Keysight N1913, 1914A User Manual

Keysight N1913/1914A EPM Series Power Meters
User’s Guide
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 2009–2018 No part of this manual may be
reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Manual Part Number
N1913-90001
Edition
Edition 20, February 12, 2018
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/find/
sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight N1913/1914A User’s Guide

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standard and Technology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members.

General Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part of) other Keysight products. During the warranty period, Keysight will, at its option, either repair or replace products which prove to be defective. The warranty period begins on the date of delivery or on the date of installation if installed by Keysight.

Warranty Service

For warranty service or repair, this product must be returned to a service facility designated by Keysight. For products returned to Keysight for warranty service, the Buyer shall prepay shipping charges to Keysight and Keysight shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Keysight from another country.
Keysight N1913/1914A User’s Guide 3

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Keysight does not warrant the Buyer’s circuitry or malfunctions of Keysight products that result from the Buyer’s circuitry. In addition, Keysight does not warrant any damage that occurs as a result of the Buyer’s circuit or any defects that result from Buyer-supplied products.
To the extent allowed by local law, Keysight makes no other warranty, expressed or implied, whether written or oral with respect to this product and specifically disclaims any implied warranty or condition of merchantability, fitness for a particular purpose or satisfactory quality.

Exclusive Remedies

To the extent allowed by local law, the remedies provided herein are the Buyer’s sole and exclusive remedies. Keysight shall not be liable for any direct, indirect, special, incidental, or consequential damages (including lost profit or data), whether based on warranty, contract, tort, or any other legal theory.
4 Keysight N1913/1914A User’s Guide

Restricted Rights Legend

The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227-7013 (Oct 1988), DFARS 252.211-7015 (May 1991), or DFARS
252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR 2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun 1987) (or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Keysight standard software agreement for the product involved.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Keysight N1913/1914A User’s Guide 5

Safety Summary

WARNING
CAUTION
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.

Safety Notices

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING notice until the indicated cond itions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
6 Keysight N1913/1914A User’s Guide

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of danger. The Instruction Documentation Symbol. The instrument is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
Alternating current (AC)
Direct current (DC) On (Supply)
Both direct and alternating current Off (Supply)
Three-phase alternating current Caution, risk of electric shock
Earth (ground) TERMINAL Caution, hot surface
PROTECTIVE CONDUCTOR TERMINAL In position of bi-stable push control
This symbol indicates the operating switch for ‘Stand-by’ mode. Note, this instrument is NOT isolated from the mains when the switch is pressed. To isolate the instrument, the mains coupler (mains input cord) should be removed from the power supply.
Instrument protected throughout by DOUBLE INSULATION or REINFORCED INSULATION.
Keysight N1913/1914A User’s Guide 7
Frame or chassis TERMINAL Out position of bi-stable push control
This symbol indicates that a device, or part of a device, may be susceptible to electrostatic discharges (ESD) which
Equipotentiality
can result in damage to the product. Observe ESD precautions given on the product, or its user documentation,
when handling equipment bearing this mark.
8 Keysight N1913/1914A User’s Guide

General Safety Information

WARNING
This is a Safety Class I instrument (provided with a protective earthing ground, incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the instrument is likely to damage the meter. Intentional interruption is prohibited.
– Do not operate the instrument in an explosive atmosphere or in the
presence of flammable gasses or fumes.
– Do not use repaired fuses or short-circuited fuseholders: For continued
protection against fire, replace the line fuse(s) only with fuse(s) of the same voltage and current rating and type.
– Do not perform procedures involving cover or shield removal unless you
are qualified to do so: Operating personnel must not remove the meter covers or shields. Procedures involving the removal of covers and shields are for use by service-trained personnel only.
– Do not service or adjust alone: Under certain conditions, dangerous
voltages may exist even with the instrument switched off. To avoid electrical shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Keysight N1913/1914A User’s Guide 9
– Do not operate damaged instrument: Whenever it is possible that the
safety protection features built into this instrument have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the instrument until safe operation can be verified by service-trained personnel. If necessary, return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
– Do not substitute parts or modify the instrument: Because of the danger
of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.

Environmental Conditions

The N1913/1914A is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Temperature
Humidity
Altitude Up to 4600 m
Pollution degree 2

Regulatory Information

Operating condition – 0 °C to 55 °C Storage condition – –40 °C to 70 °C
Operating condition – Up to 95% RH at 40°C (non-condensing) Storage condition – Up to 90% RH at 65°C (non-condensing)
The N1913/1914A EPM Series power meters comply with the following safety and Electromagnetic Compatibility (EMC) compliances:
Safety compliance
– IEC 61010-1:2010/EN 61010-1:2010 (3rd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-12
– USA: ANSI/UL 61010-1 (3rd Edition)
EMC compliance
– IEC 61326-1:2005/EN 61326-1:2006
– CISPR11:2003/EN 55011:2007, Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004
10 Keysight N1913/1914A User’s Guide

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this is an Industrial Scientific and Medical Group 1 Class A product.
This symbol is a South Korean Class A EMC Declaration. This is a Class A instrument suitable for professional use and in electromagnetic environment outside of the home.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
The CSA mark is a registered trademark of the Canadian Standards Association.
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Keysight N1913/1914A User’s Guide 11

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/ 96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/epm
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
12 Keysight N1913/1914A User’s Guide

Table of Contents

Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
General Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Technology Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Safety Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/
EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
1Introduction
LXI Class-C Compliant Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Power Meter and Sensor Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Conventions Used in this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Front Panel Keys and Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
The Display Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Window Symbols and Pop-ups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Keysight N1913/1914A User’s Guide 13
Warning Symbol Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Wait Symbol Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Confirm Symbol Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Numeric Entry Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Text Entry Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
List Pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Rear Panel Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Using the Instrument Web Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Using the Remote Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Editing the Instrument’s LAN Settings . . . . . . . . . . . . . . . . . . . . . . . 46
Capturing the Screen Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Getting the Instrument Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Making Socket Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Programming Language Selection (Option 200) . . . . . . . . . . . . . . . . . . 53
2 General Power Meter Functions
Setting the Units of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Setting the Measurement Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Setting the Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Making Relative Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Setting Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Setting Channel Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Setting Display Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Setting Frequency Dependent Offsets . . . . . . . . . . . . . . . . . . . . . . . 65
Editing Frequency Dependent Offset Tables . . . . . . . . . . . . . . . . . . 68
Selectable Frequency Dependent Offset Unit (dB or %) . . . . . . . . . 71
Setting Measurement Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Step Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Measuring Pulsed Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Setting External Trigger for Average Power Measurement . . . . . . . . . . 79
Power Sweep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Frequency Sweep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Determine the Right Step to be Set . . . . . . . . . . . . . . . . . . . . . . . . . 86
14 Keysight N1913/1914A User’s Guide
Setting Measurement Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Setting Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
Checking for Limit Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
Numeric Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
Single Function Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Combined Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
Max Hold/Min Hold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
Recorder Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
Saving and Recalling Power Meter States . . . . . . . . . . . . . . . . . . . . . .100
Editing a Register’s Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
Recalling a Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . .101
Zeroing and Calibrating the Power Meter . . . . . . . . . . . . . . . . . . . . . .102
Zeroing the Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Zero/Cal Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Calibration Procedure Using E-Series Power Sensors and N8480 Series
Power Sensors (excluding Option CFT) . . . . . . . . . . . . . . . . . . .105
Calibration Procedure Using 8480 Series Power Sensors and N8480 Se-
ries Power Sensors (with Option CFT) . . . . . . . . . . . . . . . . . . . .106
Adapter Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
Blank Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Secure Blank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Backlight Intensity Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118
Memory Erase/Secure Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
VGA Output (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Warm Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Battery Information (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .124
Running Under Battery Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . .124
Battery Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Battery General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128
Setting the Cable Short/Long . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .130
3 Using E9300 E-Series Power Sensors
Keysight N1913/1914A User’s Guide 15
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Measuring Spread Spectrum and Multitone Signals . . . . . . . . . . . . . 138
CDMA Signal Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Multitone Signal Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Measuring TDMA Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Power Meter and Sensor Operation . . . . . . . . . . . . . . . . . . . . . . . . 141
Achieving Stable Results with TDMA Signals . . . . . . . . . . . . . . . . . 141
Achieving Stable Results with GSM Signals . . . . . . . . . . . . . . . . . . 142
Electromagnetic Compatibility (EMC) Measurements . . . . . . . . . . . . 143
Measurement Accuracy and Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Setting the Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
4 Using E4410 E-Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
5 Using 8480 Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
8480 Series Sensors Connection Requirements . . . . . . . . . . . . . . 157
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Frequency Specific Calibration Factors . . . . . . . . . . . . . . . . . . . . . . . . 159
Sensor Calibration Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Editing/Generating Sensor Calibration Tables . . . . . . . . . . . . . . . . 166
Pre-installed Calibration Table Contents . . . . . . . . . . . . . . . . . . . . 170
16 Keysight N1913/1914A User’s Guide
6 Using N8480 Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .176
Power Meter Configuration Changes . . . . . . . . . . . . . . . . . . . . . . . . . .177
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .178
N8480 Series Sensors Connection Requirements . . . . . . . . . . . . . . . .179
N8480 Series Power Sensors (excluding Option CFT) . . . . . . . . . . . . .180
N8480 Series Power Sensors with Option CFT . . . . . . . . . . . . . . . . . .182
Frequency Specific Calibration Factors . . . . . . . . . . . . . . . . . . . . .182
Sensor Calibration Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187
Selecting a Sensor Calibration Table . . . . . . . . . . . . . . . . . . . . . . .188
Editing/Generating Sensor Calibration Tables . . . . . . . . . . . . . . . .191
7 Using U2000 Series USB Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .197
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .199
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200
Electromagnetic Compatibility (EMC) Measurements . . . . . . . . . . . . .202
Measurement Accuracy and Speed . . . . . . . . . . . . . . . . . . . . . . . . . . .203
Setting the Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204
8 Using U8480 Series USB Thermocouple Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .208
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .211
Zeroing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .213
Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .215
Zero+Cal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217
FDO Table Editing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217
Reference Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218
Keysight N1913/1914A User’s Guide 17
9Maintenance
Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220
Front Panel Selection of Self Tests . . . . . . . . . . . . . . . . . . . . . . . . . 220
Instrument Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Remote Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Test Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Error Message List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226
Operator Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Replacing the Power Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Connector Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234
Contacting Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
Before calling Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . 235
Check the Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
Instrument serial numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
Recommended Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . 237
Erasing Memory Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
Returning Your Power Meter for Service . . . . . . . . . . . . . . . . . . . . . . . 239
Packaging the Power Meter for Shipment . . . . . . . . . . . . . . . . . . . 239
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
10 Characteristics and Specifications
18 Keysight N1913/1914A User’s Guide

