Keysight 5500 User Manual

Keysight 5500 Scanning Probe Microscope

User’s Guide

Notices
© Keysight Technologies 2015
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Key­sight Technologies as governed by United States and international copyright laws.
Manual Part Number
N9410-90001
Edition
Rev H, December 2015
Printed in USA
Keysight Technologies 5301 Stevens Creek Blvd. Santa Clara, CA 95051
Warranty
The material contained in this doc­ument is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight dis­claims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for inci­dental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agree­ment with warranty terms covering the material in this document that conflict with these terms, the war­ranty terms in the separate agree­ment shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ stan­dard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indi­cated conditions are fully understood and met.

Read This First

Read This First
Warranty
Keysight warrants Keysight hardware, accessories and supplies against defects in material and workmanship for a period of one year from date of shipment. If Keysight receives notice of such defects during the warranty period, Keysight will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
Keysight warrants that Keysight software will not fail to execute its programming instructions for the period specified above due to defects in material and workmanship when properly installed and used. If Keysight receives notice of such defects during the warranty period, Keysight will replace software media which does not execute its programming instructions due to such defects. For detailed warranty information, see back matter.
Safety Considerations
General - This product and related documentation must be reviewed
for familiarization with these safety markings and instructions before operation.
This product is a safety Class I instrument (provided with a protective earth terminal).
Before Applying Power - Verify that the product is set to match the available line voltage and the correct fuse is installed. Refer to instructions in “Facility Requirements" on page 4 of the manual.
Before Cleaning - Disconnect the product from operating power before cleaning.
Safety Earth Ground - An uninterrupted safety earth ground must be provided from the main power source to the product input wiring terminals or supplied power cable.
N9410-90001 Keysight 5500 SPM User’s Guide iv

Specifications

Read This First
Environmental Conditions
Temperature (Operating): 5 to 40 °C
Temperature (Non-operating): -40 to 70 °C
Relative Humidity (Operating): 15 to 95 % non-condensing
Altitude: 2000 m
Power Requirements
100/120/220/240 VAC, 50/60 Hz
Mains supply voltage fluctuations are not to exceed 10 % of the nominal supply voltage.
NOTE
CAUTION
These specifications apply to the Keysight 5500 system, and do not guarantee the function of an experiment (including the cantilever) in these conditions.
Equipment Class I, Pollution Degree 2, Installation Category II.
This equipment is for indoor use only.
When the product is subjected to 8 kV air discharge or 4 kV contact discharge in accordance with IEC 61000-4-2, interruption of the laser output may occur.
If this happens, laser power must be re-cycled in order to resume normal operation. CAUT
Stop using the scanner if the scanner cable insulation is damaged in order to avoid electrical shock. Have it repaired or replaced by the factory.
N9410-90001 Keysight 5500 SPM User’s Guide v

Voltage and Laser Safety Information

The 5500 SPM will be marked with the following Instruction Manual symbol when it is necessary for the user to refer to this User’s Guide:
The following symbol indicates hazardous voltages:
Read This First
WARNING
This system is designed to be used with a Class II or Class III diode laser with an output of up to 1 mW of visible radiation at 670 nm or 980 nm. The aperture in the AFM scanning head is labeled with the logotype (shown below). DO NOT stare directly into the laser beam. To ensure safe operation, the scanner must be operated and maintained in accordance with the instructions included with the laser. The laser must only be powered by a controller that includes an on/off switch, such as the Keysight SPM Controller. DO NOT attempt to make any adjustments to the laser, the laser’s electronics, or optics. If laser malfunction is suspected, immediately return the scanner to Keysight Technologies for repair or replacement; there are no user-serviceable parts. WA
RN
Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous light exposure. Furthermore, the use of optical instruments with this product may increase eye hazard.
N9410-90001 Keysight 5500 SPM User’s Guide vi

Power Supply

Read This First
In accordance with federal FDA requirements, one of the following laser precautions is affixed to the scanner:
It is not necessary to open the Keysight AFM Controller to make changes to the power supply. However, the power cord should always be unplugged before making any adjustments to the power source. The Keysight SPM Controller has several different power supply options.
Procedure for Changing Input Voltage
1 Unplug the power cord from the Keysight AFM Controller.
2 Remove the fuse holder located on the back of the controller.
3 Underneath where the fuse holder was located is the input voltage
4 Reinsert the voltage switch with the desired voltage.
5 Replace the fuse holder.

Piezo Scanner Precautions

Piezo scanners are, by nature, very FRAGILE pieces of equipment. The piezo material that does the scanning is a ceramic and is consequently quite easily broken. Dropping a piezo scanner will result in damage to the scanner that can only be repaired by completely replacing the scanner piezo core. This can be an expensive and time-consuming process and so it is advised that the utmost care is used when handling the scanners. Keysight Technologies recommends that the scanners be stored in the padded scanner case that was supplied with the scanner and that the scanner be kept in a dry environment when not in use. Piezo scanners also perform better with consistent use. If a scanner is not used for some time it may require a short period of use before the scan range is stable and the calibration is correct. It may also be necessary to re-calibrate the scanner from time to time. The calibration can be
control switch. Pull out the switch and rotate it to the desired input voltage 100/120/220/240 V.
N9410-90001 Keysight 5500 SPM User’s Guide vii
verified using a calibration standard, and adjustments can be made using the calibration tools.

General Care Requirements

SPM equipment is sensitive scientific equipment. Care must be used when handling all parts. When removing scanners from the microscope ensure that all cable connections to the scanner are disconnected. This includes cables for photo-diode detectors. Also, the photo-diode detector should be removed from the scanner prior to the removal of the scanner from the microscope. All equipment, especially the sample plates and scanner nose modules should be kept clean and free from contamination when not in use. It is recommended, to prolong the life of these items, that after use all sample plates and noses are cleaned thoroughly and dried off prior to storage. Cleaning can be done using an organic solvent. Please refer to the appropriate sections of the manual for further information regarding the proper cleaning of equipment.
Read This First

Disclaimers

This User’s Guide, as well as the hardware herein described, is licensed and can only be used in compliance with such terms and agreements as entered in by Keysight Technologies. Users of these products understand, except where permission is given by Keysight Technologies by said license, no part of this manual may be copied, transmitted, stored in a general retrieval system, in any form or means, electronic, or mechanical, without prior written permission of Keysight Technologies. Information contained herein this User’s Guide is for general information use only. Information is subject to change without notice. Information should not be construed as a commitment by Keysight Technologies. Furthermore, Keysight Technologies assumes no responsibility or liability for any misinformation, errors, or general inaccuracies that may appear in this manual.
N9410-90001 Keysight 5500 SPM User’s Guide viii

Declaration of Conformity

Read This First
N9410-90001 Keysight 5500 SPM User’s Guide ix

Contact Information

Read This First
Keysight Technologies
5301 Stevens Creek Blvd., Santa Clara, CA 95051 U.S.A.
Tel: +1.480-756-5900 Fax: +1.480-756-5950
E-mail: AFM-info@keysight.com Web: www.keysight.com
Customer Technical Support
Tel: +1-480-756-5900
Fax: +1-480-756-5950
E-mail: AFM-Support@keysight.com
Technical Sales
Tel: +1-480-756-5900
Fax: +1-480-756-5950
E-mail: AFM-info@keysight.com
Distributors and Account Representatives
Please visit our web site for up-to-date information:
www.keysight.com/find/nano
N9410-90001 Keysight 5500 SPM User’s Guide x

