Learn the Touchscreen Controls / 50 Learn the Voice Controls / 63 Learn the Rear Panel Connectors / 64 Learn the Oscilloscope Display / 65 Access the Built-In Quick Help / 67
This chapter describes the steps you take when using the oscilloscope for the first time.

Inspect the Package Contents

• Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
31
1 Getting Started
• Verify that you received the following items and any optional accessories you may have ordered:
InfiniiVision 6000 X-Series oscilloscope.
Power cord (country of origin determines specific type).
Oscilloscope probes:
Two probes for 2-channel models.
Four probes for 4-channel models.
Digital probe kit (MSO models only).
Documentation CD-ROM.
32 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
InfiniiVision 6000 X-Series oscilloscope
o
Getting Started 1
N2894A probes (Qty 2 or 4)
Documentation CD
Power cord (Based on country
f origin)
N2756-60001 Digital Probe Kit (MSO models only)
See Also "Probes and Accessories" on page 414
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 33
1 Getting Started

Tilt the Oscilloscope for Easy Viewing

There are tabs under the oscilloscope's front feet that can be flipped out to tilt the oscilloscope.

Power-On the Oscilloscope

Power
Requirements
Ventilation
Requirements
34 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Line voltage, frequency, and power:
• 100-120 Vac, 50/60/400 Hz
• 100-240 Vac, 50/60 Hz
• 200 W max
The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is required for proper cooling. Always ensure that the air intake and exhaust areas are free from obstructions.
The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscope.
To po wer-on the
CAUTION
WARNING
oscilloscope
Getting Started 1
When using the oscilloscope in a bench-top setting, provide at least 2" clearance at the sides and 4" (100 mm) clearance above and behind the oscilloscope for proper cooling.
Use only the power cord that came with the oscilloscope
The power cord provided is matched to the country of origin of the order.
1 Connect the power cord to the rear of the oscilloscope, then to a suitable AC
voltage source. Route the power cord so the oscilloscope's feet and legs do not pinch the cord. Position the oscilloscope so that access to the power switch or power cord is not restricted.
2 The oscilloscope automatically adjusts for input line voltages in the range 100
to 240 VAC. The line cord provided is matched to the country of origin.
Always use a grounded power cord. Do not defeat the power cord ground.
3 Press the power switch.
The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds.

Connect Probes to the Oscilloscope

1 Connect the oscilloscope probe to an oscilloscope channel BNC connector. 2 Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point on the circuit.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 35
1 Getting Started
CAUTION
CAUTION
CAUTION
WARNING
Maximum input voltage at analog inputs
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50 Ω input impedance setting is ±5 V.
Input protection is enabled in 50 mode and the 50 load will disconnect if greater than 5 Vrms is detected. However the inputs could still be damaged, depending on the time constant of the signal. The 50 input protection functions only when the oscilloscope is powered on.
The maximum input voltage for the 1 M input impedance setting is 30 Vrms or ±40 Vmax (DC+Vpeak)
Probing technology allows for testing of higher voltages; the included N2894A 10:1 probe supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage allowed.
When measuring voltages over 30 V, use a 10:1 probe.
Do not float the oscilloscope chassis
Defeating the ground connection and "floating" the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage. The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to measure between two live points, use a differential probe with sufficient dynamic range.
Do not negate the protective action of the ground connection to the oscilloscope. The oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard.

Input a Waveform

The first signal to input to the oscilloscope is the Demo 2, Probe Comp signal. This signal is used for compensating probes.
36 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
1 Connect an oscilloscope probe from channel 1 to the Demo 2 (Probe Comp)
terminal on the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the Demo 2
terminal).

Recall the Default Oscilloscope Setup

To recall the default oscilloscope setup: 1 Press [Default Setup].
The default setup restores the oscilloscope's default settings. This places the oscilloscope in a known operating condition. The major default settings are:
Table 2 Default Configuration Settings
Horizontal Normal mode, 100 µs/div scale, 0 s delay, center time reference.
Vertical (Analog) Channel 1 on, 5 V/div scale, DC coupling, 0 V position, 1 M impedance.
Getting Started 1
Trigger Edge trigger, Auto trigger mode, 0 V level, channel 1 source, DC coupling, rising
edge slope, 40 ns holdoff time.
Display Persistence off, 20% grid intensity, 50% waveform intensity.
Other Acquire mode normal, [Run/Stop] to Run, cursors and measurements off.
Labels All custom labels that you have created in the Label Library are preserved (not
erased), but all channel labels will be set to their original names.
In the Save/Recall Menu, there are also options for restoring the complete factory settings (see "Recalling Default Setups" on page 369) or performing a secure erase (see "Performing a Secure Erase" on page 370).

Use Autoscale

Use [Auto Scale] to automatically configure the oscilloscope to best display the input signals.
1 Press [Auto Scale].
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 37
1 Getting Started
You should see a waveform on the oscilloscope's display similar to this:
2 If you want to return to the oscilloscope settings that existed before, press Undo
Autoscale.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, press Fast Debug, Channels, or Acq Mode.
These are the same softkeys that appear in the Autoscale Preferences Menu. See "To set Autoscale preferences" on page 386.
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure "Compensate Passive Probes" on page 39.
If you do not see the waveform, make sure the probe is connected securely to the front panel channel input BNC and to the left side, Demo 2, Probe Comp terminal.
How Autoscale
Works
38 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Auto Scale analyzes any waveforms present at each channel and at the external trigger input. This includes the digital channels, if connected.
Auto Scale finds, turns on, and scales any channel with a repetitive waveform that has a frequency of at least 25 Hz, a duty cycle greater than 0.5%, and an amplitude of at least 10 mV peak-to-peak. Any channels where no signal is found are turned off.
The trigger source is selected by looking for the first valid waveform starting with external trigger, then continuing with the lowest number analog channel up to the highest number analog channel, and finally (if digital probes are connected) the highest number digital channel.
During Autoscale, the delay is set to 0.0 seconds, the horizontal time/div (sweep speed) setting is a function of the input signal (about 2 periods of the triggered signal on the screen), and the triggering mode is set to Edge.

Compensate Passive Probes

Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected. A poorly compensated probe can introduce significant measurement errors.
1 Input the Probe Comp signal (see "Input a Waveform" on page 36). 2 Press [Default Setup] to recall the default oscilloscope setup (see "Recall the
Default Oscilloscope Setup" on page 37).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
Comp signal (see "Use Autoscale" on page 37).
4 Press the channel key to which the probe is connected ([1], [2], etc.). 5 In the Channel Menu, press Probe. 6 In the Channel Probe Menu, press Probe Check; then, follow the instructions
on-screen.
Getting Started 1
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the trimmer capacitor on the probe for the flattest pulse possible.
On N2894A probes, the trimmer capacitor is located on the probe BNC connector.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 39
1 Getting Started
Perfectly compensated
Over compensated
Under compensated
7 Connect probes to all other oscilloscope channels (channel 2 of a 2-channel
oscilloscope, or channels 2, 3, and 4 of a 4-channel oscilloscope).
8 Repeat the procedure for each channel.

