Agilent X-Series
E6607B Wireless
Communications
Test Set
Notice: This document contains references to
Agilent. Please note that Agilent’s Test and
Measurement business has become Keysight
Technologies. For more information, go to
www.keysight.com.
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Trademark
Acknowledgements
Microsoft “ is a U.S. registered trademark
of Microsoft Corporation.
Windows “ and MS Windows “ are U.S.
registered trademarks of Microsoft
Corporation.
Adobe Reader “ is a U.S. registered
trademark of Adobe System Incorporated.
Java™ is a U.S. trademark of Sun
Microsystems, Inc.
MA TLAB “ is a U.S. registered trademark
of Math Works, Inc.
Norton Ghost™ is a U.S. trademark of
Symantec Corporation.
Copyright 2008-9 Agilent Technologies
Inc.
Licensed under the Apache License, Version 2.0 (the "License");
you may not use this file except in compliance with the License.
You may obtain a copy of the License at:
http://www.apache.org/licenses/
LICENSE-2.0
Unless required by applicable law or
agreed to in writing, software
distributed under the License is
distributed on an "AS IS" BASIS,
WITHOUT WARRANTIES OR CONDITIONS OF ANY KIND, either express or
implied.
See the License for the specific
language governing permissions and limitations under the License.
Manual Part Number
E6607-90033
Print Date
December 2014
Printed in USA
Agilent Technologies, Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
The material contained in this document is provided “as is,” and is
subject to being changed, without
notice, in future editions. Further,
to the maximum extent permitted
by applicable law, Agilent disclaims
all warranties, either express or
implied, with regard to this manual
and any information contained
herein, including but not limited to
the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not
be liable for errors or for incidental
or consequential damages in connection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Agilent and the user have a
separate written agreement with
warranty terms covering the material in this document that conflict
with these terms, the warranty
terms in the separate agreement
shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance
of a U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DF AR 252.227-7014
(June 1995), or as a “commercial item” as
defined in F AR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S.
Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.22714 (June 1987) or DFAR 252.227-7015
(b)(2) (November 1995), as applicable in
any technical data.
Safety Notices
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in damage to the product or loss of
important data. Do not proceed
beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like that, if not correctly
performed or adhered to, could
result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated conditions are fully
understood and met.
Warranty
This Agilent Technologies product is warranted against defects in material and
workmanship for a period of one year from the date of shipment. during the warranty
period, Agilent Technologies will, at its option, either repair or replace products that
prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent
Technologies shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent
Technologies from another country.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about this test set,
including firmware upgrades, application information, and product information, see
the following URLs:
http://www.agilent.com/find/ext
To receive the latest updates by email, subscribe to Agilent Email Updates:
http://www.agilent.com/find/emailupdates
Information on preventing test set damage can be found at:
http://www.agilent.com/find/tips
Compliance
This product has been designed and tested in accordance with accepted industry standards, and has been
supplied in a safe condition. The documentation contains information and warnings that must be followed
by the user to ensure safe operation and to maintain the product in a safe condition.
Canadian EMC Requirements
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
Acoustic Noise Requirements
Acoustic Noise EmissionGeraeuschemission
LpA <70 dB
Operator Position
Normal Position
Per ISO 7779
Acoustic Noise -- Further
Information
Ambient Temperature <40
LpA < 70 dB
Am Arbeitsplatz
Normaler Betrieb
Nach DIN 45635 t.19
Supplemental Information
(Values given are per ISO 7779 standard in the “Operator
Sitting” position)
° CNominally under 55 dBA Sound Pressure. 55 dBA is gen-
erally considered suitable for use in quiet office environments.
Ambient Temperature >40° CNominally under 65 dBA Sound Pressure. 65 dBA is gen-
erally considered suitable for noisy office environments.
(The fan speed, and thus the noise level, increases with
increasing ambient temperature.)
In This Guide…
This guide contains the following information:
1 Quick-Start
This chapter explains how to initial ize the test se t and generate and view a
signal.
