Keysight 3458A Calibration Manual

Page 1
Keysight 3458A Multimeter
Calibration Manual
Page 2
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 1988 - 2017 No part of this manual may be repro-
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
03458-90017
Edition
Edition 7, July 1, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight 3458A Calibration Manual
Page 3
KEYSIGHT TECHNOLOGIES WARRANTY STATEMENT
KEYSIGHT PRODUCT: 3458A multimeter
DURATION OF WARRANTY: 1 year
1 Keysight Technologies warrants Keysight hardware, accessories and supplies
against defects in materials and workmanship for the period specified above. If Keysight receives notice of such defects during the warranty period, Keysight will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
2 Keysight warrants that Keysight software will not fail to execute its
programming instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If Keysight receives notice of such defects during the warranty period, Keysight will replace software media which does not execute its programming instructions due to such defects.
3 Keysight does not warrant that the operation of Keysight products will be
interrupted or error free. If Keysight is unable, within a reasonable ntime, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the product.
4 Keysight products may contain remanufactured parts equivalent to new in
performance or may have been subject to incidental use.
5 The warranty period begins on the date of delivery or on the date of installation
if installed by Keysight. If customer schedules or delays Keysight installation more than 30 days after delivery, warranty begins on the 31st day from delivery.
6 Warranty does not apply to defects resulting from (a) improper or inadequate
maintenance or calibration, (b) software, interfacing, parts or supplies not supplied by Keysight, (c) unauthorized modification or misuse, (d) operation outside of the published environmental specifications for the product, or (e) improper site preparation or maintenance.
7 TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES ARE
EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED AND KEYSIGHT SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTY OR CONDITIONS OF MERCHANTABILITY, SATISFACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
Keysight 3458A Calibration Manual 3
Page 4
8 Keysight will be liable for damage to tangible property per incident up to the
greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Keysight product.
9 TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS
WARRANTY STATEMENT ARE CUSTOMER’S SOLE AND EXLUSIVE REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL KEYSIGHT OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT, TORT, OR OTHERWISE. FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND: THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE, RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
4 Keysight 3458A Calibration Manual
Page 5
U.S. Government Restricted Rights
The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227- 7013 (Oct 1988), DFARS 252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR
2.101(a), or as “Restricted computer software” as defined in FAR 52.227-19 (Jun
1987)(or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Keysight standard software agreement for the product involved.
Keysight 3458A Calibration Manual 5
Page 6
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC) Alternating current (AC)
WARNING, RISK OF ELECTRIC SHOCK.
Indicates the field wiring terminal that must be connected to earth ground before operating the equipment — protects against electrical shock in case of fault.
Instruction manual symbol affixed to Alternating current (AC) product. Indicates that the user must refer to the manual for specific WARNING or CAUTION information to avoid personal injury or damage to the product.
Frame or chassis ground terminal—typically connects to the equipment's metal frame.
6 Keysight 3458A Calibration Manual
Page 7
Safety Considerations
WARNING
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
Ground the equipment: For Safety Class 1 equipment (equipment having
a protective earth terminal), an uninterruptible safety earth ground must be provided from the mains power source to the product input wiring terminals or supplied power cable.
DO NOT operate the product in an explosive atmosphere or in the
presence of flammable gases or fumes.
For continued protection against fire, replace the line fuse(s) only with
fuse(s) of the same voltage and current rating and type. DO NOT use repaired fuses or short-circuited fuse holders.
Keep away from live circuits: Operating personnel must not remove
equipment covers or shields. Procedures involving the removal of covers or shields are for use by service-trained personnel only. Under certain conditions, dangerous voltages may exist even with the equipment switched off. To avoid dangerous electrical shock, DO NOT perform procedures involving cover or shield removal unless you are qualified to do so.
Keysight 3458A Calibration Manual 7
DO NOT operate damaged equipment: Whenever it is possible that the
safety protection features built into this product have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to Keysight for service and repair to ensure that safety features are maintained.
Page 8
WARNING
DO NOT service or adjust alone: Do not attempt internal service or
adjustment unless another person, capable of rendering first aid and resuscitation, is present.
DO NOT substitute parts or modify equipment: Because of the danger of
introducing add itional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to Keysight for service and repair to ensure that safety features are maintained.
Measuring high voltages is always hazardous: ALL multimeter input
terminals (both front and rear) must be considered hazardous whenever inputs greater than 42V (dc or peak) are connected to ANY input terminal.
Permanent wiring of hazardous voltage or sources capable of delivering
grater than 150 VA should be labeled, fused, or in some other way protected against accidental bridging or equipment failure.
DO NOT leave measurement terminals energized when not in use.
DO NOT use the front/rear switch to multiplex hazardous signals between
the front and rear terminals of the multimeter.
8 Keysight 3458A Calibration Manual
Page 9
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/3458A
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight 3458A Calibration Manual 9
Page 10
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
10 Keysight 3458A Calibration Manual
Page 11

Table of Contents

Keysight Technologies WARRANTY STATEMENT . . . . . . . . . . . . . . . . . .3
U.S. Government Restricted Rights . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . . 9
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1 3458A Calibration Introduction
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Calibration Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Security code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Changing the security code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Hardware lock-out of calibration . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Number of calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Monitoring for CAL violations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Monitoring Calibration Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
2 Operational Verification Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Operational Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
2-wire ohms function offset test . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
4-wire ohms function gain test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
DC voltage function gain test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
DC voltage function offset test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
3 Adjustment Procedures
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Keysight 3458A Calibration Manual 11
Page 12
Preliminary Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Front Terminal Offset Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Rear Terminal Offset Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
DC Gain Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Resistance and DC current adjustment . . . . . . . . . . . . . . . . . . . . . . 43
AC Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4 Performance Verification Tests
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Test card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Calibration cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Test considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
General test procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
DC Voltage Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
DC voltage function offset test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
DC voltage function gain test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Analog AC voltage performance tests . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
AC voltage test procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
DC Current Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
DC current function offset test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
DC current function gain test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Ohms Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
2-wire ohms function offset test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4-wire ohms function offset test (rear terminals) . . . . . . . . . . . . . . . 73
12 Keysight 3458A Calibration Manual
Page 13
4-wire ohms function gain test . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Frequency Counter Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . .76
Required equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Preliminary steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Frequency counter accuracy test . . . . . . . . . . . . . . . . . . . . . . . . . . .76
5Command Summary
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
ACAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
CAL? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
CALNUM? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
CALSTR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
REV? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107
SCAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
SECURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111
TEMP? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
A Specifications
B Electronic Calibration of the 3458A (Product Note 3458A-3)
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118
Saving Calibration Time and Money . . . . . . . . . . . . . . . . . . . . . . . . . .119
The basis for auto-calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
Offset adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
DC gain adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
Resistance and DC current adjustments . . . . . . . . . . . . . . . . . . . . .122
AC flatness and gain adjustments . . . . . . . . . . . . . . . . . . . . . . . . . .123
One-time adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Traceability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126
Externally derived calibration constants . . . . . . . . . . . . . . . . . . . . .126
Keysight 3458A Calibration Manual 13
Page 14
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight 3458A Calibration Manual
Page 15

List of Figures

Figure 3-1 4-terminal short . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Figure 3-2 DC gain adjustment connections . . . . . . . . . . . . . . . . .43
Figure 3-3 Resistance and DC current adjustment connections .45
Figure 3-4 Characterizing the adjustment setup . . . . . . . . . . . . .48
Figure 3-5 AC adjustment connections . . . . . . . . . . . . . . . . . . . . .49
Figure 4-1 General test procedure . . . . . . . . . . . . . . . . . . . . . . . .59
Keysight 3458A Calibration Manual 15
Page 16
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
16 Keysight 3458A Calibration Manual
Page 17

List of Tables

Table 4-1 Offset performance tests . . . . . . . . . . . . . . . . . . . . . . .61
Table 4-2 DCV gain performance tests . . . . . . . . . . . . . . . . . . . .63
Table 4-3 AC performance tests . . . . . . . . . . . . . . . . . . . . . . . . . .67
Table 4-4 Current offset performance tests . . . . . . . . . . . . . . . . .69
Table 4-5 DCI gain performance tests . . . . . . . . . . . . . . . . . . . . .70
Table 4-6 OHMF gain performance tests . . . . . . . . . . . . . . . . . . .74
Keysight 3458A Calibration Manual 17
Page 18
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
18 Keysight 3458A Calibration Manual
Page 19
Keysight 3458A Multimeter Calibration Manual
1 3458A Calibration
Introduction
Introduction 20 Calibration Security 21 Monitoring Calibration Constants 25
19
Page 20
1 3458A Calibration Introduction
WARNING

Introduction

This manual provides operation verification procedures, adjustment procedures, and performance verification procedures for the 3458A multimeter.
The information contained in this manual is intended for the use of service-trained personnel who understand electronic circuitry and are aware of the hazards involved. Do not attempt to perform any of the procedures outlined in this section unless you are qualified to do so.
The manual contains five chapters and two appendixes.
Chapter 1: Introduction describes the manual contents and calibration
security features of the 3458A.
Chapter 2: Operational Verification provides a short test procedure to verify
that the multimeter is functioning properly.
Chapter 3: Adjustment Procedure gives the procedures for adjusting the
multimeter to obtain best accuracy.
Chapter 4: Performance Verification is comprised of test procedures used to
verify that all parts of the instrument are functioning properly and within specification. This chapter contains Test Cards for recording the results of each test.
Chapter 5: Command Summary provides an alphabetical summary of
commands that are used in adjusting and performance testing the 3458A.
Appendix A: 3458A Technical Specifications
Appendix B: Electronic Calibration of the 3458A, Product Note 3458A-3
20 Keysight 3458A Calibration Manual
Page 21

Calibration Security

The calibration security feature of the 3458A allows the person responsible for calibration to enter a security code to prevent accidental or unauthorized calibration (CAL) or autocalibration (ACAL). The SECURE command is used to change the security code of the 3458A.

Security code

The security code is an integer from –2.1E-9 to 2.1E9. If the number specified is not an integer, the multimeter rounds it to an integer value. The multimeter is shipped from the factory with its security code set to 3458. Specifying 0 for the new_code in the SECURE command disables the security feature making it no longer necessary to enter the security code to perform a calibration or autocal.

Changing the security code

The security code is changed with the SECURE command which has the following syntax:
SECURE old_code, new_code [,acal_secure]
The procedure for changing the security code is as follows: 1 Access the SECURE command. (Press the blue SHIFT key, then the
S menu key. If using the full command menu use the SECURE command).
3458A Calibration Introduction 1
scroll key to display the
2 Enter the old security code, the delimiter (,) and the new security code. If you
want to control the auto calibration of the multimeter, enter another delimiter (,) and the acal_secure parameter ON. The instrument is shipped from the factory with the security code set to 3458 and the acal_secure parameter set to ON (security code required to do an acal).
3 Press the Enter key. The instrument will now respond to the new security code.
In the event the security code is unknown, the security feature can be disabled to permit a new code to be entered. Perform the following procedure to disable the current unknown security code and enter a known code:
a Turn the instrument off and remove the line cord and all external inputs to
the 3458A.
Keysight 3458A Calibration Manual 21
Page 22
1 3458A Calibration Introduction
NOTE
b Remove the top cover.
i Remove both rear handle brackets with a #2 Pozidrive.
ii Remove the rear bezel by loosening the four #15 TORX screws.
iii With the back of the instrument facing you, remove the #10 TORX screw
securing the top cover to the right side.
iv Remove the top cover.
c Change the position of jumper JM600 on the 03458-66505 assembly, or
option 001 03458-66515 assembly from the left position to the right position (front of instrument facing you).
d Reconnect the power and turn the instrument on.
e Access the SECURE command (Press the blue SHIFT key then the S MENU
key. Use the SECURE command).
f Enter the number 0 followed by the delimiter (,) and the security code you
want to use.
g Press the ENTER key.
h Turn the instrument off, disconnect power, and return jumper JM600 to the
left position (front of instrument facing you).
i Replace the top cover and reconnect power. The instrument will now
respond to the new security code you just entered.
scroll key if in the full command menu to display the
When jumper JM600 is in the right position, the security feature is disabled (i.e. old_code = 0). It is possible to calibrate the instrument without entering a security number under these conditions. If a new security number (new_code of SECURE command) is not entered while the jumper is in the right position the original number will again be in effect when jumper JM600 is returned to the left position.
22 Keysight 3458A Calibration Manual
Page 23

Hardware lock-out of calibration

NOTE
You can set jumper J132 on the 03458-66505 or -66515 (option 001) assembly to require removing the instrument cover and repositioning this jumper whenever adjustments (CAL command) are to be made. Use the following procedure to set hardware “lock-out” of the CAL and ACAL commands.
1 Remove the instrument top cover as described in steps a and b of the previous
section.
2 With the instrument front facing you, set jumper J132 to the right position.
Neither the CAL or ACAL command can be executed when the jumper is in this position even when the correct security code is entered.
3 Replace the top cover.
To perform an adjustment with the CAL command or do an auto-calibration with the ACAL command, you must remove the top cover and set jumper J132 to the left position (instrument front facing you). You may attach a seal to the top cover that must be broken to remove the cover indicating if unauthorized access to the hardware has occurred.