List of Figures

Figure 1-1 Dual numeric display . . . . . . . . . . . . . . . . . . . . . . . . . .34
Figure 1-2 Single numeric and analog display . . . . . . . . . . . . . . .35
Figure 1-3 Full screen numeric display . . . . . . . . . . . . . . . . . . . . .36
Figure 1-4 Accessing the instrument Web interface . . . . . . . . . . .41
Figure 1-5 N1914A EPM Series power meter Web interface
Figure 1-6 Message to identify the instrument . . . . . . . . . . . . . . .43
Figure 1-7 Example of Lan Status message . . . . . . . . . . . . . . . . .44
Figure 1-8 Viewing LAN configuration settings from the Web
Figure 1-9 Password security dialog box . . . . . . . . . . . . . . . . . . . .47
Figure 1-10 Changing the instrument LAN interface configuration 48
Figure 1-11 Manual Configuration window . . . . . . . . . . . . . . . . . . .51
Figure 1-12 Power meter successfully connected via socket
Figure 2-1 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Figure 2-2 Typical relative measurement display . . . . . . . . . . . . .59
Figure 2-3 Numeric display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
Figure 2-4 Simplified measurement path . . . . . . . . . . . . . . . . . . .61
Figure 2-5 Typical channel offset display . . . . . . . . . . . . . . . . . . .62
Figure 2-6 Channel offset indicator . . . . . . . . . . . . . . . . . . . . . . . .63
Figure 2-7 Typical display offset display . . . . . . . . . . . . . . . . . . . .64
Figure 2-8 Frequency dependent offset tables display . . . . . . . . .66
Figure 2-9 Frequency dependent offset indicator . . . . . . . . . . . . .67
Figure 2-10 “Edit Offset” display with data added . . . . . . . . . . . . .69
Figure 2-11 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . . .69
Figure 2-12 Changing offset unit . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Figure 2-13 Offset display in engineering unit (when the selected unit is
Figure 2-14 Typical averaged readings . . . . . . . . . . . . . . . . . . . . . .73
Figure 2-15 Meas Avg Count pop-up . . . . . . . . . . . . . . . . . . . . . . .74
Figure 2-16 Pulsed signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Figure 2-17 Duty cycle setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Figure 2-18 Duty cycle indicator . . . . . . . . . . . . . . . . . . . . . . . . . . .77
(Welcome Page) . . . . . . . . . . . . . . . . . . . . . . . . . . .42
interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
%) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Keysight N1913/1914A User’s Guide 19
Figure 2-19 TRIG IN and TRIG OUT connection diagram between power
meter and power source . . . . . . . . . . . . . . . . . . . . . 80
Figure 2-20 Channel Setup display . . . . . . . . . . . . . . . . . . . . . . . . . 81
Figure 2-21 Trigger setting menu 1 of 2 . . . . . . . . . . . . . . . . . . . . . 82
Figure 2-22 Trigger setting menu 2 of 2 . . . . . . . . . . . . . . . . . . . . . 82
Figure 2-23 Limits checking applications . . . . . . . . . . . . . . . . . . . . 87
Figure 2-24 Limits checking results . . . . . . . . . . . . . . . . . . . . . . . . 87
Figure 2-25 Minimum limit pop-up . . . . . . . . . . . . . . . . . . . . . . . . . 89
Figure 2-26 Limit failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Figure 2-27 Measurement Setup showing single configuration . . 91
Figure 2-28 Function pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Figure 2-29 Measurement Setup showing combined configuration .
93
Figure 2-30 Measurement example display . . . . . . . . . . . . . . . . . . 93
Figure 2-31 Max hold/min hold measurement is performed on the
‘HOLD’ block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Figure 2-32 Hold pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Figure 2-33 Min Hold and Max Hold measurement mode indicated in
the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Figure 2-34 Measurement mode in full word . . . . . . . . . . . . . . . . . 96
Figure 2-35 Recorder Minimum pop-up . . . . . . . . . . . . . . . . . . . . . 98
Figure 2-36 Recorder Maximum pop-up . . . . . . . . . . . . . . . . . . . . 99
Figure 2-37 Save/Recall screen . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Figure 2-38 Save confirm pop-up . . . . . . . . . . . . . . . . . . . . . . . . . 100
Figure 2-39 File name pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Figure 2-40 Recall pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Figure 2-41 Zeroing pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Figure 2-42 Please zero and calibrate window . . . . . . . . . . . . . . . 103
Figure 2-43 Calibration wait pop-up . . . . . . . . . . . . . . . . . . . . . . . 104
Figure 2-44 System calibration setup . . . . . . . . . . . . . . . . . . . . . . 110
Figure 2-45 Blank screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Figure 2-46 Secure Blank feature . . . . . . . . . . . . . . . . . . . . . . . . . 113
Figure 2-47 Enter 6-digit Password pop-up . . . . . . . . . . . . . . . . . 114
Figure 2-48 6-digit password entered . . . . . . . . . . . . . . . . . . . . . 114
Figure 2-49 Warning message . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Figure 2-50 Reconfirm password . . . . . . . . . . . . . . . . . . . . . . . . . 115
Figure 2-51 Warning message . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
20 Keysight N1913/1914A User’s Guide
Figure 2-52 Password reconfirmation pop-up . . . . . . . . . . . . . . .116
Figure 2-53 Enter password to restore display . . . . . . . . . . . . . . .117
Figure 2-54 Backlight intensity control . . . . . . . . . . . . . . . . . . . . .118
Figure 2-55 Increase or decrease the backlight brightness . . . . .118
Figure 2-56 Activate memory erase using combo keys . . . . . . . .119
Figure 2-57 Secure Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120
Figure 2-58 Confirmation to begin secure erase . . . . . . . . . . . . . .120
Figure 2-59 Warning pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
Figure 2-60 VGA toggle ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . .122
Figure 2-61 Enable/disable warm start feature . . . . . . . . . . . . . .123
Figure 2-62 “Running under battery power” pop-up . . . . . . . . . .124
Figure 2-63 Battery indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Figure 2-64 Battery menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Figure 2-65 Battery status display . . . . . . . . . . . . . . . . . . . . . . . . .126
Figure 2-66 Low battery indicator . . . . . . . . . . . . . . . . . . . . . . . . .126
Figure 2-67 Display backlight control . . . . . . . . . . . . . . . . . . . . . .127
Figure 2-68 Short/long cable option . . . . . . . . . . . . . . . . . . . . . . .130
Figure 3-1 E9300 E-Series auto-averaging settings . . . . . . . . . .133
Figure 3-2 . . . . . E9300 E-Series sensor default channel setup 134
Figure 3-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . .137
Figure 3-4 Spread spectrum signal . . . . . . . . . . . . . . . . . . . . . . .138
Figure 3-5 Wideband CDMA error of E-Series E9300 power sensor
versus corrected CW sensor . . . . . . . . . . . . . . . . .139
Figure 3-6 CDMA (IS-95A): 9Ch Fwd . . . . . . . . . . . . . . . . . . . . .139
Figure 3-7 Calibration factors versus frequency . . . . . . . . . . . . .140
Figure 4-1 E-Series CW sensor auto-averaging settings . . . . . .149
Figure 4-2 E-Series E4410 sensor default channel setup . . . . . .150
Figure 4-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . .152
Figure 5-1 8480 Series auto-averaging settings . . . . . . . . . . . . .155
Figure 5-2 8480 Series sensor default channel setup . . . . . . . . .156
Figure 5-3 Reference Calibration Factor pop-up window . . . . .160
Figure 5-4 Calibration factor pop-up window . . . . . . . . . . . . . . .161
Figure 5-5 Calibration factor display . . . . . . . . . . . . . . . . . . . . . .161
Figure 5-6 Sensor table selected . . . . . . . . . . . . . . . . . . . . . . . . .164
Figure 5-7 Frequency dependent offset indicator . . . . . . . . . . . .164
Figure 5-8 Frequency/calibration table display . . . . . . . . . . . . .165
Figure 5-9 “Sensor Tbls” screen . . . . . . . . . . . . . . . . . . . . . . . . .167
Keysight N1913/1914A User’s Guide 21
Figure 5-10 “Edit Cal” display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Figure 5-11 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . . 168
Figure 6-1 Auto-averaging settings . . . . . . . . . . . . . . . . . . . . . . 177
Figure 6-2 N8480 Series sensor (excluding Option CFT) default chan-
nel setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Figure 6-3 N8480 Series sensor with Option CFT default channel
setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Figure 6-4 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Figure 6-5 Reference calibration factor pop-up window . . . . . . 183
Figure 6-6 Calibration factor pop-up window . . . . . . . . . . . . . . 184
Figure 6-7 Calibration factor displayed . . . . . . . . . . . . . . . . . . . . 185
Figure 6-8 Sensor table selected . . . . . . . . . . . . . . . . . . . . . . . . 188
Figure 6-9 Frequency dependent offset indicator . . . . . . . . . . . 189
Figure 6-10 Frequency/calibration table display . . . . . . . . . . . . . 190
Figure 6-11 “Sensor Tbls” screen . . . . . . . . . . . . . . . . . . . . . . . . . 192
Figure 6-12 “Edit Cal” display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Figure 6-13 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . . 193
Figure 7-1 U2000 Series auto-averaging settings . . . . . . . . . . . 198
Figure 7-2 . . U2000 Series USB power sensor default channel setup
199
Figure 7-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Figure 8-1 U8480 Series auto-averaging settings . . . . . . . . . . . 209
Figure 8-2 U8480 Series USB power sensor default channel setup
210
Figure 8-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 212
Figure 8-4 Zeroing pop-up message . . . . . . . . . . . . . . . . . . . . . 213
Figure 8-5 Zeroing error pop-up message . . . . . . . . . . . . . . . . . 214
Figure 8-6 Calibration type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
Figure 8-7 Calibration pop-up message . . . . . . . . . . . . . . . . . . . 216
Figure 8-8 Calibration error pop-up message . . . . . . . . . . . . . . 216
Figure 8-9 FDO table editing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
Figure 8-10 Reference manual softkey . . . . . . . . . . . . . . . . . . . . . 218
Figure 8-11 QR code screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218
Figure 9-1 Self test complete . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Figure 9-2 Error indicator position . . . . . . . . . . . . . . . . . . . . . . . 224
Figure 9-3 Replacing the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Figure 9-4 Secure erase status pop-up . . . . . . . . . . . . . . . . . . . 238
22 Keysight N1913/1914A User’s Guide