2 Front and Rear Panel Features
Refer to this chapter for information on front- and rear-panel key
functionality, and display annotations.
3 Test Set Operating System
This chapter describes the Microsoft Windows OS configuration and the
settings used with the Agilent test set software.
4 Using Microsoft Windows
The information in this chapter provides some guidelines for using the
Microsoft Windows feature capabilities with the test set.
5 Troubleshooting
This chapter details some basic steps that may solve any problems you are
experiencing with either the test set or Microsoft Windows OS.
Contents
1 Quick-Start
Initial Inspection12
Table. Verifying the contents12
Shipping problems?13
Test Set Location and Rack Mounting Requirements13
Turning on the test set the first time14
Initializing the test set using a USB mouse14
Initializing the test set using front-panel navigation 19
Anti-Virus Software and Firewalls23
Shortening the Test Set Power-On Time23
Instrument Information24
Power requirements24
Environmental Conditions (Operating)25
Ventilation 26
Instrument Maintenance26
Protecting against electrostatic discharge28
2 Front and Rear Panel Features
Front-Panel Features (E6607A/B)30
V irtual Front Panel33
When the Test Set is Controlled Remotely34
Display Annotations35
Rear-Panel Features37
Front and Rear Panel Symbols39
3 Test Set Operating System
Agilent Software Installed42
Agilent U9065A & U9060A software42
Customer Installation of Software43
3rd party software verified by Agilent43
Getting Started Guide7
Installation of other 3rd party software43
User Accounts44
Administrator login44
User login44
AgilentOnly user account44
Agilent service user accounts44
Customer creation of accounts45
Agilent X-Series Licensing Options46
Fixed Perpetual46
Transportable Perpetual 46
Time-Based Licenses47
Licensing New Measurement Application Software - After Initial Purchase48
Transporting a License Between X-Series Test Sets50
Windows Configuration53
Settings that can be changed53
Settings that must not be changed54
Configuring Printers56
Configuring LAN57
Hostname57
IP Address & Gateway57
Windows Security58
Windows Firewall59
Automatic updates59
Virus protection60
Spyware protection61
System Maintenance62
Backup62
System restore62
Disk defragmenting62
USB Connections63
Hard Drive Partitioning and Use64
Backing-up the factory calibration data65
Hard Drive Recovery Process67
Updating the software69
Configuring recovery prompt timing71
8Getting Started Guide
4 Using Microsoft Windows Operating System
Navigating Windows Without a Mouse74
Remote Desktop: Using the X-Series test set remotely78
Overview of Remote Desktop operation78
Setting up Remote Desktop operation78
How to locate the computer name of the test set80
Running a Remote Desktop session82
The Virtual Front Panel91
Embedded Web Server: Using the X-Series Test Set Remotely92
Accessing the test set through the Internet92
Accessing the Web Control tab96
Selecting the Get Data tab97
Selecting the Get Image tab98
Selecting the SCPI Telnet tab100
Selecting the Help Tab102
Capturing/Printing Displays and Windows103
Windows Shortcuts and Miscellaneous Tasks104
Windows shortcuts (key combinations)104
Windows taskbar auto-hide feature105
Windows startup folder107
5 Troubleshooting
Check the Basics110
Problems with Microsoft Windows Operating System112
Returning a test set for Service113
Calling Agilent Technologies 113
Locations for Agilent Technologies113
Read the Warranty 114
Service Options 114
Service Tag114
Packaging the T e st Set116
Index
Getting Started Guide9
10Getting Started Guide
Agilent X-Series Wireless Communications Test Set
Getting Started Guide
1
Quick-Start
This section explains how to initialize the test set and generate and view a signal.