Table of Contents

I Read This First
Specifications I-v
Voltage and Laser Safety Information I-vi
Power Supply I-vii
Piezo Scanner Precautions I-vii
General Care Requirements I-viii
Disclaimers I-viii
Declaration of Conformity I-ix
Contact Information I-x
Contents
II Table of Contents
1 Introduction to the Keysight 5500
Overview of Keysight SPM System 1-2
SPM Basics 1-3
SPM Techniques 1-5
Scanning Tunneling Microscopy (STM) 1-5
Atomic Force Microscopy (AFM) 1-6
Contact Mode AFM 1-8
Intermittent Contact AFM 1-9
Current Sensing Mode (CSAFM) 1-12
Force Modulation Microscopy (FMM) 1-13
Lateral Force Microscopy (LFM) 1-14
Dynamic Lateral Force Microscopy (DLFM) 1-14
Magnetic Force Microscopy (MFM) 1-15
Electrostatic Force Microscopy (EFM) 1-15
Kelvin Force Microscopy (KFM) 1-15
N9410-90001 Keysight 5500 SPM User’s Guide 1
2 Keysight 5500 SPM Components
Microscope 2-3
Probes 2-4
Nose Assembly 2-5
One-Piece Nose Assemblies 2-5
Two-Piece Nose Assemblies 2-6
Scanner 2-7
Detector 2-9
Sample Plates 2-10
Video System 2-12
Head Electronics Box (HEB) 2-13
AFM Controller 2-15
Breakout Box 2-16
Contents
Vibration Isolation Chamber 2-21
Software 2-22
System Options 2-23
MAC Mode 2-23
MAC III Mode 2-23
Liquid Cell 2-23
Temperature Control 2-24
Thermal K 2-24
Environmental Chamber 2-24
Glove Box 2-25
Electrochemistry 2-26
PicoTREC 2-26
PicoLITH 2-26
3 Setting Up the Keysight 5500 SPM
Component and Facility Dimensions 3-1
Facility Requirements 3-4
Utilities 3-5
Noise and Facility Specifications 3-5
Assembling the Vibration Isolation Chamber 3-7
Switching Door Orientation 3-9
N9410-90001 Keysight 5500 SPM User’s Guide 2
Managing Cables 3-10
Connecting the Components 3-12
Guidelines for Moving the System 3-13
Care and Handling of the Probes and Scanner 3-14
Probes 3-14
Nose Assembly 3-14
Scanner 3-14
4 Preparing for Imaging
Setting Up the Scanner Assembly 4-1
One-Piece Nose Assembly 4-2
Two-Piece Nose Assembly 4-9
Inserting the Scanner and Connecting Cables 4-12
Contents
Aligning the Laser 4-14
Inserting and Aligning the Detector 4-22
Mounting the Sample 4-26
Using the Video System 4-29
5 Contact Mode Imaging
Setting Up for Contact Mode Imaging 5-2
Constant Force Mode 5-2
Constant Height Mode 5-9
Fine-Tuning the Image 5-9
6 AC Modes
Acoustic AC Mode (AAC) 6-3
AAC Mode 6-3
Constant Height Mode 6-8
Magnetic AC (MAC) Mode 6-9
Standard MAC Mode 6-10
Top MAC Mode 6-12
Q Control 6-13
N9410-90001 Keysight 5500 SPM User’s Guide 3
7 Additional Imaging Modes
Scanning Tunneling Microscopy (STM) 7-1
STM Gain and Preamp Settings 7-3
STM Imaging Procedure 7-6
Current Sensing AFM (CSAFM) 7-9
Lateral Force Microscopy (LFM) 7-13
Dynamic Lateral Force Microscopy (DLFM) 7-14
Force Modulation Microscopy (FMM) 7-16
Magnetic Force Microscopy 7-19
Electrostatic Force Microscopy (EFM) 7-25
Kelvin Force Microscopy (KFM) 7-29
Force Spectroscopy 7-31
Force Spectroscopy Procedure for Contact Mode 7-32
Contents
FlexGrid Spectroscopy 7-36
FlexGrid Spectroscopy Procedure for Contact Mode 7-36
Volume Spectroscopy 7-41
Volume Spectroscopy Procedure for Contact Mode 7-41
8 Scanner Maintenance and Calibration
Care and Handling of the Probes and Scanner 8-3
Probes 8-3
Nose Assembly 8-3
Scanner 8-4
Scanner Characteristics 8-4
Non-Linearity 8-5
Sensitivity 8-5
Hysteresis 8-5
Other Characteristics 8-6
Calibrating the Multi-Purpose Scanner 8-8
X Calibration 8-9
Y Calibration 8-14
Non-Orthogonality 8-17
Z Calibration 8-18
Servo Gain Multiplier 8-19
N9410-90001 Keysight 5500 SPM User’s Guide 4
Archive the Calibration Files 8-19
9 Closed-Loop Scanners
Scanner Types 9-1
Z-Axis Closed-Loop Scanner 9-1
X/Y/Z Closed-Loop Scanner 9-2
Calibration 9-2
X and Y Sensor Calibration 9-3
Z Sensor Calibration 9-11
10 MAC Mode
List of MAC Mode Components 10-1
Contents
11 MAC III
Connections 10-2
Hardware and Sample Setup 10-4
Initial Setup 11-2
List of MAC III Components 11-2
Hardware and Sample Setup 11-6
Auxiliary Signal Access Box 11-6
MAC III Software Controls 11-11
Simplified Software Control Options 11-12
Advanced Software Control Options 11-24
Amplitude and Frequency Modulation Techniques 11-33
Setting Parameters for AM Mode Operation 11-36
Setting Parameters for FM Mode Operation 11-40
Piezoresponse Force Microscopy 11-43
12 Liquid Cell
Liquid Cell with Standard Sample Plate 12-2
Liquid Cell with MAC Mode 12-4
Flow-Through Liquid Cell 12-4
N9410-90001 Keysight 5500 SPM User’s Guide 5
13 Temperature Control
Cantilevers for Temperature Controlled Imaging 13-1
High Temperature Sample Plates 13-2
Connections 13-5
Imaging 13-11
Peltier (Cold MAC) Sample Plate 13-13
Connections 13-15
Water Cooling 13-20
Imaging 13-21
Tips for Temperature Controlled Imaging 13-22
14 Thermal K
Thermal K Calibration 14-3
Thermal K Setup 14-3
Contents
Spring Constant Calibration 14-4
15 Environmental Control
Environmental Chamber 15-1
Glove Box 15-3
16 Electrochemistry
Equipment 16-3
Liquid Cell 16-3
Electrodes 16-3
Cleaning 16-5
Liquid Cell Cleaning 16-5
Electrode Cleaning 16-6
Sample Plate Cleaning 16-6
Substrate Cleaning 16-6
Assembling and Loading the Liquid Cell 16-6
Troubleshooting 16-8
Electrochemistry Definitions 16-8
Software Controls 16-9
N9410-90001 Keysight 5500 SPM User’s Guide 6
Guidelines for Electochemistry 16-10
SECM - Scanning Electrochemical Microscopy 16-12
SECM System Setup 16-12
Assembling the SECM Nose Cone 16-14
SECM Software Interface 16-17
A Wiring Diagrams
Keysight 5500 SPM Standard Wiring Diagram A-2
Keysight 5500 SPM with MAC Mode Controller A-3
Keysight 5500 SPM with MAC Mode, Force Modulation
Imaging A-4
Keysight 5500 SPM with MAC III Option A-5
Keysight 5500 SPM with MAC III Option and Closed Loop
Scanner A-6
Contents
Index
N9410-90001 Keysight 5500 SPM User’s Guide 7
Keysight 5500 SPM User’s Guide
1 Introduction to the Keysight 5500
Overview of Keysight SPM System 1-2
SPM Basics 1-3
SPM Techniques 1-5
Scanning Tunneling Microscopy (STM) 1-5
Atomic Force Microscopy (AFM) 1-6
Intermittent Contact AFM 1-9
Acoustic AC (AAC) AFM 1-10
Magnetic AC (MAC) Mode 1-11
Top MAC Mode 1-12
Current Sensing Mode (CSAFM) 1-12
Force Modulation Microscopy (FMM) 1-13
Lateral Force Microscopy (LFM) 1-14
Dynamic Lateral Force Microscopy (DLFM) 1-14
Magnetic Force Microscopy (MFM) 1-15
Electrostatic Force Microscopy (EFM) 1-15
Kelvin Force Microscopy (KFM) 1-15
The Keysight 5500 SPM is the ideal multiple-user research system for Scanning Probe Microscopy (SPM). As the high-performance Atomic Force Microscope (AFM) flagship of Keysight’s product line, the 5500 SPM provides a wealth of unique technological features, including precision temperature control and industry-leading environmental control.
The Keysight 5500 SPM offers features and software for research in materials science, polymers, nanolithography and general surface characterization. With excellent ease of use, the 5500 SPM also affords educators an unprecedented opportunity to introduce students to AFM technology.

Overview of Keysight SPM System

The main component of the Keysight 5500 SPM system is the microscope (Figure 1-1), which includes the X/Y motion controls, scanner, high-resolution probe/tip, and detector. The control system for the microscope includes, at minimum, a high-speed computer, AFM controller and Head Electronics Box. Optional components include additional electronics, specialized scanners and probes for particular SPM techniques, and an environmental enclosure to control acoustic and vibration noise.
Introduction to the Keysight 5500 5
Figure 1-1 The Keysight 5500 SPM microscope, shown with optional
environmental chamber
In this User’s Guide we will begin with a brief introduction to Scanning Probe Microscopy techniques. The sections that follow will show you how to handle the 5500 SPM components and how to image in the available modes.
Keysight 5500 SPM User’s Guide 5-2

SPM Basics

Introduction to the Keysight 5500 5
Scanning Probe Microscopy (SPM) is a large and growing collection of techniques for investigating the properties of a sample, at or near the sample surface. The SPM instrument has a sharp probe (with radius of curvature typically in the nanometers or tens of nanometers) that is in near-contact, intermittent contact, or perpetual contact with the sample surface.
An SPM is used to investigate sample properties at or near the sample surface; that is, immediately beneath the surface (typically several nanometers deep) and immediately above the surface (typically several tens of nanometers high).
In SPM techniques, the sharp probe (tip) is scanned across a sample surface, or the surface is scanned beneath the tip (Figure 1-2). Interactions between the tip and sample are detected and mapped. Different techniques sense different interactions, which can be used to describe surface topography, adhesion, elasticity, electrostatic charge, etc.
Figure 1-2 Scanning Probe Microscopy diagram
The small size of the probe tip is key to the SPM’s high resolution. However, its small size also means that the tip must be scanned in order to image a significant area of the sample. SPM techniques use “raster scanning,” in which high resolution actuators, usually made of piezoelectric materials, move the probe across the sample and back over each line of the image area. For each X/Y coordinate pair, the interaction of the tip and sample is recorded as one data point. The
Keysight 5500 SPM User’s Guide 5-3
Introduction to the Keysight 5500 5
collection of data points is then synthesized into the “SPM image,” a 3-dimensional map of the surface characteristic being examined.
The most common SPM images are topography images, in which the third dimension, Z, for any given X/Y coordinates, is the relative height of the sample surface. This interpretation implies that the sharp probe does not deform the sample surface—the harder the sample surface, the more accurate is this interpretation. In other words, the tip follows the height variations of hard surface with higher fidelity than it does soft surfaces.
Topography measurements are in general calibrated against height standards.Therefore, topography images may be compared for quantitative information, provided the systems have been correctly calibrated and operated, and that the data is properly interpreted.
In other types of SPM images, the third dimension is a measure of the relative strength of a detectable interaction between the probe and sample. The image is usually recorded simultaneously with, and displayed along side, the topography image of the same sample area. This helps reveal any correlation between topography and the interaction.
In some instances, the signal from the SPM’s detector is mapped directly; for example, the deflection of the probe cantilever, or the current through a metal tip. In other instances, the signal from the detector serves as the input of a feedback system which attempts to maintain the detector signal at a user-defined setpoint. The output of the feedback system can then be mapped to construct the image.
SPM can also be used for “non-imaging techniques,” or “nano-manipulation,” in which the probe is used to modify the sample surface. For example, one can use the probe or tip to rearrange nanometer-scale objects physisorbed on that surface. Essentially, the tip serves as a nano-scale finger to interact with the sample.
Nano-manipulation is sometimes performed in the plane of the sample surface (in-plane) and sometimes at right angles to this plane (out-of-plane nano-manipulation). An example of out-of-plane nano-manipulation is attaching the probe tip to the end of a macromolecule on the sample surface, and pulling the molecule so that its secondary or tertiary structure unfolds. This is now an extremely active area of research, with applications extending to fields as diverse as drug discovery and composite materials design.
Keysight 5500 SPM User’s Guide 5-4

SPM Techniques

Scanning Tunneling Microscopy (STM)

Introduction to the Keysight 5500 5
The earliest, widely-adopted SPM technique was Scanning Tunneling Microscopy (STM). In STM, a bias voltage is applied between a sharp, conducting tip and the sample. When the tip approaches the sample, electrons “tunnel” through the narrow gap, either from the sample to the tip or vice versa, depending on the bias voltage. Changes of only 0.1 nm in the separation distance cause an order of magnitude difference in the tunneling current, giving STM remarkably high precision. The basic STM schematic is shown in Figure 1-3.
Figure 1-3 Basic STM schematic
STM can image a sample surface in either constant current or constant height mode, as described in Figure 1-4. In constant height mode, the tip remains in a constant plane above the sample, and the tunneling current varies depending on topography and local surface properties. The tunneling current measured at each location constitutes the image. The sample surface, however, must be relatively smooth in order for the system to acquire useful information.
In constant current mode, a feedback loop is used to adjust the height of the tip in order to hold the tunneling current at a setpoint value. The scanner height measured at each location is then used to map the surface topography. Because the feedback response requires time, constant
Keysight 5500 SPM User’s Guide 5-5
Introduction to the Keysight 5500 5
current mode is typically slower than constant height mode. However, greater variations in height can be accommodated.
Figure 1-4 Constant Height mode STM (above) is faster but is limited
to smooth surfaces; Constant Current mode (below) is capable of mapping larger variation in Z
For electron tunneling to occur, both the sample and tip must be conductive or semi-conductive. Therefore, STM cannot be used on insulating materials. This is one of the significant limitations of STM, which led to the development of other SPM methods described below.

