Keysight DSOX1204A, DSOX1204G Service Manual

Page 1
Keysight InfiniiVision DSOX1204A/G Oscilloscopes
Service Guide
Page 2
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2008-2019
Manual Part Number
N2137-97013
Edition
February 2019
Available in electronic format only
Published by: Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
A newer version of this manual may be available at
www.keysight.com/manuals/ DSOX1204G
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Key­sight shall not be liable for errors or for inci­dental or consequential damages in connection with the furnishing, use, or per­formance of this document or of any infor­mation contained herein. Should Keysight and the user have a separate written agree­ment with warranty terms covering the
material in this document that conflict with these terms, the warranty terms in the sepa­rate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial
computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 DSOX1204A/G Oscilloscopes Service Guide
Page 3
In This Service Guide
This book provides the service information for the Keysight InfiniiVision DSOX1204A/G oscilloscopes. This manual is divided into these chapters:
1 Characteristics and Specifications
This chapter contains a partial list of characteristics and specifications for the Keysight InfiniiVision DSOX1204A/G oscilloscopes.
2 Testing Performance
This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
3 Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating performance.
4 Troubleshooting
This chapter begins with suggestions for solving general problems that you may encounter with the oscilloscope. Procedures for troubleshooting the oscilloscope follow the problem solving suggestions.
5 Replacing Assemblies
The service policy for DSOX1204A/G oscilloscopes is unit replacement, so there are no instructions for replacing internal assemblies in this service guide.
6Replaceable Parts
Because the service policy for the DSOX1204A/G oscilloscopes is unit replacement, no replaceable parts are available.
7Safety Notices
At the front of the book you will find safety notice descriptions and document warranties.
DSOX1204A/G Oscilloscopes Service Guide 3
Page 4
Abbreviated instructions for pressing a series of keys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing Key1, then pressing Softkey2, then pressing Softkey3 are abbreviated as follows: Press [Key1] & Softkey2 & Softkey3. The keys may be front panel keys, or softkeys, which are located directly below the oscilloscope display.
4 DSOX1204A/G Oscilloscopes Service Guide
Page 5

Contents

In This Service Guide / 3
Figures / 7
Tables / 9
1 Characteristics and Specifications
Power Requirements / 11
Measurement Category / 12
Measurement Category / 12 Measurement Category Definitions / 12
Maximum Input Voltages / 13
Environmental Conditions / 13
Specifications / 14
Contact Us / 14
2 Testing Performance
Overview / 16
List of Test Equipment / 17
Conventions / 18
To verify DC vertical gain accuracy / 19
To verify bandwidth (-3 dB) / 24
To verify time base accuracy / 28
To verify trigger sensitivity / 30
Test Internal Trigger Sensitivity / 31
DSOX1204A/G Oscilloscopes Service Guide 5
Page 6
Contents
Keysight DSOX1204A/G Oscilloscopes Performance Test
Record / 34
3 Calibrating and Adjusting
User Calibration / 38
To perform User C al / 38 User Cal Status / 38
4 Troubleshooting
Solving General Problems with the Oscilloscope / 40
If there is no display / 40 If there is no trace display / 40 If the trace display is unusual or unexpected / 40 If you cannot see a channel / 41
Verifying Basic Operation / 42
To power-on the oscilloscope / 42 To perform hardware self test / 42 To perform front panel self test / 43 To verify default setup / 43 To perform an Auto Scale on the Probe Comp signal / 45 To compensate passive probes / 46
5 Replacing Assemblies
6 Replaceable Parts
7 Safety Notices
Warnings / 52
To clean the instr ument / 53
Safety Symbols / 54
Index
6 DSOX1204A/G Oscilloscopes Service Guide
Page 7

Figures

Figure 1. Setting up Equipment for DC Vertical Gain Accuracy Test / 22 Figure 2. Using a Blocking Capacitor to Reduce Noise / 23 Figure 3. Setting Up Equipment for Bandwidth (-3 dB) Verification Test / 25 Figure 4. Setting Up Equipment for Internal Trigger Sensitivity Test / 32 Figure 5. Default setup screen / 44 Figure 6. Example pulses / 46
DSOX1204A/G Oscilloscopes Service Guide 7
Page 8
Figures
8 DSOX1204A/G Oscilloscopes Service Guide
Page 9

Tables

Table 1. List of test equipment / 17 Table 2. Conventions / 18 Table 3. DC Vertical Gain Accuracy Test Limits / 19 Table 4. Equipment Required to Verify DC Vertical Gain Accuracy / 19 Table 5. Settings Used to Verify DC Vertical Gain Accuracy / 20 Table 6. Bandwidth (-3 dB) Test Limits / 24 Table 7. Equipment Required to Verify Bandwidth (-3 dB) / 24 Table 8. Time Base Accuracy Test Limits / 28 Table 9. Equipment Required to Verify Time Base Accuracy / 28 Table 10. Equipment Required to Verify Trigger Sensitivity / 30 Table 11. Internal Trigger Sensitivity Specifications / 31
DSOX1204A/G Oscilloscopes Service Guide 9
Page 10
Tables
10 DSOX1204A/G Oscilloscopes Service Guide
Page 11
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide
1 Characteristics and
Specifications
Power Requirements / 11 Measurement Category / 12 Maximum Input Voltages / 13 Environmental Conditions / 13 Specifications / 14 Contact Us / 14
This chapter contains a partial list of characteristics and specifications for the Keysight InfiniiVision DSOX1204A/G oscilloscopes.
For a full list of Keysight InfiniiVision DSOX1204A/G oscilloscopes characteristics and specifications see the data sheets.
The data sheets are available at www.keysight.com/products/DSOX1204G.