The following topics can be found in this section:
“Initial Inspection" on page 12
“Test Set Location and Rack Mounting Requirements" on page 13
“Turning on the test set the first time" on page 14
“Anti-V irus Software and Firewalls" on page 23
“Shortening the Test Set Power-On Time" on page 23
“Instrument Information" on page 24
Agilent Technologies
11
1Quick-Start
Initial Inspection
Inspect the shipping container and the cushioning material for signs of stress. Retain the
shipping materials for future use, as you may wish to ship the test set to another location
or to Agilent Technologies for service.
Verifying the contents
ItemDeliverableDescription
Getting Started GuideProvides first-time power on instructions,
Agilent IO Libraries diskSoftware for establishing and configuring
licensing information, operating system
information, and general hardware
information.
PC to test set interfaces.
Optical mouseAn optical USB mouse to simplify screen
navigation.
E6607B Safety and
Regulations Guide
Power cableConnection for test set power.
For use with E6607B Wireless
Communications Test Set.
12Getting Started Guide
Shipping problems?
If the shipping materials are damaged or the contents of the container are incomplete:
•Contact the nearest Agilent Technologies office.
•Keep the shipping materials for the carrier’s inspection.
•If you must return an test set to Agilent Technologies, use the original (or
comparable) shipping materials. See “Returning a test set for Service" on page 113.
Test Set Location and Rack Mounting Requirements
Locating the test set
Make sure that the fan inlet and exhaust vent areas on the sides of the test set are not
obstructed. The minimal required clearance is 2 inches. Airflow restrictions cause
additional airflow noise and cause the fans to speed up so they can draw in enough air for
the required cooling resulting in excessive audible noise.
Cooling and rack mounting
Quick-Start1
Do not rack mount the test set side-by-side with any other instrument with side-by-side
ventilation. Make sure the exhaust air from the first instrument is directed away from the
inlet of the second unit. If the pre-heated air from the first instrument is directed into the
second instrument, it can cause excessive operating temperatures in the second unit and
can cause instrument failures. The test set draws air in from the left side and exhausts air
from the right side.
Getting Started Guide13
1Quick-Start
Turning on the test set the first time
Initial power-on of the test set can be accomplished using either of the following
methods:
•“Initializing the test set using a USB mouse" on page 14
•“Initializing the test set using front-panel navigation" on page 19
During the initial power-on process, the test set turns off and restarts
several times. This only happens during the initial installation process.
Initializing the test set using a USB mouse
StepsActionsNotes
1 Connect the mouse
2 Power on the test seta Position the test set so you
• Connect the mouse to one
of the USB ports on the test
set.
have easy access to the
power cord and plug it in.
b Press the power switch
(located in the lower left
corner of the test set front
panel) to turn the test set
on.
• See “Test Set Location and Rack Mounting
Requirements and “Power requirements" on page 24
for more details.
• The test set can require more than 5 minutes to
power-on. This is affected by Windows start-up
requirements.
• The Agilent Technologies screen appears followed
by a screen that allows you to select the operating
system.
• At this time, it is safe to turn off the test set before
starting the Setup Wizard.
14Getting Started Guide
StepsActionsNotes
Quick-Start1
3 Launch Microsoft Setup
Wizard
• Highlight Microsoft
Windows and press
to start the pre-installed
operating system.
a The analyzer performs the following steps:
Enter
• Windows S tartup window
• Black screen
• Windows blue please wait message window, “Please wait wh ile windows prepares
to setup”
• Blue Agilent window
• The following Welcome window appears giving you info rmation about the End-
User License Agreement.
• At this time, it is safe to turn off the instrument before initializing the software.
After launching the setup, do not turn off the instrument or remove power before the
Setup Wizard comple tes and the system restarts. Turning off the instrument may corrupt
the system and the application may not operate.
4 Reboot and logona Click
Getting Started Guide15
OK• When the instrument restarts, the following message
window appears:
1Quick-Start
StepsActionsNotes
• This window appears and covers the LaunchXSA window.
If you do not check the “Do not show this message again” check box, this message will
be displayed each time the analyzer is turned on. No application will start while this
message is displayed. Before continuing, make sure that you carefully read the AntiVirus message and determine what action is appropriate.