Number of calibrations

3458A Calibration Introduction 1
You can monitor the number of times calibration has been performed (CAL and ACAL combined if ACAL is secured by the SECURE command) by using the CALNUM? command. CALNUM? (calibration number query) returns a decimal number indicating the number of times the multimeter has been unsecured and adjusted. The number of calibrations is stored in cal-protected memory and is not lost when power is removed. The calibration number is incremented by 1 whenever the multimeter is unsecured and a CAL. (or ACAL if secured) is executed. If autocal is secured, the calibration number is also incremented by 1 whenever an autocal is performed; if unsecured, autocal does not affect the calibration number.
The multimeter was adjusted before it left the factory. This has incremented the calibration number. When you receive the multimeter, read the calibration number to determine the initial value you start with. The procedure for reading the number of calibrations is presented after this note.
Keysight 3458A Calibration Manual 23
Page 24
1 3458A Calibration Introduction
Read the number of calibrations with the following procedure:
1 Access the CALNUM? command. In the full command menu, press the blue
SHIFT key then the C menu key. Use the command. (Full command menu is obtained by pressing the blue SHIFT key, the “menu” key, the
2 Press the ENTER key.
3 The display indicates CALNUM and the current number of calibrations.

Monitoring for CAL violations

You can use the CALSTR command in conjunction with the CALNUM? command to monitor for calibration (CAL) violations. After each authorized calibration has taken place, use the CALNUM? command to access the current number of calibrations as described in the previous section. Store this number in the calibration string (must be done remotely) with the CALSTR command i.e. OUTPUT 722;“CALSTR 'calnum = 270'”. At any later time, you can execute the CALNUM? and CALSTR? commands and compare the two calibration numbers. If the CALNUM is greater than the CALSTR entry, calibration security has been violated and unauthorized adjustments performed.
scroll key to display the CALNUM?
scroll key, and ENTER key.)
The following example illustrates monitoring for CAL violations:
1 After adjustments are performed, execute CALNUM?. Display shows “CALNUM
270”
2 Remotely execute OUTPUT 722;“CALSTR 'CALNUM=270'”
3 At a later time you can verify if CAL has been violated by executing CALNUM?
and CALSTR shift? to see if CALNUM is greater than that stored in CALSTR.
24 Keysight 3458A Calibration Manual
Page 25

Monitoring Calibration Constants

WARNING
Each time you do an ACAL, most calibration constants are recalculated. Executing an ACAL ALL recalculates 197 of the 253 calibration constants. The remaining constants (such as internal reference and offset constants) are externally derived and not changed by an ACAL. Periodically you may want to monitor a particular constant and track its movement within the lower and upper limits (see CAL? command, cal_item parameter). This may give you an indication of the calibration cycle you want to establish for your 3458A. Information on the externally derived calibration constants and the 197 internally derived calibration constants is presented on the last page of Appendix B. Detailed information about each constant appears in the CAL? command located in Chapter 5, "Command
Summary".
– Only qualified, service trained personnel who are aware of the hazards
involved should remove or install the multimeter or connect wiring to the multimeter. Disconnect the multimeter's power cord before removing any covers, changing the line voltage selector switches, or installing or changing the line power fuse.
– Measuring high voltage is al ways hazardous. All multimeter input
terminals (both front and rear) must be considered as hazardous whenever inputs in excess of 42 V are connected to any terminal. Regard all terminals as being at the same potential as the highest voltage applied to any terminal.
3458A Calibration Introduction 1
Keysight 3458A Calibration Manual 25
– Keysight Technologies recommends that the wiring installer attach a
label to any wiring having hazardous voltages. This label should be as close to the input terminals as possible and should be an eye-catching color, such as red or yellow. Clearly indicate on the label that high voltages may be present.
Page 26
1 3458A Calibration Introduction
CAUTION
Rated input Maximum non-destructive input
HI to LO Input: ±l000 V peak ±1200 V peak
HI/LO W Sense to LO Input: ±200 V peak ±350 V peak
HI to LO W Sense: ±200 V peak ±350 V peak
LO Input to Guard: ±200 V peak ±350 V peak
Guard to Earth Ground: ±500 V peak ±l000 V peak
HI/LO Input, HI/LO W Sense, or I terminal to earth ground:
±l000 V peak ±1500 V peak
Front terminals to rear terminals:
±1000 V peak ±1500 V peak
– The current input terminals (I) are rated at ±1.5 A peak with a maximum
non-destructive input of <1.25 A RMS. Current inputs are fuse protected. The multimeter's input voltage ratings are:
– The multimeter will be damaged if any of the above maximum
non-destructive inputs are exceeded.
26 Keysight 3458A Calibration Manual
Page 27
Keysight 3458A Multimeter Calibration Manual
2 Operational Verification
Tests
Introduction 28 Operational Tests 29
27
Page 28
2 Operational Verification Tests

Introduction

This section contains Operational Verification Tests which provide an abbreviated method of testing the operation and accuracy of the unit. The Operational Verification Tests are designed to provide a 90% confidence that the 3458A is operational and meets the specifications listed in Appendix A.
Operational Verification Tests perform a three point verification. These three points are the basis for all internal electronic adjustments (see the section titled
The basis for auto-calibration in Appendix B, "Electronic Calibration of the 3458A (Product Note 3458A-3)"). Prior to the three point test, a self test verifies that all
calibration constants are within their upper and lower limits, an indicator of proper operation.
28 Keysight 3458A Calibration Manual
Page 29

Operational Tests

NOTE

Required equipment

The following equipment or its equivalent is required for these operational tests.
– Stable DC voltage/resistance standard (Fluke 5700A or equivalent)
– Transfer standard DMM (3458A Opt. 002 within 90 days of CAL)
– Low thermal short (copper wire)
– Low thermal test leads (such as Keysight 11053A, 11174A, or 11058A)
– Equipment required for the adjustment procedures can be used for
operational tests since the three-point test verifies the external adjustment points of the adjustment procedure.
– To have your transfer standard 3458A OPT. 002 calibrated to 90 day
specifications, contact your Keysight Technologies sales and service office.

Preliminary steps

Operational Verification Tests 2
1 Verify that the DC voltage/resistance standard is properly warmed up.
2 The 3458A requires a 4 hour warm-up period. If this has not occurred turn the
instrument ON and allow it to warm up before proceeding.
3 The internal temperature of the 3458A under test must be within 5
temperature when last calibrated. Use the TEMP? command to obtain the current internal temperature and compare it to the calibration temperature obtained by executing the command CAL? 59 for DCV and CAL? 60 for OHMS.
You can use the the temperatures on the Test Card.
4 If the instrument self test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
5 Execute the ACAL OHMS function. This auto calibration will take approximately
ten minutes to complete.
6 Configure the transfer standard DMM as follows:
Keysight 3458A Calibration Manual 29
and scroll keys to view entire CAL? message. Record
°C of its
Page 30
2 Operational Verification Tests
–OHM
–NDIG 8
– NPLC 100
–TRIG SGL
7 Configure the DMM under test as follows:
–OHM
–NDIG 8
– NPLC 100
–TRIG SGL

2-wire ohms function offset test

This procedure operation verifies 2-wire ohms offset for both the front and rear terminals.
1 Connect a low thermal short across the 3458A front HI and LO terminals (see
Figure 3-1 on page 39).
2 Verify that the Terminals switch is in the Front position.
3 Set the 3458A under test to the 10 W range (function = OHM). Allow five
minutes for the range relays to thermally stabilize.
4 Execute Trig from the front panel two times and use the Operational Test Card
to record the offset reading. Reading must be less than the limit specified on the test card.
5 Remove the short from the front panel input terminals and connect it to the
rear input HI and LO terminals.
6 Change the Terminals switch to the Rear position.
7 Allow 5 minutes for thermal stabilization.
8 Execute Trig from the front panel two times and record the rear terminal offset
reading on the Operational Test Card. Reading must be less than the limit specified on the test card. If reading is greater than the limit, refer to Chapter 3 to make adjustments.
9 Remove the short from the rear input terminals.
30 Keysight 3458A Calibration Manual
Page 31

4-wire ohms function gain test

The following procedure verifies the gain of the ohms function. The 10 kW point is used for internal electronic calibration using ACAL. The procedure requires alternately connecting the transfer standard DMM and then the 3458A under test to the resistance verification standard as described in the Chapter 4 section titled
General test procedure.
1 Connect the resistance standard to the transfer standard DMM 4-wire ohms
front input terminals.
2 Set the Terminals switch of both DMMs to the Front position.
3 Set the range of the transfer standard DMM to 10 kW (function = OHMF).
4 Set the range of the 3458A under test to 10 kW (function = OHMF).
5 Set the resistance standard to 10 kW.
6 Execute Trig from the front panel two times and read the value of the
resistance standard as measured with the transfer standard DMM and record this reading in the “Transfer standard reading” column of the ohms gain operational test card.
7 Remove the connection between the transfer standard DMM and the
resistance standard.
Operational Verification Tests 2
8 Connect the resistance standard to the 4-wire ohms input terminals of the
3458A under test.
9 Execute Trig from the front panel two times and read the value as measured
with the 3458A under test and record this value in the “Unit under test reading” column of the 4-wire ohms function gain operational test card.
10 Calculate and record in the column provided, the difference (absolute value)
between the transfer standard DMM reading and the unit under test reading for the test.
11 If the difference calculated is greater than the specified limits, refer to
Chapter 3, "Adjustment Procedures", to make adjustments.
12 Disconnect the resistance standard from the 3458A input terminals.
Keysight 3458A Calibration Manual 31
Page 32
2 Operational Verification Tests

DC voltage function gain test

The following procedure verifies the 10 V input on the 10 V range. This test verifies the gain of the DC voltage function and checks the point used for internal adjustments by ACAL. The procedure requires alternately connecting the transfer standard DMM and then the 3358A under test to the DC verification source as described in the general test description of Chapter 4, "Performance Verification
Tests".
1 Execute the ACAL DCV command using the front panel “Auto Cal” key and
scroll keys. This auto calibration will take approximately two minutes to complete.
2 Configure the transfer standard DMM as follows:
–DCV
–NDIG 8
– NPLC 100
–Trig SGL
3 Configure the DMM under test as follows:
–DCV
–NDIG 8
– NPLC 100
–Trig SGL
4 Set the range of the transfer standard DMM to 10 V (function = DCV).
5 Set the range of the 3458A under test to 10 V (function = DCV).
6 Connect the DC voltage source to the transfer standard DMM.
7 Set the DC voltage source to 10 V.
8 Execute Trig SGL and read the output of the DC voltage source as measured
with the transfer standard DMM and record this reading in the “Transfer standard reading” column of the DC voltage operational test record.
9 Remove the connection from the transfer standard DMM to the DC voltage
source.
10 Connect the 3458A under test to the DC voltage source.
32 Keysight 3458A Calibration Manual
Page 33
11 Execute Trig SGL and read the value as measured with the 3458A under test
and record this value in the “Unit under test reading” column of the DC voltage operational test record.
12 Connect the DC voltage source to the transfer standard DMM.
13 Repeat steps 8 through 11 for a –10 V DC voltage source output.
14 Calculate, and record in the column provided, the difference (absolute value)
between the transfer standard DMM reading and the unit under test reading.
15 If the difference calculated is greater than the specified limits, refer to
Chapter 3, "Adjustment Procedures" to make adjustments.