List of Tables

Table 1-1 Type of Lan Status message . . . . . . . . . . . . . . . . . . . .43
Table 2-1 Measurement units - Single channel meters . . . . . . .56
Table 2-2 Measurement units - Dual channel meters . . . . . . . . .56
Table 2-3 Range of values for window limits . . . . . . . . . . . . . . . .88
Table 2-4 Ranges of recorder output setting . . . . . . . . . . . . . . . .99
Table 2-5 Power Sensor Connection Requirements . . . . . . . . .107
Table 2-6 Equipment list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
Table 2-7 Typical specifications . . . . . . . . . . . . . . . . . . . . . . . . .109
Table 3-1 Power sensor connection requirements . . . . . . . . . .135
Table 5-1 8480 Series connection requirements . . . . . . . . . . . .157
Table 5-2 Installed power sensor models . . . . . . . . . . . . . . . . .166
Table 6-1 Power range in the Range setting . . . . . . . . . . . . . .176
Table 6-2 N8480 Series connection requirements . . . . . . . . . .179
Table 6-3 Installed power sensor models . . . . . . . . . . . . . . . . .191
Table 7-1 Power sensor connection requirements . . . . . . . . . .200
Keysight N1913/1914A User’s Guide 23
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 Keysight N1913/1914A User’s Guide
Keysight N1913/1914A EPM Series Power Meters User’s Guide

1 Introduction

LXI Class-C Compliant Power Meter 26 Rack Mounting 26 Power Meter and Sensor Capability 27 Conventions Used in this Guide 28 Front Panel Keys and Connections 29 The Display Layout 34 Window Symbols and Pop-ups 37 Rear Panel Connections 40 Using the Instrument Web Interface 41 Making Socket Connection 51 Programming Language Selection (Option 200) 53
This chapter introduces you to the front panel display and instrument Web browser of the N1913/1914A EPM Series power meter.
25
1Introduction

LXI Class-C Compliant Power Meter

The N1913/1914A EPM Series power meter is a LXI Class C compliant instrument, developed using LXI Technology. LXI, an acronym for LAN eXtension for Instrumentation, is an instrument standard for devices that use the Ethernet (LAN) as their primary communication interface.
Hence, it is an easy-to-use instrument especially with the usage of an integrated Web browser that provides a convenient way to configure the instrument’s functionality.

Rack Mounting

The N1913/1914A can be mounted in a standard 19-inch rack. Rack mount kits are available as listed below. Support rails are also required for rack mounting. These are normally supplied with the rack and are not included with the rack mount options.
If you are installing an instrument on top of the N1913/1914A, ensure that the instrument does not obstruct the ventilation holes at the top of the N1913/1914A. If required, use a filler panel above the N1913/1914A to ensure adequate space for air circulation.
Option Description
N1913A Option 908 Rack mount kit for one instrument
N1913A Option 909 Rack mount kit for two instruments
N1914A Option 908 Rack mount kit for one instrument
N1914A Option 909 Rack mount kit for two instruments
26 Keysight N1913/1914A User’s Guide

Power Meter and Sensor Capability

Your N1913/1914A EPM Series power meter is compatible with the Keysight E9300 E-Series, E4410 E-Series, 8480 Series, N8480 Series, U2000 Series, and the U8480 Series thermocouple sensor. However, not all sensor and meter combinations have the same features or capabilities. The main differences are as below:
Introduction 1
Features E-Series
E9300
Average power of CW signal ••••••
Average power of modulated signal ••••
Cal factors stored on EEPROM
Correction factors stored in a 3 MB Flash memory
>
200 readings/sec ••
[a] Not applicable for N8480 Series power sensors with Option CFT
••
E-Series
E4410
8480
Series
N8480
Series
[a]
U2000
Series
••
••
U8480
Series

Specifications

The specifications for the power meter are listed in Chapter 10, "Characteristics
and Specifications," starting on page 241.
Keysight N1913/1914A User’s Guide 27
1Introduction
Softkey
Channel
Channel A
Channel B
Softkey

Conventions Used in this Guide

The following conventions are used throughout this guide.
This symbol and text represents a labeled key on the power meter front panel.
This symbol and text represents a labeled softkey and is used to indicate that you should press the unmarked key beside the displayed text.
Message This text represents a displayed message.
Parameter This is used to represent a parameter, value, or title.
“Channel” This User’s Guide describes the operation for both the single channel and the dual channel power
meter. To identify channels on a dual channel meter a softkey on an N1913A meter becomes and on an N1914A.
When you are asked to press “the channel” in a procedure, make sure you select the relevant channel.
28 Keysight N1913/1914A User’s Guide

Front Panel Keys and Connections

This section briefly describes the functions of the front panel keys and connectors.
These keys are located to the left of the display.
Key Function
Press this key to preset the power-meter to the default setting.
Introduction 1
Press this key to control the power meter from the front panel when it is operating via the remote interfaces (when Local Lock Out is not enabled).
Press this key to select the upper or lower measurement window. The selected window is highlighted by a blue line on the right side of the window. Any measurement setup you create is performed in the selected window.
Press this key to choose windowed, expanded, or full-screen display of a numeric measurement.
Press this key to switch the meter between on and standby. When power is supplied, the background LED is red. Pressing the key, switches the power meter on and the background LED is green. When the meter is powered on, the start-up will take approximately 25 seconds.
Keysight N1913/1914A User’s Guide 29
1Introduction
These keys are located along the lower edge of the display.
Key Function
Press this key to access general configuration-menus, such as GPIB address. You can also access some measurement configuration menus. The measurement screen remains visible.
Press this key to access the channel configuration menus. Channel parameters such as averaging and offsets are configured from this menu.
Press this key to access the triggering menu. The triggering feature will be made available in future for power sensors with triggering capability.
Press this key to setup relative measurements or set display offsets. Use this key to configure the selected measurement.
Press this key to access the measurement display menu. You can choose the displayed measurement resolution, units and display format.
Use this key together with to configure measurement displays.
30 Keysight N1913/1914A User’s Guide
These keys are all associated with the menu labels and data entry. They are
Confirm
located to the right of the display.
Key Function
Press this key to return to the previous screen. This key also cancels pop-up entry.
Introduction 1
These unmarked keys are called ‘softkeys’ and are referred to by the text on the display next to them.
For example, during a Preset, you are given an option to confirm the command. Press to continue, that is, press the softkey beside the displayed word ‘confirm’.
The lowest of the unmarked softkeys is used when there is a two page menu to be displayed. For example, a 1 of 2 is displayed beside the key indicating the first page of a two page menu. Press the key to access the next page or second page. (A 2 of 2 is displayed).
Keysight N1913/1914A User’s Guide 31
1Introduction
These keys and connectors are associated with the measurement channels and are located on the right-hand side of the front panel.
Key Function
The arrow keys are used for navigation around the parameter entry screens. The up and down arrows are used for selecting values from a pop-up list. They are also used to enter text, for example, table names.
Press this key to select a highlighted field to allow data entry, check a checkbox and terminate entry of a popup list.
Press this key to access the zero and calibration menus.
Press this key to reset the MAX HOLD and MIN HOLD measurement.
Press these keys to enter numeric values in the pop-up fields, for example, the offset values. To complete the entry, use the softkey.
32 Keysight N1913/1914A User’s Guide
Connector Function
The power reference is a 1 mW (0 dBm) 50 MHz signal available from a 50 Ω type-N connector. It is used for calibrating an 8480 or E-Series power sensor and meter system. If the meter is configured with Option C03, the connector is fitted to the rear panel. The Green LED beside the connector is lit when the calibrator is turned on.
The sensor input connectors (N1914A shown, the N1913A has one input). If the meter is configured with Option C02 or C03, the connectors are fitted to the rear panel and the front panel connectors are retained.
If the meter is configured with Option 201, one USB Type A port is fitted to the front panel (Channel C) and another Type A port (Channel D) to the rear panel.
Introduction 1
Keysight N1913/1914A User’s Guide 33
1Introduction
Disp Type
1
2
3
4
5
7
6