Atomic Force Microscopy (AFM)

Atomic Force Microscopy (AFM) can resolve features as small as an atomic lattice, for either conductive or non-conductive samples. AFM provides high-resolution and three-dimensional information, with little sample preparation. The technique makes it possible to image in-situ, in fluid, under controlled temperature and in other controlled environments. The potential of AFM extends to applications in life science, materials science, electrochemistry, polymer science, biophysics, nanotechnology, and biotechnology.
In AFM, as shown in Figure 1-5, a sharp tip cantilever (the “probe”) is brought into contact with the sample surface. The tip interacts with the surface, causing the cantilever to bend. A laser spot is reflected from the cantilever onto a position-sensitive photodiode detector. As the cantilever bends, the position of the laser spot changes. The resulting signal from the detector is the Deflection, in volts. The
at the free end of a
Keysight 5500 SPM User’s Guide 5-6
Introduction to the Keysight 5500 5
difference between the Deflection value and the user-specified Set Point is called the “error signal.”
Figure 1-5 Basic AFM principles
Figure 1-6 shows the force interaction as the tip approaches the sample.
At the right side of the curve the tip and sample are separated by large distance. As they approach, tip and sample atoms first weakly attract each other. This zone of interaction is known as the “non-contact” regime. Closer still, in the “intermittent contact” regime, the repulsive van der Waals force predominates. When the distance between tip and sample is just a few angstroms, the forces balance, and the net force drops to zero. When the total force becomes positive (repulsive), the atoms are in the “contact” regime.The various AFM techniques
Keysight 5500 SPM User’s Guide 5-7
Introduction to the Keysight 5500 5
described below, can be generally described by their function within these three domains.
Figure 1-6 Zones of interaction as the tip approaches the sample
The tip-sample interaction is complicated by additional forces, including strong capillary and adhesive forces that attract the tip and sample. The capillary force arises when water, often present when imaging in the ambient environment, wicks around the tip, holding the tip in contact with the surface. As long as the tip is in contact with the sample, the capillary force should be constant because the fluid between the tip and the sample is virtually incompressible. The total force that the tip exerts on the sample is the sum of the capillary, adhesive and van der Waals forces.
The van der Waals force counters almost any force that attempts to push the atoms closer together. When the cantilever pushes the tip against the sample, the cantilever bends rather than forcing the tip closer to the sample atoms. The deflection, therefore, can be used as a reliable indicator of surface topography.

Contact Mode AFM

In Contact Mode AFM, the AFM tip is attached to the end of a cantilever with a low spring constant (typically 0.001 - 5 nN/nm). The
Keysight 5500 SPM User’s Guide 5-8
Introduction to the Keysight 5500 5
tip makes gentle contact with the sample, exerting from ~0.1-1000 nN force on the sample.
AFM can be conducted in either constant height or constant force
modes.
In constant height mode, the height of the scanner is fixed as it scans. For small cantilever deflections (<500 nm) on hard surfaces, the error signal (in volts) is used to generate an image that is sensitive to small changes in topography, though actual topographic information is not obtained. Constant height mode is often used for generating atomic-resolution images of atomically flat surfaces, where the cantilever deflections, and thus variations in applied force, are small. It is also used for recording real-time images of changing surfaces, where high scan speed is essential.
In constant force mode, the error signal is used as the input to a feedback circuit which, after amplification, controls the z-height piezo actuator. The feedback circuit responds to the surface topography, keeping the cantilever deflection constant, and thus holding the total force applied to the sample constant as well. The output of the feedback circuit is used to generate the topography image.
Constant force mode is more typically used than constant height mode as it enables imaging of greater surface height variability. The speed of scanning is limited by the response time of the feedback circuit, however. The resolution is lower than constant height mode as well, due to inherent noise in the piezo feedback circuit itself.
NOTE
The signal path is actually the same for constant height and constant force mode. In both cases, the error signal from the detector is the input
to the feedback loop, and the output of the feedback loop is the actual deflection signal. In constant height mode, however, the gain for the feedback loop is set to zero, effectively turning it off. Thus, the error signal is passed through and read directly.

Intermittent Contact AFM

Intermittent Contact Mode AFM is typically referred to as AC Mode due to the alternating contact of the tip to the surface. In AC Mode, the cantilever is driven to oscillate, typically in sinusoidal motion, at or near one of its resonance frequencies. When the cantilever and sample are close during each oscillation cycle, the tip moves through an interaction potential that includes long-range attractive and short term repulsive
Keysight 5500 SPM User’s Guide 5-9
Introduction to the Keysight 5500 5
components.The complex tip-sample forces cause changes in the amplitude, phase and resonance frequency of the oscillating cantilever. Thus, topography, amplitude and phase can be collected simultaneously. The phase and amplitude images may highlight physical properties that are not readily discernible in the topographic map. For example, fine morphological features are, in general, better distinguished in amplitude and phase images.
The force of the oscillating tip is directed almost entirely in the Z axis; thus, very little lateral force is developed and tip/sample degradation is minimized. This benefit also makes it possible to obtain clear images of soft samples.
A feedback system is employed to maintain the oscillation amplitude at a setpoint value. The difference between the amplitude and set point, called the “error signal,” is used as the input to the feedback system. The output of the feedback loop is amplified and drives the Z-actuator. The map of this output signal is called the “Amplitude Image,” which is typically plotted side-by-side with the topography image. The topography image is the voltage applied to the piezo required to keep the oscillation amplitude constant, multiplied by the sensitivity of the piezo in nanometers/volt.
AC Mode can operate in either the intermittent contact (net repulsive) regime or the non-contact (net attractive) regime. During intermittent contact, the tip is brought close to the sample so that it lightly contacts the surface at the bottom of its travel, causing the oscillation amplitude to drop.
The tip is usually driven by a sinusoidal force, with the drive frequency typically at or near one of the cantilever’s resonance frequencies (eigenfrequencies), and most often at the fundamental frequency. Absent any tip-sample interactions, the cantilever oscillations are also sinusoidal if the drive amplitude is small enough to keep the cantilever motion small compared with the cantilever thickness.
Two methods are used to drive the cantilever oscillation: by indirect, acoustic vibration (Acoustic Mode), or by direct vibration in a magnetic field (MAC Mode).
Acoustic AC (AAC) AFM
In Acoustic AC (AAC) Mode AFM, shown schematically in Figure 1-7 on page 11, a piezoelectric transducer (1) shakes the cantilever holder (2) at or near its resonant frequency, typically 100 to 400 kHz. Interaction between the probe (3) and the sample reduces the oscillation
Keysight 5500 SPM User’s Guide 5-10
Introduction to the Keysight 5500 5
amplitude—this reduction (c.) is used as a feedback signal to maintain constant amplitude of the cantilever motion.
NOTE
Acoustic AC Mode is an option for the 5500 SPM and requires the additional Mac Mode or MAC III controller.
Figure 1-7 Acoustic AC mode (AAC) (a.) Overview (b.) The free
amplitude before contact and (c.) The amount of amplitude dampening from the tip-sample interaction
Magnetic AC (MAC) Mode
In Magnetic AC (MAC) Mode AFM, the back side of the cantilever is coated with magnetic material. A solenoid applies an AC magnetic field which is used to oscillate the cantilever. (Figure 1-8). MAC Mode is typically cleaner and gentler than Acoustic AC Mode and is free from spurious background signals that are somewhat common when AAC Mode. The benefits are particularly pronounced when imaging in liquid.
NOTE
Keysight 5500 SPM User’s Guide 5-11
MAC Mode is an option for the 5500 SPM and requires the additional MAC Mode or MAC III controller.
Introduction to the Keysight 5500 5
Figure 1-8 Magnetic AC mode (MAC mode)
Top MAC Mode
In standard MAC Mode the magnetic coil is located in the sample plate, below the sample. A variant of MAC Mode, known as Top MAC, places the drive coil above the cantilever. This enables MAC Mode to be used with or without a sample plate, for large samples, or for samples which tend to dissipate the magnetic field enough to affect the resolution of regular MAC Mode.
NOTE
Top MAC Mode is an option for the 5500 SPM and requires the additional MAC Mode or MAC III controller, as well as a Top MAC nose assembly.

Current Sensing Mode (CSAFM)

Current Sensing AFM (CSAFM) uses standard AFM Contact Mode, including a special nose cone containing a pre-amp along with an ultra-sharp AFM cantilever coated with a conducting film, to probe the conductivity and topography of the sample. By applying a voltage bias
Keysight 5500 SPM User’s Guide 5-12
Introduction to the Keysight 5500 5
between the conducting cantilever and sample, a current is generated which is used to construct a conductivity image.
CSAFM is compatible with measurements in air, under controlled environments, and measurements with temperature control. The technique is useful in molecular recognition studies and can be used to spatially resolve electronic and ionic processes across cell membranes. It has proven useful in joint I/V spectroscopy and contact force experiments as well as contact potential studies.
NOTE
CAUTION
CSAFM requires a 10 ° nose cone with a pre-amp and ultra-sharp, conducting cantilevers.
The CSAFM nose assembly cannot be used for imaging in liquid.

Force Modulation Microscopy (FMM)

In Force Modulation Mode (FMM), the AFM tip is scanned in contact with the sample. As in Contact Mode, a feedback loop is used to maintain a constant cantilever deflection, and an additional, periodic vertical oscillation applied to the tip. The amplitude and phase of cantilever modulation resulting from the cantilever’s interaction with the sample varies according to the elastic properties of the sample (Figure 1-9), with particular sensitivity to elasticity and viscoelasticity.
Figure 1-9 Cantilever response to the applied modulation changes with
surface elasticity, as well as other characteristics
Keysight 5500 SPM User’s Guide 5-13
Introduction to the Keysight 5500 5
NOTE
FMM requires MAC Mode or MAC III.

Lateral Force Microscopy (LFM)

Lateral Force Microscopy (LFM) is a derivative of Contact AFM with the scan direction perpendicular to the long axis of the cantilever. In LFM, the tip is constantly in contact with the sample surface. In addition to its vertical deflection, the cantilever also twists in the scan direction. As a result, along with the near-vertical deflection signal which is usually present during Contact Mode AFM, the detector can also collect a sizeable lateral defection (Friction) signal from the cantilever‘s twisting motion. The strength of the lateral deflection signal is related to the friction force between the sample surface and the tip; thus, LFM is sometimes called Friction Force Microscopy.
The LFM signal is highly affected by topographic variations: the rougher the sample surface, the more the topography will affect the friction signal. To differentiate between friction and topography, two images are typically captured side-by-side. One is constructed from the detector signal during the trace (left-to-right tip motion) of each line in the raster scan, and the other is mapped during retrace (right-to-left tip motion). Then one of the two images is inverted and subtracted from the other. This reduces the topographic artifacts in the LFM signal, leaving an image of primarily frictional forces.

