Keysight MSOX3012T, DSOX3024T, MSOX3014T, DSOX3014T, MSOX3022T User Manual

...
Page 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide
Page 2
CAUTION
© Keysight Technologies, Inc. 2005-2017
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
75037-97038
Edition
Fourth edition, April 2017
Available in electronic format only Published by:
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Print History
75037-97000, November 2014
75037-97013, August 2015
75037-97025, July 2016
75037-97038, April 2017
Warranty
The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, incl uding but not l imited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate
written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology License
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Rel inquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
2 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 3
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 3
Page 4

InfiniiVision 3000T X-Series Oscilloscopes—At a Glance

Table 1 InfiniiVision 3000T X-Series Oscilloscope Models, Bandwid ths, Sample Rates
Bandwid th 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Sample Rate (interleaved, non-interleaved)
2-Channel + 16 Logic Channels MSO
4-Channel + 16 Logic Channels MSO
2-Channel DSO DSO-X 3012T DSO-X 3022T DSO-X 3032T DSO-X 3052T DSO-X 3102T
4-Channel DSO DSO-X 3014T DSO-X 3024T DSO-X 3034T DSO-X 3054T DSO-X 3104T
5 GSa/s,
2.5 GSa/s
MSO-X 3012T MSO-X 3022T MSO-X 3032T MSO-X 3052T MSO-X 3102T
MSO-X 3014T MSO-X 3024T MSO-X 3034T MSO-X 3054T MSO-X 3104T
5 GSa/s,
2.5 GSa/s
5 GSa/s,
2.5 GSa/s
5 GSa/s,
2.5 GSa/s
5 GSa/s,
2.5 GSa/s
The Keysight InfiniiVision 3000T X-Series oscilloscopes deliver these features:
100 MHz, 200 MHz, 350 MHz, 500 MHz, and 1 GHz bandwidth models.
4 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 5
2- and 4-channel digital storage oscilloscope (DSO) models.
2+16-channel and 4+16-channel mixed-signal oscilloscope (MSO) models.
An MSO lets you debug your mixed-signal designs using analog signals and tightly correlated digital signals simultaneously. The 16 digital channels have a
1.25 GSa/s sample rate, with a 200 MHz toggle rate.
8.5 inch WVGA touchscreen display. The touchscreen makes the oscilloscope easier to use:
You can "touch" inside alpha-numeric keypad dialogs to enter file, label,
network, and printer names, etc., instead of using softkeys and the Entry knob.
You can drag a finger across the screen to draw rectangular boxes for
zooming in on waveforms or setting up Zone triggers.
You can touch the blue menu icon in the sidebar to display information or
control dialogs. You can drag (undock) these dialogs out of the sidebar, for example, to view cursor values and measurements at the same time.
You can touch other areas of the screen as substitutes for using front panel
keys, softkeys, and knobs.
Interleaved 4 Mpts or non-interleaved 2 Mpts MegaZoom IV memory for the fastest waveform update rates, uncompromised.
All knobs are pushable for making quick selections.
Trigger types: edge, edge then edge, pulse width, pattern, OR, rise/fall time, Nth edge burst, runt, setup & hold, video, NFC, and zone.
Serial decode/trigger options for: CAN/CAN FD/LIN, CXPI, FlexRay, I
2
I
S, Manchester/NRZ, MIL-STD-1553/ARINC 429, SENT, and UART/RS232.
There is a Lister for displaying serial decode packets.
Dedicated [FFT] key and FFT waveform math function.
Two additional math waveforms: add, subtract, multiply, divide, d/dt, integrate, FFT (Magnitude), FFT (Phase), Ax+B, square, square root, absolute value, common logarithm, natural logarithm, exponential, base 10 exponential, low pass filter, high pass filter, averaged value, smoothing, envelope, magnify, maximum, minimum, peak-peak, max hold, min hold, measurement trend, chart logic bus timing, and chart logic bus state.
Reference waveform locations (2) for comparing with other channel or math waveforms.
Many built-in measurements and a measurement statistics display.
2
C/SPI,
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 5
Page 6
Built-in, license-enabled 1-channel waveform generator with: arbitrary, sine, square, ramp, pulse, DC, noise, sine cardinal, exponential rise, exponential fall, cardiac, and Gaussian pulse. Modulated waveforms on WaveGen except for arbitrary, pulse, DC, and noise waveforms.
USB ports make printing, saving and sharing data easy.
Optional LAN/VGA module for connecting to a network and displaying the screen on a different monitor.
Optional GPIB module.
A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help. Complete instructions for using the quick help system are given in “Access the Built-In Quick Help" on page 61.
For more information about InfiniiVision oscilloscopes, see:
www.keysight.com/find/scope
6 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 7

In This Guide

This guide shows how to use the InfiniiVision 3000T X-Series oscilloscopes.
When unpacking and using the oscilloscope for the first time, see:
When displaying waveforms and acquired data, see:
When setting up triggers or changing how data is acquired, see:
Making measurements and analyzing data:
Chapter 1, “Getting Started,” starting on page 27
Chapter 2, “Horizontal Controls,” starting on page 63
Chapter 3, “Vertical Controls,” starting on page 79
Chapter 4, “FFT Spectral Analysis,” starting on page 89
Chapter 5, “Math Waveforms,” starting on page 99
Chapter 6, “Reference Waveforms,” starting on page
127
Chapter 7, “Digital Channels,” starting on page 131
Chapter 8, “Serial Decode,” starting on page 149
Chapter 9, “Display Settings,” starting on page 155
Chapter 10, “Labels,” starting on page 161
Chapter 11, “Triggers,” starting on page 169
Chapter 12, “Trigger Mode/Coupling,” starting on
page 207
Chapter 13, “Acquisition Control,” starting on page
215
Chapter 14, “Cursors,” starting on page 233
Chapter 15, “Measurements,” starting on page 243
Chapter 16, “Mask Testing,” starting on page 275
Chapter 17, “Digital Voltmeter and Counter,” starting
on page 289
Chapter 18, “Frequency Response Analysis,” starting
on page 295
When using the built-in license enabled waveform generator, see:
When saving, recalling, or printing, see:
Chapter 19, “Waveform Generator,” starting on page
301
Chapter 20, “Save/Email/Recall (Setups, Screens,
Data),” starting on page 321
Chapter 21, “Print (Screens),” starting on page 335
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 7
Page 8
When using the oscilloscope's utility
NOTE
functions or web interface, see:
For reference information, see: Chapter 24, “Reference,” starting on page 375
Chapter 22, “Utility Settings,” starting on page 341
Chapter 23, “Web Interface,” starting on page 359
When using licensed serial bus triggering and decode features, see:
Chapter 25, “CAN/LIN Triggering and Serial Decode,”
starting on page 395
Chapter 26, “CXPI Triggering and Serial Decode,”
starting on page 417
Chapter 27, “FlexRay Triggering and Serial Decode,”
starting on page 427
Chapter 28, “I2C/SPI Triggering and Serial Decode,”
starting on page 437
Chapter 29, “I2S Triggering and Serial Decode,”
starting on page 455
Chapter 30, “Manchester/NRZ Triggering and Serial
Decode,” starting on page 465
Chapter 31, “MIL-STD-1553/ARINC 429 Triggering
and Serial Decode,” starting on page 481
Chapter 32, “SENT Triggering and Serial Decode,”
starting on page 497
Chapter 33, “UART/RS232 Triggering and Serial
Decode,” starting on page 511
Abbreviated instructions for pressing a series of keys and softkeys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:
Press [Key1] > Softkey2 > Softkey3.
The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display.
8 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 9