The Display Layout

Figure 1-1 shows the display layout when two windows are configured in dual
numeric mode.
Other display formats are available by pressing , .
Figure 1-1 Dual numeric display
1 The status reporting line displays messages and the control status of the
power meter.
For example, the status can be either RMT (remote, GPIB, USB or LAN operation) or LCL (local, front panel operation). The message fields indicate ERR for any error conditions that occur or informing you to Please Zero the power sensor.
2 The measured channel is shown with a 8480 Series or E-Series power sensor
connected.
3 This field displays the menu title.
For example, Channel Setup or press and the Zero/Cal menu is
displayed.
4 The blue highlight on the right hand side of the window shows it is the
currently selected measurement display line. This measurement line is the Upper Window/Upper Measurement.
5 The available softkey labels are displayed in these three fields. Additionally,
settings associated with the labeled function are displayed under the label.
34 Keysight N1913/1914A User’s Guide
Introduction 1
8
9
10
11
12
Softkeys labels that are grayed out cannot be selected.
6 This displays the measurement units, either dBm or Watts (W).
7 This displays the number of pages in the current menu. For example, 1 of 2
indicates that there are two pages in the menu and the first page is currently displayed. Pressing the softkey displays the next page, indicated by 2 of 2 (press the softkey to display the previous menu page).
Figure 1-2 Single numeric and analog display
Figure 1-2 shows the default display mode of two measurement windows.
8 The channel measurement frequency.
9 The upper window is configured to show a single numeric display.
10 The lower window is configured to show an analog meter which displays the
measurement result and the meter scaling.
11 This displays the connected sensor, the offset value, and the acquisition mode
on the channel. On dual channel models, it shows for both channels.
12 The blue highlight on the right hand side of the window shows it is the
currently selected measurement display line.
Using the , , or keys, you can change the measurement window
selection.
Using the key on numeric measurement results window, you can choose either two rectangular windows, a single enlarged window, or a full
screen display. The display style is applied to the currently selected window or measurement line.
Keysight N1913/1914A User’s Guide 35
1Introduction
18
13
17
14
15
16
Figure 1-3 Full screen numeric display
Figure 1-3 shows a single numeric full screen displaying a relative result.
13 This field displays Minimum Hold if range hold is set to minimum.
14 The information in this field is displayed on two lines and depends on the
sensor type, sensor calibration table, frequency dependent offset table currently selected, and the measurement frequency.
15 This field displays Dty Cyc if a duty cycle is set.
16 This field displays Ofs if an offset is set.
17 This field displays Rel if relative mode is on.
18 This field indicates the measurement result is beyond the configured upper or
lower limit. If the measurement is within the limits this field is empty. If the measurement result is less than the minimum limit set, Undr Lmt is displayed. If the measurement result is more than the maximum limit set, Over Lmt is displayed.
36 Keysight N1913/1914A User’s Guide

Window Symbols and Pop-ups

There are several different graphic symbols and pop-up windows that can occur on the power meter display. These can occur for a variety of reasons, for example:
– An error or warning occurs
– You are required to wait while the power meter carries out a procedure
– You are required to select an entry from a list
– You are required to enter a numeric value
There are three different colors used to signify the pop-up status:
– Green - used to allow data entry
– Orange - used to display information
– Red - used to display an error

Warning Symbol Pop-up

The warning symbol is displayed either in a pop-up window or directly in the measurement window when such an event occurs. A pop-up window is displayed for approximately two seconds. The text in the pop-up window gives details of the warning type, for example, to indicate that a power sensor has insufficient bandwidth or a previous entered frequency value in a table. Depending on the severity of the warning, the pop-up may be displayed in orange or red.
Introduction 1
Keysight N1913/1914A User’s Guide 37
1Introduction

Wait Symbol Pop-up

The wait symbol is displayed when the power meter is carrying out a procedure and no action is required from you. The symbol appears in a pop-up window. It may appear, for example, during a calibration.

Confirm Symbol Pop-up

This type of pop-up window is displayed when you are required to press Confirm to verify your previous selection. For example, prior to a Save being carried out.

Numeric Entry Pop-up

This type of pop-up window is displayed when you need to modify numeric data. The numeric keys allow you to enter the value.
38 Keysight N1913/1914A User’s Guide

Text Entry Pop-up

This type of pop-up window is displayed when you need to modify alphanumeric data such as table names. The up/down arrow keys increment and decrement the alphanumeric digit that the cursor is currently positioned. The left/right arrow keys move the cursor to another alphanumeric digit.

List Pop-up

This pop-up window is displayed when you are required to select an entry from a list. Use the up/down arrow keys to highlight your choice. Press select to complete the entry.
Introduction 1
Keysight N1913/1914A User’s Guide 39
1Introduction
7
8
6
9
4
3
2
1
5

Rear Panel Connections

No. Connections
1 VGA Output (Option 201)
2 Ground Connector
3 USB Type A port
4 Recorder 1/2
Recorder output (two outputs are fitted to dual channel meters) connections are made via BNC connectors. This output produces a DC voltage that corresponds to the power level of the channel input.
5AC Inlet
This power meter has an auto configuring power supply. This allows it to operate over a range of voltages without manually being set to a certain voltage.
6 Trig In/Trig Out
Trigger input and output connections are made via BNC connectors.
7 USB Mini-B port
This USB port is used only for remote interface connection.
8LAN
9GPIB
This connector allows the power meter to be controlled remotely using the General Purpose Interface Bus.
40 Keysight N1913/1914A User’s Guide

Using the Instrument Web Interface

NOTE
You can communicate with the N1913/1914A EPM Series power meters using the Web interface.
The instrument Web interface can be accessed from Keysight Connection Expert as shown in Figure 1-4.
Alternatively, the instrument Web interface can also be accessed directly from a Web broswer by entering the instrument’s IP address or hostname in the browser’s ‘address’ window.
Introduction 1
Figure 1-4 Accessing the instrument Web interface
Keysight N1913/1914A User’s Guide 41
1Introduction
NOTE
An example of the instrument Web interface (Welcome Page) is shown in
Figure 1-5.
Figure 1-5 N1914A EPM Series power meter Web interface
(Welcome Page)
You can control the instrument via GPIB, LAN, and USB connection. The connection parameters can be found on the Welcome Page. For example, SCPI TCPIP socket port (5025), SCPI Telnet port (5024), VISA TCPIP Connect String, VISA USB Connect String, and GPIB address are shown. Click Advanced Information… to display more information about the instrument.
– The instrument has an embedded Web server, which is listening on port 80 to
serve Web pages.
– The Web pages can be browsed using Web browser such as Internet Explorer
and Mozilla Firefox.
Instrument on the network can be physically identified from the blinking message on front panel screen by clicking Turn On Front Panel Identification Indicator on the Welcome Page.
42 Keysight N1913/1914A User’s Guide
Introduction 1
IDENTIFY” message to blink when the front panel identification indicator is turned on.
Remote Interfaces
When the front panel identification indicator is turned on, a blinking “IDENTIFY” message is displayed on the screen of the front panel. See Figure 1-6. The “IDENTIFY” message will blink to identify the instrument until you click Turn Off
Front Panel Identification Indicator.
Figure 1-6 Message to identify the instrument
When the instrument is configured to LAN, the Lan Status on screen will show the LAN error condition and status of the LAN configuration connection. There are six types of Lan Status messages that may occur. See
Figure 1-1. See also Figure 1-7 for the example of Lan Status message.
Type of message Description
Lan: No Fault Status: Initialized
Lan: No Fault Status: Running
Lan: Faul t Status: Initialization failed
Lan: Faul t Status: Disconnected
Keysight N1913/1914A User’s Guide 43
Table 1-1 Type of Lan Status message
– A valid IP address is successfully obtained using selected LAN configuration
and the network state is initialized.
– A valid IP address is successfully obtained using selected LAN configuration,
while network is running.
– IP conflict occurs, or – IP address failed to be obtained using selected LAN configuration, or – No LAN configuration is selected.
– LAN cable is unplugged.
1Introduction
NOTE
Lan: – Status: Restarting
Lan: DHCP Not Available Status: Running
Figure 1-7 Example of Lan Status message
For more details on remote interface configuration, refer to N1913/1914A EPM Series Power Meters Installation Guide.
– Restart the network and try to obtain an IP address using selected LAN
configuration.
– Unable to obtain IP address from DHCP server (if user select DHCP
configuration)
– IP address obtained from Auto-IP or manual configuration.
44 Keysight N1913/1914A User’s Guide

Using the Remote Front Panel

NOTE
The instrument Web interface also provides a virtual front panel interface that can be used to control the power meter remotely.
1 On the left of the Welcome Page, select Browser Web Control. The remote
front panel appears.
Introduction 1
2 Click the front panel keys to control the instrument.
Java™ must be installed on the controlling PC for remote front panel operation.
Keysight N1913/1914A User’s Guide 45
1Introduction

Editing the Instrument’s LAN Settings

Once communication path to the instrument has been established, the instrument’s LAN configuration can be viewed and modified using the Web interface.
On the Welcome Page, click View and Modify Configuration. This opens the configuration window shown in Figure 1-8.
Figure 1-8 Viewing LAN configuration settings from the Web interface
To edit parameters shown, click Mod ify Configuration. The Enter Password dialog box appears as shown in Figure 1-9.
46 Keysight N1913/1914A User’s Guide
Introduction 1
NOTE
Remote Interfaces
1 of 2
LAN Reset
NOTE
LAN Reset
NOTE
Figure 1-9 Password security dialog box
Click the Submit (accept the default password) and the window opens as shown in Figure 1-10. The default password is “keysight”.
A LAN reset needs to be performed to ensure that the password is reset to default. See LAN reset procedure as below.
Procedure:
– Press , to display the Remote Interfaces screen.
– Press softkey to display the second page of the Remote I/F menu.
– Press softkey to reset the LAN settings to default.
When softkey is pressed, it will also change the GPIB address to default.
You can change the password from the Configuring your N1914A Power Meter window as shown in Figure 1-10. Scroll down the Parameter column until you locate the Change Password parameter.
Keysight N1913/1914A User’s Guide 47
1Introduction
Figure 1-10 Changing the instrument LAN interface configuration
48 Keysight N1913/1914A User’s Guide