DC voltage function offset test

This procedure tests the DCV offset voltage specification on the 10 V range. This reading and the 10 V and –10 V readings from the previous DCV gain test are used to do a turnover check of the A-D converter and verify its linearity.
1 Connect a low thermal short across the front panel HI and LO input terminals
of the DMM under test (see Figure 3-1 on page 39).
2 Set the range of the 3458A under test to 10 V.
3 Let the instrument sit for five minutes before taking a reading to allow the
short and relays to thermally stabilize.
4 Execute Trig and record the offset reading on the Test Card.
Operational Verification Tests 2
Turnover check
The turnover check is a calculation using the unit under test readings from tests 4, 5, and 6 on the Test Card. This check verifies the linearity of the A-to-D converter which is fundamental to the 3458A's calibration technique. Calculate the following:
(UUT Reading #4) - (UUT Reading #6) = A
(UUT Reading #5) - (UUT Reading #6) = B If the A-to-D converter is linear. the difference in the absolute values of
A and B will be less than or equal to 4 µV.
Keysight 3458A Calibration Manual 33
Page 34
2 Operational Verification Tests
OPERATIONAL TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
CAL? 60 _________________________ Date _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL OHMS
Test # 3458A input
3458A
range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
2-wire ohms function offset test
1Short10 W N/A __________ N/A 00.25007 00.25007 __________ __________
(Front terminals)
2Short10 W N/A __________ N/A 00.25007 00.25007 __________ __________
(Rear terminals)
4-wire ohms function gain test
3 10 kW 10 kW __________ __________ __________ 00.000142 00.000142 __________ __________
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL DCV
DC voltage function gain test
4 10 V 10 V __________ __________ __________ 00.0000892 00.0000624 __________ __________
5 –10 V 10 V __________ __________ __________ 00.0000892 00.0000624 __________ __________
DC voltage function offset test
6 Short 10 V N/A __________ N/A 00.0000023 00.0000023 __________ __________
34 Keysight 3458A Calibration Manual
Page 35
Keysight 3458A Multimeter Calibration Manual

3 Adjustment Procedures

Introduction 36 Required Equipment 37 Preliminary Adjustment Procedure 38 Front Terminal Offset Adjustment 40 Rear Terminal Offset Adjustment 41 DC Gain Adjustment 42 AC Adjustment 46
35
Page 36
3 Adjustment Procedures

Introduction

This section contains procedures for adjusting the 3458A multimeter. The 3458A uses closed-box electronic adjustment. No potentiometers or other electro-mechanical adjustments are used and the complete adjustment is done without removing any of the multimeter's covers. Only a voltage standard. A resistance standard, a low-thermal short, and an AC signal source are needed to perform all of the adjustments. This chapter contains the following adjustment procedures.
1 Front Terminal Offset Adjustment
2 Rear Terminal Offset Adjustment
3 DC Gain Adjustment
4 Resistance and DC Current Adjustment
5 AC Adjustment
You must perform the adjustments in the order presented in this chapter. All of the adjustments can be performed in approximately one hour (you must allow 4 hours of warm-up time from the time power is applied to the multimeter before performing any adjustments). Whenever adjusting the multimeter, always perform the adjustments numbered 1 through 4 in the above list. Adjustment number 5 (AC Adjustment) is required only once every 2 years or whenever the 03458-60502 or 03458-66503 PC assembly has been replaced or repaired. Product Note 3458A-3 in Appendix B discusses the purpose of each adjustment in detail.
An Adjustment Record is located at the back of this chapter. You should make photocopies of this record and complete the record whenever the multimeter is adjusted. The record contains information such as the date, which adjustments were performed, the calibration number, and the multimeter's adjustment temperature. You can then file the adjustment records to maintain a complete adjustment history for the multimeter.
36 Keysight 3458A Calibration Manual
Page 37

Required Equipment

You will need the following equipment to perform the adjustments:
– A low-thermal 4-terminal short for the offset adjustments (this is typically a
bent piece of copper wire as shown in Figure 3-1 on page 39).
– 10 VDC Voltage Standard--Fluke 732A or equivalent (for the DC Gain
Adjustment).
– 10 kW Resistance Standard--Fluke 742-10 K or equivalent (for the Resistance
and DC Current Adjustment).
– AC Source-Keysight 3325A Synthesizer/Function Generator or equivalent (for
the AC adjustment).
The resultant accuracy of the multimeter depends on the accuracy of the equipment used, the thermal characteristics of the short. and the type of cabling used. We recommend high impedance, low dielectric absorption cables for all connections.
Adjustment Procedures 3
Keysight 3458A Calibration Manual 37
Page 38
3 Adjustment Procedures

Preliminary Adjustment Procedure

Perform the following steps prior to adjusting the 3458A:
1 Select the adjustment area. You can adjust the 3458A on the bench or in a
system cabinet. The temperature of the adjustment environment should be between 15 more accurate the adjustment.
2 Connect the 3458A to line power and turn the multimeter on. Refer to
“Installing the Multimeter” in Chapter 1 of the 3458A User’s Guide for information on setting the line voltage switches and installing the line power fuse.
3 Remove all external input signals from the front and rear input terminals.
4 Select the DCV function (use the DCV key) and the 100 mV range (repeatedly
press the down arrow key until the range no longer- changes). (Refer to Chapter 2 of the 3458A User’s Guide for more information on front panel operation.)
5 Set the front panel Terminals switch to the Front position.
6 Allow the multimeter to warm up for 4 hours from the time power was applied.
(At this point. you can connect the 4-terminal short to the front terminals as shown in Figure 3-1 on page 39 to prepare for the Front Terminal Offset Adjustment.)
°C and 30 °C. The more thermally stable the environment is, the
38 Keysight 3458A Calibration Manual
Page 39
Figure 3-1 4-terminal short
Adjustment Procedures 3
Keysight 3458A Calibration Manual 39
Page 40
3 Adjustment Procedures
NOTE

Front Terminal Offset Adjustment

This adjustment uses an external 4-terminal short. The multimeter makes offset measurements and stores constants for the DCV, DCI, OHM, and OHMF functions. These constants compensate for internal offset errors for front terminal measurements.
Equipment required: A low-thermal short made of 12 or 14 gauge solid copper wire as shown in Figure 3-1 on page 39.
1 Make sure you have performed the steps described previously under
“Preliminary Adjustment Procedures”.
2 Connect a 4-terminal short across the front panel HI and LO Input terminals
and the HI and LO W Sense terminals as shown in Figure 3-1 on page 39.
3 After connecting the 4-terminal short, allow 5 minutes for thermal equilibrium.
Take precautions to prevent thermal changes near the 4-wire short. You should not touch the short after it is installed. If drafts exist, you should cover the input terminals/short to minimize the thermal changes.
4 Execute the CAL 0 command. The multimeter automatically performs the front
terminal offset adjustment and the display shows each of the various steps being performed. This adjustment takes about 5 minutes. When the adjustment is complete, the multimeter returns to displaying DC voltage measurements.
40 Keysight 3458A Calibration Manual
Page 41

Rear Terminal Offset Adjustment

NOTE
This adjustment compensates for internal offset errors for rear terminal measurements.
1 Connect the 4-terminal short to the rear terminals.
2 Set the front panel Terminals switch to Rear.
3 After connecting the 4-terminal short, allow 5 minutes for thermal equilibrium.
Take precautions to prevent thermal changes near the 4-wire short. You should not touch the short after it is installed. If drafts exist, you should cover the input terminals/short to minimize the thermal changes.
4 Execute the CAL 0 command. The multimeter automatically performs the rear
terminal offset adjustment and the display shows each of the various steps being performed. This adjustment takes about 5 minutes. When the adjustment is complete, the multimeter returns to displaying DC voltage measurements.
5 Remove the 4-terminal short from the rear terminals.
Adjustment Procedures 3
Keysight 3458A Calibration Manual 41
Page 42
3 Adjustment Procedures
NOTE

DC Gain Adjustment

In this adjustment, the multimeter measures the standard voltage using its 10 V range. The multimeter then adjusts its gain so that the measured value agrees with the standard's exact value (specified using the CAL command). The multimeter then measures its 7 V internal reference voltage using the 10 V range and stores both the 10 V gain adjustment constant and the value of the internal 7 V reference. This adjustment also automatically performs the DCV autocalibration with computes DC gain constants.
Equipment required: A DC voltage standard capable of providing 10 VDC (the resultant accuracy of the 3458A depends on the accuracy of the voltage standard).
Voltage standards from 1 V DC to 12 V DC can be used for this procedure. However, using a voltage standard <10 V DC will degrade the multimeter's accuracy specifications.
1 Select the DC Voltage function.
2 Set the front panel Terminals switch to Front.
3 Connect the voltage standard to the multimeter's front panel HI and LO Input
terminals as shown in Figure 3-2. If using a Guard wire (as shown in Figure 3-2). set the Guard switch to the Open position. If not using a Guard wire, set the Guard switch to the To LO position.
4 Execute the CAL command specifying the exact output voltage of the
standard. For example, if the standard's voltage is 10.0001 VDC, execute CAL
10.0001. The multimeter automatically performs the DC gain adjustment and the display shows each of the various steps being performed. This adjustment takes about 2 minutes. When the adjustment is complete, the multimeter returns to displaying DC voltage measurements.
5 Disconnect the voltage standard from the multimeter.
42 Keysight 3458A Calibration Manual
Page 43
Adjustment Procedures 3
Figure 3-2 DC gain adjustment connections