Power Requirements

Line voltage, frequency, and power:
• ~Line 100-120 Vac, 50/60/400 Hz
• 100-240 Vac, 50/60 Hz
•50W max
11
Page 12
1 Characteristics and Specifications
WARNING

Measurement Category

Measurement Category

The InfiniiVision DSOX1204A/G oscilloscope are not intended to be used for measurements in Measurement Category II, III, or IV.
Use this instrument only for measurements within its specified measurement category (not rated for CAT II, III, IV). No transient overvoltages allowed.

Measurement Category Definitions

The "Not rated for CAT II, III, IV" measurement category is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits. In the latter case, transient stresses are variable; for that reason, the transient withstand capability of the equipment is made known to the user.
Measurement category II is for measurements performed on circuits directly connected to the low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
12 DSOX1204A/G Oscilloscopes Service Guide
Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units.
Page 13

Maximum Input Voltages

CAUTION
CAUTION
Maximum input voltage for analog inputs:
150 Vrms, 200 Vpk
Maximum input voltage for external trigger input:
30 Vrms

Environmental Conditions

Environment Indoor use only.
Characteristics and Specifications 1
Ambient temperature
Humidity Operating: Up to 95% RH, non-condensing to temperatures up to +40 °C
Altitude Operating: Up to 3,000 m
Overvoltage Category
Pollution Degree The InfiniiVision DSOX1204A/G oscilloscopes may be operated in environments
Pollution Degree Definitions
Operating: 0 Non-operating: –40 °C to +70 °C
decreasing linearly to 50% RH at +50 °C
Non-operating: Up to 90% RH up to +65 °C (non condensing)
Non-operating: Up to 15,300 m
This product is intended to be powered by MAINS that comply to Overvoltage Category II, which is typical of cord-and-plug connected equipment.
of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. Example: A clean room or climate controlled office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. Example: Sheltered outdoor environment.
°C to +50 °C
DSOX1204A/G Oscilloscopes Service Guide 13
Page 14
1 Characteristics and Specifications

Specifications

Please see the InfiniiVision DSOX1204A/G Oscilloscopes Data Sheet for complete, up-to-date specifications and characteristics.
To download a copy of the data sheet please visit:
www.keysight.com/products/DSOX1204G.
Or go to the Keysight home page at www.keysight.com and search for DSOX1204A/G oscilloscopes data sheet.
To order a data sheet by phone, please contact your local Keysight office.

Contact Us

To contact Keysight, see: www.keysight.com/find/contactus
14 DSOX1204A/G Oscilloscopes Service Guide
Page 15
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

2 Testing Performance

Overview / 16 List of Test Equipment / 17 To verify DC vertical gain accuracy / 19 To verify bandwidth (- 3 dB) / 24 To verify time base accuracy / 28 To verify trigger sensitivity / 30 Keysight DSOX1204A/G Oscilloscopes Performance Test Record / 34
This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
15
Page 16
2 Testing Performance

Overview

Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the oscilloscope warm up 30 minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance Test Record on page 34. To verify whether a test passes, verify that the reading is within the limits in the Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure. Press the following keys to access User Cal: [Utility]→Service→Start User Cal.
16 DSOX1204A/G Oscilloscopes Service Guide
Page 17
Testing Performance 2

List of Test Equipment

Below is a list of test equipment and accessories required to perform the performance test verification procedures.
Table 1 List of test equipment
Equipment Critical Specifications Recommended Model/
Part Number
Digital Multimeter 0.1 mV resolution, 0.005% accuracy Keysight 34401A
Power Splitter Outputs differ by 0.15 dB Keysight 11667B
Oscilloscope Calibrator DC offset voltage of -5.5 V to 70.5 V, 0.1 V resolution
25 MHz—500 MHz sine wave, 5 ppm
Signal Generator 25 MHz, 100 MHz, 350 MHz, 500 MHz, and 1 GHz
sine waves
Power Meter 1 GHz Power Sensor 1 GHz ±3% accuracy Keysight E9304A or
BNC banana cable BNC (m) to dual banana Pomona 2BC-BNC-36 or
BNC cable (qty 3) BNC - BNC, 48” length Keysight 10503A
Cable Type N (m) 609.6 mm (24 in.) Keysight 11500B
Adapter BNC(f) to banana(m) Keysight 1251-2277
Adapter BNC Tee (m) (f) (f) Keysight 1250-0781† or
Adapter Type N (m) to BNC (m) Keysight 1250-0082 or
Blocking capacitor and shorting cap Note: if a BNC blocking capacitor is not available use an
Adapter (qty 3) N(m) to BNC(f) Keysight 1250-0780
50 Ohm Feedthrough Termination 50
Most parts and equipment are available at www.keysight.com. See respective manufacturer’s websites for their equipment.
These parts available at www.parts.keysight.com at the time this manual was published.
±3% accuracy Keysight N1914A
SMA blocking capacitor.
Ω BNC (f) to BNC (m) Keysight 0960-0301
Fluke 5820A
Keysight N5181A
N8482A
Keysight 11001-66001
Pomona 3285
Pomona 3288 with Pomona 3533
Keysight 11742A + Pomona 4288 + Pomona 5088
DSOX1204A/G Oscilloscopes Service Guide 17
Page 18
2 Testing Performance

