Keysight 85052B Service Guide

Page 1
Keysight Technologies 85052B
3.5 mm Calibration Kit
User’s and Service Guide
Page 2
Page 3
85052B
Page 4
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Keysight web site:
•Go to www.keysight.com.
Enter the product model number in the Search function and click the Search icon (magnifying glass).
•Click on the Manuals hyperlink.
Open the PDF of your choice and print the document.
85052B
Page 5
Contents
1 General Information
Calibration Kit Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Broadband Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Offset Opens and Shorts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Sliding Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Installation of the Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Recording the Device Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Precision Slotless Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Clarifying the Terminology of a Connector Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
When to Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2 Specifications
Environmental Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Temperature—What to Watch Out For . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Pin Depth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Residual Errors after Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
3 Use, Maintenance, and Care of the Devices
Electrostatic Discharge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Look for Obvious Defects and Damage First . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
What Causes Connector Wear? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect the Mating Plane Surfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect Female Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Gaging Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Connector Gage Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
When to Gage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Gaging 3.5 mm Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Gaging the 3.5 mm Sliding Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Adjusting the Sliding Load Pin Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
How to Make a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Preliminary Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Final Connection Using a Torque Wrench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Connecting the Sliding Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
How to Separate a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Using the Sliding Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
85052B
Page 6
Contents
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
4 Performance Verification
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
How Keysight Verifies the Devices in Your Kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
How Often to Recertify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Where to Send a Kit for Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
5 Troubleshooting
Troubleshooting Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Where to Look for More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Returning a Kit or Device to Keysight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Contacting Keysight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
6 Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
A Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web. . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
85052B
Page 7
1 General Information
Page 8
General Information
Calibration Kit Overview
Calibration Kit Overview
The Keysight 85052B 3.5 mm calibration kit is used to calibrate Keysight network analyzers up to 26.5 GHz for measurements of components with 3.5-mm connectors.
Kit Contents
The 85052B calibration kit includes the following items:
calibration kit storage case
user’s and service guide
offset opens and shorts, broadband loads and sliding load terminations
3.5 mm gage sets
three 3.5 mm adapters
5/16 in, 90 N-cm (8 in-lb) torque wrench
centering bead (for gaging 3.5 mm sliding loads
open-end wrench
Refer to Table 6-1 on page 6-2 and Table 6-2 on page 6-4 for a complete list of kit contents and their associated part numbers.
For measurement convenience, the kit also contains three 3.5 mm adapters. The adapters are primarily intended for use in measuring non-insertable devices, but can also be used as a connector saver.
Broadband Loads
The broadband loads are metrology-grade terminations that have been optimized for performance up to
26.5 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current state-of-the-art in precision machining.
The offset short’s inner conductors have a one-piece construction, common with the shorting plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong, low-dielectricconstant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.
1- 2 85052B
Page 9
General Information
Calibration Kit Overview
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance and to ensure stable, repeatable connections.
The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized. The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.
Sliding Loads
The sliding loads in this kit are designed to provide excellent performance from 3 GHz to 26.5 GHz. The inner and outer conductors of the airline portion are precision machined to state-of-the-art tolerances. Although the sliding load has exceptional return loss, its superior load stability qualifies it as a high-performance device.
The sliding load was designed with the ability to extend the inner conductor for connection purposes and then pull it back to a preset pin depth. This feature is critical since it minimizes the possibility of damage during the connection, while maintaining a minimum pin depth to optimize performance.
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
resident within the analyzer
entered from the front panel
Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at
www.na.support.keysight.com/pna/caldefs/stddefs.html
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.
Equipment Required but Not Supplied
Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are not supplied with the calibration kit but are required to ensure successful operation of the kit. Refer to Table 6-2 on page
6-4 for ordering information.
85052B 1-3
Page 10
General Information
Incoming Inspection
Incoming Inspection
Refer to “Kit Contents” on page 1-2 to verify a complete shipment. Use Ta b l e 1- 1 to record the serial numbers of all serialized devices in your kit.
Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, refer to “Contacting Keysight” on page 5-5. Keysight will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. See “Returning a Kit or Device to Keysight” on page 5-4.
1- 4 85052B
Page 11
General Information
Recording the Device Serial Numbers
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in
Ta b l e 1 - 1 . Recording the serial numbers will prevent confusing the devices in this kit with similar devices
from other kits.
The adapters included in the kit are for measurement convenience only and are not serialized.
Table 1-1 Serial Number Record for the 85052B
Device Serial Number
Calibration kit
-m- broadband load
-f- broadband load
-m- open
-f- open
-m- short
-f- short
-m- sliding load
-f- sliding load
-m- connector Gage
-m- gage Master
-f- connector Gage
-f- gage Master
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
85052B 1-5
Page 12
General Information
Precision Slotless Connectors
Precision Slotless Connectors
The female 3.5 mm connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to national measurement standards through their well-defined mechanical dimensions.
Conventional female center conductors are slotted. When mated, the female center conductor is flared by the male pin. Because physical dimensions determine connector impedance, electrical characteristics of the female connector (and connection pair) are dependent upon the mechanical dimensions of the male pin. While connectors are used in pairs, their male and female halves are always specified separately as part of a standard, instrument, or device under test. Because of these facts, making precision measurements with the conventional slotted connector is very difficult, and establishing a direct traceability path to primary dimensional standards is nearly impossible.
The precision slotless connector was developed to eliminate these problems. All PSCs are female. A PSC incorporates a center conductor with a solid cylindrical shell that defines the outside diameter of the female center pin. Its outside diameter and, therefore, the impedance in its region does not change. The inner part provides an internal contact that flexes to accept the allowed range of male pin diameters.
