Keysight E6607B, Agilent X-Series Getting Started Manual

Agilent X-Series E6607B Wireless Communications Test Set
Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com.
Getting Started Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2012-2014 No part of this manual may be reproduced
in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Agi­lent Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgements
Microsoft “ is a U.S. registered trademark of Microsoft Corporation.
Windows “ and MS Windows “ are U.S. registered trademarks of Microsoft Corporation.
Adobe Reader “ is a U.S. registered trademark of Adobe System Incorporated.
Java™ is a U.S. trademark of Sun Microsystems, Inc.
MA TLAB “ is a U.S. registered trademark of Math Works, Inc.
Norton Ghost™ is a U.S. trademark of Symantec Corporation.
Copyright 2008-9 Agilent Technologies Inc.
Licensed under the Apache License, Ver­sion 2.0 (the "License");
you may not use this file except in com­pliance with the License.
You may obtain a copy of the License at: http://www.apache.org/licenses/ LICENSE-2.0
Unless required by applicable law or agreed to in writing, software
distributed under the License is distributed on an "AS IS" BASIS,
WITHOUT WARRANTIES OR CONDI­TIONS OF ANY KIND, either express or implied.
See the License for the specific language governing permissions and lim­itations under the License.
Manual Part Number
E6607-90033
Print Date
December 2014
Printed in USA Agilent Technologies, Inc.
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
The material contained in this doc­ument is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of mer­chantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in con­nection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the mate­rial in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer soft­ware” as defined in DF AR 252.227-7014 (June 1995), or as a “commercial item” as defined in F AR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Gov­ernment users will receive no greater than Limited Rights as defined in FAR 52.227­14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi­cated conditions are fully understood and met.
Warranty
This Agilent Technologies product is warranted against defects in material and workmanship for a period of one year from the date of shipment. during the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about this test set, including firmware upgrades, application information, and product information, see the following URLs:
http://www.agilent.com/find/ext
To receive the latest updates by email, subscribe to Agilent Email Updates:
http://www.agilent.com/find/emailupdates
Information on preventing test set damage can be found at:
http://www.agilent.com/find/tips
Compliance
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Canadian EMC Requirements
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada.
Acoustic Noise Requirements
Acoustic Noise Emission Geraeuschemission
LpA <70 dB Operator Position Normal Position Per ISO 7779
Acoustic Noise -- Further Information
Ambient Temperature <40
LpA < 70 dB Am Arbeitsplatz Normaler Betrieb Nach DIN 45635 t.19
Supplemental Information
(Values given are per ISO 7779 standard in the “Operator Sitting” position)
° C Nominally under 55 dBA Sound Pressure. 55 dBA is gen-
erally considered suitable for use in quiet office environ­ments.
Ambient Temperature >40° C Nominally under 65 dBA Sound Pressure. 65 dBA is gen-
erally considered suitable for noisy office environments. (The fan speed, and thus the noise level, increases with increasing ambient temperature.)
In This Guide…
This guide contains the following information:
1 Quick-Start
This chapter explains how to initial ize the test se t and generate and view a signal.
2 Front and Rear Panel Features
Refer to this chapter for information on front- and rear-panel key functionality, and display annotations.
3 Test Set Operating System
This chapter describes the Microsoft Windows OS configuration and the settings used with the Agilent test set software.
4 Using Microsoft Windows
The information in this chapter provides some guidelines for using the Microsoft Windows feature capabilities with the test set.
5 Troubleshooting
This chapter details some basic steps that may solve any problems you are experiencing with either the test set or Microsoft Windows OS.