Resistance and DC current adjustment

This adjustment calculates gain corrections for the resistance and DC current ranges. The DC Gain Adjustment must be performed prior to this adjustment because this adjustment relies on the values calculated by the DC Gain Adjustment.
Keysight 3458A Calibration Manual 43
Page 44
3 Adjustment Procedures
NOTE
NOTE
NOTE
When offset compensated ohms is enabled (OCOMP ON command), the default delay time used by the multimeter for this adjustment is 50 ms (50 ms is the settling time used after the current source is switched on or off). For most resistance standards and cabling, this provides adequate settling time for the measurements made during the adjustment. If, however, the resistance standard and/or cabling has slow transient response or high dielectric absorption you should specify a longer delay. You can determine this experimentally prior to performing the following adjustment by measuring the resistance standard using a 50 ms delay and then measuring it using a much longer delay (e.g., 1 second). If the two measurements are significantly different, you should use a longer delay in the adjustment procedure. You must specify the longer delay using the DELAY command prior to executing the CAL command (step 5). For example, to specify a 200 ms delay execute: DELAY 200E-3. The multimeter will then use the specified delay in the adjustment. If a value of less than 50 ms is specified, the multimeter will automatically use a delay of 50 ms. Do not specify a delay longer than 60 seconds; a delay >60 seconds will adversely affect the adjustment.
Equipment required: A 10 kW resistance standard (the resultant accuracy of the multimeter depends on the accuracy of the resistance standard used).
Resistance standards from 1 kW to 12 kW can be used for the procedure. However, using a resistance standard <10 kW will degrade the multimeter's accuracy specifications.
1 Select the 4-wire ohms measurement function (use the shifted OHM key).
2 Execute the OCOMP ON command (use the front panel Offset Comp W key).
You can perform this adjustment with offset compensation disabled (OCOMP OFF command). This eliminates the settling time requirement (DELAY command) when dealing with high dielectric absorption in the adjustment setup (see note as the beginning of this adjustment). However, with offset compensation disabled, any offset voltages present will affect the adjustment. For most applications, we recommend enabling offset compensation for this adjustment.
44 Keysight 3458A Calibration Manual
Page 45
Adjustment Procedures 3
3 Set the front panel Terminals switch to Front.
4 Connect the resistance standard to the multimeter's front panel HI and LO
Input and HI and LO Sense terminals as shown in Figure 3-3. If using a Guard wire (as shown in Figure 3-2), set the Guard switch to the Open position. If not using a Guard wire, set the Guard switch to the To LO position.
5 Execute the CAL command specifying the exact value of the resistance
standard. For example, if the standard's value is 10.003 kW, execute CAL
10.003E3. The multimeter automatically performs the resistance and DC current adjustment and the display shows each of the various steps being performed. This adjustment takes about 12 minutes. When the adjustment is complete, the multimeter returns to displaying resistance readings.
6 Disconnect the resistance standard from the multimeter.
7 Execute the ACAL AC command (use the AUTO CAL key). This autocalibrates
the multimeter's AC section since the following AC Adjustment is normally performed only once every two years or whenever the 03458-66502 or 03458-66503 PC Assembly has been replaced or repaired. The AC autocalibration takes about 2 minute to complete.
Figure 3-3 Resistance and DC current adjustment connections
Keysight 3458A Calibration Manual 45
Page 46
3 Adjustment Procedures
CAUTION

AC Adjustment

This adjustment is only required once every two years or whenever the 03458-66502 PC Assembly or the 03458-66503 PC Assembly has been replaced or repaired. This adjustment sets the internal crystal frequency for the frequency and period measurement functions: adjusts the attenuator and amplifier high frequency response; and adjusts the Time Interpolator timing accuracy. Following this adjustment, the internal circuits have constant gain versus frequency.
Equipment required:
– Keysight 3325A Synthesizer/Function Generator or equivalent.
– 3 V Thermal Converter, Ballantine 1395A-3 or equivalent.
– 1 V Thermal Converter, Ballantine 1395A-1 or equivalent.
– 0.5 V Thermal Converter, Ballantine 1395A-0.4 or equivalent.
– 50 W BNC cable (keep this cable as short as possible)
– 50 W resistive load (typically a 50 W carbon composition or metal film resistor).
– BNC to Banana Plug Adapter--Keysight 1251-2277 or equivalent.
In the following procedure, the output voltage of the synthesizer is adjusted with the thermal converters in-circuit. Thermal converters are typically easily damaged by vol tage overload. Use extreme care to ensure the voltage applied to the thermal converters does not exceed the thermal converter's maximum voltage rating.
Procedure
In the following procedure, steps 1 through 12 characterize the frequency flatness of the synthesizer and cabling configuration. The equipment setting determined from this characterization are then used in the remaining steps to precisely adjust the multimeter.
46 Keysight 3458A Calibration Manual
Page 47
Adjustment Procedures 3
NOTE
The voltages referenced in this procedure are 3 V, 1 V and 100 mV rms for the SCAL 10, SCAL 1, and SCAL .1 commands, respectively. If necessary, you can use any value between 3 V and 10 V rms wherever 3 V is referenced, 300 mV to 1 V rms wherever 1 V is referenced, and 30 mV to 100 mV wherever 100 mV is referenced (make sure not to exceed the vol tage rating of the thermal converters). (You still execute the SCAL 10, SCAL 1, and SCAL.1 commands regardless of the rms voltage value used). Whenever making low-level measurements, take precautions to minimize noise and interference in the test setup. Refer to Test considerations in Chapter 4 for more information.
1 Execute the ACAL AC command. Following the autocal, execute the RESET
command.
2 Set the front panel Terminals switch to Front. Set the Guard switch to the To
LO position.
3 Set the synthesizer to deliver a 3 V rms sinewave at a frequency of
100 kHz. Connect the synthesizer, the 3 V thermal converter, and the multimeter as shown in Figure 3-4. Record the exact DC voltage measured by the multimeter on Line A of the Adjustment Record.
Keysight 3458A Calibration Manual 47
Page 48
3 Adjustment Procedures
Figure 3-4 Characterizing the adjustment setup
4 Set the synthesizer to deliver a 3 V rms sinewave at a frequency of
2 MHz. Adjust the synthesizer's output voltage until the voltage displayed on the multimeter is within 0.2% of the voltage recorded on Line A. Record the synthesizer's voltage setting on Line C of the Adjustment Record.
5 Set the synthesizer to deliver a 3 V rms sinewave at a frequency of
8 MHz. Adjust the synthesizer until the voltage displayed on the multimeter is within 0.2% of the voltage recorded on Line A. Record the synthesizer's voltage setting on Line D of the Adjustment Record.
6 Set the synthesizer to deliver a 1 V rms sinewave at a frequency of
100 kHz. Replace the 3 V thermal converter with the 1 V thermal converter. Record the exact DC voltage measured by the multimeter on Line E of the Adjustment Record.
48 Keysight 3458A Calibration Manual
Page 49
Adjustment Procedures 3
7 Set the synthesizer to deliver a 1 V rms sinewave at a frequency of
8 MHz. Adjust the synthesizer until the voltage displayed on the multimeter is within 0.2% of the voltage recorded on Line E. Record the synthesizer's voltage setting on Line F of the Adjustment Record.
8 Set the synthesizer to deliver a 100 mV rms sinewave at a frequency of 100
kHz. Replace the 1 V thermal converter with the 0.5 V thermal converter. Record the exact DC voltage measured by the multimeter on Line G of the Adjustment Record.
9 Set the synthesizer to deliver a 100 mV rms sinewave at a frequency of 8 MHz.
Adjust the synthesizer until the voltage displayed on the multimeter is within
0.2% of the voltage recorded on Line G. Record the synthesizer's voltage setting on Line H of the Adjustment Record.
10 Disconnect the thermal converter and connect the synthesizer. 50 W resistive
load. and multimeter as shown in Figure 3-5.
Figure 3-5 AC adjustment connections
Keysight 3458A Calibration Manual 49
Page 50
3 Adjustment Procedures
11 Set the synthesizer to output 3 V rms at 100 kHz. Execute the
SCAL 1E5 command. The multimeter automatically performs the adjustment. When the adjustment is complete, the multimeter returns to displaying DC voltage readings.
12 Without changing the synthesizer settings, execute the SCAL 10 command as
shown on Line B of the Adjustment Record.
13 Set the synthesizer to the voltage and frequency shown on Line C of the
Adjustment Record. Execute the SCAL command as shown on Line C of the Adjustment Record.
14 Repeat step 13 for each synthesizer setting and SCAL command shown on
Lines D through H on the Adjustment Record.
15 Disconnect all equipment from the multimeter.
16 Execute the ACAL AC command.
50 Keysight 3458A Calibration Manual
Page 51
Adjustment Procedures 3
3458A ADJUSTMENT RECORD
Adjusted by _________________________ Date _________________________ 3458A serial number or other device ID number _________________________ Previous calibration number (CALNUM? command) _______________ (record this number before adjusting the multimeter)
Adjustments performed Perform an ACAL OHMS
1. _____ Front terminal offset adjustment
2. _____ Rear terminal offset adjustment
3. _____ DC gain adjustment (DCV standard uncertainty = _______________)
4. _____ Resistance and DC current adjustment (resistance standard uncertainty = _______________)
5. _____ AC adjustment
[a]
:
Multimeter reading Synthesizer setting
AC source frequency
Execute
command
Adjustment description
Line A _____________________ V 3 V 100 kHz SCAL 1E5 Frequency adjustment Line B 3 V 100 kHz SCAL 10 Low-freq. voltage reference Line C _____________________ V 2 MHz SCAL 10 Time interpolator and flatness Line D _____________________ V 8 MHz SCAL 10 Flatness adjustment Line E _____________________ V 1 V 100 kHz SCAL 1 Low-freq. voltage reference Line F _____________________ V 8 MHz SCAL 1 Flatness adjustment Line G 100 mV 100 kHz SCAL .1 Low-freq. voltage reference Line H _____________________ V _____________________ V 8 MHz SCAL .1 Flatness adjustment
Internal adjustment temperature (TEMP? command) _____ °C Calibration number (CALNUM? command) _______________ (record this number after adjusting the multimeter) Calibration secured _________________________ Unsecured _________________________
[a] Always perform the above adjustments numbered 1 through 4: adjustment number 5 is only required once every 2 years or whenever the
03458-66502 or 03458-66503 PC Assembly has been replace or repaired.
Keysight 3458A Calibration Manual 51
Page 52
3 Adjustment Procedures
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
52 Keysight 3458A Calibration Manual
Page 53
Keysight 3458A Multimeter Calibration Manual
4 Performance Verification
Tests
Introduction 54 DC Voltage Performance Tests 60 Analog AC voltage performance tests 65 DC Current Performance Tests 68 Ohms Performance Tests 72 Frequency Counter Performance Tests 76
53
Page 54
4 Performance Verification Tests
NOTE

Introduction

This chapter contains performance tests designed to verify that the 3458A multimeter is operating within the specifications listed in Appendix A. The Performance Tests are performed without access to the interior of the instrument.

Required equipment

The equipment required for the performance tests is listed below. Equipment other than that recommended can be used as long as the specifications of the substituted equipment is equivalent to that recommended.
– Fluke 5700A AC/DC Standard
– Keysight 3325A Function Generator/Frequency Synthesizer
– Transfer standard DMM (3458A Opt. 002 within 90 days of CAL)
– Low thermal short (see Figure 3-1 on page 39)
– Low thermal test leads (such as Keysight 11053A, 11174A, or 11058A)
– Shielded test leads (such as Keysight 11000-60001)
To have your transfer standard 3458A Opt. 002 calibrated to 90 day specifications, contact your Keysight Technologies sales and service office.

Test card

Results of the performance tests may be tabulated on the appropriate Test Card located at the end of the test procedures. Make copies of the Test Cards for performance test tabulations and retain the originals to copy for use in future performance testing. The Test Cards list all of the tested functions and the acceptable limits for the test results.
54 Keysight 3458A Calibration Manual
Page 55

Calibration cycle

The frequency of performance verification depends on the instrument's usage and the environmental operating conditions. To maintain 24 hour or 90-day specifications, the instrument should be checked at these intervals by a metrology lab with test capability for these accuracies. For normal operation, it is recommended you perform performance verification every year.