The calibration of a network analyzer having a conventional slotted female connector on the test port remains valid only when the device under test and all calibration standards have identical male pin diameters. For this reason PSC test port adapters are supplied in most calibration kits.
Precision slotless connectors have the following characteristics:
There is no loss of traceable calibration on test ports when the male pin diameter of the connector on the device under test is different from the male pin diameter of the calibration standard.
The female PSC and its mating male connector can be measured and specified separately as part of the device either is attached to.
All female connectors can have a known, stable impedance based only on the diameters of their inner and outer conductors.
Female calibration standards can be fully specified. Their specifications and traceability are unaffected by the diameter of the male mating pin.
A fully traceable performance verification is made using a precision 50 ohm airline having a PSC.
Measurement repeatability is enhanced due to non-changing connector characteristics with various pin diameters.
With PSCs on test ports and standards, the percentage of accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
1- 6 85052B
Page 13
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the gender of cable connectors and adapters is referred to in terms of the center conductor. For example, a connector or device designated as
1.85 mm –f– has a 1.85 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in the kit include:
routine visual inspection
cleaning
proper gaging
proper connection techniques
All of these are described in page 3-1. Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-2 on page 2-4) or from bad connection techniques, can also damage these devices.
85052B 1-7
Page 14
General Information
When to Calibrate
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the systematic error are insignificant. This means that changes to the uncorrected leakages (directivity and isolation), mismatches (source match and load match), and frequency response of the system are small (<10%) relative to accuracy specifications.
Change in the environment (especially temperature) between calibration and measurement is the major cause in calibration accuracy degradation. The major effect is a change in the physical length of external and internal cables. Other important causes are dirty and damaged test port connectors and calibration standards. If the connectors become dirty or damaged, measurement repeatability and accuracy is affected. Fortunately, it is relatively easy to evaluate the general validity of the calibration. To test repeatability, remeasure one of the calibration standards. If you can not obtain repeatable measurements from your calibration standards, maintenance needs to be performed on the test port connectors, cables and calibration standards. Also, maintain at least one sample of the device under test or some known device as your reference device. A verification kit may be used for this purpose. After calibration, measure the reference device and note its responses. Periodically remeasure the device and note any changes in its corrected response which can be attributed to the test system. With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again.
1- 8 85052B
Page 15
2 Specifications
Page 16
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Te m pe r a t ur e
Operating
Storage 40 C to +75 C
Error-corrected range
Relative humidity Type tested at 95%, 40 C (non-condensing)
a. The temperature range over which the calibration standards maintain conformance to their specifications. b. The allowable network analyzer ambient temperature drift during measurement calibration and during
a
b
measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.
+20 C to +26 C
1 C of measurement calibration temperature
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in
Ta b l e 2 - 1 .
IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a
heat source.
2- 2 85052B
Page 17
Specifications
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-6 explains how to use gages to determine if the kit devices have
maintained their mechanical integrity. Refer to Table 2-2 on page 2-4 for typical and observed pin depth limits.
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. See Figure 2-1. The pin depth of a connector can be in one of two states: either protruding or recessed.
Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane. This condition will indicate a positive value on the connector gage.
Recession is the condition in which the center conductor is set back from the outer conductor mating plane. This condition will indicate a negative value on the connector gage.
Figure 2-1 Connector Pin Depth
85052B 2-3
Page 18
Specifications
Mechanical Characteristics
The pin depth value of each calibration device in the kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in the kit take into account the effect of pin depth on the device’s performance. Tab le 2 - 2 lists the typical pin depths and measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to
Figure 2-1 for a visual representation of proper pin depth (slightly recessed).
Table 2-2 Pin Depth Limits
Device
Opens 0 to 0.0127 mm
Shorts 0 to 0.0127 mm
Fixed loads 0.0025 to 0.0254 mm
Adapter 0.0025 to 0.0254 mm
Sliding loads 0 to 0.0076 mm
a. Approximately +2 sigma to 2 sigma of gage uncertainty based on studies done at the factory according to
recommended procedures.
b. Observed pin depth limits are the range of observation limits seen on the gage reading due to measurement
uncertainty. The depth could still be within specifications.
Typical Pin Depth
0 to 0.00050 in
0 to 0.00050 in
0.0001 to 0.0010 in
0.0001 to 0.0010 in
0 to 0.00030 in
Measurement Uncertainty
+0.0064 to 0.0064 mm +0.00025 to 0.00025 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in
+0.0041 to +0.00016 to 0.00016 in
0.0041 mm
a
Observed Pin Depth Limitsb
+0.0064 to 0.0191 mm +0.00025 to 0.00075 in
+0.0041 to 0.0168 mm +0.00016 to 0.00066 in
+0.0016 to 0.0295 mm +0.00006 to 0.00116 in
+0.0016 to 0.0295 mm +0.00006 to 0.00116 in
+0.0041 to 0.0117 mm +0.00016 to 0.00046 in
2- 4 85052B
Page 19
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications in Ta b l e 2 - 3 apply to the devices in your calibration kit when connected with a Keysight precision interface.
Table 2-3 Electrical Specifications for 85052B 3.5 mm Devices
Device Specification Frequency (GHz)
Broadband loads Return loss 46dB 0.00501) dc to  2
(male and female) Return loss 44 dB 0.00631) > 2 to  3
Return loss 38dB 0.01259) > 3 to 8
Return loss 36dB 0.01585) > 8 to 20
Return loss 34dB 0.01995) > 20 to 26.5
Sliding loads
Offset opens
(male and female) 1.20 deviation from nominal > 3 to  8
Offset shorts
(male and female) 1.00 deviation from nominal > 3 to  8
Adaptersc
a. Assuming proper usage, the specifications for the residual return loss after calibration for the sliding
a
b
b
load termination include:
• the quality of the airline portions within the sliding load, combined with
• the effective stability of the sliding element. Proper usage includes the following practices:
Connector mating surfaces are clean.