Contents

1 Quick-Start
Initial Inspection 12
Table. Verifying the contents 12 Shipping problems? 13 Test Set Location and Rack Mounting Requirements 13
Turning on the test set the first time 14
Initializing the test set using a USB mouse 14
Initializing the test set using front-panel navigation 19 Anti-Virus Software and Firewalls 23 Shortening the Test Set Power-On Time 23 Instrument Information 24
Power requirements 24
Environmental Conditions (Operating) 25
Ventilation 26
Instrument Maintenance 26
Protecting against electrostatic discharge 28
2 Front and Rear Panel Features
Front-Panel Features (E6607A/B) 30
V irtual Front Panel 33
When the Test Set is Controlled Remotely 34 Display Annotations 35 Rear-Panel Features 37 Front and Rear Panel Symbols 39
3 Test Set Operating System
Agilent Software Installed 42
Agilent U9065A & U9060A software 42 Customer Installation of Software 43
3rd party software verified by Agilent 43
Getting Started Guide 7
Installation of other 3rd party software 43 User Accounts 44
Administrator login 44
User login 44
AgilentOnly user account 44
Agilent service user accounts 44
Customer creation of accounts 45 Agilent X-Series Licensing Options 46
Fixed Perpetual 46
Transportable Perpetual 46
Time-Based Licenses 47 Licensing New Measurement Application Software - After Initial Purchase 48 Transporting a License Between X-Series Test Sets 50 Windows Configuration 53
Settings that can be changed 53
Settings that must not be changed 54 Configuring Printers 56 Configuring LAN 57
Hostname 57
IP Address & Gateway 57 Windows Security 58
Windows Firewall 59
Automatic updates 59
Virus protection 60
Spyware protection 61 System Maintenance 62
Backup 62
System restore 62
Disk defragmenting 62 USB Connections 63 Hard Drive Partitioning and Use 64
Backing-up the factory calibration data 65 Hard Drive Recovery Process 67
Updating the software 69
Configuring recovery prompt timing 71
8 Getting Started Guide
4 Using Microsoft Windows Operating System
Navigating Windows Without a Mouse 74 Remote Desktop: Using the X-Series test set remotely 78
Overview of Remote Desktop operation 78
Setting up Remote Desktop operation 78
How to locate the computer name of the test set 80
Running a Remote Desktop session 82
The Virtual Front Panel 91 Embedded Web Server: Using the X-Series Test Set Remotely 92
Accessing the test set through the Internet 92
Accessing the Web Control tab 96
Selecting the Get Data tab 97
Selecting the Get Image tab 98
Selecting the SCPI Telnet tab 100
Selecting the Help Tab 102 Capturing/Printing Displays and Windows 103 Windows Shortcuts and Miscellaneous Tasks 104
Windows shortcuts (key combinations) 104
Windows taskbar auto-hide feature 105
Windows startup folder 107
5 Troubleshooting
Check the Basics 110 Problems with Microsoft Windows Operating System 112 Returning a test set for Service 113
Calling Agilent Technologies 113
Locations for Agilent Technologies 113
Read the Warranty 114
Service Options 114
Service Tag 114
Packaging the T e st Set 116
Index
Getting Started Guide 9
10 Getting Started Guide
Agilent X-Series Wireless Communications Test Set Getting Started Guide
1 Quick-Start
This section explains how to initialize the test set and generate and view a signal. The following topics can be found in this section:
“Initial Inspection" on page 12
“Test Set Location and Rack Mounting Requirements" on page 13
“Turning on the test set the first time" on page 14
“Anti-V irus Software and Firewalls" on page 23
“Shortening the Test Set Power-On Time" on page 23
“Instrument Information" on page 24
Agilent Technologies
11
1 Quick-Start

Initial Inspection

Inspect the shipping container and the cushioning material for signs of stress. Retain the shipping materials for future use, as you may wish to ship the test set to another location or to Agilent Technologies for service.
Verifying the contents
Item Deliverable Description
Getting Started Guide Provides first-time power on instructions,
Agilent IO Libraries disk Software for establishing and configuring
licensing information, operating system information, and general hardware information.
PC to test set interfaces.
Optical mouse An optical USB mouse to simplify screen
navigation.
E6607B Safety and Regulations Guide
Power cable Connection for test set power.
For use with E6607B Wireless Communications Test Set.
12 Getting Started Guide

Shipping problems?

If the shipping materials are damaged or the contents of the container are incomplete:
Contact the nearest Agilent Technologies office.
Keep the shipping materials for the carrier’s inspection.
If you must return an test set to Agilent Technologies, use the original (or
comparable) shipping materials. See “Returning a test set for Service" on page 113.