Contents

InfiniiVision 3000T X-Series Oscilloscopes—At a Glance / 4
In This Guide / 7
1Getting Started
Inspect the Package Contents / 27
Install the Optional LAN/VGA or GPIB Module / 30
Tilt the Oscilloscope for Easy Viewing / 30
Power-On the Oscilloscope / 31
Connect Probes to the Oscilloscope / 32
Maximum input voltage at analog inputs / 32
Do not float the oscilloscope chassis / 33
Input a Waveform / 33
Recall the Default Oscilloscope Setup / 33
Use Autoscale / 34
Compensate Passive Probes / 36
Learn the Front Panel Controls and Connectors / 37
Front Panel Overlays for Different Languages / 44
Learn the Touchscreen Controls / 45
Draw Rectangles for Waveform Zoom or Zone Trigger Set
Up / 46 Flick or Drag to Scale, Position, and Change Offset / 47 Select Sidebar Information or Controls / 49
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 9
Page 10
Undock Sidebar Dialogs by Dragging / 49 Re-dock Sidebar Dialogs to Split Sidebar / 50 Select Dialog Menus and Close Dialogs / 51 Drag Cursors / 51 Touch Softkeys and Menus On the Screen / 51 Enter Names Using Alpha-Numeric Keypad Dialogs / 52 Change Waveform Offsets By Dragging Ground Reference
Icons / 53 Access Controls and Menus Via the Menu Icon / 54 Turn Channels On/Off and Open Scale/Offset Dialogs / 55 Access the Horizontal Menu and Open the Scale/Delay
Dialog / 56 Access the Trigger Menu, Change the Trigger Mode, and Open the
Trigger Level Dialog / 57 Use a USB Mouse and/or Keyboard for Touchscreen
Controls / 58
Learn the Rear Panel Connectors / 58
Learn the Oscilloscope Display / 59
Access the Built-In Quick Help / 61
2Horizontal Controls
To adjust the horizontal (time/div) scale / 64
To adjust the horizontal delay (position) / 65
Panning and Zooming Single or Stopped Acquisitions / 66
To change the horizontal time mode (Normal, XY, or Roll) / 67
XY Time Mode / 68
To display the zoomed time base / 71
To change the horizontal scale knob's coarse/fine adjustment
setting / 72
To position the time reference (left, center, right, custom) / 72
10 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 11
Searching for Events / 73
Navigating the Time Base / 75
3Vertical Controls
To turn waveforms on or off (channel or math) / 80
To adjust the vertical scale / 81
To adjust the vertical position / 81
To specify channel coupling / 81
To specify channel input impedance / 82
To specify bandwid th limiting / 83
To change the vertical scale knob's coarse/fine adjustment
To set up searches / 74 To copy search setups / 74
To navigate time / 75 To navigate search events / 75 To navigate segments / 76
setting / 83
To invert a waveform / 84
Setting Analog Channel Probe Options / 84
To specify the channel units / 85 To specify the probe attenuation / 85 To specify the probe skew / 86 To calibrate a probe / 86
4 FFT Spectral Analysis
Searching for FFT Peaks / 93
FFT Measurement Hints / 93
FFT Units / 95
FFT DC Value / 95
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 11
Page 12
FFT Aliasing / 95
FFT Spectral Leakage / 97
5Math Waveforms
To display math waveforms / 99
To adjust the math waveform scale and offset / 101
Units for Math Waveforms / 101
Math Operators / 102
Math Transforms / 104
Add or Subtract / 102 Multiply or Divide / 103
Differentiate / 105 Integrate / 106 FFT Magnitude, FFT Phase / 108 Square Root / 113 Ax + B / 114 Square / 115 Absolute Value / 115 Common Logarithm / 116 Natural Logarithm / 116 Exponential / 117 Base 10 Exponential / 117
Math Filters / 118
High Pass and Low Pass Filter / 118 Averaged Value / 119 Smoothing / 120 Envelope / 120
Math Visualizations / 121
Magnify / 121 Maximum/Minimum / 122
12 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 13
Peak-Peak / 122 Max/Min Hold / 122 Measurement Trend / 122 Chart Logic Bus Timing / 124 Chart Logic Bus State / 125
6 Reference Waveforms
To save a waveform to a reference waveform location / 127
To display a reference waveform / 128
To scale and position reference waveforms / 129
To adjust reference waveform skew / 129
To display reference waveform information / 130
To save/recall reference waveform files to/from a USB storage
device / 130
7 Digital Channels
To connect the digital probes to the device under test / 131
Probe cable for digital channels / 132
Acquiring waveforms using the digital channels / 135
To display digital channels using Autoscale / 135
Interpreting the digital waveform display / 136
To change the displayed size of the digital channels / 137
To switch a single channel on or off / 138
To switch all digital channels on or off / 138
To switch groups of channels on or off / 138
To change the logic threshold for digital channels / 138
To reposition a digital channel / 139
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 13
Page 14
8Serial Decode
9 Display Settings
To display digital channels as a bus / 140
Digital channel signal fidelity: Probe impedance and
grounding / 143 Input Impedance / 143
Probe Grounding / 145 Best Probing Practices / 147
Serial Decode Options / 149
Lister / 151
Searching Lister Data / 152
To adjust waveform intensity / 155
To set or clear persistence / 157
To clear the display / 158
To select the grid type / 158
To adjust the grid intensity / 159
To freeze the display / 159
10 Labels
To turn the label display on or off / 161
To assign a predefined label to a channel / 162
To define a new label / 163
To load a list of labels from a text file you create / 164
To reset the label library to the factory default / 165
To add an annotation / 166
14 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 15
11 Triggers
Adjusting the Trigger Level / 170
Forcing a Trigger / 171
Edge Trigger / 172
Edge then Edge Trigger / 174
Pulse Width Trigger / 175
Pattern Trigger / 178
Hex Bus Pattern Trigger / 180
OR Trigger / 181
Rise/Fall Time Trigger / 182
Near Field Communication (NFC) Trigger / 183
Nth Edge Burst Trigger / 186
Runt Trigger / 188
Setup and Hold Trigger / 190
Video Trigger / 191
To set up Generic video triggers / 195 To trigger on a specific line of video / 196 To trigger on all sync pulses / 197 To trigger on a specific field of the video signal / 198 To trigger on all fields of the video signal / 199 To trigger on odd or even fields / 200
Serial Trigger / 203
Zone Qualified Trigger / 204
12 Trigger Mode/Coupling
To select the Auto or Normal trigger mode / 208
To select the trigger coupling / 209
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 15
Page 16
To enable or disable trigger noise rejection / 211
To enable or disable trigger HF Reject / 211
To set the trigger holdoff / 212
External Trigger Input / 213
13 Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run
Overview of Sampling / 217
Selecting the Acquisition Mode / 221
Maximum voltage at oscilloscope external trigger input / 213
Control) / 215
Sampling Theory / 217 Aliasing / 217 Oscilloscope Bandwidth and Sample Rate / 218 Oscilloscope Rise Time / 219 Oscilloscope Bandwidth Required / 220 Memory Depth and Sample Rate / 221
Normal Acquisition Mode / 222 Peak Detect Acquisition Mode / 222 Averaging Acquisition Mode / 225 High Resolution Acquisition Mode / 227
Realtime Sampling Option / 227
Realtime Sampling and Oscilloscope Bandwid th / 228
Acquiring to Segmented Memory / 229
Navigating Segments / 230 Measurements, Statistics, and Infinite Persistence with
Segmented Memory / 231 Segmented Memory Re-Arm Time / 231 Saving Data from Segmented Memory / 231
16 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 17
14 Cursors
15 Measurements
To make cursor measurements / 234
Cursor Examples / 237
To make automatic measurements / 244
To edit measurements / 246
Measurements Summary / 246
Snapshot All / 249
Voltage Measurements / 250
Peak-Peak / 251 Maximum / 251 Minimum / 251 Amplitude / 251 Top / 251 Base / 252 Overshoot / 252 Preshoot / 254 Average / 254 DC RMS / 255 AC RMS / 255 Ratio / 257
Time Measurements / 257
Period / 258 Frequency / 258 Counter / 259 + Width / 260 – Width / 260 Burst Width / 260 Duty Cycle / 260 Bit Rate / 261
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 17
Page 18
Rise Time / 261 Fall Time / 261 Delay / 261 Phase / 262 X at Min Y / 264 X at Max Y / 264
Count Measurements / 265
Positive Pulse Count / 265 Negative Pulse Count / 265 Rising Edge Count / 266 Falling Edges Count / 266
Mixed Measurements / 266
Area / 266
FFT Analysis Measurements / 267
Channel Power / 267 Occupied Bandwidth / 268 Adjacent Channel Power Ratio (ACPR) / 268 Total Harmonic Distortion (THD) / 269
Measurement Thresholds / 270
Measurement Window / 272
Measurement Statistics / 272
16 Mask Testing
To create a mask from a "golden" waveform (Automask) / 275
Mask Test Setup Options / 277
Mask Statistics / 280
To manually modify a mask file / 281
Building a Mask File / 284
How is mask testing done? / 287
18 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 19
17 Digital Voltmeter and Counter
Digital Voltmeter / 290
Counter / 291
18 Frequency Response Analysis
To make connections / 295
To set up and run the analysis / 296
To view and save the analysis results / 298
19 Waveform Generator
To select generated waveform types and settings / 301
To edit arbitrary waveforms / 305
Creating New Arbitrary Waveforms / 306 Editing Existing Arbitrary Waveforms / 307 Capturing Other Waveforms to the Arbitrary Waveform / 311
To output the waveform generator sync pulse / 312
To specify the expected output load / 312
To use waveform generator logic presets / 313
To add noise to the waveform generator output / 313
To add modulation to the waveform generator output / 314
To set up Amplitude Modulation (AM) / 315 To set up Frequency Modulation (FM) / 316 To set up Frequency-Shift Keying Modulation (FSK) / 317
To restore waveform generator defaults / 318
20 Save/Email/Recall (Setups, Screens, Data)
Saving Setups, Screen Images, or Data / 321
To save setup files / 323 To save BMP or PNG image files / 324
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 19
Page 20
21 Print (Screens)
To save CSV, ASCII XY, or BIN data files / 324 Length Control / 326 To save Lister data files / 327 To save reference waveform files to a USB storage device / 327 To save masks / 328 To save arbitrary waveforms / 328 To navigate storage locations / 329 To enter file names / 329
Emailing Setups, Screen Images, or Data / 330
Recalling Setups, Masks, or Data / 331
To recall setup files / 332 To recall mask files / 332 To recall reference waveform files from a USB storage
device / 332 To recall arbitrary waveforms / 333
Recalling Default Setups / 333
Performing a Secure Erase / 334
To print the oscilloscope's display / 335
To set up network printer connections / 337
To specify the print options / 338
To specify the palette option / 339
22 Utility Settings
I/O Interface Settings / 341
Setting up the Oscilloscope's LAN Connection / 342
To establish a LAN connection / 343 Stand-alone (Point-to-Point) Connection to a PC / 344
File Explorer / 344
20 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 21
Setting Oscilloscope Preferences / 346
To choose "expand about" center or ground / 347 To disable/enable transparent backgrounds / 347 To load the default label library / 347 To set up the screen saver / 348 To set Autoscale preferences / 349
Setting the Oscilloscope's Clock / 349
Setting the Rear Panel TRIG OUT Source / 350
Enabling Remote Command Logging / 351
Performing Service Tasks / 352
To perform user calibration / 352 To perform hardware self test / 355 To perform front panel self test / 356 To display oscilloscope information / 356 To display the user calibration status / 356 To clean the oscilloscope / 356 To check warranty and extended services status / 356 To contact Keysight / 357 To return the instrument / 357
Configuring the [Quick Action] Key / 357
23 Web Interface
Accessing the Web Interface / 360
Browser Web Control / 361
Full Scope Remote Front Panel / 362 Screen Only Remote Front Panel / 363 Tablet Remote Front Panel / 364 Remote Programming via the Web Interface / 365 Remote Programming with Keysight IO Libraries / 366
Save/Recall / 367
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 21
Page 22
24 Reference
Saving Files via the Web Interface / 367 Recalling Files via the Web Interface / 368
Get Image / 369
Identification Function / 370
Instrument Utilities / 370
Setting a Password / 372
Specifications and Characteristics / 375
Measurement Category / 375
Oscilloscope Measurement Category / 375 Measurement Category Definitions / 376 Maximum Input Voltages / 376
Maximum input voltage at analog inputs / 376
Maximum input voltage at digital channels / 377
Environmental Conditions / 377
Probes and Accessories / 378
Loading Licenses and Displaying License Information / 378
Licensed Options Available / 379 Other Options Available / 380 Upgrading to an MSO / 380
Software and Firmware Updates / 380
Binary Data (.bin) Format / 381
Binary Data in MATLAB / 382 Binary Header Format / 382 Example Program for Reading Binary Data / 385 Examples of Binary Files / 385
22 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 23
CSV and ASCII XY files / 388
CSV and ASCII XY file structure / 389 Minimum and Maximum Values in CSV Files / 389 Frequency Data in CSV Files / 390
Acknowledgements / 390
Product Markings and Regulatory Information / 392
25 CAN/LIN Triggering and Serial Decode
Setup for CAN/CAN FD Signals / 395
Loading and Displaying CAN Symbolic Data / 398
CAN/CAN FD Triggering / 399
CAN/CAN FD Serial Decode / 402
Interpreting CAN/CAN FD Decode / 403 CAN Totalizer / 404 Interpreting CAN Lister Data / 405 Searching for CAN Data in the Lister / 406
Setup for LIN Signals / 407
Loading and Displaying LIN Symbolic Data / 409
LIN Triggering / 409
LIN Serial Decode / 411
Interpreting LIN Decode / 413 Interpreting LIN Lister Data / 414 Searching for LIN Data in the Lister / 414
26 CXPI Triggering and Serial Decode
Setup for CXPI Signals / 417
CXPI Triggering / 418
CXPI Serial Decode / 421
Interpreting CXPI Decode / 423
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 23
Page 24
Interpreting CXPI Lister Data / 425
27 FlexRay Triggering and Serial Decode
Setup for FlexRay Signals / 427
FlexRay Triggering / 428
Triggering on FlexRay Frames / 429 Triggering on FlexRay Errors / 430 Triggering on FlexRay Events / 430
FlexRay Serial Decode / 431
Interpreting FlexRay Decode / 432 FlexRay Totalizer / 433 Interpreting FlexRay Lister Data / 434 Searching for FlexRay Data in the Lister / 434
28 I2C/SPI Triggering and Serial Decode
Setup for I2C Signals / 437
I2C Triggering / 438
I2C Serial Decode / 442
Interpreting I2C Decode / 443 Interpreting I2C Lister Data / 444 Searching for I2C Data in the Lister / 444
Setup for SPI Signals / 445
SPI Triggering / 449
SPI Serial Decode / 451
Interpreting SPI Decode / 452 Interpreting SPI Lister Data / 453 Searching for SPI Data in the Lister / 453
29 I2S Triggering and Serial Decode
Setup for I2S Signals / 455
24 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 25
I2S Triggering / 458
I2S Serial Decode / 461
Interpreting I2S Decode / 462 Interpreting I2S Lister Data / 463 Searching for I2S Data in the Lister / 463
30 Manchester/NRZ Triggering and Serial Decode
Setup for Manchester Signals / 465
Manchester Triggering / 468
Manchester Serial Decode / 470
Interpreting Manchester Decode / 470 Interpreting Manchester Lister Data / 472
Setup for NRZ Signals / 473
NRZ Triggering / 475
NRZ Serial Decode / 477
Interpreting NRZ Decode / 478 Interpreting NRZ Lister Data / 479
31 MIL-STD-1553/ARINC 429 Triggering and Serial Decode
Setup for MIL-STD-1553 Signals / 481
MIL-STD-1553 Triggering / 483
MIL-STD-1553 Serial Decode / 484
Interpreting MIL-STD-1553 Decode / 485 Interpreting MIL-STD-1553 Lister Data / 486 Searching for MIL-STD-1553 Data in the Lister / 487
Setup for ARINC 429 Signals / 488
ARINC 429 Triggering / 489
ARINC 429 Serial Decode / 491
Interpreting ARINC 429 Decode / 493
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 25
Page 26
ARINC 429 Totalizer / 494 Interpreting ARINC 429 Lister Data / 495 Searching for ARINC 429 Data in the Lister / 495
32 SENT Triggering and Serial Decode
Setup for SENT Signals / 497
SENT Triggering / 502
SENT Serial Decode / 504
Interpreting SENT Decode / 505 Interpreting SENT Lister Data / 507 Searching for SENT Data in the Lister / 508
33 UART/RS232 Triggering and Serial Decode
Setup for UART/RS232 Signals / 511
UART/RS232 Triggering / 513
UART/RS232 Serial Decode / 515
Interpreting UART/RS232 Decode / 516 UART/RS232 Totalizer / 517 Interpreting UART/RS232 Lister Data / 518 Searching for UART/RS232 Data in the Lister / 518
Index
26 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 27
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide

1 Getting Started

Inspect the Package Contents / 27 Tilt the Oscilloscope for Easy Viewing / 30 Power-On the Oscilloscope / 31 Connect Probes to the Oscilloscope / 32 Input a Waveform / 33 Recall the Default Oscilloscope Setup / 33 Use Autoscale / 34 Compensate Passive Probes / 36 Learn the Front Panel Controls and Connectors / 37 Learn the Touchscreen Controls / 45 Learn the Rear Panel Connectors / 58 Learn the Oscilloscope Display / 59 Access the Built-In Quick Help / 61
This chapter describes the steps you take when using the oscilloscope for the first time.

Inspect the Package Contents

• Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
27
Page 28
1 Getting Started
• Verify that you received the following items and any optional accessories you may have ordered:
InfiniiVision 3000T X-Series oscilloscope.
Power cord (country of origin determines specific type).
Oscilloscope probes:
Two probes for 2-channel models.
Four probes for 4-channel models.
Digital probe kit (MSO models only).
Documentation CD-ROM.
28 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 29
o
Documentation CD
Getting Started 1
InfiniiVision 3000T X-Series oscilloscope
N2843A probes (Qty 2 or 4)
Power cord (Based on country
f origin)
N2756-60001 Digital Probe Kit (MSO models only)
See Also “Probes and Accessories" on page 378
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 29
Page 30
1 Getting Started
NOTE

Install the Optional LAN/VGA or GPIB Module

If you need to install a DSOXLAN LAN/VGA module or a DSOXGPIB GPIB module, perform this installation before you power on the oscilloscope.
1 If you need to remove a module before installing a different module, pinch the
module's spring tabs, and gently remove the module from the slot.
2 To install a module, slide the module into the slot on the back until it is fully
seated.
The module's spring tabs will latch into the slot, keeping the module in place.
LAN/VGA Module
Module Slot
GPIB Module
WARNING: MAIN TAIN GROUND TO AVOID ELECTRIC SHOCK
100-120V, 50/60/400Hz
~
100-240V, 50/60Hz
~
100 Watts MAX
The LAN/VGA or GPIB module must be installed before powering on the oscilloscope.

Tilt the Oscilloscope for Easy Viewing

There are tabs under the oscilloscope's front feet that can be flipped out to tilt the oscilloscope.
30 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 31

Power-On the Oscilloscope

Flip-Out Tabs
Getting Started 1
Power
Requirements
Ventilation
Requirements
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 31
Line voltage, frequency, and power:
• ~Line 100-120 Vac, 50/60/400 Hz
• 100-240 Vac, 50/60 Hz
• 100 W max
The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is required for proper cooling. Always ensure that the air intake and exhaust areas are free from obstructions.
The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscope.
When using the oscilloscope in a bench-top setting, provide at least 2" clearance at the sides and 4" (100 mm) clearance above and behind the oscilloscope for proper cooling.
Page 32
1 Getting Started
WARNING
CAUTION
CAUTION
To power-on the
oscilloscope
1 Connect the power cord to the rear of the oscilloscope, then to a suitable AC
voltage source. Route the power cord so the oscilloscope's feet and legs do not pinch the cord.
2 The oscilloscope automatically adjusts for input line voltages in the range 100
to 240 VAC. The line cord provided is matched to the country of origin.
Always use a grounded power cord. Do not defeat the power cord ground.
3 Press the power switch.
The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds.

Connect Probes to the Oscilloscope

1 Connect the oscilloscope probe to an oscilloscope channel BNC connector. 2 Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point on the circuit.
32 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Maximum input voltage at analog inputs
135 Vrms
Ω input: 5 Vrms Input protection is enabled in 50 Ω mode and the 50 Ω load will
50 disconnect if greater than 5 Vrms is detected. However the inputs could still be damaged, depending on the time constant of the signal. The 50 when the oscilloscope is powered on.
When measuring voltages over 30 V, use a 10:1 probe.
Ω input protection only functions
Page 33
CAUTION
Do not float the oscilloscope chassis
WARNING
Defeating the ground connection and "floating" the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage. The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to measure between two live points, use a differential probe with sufficient dynamic range.
Do not negate the protective action of the ground connection to the oscilloscope. The oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard.

Input a Waveform

The first signal to input to the oscilloscope is the Demo 2, Probe Comp signal. This signal is used for compensating probes.
1 Connect an oscilloscope probe from channel 1 to the Demo 2 (Probe Comp)
terminal on the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the Demo 2
terminal).
Getting Started 1

Recall the Default Oscilloscope Setup

Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 33
To recall the default oscilloscope setup: 1 Press [Default Setup].
The default setup restores the oscilloscope's default settings. This places the oscilloscope in a known operating condition. The major default settings are:
Page 34
1 Getting Started
Table 2 Default Configuration Settings
Horizontal Normal mode, 100 µs/div scale, 0 s delay, center time reference.

Use Autoscale

Vertical (Analog)
Trigger Edge trigger, Auto trigger mode, 0 V level, channel 1 source, DC coupling, rising
Display Persistence off, 20% grid intensity, 50% waveform intensity.
Other Acquire mode normal, [Run/Stop] to Run, cursors and measurements off.
Labels All custom labels that you have created in the Label Library are preserved (not
In the Save/Recall Menu, there are also options for restoring the complete factory settings (see “Recalling Default Setups" on page 333) or performing a secure erase (see “Performing a Secure Erase" on page 334).
Use [Auto Scale] to automatically configure the oscilloscope to best display the input signals.
1 Press [Auto Scale].
You should see a waveform on the oscilloscope's display similar to this:
Channel 1 on, 5 V/div scale, DC coupling, 0 V position, 1 M
edge slope, 40 ns holdoff time.
erased), but all channel labels will be set to their original names.
Ω impedance.
34 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 35
Getting Started 1
2 If you want to return to the oscilloscope settings that existed before, press Undo
Autoscale.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, press Fast Debug, Channels, or Acq Mode.
These are the same softkeys that appear in the Autoscale Preferences Menu. See “To set Autoscale preferences" on page 349.
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure “Compensate Passive Probes" on page 36.
If you do not see the waveform, make sure the probe is connected securely to the front panel channel input BNC and to the left side, Demo 2, Probe Comp terminal.
How Autoscale
Works
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 35
Autoscale analyzes any waveforms present at each channel and at the external trigger input. This includes the digital channels, if connected.
Autoscale finds, turns on, and scales any channel with a repetitive waveform that has a frequency of at least 25 Hz, a duty cycle greater than 0.5%, and an amplitude of at least 10 mV peak-to-peak. Any channels where no signal is found are turned off.
Page 36
1 Getting Started
The trigger source is selected by looking for the first valid waveform starting with external trigger, then continuing with the lowest number analog channel up to the highest number analog channel, and finally (if digital probes are connected) the highest number digital channel.
During Autoscale, the delay is set to 0.0 seconds, the horizontal time/div (sweep speed) setting is a function of the input signal (about 2 periods of the triggered signal on the screen), and the triggering mode is set to Edge.

Compensate Passive Probes

Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected. A poorly compensated probe can introduce significant measurement errors.
1 Input the Probe Comp signal (see “Input a Waveform" on page 33). 2 Press [Default Setup] to recall the default oscilloscope setup (see “Recall the
Default Oscilloscope Setup" on page 33).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
Comp signal (see “Use Autoscale" on page 34).
4 Press the channel key to which the probe is connected ([1], [2], etc.). 5 In the Channel Menu, press Probe. 6 In the Channel Probe Menu, press Probe Check; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the trimmer capacitor on the probe for the flattest pulse possible.
On N2894A probes, the trimmer capacitor is located on the probe BNC connector.
36 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 37
7 Connect probes to all other oscilloscope channels (channel 2 of a 2-channel
Perfectly compensated
Over compensated
Under compensated
oscilloscope, or channels 2, 3, and 4 of a 4-channel oscilloscope).
8 Repeat the procedure for each channel.