• The changes in slide positioning are NOT done in equal steps since this results in very poor calibration for some portions of the frequency range.
• The center conductor of test port connectors are nominally set back from the outer conductor. Sliding loads are designed to allow the center conductor to be moved. The position of the sliding load center conductor should be set by a reference block and not positioned flush against the center conductor of the test port.
Return loss 44 dB 0.00631) > 3 to 26.5
0.65 deviation from nominal
2.00 deviation from nominal > 8 to 20
2.00 deviation from nominal > 20 to 26.5
0.50 deviation from nominal
1.75 deviation from nominal > 8 to 20
1.75 deviation from nominal > 20 to 26.5
Return loss 30 dB 0.03162) dc to  8
Return loss 28 dB 0.03981) > 8 to  18
Return loss 26 dB 0.05012) > 18 to 26.5
dc to  3
dc to  3
<See the following page for more footnotes.>
85052B 2-5
Page 20
Specifications
Electrical Specifications
b. The specifications for the opens and shorts are given as allowed deviation from the nominal model
as defined in the standard definitions.
c. The offset delay value is 94.75 psec.
Residual Errors after Calibration
The 8510 “Specifications and Performance Verification” software can be used to obtain a printout of the residual errors after a calibration has been performed. Refer to the “Specifications and Performance Verification” section of the 8510C On-Site Service Manual for information on how to use the software.
Certification
Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Keysight Verifies the Devices in Your Kit” on page 4-2 for more information.
2- 6 85052B
Page 21
3 Use, Maintenance, and Care of the Devices
Page 22
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test, can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
Always wear a grounded wrist strap having a 1 M resistor in series with it when handling components and devices or when making connections to the test set.
Always use a grounded antistatic mat in front of your test equipment.
Always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
Always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
Always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to
the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to Chapter 6
for information on ordering supplies for ESD protection.
Figure 3-1 ESD Protection Setup
3- 2 85052B
Page 23
Use, Maintenance, and Care of the Devices
Visual Inspection
Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be misleading. Defects and damage that cannot be seen without magnification generally have no effect on electrical or mechanical performance. Magnification is of great use in analyzing the nature and cause of damage and in cleaning connectors, but it is not required for inspection.
Look for Obvious Defects and Damage First
Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are not connected properly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life if their use is on the order of a few times per week.
The test port connectors on the network analyzer test set may have many connections each day, and are, therefore, more subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the test set’s test port connectors. Replace devices with worn connectors.
Inspect the Mating Plane Surfaces
Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.
Inspect Female Connectors
Inspect the contact fingers in the female center conductor carefully. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will not make good electrical contact and must be replaced.
85052B 3-3
Page 24
Use, Maintenance, and Care of the Devices
Cleaning Connectors
NOTE This is particularly important when mating nonprecision to precision devices.
The female 3.5 mm connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). Precision slotless connectors are used to improve accuracy. With PSCs on test ports and standards, the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
Conventional female center conductors are slotted and, when mated, are flared by the male pin. Because physical dimensions determine connector impedance, this change in physical dimension affects electrical performance, making it very difficult to perform precision measurements with conventional slotted connectors.
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, this center conductor has an internal contact that flexes to accept the male pin.
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
1. Use Compressed Air or Nitrogen
WARNING Always use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. Clean air cannot damage a connector or leave particles or residues behind.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.
2. Clean the Connector Threads
3- 4 85052B
Page 25
Use, Maintenance, and Care of the Devices
Cleaning Connectors
WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed
container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Tab l e
6-2 on page 6-4 for a part number for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on page 2-3. When
cleaning a female connector, avoid snagging the swab on the center conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Inspect
Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on
page 3-3.
85052B 3-5
Page 26
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Connectors
The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for precise pin depth measurements. See Table 6-1 on page 6-2 for part number information.
Connector Gage Accuracy
The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the repeatability uncertainties that are associated with the measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteristics of the devices.
With proper technique, the gages are useful in detecting gross pin depth errors on device connectors. To achieve maximum accuracy, random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as 0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-2 on page 2-4 assumes new gages and gage masters. Therefore, these systematic errors were not included in the uncertainty analysis. As the gages undergo more use, the systematic errors can become more significant in the accuracy of the measurement.
The measurement uncertainties in Ta b le 2 - 2 are primarily a function of the assembly materials and design, and the unique interaction each device type has with the gage. Therefore, these uncertainties can vary among the different devices. For example, note the difference between the uncertainties of the opens and shorts.
The observed pin depth limits in Tab l e 2 - 2 add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages. See “Pin Depth” on page 2-3. Refer to
“Kit Contents” on page 1-2 for more information on the design of the calibration devices in the kit.
NOTE When measuring pin depth, the measured value (resultant average of three or more
measurements) contains measurement uncertainty and is not necessarily the true value. Always compare the measured value with the observed pin depth limits (which account for measurement uncertainties) in Table 2-2 on page 2-4 to evaluate the condition of device connectors.
3- 6 85052B
Page 27
Use, Maintenance, and Care of the Devices
Gaging Connectors
When to Gage Connectors
Gage a connector at the following times:
Prior to using a device for the first time, record the pin depth measurement so that it can be compared with future readings. (It will serve as a good troubleshooting tool when you suspect damage may have occurred to the device.)
If either visual inspection or electrical performance suggests that the connector interface may be out of typical range (due to wear or damage, for example).