Test Set Location and Rack Mounting Requirements

Locating the test set
Make sure that the fan inlet and exhaust vent areas on the sides of the test set are not obstructed. The minimal required clearance is 2 inches. Airflow restrictions cause additional airflow noise and cause the fans to speed up so they can draw in enough air for the required cooling resulting in excessive audible noise.
Cooling and rack mounting
Quick-Start 1
Do not rack mount the test set side-by-side with any other instrument with side-by-side ventilation. Make sure the exhaust air from the first instrument is directed away from the inlet of the second unit. If the pre-heated air from the first instrument is directed into the second instrument, it can cause excessive operating temperatures in the second unit and can cause instrument failures. The test set draws air in from the left side and exhausts air from the right side.
Getting Started Guide 13
1 Quick-Start

Turning on the test set the first time

Initial power-on of the test set can be accomplished using either of the following methods:
“Initializing the test set using a USB mouse" on page 14
“Initializing the test set using front-panel navigation" on page 19
During the initial power-on process, the test set turns off and restarts several times. This only happens during the initial installation process.

Initializing the test set using a USB mouse

Steps Actions Notes 1 Connect the mouse
2 Power on the test set a Position the test set so you
Connect the mouse to one
of the USB ports on the test set.
have easy access to the power cord and plug it in.
b Press the power switch
(located in the lower left corner of the test set front panel) to turn the test set on.
See “Test Set Location and Rack Mounting
Requirements and “Power requirements" on page 24
for more details.
The test set can require more than 5 minutes to
power-on. This is affected by Windows start-up requirements.
The Agilent Technologies screen appears followed
by a screen that allows you to select the operating system.
At this time, it is safe to turn off the test set before
starting the Setup Wizard.
14 Getting Started Guide
Steps Actions Notes
Quick-Start 1
3 Launch Microsoft Setup
Wizard
Highlight Microsoft
Windows and press to start the pre-installed operating system.
a The analyzer performs the following steps:
Enter
Windows S tartup window
Black screen
Windows blue please wait message window, “Please wait wh ile windows prepares
to setup”
Blue Agilent window
The following Welcome window appears giving you info rmation about the End-
User License Agreement.
At this time, it is safe to turn off the instrument before initializing the software.
After launching the setup, do not turn off the instrument or remove power before the Setup Wizard comple tes and the system restarts. Turning off the instrument may corrupt the system and the application may not operate.
4 Reboot and logon a Click
Getting Started Guide 15
OK When the instrument restarts, the following message
window appears:
1 Quick-Start
Steps Actions Notes
This window appears and covers the LaunchXSA window.
If you do not check the “Do not show this message again” check box, this message will be displayed each time the analyzer is turned on. No application will start while this message is displayed. Before continuing, make sure that you carefully read the Anti­Virus message and determine what action is appropriate.
5 Disable the Anti-Virus
message
6 Verify the installation a On the test set, press
7 View a signal a Press
a Select the check box and
click
Continue.
System, Show, System.
b Verify that the purchased
application(s) appear in the list.
Input/Output,
RF Calibrator, 50 MHz.
b Press
c Press
d Press
AMPTD Y Scale, Ref
Value
, 10, dBm.
FREQ, Center Freq,
50, MHz.
SPAN X Scale,
Span, 8, MHz.
Messages similar to the following continue to
appears:
The system reboots.
Several required processes continue.
The system reboots.
The XSA application initializes
If you require further assistance, contact the Agilent
support team. Online assistance: http://www.agilent.com/find/assist
To view a signal without attaching any RF cables to
the test set, route the internal 50 MHz calibration signal to the test set input.
Set the reference level to 10 dBm.
Set the analyzer center frequency to 50 MHz.
Set the analyzer span to 8 MHz.
16 Getting Started Guide
Steps Actions Notes
The 50 MHz reference signal appears on the display.
Quick-Start 1
8 Generate and view signal a Connect the ends of a short
RF cable to the RF Input and RF Output connectors on the front panel.
b Press
c Press
d Press
e Press
f Press
g Press
h Press
i Press
Input/Output, RF Input, RF Input Port, RF In.
Input/Output, More, RF Output Port, RF Output.
Source, RF Output, On.
Source, Amplitude, RF Power, 0, dBm.
Freq, Center Freq, 1,
GHz.
AMPTD Y Scale,
Ref Value, 10, dBm.
FREQ, Center Freq,
1, GHz.
SPAN X Scale,
Span, 8, MHz.
The purpose is to use a loopback cable to verify that
the test set’s source can generate a signal and the test set’s analyzer can measure it.
Because there are four RF ports on the front panel, it
is necessary to specify which ports are used for the input and output.
You need to choose frequency and power-level
settings for the source; you also need to specify that the source output is on.
You need to specify the center frequency, span, and
reference power level for the analyzer.
Getting Started Guide 17
1 Quick-Start
Steps Actions Notes
The 1 GHz signal appears on the display.
18 Getting Started Guide