Learn the Front Panel Controls and Connectors

On the front panel, key refers to any key (button) you can press.
Getting Started 1
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 37
Softkey specifically refers to the six keys that are directly below the display. The legend for these keys is directly above them, on the display. Their functions change as you navigate through the oscilloscope's menus.
For the following figure, refer to the numbered descriptions in the table that follows.
Page 38
1 Getting Started
14. Tools keys
1. Power switch
2. Softkeys
3. [Intensity] key
4. Entry knob
6. Trigger controls
5. Waveform keys
18. Demo 2, Ground, and Demo 1
terminals
17. Analog channel
inputs
19. USB Host
port
15. [Help] key
13. File keys
8. Run Control keys
12. Measure controls
11. Additional waveform controls
7. Horizontal controls
10. [Auto Scale] key
9. [Default Setup] key
16. Vertical controls
21. Waveform generator
output
20. Digital channel
inputs
Back
Back
1. Power switch Press once to switch power on; press again to switch power off. See “Power-On the
2. Softkeys The functions of these keys change based upon the menus shown on the display directly
3. [Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to adjust waveform
38 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Oscilloscope" on page 31.
above the keys.
The Back/Up key moves up in the softkey menu hierarchy. At the top of the hierarchy, the
Back/Up key turns the menus off, and oscilloscope information is shown instead.
intensity.
You can vary the intensity control to bring out signal detail, much like an analog oscilloscope.
Digital channel waveform intensity is not adjustable.
More details about using the Intensity control to view signal detail are on “To adjust
waveform intensity" on page 155.
Page 39
Getting Started 1
4. Entry knob The Entry knob is used to select items from menus and to change values. The function of the Entry knob changes based upon the current menu and softkey selections.
Note that the curved arrow symbol above the entry knob illuminates whenever the entry
knob can be used to select a value. Also, note that when the Entry knob symbol appears on a softkey, you can use the Entry knob, to select values.
Often, rotating the Entry knob is enough to make a selection. Sometimes, you can push the Entry knob to enable or disable a selection. Pushing the Entry knob also makes popup menus disappear.
5. Waveform keys The [Acquire] key lets you select Normal, Peak Detect, Averaging, or High Resolution acquisition modes (see “Selecting the Acquisition Mode" on page 221) and use segmented memory (see “Acquiring to Segmented Memory" on page 229).
The [Display] key lets you access the menu where you can enable persistence (see “To set or
clear persistence" on page 157), clear the display, and adjust the display grid (graticule)
intensity (see “To ad just the grid intensity" on page 159).
[Touch] key (to right of channel 2 key) — Press this key to disable/enable the touchscreen.
6. Trigger controls These controls determine how the oscilloscope triggers to capture data. See Chapter 11, “Triggers,” starting on page 169 and Chapter 12, “Trigger Mode/Coupling,” starting on page
207.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 39
Page 40
1 Getting Started
7. Horizontal controls The Horizontal controls consist of:
Horizontal scale knob — Turn the knob in the Horizontal section that is marked
to adjust the time/div (sweep speed) setting. The symbols under the knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale.
Horizontal position knob — Turn the knob marked to pan through the waveform data
horizontally. You can see the captured waveform before the trigger (turn the knob clockwise) or after the trigger (turn the knob counterclockwise). If you pan through the waveform when the oscilloscope is stopped (not in Run mode) then you are looking at the waveform data from the last acquisition taken.
•[Horiz] key — Press this key to open the Horizontal Menu where you can select XY and Roll
modes, enable or disable Zoom, enable or disable horizontal time/d ivision fine adjustment, and select the trigger time reference point.
Zoom key — Press the zoom key to split the oscilloscope display into Normal and
Zoom sections without opening the Horizontal Menu.
•[Search] key — Lets you search for events in the acquired data.
•[Navigate] keys — Press these keys to navigate through captured data via time, search
events, or segmented memory acquisition. See “Navigating the Time Base" on page 75.
For more information see Chapter 2, “Horizontal Controls,” starting on page 63.
8. Run Control keys When the [Run/Stop] key is green, the oscilloscope is running, that is, acquiring data when trigger conditions are met. To stop acquiring data, press [Run/Stop].
When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring data, press [Run/Stop].
To capture and display a single acquisition (whether the oscilloscope is running or stopped), press [Single]. The [Single] key is yellow until the oscilloscope triggers.
For more information, see “Running, Stopping, and Making Single Acquisitions (Run
Control)" on page 215.
9. [Default Setup] key Press this key to restore the oscilloscope's default settings (details on “Recall the Defaul t
Oscilloscope Setup" on page 33).
10. [Auto Scale] key When you press the [Auto Scale] key, the oscilloscope will quickly determine which channels have activity, and it will turn these channels on and scale them to d isplay the input signals. See “Use Autoscale" on page 34.
40 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 41
Getting Started 1
11. Additional waveform
controls
The additional waveform controls consist of:
• [FFT] key — provides access to FFT spectrum analysis function. See Chapter 4, “FFT
Spectral Analysis,” starting on page 89.
•[Math] key — provides access to math (add, subtract, etc.) waveform functions. See
Chapter 5, “Math Waveforms,” starting on page 99.
•[Ref] key — provides access to reference waveform functions. Reference waveforms are
saved waveforms that can be displayed and compared against other analog channel or math waveforms. Also, measurements can be made on reference waveforms. See
Chapter 6, “Reference Waveforms,” starting on page 127.
• [Digital] key — Press this key to turn the digital channels on or off (the arrow to the left will
illuminate).
When the arrow to the left of the [Digital] key is illuminated, the upper multiplexed knob selects (and highlights in red) individual digital channels, and the lower multiplexed knob positions the selected digital channel.
If a trace is repositioned over an existing trace the indicator at the left edge of the trace will change from Dnn designation (where nn is a one or two d igit channel number from 0 to 15) to D*. The "*" indicates that two or more channels are overlaid.
You can rotate the upper knob to select an overlaid channel, then rotate the lower knob to position it just as you would any other channel.
For more information on digital channels see Chapter 7, “Digital Channels,” starting on page 131.
Multiplexed scale knob — This scale knob is used with Math, Ref, or Digital waveforms,
whichever has the illuminated arrow to the left. For math and reference waveforms, the scale knob acts like an analog channel vertical scale knob.
Multiplexed position knob — This position knob is used with Math, Ref, or Digital
waveforms, whichever has the illuminated arrow to the left. For math and reference waveforms, the position knob acts like an analog channel vertical position knob.
•[Serial] key (between channel 1 and 2 keys) — This key is used to enable serial decode. The
multiplexed scale and position knobs are not used with serial decode. For more information on serial decode, see Chapter 8, “Serial Decode,” starting on page 149.
12. Measure controls The measure controls consist of:
Cursors knob — Push this knob to select cursors from a popup menu. Then, after the popup
menu closes (either by timeout or by pushing the knob again), rotate the knob to adjust the selected cursor position.
•[Cursors] key — Press this key to open a menu that lets you select the cursors mode and
source.
•[Meas] key — Press this key to access a set of predefined measurements. See Chapter 15,
“Measurements,” starting on page 243.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 41
Page 42
1 Getting Started
13. File keys Press the [Save/Recall] key to save or recall a waveform or setup. See Chapter 20, “Save/Email/Recall (Setups, Screens, Data),” starting on page 321.
The [Print] key opens the Print Configuration Menu so you can print the displayed waveforms. See Chapter 21, “Print (Screens),” starting on page 335.
14. Tools keys The Tools keys consist of:
• [Utility] key — Press this key to access the Utility Menu, which lets you configure the
oscilloscope's I/O settings, use the file explorer, set preferences, access the service menu, and choose other options. See Chapter 22, “Utility Settings,” starting on page 341.
• [Quick Action] key — Press this key to perform the selected quick action: measure all
snapshot, print, save, recall, freeze display, and more. See “Configuring the [Quick
Action] Key" on page 357.
•[Wave Gen] key — Press this key to access waveform generator functions. See
Chapter 19, “Waveform Generator,” starting on page 301.
•[Analyze] key — Press this key to access analysis features like:
Measurement threshold setting.
Trigger level setting.
Video trigger automatic set up and display.
Counter (DVMCTR) (see “Counter" on page 291).
Digital voltmeter (DVMCTR) (see “Digital Voltmeter" on page 290).
Mask testing (see Chapter 16, “Mask Testing,” starting on page 275).
The DSOX3PWR power measurement and analysis application.
15. [Help] key Opens the Help Menu where you can d isplay overview help topics and select the Language. See also “Access the Built-In Quick Help" on page 61.
16. Vertical controls The Vertical controls consist of:
Analog channel on/off keys — Use these keys to switch a channel on or off, or to access a
channel's menu in the softkeys. There is one channel on/off key for each analog channel.
Vertical scale knob — There are knobs marked for each channel. Use these knobs
to change the vertical sensitivity (gain) of each analog channel.
Vertical position knobs — Use these knobs to change a channel's vertical position on the
display. There is one Vertical Position control for each analog channel.
For more information, see Chapter 3, “Vertical Controls,” starting on page 79.
42 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 43
Getting Started 1
17. Analog channel
inputs
18. Demo 2, Ground, and
Demo 1 terminals
19. USB Host port This port is for connecting a USB mass storage device, printer, mouse, or keyboard to the
Attach oscilloscope probes or BNC cables to these BNC connectors.
With the InfiniiVision 3000T X-Series oscilloscopes, you can set the input impedance of the analog channels to either 50 Ω or 1 MΩ. See “To specify channel input impedance" on page 82.
The InfiniiVision 3000T X-Series oscilloscopes also provide the AutoProbe interface. The AutoProbe interface uses a series of contacts directly below the channel's BNC connector to transfer information between the oscilloscope and the probe. When you connect a compatible probe to the oscilloscope, the AutoProbe interface determines the type of probe and sets the oscilloscope's parameters (units, offset, attenuation, coupling, and impedance) accordingly.
Demo 2 terminal — After a [Default Setup], this terminal outputs the Probe Comp signal
which helps you match a probe's input capacitance to the oscilloscope channel to which it is connected. See “Compensate Passive Probes" on page 36. With certain licensed features, the oscilloscope can also output demo or training signals on this terminal.
Ground terminal — Use the ground terminal for oscilloscope probes connected to the Demo
1 or Demo 2 terminals.
Demo 1 terminal — The oscilloscope can output training signals on this and the Demo 2
terminal. See: [Help] > Training Signals > Training Signals.
oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to save or recall oscilloscope setup files and reference waveforms or to save data and screen images. See
Chapter 20, “Save/Email/Recall (Setups, Screens, Data),” starting on page 321.
To print, connect a USB compliant printer. For more information about printing see
Chapter 21, “Print (Screens),” starting on page 335.
You can also use the USB port to update the oscilloscope's system software when updates are available.
You do not need to take special precautions before removing the USB mass storage device from the oscilloscope (you do not need to "eject" it). Simply unplug the USB mass storage device from the oscilloscope when the file operation is complete.
CAUTION: Do not connect a host computer to the oscilloscope's USB host port. Use the device port. A host computer sees the oscilloscope as a device, so connect the host computer to the oscilloscope's device port (on the rear panel). See “I/O Interface Settings" on page 341.
There is a second USB host port on the back panel.
20. Digital channel
inputs
Connect the digital probe cable to this connector (MSO models only). See Chapter 7, “Digital Channels,” starting on page 131.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 43
Page 44
1 Getting Started
21. Waveform generator
outputs

Front Panel Overlays for Different Languages

Built-in, license-enabled waveform generator can output arbitrary, sine, square, ramp, pulse, DC, noise, sine cardinal, exponential rise, exponential fall, cardiac, or Gaussian pulse waveforms on the Gen Out BNC connectors. Modulated waveforms are available on Wave Gen except for arbitrary, pulse, DC, and noise waveforms. Press the [Wave Gen] key to set up the waveform generator. See Chapter 19, “Waveform Generator,” starting on page 301.
Front panel overlays, which have translations for the English front panel keys and label text, are available in several languages. The appropriate overlay is included when the localization option is chosen at time of purchase.
To install a front panel overlay:
1 Gently pull on the front panel knobs to remove them. 2 Insert the overlay's side tabs into the slots on the front panel.
3 Reinstall the front panel knobs.
Front panel overlays may be ordered from www.keysight.com/find/parts using the following part numbers:
44 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 45
Getting Started 1
Language 2 Channel Overlay 4 Channel Overlay
Czech 75037-94327 75037-94328
French 75037-94321 75037-94322
German 75037-94317 75037-94318
Italian 75037-94325 75037-94326
Japanese 75037-94323 75037-94324
Korean 75037-94307 75037-94308
Polish 75037-94329 75037-94330
Portuguese 75037-94315 75037-94316
Russian 75037-94331 75037-94332
Simplified Chinese 75037-94309 75037-94310
Spanish 75037-94319 75037-94320
Thai 75037-94312 75037-94311
Traditional Chinese 75037-94305 75037-94306
Turkish 75037-94313 75037-94314

Learn the Touchscreen Controls

When the [Touch] key is lit, you can control the oscilloscope by touching different areas of the screen. You can:
“Draw Rectangles for Waveform Zoom or Zone Trigger Set Up" on page 46
“Flick or Drag to Scale, Position, and Change Offset" on page 47
“Select Sidebar Information or Controls" on page 49
“Undock Sidebar Dialogs by Dragging" on page 49
“Re-dock Sidebar Dialogs to Split Sidebar" on page 50
“Select Dialog Menus and Close Dialogs" on page 51
“Drag Cursors" on page 51
“Touch Softkeys and Menus On the Screen" on page 51
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 45
Page 46
1 Getting Started