If a calibration device is used by someone else or on another system or piece of equipment.
Initially, after every 100 connections, and after that, as often as experience indicates.
85052B 3-7
Page 28
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Procedures
Gaging 3.5 mm Connectors
NOTE Always hold a connector gage by the gage barrel, below the dial indicator. This gives the
best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
1. Select the proper gage for your connector. Refer to Table 6-1 on page 6-2 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual Inspection” and
“Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage
to the gage master by interconnecting the male and female connectors. Connect the nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in
the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the device connector (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage
to the device by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in
the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read only the black signs;not the red signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4.
3- 8 85052B
Page 29
Figure 3-2 Gaging 3.5 mm Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85052B 3-9
Page 30
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging the 3.5 mm Sliding Loads
Gage the sliding load before each use. If the sliding load pin depth is out of the observed pin depth limits listed in Table 2-2 on page 2-4, refer to “Adjusting the Sliding Load Pin Depth” on page 3-12.
NOTE Always hold a connector gage by the gage barrel, below the dial indicator. This gives the
best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
NOTE The sliding load uses a plastic centering bead to support its center conductor when pin
depth is adjusted and gaged and when the load is stored. Remove this support bead from the sliding load before you connect the load for an electrical calibration. Reinsert this support bead when you’ve finished using the sliding load.
1. Select the proper gage for your connector. Refer to Table 6-1 on page 6-2 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual Inspection” on page
3-3 and “Cleaning Connectors” on page 3-4 earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2 on page 3-9):
a. While holding the gage by the barrel, and without turning the gage or the device, connect the gage
to the gage master by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating, use the torque wrench included
in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the sliding load connector (refer to Figure 3-3):
a. Unlock the center conductor pullback mechanism by raising the pullback handle to the unlocked
position.
b. Carefully move the pullback mechanism toward the connector end of the sliding load. The center
conductor will extend beyond the end of the connector. Continue to hold the pullback mechanism in this position.
c. Pull the sliding ring back approximately 0.5 in and install a centering bead (if not already installed) in
the connector end of the sliding load.
CAUTION The sliding load center conductor can be damaged if the sliding load is not in alignment
with the mating connector while making the connection.
3- 10 85052B
Page 31
Use, Maintenance, and Care of the Devices
Gaging Connectors
d. Keep the center conductor extended by holding the center conductor pullback mechanism toward
the connector end of the sliding load. Align the sliding load with the mating connector on the gage and mate the sliding load center conductor with the gage center conductor.
e. Release the center conductor pullback mechanism and move the body of the sliding load toward the
gage to mate the outer conductor of the sliding load connector with the outer conductor of the gage connector.
f. Without turning the gage or the sliding load, connect the gage to the sliding load being measured by
interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
g. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in
the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-14 for additional information.
CAUTION Always move the center conductor pullback mechanism back before locking the handle. Do
not force the handle past the locked position.
h. Move the center conductor pullback mechanism back (away from the connector end of the sliding
load), and place the pullback handle in its locked position.
i. Gently tap the barrel of the gage with your finger to settle the gage reading.
j. Read the gage indicator dial. Read only the black signs;not the red signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. Use different orientations of the gage within the connector. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
k. Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4. If the pin
depth is outside the limits, it must be adjusted before proceeding. Refer to “Adjusting the Sliding
Load Pin Depth” on page 3-12.
85052B 3-11
Page 32
Use, Maintenance, and Care of the Devices
Gaging Connectors
Figure 3-3 Gaging the 3.5 mm Sliding Loads
l. Without turning the gage or the sliding load, loosen the connection between the gage and the
sliding load and remove the sliding load from the gage.
CAUTION Damage can occur to the sliding load during the removal of a centering bead that has
slipped too far into the sliding load. If you’re going to perform an electrical calibration, prevent damage by removing the centering bead immediately after gaging the sliding load pin depth. The sliding load will not perform to its specifications if the centering bead is not removed from the sliding load before an electrical calibration.
m. Leave the centering bead installed on the sliding load if you are going to adjust the pin depth.
Carefully remove the centering bead from the sliding load if you’re going to use it for an electrical calibration. If the centering bead does not come out of the sliding load easily:
i. Unlock the center conductor pullback handle and move the center conductor pullback
mechanism toward the connector end of the sliding load to extend the center conductor.
ii. While holding the center conductor pullback mechanism toward the connector end of the sliding
load, remove the centering bead.
If the centering bead still will not come out:
i. Hold the sliding load with the connector end pointed downward.
ii. Move the sliding ring up, then quickly down. The trapped air behind the centering bead should
eject it.
Return the center conductor pullback mechanism to the rear of the sliding load and return the pullback handle to its locked position.
Adjusting the Sliding Load Pin Depth
The sliding loads in this kit have a setback mechanism that allows the pin depth to be set to any desired
3- 12 85052B
Page 33
Use, Maintenance, and Care of the Devices
Gaging Connectors
value. The pin depth of the sliding load is preset at the factory. The pin depth should not have to be reset each time the sliding load is used, but it should be checked before each use.
If the pin depth is outside the observed limits listed in Table 2-2 on page 2-4, use the following procedure to reset it to the nominal value of
0.00381 mm (0.00015 in).
This procedure assumes that you were directed here from “Gaging the 3.5 mm Sliding Loads” on page 3-10. If not, perform the steps in that procedure before performing this procedure.
1. The gage should be attached to the sliding load. The sliding load should have its centering bead installed. Refer to “Gaging the 3.5 mm Sliding Loads” on page 3-10 if necessary.