Conventions

The following conventions will be used when referring to oscilloscope models throughout this chapter.
Table 2 Conventions
Models Referred to as:
DSOX1204A, DSOX1204G 70 MHz Models
DSOX1204A, DSOX1204G with D1200BW1A upgrade 100 MHz Models
DSOX1204A, DSOX1204G with D1200BW2A upgrade (from 70 MHz) or D1200BW3A upgrade (from 100 MHz)
200 MHz Models
18 DSOX1204A/G Oscilloscopes Service Guide
Page 19

To verify DC vertical gain accuracy

This test verifies the accuracy of the analog channel DC vertical gain for each channel.
In this test, you will measure the dc voltage output of an oscilloscope calibrator using the oscilloscope’s Average - Full Screen voltage measurement and compare the results with the multimeter reading.
Table 3 DC Vertical Gain Accuracy Test Limits
Test Limits Notes
Testing Performance 2
±3% of full scale (>= 10 mV/div); ±4% of full scale (< 10 mV/div)
Full scale is defined as 8 mV on the 500 uV/div range.
Full scale on all other ranges is defined as 8 divisions
times the V/div setting.
Table 4 Equipment Required to Verify DC Vertical Gain Accuracy
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator 3.5 mV to 70 Vdc,
0.1 V resolution
Digital multimeter Better than 0.01% accuracy Keysight 34401A
Cable BNC, Qty 2 Keysight 10503A
Shorting cap BNC Keysight 1250-0774
Adapter BNC (f) to banana (m) Keysight 1251-2277
Adapter BNC tee (m) (f) (f ) Keysight 1250-0781 or Pomona
Blocking capacitor Keysight 11742A +
Fluke 5820A
3285
Pomona 4288 + Pomona 5088
1 Press [Save/Recall] > Default/Erase > Factory Default to recall the factory default
setup.
2 Set the probe attenuation to 1:1 on the analog channel you are testing (for
example, [1] > Probe > Probe; then, turn the Entry knob to select 1.00 : 1).
DSOX1204A/G Oscilloscopes Service Guide 19
Page 20
2 Testing Performance
3 Set up the oscilloscope.
Table 5 Settings Used to Verify DC Vertical Gain Accuracy
a Adjust the horizontal scale to 200.0 us/div. b Set the Volts/Div setting to the value in the first line and column of Table 5 . c Adjust the channel’s vertical position knob to place the baseline (reference
level) at 0.5 major division from the bottom of the display.
Volts/Div Setting Oscilloscope Calibrator
Setting
10 V/Div 70 V 67.6 V to 72.4 V
5 V/Div 35 V 33.8 V to 36.2 V
2 V/Div 14 V 13.52 V to 14.48 V
1V/Div 7V 6.76V to 7.24V
500 mV/Div 3.5 V 3.38 V to 3.62 V
200 mV/Div 1.4 V 1.352 V to 1.448 V
100 mV/Div 700 mV 676 mV to 724 mV
50 mV/Div 350 mV 338 mV to 362 mV
20 mV/Div 140 mV 135.2 mV to 144.8 mV
10 mV/Div 70 mV 67.6 mV to 72.4 mV
5mV/Div
2mV/Div
1mV/Div
0.5 mV/Div
1
Reduce Noise" on page 23.
1
1
1
1, 2
A blocking capacitor is required at this range to reduce noise. See “Use a Blocking Capacitor to
35 mV 33.4 mV to 36.6 mV
14 mV 13.36 mV to 14.64 mV
7 mV 6.68 mV to 7.32 mV
3.5mV 3.18mV to 3.82mV
Test Limits
2
Full scale is defined as 8 mV on the 500 uV/div range. Full scale on all other ranges is defined as 8
divisions times the V/div setting.
20 DSOX1204A/G Oscilloscopes Service Guide
Page 21
Testing Performance 2
d Press the [Acquire] key. e Then press the Acq Mode softkey and select Averaging. f Then press the #Avgs softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4 Add a measurement for the average voltage:
a Press the [Meas] key. b Press Source; then, turn the Entry knob (labeled on the front panel) to
select the channel you are testing.
c Press Type:; then, turn the Entry knob to select Average - Full Screen, and press
Add Measurement. 5 Read the “current” average voltage value as V1. 6 Use the BNC tee and cables to connect the oscilloscope calibrator/power
supply to both the oscilloscope and the multimeter (see Figure 1).
DSOX1204A/G Oscilloscopes Service Guide 21
Page 22
2 Testing Performance
Oscilloscope
Oscilloscope Calibrator
Digital Multimeter
BNC Tee
BNC (f) to dual bananna adapter

Figure 1 Setting up Equipment for DC Vertical Gain Accuracy Test

7 Adjust the output so that the multimeter reading displays, as close as possible,
the first Volts/div calibrator setting value in Tab le 5.
8 Wait until the measurement settles. 9 Read the “current” average voltage value again as V2. 10 Calculate the difference V2 - V1.
22 DSOX1204A/G Oscilloscopes Service Guide
Page 23
Testing Performance 2
To oscilloscope input
BNC shorting cap
Blocking Capacitor
The difference in average voltage readings should be within the test limits of
Table 5.
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then return here.
11 Disconnect the oscilloscope calibrator from the oscilloscope. 12 Repeat this procedure to check the DC vertical gain accuracy with the
remaining Volts/div setting values in Table 5.
13 Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 0.5 mV/div, 1 mV/div, 2 mV/div, and 5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and shorting cap at the oscilloscope channel input to shunt the noise to ground. See Figure 2. If a BNC capacitor is not available, use an SMA blocking capacitor, adapter, and cap. See “Blocking capacitor and shorting cap in the equipment list on page 17 for details.