Draw Rectangles for Waveform Zoom or Zone Trigger Set Up

“Enter Names Using Alpha-Numeric Keypad Dialogs" on page 52
“Change Waveform Offsets By Dragging Ground Reference Icons" on page 53
“Access Controls and Menus Via the Menu Icon" on page 54
“Turn Channels On/Off and Open Scale/Offset Dialogs" on page 55
“Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 56
“Access the Trigger Menu, Change the Trigger Mode, and Open the Trigger
Level Dialog" on page 57
“Use a USB Mouse and/or Keyboard for Touchscreen Controls" on page 58
1 Touch the upper-right corner to select the rectangle draw mode.
2 Drag your finger across the screen to draw a rectangle. 3 Take your finger off the screen. 4 Touch the desired option from the popup menu.
46 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 47
Getting Started 1

Flick or Drag to Scale, Position, and Change Offset

1 Touch the upper-right corner to select the horizontal drag mode.
2 When the waveform drag mode is selected, you can use these touch gestures:
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 47
Page 48
1 Getting Started
Flick — allows very fast browsing of waveforms. It is similar to browsing on
tablets and smartphones. It is much easier to flick than to continually turn a knob.
Drag — drag your finger across the screen to change the horizontal delay.
Drag your finger up or down to change the vertical offset.
To select waveforms, tap them. The waveform closest horizontally to the tap location is selected. The selected waveform is indicated by the ground marker with the filled background (channel 1 in the following example).
48 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 49

Select Sidebar Information or Controls

1 Touch the blue menu icon in the sidebar. 2 In the popup menu, touch the type of information or controls you want to see in
the sidebar.
Getting Started 1

Undock Sidebar Dialogs by Dragging

Sidebar dialogs can be undocked and placed anywhere on the screen. 1 Drag the sidebar dialog title wherever you like.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 49
Page 50
1 Getting Started

Re-dock Sidebar Dialogs to Split Sidebar

This lets you view multiple types of information or controls at the same time.
Sidebar dialogs can be re-docked to either half or full height of the sidebar. 1 Drag the dialog title back to the desired sidebar target.
You can display two dialogs in the sidebar at the same time.
50 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 51

Select Dialog Menus and Close Dialogs

• Touch the blue menu icon in the dialog for options.
• Touch the red "X" icon to close a dialog.

Drag Cursors

When cursors are displayed, you can drag the name handles to position them.
Getting Started 1

Touch Softkeys and Menus On the Screen

• Touch onscreen softkey labels to select them.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 51
Page 52
1 Getting Started
This is the same as pressing the softkey keys.
• When softkeys provide menus, double-touch to select a menu item.
This may be an easier than selecting a menu item via the Entry knob.

Enter Names Using Alpha-Numeric Keypad Dialogs

Some softkeys open alpha-numeric dialogs that let you touch to enter names.
52 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 53
Getting Started 1

Change Waveform Offsets By Dragging Ground Reference Icons

You can drag ground icons to change a waveform's vertical offset.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 53
Page 54
1 Getting Started
54 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide

Access Controls and Menus Via the Menu Icon

1 Touch the upper-left menu icon to open the main menu.
2 Touch left side controls to perform oscilloscope operations.
Page 55
Getting Started 1
3 Touch menu items and submenu items to access menus and additional
controls.

Turn Channels On/Off and Open Scale/Offset Dialogs

• Touch channel numbers to turn them on or off.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 55
Page 56
1 Getting Started

Access the Horizontal Menu and Open the Scale/Delay Dialog

• When channels are on, touch the scale and offset values to access a dialog for changing them.
• Touch "H" to access the Horizontal Menu.
• Touch the horizontal scale and delay values to access a dialog for changing them.
56 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 57
Getting Started 1

Access the Trigger Menu, Change the Trigger Mode, and Open the Trigger Level Dialog

• Touch "T" to access the Trigger Menu.
• Touch the trigger level value(s) to access a dialog for changing the level(s).
• Touch "Auto" or "Trig'd" to quickly toggle the trigger mode.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 57
Page 58
1 Getting Started

Use a USB Mouse and/or Keyboard for Touchscreen Controls

Connecting a USB mouse gives you a mouse pointer on the display. Mouse clicks and drags behave the same as screen touches and finger drags.
If you connect a USB keyboard, you can use it to enter values in alpha-numeric keypad dialogs.

Learn the Rear Panel Connectors

For the following figure, refer to the numbered descriptions in the table that follows.
WARNING: MAIN TAIN GROUND TO AVOID ELECTRIC SHOCK
100-120V, 50/60/400Hz
~
100-240V, 50/60Hz
~
100 Watts MAX
8. USB Device port
7. USB Host port
6. EXT TRIG IN connector
5. Calibration
protect
button
4. TRIG OUT
connector
3. Module slot
2. Kensington lock hole
1. Power cord connector
3. LAN/VGA
option module
3. GPIB
option module
58 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 59
Getting Started 1
1. Power cord connector
2. Kensington lock hole
3. Module slot A DSOXLAN LAN/VGA module may be ordered and installed separately.
4. TRIG OUT connector
5. Calibration protect button
6. EXT TRIG IN connector
Attach the power cord here.
This is where you can attach a Kensington lock for securing the instrument.
LAN port — lets you communicate with the oscilloscope and use the
Remote Front Panel feature using the LAN port. See Chapter 23, “Web Interface,” starting on page 359 and “Accessing the Web
Interface" on page 360.
VGA video output — lets you connect an external monitor or projector to
provide a larger display or to provide a display at a viewing position away from the oscilloscope.
The oscilloscope's built-in d isplay remains on even when an external display is connected. The video output connector is always active.
For optimal video quality and performance, we recommend you use a shielded video cable with ferrite cores.
A DSOXGPIB GPIB module may be ordered and installed separately.
Trigger output BNC connector. See “Setting the Rear Panel TRIG OUT
Source" on page 350.
See “To perform user calibration" on page 352.
External trigger input BNC connector. See “External Trigger Input" on page 213 for an explanation of this feature.
7. USB Host port This port functions identically to the USB host port on the front panel. USB
Host Port is used for saving data from the oscilloscope and loading software updates. See also USB Host port (see page 43).
8. USB Device port
This port is for connecting the oscilloscope to a host PC. You can issue remote commands from a host PC to the oscilloscope via the USB device port. See “Remote Programming with Keysight IO Libraries" on page 366.

Learn the Oscilloscope Display

The oscilloscope display contains acquired waveforms, setup information, measurement results, and the softkey definitions.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 59
Page 60
1 Getting Started
Analog channel sensitivity
Status line
Analog channels and ground levels
Trigger level
Digital channels
Softkeys
Menu line
Trigger point, time reference
Delay time
Time/ div
Run/Stop status
Trigger type
Trigger source
Measurements
Trigger level or digital threshold
Sidebar information and controls area
Cursors defining measurement
Measurement statistics
60 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide

Figure 1 Interpreting the oscilloscope display

Status line The top line of the d isplay contains vertical, horizontal, and trigger setup
information.
Display area The d isplay area contains the waveform acquisitions, channel identifiers, and
analog trigger, and ground level indicators. Each analog channel's information appears in a different color.
Signal detail is displayed using 256 levels of intensity. For more information about viewing signal detail see “To adjust waveform intensity" on page 155.
For more information about display modes see Chapter 9, “Display Settings,” starting on page 155.
Page 61
Getting Started 1
Back
Sidebar information and controls area
Menu line This line normally contains menu name or other information associated with the
Softkey labels These labels describe softkey functions. Typically, softkeys let you set up
The sidebar information area can contain summary, cursors, measurements, or digital vol tmeter information dialogs or it can contain navigation and other control dialogs.
For more information, see:
“Select Sidebar Information or Controls" on page 49
“Undock Sidebar Dialogs by Dragging" on page 49
selected menu.
additional parameters for the selected mode or menu.
Pressing the Back/Up key at the top of the menu hierarchy turns off softkey labels and d isplays additional status information describing channel offset and other configuration parameters.

Access the Built-In Quick Help

To view Quick Help 1 Press and hold the key or softkey for which you would like to view help.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 61
Page 62
1 Getting Started
Quick Help message
Press and hold front panel key or softkey (or right-click softkey when using web browser remote front panel).
Quick Help remains on the screen until another key is pressed or a knob is turned.
To select the user
interface language
62 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
To select the user interface language:
1 Press [Help], then press the Language softkey. 2 Repeatedly press and release the Language softkey or rotate the Entry knob until
the desired language is selected.
The following languages are available: Czech, English, French, German, Italian, Japanese, Korean, Polish, Portuguese, Russian, Simplified Chinese, Spanish, Thai, Traditional Chinese, and Turkish.
Page 63
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide

2 Horizontal Controls

To adjust the horizontal (time/div) scale / 64 To adjust the horizontal delay (position) / 65 Panning and Zooming Single or Stopped Acquisitions / 66 To change the horizontal time mode (Normal, XY, or Roll) / 67 To display the zoomed time base / 71 To change the horizontal scale knob's coarse/fine adjustment setting / 72 To position the time reference (left, center, right, custom) / 72 Searching for Events / 73 Navigating the Time Base / 75
The horizontal controls include:
• The horizontal scale and position knobs.
•The [Horiz] key for accessing the Horizontal Menu.
• The zoom key for quickly enabling/disabling the split-screen zoom display.
•The [Search] key for finding events on analog channels or in serial decode.
•The [Navigate] keys for navigating time, search events, or segmented memory
acquisitions.
• Touchscreen controls for setting the horizontal scale and position (delay), accessing the Horizontal Menu, and navigating.
The following figure shows the Horizontal Menu which appears after pressing the [Horiz] key.
63
Page 64
2 Horizontal Controls
Trigger point
Sample rate
Time reference
Delay time
Time/ div
Trigger source
Trigger level or threshold
XY or Roll mode
Normal time mode
Zoomed time base
Fine control
Time reference

Figure 2 Horizontal Menu

The Horizontal Menu lets you select the time mode (Normal, XY, or Roll), enable Zoom, set the time base fine control (vernier), and specify the time reference.
The current sample rate is displayed in the Summary box in the right-side information area.

To adjust the horizontal (time/div) scale

1 Turn the large horizontal scale (sweep speed) knob marked to
change the horizontal time/div setting.
64 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 65
You can also make this adjustment using the touchscreen. See “Access the
Horizontal Menu and Open the Scale/Delay Dialog" on page 56.
Notice how the time/div information in the status line changes. The symbol at the top of the display indicates the time reference point.
The horizontal scale knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See “Panning and Zooming Single or Stopped
Acquisitions" on page 66.
Note that the horizontal scale knob has a different purpose in the Zoom display. See “To display the zoomed time base" on page 71.

To adjust the horizontal delay (position)

1 Turn the horizontal delay (position) knob ( ).
Horizontal Controls 2
The trigger point moves horizontally, pausing at 0.00 s (mimicking a mechanical detent), and the delay value is displayed in the status line.
You can also make this adjustment using the touchscreen. See “Flick or Drag to
Scale, Position, and Change Offset" on page 47 and “Access the Horizontal Menu and Open the Scale/Delay Dialog" on page 56.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and indicates how far it is from the time reference point (hollow inverted triangle ). These reference points are indicated along the top of the display grid.
Figure 2 shows the trigger point with the delay time set to 200 µs. The delay time
number tells you how far the time reference point is located from the trigger point. When delay time is set to zero, the delay time indicator overlays the time reference indicator.
All events displayed left of the trigger point happened before the trigger occurred. These events are called pre-trigger information, and they show events that led up to the trigger point.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 65
Page 66
2 Horizontal Controls
Everything to the right of the trigger point is called post-trigger information. The amount of delay range (pre-trigger and post-trigger information) available depends on the time/div selected and memory depth.
The horizontal position knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See “Panning and Zooming Single or
Stopped Acquisitions" on page 66.
Note that the horizontal position knob has a different purpose in the Zoom display. See “To display the zoomed time base" on page 71.