5 Disable the Anti-Virus
message
6 Verify the installationa On the test set, press
7 View a signala Press
a Select the check box and
click
Continue.
System, Show, System.
b Verify that the purchased
application(s) appear in the
list.
Input/Output,
RF Calibrator, 50 MHz.
b Press
c Press
d Press
AMPTD Y Scale, Ref
Value
, 10, dBm.
FREQ, Center Freq,
50, MHz.
SPAN X Scale,
Span, 8, MHz.
• Messages similar to the following continue to
appears:
• The system reboots.
• Several required processes continue.
• The system reboots.
• The XSA application initializes
• If you require further assistance, contact the Agilent
support team.
Online assistance: http://www.agilent.com/find/assist
• To view a signal without attaching any RF cables to
the test set, route the internal 50 MHz calibration
signal to the test set input.
• Set the reference level to 10 dBm.
• Set the analyzer center frequency to 50 MHz.
• Set the analyzer span to 8 MHz.
16Getting Started Guide
StepsActionsNotes
The 50 MHz reference signal appears on the display.
Quick-Start1
8 Generate and view signal a Connect the ends of a short
RF cable to the RF Input
and RF Output connectors
on the front panel.
b Press
c Press
d Press
e Press
f Press
g Press
h Press
i Press
Input/Output,
RF Input, RF Input Port,
RF In.
Input/Output, More,
RF Output Port,
RF Output.
Source, RF Output,
On.
Source, Amplitude,
RF Power, 0, dBm.
Freq, Center Freq, 1,
GHz.
AMPTD Y Scale,
Ref Value, 10, dBm.
FREQ, Center Freq,
1, GHz.
SPAN X Scale,
Span, 8, MHz.
• The purpose is to use a loopback cable to verify that
the test set’s source can generate a signal and the test
set’s analyzer can measure it.
• Because there are four RF ports on the front panel, it
is necessary to specify which ports are used for the
input and output.
• You need to choose frequency and power-level
settings for the source; you also need to specify that
the source output is on.
• You need to specify the center frequency, span, and
reference power level for the analyzer.
Getting Started Guide17
1Quick-Start
StepsActionsNotes
The 1 GHz signal appears on the display.
18Getting Started Guide
Initializing the test set using front-panel navigation
The following keys are used in the initialization process:
Quick-Start1
StepsActionsNotes
1 Power on the test seta Position the test set so you
have easy access to the
power cord and plug it in.
b Press the power switch
(located in the lower left
corner of the test set front
panel) to turn the test set
on.
• See “Test Set Location and Rack Mounting
Requirements and “Power requirements" on page 24
for more details.
• The test set can require more than 5 minutes to power-
on. This is affected by Windows start-up
requirements.
• The Agilent T echnologies sc reen appears followed by
a screen that allows you to select the operating
system.
• At this time, it is safe to turn off the test set before
starting the Setup Wizard.
After launching the Setup Wizard, do not turn off the test set or remove power before the
Setup Wizard completes and the system restarts. Turning off the test set may corrupt the
system and the application may not operate.
Getting Started Guide19
1Quick-Start
StepsActionsNotes
2 Launch the Microsoft
Setup Wizard
a Highlight Microsoft
Windows and press
to start the pre-installed
operating system.
a The analyzer performs the following steps:
Enter
• Windows S tartup window
• Black screen
• Windows blue please wait message window, “Please wait while windows prepares
to setup”
• Blue Agilent window
• The following Welcome window appears giving you info rmation about the End-
User License Agreement.
• At this time, it is safe to turn off the instrument before initializing the software.
After launching the setup, do not turn off the instrument or remove power before the
Setup Wizard completes and the system restarts. Turning off the instrument may corrupt
the system and the application may not operate.
3 Reboot and logona Click
20Getting Started Guide
OK• When the instrument restarts, the following message
window appears:
StepsActionsNotes
• This window appears and covers the LaunchXSA window.