Dynamic Lateral Force Microscopy (DLFM)

In Dynamic Lateral Force Microscopy (DLFM), the tip is in contact with the sample, and the cantilever is oscillated parallel to the sample surface (as opposed to perpendicular oscillation in AC Mode). The topography is determined by cantilever deflection, as in contact mode. However, the lateral oscillation is also monitored, such that the amplitude and phase can be imaged, as in standard AAC Mode. DLFM is used in polymer studies as it is very sensitive to changes in surface properties such as friction and adhesion.
NOTE
Keysight 5500 SPM User’s Guide 5-14
DLFM requires MAC Mode or MAC III.

Magnetic Force Microscopy (MFM)

Magnetic Force Microscopy (MFM) probes the force between a ferromagnetic tip and a ferromagnetic or paramagnetic sample to image domain structures. The system detects changes in the phase of the cantilever due to interatomic magnetic force that persists for greater tip-sample separation than the van der Waals force.
A standard topography image can be collected for the same scanned area, using AAC in Intermittent Contact mode. The two images can then be displayed side-by-side to highlight any correlation between the magnetic structure and topography.
Introduction to the Keysight 5500 5
NOTE
MFM requires AAC, MAC, AAC III or MAC III controller.

Electrostatic Force Microscopy (EFM)

Electrostatic Force Microscopy (EFM) is a qualitative method for examining changes in the intrinsic, or applied, electrostatic field of a sample surface. A voltage bias is applied between the tip and the sample, allowing local static charge domains and charge carrier density to be measured.
The system detects changes in the phase response of the cantilever which are induced by the interaction of the conducting tip and the electrostatic field of the sample surface. EFM images are usually obtained by monitoring the phase change of the cantilever oscillation at the applied frequency.
A standard topography image can be collected for the same scanned area, using AAC (or MAC) in Intermittent Contact Mode. The two images can then be displayed side-by-side to highlight any correlation between the electrostatic response and topography.
NOTE
EFM is an option for the 5500 SPM and requires the additional MAC III controller.

Kelvin Force Microscopy (KFM)

Kelvin Force Microscopy (KFM) is similar to EFM, but with the addition of a feedback loop to maintain a DC tip bias that counteracts
Keysight 5500 SPM User’s Guide 5-15
Introduction to the Keysight 5500 5
the surface electrostatic force. The output from this feedback loop provides a quantitative analysis of changes in the applied or intrinsic electrostatic field of the sample.
As in EFM mode, KFM uses a conductive tip and either standard AAC or MAC Modes.
NOTE
KFM is an option for the 5500 SPM and requires the additional MAC III controller.
Keysight 5500 SPM User’s Guide 5-16
Keysight 5500 SPM User’s Guide
2 Keysight 5500 SPM Components
Microscope 2-3
Probes 2-4
Nose Assembly 2-5
One-Piece Nose Assemblies 2-5
Two-Piece Nose Assemblies 2-6
Scanner 2-7
Detector 2-9
Sample Plates 2-10
Video System 2-12
Head Electronics Box (HEB) 2-13
AFM Controller 2-15
Breakout Box 2-16
Vibration Isolation Chamber 2-21
Software 2-22
System Options 2-23
MAC Mode 2-23
MAC III Mode 2-23
Liquid Cell 2-23
Temperature Control 2-24
Thermal K 2-24
Environmental Chamber 2-24
Glove Box 2-25
Electrochemistry 2-26
PicoTREC 2-26
PicoLITH 2-26
Keysight 5500 SPM Components 5
The major components for the Keysight 5500 SPM are shown in
Figure 2-1.
Figure 2-1 Components of the Keysight 5500 SPM
Keysight 5500 SPM User’s Guide 5-2

Microscope

Keysight 5500 SPM Components 5
The microscope (Figure 2-2) includes the hinged support stand, coarse z-axis motors, manual X/Y positioning micrometers, magnetic supports for the sample plates, and interconnections for all electronics. The support stand is hinged to allow easy access to the sample plate area.
Figure 2-2 5500 SPM with support stand
Keysight 5500 SPM User’s Guide 5-3

Probes

Keysight 5500 SPM Components 5
The SPM techniques described in Chapter 1, “Introduction to the Keysight 5500,” are performed using either a wire tip (for STM) or, for AFM imaging, a tip at the free end of a cantilever (a “probe”). STM tips are made by cutting or electrochemical etching Platinum-Iridium or tungsten wire. AFM cantilevers are fabricated from silicon or silicon nitride with an integrated sharp tip at the end.
The selection of probe and tip geometry, cantilever spring constant, and cantilever resonance frequencies will vary depending on application, type of sample surface, imaging environment, and type of image being generated. Tip geometry may be tetrahedral, pyramidal or conical. Tip sharpness, defined by radius of curvature and sidewall angles, greatly affects the resolution available with the probe.
Common cantilever shapes are triangular (V-shaped) and rectangular (beam-shaped). Cantilever spring constants vary from a fraction of N/m (soft) to tens or hundreds of N/m (stiff). Cantilevers for any type of AC Mode imaging (AAC, MAC, etc.) will have resonance frequencies ranging from tens to hundreds of kilohertz.
Cantilevers for particular imaging modes may be coated with reflecting, conductive or magnetic materials.
Keysight 5500 SPM User’s Guide 5-4

Nose Assembly

One-Piece Nose Assemblies

Keysight 5500 SPM Components 5
The nose assembly retains the cantilever and enables its motion. A spring clip on the nose assembly secures the probe in place. The nose assembly is held securely in the scanner by an O-ring.
The most widely used nose assemblies consist of a single unit which is installed in the scanner (Figure 2-3). One-piece nose assemblies are available for different modes and may include additional electronics and/or components. For example, the Top MAC nose assembly includes a coil that provides an oscillating magnetic field.
Additionally, nose assemblies are designed to hold the probe at either nine degrees or eight degrees from horizontal. Nine-degree nose assemblies are used for general purpose imaging. Nine and eight-degree nose assemblies may be used for imaging in liquid. A Top MAC nose assembly is available with an eight-degree nose assembly option.
Figure 2-3 One-piece nose assemblies. Clockwise from upper left: Top
MAC, CSAFM, Contact Mode, AAC, STM
CAUTION
Keysight 5500 SPM User’s Guide 5-5
The CSAFM nose assembly cannot be used for imaging in liquid.

Two-Piece Nose Assemblies

The all-metal, two-piece nose assembly (Figure 2-4) was designed to simplify the process of inserting a cantilever. It also helps prevent damage to the scanner during installation of the nose assembly. Currently the two-piece nose assembly is only available for AC Mode/Contact Mode imaging.
Keysight 5500 SPM Components 5
CAUTION
The two-piece nose assembly cannot be used for imaging in liquid.
Figure 2-4 Two-piece nose cone with removal tool and assembly fixture
Keysight 5500 SPM User’s Guide 5-6

Scanner

Keysight 5500 SPM Components 5
The Keysight 5500 is a tip-scanning system, in which the cantilever sits on a scanner and is moved in raster fashion across the stationary sample. The scanner (Figure 2-5) includes one or more elements made from piezoceramic material. When an electric field is applied to the piezo elements, they elongate or contract. The motion of the tip in the Z axis, and raster scanning in the X and Y axes, are all achieved by applying high voltages to the scanner’s piezo element(s).
Keysight’s multi-purpose scanner modules contain the piezo elements, the socket for the nose assembly, mounting for the detector, and interconnections. The scanners are considered “multi-purpose” because nose assemblies can be switched in and out of the scanner for different imaging modes or environments.
Figure 2-5 A-type Scanner Module
Keysight SPMs use two types of scanners: A and B. A-type scanners are most typically used with the Keysight 5500 SPM. B-type scanners are designed for Keysight SPM models that have a slightly different video system and a differently shaped scanner holder. The B-type scanners are equipped with a replaceable lens assembly so they can be used in the 5500 SPM.
A-type scanners that use a 980 nm IR wavelength laser diode are also available for those using inverted microscopes or who have
Keysight 5500 SPM User’s Guide 5-7
Keysight 5500 SPM Components 5
experimental needs for an IR light source rather than the standard source. Additional details for aligning such lasers are in Chapter 4.
There are four A-type scanners:
The small multi-purpose scanner includes four piezo plates (two each for X and Y motion) and a small piezo tube (for Z motion). The scanner provides scans up to 10 microns square. It is capable of atomic-level resolution imaging.
The large multi-purpose scanner includes four piezo elements for X and Y and provides scans up to 90 microns square. There is a larger, two-element staked piezo tube to provide a larger Z range of motion. The scanner provides high resolution and speed for general use applications.
The large multi-purpose scanner is also available with closed-loop positioning, in which ultra-precise positioning sensors measure displacement in the Z axis only, or all three (X, Y, Z) axes. Closed-loop scanning provides superior lateral positioning control and more accurate Z-position.
An STM-specific scanner is purpose-built for Scanning Tunneling Microscopy.
Keysight 5500 SPM User’s Guide 5-8

Detector

Keysight 5500 SPM Components 5
The photodiode detector, composed of four separate quadrants, receives the reflection of the laser spot off the back of the cantilever. The top and bottom halves of the detector monitor the cantilever deflection (the Deflection signal) for AFM imaging, while the two side halves measure the twist of the cantilever (the Friction signal) for lateral force imaging.
The detector is held in the scanner by magnets on the detector housing and in the scanner. Two thumbwheels enable alignment of the detector in both directions. There are also four DIP switches to enable or disable the gain of the signal from the detector. Enabling the gain is useful when using very non-reflective cantilevers to ensure a high enough Sum value is provided to the control system.
Figure 2-6 Detector assembly, top and bottom views
Keysight 5500 SPM User’s Guide 5-9

Sample Plates

Keysight 5500 SPM Components 5
The Keysight 5500 SPM is designed to allow scanning from above the sample. A variety of sample plates provide mounting options and micro-environments for imaging (Figure 2-7). The standard sample plate has a magnetic core that will securely hold samples mounted on magnetic backings. Other plates are available for measurement in liquid, temperature controlled imaging, for MAC and other applications. More information for imaging in liquid is in Chapter 12, “Liquid Cell.”
Figure 2-7 Three sample plates: MAC Mode with heating, liquid
imaging, and Petri dish.
Keysight 5500 SPM User’s Guide 5-10
Keysight 5500 SPM Components 5
The microscope stand is equipped with three magnetic posts from which a sample plate is mounted (Figure 2-8). Micrometers enable manual X/Y positioning with total travel of approximately ±5 mm.
Figure 2-8 Sample plate on microscope stand
Keysight 5500 SPM User’s Guide 5-11