2. The face of the gage and the label on the sliding load should be facing up.
3. The center conductor pullback handle should be in the locked position.
4. With a small screwdriver, gently turn the center conductor pin depth adjustment screw until the gage pointer reads
0.00381 mm (0.00015 in). Refer to Figure 3-4 for the location of the adjustment screw.
5. Wait approximately five minutes to allow the temperature to stabilize. Do not touch either the gage or the sliding load during this time.
6. Note the gage reading. If it is no longer within the allowable range, perform step 4 again.
7. Move the center conductor pullback handle to the unlocked position and then back to the locked position. The gage reading should return to the value previously set. If not, repeat steps 4 through 7.
8. Return to “Gaging the 3.5 mm Sliding Loads” on page 3-10.
Figure 3-4 Adjusting the Sliding Load Pin Depth
85052B 3-13
Page 34
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4.
4. Use a connector gage to verify that all center conductors are within the observed pin depth values in
Table 2-2 on page 2-4. Refer to “Gaging Connectors” on page 3-6.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
6. Push the connectors straight together and tighten the connector nut finger tight.
CAUTION Do not turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Tab l e 3 - 1 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Refer to
Table 6-1 on page 6-2 for replacement part number and ordering information.
Table 3-1 Torque Wrench Information
Co n n e c to r Typ e Torq u e S e t t i n g Torq u e To l e r an c e
3.5 mm 90 N-cm (8 in-lb) 9.0 N-cm (0.8 in-lb)
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector
3- 14 85052B
Page 35
Use, Maintenance, and Care of the Devices
Connections
damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). However, it is recommended that you use an open-end wrench to keep the body of the device from turning.
1. Position both wrenches within 90 degrees of each other before applying force. See Figure 3-5. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.
Figure 3-5 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-6.
Figure 3-6 Using the Torque Wrench
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise, you apply an unknown amount of torque.
4. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-6 on page 3-15. Do not tighten the connection further.
85052B 3-15
Page 36
Use, Maintenance, and Care of the Devices
Connections
CAUTION You don’t have to fully break the handle of the torque wrench to reach the specified torque;
doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.
Connecting the Sliding Load
NOTE The sliding load uses a plastic centering bead to support its center conductor when pin
depth is adjusted and gaged and when the load is stored. Remove this support bead from the sliding load before you connect the load for an electrical calibration. Reinsert this support bead when you’ve finished using the sliding load.
1. Unlock the center conductor pullback mechanism by raising the pullback handle to the unlocked position. Refer to Figure 3-7.
2. Carefully move the pullback mechanism toward the connector end of the sliding load. The center conductor will extend beyond the end of the conductor. Continue to hold the pullback mechanism in this position.
CAUTION The sliding load center conductor can be damaged if the sliding load is not in alignment
with the mating connector while making the connection.
3. Keep the center conductor extended by holding the center conductor pullback mechanism toward the connector end of the sliding load. Align the sliding load with the mating connector on the cable or test port to which it is being connected and mate the sliding load center conductor with the center conductor of the cable or test port.
4. Release the center conductor pullback mechanism and move the body of the sliding load toward the cable or test port to mate the outer conductor of the sliding load connector to the outer conductor of the cable or test port connector.
5. Without turning the sliding load, connect the sliding load to the cable or test port by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
6. Using an open-end wrench to keep the device body from rotating, use the torque wrench included in the kit to tighten the connecting nut to the specified torque. Refer to
“Final Connection Using a Torque Wrench” on page 3-14 for additional information.
CAUTION Always move the center conductor pullback mechanism back before locking the handle. Do
not force the handle past the locked position.
7. Move the center conductor pullback mechanism back (away from the connector end of the sliding load), and place the pullback handle in its locked position.
3- 16 85052B
Page 37
Figure 3-7 Connecting the Sliding Load
Use, Maintenance, and Care of the Devices
Connections
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.
CAUTION Do not turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is twisted.
If disconnecting a sliding load, leave the center conductor pullback handle in the locked position.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connecting nut.
3. Complete the separation by hand, turning only the connecting nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
85052B 3-17
Page 38
Use, Maintenance, and Care of the Devices
Using the Sliding Load
Using the Sliding Load
When performing a sliding load calibration, it is recommended that the sliding ring be set at the marked positions (rings) along the sliding load body. Using the set marks ensures that a broad distribution of phase angles is selected, thereby optimizing the calibration.
The set marks function as detents so that the internal center of the sliding ring can mate with them. Because of this, the set mark being used cannot be seen but is felt as the sliding ring is moved from mark to mark during a calibration. Moving the sliding ring with only the index fingers of both hands will increase your ability to detect the sliding ring detent at each position.
NOTE The sliding load uses a plastic bead to support its center conductor when it is stored.
Remove this support bead from the sliding load before you connect the load for use. Reinsert this support bead when you are finished using the sliding load.
Preferred Method:
1. Move the sliding ring forward as far as possible toward the connector end of the load.
2. Move the sliding ring back until you feel it detent at the first set mark.
3. Follow the network analyzer instructions to measure the sliding load at each detent position.
NOTE After a calibration has begun, always move the sliding ring toward the center conductor
pullback end of the sliding load. If you slightly overshoot the desired mark by less than 0.5 mm (0.02 inch), do not move the sliding ring, but continue with the calibration as if the sliding ring is set to the proper position. If the sliding ring is moved toward the connector end of the load during the calibration sequence, the calibration may be unstable and poor measurements may result. If the desired position is overshot by more than 0.5 mm (0.02 inch), restart the calibration sequence from step 1.
It does not matter if you start with the slide at the front and move it backwards, or start at the back and move it forward. To minimize stability errors, it is important to start at one end and move it in the same direction for each of the measurement steps.