Figure 2 Using a Blocking Capacitor to Reduce Noise

DSOX1204A/G Oscilloscopes Service Guide 23
Page 24
2 Testing Performance

To verify bandwidth (-3 dB)

This test checks the bandwidth (-3 dB) of the oscilloscope. In this test you will use a signal generator and a power meter.
Table 6 Bandwidth (-3 dB) Test Limits
Models Test Limits
200 MHz Models All channels (-3 dB), dc to 200 MHz
100 MHz Models All channels (-3 dB), dc to 100 MHz
70 MHz Models All channels (-3 dB), dc to 70 MHz
Table 7 Equipment Required to Verify Bandwidth (-3 dB)
Equipment Critical Specifications Recommended Model/Part
Signal Generator 100 kHz - 200 MHz at 200 mVrms Keysight N5181A
Power Meter 1 MHz - 200 MHz ±3% accuracy Keysight N1914A
Power Sensor 1 MHz - 200 MHz ±3% accuracy Keysight E9304A or N8482A
Power Splitter outputs differ by < 0.15 dB Keysight 11667A
Cable Type N (m) 24 inch Keysight 11500B
Adapter Type N (m) to BNC (m) Keysight 1250-0082 or
Pomona 3288 with Pomona 3533
50 Ohm Feedthrough Termination
1 Connect the equipment (see Figure 3).
a Use the N cable to connect the signal generator to the input of the power
splitter input.
b Connect the power sensor to one output of the power splitter. c Use an N-to-BNC adapter to connect the other splitter output to the
channel 1 input using a 50 ohm feedthrough terminator at the oscilloscope
input BNC.
24 DSOX1204A/G Oscilloscopes Service Guide
Ω BNC (f) to BNC (m) Keysight 0960-0301
50
Page 25
50 Ohm Feedthrough
N to BNC Adapter
Testing Performance 2
Oscilloscope
Signal Generator
Power Splitter
Power Sensor
N Cable
Power Meter

Figure 3 Setting Up Equipment for Bandwidth (-3 dB) Verification Test

2 Set up the power meter.
Set the power meter to display measurements in units of watts.
DSOX1204A/G Oscilloscopes Service Guide 25
Page 26
2 Testing Performance
Vin
1MHz
Pmeas
1MHz
50Ω×=
3 Set up the oscilloscope.
a Press the [Default Setup] key. b Set channel 1 Coupling to DC. c Set the time base to 500 ns/div. d Set the Volts/Div for channel 1 to 200 mV/div. e Press the [Acquire] key, then press the Averaging softkey. f Turn the Entry knob to set # Avgs to 8 averages.
4 Set the signal generator for 1 MHz and six divisions of amplitude.
The signal on the oscilloscope screen should be about five cycles at six divisions amplitude.
5 Set up the Amplitude measurement
a Press the [Meas] key. b Press the Clear Meas softkey and then the Clear All softkey. c Press the Type: softkey and use the Entry knob to select AC RMS - N Cycles
within the select menu.
d Press the Add Measurement softkey.
6 Note the oscilloscope AC RMS - Cyc(1) reading at the bottom of the screen.
(This is the RMS value with any dc offset removed.)
7 Set the power meter Cal Factor % to the 1 MHz value on the calibration chart
on the power sensor.
8 Note the reading on the power meter and covert to Vrms using the expression:
For example, if the power meter reading is 892 uW, then Vin
1/2
50Ω)
9 Change the signal generator output frequency according to the maximum
frequency for the oscilloscope using the following:
200 MHz Models: 200 MHz
100 MHz Models: 100 MHz
70 MHz Models: 70 MHz
10 Referencing the frequency from step 9, set the power meter Cal Factor % to the
frequency value on the calibration chart on the power sensor.
26 DSOX1204A/G Oscilloscopes Service Guide
= 211.2 mV
rms
.
= (892*10-6 *
1MHz
Page 27
Example If:
Vin
maxfreq
Pmeas
maxfreq
50Ω×=
response(dB) = 20 log
10
 
 
MHz 1 MHz 1
freqmax freqma x
Vin / Vout
Vin / Vout
response(dB) = 20 log
10
mV 185.3 / mV 161.6
= -1.16 dB
Testing Performance 2
11 Set the oscilloscope sweep speed according to the following:
200 MHz Models: 2 ns/div
100 MHz Models: 5 ns/div
70 MHz Models: 5 ns/div
12 Note the oscilloscope AC RMS - Cyc(1) reading at the bottom of the screen. 13 Note the reading on the power meter and covert to Vrms using the expression:
14 Calculate the response using the expression:
Pmeas Std Dev(n) Pmeas Std Dev(n)
Then after converting the values from the power meter to Vrms:
1_MHz
1MHz
max_freq
max freq
= 892 uW
= 210.4 mV
= 687 uW
= 161.6 mV
DSOX1204A/G Oscilloscopes Service Guide 27
mV211.2mV/ 210.4
15 The result from step 14 should be between +3.0 dB and -3.0 dB. Record the
result in the Performance Test Record (see page 34).
16 Move the power splitter from the channel 1 to the channel 2 input. 17 Turn off the current channel and turn on channel 2 using the channel keys. 18 Set the trigger source to channel 2. 19 Repeat steps 3 through 15 for the remaining channels, setting the parameters
of the channel being tested where appropriate.
Page 28
2 Testing Performance