Test considerations

This section discusses many of the major problems associated with low-level measurements. Many of the measurements in this manual fall into this category. It is beyond the scope of this manual to go into great detail on this subject. For more information, refer to a textbook dealing with standard metrology practices.
– Test leads: Using the proper test leads is critical for low-level measurements.
We recommend using cable or other high impedance, low dielectric absorption cable for all measurement
– Connections. It is important to periodically clean all connection points
(including the multimeter terminals) using a cotton swab dipped in alcohol.
– Noise Rejection: For DC voltage, DC current, and resistance measurements,
the multimeter achieves normal mode noise rejection (NMR) A/D converter's reference frequency (typically the same as the power line frequency) when the integration time is 1 power line cycles. You can specify integration time in terms of power line cycles (PLCs) using the NPLC command. For maximum NMR of 80dB, set the power line cycles to 1000 (NPLC 1000 command).
– Guarding: Whenever possible, make measurements with the multimeter's
Guard terminal connected to the low side of the measurement source and the Guard switch set to the Open position (guarded measurements). This provides the maximum effective common mode rejection (ECMR).
Performance Verification Tests 4
[1]
for noise at the
– Thermoelectric Voltages (Thermal EMF): This is a common source of errors in
low-level measurements. Thermal EMF occurs when conductors of dissimilar metals are connected together or when different parts of the circuit being measured are at different temperatures. Thermal EMF can become severe in high-temperature environments. To minimize thermal EMF, minimize the
[1] Normal mode noise rejection is the multimeter’s ability to reject noise at the power line frequency
from DC voltage, DC current, or resistance measurements.
Keysight 3458A Calibration Manual 55
Page 56
4 Performance Verification Tests
number of connections: use the same type of metal for all connections; minimize the temperature variations across the measurement wiring: try to keep the multimeter and the wiring at the same temperature: and avoid high-temperature environments whenever possible.
– Electromagnetic Interference (EMI): This type of interference is generally
caused by magnetic fields or high levels of radio frequency (RF) energy. Magnetic fields can surround all types of equipment operating off of AC line power, especially electric motors. RF energy from nearby radio or television stations or communications equipment can also be a problem. Use shielded wiring whenever the measurement setup is in the presence of high EMI. If possible, move farther away or turn off sources of high EMI. It may be necessary to test in a shielded room.
– Ground Loops: Ground loops arise when the multimeter and the circuit under
test are grounded at physically different points. A typical example of this is when a number of instruments are plugged into a power strip in an equipment rack. If there is a potential difference between the ground points, a current will flow through this ground loop. This generates an unwanted voltage in series with the circuit under test. To eliminate ground loops, ground all equipment/ circuits at the same physical point.
– Internal Temperature: The internal temperature of the 3458A under test must
be within 5 temperature is not within 5
clean the filter. Also, make sure that you adjust the operating environment such that the ambient temperature is at or very near 25 best results if you maintain your environment close to 25 verify performance when the temperature is not within 5 recalculate all test limits based on the temperature variation beyond the 5 limitation. The published test limits were calculated without additional
temperature coefficient errors added.
°C of its temperature when last adjusted. If the multimeter's
°C first check the multimeter's fan operation and
°C. You will achieve the
°C. If you choose to
°C, you must
°C
56 Keysight 3458A Calibration Manual
Page 57

General test procedure

The following performance tests utilize a transfer standard DMM to precisely measure the verification source. The transfer standard DMM recommended is an 3458A option 002 (high stability) that is within a 90-day calibration. The verification source is first measured by the transfer standard DMM and then connected to the unit under test. The general test procedure is as follows:
A. Performed one time prior to testing (preliminary steps)
1 Verify that the Verification Source is properly warmed up.
2 The 3458A requires a 4 hour warm-up period. Verify that the transfer standard
DMM and the 3458A unit under test (UUT) are properly warmed up.
3 The internal temperature of the 3458A under test must be within
5 degrees C of its temperature when last adjusted (CAL 0, CAL 10, and CAL 10K). These temperatures can be determined by executing the commands CAL? 58, CAL? 59, CAL? 60.
4 If the instrument self test has not been run, verify all inputs are disconnected
and execute the TEST function. The display must read “SELF TEST PASSED”.
Performance Verification Tests 4
B. Repeated for each function and range tested
5 Execute the ACAL command for the function being tested on both the transfer
standard and the unit under test (UUT).
6 Configure the transfer DMM as specified in each test.
7 Configure the DMM under test as specified in each test.
8 Connect the Verification source to the transfer standard DMM and determine
the source output (see Figure 4-1 (A)). Record this value on the Test Card under “Transfer Standard Reading”.
9 Disconnect the Verification Source from the transfer standard DMM and
connect it to the 3458A under test (see Figure 4-1 (B)). Record this value on the Test Card under “Unit Under Test Reading”.
10 Calculate the difference between the transfer standard DMM reading and the
UUT reading. Record the absolute value (ignore sign) of the difference on the Test Card under “Difference”.
Keysight 3458A Calibration Manual 57
Page 58
4 Performance Verification Tests
11 Compare this difference to the allowable difference specified on the Test Card.
If less than the specified difference, note that the test passed. If greater than the specified difference, note that the test failed.
58 Keysight 3458A Calibration Manual
Page 59
Performance Verification Tests 4
Figure 4-1 General test procedure
Keysight 3458A Calibration Manual 59
Page 60
4 Performance Verification Tests

DC Voltage Performance Tests

Required equipment

The following equipment or its equivalent is required for these performance tests.
– Stable DC voltage source (Fluke 5700A or equivalent)
– Transfer standard DMM (3458A Opt. 002 within 90 days of CAL)
– Low thermal short (copper wire)
– Low thermal test leads (such as Keysight 11053A, 11174A, 11058A)

Preliminary steps

1 Verify that the DC source is properly warmed up.
2 The 3458A requires a 4-hour warm-up period. If this has not been done, turn
the instrument ON and allow it to warm up before proceeding.
3 The internal temperature of the 3458A under test must be within
5 degrees C of its temperature when last adjusted. Use the TEMP? command to obtain the current internal temperature and compare it to the calibration temperature obtained by executing the command CAL? 59. Record the temperatures obtained on the DC VOLTAGE TESTS test card.
4 If the instrument self test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
5 Execute the ACAL DCV command on both the transfer standard DMM and the
UUT using the front panel “Auto Cal” key and scroll keys. This auto calibration will take approximately two minutes to complete.
6 Configure the transfer standard DMM as follows:
–DCV
–NDlG 8
– NPLC 100
–Trig SGL
7 Configure the DMM under test as follows:
60 Keysight 3458A Calibration Manual
Page 61
–DCV
–NDlG 8
– NPLC 100
–Trig SGL

DC voltage function offset test

The following procedure tests the offset voltage specification with the input terminals shorted. A low-thermal short must be used to minimize thermally induced errors. Also, you must allow five minutes before making the first measurement to allow for thermal stabilization of the range relays.
1 Connect a low thermal short across the front panel HI and LO input terminals
of the DMM under test (see Figure 3-1 on page 39).
2 Set the range of the 3458A under test as specified in Ta ble 4-1.
3 Let the instrument sit for five minutes before taking the first reading to allow
the range relay and short to thermally stabilize. NOTE: The thermal stabilization achieved for the 100 mV range is present for the 1V and 10V ranges since these ranges use the same relays. The range relays are opened for the 100V and 1000V ranges and therefore, have no thermal impact on the measurement.
4 Execute Trig and record the offset reading (absolute value) for each range
listed in Table 4-1 and on the DC VOLTAGE TESTS Test Card provided at the end of this chapter.
Performance Verification Tests 4
Table 4-1 Offset performance tests
Offset test number DMM range
1 100 mV
21 V
3 10 V
4 100 V
5 1000 V
Keysight 3458A Calibration Manual 61
Page 62
4 Performance Verification Tests
5 If any of the offset readings are greater than the limits specified on the DC
VOLTAGE TESTS Test Card, the instrument should be adjusted. Refer to Chapter 3, “Adjustment Procedures”, to make adjustments.
6 Remove the short from the front panel terminals.

DC voltage function gain test

The following is a step-by-step procedure for all test points that verify gain of the DC voltage function. The procedure requires alternately connecting the transfer standard DMM and then the 3458A under test to the DC voltage source as described in the general test description.
1 Set the output of the DC voltage source to standby/off and short the HI and
LO output terminals of the source using a shorting strap.
2 Connect the leads of a shielded low-thermal EMF cable from the source output
terminals to the Input HI and LO terminals of the transfer standard DMM.
3 Connect the Guard terminal of the voltage source to the guard terminal of the
DMM.
4 Set the range of the DMM to 100 mV. Wait five minutes for the DMM to
thermally stabilize.
5 Set MATH NULL on the transfer standard DMM.
6 Execute Trig SGL to trigger the NULL reading.
7 Remove the EMF cable leads from the DMM and connect them to the HI and
LO Input terminals 3458A under test. Connect the Guard terminal of the voltage source to the guard terminal of the 3458A under test.
8 Set the range of the 3458A under test to 100 mV. Wait five minutes for the
3458A under test to thermally stabilize.
9 Set MATH NULL on the 3458A under test and execute Trig SGL to trigger the
NULL reading
10 REMOVE THE SHORTING STRAP FROM THE VOLTAGE SOURCE HI and LO
OUTPUT TERMINALS.
62 Keysight 3458A Calibration Manual
Page 63
Performance Verification Tests 4
NOTE
Connect the output of the voltage source to the HI and LO Input terminals of the transfer standard DMM. With the DMM range set to 100 mV, set the output of the voltage source to 100 mV. Wait five minutes for the transfer standard DMM to thermally stabilize.
The thermal stabilization achieved for the 100 mV range is present for the 1V and 10V ranges since these ranges use the same relays. The range relays are opened for the 100V and 1000V ranges and therefore, have no thermal impact on the measurement.
11 Execute Trig SGL and read the output of the voltage source as measured with
the transfer standard DMM and record this reading in the “Transfer standard reading” column of the DC VOLTAGE TESTS test card.
12 Move the connection from the transfer standard DMM HI and LO Input
terminals to the HI and LO Input terminals of the 3458A under test.
13 Execute Trig SGL and read the value as measured with the 3458A under test
and record this value in the “Unit under test reading” column of the DC voltage Test Record.
14 Repeat steps 11 through 14 for each of the remaining DC voltage test points
specified in Table 4-2.
Table 4-2 DCV gain performance tests
DC gain test number DMM range Source output
1 100 mV 100 mV
21 V 1 V
310 V 1 V
4 10 V -1 V
5 10 V -10 V
610 V10 V
7 100 V 100 V
[a]
8
[a] NOTE: After completing test 8, decrease the 1000 V verification source output to 0 V before disconnecting.
1000 V 1000 V
Keysight 3458A Calibration Manual 63
Page 64
4 Performance Verification Tests
15 After all DC gain tests have been performed, calculate and record in the
column provided. The difference (absolute value) between the transfer standard DMM reading and the unit under test reading for each of the test points.
16 If any of the differences calculated are beyond the specified limits, refer to
Chapter 3, “Adjustment Procedures”, to make adjustments.
64 Keysight 3458A Calibration Manual
Page 65

Analog AC voltage performance tests

Required equipment

The following list of equipment is required to test the analog AC performance of the 3458A.
– Stable AC voltage source (Fluke 5700A or equivalent).
– Transfer Standard DMM (3458A Opt. 002 within 90 days of Cal.)
– Shielded test leads terminated with dual banana plugs (such as
Keysight 11000-60001).