Initializing the test set using front-panel navigation

The following keys are used in the initialization process:
Quick-Start 1
Steps Actions Notes 1 Power on the test set a Position the test set so you
have easy access to the power cord and plug it in.
b Press the power switch
(located in the lower left corner of the test set front panel) to turn the test set on.
See “Test Set Location and Rack Mounting
Requirements and “Power requirements" on page 24
for more details.
The test set can require more than 5 minutes to power-
on. This is affected by Windows start-up requirements.
The Agilent T echnologies sc reen appears followed by
a screen that allows you to select the operating system.
At this time, it is safe to turn off the test set before
starting the Setup Wizard.
After launching the Setup Wizard, do not turn off the test set or remove power before the Setup Wizard completes and the system restarts. Turning off the test set may corrupt the system and the application may not operate.
Getting Started Guide 19
1 Quick-Start
Steps Actions Notes 2 Launch the Microsoft
Setup Wizard
a Highlight Microsoft
Windows and press to start the pre-installed operating system.
a The analyzer performs the following steps:
Enter
Windows S tartup window
Black screen
Windows blue please wait message window, “Please wait while windows prepares
to setup”
Blue Agilent window
The following Welcome window appears giving you info rmation about the End-
User License Agreement.
At this time, it is safe to turn off the instrument before initializing the software.
After launching the setup, do not turn off the instrument or remove power before the Setup Wizard completes and the system restarts. Turning off the instrument may corrupt the system and the application may not operate.
3 Reboot and logon a Click
20 Getting Started Guide
OK When the instrument restarts, the following message
window appears:
Steps Actions Notes
This window appears and covers the LaunchXSA window.
If you do not check the “Do not show this message again” check box, this message will be displayed each time the analyzer is turned on. No application will start while this message is displayed. Before continuing, make sure that you carefully read the Anti-Virus message and determine what action is appropriate.
Quick-Start 1
4 Disable the Anti-Virus
message
5 Verify the installation a Press
6 View a signal a Press
a Press Select to select the
check box and continue.
System.
b Verify that the purchased
application(s) appear in the list.
RF Calibrator, 50 MHz.
b Press
Value
c Press
50, MHz.
d Press
Span, 8, MHz.
Enter to
System, Show,
Input/Output,
AMPTD Y Scale, Ref
, 10, dBm.
FREQ, Center Freq,
SPAN X Scale,
Messages similar to the following continue to
appears:
The system reboots.
Several required processes continue.
The system reboots.
The XSA application initializes
If you require further assistance, contact the Agilent
support team. Online assistance: http://www.agilent.com/find/assist
To view a signal without attaching any RF cables to
the test set, route the internal 50 MHz calibration signal to the test set input.
Set the reference level to 10 dBm.
Set the analyzer center frequency to 50 MHz.
Set the analyzer span to 8 MHz.
Getting Started Guide 21
1 Quick-Start
Steps Actions Notes
The 50 MHz reference signal appears on the display.
7 Generate and view signal a Connect the ends of a short
RF cable to the RF Input and RF Output connectors on the front panel.
b Press
c Press
d Press
e Press
f Press
g Press
h Press
i Press
Input/Output, RF Input, RF Input Port, RF In.
Input/Output, More, RF Output Port, RF Output.
Source, RF Output, On.
Source, Amplitude, RF Power, 0, dBm.
Freq, Center Freq, 1,
GHz.
AMPTD Y Scale, Ref Value, 10, dBm.
FREQ, Center Freq, 1, GHz.
SPAN X Scale, Span, 8, MHz.
The purpose is to use a loopback cable to verify that
the test set’s source can generate a signal and the test set’s analyzer can measure it.
Because there are four RF ports on the front panel, it
is necessary to specify which ports are used for the input and output.
You need to choose frequency and power-level
settings for the source; you also need to specify that the source output is on.
You need to specify the center frequency, span, and
reference power level for the analyzer.
22 Getting Started Guide
Steps Actions Notes
The 1 GHz signal appears on the display.
Quick-Start 1

Anti-Virus Software and Firewalls

No anti-virus software is shipped with the test set. It is recommended that you install anti-virus software if your test set is connected to the LAN. Check with your IT department to see what they recommend.
The test set is shipped with the Windows firewall enabled. Do not modify the default network settings as this may cause problems with the operating system of the test set.