Learn the Front Panel Controls and Connectors

On the front panel, key refers to any key (button) you can press.
40 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Softkey specifically refers to the six keys that are directly below the display. The legend for these keys is directly above them, on the display. Their functions change as you navigate through the oscilloscope's menus.
For the following figure, refer to the numbered descriptions in the table that follows.
Getting Started 1
14. Tools keys
1. Power switch
2. Softkeys
3. [Intensity] key
4. Entry knob
6. Trigger controls
5. Waveform keys
18. Demo 2, Ground, and Demo 1
terminals
17. Analog channel
inputs
19. USB Host
ports
15. [Help] key
13. File keys
8. Run Control keys
12. Measure controls
11. Additional waveform controls
7. Horizontal controls
10. [Auto Scale] key
9. [Default Setup] key
16. Vertical controls
21. Waveform generator
outputs
20. EXT TRIG IN connector
Back
Back
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 41
1. Power switch Press once to switch power on; press again to switch power off. See
2. Softkeys The functions of these keys change based upon the menus shown on the
3. [Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to
"Power-On the Oscilloscope" on page 34.
display directly above the keys.
The Back/Up key moves up in the softkey menu hierarchy. At the top
of the hierarchy, the Back/Up key turns the menus off, and oscilloscope information is shown instead.
adjust waveform intensity.
You can vary the intensity control to bring out signal detail, much like an analog oscilloscope.
Digital channel waveform intensity is not adjustable.
More details about using the Intensity control to view signal detail are on
"To adjust waveform intensity" on page 161.
1 Getting Started
4. Entry knob The Entry knob is used to select items from menus and to change values. The function of the Entry knob changes based upon the current menu and softkey selections.
Note that the curved arrow symbol above the entry knob illuminates whenever the entry knob can be used to select a value. Also, note that
when the Entry knob symbol appears on a softkey, you can use the Entry knob, to select values.
Often, rotating the Entry knob is enough to make a selection. Sometimes, you can push the Entry knob to enable or disable a selection. Pushing the Entry knob also makes popup menus disappear.
42 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
5. Waveform keys [Analyze] key — Press this key to access analysis features like:
Trigger level setting.
Measurement threshold setting.
The USB 2.0 signal quality analysis application.
Video trigger automatic set up and display.
Color grade waveform display (see Chapter 16, “Color Grade
Waveform,” starting on page 295).
Counter (see "Counter" on page 328).
Digital voltmeter (see "Digital Voltmeter" on page 326).
Histogram of waveforms or measurements (see Chapter 15,
“Histogram,” starting on page 285).
Mask testing (see Chapter 18, “Mask Testing,” starting on page 311).
The power measurement and analysis application.
Precision measurements and math functions (see "Precision
Measurements and Math" on page 284).
Real-time eye analysis (included with the jitter analysis application, see
"Real-Time Eye Analysis" on page 307).
The [Acquire] key lets you select Normal, Peak Detect, Averaging, or High Resolution acquisition modes (see "Selecting the Acquisition
Mode" on page 229) and use segmented memory (see "Acquiring to Segmented Memory" on page 238).
The [Jitter] key lets you set up jitter analysis (see Chapter 17, “Jitter and Real-Time Eye Analysis,” starting on page 299).
[Clear Display] key — Press this key to clear acquisition data from the oscilloscope display.
The [Display] key lets you access the menu where you can enable persistence (see "To set or clear persistence" on page 163), clear the display, and adjust the display grid (graticule) intensity (see "To adjust
the grid intensity" on page 165).
6. Trigger controls These controls determine how the oscilloscope triggers to capture data. See Chapter 10, “Triggers,” starting on page 177 and Chapter 11, “Trigger Mode/Coupling,” starting on page 215.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 43
1 Getting Started
7. Horizontal
controls
The Horizontal controls consist of:
Horizontal scale knob — Turn the knob in the Horizontal section that is
marked to adjust the time/div (sweep speed) setting. The symbols under the knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale.
Horizontal position knob — Turn the knob marked to pan through
the waveform data horizontally. You can see the captured waveform before the trigger (turn the knob clockwise) or after the trigger (turn the knob counterclockwise). If you pan through the waveform when the oscilloscope is stopped (not in Run mode) then you are looking at the waveform data from the last acquisition taken.
•[Horiz] key — Press this key to open the Horizontal Menu where you can
select XY and Roll modes, enable or disable Zoom, enable or disable horizontal time/division fine adjustment, and select the trigger time reference point.
Zoom key — Press the zoom key to split the oscilloscope
display into Normal and Zoom sections without opening the Horizontal Menu.
•[Search] key — Lets you search for events in the acquired data.
•[Navigate] keys — Press these keys to navigate through captured data
via time, search events, or segmented memory acquisition. See
"Navigating the Time Base" on page 81.
For more information see Chapter 2, “Horizontal Controls,” starting on page 69.
8. Run Control
keys
9. [Default Setup]
key
When the [Run/Stop] key is green, the oscilloscope is running, that is, acquiring data when trigger conditions are met. To stop acquiring data, press [Run/Stop].
When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring data, press [Run/Stop].
To capture and display a single acquisition (whether the oscilloscope is running or stopped), press [Single]. The [Single] key is yellow until the oscilloscope triggers.
For more information, see "Running, Stopping, and Making Single
Acquisitions (Run Control)" on page 223.
Press this key to restore the oscilloscope's default settings (details on
"Recall the Default Oscilloscope Setup" on page 37).
44 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
10. [Auto Scale]
key
11. Additional
waveform controls
When you press the [Auto Scale] key, the oscilloscope will quickly determine which channels have activity, and it will turn these channels on and scale them to display the input signals. See "Use Autoscale" on page 37.
The additional waveform controls consist of:
•[Math] key — provides access to math (add, subtract, etc.) waveform
functions. See Chapter 4, “Math Waveforms,” starting on page 95.
•[Ref] key — provides access to reference waveform functions. Reference
waveforms are saved waveforms that can be displayed and compared against other analog channel or math waveforms. Also, measurements can be made on reference waveforms. See Chapter 5, “Reference Waveforms,” starting on page 133.
•[Digital] key — Press this key to turn the digital channels on or off (the
arrow to the left will illuminate).
When the arrow to the left of the [Digital] key is illuminated, the upper multiplexed knob selects (and highlights in red) individual digital channels, and the lower multiplexed knob positions the selected digital channel.
If a trace is repositioned over an existing trace the indicator at the left edge of the trace will change from Dnn designation (where nn is a one or two digit channel number from 0 to 15) to D*. The "*" indicates that two or more channels are overlaid.
You can rotate the upper knob to select an overlaid channel, then rotate the lower knob to position it just as you would any other channel.
For more information on digital channels see Chapter 6, “Digital Channels,” starting on page 137.
•[Serial] key — This key is used to enable serial decode. The multiplexed
scale and position knobs are not used with serial decode. For more information on serial decode, see Chapter 7, “Serial Decode,” starting on page 155.
Multiplexed scale knob — This scale knob is used with Math, Ref, or
Digital waveforms, whichever has the illuminated arrow to the left. For math and reference waveforms, the scale knob acts like an analog channel vertical scale knob.
Multiplexed position knob — This position knob is used with Math, Ref,
or Digital waveforms, whichever has the illuminated arrow to the left. For math and reference waveforms, the position knob acts like an analog channel vertical position knob.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 45
1 Getting Started
12. Measure
controls
13. File keys Press the [Save/Recall] key to save or recall a waveform or setup. See
14. Tools keys The Tools keys consist of:
The measure controls consist of:
Cursors knob — Push this knob to select cursors from a popup menu.
Then, after the popup menu closes (either by timeout or by pushing the knob again), rotate the knob to adjust the selected cursor position.
•[Cursors] key — Press this key to open a menu that lets you select the
cursors mode and source.
•[Meas] key — Press this key to access a set of predefined
measurements. See Chapter 14, “Measurements,” starting on page
251.
Chapter 22, “Save/Email/Recall (Setups, Screens, Data),” starting on
page 357.
The [Print] key opens the Print Configuration Menu so you can print the displayed waveforms. See Chapter 23, “Print (Screens),” starting on page
371.
• [Utility] key — Press this key to access the Utility Menu, which lets you
configure the oscilloscope's I/O settings, use the file explorer, set preferences, access the service menu, and choose other options. See
Chapter 24, “Utility Settings,” starting on page 377.
•[Quick Action] key — Press this key to perform the selected quick
action: measure all snapshot, print, save, recall, freeze display, and more. See "Configuring the [Quick Action] Key" on page 395.
•[Wave Gen1], [Wave Gen2] keys — Press these keys to access
waveform generator functions. See Chapter 21, “Waveform Generator,” starting on page 337.
15. [Help] key Opens the Help Menu where you can display overview help topics and select the Language. See also "Access the Built-In Quick Help" on page 67.
46 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
16. Vertical
controls
17. Analog channel
inputs
The Vertical controls consist of:
Analog channel on/off keys — Use these keys to switch a channel on or
off, or to access a channel's menu in the softkeys. There is one channel on/off key for each analog channel.
Vertical scale knob — There are knobs marked for each
channel. Use these knobs to change the vertical sensitivity (gain) of each analog channel.
Vertical position knobs — Use these knobs to change a channel's
vertical position on the display. There is one Vertical Position control for each analog channel.
•[Label] key — Press this key to access the Label Menu, which lets you
enter labels to identify each trace on the oscilloscope display. See
Chapter 9, “Labels,” starting on page 171.
•[Touch] key — Press this key to disable/enable the touchscreen.
For more information, see Chapter 3, “Vertical Controls,” starting on page
85.
Attach oscilloscope probes or BNC cables to these BNC connectors.
With the InfiniiVision 6000 X-Series oscilloscopes, you can set the input impedance of the analog channels to either 50 or 1 M. See "To
specify channel input impedance" on page 89.
The InfiniiVision 6000 X-Series oscilloscopes also provide the AutoProbe interface. The AutoProbe interface uses a series of contacts directly below the channel's BNC connector to transfer information between the oscilloscope and the probe. When you connect a compatible probe to the oscilloscope, the AutoProbe interface determines the type of probe and sets the oscilloscope's parameters (units, offset, attenuation, coupling, and impedance) accordingly.
18. Demo 2,
Ground, and Demo 1 terminals
Demo 2 terminal — After a [Default Setup], this terminal outputs the
Probe Comp signal which helps you match a probe's input capacitance to the oscilloscope channel to which it is connected. See
"Compensate Passive Probes" on page 39. With certain licensed
features, the oscilloscope can also output demo or training signals on this terminal.
Ground terminal — Use the ground terminal for oscilloscope probes
connected to the Demo 1 or Demo 2 terminals.
Demo 1 terminal — The oscilloscope can output training signals on this
and the Demo 2 terminal. See: [Help] > Training Signals > Training Signals.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 47
1 Getting Started
19. USB Host ports These ports are for connecting a USB mass storage device, printer, mouse, or keyboard to the oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to save or recall oscilloscope setup files and reference waveforms or to save data and screen images. See Chapter 22, “Save/Email/Recall (Setups, Screens, Data),” starting on page 357.
To print, connect a USB compliant printer. For more information about printing see Chapter 23, “Print (Screens),” starting on page 371.
You can also use the USB port to update the oscilloscope's system software when updates are available.
You do not need to take special precautions before removing the USB mass storage device from the oscilloscope (you do not need to "eject" it). Simply unplug the USB mass storage device from the oscilloscope when the file operation is complete.
CAUTION: Do not connect a host computer to the oscilloscope's USB host port. Use the device port. A host computer sees the oscilloscope as a
device, so connect the host computer to the oscilloscope's device port (on the rear panel). See "I/O Interface Settings" on page 377.
There is a third USB host port on the back panel.
20. EXT TRIG IN
connector
External trigger input BNC connector. See "External Trigger Input" on page 221 for an explanation of this feature.
21. Waveform
generator outputs
Built-in, license-enabled 2-channel waveform generator can output arbitrary, sine, square, ramp, pulse, DC, noise, sine cardinal, exponential rise, exponential fall, cardiac, or Gaussian pulse waveforms on the Gen Out 1 or Gen Out 2 BNC connectors. Modulated waveforms are available on WaveGen1 except for arbitrary, pulse, DC, and noise waveforms. Press the [Wave Gen1] or [Wave Gen2] keys to set up the waveform generator. See Chapter 21, “Waveform Generator,” starting on page 337.