This manual contains limited information about a sliding load calibration. For detailed information on this calibration using a VNA, ENA or PNA series network analyzer, refer to the appropriate user guide or online Help.
To view the ENA or PNA online Help, press the Help key on the front panel of the network analyzer.
To view an online VNA user guide, use the following steps:
1. Go to www.keysight.com.
2. Enter your VNA model number (Ex: N5242A) in the Search box and click Search.
3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF you want to
view.
If you need additional information, see “Where to Look for More Information” on page 5-3.
3- 18 85052B
Page 39
Figure 3-8 Sliding Load Set Marks
Use, Maintenance, and Care of the Devices
Handling and Storage
Handling and Storage
Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use.
Never store connectors loose in a box, or in a desk or bench drawer. This is the most common cause of connector damage during storage.
Keep connectors clean.
Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
85052B 3-19
Page 40
Use, Maintenance, and Care of the Devices
Handling and Storage
3- 20 85052B
Page 41
4 Performance Verification
Page 42
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
How Keysight Verifies the Devices in Your Kit
Keysight verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to the National Institute of Standards and Technology (NIST). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Keysight includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the Institute's calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies.
4- 2 85052B
Page 43
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
a new calibration sticker affixed to the case
a certificate of calibration
a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTE A list of NIST traceable numbers may be purchased upon request to be included in the
calibration report.
Keysight Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Keysight Technologies. Refer to “Contacting Keysight” on page 5-5 for a list of offices.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.
NOTE The recertification interval should begin on the date the kit is first used after the
recertification date.
Where to Send a Kit for Recertification
Contact Keysight Technologies for information on where to send your kit for recertification. Contact information is listed on page 5-5. Refer to “Returning a Kit or Device to Keysight” on page 5-4 for details on sending your kit.
85052B 4-3
Page 44
Performance Verification
Recertification
4- 4 85052B
Page 45
5 Troubleshooting
Page 46
Troubleshooting
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5- 2 85052B
Page 47
Troubleshooting
Where to Look for More Information
Where to Look for More Information
This manual contains limited information about network analyzer system operation. For detailed information on using an ENA or PNA series network analyzer, refer to the appropriate user guide or Help file.
To view the ENA or PNA Help, press the Help key on the front panel of the network analyzer.
To view Help or a user guide online, use the following steps:
1. Go to www.keysight.com.
2. Enter your analyzer model number (Ex: N5242A) in the Search box and click Search.
3. Under the heading Manuals, click on the title/hyperlink for the document PDF you want to view.
If you need additional information, see “Contacting Keysight” on page 5-5.
85052B 5-3
Page 48
Troubleshooting
Returning a Kit or Device to Keysight
Returning a Kit or Device to Keysight
If your kit or device requires service, contact Keysight Technologies for information on where to send it. See
“Contacting Keysight” on page 5-5 for contact information. Include a service tag (located near the end of
this manual) on which you provide the following information:
your company name and address
a technical contact person within your company, and the person's complete telephone number
the model number and serial number of the kit
the part number and serial number of each device
the type of service required
•a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)
5- 4 85052B
Page 49
Troubleshooting
Contacting Keysight
Contacting Keysight
Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at:
www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
NOTE In any correspondence or telephone conversation, refer to the Keysight product by its
model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.
85052B 5-5
Page 50
Troubleshooting
Contacting Keysight
5- 6 85052B
Page 51
6 Replaceable Parts
Page 52
Replaceable Parts
Introduction
Introduction
Ta b l e 6 - 1 lists the replacement part numbers for the 85052B calibration kit. Ta b le 6 - 2 lists the replacement
part numbers for items not included in the calibration kit that are either required or recommended for successful operation of the kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Keysight Technologies. See “Contacting Keysight” on page 5-5 for contact information.
Table 6-1 Replaceable Parts for the 85052B Calibration Kit
Descriptiona
Calibration Devices (3.5 mm)
-m- sliding load 1 00911-60019
-f- sliding load 1 00911-60020
-m- broadband load
-f- broadband load
-m- offset short 1 85052-60006
-f- offset short 1 85052-60007
-m- offset open 1 85052-60008
-f- offset open 1 85052-60009
-f- to -f- 1 85052-60012
-m- to -f- 1 85052-60013
-m- to -m- 1 85052-60014
b
b
Adapters (3.5 mm)
Qty Per
Kit
1
1
Keysight Part Number
00902-60005
00902-60003
00902-60006
00902-60004
c
c
Protective End Caps for Connectors
For 3.5-mm -f- connectors
For 3.5-mm -m- connectors
Wrench
5/16 in, 90 N-cm (8 in-lb) torque wrench 1 8710-1765
7 mm open-end wrench 1 8710-1761
Calibration Kit Storage Case
Storage box assembly (box with foam pads) 1 85052-60044
as
required
as
required
1401-0202
1401-0208
6- 2 85052B
Page 53
Table 6-1 Replaceable Parts for the 85052B Calibration Kit
Replaceable Parts
Introduction
Descriptiona
Box (without foam pads) 1 5180-7900
Foam pad (for lid) 1 5181-5543
Foam pad (for lower case) 1 85052-80031
Connector Gages (3.5 mm)
Gage Set (for female connectors) 1 11752-60105
Gage Set (for male connectors) 1 11752-60106
Centering bead (for gaging 3.5 mm sliding load) 2 85052-20057
Miscellaneous Items
User’s and service guide 1 85052-90077
“Clarifying the Terminology of a Connector Interface” on page 1-7.
a. See b. Broadband load has replaced lowband load. c. Part numbers in italic typeface are for production assemblies and cannot be ordered by customers. Both the
replacement part number and the production part number are interchangeable and have identical performance.