Shortening the Test Set Power-On Time

The test set desktop includes a shortcut to the ConfigureApplications.exe application that lets you control your test set power-on configuration. You can set the configuration to pre-load only the applications that you typically run. This can significantly shorten the time it takes for your test set to power-on. You can also access this functionality by pressing:
If (after power-on) you want to access an application that you did not pre-load, it will take a little longer to bring it up once you press the application softkey . This longer time­frame is only for the first access after power-on. After the initial access, mode (application) switching will operate as quickly as if the application was pre-loaded.
System, Power On, Configure Applications.
Getting Started Guide 23
1 Quick-Start

Instrument Information

Power requirements

The only physical installation of your Agilent test set is a connection to a power source. Line voltage does not need to be selected.
This test set does not contain customer serviceable fuses.
This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited. (IEC 348 clauses 17.3.3c & 17.3.4)
Failure to ground the test set properly can result in personal injury. Before turning on the test set, you must connect its protective earth terminals to the protective conductor of the main power cable. Insert the main power cable plug into a socket outlet that has a protective earth contact only . DO NOT defeat the earth-grounding pr otection by using an extension cable, power cable, or autotransformer without a protective ground conductor.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 Second Edition and IEC 664 respectively. This instrument has autoranging line voltage input. Before switching on the instrument, be sure the supply voltage is within the specified range.
The Mains wiring and connectors shall be compatible with the connector used in the premise electrical system. Failure, to ensure adequate earth grounding by not using the correct components may cause product damage, and serious injury.
24 Getting Started Guide
Quick-Start 1
AC power cord
The test set is equipped with a three-wire power cord, in accordance with international safety standards. This cable grounds the test set cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the test set. See:
http://www.agilent.com/find/powercords
Always use the three-prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord can cause product damage.
This instrument has auto-ranging line voltage input; be sure the supply voltage is within the specified range and voltage fluctuations do not to exceed 10 percent of the nominal supply voltage:
100/120 VAC 50/60 Hz, 220/240 VAC 50/60 Hz, 270 W Max.
If this product is not used as specifi ed, the pr otect ion pr ovid ed by the equipment could be impaired. This product mu st be used in a normal condition (in which all means for protection are intact) only. Install the instrument so that the detachable power cord is readily identifiable and easily reached by the operator. The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is only a standby switch and is not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.

Environmental Conditions (Operating)

This product is designed for use in the following conditions:
For indoor use only
Temperature 5°C to 50°C
Altitude up to 3048 meters
0% Relative Humidity to 90% Relative Humidity up to 40°C, and li nearly decreasing to 50% Relative Humidity up to 50°C; non-condensing..
This product is designed for use in Installation Category II and Pollution Degree 2.
Getting Started Guide 25
1 Quick-Start

Ventilation

Instrument Maintenance

VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the product by 4 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Cleaning the instrument
To prevent electrical shock, disconnect the test set from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
o
C for every 100
Cleaning the connectors
Cleaning connectors with alcohol shall only be done with the instrument power cord removed, and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumers to dissapate prior to energizing the instrument.
Battery Information
The test set uses a lithium battery located on the CPU board. This is not an operator replaceable part. See “Returning a test set for Service" on page 113. Replaceable parts must be approved or supplied by Agilent Technologies.
You can order the service documentation for the instrument through your Agilent Sales and Service office.
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to the manufacturer’s instructions.
Do not throw batteries away but collect as small chemical waste.
26 Getting Started Guide
Quick-Start 1
Getting Started Guide 27
1 Quick-Start