Capturing the Screen Image

To save the instrument’s display from the Web interface:
1 On the left of the Welcome Page, select Get Image. The screen image will be
displayed.
Introduction 1
2 Right-click on the image and select Save Picture As….
3 Select a storage location for the image file and click Save.
The image is captured as a Bitmap (BMP) file, to the default file name display.bmp.
Keysight N1913/1914A User’s Guide 49
1Introduction

Getting the Instrument Data

The instrument Web interface allows you to transport measurement readings from the instrument to PC applications such as word and spreadsheet applications.
To get the instrument data:
1 On the left of the Welcome Page, select Get Data. The Get Data Web page will
be displayed.
2 Select the window/measurement type of the instrument.
3 Enter your desired count value (up to 1000 only) of the data and click Get
Data. The data will be displayed in a text box.
4 Copy and paste the data in your intended PC application.
50 Keysight N1913/1914A User’s Guide

Making Socket Connection

2
3
4
5
6
7
The following steps describe how the power meter can be remotely connected via socket connection:
1 Right-click the Keysight IO Libraries icon on the taskbar and select
Keysight Connection Expert.
Introduction 1
Figure 1-11 Manual Configuration window
2 Click Manual Configuration on the Keysight Connection Expert window.
3 Click LAN instrument.
4 Enter the Hostname or IP Ad dress and TCPIP Interface ID.
5 Select Socket.
6 Click Test This VISA Address to test if the instrument is present.
Keysight N1913/1914A User’s Guide 51
7 Click Accept. You will then return to the Keysight Connection Expert window.
Your power meter is successfully connected via socket connection.
1Introduction
Figure 1-12 Power meter successfully connected via socket connection
52 Keysight N1913/1914A User’s Guide

Programming Language Selection (Option 200)

NOTE
The programming language selection is available as an orderable option. For N1913A, you can use SCPI, HP 436A, or HP 437B programming language to program the power meter from the remote interface. For N1914A, you can use either SCPI or HP 438A programming language to program the power meter from the remote interface. The default language is SCPI when the power meter is shipped from the factory.
The power meter complies with the rules and regulations of the 1996.0 version of SCPI (Standard Commands for Programmable Instruments). You can determine the SCPI version with which the power meter is in compliance by sending the SYSTem:VERSion? command from the remote interface. You cannot query the SCPI version from the front panel.
The language selection is stored in non-volatile memory and does not change when power has been off or after a remote interface reset.
To select the interface language from the front panel (N1913A),
1 Press , Remote Interfaces, 1 of 2, and Command Set.
2 Select the language from HP 436A
[1]
, HP 437B
[1]
, and SCPI
Introduction 1
[2]
.
To select the interface language from the front panel (N1914A),
1 Press , Remote Interfaces, 1 of 2, and Command Set.
2 Select the language from HP 438A
[1]
, and SCPI
[2]
.
To select the interface language from the remote interface, use the SYSTem:LANGuage command.
Option 200 can only support 8480 series, N8480 series CFT-option, and E4412/ 3A power sensors.
[1] HP 436A, HP 437B, and HP 438A language modes are not compatible with LAN or USBTM remote
interface.
[2] SCPI is the default factory setting. License is needed for the other languages to be available.
Keysight N1913/1914A User’s Guide 53
1Introduction
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
54 Keysight N1913/1914A User’s Guide
Keysight N1913/1914A EPM Series Power Meters User’s Guide
2 General Power Meter
Functions
Setting the Units of Measurement 56 Setting the Measurement Frequency 57 Setting the Resolution 58 Making Relative Measurements 59 Setting Offsets 61 Setting Measurement Averaging 73 Step Detection 75 Measuring Pulsed Signals 76 Setting External Trigger for Average Power Measurement 79 Setting Measurement Limits 87 Single Function Measurement 92 Combined Measurement 93 Max Hold/Min Hold 94 Recorder Output 97 Saving and Recalling Power Meter States 100 Zeroing and Calibrating the Power Meter 102 Adapter Test Procedure 109 Secure Blank 113 Backlight Intensity Control 118 Memory Erase/Secure Erase 119 VGA Output (Optional) 122 Warm Start 123 Battery Information (Optional) 124 Setting the Cable Short/Long 130
This chapter describes the general operation of the N1913/1914A EPM Series power meters.
55
2 General Power Meter Functions
Units
dBmWdB
%
NOTE

Setting the Units of Measurement

The Units menu is used to select the measurement units for the currently selected window. These can either be logarithmic (dBm or dB) or linear (Watt or %) units.
Presetting ( ) the power meter sets the measurement units to dBm (logarithmic units). Table 2-1 and Table 2-2 show units that are applicable to each
measurement mode.
Press , . Select the unit of measurement from , , , and
. Softkeys which cannot be selected in your particular mode of operation are
grayed out.
When the measurement unit is set to Watt (W), it is possible that negative power results are displayed when measuring low power levels.
Tab le 2-1 Measurement units - Single channel meters
Measurement Mode Relative Mode Off Relative Mode On
Log dBm dB
Linear Watt %
Tab le 2-2 Measurement units - Dual channel meters
Measurement Mode Relative Mode Off Relative Mode On
Ratio Log dB dB
Linear % %
Difference Log dBm dB
Linear Watt %
56 Keysight N1913/1914A User’s Guide

Setting the Measurement Frequency

MHz
GHz
Entering the frequency of the RF signal you are measuring optimizes the accuracy and minimizes measurement uncertainty, especially when making comparative measurements between signals.
Procedure
Set the measurement frequency as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to highlight the Frequency value field and press
to display the Frequency pop-up. Use the numeric keypad to enter the
required value in the Frequency pop-up window.
General Power Meter Functions 2
Figure 2-1 Frequency pop-up
3 Confirm your choice by pressing or .
4 Press key to close the Channel Setup screen.
Keysight N1913/1914A User’s Guide 57
2 General Power Meter Functions
Resolution
Resolution

Setting the Resolution

The resolution of each of the power meter’s numeric type windows can be set to four different levels (1, 2, 3 or 4).
These four levels represent:
– 1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
– 1, 2, 3 or 4 significant digits respectively if the measurement suffix is W or %.
The default value is 0.01 dB (3 digits).
To set the resolution on the currently selected window:
1 Press . The current setting of the resolution is highlighted under the
softkey.
2 To change this setting, press repeatedly until the required
resolution setting is highlighted.
58 Keysight N1913/1914A User’s Guide

Making Relative Measurements

Meas Select
Relative
setting field
Relative field
Original result
Result field
Selected window/ measurement
Relative mode enables comparison of a measurement result to a reference value. The relative reading, or difference, can be displayed in either dB or % terms. When the measurement result is displayed in % a prefix multiplier may be shown.
Procedure
1 Press to display the Measurement Setup menu.
Figure 2-2 shows a Measurement Setup display and the relative measurement
items labeled.
2 Select the window you wish to set a reference value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Relative setting field.
General Power Meter Functions 2
Figure 2-2 Typical relative measurement display
4 Press to check the Relative setting field.
5 Confirm that the power meter is measuring the signal you want to use as the
reference. This is displayed under the Result field.
6 Press the , , key to highlight the Rel field.
Keysight N1913/1914A User’s Guide 59
2 General Power Meter Functions
NOTE
Relative indicator
NOTE
7 Press and the original result value will appear on the right hand side of
the Rel check field.
8 The relative value displayed under Result field will change as the measured
signal varies.
If you return the power meter to display the numeric display, a Rel symbol will be displayed in the measurement window it is applied to.
Figure 2-3 Numeric display
The Rel symbol is not displayed when the associated measurement is displayed in Dual Numeric or Analog format.
60 Keysight N1913/1914A User’s Guide

Setting Offsets

Channel Functions
Window Functions
Display
Relative
Offset
Channel A
Upper Window
Lower Window
Display
Relative
Offset
Display
Relative
Offset
Display
Relative
Offset
Measurement Highway
Freq Dep Offset
Channel Offset
Cal Factor
Channel B
Freq Dep Offset
Channel Offset
Cal Factor
Channel C
Freq Dep Offset
Channel Offset
Cal Factor
Channel D
Freq Dep Offset
Channel Offset
Cal Factor
The power meter can be configured to compensate for a signal loss or gain in your test setup. The power meter allows you to apply offsets at three different points in the measurement path.
General Power Meter Functions 2
Figure 2-4 Simplified measurement path
Keysight N1913/1914A User’s Guide 61
2 General Power Meter Functions
Offsets
Offset value field
Offset setting field
Figure 2-4 shows that how you can apply a Channel Offset or a Frequency
Dependent Offset prior to any mathematical functions. These allow you to
compensate each channel individually. An overall offset can be applied if required using the Display Offset.

Setting Channel Offsets

This gain or loss is applied to the measured power before any mathematical functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
A positive value compensates for a loss and a negative value compensates for a gain.
Procedure
To enter a channel offset:
1 Press to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
2 Press to display the Offsets Setup.
3 Use the and keys to highlight the Offset setting field.
4 Press to check the Offset setting field.
Figure 2-5 Typical channel offset display
5 Press to highlight the Offset value field and press to display the
Offset pop-up. Use the numeric keypad to enter the required value in the Offset pop-up window.
62 Keysight N1913/1914A User’s Guide
6 Confirm your choice by pressing .
dB
Offset indicator
NOTE
Meas Select
7 Press key to complete the offset entry. If either a channel or a display
offset is set, the Ofs indicator is displayed.
Figure 2-6 Channel offset indicator
The Ofs symbol is not displayed when the associated measurement is displayed in Dual Numeric or Analog format.