Front Panel Overlays for Different Languages

Front panel overlays, which have translations for the English front panel keys and label text, are available in 10 languages. The appropriate overlay is included when the localization option is chosen at time of purchase.
To install a front panel overlay: 1 Gently pull on the front panel knobs to remove them.
48 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
2 Insert the overlay's side tabs into the slots on the front panel.
Getting Started 1
3 Reinstall the front panel knobs.
Front panel overlays may be ordered from www.keysight.com/find/parts using the following part numbers:
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 49
1 Getting Started
Language 2 Channel Overlay 4 Channel Overlay
French 54608-94314 54609-94314
German 54608-94313 54609-94313
Italian 54608-94315 54609-94315
Japanese 54608-94317 54609-94317
Korean 54608-94312 54609-94312
Portuguese 54608-94318 54609-94318
Russian 54608-94319 54609-94319
Simplified Chinese 54608-94310 54609-94310
Spanish 54608-94316 54609-94316
Traditional Chinese 54608-94311 54609-94311

Learn the Touchscreen Controls

When the [Touch] key is lit, you can control the oscilloscope by touching different areas of the screen. You can:
"Draw Rectangles for Waveform Zoom or Zone Trigger Set Up" on page 51
"Pinch, Flick, or Drag to Scale, Position, and Change Offset" on page 52
"Select Sidebar Information or Controls" on page 54
"Undock Sidebar Dialogs by Dragging" on page 55
"Select Dialog Menus and Close Dialogs" on page 56
"Drag Cursors" on page 56
"Touch Softkeys and Menus On the Screen" on page 56
"Enter Names Using Alpha-Numeric Keypad Dialogs" on page 57
"Change Waveform Offsets By Dragging Ground Reference Icons" on page 58
"Access Controls and Menus Using the Spark Icon" on page 59
"Turn Channels On/Off and Open Scale/Offset Dialogs" on page 61
"Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 61
50 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
"Access the Trigger Menu, Change the Trigger Mode, and Open the Trigger
Level Dialog" on page 62
"Use a USB Mouse and/or Keyboard for Touchscreen Controls" on page 63

Draw Rectangles for Waveform Zoom or Zone Trigger Set Up

1 Touch the upper-right corner to select the rectangle draw mode.
2 Drag your finger across the screen to draw a rectangle. 3 Take your finger off the screen. 4 Touch the desired option from the popup menu.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 51
1 Getting Started

Pinch, Flick, or Drag to Scale, Position, and Change Offset

1 Touch the upper-right corner to select the waveform drag mode.
2 When the waveform drag mode is selected, you can use these touch gestures:
52 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
Pinch — used to zoom in on a waveform of interest. A horizontal pinch
adjusts the oscilloscope's time/div and delay settings at once — for "off center zooming", this is more efficient than using knobs. A vertical pinch adjusts a waveform's V/div and offset settings at once.
To select waveforms, tap them. The waveform closest horizontally to the tap location is selected. The selected waveform is indicated by the ground marker with the filled background (channel 1 in the following example).
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 53
1 Getting Started
Flick — allows very fast browsing of waveforms. It is similar to browsing on
tablets and smartphones. It is much easier to flick than to continually turn a knob.
Drag — drag your finger across the screen to change the horizontal delay.
54 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Select Sidebar Information or Controls

1 Touch the blue menu icon in the sidebar. 2 In the popup menu, touch the type of information or controls you want to see in
the sidebar.

Undock Sidebar Dialogs by Dragging

Sidebar dialogs can be undocked and placed anywhere on the screen. 1 Drag the sidebar dialog title wherever you like.
Getting Started 1
This lets you view multiple types of information or controls at the same time.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 55
1 Getting Started

Select Dialog Menus and Close Dialogs

Drag Cursors

• Touch the blue menu icon in the dialog for options.
• Touch the red "X" icon to close a dialog.
When cursors are displayed, you can drag the name handles to position them.

Touch Softkeys and Menus On the Screen

• Touch onscreen softkey labels to select them.
This is the same as pressing the softkey keys.
56 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Getting Started 1
• When softkeys provide menus, double-touch to select a menu item.
This may be an easier than selecting a menu item via the Entry knob.