Qty Per
Kit
Keysight Part Number
85052B 6-3
Page 54
Replaceable Parts
Introduction
Table 6-2 Items Not Included in the Calibration Kit
Description Qty Keysight Part Number
ESD Protection Devices
Grounding wrist strap 1 9300-1367
5 ft grounding cord for wrist strap 1 9300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797
ESD heel strap 1 9300-1308
Connector Cleaning Supplies
Anhydrous isopropyl alcohol (92% pure)
Cleaning swabs 100 9301-1243
3.5 mm slotless connector contact repair kit
Sliding load handle replacement kit 1 85052-60047
Keysight Application Note An 1287-11 1 5989-4840EN
a. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally. b. All female connectors on the precision devices in this kit are slotless connectors. Refer to “Inspect Female
Connectors” on page 3-3
a
Miscellaneous Items
b
.
-- --
1 85052-60049
6- 4 85052B
Page 55
Figure 6-1. Replaceable Parts for the 85052B Calibration Kit
Replaceable Parts
Introduction
85052B 6-5
Page 56
Replaceable Parts
Introduction
Figure 6-2 Replaceable Parts for the 85052B Calibration Kit
6- 6 85052B
Page 57
A Standard Definitions
Page 58
Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web
Class Assignments and Standard Definitions Values are Available on the Web
Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at
http://na.support.keysight.com/pna/caldefs/stddefs.html.
For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying Calibration Standards and Kits for Keysight Vector Network Analyzers.” This application note covers calibration standard definitions, calibration kit content and its structure requirements for Keysight vector network analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an existing calibration kit definition file. To download a free copy, go to www.keysight.com and enter literature number 5989-4840EN in the Search window.
A- 2 85052B
Page 59

Index

A
accuracy
connector gage, 3-6
adapters, 1-3
part numbers, 6-2 specifications, 2-5
agreements
customer assistance, -5 maintenance, -5
alcohol
isopropyl
as cleaning solvent, 3-5 precautions for use of, 3-5
assistance
customer, -5 who to contact, -5
B
box, part number, 6-2 broadband loads, 1-2
part numbers, 6-2 specifications, 2-5
C
cal kit
contents, 1-2 verifying, 4-2
calibration
bad, 5-2 certificate of, 4-3 constants, See calibration definitions definitions
entering, 1-3
permanently stored, 1-3 frequency, 1-8 report, 4-3 temperature, 2-2 when to perform, 1-8
calibration kit
contents, 1-2
drawing of, 6-5, 6-6 Keysight Application Note, A-2 modifying definition files, A-2 overview, 1-2 performance
how Keysight verifies, 4-2
case, part number, 6-2 certificate of calibration, 4-3 certification
of device specifications, 2-6
characteristics
mechanical, 2-3
class assignments
downloading from Keysight Web site,
A-2
cleaning connectors, 3-4
cleaning supplies, 1-3
part number, 6-4
compressed air
for cleaning, 3-4 conductive mat, part number, 6-4 connections, 3-2, 3-14
ESD protection, 3-2
final, 3-14
preliminary, 3-14
separating, 3-17
sliding load, 3-17 sliding load, 3-16 using torque wrench, 3-14
connector
cleaning, 3-4 damage, 3-3 defects, 3-3 female, 3-3 gage
accuracy, 3-6
handling, 3-6, 3-8, 3-10
use of, 3-6
zeroing, 3-6, 3-8, 3-10 gaging, 3-6, 3-8
to determine pin depth, 3-6
when to do, 3-7 gender, 1-7 mating plane surfaces
cleaning, 3-5 slotless, 1-6, 3-4, 6-4 terminology, 1-7 threads
cleaning, 3-4
inspecting, 3-3 visual inspection, 3-3 wear, 3-3
affect on electrical performance, 3-3
constants, calibration, See calibration
definitions
contacting Keysight Technologies, 5-4 contents
calibration kit, 1-2 drawing of, 6-5, 6-6 incomplete
what to do, 1-4
D
damage
caused by electrostatic discharge, 3-2 device, 3-3 inspecting for, 3-3 to connectors, 3-3
data
recertification, 4-3
defective connectors, 3-3 definitions
calibration, 1-3
part numbers, 6-2
deviation from nominal phase, 2-5 device
cleaning, 3-4 connecting, 3-14 damage, 3-3 disconnecting, 3-17 handling, 3-19 maintenance, 1-7 performance
verifying, 4-2
specifications, 2-5
certification of, 2-6 electrical, 2-5
traceability, 4-2, 4-3 storage, 3-19 temperature, 2-2 visual inspection, 3-3
disconnections, 3-17 documentation, part number, 6-3 downloading class assignments & std
definitions from the Web
E
electrical characteristics
effects of temperature, 2-2
electrical specifications, 2-5 electrostatic discharge, See ESD end cap, part numbers, 6-2 environmental
regulations, 3-5 requirements, 2-2 specifications, 2-2
equipment required, 1-3 ESD, 3-2
protection, 3-2 supplies
part numbers, 6-4
F
female connectors, 3-3
inspection of, 3-3
flowchart, troubleshooting, 5-2 frequency of calibration, 1-8 frequency specifications, 2-5