Setting Display Offsets

General Power Meter Functions 2
This gain or loss is applied to the measured power after any channel offsets or mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB. A positive value compensates for a loss, and a negative value compensates for a gain.
Procedure
Enter a display offset on the currently selected window:
1 Press to display the Measurement Setup screen.
2 Select the window you wish to set the offset value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Offset setting field.
4 Press to check the Offset setting field.
Keysight N1913/1914A User’s Guide 63
2 General Power Meter Functions
Selected window/ measurement
Offset value field
Offset setting field
dB
NOTE
Figure 2-7 Typical display offset display
5 Press to highlight the Offset value field and press to display the
Display Offset pop-up. Use the numeric keypad to enter the required value in
the Offset pop-up window.
64 Keysight N1913/1914A User’s Guide
6 Confirm your choice by pressing . Press key to complete the offset
entry.
The Ofs indicator is displayed if Display Offset is selected.

Setting Frequency Dependent Offsets

Tab les
Meter
Freq. Dep. Offset
Tab les
Sensor ChC
Freq. Dep. Offset
Tab les
Sensor ChD
Freq. Dep. Offset
Frequency dependent offset tables provide a quick and convenient method of compensating for frequency related changes in the response of your test system. Note that when selected, frequency dependent offset corrections are applied IN ADDITION to any correction for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables with a maximum of 80 frequency points each.
To use frequency dependent offset tables:
1 Select the table to be applied to a channel. Refer to “Setting Frequency
Dependent Offsets” on page 65 for further information. If you require to edit
the table refer to “Editing Frequency Dependent Offset Tables” on page 68 for further information.
2 If an 8480 Series, N8480 Series, an E-Series, or an U2000 Series sensor is
used, zero and calibrate the power sensor. The reference calibration factor used during the calibration is automatically set by the power meter from the sensor calibration table (if selected).
3 Specify the frequency of the signal you want to measure. The calibration
factor/offset is automatically set by the power meter from the sensor calibration table (if selected) and the frequency dependent offset table. Refer to “Procedure” on page 66 for further information.
4 Make the measurement.
General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 65
Selecting a Frequency Dependent Offset Table
You can select a frequency dependent offset table from the key menu followed by , , and for Channel A or Channel B.
For Channel C, you can select the frequency dependent offset table from the
key menu followed by , , and .
For Channel D, you can select the frequency dependent offset table from the
key menu followed by , , and .
2 General Power Meter Functions
NOTE
Offsets
Tab les
Freq. Dep. Offset
Offsets
Tab le
A Table
B Table
On
The State column indicates if any frequency dependent offset tables are currently selected. The Offset Tables screen is shown in Figure 2-8.
You can also view which FDO table is being used by pressing the ,
and use the and keys to highlight the FDO Table setting field
and press to display the table.
Procedure
Select an offset table as follows:
1 Press, either:
a , , .
b , and use the and keys to highlight the FDO Table
setting field and press to display the table.
Figure 2-8 Frequency dependent offset tables display
2 Use the and keys to highlight one of the 10 table titles and press
( or for dual channel) to highlight .
66 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE
Tab le
Offset Table A Selected
NOTE
When no data is contained in the highlighted table, the key is disabled (grayed out).
3 Press to complete the selection of the offset table.
4 Press again to display the measurement screen. Figure 2-9 shows
which offset table is selected.
Figure 2-9 Frequency dependent offset indicator
Keysight N1913/1914A User’s Guide 67
5 To change the frequency, press and use the and keys to
highlight the Frequency field.
6 Press to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The measurement result, including offset, is now displayed.
If the measurement frequency does not correspond directly to a frequency in the sensor calibration table (if selected) and the frequency dependent offset table being used, the power meter calculates the calibration factor and offset using linear interpolation. If you enter a frequency outside the frequency range defined in the sensor calibration table or the frequency dependent offset table, the power meter uses the highest or lowest frequency point in the appropriate table to set the calibration factor and offset.
2 General Power Meter Functions
Tab les
Freq. Dep. Offset
NOTE
Tab les
Freq. Dep. Offset
Edit Table

Editing Frequency Dependent Offset Tables

There are ten frequency dependent offset tables named CUSTOM_A through CUSTOM_J. They are empty of any data when the power meter is shipped from the factory.
You cannot delete any of the 10 existing frequency dependent offset tables or create any additional tables. However, you can enter values into the 10 existing tables. Each frequency dependent offset table can contain a maximum of 80 frequency points.
To view the frequency dependent offset tables currently stored in the power
meter, press , , . The Frequency Dependent Offset Tables screen is displayed as shown in Figure 2-8.
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
68 Keysight N1913/1914A User’s Guide
Procedure
First, select the table you want to edit as follows:
A frequency in the range of 0.001 MHz to 1000.0 GHz can be entered. A calibration factor in the range of 0.0000000001 to 10000000000 can be entered.
1 Press , , to display the Offset Tbls screen.
2 Choose the table you want to edit using the and keys. Press
to display the Ed it Offset screen as shown in Figure 2-10.
General Power Meter Functions 2
Change
Insert Char
Delete Char
Enter
Figure 2-10 “Edit Offset” display with data added
3 Highlight the table title using the and keys. Press and use
the , , and keys to select and change the characters in the
Tab le Name pop-up to create the name you want to use.
Figure 2-11 Edit table title pop-up
– Pressing adds a new character to the right of the selected
character.
– Pressing removes the selected character.
4 Press to complete the entry.
Keysight N1913/1914A User’s Guide 69
2 General Power Meter Functions
NOTE
Insert
Change
GHz
MHz
Change
%
Done
The following rules apply to naming sensor calibration tables:
– The name must consist of no more than 12 characters.
– All characters must be upper or lower case alphabetic characters, or numeric
(0-9), or an underscore (_).
– No other characters are allowed.
– No spaces are allowed in the name.
Enter (or edit) the frequency and offset pairs as follows:
1 Press to add a new frequency value (or press to edit). Use the
numeric keypad to enter the required value in the Frequency pop-up window. Complete the entry by pressing the , keys.
2 Enter the new offset value (or press to edit). Use the numeric keypad
to enter the required value in the Offset pop-up window. Complete the entry by pressing the key.
Figure 2-12 Changing offset unit
3 Continue adding/editing values until you have entered all the data you require.
4 When you have finished editing the table press to save the table.
70 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE
1 of 2
Offset Unit
dB
1 of 2
Offset Unit
%
If you measure a signal with a frequency outside the frequency range defined in the frequency dependent offset table, the power meter uses the highest or lowest frequency point in the frequency dependent offset table to calculate the offset

Selectable Frequency Dependent Offset Unit (dB or %)

The offset in dB range is from –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
Once the offset unit is selected, it will be applied to all the offsets in the selected table and also to the remaining nine tables.
To change the offset unit to dB, press , , and .
To change the offset unit to %, press , , and .
Keysight N1913/1914A User’s Guide 71
2 General Power Meter Functions
If the selected offset unit is %, the display of the offset will be in engineering unit only if the percentage value is less than 0.01% or more than 999%. The format of the display in engineering unit for an offset greater than 999% will be shown as follows,
xxx. yyyyyyyyy e+z
– There is a maximum of three numbers before the decimal point indicated
by x.
y is optional if there are non-zero numbers after the decimal point.
For example,
– 123478202 will be displayed as 123.478202 e+6
– 10000 will be displayed as 100 e+2.
The format of the display in engineering unit for an offset less than 0.01% will be as follows,
x. yyyyyyyyy e-z
– The first leading non-zero number, x, is placed before the decimal point.
y is optional if there are non-zero numbers after the decimal point.
For example,
– 0.009876 will be displayed as 9.876 e-3
– 0.0001 will be displayed as 1e-4.
Figure 2-13 Offset display in engineering unit (when the selected unit is %)
72 Keysight N1913/1914A User’s Guide

Setting Measurement Averaging

10 dB
10 dB
Minimum sensor power
Maximum sensor power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
1 8 128 128
Resolution Setting
Number of Averages
1 1 16 256
11232
11116
1118
The power meter uses a digital filter to average power readings. The number of readings averaged can range from 1 to 1024. This filter is used to reduce noise, obtain the desired resolution and to reduce the jitter in the measurement results. Increasing the value of the measurement average reduces measurement noise. However, the measurement time is increased. You can manually select the measurement average or you can set the power meter to auto measurement average mode. The default is AUTO.
When the auto measurement average mode is enabled, the power meter automatically sets the number of readings averaged together to satisfy the filtering requirements for most power measurements. The number of readings averaged together depends on the resolution setting and the power level currently being measured.
General Power Meter Functions 2
Figure 2-14 Typical averaged readings
Keysight N1913/1914A User’s Guide 73
Figure 2-14 shows the typical number of averages for each range and resolution
when the power meter is in auto filter mode and is set to normal speed mode. N1913/1914A EPM Series power meters recognize different sensor types when they are connected, and configure suitable averaging automatically.
2 General Power Meter Functions
Enter
Resolution is a measurement display function and not a channel function. In the case where a channel is set up in both the upper and lower window and the resolution settings are different, the highest resolution setting is taken to calculate the averaging number.
These four resolution levels represent:
– 1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
– 1, 2, 3 or 4 significant digits respectively if the measurement suffix is Ω or %.
Procedure
Set measurement averaging as follows:
1 Press . On dual channel meters select the required channel. The
current setting is shown in the Meas Avg field (
Channel Setup screen. The default is
2 Use the and keys to select the Filter setting field.
3 Press and use the and to step through the available settings.
If you have selected AUTO or OFF proceed at step 7. If you have selected MAN proceed as follows.
AUTO, MAN, or OFF) on the
AUTO.
4 Use the key to select the Meas Avg: value field.
5 Press to display the Meas Avg Count pop-up.
Figure 2-15 Meas Avg Count pop-up
6 Use the numeric keys to enter the required value and press .
7 Press key to close the Channel Setup screen.
74 Keysight N1913/1914A User’s Guide