Video System

Keysight 5500 SPM Components 5
The video system enables the location of regions of interest and align the laser on the probe tip. It includes a camera and optics on an adjustable stand (Figure 2-9) along with a separate illumination source (Figure 2-10). A USB cable connects the camera to the computer.
Figure 2-9 Video system
Figure 2-10 Video system light source
Keysight 5500 SPM User’s Guide 5-12

Head Electronics Box (HEB)

The Head Electronics Box (HEB) (Figure 2-11) reads the signals from the detector and can display the Sum signal (sum of all four quadrants) and the Deflection or Friction signals. Meter A (upper left) is the Sum signal reading and Meter B (upper right) shows Deflection or Friction (LFM) depending on the state of the switch directly below the meter.
The Open/Close switch is used to move the sample and tip further apart (Open) or closer together (Close).
Keysight 5500 SPM Components 5
Figure 2-11 Head Electronics Box
Keysight 5500 SPM User’s Guide 5-13
Keysight 5500 SPM Components 5
The HEB back panel connections are shown in Figure 2-12:
Figure 2-12 HEB back panel connectors and controls
Keysight 5500 SPM User’s Guide 5-14

AFM Controller

Keysight 5500 SPM Components 5
The AFM Controller (Figure 2-13) provides the high voltage to the piezoes and other control functions. Model N9605A is standard; Model N9610A provides optional closed-loop scanning control and extra data channels required for modes such as EFM and KFM.
Figure 2-13 AFM Controller (Model N9610A)
The Aux In BNC input is added on to the current signal, which is used in STM and CSAFM. The External Input BNC to the Z piezo is limited to ±10 V and the input is scaled up by the controller to ±215 V. The External Sync BNC is a signal that is sent each time a frame, line, or pixel is completed. This can be used in Custom Spectroscopy or
Keysight 5500 SPM User’s Guide 5-15

Breakout Box

Keysight 5500 SPM Components 5
Scripting. The Sync signal is not accessible with some previous versions of PicoView.
There is an optional breakout box (N9447A) to access various signals associated with the operation of the SPM. This box is shown in
Figure 2-14 and described in Table 1.
“Breakout Box Connections must be made properly and are shown in Figure 2-15 on page 20:
Figure 2-14 Breakout box for AFM controller
Keysight 5500 SPM User’s Guide 5-16
Keysight 5500 SPM Components 5
Table 1 List of signals and descriptions for AFM controller breakout
box
Signal Description
Error Signal This is the servo signal from the Head Electronics Box to the controller for
AFM/STM. This signal is the photo detector A-B signal with the force setpoint subtracted in AFM mode or the tunneling current in STM mode. If the breakout box is placed between the AC Mode controller and the AFM controller and AC Mode is On, then the signal will be the Amplitude signal from the AC Mode controller. This signal is the Amplitude of the cantilever minus the Amplitude setpoint. This input has a range of ±10 V.
Friction This is the LFM signal from the Electronics box to the controller for AFM. This
signal is the photo detector C-D signal. If the breakout box is placed between the AC Mode controller and the AFM controller and AC Mode is On, then the signal will be the Phase signal from the AC Mode controller. This input has a range of ±10 V.
VEC This is the potential signal from the PicoStat (Potentiostat) board if the
microscope has the electrochemistry option. If there is no electrochemistry option, this is a floating input into the AFM controller and can be used as an auxiliary data input channel. This input is displayed by the software with units of volts. This signal is not affected by the presence of the AC Mode controller in the system. This input has a range of ±10 V.
IEC This is the current signal from the PicoStat (Potentiostat) board if the microscope
has the electrochemistry option. If there is no electrochemistry option, this is a floating input into the AFM controller and can be used as an auxiliary data input channel. This input is displayed by the software with units of current. This signal is not affected by the presence of the AC Mode controller in the system. This input has a range of ±10 V.
Keysight 5500 SPM User’s Guide 5-17
Keysight 5500 SPM Components 5
Current This is the current signal from the STM or CSAFM pre-amp. This signal is not
affected by the AC Mode controller. If there is no pre-amp installed in the scanner, this signal is still driven by a buffer in the HEB so it is only available for use as an auxiliary input if the switch is placed in the IN position and the auxiliary signal would then be fed into the IN BNC only. This input channel is summed into the Aux In BNC on the front of the controller.
Ref Set This signal is an output from the controller and is the Reference Setpoint DAC
output. This signal has a range of ±10 V and is a 24 bit DAC output.
Bias This signal is an output from the controller. It is the Sample Bias DAC output or
Force Setpoint DAC output (Amplitude Setpoint in AC Mode). This signal has a range of ±10 V and is a 24 bit DAC output.
+5 V This is the 5 V power supply output from the controller.
+15 V This is the +15 V power supply output from the controller.
-15 V This is the -15 V power supply output from the controller.
Force Setpoint This signal is an output from the controller and is the Force Setpoint DAC output.
This signal has a range of ±10 V and is a 24 bit DAC output. This signal is the Amplitude Setpoint in AC Mode.
Tip Bias This signal is an output from the controller and is the Tip Bias DAC output. This
signal has a range of ±10 V and is a 24 bit DAC output.
X HV This is the +X High Voltage signal used to drive the piezo for scanning in the X
direction. This output is ±215 V and is current limited to 70 mA.
-X HV This is the -X High Voltage signal used to drive the piezo for scanning in the X direction. This output is ±215 V and is current limited to 70 mA.
Z HV This is the Z High Voltage signal used to drive the piezo for scanning in the Z
direction. This output is ±215 V and is current limited to 70 mA.
Keysight 5500 SPM User’s Guide 5-18
Keysight 5500 SPM Components 5
Y HV This is the +Y High Voltage signal used to drive the piezo for scanning in the Y
direction. This output is ±215 V and is current limited to 70 mA.
-Y HV This is the -Y High Voltage signal used to drive the piezo for scanning in the Y direction. This output is ±215 V and is current limited to 70 mA.
The signals are all passed through the box when the switches are in the OUT position and the signals are all accessible from the IN or OUT BNC's. If a switch is in the IN position, the signal on the OUT BNC is the signal from the source and the connection between the IN BNC and the OUT BNC is broken along with the connection through the box. This allows for some external modification of the signal. The modified signal would then be fed into the system using the IN BNC. The source of the signal is from the Controller for all outputs and from either the microscope electronics or the AC Mode controller for all of the inputs.
Breakout Box Connections
It is important to connect the breakout box correctly as shown in
Figure 2-15.
To access the low voltage ports of the breakout box, connect one end of a low voltage DB-44 cable to the front panel of the AFM controller. Connect the other end of the cable to the N9744A connector labeled Controller. Connect one end of a second DB-44 cable (included with the breakout box) to the N9744A connector labeled Microscope. Connect the other end of the second cable to the connector labeled Controller on the first microscope component in the series configuration. For example, this could be a MAC or AC mode controller, or possibly the Head Electronics Box.
To access the high voltage ports of the breakout box, connect one end of a high voltage DB-9 cable to the front panel of the AFM controller. Connect the other end of the cable to the appropriate 9-pin connector located on the side of the N9744A breakout box. Connect one end of a second DB-9 cable (included with the breakout box) to the second 9-pin
Keysight 5500 SPM User’s Guide 5-19
Keysight 5500 SPM Components 5
connector on the side of the breakout box. Connect the other end of the second cable to the 9-pin connector on the 5500 microscope base.
Figure 2-15 Wiring diagram for breakout box
Keysight 5500 SPM User’s Guide 5-20

Vibration Isolation Chamber

The isolation chamber (Figure 2-16) isolates the 5500 SPM from vibration, air turbulence and acoustic noise which would adversely affect imaging. It also, to an extent, helps control temperature variability.
The chamber provides acoustic isolation for the instrument with multiple layers of sound-damping materials. The vibration isolation system damps incoming vibrations through the use of bungees and suspended mass. The system is designed to be accessed from either the left or right side as both the door and cable ports are reversible.
The enclosure is considered a “mandatory option,” as the improvements it provides for imaging are essential for all but the most stringently controlled environments.
The details for initial assembly are in “Assembling the Vibration
Isolation Chamber" on page 7 of Chapter 3, “Setting Up the Keysight
5500 SPM.
Keysight 5500 SPM Components 5
Figure 2-16 Vibration isolation chamber
Keysight 5500 SPM User’s Guide 5-21

Software

Keysight 5500 SPM Components 5
The Keysight 5500 SPM includes PicoView, a powerful software package for controlling all aspects of alignment, calibration, imaging and more. Integrated in the software are windows for controlling digital cameras and displaying video or still-image output. The post-acquisition PicoImage software for image analysis and data manipulation offline is an available option and highly recommended.
To accomplish the steps in the following chapters some familiarity with PicoView is required. The software steps will be documented briefly in this manual. For more information on the software, please review the online PicoView User Guide. The online User Guide is intended to explain the use of the PicoView screens and controls. Click the keyboard F1 button from most PicoView windows to access this Help file. In most cases, clicking F1 will directly open the pertinent topic. There is also a "breadcrumb" trail along the top of the PicoView User Guide window to aid in navigation.
NOTE
The procedures and descriptions in this User Guide are for PicoView
1.18x and higher.
Keysight 5500 SPM User’s Guide 5-22

System Options

MAC Mode

Keysight 5500 SPM Components 5
Many options are available for the Keysight 5500 SPM. As discussed above, probes, nose assemblies and sample plates are available for particular applications. Scanner options include large and small scan ranges, closed-loop scanning, and a dedicated STM scanner. Other options include:
The MAC Mode options includes the hardware required for MAC mode, which greatly improves imaging in fluid. The options include the MAC controller (Figure 2-17), the Top MAC nose assembly, AAC nose assembly and/or MAC sample plate.

MAC III Mode

Liquid Cell

Figure 2-17 Mac Mode Module
MAC III Mode provides the benefits of regular MAC mode, provides three lock-in amplifiers for flexibility, enables EFM and KFM imaging, and provides “Q control” for more precise control of cantilever oscillation.
A standard liquid cell is available for AFM and STM. A flow-through liquid cell is also available with connections for tubing. Additional
Keysight 5500 SPM User’s Guide 5-23
Keysight 5500 SPM Components 5
information is discussed in Chapter 12, “Liquid Cell.” A schematic drawing of the liquid cell is shown below in Figure 2-18.
Figure 2-18 Schematic drawing showing scanner with nose assembly,
liquid cell, O-ring, and sample plate

Temperature Control

This option includes low and/or high temperature sample plates, a temperature controller and related hardware for maintaining sample temperature during imaging. Additional information is provided in
Chapter 13.

Thermal K

Thermal K provides a method for accurately determining the force constant of the cantilever for highly accurate force measurements. By measuring the thermal oscillation of the cantilever with no drive signal applied, the cantilever force constant can be determined. The option includes a separate acquisition card that is installed in an empty slot in the system computer. Additional information is provided in Chapter 14.