Preliminary steps

1 Make certain that the AC source is properly warmed up.
2 The 3458A requires a 4 hour warm up period. If this has not been done, turn
the instrument ON and allow it to warm up.
3 Execute the ACAL AC function on both the transfer standard DMM and the
UUT. This auto calibration will take approximately 1 minute to complete.
4 If the instrument Self Test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
5 Configure the transfer standard DMM as follows:
Performance Verification Tests 4
–ACV
–SETACV SYNC
– ACBAND 10,2E6
–RANGE10
–RES .001
–TRIG SGL
– LFILTER ON
6 Configure the DMM under test as follows:
–ACV
Keysight 3458A Calibration Manual 65
Page 66
4 Performance Verification Tests
–SETACV ANA
– ACBAND 10,2E6
–RANGE10
–RES .01
–TRIG SGL
–LFILTER ON

AC voltage test procedure

The following is a step-by-step procedure for all test points in the AC performance verification section. The procedure requires alternately connecting the transfer standard DMM and then the 3458A under test to the AC source. Because of this and because the accuracy of AC coupled measurements does not suffer due to small thermal induced offsets, the test connection can be made using shielded test leads terminated with dual banana plugs. Refer to the general test procedure for test connections.
1 Connect the AC voltage source to the transfer standard DMM.
2 Set the range of the transfer standard DMM as specified in Table 4-3.
3 Set the range of the 3458A under test as specified in Table 4-3.
4 Set the AC source to the voltage level and frequency specified in Table 4-3
5 Execute Trig SGL and read the output of the AC source as measured with the
transfer standard DMM and record this reading in the “Transfer standard reading” column of the AC VOLTAGE TESTS Test Card.
6 Remove the connection from the transfer standard DMM to the AC source.
7 Connect the 3458A under test to the AC source.
8 Execute Trig SGL and read the value as measured with the 3458A under test
and record this value in the “Unit under test reading” column of the AC
VOLTAGE TESTS Test Card.
9 Repeat steps 1 through 8 for each of the remaining AC voltage test points as
specified in Table 4-3.
66 Keysight 3458A Calibration Manual
Page 67
Performance Verification Tests 4
Table 4-3 AC performance tests
AC test number DMM range Source level Source frequency
1 100 mV 100 mV 1 kHz
21 V1 V1 kHz
3 10 V 1 V 1 kHz
4 10 V 10 V 20 Hz
5 10 V 10 V 1 kHz
6 10 V 10 V 20 kHz
7 10 V 10 V 100 kHz
8 10 V 10 V 1 MHz
9 100 V 100 V 1 kHz
[a]
10
[a] NOTE: After completing test 10, reduce the ACV standard voltage to 0 V before disconnecting.
1000 V 700 V 1 kHz
10 After all AC voltage tests have been performed, calculate and record in the
column provided, the difference between the transfer standard DMM reading and the unit under test reading for each of the test points.
11 If any of the differences calculated are greater than the specified limits, refer to
Chapter 3, “Adjustment Procedures”, to make adjustments.
Keysight 3458A Calibration Manual 67
Page 68
4 Performance Verification Tests

DC Current Performance Tests

Required equipment

The following equipment or its equivalent is required for these performance tests.
– Stable DC current source (Fluke 5700A or equivalent)
– Transfer standard DMM (3458A Opt. 002 within 90 days of CAL)
– Low thermal test leads (such as Keysight 11053A, 11174A, or 11058A)

Preliminary steps

1 Verify that the DC current source is properly warmed up.
2 The 3458A requires a 4 hour warm-up period. If this has not been done, turn
the instrument ON and allow it to warm up before proceeding.
3 The internal temperature of the 3458A under test must be within
5 degrees C of its temperature when last adjusted. The current internal temperature is obtained by executing TEMP?. Compare this temperature to the calibration temperature obtained by executing the command CAL? 60. Record these temperatures on the DC CURRENT TESTS Test Card.
4 If the instrument self test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
5 Execute the ACAL OHMS function on both the transfer standard DMM and the
UUT. This auto calibration will take approximately ten minutes to complete.
6 Configure the transfer standard DMM as follows:
–DCI
–NDIG 8
– NPLC 100
–Trig SGL
7 Configure the DMM under test as follows:
–DCI
–NDIG 8
68 Keysight 3458A Calibration Manual
Page 69
– NPLC 100
–Trig SGL

DC current function offset test

The following procedure tests the DC current offset specifications with the input terminals open.
1 Set the 3458A under test to the DC Current Function (DCI).
2 Set the range of the 3458A under test as specified in Ta ble 4-4.
3 Let the instrument sit for 5 minutes to allow the range relays to thermally
stabilize.
4 Execute Trig and record the absolute value of the offset reading of each range
listed in Table 4-4 on the DC CURRENT TESTS Test Card provided at the end of this section.
Table 4-4 Current offset performance tests
Offset test number DMM range
1 100 µA
Performance Verification Tests 4
21 mA
310 mA
4100 mA
51 A
5 If the offset tests are out of specification, perform another ACAL before
performing step 6 below.
6 If any of the offset readings are beyond the limits specified in the Test Record,
the instrument should be adjusted. Refer to Chapter 3, “Adjustment
Procedures” to make adjustments.
Keysight 3458A Calibration Manual 69
Page 70
4 Performance Verification Tests

DC current function gain test

The following is a step-by-step procedure for all test points that performance verify gain of the DC current function. The procedure requires alternately connecting the transfer standard DMM and then the 3458A under test to the DC verification source as described in the section titled General test procedure.
1 Connect the DC current source to the transfer standard DMM I and LO input
terminals using low thermal test leads.
2 Set the range of the transfer standard DMM as specified in Table 4-5.
3 Set the range of the 3458A under test as specified in Table 4-5.
4 Set the DC source to the current level specified in Table 4-5.
5 Execute Trig SGL and read the output of the DC current source as measured
with the transfer standard DMM and record this reading in the “Transfer standard reading” column of the DC CURRENT TESTS Test Card.
6 Remove the connection from the transfer standard DMM to the DC current
source.
7 Connect the DC current source to the 3458A under test HI and LO input
terminals.
8 Execute Trig and read the value as measured with the 3458A under test and
record this value in the “Unit under test reading” column of the DC CURRENT
TESTS Test Card.
9 Repeat steps 1 through 8 for each of the remaining DC current test points as
specified in Table 4-5.
Tab le 4-5 DCI gain performance tests
DCI gain test number Source and DMM range Source output
1 100 µA 100 µA
2 1 mA 1 mA
3 10 mA 10 mA
4 100 mA 100 mA
51 A1 A
70 Keysight 3458A Calibration Manual
Page 71
Performance Verification Tests 4
10 After all DC current gain tests have been performed, calculate and record in
the column provided, the difference (absolute value) between the transfer standard DMM reading and the unit under test reading for each test point.
11 If any of the differences calculated are beyond the specified limits, refer to
Chapter 3, “Adjustment Procedures”, to make adjustments.
12 Reduce the output of the DC Current Source and disconnect it from the 3458A
input terminals.
Keysight 3458A Calibration Manual 71
Page 72
4 Performance Verification Tests

Ohms Performance Tests

Required equipment

The following list of equipment is required to test the ohms performance of the 3458A.
– Stable resistance standard (Fluke 5700A or equivalent)
– Transfer standard DMM (3458A Opt. 002 within 90 days of CAL)
– Low thermal short (copper wire)
– Low thermal test leads (such as Keysight 11053A, 11174A, or 11058A)

Preliminary steps

1 Verify that the resistance standard is properly warmed up.
2 The 3458A requires a 4-hour warm-up period. If this has not been done, turn
the instrument ON and allow it to warm up before proceeding.
3 The internal temperature of the 3458A under test must be within 5 degrees C
of its temperature when last ohms adjusted. The current internal temperature can be obtained by executing TEMP?. Compare this temperature to adjustment temperature obtained by executing the command CAL? 60 and record both temperatures on the OHMS TESTS Test Card.
4 If the instrument self test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
5 If you have just performed DCI tests, you have done an ACAL OHMS which
takes approximately ten minutes to complete. Compare the TEMP? temperatures recorded on the DC CURRENT TESTS and OHMS TESTS Test
Cards. If they differ by more than 1 have not been done previously, execute ACAL OHMS.
6 Configure the transfer standard DMM as follows:
–OHMF
–NDIG 8
– NPLC 100
72 Keysight 3458A Calibration Manual
°C. execute ACAL OHMS again. If DCI tests
Page 73
–OCOMP ON
–Trig SGL
7 Configure the DMM under test as follows:
–OHM
–NDIG 8
– NPLC 100
–OCOMP ON
–Trig SGL

2-wire ohms function offset test

The following procedure performance verifies the front terminal ohms offset.
1 Connect a low thermal short across the front panel HI and LO input terminals
of the 3458A under test as shown in Figure 3-1 on page 39.
2 Set the 3458A under test to the 10 Ω range. Allow 5 minutes for the range
relays to thermally stabilize.
3 Execute Trig and use the OHMS TESTS Test Card to record the offset reading.
4 Remove the short from the front panel input terminals.
Performance Verification Tests 4

4-wire ohms function offset test (rear terminals)

This procedure performance verifies the rear terminal ohms offset.
1 Connect a low thermal short across the rear terminals of the 3458A as shown
for the front terminals in Figure 3-1 on page 39.
2 On the 3458A under test, select 4-wire ohms and the 10 W range by executing
OHMF, 10.
3 Execute Trig and use the OHMS TESTS Test Card to record the offset reading.
4 Remove the short from the rear panel input terminals.
Keysight 3458A Calibration Manual 73
Page 74
4 Performance Verification Tests

4-wire ohms function gain test

The following is a step-by-step procedure for all test points that performance verify gain of the ohms function. The procedure requires alternately connecting the transfer standard DMM and then the 3458A under test to the resistance verification source as described in the section titled General test procedure.
1 Connect the resistance standard to the transfer standard DMM 4-wire ohms
front input terminals.
2 Set the range of the transfer standard DMM as specified in Table 4-6.
3 Set the range of the 3458A under test as specified in Table 4-6.
4 Set the resistance standard to the ohms level specified in Table 4-6.
5 Execute Trig and read the output of the resistance standard as measured with
the transfer standard DMM and record this reading in the “Transfer standard reading” column of the OHMS TESTS Test Card.
6 Remove the connection from the transfer standard DMM to the resistance
standard.
7 Connect the resistance standard to the front panel 4-wire ohms input
terminals of the 3458A under test.
8 Execute Trig two times and read the value as measured with the
3458A under test and record this value in the “Unit under test reading” column of the OHMS TESTS Test Card.
9 Repeat steps 1 through 8 for each of the remaining resistance test points as
specified in Table 4-6.
Tab le 4-6 OHMF gain performance tests
OHMF gain test number Source and DMM range Source output
1 10 W 10 W
2 100 W 100 W
31 kW 1 kW
4 10 kW 10 kW
5 100 kW 100 kW
74 Keysight 3458A Calibration Manual
Page 75
Performance Verification Tests 4
Table 4-6 OHMF gain performance tests
OHMF gain test number Source and DMM range Source output
61 MW 1 MW
[a]
7
[a] NOTE: At 10 MW, leakage current from the Sense leads introduce additional uncertainty in the measurement. This
uncertainty, however, is accounted for in the published performance test limit. For best accuracy and consistency across calibration sites and environments, Test 7 can be performed using a 2-wire ohms measurement. The performance test limit remains the same for both 4-wire and 2-wire measurements of 10 MW.
10 MW 10 MW
10 After all OHMF gain tests have been performed, calculate and record in the
column provided, the difference (absolute value) between the transfer standard DMM reading and the unit under test reading for each of the test points.
11 If any of the differences calculated are beyond the specified limits, refer to
Chapter 3, “Adjustment Procedures” to make adjustments.
12 Disconnect the resistance standard from the 3458A input terminals.
Keysight 3458A Calibration Manual 75
Page 76
4 Performance Verification Tests

Frequency Counter Performance Tests

Required equipment

The following equipment is required for testing the frequency counter performance of the 3458A.
– Stable frequency source (Keysight 3325A Frequency Synthesizer or equivalent)
– Shielded test leads, BNC to dual banana plug (such as Keysight 11001-60001)

Preliminary steps

1 Verify that the frequency source is properly warmed up.
2 The 3458A requires a 4-hour warm-up period. If this has not been done, turn
the instrument ON and allow it to warm up before proceeding.
3 If the instrument self test has not been run, make certain all inputs are
disconnected and execute the TEST function. The display must read “SELF TEST PASSED”.
4 Configure the DMM under test as follows:
–FREQ
–Trig SGL
–FSOURCE ACDCV
– LEVEL 0,DC