If you do not check the “Do not show this message again” check box, this message will
be displayed each time the analyzer is turned on. No application will start while this
message is displayed. Before continuing, make sure that you carefully read the Anti-Virus
message and determine what action is appropriate.
Quick-Start1
4 Disable the Anti-Virus
message
5 Verify the installationa Press
6 View a signala Press
a Press Select to select the
check box and
continue.
System.
b Verify that the purchased
application(s) appear in the
list.
RF Calibrator, 50 MHz.
b Press
Value
c Press
50, MHz.
d Press
Span, 8, MHz.
Enter to
System, Show,
Input/Output,
AMPTD Y Scale, Ref
, 10, dBm.
FREQ, Center Freq,
SPAN X Scale,
• Messages similar to the following continue to
appears:
• The system reboots.
• Several required processes continue.
• The system reboots.
• The XSA application initializes
• If you require further assistance, contact the Agilent
support team.
Online assistance: http://www.agilent.com/find/assist
• To view a signal without attaching any RF cables to
the test set, route the internal 50 MHz calibration
signal to the test set input.
• Set the reference level to 10 dBm.
• Set the analyzer center frequency to 50 MHz.
• Set the analyzer span to 8 MHz.
Getting Started Guide21
1Quick-Start
StepsActionsNotes
The 50 MHz reference signal appears on the display.
7 Generate and view signal a Connect the ends of a short
RF cable to the RF Input
and RF Output connectors
on the front panel.
b Press
c Press
d Press
e Press
f Press
g Press
h Press
i Press
Input/Output,
RF Input, RF Input Port,
RF In.
Input/Output, More,
RF Output Port,
RF Output.
Source, RF Output,
On.
Source, Amplitude,
RF Power, 0, dBm.
Freq, Center Freq, 1,
GHz.
AMPTD Y Scale,
Ref Value, 10, dBm.
FREQ, Center Freq,
1, GHz.
SPAN X Scale,
Span, 8, MHz.
• The purpose is to use a loopback cable to verify that
the test set’s source can generate a signal and the test
set’s analyzer can measure it.
• Because there are four RF ports on the front panel, it
is necessary to specify which ports are used for the
input and output.
• You need to choose frequency and power-level
settings for the source; you also need to specify that
the source output is on.
• You need to specify the center frequency, span, and
reference power level for the analyzer.
22Getting Started Guide
StepsActionsNotes
The 1 GHz signal appears on the display.
Quick-Start1
Anti-Virus Software and Firewalls
No anti-virus software is shipped with the test set. It is recommended that you install
anti-virus software if your test set is connected to the LAN. Check with your IT
department to see what they recommend.
The test set is shipped with the Windows firewall enabled. Do not modify the default
network settings as this may cause problems with the operating system of the test set.
Shortening the Test Set Power-On Time
The test set desktop includes a shortcut to the ConfigureApplications.exe application that
lets you control your test set power-on configuration. You can set the configuration to
pre-load only the applications that you typically run. This can significantly shorten the
time it takes for your test set to power-on. You can also access this functionality by
pressing:
If (after power-on) you want to access an application that you did not pre-load, it will
take a little longer to bring it up once you press the application softkey . This longer timeframe is only for the first access after power-on. After the initial access, mode
(application) switching will operate as quickly as if the application was pre-loaded.
System, Power On, Configure Applications.
Getting Started Guide23
1Quick-Start
Instrument Information
Power requirements
The only physical installation of your Agilent test set is a connection to a power source.
Line voltage does not need to be selected.
This test set does not contain customer serviceable fuses.
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact.
Any interruption of the protective conductor inside or outside of the
product is likely to make the product dangerous. Intentional
interruption is prohibited. (IEC 348 clauses 17.3.3c & 17.3.4)
Failure to ground the test set properly can result in personal injury.