To verify time base accuracy

This test verifies the accuracy of the time base.
Table 8 Time Base Accuracy Test Limits
Models Test Limits
All Models 50 ppm ±5 ppm per year (aging)
In this test you will measure the absolute error of the time base oscillator and compare the results to the specification.
Table 9 Equipment Required to Verify Time Base Accuracy
Equipment Critical Specifications Recommended Model/Part
Signal Generator 100 kHz - 200 MHz, 0.01 Hz frequency
resolution, jitter: < 2ps
Cable BNC, 3 feet Keysight 10503A
50 Ohm Feedthrough Termination
1 Set up the signal generator.
a Set the output to 10 MHz, approximately 1 V
2 Connect the output of the signal generator to oscilloscope channel 1 using the
BNC cable. Also, connect a 50 ohm feedthrough termination between the channel 1 input and the BNC cable.
Ω BNC (f) to BNC (m) Keysight 0960-0301
50
Keysight N5181A
sine wave.
pp
28 DSOX1204A/G Oscilloscopes Service Guide
Page 29
Testing Performance 2
3 Set up the oscilloscope:
a Press [AutoScale]. b Set the oscilloscope Channel 1 vertical sensitivity to 200 mv/div. c Set the oscilloscope horizontal sweep speed control to 10 ns/div. d Adjust the intensity to get a sharp, clear trace. e Adjust the oscilloscope’s trigger level so that the rising edge of the
waveform at the center of the screen is located where the center horizontal
and vertical grid lines cross (center screen).
f Ensure the horizontal position control is set to 0.0 seconds.
4 Make the measurement.
a Set oscilloscope horizontal sweep speed control to 1 ms/div. b Set horizontal position control to +1 ms (rotate control CCW). c Set the oscilloscope horizontal sweep speed control to 10 ns/div. d Record the number of nanoseconds from where the rising edge crosses the
center horizontal grid line to the center vertical grid line. The number of
nanoseconds is equivalent to the time base error in ppm.
e Derive the date code from the oscilloscope’s serial number to calculate the
number of years since manufacture. Include any fractional portion of a year.
i First, get the 3rd to 6th digits of the serial number. For example, for the
serial number CN47470001, get “4747”.
ii If the number is greater than 4000, subtract 4000 to get the Date Code.
For example, 4747-4000 is “0747”.
iii If the number is smaller than 4000, add 6000 to get the Date Code.
iv In the Date Code, the first two digits represent the year, and the second
two digits represent the week in the year. For example, for “0747”, the year is “07” and the week is “47”.
f Use the following formula to calculate the test limits.
Time base accuracy limit: 50 ppm ±5 ppm per year (aging)
g Record the result and compare it to the limits in the Performance Test
Record (see page 34).
DSOX1204A/G Oscilloscopes Service Guide 29
Page 30
2 Testing Performance

To verify trigger sensitivity

These tests verify the trigger sensitivity. In these tests, you will apply a sine wave to the oscilloscope at two bandwidths: 10 MHz and the maximum bandwidth of the oscilloscope. For each sine wave, you will decrease the amplitude of the signal to a specified level and check to see if the oscilloscope still triggers.
Table 10 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Signal Generator 10 MHz and 70 MHz, 100 MHz, or
200 MHz sine waves
Power splitter Outputs differ < 0.15 dB Keysight 11667A
Power Meter Keysight N1914A
Power Sensor Keysight E9304A or N8482A
Cable BNC, Qty 3 Keysight 10503A
Adapter N (m) to BNC (f), Qty 3 Keysight 1250-0780
Feedthrough 50
Ω BNC (f) to BNC (m) Keysight 0960-0301
Keysight N5181A
30 DSOX1204A/G Oscilloscopes Service Guide
Page 31

Test Internal Trigger Sensitivity

Table 11 Internal Trigger Sensitivity Specifications
Testing Performance 2
Oscilloscope Models
70 MHz greater of 0.6 div or 2.5 mV greater of 0.9 div or 3.8 mV
100 MHz greater of 0.6 div or 2.5 mV greater of 1.2 div or 5 mV
200 MHz greater of 0.6 div or 2.5 mV greater of 1.2 div or 5 mV
Follow these steps to perform the internal trigger sensitivity test:
1 On the oscilloscope, press the [Default Setup] key. 2 Press the [Mode/Coupling] key; then, press the Mode softkey to select Normal. 3 Connect the equipment (see Figure 4).
a Connect the signal generator output to the oscilloscope channel 1 input. b Connect a 50 ohm feedthrough termination between the channel 1 input
and the BNC cable.
Internal Trigger Sensitivity Specification Bandwidth
<= 10 MHz Maximum Bandwidth
DSOX1204A/G Oscilloscopes Service Guide 31
Page 32
2 Testing Performance
Oscilloscope
Signal Generator
50 Ohm Feedthrough
N to BNC Adapter
N Cable
Figure 4 Setting Up Equipment for Internal Trigger Sensitivity Test
4 Set the output frequency of the signal generator to 10 MHz. 5 Perform these steps to test at the 10 MHz setting:
a Set the signal generator amplitude to about 20 mV b Press the [AutoScale] key. c Set the time base to 10 ns/div. d Set channel 1 to 10 mV/div. e Decrease the amplitude from the signal generator until 0.6 vertical divisions
of the signal (about 6 mV
32 DSOX1204A/G Oscilloscopes Service Guide
) is displayed.
pp
pp
.
Page 33
Testing Performance 2
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
f Record the result as Pass or Fail in the Performance Test Record (see
page 34).
g Repeat this step for the remaining oscilloscope channels.
6 Set the output frequency of the signal generator to the maximum bandwidth of
the oscilloscope:
200 MHz models: 200 MHz
100 MHz models: 100 MHz
70 MHz models: 70 MHz
7 Perform these steps to test at the maximum bandwidth setting:
a Set the signal generator amplitude to about 20 mV b Press the [AutoScale] key. c Set the time base to 10 ns/div. d Set channel 1 to 10 mV/div. e Decrease the amplitude from the signal generator as described in the
following table according to your oscilloscope’s maximum bandwidth:
pp
.
Oscilloscope’s maximum bandwidth:
70 MHz 0.9 vertical divisions 9 mVpp
100 MHz 1.2 vertical divisions 12 mVpp
200 MHz 1.2 vertical divisions 12 mVpp
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
f Record the result as Pass or Fail in the Performance Test Record (see
page 34).
g Repeat this step for the remaining oscilloscope channels.
DSOX1204A/G Oscilloscopes Service Guide 33
Decrease amplitude until these vertical divisions are displayed:
Which is about this V:
Page 34
2 Testing Performance