Panning and Zooming Single or Stopped Acquisitions

When the oscilloscope is stopped, use the horizontal scale and position knobs to pan and zoom your waveform. The stopped display may contain several acquisitions worth of information, but only the last acquisition is available for pan and zoom.
The ability to pan (move horizontally) and scale (expand or compress horizontally) an acquired waveform is important because of the additional insight it can reveal about the captured waveform. This additional insight is often gained from seeing the waveform at different levels of abstraction. You may want to view both the big picture and the specific little picture details.
The ability to examine waveform detail after the waveform has been acquired is a benefit generally associated with digital oscilloscopes. Often this is simply the ability to freeze the display for the purpose of measuring with cursors or printing the screen. Some digital oscilloscopes go one step further by including the ability to further examine the signal details after acquiring them by panning through the waveform and changing the horizontal scale.
There is no limit imposed on the scaling ratio between the time/div used to acquire the data and the time/div used to view the data. There is, however, a useful limit. This useful limit is somewhat a function of the signal you are analyzing.
66 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 67
Horizontal Controls 2
NOTE
Zooming into stopped acquisitions
The screen will still contain a relatively good display if you zoom-in horizontally by a factor of 1000 and zoom-in vertically by a factor of 10 to display the information from where it was acquired. Remember that you can only make automatic measurements on displayed data.

To change the horizontal time mode (Normal, XY, or Roll)

1 Press [Horiz]. 2 In the Horizontal Menu, press Time Mode; then, select:
Normal — the normal viewing mode for the oscilloscope.
In the Normal time mode, signal events occurring before the trigger are plotted to the left of the trigger point () and signal events after the trigger plotted to the right of the trigger point.
XY — XY mode changes the display from a volts-versus-time display to a
volts-versus-volts display. The time base is turned off. Channel 1 amplitude is plotted on the X-axis and Channel 2 amplitude is plotted on the Y-axis.
You can use XY mode to compare frequency and phase relationships between two signals. XY mode can also be used with transducers to display strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
Use the cursors to make measurements on XY mode waveforms.
For more information about using XY mode for measurements, refer to “XY
Time Mode" on page 68.
Roll — causes the waveform to move slowly across the screen from right to
left. It only operates on time base settings of 50 ms/div and slower. If the current time base setting is faster than the 50 ms/div limit, it will be set to 50 ms/div when Roll mode is entered.
In Roll mode there is no trigger. The fixed reference point on the screen is the right edge of the screen and refers to the current moment in time. Events that have occurred are scrolled to the left of the reference point. Since there is no trigger, no pre-trigger information is available.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 67
Page 68
2 Horizontal Controls
sinθ =Aor
C

XY Time Mode

Example This exercise shows a common use of the XY display mode by measuring the
If you would like to pause the display in Roll mode press the [Single] key. To clear the display and restart an acquisition in Roll mode, press the [Single] key again.
Use Roll mode on low-frequency waveforms to yield a display much like a strip chart recorder. It allows the waveform to roll across the display.
The XY time mode converts the oscilloscope from a volts-versus-time display to a volts-versus-volts display using two input channels. Channel 1 is the X-axis input, channel 2 is the Y-axis input. You can use various transducers so the display could show strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
phase difference between two signals of the same frequency with the Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the same
frequency but out of phase to channel 2.
2 Press the [Auto Scale] key, press the [Horiz] key; then, press Time Mode and select
"XY".
3 Center the signal on the display with the channel 1 and 2 position ( ) knobs.
Use the channel 1 and 2 volts/div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing.
The phase difference angle (θ) can be calculated using the following formula (assuming the amplitude is the same on both channels):
68 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 69
Horizontal Controls 2
A
B
D
C
Signal must be cent ered i n “X”
Measuring
phase difference
Signals 90 degrees
out of phase
Signals
in phase
Figure 3 XY time mode signals, centered on display
4 Press the [Cursors] key. 5 Set the Y2 cursor to the top of the signal, and set Y1 to the bottom of the
signal. Note the ΔY value at the bottom of the display. In this example, we are using
the Y cursors, but you could have used the X cursors instead.
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis.
Again, note the ΔY value.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 69
Page 70
2 Horizontal Controls
sinθ =
second Δ Y
first Δ Y
=
1.319
2.297
; θ = 35.05 degrees of phase shift
NOTE
Figure 4 Phase difference measurements, automatic and using cursors
7 Calculate the phase difference using the formula below.
For example, if the first ΔY value is 2.297 and the second ΔY value is 1.319:
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is the X-axis input, channel 2 is the Y-axis input, and the EXT TRIG IN is the Z-axis input. If you only want to see portions of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off (analog oscilloscopes called this Z-axis blanking because it turned the beam on and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off.
70 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 71

To display the zoomed time base

These markers show the beginning and end of the Zoom window
Normal window
Time/div for zoomed window
Time/div for normal window
Delay time momentarily displays when the Horizontal position knob is turned
Zoom window
Signal anomaly expanded in zoom window
Select Zoom
Zoom, formerly called Delayed sweep mode, is a horizontally expanded version of the normal display. When Zoom is selected, the display divides in half. The top half of the display shows the normal time/div window and the bottom half displays a faster Zoom time/div window.
The Zoom window is a magnified portion of the normal time/div window. You can use Zoom to locate and horizontally expand part of the normal window for a more detailed (higher-resolution) analysis of signals.
To turn on (or off) Zoom:
1 Press the zoom key (or press the [Horiz] key and then the Zoom softkey).
Horizontal Controls 2
The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted. The box shows the portion of the normal sweep that is expanded in the lower half.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 71
Page 72
2 Horizontal Controls
To change the time/div for the Zoom window, turn the horizontal scale (sweep speed) knob. As you turn the knob, the zoomed window time/div is highlighted in the status line above the waveform display area. The Horizontal scale (sweep speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left-to-right position of the zoom window. The delay value, which is the time displayed relative to the trigger point) is momentarily displayed in the upper-right portion of the display when the delay time ( ) knob is turned.
Negative delay values indicate you're looking at a portion of the waveform before the trigger event, and positive values indicate you're looking at the waveform after the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn the horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to “To isolate a
pulse for Top measurement" on page 252 and “To isolate an event for frequency measurement" on page 258.

To change the horizontal scale knob's coarse/fine adjustment setting

1 Push the horizontal scale knob (or press [Horiz] > Fine) to toggle between fine
and coarse adjustment of the horizontal scale.
When Fine is enabled, turning the horizontal scale knob changes the time/div (displayed in the status line at the top of the display) in smaller increments. The time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in a 1-2-5 step sequence.

To position the time reference (left, center, right, custom)

Time reference is the reference point on the display for delay time (horizontal position).
1 Press [Horiz].
72 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 73
Horizontal Controls 2
2 In the Horizontal Menu, press Time Ref; then, select:
Left — the time reference is set to one major division from the left edge of the
display.
Center — the time reference is set to the center of the display.
Right — the time reference is set to one major division from the right edge of
the display.
Custom Location — lets you place the time reference location at a percent of
the graticule width (where 0% is the left edge and 100% is the right edge).
A small hollow triangle () at the top of the display grid marks the position of the time reference. When delay time is set to zero, the trigger point indicator (▼) overlays the time reference indicator.
The time reference position sets the initial position of the trigger event within acquisition memory and on the display, with delay set to 0.
Turning the Horizontal scale (sweep speed) knob expands or contracts the waveform about the time reference point (). See “To adjust the horizontal
(time/div) scale" on page 64.
Turning the Horizontal position ( ) knob in Normal mode (not Zoom) moves the trigger point indicator () to the left or right of the time reference point (). See
“To adjust the horizontal delay (position)" on page 65.

Searching for Events

You can use the [Search] key and menu to search for Edge, Pulse Width, Rise/Fall Time, Runt, Frequency Peaks, and Serial events on the analog channels.
Setting up searches (see “To set up searches" on page 74) is similar to setting up triggers. In fact, except for Frequency Peaks and Serial events, you can copy search setups to trigger setups and vice-versa (see “To copy search setups" on page 74).
Searches are different than triggers in that they use the measurement threshold settings instead of trigger levels.
Found search events are marked with white triangles at the top of the graticule, and the number of events found is displayed in the menu line just above the softkey labels.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 73
Page 74
2 Horizontal Controls

To set up searches

1 Press [Search]. 2 In the Search Menu, press Search; then, turn the Entry knob to select the search
type.
3 Use the remaining softkeys to set up the selected search type.
In most cases, setting up searches is similar to setting up triggers:
For setting up Edge searches, see “Edge Trigger" on page 172.
For setting up Pulse Width searches, see “Pulse Width Trigger" on
page 175.
For setting up Rise/Fall Time searches, see “Rise/Fall Time Trigger" on
page 182.
For setting up Runt searches, see “Runt Trigger" on page 188.
For setting up Frequency Peak searches, see “Searching for FFT Peaks" on
page 93.
For setting up Serial searches, see “Serial Trigger" on page 203 and
“Searching Lister Data" on page 152.
Remember that searches use the measurement threshold settings instead of trigger levels. Use the Thresholds softkey in the Search Menu to access the Measurement Threshold Menu. See “Measurement Thresholds" on page 270.

To copy search setups

Except for Frequency Peak and Serial event search setups, you can copy search setups to trigger setups and vice-versa.
1 Press [Search]. 2 In the Search Menu, press Search; then, turn the Entry knob to select the search
type.
3 Press Copy. 4 In the Search Copy Menu:
Press Copy to Trigger to copy the setup for the selected search type to the
same trigger type. For example, if the current search type is Pulse Width, pressing Copy to Trigger copies the search settings to the Pulse Width trigger settings and selects the Pulse Width trigger.
74 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 75
Press Copy from Trigger to copy the trigger setup for the selected search type
to the search setup.
To undo a copy, press Undo Copy.
The softkeys in the Search Copy Menu may not be available when one of the settings cannot be copied or there is no trigger type that corresponds to the search type.

Navigating the Time Base

You can use the [Navigate] key and controls to navigate through:
• Captured data (see “To navigate time" on page 75).
• Search events (see “To navigate search events" on page 75).
• Segments, when segmented memory acquisitions are turned on (see “To
navigate segments" on page 76).
You can also access navigation controls on the touchscreen. See “Select Sidebar
Information or Controls" on page 49.
Horizontal Controls 2

To navigate time

When acquisitions are stopped, you can use the navigation controls to play through the captured data.
1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Time.
3 Press the navigation keys to play backward, stop, or play forward in
time. You can press the or keys multiple times to speed up the playback. There are three speed levels.
You can also access navigation controls on the touchscreen. See “Select Sidebar
Information or Controls" on page 49.