Protecting against electrostatic discharge

Electrostatic discharge (ESD) can damage or destroy electronic components (the possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used).
Test equipment and ESD
To help reduce ESD damage that can occur while using test equipment:
Do not use these first three techniques when working on circuitry with a voltage potential greater than 500 volts.
Before connecting any coaxial cable to a test set connector for the first time each day , momentarily short the center and outer conductors of the cable together.
Personnel should be grounded with a 1 MΩ resistor-isolated wrist-strap before touching the center pin of any connector and before removing any assembly from the test set.
Be sure that all instruments are properly earth-grounded to prevent build-up of st atic charge.
Perform work on all components or assemblies at a static-safe workstation.
Keep static-generating materials at least one meter away from all components.
Store or transport components in static-shielding containers.
Always handle printed circuit board assemblies by the edges. This reduces the possibility of ESD damage to components and prevent contamination of exposed plating.
Additional information about ESD
For more information about ESD and how to prevent ESD damage, contact the Electrostatic Discharge Association (http://www.esda.org). The ESD standards developed by this agency are sanctioned by the American National Standards Institute (ANSI).
28 Getting Started Guide
Agilent X-Series Wireless Communications Test Set Getting Started Guide
2 Front and Rear Panel Features
This section describes the following features:
“Front-Panel Features (E6607A/B)" on page 30 “Display Annotations" on page 35 “Rear-Panel Features" on page 37 “Front and Rear Panel Symbols" on page 39
Agilent Technologies
29
2 Front and Rear Panel Features

Front-Panel Features (E6607A/B)

Item
Description
#Name
1 Menu Keys Key labels appear to the left of the menu keys to identify the current function
of each key. The displayed functions are dependent on the currently selected Mode and Measurement, and are directly related to the most recent key press.
2 Measurement Keys These keys select the Mode, and the Measurement within the mode. They also
control the initiation and rate of recurrence of measurements.
3 Setup Keys These keys set the parameters used for setting up the test set in the current
Mode.
4 Marker Keys Markers are often available for a measurement, to measure a very specific
point/segment of data within the range of the current measurement data.
5 Utility Keys These keys control system-wide functionality such as
test set configuration information and I/O setup,
printer setup and printing,
file management, save and recall,
test set presets. 6 Probe Power Supplies power for external high frequency probes and accessories. 7 Headphones Output Headphones can be used to hear any available audio output. 8 Back Space Key Press this key to delete the previous character when entering alphanumeric
information. It also works as the Back key in Help and Explorer windows.
9 USB Connectors Standard USB 2.0 ports, Type A. Connect to external peripherals such as a
mouse, keyboard, DVD drive or hard drive.
10 Delete Key Press this key to delete files, or to perform other deletion tasks.
30 Getting Started Guide
Front and Rear Panel Features 2
Item
Description
#Name
11 Local/Cancel/(Esc) Key If you are in remote operation the Local key
returns test set control from remote back to local (the front panel).
turns the display on (if it was turned off for remote operation).
can be used to clear errors. (Press the key once to return to local control,
and a second time to clear error message line.)
If you have not already pressed the units or Enter key, Cancel exits the currently selected function without changing its value.
Esc works the same as it does on a pc keyboard. It
exits Windows dialogs
clears errors
aborts printing
cancels operations.
This key also exits the help system if it has been accessed.
12 Numeric Keypad Enters a specific numeric value for the current function. Entries appear on the
upper left of the display, in the measurement information area.
13 Enter and Arrow Keys The Enter key terminates data entry when either no unit of measure is needed,
or you want to use the default unit. The arrow keys
Increment and decrement the value of the current measurement selection.
Navigate help topics.
Navigate, or make selections, within Windows dialogs.
Navigate within forms used for setting up measurements.
Navigate within tables.
The arrow keys cannot be used to move a mouse
pointer around on the display.
14 Menu/ (Alt) Key Alt works the same as a pc keyboard. Use it to change control focus in
Windows pull-down menus.
15 Ctrl Key Ctrl works the same as a pc keyboard. Use it to navigate in W i ndows
applications, or to select multiple items in lists.
16 Select / Space Key Select is also the Space key and it has typical pc functionality. For example, in
Windows dialogs, it selects files, checks and unchecks check boxes, and picks
radio button choices. It opens a highlighted Help topic. 17 Tab Keys Use these keys to move between fields in Windows dialogs. 18 Knob Increments and decrements the value of the current active function. 19 Return Key Exits the current menu and returns to the previous menu. Has typical pc
functionality. 20 Full Screen Key Pressing this key turns off the softkeys to maximize the graticule display area.
Getting Started Guide 31
Loading...
+ 89 hidden pages