Enter Names Using Alpha-Numeric Keypad Dialogs

Some softkeys open alpha-numeric dialogs that let you touch to enter names.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 57
1 Getting Started

Change Waveform Offsets By Dragging Ground Reference Icons

58 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
When the waveform drag mode is selected, you can drag waveforms up or down to change the vertical offset.
You can always change waveform vertical offsets by dragging ground markers and labels, even when in the rectangle draw mode.
Getting Started 1

Access Controls and Menus Using the Spark Icon

1 Touch the upper-left spark icon to open the main menu.
2 Touch left side controls to perform oscilloscope operations.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 59
1 Getting Started
3 Touch menu items and submenu items to access menus and additional
controls.
60 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Turn Channels On/Off and Open Scale/Offset Dialogs

• Touch channel numbers to turn them on or off.
• When channels are on, touch the scale and offset values to access a dialog for changing them.
Getting Started 1

Access the Horizontal Menu and Open the Scale/Delay Dialog

• Touch "H" to access the Horizontal Menu.
• Touch the horizontal scale and delay values to access a dialog for changing them.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 61
1 Getting Started

Access the Trigger Menu, Change the Trigger Mode, and Open the Trigger Level Dialog

• Touch "T" to access the Trigger Menu.
• Touch the trigger level value(s) to access a dialog for changing the level(s).
• Touch "Auto" or "Trig'd" to quickly toggle the trigger mode.
62 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Use a USB Mouse and/or Keyboard for Touchscreen Controls

Connecting a USB mouse gives you a mouse pointer on the display. Mouse clicks and drags behave the same as screen touches and finger drags.
If you connect a USB keyboard, you can use it to enter values in alpha-numeric keypad dialogs.

Learn the Voice Controls

Voice control is a valuable feature when browsing signals in your device under test and using your hands to hold down probes. The set of voice commands is limited so they are easy to remember.
On the upper right corner of the display is an icon that describes the state of voice recognition:
means voice recognition is not running at all.
Getting Started 1
means voice recognition is running but it is only listening for the wakeup command.
means voice recognition is listening for commands.
To enable voice recognition, tap the icon and (in the Audio Menu) press the Voice Recognition softkey.
The initial screen shows the list of commands and includes a few brief explanations. There is also a smaller help screen that simply lists the commands.
The best way to recall the list of voice commands is to say the command "Help".
Notes on Voice
Recognition
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 63
Most equipment noise will not bother recognition but other people speaking will. Use a USB headset if other people are speaking nearby.
It is best to speak at a normal speed. Speaking slowly or with long pauses will make recognition worse just like speaking too quickly.
The voice recognition uses Speaker Adaption. This is reset each time a different person speaks, so it is best if only one person speaks to the oscilloscope.
1 Getting Started
9. USB Device port
8. USB Host port
1. Power cord connector
2. Kensington lock hole
3. TRIG OUT connector
5. Calibration protect switch
4. 10 MHz REF connector
6. Digital channel inputs
7. VGA Video Out
10. LAN port
The recognition is best when using the accent of the selected language. See "To
select the language" on page 68.

Learn the Rear Panel Connectors

For the following figure, refer to the numbered descriptions in the table that follows.
64 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
1. Power cord connector
2. Kensington lock hole
3. TRIG OUT connector
Attach the power cord here.
This is where you can attach a Kensington lock for securing the instrument.
Trigger output BNC connector. See "Setting the Rear Panel TRIG OUT
Source" on page 388.
Getting Started 1
4. 10 MHz REF connector
5. Calibration protect switch
6. Digital channel inputs
7. VGA video output
8. USB Host port This port functions identically to the USB host port on the front panel. USB
9. USB Device port
10. LAN port Lets you print to network printers (see Chapter 23, “Print (Screens),”
For synchronizing the timebase of multiple instruments. See "Setting the
Reference Signal Mode" on page 389.
See "To perform user calibration" on page 392.
Connect the digital probe cable to this connector (MSO models only). See
Chapter 6, “Digital Channels,” starting on page 137.
Lets you connect an external monitor or projector to provide a larger display or to provide a display at a viewing position away from the oscilloscope.
The oscilloscope's built-in display remains on even when an external display is connected. The video output connector is always active.
For optimal video quality and performance, we recommend you use a shielded video cable with ferrite cores.
Host Port is used for saving data from the oscilloscope and loading software updates. See also USB Host port (see page 48).
This port is for connecting the oscilloscope to a host PC. You can issue remote commands from a host PC to the oscilloscope via the USB device port. See "Remote Programming with Keysight IO Libraries" on page 402.
starting on page 371) and access the oscilloscope's built-in web server. See Chapter 25, “Web Interface,” starting on page 397 and "Accessing
the Web Interface" on page 398.

Learn the Oscilloscope Display

The oscilloscope display contains acquired waveforms, setup information, measurement results, and the softkey definitions.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 65
1 Getting Started
Analog channel sensitivity
Status line
Analog channels and ground levels
Trigger level
Digital channels
Softkey labels
Menu line
Trigger point, time reference
Delay time
Time/ div
Run/Stop status
Trigger type
Trigger source
Measurements
Trigger level or digital threshold
Sidebar information and controls area
Cursors defining
measurement Measurement statistics
Voice recognition
Draw/drag
touch mode
66 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Figure 1 Interpreting the oscilloscope display

Status line The top line of the display contains vertical, horizontal, and trigger setup
Display area The display area contains the waveform acquisitions, channel identifiers, and
information, as well as the voice recognition icon and the rectangle draw mode or waveform drag mode selector.
analog trigger, and ground level indicators. Each analog channel's information appears in a different color.
Signal detail is displayed using 256 levels of intensity. For more information about viewing signal detail see "To adjust waveform intensity" on page 161.
For more information about display modes see Chapter 8, “Display Settings,” starting on page 161.
Getting Started 1
Back
Sidebar information and controls area
Menu line This line normally contains menu name or other information associated with the
Softkey labels These labels describe softkey functions. Typically, softkeys let you set up
The sidebar information area can contain summary, cursors, measurements, or digital voltmeter information dialogs or it can contain navigation and other control dialogs.
For more information, see:
"Select Sidebar Information or Controls" on page 54
"Undock Sidebar Dialogs by Dragging" on page 55
selected menu.
additional parameters for the selected mode or menu.
Pressing the Back/Up key at the top of the menu hierarchy turns off softkey labels and displays additional status information describing channel offset and other configuration parameters.

Access the Built-In Quick Help

To view Quick Help 1 Press and hold the key or softkey for which you would like to view help.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 67
1 Getting Started
Quick Help message
Press and hold front panel key or softkey (or right-click softkey when using web browser remote front panel).
Quick Help remains on the screen until another key is pressed or a knob is turned.
To select the
language
68 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
To select the language of the user interface, Quick Help, and voice control:
1 Press [Help], then press the Language (Voice) softkey. 2 Repeatedly press and release the Language (Voice) softkey or rotate the Entry
knob until the desired language is selected.
The "Voice" in parentheses specifies the accent of the selected language for voice recognition. See "Learn the Voice Controls" on page 63.
Keysight InfiniiVision 6000 X-Series Oscilloscopes
User's Guide

2 Horizontal Controls

To adjust the horizontal (time/div) scale / 70 To adjust the horizontal delay (position) / 71 Panning and Zooming Single or Stopped Acquisitions / 72 To change the horizontal time mode (Normal, XY, or Roll) / 73 To display the zoomed time base / 77 To change the horizontal scale knob's coarse/fine adjustment setting / 78 To position the time reference (left, center, right, custom) / 79 Searching for Events / 79 Navigating the Time Base / 81
The horizontal controls include:
• Touchscreen controls for setting the horizontal scale and position (delay),
accessing the Horizontal Menu, and navigating (see "Pinch, Flick, or Drag to
Scale, Position, and Change Offset" on page 52 and "Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 61).
• The horizontal scale and position knobs.
•The [Horiz] key for accessing the Horizontal Menu.
• The zoom key for quickly enabling/disabling the split-screen zoom display.
•The [Search] key for finding events on analog channels or in serial decode.
•The [Navigate] keys for navigating time, search events, or segmented memory
acquisitions.
The following figure shows the Horizontal Menu which appears after pressing the [Horiz] key.
69
2 Horizontal Controls
Trigger point
Sample rate
Time reference
Delay time
Time/ div
Trigger source
Trigger level or threshold
XY or Roll mode
Normal time mode
Zoomed time base
Fine control
Time reference

Figure 2 Horizontal Menu

The Horizontal Menu lets you select the time mode (Normal, XY, or Roll), enable Zoom, set the time base fine control (vernier), and specify the time reference.
The current sample rate is displayed in the right-side information area.