Environmental Chamber

The environmental chamber (Figure 2-19) allows imaging in controlled atmospheres. Ports and fittings enable gases, liquids and probes to be
Keysight 5500 SPM User’s Guide 5-24
Keysight 5500 SPM Components 5
introduced to the chamber. Additional information is provided in
Chapter 15.
Figure 2-19 Environmental chamber

Glove Box

The 5500 SPM body can be place directly on top of this small glove box, shown in Figure 2-20, offering greater environmental control. Since the piezo and electronic parts are totally isolated from the imaging environment, it is possible to perform experiments under very reactive conditions without damaging the system or the sample. Additional information is provided in Chapter 15.
Figure 2-20 Glove box
Keysight 5500 SPM User’s Guide 5-25

Electrochemistry

The electrochemical SPM option includes a low-noise potentiostat/galvanostat for in-situ EC-STM and EC-AFM. When combined with temperature control, it is possible to obtain valuable information about electrochemical processes that would otherwise be inaccessible. Additional environmental controls allow imaging with no dissolved oxygen in either aqueous or non-aqueous solutions. Additional information is available in Chapter 16.

PicoTREC

The PicoTREC molecular recognition tool kit (Figure 2-21) provides a faster method than force-volume spectroscopy for distinguishing molecular binding events. You can also use PicoTREC to explore dynamic properties of biological systems by imaging patterns of molecular binding and adhesion on surfaces.
Keysight 5500 SPM Components 5
Figure 2-21 PicoTREC controller

PicoLITH

PicoLITH is an optional package for nanoscale positioning and manipulation, and nanolithography. The PicoLITH option includes its own documentation and is not covered in this manual.
Keysight 5500 SPM User’s Guide 5-26
Keysight 5500 SPM User’s Guide
3 Setting Up the Keysight 5500 SPM
Component and Facility Dimensions 3-1
Facility Requirements 3-4
Utilities 3-5
Noise and Facility Specifications 3-5
Acoustic Noise 3-5
Temperature and Humidity Variation 3-6
Assembling the Vibration Isolation Chamber 3-7
Switching Door Orientation 3-9
Managing Cables 3-10
Connecting the Components 3-12
Guidelines for Moving the System 3-13
Care and Handling of the Probes and Scanner 3-14
Probes 3-14
Nose Assembly 3-14
Scanner 3-14
The Keysight 5500 SPM is typically installed by trained Keysight technical staff. This chapter includes information on the facilities requirements and preparation needed prior to installation. It also offers suggestions on how to handle and re-connect the components should you ever need to relocate the system after installation.

Component and Facility Dimensions

The Keysight 5500 SPM system includes, at minimum, the microscope, computer, Head Electronics Box, and AFM controller. Options purchased with your system may include additional hardware.
Here are the approximate dimensions of some of the components of the Keysight 5500 SPM:
5500 Microscope: 210 mm W x 190 mm H x 160 mm D (8 in W x 8 in H x 6 in D)
Setting Up the Keysight 5500 SPM 5
Head Electronics Box: 203 mm W x 102 mm H x 203 mm D
(8 in W x 4 in H x 8 in D)
Computer: 203 mm W × 432 mm H × 457 mm D (8 in W x 17 in H x 18 in D)
AFM Controller: 178 mm W x 483 mm H x 406 mm D (7 in W x 19 in H x 16 in D)
MAC Mode Controller: 254 mm W x 127 mm H x 254 mm D (10 in W x 5 in H x 10 in D)
MAC III Controller: 254 mm W x 127 mm H x 254 mm D
(10 in W x 5 in H x 10 in D)
Vibration Isolation Chamber: 521 mm W × 915 mm H × 603 mm D (20.5 in W x 36 in H x 23.75 in D)
The most common system configuration includes the 5500 SPM within a vibration isolation chamber, with the controls on a separate table from the rest of the components, as shown in Figure 3-1 on page 3 (top and front views). Keeping the chamber and microscope on a separate table helps to minimize acoustic coupling from the control station. The isolation chamber and microscope together weight approximately 250
Keysight 5500 SPM User’s Guide 5-2
Setting Up the Keysight 5500 SPM 5
pounds; therefore, a solid table that can easily accommodate 300 pounds is required.
Figure 3-1 Top and front views of Keysight 5500 SPM suspended
inside the isolation chamber, on the left. The control station is on the right.
Keysight 5500 SPM User’s Guide 5-3

Facility Requirements

Setting Up the Keysight 5500 SPM 5
Following these guidelines for preparing the Keysight 5500 SPM facility will ensure a smooth installation, will make using the system more convenient and will improve system performance for the life of the SPM:
Minimize the acoustic noise level from all possible sources, such as
paging speakers, telephone ringer, air conditioner, especially during data acquisition.
Install the equipment as far away as possible from facility equipment
such as air handlers and pumps.
Keep far away from any source of strong electromagnetic fields.
Reduce the exposure of the SPM to air flow or dramatic temperature
changes. Minimal temperature variation is desirable to minimize thermal drift during measurements and to minimize settling time..
Use dedicated power outlets with surge protection (strongly
recommended).
Include a set of organized shelves and drawers for system
components.
If gold substrates are frequently used, a hydrogen flame-annealing
station is recommended.
Appropriate water sources should be available for temperature
control and biological experiments.
Keysight 5500 SPM User’s Guide 5-4

Utilities

The following table summarizes the utility requirements for the Keysight 5500 SPM.
Table 2 Keysight 5500 SPM utility requirements
Configuration Keysight 5500 SPM
Electrical 1600 W; single phase; 100-120 V or
Surge protection Strongly recommended; minimum 7 outlets
Air for isolation chamber Not required
Internet connection Recommended

Noise and Facility Specifications

Setting Up the Keysight 5500 SPM 5
220-240 VAC; 5 A; 50-60 Hz
Acoustic Noise
The semiconductor manufacturing industry has developed a standardized set of one-third octave band velocity spectra, called vibration criterion curves (Figure 3-2), to define acceptable environmental noise. For operation of the Keysight 5500 SPM, facility acoustic noise should be less than 75 dBc. In order to achieve the
Keysight 5500 SPM User’s Guide 5-5
Setting Up the Keysight 5500 SPM 5
specified noise level, the vibration isolation chamber installation surface must meet Criterion D.
Figure 3-2 Vibration criterion curves and ISO guidelines
Temperature and Humidity Variation
Changes in temperature and humidity will affect both resolution and repeatability of imaging. Temperature variation should be limited to ±2 degrees Fahrenheit. Humidity variation should not exceed ±20 % RH. Locating the instrument away from vents and air handlers will help meet this goal.
Keysight 5500 SPM User’s Guide 5-6

Assembling the Vibration Isolation Chamber

Unpack the vibration isolation chamber from the box and remove the packing bag. Save the packing materials for subsequent use.
Setting Up the Keysight 5500 SPM 5
CAUTION
The chamber is very heavy. Using the lifting handles (see Figure 3-3) on the sides of the chamber, two people should move the enclosure to a rigid desk or table. Do not use the chamber door to support any weight.
Figure 3-3 Vibration isolation chamber with labeled components
Keysight 5500 SPM User’s Guide 5-7
Setting Up the Keysight 5500 SPM 5
Remove the accessory bag from the chamber. This bag contains four adjustable bungee mounts, four bungee cords and a 2.5 mm hex wrench for assembly.
Place the instrument platform, included in a smaller box, on the floor of the chamber.
After unpacking is complete, proceed with assembling the vibration isolation chamber:
1 Attach bungee mounts to screws mounted on enclosure ceiling by
inserting screw into open end of bungee mount and turning the mounts' pin to the right.
2 Attach bungee cords to eyebolts on adjustable bungee mount using
bungee cord hook.
3 Lift back side of instrument platform and tilt up towards bungee
hooks. Attach the back two bungee cord hooks, one at a time, to recessed attachment points as shown in Figure 3-4 on page 8:
a Insert hook into recess hole parallel to the attachment bar.
b Turn hook ninety degrees.
Figure 3-4 Attaching bungee hook to instrument platform
4 Repeat step 3 for the front side of the instrument platform.
CAUTION
Do not adjust bungee mounts with SPM instrument on platform.
5 Adjust bungee mounts to ensure that the instrument platform is
resting level. Turn the standoff using the center pin as shown in
Keysight 5500 SPM User’s Guide 5-8
Figure 3-5. Turning the mount to the right will raise the bungee
mount; turning the mount to the left will lower the bungee mount.
Figure 3-5 Adjusting bungee mounts to level the instrument platform

Switching Door Orientation

The door may be hinged on either the left or right side as needed.
Setting Up the Keysight 5500 SPM 5
1 Using the 2.5 mm hex wrench from the accessory bag, detach the 12
fill screws (4 of 12 shown in Figure 3-6) from the door and chamber body. Keep the screws for reattachment.
Figure 3-6 Fill screws on door and chamber body
Keysight 5500 SPM User’s Guide 5-9
Setting Up the Keysight 5500 SPM 5
2 Use wrench to detach latch attachment screws (Figure 3-7). Note
orientation of latch parts. Keep screws and latch parts for subsequent reattachment.
Figure 3-7 Latch with latch attachment screws
3 Support weight of enclosure door from below. Use wrench to detach
screws which attach the hinge to the enclosure body. Remove door.
4 Use wrench to detach hinges from door. Attach hinges to enclosure
5 Put enclosure door back into place and support weight from below.
6 Use wrench to reattach door latch and to reattach fill screws.

Managing Cables

There are cable ports on both sides of the enclosure. Before installing instrument, select which side of the enclosure the instrument cables should be ported through. If necessary, remove port cover and place in port which will not be used.
1 Remove damping foam from cable port to be used. Feed instrument
2 Outside of enclosure, compress two pieces of damping foam around
3 Adjust as necessary to ensure a good seal around the cables, as
body on opposite side.
Use wrench to attach hinge to door.
cables out through port.
cables. Push foam and cables into cable port.
shown in Figure 3-8 on page 11.
Keysight 5500 SPM User’s Guide 5-10
Setting Up the Keysight 5500 SPM 5
Figure 3-8 Cables installed with good seal.
Keysight 5500 SPM User’s Guide 5-11

Connecting the Components

The cabling for the standard 5500 SPM is shown in Figure 3-9. Other cabling configurations are included in Appendix A.
Setting Up the Keysight 5500 SPM 5
Figure 3-9 Cabling for basic 5500 SPM configuration
CAUTION
Keysight 5500 SPM User’s Guide 5-12
Always make sure that all cables are connected before turning on any of the components. Failure to do so can result in damage to equipment.