Frequency counter accuracy test

1 Execute FSOURCE ACDCV (specifies the type of signal to be used as the input
signal for frequency measurement).
2 Set the Frequency Standard to output a 1 volt p-p, 1 Hz sine-wave. Record the
exact Frequency Standard Value on the FREQUENCY TESTS Test Card. Connect the output of the Frequency Standard to the HI and LO input terminals of the 3458A under test.
3 Execute Trig and record the Unit Under Test Reading on the FREQUENCY
TESTS Test Card.
76 Keysight 3458A Calibration Manual
Page 77
Performance Verification Tests 4
4 Subtract the 1 Hz Unit Under Test Readings from the 1 Hz Frequency Standard
Value. Record the difference on the FREQUENCY TESTS Test Card.
5 Change the Frequency Standard to 10 MHz and record the exact Frequency
Standard Value on the FREQUENCY TESTS Test Card. Execute Trig, and record the Unit Under Test Reading on the FREQUENCY TESTS Test Card.
6 Subtract the 10 MHz Unit Under Reading from the 10 MHz Frequency
Standard Value. Record the difference on the FREQUENCY TESTS Test Card.
7 If either of the differences are beyond the limits specified, the instrument
should be adjusted. See Chapter 3, “Adjustment Procedures” to make adjustments.
Keysight 3458A Calibration Manual 77
Page 78
4 Performance Verification Tests
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
DC VOLTAGE TESTS
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL DCV
Test #
3458A
input
3458A
range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
Offset test (NOTE: Math null is disabled)
1 Short 100 mV N/A __________ N/A 000.00106 mV 000.00106 mV __________ __________
2 Short 1 V N/A __________ N/A 0.00000106 V 0.00000106 V __________ __________
3 Short 10 V N/A __________ N/A 00.0000023 V 00.0000023 V __________ __________
4 Short 100 V N/A __________ N/A 000.000036 V 000.000036 V __________ __________
5 Short 1000 V N/A __________ N/A 0000.00010 V 0000.00010 V __________ __________
Gain test
1 100 mV 100 mV __________ __________ __________ 000.00212 mV 000.00188 mV __________ __________
2 1 V 1 V __________ __________ __________ 0.00000998 V 0.00000740 V __________ __________
3 1 V 10 V __________ __________ __________ 00.0000111 V 00.0000085 V __________ __________
4 –1 V 10 V __________ __________ __________ 00.0000111 V 00.0000085 V __________ __________
5 –10 V 10 V __________ __________ __________ 00.0000892 V 00.0000624 V __________ __________
6 10 V 10 V __________ __________ __________ 00.0000892 V 00.0000624 V __________ __________
7 100 V 100 V __________ __________ __________ 000.001114 V 000.000853 V __________ __________
8 1000 V 1000 V __________ __________ __________ 0000.02396 V 0000.01934 V __________ __________
78 Keysight 3458A Calibration Manual
Page 79
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
AC VOLTAGE TESTS
Perform an ACAL AC
Test # 3458A input 3458A range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
1 100 mV, 1kHz 100 mV __________ __________ __________ 000.0250 mV 000.0250 mV __________ __________
2 1 V, 1 kHz 1 V __________ __________ __________ 0.000250 V 0.000250 V __________ __________
3 1 V, 1 kHz 10 V __________ __________ __________ 00.00096 V 00.00096 V __________ __________
4 10 V, 20 Hz 10 V __________ __________ __________ 00.01338 V 00.01338 V __________ __________
5 10 V, 1 kHz 10 V __________ __________ __________ 00.00250 V 00.00250 V __________ __________
6 10 V, 20 kHz 10 V __________ __________ __________ 00.00272 V 00.00272 V __________ __________
7 10 V, 100 kHz 10 V __________ __________ __________ 00.05372 V 00.05372 V __________ __________
8 10 V, 1 MHz 10 V __________ __________ __________ 00.55450 V 00.55450 V __________ __________
9 100 V, 1 kHz 100 V __________ __________ __________ 000.0364 V 000.0364 V __________ __________
10 700 V, 1 kHz 1000 V __________ __________ __________ 0000.544 V 0000.544 V __________ __________
Performance Verification Tests 4
Keysight 3458A Calibration Manual 79
Page 80
4 Performance Verification Tests
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
DC CURRENT TESTS
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL OHMS
Test #
3458A
input
3458A
range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
Offset test (NOTE: Math null is disabled)
1 Open 100 µA N/A __________ N/A 000.00095 µA 000.00095 µA __________ __________
2 Open 1 mA N/A __________ N/A 0.0000065 mA 0.0000065 mA __________ __________
3 Open 10 mA N/A __________ N/A 00.000065 mA 00.000065 mA __________ __________
4 Open 100 mA N/A __________ N/A 000.00065 mA 000.00065 mA __________ __________
5 Open 1 A N/A __________ N/A 0.0000115 A 0.0000115 A __________ __________
Gain test
1 100 µA 100 µA __________ __________ __________ 000.00356 µA 000.00356 µA __________ __________
2 1 mA 1 mA __________ __________ __________ 0.0000323 mA 0.0000323 mA __________ __________
3 10 mA 10 mA __________ __________ __________ 00.000323 mA 00.000323 mA __________ __________
4 100 mA 100 mA __________ __________ __________ 000.00489 mA 000.00489 mA __________ __________
5 1 A 1 A __________ __________ __________ 0.0001349 A 0.0001349 A __________ __________
80 Keysight 3458A Calibration Manual
Page 81
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
OHMS TESTS
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL OHMS
Test #
3458A
input
3458A
range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
2-wire function offset test
1 Short 10 W N/A __________ N/A 00.25007 W 00.25007 W __________ __________
4-wire function offset test (rear terminals)
1 Short 10 W N/A __________ N/A 00.00007 W 00.00007 W __________ __________
4-wire function gain test
110 W 10 W __________ __________ __________ 00.00028 W 00.00028 W __________ __________
2 100 W 100 W __________ __________ __________ 000.00231 W 000.00231 W __________ __________
31 kW 1 kW __________ __________ __________ 0.0000142 kW 0.0000142 kW __________ __________
410 kW 10 kW __________ __________ __________ 0.0000142 kW 0.0000142 kW __________ __________
5 100 kW 100 kW __________ __________ __________ 0.0000142 kW 0.0000142 kW __________ __________
61 MW 1 MW __________ __________ __________ 0.0000209 MW 0.0000209 MW __________ __________
710 MW 10 MW __________ __________ __________ 00.000703 MW 00.000703 MW __________ __________
7
[a]
10 MW 10 MW __________ __________ __________ 00.000703 MW 00.000703 MW __________ __________
Performance Verification Tests 4
Keysight 3458A Calibration Manual 81
Page 82
4 Performance Verification Tests
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
FREQUENCY TESTS
Test #
3458A
input
3458A
range
Transfer
standard
read ing
Unit
under test
reading
Difference
Limit
(Std)
Limit
(Opt 002)
Pass Fail
1 1 Hz N/A __________ __________ __________ ±0.000500 Hz ±0.000500 Hz __________ __________
2 10 MHz N/A __________ __________ __________ ±00.00100 MHz ±00.00100 MHz __________ __________
82 Keysight 3458A Calibration Manual
Page 83
Keysight 3458A Multimeter Calibration Manual

5 Command Summary

Introduction 84 ACAL 86 CAL 88 CAL? 90 CALNUM? 104 CALSTR 105 REV? 107 SCAL 108 SECURE 111 TEMP? 113 TEST 114
83
Page 84
5 Command Summary

Introduction

This section provides an alphabetical summary of commands that are used in calibrating the 3458A (adjustments or performance verification). Detailed command reference pages for each command are also included in this chapter.
ACAL
CAL
CAL?
CALNUM?
CALSTR
REV?
SCAL
Autocal. Instructs the multimeter to perform one or all of its automatic calibrations.
Calibration. Calibrates the internal 7 V reference to an external 10V standard (CAL 10) followed by an ACAL DCV. It also calibrates the internal 40 kW reference to an external 10 kW standard (CAL 10E3) followed by an ACAL OHMS. Offset for the front and rear terminals are also calculated (CAL 0).
Calibration query. Returns one of four values for the calibration constant specified; the initial (nominal) value, low limit, high limit, or actual value of the specified constant.
Calibration number query. Returns a decimal number indicating the number of times the multimeter has be adjusted.
Calibration string (remote only). Stores a string in the multimeter's nonvolatile calibration RAM. Typical uses for this string include the date or place of adjustment/verification, technician's name, or the scheduled date for the next adjustment.
Revision query. Returns two numbers separated by a comma. The first number is the multimeter's outguard firmware revision. The second number is the inguard firmware revision.
Service calibration. Adjusts the AC section of the instrument. Calculates the corrections to accurately measure frequency and calibrates the ac ranges.
SECURE
84 Keysight 3458A Calibration Manual
Security code. Allows the person responsible for calibration to enter a security code to prevent accidental or unauthorized adjustment or autocalibration.
Page 85
Command Summary 5
TEMP?
TEST
Temperature query. Returns the multimeter's internal temperature in degrees Centigrade.
Self-test. Causes the multimeter to perform a series of internal self-tests. If all constants are within their lower and upper limits, the self-test passes.
Keysight 3458A Calibration Manual 85
Page 86
5 Command Summary

ACAL

Description
Autocal. Instructs the multimeter to perform one or all of its automatic calibrations.
Syntax
ACAL [type][,security_code]
type
The type parameter choices are:
type Parameter
ALL 0 Performs the DCV, AC, and OHMS autocals
DCV 1 DC voltage gain and offset (see first Remark)
AC 2 ACV flatness, gain, and offset (see second Remark)
OHMS 4 OHMS gain and offset (see third Remark)
Numeric query
equivalent
Description
security_code
When autocal is secured, you must enter the correct security code to perform an autocal (when shipped from the factory, autocal is secured with the security code
3458). When autocal is not secured, no security code is required. Refer to the SECURE command for more information on the security code and how to secure or unsecure autocal.
Remarks
– Since the DCV autocal applies to all measurement functions, you should
perform it before performing the AC or OHMS autocal. When ACAL ALL is specified, the DCV autocal is performed prior to the other autocals.
– The AC autocal performs specific enhancements for ACV or ACDCV (all
measurement methods), ACI or ACDCI, DSAC, DSDC, SSAC, SSDC, FREQ, and PER measurements.
86 Keysight 3458A Calibration Manual
Page 87
Command Summary 5
– The OHMS autocal performs specific enhancements for 2- or 4-wire ohms,
DCI, and ACI measurements.
– Always disconnect any AC input signals before you perform an autocal. If you
leave an input signal connected to the multimeter, it may adversely affect the autocal.
– The autocal constants are stored in continuous memory (they remain intact
when power is removed). You do not need to perform autocal simply because power has been cycled.
– The approximate time required to perform each autocal routine is:
ALL: 14 minutes
DCV: 2 minutes and 20 seconds
AC: 2 minutes and 20 second s
OHMS: 10 minutes
– If power is turned off or the Reset button is pushed during an ACAL, an error is
generated. You must perform an ACAL ALL to recalculate new calibration constants.
Related command s: CAL, SCAL, SECURE
Example
OUTPUT 722;"ACAL ALL,3458" !RUNS ALL AUTOCALS, SECURITY CODE IS 3458! (FACTORY SECURITY CODE SETTING)
Keysight 3458A Calibration Manual 87
Page 88
5 Command Summary
CAL
Description
Calibration. Calibrates the internal 7 V reference to an external 10V standard (CAL10) and does the equivalent of ACAL DCV. Also calibrates the internal 40 K reference to an external 10 K standard (CAL 10E3) and does the equivalent of ACAL OHMS. Alternate CAL standard values can be used as described in the first remark. It also calculates the offset for the front and rear terminals (CAL 0).
Syntax
CAL value [,security _ code]
value
Specifies the value of the adjustment source that will be used to adjust the multimeter. For highest accuracy. 10 V and 10 K ohm standards are recommended and the value sent must be the exact output value of the adjustment source. If the 10 V source actually outputs 10.0001, then specify a value of 10.0001 in the CAL command.
security_code
When a security code is set to a number other than 0 by the SECURE command, you must enter the correct security code to perform a CAL. If CAL is not secured (security code = 0), no security code is required to execute CAL. Refer to the SECURE command for more information on the security code and how to secure the calibration of the 3458A.
Remarks
– For highest accuracy, the value sent with the CAL command must exactly
equal the actual output value of the adjustment source. It is recommended that 10V be used for CAL 10 and 10 K ohms be used for CAL 10E3. NOTE: Any
standard value between 1 V and 12 V or 1 k less than 10 V or less than 10 k multimeter's accuracy specifications. For example, a 1 V DC standard can be used instead of 10 V (you would execute CAL 1.0000). A 1 k used instead of 10 k accuracy specifications of the instrument.
88 Keysight 3458A Calibration Manual
W
(you would execute CAL 1E3). Each case degrades the
W
will introduce additional uncertainty to the
W
and 12 kW can be used. A value
W
standard can
Page 89
Command Summary 5
– For highest accuracy when performing a CAL 0, a four-point short must be
used. Also, CAL 0 must be performed twice, once for the front terminals and again for the rear terminals. You must manually switch the terminals to be calibrated using the front panel switch.
– It is recommended that the OCOMP command be executed prior to adjusting
with the 10 K source and OCOMP be set to ON. This will account for any thermals and result in a more accurate adjustment.
Related command s: ACAL, SCAL, SECURE
Example
OUTPUT 722; "CAL 10.0011" !DCV ADJUSTMENT SOURCE = 10.0011 VOLTS
OUTPUT 722;"OCOMP ON"
OUTPUT 722:"CAL 10000.001" !RESISTANCE ADJUSTMENT SOURCE = 10000.001 OHMS
Keysight 3458A Calibration Manual 89
Page 90
5 Command Summary