To adjust the horizontal (time/div) scale

You can:
70 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Horizontal Controls 2
• Use the touchscreen pinch gesture (see "Pinch, Flick, or Drag to Scale,
Position, and Change Offset" on page 52).
• Use the touchscreen controls to open a horizontal scale/delay dialog (see
"Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 61).
• Turn the large horizontal scale (sweep speed) knob marked to change the horizontal time/div setting.
Notice how the time/div information in the status line changes. The symbol at the top of the display indicates the time reference point.
The horizontal scale knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See "Panning and Zooming Single or Stopped
Acquisitions" on page 72.
Note that the horizontal scale knob has a different purpose in the Zoom display. See "To display the zoomed time base" on page 77.

To adjust the horizontal delay (position)

You can:
• Use the touchscreen flick and drag gestures (see "Pinch, Flick, or Drag to
Scale, Position, and Change Offset" on page 52).
• Use the touchscreen controls to open a horizontal scale/delay dialog (see
"Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 61).
• Turn the horizontal delay (position) knob ( ).
The trigger point moves horizontally, pausing at 0.00 s (mimicking a mechanical detent), and the delay value is displayed in the status line.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and indicates how far it is from the time reference point (hollow inverted triangle ). These reference points are indicated along the top of the display grid.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 71
2 Horizontal Controls
Figure 2 shows the trigger point with the delay time set to 200 µs. The delay time
number tells you how far the time reference point is located from the trigger point. When delay time is set to zero, the delay time indicator overlays the time reference indicator.
All events displayed left of the trigger point happened before the trigger occurred. These events are called pre-trigger information, and they show events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The amount of delay range (pre-trigger and post-trigger information) available depends on the time/div selected and memory depth.
The horizontal position knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See "Panning and Zooming Single or
Stopped Acquisitions" on page 72.
Note that the horizontal position knob has a different purpose in the Zoom display. See "To display the zoomed time base" on page 77.

Panning and Zooming Single or Stopped Acquisitions

When the oscilloscope is stopped, use the touchscreen pinch, flick, and drag gestures or use the horizontal scale and position knobs to pan and zoom your waveform. The stopped display may contain several acquisitions worth of information, but only the last acquisition is available for pan and zoom.
The ability to pan (move horizontally) and scale (expand or compress horizontally) an acquired waveform is important because of the additional insight it can reveal about the captured waveform. This additional insight is often gained from seeing the waveform at different levels of abstraction. You may want to view both the big picture and the specific little picture details.
The ability to examine waveform detail after the waveform has been acquired is a benefit generally associated with digital oscilloscopes. Often this is simply the ability to freeze the display for the purpose of measuring with cursors or printing the screen. Some digital oscilloscopes go one step further by including the ability to further examine the signal details after acquiring them by panning through the waveform and changing the horizontal scale.
72 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Horizontal Controls 2
NOTE
There is no limit imposed on the scaling ratio between the time/div used to acquire the data and the time/div used to view the data. There is, however, a useful limit. This useful limit is somewhat a function of the signal you are analyzing.
Zooming into stopped acquisitions
The screen will still contain a relatively good display if you zoom-in horizontally by a factor of 1000 and zoom-in vertically by a factor of 10 to display the information from where it was acquired. Remember that you can only make automatic measurements on displayed data.

To change the horizontal time mode (Normal, XY, or Roll)

1 Press [Horiz]. 2 In the Horizontal Menu, press Time Mode; then, select:
Normal — the normal viewing mode for the oscilloscope.
In the Normal time mode, signal events occurring before the trigger are plotted to the left of the trigger point () and signal events after the trigger plotted to the right of the trigger point.
XY — XY mode changes the display from a volts-versus-time display to a
volts-versus-volts display. The time base is turned off. Channel 1 amplitude is plotted on the X-axis and Channel 2 amplitude is plotted on the Y-axis.
You can use XY mode to compare frequency and phase relationships between two signals. XY mode can also be used with transducers to display strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
Use the cursors to make measurements on XY mode waveforms.
For more information about using XY mode for measurements, refer to "XY
Time Mode" on page 74.
Roll — causes the waveform to move slowly across the screen from right to
left. It only operates on time base settings of 50 ms/div and slower. If the current time base setting is faster than the 50 ms/div limit, it will be set to 50 ms/div when Roll mode is entered.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 73
2 Horizontal Controls
sinθ =
A B
or
C D

XY Time Mode

Example This exercise shows a common use of the XY display mode by measuring the
In Roll mode there is no trigger. The fixed reference point on the screen is the right edge of the screen and refers to the current moment in time. Events that have occurred are scrolled to the left of the reference point. Since there is no trigger, no pre-trigger information is available.
If you would like to pause the display in Roll mode press the [Single] key. To clear the display and restart an acquisition in Roll mode, press the [Single] key again.
Use Roll mode on low-frequency waveforms to yield a display much like a strip chart recorder. It allows the waveform to roll across the display.
The XY time mode converts the oscilloscope from a volts-versus-time display to a volts-versus-volts display using two input channels. Channel 1 is the X-axis input, channel 2 is the Y-axis input. You can use various transducers so the display could show strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
phase difference between two signals of the same frequency with the Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the same
frequency but out of phase to channel 2.
2 Press the [Auto Scale] key, press the [Horiz] key; then, press Time Mode and select
"XY".
3 Center the signal on the display with the channel 1 and 2 position ( ) knobs.
Use the channel 1 and 2 volts/div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing.
The phase difference angle (θ) can be calculated using the following formula (assuming the amplitude is the same on both channels):
74 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Horizontal Controls 2
A
B
D
C
Signal must be cent ered i n “X”
Measuring
phase difference
Signals 90 degrees
out of phase
Signals
in phase
Figure 3 XY time mode signals, centered on display
4 Press the [Cursors] key. 5 Set the Y2 cursor to the top of the signal, and set Y1 to the bottom of the
signal. Note the ΔY value at the bottom of the display. In this example, we are using
the Y cursors, but you could have used the X cursors instead.
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis.
Again, note the ΔY value.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 75
2 Horizontal Controls
sinθ =
second Δ Y
first Δ Y
=
1.319
2.297
; θ = 35.05 degrees of phase shift
NOTE
Figure 4 Phase difference measurements, automatic and using cursors
7 Calculate the phase difference using the formula below.
For example, if the first ΔY value is 2.297 and the second ΔY value is 1.319:
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is the X-axis input, channel 2 is the Y-axis input, and the EXT TRIG IN is the Z-axis input. If you only want to see portions of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off (analog oscilloscopes called this Z-axis blanking because it turned the beam on and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off.
76 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To display the zoomed time base