To navigate search events

When acquisitions are stopped, you can use the navigation controls to go to found search events (set using the [Search] key and menu, see “Searching for Events" on page 73).
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 75
Page 76
2 Horizontal Controls

To navigate segments

1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Search.
3 Press the back and forward keys to go to the previous or next search
event.
When searching Serial decode:
• You can press the stop key to set or clear a mark.
•The Auto zoom softkey specifies whether the waveform display is automatically zoomed to fit the marked row as you navigate.
• Pressing the Scroll Lister softkey lets you use the Entry knob to scroll through data rows in the Lister display.
You can also access navigation controls on the touchscreen. See “Select Sidebar
Information or Controls" on page 49.
When the segmented memory acquisition is enabled and acquisitions are stopped, you can use the navigation controls to play through the acquired segments.
1 Press [Navigate]. 2 In the Navigate Menu, press Navigate; then, select Segments. 3 Press Play Mode; then, select:
Manual — to play through segments manually.
In the Manual play mode:
Press the back and forward keys to go to the previous or next
segment.
Press the softkey to go to the first segment.
Press the softkey to go to the last segment.
Auto — to play through segments in an automated fashion.
In the Auto play mode:
76 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 77
Horizontal Controls 2
Press the navigation keys to play backward, stop, or play
forward in time. You can press the or keys multiple times to speed up the playback. There are three speed levels.
You can also access navigation controls on the touchscreen. See “Select Sidebar
Information or Controls" on page 49.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 77
Page 78
2 Horizontal Controls
78 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 79
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide

3 Vertical Controls

To turn waveforms on or off (channel or math) / 80 To adjust the vertical scale / 81 To adjust the vertical position / 81 To specify channel coupling / 81 To specify channel input impedance / 82 To specify bandwidth limiting / 83 To change the vertical scale knob's coarse/fine adjustment setting / 83 To invert a waveform / 84 Setting Analog Channel Probe Options / 84
The vertical controls include:
• The vertical scale and position knobs for each analog channel.
• The channel keys for turning a channel on or off and accessing the channel's softkey menu.
• Touchscreen controls for setting the vertical scale and position (offset) and accessing the Channel menus.
The following figure shows the Channel 1 Menu that appears after pressing the [1] channel key.
79
Page 80
3 Vertical Controls
Channel, Volts/div
Channel 1 ground level
Trigger source
Trigger level or threshold
Channel 2 ground level
NOTE
The ground level of the signal for each displayed analog channel is identified by the position of the icon at the far-left side of the display.

To turn waveforms on or off (channel or math)

80 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
1 Press an analog channel key turn the channel on or off (and to display the
channel's menu).
When a channel is on, its key is illuminated.
You can also do this using the touchscreen. See “Turn Channels On/Off and
Open Scale/Offset Dialogs" on page 55.
Turning channels off
You must be viewing the menu for a channel before you can turn it off. For example, if channel 1 and channel 2 are turned on and the menu for channel 2 is being displayed, to turn channel 1 off, press [1] to display the channel 1 menu; then, press [1] again to turn channel 1 off.
Page 81

To adjust the vertical scale

1 Turn the large knob above the channel key marked to set the vertical
scale (volts/division) for the channel.
You can also do this using the touchscreen. See “Turn Channels On/Off and
Open Scale/Offset Dialogs" on page 55.
The vertical scale knob changes the analog channel scale in a 1-2-5 step sequence (with a 1:1 probe attached) unless fine adjustment is enabled (see “To
change the vertical scale knob's coarse/fine adjustment setting" on page 83).
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the volts/division knob is vertical expansion about the ground level of the channel; however, you can change this to expand about the center of the display. See “To choose "expand
about" center or ground" on page 347.
Vertical Controls 3

To adjust the vertical position

1 Turn the small vertical position knob ( ) to move the channel's waveform up or
down on the display.
You can also make this adjustment using the touchscreen. See “Change
Waveform Offsets By Dragging Ground Reference Icons" on page 53 and “Turn Channels On/Off and Open Scale/Offset Dialogs" on page 55.
The offset voltage value represents the voltage difference between the vertical center of the display and the ground level ( ) icon. It also represents the voltage
at the vertical center of the display if vertical expansion is set to expand about ground (see “To choose "expand about" center or ground" on page 347).

To specify channel coupling

Coupling changes the channel's input coupling to either AC (alternating current) or DC (direct current).
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 81
Page 82
3 Vertical Controls
TIP
NOTE
If the channel is DC coupled, you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is removed, allowing you to use greater sensitivity to display the AC component of the signal.
1 Press the desired channel key. 2 In the Channel Menu, press the Coupling softkey to select the input channel
coupling:
DC — DC coupling is useful for viewing waveforms as low as 0 Hz that do not
have large DC offsets.
AC — AC coupling is useful for viewing waveforms with large DC offsets.
When AC coupling is chosen, you cannot select 50Ω mode. This is done to prevent damage to the oscilloscope.
AC coupling places a 10 Hz high-pass filter in series with the input waveform that removes any DC offset voltage from the waveform.
Note that Channel Coupling is independent of Trigger Coupling. To change trigger coupling see “To select the trigger coupling" on page 209.

To specify channel input impedance

When you connect an AutoProbe, self-sensing probe, or a compatible InfiniiMax probe, the oscilloscope automatically configures the analog input channels to the correct impedance.
1 Press the desired channel key. 2 In the Channel Menu, press Imped (impedance); then, select either:
50 Ohm — matches 50 ohm cables commonly used in making high frequency
measurements, and 50 ohm active probes.
When 50 Ohm input impedance is selected, it is displayed with the channel information on-screen.
82 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 83
When AC coupling is selected (see “To specify channel coupling" on page 81) or excessive voltage is applied to the input, the oscilloscope automatically switches to 1M Ohm mode to prevent possible damage.
1M Ohm — is for use with many passive probes and for general-purpose
measurements. The higher impedance minimizes the loading effect of the oscilloscope on the device under test.
This impedance matching gives you the most accurate measurements because reflections are minimized along the signal path.
See Also • For more information on probing, visit: www.keysight.com/find/scope_probes
• Information about selecting a probe can be found in document number
Keysight Oscilloscope Probes and Accessories Selection Guide (part number
5989-6162EN), available at www.keysight.com.

To specify bandwidth limiting

1 Press the desired channel key. 2 In the Channel Menu, press the BW Limit softkey to enable or disable bandwidth
limiting.
Vertical Controls 3
When bandwidth limit is on, the maximum bandwidth for the channel is approximately 20 MHz. For waveforms with frequencies below this, turning bandwidth limit on removes unwanted high frequency noise from the waveform. The bandwidth limit also limits the trigger signal path of any channel that has BW Limit turned on.

To change the vertical scale knob's coarse/fine adjustment setting

1 Push the channel's vertical scale knob (or press the channel key and then the
Fine softkey in the Channel Menu) to toggle between fine and coarse
adjustment of the vertical scale.
You can also do this using the touchscreen. See “Turn Channels On/Off and
Open Scale/Offset Dialogs" on page 55.
When Fine adjustment is selected, you can change the channel's vertical sensitivity in smaller increments. The channel sensitivity remains fully calibrated when Fine is on.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 83
Page 84
3 Vertical Controls
The vertical scale value is displayed in the status line at the top of the display.
When Fine is turned off, turning the volts/division knob changes the channel sensitivity in a 1-2-5 step sequence.

To invert a waveform

1 Press the desired channel key. 2 In the Channel Menu, press the Invert softkey to invert the selected channel.
When Invert is selected, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed. However, when using basic triggers, the oscilloscope attempts to maintain the same trigger point by changing trigger settings.
Inverting a channel also changes the result of any math function selected in the Waveform Math Menu or any measurement.

Setting Analog Channel Probe Options

1 Press the probe's associated channel key. 2 In the Channel Menu, press the Probe softkey to display the Channel Probe
Menu.
This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe.
The Channel Probe Menu changes depending on the type of probe connected.
For passive probes (such as the N2862A/B, N2863A/B, N2889A, N2890A, 10073C, 10074C, or 1165A probes), the Probe Check softkey appears; it guides you through the process of compensating probes.
84 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 85
For some active probes (such as InfiniiMax probes), the oscilloscope can accurately calibrate its analog channels for the probe. When you connect a probe that can be calibrated, the Calibrate Probe softkey appears (and the probe attenuation softkey may change). See “To calibrate a probe" on page 86.
See Also “To specify the channel units" on page 85
“To specify the probe attenuation" on page 85
“To specify the probe skew" on page 86

To specify the channel units

1 Press the probe's associated channel key. 2 In the Channel Menu, press Probe. 3 In the Channel Probe Menu, press Units; then, select:
Volts — for a voltage probe.
Amps — for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions will reflect the measurement units you have selected.
Vertical Controls 3

To specify the probe attenuation

This is set automatically if the oscilloscope can identify the connected probe. See Analog channel inputs (see page 43).
The probe attenuation factor must be set properly for accurate measurement results.
If you connect a probe that is not automatically identified by the oscilloscope, you can manually set the attenuation factor as follows:
1 Press the channel key. 2 Press the Probe softkey until you have selected how you want to specify the
attenuation factor, choosing either Ratio or Decibels.
3 Turn the Entry knob to set the attenuation factor for the connected probe.
When measuring voltage values, the attenuation factor can be set from 0.001:1 to 10000:1 in a 1-2-5 sequence.
When measuring current values with a current probe, the attenuation factor can be set from 10 V/A to 0.0001 V/A.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 85
Page 86
3 Vertical Controls

To specify the probe skew

When specifying the attenuation factor in decibels, you can select values from
-20 dB to 80 dB.
If Amps is chosen as the units and a manual attenuation factor is chosen, then the units as well as the attenuation factor are displayed above the Probe softkey.
When measuring time intervals in the nanoseconds (ns) range, small differences in cable length can affect the measurement. Use Skew to remove cable-delay errors between any two channels.
1 Probe the same point with both probes. 2 Press one of the probes associated channel key. 3 In the Channel Menu, press Probe. 4 In the Channel Probe Menu, press Skew; then, select the desired skew value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a total of 200 ns difference.
The skew setting is not affected by pressing [Default Setup] or [Auto Scale].

To calibrate a probe

The Calibrate Probe softkey guides you through the process of calibrating probes.
For certain active probes, such as InfiniiMax probes, the oscilloscope can accurately calibrate its analog channels for the probe. When you connect a probe that can be calibrated, the Calibrate Probe softkey in the Channel Probe Menu becomes active.
To calibrate one of these probes: 1 First, plug your probe into one of the oscilloscope channels.
This could be, for example, an InfiniiMax probe amplifier/probe head with attenuators attached.
2 Connect the probe to the left side, Demo 2, Probe Comp terminal, and the
probe ground to the ground terminal.
86 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 87
Vertical Controls 3
NOTE
When calibrating a differential probe, connect the positive lead to the Probe Comp terminal and the negative lead to the ground terminal. You may need to connect an alligator clip to the ground lug to allow a differential probe to span between the Probe Comp test point and ground. A good ground connection ensures the most accurate probe calibration.
3 Press the Channel on/off key to turn the channel on (if the channel is off). 4 In the Channel Menu, press the Probe softkey. 5 In the Channel Probe Menu, the second softkey from the left is for specifying
your probe head (and attenuation). Repeatedly press this softkey until the probe head selection matches the attenuator you are using.
The choices are:
10:1 single-ended browser (no attenuator).
10:1 differential browser (no attenuator).
10:1 (+6 dB Atten) single-ended browser.
10:1 (+6 dB Atten) differential browser.
10:1 (+12 dB Atten) single-ended browser.
10:1 (+12 dB Atten) differential browser.
10:1 (+20 dB Atten) single-ended browser.
10:1 (+20 dB Atten) differential browser.
6 Press the Calibrate Probe softkey and follow the instructions on the display.
For more information on InfiniiMax probes and accessories, see the probe's User's Guide.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 87
Page 88
3 Vertical Controls
88 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 89
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide

4 FFT Spectral Analysis

FFT is used to compute the fast Fourier transform using analog input channels or a lower math function. FFT takes the digitized time record of the specified source and transforms it to the frequency domain. When the FFT function is selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency. The readout for the horizontal axis changes from time to frequency (Hertz) and the vertical readout changes from volts to dB.
Use the FFT function to find crosstalk problems, to find distortion problems in analog waveforms caused by amplifier non-linearity, or for adjusting analog filters.
To display a FFT waveform: 1 Press the [FFT] key.
Source 1 — selects the source for the FFT.
Display — selects one of these FFT waveform display modes:
Normal — this is the FFT waveform without any averaging or hold functions
applied. This is how FFT math function waveforms are displayed.
Averaged Value — the FFT waveform is averaged the selected number of
times. Averages are calculated using a "decaying average" approximation, where:
next_average = current_average + (new_data - current_average)/N
Where N starts at 1 for the first acquisition and increments for each following acquisition until it reaches the selected number of averages, where it holds.
89
Page 90
4 FFT Spectral Analysis
Max Hold — records the maximum vertical values found at each horizontal
bucket across multiple analysis cycles and uses those values to build a waveform. This display mode is often referred to as Max Envelope.
Min Hold — records the minimum vertical values found at each horizontal
bucket across multiple analysis cycles and uses those values to build a waveform. This display mode is often referred to as Min Envelope.
Span/Center or Start Freq/Stop Freq — this pair of softkeys let you define the
displayed frequency range. Press the softkeys to toggle between:
Span/CenterSpan specifies the frequency range represented by the
width of the display. Divide span by 10 to calculate the frequency scale per division. Center specifies the frequency at the center vertical grid line of the display.
Start Freq/Stop FreqStart Freq specifies the frequency at the left side of
the display. Stop Freq specifies the frequency at the right side of the display.
To set desired values, tap the softkey label on screen for a keypad entry dialog or turn the Entry knob.
More FFT — displays the More FFT Settings Menu.
2 Press the More FFT softkey to display additional FFT settings.
Window— selects a window to apply to your FFT input signal:
Hanning — window for making accurate frequency measurements or for
resolving two frequencies that are close together.
Flat Top — window for making accurate amplitude measurements of
frequency peaks.
Rectangular — good frequency resolution and amplitude accuracy, but use
only where there will be no leakage effects. Use on self-windowing waveforms such as pseudo-random noise, impulses, sine bursts, and decaying sinusoids.
Blackman Harris — window reduces time resolution compared to a
rectangular window, but improves the capacity to detect smaller impulses due to lower secondary lobes.
90 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 91
FFT Spectral Analysis 4
NOTE
Bartlett — (triangular, with end points at zero) window is similar to the
Hanning window in that it is good for making accurate frequency measurements, but its higher and wider secondary lobes make it not quite as good for resolving frequencies that are close together.
Vertical Units — lets you select Decibels or V RMS as the units for the FFT
vertical scale.
FFT Gating — when the zoomed time base is displayed, press this softkey to
select:
No Gating — the FFT is performed on the source waveform in the upper
Main time base window.
Gate By Zoom — the FFT is performed on the source waveform in the lower
Zoom window.
Auto Setup — sets the frequency Span and Center to values that will cause
the entire available spectrum to be displayed. The maximum available frequency is half the FFT sample rate, which is a function of the time per division setting. The FFT resolution is the quotient of the sampling rate and the number of FFT points (f
above the softkeys.
/N). The current FFT Resolution is displayed
S
Scale and offset considerations
If you do not manually change the FFT scale or offset settings, when you turn the horizontal scale knob, the span and center frequency settings will automatically change to allow optimum viewing of the full spectrum.
If you do manually set scale or offset, turning the horizontal scale knob will not change the span or center frequency settings, allowing you see better detail around a specific frequency.
Pressing the FFT Auto Setup softkey will automatically rescale the waveform and span and center will again automatically track the horizontal scale setting.
3 To make cursor measurements, press the [Cursors] key and set the Source
softkey to Math N.
Use the X1 and X2 cursors to measure frequency values and difference between two frequency values (ΔX). Use the Y1 and Y2 cursors to measure amplitude in dB and difference in amplitude (ΔY).
4 To make other measurements, press the [Meas] key and set the Source softkey to
Math N.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 91
Page 92
4 FFT Spectral Analysis
You can make peak-to-peak, maximum, minimum, and average dB measurements on the FFT waveform. You can also find the frequency value at the first occurrence of the waveform maximum by using the X at Max Y measurement.
To adjust the FFT waveform scale and offset 1 Make sure the multiplexed scale and position knobs above and below the [FFT]
key are selected for the FFT waveform.
If the arrow to the left of the [FFT] key is not illuminated, press the key.
2 Use the multiplexed scale and position knobs next to the [Math] key to re-size
and re-position the math waveform.
The following FFT spectrum was obtained by connecting a 2.5 V, 100 kHz square wave to channel 4. Set the horizontal scale to 50 µs/div, vertical sensitivity to 1 V/div, Units/div to 20 dBV, Offset to -40.0 dBV, Center frequency to 500 kHz, frequency Span to 1 MHz, and window to Hanning.
See Also “Searching for FFT Peaks" on page 93
“FFT Measurement Hints" on page 93
“FFT Units" on page 95
“FFT DC Value" on page 95
92 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 93
“FFT Aliasing" on page 95
“FFT Spectral Leakage" on page 97
“Units for Math Waveforms" on page 101

Searching for FFT Peaks

To search for FFT math function frequency peaks:
1 Press [Search]. 2 In the Search Menu, press Search; then, turn the Entry knob to select Frequency
Peaks. 3 Press Source and select the FFT math function waveform to search. 4 Press Max # Peaks and specify the maximum number of FFT peaks to find. 5 Press Threshold and turn the Entry knob to specify the threshold level necessary
to be considered a peak. 6 Press Excursion to specify the amplitude above the FFT waveform's noise floor
necessary to be recognized as a peak.
Note that the FFT waveform's noise floor level differs when additional math
functions are applied to the FFT:
When Averaged Value, Max Hold, or Min Hold are applied, the FFT waveform's
noise floor is more stable, and excursion level settings are more accurate.
When no additional math functions are applied (normal), the FFT waveform's
noise floor is less stable and excursion level settings are less accurate.
FFT Spectral Analysis 4
White arrowheads at the top of the graticule show where FFT peaks are found.
When acquisitions are stopped, you can use the [Navigate] keys and cursors to look at the search events found.

FFT Measurement Hints

The number of points acquired for the FFT record can be up to 65,536, and when frequency span is at maximum, all points are displayed. Once the FFT spectrum is displayed, the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 93
Page 94
4 FFT Spectral Analysis
NOTE
detail. Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution. As frequency span is decreased, the number of points shown is reduced, and the display is magnified.
While the FFT spectrum is displayed, use the [Math] and [Cursors] keys to switch between measurement functions and frequency domain controls in FFT Menu.
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N). With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the resolution.
Decreasing the effective sampling rate by selecting a greater time/div setting will increase the low frequency resolution of the FFT display and also increase the chance that an alias will be displayed. The resolution of the FFT is the effective sample rate divided by the number of points in the FFT. The actual resolution of the display will not be this fine as the shape of the window will be the actual limiting factor in the FFTs ability to resolve two closely space frequencies. A good way to test the ability of the FFT to resolve two closely spaced frequencies is to examine the sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
• Make sure the probe attenuation is set correctly. The probe attenuation is set from the Channel Menu if the operand is a channel.
• Set the source sensitivity so that the input signal is near full screen, but not clipped.
•Use the Flat Top window.
• Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
• Use the Hanning window.
• Use Cursors to place an X cursor on the frequency of interest.
• Adjust frequency span for better cursor placement.
• Return to the Cursors Menu to fine tune the X cursor.
94 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 95

FFT Units

NOTE
FFT Spectral Analysis 4
For more information on the use of FFTs please refer to Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pd f. Additional
information can be obtained from Chapter 4 of the book Spectrum and Network Measurements by Robert A. Witte.
FFT (Magnitude)
Units
FFT (Phase) Units In this case, the vertical units are degrees or radians.
FFT units are displayed as dB.

FFT DC Value

The FFT computation produces a DC value that is incorrect. It does not take the offset at center screen into account. The DC value is not corrected in order to accurately represent frequency components near DC.

FFT Aliasing

When using FFTs, it is important to be aware of frequency aliasing. This requires that the operator have some knowledge as to what the frequency domain should contain, and also consider the sampling rate, frequency span, and oscilloscope vertical bandwidth when making FFT measurements. The FFT resolution (the quotient of the sampling rate and the number of FFT points) is displayed directly above the softkeys when the FFT Menu is displayed.
Nyquist Frequency and Aliasing in the Frequency Domain
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. This frequency is half of the sample rate. Frequencies above the Nyquist frequency will be under sampled, which causes aliasing. The Nyquist frequency is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain.
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 95
Page 96
4 FFT Spectral Analysis
Aliasing happens when there are frequency components in the signal higher than half the sample rate. Because the FFT spectrum is limited by this frequency, any higher components are displayed at a lower (aliased) frequency.
The following figure illustrates aliasing. This is the spectrum of a 990 Hz square wave, which has many harmonics. The horizontal time/div setting for the square wave sets the sample rate and results in a FFT resolution of 1.91 Hz. The displayed FFT spectrum waveform shows the components of the input signal above the Nyquist frequency to be mirrored (aliased) on the display and reflected off the right edge.

Figure 5 Aliasing

Because the frequency span goes from 0 to the Nyquist frequency, the best way to prevent aliasing is to make sure that the frequency span is greater than the frequencies of significant energy present in the input signal.
96 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 97

FFT Spectral Leakage

The FFT operation assumes that the time record repeats. Unless there is an integral number of cycles of the sampled waveform in the record, a discontinuity is created at the end of the record. This is referred to as leakage. In order to minimize spectral leakage, windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT. The FFT Menu provides these windows: Hanning, Flat Top, Rectangular, Blackman-Harris, and Bartlett. For more information on leakage, see Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pd f.
FFT Spectral Analysis 4
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide 97
Page 98
4 FFT Spectral Analysis
98 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
Page 99
Keysight InfiniiVision 3000T X-Series Oscilloscopes
User's Guide

5 Math Waveforms

To display math waveforms / 99 To adjust the math waveform scale and offset / 101 Units for Math Waveforms / 101 Math Operators / 102 Math Transforms / 104 Math Filters / 118 Math Visualizations / 121
In addition to the dedicated FFT function (see Chapter 4, “FFT Spectral Analysis,” starting on page 89), you can define two other math functions. One math function waveform can be displayed at a time. The math function waveform is displayed in light purple.
Math functions can be performed on analog channels or they can be performed on lower math functions when using operators other than add, subtract, multiply, or divide.

To display math waveforms

1 Press the [Math] key on the front panel to display the Waveform Math Menu.
2 Press the Display Math softkey and turn the Entry knob to select the math
function you want to display. Then, either push the Entry knob or press the Display Math softkey again to display the selected math function.
99
Page 100
5 Math Waveforms
TIP
3 Use the Operator softkey to select an operator, transform, filter, or visualization.
For more information on the operators, see:
“Math Operators" on page 102
“Math Transforms" on page 104
“Math Filters" on page 118
“Math Visualizations" on page 121
4 Use the Source 1 softkey to select the analog channel, lower math function, or
reference waveform on which to perform math. You can rotate the Entry knob or repetitively press the Source 1 softkey to make your selection.
Higher math functions can operate on lower math functions when using operators other than the simple arithmetic operations (+, -, *, /). For example, if Math 1 is set up as a subtract operation between channels 1 and 2, the Math 2 function could be set up as a FFT operation on the Math 1 function. These are called cascaded math functions.
To cascade math functions, select the lower math function using the Source 1 softkey.
100 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
When cascading math functions, to get the most accurate results, be sure to vertically scale lower math functions so that their waveforms take up the full screen without being clipped.
5 If you selected an arithmetic operator for the math function, use the Source 2
softkey to select the second source for the arithmetic operation.
6 To re-size and re-position the math waveform, see “To adjust the math
waveform scale and offset" on page 101.
Loading...