Keysight DSOX1204A/G Oscilloscopes Performance Test Record

Serial No. ______________________________________ Test by _____________________________ Test Interval ____________________________________ Work Order No. ______________________ Recommended Next Testing ________________________ Temperature ____________
DC Vertical Gain Accuracy Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3 Channel 4
10 V/Div 70 V 67.6 V to 72.4 V ________ ________ ________ ________ 5 V/Div 35 V 33.8 V to 36.2 V ________ ________ ________ ________ 2 V/Div 14 V 13.52 V to 14.48 V ________ ________ ________ ________ 1 V/Div 7 V 6.76 V to 7.24 V ________ ________ ________ ________ 500 mV/Div 3.5 V 3.38 V to 3.62 V ________ ________ ________ ________ 200 mV/Div 1.4 V 1.352 V to 1.448 V ________ ________ ________ ________ 100 mV/Div 700 mV 676 mV to 724 mV ________ ________ ________ ________ 50 mV/Div 350 mV 338 mV to 362 mV ________ ________ ________ ________ 20 mV/Div 140 mV 135.2 mV to 144.8 mV ________ ________ ________ ________ 10 mV/Div 70 mV 67.6 mV to 72.4 mV ________ ________ ________ ________ 5 mV/Div 35 mV 33.4 mV to 36.6 mV ________ ________ ________ ________ 2 mV/Div 14 mV 13.36 mV to 14.64 mV ________ ________ ________ ________ 1 mV/Div 7 mV 6.68 mV to 7.32 mV ________ ________ ________ ________
0.5 mV/Div 3.5 mV 3.18 mV to 3.82 mV ________ ________ ________ ________
Bandwidth (-3 dB) Model Test Limits Channel 1 Channel 2 Channel 3 Channel 4
200 MHz -3 dB at 200 MHz ________ ________ ________ ________ 100 MHz -3 dB at 100 MHz ________ ________ ________ ________ 70 MHz -3 dB at 70 MHz ________ ________ ________ ________
Time Base Accuracy Limits Calculated
time base accuracy limit (ppm)
Time Base Accuracy Limit: 50 ppm ±5 ppm per year (aging)
_________ ________ ________
Measured time base error (ppm)
Pass/Fail
34 DSOX1204A/G Oscilloscopes Service Guide
Page 35
Testing Performance 2
Internal Trigger Sensitivity
Generator Setting Test Limits, greater of Channel 1 Channel 2 Channel 3 Channel 4
200 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (200 MHz) 1.2 div or 5 mV ________ ________ ________ ________
100 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (100 MHz) 1.2 div or 5 mV ________ ________ ________ ________
70 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (70 MHz) 0.9 div or 3.8 mV ________ ________ ________ ________
DSOX1204A/G Oscilloscopes Service Guide 35
Page 36
2 Testing Performance
36 DSOX1204A/G Oscilloscopes Service Guide
Page 37
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

3 Calibrating and Adjusting

This chapter explains how to adjust the oscilloscope for optimum operating performance. You should perform user calibration according to the following recommendations:
• Every five years or after 10000 hours of operation
• If the ambient temperature is >10 °C from the calibration temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope warm up for at least 30 minutes.
37
Page 38
3 Calibrating and Adjusting

User Calibration

Perform user-calibration:
• Every five years or after 10000 hours of operation.
• If the ambient temperature is >10° C from the calibration temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter User Cal intervals.
User Cal performs an internal self-alignment routine to optimize the signal path in the oscilloscope. The routine uses internally generated signals to optimize circuits that affect channel sensitivity, offset, and trigger parameters. Disconnect all inputs and allow the oscilloscope to warm up before performing this procedure.
Performing User Cal will invalidate your Certificate of Calibration. If NIST (National Institute of Standards and Technology) traceability is required perform the procedures in Chapter 2 in this book using traceable sources.

To perfo rm Use r Cal

1 Disconnect all inputs from the front panel and allow the oscilloscope to warm
up before performing this procedure.
2 Make sure user calibration protection is not on by pressing the [Utility] > Options
> Auxiliary; then, press the Cal Protect softkey to disable calibration protection.
3 Press the [Utility] key; then, press the Service softkey. 4 Begin the User Cal by pressing the Start User Cal softkey.