Before turning on the test set, you must connect its protective earth
terminals to the protective conductor of the main power cable. Insert
the main power cable plug into a socket outlet that has a protective
earth contact only . DO NOT defeat the earth-grounding pr otection by
using an extension cable, power cable, or autotransformer without a
protective ground conductor.
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010 Second Edition and IEC 664 respectively.
This instrument has autoranging line voltage input. Before switching on
the instrument, be sure the supply voltage is within the specified range.
The Mains wiring and connectors shall be compatible with the connector
used in the premise electrical system. Failure, to ensure adequate earth
grounding by not using the correct components may cause product
damage, and serious injury.
24Getting Started Guide
Quick-Start1
AC power cord
The test set is equipped with a three-wire power cord, in accordance with international
safety standards. This cable grounds the test set cabinet when connected to an
appropriate power line outlet. The cable appropriate to the original shipping location is
included with the test set. See:
http://www.agilent.com/find/powercords
Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can
cause product damage.
This instrument has auto-ranging line voltage input; be sure the supply
voltage is within the specified range and voltage fluctuations do not to
exceed 10 percent of the nominal supply voltage:
If this product is not used as specifi ed, the pr otect ion pr ovid ed by the
equipment could be impaired. This product mu st be used in a normal
condition (in which all means for protection are intact) only.
Install the instrument so that the detachable power cord is readily
identifiable and easily reached by the operator. The detachable power
cord is the instrument disconnecting device. It disconnects the mains
circuits from the mains supply before other parts of the instrument.
The front panel switch is only a standby switch and is not a LINE
switch. Alternatively, an externally installed switch or circuit breaker
(which is readily identifiable and is easily reached by the operator)
may be used as a disconnecting device.
Environmental Conditions (Operating)
This product is designed for use in the following conditions:
•For indoor use only
•Temperature 5°C to 50°C
•Altitude up to 3048 meters
•0% Relative Humidity to 90% Relative Humidity up to 40°C, and li nearly decreasing
to 50% Relative Humidity up to 50°C; non-condensing..
This product is designed for use in Installation Category II and
Pollution Degree 2.
Getting Started Guide25
1Quick-Start
Ventilation
Instrument Maintenance
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be restricted.
The ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the product by 4
watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
Cleaning the instrument
To prevent electrical shock, disconnect the test set from mains before
cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
o
C for every 100
Cleaning the connectors
Cleaning connectors with alcohol shall only be done with the
instrument power cord removed, and in a well-ventilated area. Allow
all residual alcohol moisture to evaporate, and the fumers to
dissapate prior to energizing the instrument.
Battery Information
The test set uses a lithium battery located on the CPU board. This is not an operator
replaceable part. See “Returning a test set for Service" on page 113. Replaceable parts
must be approved or supplied by Agilent Technologies.
You can order the service documentation for the instrument through your Agilent Sales
and Service office.
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommended. Discard used
batteries according to the manufacturer’s instructions.
Do not throw batteries away but collect as small chemical waste.
26Getting Started Guide
Quick-Start1
Getting Started Guide27
1Quick-Start
Protecting against electrostatic discharge
Electrostatic discharge (ESD) can damage or destroy electronic components (the
possibility of unseen damage caused by ESD is present whenever components are
transported, stored, or used).
Test equipment and ESD
To help reduce ESD damage that can occur while using test equipment:
Do not use these first three techniques when working on circuitry with
a voltage potential greater than 500 volts.
•Before connecting any coaxial cable to a test set connector for the first time each day ,
momentarily short the center and outer conductors of the cable together.
•Personnel should be grounded with a 1 MΩ resistor-isolated wrist-strap before
touching the center pin of any connector and before removing any assembly from the
test set.
•Be sure that all instruments are properly earth-grounded to prevent build-up of st atic
charge.
•Perform work on all components or assemblies at a static-safe workstation.
•Keep static-generating materials at least one meter away from all components.
•Store or transport components in static-shielding containers.