These markers show the beginning and end of the Zoom window
Normal window
Time/div for zoomed window
Time/div for normal window
Delay time momentarily displays when the Horizontal position knob is turned
Zoom window
Signal anomaly expanded in zoom window
Select Zoom
Zoom, formerly called Delayed sweep mode, is a horizontally expanded version of the normal display. When Zoom is selected, the display divides in half. The top half of the display shows the normal time/div window and the bottom half displays a faster Zoom time/div window.
The Zoom window is a magnified portion of the normal time/div window. You can use Zoom to locate and horizontally expand part of the normal window for a more detailed (higher-resolution) analysis of signals.
To turn on (or off ) Zoom:
1 Press the zoom key (or press the [Horiz] key and then the Zoom softkey).
Horizontal Controls 2
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 77
2 Horizontal Controls
The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted. The box shows the portion of the normal sweep that is expanded in the lower half.
To change the time/div for the Zoom window, turn the horizontal scale (sweep speed) knob. As you turn the knob, the zoomed window time/div is highlighted in the status line above the waveform display area. The Horizontal scale (sweep speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left-to-right position of the zoom window. The delay value, which is the time displayed relative to the trigger point) is momentarily displayed in the upper-right portion of the display when the
delay time ( ) knob is turned.
Negative delay values indicate you're looking at a portion of the waveform before the trigger event, and positive values indicate you're looking at the waveform after the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn the horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to "To isolate a
pulse for Top measurement" on page 260 and "To isolate an event for frequency measurement" on page 267.

To change the horizontal scale knob's coarse/fine adjustment setting

1 Push the horizontal scale knob (or press [Horiz] > Fine) to toggle between fine
and coarse adjustment of the horizontal scale.
When Fine is enabled, turning the horizontal scale knob changes the time/div (displayed in the status line at the top of the display) in smaller increments. The time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in a 1-2-5 step sequence.
78 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Horizontal Controls 2

To position the time reference (left, center, right, custom)

Time reference is the reference point on the display for delay time (horizontal position).
1 Press [Horiz]. 2 In the Horizontal Menu, press Time Ref; then, select:
Left — the time reference is set to one major division from the left edge of the
display.
Center — the time reference is set to the center of the display.
Right — the time reference is set to one major division from the right edge of
the display.
Custom Location — lets you place the time reference location at a percent of
the graticule width (where 0% is the left edge and 100% is the right edge).
A small hollow triangle () at the top of the display grid marks the position of the time reference. When delay time is set to zero, the trigger point indicator (▼) overlays the time reference indicator.
The time reference position sets the initial position of the trigger event within acquisition memory and on the display, with delay set to 0.
Turning the Horizontal scale (sweep speed) knob expands or contracts the waveform about the time reference point (). See "To adjust the horizontal
(time/div) scale" on page 70.
Turning the Horizontal position ( ) knob in Normal mode (not Zoom) moves the trigger point indicator () to the left or right of the time reference point (). See
"To adjust the horizontal delay (position)" on page 71.

Searching for Events

You can use the [Search] key and menu to search for Edge, Pulse Width, Rise/Fall Time, Runt, Frequency Peaks, and Serial events on the analog channels.
Setting up searches (see "To set up searches" on page 80) is similar to setting up triggers. In fact, except for Frequency Peaks and Serial events, you can copy search setups to trigger setups and vice-versa (see "To copy search setups" on page 80).
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 79
2 Horizontal Controls

To set up searches

Searches are different than triggers in that they use the measurement threshold settings instead of trigger levels.
Found search events are marked with white triangles at the top of the graticule, and the number of events found is displayed in the menu line just above the softkey labels.
1 Press [Search]. 2 In the Search Menu, press Search; then, turn the Entry knob to select the search
type.
3 Use the remaining softkeys to set up the selected search type.
In most cases, setting up searches is similar to setting up triggers:
For setting up Edge searches, see "Edge Trigger" on page 180.
For setting up Pulse Width searches, see "Pulse Width Trigger" on
page 184.
For setting up Rise/Fall Time searches, see "Rise/Fall Time Trigger" on
page 191.
For setting up Runt searches, see "Runt Trigger" on page 194.
For setting up Frequency Peak searches, see "Searching for FFT Peaks" on
page 110.
For setting up Serial searches, see "Serial Trigger" on page 211 and
"Searching Lister Data" on page 159.
Remember that searches use the measurement threshold settings instead of trigger levels. Use the Thresholds softkey in the Search Menu to access the Measurement Threshold Menu. See "Measurement Thresholds" on page 279.

To copy search setups

Except for Frequency Peak and Serial event search setups, you can copy search setups to trigger setups and vice-versa.
1 Press [Search]. 2 In the Search Menu, press Search; then, turn the Entry knob to select the search
type.
3 Press Copy.
80 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
4 In the Search Copy Menu:
Press Copy to Trigger to copy the setup for the selected search type to the
same trigger type. For example, if the current search type is Pulse Width, pressing Copy to Trigger copies the search settings to the Pulse Width trigger settings and selects the Pulse Width trigger.
Press Copy from Trigger to copy the trigger setup for the selected search type
to the search setup.
To undo a copy, press Undo Copy.
The softkeys in the Search Copy Menu may not be available when one of the settings cannot be copied or there is no trigger type that corresponds to the search type.

Navigating the Time Base

You can use the [Navigate] key and controls to navigate through:
• Captured data (see "To navigate time" on page 81).
• Search events (see "To navigate search events" on page 82).
• Segments, when segmented memory acquisitions are turned on (see "To
navigate segments" on page 82).
Horizontal Controls 2
You can also access navigation controls on the touchscreen. See "Select Sidebar
Information or Controls" on page 54.

To navigate time

When acquisitions are stopped, you can use the navigation controls to play through the captured data.
1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Time.
3 Press the navigation keys to play backward, stop, or play forward in
time. You can press the or keys multiple times to speed up the playback. There are three speed levels.
You can also access navigation controls on the touchscreen. See "Select Sidebar
Information or Controls" on page 54.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 81
2 Horizontal Controls

To navigate search events

When acquisitions are stopped, you can use the navigation controls to go to found search events (set using the [Search] key and menu, see "Searching for Events" on page 79).
1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Search.
3 Press the back and forward keys to go to the previous or next search
event.
When searching Serial decode:
•The Auto zoom softkey specifies whether the waveform display is automatically zoomed to fit the marked row as you navigate.
• Pressing the Scroll Lister softkey lets you use the Entry knob to scroll through data rows in the Lister display.
You can also access navigation controls on the touchscreen. See "Select Sidebar
Information or Controls" on page 54.
82 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To navigate segments

When the segmented memory acquisition is enabled and acquisitions are stopped, you can use the navigation controls to play through the acquired segments.
1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Segments. 3 Press Play Mode; then, select:
Manual — to play through segments manually.
In the Manual play mode:
Press the back and forward keys to go to the previous or next
segment.
Press the softkey to go to the first segment.
Press the softkey to go to the last segment.
Auto — to play through segments in an automated fashion.
In the Auto play mode:
Horizontal Controls 2
Press the navigation keys to play backward, stop, or play
forward in time. You can press the or keys multiple times to speed up the playback. There are three speed levels.
You can also access navigation controls on the touchscreen. See "Select Sidebar
Information or Controls" on page 54.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 83
2 Horizontal Controls
84 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Keysight InfiniiVision 6000 X-Series Oscilloscopes
User's Guide

3 Vertical Controls

To turn waveforms on or off (channel or math) / 86 To adjust the vertical scale / 87 To adjust the vertical position / 87 To specify channel coupling / 88 To specify channel input impedance / 89 To specify bandwidth limiting / 89 To change the vertical scale knob's coarse/fine adjustment setting / 90 To invert a waveform / 91 Setting Analog Channel Probe Options / 91
The vertical controls include:
• Touchscreen controls for setting the vertical scale and position (offset) and
accessing the Channel menus (see "Pinch, Flick, or Drag to Scale, Position,
and Change Offset" on page 52 and "Turn Channels On/Off and Open Scale/Offset Dialogs" on page 61).
• The vertical scale and position knobs for each analog channel.
• The channel keys for turning a channel on or off and accessing the channel's softkey menu.
The following figure shows the Channel 1 Menu that appears after pressing the [1] channel key.
85
3 Vertical Controls
Channel, Volts/div
Channel 1 ground level
Trigger source
Trigger level or threshold
Channel 2 ground level
The ground level of the signal for each displayed analog channel is identified by the position of the icon at the far-left side of the display.