User Cal Status

Pressing the User Cal Status softkey displays the following summary results of the previous User Cal, and the status of probe calibrations for probes that can be calibrated. Note that AutoProbes do not need to be calibrated, but InfiniiMax probes can be calibrated.
Results: User Cal date: Change in temperature since last User Cal: Failure: Comments: Probe Cal Status:
38 DSOX1204A/G Oscilloscopes Service Guide
Page 39
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

4 Troubleshooting

Solving General Problems with the Oscilloscope / 40 Verifying Basic Operation / 42
Read All Cautions and Warnings
Before you begin any troubleshooting, read all Warning and Cautions in the “Troubleshooting” section.
This chapter begins with “Solving General Problems with the Oscilloscope. It tells you what to do in these cases:
If there is no display.
If there is no trace display.
If the trace display is unusual or unexpected.
If you cannot see a channel.
Next, this chapter describes procedures for “Verifying Basic Operation of the oscilloscope:
To power-on the oscilloscope.
To perform hardware self test.
To perform front panel self test.
To verify default setup.
To perform an Auto Scale on the Probe Comp signal.
To compensate passive probes.
The service policy for all DSOX1204A/G oscilloscopes is unit replacement, so there are no internal assembly troubleshooting instructions in this service guide.
39
Page 40
4 Troubleshooting

Solving General Problems with the Oscilloscope

This section describes how to solve general problems that you may encounter while using the Keysight DSOX1204A/G oscilloscopes.

If there is no display

Check that the power cord is firmly seated in the oscilloscope power
receptacle.
Check that the power source is live.
Check that the front-panel power switch is on.
If there is still no display, go to the troubleshooting procedures in this chapter.

If there is no trace display

Check that the Intensity (on the front panel) is adjusted correctly.
Recall the default setup by pressing [Default Setup]. This will ensure that the
trigger mode is Auto.
Check that the probe clips are securely connected to points in the circuit under
test, and that the ground is connected.
Check that the circuit under test is powered on.
Press the [Auto Scale] key.
Obtain service from Keysight Technologies, if necessary.

If the trace display is unusual or unexpected

Check that the Horizontal time/division setting is correct for the expected
frequency range of the input signals.
The sampling speed of the oscilloscope depends on the time/division setting. It
may be that when time/division is set to slower speeds, the oscilloscope is sampling too slowly to capture all of the transitions on the waveform. Use peak detect mode.
Check that all oscilloscope probes are connected to the correct signals in the
circuit under test.
40 DSOX1204A/G Oscilloscopes Service Guide
Page 41
Ensure that the probe’s ground lead is securely connected to a ground point in
the circuit under test. For high-speed measurements, each probe’s individual ground lead should also be connected to a ground point closest to the signal point in the circuit under test.
Check that the trigger setup is correct.
A correct trigger setup is the most important factor in helping you capture the
data you desire. See the User’s Guide for information about triggering.
Check that persistence in the Display menu is turned off, then press the
Clear Display softkey.
Press the [Auto Scale] key.

If you cannot see a channel

Recall the default setup by pressing [Default Setup]. This will ensure that the
trigger mode is Auto.
Check that the oscilloscope probe’s BNC connector is securely attached to the
oscilloscope’s input connector.
Troubleshooting 4
Check that the probe clips are securely connected to points in the circuit under
test.
Check that the circuit under test is powered on.
You may have pressed the [Auto Scale] key before an input signal was available.
Performing the checks listed here ensures that the signals from the circuit under test will be seen by the oscilloscope. Perform the remaining checks in this topic to make sure the oscilloscope channels are on, and to obtain an automatic setup.
Check that the desired oscilloscope channels are turned on.
a Press the analog channel key until it is illuminated.
Press the [Auto Scale] key to automatically set up all channels.
DSOX1204A/G Oscilloscopes Service Guide 41
Page 42
4 Troubleshooting
WARNING

Verifying Basic Operation

To power-on the oscilloscope

1 Connect the power cord to the rear of the oscilloscope, then to a suitable ac
voltage source.
The oscilloscope power supply automatically adjusts for input line voltages in the range of 100 to 240 VAC. Ensure that you have the correct line cord. The power cord provided is matched to the country of origin.
AVOID INJURY.
Always operate the oscilloscope with an approved three conductor power cable. Do not negate the protective action of the three conductor power cable.
• Press the power switch.
When the oscilloscope is turned on, the front panel LEDs will briefly light up
in groups from bottom to top.
Next the Keysight logo appears on the display.
Next a message will appear with tips on getting started using the
oscilloscope. At this time you can press any key to remove the message and view the display. Or you can wait and the message will automatically disappear.
It will take a total of about 20-30 seconds for the oscilloscope to go through
its basic self test and power-up routine.
2 Proceed to “To perform hardware self test" on page 42.
42 DSOX1204A/G Oscilloscopes Service Guide

To perform hardware self test

Pressing [Utility] > Service > Hardware Self Test performs a series of internal procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
• After experiencing abnormal operation.
• For additional information to better describe an oscilloscope failure.
• To verify proper operation after the oscilloscope has been repaired.
Page 43
Successfully passing Hardware Self Test does not guarantee 100% of the oscilloscope's functionality. Hardware Self Test is designed to provide an 80% confidence level that the oscilloscope is operating properly.

To perform front panel self test

Pressing [Utility] > Service > Front Panel Self Test lets you test the front panel keys and knobs as well as the oscilloscope display.
Follow the on-screen instructions.
Failures in the front panel self test indicate problems with the keyboard, keypad, or display.