Step Detection

To reduce the filter settling time after a significant step in the measured power the filter can be set to re-initialize upon detection of a step increase or decrease in the measured power. Step detection can be set in both manual and automatic measurement average modes.
Procedure
Set step detection as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Step Detect setting field.
3 Press to check the step detection to on or off.
4 Press key to close the Channel Setup screen.
General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 75
2 General Power Meter Functions
NOTE
Power
Time
A
B
Duty Cycle =
A B

Measuring Pulsed Signals

The power meter can be used to measure the power of a pulsed signal. The measurement result is a mathematical representation of the pulse power rather than an actual measurement (assumes constant peak power). The power meter measures the average power of the pulsed input signal and then divides the measurement result by the duty cycle value to obtain the pulse power reading. The allowable range of values is 0.001% to 100%. The default value is 1.000%.
If duty cycle is enabled, then Dty Cyc is displayed.
Pulse measurements are not recommended using Keysight E4412A and E4413A power sensors.
An example of a pulsed signal is shown in Figure 2-16.
Figure 2-16 Pulsed signal
76 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
Offsets
Duty cycle setting field
Duty cycle value field
%
Duty cycle indicator
Procedure
Set duty cycle as follows:
1 Press to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
2 Press to display the Offsets Setup.
3 Use the and keys to highlight the Duty Cycle setting field.
4 Press to check the Duty Cycle setting field.
Figure 2-17 Duty cycle setting
5 Press to highlight the Duty Cycle value field and press to display
the Duty Cycle pop-up. Use the numeric keypad to enter the required value in the Duty Cycle pop-up window.
Keysight N1913/1914A User’s Guide 77
6 Confirm your choice by pressing .
7 Press key to complete the duty cycle entry. The Dty Cyc indicator is
displayed as shown in Figure 2-18.
Figure 2-18 Duty cycle indicator
2 General Power Meter Functions
NOTE
Pulse power averages out any aberrations in the pulse such as overshooting or ringing. For this reason, it is called pulse power and not peak power or peak pulse power.
In order to ensure accurate pulse power readings, the input signal must be pulsed with a rectangular pulse. Other pulse shapes (such as triangle, chirp, or Gaussian) will cause erroneous results.
The pulse power on/off ratio must be much greater than the duty cycle ratio.
78 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE

Setting External Trigger for Average Power Measurement

There are two modes featured in triggered average power measurement
– Power sweep mode
– Frequency sweep mode
These modes are used to eliminate the need for lengthy test routines, while increasing measurement throughput by reducing overhead of communication with the controller. The sweep feature allows you to make power measurement by quickly stepping through a series of frequencies or power levels. Configuration shown in Figure 2-19 illustrates the triggering connection required to synchronize power meter measurement to the power source settings.
Trigger delay and trigger holdoff are both not applicable when the power meter is set to power sweep mode or frequency sweep mode.
Keysight N1913/1914A User’s Guide 79
2 General Power Meter Functions
NOTE
TRIG
IN
TRIG
IN
TRIG OUT
TRIG OUT
Power Meter
Power Source
Average Power Sensor
(optional connection)

Power Sweep Mode

Power sweep is generally used in power level calibration setup where the frequency is fixed (CW frequency), and the amplitude of the power source signal is swept. This mode can be used to characterize the flatness, linearity or gain compression of a device under test.
This feature is only available when 8480 Series, N8480 Series, E-Series E4410, or E-Series E9300 sensor is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a BNC cable.
The same connection is applied for power source TRIG OUT to power meter TRIG IN.
Figure 2-19 TRIG IN and TRIG OUT connection diagram between power
3 Press . The Channel Setup screen displays as below.
meter and power source
80 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE
Acqn
Sing Trig
Cont Trig
Sing Trig
Cont Trig
Settings
Figure 2-20 Channel Setup display
When 8480 Series, N8480 Series, E-Series E4410, E-Series E9300 or U2000 Series sensor is connected, Sensor Mode is set to AVG only by default.
4 Press . The Trigger menu is displayed.
5 Press softkey to configure the trigger.
6 Select either or .
is a single shot mode. After triggering, the measurement is
halted, and the symbol is displayed. You can start another
measurement by pressing the key.
is a continuous trigger mode. The symbol or is displayed.
7 Press to configure the remaining trigger parameters. The trigger
Settings menu consists of two pages. Figure 2-21 shows page 1 and
Figure 2-22 shows page 2.
Keysight N1913/1914A User’s Guide 81
2 General Power Meter Functions
External trigger source
Source
Ext
1 of 2
Output
On
Slope
+
Figure 2-21 Trigger setting menu 1 of 2
Figure 2-22 Trigger setting menu 2 of 2
8 Press and will be automatically enabled.
9 Press , and to allow a TTL level high to be produced at the
rear panel TRIG OUT BNC port when the meter is triggered upon a completed measurement.
10 Press and select or to set the trigger edge.
11 Set the trigger buffer size by sending SENSE:BUFFer:COUNt <buffer_size>
command to meter using remote interface.
82 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE
NOTE
Refer to the N1913/1914A EPM Series Power Meters Programming Guide for the commands usage details.
Example of command set:
BUFF:COUN 100
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488) programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation Guide for details.
12 Press or to setup measurement setting such as measurement
averaging, measurement frequency, offsets, duty cycle and so forth. Refer to
“General Power Meter Functions” on page 55 for the setup procedures.
13 Send *OPC (OPeration Complete) command to meter to set the operation
complete bit in the Standard Event Status register when all pending device operations are completed.
14 Set the meter to continuous trigger mode by sending INITiate:CONTinuous
ON command to meter.
15 Configure and set the required power sweep range and step on the power
source accordingly.
16 Set the power source trigger input and trigger output, then start sweeping.
17 Poll the status of the power meter by sending *ESR?. *ESR? will return a 1
when buffering is completed. Use FETCh? to retrieve all the buffered measurement.
Keysight N1913/1914A User’s Guide 83
2 General Power Meter Functions
NOTE
NOTE
Acqn
Sing Trig
Cont Trig
Settings
Source
Ext
1 of 2
Output
On

Frequency Sweep Mode

Frequency sweep is generally used in a frequency response calibration system where the amplitude is fixed, and the frequency of the power source signal is swept. This mode can be used to determine the frequency response of a device under test.
This feature is only available when N8480 Series, E-Series E4410, or E-Series E9300 sensor is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a BNC cable.
The same connection is applied for power source TRIG OUT to power meter TRIG IN. See Figure 2-19.
3 Press . The Channel Setup displays as shown in Figure 2-20.
When N8480 Series, E-Series E4410, E-Series E9300 or U2000 Series sensor is connected, Sensor Mode is set to AVG only by default.
4 Press . The Trigger menu is displayed.
5 Press softkey to configure a trigger.
6 Select either or .
7 Press to configure the remaining trigger parameters. The trigger
Settings menu consists of two pages. Figure 2-21 shows page 1 and
Figure 2-22 shows page 2.
8 Press and will be automatically enabled.
9 Press , and to allow a TTL level high to be produced at the
rear panel TRIG OUT BNC port when the meter is triggered upon a completed measurement.
84 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
NOTE
NOTE
10 Press or to setup measurement settings such as
measurement averaging, measurement frequency, offsets, duty cycle and so forth. Refer to “General Power Meter Functions” on page 55 for the setup procedures.
11 Set the frequency range and step by sending the below commands to meter
using remote interface.
– SENSe1:FREQuency:STARt <start_frequency><frequency_unit>
– SENSe1:FREQuency:STOP <stop_frequency><frequency_unit>
– SENSe1:FREQuency:STEP <frequency_step_size>
Refer to the EPM Series Power Meters Programming Guide for the commands usage details.
Example of commands set:
FREQ:STAR 10MHz
FREQ:STOP 500MHz
FREQ:STEP 10
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488) programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation Guide for details.
12 Send *OPC (OPeration Complete) command to meter; to set the operation
complete bit in the Standard Event Status register when all pending device operations are completed.
13 Set the meter to continuous trigger cycles by sending INITiate:CONTinuous
ON command to meter.
14 Configure and set the required power sweep range and step of the power
source.
15 Set the power source trigger input and trigger output.
16 Set the power source to sweep.
Keysight N1913/1914A User’s Guide 85
2 General Power Meter Functions
NOTE
17 Poll the status of the power meter by sending *ESR?. *ESR? will return a 1
when buffering is completed. Use FETCh? to retrieve all the buffered measurement.
In both power sweep and frequency sweep mode, the meter TRIG OUT to power source TRIG IN connection is optional. You may choose to setup the Dwell time in the power source step setting to cater the maximum settling time required by the power meter.