To turn waveforms on or off (channel or math)

You can:
• Use the touchscreen controls to turn channels on or off (see "Turn Channels
On/Off and Open Scale/Offset Dialogs" on page 61).
• Press an analog channel key turn the channel on or off (and to display the channel's menu).
When a channel is on, its key is illuminated.
86 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
NOTE
Turning channels off
You must be viewing the menu for a channel before you can turn it off. For example, if channel 1 and channel 2 are turned on and the menu for channel 2 is being displayed, to turn channel 1 off, press [1] to display the channel 1 menu; then, press [1] again to turn channel 1 off.

To adjust the vertical scale

You can:
• Use the touchscreen pinch gesture (see "Pinch, Flick, or Drag to Scale,
Position, and Change Offset" on page 52).
• Use the touchscreen controls to open a vertical scale/offset dialog (see "Turn
Channels On/Off and Open Scale/Offset Dialogs" on page 61).
• Turn the large knob above the channel key marked to set the vertical scale (volts/division) for the channel.
The vertical scale knob changes the analog channel scale in a 1-2-5 step sequence (with a 1:1 probe attached) unless fine adjustment is enabled (see "To
change the vertical scale knob's coarse/fine adjustment setting" on page 90).
Vertical Controls 3
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the volts/division knob is vertical expansion about the ground level of the channel; however, you can change this to expand about the center of the display. See "To choose "expand
about" center or ground" on page 383.

To adjust the vertical position

You can:
• Use the touchscreen pinch and drag gestures (see "Pinch, Flick, or Drag to
Scale, Position, and Change Offset" on page 52).
• Use the touchscreen controls to open a vertical scale/offset dialog (see "Turn
Channels On/Off and Open Scale/Offset Dialogs" on page 61).
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 87
3 Vertical Controls
TIP
• Turn the small vertical position knob ( ) to move the channel's waveform up or down on the display.
The offset voltage value represents the voltage difference between the vertical center of the display and the ground level ( ) icon. It also represents the voltage
at the vertical center of the display if vertical expansion is set to expand about ground (see "To choose "expand about" center or ground" on page 383).

To specify channel coupling

Coupling changes the channel's input coupling to either AC (alternating current) or DC (direct current).
If the channel is DC coupled, you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is removed, allowing you to use greater sensitivity to display the AC component of the signal.
1 Press the desired channel key. 2 In the Channel Menu, press the Coupling softkey to select the input channel
coupling:
DC — DC coupling is useful for viewing waveforms as low as 0 Hz that do not
have large DC offsets.
AC — AC coupling is useful for viewing waveforms with large DC offsets.
When AC coupling is chosen, you cannot select 50mode. This is done to prevent damage to the oscilloscope.
AC coupling places a 10 Hz high-pass filter in series with the input waveform that removes any DC offset voltage from the waveform.
Note that Channel Coupling is independent of Trigger Coupling. To change trigger coupling see "To select the trigger coupling" on page 217.
88 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To specify channel input impedance

NOTE
When you connect an AutoProbe, self-sensing probe, or a compatible InfiniiMax probe, the oscilloscope automatically configures the analog input channels to the correct impedance.
1 Press the desired channel key. 2 In the Channel Menu, press Imped (impedance); then, select either:
50 Ohm — matches 50 ohm cables commonly used in making high frequency
measurements, and 50 ohm active probes.
When 50 Ohm input impedance is selected, it is displayed with the channel information on-screen.
When AC coupling is selected (see "To specify channel coupling" on page 88) or excessive voltage is applied to the input, the oscilloscope automatically switches to 1M Ohm mode to prevent possible damage.
1M Ohm — is for use with many passive probes and for general-purpose
measurements. The higher impedance minimizes the loading effect of the oscilloscope on the device under test.
Vertical Controls 3
This impedance matching gives you the most accurate measurements because reflections are minimized along the signal path.
See Also • For more information on probing, visit: www.keysight.com/find/scope_probes
• Information about selecting a probe can be found in document number
Keysight Oscilloscope Probes and Accessories Selection Guide (part number
5989-6162EN), available at www.keysight.com.

To specify bandwidth limiting

1 Press the desired channel key. 2 In the Channel Menu, press the BW Limit softkey to specify the channel's
bandwidth limit or to turn off the bandwidth limit.
For waveforms with frequencies below the bandwidth limit, turning the bandwidth limit on removes unwanted high frequency noise from the waveform.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 89
3 Vertical Controls
The bandwidth limit also limits the trigger signal path of any channel that has BW Limit turned on.
The limits that can be selected depend on the bandwidth of oscilloscope, the channel's input impedance setting, and the channel's V/div setting:
Input Impedance V/div Oscilloscope
Bandwidth
1M Ohm >2 mV/
div
<=2 mV /div
50 Ohm (all) 1 GHz 20 MHz or 200 MHz
(all) 20 MHz
(all) 20 MHz or 200 MHz
2.5 GHz 20 MHz, 200 MHz, or 1.5 GHz
4 GHz 20 MHz, 200 MHz, 1.5 GHz, or 3 GHz
6 GHz 20 MHz, 200 MHz, 1.5 GHz, or 3 GHz
Limits Available

To change the vertical scale knob's coarse/fine adjustment setting

1 Push the channel's vertical scale knob (or press the channel key and then the
Fine softkey in the Channel Menu) to toggle between fine and coarse
adjustment of the vertical scale.
You can also do this using the touchscreen. See "Turn Channels On/Off and
Open Scale/Offset Dialogs" on page 61.
When Fine adjustment is selected, you can change the channel's vertical sensitivity in smaller increments. The channel sensitivity remains fully calibrated when Fine is on.
The vertical scale value is displayed in the status line at the top of the display.
When Fine is turned off, turning the volts/division knob changes the channel sensitivity in a 1-2-5 step sequence.
90 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To invert a waveform

1 Press the desired channel key. 2 In the Channel Menu, press the Invert softkey to invert the selected channel.
When Invert is selected, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed. However, when using basic triggers, the oscilloscope attempts to maintain the same trigger point by changing trigger settings.
Inverting a channel also changes the result of any math function selected in the Waveform Math Menu or any measurement.

Setting Analog Channel Probe Options

1 Press the probe's associated channel key. 2 In the Channel Menu, press the Probe softkey to display the Channel Probe
Menu.
This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe.
Vertical Controls 3
The Channel Probe Menu changes depending on the type of probe connected.
For passive probes, the Probe Check softkey appears; it guides you through the process of compensating probes.
For some active probes (such as InfiniiMax probes), the oscilloscope can accurately calibrate its analog channels for the probe. When you connect a probe that can be calibrated, the Calibrate Probe softkey appears (and the probe attenuation softkey may change). See "To calibrate a probe" on page 93.
See Also "To specify the channel units" on page 92
"To specify the probe attenuation" on page 92
"To specify the probe skew" on page 93
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 91
3 Vertical Controls