•Always handle printed circuit board assemblies by the edges. This reduces the
possibility of ESD damage to components and prevent contamination of exposed
plating.
Additional information about ESD
•For more information about ESD and how to prevent ESD damage, contact the
Electrostatic Discharge Association (http://www.esda.org). The ESD standards
developed by this agency are sanctioned by the American National Standards
Institute (ANSI).
28Getting Started Guide
Agilent X-Series Wireless Communications Test Set
Getting Started Guide
2
Front and Rear Panel Features
This section describes the following features:
“Front-Panel Features (E6607A/B)" on page 30
“Display Annotations" on page 35
“Rear-Panel Features" on page 37
“Front and Rear Panel Symbols" on page 39
Agilent Technologies
29
2Front and Rear Panel Features
Front-Panel Features (E6607A/B)
Item
Description
#Name
1Menu KeysKey labels appear to the left of the menu keys to identify the current function
of each key. The displayed functions are dependent on the currently selected
Mode and Measurement, and are directly related to the most recent key press.
2Measurement KeysThese keys select the Mode, and the Measurement within the mode. They also
control the initiation and rate of recurrence of measurements.
3Setup KeysThese keys set the parameters used for setting up the test set in the current
Mode.
4Marker KeysMarkers are often available for a measurement, to measure a very specific
point/segment of data within the range of the current measurement data.
5Utility KeysThese keys control system-wide functionality such as
• test set configuration information and I/O setup,
• printer setup and printing,
• file management, save and recall,
• test set presets.
6Probe PowerSupplies power for external high frequency probes and accessories.
7Headphones OutputHeadphones can be used to hear any available audio output.
8Back Space KeyPress this key to delete the previous character when entering alphanumeric
information. It also works as the Back key in Help and Explorer windows.
9USB ConnectorsStandard USB 2.0 ports, Type A. Connect to external peripherals such as a
mouse, keyboard, DVD drive or hard drive.
10Delete KeyPress this key to delete files, or to perform other deletion tasks.
30Getting Started Guide
Front and Rear Panel Features2
Item
Description
#Name
11 Local/Cancel/(Esc) KeyIf you are in remote operation the Local key
•returns test set control from remote back to local (the front panel).
•turns the display on (if it was turned off for remote operation).
•can be used to clear errors. (Press the key once to return to local control,
and a second time to clear error message line.)
If you have not already pressed the units or Enter key, Cancel exits the
currently selected function without changing its value.
Esc works the same as it does on a pc keyboard. It
•exits Windows dialogs
•clears errors
•aborts printing
•cancels operations.
This key also exits the help system if it has been accessed.
12Numeric KeypadEnters a specific numeric value for the current function. Entries appear on the
upper left of the display, in the measurement information area.
13Enter and Arrow KeysThe Enter key terminates data entry when either no unit of measure is needed,
or you want to use the default unit.
The arrow keys
•Increment and decrement the value of the current measurement selection.
•Navigate help topics.
•Navigate, or make selections, within Windows dialogs.
•Navigate within forms used for setting up measurements.
•Navigate within tables.
The arrow keys cannot be used to move a mouse
pointer around on the display.
14Menu/ (Alt) KeyAlt works the same as a pc keyboard. Use it to change control focus in
Windows pull-down menus.
15Ctrl KeyCtrl works the same as a pc keyboard. Use it to navigate in W i ndows
applications, or to select multiple items in lists.
16Select / Space KeySelect is also the Space key and it has typical pc functionality. For example, in
Windows dialogs, it selects files, checks and unchecks check boxes, and picks
radio button choices. It opens a highlighted Help topic.
17Tab KeysUse these keys to move between fields in Windows dialogs.
18KnobIncrements and decrements the value of the current active function.
19Return KeyExits the current menu and returns to the previous menu. Has typical pc
functionality.
20Full Screen KeyPressing this key turns off the softkeys to maximize the graticule display area.
Getting Started Guide31
Loading...
+ 89 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.