Guidelines for Moving the System

Should you ever need to relocate the system, here are important guidelines which must be followed to ensure safe operation:
The new location must meet all of the facility specifications
described above.
Turn off all components before disconnecting cables.
Disconnect all cables before moving any components.
Remove the scanner, detector and sample from the 5500 SPM before
moving the microscope base.
Remove the microscope base from the vibration isolation chamber
and transport both separately.
Follow the cabling diagrams exactly, being sure to connect all cables
before powering up the components.
Setting Up the Keysight 5500 SPM 5
Keysight 5500 SPM User’s Guide 5-13

Care and Handling of the Probes and Scanner

Probes

Always store probes at room temperature in their protective cases.
Handle probes gently with tweezers, following the procedures described earlier in this chapter.
If a probe is dropped it may very well be damaged. You can check whether the cantilever is intact by viewing it through a magnifier.
If you are using more than one type of probe, be sure to store them separately in well-marked cases to avoid confusion.

Nose Assembly

Put nose assemblies in the specially tailored scanner container box and store in a clean, dry location where they will not be subject to excessive humidity, temperature changes or contact.
Setting Up the Keysight 5500 SPM 5

Scanner

Dirt, grease or spots on the glass window of the nose assembly can interfere with the optical path of the laser. Regularly clean the window with cotton or a soft tissue (dry, wetted with water, or with ethanol).
The glass is glued to the nose cone with chemically resistive epoxy, so if the window breaks there is no easy way to replace it and the entire nose assembly will likely need to be replaced.
Only remove the nose assembly from the scanner using the Nose Assembly Removal Tool, with the scanner placed upright in its fixture.
Do NOT use the removal tool to install the nose assembly.
Two-Piece Nose Cone Cleaning
The two-piece nose cone is not to be used in liquid because it does not have a glass window to prevent liquid from getting to the scanner. After it is removed from a scanner, the two-piece nose cone may be cleaned with a low oxidizing organic solvent such as ethyl alcohol.
Between uses, remove the scanner from the microscope and store it on its fixture or in the storage case, in a location where it will not be subject
Keysight 5500 SPM User’s Guide 5-14
Setting Up the Keysight 5500 SPM 5
to excessive humidity, temperature changes or contact. Keysight recommends that scanners be stored in a desiccator.
The scanner contains very brittle and fragile piezoelectric ceramics. Applying excessive lateral force while exchanging nose assemblies, or dropping the scanner even a short distance onto a hard surface, will damage the scanner. If the nose assembly housing becomes loose or can be wiggled when in place, contact Keysight support for assistance.
Cracked or broken piezoelectrodes will result in abnormal imaging. Damage to the scanner such as those described above are NOT covered by the standard warranty.
Keysight 5500 SPM User’s Guide 5-15
Keysight 5500 SPM User’s Guide
4 Preparing for Imaging
Setting Up the Scanner Assembly 4-1
One-Piece Nose Assembly 4-2
Inserting the One-Piece Nose Assembly 4-2
Removing the One-Piece Nose Assembly 4-4
Inserting a Probe in the One-Piece Nose Assembly 4-6
Two-Piece Nose Assembly 4-9
Inserting the Body of the Two-Piece Nose Assembly 4-9
Removing the Body of the Two-Piece Nose Assembly 4-10
Inserting a Probe in the Two-Piece Nose Assembly 4-11
Inserting the Scanner and Connecting Cables 4-12
Aligning the Laser 4-14
Inserting and Aligning the Detector 4-22
Mounting the Sample 4-26
Using the Video System 4-29
The Keysight 5500 SPM is capable of imaging in many different modes. Several steps of the imaging process are similar or identical, however, for all modes. This chapter will cover the steps that are common to most imaging procedures.

Setting Up the Scanner Assembly

As mentioned earlier, the Keysight 5500 SPM is a tip-scanning system, in which the probe is raster-scanned across the stationary sample. When an electric field is applied to the scanner’s piezo elements, they elongate or contract, depending on the direction of the field.The Z-motion of the tip is achieved by elongation or contraction of the piezo element in the scanner. X/Y raster scanning is achieved by applying alternating
Preparing for Imaging 5
voltages to opposite piezo elements in the scanner so that one element elongates and the other contracts.
CAUTION
The thickness of the piezo elements determines how much they will expand or contract per applied unit voltage. They are necessarily thin to provide scanning resolution. If dropped, the piezo elements inside the scanner WILL break. Cracked or broken piezoelectrodes will result in abnormal imaging. Proper handling is essential to preserve the long expected life of your multi-purpose scanner.
The scanner mounting fixture supplied with your system is designed to keep the scanner and its components safe during handling (Figure 4-1). The main cutout safely holds the scanner, while the smaller cutout safely holds a nose assembly. A magnetic disk keeps additional tools close at hand.
Figure 4-1 Scanner mounting fixture with nose assembly and spring
key; scanner in mounting fixture
The next sections will describe how to safely handle the scanner components for long life and excellent imaging.

One-Piece Nose Assembly

Inserting the One-Piece Nose Assembly
The nose assembly is held in the scanner by an O-ring around its circumference. To insert a nose assembly, first place the scanner in the scanner mounting fixture (Figure 4-1). Place the nose assembly in the socket on top of the scanner, aligning its contact pins if applicable.
CAUTION
Ensure that the ends of the metal spring for retaining the cantilever are not caught in between the nose cone and the socket.
Keysight 5500 SPM User’s Guide 5-2
Preparing for Imaging 5
Applying even, steady, vertical pressure at the edges of the nose assembly, seat it into the socket, as shown in Figure 4-2.
Figure 4-2 Apply even, vertical pressure at the edges to insert the nose
assembly (as shown on left). O-ring should be fully inside socket (as shown on right).
CAUTION
Push evenly and straight down when inserting the nose assembly. Small off-axis forces will create LARGE torques about the anchor point for the piezoes, where most breakage occurs.
Do NOT push down on the top of the nose assembly as this will damage the spring clip and/or glass window.
Keysight 5500 SPM User’s Guide 5-3
Preparing for Imaging 5
Removing the One-Piece Nose Assembly
A removal tool is included with your system to limit damaging lateral forces on the scanner while removing the nose assembly. The following is the only acceptable procedure for removing the nose assembly:
Figure 4-3 Nose assembly removal tool
1 Place the scanner in the scanner mounting fixture.
2 Carefully slide the removal tool onto the nose assembly, ensuring
that the opening seats on both sides of the nose.
3 Position your thumb on the flat surface of the removal tool and your
fingers on BOTH sides of the extraction arm.
4 Gently pull up with your fingers while pushing down with your
thumb (Figure 4-4).
Keysight 5500 SPM User’s Guide 5-4
Preparing for Imaging 5
5 Once the nose assembly is clear of the scanner you can remove it
from the tool.
CAUTION
CAUTION
Figure 4-4 Using the nose assembly removal tool
Do not use the nose removal tool to insert a nose assembly. It is not designed for this purpose.
DO NOT use tweezers to remove the nose assembly (Figure 4-5).Tweezers can create a pivot point to lever the nose out of the scanner, placing large lateral forces on the piezo assembly. The nose assembly removal tool is the only acceptable method for extracting the nose from the scanner.
Keysight 5500 SPM User’s Guide 5-5
Preparing for Imaging 5
Figure 4-5 Do not use tweezers to remove a nose assembly. Doing so
can place damaging lateral forces on the scanner.
Inserting a Probe in the One-Piece Nose Assembly
Keysight nose assemblies are designed with a spring and guides to retain the probe in the proper position for imaging. A spring key (Figure 4-6) is included with the system to let you safely hold back the spring while inserting the probe. Figure 4-7 shows a properly positioned
probe.
Figure 4-6 Spring key
Keysight 5500 SPM User’s Guide 5-6
Preparing for Imaging 5
Figure 4-7 Probe properly situated on AFM nose assembly
CAUTION
AFM probe tips are extremely delicate and can break when dropped even a short distance. The following instructions include several helpful tips that will simplify the process of inserting a probe in the nose assembly:
1 Mount the nose assembly in the scanner.
2 Place the scanner into the scanner mounting fixture.
3 Grasp the tweezers in the orientation shown in Figure 4-8.
4 Gently grasp the probe from its sides, applying just enough pressure
to secure it in the tweezers. It is often easier to take the probe from the case, as demonstrated in Figure 4-8, than to try to grasp the probe with the case sitting on the desk or table. This method allows the probe to be held at an angle, making it easier to insert it into the nose assembly.
Ensure that the ends of the metal spring for retaining the cantilever are not caught in between the nose cone and the socket.
Keysight 5500 SPM User’s Guide 5-7
Preparing for Imaging 5
Figure 4-8 Holding the tweezers as shown, remove a probe from the
protective case
5 With the free hand, use the spring key to rock back the retainer
spring (Figure 4-9), pulling it back until it contacts the scanner body.
6 Place the probe between the guides such that a little more than half
of the probe extends over the lens (placement will vary depending on the type and shape of the probe). Figure 4-9 shows this process. The final probe position should be as shown in Figure 4-7 on page 7.
Figure 4-9 Hold back the retaining spring while placing the probe
7 Gently lower the spring clip to hold the probe in place.
Keysight 5500 SPM User’s Guide 5-8
Preparing for Imaging 5
CAUTION
The retainer spring can snap back with enough force to damage the probe, so be sure to release the spring slowly and gently.

Two-Piece Nose Assembly

The two-piece nose assembly was designed to simplify the process of inserting a probe through the use of an assembly fixture (Figure 4-10). The nose assembly consists of a body, which inserts into the scanner, and a flat, stainless steel disk which holds the cantilever. The disk is held to the body magnetically and can be separated by holding the disk at its edges and gently pulling it from the body. The proper attachment of the disk is ensured by aligning the three positioning guides on the body of the nose cone with the tabs on the disk.
Figure 4-10 Two-piece nose assembly, removal tool and assembly
fixture. The nose assembly disk is shown on the fixture; the body is shown to its left.
Inserting the Body of the Two-Piece Nose Assembly
As with the one-piece nose assembly, the body of the two-piece assembly is held in the scanner by an O-ring. To insert the body, first place the scanner in the scanner mounting fixture (Figure 4-1 on
Keysight 5500 SPM User’s Guide 5-9
Preparing for Imaging 5
page 2). Place the body in the socket on top of the scanner, aligning its contact pins.
Applying even, steady, vertical pressure with your fingers to seat the body into its socket.
CAUTION
It is essential to push evenly and straight down when inserting the nose assembly. Small off-axis forces will create LARGE torques about the anchor point for the piezoes, where most breakage occurs.
Removing the Body of the Two-Piece Nose Assembly
As with the one-piece nose assembly, a tool (Figure 4-11) is included to remove the body of the two-piece assembly from the scanner. The tool limits damaging, lateral forces on the scanner during the removal process. The following is the only acceptable procedure for removing the nose assembly body:
Figure 4-11 Two-piece nose assembly removal tool
1 Place the scanner in the scanner mounting fixture.
2 Remove the nose assembly disk from the body by gently pulling it
up from its edges.
3 Carefully slide the removal tool onto the nose assembly, ensuring
that the opening seats on both sides.
4 Position thumb on the flat surface of the removal tool and fingers on
BOTH sides of the extraction arm.
5 Gently pull up with fingers while pushing down the thumb.
Keysight 5500 SPM User’s Guide 5-10
Preparing for Imaging 5
Inserting a Probe in the Two-Piece Nose Assembly
AFM probe tips are extremely delicate and can break when dropped. Follow these instructions to safely insert a probe in the nose assembly:
1 Place the nose assembly disk on the assembly fixture, as shown in
Figure 4-12. Make sure that it aligns with the center disk and two
small alignment pins.
Figure 4-12 Two-piece nose assembly disk on fixture
2 Using tweezers, gently grasp a probe from its sides, applying just
enough pressure to secure it in the tweezers.
3 Move the assembly fixture lever to the right, which will slightly
separate the nose assembly disk (Figure 4-13).
Keysight 5500 SPM User’s Guide 5-11
Preparing for Imaging 5
Figure 4-13 Move the lever to open the nose assembly disk.
4 Place the probe under the copper-colored spring clip on the nose
assembly disk. Use the alignment guides in the fixture to help locate the probe laterally.
5 A small alignment spot on the fixture (Figure 4-12 on page 11)
indicates the proper location for the cantilever tip. Place the probe such that the tip is close as possible to this spot.
6 Move the lever to the left to close the nose assembly disk.
7 Use the tweezers to finely adjust the probe such that the cantilever is
aligned over the alignment spot. Only grasp the probe from the sides to avoid damaging the cantilever.
8 Grasping the nose assembly disk from the edges, remove it from the
fixture and align it on the nose assembly body already in the scanner.