To specify the channel units

To specify the probe attenuation

1 Press the probe's associated channel key. 2 In the Channel Menu, press Probe. 3 In the Channel Probe Menu, press Units; then, select:
Volts — for a voltage probe.
Amps — for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions will reflect the measurement units you have selected.
This is set automatically if the oscilloscope can identify the connected probe. See Analog channel inputs (see page 47).
The probe attenuation factor must be set properly for accurate measurement results.
If you connect a probe that is not automatically identified by the oscilloscope, you can manually set the attenuation factor as follows:
1 Press the channel key. 2 Press the Probe softkey until you have selected how you want to specify the
attenuation factor, choosing either Ratio or Decibels.
3 Turn the Entry knob to set the attenuation factor for the connected probe.
When measuring voltage values, the attenuation factor can be set from 0.1:1 to 10000:1 in a 1-2-5 sequence.
When measuring current values with a current probe, the attenuation factor can be set from 10 V/A to 0.0001 V/A.
When specifying the attenuation factor in decibels, you can select values from
-20 dB to 80 dB.
If Amps is chosen as the units and a manual attenuation factor is chosen, then the units as well as the attenuation factor are displayed above the Probe softkey.
92 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To specify the probe skew

When measuring time intervals in the nanoseconds (ns) range, small differences in cable length can affect the measurement. Use Skew to remove cable-delay errors between any two channels.
1 Probe the same point with both probes. 2 Press one of the probes associated channel key. 3 In the Channel Menu, press Probe. 4 In the Channel Probe Menu, press Skew; then, select the desired skew value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a total of 200 ns difference.
The skew setting is not affected by pressing [Default Setup] or [Auto Scale].
Vertical Controls 3

To calibrate a probe

The Calibrate Probe softkey guides you through the process of calibrating probes.
For certain active probes, such as InfiniiMax probes, the oscilloscope can accurately calibrate its analog channels for the probe. When you connect a probe that can be calibrated, the Calibrate Probe softkey in the Channel Probe Menu becomes active.
To calibrate one of these probes: 1 First, plug your probe into one of the oscilloscope channels.
This could be, for example, an InfiniiMax probe amplifier/probe head with attenuators attached.
2 Connect the probe to the left side, Demo 2, Probe Comp terminal, and the
probe ground to the ground terminal.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 93
3 Vertical Controls
NOTE
When calibrating a differential probe, connect the positive lead to the Probe Comp terminal and the negative lead to the ground terminal. You may need to connect an alligator clip to the ground lug to allow a differential probe to span between the Probe Comp test point and ground. A good ground connection ensures the most accurate probe calibration.
3 Press the Channel on/off key to turn the channel on (if the channel is off). 4 In the Channel Menu, press the Probe softkey. 5 In the Channel Probe Menu, the second softkey from the left is for specifying
your probe head (and attenuation). Repeatedly press this softkey until the probe head selection matches the attenuator you are using.
The choices are:
10:1 single-ended browser (no attenuator).
10:1 differential browser (no attenuator).
10:1 (+6 dB Atten) single-ended browser.
10:1 (+6 dB Atten) differential browser.
10:1 (+12 dB Atten) single-ended browser.
10:1 (+12 dB Atten) differential browser.
10:1 (+20 dB Atten) single-ended browser.
10:1 (+20 dB Atten) differential browser.
6 Press the Calibrate Probe softkey and follow the instructions on the display.
For more information on InfiniiMax probes and accessories, see the probe's User's Guide.
94 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Keysight InfiniiVision 6000 X-Series Oscilloscopes
User's Guide

4 Math Waveforms

To display math waveforms / 95 To adjust the math waveform scale and offset / 97 Units for Math Waveforms / 97 Math Operators / 98 Math Transforms / 100 Math Filters / 119 Math Visualizations / 122 The Measurement Record and Waveform Math / 130
You can define and display up to four math functions. Math function waveforms display in shades of light purple.
Math functions can be performed on analog channels or they can be performed on lower math functions when using operators other than add, subtract, multiply, or divide.

To display math waveforms

1 Press the [Math] key on the front panel to display the Waveform Math Menu.
2 Press the Display Math softkey and turn the Entry knob to select the math
function you want to display. Then, either push the Entry knob or press the Display Math softkey again to display the selected math function.
95
4 Math Waveforms
TIP
TIP
3 Use the Operator softkey to select an operator, transform, filter, or visualization.
For more information on the operators, see:
"Math Operators" on page 98
"Math Transforms" on page 100
"Math Filters" on page 119
"Math Visualizations" on page 122
4 Use the Source 1 softkey to select the analog channel, lower math function, or
reference waveform on which to perform math. You can rotate the Entry knob or repetitively press the Source 1 softkey to make your selection.
Higher math functions can operate on lower math functions. For example, if Math 1 is set up as a subtract operation between channels 1 and 2, the Math 2 function could be set up as a FFT operation on the Math 1 function. These are called cascaded math functions.
To cascade math functions, select the lower math function using the Source 1 softkey.
When cascading math functions, to get the most accurate results, be sure to vertically scale lower math functions so that their waveforms take up the full screen without being clipped.
5 If you selected an arithmetic operator for the math function, use the Source 2
softkey to select the second source for the arithmetic operation.
6 To re-size and re-position the math waveform, see "To adjust the math
waveform scale and offset" on page 97.
Math Operating Hints
If the analog channel or math function is clipped (not fully displayed on screen) the resulting displayed math function will also be clipped.
Once the function is displayed, the analog channel(s) may be turned off for better viewing of the math waveform.
The vertical scaling and offset of each math function can be adjusted for ease of viewing and measurement considerations.
The math function waveform can be measured using [Cursors] and/or [Meas].
96 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

To adjust the math waveform scale and offset

NOTE
1 Make sure the multiplexed scale and position knobs above and below the [Math]
key are selected for the math waveform.
If the arrow to the left of the [Math] key is not illuminated, press the key.
2 Use the multiplexed scale and position knobs above and below the [Math] key to
re-size and re-position the math waveform.
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is automatically scaled for optimum vertical scale and offset. If you manually set scale and offset for a function, select a new function, then select the original function, the original function will be automatically rescaled.
See Also "Units for Math Waveforms" on page 97
Math Waveforms 4

Units for Math Waveforms

Units for each input channel can be set to Volts or Amps using the Units softkey in the channel's Probe Menu. Units for math function waveforms are:
Math function Units
add or subtract V or A
multiply
d/dt V/s or A/s (V/second or A/second)
dt
FFT dB* (decibels). See also "FFT Units" on
(square root)
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 97
2
, A2, or W (Volt-Amp)
V
Vs or As (V-seconds or A-seconds)
page 112.
1/2
1/2
, or W
1/2
(Volt-Amp)
, A
V
4 Math Waveforms
Math function Units
* When the FFT source is channel 1, 2, 3 or 4, FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 M. FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50. FFT units will be displayed as dB for all other FFT sources or when a source channel's units has been set to Amps.
A scale unit of U (undefined) will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved.

Math Operators

Math operators perform arithmetic operations (like add, subtract, or multiply) on analog input channels.
"Add or Subtract" on page 98
"Multiply or Divide" on page 99
98 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide

Add or Subtract

When you select add or subtract, the Source 1 and Source 2 values are added or subtracted point by point, and the result is displayed.
You can use subtract to make a differential measurement or to compare two waveforms.
If your waveforms' DC offsets are larger than the dynamic range of the oscilloscope's input channels you will need to use a differential probe instead.
Math Waveforms 4
Figure 5 Example of Subtract Channel 2 from Channel 1
See Also "Units for Math Waveforms" on page 97

Multiply or Divide

When you select the multiply or divide math function, the Source 1 and Source 2 values are multiplied or divided point by point, and the result is displayed.
The divide by zero case places holes (that is, zero values) in the output waveform.
Multiply is useful for seeing power relationships when one of the channels is proportional to the current.
Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide 99
4 Math Waveforms
Figure 6 Example of Multiply Channel 1 by Channel 2
See Also "Units for Math Waveforms" on page 97

Math Transforms

Math transforms perform a transform function (like differentiate, integrate, FFT, or square root) on an analog input channel or on the result of an arithmetic operation.
"Differentiate" on page 101
"Integrate" on page 102
"FFT Magnitude, FFT Phase" on page 105
"Square Root" on page 115
"Ax + B" on page 115
100 Keysight InfiniiVision 6000 X-Series Oscilloscopes User's Guide
Loading...