CAL?

Description
Calibration query. Returns a string containing one of four values for the calibration constant specified; the initial (nominal) value, low limit, high limit, or actual value of the specified constant. The returned string also contains a description of the constant. This command is in the full command menu; it is not in the short command menu.
Syntax
CAL? const_id [,cal_item]
cal_item
Specifies which of the four calibration constant values is to be returned. The cal_item parameter choices are:
cal_item Description
0 Initial (nominal) value
1
3 Upper limit
Actual value
[a]
5 Lower limit
[a] The default for cal_item is the actual value.
90 Keysight 3458A Calibration Manual
Page 91
Command Summary 5
const_id
Specifies the identifier number for the calibration constant of interest. Each const_id and the associated calibration constant description is listed below.
const_id Description Constant derived from
1 40 K Reference External gain adjustment
2 7 V Reference External gain adjustment
3 dcv zero front 100 mV
4 dcv zero rear 100 mV
5 dcv zero front 1 V
6 dcv zero rear 1 V
7 dcv zero front 10 V
8 dcv zero rear 10 V
9 dcv zero front 100 V
10 dcv zero rear 100 V
11 dcv zero front 1 kV
12 dcv zero rear 1 kV -External zero adjustment
13 ohm zero front 10
14 ohm zero front 100
15 ohm zero front 1 K
16 ohm zero front 10 K
17 ohm zero front 100 K
18 ohm zero front 1 M
19 ohm zero front 10 M
20 ohm zero front 100 M
21 ohm zero front 1 G
Keysight 3458A Calibration Manual 91
Page 92
5 Command Summary
const_id Description Constant derived from
22 ohm zero rear 10
23 ohm zero rear 100
24 ohm zero rear 1 K
25 ohm zero rear 10 K
26 ohm zero rear 100 K
27 ohm zero rear 1 M
28 ohm zero rear 10 M
29 ohm zero rear 100 M
30 ohm zero rear 1 G
31 ohmf zero front 10
32 ohmf zero front 100
33 ohmf zero front 1 K
34 ohmf zero front 10 K -External zero adjustment
35 ohmf zero front 100 K
36 ohmf zero front 1 M
37 ohmf zero front 10 M
38 ohmf zero front 100 M
39 ohmf zero front 1 G
40 ohmf zero rear 10
41 ohmf zero rear 100
42 ohmf zero rear 1 K
43 ohmf zero rear 10 K
44 ohmf zero rear 100 K
45 ohmf zero rear 1 M
46 ohmf zero rear 10 M
92 Keysight 3458A Calibration Manual
Page 93
Command Summary 5
const_id Description Constant derived from
47 ohmf zero rear 100 M
48 ohmf zero rear 1 G
49 offset ohm 10
50 offset ohm 100
51 offset ohm 1 K
52 offset ohm 10 K -External zero adjustment
53 offset ohm 100 K
54 offset ohm 1 M
55 offset ohm 10 M
56 offset ohm 100 M
57 offset ohm 1 G
58 cal 0 temperature
59 cal 10 temperature -Internal temperatures at time of last
60 cal 10 K temperature CAL adjustment
61 vos dac (Dac count to zero boot-strap amp Q7, U12) -External zero adj
62 dci zero rear 100 nA
63 dci zero rear 1 µA
64 dci zero rear 10 µA
65 dci zero rear 100 µA
66 dci zero rear 1 mA -ACAL OHMS
67 dci zero rear 10 mA
68 dci zero rear 100 mA
69 dci zero rear 1 A
70 dcv gain 100 mV
Keysight 3458A Calibration Manual 93
Page 94
5 Command Summary
const_id Description Constant derived from
71 dcv gain 1 V
72 dcv gain 10 V -ACAL DCV
73 dcv gain 100 V
74 dcv gain 1 kV
75 ohm gain 10
76 ohm gain 100
77 ohm gain 1 K
78 ohm gain 10 K
79 ohm gain 100 K -ACAL OHMS
80 ohm gain 1 M
81 ohm gain 10 M
82 ohm gain 100 M
83 ohm gain 1 G
84 ohm ocomp gain 10
85 ohm ocomp gain 100
86 ohm ocomp gain 1 K
87 ohm ocomp gain 10 K
88 ohm ocomp gain 100 K
89 ohm ocomp gain 1 M -ACAL OHMS
90 ohm ocomp gain 10 M
91 ohm ocomp gain 100 M
92 ohm ocomp gain 1 G
93 dci gain 100 nA
94 dci gain 1 µA
95 dci gain 10 µA
94 Keysight 3458A Calibration Manual
Page 95
Command Summary 5
const_id Description Constant derived from
96 dci gain 100 µA
97 dci gain 1 mA
98 dci gain 10 mA -ACAL OHMS
99 dci gain 100 mA
100 dci gain 1 A
101 precharge dac
102 mc dac (dac settings to minimize charge coupling from input fets)
103 high speed gain
-ACAL OHMS
104 il (OFF leakage of ohmmeter current source)
105 il2 (input leakage correction used on 1 MW and higher)
106 rin (value of 10 MW attenuator RP7)
107 low aperture
108 high aperture
109 high aperture slope .01 PLC
110 high aperture slope .1 PLC
111 high aperture null .01 PLC -ACAL DCV
112 high aperture null .1 PLC
113 underload dcv 100 mV
114 underload dcv 1 V
115 underload dcv 10 V
Keysight 3458A Calibration Manual 95
Page 96
5 Command Summary
const_id Description Constant derived from
116 underload dcv 100 V -ACAL DCV
117 underload dcv 1000 V
118 overload dcv 100 mV
119 overload dcv 1 V
120 overload dcv 10 V -ACAL DCV
121 overload dcv 100 V
122 overload dcv 1000 V
123 underload ohm 10
124 underload ohm 100
125 underload ohm 1 K
126 underload ohm 10 K
127 underload ohm 100 K
128 underload ohm 1 M
129 underload ohm 10 M
130 underload ohm 100 M
131 underload ohm 1 G -ACAL OHMS
132 overload ohm 10
133 overload ohm 100
134 overload ohm 1 K
135 overload ohm 10 K
136 overload ohm 100 K
137 overload ohm 1 M
138 overload ohm 10 M
139 overload ohm 100 M
96 Keysight 3458A Calibration Manual
Page 97
Command Summary 5
const_id Description Constant derived from
140 overload ohm 1 G
141 underload ohm ocomp 10
142 underload ohm ocomp 100
143 underload ohm ocomp 1 K
144 underload ohm ocomp 10 K
145 underload ohm ocomp 100 K
146 underload ohm ocomp 1 M
147 underload ohm ocomp 10 M
148 underload ohm ocomp 100 M
149 underload ohm ocomp 1 G
150 overload ohm ocomp 10
151 overload ohm ocomp 100
152 overload ohm ocomp 1 K -ACAL OHMS
153 overload ohm ocomp 10 K
154 overload ohm ocomp 100 K
155 overload ohm ocomp 1M
156 overload ohm ocomp 10 M
157 overload ohm ocomp 100 M
158 overload ohm ocomp 1 G
159 underload dci 100 nA
160 underload dci 1 µA
161 underload dci 10 µA
162
163 underload dci 1 mA
underload dci 100
µA
Keysight 3458A Calibration Manual 97
Page 98
5 Command Summary
const_id Description Constant derived from
164 underload dci 10 mA
165 underload dci 100 mA
166 underload dci 1 A
167 overload dci 100 nA
168 overload dci 1 µA -ACAL OHMS
169 overload dci 10 µA
170 overload dci 100 µA
171 overload dci 1 mA
172 overload dci 10 mA
173 overload dci 100 mA
174 overload dci 1 A
175 acal dcv temperature
176 acal ohm temperature -Last ACAL temperatures
177 acal acv temperature
178 ac offset dac 10 mV
179 ac offset dac 100 mV
180 ac offset dac 1 V
181 ac offset dac 10 V
182 ac offset dac 100 V -ACAL AC
183 ac offset dac 1 kV
184 acdc offset dac 10 mV
185 acdc offset dac 100 mV
186 acdc offset dac 1 V
98 Keysight 3458A Calibration Manual
Page 99
Command Summary 5
const_id Description Constant derived from
187 acdc offset dac 10 V
188 acdc offset dac 100 V
189 acdc offset dac 1 kV
190 acdci offset dac 100 µA
191 acdci offset dac 1 mA
192 acdci offset dac 10 mA
193 acdci offset dac 100 mA
194 acdci offset dac 1 A
195 flatness dac 10 mV
196 flatness dac 100 mV
197 flatness dac 1 V
198 flatness dac 10 V
199 flatness dac 100 V -ACAL AC
200 flatness dac 1 kV
201 level dac dc 1.2 V
202 level dac dc 12 V
203 level dac ac 1.2 V
204 level dac dc 12 V
205 dcv trigger offset 100 mV
206 dcv trigger offset 1 V
207 dcv trigger offset 10 V
208 dcv trigger offset 100 V
209 dcv trigger offset 1000 V
Keysight 3458A Calibration Manual 99
Page 100
5 Command Summary
const_id Description Constant derived from
210 acdcv sync offset 10 mV
211 acdcv sync offset 100 mV
212 acdcv sync offset 1 V
213 acdcv sync offset 10 V
214 acdcv sync offset 100 V
215 acdcv sync offset 1 kV
216 acv sync offset 10 mV
217 acv sync offset 100 mV
218 acv sync offset 1 V
219 acv sync offset 10 V
220 acv sync offset 100 V
221 acv sync offset 1 kV -ACAL AC
222 acv sync gain 10 mV
223 acv sync gain 100 mV
224 acv sync gain 1 V
225 acv sync gain 10 V
226 acv sync gain 100 V
227 acv sync gain 1 kV
228 ab ratio
229 gain ratio
230 acv ana gain 10 mV
231 acv ana gain 100 mV
100 Keysight 3458A Calibration Manual
Loading...