G
gage
connector, 1-3
handling, 3-8, 3-10 part numbers, 6-3 zeroing, 3-8, 3-10
gaging
connectors, 3-6, 3-8
when to do, 3-7 procedures, 3-8
, A-2
85052B Index-1
Page 60
Index
sliding loads, 3-10
to determine pin depth, 3-6 gender, connector, 1-7 general information, 1-1
H
handling, 3-19 heel strap, part number, 6-4 how often to calibrate, 1-8 humidity specifications, 2-2
I
incoming inspection, 1-4 information, troubleshooting, 5-3 inspection
damage, 3-3
defects, 3-3
female connectors, 3-3
incoming, 1-4
mating plane surfaces, 3-3
visual, 3-3 isopropyl alcohol
as cleaning solvent, 3-5
precautions for use of, 3-5
K
Keysight Technologies
application note, A-2
contacting, 5-4, 5-5 kit
contents, 1-2, 6-5, 6-6
drawing of, 6-5, 6-6
overview, 1-2
L
limits
pin depth, 2-4 loads
broadband, 1-2
part numbers, 6-2
M
maintenance, 3-2
agreements, -5
of devices, 1-7
preventive, 1-7 making connections, 3-14
ESD protection, 3-2
precautions, 3-14 manual
copyright dates, -4
part number, -4, 6-2
print date, -4
printing, -5
title, -4
mat
conductive
part number, 6-4
mating plane surfaces
cleaning, 3-5 connector, 3-5 inspection of, 3-3
mechanical characteristics, 2-3
affect on electrical performance, 2-3 verifying, 3-6
modifying calibration kit definition files,
A-2
N
National Institute of Standards and
Tec h n o l o g y (N I ST )
nitrogen, for cleaning, 3-4 numbers
replaceable parts, 6-2, 6-4 serial, 1-5
O
observed limits
pin depth, 2-4
offset opens and shorts, 1-2
part numbers, 6-2 specifications, 2-5
open-end wrench, 1-2, 3-17
part number, 6-2
opens, 1-2
part numbers, 6-2 specifications, 2-5
ordering parts, 6-2, 6-4
P
part numbers, 6-2
of items in kit, 6-2 of items not in kit, 6-4
parts
included in kit, 1-2, 6-2 not included in kit, 6-2, 6-4 ordering, 6-2 replaceable, 6-2
performance verification
failure, 5-2
permanently stored calibration
definitions
pin depth, 2-3
adjusting sliding load, 3-12 affect on electrical performance, 2-4 gaging to determine, 3-6 observed limits, 2-4, 3-6 protrusion, 2-3 recession, 2-3 typical values, 2-4
precision adapters, 1-3
, 2-6, 4-2
, 1-3
part numbers, 6-2 precision slotless connectors, 1-6 preventive maintenance, 1-7 procedures
cleaning, 3-4
gaging, 3-8
zeroing, 3-8 protective end caps, part numbers, 6-2 protrusion
pin depth, 2-3
R
recertification
how to order, 4-3
interval, 4-3
locations, 4-3
services included, 4-3 recession
pin depth, 2-3 regulations
environmental, 2-2, 3-5 replaceable parts, 6-2, 6-4
drawing of, 6-5, 6-6 report, calibration, 4-3 requirements
environmental, 2-2 return kit or device to Keysight, 1-4, 5-4 return loss, specifications, 2-5
S
separating connections, 3-17 serial numbers, 1-5
devices, 1-5
recording, 1-5 service, 5-4 service tag, 4-3, 5-4 set marks
sliding load, 3-18 shorts, 1-2
part numbers, 6-2
specifications, 2-5 sliding load, 1-3
calibration, 3-18
connecting, 3-16
gaging, 3-10
pin depth
adjusting, 3-12 separating connection, 3-17 set marks, 3-18 sliding ring, 3-18 specification, 2-5 using, 3-18
specifications, 2-2
adapters, 2-5 broadband loads, 2-5 certification of, 2-6 deviation from nominal phase, 2-5
Index-2 85052B
Page 61
Index
device, 2-5 electrical, 2-5 environmental, 2-2 frequency, 2-5 humidity
operating, 2-2
storage, 2-2 offset opens and shorts, 2-5 return loss, 2-5 sliding loads, 2-5 temperature, 2-2 torque wrench, 3-14 traceability, 4-2, 4-3
standard definitions
downloading from Keysight Web site,
A-2
standards
international, 2-6 National Institute of Standards and
Technology (NIST)
static
discharge, 3-2 electricity, 3-2
storage, 3-19 storage case, part number, 6-2 strap
heel and wrist, part number, 6-4
supplies
cleaning, 1-3
part number, 6-4
swabs
cleaning, 3-5
, 2-6, 4-2
troubleshooting, flowchart, 5-2
U
user’s and service guide
part number, 6-3
V
verification
performance, 4-3 temperature, 2-2
visual inspection, 3-3
W
wear
connector, 3-3
affect on electrical performance, 3-3
when to calibrate, 1-8 wrench
open-end, 1-3, 3-15, 3-17 proper positioning of, 3-15 torque, 1-3, 3-14, 3-15
precautions for use of, 3-15 proper use of, 3-15
wrist strap, part number, 6-4
Z
zeroing
connector gage, 3-8, 3-10
T
tag
service, 4-3, 5-4
temperature
affect on electrical performance, 2-2 calibration, 2-2 cautions about, 2-2 changes in, 2-2 device, 2-2 error-corrected, 2-2 measurement, 2-2 specifications, 2-2
operating, 2-2
storage, 2-2 verification and measurement, 2-2
test data, 4-3 threads
connector
cleaning, 3-4
inspecting, 3-3
torque wrench, 1-3
specifications, 3-14
traceability
of device specifications, 4-2, 4-3
85052B Index-3
Page 62
Index
Index-4 85052B
Page 63
This information is subject to change without notice.
© Keysight Technologies 1994 - 2014