To verify default setup

The oscilloscope is designed to turn on with the setup from the last turn on or previous setup.
To recall the default setup: 1 Press the [Default Setup] key.
Troubleshooting 4
This returns the oscilloscope to its default settings and places the oscilloscope in a known operating condition. The major default settings are:
Horizontal:
main mode.
100 us/div scale.
0s delay.
center time reference.
Vertical:
Channel 1 on.
5V/div scale.
dc coupling.
0V position.
probe attenuation factor to 10.0:1.
Trigger:
Edge trigger.
Auto sweep mode.
DSOX1204A/G Oscilloscopes Service Guide 43
Page 44
4 Troubleshooting
0 V level.
channel 1 source.
dc coupling.
rising edge slope.
60 ns holdoff time.
Display:
20% grid intensity.
persistence off.
Other:
Acquire mode normal.
Run/Stop to Run.
cursor measurements off.
Figure 5 Default setup screen
2 If your screen looks substantially different, replace the system board.
44 DSOX1204A/G Oscilloscopes Service Guide
Page 45

To perform an Auto Scale on the Probe Comp signal

1 Press the [Default Setup] key. The oscilloscope is now configured to its default
settings.
2 Connect an oscilloscope probe from channel 1 to the Probe Comp signal terminal
on the front panel.
3 Connect the probe’s ground lead to the ground terminal that is next to the Demo
2 (Probe Comp) terminal. 4 Press [AutoScale]. 5 You should see a waveform on the oscilloscope’s display similar to this:
Troubleshooting 4
If you see the waveform, but the square wave is not shaped correctly as shown
above, perform the procedure “To compensate passive probes" on page 46.
If you do not see the waveform, ensure your power source is adequate, the
oscilloscope is properly powered-on, and the probe is connected securely to
the front-panel analog channel input BNC and to the Demo 2 (Probe Comp)
terminal.
DSOX1204A/G Oscilloscopes Service Guide 45
Page 46
4 Troubleshooting
comp.cdr
Perfectly compensated
Over compensated
Under compensated

To compensate passive probes

You should compensate your passive probes to match their characteristics to the oscilloscope’s channels. A poorly compensated probe can introduce measurement errors.
1 Perform the procedure “To perform an Auto Scale on the Probe Comp
signal" on page 45
2 Press the channel key to which the probe is connected ([1], [2]). 3 In the Channel Menu, press Probe. 4 In the Channel Probe Menu, press Probe Check; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On the some probes, the trimmer capacitor is the yellow adjustment on the
probe tip. On other probes, the trimmer capacitor is located on the probe BNC
connector.
Figure 6 Example pulses
5 Connect probes to all other oscilloscope channels (channel 2 of a 2-channel
oscilloscope).
46 DSOX1204A/G Oscilloscopes Service Guide
6 Repeat the procedure for each channel.
The process of compensating the probes serves as a basic test to verify that the oscilloscope is functional.
Page 47
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

5 Replacing Assemblies

The service policy for all DSOX1204A/G oscilloscopes is unit replacement, so there are no instructions for replacing internal assemblies in this service guide.
47
Page 48
5 Replacing Assemblies
48 DSOX1204A/G Oscilloscopes Service Guide
Page 49
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

6 Replaceable Parts

Because the service policy for DSOX1204A/G oscilloscopes is unit replacement, no replaceable parts are available for the Keysight DSOX1204A/G oscilloscopes.
49
Page 50
6 Replaceable Parts
50 DSOX1204A/G Oscilloscopes Service Guide
Page 51
Keysight InfiniiVision DSOX1204A/G Oscilloscope
Service Guide

7 Safety Notices

This apparatus has been designed and tested in accordance with IEC Publication 61010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under “Safety Symbols.”
51
Page 52
7 Safety Notices

Warnings

Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source.
Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation.
Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Do not install substitute parts or perform any unauthorized modification to the instrument.
Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not use the instrument in a manner not specified by the manufacturer.
52 DSOX1204A/G Oscilloscopes Service Guide
Page 53

To clean the instrument

If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.
Safety Notices 7
DSOX1204A/G Oscilloscopes Service Guide 53
Page 54
7 Safety Notices
!

Safety Symbols

Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product.
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
54 DSOX1204A/G Oscilloscopes Service Guide
Page 55

Index

A
accuracy, DC vertical gain, 19 assemblies
replacing, 47
B
bandwidth
verification, 24
C
calibration, 38 certificate of calibration, 38 channels
problem solving, 41 cleaning the instrument, 53 compensating your probe, 46 contact Keysight, 14
D
DC vertical gain accuracy, 19 display
problem solving, 40
E
equipment, test, 17
F
front panel self test, 43
H
hardware self test, 42
I
input voltages, maximum, 13
internal self-tests, 42
K
Keysight
contact information, 14
L
LED, startup sequence, 42
M
measurement category, 12
definitions, 12
N
NIST, 38
O
overvoltage category, 13
P
perform user cal, 38 performance
test record, 34 pollution degree, 13 pollution degree, definitions, 13 power requirements, 11 probes
compensating, 46 problem solving, 39
R
record tests, 34 replacing assemblies, 47 run self-tests, 42
S
self test, front panel, 43 self test, hardware, 42 self-alignment, user cal, 38 self-tests, 42 specifications, 14 startup sequence, 42 status, User Cal, 38 status, user cal, 38
T
test
equipment, 17 record, 34
trace display
problem solving, 40
trigger
sensitivity, 30 sensitivity, internal, 31
troubleshooting, 39
U
user calibration, 38
V
verify
bandwidth, 24 trigger, 30
W
warranted specifications, 14
DSOX1204A/G Oscilloscopes Service Guide 55
Page 56
Index
56 DSOX1204A/G Oscilloscopes Service Guide
Loading...