Determine the Right Step to be Set

Number of frequency step can be calculated by using equation below:
Step = (f
where,
Step = Number of frequency step
f
= Frequency sweep’s start point
start
= Frequency sweep’s stop point
f
stop
Interval = Frequency step size
stop
– f
+ Interval)/Interval
start
Example
When f
= 1 GHz and f
start
= 5 GHz with given interval of 0.5 GHz, the Step
stop
should be set to
Step
= (f
= (5 GHz – 1 GHz + 0.5 GHz)/0.5 GHz
= 9
stop
– f
+ Interval)/Interval
start
86 Keysight N1913/1914A User’s Guide

Setting Measurement Limits

Power Meter
Swept Source
OUT
OUT
Device Under Test
+4 dBm
+10 dBm
Amplitude
Frequency
o
o
o
o
o
o
o
Fail
Fail
Upper limit l ine
Lower limit line
You can configure the power meter to detect when a measurement has crossed over a predefined upper and/or lower limit value.
Limits are boundaries set for a certain power range and it can be applied to power, ratio or difference measurement.
Figure 2-23 Limits checking applications
In this application a swept frequency signal is applied to the input of the Device Under Test. The power meter measures the output power. The limits have been set at +4 dBm and +10 dBm. A fail occurs each time the output power is outside these limits as shown in Figure 2-24.
General Power Meter Functions 2
Figure 2-24 Limits checking results
Keysight N1913/1914A User’s Guide 87
2 General Power Meter Functions
NOTE
Meas Select

Setting Limits

The power meter can be configured to verify the current measurement in any measurement line against predefined upper and/or lower limit values. The range of values that can be set for the upper and lower limits and the default values depends on the measurement units in the currently selected measurement line (see Table 2-3 ).
Tab le 2-3 Range of values for window limits
Window Units Maximum Minimum Default Maximum Defaul t Minimum
dB +200 dB –180 dB 60 dB –120 dB
dBm +230 dBm –150 dBm 90 dBm –90 dBm
% 999.9 X% 100.0 a% 100.0 M% 100.0 p%
W 100.000 XW 1.000 aW 1.000 MW 1.000 pW
Procedure
Set the limits as follows:
Ensure you have selected the channel you wish to set up.
1 Press , to display the Measurement Setup menus.
2 Use the and keys to highlight the Limits: setting field.
3 Press to check the Limits: setting field.
4 Use the key to highlight the Minimum Limits: value field.
5 Press to display the Minimum Limit pop-up.
88 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
dBm
dBm
NOTE
Figure 2-25 Minimum limit pop-up
6 Use the numeric keys to enter the required value and press .
7 Use the key to highlight the Maximum Limits: value field.
8 Press to display the Maximum Limit pop-up.
9 Use the numeric keys to enter the required value and press .
10 Press key to close the Measurement Setup screen.
The Limits can be disabled and re-enabled by checking the Limits: setting field.
Keysight N1913/1914A User’s Guide 89
2 General Power Meter Functions
This measurement has failed as the result is greater than the set limit level.
This measurement has failed as the result is less than the set limit level.

Checking for Limit Failures

Limit failures are displayed in the appropriate field in the measurement window on the power meter’s display as shown in Figure 2-26.
Figure 2-26 Limit failures
90 Keysight N1913/1914A User’s Guide

Numeric Format

Single Numeric
Dual Numeric
Meas Select
Selected Window/
Function Field
Measurement Field
Measurements
Channel Field
Configure a measurement displayed in or format as follows:
– Press , to select the measurement window or
measurement line you want to configure.
General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 91
Figure 2-27 Measurement Setup showing single configuration
2 General Power Meter Functions
NOTE

Single Function Measurement

Figure 2-27 shows an average measurement assigned in the upper measurement
line of the upper window. (For single channel power meter, N1913A, the Channel field will be disabled, as shown in Figure 2-27).
The gate field is disabled if trigger acquisition is Free Run.
1 Use the , , , , to highlight the Combination function field.
2 Press to display the Function pop-up, and use the and to
highlight Single.
Figure 2-28 Function pop-up
3 Press to complete the entry.
4 The Meas field is set to Avg by default.
5 Press key to complete the setup and display the measurement results.
92 Keysight N1913/1914A User’s Guide

Combined Measurement

Selected window/
Function field
Gate fields
Measurement fields
Combination field
measurement
Channel fields
Figure 2-29 shows a Combined Measurement configuration; Channel A and
Channel C to be displayed in the upper measurement line of the upper display window. (For single channel power meter, N1913A, the Channel field will be disabled, as shown in Figure 2-29).
Figure 2-29 Measurement Setup showing combined configuration
General Power Meter Functions 2
Keysight N1913/1914A User’s Guide 93
1 Use the , , , , to highlight the Combination function field.
2 Press to display the Function pop-up (see Figure 2-28) and use the
and to highlight Combined.
3 Press to complete the entry.
4 Press key to complete the setup and display the measurements
results.
Figure 2-30 Measurement example display
2 General Power Meter Functions
SENSe1: Input from SENSe1 block
SENSe2: Input from SENSe2 block
(N1914A only)
:FEED
FEED1
FEED2
:MATH
“A”|”B”
“A- A”|“A/ A“
“B-A”|“B/A“
“A- B”|“A /B“
“B-B”|“B/B“
A
B
:GAIN
:HOLD
:REL
CALCulate Block
AVG
AVG

Max Hold/ Min Hold

Max Hold/Min Hold
The max hold/min hold setting on the measurement setup window can be set on the front panel or via SCPI.
Min hold is the minimum of all measurements since the start time. The min hold reading will be updated as new minimum reading appears.
Max hold is the maximum of all measurements since the start time. The max reading will be updated as new maximum reading appears.
The max hold/min hold reading can be resettable by pressing the key.
Figure 2-31 Max hold/min hold measurement is performed on the ‘HOLD’
block
94 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
To set max hold/min hold,
1 Press to display the Measurement Setup menu.
2 Use the , , , , to highlight the Hold function field.
3 Press to display the Hold pop-up, and use the and to
highlight Min or Max (see Figure 2-32).
Figure 2-32 Hold pop-up
4 Press to complete the entry.
5 Press key to complete the setup and display the measurements
results.
Figure 2-33 Min Hold and Max Hold measurement mode indicated in the
display
Keysight N1913/1914A User’s Guide 95
2 General Power Meter Functions
6 Press to display the measurement window in full screen mode. The
measurement mode will be indicated in full word in the expanded window (see
Figure 2-34).
Figure 2-34 Measurement mode in full word
96 Keysight N1913/1914A User’s Guide

Recorder Output

Max Power
Min Power
NOTE
Meas Select
The rear panel Recorder Output connectors (1 and 2) produce a dc voltage that corresponds to the power level in Watts of the selected measurement window. This dc voltage ranges from 0 to +1 Vdc. The output impedance is typically 1 kΩ.
For example, the Recorder Outputs can be used to:
– Record swept measurements
– Level an output from a source using external leveling, or
– Monitor the output power
To access the Recorder menu press , and enable Rec o/p. This allows you to switch the Recorder Output signal on or off and set the scaling of power levels
automatically or manually. The and softkeys allow you to manually scale the power levels to represent the 1 Vdc maximum and 0 Vdc minimum output voltage of the Recorder Output. Otherwise, you can set the Recorder Output to Auto to enable auto-scaling.
In auto-scaling, the minimum power is fixed at –150 dBm and the maximum power is automatically scaled based on the current power level. The auto-scaled maximum power ranges from –140 dBm, steps every 10 dBm until +230 dBm depending on the current power level. The auto-scaling is not applicable if the measurement of the selected window is a difference, ratio, channel C, or D measurement.
General Power Meter Functions 2
Procedure
Set the recorder output as follows:
Ensure you have selected the channel you want to set up.
1 Press , to display the Measurement Setup menus.
2 Use the and keys to highlight the Rec o/p: setting field.
Keysight N1913/1914A User’s Guide 97
2 General Power Meter Functions
dBm
3 Press to reveal the Recorder option menu.
4 Use the and keys to highlight the option you require.
– The N1913A has only three options, On, Auto, or Off.
– The N1914A has five options: 1, 2, Auto 1, Auto 2, or Off.
5 Press .
6 Use the key to highlight the Recorder Minimum: value field.
7 Press to display the Recorder Minimum pop-up.
Figure 2-35 Recorder Minimum pop-up
8 Use the numeric keys to enter the power level you want to generate a 1 Vdc
output in the Recorder Maximum pop-up and press .
9 Use the key to highlight the Recorder Maximum: value field.
10 Press to display the Recorder Maximum pop-up.
98 Keysight N1913/1914A User’s Guide
General Power Meter Functions 2
dBm
NOTE
Figure 2-36 Recorder Maximum pop-up
11 Use the numeric keys to enter the power level you want to generate a 0 Vdc
output in the Recorder Minimum pop-up and press .
12 Press key to close the Measurement Setup screen.
The recorder output can be disabled and re-enabled by checking the Rec o/p: setting field.
The highest power you are going to measure is used to determine the value which you should set for the Recorder Output maximum setting. For example, if you are measuring a power less than 1 mW and greater than 100 mW, then set the recorder maximum value to 1 mW.
Table 2-4 Ranges of recorder output setting
Log5040302010 0
Lin 100 W 10 W 1 W 100 mW 10 mW 1 mW
Log –10 –20 –30 –40 –50 –60
Lin 100 mW 10 mW 1 mW 100 nW 10 nW 1 nW
Keysight N1913/1914A User’s Guide 99
2 General Power Meter Functions
Save/Recall
Save
Confirm

Saving and Recalling Power Meter States

To reduce repeated setup sequences, you can save a maximum of ten power meter states in the non-volatile memory.
The save/recall functions are part of the Sys/Inputs menu, accessed by pressing
the key.
To save a measurement setup:
1 Press , to display the Save/Recall screen as shown in
Figure 2-37.
Figure 2-37 Save/Recall screen
2 Using the and keys, select an available name from the displayed list.
To change the name of a register -See Chapter 2, “Editing a Register’s Name,” starting on page 101, otherwise press .
3 The power meter prompts you to press to proceed.
Figure 2-38 Save confirm pop-up
100 Keysight N1913/1914A User’s Guide
Loading...