Inserting the Scanner and Connecting Cables

At this point the probe, nose assembly and scanner should all be assembled into one unit.
1 Make sure there is adequate clearance below the scanner socket in
the middle of the microscope.
2 Place the scanner assembly into the scanner socket, with the
scanner’s frosted screen facing up and forward (Figure 4-14).
Keysight 5500 SPM User’s Guide 5-12
Preparing for Imaging 5
CAUTION
Figure 4-14 Placing scanner assembly into microscope
3 Finger-tighten the knob on the right side of the microscope base to
lock the scanner in position.
4 Attach the high voltage (red) and low voltage (blue) cables on either
side of the scanner to the sockets on the microscope base. The cables are color coded to avoid confusion. If you are using a closed-loop scanner, connect its third cable to the C/L socket on the rear of the Head Electronics Box.
Be sure to withdraw tip from sample surface before disconnecting cables and removing scanner from microscope base.
Keysight 5500 SPM User’s Guide 5-13

Aligning the Laser

Preparing for Imaging 5
The next step is to ensure that the scanner laser spot is aligned to reflect off of the cantilever. Several methods can be used to do so.
In most cases, particularly with highly reflective samples, you can use the 5500 SPM video system to focus on the cantilever and align the laser spot (Figure 4-15). If the sample is not reflective, use a piece of gold film on mica or other reflective sample.
The laser spot will be visible in the video image until it crosses the cantilever, so you can use a similar procedure to the paper method below. See “Using the Video System” later in this chapter.
Figure 4-15 Using video system to align laser
Figure 4-16 shows how the position of the laser with respect to the
cantilever affects the position of the laser reflection relative to the
Keysight 5500 SPM User’s Guide 5-14
Preparing for Imaging 5
detector. Due to the variation of cantilever types and vendors, the position of the laser on the cantilever needs to be optimized for each tip.
Figure 4-16 Laser alignment
Keysight 5500 SPM User’s Guide 5-15
Preparing for Imaging 5
Note that the IR sensor card should be used for coarse positioning of the laser if using the 980 nm IR scanners.
Figure 4-17 Coarse laser alignment for IR scanners using sensor card.
Scanner removed from microscope base for clarity.
Another method is to place a white card or piece of paper under the scanner as it sits in the base to make the laser spot visible. By moving the laser you can then align it on the probe—when this happens the probe will block the laser spot, and the spot will no longer be visible on the paper.
1 Place a white piece of paper or business card on the table below the
microscope. If using a 980 nm IR scanner, use an IR sensor card in
Keysight 5500 SPM User’s Guide 5-16
Preparing for Imaging 5
place of a business card (see Figure 4-17). The operational range of the sensor card is 700 nm to 1400 nm.
2 Turn on the Head Electronics Box, which will activate the laser. You
should be able to see the red laser spot on the paper (Figure 4-18).
Figure 4-18 Aligning laser spot over white paper.
The laser alignment knobs are located on the top of the scanner (Figure 4-19). The front-to-back knob moves the laser spot toward the
Keysight 5500 SPM User’s Guide 5-17
Preparing for Imaging 5
cantilever tip (counterclockwise) or away from it. The left-to-right knob adjusts the lateral position.
Figure 4-19 Use the scanner knobs to position the laser spot
3 Rotate the front-to-back knob clockwise to move the laser spot onto
the cantilever chip (Diagram B in Figure 4-20). When the laser reaches the chip it will be blocked and will no longer be visible on the paper. You should only need to turn the knob a few rotations.
Keysight 5500 SPM User’s Guide 5-18
Preparing for Imaging 5
Figure 4-20 Steps to aligning laser on cantilever beam tip
4 Rotate the front-to-back knob counterclockwise until the spot just
reappears on the paper. The spot is now at the edge of the chip (Diagram C in Figure 4-20).
5 Rotate the left-to-right knob to position the laser on the cantilever
(Diagram D in Figure 4-20). As the laser passes over the cantilever it will disappear and reappear from the paper in rapid succession. When the laser is on the cantilever the reflection will be visible on the frosted screen.
6 Turn the front-to-back knob counterclockwise to move the spot
down the cantilever, toward the tip until the spot on the frosted glass disappears (and the spot reappears on the paper) (Diagram E in
Figure 4-20).
7 Turn the front-to-back knob clockwise slightly to position the laser
just on the cantilever tip (Diagram F in Figure 4-20). The spot will reappear on the ground glass.
Keysight 5500 SPM User’s Guide 5-19
Preparing for Imaging 5
The process is similar for triangular-shaped cantilevers, with the exception that the laser will be obscured twice as it moves left to right (over the two beams). The process is shown in Figure 4-21.
Figure 4-21 Steps to aligning the laser on triangular cantilevers
A potential error during the alignment process is to turn either of the positioning controls too far in the wrong direction and to thereby lose the laser spot altogether. Figure 4-22 shows the positioning controls when they are well out of alignment. The left-to-right knob will be visibly tilted when the lateral alignment is far out, and the laser housing will be moved to one side when the front-to-back alignment is out. The easiest way to recover is to roughly center both controls again, moving
Keysight 5500 SPM User’s Guide 5-20
Preparing for Imaging 5
the laser back to the center of its travel in both directions. Doing so should make the laser spot reappear.
Figure 4-22 Laser alignment control when far out of alignment
Keysight 5500 SPM User’s Guide 5-21

Inserting and Aligning the Detector

The photodiode detector records changes in the position of the laser spot as the cantilever passes over the sample surface. As shown in
Figure 4-23, the detector senses the movement of the laser spot between
the four quadrants, reporting the AFM (vertical deflection), LFM (lateral, or friction), and SUM signals.
Preparing for Imaging 5
Figure 4-23 Photodiode detector operation
To install the photodiode detector, insert it into the scanner until it stops (Figure 4-24). Plug the detector cable into the Detector socket on the
Keysight 5500 SPM User’s Guide 5-22
Preparing for Imaging 5
microscope base. Be sure to align the laser on the cantilever before installing the detector in the scanner.
Figure 4-24 Inserting the detector module into the scanner
The Gain Switches on the detector determine whether the laser signal is amplified before going to the rest of the electronics. Gain is disabled when all four switches are pushed up, away from the adjustment knob. Each switch represents one of the four quadrants in the photodetector, therefore all switches should be either up for normal operation, or down to increase the signal when using less reflective cantilevers.
Detector alignment is completed through the PicoView software:
1 Launch PicoView. The Laser Alignment window (as well as other
windows) will open, displaying the position of the laser spot on the
Keysight 5500 SPM User’s Guide 5-23
Preparing for Imaging 5
photodiode detector (Figure 4-25). You can also click the Laser Alignment toolbar button to open the window.
NOTE
Figure 4-25 Laser Alignment window in PicoView
The vertical red bar on the left shows the Sum of all four quadrants. The Deflection signal is the difference between the top and bottom halves divided by the Sum. The Friction signal is the difference between the left and right halves divided by the Sum. The Laser On and Off can be controlled through the software by selecting or clearing, respectively, the Laser on check box. This software switch is in series with the mechanical switch on the Head Electronics Box (HEB), so if the check box is cleared, the laser will not turn on regardless of the position of the HEB switch, and vice versa.
These signals can also be seen on the Head Electronics Box where Meter A is the Sum signal reading and Meter B shows Deflection and Friction (LFM) depending on the state of the switch directly below the meter.
2 Use the knobs on the detector to move the laser spot to the center of
the four quadrants. The front (deflection) knob moves the spot up (clockwise) or down (counterclockwise). The left (friction) knob moves the spot to the left (clockwise) or to the right (counterclockwise).
3 If the laser spot is off the detector, the SUM signal will be close to
0 V and the solid laser spot in the Laser Alignment window will be displayed as an unfilled circle. Take the detector out of the scanner
Keysight 5500 SPM User’s Guide 5-24
Preparing for Imaging 5
and adjust the position of the detector using the laser spot on the frosted screen on the scanner.
4 For Contact Mode, the dotted yellow line (Figure 4-26) shows the
recommended vertical alignment of the laser prior to approaching the sample.
Figure 4-26 Align spot to yellow, dotted line for Contact Mode
Keysight 5500 SPM User’s Guide 5-25

Mounting the Sample

Preparing for Imaging 5
The Keysight 5500 SPM accepts a wide variety of sample plates, including specialized plates for imaging in liquid, in controlled temperature, etc. To use a sample plate:
1 Mount the sample to the sample plate.
In general, samples should be held in place securely enough to prevent drift or creep during measurement, but not so firmly as to induce stress in the sample. Several mounting methods are available. A common approach is to mount the sample on a magnetically attractive backing which can then be held by the magnet on the standard sample plate. Large, flat samples can be held down using the clips from the liquid cell plate. Double-back tape can also be used, though the tape tends to deform easily and can lead to creep during imaging.
CAUTION
Verify that there is enough space between the scanner and sample plate such that the tip will not contact the sample plate once the plate is mounted. Contact with the plate will certainly damage the probe and perhaps the sample.
It is recommended that the scanner be moved up 100 microns or more whenever changing a probe or plate to avoid damage. Simply hold the switch on the front of the HEB in the “Open” for 5-10 seconds to provide adequate clearance.
2 Place the front alignment tab of the sample plate over the front
alignment pin, as in image A of Figure 4-27.
Keysight 5500 SPM User’s Guide 5-26
Loading...